A&D UA-767 User Manual

Page 1
Digital Blood Pressure Monitor
Model UA-767 30
Instruction Manual
Manuel d
instructions
Manuale di Istruzioni
Bedienungsanleitung
WM+PD4001183
使用手册
Page 2
33
Contents
Dear Customers.................................................................... 2
Preliminary Remarks ............................................................. 2
Precautions........................................................................... 2
Parts Identification................................................................ 3
Symbols................................................................................ 4
Operation Mode.................................................................... 5
1. Normal Measurement With Storing Data ..................... 5
2. Recalling Data.............................................................. 5
3. Clearing Data............................................................... 5
4. Measurement With The Desired Systolic Pressure ......... 5
Using the Monitor ................................................................. 6
Installing / Changing The Batteries .................................. 6
Connecting The Air Hose ................................................. 6
Selecting The Correct Cuff ............................................... 6
Attaching The Arm Cuff ................................................... 7
How To Take Proper Measurements ................................... 7
Measurement................................................................... 7
After Measurement .......................................................... 7
Measurements ...................................................................... 8
Normal Measurement....................................................... 8
Measurement With The Desired Systolic Pressure ............. 9
Notes For Proper Measurement ........................................ 9
What Is An Irregular Heartbeat ............................................ 10
Pressure Bar Indicator ......................................................... 10
WHO Classification Indicator ............................................... 10
About Blood Pressure.......................................................... 11
What Is Blood Pressure? ................................................. 11
What Is Hypertension And How Is It Controlled? .............11
Why Measure Blood Pressure At Home? .......................... 11
WHO Blood Pressure Classification ................................. 11
Blood Pressure Variations............................................... 11
Troubleshooting ................................................................. 12
Maintenance ....................................................................... 13
Technical Data.................................................................... 13
English 1
Page 3
Dear Customers
Congratulations on purchasing a state-of-the-art A&D blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen.
We recommend that you read through this manual carefully before using the device for the first time.
Preliminary Remarks
This device conforms to the European Directive 93/42 EEC for Medical
Products. This is made evident by the
(0366: The reference number of the designated authority).
The device is designed for use on adults only, not newborns or infants. Environment for use
The device is for use indoors.
mark of conformity.
Precautions
Precision components are used in the construction of this device.
Extremes in temperature, humidity, direct sunlight, shock or dust should be avoided.
Clean the device with a soft, dry cloth. Never use thinner, alcohol,
benzine, or wet cloth.
Avoid tightly folding the cuff or storing the hose tightly twisted for long
periods, as such treatment may shorten the life of the components.
The device and cuff are not water resistant. Prevent rain, sweat and
water from soiling the device and cuff.
Measurements may be distorted if the device is used close to televisions,
microwave ovens, cellular telephones, X-ray or other devices with strong electrical fields.
Used equipment, parts and batteries are not treated as ordinary
household waste, and must be disposed of according to the applicable local regulations.
English 2
Page 4
Parts Identification
Air Connector Plug
Part of Display
Arm Cuff
Battery Cover
WHO Classification Indicator
Display
Air Socket
Battery Compartment
MEMORY
Pressure Bar Indicator
and
Battery Indicator
I.H.B. Indicator (Irregular heartbeat indicator)
DC Jack
START button (Orange illumination)
Air Hose
Proper Fit Range
Index Mark
1.5V Batteries
(R6P, LR6 or AA)
Systolic Pressure
Diastolic Pressure
Pulse Rate
Heart Mark
English 3
Page 5
Symbols
Symbols Function / Meaning Recommended Action
SN
Standby and Turn the device on. Battery installation guide Direct current Serial number Date of manufacture Type BF: Device, cuff and tubing are
designed to provide special protection against electrical shocks. The indicator while measurement is in progress. It blinks while detecting the pulse. Irregular Heartbeat indicator. (I.H.B.) The indicator that detects irregular heartbeats or any excessive body movement during the measurement. Previous measurements stored in MEMORY. The battery power indicator during
Full Battery
Low Battery
measurement. The battery is low when it blinks.
Unstable blood pressure due to movement during the measurement.
The systolic and diastolic values are within 10 mmHg of each other. The pressure value did not increase during inflation.
PUL. DISPLAY ERROR
SYS
DIA
PUL./min
The cuff is not fastened correctly. The pulse is not detected correctly.
Systolic blood pressure in mmHg Diastolic blood pressure in mmHg. Pulse per minute
Measurement is in progress. Remain as still as possible.
Replace all batteries with new ones, when the indicator blinks. Try measurement again. Remain very still during the measurement.
Fasten the cuff correctly, and try the measurement again.
EC directive medical device label
English 4
WEEE label
Page 6
Operation Mode
1. Normal Measurement With Storing Data ( Refer to page 8)
Press the START button. Blood pressure is measured and data is stored
within the device. The device can store the last thirty sets of data in memory automatically.
2. Recalling Data
Press and hold the START button to recall the
data. Stored data are automatically displayed in
Press and hold
Data number
order from the last data.
The device can store the last thirty sets of data
in memory automatically.
Alternation
Stored Data
3. Clearing Data
Remove batteries to clear all data. Re-install them again.
4. Measurement With The Desired Systolic Pressure
Refer to page 9 for t he measurement with the desired systolic pressure.
English 5
Page 7
Using the Monitor
A
A
A
Installing / Changing The Batteries
1. Slide the battery cover up to open it.
2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+) and (-) are correct.
3. Slide the battery cover down to close it.
Use only R6P, LR6 or AA batteries.
CAUTION
Insert the batteries as shown in the battery compartment. If not, the device will
not work.
When
(LOW BATTERY mark)
blinks in the display, replace all batteries with new ones. Do not mix old and new batteries. It may shorten the battery life, or cause the device to malfunction.
(LOW BATTERY mark)
does not appear when the batteries are drained.
Battery life varies with the ambient temperature and may be shorter at low
temperatures.
Use the specified batteries only. The batteries provided with the device are for
testing monitor performance and may have a limited life.
Remove the batteries if the device is not to be used for a long time.
The batteries may leak and cause a malfunction.
Stored data are cleared when the batteries are removed.
Step 3
Connecting The Air Hose
Insert the air connector plug into
the air socket firmly.
ir Connector Plug
Selecting The Correct Cuff
Using the correct cuff size is important for an accurate reading. If the cuff is not
the proper size, the reading may yield an incorrect blood pressure value.
The index and proper fit range, on the cuff, tell you if you are attaching the
correct cuff. (Refer to "Attaching The Arm Cuff" on next page)
If the index points outside of the range, contact your local dealer for replacement.
Arm Size Recommended Cuff Size Order No. 32 cm to 42 cm Large cuff UA-CUFDBKLA-EC 22 cm to 32 cm Medium cuff UA-CUFDBKAU-EC 18 c m t o 2 2 c m Sm a l l c u f f UA-CUFDBKCH-EC
Arm size: The circumference at the biceps.
Step 1
Step 2
ir Hose
ir Socket
English 6
Page 8
A
x
Using the Monitor
Attaching The Arm Cuff
1. Wrap the cuff around the upper arm, about 2 - 3 cm above the elbow, as shown. Place the cuff directly against the skin, as clothing may cause a faint pulse and result in a measurement error.
2. Constriction of the upper arm, caused by rolling up a shirtsleeve, may prevent accurate readings.
3. Confirm that the index points within the proper fit range.
How To Take Proper Measurement s
For the most accurate blood pressure
measurement:
Sit comfortably around the table.
Rest your arm on the table.
Relax for about five to ten minutes
before measurement.
Place the center of the cuff at the same height as your heart. Remain still and keep quiet during measurement. Do not measure right after physical exercise or a bath. Rest for twenty or
thirty minutes before taking the measurement.
Try to measure your blood pressure at the same time every day.
Rolled up shirtsleeve
Cuff
ir Hose
2 -3 cm
Constriction of the Upper Arm
Inde
and Proper fit range
Measurement
During measurement, it is normal for the cuff to feel very tight. (Do not be
alarmed)
After Measurement
After measurement, press the START button to turn off the power.
Remove the cuff and record your data.
Note: The device has an automatic power shut-off function, which turns the
power off approximately one minute after measurement.
Allow at least ten minutes between measurements on the same person.
English 7
Page 9
A
Measurements
Model UA-767 30 is designed to detect the pulse and to inflate the
cuff to a systolic pressure level automatically.
If your systolic pressure is expected to exceed 230 mmHg or you use the
optional small cuff, read "Measurement With The Desired Systolic Pressure" on next page.
Normal Measurement
1. Place the cuff on the arm (preferably the left arm).
Sit quietly during measurement.
2. Press the START button.
The average blood pressures and last data
number are displayed briefly. Then the display changes, as indicated in the figure at the right, as the measurement begins. The cuff starts to inflate. It is normal for the cuff to feel very tight. A pressure bar indicator is displayed, as in the figure at the right, during inflation.
Note: If you wish to stop inflation at any time,
press the
START button again.
3. When inflation is complete, deflation starts
automatically and the
(heart mark) blinks, indicating that the measurement is in progress. Once the pulse is detected, the mark flashes with each pulse beat.
Note: If an appropriate pressure is not
obtained, the device starts to inflate again automatically.
4. When the measurement is complete, the systolic and diastolic pressure readings and pulse rate are displayed. The cuff exhausts the remaining air and deflates completely.
5. Press the START button again to turn off the power.
Note: Model UA-767 30 is provided with an automatic power
shut-off function. Allow at least ten minutes between measurements on the same person.
At heart level
START button
veraged data
Data number
Starts inflation
Pressurizing
Measurement
in progress
Systolic pressure Diastolic pressure WHO classification Pulse rate
Exhausts remaining
air automatically
English 8
Page 10
Measurements
Measurement With The Desired Systolic Pressure
If your systolic pressure is expected to exceed 230 mmHg or you use the
optional small cuff, use this procedure.
1. Place the cuff on the arm (preferably the left arm).
2. Press and hold the START button until a number about 30 to 40 mmHg higher than your expected systolic
Press and hold the button
pressure appears.
3. Release the START button to start measurement, when the desired number is reached. Then continue to
Refer to the previous page
measure your blood pressure as described on the previous page.
At heart level
Release the button
at the desired
systolic pressure
for measurement
Notes For Proper Measurement
Sit down in a comfortable position. Place the arm to be used for the
measurement on a table or other support so that the center of the cuff will be at the same height as your heart.
Relax for about five or ten minutes before taking a measurement. If you
are excited or depressed by emotional stress, the measurement will reflect this stress as a higher (or lower) than normal blood pressure reading and the pulse reading will usually be faster than normal.
An individual's blood pressure varies constantly, depending on what you
are doing and what you have eaten. What you drink can have a very strong and rapid effect on your blood pressure.
This device bases its measurements on the heartbeat. If you have a very
weak or irregular heartbeat, the device may have difficulty determining your blood pressure.
Should the device detect a condition that is abnormal, it will stop the
measurement and display an error symbol. See page 4 for the description of symbols.
This blood pressure monitor is intended for use by adults only. Consult
with your physician before using this device on a child. A child should not use this device unattended.
English 9
Page 11
What Is An Irregular Heartbeat
Model UA-767 30 blood pressure monitor provides a blood
pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important that you be relaxed, remain still and do not talk during measurements.
Note: We recommend contacting your physician if you see this
indicator frequently.
Pressure Bar Indicator
The indicator monitors the progress of pressure during measurement.
Inflation in progress Measurement in progress
Pressurizing
Releasing air
Inflation complete
WHO Classification Indicator
Each six segments of the bar indicator
correspond to the WHO blood pressure classification described on the next page.
:
English 10
WHO Classification Indicator
Severe hypertension Moderate hypertension Mild hypertension High normal Normal Optimal
The indicator displays a segment, based on the current data, corresponding to the WHO classification.
Example:
Moderate hypertension
Mild hypertension
High normal
Page 12
About Blood Pressure
What Is Blood Pressure?
Blood pressure is the force exerted by blood against the walls of the arteries.
Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg). One's natural blood pressure is represented by the fundamental pressure, which is measured first thing in the morning while one is still at rest and before eating.
What Is Hypertension And How Is It Controlled?
Hypertension, an abnormally high arterial blood pressure, if left unattended,
can cause many health problems including stroke and heart attack. Hypertension can be controlled by altering lifestyle, avoiding stress, and with medication under a doctor’s supervision.
To prevent hypertension or keep it under control:
Do not smoke Exercise regularly Reduce salt and fat intake Have regular physical checkups Maintain proper weight
Why Measure Blood Pressure At Home?
Blood pressure measured at a clinic or doctor's office may cause
apprehension and can produce an elevated reading, 25 to 30 mmHg higher than that measured at home. Home measurement reduces the effects of outside influences on blood pressure readings, supplements the doctor's readings and provides a more accurate, complete blood pressure history.
WHO Blood Pressure Classification
Standards to assess high
blood pressure, without regard to age, have been established by the World Health Organization (WHO), as shown in the chart .
Blood Pressure Variations
An individual’s blood pressure
varies greatly on a daily and seasonal basis. It may vary by 30 to 50 mmHg due to various conditions during the day. In hypertensive individuals’ variations are even more pronounced. Normally, the blood pressure rises while at work or play and falls to its lowest levels during sleep. So, do not be overly concerned by the results of one measurement.
Take measurements at the same time every day using the procedure
English 11
Page 13
described in this manual to get to know your normal blood pressure. Regular readings give a more comprehensive blood pressure history. Be sure to note date and time when recording your blood pressure. Consult your doctor to interpret your blood pressure data.
Troubleshooting
Problem Possible Reason Recommended Action
Nothing appears in the display, even when the power is turned on.
The cuff does not inflate.
The unit does not measure. Readings are too high or too low.
Other
Batteries are drained.
Battery terminals are not in the correct position.
Battery voltage is too low.
(LOW BATTERY
mark)
blinks. [ If the batt eri es are drained completely, the mark does not appear.] The cuff is not fastened properly. You moved your arm or body during the measurement.
The cuff position is not correct.
The value is different from that measured at a clinic or doctor’s office.
Replace all batteries with new ones. Reinstall the batteries with negative and positive terminals matching those indicated on the battery compartment.
Replace all batteries with new ones.
Fasten the cuff correctly. Make sure you remain very still
and quiet during the measurement. Sit comfortably and still. Raise your hand so that the cuff is at the same level as your heart. If you have a very weak or irregular heat beat, the device may have difficulty in determining your blood pressure.
See “Why measure blood pressure at home”.
Remove the batteries. Place them back properly and try the measurement again.
Note: If the actions described above do not solve the problem, contact the
dealer. Do not attempt to open or repair this product, as any attempt to do so will make your warranty invalid.
English 12
Page 14
Maintenance
Do not open the device. It uses delicate electrical components and an
intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, request service from your dealer or from the A&D service group. The A&D service group will provide technical information, spare parts and units to authorized dealers.
The device was designed and manufactured for a long service life.
However it is generally recommended to have the monitor inspected every 2 years, to ensure proper functioning and accuracy. Please contact either your authorized dealer or A&D for maintenance.
Technical Data
Type UA-767 30 Measurement method Oscillometric measurement Measurement range Pressure: 20 - 280 mmHg Pulse: 40 - 200 beats / minute Measurement accuracy Pressure: ±3 mmHg or 2%, whichever is greater Pulse: ±5% Power supply 4 x 1.5V batteries (R6P, LR6 or AA) Upper arm circumference 22 - 32 cm Classification Type BF Clinical test According to ANSI / AAMI SP-10 1987 EMC IEC 60601-1-2: 2001 Memory Last 30 measurements Operating condition +10°C to +40°C / 30%RH to 85 %RH Storage condition -10°C to +60°C / 30%RH to 85 %RH Dimensions Approx. 147 [W] x 64 [H] x 110 [D] mm Weight Approx. 300 g, excepting batteries
Note: Specifications are subject to change without prior notice.
English 13