Measure and calibrate SPRTs, RTDs, thermistors and thermocouples
1 PPM resistance ratio accuracy (channel 1)
8 1/2-digit DC multimeter
Measure up to 82 channels
Sample rates up to 10 channels per second
Bluetooth, WIFI, USB& Ethernet (RJ-45) capable
Auto temperature control of Additel and other manufacture's heat sources
Built-in automatic temperature control, data collection, and coefficient generation
Support for creating custom control of heat sources with RS-232
Auto - zero power feature (self heating compensation)
10.1 in. touch screen display
Supports fully automated temperature calibrations with data collection and report
generation (no software required)
OVERVIEW
The Additel 286 Multifunction Reference Thermometer Readout is an industry first!
We have combined the capabilities of a high-end reference thermometer with a highly
capable data acquisition system and 8.5 digit multimeter. The ADT286 is capable
of scanning and recording up to 82 channels at 10 channels per second. Users can
easily congure the ADT286 to perform eld calibrations and uniformity studies as well
as use the unit in the lab as a precision thermometer and 8.5 digit multimeter. Get
more for less with this newest game changer from Additel!
If you're in need of a precision reference thermometer for your laboratory, then look no further than the Additel 286.
The base unit comes with two precision readout channels that can be used to measure your SPRT. Need to calibrate
RTDs, PRTs, thermistors or thermocouples? Add a scanner module and you now have the ability to measure 10
RTDs, PRTs, or thermistors and 20 thermocouples. Expand up to 82 channels with our unique easy to use scanner
modules. Each 20 channel module is outfitted with our proprietary universal terminals with an industries best cold
junction capability second to none. Utilize the module docked atop the ADT286, or connect remotely with cables
to suit nearly any unique setup/configuration. Additel also has a process module specifically designed to measure
process instrumentation like transmitters and switches. This scanner will also supply loop power for the transmitters.
Temperature Calibration Equipment
Designed to make your job easier, the ADT286 has a large sensor library supporting 15 TC types, both standard
and special limits, 18 different thermal resistors, CVD, ITS-90, and a large variety of standard curves for RTDs and
thermistors. The ADT286 is loaded with special applications such as probe calibration, SPRT calibration, chamber
mapping and more. And we continue to add applications on a regular basis!
Automatic Temperature Control and Probe Calibration
The Additel 286 Multifunction Reference Thermometer Readout has preinstalled drivers to control Additel and other
manufacturer’s heat sources. Simply connect to one or more heat sources via a communication cable, Ethernet or
wireless and now it will automatically control to the set point and desired stability. If your heat source is not on the list,
you can easily add the driver yourself so you can run automated calibrations with any heat source.
Now combine the heat source control feature with our probe calibration application and you have a very powerful
automatic calibration solution. The probe calibration app allows you to automatically setup and run calibration routines
with multiple set points and multiple heat sources, collect data, and develop calibration coefcients — all with one device
and without the need of software! Simply place a batch of sensors of any mix and type into your heat source, connect
it to the ADT286, run the probe calibration app and come back to a completed test. All that is left to do is generate
and export all the calibration data. There’s no need to work with complicated software for communication, set up or
coefcient generation. There is no reason to have a calibration technician manually monitor the process and record the
data. This Multifunction Reference Thermometer Readout will do all the work for you.
ADT286-RS Resistance Standards
Available in 25 & 100 Ohm values, users can enjoy improved resistance ratio performance by easily plugging one of our
reference resisters into channel 2 on the new ADT286. Perfect for calibrating your SPRT’s and high end PRT’s.
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Page 5
SPECIFICATIONS
General Specifications
SpecificationDescription
Voltage
Frequency47Hz to 440Hz.Automatically sensed at power-on
Temperature Calibration Equipment
Power Consumption40VA peak (30Watt average)
Temperature
Warm-up90 mins for full uncertainty specications
Relative Humidity (non-
condensing)
Altitude
Vibration and ShockComplies with MIL-28800F Class 3
Input Protection50V all functions, ranges and terminals
CommunicationUSB-A , USB-B , RJ45 , WiFi , Bluetooth
Memory10G - All data stored with time stamps
LocalizationEnglish , Chinese
Display10.1 in (256 mm) TFT color display
Size (H x W x D)9.8 in (250 mm) x 16.5 in (420 mm) x 7.9 in (200 mm)
Weight18.5 lb ( 8.39 kg)
Other ConformitiesCE
100V Setting 90V to 110V
120V Setting 108V to 132V
220V Setting 198V to 242V
240V Setting 216V to 264V
Operating : 0°C to 50°C
Full accuracy :18°C to 28°C
Storage : -20°C to 70°C
Operating : 0°C to 28°C < 90%
28°C to 40°C < 75%
40°C to 50°C < 50%
Storage : -20°C to 70°C < 95%
Operating : 2000 m
Storage : 12000 m
Measurement Specifications
Specication Conditions: 90 mins Warm-Up Time / Environment Temperature (18 - 28) °C.
The following specications apply for front panel, after at least 90 minutes warm-up.
24-hour specications are relative to calibration standards and assume a controlled electromagnetic environment per EN 61326.
Resistance Ratio Accuracy (Rx/Rs) using External Rs
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Page 6
SPRT/PRT Measurement Accuracy using External Rs
SPRT/PRT Type
PT25 25 Ω
External Reference
Resistance
Temperature Calibration Equipment
PT100100 Ω
[1] The PT25 indicator is based on a nominal resistance of 25 Ω for Rx.
[2] The PT100 indicator is based on a nominal resistance of 100 Ω for Rx.
[3] The uncertainty of external Rs is not included. The user may choose the ADT280-RS-25/100 standard resistor as external Rs, which has an accuracy
of 5 ppm at (23±2) °C.
Resistance Accuracy using Internal Rs
Measurement
Range
( 0~100 ) Ω
( 0~400 ) Ω
( 0~4000 ) Ω
[1] Accuracy Index: ± (ppm of reading or xxmΩ , whichever is greater).
[2] Temperature coefcient index: exceeds (18-28) °C range, increase per degree (ppm reading +xxmΩ).
[3] Specications are for 4-wire function. For 3-wire, add 0.005 Ω for internal resistance mismatch. For 2-wire, add 0.005 Ω for internal resistance
[4] Automatic current reversal.
Scanning Speed Resolution
Slow Speed0.01 mΩ 3 ppm or 0.2 mΩ12 ppm or 0.35 mΩ15 ppm or 0.35 mΩ
Fast Speed0.1 mΩ 3.6 ppm or 1.7 mΩ 12.6 ppm or 1.85 mΩ 15.6 ppm or 1.85 mΩ
Slow Speed0.01 mΩ 3 ppm or 0.3 mΩ12 ppm or 0.4 mΩ15 ppm or 0.4 mΩ
Fast Speed0.1 mΩ 3.6 ppm or 1.9 mΩ 12.6 ppm or 2 mΩ15.6 ppm or 2 mΩ
Slow Speed0.1 mΩ 3 ppm or 4 mΩ12 ppm or 5 mΩ15 ppm or 5 mΩ
Fast Speed1 mΩ 3.6 ppm or 20 mΩ 12.6 ppm or 21 mΩ15.6 ppm or 21 mΩ
Resistance Ratio
Temperature (
°C) °
-189.34420.222.50.13
-38.83440.841.50.32
0.0110.60.15
231.9281.890.850.44
419.5272.571.51.11
660.3233.371.51.58
-189.34420.221.20.07
-38.83440.840.950.20
0.0110.260.07
231.9281.890.810.42
419.5272.5721.47
660.3233.3722.11
(23
24 Hour
±1) °°C
(Rx/Rs)
90 Days
(23
1 Year(23
ppm of reading
± 5) °C
± 5) °°C
Temperature (mK)
(23
Equivalent to
1 year
± 5) °°C
Excitation
Currert
±1 mA/±12 V3 ppm + 0.01 mΩMedium Speed0.01 mΩ 3 ppm or 0.55 mΩ12 ppm or 0.7 mΩ15 ppm or 0.7 mΩ
±1 mA/±12 V3 ppm + 0.02 mΩMedium Speed0.01 mΩ 3 ppm or 0.7 mΩ12 ppm or 0.8 mΩ15 ppm or 0.8 mΩ
±0.1 mA/±12 V3 ppm + 0.2 mΩMedium Speed0.1 mΩ 3 ppm or 8 mΩ12 ppm or 9 mΩ15 ppm or 9 mΩ
Temperature
Coefficient
06
PRT Measurement Accuracy using Internal Rs
°C
Scanning SpeedTemperature
-200 °C0.0005 0.0008 0.0008 0.0002
Slow Speed
Medium Speed
Fast Speed
[1] The indicator is based on the electrical accuracy of the 4-wire PT100 PRT and does not include the accuracy of the PRT itself.
[2] Temperature maximum Resolution is 0.0001 °C.
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature coefcient index: Exceed the range of (18-28) °C, increase (ppm of reading + xxμV)/°°C.
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Page 9
Thermistor Accuracy
Measurement
Range
( 0~12 ) kΩ
( 10~120 ) kΩ
Temperature Calibration Equipment
( 100~1000 ) kΩ
[1] Accuracy Index: ± (ppm of reading or xxmΩ , whichever is greater).
[2] Temperature coefcient index: exceeds (18-28) °C range, increase (ppm reading +xxmΩ) / °C.
[3] Specications are for 4-wire function.
Scanning Speed Resolution
Slow Speed1 mΩ 10 ppm or 60 mΩ30 ppm or 80 mΩ40 ppm or 80 mΩ
Fast Speed10 mΩ 10 ppm or 210 mΩ30 ppm or 230 mΩ40 ppm or 230 mΩ
[1] The indicator is based on the electrical accuracy of the 4-wire thermistor and does not include the accuracy of the thermistor itself.
[2] Temperature maximum Resolution is 0.0001 °C.
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature Coefcient index: Exceed the range of (18-28) °C, increase (ppm reading + xxμV)/°°C.
[3] Any range, the maximum input voltage is 50 V.
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature Coefcient index:Exceed the range of range of (18-28) °C, increase (ppm reading+ xxnA)/°°C.
[3] Input Protection 0.3A/600V Resettable PTC.
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature Coefcient index:Exceed the range of range of (18-28) °C, increase (ppm reading+ xx Ω)/°°C.
[3] The above is a 4-wire measurement index.
[4] When the range is less than or equal to 10 kΩ, the default is automatic current reversal.
[5] Max Lead Resistance(4-wire ohms): 10 Ω per lead for 100 Ω & 1 kΩ ranges; 100 Ω per lead for 10 kΩ &100 kΩ ranges; 1 kΩ per lead on all other ranges.