AEMC AEMC 505 Manual

Page 1
CLAMP-ON METERS
505
MEDIDORES DE ABRAZADERA
PINCE MULTIMETRE
99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
ENGLISH
Page 2
CONTENTS
1 DESCRIPTION ................................................................................................. 7
1.1 Interface .................................................................................................................. 7
1.2 Display .................................................................................................................... 8
1.3 Rotary Switch ......................................................................................................... 9
2 OPERATION .................................................................................................. 10
2.1 Battery Installation ................................................................................................ 10
2.2 Instrument Check ................................................................................................. 10
2.3 Measurements ..................................................................................................... 11
2.3.1 DC Current ......................................................................................... 11
2.3.2 AC Current ......................................................................................... 12
2.3.3 AC Voltage ......................................................................................... 13
2.3.4 DC Voltage ......................................................................................... 14
2.3.5 Resistance ......................................................................................... 15
2.3.6 Continuity Check ................................................................................ 16
2.4 ∆ ZERO Mode ...................................................................................................... 17
2.5 HOLD Function .................................................................................................... 17
2.6 PEAK Mode .......................................................................................................... 17
2.7 Auto Shutdown ..................................................................................................... 17
3 SPECIFICATIONS ......................................................................................... 18
3.1 Reference Conditions .......................................................................................... 18
3.2 Electrical Specifications ....................................................................................... 18
3.2.1 DC Voltage ......................................................................................... 18
3.2.2 AC Voltage ......................................................................................... 18
3.2.3 DC Current ......................................................................................... 18
3.2.4 AC Current ......................................................................................... 19
3.2.5 Resistance ......................................................................................... 19
3.2.6 Continuity ........................................................................................... 19
3.3 General Specifications ......................................................................................... 19
3.3.1 Display ............................................................................................... 19
3.3.2 True RMS Measurements .................................................................. 19
3.3.3 Environmental Conditions .................................................................. 19
3.3.4 Power Supply ..................................................................................... 19
5 Mechanical ......................................................................................... 20
3.3.
3.3.6 Compliance with International Standards ........................................... 20
3.3.7 Safety ................................................................................................. 20
4 MAINTENANCE ............................................................................................. 21
4.1 Cleaning ................................................................................................................ 21
REPAIR AND CALIBRATION ........................................................................... 22
TECHNICAL AND SALES ASSISTANCE ......................................................... 22
LIMITED WARRANTY ....................................................................................... 23
Warranty Repairs ....................................................................................................... 23
Clamp-On Meter Model 505 3
Page 3
Thank you for purchasing the AEMC Clamp-On Meter Model 505. For best
regulations covering EMC.
results from your instrument and for your safety, read the enclosed operating instructions carefully and comply with the precautions for use. This instrument is
compliant with safety standard IEC 61010- 2-032; the leads are compliant with IEC 61010-2-031 for voltages up to 600 V in category III. Failure to observe the
safety instructions may result in electric shock, fire, explosion, or destruction of the instrument and of the installations.
Symbols
WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.
Ground/earth.
Equipment is protected by double insulation.
Application or withdrawal authorized on conductors carrying dangerous
voltages. Type A current sensor as per IEC 61010-2-032.
The trash can with a line through it means that in the European Union, the product must undergo selective disposal for the recycling of electric and electronic material, in compliance with Directive WEEE 2002/96/EC.
The CE marking guarantees conformity with European directives and with
Battery.
Alternating current.
Direct current.
AC or DC.
Useful information or tip to read.
Def
inition of Measurement Categories (CAT)
CAT IV Corresponds to measurements taken at the source of low-voltage
installations. Example: power feeders, counters and protection devices.
CAT III Corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, machines or fixed industrial devices.
CAT II Corresponds to measurements taken on circuits directly connected to low-voltage installations. Example: power supply to domestic electrical appliances and portable tools.
4 Clamp-On Meter Model 505
Page 4

PRECAUTIONS FOR USE

These safety instructions are intended to ensure the safety of persons and proper operation of the device.
The operator and/or the responsible authority must carefully read and clearly
understand the various precautions to be taken in use.
If this instrument is used other than as specified, the protection it provides
may be compromised, thereby endangering you.
Do not use the instrument in an explosive atmosphere or in the presence of
flammable gases or fumes.
Do not use the instrument on networks of which the voltage or category
exceeds those mentioned.
Do not exceed the rated maximum voltages and currents between terminals
or with respect to earth.
Do not use the instrument if it appears to be damaged, incomplete, or not
properly closed.
Before each use, check the condition of the insulation on the leads, housing,
and accessories. Any element of which the insulation is deteriorated (even partially) must be set aside for repair or scrapped.
Use leads and accessories rated for voltages and categories at least equal
to those of the instrument. If not, an accessory of a lower category lowers the category of the combined clamp + accessory to that of the accessory.
Observe the environmental conditions of use.Do not modify the instrument and only use factory replacement parts.
Repairs and adjustments must be done by approved qualified personnel.
Replace the batteries as soon as the symbol appears on the display of
the unit. Disconnect all leads before opening the battery compartment cover.
Use personal protective equipment when conditions require.Keep your hands away from the unused terminals of the instrument.When handling the test probes, alligator clips, and clamp ammeters, keep
your fingers behind the physical guard.
As a safety measure, and to avoid repeated overloads on the inputs of the
device, configuration operations should only be performed when the device is disconnected from all dangerous voltages.
Clamp-On Meter Model 505 5
Page 5

RECEIVING YOUR SHIPMENT

Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the carrier and notify your distributor at once, giving a detailed description of any damage. Save the damaged packing container to substantiate your claim.

Ordering Information

Clamp-on Meter Model 505 .......................................................... Cat. #2139.82
Includes meter, pair of test leads (red/black with probe tips), two 1.5V AAA batteries, soft carrying case and a user manual.
Replacement Parts:
Lead - Replacement Set of 2, Needle Tip Color-coded (Red/Black) w/4mm Right-angle Plug (600V CAT IV w/shiel w/o Shield) for any meter that accepts 4mm banana plugs
Replacement pouch for Model 505 .................................................. Cat. #2118.74
d on Needle Tip, 1000V CAT II
.......... Cat. #2154.74
6 Clamp-On Meter Model 505
Page 6

1 DESCRIPTION

The Model 505 is a clamp-on multimeter that measures electrical and physical quantities. It can perform the following functions:
AC voltage measurementDC voltage measurementAC current measurementDC current measurementResistance measurementContinuity test with buzzer

1.1 Interface

“Delta Zero” mode“Peak” modeAuto Shutdown feature“Hold” featureHazardous voltage warning
Item Function
1 Jaws
HOLD button
2
(>2 sec) PEAK button
ZERO button
3
4 Rotary switch
5 LCD display screen
+ input terminal
6
COM (common) terminal
7
8 Trigger
Figure 1
Clamp-On Meter Model 505 7
Page 7

1.2 Display

Item Indicator See §
1 Auto Shutdown activated 2.7
2
3
4 Continuity mode 2.3.6
5
6 Measurement display area --
7
8 Low battery 2.1
9 Hazardous voltage present --
10 Negative value --
11
Figure 2
∆ ZERO mode
PEAK mode
Measurement unit - A, V, or (amp, volt, or ohm)
HOLD function
DC or AC
2.4
2.6
--
2.5
--
8 Clamp-On Meter Model 505
Page 8

1.3 Rotary Switch

Figure 3
Setting
OFF
Function See §
Turns the meter OFF --
rent measurement (A
DC cur
AC curr
ent measurement (A
AC voltage measurement (VAC)
oltage measurement (V
DC v
Conti
nuity test
Resistance measurement
DC
AC
DC
)
)
) 2.3.4
2.3.1
2.3.2
2.3.3
2.3.6
2.3.5
Clamp-On Meter Model 505 9
Page 9

2 OPERATION

Before changing batteries: set the switch to OFF, disconnect the

2.1 Battery Installation

measuring leads, and remove the clamp from the circuit under measurement.
1. Using a screwdriver, remove the battery compartment cover from the
back of the housing (see Figure 4).
2. Insert the two supplied 1.5V AA batteries, observing polarities.
3. Replace the battery compartment cover and screw it onto the housing.
Figure 4

2.2 Instrument Check

Start the meter by turning the rotary switch to any setting other than OFF.
Observe the LCD and ensure all segments are displayed (see Figure 2).
Turn the switch to the (continuity) setting. The LCD should display
OL.
Connect the two test leads to the instrument and touch the tips together. The
short-circuit buzzer should sound.
Turn the switch to a voltage setting (either or ) and check a known
voltage (for example a battery). Ensure the displayed voltage reading is
correct.
If all previous steps function normally, the meter is ready for operation.
10 Clamp-On Meter Model 505
Page 10

2.3 Measurements

For safety, disconnect the measurement leads before performing this
2.3.1 DC Current
operation.
The clamp must be positioned around a single conductor in a circuit; otherwise the measurement will be invalid.
To maximize accuracy, center the jaws around the conductor.
Fig
ure 5
1. Set
2. Press the trigger to open the clamp.
3. Place the clamp around the conductor to be measured and release
4. Read the result on the LCD.
the switch to .
the trigger. Ensure the clamp is completely closed.
Clamp-On Meter Model 505 11
Page 11
2.3.2 AC Current
For safety, disconnect the measurement leads before performing this
operation.
The clamp must be positioned around a single conductor in a circuit; otherwise the measurement will be invalid.
To maximize accuracy, center the jaws around the conductor.
ure 6
Fig
1. Set the switch to .
2. Press the trigger to open the clamp.
3. Place the clamp around the conductor to be measured and release
the trigger. Ensure the clamp is completely closed.
4. Read the result on the LCD.
12 Clamp-On Meter Model 505
Page 12
2.3.3 AC Voltage
ure 7
Fig
1. Set the switch to .
2. Connect the red test lead to the + input terminal and the black test lead
to the COM input terminal.
3. Touch the lead probe tips to the AC circuit to be measured.
4. Read the result on the LCD.
If
the meter detects voltage >30V the LCD. In this case, immediately remove the probes from the circuit under test. Also remove the probes from the circuit if the reading displays OL.
, the (hazardous) icon appears on
AC
Clamp-On Meter Model 505 13
Page 13
2.3.4 DC Voltage
Figure 8
1. Set the switch to .
2. Connect the red test lead to the + input terminal and the black test lead
to
the COM input terminal.
3. Touch the le
4. R
ead the result on the LCD.
ad probe tips to the DC voltage to be measured.
If the m the LCD. Remove the probes from the circuit if the reading displays OL.
eter detects voltage >30V
, the (hazardous) icon appears on
DC
14 Clamp-On Meter Model 505
Page 14
2.3.5 Resistance
Figure 9
1. Set the sw
2. Connect the red test lead to the + input terminal and the black test lead to the COM input terminal.
3. Touch the le to be m
4. Read the r
itch to .
ad probe tips to the points across which the resistance is
easured.
esult on the LCD.
Clamp-On Meter Model 505 15
Page 15
2.3.6 Continuity Check
Figure 10
1. Set the switch to .
2. Connect the red test lead to the + input terminal and the black test lead to the COM input terminal.
3. Touch the le
4. If the resistance is less than
ad probe tips t
o the circuit to be tested.
35Ω, the buzzer sounds continuously.
16 Clamp-On Meter Model 505
Page 16

2.4 ZERO Mode

ZERO mode enables you to view a measurement as the difference between it
and a selected “reference” value.
To enable
The measurement becomes the reference, and will be subtracted from subsequent measurements. This difference will then be displayed on the LCD.
When t
Press ZE
this mode, press the ∆ ZERO button after making a measurement.
he meter is in this mode, the symbol appears on the LCD.
RO a second time to exit this mode.

2.5 HOLD Function

Pressing the HOLD button “freezes” the measurement on the LCD. In this mode, the word HOLD appears on the LCD.
Press HOLD again to exit this mode and resume updating the measurement display.

2.6 PEAK Mode

In PEAK mode the instrument only updates the LCD when the instantaneous measurement exceeds the displayed value.
To enable this mode, press the PEAK button for >2 seconds. The measurement remains on the screen until an instantaneous measurement is made that exceeds the displayed value, at which point the LCD displays the new
measurement.
When t
he meter is in this mode, the word PEAK appears on the LCD.
Press PEAK a s
econd time to exit this mode.
In PEAK mode, pressing HOLD has no effect.

2.7 Auto Shutdown

By default, the Model 505 automatically turns OFF after 20 minutes of inactivity. The instrument can then be re-activated by either pressing the HOLD button or changing the switch setting.
When the meter is in this mode, the icon appears on the LCD.
To disable the Auto Shutdown feature, press and hold down the HOLD button
when turning the instrument ON.
Clamp-On Meter Model 505 17
Page 17

3 SPECIFICATIONS

3.1 Reference Conditions

Quantities of Influence Reference Conditions
Temperature 73.4°F ± 9°F (23°C ± 5°C)
Frequency range of the applied signal 45 to 65Hz
Magnetic field <40A/m
Electric field <1V/m

3.2 Electrical Specifications

Out-of-range measurements are identified by OL displayed on the LCD, accompanied by an audible beep.
3.2.1 DC Voltage
Display range Range Resolution Accuracy
60V 0.03 – 60.00V 0.01V
600V 60.0 – 600.0V 0.1V
1% R + 3ct
3.2.2 AC Voltage
Display range Range Frequency Resolution Accuracy
60V 0.05 – 60.00V
600V 60.0 – 600.0V
48 – 65Hz
65 – 400Hz 3.8% R + 5ct
48 – 65Hz
65 – 400Hz 3.8% R + 5ct
0.01V
0.1V
1.9% R + 5ct
1.9% R + 5ct
Note: When the measurement exceeds 630V RMS (900V in PEAK mode), the LCD displays OL and the audible alarm beeps.
3.2.3 DC Current
Display range Range Resolution Accuracy
60A 0.10 – 60.00A 0.01A
400A 60.0 – 400.0A 0.1A
2.5% R + 10ct
18 Clamp-On Meter Model 505
Page 18
3.2.4 AC Current
Display range Range Frequency Resolution Accuracy
60A 0.05 – 60.00A
400A 60.0 – 400.0A
Note: When the measurement exceeds 420A RMS (600A in PEAK mode), the LCD displays OL and the audible alarm beeps.
48 – 65Hz
65 – 400Hz 3.8% R + 5ct
48 – 65Hz
65 – 400Hz 3.8% R + 5ct
0.01A
0.1A
1.9% R + 5ct
1.9% R + 5ct
3.2.5 Resistance
Display range Range Resolution Accuracy
600 0.2 – 600.0 0.1 1% R + 2ct
3.2.6 Continuity
Detection threshold: <35
3.

3 General Specifications

3.3.1 Display
LCD 6000 counts
3.3.2 True RMS Measurements
Sampling rate: 2 measurements per second Accuracy: True R
Peak factor: ≥1.5 full scale
MS value measured up to 1kHz (-3dB), pass band ≤3.5kHz
3.3.3 Environmental Conditions
Operation temperature: 32 to 104°F (0 to 40°C) Storage temperature: 14 to 140°F (-10 to 60°C) Relative humidity: < 70% RH
3.3.4 Power Supply
Battery: 2 x 1.5V AAA or LR03 Battery life: approximately 40 hours
Clamp-On Meter Model 505 19
Page 19
3.3.5 Mechanical
Dimensions: 7.83 x 2.95 x 1.42” (199 x 75 x 36mm) Jaws opening: 1.18” (30mm) Weight: 8.57oz (243g) with batteries
Protection rating: IP30
3.3.6 Compliance with International Standards
Safety: IEC 61010-1; IEC 61010-2-032
IEC 61010-2-033; IEC 61010-2-031
EMC: IEC 61326-1
3.3.7 Safety
Insulation: double insulation - class II Degree of pollution: 2 Altitude: <6500’ (< 2000m) Category: CAT III 600V
20 Clamp-On Meter Model 505
Page 20

4 MAINTENANCE

Except for the batteries, the instrument contains no parts that can be
replaced by personnel who have not been specially trained and accredited. Any unauthorized repair or replacement of a part by an
"equivalent" may gravely impair safety.

4.1 Cleaning

Disconnect the instrument completely and turn the rotary switch to OFF.
Use a soft cloth, dampened with soapy water.
Rinse with a damp cloth and dry rapidly with a dry cloth or forced air.
Do not use alcohol, solvents, or hydrocarbons.
Keep the clamp jaws as clean as possible.
Clamp-On Meter Model 505 21
Page 21

REPAIR AND CALIBRATION

To ensure that your instrument meets factory specifications, we recommend that it be submitted to our factory Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus recorded calibration data).
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: All customers must obtain a CSA# before returning any instrument.

TECHNICAL AND SALES ASSISTANCE

If you are experiencing any technical problems, or require any assistance with the proper operation or application of your instrument, please call, mail, fax or e-mail our technical support hotline:
NOTE: Do not ship instruments to our Foxborough, MA address.
22 Clamp-On Meter Model 505
Page 22

LIMITED WARRANTY

The instrument is warranted to the owner for a period of two years from the date of original purchase against defects in manufacture. This limited warranty is
given by AEMC
Thi
s warranty is void i
related to service not
Please print the online Warranty Coverage Information for your records.
If a malfunction to us
for repair, provided we have your warranty registration information on file
or a proof of purchase. AEMC® Instruments will, at its the faulty material.
®
Instruments, not by the distributor from whom it was purchased.
f the unit has been tampered with, abused or if the defect is
performed by AEMC® Instruments.
occurs within the warranty period, you may return the instrument
option, repair or replace
Clamp-On Meter Model 505 23
Page 23
10/19
99-MAN 1
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
00520 v1
Loading...