A&D Weighing AD-4212C-301, AD-4212C-3000, AD-4212C-300, AD-4212C-6000, AD-4212C-600 Instruction Manual

...
AD-4212C-300 A
D-4212C-600
A
D-4212C-3000
A
D-4212C-6000
A
D-4212C-301
A
D-4212C-3100
Production Weighing Unit
1WMPD4001992B
© 2011 A&D Company, Limited. All rights reserved.
No part of this publication may be reproduced, transmitted, transcribed, or translated
into any language in any form by any means without the written permission of A&D
Company, Limited.
The contents of this manual and the specifications of the instrument covered by this
manual are subject to change for improvement without notice.
Windows and Excel are registered trademarks of Microsoft Corporation.
.
CONTENTS
1. INTRODUCTION..........................................................................................................................2
1.1. Features ........................................................................................................................................ 2
1.2. Compliance ................................................................................................................................... 5
2. UNPACKING THE BALANCE .....................................................................................................6
2.1. Unpacking...................................................................................................................................... 6
2.2. Installing the Balance .................................................................................................................... 8
3. PRECAUTIONS .........................................................................................................................10
3.1. Before Use .................................................................................................................................. 10
3.2. When Building Into a System .......................................................................................................11
3.3. During Use .................................................................................................................................. 14
3.4. After Use...................................................................................................................................... 14
3.5. Power Supply .............................................................................................................................. 14
4. OPERATION OF WinCT-AD4212C ...........................................................................................15
4.1. Changing the Weighing Speed Using WinCT-AD4212C ................................................................... 16
4.2. Calibration ................................................................................................................................... 16
4.3. Changing the Baud Rate Using WinCT-AD4212C............................................................................ 18
5. SMART RANGE FUNCTION .....................................................................................................19
6. COMMUNICATION SPECIFICATIONS......................................................................................20
7. MAINTENANCE.........................................................................................................................25
8. TROUBLESHOOTING ...............................................................................................................26
8.1. Checking the Balance Performance and Environment ............................................................... 26
8.2. Asking For Repair........................................................................................................................ 27
9. SPECIFICATIONS......................................................................................................................28
10. EXTERNAL DIMENSIONS .....................................................................................................30
11. OPTIONS................................................................................................................................31
12. TERMS/INDEX .......................................................................................................................33
1
1. INTRODUCTION
)
)
This manual describes how the AD-4212C series balance works and how to get the most out of it in
terms of performance.
Read this manual thoroughly before using the balance and keep it at hand for future reference.
1.1. Features
z A Weighing Unit, suitable for building into a production line system. The weighing unit is
compact, with a width of 59 mm.
z A Super Hybrid Sensor (SHS) is used to provide high resolution and high response speed.
Model
AD-4212C-300 320 g 0.001 g 0.5 second (0 - 30 g), 1.0 second (30 - 320 g) AD-4212C-600 620 g 0.001 g 0.5 second (0 - 30 g), 1.0 second (30 - 620 g) AD-4212C-3000 3200 g 0.01 g 0.5 second (0 - 300 g), 1.0 second (300 - 3200 g AD-4212C-6000 6200 g 0.01 g 0.5 second (0 - 300 g), 1.0 second (300 - 6200 g
AD-4212C-301
AD-4212C-3100
*1 With FAST selected under good environment *2 Smart range (Refer to “5. SMART RANGE FUNCTION”.)
z The AD-4212C series can output the weighing digital data directly. Therefore, the AD-4212C
series (the weighing unit) can be connected to a personal computer or a PLC directly.
z Dust-protected and Protected Against Splashing Water (Complying with IP65). The AD-4212C
series has a patented shock absorber under the weighing pan and can cope with movement in
all directions, protecting the weighing unit from an actuator malfunction.
z Windows Communication Tools (WinCT-AD4212C), allows easy confirming of the weighing
data by using a Windows-based personal computer.
Windows is a registered trademark of the Microsoft Corporation.
Weighing
capacity
51 g*2 0.0001 g 1.3 seconds
320 g 0.001 g 1.0 second 510 g*2 0.001 g 1.3 seconds 3200 g 0.01 g 1.0 second
Minimum
weighing
value
Stabilization time
*1
z When connected to an optional remote controller, an AD-8922A, AD-8923-BCD or AD-8923-CC,
the remote controller can receive the weighing data and manually perform re-zero and
calibration.
The AD-8922A can output the data using BCD, comparator or analog out by selection of the
option installed in the AD-8922A.
The AD-8923-BCD outputs the data using BCD and the AD-8923-CC outputs the data using
CC-Link.
Note: CC-Link is a high-speed field network able to simultaneously handle both control and
information data.
z When connected to an optional AD-8526 Ethernet converter, it can convert the AD-4212C
RS-232C data, for output to a local area network (LAN).
2
A
(
)
P
l
A
A
A
(
)
A
P
l
Example 1 (Connecting to a personal computer or an AD-8922A)
D-4212C
D-4212C
D-8922A
ccessory
RS-232C cable
ccessory
RS-232C cable
USB converter
Option: AX-USB-9P)
10 m
10 m
USB port
Communications cable
Option: AX-KO1786-200)
Communication software “
WinCT-AD4212C”
ersona
computer
USB converter
Option: AX-USB-9P)
ersona
computer
3
A
(
)
A
A
A
A
*
A
A
A
(
)
A
A
A
Example 2 (Connecting to a PLC)
.
D-4212C
D-4212C
ccessory
RS-232C cable
D-8922A-01
Connection cable
D-8922A
BCD output board
(Option
AD-8922A-01)
10 m
D-sub 9-
BCD cable (Fabricate locally)
pin (female)
PLC
BCD input
PLC
* Specify the AD-8922A-01 connection cable when ordering the AD-8922A-01.
AX-KO3705-200 2 m AX-KO3705-500 5 m AX-KO3705-1000 10 m
PLC
D-4212C
D-8923-CC
CC-Link network
ccessory
RS-232C cable
10 m
ccessory
RS-232C cable (10 m) To ot h e r remote stations
D-4212C
LAN port
D-8526
LAN
cable
4
PLC
1.2. Compliance
Compliance with FCC Rules
Please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when the equipment is operated in a commercial environment. If this unit is operated in a residential area, it may cause some interference and under these circumstances the user would be required to take, at his own expense, whatever measures are necessary to eliminate the interference.
(FCC = Federal Communications Commission in the U.S.A.)
Compliance with EMC Directives
This device features radio interference suppression in compliance with valid EC
Regulation 2004/108/EEC.
5
2. UNPACKING THE BALANCE
2.1. Unpacking
The balance is a precision instrument. Unpack the balance carefully. Keep the packing material
to be used for transporting the balance in the future.
When unpacking, see the illustration below to confirm that everything is contained.
Screws (Flat head: M4 X 6)
Pan support boss
Breeze break
Weighing pan
Pan support
RS-232C connector
Grounding terminal
(See note below.)
Arrow mark face up
AC adapter jack
AC adapter plug
RS-232C cable (10 m)
Weighing unit
Windows communication tools (WinCT-AD4212C)
Leveling foot
Accessories
1
AC adapter
*
Calibration weight 200 g OIML, Class E2 or equivalent
2
RS-232C cable (10 m)
Weighing pan
Pan support
Screws (Flat head: M4 X 6)
*1 Please confirm that the AD adapter type is correct for your local voltage and receptacle type.
*2 2-m and 5-m cables are available as options. Refer to "10. OPTIONS".
*
AC adapter ID label
Note
When the AD-4212C series weighing unit is built into a system, be sure to
earth ground the weighing unit using the grounding terminal.
6
ATTACHING THE AD-4212C SERIES WEIGHING UNIT DIRECTLY ON THE MOUNTING BASE
When the AD-4212C series weighing unit is built into a system, remove the three leveling feet and use the screw holes to secure the weighing unit to the base.
Attachment Procedure
1. Remove the three leveling feet.
Caution Do not remove any other screws from the bottom of the weighing unit. Doing so could damage the weighing sensor.
2. Secure the weighing unit from above. (Prepare appropriate screws with a size corresponding to an M5 screw. The screw hole diameter: 6 mm).
In case of securing the weighing unit from the under side, use M6 screws with 1-mm pitch.
Notes
The screws to secure the weighing unit
to the mounting base are not provided.
For the position of the holes, refer to “10.
EXTERNAL DIMENSIONS”.
Weighing unit
Leveling foot
To secure
When the leveling adjustment is difficult to perform due to the installation conditions, place a shim between the lower surface of the weighing unit and the securing surface, or use two nuts. If the adjustment is still difficult to perform, perform calibration before use. Then, the balance will function normally.
7
2.2. Installing the Balance
Install the balance as follows:
1. Refer to “3. PRECAUTIONS” for installing the balance.
2. Refer to "2-1 Unpacking" on page 6, to attach the pan support and the weighing pan on the weighing unit.
3. Insert the RS-232C cable into the jack located on the rear of the weighing unit.
When confirming the performance of the weighing unit using WinCT-AD4212C, connect the cable into RS-232C terminal of the computer.
D-sub 9-pin
Rear of the weighing unit
Arrow mark face up
RS-232C cable
How to disconnect the cable from the weighing unit Slide the connector sleeve in the direction of the arrow to unlock and gently pull the connector
out.
Slide the connector sleeve
Unlocked
in the direction of the arrow
4. Confirm that the AC adapter type is correct for the local voltage and power receptacle type.
5. Plug the AC adapter plug into the AC adapter jack located on the rear of the weighing unit and plug the AC adapter into the electrical outlet. Warm up the balance for 30 minutes or more with nothing on the weighing pan.
6. Set the pan unit and I/O unit to adapt to the peripheral system. Set the following for the I/O unit.
 RS-232C (Refer to“6. COMMUNICATION SPECIFICATIONS”)
Set the weighing speed to adapt to the ambient conditions.
Note
Make sure that the personal computer is in operation before connecting to the AD-4212C.
7. After the balance has been installed, calibrate the balance using the 200 g calibration weight provided with the balance. For details, refer to “4.2. Calibration”. (The AD-4212C-3000/6000/3100 can be calibrated using the 200 g weight.)
8. By removing the provided weighing pan, a locally fabricated weighing pan can be attached to the
AD-4212C series by using the four screw holes (M4 screw x pitch 0.7 mm) on the pan support.
8
MASS OF THE LOCALLY FABRICATED WEIGHING PAN
Design the weighing pan so that the mass falls within the ranges shown in the table below:
Model Status
Mass of
the weighing pan (g)
Pan support is used (with weighing pan removed) 20 to 130
AD-4212C-300/301
Pan support boss is used (with the weighing pan and pan support removed)
70 to 180
Pan support is used (with weighing pan removed) 20 to 410
AD-4212C-600
Pan support boss is used (with the weighing pan and pan support removed)
70 to 460
Pan support is used (with weighing pan removed) 20 to 3010
AD-4212C-3000/3100
Pan support boss is used (with the weighing pan and pan support removed)
70 to 3060
Pan support is used (with weighing pan removed) 20 to 6010
AD-4212C-6000
Pan support boss is used (with the weighing pan and pan support removed)
70 to 6060
AD-4212C-300/301 AD-4212C-600
When the weighing pan
is removed
Mass range to be weighed
320 g
210 g
When the weighing pan and
pan support are removed
Mass range to be weighed
320 g
210 g
When the weighing pan
is removed
Mass range to be weighed
620 g
210 g
When the weighing pan and
pan support are removed
Mass range to be weighed
620 g
210 g
20 g
fabricated pan
Mass
of locally
130 g
70 g
Mass of
locally
fabricated pan
180 g
20 g
Mass of
locally
fabricated pan
410 g
70 g
460 g
Mass of
locally
fabricated pan
AD-4212C-3000/3100 AD-4212C-6000
When the weighing pan
is removed
Mass range
3200 g
to be weighed
Mass
of locally
20 g
fabricated pan
Notes
210 g
3010 g
When the weighing pan and
pan support are removed
Mass range
70 g
to be weighed
Mass of
locally
fabricated pan
3200 g
210 g
3060 g
When the weighing pan
is removed
Mass range
20 g
to be weighed
Mass of
locally
fabricated pan
6200 g
210 g
6010 g
When the weighing pan and
pan support are removed
Mass range
70 g
to be weighed
Mass of
locally
fabricated pan
6200 g
210 g
6060 g
If the balance is to be used in a range other than shown above, contact the local A&D dealer.
To avoid the effects of static electricity or magnetism, use materials other than resin or
magnetic materials when designing a weighing pan.
When a locally fabricated weighing pan is used, the zero point may be shifted greatly
right after the AC adapter is plugged in. Perform re-zero before weighing to cancel the amount of zero drift as necessary.
9
3. PRECAUTIONS
To get the optimum performance from the balance and acquire accurate weighing data, note the
following:
3.1. Before Use
Install the weighing unit in an environment where the temperature and humidity are not
excessive. The best operating temperature is about 20°C / 68°F at about 50% relative
humidity.
Install the weighing unit where it is not exposed to direct sunlight and it is not affected by
heaters or air conditioners.
Install the weighing unit where it is free of dust.
Install the weighing unit away from equipment which produces magnetic fields.
Install the weighing unit in a stable place avoiding vibration and shock. Corners of rooms on
the first floor are best, as they are less prone to vibration.
Ensure a stable power source when using the AC adapter.
Plug in the AC adapter as usual, and warm up the balance for 30 minutes or more.
Calibrate the balance before use or after having moved it to another location.
In addition, calibrate it periodically to maintain the accuracy.
Caution
Do not install the balance where flammable or corrosive gas is present.
10