Analytical Weighing Sens o rs
Series
At last, 1 μg readability in
production-line weighing!
ISO
CERTIFIED
9000
http://www.aandd.jp
...Clearly a Better Value

Looking for the Highest Precision
Possible for Your Weighing System?
Display unit
2 m (removable)
Quick reference sheet
Weighing unit
80mm
Exceptional Weighing Performance
GLP
Compliant
i
i 1 µg (= 1/1,000,000 g) is approximately 1 nanoliter of water volume.
*
ii
ii AD-4212B-101/102
*
ON
Counting
Function
Percentage
Function
Comparator
Function
Auto Power
ON
Smart Range
Software
Minimum Weighing Value from 1 µg
The AD-4212B realizes the precision of laboratory analytical balances (including micro-balances) and most minute
weighing in production processes.
Fast Stabilization of 4.0 Seconds for 0.01 mg Resolution
The high-speed, efficient weighing dramatically improves the throughput of your system.
Adjustable Response Characteristics
Select one of three preset combinations of weighing speed and stability (FAST, MID, or SLOW) to effectively minimize
the influence of drafts and vibrations.
Super Hybrid
Sensor
LCD
Backlight
External
Calibration
LAN-Ethernet
Interface
RS-232C
Interface

Great Installation Flexibility
Space-saving with an 80 mm width
The slim separate weighing unit facilitates the compact arrangement of your system.
Wall / Panel-Mountable Display Unit
The provided stand can be used to attach the display unit to a wall. The DIN-sized (192 × 96 mm)
display unit is also suitable for panel mounting.
Wall / panel mounti ng
Possible to Create Your Own Weighing Pan
Design and attach your preferred weighing pan when the standard pan is unfit for your specific application.
Weighing in Vacuum Chambers
You can safely install and use the weighing unit in a vacuum chamber, as it is
isolated from the delicate electric parts in the display unit.
Weighing in a vacuum chamber
Diverse Data Management Solutions
Data Memory Function
The balance stores up to 200 weighing results (100 with date & time) or the 50 latest calibration results without external
memory devices. It can also save 20 unit masses for the counting mode or 20 sets of upper/lower limit values for easy
recall afterwards.
iii
iii Not available when the BCD output (AD-4212-01) is installed
*
Various Input / Output Methods
The following three choices are available:
1. RS-232C serial interface, 3-level / 5-level comparator contact output with buzzer, and external contact input for
RE-ZERO and PRINT (standard)
2. BCD output and external contact input for RE-ZERO and ON/OFF (AD4212A-01, optional)
3. LAN-Ethernet interface with WinCT-Plus software (AD4212A-08, optional)
GLP/GMP/GCP/ISO Compliant Output
For documentation, the AD-4212B can output the balance manufacturer, model, serial number, ID number, date,
time, space for a signature, calibration records, etc.
High Durability and Protection
Guarded Against Vertical and Transverse Overloading
Our patented shock absorber under the weighing pan copes with movement in all directions so that the sensor is
protected from actuator malfunctions.
Chemically Resistant, Stainless Steel Construction
The splash-proof, IP54 weighing unit with a SUS304 body (SUS316 weighing pan) protects the sensor from accidental
spilling of medical agents, resist ink, coating materials, etc.
iv
iv Not designed to handle impact loads.
*
Other Helpful Features
Securable to a base by removing the leveling feet and using the provided
mounting fixtures
Twelve units of weighing including g, mg, oz, ozt, ct, mom, dwt, GN, tael,
tola, PC (counting mode), Pct (percent mode)
Preventive function to inhibit unintended changes to important parameters
Stainless steel breeze break for better weighing stability
Comes with a free calibration weight (OIML Class E2 equivalent)
Stainless steel breeze break

Specifications
uments India Private e Li
Models
Weighing capacity
Minimum weighing value
Repeatability (standard deviation)
Linearity
Stabilization time
(when set to FAST under a good environment)
Sensitivity drift
Operating environment
Display refresh rate
Display mode
Counting mode
Percent mode
I/O unit
Calibration weight provided
(conforming to OIML Class E2)
Applicable calibration weight value
Weighing unit
Display unit
Connection cable
Power supply
Power consumption
v Smart range function: The display will switch to the standard range automatically when the value exceeds 31 g but return to the precision range when RE-ZERO (tare) is performed.
*
vi Repeatability for a 1 g mass using an automatic loading machine.
*
Options AD4212A-01
Accessories
Minimum unit mass
Number of samples
Minimum 100% reference mass
Minimum 100% display
RS-232C
Comparator contact output
External contact input
Dimensions
Weighing pan
Wei ght
Dimensions
Wei ght
BCD output / external contact input
AD4212A-07
AD4212A-08
AD4212A-19
AD4212A-20
AD-1683
AD-1684
AD-1688
AD-1689
AD-8121B
AD-8527
AD-8951
Extension cable (3 m)
Ethernet interface
(with application software WinCT-plus on CD-ROM)
Stainless steel breeze break
(comes as standard but also available as an option)
Metal leveling foot
Static eliminator
Electrostatic field meter
Weighing data logger
Tweezers for calibration weight
Compact printer
Quick USB adapter
Comparator light
AD-4212B-23
21 g
0.001 mg
0 - 10 g (gross) : 0.003 mg
10 - 21 g (gross) : 0.004 mg
±0.01 mg
Approx. 12 sec
1 mg
1 g
1 g, 2 g, 5 g, 10 g, 20 g
AD-4212B-101
31 g / 110 g
0.01 mg / 0.1 mg
vi
0.05 mg / 0.1 mg
vi
±0.05 mg / ±0.2 mg
Approx. 4.0 sec / 2.5 sec
g, mg, oz, ozt, ct, mom, dwt, GN, tael, tola, PC (counting mode), Pct (percent mode)
v
5 °C to 40 °C (41 °F to 104 °F), 85%RH or less (no condensation)
10 times/second (5 times/second can be selected)
0.01%, 0.1%, 1% (depends on the reference mass stored)
10 g, 20 g, 50 g, 100 g
Ø34 mm
Approx. 2 m (between the weighing unit and the display unit)
AD-4212B-102
110 g
0.01 mg
0 - 30 g : 0.02 mg
30 - 100 g : 0.04 mg
±0.1 mg
Approx. 4.0 sec
±2 ppm/°C (10 °C to 30 °C/50 °F to 86 °F)
0.1 mg
10, 25, 50 or 100 pieces
10.0 mg
Bi-directional, 600-19200 bps
3-level or 5-level (with buzzer)
RE-ZERO, PRINT
50 g
80(W) × 230(D) × 90(H) mm
Approx. 2.3 kg
237(W) × 150(D) × 155(H) mm
Approx. 2 kg
AC adapter
Approx. 15 VA (supplied to the AC adapter)
AD-4212B-201
210 g
0.1 mg
0.1 mg
±0.2 mg
Approx. 2.5 sec
10 g, 20 g, 50 g
100 g, 200 g
Dimensions(mm/inch)
AD-4212B-23/101/102
25/0.98
20/0.79
80/3.1580/3.15
70/2.76
176/6.93
230/9.06
210/8.27
Ø34/1.34
AD-4212B-301
310 g
0.1 mg
0.2 mg
±0.3 mg
Approx. 2.5 sec
100 g
10 g, 20 g, 50 g
100 g, 200 g, 300 g
50 × 50 mm
85~95)
*
90(
3.54(33.5~3.74)
62/2.44
9/0.35
(98/3.86)
Adjustable range
*
AD-8121B AD-1683 AD-1684
...Clearly a Better Value
3-23-14 Higashi-lkebukuro, Toshima-ku,Tokyo 170-0013 JAPAN
Telephone:[81](3) 5391-6132 Fax:[81](3) 5391-6148
http://www.aandd.jp
1756 Automation Parkway, San Jose, CA 95131 U.S.A.
Telephone:[1](408) 263-5333 Fax:[1](408) 263-0119
32 Dew Street, Thebarton, South Australia 5031 AUSTRALIA
Telephone:[61](8) 8301-8100 Fax:[61](8) 8352-7409
ia
AD-4212B-201/301
25 176 9
20/0.79
Unit 24/26 Blacklands Way Abingdon Business Park,
Abingdon, Oxon OX14 1DY UNITED KINGDOM
Telephone:[44](1235) 550420 Fax:[44](1235) 550485
<German Sales Office>
Hamburger Straße 30 D-22926 Ahrensburg GERMANY
Telephone:[49](0) 4102 459230 Fax:[49](0) 4102 459231
Manhattan Bldg. 8F, 36-2 Yoido-dong, Youngdeungpo-gu, Seoul, KOREA
Telephone:[82](2) 780-4101 Fax:[82](2) 782-4280
70/2.76
230/9.06
210/8.27
50/1.97
50/1.97
Vereyskaya str.17, Moscow, 121357 RUSSIA
Telephone: [7] (495) 937-33-44 Fax: [7] (495) 937-55-66
D Inst
509 Udyog Vihar Phase V
Gurgaon-122 016, Haryana, INDIA
Telephone: [91](124) 471-5555 Fax: [91](124) 471-5599
85~95)
*
90(
3.54(33.5~3.74)
62/2.44
(98/3.86)
Adjustable range
*
uments India Pr
AD4212B-ADCC-03-PR2-11a03
*
mited