Advanced 3250, 4250 User manual

Advanced
Advanced
®
Osmometer Model 3250
®
Cryoscope Model 4250
Service Manual
3255SM Rev22 020416
Table of Contents
1. Introduction
Specifications . . . . . . . . . . . . . . . . . . . . .7
General Description and Purpose . . . .11
Service Assistance . . . . . . . . . . . . . . . .11
Safety/Regulatory Information . . . . . .12
Symbol Conventions . . . . . . . . . . . . . .12
General Cautions . . . . . . . . . . . . . . . . .12
Service & Maintenance Cautions . . . .13
Hazardous Material Cautions . . . . . . .14
Regulatory Notices . . . . . . . . . . . . . . .14
Symbol Conventions . . . . . . . . . . . . . .14
FCC Requirements . . . . . . . . . . . . . . .15
2. General Overview
General Overview . . . . . . . . . . . . . . . .21
Calibration . . . . . . . . . . . . . . . . . . . . . .23
Circuit Description . . . . . . . . . . . . . . .24
Power Supply . . . . . . . . . . . . . . . . . . . .24
Front End . . . . . . . . . . . . . . . . . . . . . . .24
325605 Control Board Set . . . . . . . . . .24
325620/425620/325621/425621
Processor Board . . . . . . . . . . . . . . . .25
PCB605 Application Board . . . . . . . . .26
Other Circuit Components . . . . . . . . . .27
Design Changes . . . . . . . . . . . . . . . . . .29
Replacement Parts . . . . . . . . . . . . . . . .30
Supplies & Accessories . . . . . . . . . . . .32
3. Maintenance
Cleaning and Maintenance . . . . . . . . .35
Removing the Head & Instrument
Covers . . . . . . . . . . . . . . . . . . . . . . .36
Osmometer/Cryoscope Well Cleaning 37 Cleaning and Lubricating the Printer .39 Probe & Stir/Freeze Wire Alignment .40
4. Troubleshooting
Introduction . . . . . . . . . . . . . . . . . . . . .43
Freeze “Buzz” Amplitude . . . . . . . . . .44
Stir Driver Noise . . . . . . . . . . . . . . . . .45
Head Sensor Adjustment . . . . . . . . . . .47
Head Up/Down Test . . . . . . . . . . . . . .48
Power On Diagnostics . . . . . . . . . . . . .49
A/D Tests . . . . . . . . . . . . . . . . . . . . . . .49
Sample Bin Setting . . . . . . . . . . . . . . .50
Block Bin Setting . . . . . . . . . . . . . . . .51
Stir Amplitude . . . . . . . . . . . . . . . . . .51
Display/Printer Test . . . . . . . . . . . . . . .51
Beeper Test . . . . . . . . . . . . . . . . . . . . .52
Keypad Test . . . . . . . . . . . . . . . . . . . . .52
Barcode Test . . . . . . . . . . . . . . . . . . . .52
Reset Serial Number . . . . . . . . . . . . . .52
Instrument Software Updates . . . . . . .54
Performing FLASH firmware update .54
Troubleshooting Table . . . . . . . . . . . . .56
5. Replacement Instructions
Block Probe (4D3340) . . . . . . .325P340
Clapper (3C2241R) . . . . . . . .3C2P241R
Control Board Set (325605R/425605R),
Application Board (PCB605R), Processor Board (325621R/
425621R) . . . . . . . . . . . . . .325P620
Cooling Assembly (3D3300R/
4D3300R) . . . . . . . . . . . .325P300
Cooling Fan (4D3360) . . . . . . .325P360
Display Board (74051R/
702041R) . . . . . . . . . . . .325P051
Driver Board (PCB523R) . . .PCB523PM
Fuse (70011/70022) . . . . . . . . . . .325P11
Head Up Relay or Head Down
Relay (3D3109) . . . . . . . . . .325P109
Heat Transfer Fluid Pump
Assembly (4D3690), Fluid Filter (4D3710R), One-way Check
Valve (4D3705R/99133R) . . . .4DP690
Keypad (325511R) . . . . . . . . . .325P511
Load Resistor (425150R) . . . . .42P150R
Mandrel (3LH500) . . . . . . . . . .3LHP500
Motor Assembly (4D3350R/
325304R) . . . . . . . . . . . . . .325P350
NVRAM Battery (71027R) . . . .71P27R
Power Supply Assembly (4D3950/
325950R) . . . . . . . . . . . . . .325P950
Printer Assembly (325400R) . . .325P401 Printer Assembly (325420R) .325420PM Printer Mechanism
(325404R) . . . . . . . . . . . . . .325404PM
Probe Alignment Tool Kit
(3LA700) . . . . . . . . . . . . . . . . .3LP700
Sample Probe (3D3700) . . . . . . .3DP700
Sample Probe (3D3700) . . . . . .325P700
Sample Probe (4D3102) . . . . . . .4D3P02
Cryoscope Sample Probe
Versions FAQ . . . . . . . . . . . . .4DP002
Stir/Freeze Coil (3D2404R) . . .325P404
3
The Advanced®Model 3250/4250 Service Manual
Stir/Freeze Wire (3LH243/
4LH243) . . . . . . . . . . . . . . . .4LHP243
Transistor (PCB520R) . . . . . . .PCP520R
Transistor (PCB522R) . . . . . . .PCP522R
Tube Ejector Spring (3D3312R/
4D3312) . . . . . . . . . . . . . . . 4DP312
Upgrade Kit, 3250/4250 Suffix
D to E (325024) . . . . . . . . .325025PM
Yoke (3LH230R) . . . . . . . . . . .3LHP230
6. Schematics
PCB402 Assembly Drawing PCB402 Printer Interface Board
Schematic PCB402 Schematic PCB523 Assembly Drawing PCB523 Driver Board Schematic PCB523 Schematic PCB605 Assembly Drawing PCB605 Application Board Schematic PCB605 Schematic PCB620 Assembly Drawing PCB620 Processor Board Schematic 210102 Assembly Drawing 210102 Schematic 3D3380 Assembly Drawing 3D3380 Schematic 325000 Interconnection Schematic
APPENDICES
A: Symbol Definitions
B: Product Disposal and Recycling C: Service Log
Index
The Advanced®Model 3250/4250 Service Manual
4
1. Introduction
Model 3250 Osmometer Specifications
Applies to Serial Suffix A - C. Refer to Design Changes for more information.
Electrical:
Voltage: 100 to 130 VAC (1.2 A) or 200 to 250
VAC (0.6 A)
Frequency: 50 or 60 Hz
Fuses: 2-Amp time delay for 100-130 VAC;
1-Amp time delay for 200-250 VAC
Power: 150 Watts
Memory Backup: integral lithium cell; 10 years
min. (not user-replaceable)
Sample Volume: 0.2 or .25 mL
Sample Capacity: single sample
Readout: 20-character digital display
Units: mOsm/Kg H2O
Range:
Low: 0 to 2000 mOsm/Kg H2O High: 1400 to 4000 mOsm/Kg H2O
Resolution: 1 mOsm/Kg H2O
Performance Over Operating Conditions
Temperature Effects: Less than 1 mOsm/Kg H
for every 5°C (9°F) ambient temperature change.
Operating Conditions
Temperature: 18°C to 35°C (64°F to 95°F)
Room Humidity: 5 to 80% relative humidity;
(non-condensing)
Storage Temperature: -40°C to +45°C
(-40°F to +113°F)
Start-up Time: From stand-by: immediately
From power on: under 5 minutes
Test Time: Approximately 2 minutes per sample
in 0 to 2000 mOsm range; moderately longer in high range.
Dimensions inches centimeters
Width: 13 32 Depth: 18 46 Height: 16 41
O
2
Communications:
RS-232C and barcode scanner ports
Weight pounds kilograms
Net: 28 13 Shipping: 39 19
Performance at Reference Conditions
1
Warranty: One-year limited warranty on work-
Linearity: Less than ±0.5% from a straight line
manship and all parts except glass, plastic and parts warranted by their
Repeatability:
makers.
0 to 400 mOsm ±2 mOsm/Kg (1 S.D.) 400 to 4000 mOsm ±0.5% (1 S.D.)
Drift: Less than 1 mOsm/Kg H
O per
2
Certification:
month
Refer to Regulatory Notices (see page 14) for appli­cable standards.
Installation Class: I Over-Voltage Category: II Pollution Degree: 2 Moisture Protection: IPX0 (Ordinary)
1
Reference Conditions: 20 to 25°C (68 to 77°F); 40 to 60% Relative Humidity; tolerances of reference or calibration solutions
excluded.
7
The Advanced®Model 3250/4250 Service Manual
Model 3250 Osmometer Specifications
Applies to Serial Suffix D and Higher. Refer to Design Changes for more information.
Electrical:
Voltage: 100 to 250 VAC (0.6 A)
Frequency: 50 or 60 Hz
Fuses: 2-Amp time delay (Type T)
Power: 95 Watts
Memory Backup: integral lithium cell; 10 years
min. (not user-replaceable)
Sample Volume: 0.2 or .25 mL
Sample Capacity: single sample
Readout: 20-character digital display
Units: mOsm/Kg H2O
Range:
Low: 0 to 2000 mOsm/Kg H2O High: 1400 to 4000 mOsm/Kg H2O
Resolution: 1 mOsm/Kg H2O
Communications:
RS-232C and barcode scanner ports
Performance at Reference Conditions
1
Linearity: Less than ±0.5% from a straight line
Repeatability:
0 to 400 mOsm: Std. Deviation < 2 mOsm/Kg H2O 400 to 4000 mOsm: Std. Deviation < 0.5% of value
mOsm/Kg H2O
Drift: Less than 1 mOsm/Kg H
O per
2
month
Performance Over Operating Conditions
Temperature Effects: Less than 1 mOsm/Kg H
2
for every 5°C (9°F) ambient temperature change.
Operating Conditions
Temperature: 18°C to 35°C (64°F to 95°F)
Room Humidity: 5 to 80% relative humidity;
(non-condensing)
Storage Temperature: -40°C to +45°C
(-40°F to +113°F)
Start-up Time: From stand-by: immediately
From power on: under 5 minutes
Test Time: Approximately 2 minutes per sample
in 0 to 2000 mOsm range; moderately longer in high range.
Dimensions inches centimeters
Width: 13 32 Depth: 18 46 Height: 16 41
Weight pounds kilograms
Net: 23 10 Shipping: 34 15
Warranty: One-year limited warranty on work-
manship and all parts except glass, plastic and parts warranted by their makers.
Certification:
O
Refer to Regulatory Notices (see page 14) for appli­cable standards.
Installation Class: I Over-Voltage Category: II Pollution Degree: 2 Moisture Protection: IPX0 (Ordinary)
1
Reference Conditions: 20 to 25°C (68 to 77°F); 40 to 60% Relative Humidity; tolerances of reference or calibration solutions
excluded.
The Advanced®Model 3250/4250 Service Manual
8
Model 4250 Cryoscope Specifications
Applies to Serial Suffix A - C. Refer to Design Changes for more information.
Electrical:
Voltage: 100 to 130 VAC (1.2 A) or 200 to 250
VAC (0.6 A)
Frequency: 50 or 60 Hz
Fuses: 250V time delay (Type T):
2-Amp time delay for 100-130 VAC; 1-Amp time delay for 200-250 VAC
Power: 150 Watts
Memory Backup: integral lithium cell; 10-year
life (typical); (not user­replaceable)
Sample Volume: 2.0 to 2.5 mL
Sample Capacity: single sample
Readout: 20-character vacuum fluorescent
display
Units: mºC or mºH
Range: 0 to 1000 mºC or mºH
Resolution: 1 mºC or mºH
Communications: On-board printer, DTE EIA-
232/V.24 (RS-232) serial port and optional barcode scanner
Performance at Reference Conditions
Linearity: Less than ±0.5% from a straight line
1
Performance Over Operating Conditions
Temperature Effects: Less than 1 unit for every
5°C (9°F) ambient tempera ture change.
Operating Conditions
Temperature: 18°C to 35°C (64°F to 95°F)
Room Humidity: 5 to 80% relative humidity;
(non-condensing)
Storage Temperature: -40°C to +45°C
(-40°F to +113°F)
Start-up Time: From stand-by: immediately
From power on: under 5 minutes
Test Time: Approximately 90 seconds in 30-
second timed mode.
Dimensions inches centimeters
Width: 13.0 33.0 Depth: 18.0 45.7 Height: 12.0 30.5
with Head Up: 16.0 40.6
Weight pounds kilograms
Net: 28.0 12.7 Shipping: 39.0 17.7
Warranty: One-year limited warranty on work-
manship and all parts except glass, plastic and parts warranted by their makers.
Repeatability:
± 2 mºC or mºH (1 S.D.) between 0 and -400 mºC ± 0.5% (1 S.D.) between -400 and -1000 mºC
Drift: Less than 1 unit per month
Certification:
Refer to Regulatory Notices (see page 14) for appli­cable standards.
Installation Class: I
Over-Voltage Category: II
Pollution Degree: 2
Moisture Protection: IPX0 (ordinary)
9
The Advanced®Model 3250/4250 Service Manual
Model 4250 Cryoscope Specifications
Applies to Serial Suffix D and Higher. Refer to Design Changes for more information.
Electrical:
Voltage: 100 to 250 VAC (0.6 A)
Frequency: 50 or 60 Hz
Fuses: 2-Amp time delay (Type T)
Power: 95 Watts
Memory Backup: integral lithium cell; 10-year
life (typical); (not user­replaceable)
Sample Volume: 2.0 to 2.5 mL
Sample Capacity: single sample
Readout: 20-character vacuum fluorescent
display
Units: mºC or mºH
Range: 0 to 1000 mºC or mºH
Resolution: 1 mºC or mºH
Communications: On-board printer, DTE EIA-
232/V.24 (RS-232) serial port and optional barcode scanner
Performance at Reference Conditions
1
Performance Over Operating Conditions
Temperature Effects: Less than 1 unit for every
5°C (9°F) ambient tempera ture change.
Operating Conditions
Temperature: 18°C to 35°C (64°F to 95°F)
Room Humidity: 5 to 80% relative humidity;
(non-condensing)
Storage Temperature: -40°C to +45°C
(-40°F to +113°F)
Start-up Time: From stand-by: immediately
From power on: under 5 minutes
Test Time: Approximately 90 seconds in 30-
second timed mode.
Dimensions inches centimeters
Width: 13.0 33.0 Depth: 18.0 45.7 Height: 12.0 30.5
with Head Up: 16.0 40.6
Weight pounds kilograms
Net: 23 10 Shipping: 34 15
Linearity: Less than ±0.5% from a straight line
Repeatability:
0 to -400 mºC: Std. Deviation <
-400 to -1000 mºC: Std. Deviation < 0.5% of value mºC or mºH
Drift: Less than 1 unit per month
The Advanced®Model 3250/4250 Service Manual
2 mºC or mºH
Warranty: One-year limited warranty on work-
manship and all parts except glass, plastic and parts warranted by their makers.
Certification:
Refer to Regulatory Notices (see page 14) for appli­cable standards.
Installation Class: I
Over-Voltage Category: II
Pollution Degree: 2
Moisture Protection: IPX0 (ordinary)
10
General Description and Purpose
Service Assistance
®
Advanced
Osmometers and Cryoscopes
determine the concentration of solutions through freezing-point depression measure­ment. Osmometers are used primarily to measure body fluids such as serum and urine, and cryoscopes are used to determine the freezing points of various liquids and to detect added water in milk. Application infor­mation is available upon request.
Except for different cooling assemblies, stir/freeze wires, sample sizes and software, the 3250 and the 4250 are identical to one another. For clarity when replacing parts or troubleshooting, the instructions indicate any significant differences between the two instru­ments.
The Service Manual
This Service Manual contains:
To contact Advanced Instruments Product Service:
• Call (800) 225-4034 (toll-free within the USA and Canada; after normal business hours, dial extension 2191)
• Call +US (781) 320-9000 (elsewhere)
• Or fax (781) 320-3669
When calling Advanced Instruments, have the model and serial number from the label on the back of the instrument, and a description of the problem. Use a telephone as close to the instrument as possible in order to facilitate making recommended diagnostic checks.
A service technician may assist in making minor repairs over the phone, providing you with recommended parts (or part numbers), or may issue an authorization (RMA) to ship the instrument for factory repair.
• instructions for performing repairs.
• guides for troubleshooting operational and mechanical problems.
• circuit system illustrations.
• information about replacement parts and service repairs.
This manual is designed to assist service technicians, and does not imply a license to perform repairs without proper qualifications.
The RMA procedure is as follows:
1. Contact Advanced Instruments to obtain an RMA.
2. Pack and ship the instrument in its origi­nal carton. Do not send instrument sup­ply items.
3. Prepay shipment to the factory. Advanced Instruments cannot accept collect ship-
ments. Insure the shipment or accept the damage risk.
11
The Advanced®Model 3250/4250 Service Manual
Safety / Regulatory Information
To reduce the risk of bodily injury, electric shock, fire, and damage to your instrument, please read and observe the following precautions.
• If the product is used in a manner not in accordance with the equipment design, operating instructions or manufacturer's recommendations, the operation of the product may be impaired to the extent that a safety hazard is created.
• Do not attempt to perform electrical work if you are not fully qualified. This manual is not a substitute for electrical training.
Symbol Conventions
The exclamation point within an equilateral triangle is intended to alert the user to the presence of important operating and maintenance (servicing) instructions in the litera­ture accompanying this product.
The lightning flash with arrowhead symbol within an equilateral triangle is intended to alert the user to the presence of uninsulated dangerous voltage within the product's enclo­sure that may be of sufficient magni­tude to constitute risk of electric shock to persons.
The static symbol within an equilat­eral triangle is intended to alert the user to the presence of internal com­ponents that could be damaged by static electricity.
This symbol indicates the presence of a fuse.
This symbol indicates the presence of protective earth ground.
This symbol indicates the power is ON.
This symbol indicates the power is OFF.
NOTE Additional symbol definitions are
available in the Symbols section.
General Cautions
• This product should be operated only with the type of power source indicated on the product’s electrical ratings label. Refer to the installation instructions included with the product.
• If the power cord provided is replaced for any reason or if an alternate cord is used, the cord must be approved for use in the local country. The power cord must be approved for the product’s listed operating voltage and be rated at least 20% greater than the ampere ratings marked on the product’s electrical ratings label. The cord end that connects to the product must have an IEC 60320 connector.
• Plug the product into an approved ground­ed electrical outlet.
This static symbol is intended to alert the user to the presence of a specific component that could be damaged by static electricity.
This symbol indicates the presence of alternating current (AC).
The Advanced®Model 3250/4250 Service Manual
• Do not disable the power cord’s grounding plug.
• If an extension cord or power strip is used, make sure that the cord or strip is rated for the product, and that the total ampere rat­ings of all products plugged into the exten­sion cord or strip do not exceed 80% of the cord’s or strip’s rating limit.
12
• Route power cords so that they will not be walked on, tripped on, or pinched by items placed upon or against them. Pay particu­lar attention to the plug, electrical outlet, and the point where the cord exits the product.
• Do not pull on cords and cables. When unplugging cords or cables, grasp the cor­responding connector.
• Do not install or use this product in any area subject to extreme short-term temper­ature variations, or locations that exceed the specified operating environment tem­peratures.
• Never use this product in a wet area.
- The power cord, extension cord, power strip or power input module is dam­aged.
- Liquid has been spilled into the interior of the product.
- A foreign object has fallen into the product.
- The product has been dropped or dam­aged by a falling object.
- There are noticeable signs of overheat­ing or a burning odor.
- The product does not operate normally when you follow the operating proce­dures.
• To avoid injury or fire hazard, do not oper­ate this product in an explosive atmos­phere.
• Do not install or use the product on an unstable, non-level work surface.
• Do not operate this product with the covers removed or unsecured.
Service & Maintenance Cautions
Unplug the power cord prior to opening or removing covers, or else you may be exposed to electric shock, excessive tem­peratures, or mechanical hazards.
Performing service or maintenance not detailed in the User’s Guide, with or with­out this Service Manual, should only be done by a qualified service technician.
Never restrict airflow into or out of the product. Occasionally, check the air vents
for blockage.
Wipe the exterior of the product with a
soft, damp cloth as needed. Using cleaning products other than those specified, may discolor or damage the finish.
If the product requires service for any of
the following reasons, unplug the product from the electrical outlet and refer service to a qualified service technician.
- The main supply fuse(s) or any internal fuse(s) continually fail.
A discharge of static electricity from con­tact with the human body or other conduc­tor may damage system boards or static sensitive devices. Never perform internal maintenance without following recom­mended static protection procedures.
• The product is equipped with operator accessible fuses. If a fuse blows, it may be due to a power surge or failure of a com­ponent. Replace the fuse only once. If the fuse blows a second time, it is probably caused by failure of a component part. If this occurs, refer service to qualified serv­ice personnel. Always replace the fuse with one of the same rating, voltage, and type. Never replace the fuse with one of a higher current rating.
• When servicing the product, use only fac­tory-specified parts.
• WARNING: When returning this product for service, or shipping this product to a second location, remove all hazardous specimens and decontaminate the product before packaging for shipment. If the product cannot be decontaminated, consult with your shipping agent on appropriate packaging and marking.
13
The Advanced®Model 3250/4250 Service Manual
Hazardous Material Cautions
WARNING: Handle all biohazardous materials according to established good laboratory practices and follow your insti­tution’s exposure control plan. Persons handling human blood and body fluid samples must be trained in blood-borne hazards and observe universal precautions. Universal precautions is an approach to infection control, where all human blood and body fluids are treated as if known to be infectious. Use personal protective equipment such as gloves, gowns, etc., to prevent exposure. Store biohazardous materials in regulated waste containers and dispose of these materials in a safe and acceptable manner that is in compliance with all country, state and local require­ments.
If a biohazardous material is spilled on or inside the equipment, decontaminate the equipment using a 1% bleach solution, or as outlined by those policies and proce­dures established within your institution.
To avoid injury or fire hazard, do not oper­ate this product in an explosive atmos­phere.
Regulatory Notices
This product has been designed and manu-
factured in accordance with U.S., Canadian, and European regulatory requirements as outlined below. Modifications made to this product that are not expressly approved in writing by the manufacturer will void the user’s authority to operate this product, previously issued factory approvals, and the user’s rights under the warranty.
The distributor or dealer may have applied
additional local, national, or international approvals to this product. Consult the dis­tributor or dealer for more information and documentation.
Connections to this product must be made with shielded cables. Use of non-shielded cables may violate RFI/EMI limits.
Symbol Conventions
This symbol indicates conformity to relevant European directives.
This symbol indicates the product was tested to conform to relevant Canadian and U.S. safety standards by Intertek Testing Services NA, Inc. The ETL mark is approved in the United States as a Nationally Recognized Testing Lab (NRTL) by OSHA, and in Canada by the Standards Council of Canada.
In Vitro Diagnostic Medical Device complying with EU Directive 98/79/EC.
This symbol, depending upon its location, indicates the assembly, component, and/or product are free of lead. In respect to its inclusion on a circuit component or assembly, the following must be adhered to when servicing. Failure to do so may result in defects caused by dis­similar metals.
1. Do not repair or modify using a lead-containing solder.
2. Do not repair or modify using any soldering tools that have been in contact with lead-containing sol­ders.
3. Do not repair or modify with any lead-free solder other than that specified by the manufacturer or as indicated by the symbology of standard IPC-1066.
4. Do not repair or modify using any soldering tools that have been in contact with any other incompati­ble lead-free solders.
5. Contact factory service for further information.
The Advanced®Model 3250/4250 Service Manual
14
The number in this symbol indicates the type of lead-free solder used in the assembly of printed circuit boards, per standard IPC-1066. Repair or modification must be per-
XAMPLES
E
ONLY
formed using the same or compatible type lead-free solder and matching tools. Failure to do so may result in defects caused by dissimilar metals.
Contact factory service for further information.
FCC Requirements
WARNING: Changes or modifications to this unit not expressly approved by Advanced Instruments could void the user’s authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential instal­lation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accor­dance with the instructions, may cause harmful interference to radio communica­tions. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interfer­ence by one or more of the following measures:
- Re-orient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio TV technician for help.
15
The Advanced®Model 3250/4250 Service Manual
Regulatory Description
Approval Type Applies to Serial Suffix A - C Applies to Serial Suffix D and Higher
U.S. Safety
Canadian Safety
CE Declaration of Conformity - EMC
CE Declaration of Conformity - IVD
(3250 only)
This product has been listed by ETL testing labo­ratories as being in compliance with the require­ments of UL 61010A-1, 1st Edition, "Electrical Equipment for Laboratory Use; Part 1: General Requirements". The "US" in the lower right of the ETL mark demonstrates this listing.
This product has been listed by ETL testing labo­ratories as being in compliance with the require­ments of CAN/CSA C22.2 No.1010.1-92, "Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 1: General Requirements"; Including Amendment Two. The "C" in the lower left of the ETL mark demonstrates this listing.
This product meets the intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was demonstrated using the following standards, as listed in the Official Journal of the European Communities: Consult the Declaration of Conformance certificate shipped with the prod­uct for the latest update.
• EN 61326: 1997 with A1 & A2, Group 1, Class B, "Electrical Equipment for Measurement, Control, and Laboratory Use"
This product meets the intent of Directive 98/79/EC for In Vitro Diagnostic Medical Devices. Consult the Declaration of Conformance certificate shipped with the product (if required) for the latest update.
This product has been listed by ETL testing labo­ratories as being in compliance with the require­ments of UL 61010-1, "Electrical Equipment for Laboratory Use; Part 1: General Requirements". The "US" in the lower right of the ETL mark demonstrates this listing.
This product has been listed by ETL testing labo­ratories as being in compliance with the require­ments of CAN/CSA C22.2 No.61010.1, "Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 1: General Requirements"; Including Amendment Two. The "C" in the lower left of the ETL mark demonstrates this listing.
This product meets the intent of Directive 2004/108/EC Conformity - EMC for Electromag­netic Compatibility. Compliance was demonstrat­ed using the following standards, as listed in the Official Journal of the European Communities: Consult the Declaration of Conformance certifi­cate shipped with the product for the latest update.
• EN 61326-1:2006 & EN 55011:1998, Group 1, Class B, "Electrical Equipment for Measurement, Control, and Laboratory Use"
This product meets the intent of Directive 98/79/EC for In Vitro Diagnostic Medical Devices. Consult the Declaration of Conformance certificate shipped with the product (if required) for the latest update.
CE Declaration of Conformity - WEEE
CE Declaration of Conformity - Low Voltage
(3250 only)
This product meets the intent of Directive 2002/96/EC as amended by 2003/108/EC for Waste Electrical and Electronic Equipment (WEEE). Consult the Declaration of Conformance certificate shipped with the product (if required) for the latest update.
This product meets the intent of Directive 73/23/EEC, the Low Voltage Directive. Compliance was demonstrated using the following standards, as listed in the Official Journal of the European Communities: Consult the Declaration of Conformance certificate shipped with the prod­uct (if required) for the latest update.
• EN 61010-1 (2001), “Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 1: General Requirements”.
• EN 61010-2-101: 2002, “Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 2-101: Particular Requirements for In Vitro Diagnostic (IVD) Medical Equipment”.
This product meets the intent of Directive 2002/96/EC as amended by 2003/108/EC for Waste Electrical and Electronic Equipment (WEEE). Consult the Declaration of Conformance certificate shipped with the product (if required) for the latest update.
This product meets the intent of Directive 2006/95/EC, the Low Voltage Directive. Compliance was demonstrated using the follow­ing standards, as listed in the Official Journal of the European Communities: Consult the Declaration of Conformance certificate shipped with the product (if required) for the latest update.
• EN 61010-1, “Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 1: General Requirements”.
• EN 61010-2-101, “Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 2-101: Particular Requirements for In Vitro Diagnostic (IVD) Medical Equipment”.
The Advanced®Model 3250/4250 Service Manual
16
Regulatory Description
Approval Type Applies to Serial Suffix A - C Applies to Serial Suffix D and Higher
FCC - Part 15 Subpart B, Class B
Canadian ICES-003
CB Report CB Certificate
U.S. FDA Listing
(3250 only)
Health Canada License
(3250 only)
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
This Class B digital apparatus complies with Canadian ICES-003
Cet appareil numérique de la classe B est con­forme à la norme NMB-003 du Canada.
-------------------------------
The osmometer, along with the calibrators and controls manufactured by Advanced Instruments, are listed with a U.S. Department of Health and Human Services, Food and Drug Administration, as:
Osmometer Class 1 Calibrators Class 2 Controls Class 1
The osmometer, along with the calibrators and controls manufactured by Advanced Instruments, are licensed with Health Canada, Therapeutic Products Directorate, Medical Devices Bureau, as:
Osmometer Class 2 Calibrators Class 2 Controls Class 2
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
This Class B digital apparatus complies with Canadian ICES-003
Cet appareil numérique de la classe B est con­forme à la norme NMB-003 du Canada.
A CB report and certificate have been issued for this product. The standard used was IEC 61010-1.
The osmometer, along with the calibrators and controls manufactured by Advanced Instruments, are listed with a U.S. Department of Health and Human Services, Food and Drug Administration, as:
Osmometer Class 1 Calibrators Class 2 Controls Class 1
The osmometer, along with the calibrators and controls manufactured by Advanced Instruments, are licensed with Health Canada, Therapeutic Products Directorate, Medical Devices Bureau, as:
Osmometer Class 2 Calibrators Class 2 Controls Class 2
CE Declaration of Conformity - RoHS
This product meets the intent of Directive 2002/95/EC for the Restriction of Use of Certain Hazardous Substances in Electrical and Electronic Equipment” as an exempt medical device per Article 2, Paragraph 1 and per Annex 1, Category 8 and/or 9 of Directive 2002/96/EC.
This product meets the intent of Directive 2002/95/EC for the Restriction of Use of Certain Hazardous Substances in Electrical and Electronic Equipment” as an exempt medical device per Article 2, Paragraph 1 and per Annex 1, Category 8 and/or 9 of Directive 2002/96/EC.
17
The Advanced®Model 3250/4250 Service Manual
Notes:
The Advanced®Model 3250/4250 Service Manual
18
2. General Overview
General Overview
(Intel Processor Systems)
Serial Suffix A - C
L
K
J
I
H
G
F
E
D
M
N
O
Q
P
R
S
T
C
B
A. Head B. Motor C. Driver Transistors (Stir, Thermoelectric) D. Motor Capacitor E. Head Up/Head Down Relays F. Power Entry Module G. RS-232C Port H. Supervisor/Operator Keyswitch I. Barcode Port J. Power Supply K. Sample Probe Plugin
21
U
A
L. Block Probe Plugin M. Main Board Set, Application PCB N. Main Board Set, Processor PCB O. Printer P. Keypad Transition Board Q. Keypad R. Display S. Deck T. Heat Transfer Return Tube U. Heat Transfer Inlet Tube
The Advanced®Model 3250/4250 Service Manual
General Overview
(Intel Processor Systems)
Serial Suffix D and Higher
H
(underneath printer)
G
F
E
D
C
J
I
K
L
M
N
O
P
Q
R
B
A. Head B. Motor C. Driver PCB D. Power Entry Module E. RS-232C Port F. Supervisor/Operator Keyswitch G. Barcode Port H. Power Supply (underneath printer) I. Sample Probe Plugin J. Block Probe Plugin
The Advanced®Model 3250/4250 Service Manual
A
S
K. Main Board Set, Processor PCB L. Printer M. Main Board Set, Application PCB N. Keypad Transition Board O. Keypad P. Display Q. Heat Transfer Return Tube R. Heat Transfer Inlet Tube S. Deck
22
Calibration
The 3250 has two ranges, each with its own calibration. To calibrate the low range, three freezing-point tests are required at the 100 mOsm/kg level, and six tests are required at the 1500 mOsm/kg level. To calibrate the high range, six freezing-point tests are required at both the 1500 mOsm/kg level and the 3000 mOsm/kg level.
The 4250 allows timed or plateau-seeking cal­ibration in either the Hortvet (m°H) or the Celsius (m°C) scale. Six freezing-point tests are required at each of the 4250’s two calibra­tion levels:
-408 m°C (-422 m°H), and
-600 m°C (-621 m°H).
Note: If you see the message “Recalibration
Needed”, follow the instrument prompts to complete the initial calibra­tion. Your calibration may vary slightly from that mentioned here.
Calibrate the instrument according to the instructions below:
1. Turn the instrument on.
2. If there is a Supervisor/Operator key­switch, turn it to the Supervisor position. It must remain there throughout calibra­tion.
3. Press CALIB to begin the calibration sequence.
4. The display will prompt you at each cali­bration level. Place a sample of the first calibration standard into the freezing chamber and press START.
5. Continue loading and testing samples at the first calibration level until the display prompts you for samples at the second calibration level.
Refer to the instrument User’s Guide for addi­tional information on calibration.
6. Place the sample of the second calibration standard into the freezing chamber and press START.
7. When the instrument has completed cali­bration, the display will read "Calibration Complete".
23
The Advanced®Model 3250/4250 Service Manual
3250/4250 Circuit Description
The instrument is comprised of the following functional blocks:
Power Supply
Serial Suffix A - C
The power entry assembly interfaces the AC voltage presented to the instrument with the transformer. The transformer provides 120 VAC to the power supply board and the head motor. The power supply provides the power to run the main printed circuit board (±12 VDC, ±5 VDC), the thermoelectrics (5 VDC), the fan (12 VDC), the stir/freeze coil (12 VDC), and the internal printer (5 VDC).
Note: The internal printer may be powered by
+5 VDC (red/orange wire) or +12 VDC (violet wire). Consult system schematics for proper service.
Power Supply
Serial Suffix D and Higher
VDC), the stir/freeze coil (12 VDC), the inter­nal printer (12 VDC), and the motor and con­trol circuits (+24 V).
Front End
The sample thermistor measures the tempera­ture of the sample. The thermoelectrics (whose duty cycle is controlled through the block probe/thermistor) supercool the sample. The stir/freeze wire constantly stirs the sample to guarantee uniform cooling. When the supercooled sample reaches a set temperature (3000 mOsm for the low range and 4800 mOsm for the high range, by default, in the 3250, and 3000 mºC in the 4250), the stir/freeze wire "buzzes" causing the sample to freeze. During a test, the fan duty cycle tracks the thermoelectrics duty cycle -10%. If duty cycle drops below 20%, the fan is off. In between tests, the fan runs as required to maintain the software controlled target tem­peratures.
The power entry assembly interfaces the AC voltage presented to the instrument with the AC/DC switching power supply. The power supply provides the power to run the main printed circuit board set (±12 VDC, +5 VDC), the thermoelectrics (5 VDC), the fan (12
Barcode
Port
Serial Port
Head Motor
Controls
Processor
Board
Applications
Board
325605 Control Board Set Description
General: The control board set is made up of two printed circuit boards in a motherboard/ daughterboard configuration.
Power
Supply
Cooling
System
PrinterKeypadDisplay
The Advanced®Model 3250/4250 Service Manual
24
325620/425620/325621/425621 Processor Board
Processor: The processor used is an Intel 80C186 16-bit embedded processor. The 80C186 contains three programmable 16-bit timers, two serial ports, programmable inter­rupts, 1 mega-byte of memory address space, and 64 kilo-bytes of input/output (I/O) address space. The processor uses an external 32 MHz crystal to generate the internal 16 MHz system clock. The processor controls access to all memory and all I/O.
Memory Map: The 1 mega-byte of memory address space contains read-only memory (Flash EPROM), read/write memory (static RAM), and the real time clock. About half of the address space is unused, allowing for future expansion. The read-only memory is divided into four sections: reset vector, boot code, parameter blocks (unused), and appli­caiton code. The reset vector tells the proces­sor where to first start executing code; in this case, the boot code is executed first after reset. The boot code determines if new soft­ware will be downloaded by checking if the dip switch is in the “PROGRAM” position. If it is, the boot code waits for new software to be downloaded through the serial port. If not, the application code starts.
The real time clock maintains the current date and time, and contains nonvolatile memory where the instrument’s operation parameters are stored. A internal lithium battery pre­serves the contents of the nonvolatile memory when the instrument is turned off. Finally, the read-write memory is used for stacks, vari­ables, etc., except for the small section allo­cated for the interrupt vectors. The interrupt vectors tell the processor what code to run when an interrupt occurs, and functions simi­larly to the reset vector.
Input/Output Map: The 64 kilo-bytes of I/O address space control access to all periph­erals such as the serial ports, application board, etc. The processor uses 256 bytes, called the Peripheral Control Block, to control the integrated serial ports, timers, interrupts, chip selects, and I/O pins. The Peripheral Control Block is fully described in the 80C186EB/80C188EB Microprocessor User’s Manual. The application board’s registers and the CPU board’s registers control access to various functions and are described later.
Memory: Two types of memory are avail­able: read-only and read/write. The chips used here have 8-bit wide data buses. Used in pairs, the memory becomes compatible with the processor’s 16-bit wide data bus. The read-only memory is implemented using two 128 kilo-bytes x 8 bits Flash EPROMs. Flash memory is only programmable when a pro­gramming voltage (in this case, 12 volts DC) is applied to the memory; otherwise, it behaves like read-only memory. The boot code controls the programming voltage, since this is the only place where new software can be downloaded. The read/write memory is implemented using two 128 kilo-bytes x 8 bits static RAMs.
Watchdog Timer: The watchdog timer pro­vides a mechanism to reset the processor when the software is not behaving normally. The software “pets” the watchdog every 100 ­200 milliseconds. As long as the watchdog is “petted”, nothing happens. However, if the software stops “petting” the watchdog, the watchdog “bites” the processor by activating the processor’s RESET pin, causing the processor to reset. The watchdog will also reset the processor if the power supply falls below 4.75 VDC. These two features allow the instrument to recover from abnormal soft­ware and hardware conditions.
Please note that all memory addresses and sizes are always defined in terms of bytes, even though the processor handles words (1 word = 2 bytes, 1 byte = 8 bits; therefore, 16 bits = 1 word).
25
Real Time Clock (RTC): The real time
clock maintains the current date and time. The RTC uses an internal clock circuit with an internal, 10-year life, lithium battery to
The Advanced®Model 3250/4250 Service Manual
perform timekeeping. The battery also pre-
erves the contents of internal memory.
s
Dip Switch: The three-position or four­position dip switch allows the user to down­load new software via one switch. The other two switches are available for selecting differ­ent operating modes. These operating modes and the process for downloading software are described later in this manual.
Voltages: Three voltages are present on the CPU board: +5 VDC, +12 VDC, and V
.
PP
+5 VDC powers all the logic. +12 VDC is switched on and off to create the flash memo­ry’s programming voltage, V
.
PP
Glue Logic: The glue logic performs the functions of creating the RAM memory chip selects, the Flash memory write signals, the V
control signal, the watchdog control sig-
PP
nal, and accessing the dip switch. The CPU board has two software accessible hardware registers to read the dip switch and to control watchdog and V
.
PP
Connectors: A 64-pin connector provides address, data and interrupt signals. A 16-pin connector provides general chip selects and serial receive/transmit signals. The applica­tion PCB uses a subset of the available sig­nals.
to a 1 mOsm/1.86mºC change. A separate
heatstone bridge circuit is used to measure
W the voltage across each thermistor probe. This voltage is first filtered and then sampled by the analog-to-digital converter. The ana­log-to-digital converter uses a sigma-delta conversion technique with on-chip filtering and a 6.25 VDC reference voltage.
I/O Ports: Circuitry for the two serial ports (RS-232 and barcode) and the parallel printer is present on the application board, with cabling to the actual connectors on the back panel and the internal printer. Display & Keypad: The display and keypad are interfaced to the processor via program­mable logic devices (PLDs). The keypad is polled in a row/column process which identi­fies the key pressed to the microprocessor. The application logic, in response to proces­sor commands, controls the LEDs located on the keypad.
Light Emitting Diodes (LED): LEDs are provided to indicate when a high current load is active, a sensor is tripped, or a voltage sup­ply is active. Green LEDs are used on the power supply voltages, yellow LEDs are used on driver signals such as the fan and head motor controls, and red LEDs are used to indicate that the head up and down optical sensors have been tripped.
PCB605 Application Board
The application PCB contains all circuitry specific to the instrument. The board includes voltage supplies, indicators, analog-to-digital converters, drive circuitry, parallel ports, seri­al ports, and various switches.
Analog-to-Digital Converters: Two analog­to-digital converters are provided; one for the sample thermistor probe, and one for the block thermistor probe. The thermistors vary in resistance from approximately 2 kilo-ohms at room temperature, to approximately 10 kilo-ohms at -12ºC. Typically, a 0.6 ohm change in the thermistor’s resistance equates
The Advanced®Model 3250/4250 Service Manual
Application Logic: The application logic is made up mainly of two programmable logic devices (PLDs). These provide software accessible hardware registers, enabling the software to read the keypad, sensors, and ana­log-to-digital converters. These also allow manipulation of the display, external compo­nents such as the fan, head motor, etc., and interface to the RS-232 ports. The printer is controlled directly from the microprocessor through an 8-pin D-latch.
Connectors: Connectors are supplied for the CPU PCB, display, keypad, power, drives, probes, and back panel I/O ports.
26
Serial (RS-232) Port Interface: An RS-232
ine driver/receiver provides the microproces-
l sor with a serial port interface that supports both hardware and software handshaking.
The DB-9 RS-232 port conforms to the DTE RS-232C standard and has the following pin assignments:
The stir wire coil is controlled by the rear
anel-mounted darlington transistor
p (PCB522). The drive circuit consists of a programmable timer, D/A converter, and interface op-amps. The timer provides a square wave of approximately 71Hz, while the D/A converter controls the output ampli­tude to the darlington transistor.
Signal Pin Direction
Carrier Detect 1 to 3250 Receive Data 2 to 3250 Transmit Data 3 from 3250 Data Terminal Ready 4 from 3250 Signal Ground 5 common Data Set Ready 6 to 3250 Request to Send 7 from 3250 Clear to Send 8 to 3250
Note that your instrument is only designed to support unidirectional communication with an external device. At this time, there is no pro­tocol for bidirectional communication.
For a sample RS-232 Port Setup, please see the RS-232 Supplemental Information in the Appendix at the end of this user’s guide.
Note: This instrument requires the use of a
null modem RS-232C cable. There are several variations on null modem cables. Advanced Instruments recom­mends that you purchase a RS-232C cable direct from our factory.
Supervisor/Operator Keyswitch Interface:
A PLD is used to interface the supervisor/ operator keyswitch to the microprocessor. In Operator position, the user cannot change setup or calibration settings.
Other Circuit Components
Drive Circuitry, Serial Suffix A - C: Drive circuitry is provided to turn on and off the four high current loads such as the stir wire, the head motor, the thermoelectric cooler, and the fan.
The head motor is controlled by two relays located at the rear of the chassis that interface the 120 VAC motor to the DC logic. One relay raises the head, while the other relay lowers the head. The relays get their com­mands from the application board via two FETs and the application logic. LEDs are provided on the application board for moni­toring the drive signals. LEDs are also pro­vided to monitor the signals from the head sensor board, described below.
The thermoelectric cooler is controlled by the FET driver transistor mounted on the rear panel (PCB520). The FET is, in turn, con­trolled by the microprocessor through the PLD application logic. The processor varies the duty cycle square wave in response to software commands and block probe resist­ance.
Drive Circuitry, Serial Suffix D and Higher: Drive circuitry is provided to turn
on and off four system loads such as the stir wire, the head motor, the thermoelectric cool­er, and the fan.
The stir wire coil drive circuit consists of a programmable timer, D/A converter, and interface op-amps on the application board (PCB605). The timer provides a square wave of approximately 71 Hz, while the D/A con­verter controls the output amplitude to the darlington transistor mounted on the driver board (PCB523).
The thermoelectric cooler is controlled by the microprocessor through the PLD application logic. The processor varies the duty cycle square wave in response to software com-
27
The Advanced®Model 3250/4250 Service Manual
mands and block probe resistance. These sig-
als are provided to the FET driver transistor
n mounted on the driver board (PCB523).
The head motor is controlled by the application board via two FETs and the PLD logic. LEDs are provided on the application board for moni­toring the drive signals. LEDs are also provid­ed to monitor the signals from the head sensor board, described below. Two active low head up/head down signals are sent to the driver board (PCB523) where they are converted into acceleration, direction, and step commands for the motor drive chip. The motor drive chip translates these commands into current con­trolled outputs that move the DC stepper motor and provide idle torque when no movement is required. The following adjustments may be available for the motor circuit: R10 and TP16 are used to set the initial acceleration to a rise time of 4uS. R21 and TP15 are used to set the run frequency to 8 Khz.
The fan is controlled by a FET transistor on the application board, that is interfaced to the processor through the application logic.
Head Transition Board (M20450) Descrip­tion: The stir/freeze coil and the sample probe
are connected to the head transition board which allows easy sample probe replacement.
Head Sensor Board (3D3380) Description:
This board provides two optical sensors that interface to a PLD on the main board. These two sensors detect the head-up and head-down conditions.
Keypad Transition Board (FLO520) Description: This board provides the interface
to the keypad along with pull-up resistors for the keypad LEDs.
Barcode Port: A D-type, 15-pin barcode port is provided in the back of the instrument for connecting and providing power to such a device. For proper operation, the barcode port requires a 1200 bps, RS-232 signal providing asynchronous serial data containing 1 start bit, 8 data bits, 1 stop bit, and no parity.
Signal Pin Direction
+5V DC 1 to reader receive data 10 from reader gnd/earth 9 common
Barcode Port Connections
A suitable barcode scanner is available from Advanced Instruments. To interface with the instrument, the barcode scanner must be pro­grammed as follows, referring to the scanner users guide.
1200 bps CR suffix disable beep after good decode triggerless trigger mode (optional)
Internal Printer, Serial Suffix A - C: The internal printer assembly (325400) consists of the sheet metal mounting bracket, thermal print head (FL0403), thermal printer PCB (FL0402), and thermal printer PC board assembly (PCB400). All of this is connected to the appli­cation board (PCB605) via cable (M20465).
Note: The internal printer may be powered by
+5 VDC (red/orange wire) or +12 VDC (violet wire). Consult system schematics for proper service.
Internal Printer, Serial D and Higher: The internal printer assembly (325420) consists of the sheet metal mounting bracket, thermal print head (325404), main control board (325421), printer interface board (PCB402), and the con­trol cable (325419). All of this is connected to the application board (PCB605) via cable (325415), and on some models via cable (325033).
The printer interface board (PCB402) contains the power circuit converting +12 VDC into +5 VDC, a feed/self-test switch (S1), and some diagnostic LEDs. If the self-test switch is depressed as instrument power is applied, the printer will enter self-test mode and print out settings and a test character set. The LEDs indicate the operation of the paper-out detector (D1 yellow) and the door/platen-closed detector (D2 red). POWER CONNECTION TO J1
The Advanced®Model 3250/4250 Service Manual
28
SHOULD NOT BE RECONNECTED WITH INSTRUMENT POWER ON, OR DAMAGE TO THE INSTRUMENT MAY RESULT.
The printer control board (325421) translates the data from the application board (PDCB605) into control signals for the print head thermal elements and drive motor (325404). This board also contains a factory-set configuration DIP switch (DS1) and jumper (JP1).
On models equipped with cable (325033), a compatible version of main board set (325605)
board (PCB402), diagnostic LEDs, and a test switch are available in the lower right corner of the larger board (PCB605). D19 is a red LED that will indicate the printer door/platen interface is not closed or, if blinking, one of several printer errors have occurred, requiring factory assistance. D20 is a yellow LED that will indicate the paper is out or not detected when the door is closed. SW1 is a momentary switch that will cause the paper to advance when pressed or, if held down while main power is switched on, will place the printer into self-test mode.
and a properly configured printer interface
Design Changes
The serial number suffix (referred to in this manual simply as the suffix) indicates the revision of the instrument. The chart below lists the major changes made at each revision of the 3250 Osmometer and 4250 Cryoscope.
Model 3250
Osmometer Description
Serial Suffix ‘A’ Original model released. January 2004
Serial Suffix ‘B’ Base chassis and cover re­May 2007 designed.
Serial Suffix ‘C’ Up/Down relay PC boards re­October 2007 designed into a single RoHS-
compliant PC board.
Serial Suffix ‘D’ **MAJOR REVISION** September 2009 AC head motor and supporting
components replaced with a DC stepper motor, multi-driver PC board and power supply, Seiko thermal printer assembly with thread-through paper loading and supporting components replaced with a Citizen thermal printer assembly with easy-load paper loading.
Serial Suffix ‘E’ Application board redesigned to October 2010 power display and keypad LEDs
from VCC +5 VDC line. Main harness redesigned.
Model 4250 Cryoscope Description
Serial Suffix ‘A’ Original model released. November 2005
Serial Suffix ‘B’ Base chassis and cover re­May 2007 designed.
Serial Suffix ‘C’ Up/Down relay PC boards re­October 2007 designed into a single RoHS-
compliant PC board.
Serial Suffix ‘D’ **MAJOR REVISION** September 2009 AC head motor and supporting
components replaced with a DC stepper motor, multi-driver PC board and power supply, Seiko thermal printer assembly with thread-through paper loading and supporting components replaced with a Citizen thermal printer assembly with easy-load paper loading.
Serial Suffix ‘E’ Application board redesigned to October 2010 power display and keypad LEDs
from VCC +5 VDC line. Main harness redesigned.
29
The Advanced®Model 3250/4250 Service Manual
Replacement Parts
Serial Suffix D and higher has some different replacement parts from Serial Suffix A - C, although many
parts have remained common. Before selecting a replacement part, check the serial suffix of the instrument
on the serial suffix label on the rear of the instrument, then refer to the proper section. Refer to the Design
Changes section for additional information.
3250
Common Parts
Block Probe . . . . . . . . . . . . . . . . . . . .4D3340
Clapper . . . . . . . . . . . . . . . . . . . . . .3C2241R
Control Board Set . . . . . . . . . . . . . .325605R
Application Board (only) . . . . . . . .PCB605R
Processor Board (only) . . . . . . . . . .325621R
Cooling Assembly . . . . . . . . . . . . . .3D3300R
Cooling Fan . . . . . . . . . . . . . . . . . . . .4D3360
Cooling Well Inlet Check Valve . . . . .99133R
Firmware “FLASH” Update Kit . . . .SFW008
Fluid Filter . . . . . . . . . . . . . . . . . . . .4D3710R
Fuses: 2.0 Amp . . . . . . . . . . . . . . . . . .70022
1.0 Amp . . . . . . . . . . . . . . . . . .70011
Heat Transfer Fluid Intake
Check Valve . . . . . . . . . . . . . . . .4D3705R
Heat Transfer Fluid Pump Assembly .4D3690
Keypad . . . . . . . . . . . . . . . . . . . . . . .325511R
Load Resistor . . . . . . . . . . . . . . . . . .425150R
Mandrel . . . . . . . . . . . . . . . . . . . . . . .3LH500
NVRAM Battery . . . . . . . . . . . . . . . .71027R
Probe Alignment Tool Kit . . . . . . . . .3LA700
Sample Probe . . . . . . . . . . . . . . . . . . .3D3700
Stir/Freeze Coil . . . . . . . . . . . . . . . .3D2404R
Stir/Freeze Wire . . . . . . . . . . . . . . . . .3LH243
Tube Ejector Spring . . . . . . . . . . . .3D3312R
Yoke . . . . . . . . . . . . . . . . . . . . . . . .3LH230R
4250
Common Parts
Block Probe . . . . . . . . . . . . . . . . . . . .4D3340
Clapper . . . . . . . . . . . . . . . . . . . . . .3C2241R
Control Board Set . . . . . . . . . . . . . .425605R
Application Board (only) . . . . . . . .PCB605R
Processor Board (only) . . . . . . . . . .425621R
Cooling Assembly . . . . . . . . . . . . . .4D3300R
Cooling Fan . . . . . . . . . . . . . . . . . . . .4D3360
Cooling Well Inlet Check Valve . . . . .99133R
Firmware “FLASH” Update Kit . . . .SFW008
Fluid Filter . . . . . . . . . . . . . . . . . . . .4D3710R
Fuses: 2.0 Amp . . . . . . . . . . . . . . . . . .70022
1.0 Amp . . . . . . . . . . . . . . . . . .70011
Heat Transfer Fluid Intake
Check Valve . . . . . . . . . . . . . . . .4D3705R
Heat Transfer Fluid Pump Assembly .4D3690
Keypad . . . . . . . . . . . . . . . . . . . . . . .325511R
Load Resistor . . . . . . . . . . . . . . . . . .425150R
Mandrel . . . . . . . . . . . . . . . . . . . . . . .3LH500
NVRAM Battery . . . . . . . . . . . . . . . .71027R
Probe Alignment Tool Kit . . . . . . . . .3LA700
Sample Probe . . . . . . . . . . . . . . . . . .4D3102
Stir/Freeze Coil . . . . . . . . . . . . . . . .3D2404R
Stir/Freeze Wire . . . . . . . . . . . . . . . . .4LH243
Tube Ejector Spring . . . . . . . . . . . . . .4D3312
Yoke . . . . . . . . . . . . . . . . . . . . . . . .3LH230R
Serial Suffix A - C Parts
Display Board . . . . . . . . . . . . . . . . . .74051R
Head Up/Head down Relay . . . . . . . .3D3109
Motor Assembly . . . . . . . . . . . . . . .4D3350R
Power Supply Assembly . . . . . . . . . .4D3950
Printer Assembly without
Power Harness . . . . . . . . . . . . . .FL2401R
Printer Assembly with Power
Harness . . . . . . . . . . . . . . . . . . . .325400R
Stir Drive Transistor . . . . . . . . . . . .PCB522R
Thermoelectric Drive Transistor . .PCB520R
The Advanced®Model 3250/4250 Service Manual
Serial Suffix A - C Parts
Display Board . . . . . . . . . . . . . . . . . .74051R
Head Up/Head Down Relay . . . . . . .3D3109
Motor Assembly . . . . . . . . . . . . . . .4D3350R
Power Supply Assembly . . . . . . . . . .4D3950
Printer Assembly without
Power Harness . . . . . . . . . . . . . .FL2401R
Printer Assembly with Power
Harness . . . . . . . . . . . . . . . . . . . .325400R
Stir Drive Transistor . . . . . . . . . . . .PCB522R
Thermoelectric Drive Transistor . .PCB520R
30
Loading...
+ 165 hidden pages