A&D TM-2430-13 User Manual

TM-2430-13
(Doctor Pro)
Ambulatory Blood Pressure Monitor
Data Analysis Software
INSTRUCTION MANUAL
WM:PD4000019C
CONTENTS
1-1 About the Data Analysis Software...................................................................................................3
1-2 Symbols...........................................................................................................................................4
1-3 Monitoring Device Review...............................................................................................................4
1-4 Software Package Components......................................................................................................5
1-5 System Requirements .....................................................................................................................5
1-6 Installing the Software.....................................................................................................................5
1-7 Safety...............................................................................................................................................5
2. OPERATING PROCEDURE .......................................................................................6
2-1 Getting Started ................................................................................................................................6
2-2 Communication Between the Monitor and the Software.................................................................6
2-2-1 Connecting the TM-2420/TM-2421 to the computer....................................................................6
2-2-2 Connecting the TM-2430 series to the computer.........................................................................7
2-2-3 Connecting the UA-767PC to the computer.................................................................................8
2-3 Main Screen Description.................................................................................................................9
2-4 Toolbar ..........................................................................................................................................10
2-4-1 Analysis Intervals........................................................................................................................10
2-4-2 BP Mode.....................................................................................................................................11
2-5 File Menu.......................................................................................................................................12
2-5-1 Open...........................................................................................................................................13
2-5-2 Save As ......................................................................................................................................13
2-5-3 Delete .........................................................................................................................................13
2-5-4 Convert.......................................................................................................................................14
2-5-5 File Merge...................................................................................................................................14
2-5-6 Print ............................................................................................................................................15
2-5-7 Exit..............................................................................................................................................15
2-5-8 Saving a File...............................................................................................................................15
2-5-9 Closing a Data File .....................................................................................................................15
2-6 View Menu.....................................................................................................................................16
2-6-1 Summary Data............................................................................................................................17
2-6-2 BP Data List................................................................................................................................19
2-6-3 Trends Screen............................................................................................................................21
2-6-4 Correlation Plots.........................................................................................................................25
2-6-5 Histograms .................................................................................................................................27
2-6-6 Circadian Rhythm.......................................................................................................................28
2-6-7 HIB Analysis ...............................................................................................................................29
2-6-8 Patient Information .....................................................................................................................30
1
2-7 Recorder Menu.............................................................................................................................31
2-7-1 Retrieving Data from the Monitor............................................................................................... 32
2-7-2 Clearing The Monitor’s Memory ................................................................................................33
2-7-3 Programming the Monitor’s Intervals and Conditions................................................................ 34
2-7-4 Blood Pressure Measurement (Available only with the UA-767PC) ......................................... 40
2-8 Report Menu.................................................................................................................................41
2-8-1 Creating a New Report Template.............................................................................................. 42
2-8-2 Editing an Existing Report Template ......................................................................................... 43
2-8-3 Deleting an Existing Report Template....................................................................................... 43
2-8-4 Saving a Default Report Template for Printing.......................................................................... 43
2-8-5 Printing an Existing Report Template........................................................................................ 43
2-8-6 Closing the Report dialog box ................................................................................................... 43
2-9 Options Menu ...............................................................................................................................44
2-9-1 Report formats...........................................................................................................................44
2-9-2 Physician List............................................................................................................................. 44
2-9-3 Trend Display............................................................................................................................. 45
2-9-4 Default Settings ......................................................................................................................... 46
2-10 On-line Help................................................................................................................................ 47
3. TROUBLESHOOTING ............................................................................................. 48
3-1 Measurement Errors..................................................................................................................... 48
3-1-1 TM-2420/TM-2421 Error Codes ................................................................................................ 48
3-1-2 TM-2430 Series Error Codes..................................................................................................... 50
3-2 Communication Errors..................................................................................................................51
3-3 Missing Files.................................................................................................................................52
3-4 Reports Are Not Printing With the Desired BP Mode................................................................... 52
3-5 Summary Data / Sleep Tab Data can not be Seen ......................................................................52
4. PRINTOUT EXAMPLES........................................................................................... 53
5. TERMS/ABBREVIATIONS....................................................................................... 62
5-1 Terms Related to Blood Pressure ................................................................................................ 62
5-2 Terms Related to Statistics........................................................................................................... 62
5-3 Terms Related to Biological Rhythm............................................................................................ 63
5-4 Terms Related to the Risk Management of High Blood Pressure................................................. 65
2
1. INTRODUCTION
1-1 About the Data Analysis Software
Welcome to the Ambulatory Blood Pressur e Monitor (ABPM) Data Analy sis Soft war e. This software is intended for use with the following devices.
ABPM: For diagnosing high blood pressure
1) A&D TM-2430 Series Ambulatory Blood Pressure Monitor
2) A&D TM-2420/TM-2421 Ambulatory Blood Pressure Monitor and the TM-20 20/TM- 2021 Pri nter/Pro cessor/I nterfac e (herea fter re ferred to as "proc essor").
Home use monitor: For manag ing the blood pressure of a high blood pressure patient
3) A&D UA-767PC Dig it al Blood Pressure Monitor
The ABPM Data Analysis Software, that analyzes ambulatory blood pressure data, is a powerful tool for diagnosing high blood pressure and managing the blood pressure of a high blood pressure patient. The following features are incorporated in this software:
Statistical Analysis
Graphical Data
Data Conversion
Printed Reports
On-line Help
Maximum Time Period
Maximum number of readings
Statistical data may be viewed in full, partial, sleep, and awake intervals, respectively in full screen, by switching between clearly labeled tabs.
Systolic/diastolic blood pressure, mean arterial blood pressure, and pulse data are displayed graphically to quickly determine patterns or trends in the data.
Circadian rhythm and HBI analysis are displayed graphica lly and numerically.
The Data Convert feature automatically stores blood pressure and pulse data in standard (CSV) file format for use with popular spreadsheet programs.
Custom data report formats are easily defined and printed. “Mini-report” feature automatically prints a compact summary report.
Built-in Help feature provides context-sensitive help at any time.
Doctor Pro can analyze up to one week’s data starting from the oldest reading.
When used with the UA-767PC, Doctor Pro can analyze up to five years’ dat a
The maximum number of readings that Doctor Pro can process is 900.
When used with the UA-767PC, Doctor Pro can process 1827 readings.
3
1-2 Symbols
The symbols printed on the device have the following meaning:
CE marking
SN
Serial number
1-3 Monitoring Device Review
The following section briefly reviews the devices and how t hey may be used with this software.
A&D Ambulatory Blood Pressure Monitors are used to record an individual’s blood pressure over an extended period of time, normally 24-72 hours. The monitor is worn by the patient. The monitor non-invasively records the patient’s systolic/diastolic blood pressure and heart rate based on predetermined intervals and conditions.
With a UA-767PC, the blood pressure data on individuals for a long period of time can be managed and blood pressure trends are indicated clearly by using this software.
The user may divide each 24-hour period into several time intervals. Within each period the monitor may be programmed to record the patient’s blood pressure at 1 minute intervals. The monitor can also be programmed to remain OFF during any time interval.
1
, 31, 5, 10, 15, 201, 30, 60, and 1201-
These measurement interv als may be timed to coincide with the patie nt ’s d ai ly activities.
The A&D TM-2020/TM-2021 Printer/Processor/Interface is used for performing one or more of the following functions: Analyze blood pressure data stored in the monitor; program the monitor’s intervals and conditions; fun ct ion as an interface between the monit or and the software.
Note:Data that was initially analyzed using A&D TM-2020-01/02/03 or TM-2021-
01/02/03 Data Analysis Software Program for DOS™ is compatible with this software’s format and can be ana lyzed by this software.
1 The 1 and 3-minute measurement intervals are available only with the TM-2420/TM- 2421.
The 20-minute measurement interval is available only with the TM-2421. The 120-minute measurement i nt erval is available only w it h t he TM-2421/ TM-2430.
4
1-4 Software Package Components
Items included in your complete software package are:
ABPM 3.5-inch program disk Ambulatory Blood Pressure Monitor Data Analysis Software Instruction Manua l For TM-2420/TM-2421
AX-KO600 (9 pin D-SUB, socket type), Communication cable for TM-2420/TM-2421
For TM-2430 series/UA-767PC
AX-KO1502 (9 pin D-SUB, socket type), Communication cable for TM-2430 series
1-5 System Requirements
Computer RAM Memory Operating System Graphic Adapter
Disk Type Hard Disk Serial Ports
Printer Type
A 486DX or higher microproces sor Minimu m 16 MB Windows™ 95, 98, Me, 2000, NT 4. 0 SVGA video card (minimum of 256 colors, 800X600 pixels or higher
recommended)
3.5-inch floppy disk dr ive At least 2.5 MB Minimum one (9 pin D-SUB or 25 pin D-SUB)
(A cable separately available is required for the 25 pin D-SUB serial port)
Printers supported by MS Window s™ operating system
1-6 Installing the Software
The software must be run from a hard disk drive only. It may not be run from a floppy disk. To install the software on your hard disk drive, complete the following instructions:
Step 1 Insert the program disk in drive A:. Step 2 Click the Start button on the Windows taskbar and select Run. Step 3 Type “a:setup” Step 4 Click OK. Step 5 Follow the instructions on the scre en.
Note The serial number is required when installing the software. The serial number is
printed on the label of the progra m d is k.
1-7 Safety
Do not attach the monitor to a patient when the monitor, or the monitor and the processor, are being connected to a computer.
5
2. OPERATING PROCEDURE
2-1 Getting Started
Two initial action s may be executed when starting up th e software:
(1) Data may be retrieved from the monitor. (2) Stored data files may be opened and viewed for analysis.
To retrieve data from the monitor, see:
2-2 Communication Between the Mo ni tor and the Software”
To view stored files, see:
“2-5-1 Open
2-2 Communication Between the Monitor and the Software
To communicate between the monitor and the software, the connection between the monitor and the computer must be established first.
2-2-1 Connecting the TM-2420/ TM-2421 to the computer
You must use the A&D-supplied RS-232C communication cable, otherwise
communication errors will occur.
Steps 1-4 must be completed prior to accessing the software or it may not recognize that
the monitor is connected to the computer. If proper communication is not established, a cable check error may occur.
You must use AX-KO600 cable for 9 pin serial port (supplied) or AX-KO599 cable for 25
pin serial port (available se parat ely as an option).
Figure 1: Connection between the TM-2420/2421 and the computer
(The illustration above is not to scale.)
6
Step 1 Connect the monitor to the proc essor. Step 2 Connect the male end of the A&D-supplied RS-232C communication cable to the
processor.
Step 3 Connect the female end of the A&D-supplied RS-232C communication cable to the
computer.
Step 4 The display on the processor will show “0-----”. If the processor does not have a
display, the LED should be green. If the LED is not green or the “0-----” does not appear, see “3-2 Communication Errors”. If the processor shows “0-----”, go to step
5. If the processor has a green light, go to step 6.
Note:If possible, keep the cable connected to the computer unless the port is needed
for other functions.
Step 5 Press the ENTER key on the processor keypad. A series of 6 dashes “------” will
appear on the processor screen.
Step 6 Run the software. Step 7 Confirm that the correct serial p ort is set . For det ails, see “2-9-4-2 Serial Port”.
2-2-2 Connecting the TM-2430 series to the computer
You must use the A&D-supplied RS-232C communication cable, otherwise
communication errors will occur.
You must use AX-KO1502 cable for 9 pin serial port (supplied) or AX-KO1503 cable for
25 pin serial port (available separately as an option).
Figure 2 : Connection between the TM-2430 series and the computer
(The illustration above is not to scale.)
Step 1 Connect the male end of the A&D-supplied RS-232C communication cable to the
monitor.
Step 2 Connect the female end of the A&D-supplied RS-232C communication cable to the
computer.
7
Step 3 The display on the monitor will show “----”. If “----” does not appear, see “3-2
Communication Errors”. Otherwise go to step 4.
Note:If possible, keep the cable connected to the computer unless the port is needed
for other functions.
Step 4 Run the software. Step 5 Confirm that the correct serial port and the correct monitor type are set. For details,
see “2-9-4-2 Serial Port” and “2-9-4-3 Recorder Type”.
2-2-3 Connecting the UA -767PC to the computer
You must use the A&D-supplied RS-232C communication cable, otherwise
communication errors will occur.
You must use AX-KO1502 cable for 9 pin serial port (supplied) or AX-KO1503 cable for
25 pin serial port (available separately as an option).
Figure 3: Connection between the UA-767PC and the computer
(The illustration above is not to scale.)
Step 1 Connect the male end of the A&D-supplied RS-232C communication cable to the monitor. Step 2 Connect the female end of the A&D-supplied RS-232C communication cable to the
computer.
Step 3 After the connection, the display on the monito r w ill not change. The display will show
“----”, just before starting communication.
Note:If possible, keep the cable connected to the computer unless the port is needed
for other functions.
Step 4 Run the software. Step 5 Confirm that the correct serial port and the correct monitor type are set. For details,
see “2-9-4-2 Serial Port” and “2-9-4-3 Recorder Type”.
The software is now ready to:
Retrieve data from the monitor. See “2-7-1 Retrieving Data from the Monitor”. Clear the monitor’s memory. See “2-7-2 Clearing the Monitor’s Memory”. Program the monitor’s intervals and conditions. See “2-7-3 Programming the Monitor’s
Intervals and Conditions”.
8
2-3 Main Screen Description
The main screen is comprised of six menus and two toolbar buttons located permanently along the top of the screen. These menus may be used to access the various features of the software. The toolbar buttons may be used to set Part ial Analysis and Bloo d Pressure Mode parameters.
Toolbar Buttons: (Only for TM-2420/TM-2421, TM-2430)
Button See Description
Analysis Intervals 2-4-1 Analysis Intervals” The Analysis Intervals button may be used
to define specific “Full” and “Partial” time intervals within an opened data file for custom data analysis.
Note: Not available with the UA-767PC.
BP Mode 2-4-2 BP Mode” The BP Mode selection box may be used
to specify which blood pressure measurement method (Korotkoff, oscillometric or both) is used as the basis for statistical and graph ical analysis.
Note: This only applies to TM-2421 data
files.
Main Menus:
Menu See Description
File 2-5 File Menu The File menu may be used to manage the
files stored on the hard drive or a floppy disk and to exit the software.
View 2-6 View Menu The View menu may be used to access the
statistical, tabular, and graphical analysis screens.
Recorder 2-7 Recorder Menu The Recorder menu may be used to
program the monitor, clear measurement data in the monitor’s memory and to retrieve the measurement data from the mo nit or.
Report 2-8 Report Menu The Report menu may be used to specify
report formats and print reports.
Option 2-9 Options Menu The Options menu may be used to set the
default settings and user-defined parameters.
Help 2-10 On-line Help The Help menu may be used to view
information on software usage. Help may also be accessed by pressing F1 when various screens are opened or by clicking the Help button on a screen.
9
2-4 Toolbar
There are two toolbar buttons whi ch appear underneath the main me nu bar:
(1) Analysis Intervals (2) BP Mode
2-4-1 Analysis Intervals
The Analysis Intervals feature is used to define specific periods of time within an opened data file for custom data analysis. Two int ervals, Full Analysis and Part ial Analysis may be define d for ea ch data file.
Figure 4: Analysis Intervals dialog box
To set the Analysis Interv als, complete these steps:

Step 1 Click the Analysis Intervals button on the toolbar. Step 2 Set each item in the Full analysis interval box, using the scroll bar. Step 3 Check a Status button in t he Part i al analysis interval box. When “Enabled” is checked,

set each item in the box.
Step 4 Click the Save butt on. Step 5 To cancel the settings, click the Cancel button.
The Partial Analysis I nt erval affects many of the software feat ures including:
Summary Data
After the Partial Analysis Interval is enabled, the Partial tab in the Summary Data will be activated.
Trend Graphs
After Partial Analysis Interval is defined, the data specified within the parameters will appear in red on the Trend Graph.
Correlation Plots
After Partial Analysis parameters are defined, a second set of Correlation plots in red will appear based on the data speci fi ed w ithin the Partial Analys i s parameters.
10
Histograms
After Partial Analysis parameters are defined, the data specified within the parameters will appear in red overlayin g t he Full Analysis data.
2-4-2 BP Mode
The BP Mode is used to select which blood pressure method will be used as the primary and secondary choices for displaying the blood pressure measurement data. A&D monitors employ either Korotkoff, oscillometric or dual Korotkoff and oscillometric measurement methods. If the current data file contains data from a dual system, you may choose Korotkoff, oscillometric or both methods to analyze the dat a.
The measurement method(s) specified in the BP Mode will be the basis for the statistical and graphical analysis performed on the opened data file. The method noted with a one (1) is the primary method and the method noted with a two (2) is the secondary method. If a primary method measurement error occurred during the monitoring session or a measurement is manually excluded (see “2-6-2-1 Excluding/Including BP Measurements”), a valid measurement in the secondary method will be substitut ed for it with the appropriate BP Mode selection.
Method Setting Description
Oscillometric only This option sets the primary blood pressure choice to
oscillometric and disables the secondary choice. If the oscillometric reading for any measurement is invalid or has been excluded by the user in the BP Data List, the measurement will be excluded from graphical displays and statistical computat ions.
1: Oscillometric, 2: Korotkoff This option sets the primary blood pressure choice to
oscillometric and the secondary choice to Korotkoff. If an oscillometric reading is invalid, oscillometric readings will be substituted with Korotkoff readings. If both readings are invalid for a measurement, all data from the measurement is excluded.
Korotkoff only This option sets the primary blood pressure choice to
Korotkoff and disables the secondary choice. If the Korotkoff reading for any measurement is invalid or has been excluded by the user in the BP Data List, the measurement will be excluded from graphical displays and statistical computat ions.
1: Korotkoff, 2: Oscillometric This option sets the primary blood pressure choice to
Korotkoff and the secondary choice to oscillometric. If a Korotkoff reading is invalid, Korotkoff readings will be substituted with oscillometric readings. If both readings are invalid for a measurement, all data from the measurement is excluded.
11
To specify the BP Mode while v iewing an opened data file, complete these steps:
Step 1 Select the BP Mode box on the t ool bar. Step 2 Select the desired BP Mode u sed t o view and analyze the opened d ata file.
Figure 5: BP Mode selection on toolbar
To set a default BP Mode for the software at st art up, see “ 2-9-4 Default Settings”.
Note: The BP Mode for printed reports is assigned separately in the Report Format
Setup. See “2-8 Report Menu”.
2-5 File Menu
The File menu is used to manage the files stored on a hard drive or floppy disk. It may also be used to exit the software. The File menu has the following sub-menus:
Menu Description
Open This function is used to select a data file for viewing, modifying or
printing a report. After a file has been opened, the file name, patient name and date of the monitoring session are displayed in the title bar of the main screen
Save As This function is used to copy a selected file to a new file with a
different name. Delete This function is used to delet e a selected file. Convert This function is used to convert the blood pressure measurement
data of a data file to a format which is readable by most spreadsheet
applications. The convert ed file w il l have t he same na me as t he data
file with a .CSV file extension. Merge the BP data files This function is used to merge multiple data files into one file. Print This function is used to print reports for one or more data files
according to a user-selectable report f ormat. Exit This selection is used to exit t he software.
12

2-5-1 Open The File | Open function is used to access the blood pr essure data within existing data files.

Figure 6: File Open dialog box
To open a file, complete the fol lowing steps:
Step 1 Select the File menu. Step 2 Select the Open sub-menu. Step 3 Select a file from the directory shown or move to a different directory. Step 4 Click the Open button.
Note: The File | Save As, View and Report menus and Analysis Intervals buttons are
disabled until a data file is opened.
2-5-2 Save As

The File | Save As function is used to copy an open data file to another data file with a different name or to a different directory with any name.

When a file is copied, corresponding .dat and .p files are created. These two files must stay together within the same directory. The Save As function creates new .dat and .p file with the new file name in the selected directory. When entering a new file name in the Save As dialog box, it is not necessary to type the file extension. The original file will not be altered when using the Save As function.

2-5-3 Delete The File | Delete function is used t o delete a data file.

CAUTION: Data files will be permanently removed when deleted.
13
2-5-4 Convert

The File | Convert function is used to create a .CSV (comma delimited) format file from a .dat file that is compatible with most spreadsheet applications, including Microsoft Excel™. The .CSV file that is created includes DATE, TIME, SYS, DIA, MAP, PUL headings so that the data can be identified correctly when opened in a spreadsheet. The original file will not be modified in any way. The data is being converted and will be saved to the same name and directory as the original data file, but with a .CSV extension.

2-5-5 File Merge The File | Merge function is used to merge multiple data files into a new file.
Figure 7: File Merge dialog box
To merge multiple data files, complete the following steps:
Step 1 Select the File menu. Step 2 Select the Merge the BP data files sub-menu. Step 3 Select the data type to be merge d. Step 4 Open the folder that contains the dat a f iles t o be merged. Step 5 Drag the selected data files to t he Merge-File-List Box. Step 6 Click the OK button. Step 7 Save the selected data files in a new file.
The data files, of which time periods overlap, can not be merged.
Note:
The merged file will not open if the merged time period exceeds:
8 days (TM-2420/TM-2421 and TM-2430) 5 years(UA-767P C).
File merging is available only for data files of the same monitor. Data files of
different monitors can not be merged. The default data type of the file merge dialog box, when opened, is UA-767P(C).
14
2-5-6 Print

The File | Print function is used to print reports for one or more data files according to a user­selected format. This enables the user to generate multiple reports without having to open each file separately.

To print a report for multiple data fil es, complete the following steps:
Step 1 Select the File menu, then select the Print sub-menu. Step 2 In the Print dialog box, select the files that you want to print a report for by holding
the Shift or Ctrl key down while clicking t he desired files. Step 3 Click Open. Step 4 In the Report Selection dialog box, select the report format that you want to use for
the reports or click Print “M in i” r epo rt to print a condensed report. For “Mini” report
printout examples, see pages 60 and 61.

2-5-7 Exit The File | Exit selection may be used to exit the softw are.

2-5-8 Saving a File Files are automatically saved when the data is retrieved from the monitor. See “2-7-1 Retrieving Data from the Monitor”. Any modifications made to the data file are saved immediately when the
Save button is selected to close a function.
2-5-9 Closing a Data File
There is no file close feature in t his software. A file may be clos ed in two ways:
1) By opening another file. The current file will automatically be closed when you open another file.
2) By exiting the software.
15
2-6 View Menu
The View menu may be used to view numerical, statistical and graphical data screens. The View menu is disabled unless a fi le i s opened. For information on how t o open a file, see “2-5-1 Open”.
The View menu has the follow i ng sub-menus:
Menu See Description
Summary Data 2-6-1 Su mmary Data” Function for viewing statist ical data. BP Data List 2-6-2 BP Data List Function for viewing numerical data and
entering comments.
Trends 2-6-3 Trends Screen Function for viewing blood pressure, mean
arterial blood pressure, pulse (heart rate) and double product trends.
Correlation 2-6-4 Correlation Plots Function for viewing correlation plots of
systolic blood pressure to diastolic blood pressure and diastolic blood pressure to pulse.
The scatter diagram allows easy observation of the correlation between systolic and diastolic blood pressures; the minimum and maximum values, the magnitude of fluctuation and the density, of a 24-hour blood pressure data.
Histograms 2-6-5 Histograms Function for viewing the frequency distribution
of the measurement data. Allows easy observation of the data that
occurs most frequently, the frequency distribution and the distribution width during the time interval.
Circadian rhythm 2-6-6 Circadian Rhythm Function for viewing the graph showing the
estimated amplitude and peak position, and the numerical data, based on the 95% confidence interval and the cosinor analysis graph.
Note: Not available with the UA-767PC.
HBI analysis 2-6-7 HBI Analysis” Function for viewing the graph showing the
hyperbalic area and hypobalic area, and the numerical data converted into a one-day period.
Hyperbalic; that area exceeding the user­defined limits for the sleep and awake intervals.
Hypobalic; that area not reaching the user­defined limits for the sleep and awake intervals.
Note: Not available with the UA-767PC.
Patient Information 2-6-8 Patient Information Function for entering patient and physician
information, as well as fields for the sleep and awake intervals.
16
2-6-1 Summary Data
The Summary Data screen includes statistics on the data of the opened file and may be viewed directly on the screen. These st at istics are separated into four data tabs:
Full Statistics based on all t he dat a within the file. Partial Statistics based on the data specified in the Partial Analysis parameter. See “2-4-1
Analysis Intervals”.
Awake Statistics based on the data during the wa king hours.
Note:Waking hours are specified as t hose hours outside the identif ied sleeping hours.
Sleep Statistics based on the data during the specified sleeping hours. See “2-6-8 Patient
Information”.
To view, close, or print the in dividual Summary Data screen, complete the following steps:

Step 1 Select the View menu. Step 2 Select the Summary Data sub-menu. The Summary Data display appears.

Figure 8: Summary Data display
Step 3 Click one of the four data tabs, Full, Partial, Awake or Sleep. The Full tab will always
be enabled. The Partial, Sleep and Awake tabs will be enabled after they are specified. See “2-4-1 Analysis Intervals” to specify Partial Analysis and “2-6-8
Patient Information” to specify the Sleep interval.
Step 4 To print the data for an individual Summary Data tab, select the tab you want to print
and click the Print but t on on the right hand side of the screen.
Note: You may also print the Summary Data screens in the standard report, see “2-8
Report Menu”.
Step 5 Click the Close button to close the Data Summary screen.
17
Each of the four Summary Data is titled with the name of the tab, file, patient and measurement method. For detailed information on Summary Data screen listin gs, see below:
Summary Data Tab Definition
Time Interval The analysis period, start time/date – end time/date, for the
activated Summary Data tab.
Valid Readings The total number of measurements taken during the monitoring
session less any monitoring errors or manually excluded readings within the data set. See “3-1 Measurement Errors” for information on measureme nt errors.
Excluded Readings The measurement error percentage. This figure is based on the
combined total of measurement errors and excluded measurements.
Min (Minimum) The minimum measurement value for Systolic, Diastolic, MAP
and Pulse measurements during the Time Interval.
Mean The mean value for Systolic, Diastolic, MAP and Pulse
measurements during the T i me Int erval.
Max (Maximum) The maximum measurement value for Systolic, Diastolic, MAP
and Pulse measurements dur ing t he Time Interval.
SD (Standard Deviation) The Standard Deviation of the value for Systolic, Diastolic, MAP
and Pulse measurements dur ing t he Time Interval.
SE (Standard Error) The Standard Error of the value for Systolic, Diastolic, MAP and
Pulse measurements during t he Time Interval.
CV(%) (Coefficient of Variation )
SYS/DIA limits Systolic and Diastolic limits for the Awake and Sleep int e rval. Readings > SYS limit(s) The percentage of the Systolic measurement values greater than
Readings > DIA limit(s) The percentage of the Diastolic measurement values greater
Highest Systolic Value The time and date of the highest Systolic value during the Time
Highest Diastolic Value The time and date of the highest Diastolic value during the Time
Lowest Systolic Value The time and date of the lowest Systolic value during the Time
Lowest Diastolic Value The time and date of the lowest Diastolic value during the Time
The Coefficient of Variation for Systolic, Diastolic, MA P and Pulse measurements during the T i me Int erval.
the specified Systolic limit. See “2-9-4-1 BP (Blood Pressure) Settings" for information on specifying the Systolic Blood Pressure Limit.
than the specified Diastolic limit. See "2-9-4-1 BP (Blood Pressure) Settings" for information on specifying the Diastolic Blood Pressure Limit.
Interval.
Interval.
Interval.
Interval.
18
2-6-2 BP Data List
The BP Data List functio n di splays individual blo od pressure measurement withi n an open data file in tabular form. With the BP Data List, you may input measurement comments, exclude/include a measurement within the graph ics and statistics or link a measurement to the begin ning of the Trends Screen.
The following table defines each heading with in t he Blood Pressure Data List :
Item Description
# Measurement number
DATE/TIME Date and time of the measurem ent
SYS Systolic blood pressure
DIA Diastolic blood pr essure MAP Mean art erial blood pressure, obtained by MAP=DIA+(SYS-DIA)/ 3 PUL Pulse DP STATUS The status line has 3 positions. When filled with a character other than “–”,
Position 1 BP method, O or K
Position 2 E (Event)
Position 3 Error code
Exclude When this column has an “X”, the measurement has been excluded from the
Comments Space provided for user to enter patient diagnostic notations up to 20
Vertical scroll
bar
Save button Saves any edits, i.e. comments, exclusio ns or inclusions. Close button Closes the screen.
Double Product, obtained by DP=SYS∗PUL/100
each denotes a specific occurrence or condition. Denotes the BP method used for the specific measurement. “K” refers to
Korotkoff method, “O” refers to oscillometric method.
Indicates that data was measured when the monitor’s EVENT button was activated.
Denotes the specific error code that occurred during the monitoring session, see “3-1 Measurement Errors”.
statistics and graphs.
characters. The vertical scroll bar on the right hand side of the screen helps scroll through
the measurements. This may also be performed with the up and down arrow keys, as well as the mouse.
Note: Edits (comments, exclusions or inclusions) will not be saved.
Help button Opens t he Blood Pressure Data List help screen.
19
Figure 9: Blood Pressure Data List display
2-6-2-1 Excluding/Includi ng BP Measurements
It is useful to exclude a measurement from the Blood Pressure Data List in cases where the measurement is clearly incorrect. The software provides a simple way to include or exclude measurements while viewing a list. When a measurement is excluded from the Blood Pressure Data List, it will not be included in any of the software’s statisti cal or graphical analysis fu nct i ons.
A measurement can be included or excluded by clicking on the measurement in the column labeled Exclude. An “X” in the Exclude column indicates that the measurement is excluded. A “-“ indicates that the measurement is inc luded.
In some cases the software automatically excludes measurements if an error occurred during the reading. Measurements that have been automatically excluded are displayed with and “X” in the Exclude column and cannot be included by the user.
Note:Measurements that have been excluded are not deleted from the data file.
2-6-2-2 Entering Comments
Comments, such as information taken from the patient activity diary, may be entered into the data file. To enter a comment, complete the steps below:
Step 1 Click the measurement in t he column labeled Comme n t s. Step 2 Type your comments – maximum 20 ch aract ers.
Note:You may write over any of the automatic measurement error comments.
Step 3 Press the Enter key or clic k a different measurement to confirm the comments. Step 4 To save the comments, click the Save button before closing the Blood Pressure Data
List screen.
Step 5 Click the Close button if you do not want t o save the comments.
20