Adesso 2000 User Manual

Serie 2000 - Bedienungsanleitung Series 2000 - Operating Manual
Digital-Handmanometer
Serie 2000
Bedienungs-Anleitung
Portable Digital Manometer
Serie 2000
2000
Operating Manual
www.Kalibriersysteme.de
www.Kalibrier-Forum.de
www.LR-Cal.com
DRUCK & TEMPERATUR Leitenberger GmbH
Bahnhofstr. 33 C D-72138 Kirchentellinsfurt C Germany Tel.: 0 71 21 - 9 09 20 - 0 C Fax: 0 71 21 - 9 09 20 - 99 E-Mail: DT-Info@Leitenberger.de C http://www.druck-temperatur.de
Some of the options described may not apply to this instrument. Please check the functions of the instrument you have purchased before proceeding.
Particular care should be taken not to over pressure the device as this may rupture the sensor membrane. This is not covered by the manufacturers warranty.
NOTE: The differential/gauge instruments will only measure positive pressures i.e. positive pressure applied to the positive port, or negative pressure applied to the negative port.
INSTRUMENT APPLICATION
The range of 2000P instruments are suitable for gauge, differential, or absolute measurements over a wide range of pressures. They are not suitable for use with corrosive substances or cyclic hydrocarbons, e.g. motor oil, transmission fluid and freon. To use the 2000P with these components, isolation must be provided in the form of a buffer, such as a mineral oil or dry air.
GB
INSTRUMENT OPERATION
BATTERIES
ON - OFF
ON OFF
Two AAor equivalent cells (not supplied). Follow instructions on reverse of instrument for fitting/replacement. When symbol appears on display, replace batteries.
Press key for on and key for off.
If key is held when unit is switched on, automatic switch-off function will be disabled until the unit is switched off.
IP65/IP67 RATING
LOG
MAX/MIN
The waterproof rating for this product will not be maintained unless the screws holding the battery compartment are firmly tightened when inserting or replacing batteries.
Automatic switch-off after 12 minutes, unless any key is activated or the instrument is in or mode.
Models
Models
Operating Range
Overrange
Features
FEATURES
RANGE
2000P,2001P, 2002P, 2003P, 2005P
2020P,2021P, 2022P, 2023P, 2024P, 2025P,2026P
2080P,2081P, 2082P, 2083P, 2084P, 2085P,2086P
25mbar / 10HO
2
Backlight, Out of Range, Zeroing, Zeroing Absolute & Hold
350mbar / 140 H O2”
130mbar / 52 H O
2
As above plus: Smoothing, Units of Pressure, Rangelock & Max/Min
1bar / 401 H O2”
2bar / 28 PSI
As above plus: Manuallog and Autolog
4bar / 58 PSI
7bar / 101.5 PSI
10bar / 145 PSI
External Transducer
­2bar absolute 10bar / 145 PSI
10.342bar / 150 PSI
4bar
2000P,2020P, 2080P 2001P,2021P, 2081P 2002P,2022P, 2082P 2003P,2023P, 2083P 2024P,2084P 2005P,2025P, 2085P 2026P,2086P
BACKLIGHT
ONPress and hold key.
DISPLAY
OUT OF RANGE
OutIf pressure exceeds the scale range will appear on the display.
ZEROING
Models 2000P, 2001P, 2002P, 2003P, 2020P, 2021P, 2022P, 2023P
0.0
Models 2080P, 2081P, 2082P, 2083P, 2084P
0.0
In order to achieve maximum accuracy it is recommended to zero the instrument in the orientation it will be used before taking any measurements.
Ensure both ports are open to atmosphere. If indicated pressure is within 1%f.s (3% for Models 2000P, 2020Pand 2080P) of the factory calibrated zero, press key which will adjust the reading to zero.
As above but press key instead of the key.
ZEROING ABSOLUTE
Models 2005P, 2025P
0.0
Model 2085P
0.0
This feature enables correction of the indicated ambient pressure between 900mbar and 1100mbar.
Open port to atmosphere and the instrument will indicate ambient pressure. Press key and 900mbar (or the equivalent in other engineering units) will appear on the display. Press and hold this key. The display will scroll and when required atmospheric pressure is displayed (maximum 1100mbar) release key. The current pressure being measured will then be adjusted to the displayed reading and used for the rest of the session. Note: This setting will be lost when the instrument is switched off.
Press key instead of key.
DISPLAY HOLD
Pressing key freezes the display. The symbol will appear on the display. Press key again to return to normal display.
HOSE FITTING
The input port(s) on the instrument are designed to accept a hose with 4mm I.D. and 1mm wall. The hose is fitted by unscrewing the clamping nut, sliding the hose on to the stepped shaft and refitting the nut.
BATTERY
YES
TIME LOG
HH:MM
MULTIFUNCTION MULTIFUNCTION
MULTIFUNCTION
NUMERIC
ENTRY
RANGE
AVERAGE
PSI
ENTER
CAL
NE
Cal
P
1
2
2
3
mbar PSI
HO
Hg
Pa
in
in
cm
M
k
mm
OK?
A/R
Note: Do not tighten fittings attached to the pressure ports to a torque greater than 2Nm otherwise the pressure seals may be damaged.
HAND PUMP OPERATION (if applicable)
T+Use the piece to connect the hand pump to the instrument’s
port and to the equipment being tested. Exhaust the system by turning the release valve on the hand pump anti-clockwise and set the fine adjustment threaded piston to mid range. Close the release valve and apply pump action to the hand pump piston to generate the required approximate pressure. Fine adjustment of the desired pressure can then be made by rotating the threaded piston. The release valve may be used to vent the system. Note: The hand pump is capable of generating 7bar max. It is, therefore, essential that extra care is taken not to over-pressure the instrument.
A pump must not be used on models 2000P, 2020P and 2080P
PLUS FUNCTIONS Models 2020P, 2021P, 2022P, 2023P, 2024P, 2025P, 2080P, 2081P, 2082P, 2084P, 2085P
SMOOTHING MODE
Press key and the instrument continuously averages the last four measurements to give a more stable reading on the display. Press key again to return to normal measurement.
=
=
UNITS OF PRESSURE
Press key to select required units of pressure. Chosen units will be saved when the instrument is switched off. Please refer to the table on the back page. The 20x0P 0-25mbar instruments cannot show the µbar or mmHg symbols on their displays. When switching the units on they are reading in µbar and so show no unit symbol. Pushing the autorange button will put the unit into mbar and show the mbar symbol. When reading in mmHg only the symbol Hg will show on the display. This function will not operate while in store mode. Note: 2024P - 2084P K=x10 when using in H O and Pa. M=x10 when using in H O and Pa.
3
2
6
2
MAX/MIN RECORDER
Press key to start. The instrument will display alternating
NOTE: Some functions may be inhibited while information is stored in either Auto, Manual or Max/Min modes. This is to prevent inadvertent parameter mixing whilst in any of the above modes. To re-activate these functions please delete any stored information by following the “Output of data” section.
symbols and current reading. Press again to display the maximum temperature reading and symbol will be displayed. Press again to display the minimum temperature reading and symbol will be displayed. Press again to return to the actual temperature display. Press key to display the average temperature since the MAX/MIN mode was activated and symbol will appear on display. Press
key again to return to actual temperature display.
NOTE: To reset MAX/MIN recorder function switch the instrument off.
X
X
ADVANCED FUNCTIONS The following additional functions apply to models 2080P, 2081P, 2082P, 2083P, 2084P, 2085P
RANGE LOCK
This instrument has a Range Lock feature that displays readings to the best resolution, shown by / on the display. This can be turned off by pressing the key. The display will then use the same display resolution for the entire pressure range. This is useful to stop the display switching between resolutions where there are fluctuations in pressure. To cancel this mode, press the key again.
AR
//AR
AR
SETTING INTERNAL CLOCK
NEPress d (•) key and the and symbols will appear
on display with the current year and month settings Key in new year and month if desired and press key to accept. Display will show current date of month setting (DD). Key in new value if desired and press key to accept. Display will show current hours and minutes setting (24 hour clock). Key in new values if desired and press key to accept. Instrument will then return to normal mode.
ecimal point
(YYMM).
PRESET INTERVAL LOGGING
This function allows readings to be taken automatically at a preset time interval. NOTE: If data is already stored, as signified by the or symbols appearing on display, this function will not operate. Press key. symbol and will appear on the display. Enter required time interval in hh:mm (ie, in order to take a reading every 1 minute, key in 0001) and press key. Current reading will return to display and symbol will remain on display.
NE 0000
Instrument will continue logging until a maximum of 250 readings are taken or data is down-loaded.
LOGGING
These models have functions that enable the user to store and retrieve up to 250 readings and output such to a PC or Epson compatible printer via Infra-red DigiLink. We also offer DigiLog, a customised Windows™ software package. Note: Some functions may be inhibited while information is stored in either Auto, Manual or Max/Min modes. This is to prevent inadvertent parameter mixing whilst in any of the above modes. To re-activate these functions please delete any stores information by following the "Output of data" section.
LOGGING ON DEMAND
NE
0
This function allows readings to be stored as and when required. A maximum of 50 readings can be stored this way. When ready to store a reading press key. , symbols and will appear on the display. It is possible to enter up to a four digit reference number including decimal point. Press key to enter and store reading, time, date and reference number. Instrument will return to standard mode, however symbol will remain on display until data is erased.
LEAK TESTING
This function enables testing for leaks in pressure systems. The instrument has a function to automate leak testing in a variety of applications. Automatic leak test results are logged, not stored and cannot be output directly via DigiLink into your DigiLog PC software. To send leak test results to your PC please use your PC’s Hyper Terminal and output as to an Epson Printer. When entering pressure values (P1, P2, P3) you must enter numbers to exactly the same number of decimal places as full pressure range of the instrument being used. The number of decimal places required can be seen in the table of pressure ranges on the back page.
AVERAGING
MAX/MIN
MAX/MIN
Press key (when in mode) and symbol, together with average reading, will appear on display. Press
to return to mode.
SETTING UP Press key (when not in mode) and and symbols will appear on the display. Enter pre-stress pressure level (if pre-stress pressure function is not required, press key) and the display will show and symbols. Key in start pressure and press key and the and symbols will now be displayed. Key in test time in minutes and seconds (MM.SS), press key and the and symbols will now be displayed. Key in permissable change in pressure and press key. The instrument is now ready to run test and will appear on display.
MAX/MIN
NE
NE P1
P2
NE
NE P3
P1
PERFORMING A TEST This test only functions if the symbol is displayed. Apply pressure to system until changes to (Test Pressure). If a (pre­stress) pressure has been entered the system must be pressurised to this level. Bleed system. The display will then clear until test time has elapsed. or will appear on the display and be
P1
P1 P2 P1
PASS FAIL
X
X
X
OUTPUT OF DATA
1
2
3
This feature will only function if either the or symbols are displayed. To view data press key, then and keys. The first stored reading will be displayed. Press key to view successively stored readings and any other key to return to normal mode. To output data to a PC press key, then key.
Ensure PC/Printer is ready to
receive information.
After transmission, the unit will return to normal operation and any auto logging will stop. To erase stored data press key, then and keys. There will be a delay before the unit will accept the key and symbol will be displayed. This is to help stop accidental erasure. After erasure, neither or symbols will be displayed. The unit will return to normal mode.
The selected transmit option will be displayed during transmission.
If after pressing key you do not wish to delete data, you must turn instrument off and on, then the stored data will not be lost.
To output data to an Epson compatible printer press key, followed by key. Printer settings must be as follows: 9600 Baud, 8 data bits, No Parity, One stop bit, No flow control. It is also possible to print to Hyper Terminal in your PC in this way.
4
3
automatically logged with test conditions. This logged reading can be replayed on screen by pressing and keys or output to an Epson compatible printer by pressing and keys. It cannot be output to a PC other than via Hyper Terminal. Consecutive records will overwrite each other so must be output to printer before performing another test. After showing the the display will immediately revert to current pressure. The instrument must be reset manually for each new test.
1
4
PASS/FAIL
INSTRUMENT ACCURACIES
From +20°C to +30°C/+68°F to +86°F From -10°C to +50°C/+14°F to +122°F 0.15%rdg +0.15%fs +1 digit
0.1%rdg +0.1%fs +1 digit
USER DEFINED SCALE
Press key and will appear before the display clears. Enter scale percentage. Only the first two digits entered are valid ­this gives a scaling from 1 - 99%. This value will be multiplied into current display to give user defined scale. Press to accept. will stay displayed and current reading will appear. A downward arrow will indicate that this feature is active. WARNING - whilst in this mode over range indication (out) is disabled. Do not over pressure the instrument.
NE
NE
WARRANTY
This instrument has been carefully assembled and tested, and is warranted against faulty workmanship and materials for two years from the date of purchase. During the warranty period any defective instrument will be repaired or replaced at the discretion of the manufacturer. Thiswarranty does not cover damage or failure resulting from misuse or accident. Modification, adjustment or any alteration with the internal arrangement of the instrument shall absolve the manufacturer from any liability in respect of the instrument. Any instrument to be repaired should be forwarded to the supplier, carriage paid and at the owner’s risk. Abrief description of the fault should be included.
The overall performanceof the instrument is obtainedby combining the stated accuracy and anyuncertainty dueto the measurementprocess.
PRESSURE TRANSDUCER SELECTION Models 2024P, 2084P
0.0
0.0
At switch-on press and hold the key (2024P - “ ” key) and the display will scroll through the available transducer range option,
10.0 to 50.0 Bar (in 5 Bar steps) or 55.0 to 500.0 Bar (in 5 Bar steps) (release to select) --- appears on the display. To customise press and hold the key (2024P - “ ” key) again to select the required input voltage range, 10.0 to 50.0 Bar with 1 to 50mV (in 1mV steps) or 55.0 to 500.0 Bar with 10 to 50mV (in 1mV steps) (release to select). These are now the default settings.
EXTERNAL PRESSURE TRANSDUCER CONNECTION Pressure range selectable (in Bar) with Sensitivity range selectable (in mV). Power supplied
Wiring instructions
WARNING: Care should be taken to ensure the wiring instructions are followed as incorrect wiring of the cable to the transducer may damage the instrument.
+5 Volts (± 0.25 Volts equivalent to 5%)
Red Wire = + supply, Blue Wire = - supply, Green Wire = + signal, Yellow Wire = - signal
Einige der beschriebenen Optionen können mit diesem Instrument möglicherweise nicht genutzt werden. Vor der Benutzung ist zu prüfen, mit welchen Optionen das gekaufte Gerät ausgestattet ist.
Achten Sie darauf, das Gerät keinem unzulässigen Überdruck auszusetzen, da sonst die Sensor-Membrane zerstört wird. Dies fällt nicht unter Garantie Ansprüche.
Bitte beachten: die Geräte messen nur den positiven Druck; schließen Sie deshalb den positiven Druck nur am positiven Eingang und den negativen Druck (Vakuum) am negativen Eingang an.
EIN/AUSSCHALTEN
ON OFFDie Taste schaltet das Gerät ein und die Taste schaltet das
Gerät aus.
Sollten Sie die Taste beim Einschalten gedrückt
halten, wird die automatische Abschaltung ausgeschaltet.
Automatische Abschaltung nach 12 Minuten falls keine Taste aktiviert wird oder das Gerät in Registrier - oder Modus ist.
MAX/MIN
BATTERIEN
Es werden 2 Batterien AA (1,5 V) benötigt (nicht im Lieferumfang) Einsetzen/Tauschen siehe Symbole auf Geräterückseite.
Wenn im Display erscheint, sollten Sie die Batterie wechseln
IP65/IP67 SCHUTZKLASSE GEGEN STAUB UND WASSER
Die Schutzklassen IP65/67 bleiben nur dann erhalten, wenn nach dem Einbau von Batterien bzw. dem Batteriewechsel, der Deckel mit Dichtung ordnungsgemäß aufgesetzt und verschraubt wird.
D
GERÄTE- ANWENDUNGEN
Es stehen verschiedene Messbereiche für Druck-, Differendruck­und Absolutdruck-Messungen zur Aus
nicht geeignet für die Verwendung mit aggressiven Substanzen oder Hydrocarbonen z.B. Motoröl oder Freon. Um die Geräte mit diesen Stoffen zu verwenden, muss eine Trennvorlage eingesetzt werden.
wahl. Sie sind
Typen
Merkmale
BESONDERE MERKMALE
Hintergrundbeleuchtung, Messbereich Überschreitung Angleich-Modus, Absolut und Data Hold
Wie oben angegeben + Messwertglättung Einheiten, Messbereichsperre, Max/Min-Messungen
Wie oben angegeben + Data Logging u. Lecktest
Typen
Messbereich
max. zul. Überdruck
MESSBEREICH
25mbar / 10 H O
2
350mbar / 140 H O
2
130mbar / 52 H O
2
1bar / 401 H O
2
2bar / 28 PSI
4bar / 58 PSI
7bar / 101.5 PSI
10bar / 145 PSI
Externer Messumformer
-
2bar absolut
4bar
10bar / 145 PSI
10.342bar / 150 PSI
2000P,2001P, 2002P, 2003P, 2005P
2020P,2021P, 2022P, 2023P, 2024P, 2025P,2026P
2080P,2081P, 2082P, 2083P, 2084P, 2085P,2086P
2000P,2020P, 2080P 2001P,2021P, 2081P 2002P,2022P, 2082P 2003P,2023P, 2083P 2024P,2084P 2005P,2025P, 2085P 2026P,2086P
BEREICHSÜBERSCHREITUNG
Wenn der Druck über den Messbereichsendwert hinaus oder unter­halb 0 geht, wird am Display angezeigt.
NULLPUNKTKORREKTUR
Um ein Maximum an Genauigkeit zu erhalten, sollten Sie das Gerät in der Umgebung auf Null stellen in welcher Sie es auch verwenden.
MODELLE 2000P, 2001P, 2002P, 2003P, 2020P, 2021P, 2022P, 2023P
MODELLE 2080P, 2081P, 2082P, 2083P, 2084P
0.0
Beide Druckanschlüsse sind beim Nullpunktabgleich zu öffnen. Wenn der Druck im Bereich 1% v.E. des im Werk kalibrierten Bereichs ist (3% für Modelle 2000P, 2020Pund 2080P), wird durch Betätigung der 0-0 Taste eine Nullpunktkorrektur des momentanen Wertes durchgeführt.
Die Taste anstatt drücken.
0-KALIBRIERUNG BEI ABSOLUTDRUCK- VERSIONEN
MODELLE 2005P, 2025P
0.0
MODELLE 2085P
0.0
Der angezeigte atmosphärische Druck kann zwischen 900 und 1100mbar gemäß dieser Funktion geändert werden.
Druckeingang öffnen, atmosphärischer Druck wird angezeigt. Die
Taste drücken. 900mbar (oder einer equivalenten Einheit) erscheint im Display. Wenn der gewünschte Druck angezeigt wird (max. 1100mbar) Taste loslassen. Der augenblickliche Wert wird dann für den Rest der Messung zugrundegelegt. Durch das manuelle Abschalten wird der Wert gelöscht.
Die Taste anstatt drücken.
DATA HOLD
Um einen Messwert ‘einzufrieren’ drücken Sie die Taste und wird im Display angezeigt. Der Messwert bleibt auf der Anzeige bis diese Taste nochmals gedrückt wird.
BATTERIE JA
ZEITLOG
MEHRFUNKTION MEHRFUNKTION
MEHRFUNKTION
NUMERISCHE
EINGABE
MEßBEREICH
MITTELWERT
PSI
EINGABE
CAL
NE
Cal
P
1
2
2
3
mbar PSI
HO
Hg
Pa
in
in
cm
M
k
mm
OK?
A/R
ANZEIGE
HINTERGRUNDBELEUCHTUNG
ONDie Taste gedrückt halten.
ANSCHLÜSSE
Die Anschlussöffnung(en) des Gerätes sind für einen Schlauch mit 4 mm Innendurchmesser und 1 mm Wandstärke vorgesehen. Um diese Schlauchleitung zu montieren, ist die Überwurfmutter zu lösen und die Schlauchleitung über den gestuften Schaft zu schieben. Überwurfmutter wieder festschrauben. Beachten Sie: Befestigen Sie die an den Drucköffnungen befindlichen Anschlüsse nicht mit einem größeren Drehmoment als 2Nm, da die Druckdichtung ansonsten beschädigt werden könnte.
HANDPUMPE (FALLS VERWENDET)
Die Handpumpe darf nicht mit den Gerätversionen 2000P,2020P und 2080P verwendet werden.
Verwenden Sie das T-Stück, um die Handpumpe mit dem Eingang des Gerätes und die Test-Ausrüstung zu verbindern. Entleeren Sie das System mit der Handpumpe beim Drehen des Auslassventils nach links. Die Einstellung kann mit dem Feinregler an der Pumpe erfolgen; diese hat ein Feingewinde. Auslassventil schliessen und mit der Handpumpe zu pumpen, um den gewünschten Druck aufzubauen Die Feinreglung des gewünschten Druckes erfolgt durch Drehen des mit einem Gewinde versehenen Knaufs. Das Auslassventil kann zur Entlüftung des Systems benutzt werden. Achtung: das Gerät darf nicht mit zu hohem Druck beaufschlagt werden; die Handpumpe kann bis zu 7bar Druck erzeugen.
+
BEDIENUNG DER ERWEITERTEN MODELLE Die Modelle 2020P, 2021P, 2022P, 2023P, 2024P, 2025P, 2080P, 2081P, 2082P, 2084P und 2085P haben einige zusätzliche Funktionen.
MITTELWERT-BILDUNG
=Mit der Taste können Sie das Gerät in einen Angleich-Modus
schalten. In diesem Modus wird der Durchschnittswert der letzten drei Messungen angezeigt um eine unruhige Anzeige zu vermeiden.
VERSCHIEDENE EINHEITEN
Mit der Taste können Sie unter den verschiedenen Einheiten umschalten. Diese Einstellung bleibt beim Abschalten gespeichert und erscheint beim neuen Einschalten im Display. Siehe Tabelle auf der Rückseite. Der Typ 20x0P (Messbereich 0-25mbar) kann die Einheiten bar und mmHg nicht im Display darstellen. Beim Einschalten eines Gerätes, das in der bar Anzeige steht, wird keine Einheit angezeigt. Durch Drücken der Autorange-Taste wird die Messung auf mbar umgestellt. Die Einheit mbar erscheint im Display. Bei Mesungen in mmHg wird lediglich Hg im Display angezeigt. Diese Funktion ist im Speichermodus nicht in Betrieb. Beachten Sie: Typen 2024P-2084P: K = 10³ bei Benutzung in H O and Pa M = 10 bei Benutzung in H O and Pa
” ”
2
2
6
MAX/MIN MESSUNG
Dieser Modus wird durch Drücken der Taste aktivert. Das Gerät misst die aktuellen Werte und speichert gleichzeitig die Minimum und Maximalwerte. Um diese anzuzeigen drücken Sie die Taste nochmals und Sie sehen am Display den Maximalwert, steht auf dem Display. Durch erneutes Drücken erscheint der Minimumwert,
steht auf dem Display. Ein weiteres Betätigen wechselt in den
aktuellen Messmodus.
ZUSÄTZLICHE FUNKTIONEN Modelle 2080P, 2081P, 2082P, 2083P, 2084P und 2085P
AUTOMATISCHE BEREICHSEINSTELLUNG
A/R
Bei Betätigen dieser Taste wird die optimaleAuflösung für den angezeigten Wert automatisch eingestellt. erscheint im Display. Durch erneutes Drücken dieser Taste schalten Sie diesen Modus wieder ab. Diese Auflösung wird für den ganzen Meßbereich benutzt, sogar wenn es Druckschwankungen geben sollte.
ZEIT EINSTELLEN
NE• Taste für 5 Sekunden gedrückt halten. und sowie das Jahr
und der Monat sind im Display angezeigt. Das Jahr und der Monat können auf Wunsch geändert werden.
Taste drücken. Die Zeit wird angezeigt und kann auf Wunsch geändert werden. Taste drücken. Das Gerät kehrt zum normalen Modus zurück.
LOGGING AUF BEDARF
NE 0
Mit dieser Funktion können Daten erfasst und gespeichert werden. Ein Maximum von 50 Messwerten kann gespeichert werden. Um die Daten zu speichern, Taste betätigen. , und werden angezeigt. Bis zu vier Ziffern können mit dem eingegeben werden. Drücken Sie nochmals auf die Taste um die Daten, die Zeit, das Datum und die Referenz zu speichern. Das Gerät kehrt in den normalen Meßmodus zurück. wird angezeigt bis die gespeicherten Daten gelöscht sind.
Dezimalpunkt
VORWAHL-LOGGING
NE 0000.
Diese Funktion erlaubt Messungen über vorgegebene Zeitintervalle. Sie können diese Funktion nicht benutzen, wenn Daten gespeichert sind. oder erscheint im Display. Drücken Sie auf die Taste . und Zeitintervalle in ST:MM
AUFZEICHNUNGS-MODUS
Mit diesen Geräten können Sie bis 250 Daten speichern und auf PC oder Epson Drucker mittels Infrarot Digi-Link abrufen. Wir bieten auch passende Software, die unter WINDOWS lauffähig ist. Beachten Sie: Es kann sein, dass einige Funktionen während des Abspeicherns von Informationen im Auto-, Manuellen oder MAX/MIN
-Modus zeitweilig inhibiert sind. Auf diese Weise soll eine unbeabsichtigte Vermischung der Parameter in einer der obigen Betriebsarten verhindert werden. Zur Reaktivierung dieser Funktionen: Löschen Sie bitte alle gespeicherten Informationen wie im Abschnitt "Datenausgabe" beschrieben.
Beim Drücken der Taste wird der Durchschnittswert angezeigt
erscheint am Display. Drücken Sie auf die Taste um in den aktuellen Messmodus zurückzukehren. ACHTUNG: Um die MAX/MIN Werte zu löschen müssen Sie das Gerät ausschalten. ACHTUNG: Es kann sein, dass einige Funktionen während des Abspeicherns von Informationen im Auto-, Manuellen oder MAX/MIN
-Modus zeitweilig inhibiert sind. Auf diese Weise soll eine unbeabsichtigte Vermischung der Parameter in einer der obigen Betriebsarten verhindert werden. Zur Reaktivierung dieser Funktionen: Löschen Sie bitte alle gespeicherten Informationen wie im Abschnitt “Datenausgabe” beschrieben.
X
X
Loading...
+ 24 hidden pages