A&D UA-767NFC User Manual

Digital Blood Pressure Monitor
Model UA-767NFC
Instruction Manual
1WMPD4002380
Contents
Dear Customers.................................................................... 2
Preliminary Remarks ............................................................. 2
Precautions........................................................................... 2
Parts Identification................................................................ 3
Symbols................................................................................ 4
Operation Mode.................................................................... 6
Using the Monitor ................................................................. 7
Installing / Changing the Batteries ................................... 7
Connecting the Air Hose .................................................. 7
Connecting the AC Adapter.............................................. 7
Notes on the Built-in Clock .............................................. 8
Selecting the Clock Mode ................................................. 8
Adjusting the Clock ......................................................... 9
Selecting the Correct Cuff Size ....................................... 10
Applying the Arm Cuff ................................................... 10
How to Take Proper Measurements ................................ 11
Measurement................................................................. 11
After Measurement ........................................................ 11
Measurements .................................................................... 12
Normal Measurement..................................................... 12
Measurement with the Desired Systolic Pressure ............ 13
Notes for Proper Measurement....................................... 13
Near Field Communication (NFC) Function .......................... 14
Memory Data Transmission............................................ 14
Measurement Data Transmission ...................................14
Caution on Using the NFC Function..................................... 15
What is an Irregular Heartbeat............................................. 15
Pressure Bar Indicator ......................................................... 15
About Blood Pressure.......................................................... 15
What is Blood Pressure? ................................................. 15
What is Hypertension and How is it Controlled?.............. 16
Why Measure Blood Pressure at Home?........................... 16
WHO Blood Pressure Classification ................................. 16
Blood Pressure Variations............................................... 16
Troubleshooting ................................................................. 17
Maintenance ....................................................................... 18
Technical Data.................................................................... 18
Dear Customers
Congratulations on purchasing a state-of-the-art A&D blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen.
We recommend that you read through this manual carefully before using the device for the first time.
Preliminary Remarks
This device conforms to the European Directive 93/42 EEC for Medical
Products. This is made evident by the
(0366: The reference number to the involved notified body)
The device is designed for use on adults only, not newborns or infants. Environment for use
The device is intended for indoor use.
mark of conformity.
Precautions
Precision components are used in this device. Extreme temperature,
humidity, direct sunlight, shock or dust should be avoided.
Clean the device and cuff with a dry, soft cloth or a cloth dampened with
water and a neutral detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean the device or cuff.
Avoid tightly folding the cuff or storing the hose tightly twisted for long
periods, as such treatment may shorten the life of the components.
The device and cuff are not water resistant. Prevent rain, sweat and water
from soiling the device and cuff.
Measurements may be affected if the device is used near a television,
microwave oven, cellular telephone, X-ray or other devices that emit electromagnetic waves.
Used equipment, parts and batteries are not treated as ordinary
household waste, and must be disposed of according to the applicable local regulations.
When your arm is squeezed tightly by a cuff, blood collects in the lower
arm and finger tips and your arm may become congested.
2
A
r
A
k
A
Proper Fit Range
rtery Position Mar
Display
Index Mark
Arm Cuff
ir Connector Plug
Battery Compartment
Pressure Bar Indicato
Parts Identification
ir Hose
Display
Air Socket
MEMORY
Clock Button
Button
Battery Cover
Systolic Pressure
Diastolic Pressure
Pulse Rate
DC Jack
Communication Area
START Button
1.5V Batteries
(R6P, LR6 or AA)
Heart Mark
Battery Indicator
3
Clock Display
Clock Mark
I.H.B. Symbol (Irregular heartbeat symbol)
Symbols
Symbols that are printed on the device case
Symbols Function / Meaning Recommended Action
Standby and Turn the device on.
SYS. Systolic blood pressure in mmHg
DIA. Diastolic blood pressure in mmHg
PUL./min. Pulse per minute
SN
2010
Battery installation guide Direct current Serial number Date of manufacture Type BF: Device, cuff and tubing are
designed to provide special protection against electrical shocks.
EC directive medical device label WEEE label Manufacturer EU-representative
Symbols that appear on the display
Symbols Function / Meaning Recommended Action
Appears while measurement is in progress. It blinks when the pulse is detected.
Measurement is in progress. Remain as still as possible.
Irregular Heartbeat symbol (I.H.B.) Appears when an irregular heartbeat is detected. It may light when a very slight vibration like shivering or shaking is detected.
Previous measurements stored in memory.
4
Symbols
Symbols Function / Meaning Recommended Action
Full Battery
Low Battery
The battery power indicator during measurement.
The battery is low when it blinks.
Unstable blood pressure due to movement during measurement.
The systolic and diastolic values are within 10 mmHg of each other.
The pressure value did not increase during the inflation.
The cuff is not applied correctly.
Replace all batteries with new ones when the mark blinks.
Take another measurement. Remain very still during measurement.
Apply the cuff correctly, and take another measurement.
PUL. DISPLAY ERROR
The pulse is not detected correctly.
The communication device may be slanted or placed out of the communication area of the blood pressure
Cannot communicate.
monitor. Hold the communication
symbol of the NFC-enabled communication device close to the communication area of the monitor until “end” appears.
Appears during the clock setting
procedure AM Time between 0:00 and 11:59 PM Time between 12:00 and 23:59
5
Operation Mode
1. Normal Measurement
Press the START button. Blood pressure is measured and the data is stored
in memory.
2. Memory Function
This monitor can store the last 100 measurements in memory.
When the measurement exceeds 100, the oldest data will be deleted automatically and the most recent data will be stored. This monitor does not display the data stored in memory. The memory data can only be read by an NFC-enabled device.
3. Deleting all Data Stored in Memory
Make sure that the monitor displays the clock. Press and hold the The
mark appears. After a while it starts
blinking, then disappears.
Now all data stored in memory has been deleted.
Note: This operation will delete all data stored in
memory. You cannot select which data to delete.
button.
4. Measurement with the Desired Systolic Pressure
Refer to page 13 for measurement with the desired systolic pressure.
6
Using the Monitor
A
A
j
Installing / Changing the Batteries
1. Slide the battery cover up to open it.
Step 1
Step 3
2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+ and -) are correct. Use only R6P, LR6 or AA batteries.
3. Slide the battery cover down to close.
Step 2
CAUTION
Insert the batteries as shown in the battery compartment. If installed
incorrectly, the device will not work.
When
(LOW BATTERY mark)
blinks on the display, replace all batteries with new ones. Do not mix old and new batteries. It may shorten the battery life, or cause the device to malfunction. Wait two seconds or more after turning the device off, to replace the batteries. If (LOW BATTERY mark) appears even after the batteries are replaced, make a blood pressure measurement. The device may then recognize the new batteries.
(LOW BATTERY mark)
does not appear when the batteries are drained.
The battery life varies with the ambient temperature and may be shorter at
low temperatures.
Use the specified batteries only. Remove the batteries if the device is not to be used for a long time.
The batteries may leak and cause a malfunction.
Connecting the Air Hos e
Insert the air connector plug
into the air socket firmly.
ir connector plug
ir socket
Connecting the AC Adapter
Insert the AC adapter plug into
ack
the DC jack. Next, connect the AC adapter to an electrical outlet.
Use the specified AC adapter.
(Refer to page 18.)
7
DC
AC adapter plug
k
k
Using the Monitor
Notes on the Built-in Clock
The monitor displays the clock as long as power is connected. Adjust the clock prior to use. The current clock setting will be cleared if the batteries are removed, or
power is not connected for approximately 30 seconds while the AC adapter is used. In this case, adjust the clock again.
When adjusting the clock, press the
one, or press and hold the
button to change the value continuously.
The setting procedure cannot be reversed. If the wrong value is set, start
the setting procedure again.
After 30 seconds of non-operation, the setting procedure will be
automatically canceled. To cancel the operation anytime, press the START button. During measurement, the clock adjustment is not available.
Selecting the Clock Mode
Two clock modes are available, 12-hour clock or 24-hour clock.
1. Press and hold the until the year starts blinking.
2. Press and hold the “12H” appears on the display.
3. Press the
button to switch
between “12H” and “24H”.
4. Press the
button to display
the clock in the selected mode.
button
Press and hold
button.
Press and hold
button to increase the value by
Press
12-hour
cloc
24-hour
cloc
8
P
d
r
k
k
Using the Monitor
Adjusting the Clock
1. Press and hold the the year starts blinking.
2. Select the year using the
Press the
button to set the current year and move to month/day selection. The date can be set anywhere between the years 2011 and 2055.
3. Select the month using the button.
Press the
button to set the current month and move to day selection.
4. Select the day using the Press the
button to set the current day and move to hour/minute selection.
5. Select the hour using the Press the
button to set the current hour and move to minute selection.
6. Select the minute using the button.
Press the
button to set the current
minute and activate the clock.
button until
button.
button.
button.
ress an
hold
Year
Month
Day
Hou
Minute
12-hour
cloc
24-hour
cloc
9
Using the Monitor
A
prop
Selecting the Correct Cuff Size
Using the correct cuff size is important for an accurate reading. If the cuff is not
the proper size, the reading may yield an incorrect blood pressure value. The arm size is printed on each cuff. The index and proper fit range, on the cuff, tell you if you are applying the
correct cuff. (Refer to "Symbols that are printed on the cuff" on the next page) If the index points outside of the range, contact your local dealer to purchase a
replacement cuff. The arm cuff is a consumable. If it becomes worn, purchase a new one.
Arm Size Recommended Cuff Size Catalog Number 31 cm to 45 cm Large adult cuff UA-CUFFBKLA-EC 22 cm to 32 cm Adult Cuff 16 cm to 24 cm Small Adult Cuff
Arm size: The circumference of the biceps.
Applying the Arm Cuff
1. Wrap the cuff around the upper arm, about 1-2 cm above the inside of the elbow, as shown.
Place the cuff directly against the
skin, as clothing may cause a faint pulse, and result in a measurement error.
2. Constriction of the upper arm,
caused by tightly rolling up a shirtsleeve, may prevent accurate readings.
3. Confirm that the index
within the proper fit range.
points
UA-CUFFBKAU-EC UA-CUFFBKSA-EC
Do not roll up shirtsleeve tightly.
Cuff
1-2 cm
ir Hose
Artery position mark
Secure the fabric fastener firmly.
Index
and
er fit range
10
Using the Monitor
Symbols that are printed on the cuff
Symbols Function/Meaning Recommended Action
Set the
Artery Position Mark
of the upper arm or in line with the ring finger on the inside of the arm.
mark on the artery
REF
Index Catalog number
Proper fit range for Adult Cuff It's printed on the Adult Cuff.
Range to use Large Adult Cuff It's printed on the Adult Cuff.
Proper fit range for Large Adult Cuff It's printed on the Large Adult Cuff.
Range to use Small Adult Cuff It's printed on the Adult Cuff.
Proper fit range for Small Adult Cuff It's printed on the Small Adult Cuff. Range to use Adult Cuff It's printed on the Large Adult Cuff and Small Adult Cuff.
Lot number
Use Large Adult Cuff instead of Adult Cuff.
Use Small Adult Cuff instead of Adult Cuff.
Use Adult Cuff instead of Large Adult Cuff or Small Adult Cuff.
How to Take Proper Measurements
For the most accurate blood pressure measurement:
Sit comfortably at a table. Rest your arm on the table. Relax for about five to ten minutes before measurement. Place the center of the cuff at the same height as your heart. Remain still and keep quiet during measurement. Do not measure right after physical exercise or a bath. Rest for twenty or
thirty minutes before taking the measurement.
Try to measure your blood pressure at the same time every day.
Measurement
During measurement, it is normal for the cuff to feel very tight. (Do not be alarmed).
After Measurement
After measurement, press the START button to turn the device off. After one minute of non-operation, the device will turn off automatically. Remove the cuff and record your data.
Note: Allow at least three minutes between measurements on the same person.
11
Measurements
a
y
Before measurement, read “Notes for Proper Measurement” on the next page.
Normal Measurement
1. Place the cuff on the arm (preferably the left arm). Sit quietly during measurement.
2. Press the START button. The number of data stored in memory is
displayed.
Zero (0) is displayed blinking briefly.
The display changes, as indicated in the figure at the right, as the measurement begins. The cuff starts to inflate. It is normal for the cuff to feel very tight. A pressure bar indicator is displayed, on the left edge of the display, during the inflation.
Note: If you wish to stop inflation at any time,
press the
START button again.
3. When inflation is complete, deflation starts automatically and
(heart mark) blinks, indicating that the measurement is in progress. Once the pulse is detected, the mark blinks with each pulse beat.
Note:
If an appropriate pressure is not obtained, the device starts to inflate again automatically. To avoid re-inflation, see “Measurement with the Desired Systolic Pressure” on the next page.
4. When the measurement is complete, the systolic and diastolic pressure readings and pulse rate are displayed. The cuff exhausts the remaining air and deflates completely.
5. Press the START button to turn the device off. After one minute of non-operation, the device will turn off automatically.
Note: Allow at least three minutes between
measurements on the same person.
Systolic pressure Diastolic pressure Pulse rate
At heart level
Press
Number of dat
in memor
displayed
Zero display
Starts inflation
Pressurizing
Measurement
in progress
Exhausts remaining
air automatically
STA RT
12
Measurements
Measurement with the Desired Systolic Pressure
Model UA-767NFC is designed to detect the pulse and to inflate the cuff
to a systolic pressure level automatically. Use this method when re-inflation occurs repeatedly or when the results are not displayed even if the pressure decreases to 20 mmHg or less.
1. Place the cuff on the arm (preferably the left arm).
2. Press and hold the START button until a number about 30 to 40 mmHg higher than your expected systolic pressure appears.
Release the button
3. When the desired number is reached, release the
START button to start
systolic pressure
measurement. Continue to measure your blood pressure as described on the previous page.
At heart level
Press and hold
START
the button
at the desired
Refer to the previous page
for measurement
Notes for Proper Measurement
Sit down in a comfortable position. Place your arm on a table with your
palm facing upward and the cuff at the same level as your heart.
Relax for about five to ten minutes before taking a measurement. If you
are excited or depressed by emotional stress, the measurement will reflect this stress as a higher (or lower) than normal blood pressure reading and the pulse reading will usually be faster than normal.
An individual's blood pressure varies constantly, depending on what you
are doing and what you have eaten. What you drink can have a very strong and rapid effect on your blood pressure.
This device bases its measurements on the heartbeat. If you have a very
weak or irregular heartbeat, the device may have difficulty determining your blood pressure.
Should the device detect a condition that is abnormal, it will stop the
measurement and display an error symbol. Refer to page 5 for the description of symbols.
This blood pressure monitor is intended for use by adults only. Consult
with your physician before using this device on a child. A child should not use this device unattended.
13
Near Field Communication (NFC) Function
Initiat
Dat
y
Dat
Initiat
Dat
Initiat
To transmit data to or from the blood pressure monitor, communication
devices that comply with the communication specifications of the blood pressure monitor are required.
Examples of communication devices that comply Personal computer with NFC reader/writer
Access Point
Memory Data Transmission
While the blood pressure monitor displays the clock, hold the
If the clock of the blood pressure monitor is not correct, the clock will be
NFC-enabled cellular phone
communication symbol of the NFC-enabled communication device close to the communication area (
) of the monitor until “end” appears.
The memory data transmission is performed. updated at the same time.
e
communication
Number of data in memor
a
transmission
Clock updated
e
communication
a
transmission
Clock display Transmission starts Transmission ends Clock display
Measurement Data Transmission
When the measurement is complete and the monitor displays the results,
hold the communication symbol of the NFC-enabled communication device close to the communication area ( The data transmission is performed.
If the clock of the blood pressure monitor is not correct, the clock will be
updated at the same time.
e
communication
transmission
) of the monitor until “end” appears.
a
Measurement data display
14
Transmission starts Transmission ends Clock display
Caution on Using the NFC Function
The UA-767NFC blood pressure monitor h as an NFC Forum Type 3 Tag
wireless interface module built in. When the monitor is used near wireless communication devices which use the same frequency as that of the monitor (13.56 MHz) or use its harmonic frequencies, electromagnetic interference may occur. In that case, turn off the devices that are not in use, or place the monitor 1 meter or more away from the devices.
What is an Irregular Heartbeat
The UA-767NFC blood pressure monitor provides a blood pressure and pulse
rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important that you are relaxed, remain still and do not talk during measurements.
Note: We recommend contacting your physician if you see this
indicator frequently.
Pressure Bar Indicator
The indicator monitors the progress of pressure during measurement.
Pressurizing
Inflation in progress Measurement in progress
Releasing air
Inflation complete
About Blood Pressure
What is Blood Pressure?
Blood pressure is the force exerted by blood against the walls of the arteries.
Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg). One's natural blood pressure is represented by the fundamental pressure, which is measured first thing in the morning while one is still at rest and before eating.
15
About Blood Pressure
What is Hypertension and How is it Controlled?
Hypertension, an abnormally high arterial blood pressure, if left unattended,
can cause many health problems including stroke and heart attack. Hypertension can be controlled by altering lifestyle, avoiding stress, and with medication under a doctor’s supervision.
To prevent hypertension or keep it under control:
Why Measure Blood Pressure at Home?
Blood pressure measured at a clinic or doctor's office may cause
Do not smoke Exercise regularly Reduce salt and fat intake Have regular physical checkups Maintain proper weight
apprehension and can produce an elevated reading, 25 to 30 mmHg higher than that measured at home. Home measurement reduces the effects of outside influences on blood pressure readings, supplements the doctor's readings and provides a more accurate, complete blood pressure history.
WHO Blood Pressure Classification
Standards to assess high
blood pressure, without regard to age, have been established by the World Health Organization (WHO), as shown in the chart.
Blood Pressure Variations
An individual’s blood pressure
varies greatly on a daily and seasonal basis. It may vary by 30 to 50 mmHg due to various conditions during the day. In hypertensive individuals, variations are even more pronounced. Normally, the blood pressure rises while at work or play and falls to its lowest levels during sleep. So, do not be overly concerned by the results of one measurement.
Take measurements at the same
time every day using the procedure described in this manual to get to know your norm al blood pressure. Regular readings give a more comprehensive blood pressure history. Be sure to note the date and time when recording your blood pressure. Consult your doctor to interpret your blood pressure data.
16
Troubleshooting
Problem Possible Reason Recommended Action
Nothing appears on the display, even when the power is turned on.
The cuff does not inflate.
The device does not measure. Readings are too high or too low.
Other
Batteries are drained.
Battery terminals are not in the correct position.
Battery voltage is too low.
(LOW BATTERY mark) blinks. If the batteries are drained completely, the mark does not appear. The cuff is not applied properly. You moved your arm or body during measurement.
The cuff position is not correct.
The value is different from that measured at a clinic or doctor’s office.
Replace all batteries with new ones. Reinstall the batteries with negative and positive terminals matching those indicated on the battery compartment.
Replace all batteries with new ones.
Apply the cuff correctly. Make sure you remain very still
and quiet during measurement. Sit comfortably and still. Place your arm on a table with your palm facing upward and the cuff at the same level as your heart. If you have a very weak or irregular heat beat, the device may have difficulty in determining your blood pressure.
Refer to “Why Measure Blood Pressure at Home?”.
Remove the batteries. Place them back properly and take another measurement.
Note: If the actions described above do not solve the problem, contact the
dealer. Do not attempt to open or repair this product, as any attempt to do so will make your warranty invalid.
17
Maintenance
Do not open the device. It uses delicate electrical components and an
intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, contact the authorized dealer in your area or our customer service department. The A&D customer service will provide technical information, spare parts and units to authorized dealers.
The device was designed and manufactured for a long service life.
However it is generally recommended to have the device inspected every 2 years, to ensure proper functioning and accuracy. Please contact the authorized dealer in your area or A&D for maintenance.
Technical Data
Type UA-767NFC Measurement method Oscillometric measurement Measurement range Pressure: 0 - 280 mmHg Pulse: 40 - 180 beats / minute Measurement accuracy Pressure: ±3 mmHg Pulse: ±5% Power supply 4 x 1.5V batteries (R6P, LR6 or AA) or
AC adapter (TB-233) (Not included) Classification Type BF Clinical test According to ANSI / AAMI SP-10 1992 EMC IEC 60601-1-2: 2007 Memory Last 100 measurements Operating condition +10°C to +40°C / 30%RH to 85%RH Storage condition -10°C to +60°C / 30%RH to 95%RH Dimensions Approx. 147 [W] x 64 [H] x 110 [D] mm Weight Approx. 300 g, excluding the batteries
Wireless communication NFC wireless interface module Accessory AC adapter The adapter is to connect the blood pressure
monitor to a power source at home.
TB-233 Please contact your local A&D dealer for
purchasing. Note: Specifications are subject to change without prior notice. EMC table information is listed on our website:
http://www.aandd.jp/products/manual/medical/emc_acadapter_en.pdf
18
A&D Company, Limited
1-243 Asahi, Kitamoto-shi, Saitama 364-8585 JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119
REPEC
A&D INSTRUMENTS LTD.
Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485
A&D Australasia Pty Ltd.
32 Dew Street, Thebarton, South Australia 5031 AUSTRALIA Telephone: [61] (8) 8301-8100 Fax: [61] (8) 8352-7409