Additel ADT 286 Specifications

Page 1
Additel 286 Multifunction Reference Thermometer Readout
Temperature Calibration Equipment
Measure and calibrate SPRTs, RTDs, thermistors and thermocouples
1 PPM resistance ratio accuracy (channel 1)
8 1/2-digit DC multimeter
Measure up to 82 channels
Sample rates up to 10 channels per second
Bluetooth, WIFI, USB& Ethernet (RJ-45) capable
Auto temperature control of Additel and other manufacture's heat sources
Built-in automatic temperature control, data collection, and coefficient generation
Support for creating custom control of heat sources with RS-232
Auto - zero power feature (self heating compensation)
10.1 in. touch screen display
Supports fully automated temperature calibrations with data collection and report generation (no software required)
OVERVIEW
The Additel 286 Multifunction Reference Thermometer Readout is an industry first! We have combined the capabilities of a high-end reference thermometer with a highly capable data acquisition system and 8.5 digit multimeter. The ADT286 is capable of scanning and recording up to 82 channels at 10 channels per second. Users can
easily congure the ADT286 to perform eld calibrations and uniformity studies as well
as use the unit in the lab as a precision thermometer and 8.5 digit multimeter. Get more for less with this newest game changer from Additel!
01
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 2
ADT 286 Multifunction Reference Thermometer Readout Scanner Modules
If you're in need of a precision reference thermometer for your laboratory, then look no further than the Additel 286. The base unit comes with two precision readout channels that can be used to measure your SPRT. Need to calibrate RTDs, PRTs, thermistors or thermocouples? Add a scanner module and you now have the ability to measure 10 RTDs, PRTs, or thermistors and 20 thermocouples. Expand up to 82 channels with our unique easy to use scanner modules. Each 20 channel module is outfitted with our proprietary universal terminals with an industries best cold junction capability second to none. Utilize the module docked atop the ADT286, or connect remotely with cables to suit nearly any unique setup/configuration. Additel also has a process module specifically designed to measure process instrumentation like transmitters and switches. This scanner will also supply loop power for the transmitters.
Temperature Calibration Equipment
Designed to make your job easier, the ADT286 has a large sensor library supporting 15 TC types, both standard and special limits, 18 different thermal resistors, CVD, ITS-90, and a large variety of standard curves for RTDs and thermistors. The ADT286 is loaded with special applications such as probe calibration, SPRT calibration, chamber mapping and more. And we continue to add applications on a regular basis!
Automatic Temperature Control and Probe Calibration
The Additel 286 Multifunction Reference Thermometer Readout has preinstalled drivers to control Additel and other manufacturer’s heat sources. Simply connect to one or more heat sources via a communication cable, Ethernet or wireless and now it will automatically control to the set point and desired stability. If your heat source is not on the list, you can easily add the driver yourself so you can run automated calibrations with any heat source.
Now combine the heat source control feature with our probe calibration application and you have a very powerful automatic calibration solution. The probe calibration app allows you to automatically setup and run calibration routines
with multiple set points and multiple heat sources, collect data, and develop calibration coefcients — all with one device
and without the need of software! Simply place a batch of sensors of any mix and type into your heat source, connect it to the ADT286, run the probe calibration app and come back to a completed test. All that is left to do is generate and export all the calibration data. There’s no need to work with complicated software for communication, set up or
coefcient generation. There is no reason to have a calibration technician manually monitor the process and record the
data. This Multifunction Reference Thermometer Readout will do all the work for you.
ADT286-RS Resistance Standards
Available in 25 & 100 Ohm values, users can enjoy improved resistance ratio performance by easily plugging one of our reference resisters into channel 2 on the new ADT286. Perfect for calibrating your SPRT’s and high end PRT’s.
02
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 3
FEATURES
Specification Display Application
Multi-Channel
Temperature Calibration Equipment
Smart Connection
DAQ Mode
Temperature Mapping Mode
Probe Calibration
03
Environmental Temperature
Field Test
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 4
APPLICATIONS
Specification Display Application
SPRT Mode
Temperature Calibration Equipment
Switch Testing
Thermostatic Source Test
DMM
Heat Source Control
04
Snapshot
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 5
SPECIFICATIONS
General Specifications
Specification Description
Voltage
Frequency 47Hz to 440Hz.Automatically sensed at power-on
Temperature Calibration Equipment
Power Consumption 40VA peak (30Watt average)
Temperature
Warm-up 90 mins for full uncertainty specications
Relative Humidity (non-
condensing)
Altitude
Vibration and Shock Complies with MIL-28800F Class 3
Input Protection 50V all functions, ranges and terminals
Communication USB-A , USB-B , RJ45 , WiFi , Bluetooth
Memory 10G - All data stored with time stamps
Localization English , Chinese
Display 10.1 in (256 mm) TFT color display
Size (H x W x D) 9.8 in (250 mm) x 16.5 in (420 mm) x 7.9 in (200 mm)
Weight 18.5 lb ( 8.39 kg)
Other Conformities CE
100V Setting 90V to 110V 120V Setting 108V to 132V 220V Setting 198V to 242V 240V Setting 216V to 264V
Operating : 0°C to 50°C Full accuracy :18°C to 28°C Storage : -20°C to 70°C
Operating : 0°C to 28°C < 90% 28°C to 40°C < 75% 40°C to 50°C < 50% Storage : -20°C to 70°C < 95%
Operating : 2000 m Storage : 12000 m
Measurement Specifications
Specication Conditions: 90 mins Warm-Up Time / Environment Temperature (18 - 28) °C. The following specications apply for front panel, after at least 90 minutes warm-up. 24-hour specications are relative to calibration standards and assume a controlled electromagnetic environment per EN 61326.
Resistance Ratio Accuracy (Rx/Rs) using External Rs
± 5) °°C
Range Reference Resistance Ratio (Rx/Rs)
2.00-4.00 1.5
1.10-2.00 0.85
100 25 Ω
400 100 Ω
0.90-1.10 0.6
0.50-0.90 1.5
0.25-0.50 2.5
2.00-4.00 2
1.10-2.00 0.81
0.90-1.10 0.26
0.50-0.90 0.95
0.25-0.50 1.2
1 Year (23
ppm of Reading
05
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 6
SPRT/PRT Measurement Accuracy using External Rs
SPRT/PRT Type
PT25 25 Ω
External Reference
Resistance
Temperature Calibration Equipment
PT100 100 Ω
[1] The PT25 indicator is based on a nominal resistance of 25 Ω for Rx. [2] The PT100 indicator is based on a nominal resistance of 100 Ω for Rx.
[3] The uncertainty of external Rs is not included. The user may choose the ADT280-RS-25/100 standard resistor as external Rs, which has an accuracy of 5 ppm at (23±2) °C.
Resistance Accuracy using Internal Rs
Measurement
Range
( 0~100 )
( 0~400 )
( 0~4000 )
[1] Accuracy Index: ± (ppm of reading or xxmΩ , whichever is greater). [2] Temperature coefcient index: exceeds (18-28) °C range, increase per degree (ppm reading +xxmΩ). [3] Specications are for 4-wire function. For 3-wire, add 0.005 Ω for internal resistance mismatch. For 2-wire, add 0.005 Ω for internal resistance
[4] Automatic current reversal.
Scanning Speed Resolution
Slow Speed 0.01 mΩ 3 ppm or 0.2 mΩ 12 ppm or 0.35 mΩ 15 ppm or 0.35 mΩ
Fast Speed 0.1 mΩ 3.6 ppm or 1.7 mΩ 12.6 ppm or 1.85 mΩ 15.6 ppm or 1.85 mΩ
Slow Speed 0.01 mΩ 3 ppm or 0.3 mΩ 12 ppm or 0.4 mΩ 15 ppm or 0.4 mΩ
Fast Speed 0.1 mΩ 3.6 ppm or 1.9 mΩ 12.6 ppm or 2 mΩ 15.6 ppm or 2 mΩ
Slow Speed 0.1 mΩ 3 ppm or 4 mΩ 12 ppm or 5 mΩ 15 ppm or 5 mΩ
Fast Speed 1 mΩ 3.6 ppm or 20 mΩ 12.6 ppm or 21 mΩ 15.6 ppm or 21 mΩ
Resistance Ratio
Temperature (
°C) °
-189.3442 0.22 2.5 0.13
-38.8344 0.84 1.5 0.32
0.01 1 0.6 0.15
231.928 1.89 0.85 0.44
419.527 2.57 1.5 1.11
660.323 3.37 1.5 1.58
-189.3442 0.22 1.2 0.07
-38.8344 0.84 0.95 0.20
0.01 1 0.26 0.07
231.928 1.89 0.81 0.42
419.527 2.57 2 1.47
660.323 3.37 2 2.11
(23
24 Hour
±1) °°C
(Rx/Rs)
90 Days
(23
1 Year(23
ppm of reading
± 5) °C
± 5) °°C
Temperature (mK)
(23
Equivalent to
1 year
± 5) °°C
Excitation
Currert
±1 mA/±12 V 3 ppm + 0.01 mΩMedium Speed 0.01 mΩ 3 ppm or 0.55 mΩ 12 ppm or 0.7 mΩ 15 ppm or 0.7 mΩ
±1 mA/±12 V 3 ppm + 0.02 mΩMedium Speed 0.01 mΩ 3 ppm or 0.7 mΩ 12 ppm or 0.8 mΩ 15 ppm or 0.8 mΩ
±0.1 mA/±12 V 3 ppm + 0.2 mΩMedium Speed 0.1 mΩ 3 ppm or 8 mΩ 12 ppm or 9 mΩ 15 ppm or 9 mΩ
Temperature
Coefficient
06
PRT Measurement Accuracy using Internal Rs
°C
Scanning Speed Temperature
-200 °C 0.0005 0.0008 0.0008 0.0002
Slow Speed
Medium Speed
Fast Speed
[1] The indicator is based on the electrical accuracy of the 4-wire PT100 PRT and does not include the accuracy of the PRT itself. [2] Temperature maximum Resolution is 0.0001 °C.
0 °C 0.0008 0.0031 0.0038 0.0008
300 °C 0.0018 0.0089 0.0089 0.0018
600 °C 0.0029 0.0146 0.0146 0.0030
-200 °C 0.0013 0.0016 0.0016 0.0002
0 °C 0.0014 0.0031 0.0038 0.0008
300 °C 0.0020 0.0089 0.0089 0.0018
600 °C 0.0029 0.0146 0.0146 0.0030
-200 °C 0.0039 0.0043 0.0043 0.0006
0 °C 0.0044 0.0047 0.0047 0.0013
300 °C 0.0053 0.0093 0.0093 0.0024
600 °C 0.0059 0.0152 0.0152 0.0036
24 Hour /
±1) °°C
(23
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
90 Days /
± 5) °°C
(23
°C
1 year /
± 5) °°C
(23
°C
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Temperature Coefficient
°C/°C
Page 7
Thermocouple Voltage Accuracy
Test Range Scanning Speed Resolution
Slow Speed 0.01 µV 5 ppm + 2 ppm 10 ppm + 4 ppm 14 ppm + 4 ppm
(-100-100) mV
[1] Accuracy Index: ± (ppm of reading + ppm of FS). [2] Temperature coefcient index: Exceed the range of (18-28) °C, increase (ppm of reading + xxμV)/°°C.
Medium Speed 0.01 µV 5 ppm + 6 ppm 10 ppm + 8 ppm 14 ppm + 8 ppm
Fast Speed 0.1 µV 5 ppm + 22 ppm 10 ppm + 24 ppm 14 ppm + 24 ppm
Temperature Calibration Equipment
Thermocouple Cold Junction Accuracy
CJC Accuracy
Environmental Coefcient Beyond (18 ~ 28) °C, add 0.02 °C / °C
Other Each signal scanner has 10 cold Junction temperature sensors
Thermocouple Temperature Accuracy
Type Temperature
Fast Speed Medium Speed Slow Speed Fast Speed Medium Speed Slow Speed Fast Speed Medium Speed Slow Speed
°C , 1 year, 23 °C
±0.1
°°C
24 Hour/
±1) °°C
(23
24 hours
±1) °°C
(23
± 5
90 days
±5) °°C
(23
°C
90 days/
± 5) °°C
(23
°°C
(23
1 year
±5) °°C
1 year /
±5) °°C
(23
Temperature
Coefficient
°°C
Input Resistance
10 MΩ or >10 GΩ 1 ppm + 0.1 µV
-200 0.089 0.038 0.022 0.099 0.047 0.031 0.100 0.049 0.033
-100 0.049 0.021 0.012 0.054 0.026 0.017 0.055 0.026 0.017
-40 0.041 0.017 0.009 0.045 0.021 0.013 0.045 0.021 0.014
E
J
K
0 0.038 0.015 0.009 0.041 0.019 0.012 0.041 0.019 0.012
155 0.031 0.013 0.008 0.035 0.017 0.011 0.036 0.017 0.012
350 0.029 0.013 0.008 0.033 0.017 0.012 0.035 0.018 0.013
660 0.031 0.014 0.009 0.036 0.020 0.015 0.039 0.022 0.017
1000 0.034 0.017 0.012 0.042 0.025 0.019 0.046 0.029 0.024
-200 0.102 0.043 0.025 0.113 0.054 0.036 0.115 0.055 0.037
-100 0.054 0.022 0.013 0.060 0.028 0.018 0.060 0.028 0.019
-40 0.047 0.019 0.011 0.051 0.024 0.015 0.051 0.024 0.015
0 0.044 0.018 0.010 0.048 0.022 0.014 0.048 0.022 0.014
155 0.041 0.017 0.010 0.045 0.021 0.014 0.046 0.022 0.015
350 0.042 0.018 0.011 0.047 0.023 0.016 0.048 0.025 0.018
660 0.039 0.018 0.011 0.046 0.024 0.018 0.048 0.027 0.020
1200 0.044 0.022 0.015 0.054 0.031 0.024 0.059 0.036 0.029
-200 0.146 0.061 0.035 0.161 0.076 0.050 0.163 0.077 0.051
-100 0.073 0.030 0.017 0.080 0.037 0.024 0.080 0.038 0.025
-40 0.060 0.025 0.014 0.066 0.030 0.020 0.066 0.031 0.020
0 0.056 0.023 0.013 0.061 0.028 0.018 0.061 0.028 0.018
155 0.056 0.023 0.013 0.061 0.029 0.019 0.062 0.030 0.020
350 0.054 0.023 0.014 0.061 0.030 0.020 0.062 0.031 0.021
660 0.055 0.025 0.015 0.063 0.033 0.023 0.066 0.035 0.026
1372 0.073 0.035 0.023 0.087 0.049 0.037 0.093 0.055 0.043
07
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 8
Thermocouple Temperature Accuracy
Type Temperature
Fast Speed Medium Speed Slow Speed Fast Speed Medium Speed Slow Speed Fast Speed Medium Speed Slow Speed
-200 0.142 0.059 0.034 0.156 0.073 0.048 0.157 0.075 0.049
-100 0.078 0.032 0.018 0.086 0.040 0.026 0.086 0.040 0.026
-40 0.063 0.026 0.015 0.069 0.032 0.020 0.069 0.032 0.021
T
Temperature Calibration Equipment
R
S
B
N
L
U
[1] The index is based on the accuracy of the thermocouple electrical measurement of temperature scanner module, does not include the accuracy of the
thermocouple itself and the xed cold junction compensation at 0 °C.
[2] The highest temperature resolution is 0.0001 °C.
0 0.057 0.023 0.013 0.062 0.028 0.018 0.062 0.028 0.018
155 0.044 0.019 0.011 0.049 0.023 0.015 0.049 0.024 0.016
350 0.038 0.016 0.010 0.043 0.021 0.015 0.044 0.022 0.016
400 0.037 0.016 0.010 0.042 0.021 0.015 0.044 0.023 0.016
-40 0.543 0.222 0.124 0.593 0.272 0.173 0.593 0.272 0.173
0 0.416 0.170 0.095 0.454 0.208 0.132 0.454 0.208 0.132
155 0.266 0.109 0.061 0.290 0.134 0.086 0.291 0.134 0.086
350 0.220 0.091 0.051 0.241 0.112 0.072 0.242 0.113 0.073
660 0.192 0.080 0.046 0.212 0.100 0.066 0.214 0.102 0.068
1768 0.188 0.082 0.049 0.213 0.107 0.074 0.219 0.114 0.081
-40 0.515 0.211 0.117 0.562 0.258 0.164 0.562 0.258 0.164
0 0.407 0.167 0.093 0.444 0.204 0.130 0.444 0.204 0.130
155 0.275 0.113 0.063 0.300 0.138 0.089 0.301 0.139 0.089
350 0.236 0.098 0.055 0.259 0.120 0.078 0.260 0.122 0.079
660 0.214 0.089 0.051 0.236 0.111 0.073 0.239 0.114 0.075
1768 0.222 0.096 0.057 0.250 0.124 0.086 0.257 0.132 0.093
250 0.872 0.357 0.199 0.952 0.437 0.278 0.952 0.437 0.279
350 0.619 0.254 0.141 0.676 0.311 0.198 0.676 0.311 0.199
660 0.342 0.141 0.079 0.374 0.173 0.111 0.375 0.175 0.113
1820 0.199 0.085 0.050 0.222 0.108 0.073 0.227 0.113 0.078
-200 0.224 0.093 0.052 0.246 0.115 0.075 0.247 0.116 0.076
-100 0.106 0.044 0.024 0.116 0.054 0.035 0.116 0.054 0.035
-40 0.089 0.036 0.020 0.097 0.045 0.029 0.097 0.045 0.029
0 0.084 0.035 0.019 0.092 0.042 0.027 0.092 0.042 0.027
155 0.070 0.029 0.017 0.077 0.036 0.024 0.078 0.037 0.024
350 0.062 0.026 0.015 0.069 0.033 0.022 0.070 0.035 0.024
660 0.059 0.026 0.016 0.067 0.034 0.024 0.069 0.036 0.026
800 0.060 0.027 0.016 0.068 0.035 0.025 0.071 0.038 0.028
1000 0.062 0.028 0.018 0.072 0.038 0.028 0.075 0.042 0.031
1200 0.065 0.030 0.019 0.076 0.041 0.031 0.081 0.046 0.035
1300 0.068 0.032 0.020 0.080 0.044 0.033 0.085 0.049 0.038
-200 0.069 0.029 0.017 0.076 0.036 0.024 0.077 0.037 0.025
-100 0.053 0.022 0.013 0.059 0.028 0.018 0.059 0.028 0.018
-40 0.045 0.019 0.010
0 0.043 0.018 0.010 0.047 0.021 0.014 0.047 0.021 0.014
155 0.040 0.017 0.010 0.044 0.021 0.014 0.045 0.022 0.015
350 0.041 0.018 0.011 0.046 0.023 0.016 0.047 0.024 0.017
660 0.039 0.018 0.011 0.046 0.024 0.018 0.048 0.027 0.020
900 0.035 0.017 0.011 0.042 0.023 0.017 0.045 0.026 0.021
-80 0.072 0.030 0.017 0.079 0.037 0.024 0.079 0.037 0.024
-40 0.062 0.026 0.014 0.068 0.031 0.020 0.068 0.032 0.020
0 0.056 0.023 0.013 0.061 0.028 0.018 0.061 0.028 0.018
155 0.045 0.019 0.011 0.049 0.023 0.015 0.050 0.024 0.016
350 0.037 0.016 0.010 0.042 0.021 0.014 0.043 0.022 0.016
600 0.034 0.015 0.010 0.039 0.021 0.015 0.041 0.023 0.017
24 Hour/
±1) °°C
(23
°°C
°°C
90 days/
± 5) °°C
(23
0.049 0.023 0.015 0.050 0.023 0.015
1 year /
±5) °°C
(23
°°C
08
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 9
Thermistor Accuracy
Measurement
Range
( 0~12 ) k
( 10~120 ) k
Temperature Calibration Equipment
( 100~1000 ) k
[1] Accuracy Index: ± (ppm of reading or xxmΩ , whichever is greater). [2] Temperature coefcient index: exceeds (18-28) °C range, increase (ppm reading +xxmΩ) / °C. [3] Specications are for 4-wire function.
Scanning Speed Resolution
Slow Speed 1 mΩ 10 ppm or 60 mΩ 30 ppm or 80 mΩ 40 ppm or 80 mΩ
Fast Speed 10 mΩ 10 ppm or 210 mΩ 30 ppm or 230 mΩ 40 ppm or 230 mΩ
Slow Speed 10 mΩ 10 ppm 30 ppm 40 ppm
Fast Speed 100 mΩ 10.6 ppm + 200 mΩ 30.6 ppm + 200 mΩ 40.6 ppm + 200 mΩ
Slow Speed 0.1 Ω 50 ppm 80 ppm 100 ppm
Fast Speed 1 Ω 51 ppm + 2 Ω 81 ppm + 2 Ω 101 ppm + 2 Ω
Thermistor Temperature Accuracy
Type Scanning Speed Temperature
Slow Speed
10 k
[1] The indicator is based on the electrical accuracy of the 4-wire thermistor and does not include the accuracy of the thermistor itself. [2] Temperature maximum Resolution is 0.0001 °C.
Medium Speed
Fast Speed
24 Hour
±1) °°C
(23
24 Hour /
±1) °°C
(23
-40 °C 0.0007 0.0011 0.0014 0 °C 0.0002 0.0006 0.0008
50 °C 0.0004 0.0008 0.0011 100 °C 0.0030 0.0039 0.0039 150 °C 0.0130 0.0174 0.0174
-40 °C 0.0007 0.0011 0.0014
0 °C 0.0002 0.0006 0.0008
50 °C 0.0008 0.0010 0.0011 100 °C 0.0054 0.0064 0.0064 150 °C 0.0239 0.0282 0.0282
-40 °C 0.0007 0.0011 0.0014 0 °C 0.0002 0.0006 0.0008
50 °C 0.0016 0.0016 0.0016 100 °C 0.0104 0.0104 0.0104 150 °C 0.0456 0.0456 0.0456
90 Days
± 5) °C
(23
°C
90 Days /
± 5) °°C
(23
1 year
± 5) °°C
(23
°C
1 year /
± 5) °°C
(23
Excitation
°C
Currert
10 µA 5 ppm + 10 mΩMedium Speed 1 mΩ 10 ppm or 110 mΩ 30 ppm or 130 mΩ 40 ppm or 130 mΩ
10 µA 5 ppm + 20 mΩMedium Speed 10 mΩ 10 ppm + 80 mΩ 30 ppm + 80 mΩ 40 ppm + 80 mΩ
1 µA 5 ppm + 1 ΩMedium Speed 0.1 Ω 50 ppm + 1 Ω 80 ppm + 1 Ω 100 ppm + 1 Ω
Temperature
Coefficient
09
DC Voltage Accuracy
Test Range Scanning Speed Resolution
Slow Speed 0.01 µV 5 ppm + 2 ppm 10 ppm + 4 ppm 14 ppm + 4 ppm
(-100-100) mV
Fast Speed 0.1 µV 5 ppm + 22 ppm 10 ppm + 24 ppm 14 ppm + 24 ppm
Slow Speed 0.1 µV 2 ppm + 0.3 ppm 8 ppm + 0.6 ppm 14 ppm + 0.6 ppm
(-1-1) V
Fast Speed 1 µV 2.6 ppm + 3.3 ppm 8.6 ppm + 3.6 ppm 14.6 ppm + 3.6 ppm
Slow Speed 1 µV 2 ppm + 0.05 ppm 8 ppm + 0.08 ppm 14 ppm + 0.08 ppm
(-10-10) V
Fast Speed 10 µV 2.6 ppm + 1.05 ppm 8.6 ppm + 1.08 ppm 14.6 ppm + 1.08 ppm
Slow Speed 10 µV 8 ppm + 1 ppm 32 ppm + 1 ppm 38 ppm + 1 ppm
(-50-50) V
Fast Speed 100 µV 8.6 ppm + 7 ppm 32.6 ppm + 7 ppm 38.6 ppm + 7 ppm
[1] Accuracy Index: ± (ppm of reading + ppm of FS). [2] Temperature Coefcient index: Exceed the range of (18-28) °C, increase (ppm reading + xxμV)/°°C. [3] Any range, the maximum input voltage is 50 V.
24 hours
±1) °°C
(23
90 days
±5) °°C
(23
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
(23
1 year
±5) °°C
Input Resistance
>10 GΩ or 10 MΩ 1 ppm + 0.1 µVMedium Speed 0.01 µV 5 ppm + 6 ppm 10 ppm + 8 ppm 14 ppm + 8 ppm
>10 GΩ or 10 MΩ 1 ppm + 0.2 µVMedium Speed 0.1 µV 2 ppm + 1.3 ppm 8 ppm + 1.6 ppm 14 ppm + 1.6 ppm
>10 GΩ or 10 MΩ 1 ppm + 0.3 µVMedium Speed 1 µV 2 ppm + 0.35 ppm 8 ppm + 0.38 ppm 14 ppm + 0.38 ppm
10 MΩ 5 ppm + 5 µVMedium Speed 10 µV 8 ppm + 2 ppm 32 ppm + 2 ppm 38 ppm + 2 ppm
Temperature
Coefficient
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 10
DC Current Accuracy
Test Range Scanning Speed Resolution
Slow Speed 0.01 nA 15 ppm + 3 ppm 50 ppm + 6 ppm 60 ppm + 6 ppm
(-100-100) µA
Fast Speed 0.1 nA 15 ppm + 23 ppm 50 ppm +26 ppm 60 ppm + 26 ppm
Slow Speed 0.1 nA 15 ppm + 0.6 ppm 50 ppm +1 ppm 60 ppm + 1 ppm
(-1-1) mA
Temperature Calibration Equipment
(-10-10) mA
(-100-100) mA
[1] Accuracy Index: ± (ppm of reading + ppm of FS). [2] Temperature Coefcient index:Exceed the range of range of (18-28) °C, increase (ppm reading+ xxnA)/°°C. [3] Input Protection 0.3A/600V Resettable PTC.
Fast Speed 1 nA 15.6 ppm + 3.6ppm 50.6 ppm + 4 ppm 60.6 ppm + 4 ppm
Slow Speed 1 nA 30 ppm + 3 ppm 75 ppm + 6 ppm 80 ppm + 6 ppm
Fast Speed 10 nA 30 ppm + 23 ppm 75 ppm + 26 ppm 80 ppm + 26 ppm
Slow Speed 10 nA 40 ppm + 0.6 ppm 75 ppm + 1 ppm 80 ppm + 1 ppm
Fast Speed 100 nA 40.6 ppm + 3.6 ppm 75.6 ppm + 4 ppm 80.6 ppm + 4 ppm
DC Resisstance Accuracy
Test Range Scanning Speed Resolution
Slow Speed 0.01 mΩ 3 ppm + 1 ppm 13 ppm + 1.5 ppm 16 ppm + 1.5 ppm
(0-100)
Fast Speed 0.1 mΩ 3 ppm + 21 ppm 13 ppm + 21.5 ppm 16 ppm + 21.5 ppm
Slow Speed 0.1 mΩ 3 ppm + 0.2 ppm 12 ppm + 0.3 ppm 15 ppm + 0.3 ppm
(0-1) k
Fast Speed 1 mΩ 3.6 ppm + 3.2 ppm 12.6 ppm + 3.3 ppm 15.6 ppm + 3.3 ppm
Slow Speed 1 mΩ 3 ppm + 0.3 ppm 12 ppm + 0.4 ppm 15 ppm + 0.4 ppm
(0-10) k
Fast Speed 10 mΩ 3.6 ppm + 3.3 ppm 12.6 ppm + 3.4 ppm 15.6 ppm + 3.4 ppm
Slow Speed 10 mΩ 3 ppm + 0.2 ppm 12 ppm + 0.3 ppm 15 ppm + 0.3 ppm
(0-100) k
Fast Speed 100 mΩ 3.6 ppm + 1.3 ppm 12.6 ppm + 1.3 ppm 30.6 ppm + 1.3 ppm
Slow Speed 0.1 Ω 10 ppm + 0.6 ppm 30 ppm + 1 ppm 40 ppm + 1 ppm
(0-1) M
Fast Speed 1 Ω 10 ppm + 2.6 ppm 30 ppm + 3 ppm 40 ppm + 3 ppm
Slow Speed 1 Ω 50 ppm + 0.4 ppm 80 ppm + 1 ppm 100 ppm + 1 ppm
(0-10) M
Fast Speed 10 Ω 50 ppm + 4.4 ppm 80 ppm + 5 ppm 100 ppm + 5 ppm
Slow Speed 10 Ω 150 ppm + 1 ppm 400 ppm + 4 ppm 500 ppm + 4 ppm
(0-100) M
Fast Speed 100 Ω 150 ppm + 11 ppm 400 ppm + 14 ppm 500 ppm + 14 ppm
[1] Accuracy Index: ± (ppm of reading + ppm of FS). [2] Temperature Coefcient index:Exceed the range of range of (18-28) °C, increase (ppm reading+ xx Ω)/°°C. [3] The above is a 4-wire measurement index.
[4] When the range is less than or equal to 10 kΩ, the default is automatic current reversal. [5] Max Lead Resistance(4-wire ohms): 10 Ω per lead for 100 Ω & 1 kΩ ranges; 100 Ω per lead for 10 kΩ &100 kΩ ranges; 1 kΩ per lead on all other ranges.
24 hours
±1) °°C
(23
24 hours
±1) °°C
(23
90 days
(23
90 days
(23
±5) °°C
±5) °°C
(23
(23
1 year
±5) °°C
1 year
±5) °°C
Burden Voltage
<1 mV 8 ppm + 0.1 nAMedium Speed 0.01 nA 15 ppm + 7 ppm 50 ppm + 10 ppm 60 ppm + 10 ppm
<1 mV 8 ppm + 0.5 nA Medium Speed 0.1 nA 15 ppm + 1.6 ppm 50 ppm + 2 ppm 60 ppm + 2 ppm
<1 mV 8 ppm + 10 nA Medium Speed 1 nA 30 ppm + 7 ppm 75 ppm + 10 ppm 80 ppm + 10 ppm
<1 mV 8 ppm + 50 nA Medium Speed 10 nA 40 ppm + 1.6 ppm 75 ppm + 2 ppm 80 ppm + 2 ppm
Excitation
Current
1 mA 3 ppm + 0.01 mΩ Medium Speed 0.01 mΩ 3 ppm + 5 ppm 13 ppm + 5.5 ppm 16 ppm + 5.5 ppm
1 mA 3 ppm + 0.02 mΩMedium Speed 0.1 mΩ 3 ppm + 1.2 ppm 12 ppm + 1.3 ppm 15 ppm + 1.3 ppm
0.1 mA 3 ppm + 0.2 mΩMedium Speed 1 mΩ 3 ppm + 1.3 ppm 12.6 ppm + 1.3 ppm 15 ppm + 1.3 ppm
0.1 mA 3 ppm + 20 mΩMedium Speed 10 mΩ 3 ppm + 0.5 ppm 12 ppm + 0.6 ppm 15 ppm + 0.6 ppm
10 µA 5 ppm + 0.2 ΩMedium Speed 0.1 Ω 10 ppm + 1.2 ppm 30 ppm + 0.6 ppm 40 ppm + 0.6 ppm
1 µA 10 ppm + 1 Ω Medium Speed 1 Ω 50 ppm + 1.4 ppm 80 ppm + 2 ppm 100 ppm + 2 ppm
0.1 µA 50 ppm + 50 ΩMedium Speed 10 Ω 150 ppm + 6 ppm 400 ppm + 9 ppm 500 ppm + 9 ppm
Temperature Coefficient
Temperature Coefficient
10
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 11
Ordering Information
Model Number
Model Description Picture
ADT286
ADT286-TS-PKG
Temperature Calibration Equipment
ADT286-PS-PKG
Thermometer Readout base
Thermometer Readout base
unit with (1) Temperature
Scanner Module (9051 cable
Thermometer Readout base
unit with (1) Process Scanner
Accessories
Accessories (Included)
Standard Accessories Quantity Picture
Shorting Block 1 pc.
USB Cable (UK-415) 1 pc.
Test leads 4 sets (8 pcs)
9026
4-Wire Test Leads
(Only w/ ADT286-TS-PKG &
ADT286-PS-PKG)
Fuse (0213.315MXP) 1 pc.
Calibration Certicate 1 pc.
CD Manual 1 pc.
Multifunction Reference
unit only
Multifunction Reference
not included)
Multifunction Reference
Module (9051 cable not
included)
1 pc.
Standard Reference Resistor Ordering Information
ADT280-RS
25
Norminal Resistance:
25 - 25 100 - 100
ADT280-RS-X
Resistance Standards
Specification ADT280-RS-25 ADT280-RS-100
Nominal
Resistance
Stability 5 ppm/year 5 ppm/year
Operating
Temperature Temperature
Coefcient
Size
Weight 0.35 lb (160 g) 0.35 lb (160 g)
Excitation Current 1 mA 1 mA
57 mm x 57 mm x 45 mm 57 mm x 57 mm x 45 mm
25 Ω 100 Ω
23 °C±2 °C 23 °C±2 °C
0.5 ppm/°C 0.5 ppm/°C
11
Optional Accessories
Model
9026 4-wire test leads(10-Pack)
9051-10 Dsub Comm Cable=10 ft
9051-33 Dsub Comm Cable=33 ft
Carrying Case for ADT286,(2)
9916-286
ADT286-DOCK
ADT286-TS
ADT286-PS
ADT280-RS-25
ADT280-RS-100
Optional
Accessories
scanner modules and
reference probe w/wheels
Remote Module Docking
Station w/AC Adapter
ADT286 Temperature
Scanner Module
ADT286 Process Scanner
Module
25 Ω Standard Reference
Resistor
100 Ω Standard Reference
Resistor
Picture
Secondary Standard PRT Ordering Information
AM1760
Secondary PRT Model:
AM1760 AM1762
12
PRT Exterior:
12 - 12 inch straight 20 - 20 inch straight
AM17XX-X-SP
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
SP
Page 12
Secondary Standard PRT Information
Specification AM1760 Series AM1762 Series
Temperature Range -200 °C to 670 °C -200 °C to 670 °C
Resistance at 0
Temperature Coefficient 0.003925 Ω / Ω / °C
Accuracy
Temperature Calibration Equipment
Drift
Short Term Stability
Thermal Shock
Hysteresis N/A
Self-heating 0.0015 °C at 1 mA current
Response Time 9 seconds for 63% response to step change in water moving at 3 feet per second
Measurement Current 0.5 mA or 1 mA
Sensor Length 42 mm
Sensor Location 5 mm from tip
Insulation Resistance >1000 MΩ at room temperature
Sheath Material Inconel
Nominal 100 Ω Nominal 25 Ω
0.007 °C at -196 °C
±
0.006 °C at 0.01 °C
±
0.015 °C at 420 °C
±
0.025 °C at 660 °C
±
0.004 °C at TPW after 100 hours at 661 °C
±
0.002 °C
±
0.002 °C after 10 times thermal cycles from minimum to maximum temperatures
±
tm
0.007 °C at -196 °C
±
0.006 °C at 0.01 °C
±
0.015 °C at 420 °C
±
0.025 °C at 660 °C
±
Dimension
External Leads Teon
Handle Dimension 15 mm (OD) x 65 mm (L)
Handle Temperature Range
Calibration NIST traceable calibration w/ data included
[1] Handle temperatures outside the usable will cause damage to the probe. * PRT Information from www.accumac.com.
[1]
0.25 in dia X 12 in (6.35 mm X 305 mm)
0.25 in dia X 20 in (6.35 mm X 500 mm)
AM1760-12-SP
AM1760-20-SP
tm
– insulated copper wire, 4 leads, 2.5 meters
-50 °C to 160 °C -50 °C to 180° C
0.25 in dia X 12 in (6.35 mm X 305 mm)
AM1762-12-SP
AM1762-20-SP
0.25 in dia X 20 in (6.35 mm X 500 mm)
12
Corporate Headquarters
2900 Saturn Street #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Loading...