It is important to mount the AEM Monitor under the electrosurgical generator so the adapter
cable reaches the return electrode connector. The AEM Monitor is supplied with feet
spaced for compatibility with most equipment carts. Recesses are provided in the cover of
the AEM Monitor so most specified electrosurgical generator products may be located
securely upon it.
Warning
FIRE HAZARD. DONOTUSEEXTENSIONCORDS.
ELECTRICAL SHOCK HAZARD. CONNECTTHEPOWERCORDTOAPROPERLYGROUNDEDRECEPTACLE.
1. Connect the AEM Monitor power cord to a wall receptacle with the proper voltage.
2. Turn on the AEM Monitor. The system completes an automatic self-check. All of the
visual indicators illuminate and two beeps are heard. If this is not the case, please
refer to the “Troubleshooting” section.
00875-002
6 - 1
Page 2
Surgical Use
Monopolar Surgery
Active Electrode Monitoring will manage and monitor “stray” monopolar current (insulation
failure and/or capacitive coupling) that emanate either from your laparoscopic instrument
when using an Encision shield or while using a laparoscopic instrument with AEM incorporated into its construction.
Caution
THE ACTIVE ELECTRODE MONITORING SYSTEMTECHNOLOGYISDESIGNEDTOSAFELYDELIVER
ELECTROSURGICALENERGYANDTOPREVENTINJURYCAUSEDBYINSULATIONFAILUREANDCAPACITIVE
COUPLING. THE AEM MONITORISNOTINTENDEDTOTESTFORINSULATIONDAMAGEONLAPAROSCOPIC
INSTRUMENTS. DONOTATTEMPTTOUSETHISSYSTEMASANINSTRUMENTINSPECTIONTOOL.
POST is activated when the power is switched on after being off for at least 30 seconds. In
POST, each of the monopolar visual indicators illuminates for six seconds and two beeps
are heard.
Setup
1. Connect the AEM Monitor inhibit adapter to the return receptacle on the
electrosurgical generator.
2. Apply the return electrode to the patient, and plug it into the inhibit adapter already
mounted into the electrosurgical unit’s return electrode receptacle.
3. Plug AEM Cord Adapter into footswitch output of ESU.
4. Connect an AEM cord with an AEM instrument to the AEM CORDADAPTER at the
ESU.
00875-002
Page 3
Surgical Use
Checking the AEM Monitoring System
The following is a quick test of the AEM Monitoring System. A failure of this test should be
resolved before using the system. Please refer to the Troubleshooting Section to establish
the cause of any failure.
5. Verify that the setup fault indicators are off and the READY indicator illuminates on the
AEM Monitor.
6. Turn on the electrosurgical generator and enable its contact quality monitoring
system. It should be in its normal operational state.
7. Disconnect the AEM Cord from the AEM CORDADAPTER at the ESU. Verify that the
following occurs:
❏The AEM Monitor AEM CORD indicator illuminates.
❏The AEM Monitor READY indicator extinguishes.
❏The contact quality monitoring system on the electrosurgical generator alarms.
8. Reconnect the AEM Cord to the AEM CORD Adapter at the ESU. (Reset CQM
system if necessary.) Verify that the following occurs:
❏The AEM Monitor AEM CORD indicator extinguishes.
❏The AEM Monitor READY indicator illuminates.
❏The contact quality monitoring system on the electrosurgical generator no longer
alarms.
After successful completion of these steps, the system is ready for use. If the system
does not perform as described, do not use until repaired and refer to Section 4, Troubleshooting.
Bipolar Surgery
Endpoint Monitoring of the bipolar instrument will assist the surgeon in confirming the end
point of bipolar desiccation. This information is displayed on the left front panel of your
AEM Monitor as an illuminated visual graph and a volume controlled audio alarm. If you are
using the Encision Endpoint Monitor Remote Display (EMR), plug it into the receptacle
found on the left rear panel of the AEM Monitor.
ILLUMINATESDURINGBIPOLARCURRENTFLOW. THE ENDPOINT MONITORFRONTPANELONTHE AEM
MONITORDOESNOTILLUMINATE.
6 - 3
Page 4
Surgical Use
1. Plug the bipolar jumper cord into the receptacle marked Bipolar Generator on the
left front panel of the AEM Monitor and the other end into the bipolar receptacle of the
electrosurgical generator.
2. Attach bipolar jumper cord retainer bracket according to accompanying installation
instructions. (Document #00476)
3. Prepare the surgical instrument to be used for the procedure. Connect one end of the
bipolar instrument cord to the instrument and the other end into the receptacle marked
Bipolar Accessory on the left front panel of the AEM Monitor.
4. Adjust the volume of the clicks that indicate bipolar current flow. The Volume Control
switch is located on the left rear panel of the AEM Monitor.
Caution
THE ENDPOINT MONITORACTIVATIONCLICKSWHENANACCESSORYISACTIVE. DONOTTURNTHE
6. After successful completion of these steps, the bipolar activation tone on the bipolar
generator sounds upon keying and the system is ready for operation. In operation, the
bipolar indicator bar illuminates, indicating the current which is flowing between tines.
The response of the Endpoint Monitor may be tested during use.
1. Set the generator at 5 to 10 watts (depending upon the generator characteristics).
2. Touch the tines of the instrument together. There should be a smooth registration of
the current on the bar graph scale. The clicking will also change its rate to correspond
with the current.
When performing this check for the first time with a particular bipolar generator, start at a
low setting, then increase the setting to obtain a mid scale deflection of the bar graph. This
test ensures that all three components (cord, bipolar instrument and Endpoint Monitor) are
functional.
A SINGULAR AEM INSTRUMENTMUSTBETHESOLECONDUCTOROFENERGYTOTISSUE. DONOTCONDUCT
ENERGYBYTOUCHINGAN AEM INSTRUMENTTOASECONDINSTRUMENTCONTACTINGTISSUE. THE
SECONDDEVICEWILLNOTBEPROTECTEDFROMCAPACITIVECOUPLINGANDINSULATIONFAILURE.
Responding to Monitor Alarms
When using AEM Monitoring, successful electrosurgery depends upon an absence of any
critical fault conditions. Should one develop, the AEM Monitoring System disables the
attached electrosurgical generator, with contact quality monitor, from further functioning until
you correct the alarm condition.
The AEM Monitor extinguishes its READY indicator and illuminates one or more of the
alarm indicators.
❏Check that the return electrode connector is securely connected into the AEM
Monitor inhibit adapter and the inhibit adapter is connected to the return electrode
receptacle of the electrosurgical generator with contact quality monitoring.
6 - 6
If both connections have been made and the amber indicator continues to illuminate, replace the return electrode.
00875-002
Page 7
Surgical Use
AEM CORD Amber Alarm Indicator (SETUP FAULT)
❏Check the AEM cord to ensure that it is securely connected to the front panel
receptacle marked AEM CORD and that the other end is connected to an AEM
integrated instrument.
If both connections are made and the amber indicator continues to illuminate,
replace the AEM cord.
❏Replace both the instrument and Encision shield or the integrated AEM instrument,
whichever is appropriate.
❏If the INSULATION indicator continues to illuminate, use a backup AEM Monitor to
complete the surgical procedure.
If for any reason, an AEM alarm continues from your AEM Monitor, use a backup AEM
Monitor to complete the surgical procedure.
Preparing the AEM Monitor for Reuse
1. Turn off the AEM Monitor.
2. Disconnect all accessories.
Warning
ELECTRIC SHOCK HAZARD. ALWAYSUNPLUGTHE AEM MONITORBEFORECLEANING.
3. Follow the procedures approved by your institution or use a validated infection control
procedure. Use a mild cleaning solution or disinfectant and a damp cloth to
thoroughly wipe all surfaces and the power cord.