Thank you for purchasing the AEMC Clamp-On Meter Model 505. For best
regulations covering EMC.
results from your instrument and for your safety, read the enclosed operating
instructions carefully and comply with the precautions for use. This instrument is
compliant with safety standard IEC 61010- 2-032; the leads are compliant with
IEC 61010-2-031 for voltages up to 600 V in category III. Failure to observe the
safety instructions may result in electric shock, fire, explosion, or destruction of
the instrument and of the installations.
Symbols
WARNING, risk of DANGER! The operator must refer to these instructions
whenever this danger symbol appears.
Ground/earth.
Equipment is protected by double insulation.
Application or withdrawal authorized on conductors carrying dangerous
voltages. Type A current sensor as per IEC 61010-2-032.
The trash can with a line through it means that in the European Union, the
product must undergo selective disposal for the recycling of electric and
electronic material, in compliance with Directive WEEE 2002/96/EC.
The CE marking guarantees conformity with European directives and with
Battery.
Alternating current.
Direct current.
AC or DC.
Useful information or tip to read.
Def
inition of Measurement Categories (CAT)
CAT IV Corresponds to measurements taken at the source of low-voltage
installations.
Example: power feeders, counters and protection devices.
CAT III Corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, machines or fixedindustrial
devices.
CAT II Corresponds to measurements taken on circuits directly connected to
low-voltage installations.
Example: power supply to domestic electrical appliances and portable tools.
4 Clamp-On Meter Model 505
Page 4
PRECAUTIONS FOR USE
These safety instructions are intended to ensure the safety of persons and
proper operation of the device.
The operator and/or the responsible authority must carefully read and clearly
understand the various precautions to be taken in use.
If this instrument is used other than as specified, the protection it provides
may be compromised, thereby endangering you.
Do not use the instrument in an explosive atmosphere or in the presence of
flammable gases or fumes.
Do not use the instrument on networks of which the voltage or category
exceeds those mentioned.
Do not exceed the rated maximum voltages and currents between terminals
or with respect to earth.
Do not use the instrument if it appears to be damaged, incomplete, or not
properly closed.
Before each use, check the condition of the insulation on the leads, housing,
and accessories. Any element of which the insulation is deteriorated (even
partially) must be set aside for repair or scrapped.
Use leads and accessories rated for voltages and categories at least equal
to those of the instrument. If not, an accessory of a lower category lowers
the category of the combined clamp + accessory to that of the accessory.
Observe the environmental conditions of use.
Do not modify the instrument and only use factory replacement parts.
Repairs and adjustments must be done by approved qualified personnel.
Replace the batteries as soon as the symbol appears on the display of
the unit. Disconnect all leads before opening the battery compartment cover.
Use personal protective equipment when conditions require.
Keep yourhands away from the unused terminals of the instrument.
When handling the test probes, alligator clips, and clamp ammeters, keep
your fingers behind the physical guard.
As a safety measure, and to avoid repeated overloads on the inputs of the
device, configuration operations should only be performed when the device
is disconnected from all dangerous voltages.
Clamp-On Meter Model 5055
Page 5
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with
the packing list. Notify your distributor of any missing items. If the equipment
appears to be damaged, file a claim immediately with the carrier and notify your
distributor at once, giving a detailed description of any damage. Save the
damaged packing container to substantiate your claim.
Ordering Information
Clamp-on Meter Model 505 .......................................................... Cat. #2139.82
Includes meter, pair of test leads (red/black with probe tips), two 1.5V AAA batteries, soft
carrying case and a user manual.
Replacement Parts:
Lead - Replacement Set of 2, Needle Tip Color-coded (Red/Black) w/4mm
Right-angle Plug (600V CAT IV w/shielw/o Shield)for any meter that accepts 4mm banana plugs
Replacement pouch for Model 505 .................................................. Cat. #2118.74
d on Needle Tip, 1000V CAT II
.......... Cat. #2154.74
6 Clamp-On Meter Model 505
Page 6
1 DESCRIPTION
The Model 505 is a clamp-on multimeter that measures electrical and physical
quantities. It can perform the following functions:
AC voltage measurement
DC voltage measurement
AC current measurement
DC current measurement
Resistance measurement
Continuity test with buzzer
1.1 Interface
“Delta Zero” mode
“Peak” mode
Auto Shutdown feature
“Hold” feature
Hazardous voltage warning
Item Function
1 Jaws
HOLD button
2
(>2 sec) PEAK button
∆ ZERO button
3
4 Rotary switch
5 LCD display screen
+ input terminal
6
COM(common) terminal
7
8 Trigger
Figure 1
Clamp-On Meter Model 5057
Page 7
1.2 Display
Item Indicator See §
1 Auto Shutdownactivated2.7
2
3
4 Continuity mode2.3.6
5
6 Measurement display area --
7
8 Low battery 2.1
9 Hazardous voltage present --
10 Negative value --
11
Figure 2
∆ ZERO mode
PEAK mode
Measurement unit - A, V, or Ω (amp, volt, or ohm)
HOLD function
DC or AC
2.4
2.6
--
2.5
--
8 Clamp-On Meter Model 505
Page 8
1.3 Rotary Switch
Figure 3
Setting
OFF
Function See §
Turns the meter OFF --
rent measurement (A
DC cur
AC curr
ent measurement (A
AC voltage measurement (VAC)
oltage measurement (V
DC v
Conti
nuity test
Resistance measurement Ω
DC
AC
DC
)
)
) 2.3.4
2.3.1
2.3.2
2.3.3
2.3.6
2.3.5
Clamp-On Meter Model 5059
Page 9
2 OPERATION
Before changing batteries: set the switch to OFF, disconnect the
2.1 Battery Installation
measuring leads,and remove the clamp from the circuit under
measurement.
1.Using a screwdriver, remove the battery compartment cover from the
back of the housing (see Figure 4).
2.Insert the two supplied 1.5V AA batteries, observing polarities.
3.Replace the battery compartment cover and screw it onto the housing.
Figure 4
2.2 Instrument Check
Start the meter by turning the rotary switch to any setting other than OFF.
Observe the LCD and ensure all segments are displayed (see Figure 2).
Turn the switch to the (continuity) setting. The LCD should display
OL.
Connect the two test leads to the instrument and touch the tips together. The
short-circuit buzzer should sound.
Turn the switch to a voltage setting (either or ) and check a known
voltage (for example a battery). Ensure the displayed voltage reading is
correct.
If all previous steps function normally, the meter is ready for operation.
10 Clamp-On Meter Model 505
Page 10
2.3 Measurements
For safety, disconnect the measurement leads before performing this
2.3.1 DC Current
operation.
The clamp must be positioned around a single conductor in a circuit;
otherwise the measurement will be invalid.
To maximize accuracy, center the jaws around the conductor.
Fig
ure 5
1.Set
2.Press the trigger to open the clamp.
3.Place the clamp around the conductor to be measured and release
4.Read the result on the LCD.
the switch to .
the trigger. Ensure the clamp is completely closed.
Clamp-On Meter Model 50511
Page 11
2.3.2 AC Current
For safety, disconnect the measurement leads before performing this
operation.
The clamp must be positioned around a single conductor in a circuit;
otherwise the measurement will be invalid.
To maximize accuracy, center the jaws around the conductor.
ure 6
Fig
1.Set the switch to .
2.Press the trigger to open the clamp.
3.Place the clamp around the conductor to be measured and release
the trigger. Ensure the clamp is completely closed.
4.Read the result on the LCD.
12 Clamp-On Meter Model 505
Page 12
2.3.3 AC Voltage
ure 7
Fig
1.Set the switch to .
2.Connect the red test lead to the + input terminal and the black test lead
to the COM input terminal.
3.Touch the lead probe tips to the AC circuit to be measured.
4.Read the result on the LCD.
If
the meter detects voltage >30V
the LCD. In this case, immediately remove the probes from the circuit
under test. Also remove the probes from the circuit if the reading
displays OL.
, the (hazardous) icon appears on
AC
Clamp-On Meter Model 50513
Page 13
2.3.4 DC Voltage
Figure 8
1.Set the switch to .
2.Connect the red test lead to the + input terminal and the black test lead
to
the COM input terminal.
3.Touch the le
4.R
ead the result on the LCD.
ad probe tips to the DC voltage to be measured.
If the m
the LCD. Remove the probes from the circuit if the reading displays OL.
eter detects voltage >30V
, the (hazardous) icon appears on
DC
14 Clamp-On Meter Model 505
Page 14
2.3.5 Resistance
Figure 9
1.Set the sw
2.Connect the red test lead to the + input terminal and the black test lead
to the COM input terminal.
3.Touch the leto bem
4.Read the r
itch to.
ad probe tips to the points across which the resistance is
easured.
esult on the LCD.
Clamp-On Meter Model 50515
Page 15
2.3.6 Continuity Check
Figure 10
1.Set the switch to.
2.Connect the red test lead to the + input terminal and the black test lead
to the COM input terminal.
3.Touch the le
4.If the resistance is less than
ad probe tips t
o the circuit to be tested.
35Ω, the buzzer sounds continuously.
16 Clamp-On Meter Model 505
Page 16
2.4 ∆ ZERO Mode
∆ ZERO mode enables you to view a measurement as the difference between it
and a selected “reference” value.
To enable
The measurement becomes the reference, and will be subtracted from
subsequent measurements. This difference will then be displayed on the LCD.
When t
Press ∆ ZE
this mode, press the ∆ ZERO button after making a measurement.
he meter is in this mode, the ∆ symbol appears on the LCD.
RO a second time to exit this mode.
2.5 HOLD Function
Pressing the HOLD button “freezes” the measurement on the LCD. In this mode,
the word HOLD appears on the LCD.
Press HOLD again to exit this mode and resume updating the measurement
display.
2.6 PEAK Mode
In PEAK mode the instrument only updates the LCD when the instantaneous
measurement exceeds the displayed value.
To enable this mode, press the PEAK button for >2 seconds. The measurement
remains on the screen until an instantaneous measurement is made that
exceeds the displayed value, at which point the LCD displays the new
measurement.
When t
he meter is in this mode, the word PEAK appears on the LCD.
Press PEAK a s
econd time to exit this mode.
In PEAK mode, pressing HOLD has no effect.
2.7 Auto Shutdown
By default, the Model 505 automatically turns OFF after 20 minutes of inactivity.
The instrument can then be re-activated by either pressing the HOLD button or
changing the switch setting.
When the meter is in this mode, the icon appears on the LCD.
To disable the Auto Shutdown feature, press and hold down the HOLD button
when turning the instrument ON.
Clamp-On Meter Model 50517
Page 17
3 SPECIFICATIONS
3.1 Reference Conditions
Quantities of Influence Reference Conditions
Temperature 73.4°F ± 9°F (23°C ± 5°C)
Frequency range of the applied signal 45 to 65Hz
Magnetic field <40A/m
Electric field <1V/m
3.2 Electrical Specifications
Out-of-range measurements are identified by OL displayed on the LCD,
accompanied by an audible beep.
3.2.1 DC Voltage
Display range Range Resolution Accuracy
60V 0.03 – 60.00V 0.01V
600V 60.0 – 600.0V 0.1V
1% R + 3ct
3.2.2 AC Voltage
Display range Range Frequency Resolution Accuracy
60V 0.05 – 60.00V
600V 60.0 – 600.0V
48 – 65Hz
65 – 400Hz3.8% R + 5ct
48 – 65Hz
65 – 400Hz3.8% R + 5ct
0.01V
0.1V
1.9% R + 5ct
1.9% R + 5ct
Note: When the measurement exceeds 630V RMS (900V in PEAK mode), the
LCD displays OL and the audible alarm beeps.
3.2.3 DC Current
Display range Range Resolution Accuracy
60A 0.10 – 60.00A 0.01A
400A 60.0 – 400.0A 0.1A
2.5% R + 10ct
18 Clamp-On Meter Model 505
Page 18
3.2.4 AC Current
Display range Range Frequency Resolution Accuracy
60A 0.05 – 60.00A
400A 60.0 – 400.0A
Note: When the measurement exceeds 420A RMS (600A in PEAK mode), the
LCD displays OL and the audible alarm beeps.
48 – 65Hz
65 – 400Hz3.8% R + 5ct
48 – 65Hz
65 – 400Hz3.8% R + 5ct
0.01A
0.1A
1.9% R + 5ct
1.9% R + 5ct
3.2.5 Resistance
Display range Range Resolution Accuracy
600Ω0.2 – 600.0Ω0.1Ω1% R + 2ct
3.2.6 Continuity
Detection threshold: <35Ω
3.
3 General Specifications
3.3.1 Display
LCD 6000 counts
3.3.2 True RMS Measurements
Sampling rate: 2 measurements per second
Accuracy: True R
Peak factor: ≥1.5 full scale
MS value measured up to 1kHz (-3dB), pass band ≤3.5kHz
3.3.3 Environmental Conditions
Operation temperature: 32 to 104°F (0 to 40°C)
Storage temperature: 14 to 140°F (-10 to 60°C)
Relative humidity: < 70% RH
3.3.4 Power Supply
Battery: 2 x 1.5V AAA or LR03
Battery life: approximately 40 hours
Clamp-On Meter Model 50519
Page 19
3.3.5 Mechanical
Dimensions: 7.83 x 2.95 x 1.42” (199 x 75 x 36mm)
Jaws opening: 1.18” (30mm)
Weight: 8.57oz (243g) with batteries
Protection rating: IP30
3.3.6 Compliance with International Standards
Safety: IEC 61010-1; IEC 61010-2-032
IEC 61010-2-033; IEC 61010-2-031
EMC: IEC 61326-1
3.3.7 Safety
Insulation: double insulation - class II
Degree of pollution: 2
Altitude: <6500’ (< 2000m)
Category: CAT III 600V
20 Clamp-On Meter Model 505
Page 20
4 MAINTENANCE
Except for the batteries, the instrument contains no parts that can be
replaced by personnel who have not been specially trained and
accredited. Any unauthorized repair or replacement of a part by an
"equivalent" may gravely impair safety.
4.1 Cleaning
Disconnect the instrument completely and turn the rotary switch to OFF.
Use a soft cloth, dampened with soapy water.
Rinse with a damp cloth and dry rapidly with a dry cloth or forced air.
Do not use alcohol, solvents, or hydrocarbons.
Keep the clamp jaws as clean as possible.
Clamp-On Meter Model 50521
Page 21
REPAIR AND CALIBRATION
To ensure that your instrument meets factory specifications, we recommend
that it be submitted to our factory Service Center at one-year intervals for
recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization
number (CSA#). This will ensure that when your instrument arrives, it will be
tracked and processed promptly. Please write the CSA# on the outside of
the shipping container. If the instrument is returned for calibration, we need
to know if you want a standard calibration, or a calibration traceable to
N.I.S.T. (includes calibration certificate plus recorded calibration data).
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are
available.
NOTE: All customers must obtain a CSA# before returning any
instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance
with the proper operation or application of your instrument, please call, mail,
fax or e-mail our technical support hotline:
NOTE: Do not ship instruments to our Foxborough, MA address.
22 Clamp-On Meter Model 505
Page 22
LIMITED WARRANTY
The instrument is warranted to the owner for a period of two years from the date
of original purchase against defects in manufacture. This limited warranty is
given by AEMC
Thi
s warranty is void i
related to service not
Please print the online Warranty Coverage Information for your records.
If a malfunction
to us
for repair, provided we have your warranty registration information on file
or a proof of purchase. AEMC® Instruments will, at its
the faulty material.
®
Instruments, not by the distributor from whom it was purchased.
f the unit has been tampered with, abused or if the defect is
performed by AEMC® Instruments.
occurs within the warranty period, you may return the instrument
option, repair or replace
Clamp-On Meter Model 50523
Page 23
10/19
99-MAN 1
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
00520 v1
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.