ADP O2SLB, O2SLG, O2SLU, O2SOB, O2SOG Schematic Diagrams

...
Air Intake: 1" PVC Slip, Vent To Acceptable Air Source
Vacuum Vent: 2" PVC Slip To Outside With Flex Tubing
Airline: 1/2" Copper, 1/2" FPT
56"
18"
10"
Specification Sheet
Low Voltage: 18 GA-3 Wire (1 Set For Each L/V Switch)
18"
6"
36" Max. (Either Side)
Electrical: Disconnect Box (x2) (20 Amps Each)
Waterline: 1/2" FPT Valve
Vacuum: PVC Schedule 40 (See Piping Guidelines)
The Organizer
6" Max.
Drain: 1 1/2" P-Trap Reduced To 3/4" FPT (Refer Local Codes) Or Alternate Floor Sink
12"
8"
Ventilation Required Room Temperature Must Be In The Range Of 40
o
To 80
o
Custom Configurations
Are Available.
Contact Technical Support
(800) 233-4151
2' Min.
6" Min.
6" Min. 36" Max.
Exhaust Line 3/4" x 3' Flex PVC "Use PVC Glue" (Provided)
Width Depth Height Shipping Weight
24.25" 29" 56" 353 Lbs. 303 Lbs.
Lubricated Oil-Less
Model Max. Users Comp. HP Vac. HP SCFM@80PSIG db-A Volts Hertz Comp. Amps Vac. Amps Tank Size
O2SLB 2 1 1 4.7 66 208 60 7.2 8.1 10 Gal. O2SLG 2 1 1 4.7 66 208 60 7.2 8.1 10 Gal. O2SLU 2 1 1 4.7 66 208 60 7.2 8.1 10 Gal. O2SOB 2 1 1 3.7 74 208 60 5.2 8.1 10 Gal. O2SOG 2 1 1 3.7 74 208 60 5.2 8.1 10 Gal. O2SOU 2 1 1 3.7 74 208 60 5.2 8.1 10 Gal.
Note: All locations and dimensions are approximate. Systems are to be installed according to local building plumbing and electrical codes. Note: Operatory vacuum line and sewer line to terminate within three (3) feet (max) of vacuum pump(s).
Apollo Dental Products, Inc.
245 W. Dakota Ave. • Clovis, CA 93612 • (559) 233-4151 • (800) 233-4151 • Fax (559) 292-1555 • E-Mail: support@apollodental.com • www.apollodental.com
AMM40500 4/01