This document and the subject matter disclosed herein are proprietary information.
Abbott Laboratories retains all the exclusive rights of dissemination, reproduction, manufacture,
and sale. Any party using this document accepts it in confidence, and agrees not to duplicate
it in whole or in part nor disclose it to others without the written consent of Abbott Laboratories.
430-95424-001 (Rev. 09/03)Plum A+3 Infusion System
Page 4
Change History
Part NumberDescription of Change
430-95424-001 (Rev. 09/03)Original issue
Technical Service Manual i430-95424-001 (Rev. 09/03)
Page 5
CHANGE HISTORY
This page intentionally left blank.
430-95424-001 (Rev. 09/03) iiPlum A+3 Infusion System
Technical Service Manualvii430-95424-001 (Rev. 09/03)
Page 11
CONTENTS
This page intentionally left blank.
430-95424-001 (Rev. 09/03)viiiPlum A+3 Infusion System
Page 12
Section 1
INTRODUCTION
The Plum A+3™ Infusion System is designed to meet the growing demand for hospital wide
device standardization. The Plum A+3 consists of three component pumps which are
designated line 1, line 2, and line 3. By incorporating three lines into one unit, the infusion
system provides three primary lines, three secondary lines, and piggyback fluid delivery
capabilities. The Plum A+3 serves a wide range of general floor and critical care needs.
Compatibility with the LifeCare
the Plum A+3 infusion system convenient and cost-effective.
1.1
SCOPE
This manual is organized into 11 sections:
❏ Section 1 Introduction
❏ Section 2 Warranty
❏ Section 3 System Operating Manual
❏ Section 4 Theory of Operation
❏ Section 5 Maintenance and Service Tests
❏ Section 6 Troubleshooting
❏ Section 7 Replaceable Parts and Repairs
❏ Section 8 Specifications
❏ Section 9 Drawings
❏ Section 10 Index
❏ Technical Service Bulletins
®
5000 PlumSet® administration sets and accessories make
If a problem in device operation cannot be resolved using the information in this manual,
Specific instructions for operating the device are contained in the Plum A+3 SystemOperating Manual. Provision is made for the inclusion of the system operating manual
in Section 3 of this manual.
Note: Figures are rendered as graphic representations to approximate actual product.
Therefore, figures may not exactly reflect the product.
Note: Screen representations in this manual are examples only, and do not necessarily
reflect the most current software version.
Technical Service Manual1 - 1430-95424-001 (Rev. 09/03)
Page 13
SECTION 1 INTRODUCTION
1.2
CONVENTIONS
The conventions listed in Table 1-1, Conventions, are used throughout this manual.
Table 1-1. Conventions
ConventionApplicationExample
ItalicReference to a section, figure,
table, or publication
[ALL CAPS]
[Lowercase]
ALL CAPSScreen displaysCASSETTE TEST IN PROGRESS
BoldEmphasisCAUTION: Use proper ESD grounding
Throughout this manual, warnings, cautions, and notes are used to emphasize important
information as follows:
WARNING:
CAUTION: A CAUTION usually appears in front of a procedure or statement.
It contains information that could prevent hardware failure, irreversible damage to
equipment, or loss of data.
In-text references to keys
and touchswitches
In-text references to softkeys
A WARNING CONTAINS SPECIAL SAFETY EMPHASIS AND MUST
BE OBSERVED AT ALL TIMES. FAILURE TO OBSERVE A WARNING
MAY RESULT IN PATIENT INJURY AND BE LIFE-THREATENING.
(see Section 6.1, Technical Assistance)
[START]
[Choose]
techniques when handling
components.
Note: A note highlights information that helps explain a concept or procedure.
1.3
COMPONENT DESIGNATORS
Components are indicated by alpha-numeric designators, as follows:
BatteryBTDiodeDResistorR
CapacitorCFuseFSwitchSW
CrystalYIntegrated CircuitUTransistorQ
The number following the letter is a unique value for each type of component (e.g., R1, R2).
Note: Alpha-numeric designators may be followed with a dash (-) number that
indicates a pin number for that component. For example, U15-13 is pin 13 of the
encoder chip [U15] on the interface PWA.
430-95424-001 (Rev. 09/03)1 - 2Plum A+3 Infusion System
Page 14
1.4 ACRONYMS AND ABBREVIATIONS
1.4
ACRONYMS AND ABBREVIATIONS
Acronyms and abbreviations used in this manual are as follows:
430-95424-001 (Rev. 09/03)1 - 4Plum A+3 Infusion System
Page 16
1.5 USER QUALIFICATION
1.5
USER QUALIFICATION
The Plum A+3 infusion system is intended for use at the direction or under the supervision
of licensed physicians or certified healthcare professionals who are trained in the use of
the pump and the administration of parenteral and enteral fluids and drugs, and whole
blood or red blood cell components. Training should emphasize preventing related IV
complications, including appropriate precautions to prevent accidental infusion of air.
The epidural route can be used to provide anesthesia or analgesia.
1.6
ARTIFACTS
Nonhazardous, low-level electrical potentials are commonly observed when fluids are
administered using infusion devices. These potentials are well within accepted safety
standards, but may create artifacts on voltage-sensing equipment such as ECG, EMG,
and EEG machines. These artifacts vary at a rate that is associated with the infusion rate.
If the monitoring machine is not operating correctly or has loose or defective connections
to its sensing electrodes, these artifacts may be accentuated so as to simulate actual
physiological signals. To determine if the abnormality in the monitoring equipment is
caused by the pump instead of some other source in the environment, set the pump so
that it is temporarily not delivering fluid. Disappearance of the abnormality indicates that
it was probably caused by electronic noise generated by the pump. Proper setup and
maintenance of the monitoring equipment should eliminate the artifact. Refer to the
appropriate monitoring system documentation for setup and maintenance instructions.
Technical Service Manual1 - 5430-95424-001 (Rev. 09/03)
Page 17
SECTION 1 INTRODUCTION
1.7
INSTRUMENT INSTALLATION
PROCEDURE
CAUTION: Infusion pump damage may occur unless proper care is exercised during
product unpacking and installation. The battery may not be fully charged upon
receipt of the infusion pump. Do not place the infusion pump in service if it fails the
self test.
CAUTION: Infusion pump performance may be degraded by electromagnetic
interference (EMI) from devices such as electrosurgical units, cellular phones,
and two-way radios. Operation of the infusion pump under such conditions should
be avoided.
The instrument installation procedure consists of unpacking, inspection, and self test.
Note: Do not place the infusion pump in service if the battery is not fully charged.
To make certain the battery is fully charged, connect the infusion pump to AC power
for six hours (see Section 8, Specifications).
1.7.1
UNPACKING
Inspect the shipping container as detailed in Section 1.7.2, Inspection. Use care when
unpacking the infusion pump. Retain the packing slip and save all packing material in
the event it is necessary to return the Plum A+3 to the factory. Verify the shipping container
contains a copy of the system operating manual.
1.7.2
INSPECTION
Inspect the infusion pump shipping container for shipping damage. Should any damage
be found, contact the delivering carrier immediately.
CAUTION: Inspect the infusion pump for evidence of damage. Do not use the pump
if it appears to be damaged. Should damage be found, contact Abbott Laboratories
(see Section 6.1, Technical Assistance).
Inspect the infusion pump periodically for signs of defects such as worn accessories,
broken connections, or damaged cable assemblies. Also inspect the infusion pump after
repair or during cleaning. Replace any damaged or defective external parts.
430-95424-001 (Rev. 09/03)1 - 6Plum A+3 Infusion System
Page 18
1.7 INSTRUMENT INSTALLATION PROCEDURE
1.7.3
SELF TEST
CAUTION: Do not place the infusion pump in service if the self test fails.
Note: If an alarm condition occurs during the self test, cycle the power and repeat
the self test. If the alarm condition recurs, note the message and take corrective action
(see Section 6, Troubleshooting). Repeat the self test. If the alarm condition recurs,
remove the Plum A+3 infusion system from service and contact Abbott Laboratories.
Note: When performing the self test, line 1, line 2, and line 3 must be tested.
However, if appropriate, the test may be performed on all lines concurrently.
To perform the self test, refer to Figure 1-1, LCD and Keypad, and proceed as follows:
1. Connect the AC power cord to a grounded AC outlet. Verify the charging/line
indicator CHARGE illuminates and an alarm beep sounds.
2. Without a cassette installed, press [ON/OFF] to turn on the pump.
3. The LCD briefly displays the SELF TEST screen. Verify the screen display matches
Figure 1-1.
Note: If the SELF TEST screen does not appear, contact Abbott Laboratories.
4. After the self test is complete, the message “INSERT PLUM SET CLOSE LEVER”
appears. Press the decimal [.] key, then [START].
5. Using the [SELECT] arrow keys, select Set Time and Date, and press the [Choose]
softkey.
6. Verify the time, year, month, and day are correct. If any parameters are incorrect,
refer to Section 1.8.3, Setting the Time and Date.
7. Press [ON/OFF] to exit the SET TIME AND DATE screen.
8. Press [ON/OFF] to turn the pump back on.
9. Open the cassette door and insert a primed cassette. Close the cassette door.
The cassette test is complete when the “CASSETTE TEST IN PROGRESS” message
disappears.
Note: The message “MECHANISM INITIALIZATION IN PROGRESS” may briefly
appear prior to the “CASSETTE TEST IN PROGRESS” message.
10. If previously entered programming exists, the “CLEAR SETTINGS?” message
appears. Press the [Yes] softkey to clear the settings.
Technical Service Manual1 - 7430-95424-001 (Rev. 09/03)
Page 19
SECTION 1 INTRODUCTION
LINE FLOW
A
INDICATORS
B
ABBOTT Plum A+
STATUS
REGION
MESSAGE
REGION
SOFTKEY
LABEL REGION
LINE
INDICATOR
START
STOP
CHARGE
ON OFF
Release XX.X - MM/DD/YY
Copyright Abbott Laboratories
2002
System Self Test
In Pr
o
gress
1
4
7
CLEAR
2
5
8
0
3
6
9
.
WORKING
REGION
SOFTKEYS
SELECT
KEYPAD
SILENCE
98G01002
Figure 1-1. LCD and Keypad
430-95424-001 (Rev. 09/03)1 - 8Plum A+3 Infusion System
Page 20
1.8 BIOMED SETTINGS
1.8
BIOMED SETTINGS
The biomed settings screens contain the following options that can be changed or reviewed
by qualified personnel:
❏ IV screen parameters
❏ Alarms log
❏ Set time and date
Note: All Plum A+3 infusion devices (new or refurbished) are shipped with factory
settings (see Table 1-2, System Configuration Data).
Note: Biomed screens do not time out for the Infuser Idle alarm or No Action alarm.
Note: The battery will not be detected in the biomed service mode (see Section 6.2.2,
Error Codes Requiring Technical Service).
To access the service mode, refer to Figure 1-1, then proceed as follows:
1. Open the door and press [ON/OFF] to turn on the pump.
2. After the self test is complete, the message “INSERT PLUM SET CLOSE LEVER”
appears. Press the decimal [.] key, then [START], and verify the BIOMED SETTINGS
screen is displayed (see Figure 1-2, Biomed Settings).
BIOMED SETTINGS
IV Screen Parameters
Alarms Log
Set Time and Date
Select, then Choose
Choose
00H03002
Figure 1-2. Biomed Settings
Technical Service Manual1 - 9430-95424-001 (Rev. 09/03)
Page 21
SECTION 1 INTRODUCTION
DataOptions RangeFactory Setting
Table 1-2. System Configuration Data
Maximum macro IV mode delivery rate0.1 - 99.9 mL/hr and
100 - 999 mL/hr
Macro distal occlusion alarm
(pressure level)
Deliver together enableConcurrent or PiggybackPiggyback
Delayed start/standby enableYes or NoYes
Continue rateRate or KVOKVO
Nurse callback defaultYes or NoNo
Time(24 hr) 00:00 - 23:59 in
Date1/1/2002 - 12/31/2098Factory date
1 to 15 psi6.0 psi
one minute increments
999 mL/hr
Factory time
1.8.1
IV PARAMETERS
Refer to Figure 1-3, IV Parameters. The IV parameters screen contains the following:
❏ Common IV parameters
❏ Macro IV parameters
To change the IV parameters, refer to Figure 1-6, Common IV Parameters, and Figure 1-7,
Macro IV Parameters, then proceed as follows:
1. Access the biomed settings screen as described in Section 1.8.
2. Using the [SELECT] arrow keys, select IV Screen Parameters, and press [Choose].
3. Using the [SELECT] arrow keys, select the parameters to be changed, and press
[Choose].
4. Using the [Change Value] softkey, select the desired value, and press [ENTER].
5. Repeat Steps 3 and 4 for each parameter to be changed.
6. If there are no other changes, press [ON/OFF] to power off the infusion device.
430-95424-001 (Rev. 09/03)1 - 10Plum A+3 Infusion System
Page 22
BIOMED SETTINGS
IV Parameters
Common IV Parameters
Macro IV Parameters
1.8 BIOMED SETTINGS
Select, then Choose
Figure 1-3. IV Parameters
ChooseBack
00H03003
Technical Service Manual1 - 11430-95424-001 (Rev. 09/03)
Page 23
SECTION 1 INTRODUCTION
1.8.2
ALARMS LOG
To view the alarms log, refer to Figure 1-2 and Figure 1-4, Alarms Log, then proceed
as follows:
1. Access the biomed settings screen as described in Section 1.8.
2. Using the [SELECT] arrow keys, select Alarms Log, and press [Choose}.
3. Use the [Page Up] and [Page Down] softkeys to view the alarms log.
4. Press the [Back] softkey to exit the alarms log and return to the main biomed settings
screen.
Note: The alarms log will retain the latest 40 alarm and malfunction codes, listed
in order from the most current to the oldest.
ALARMS LOG
6/23/03 01:43 E437 S/W Failure # 202
6/23/03 09:18 N190 Neg. Prox. Occl. A
6/22/03 23:44 N102 Infuser Idle 2 minutes
6/22/03 21:43 N161 Line A VTBI complete
6/22/03 11:44 N106 Distal occlusion
6/22/03 09:43 N161 Line A VTBI complete
6/22/03 06:23 N160 Line B VTBI complete
6/22/03 03:40 N101 No action alarm
Page
Up
Figure 1-4. Alarms Log
Page
Down
Back
02H03008
430-95424-001 (Rev. 09/03)1 - 12Plum A+3 Infusion System
Page 24
1.8 BIOMED SETTINGS
1.8.3
SETTING THE TIME AND DATE
To set the time and date, refer to Figure 1-2 and Figure 1-5, Setting the Time and Date,
then proceed as follows:
Note: The Plum A+3 will automatically display February 29 on leap years.
Note: Daylight savings and time zone changes must be made manually.
1. Access the biomed settings screen as described in Section 1.8.
2. Using the [SELECT] arrow keys, select Set Time and Date, and press [Choose].
3. Using the [SELECT] arrow keys, select the parameter to be changed.
4. Using the numerical keypad, enter the desired value, and press [Enter].
5. Repeat Steps 3 and 4 for each parameter to be changed.
6. If there are no other changes, press [ON/OFF] to power off the infusion device.
BIOMED SETTINGS
Set Time and Date
Time: 15hr:min
Year
Month
Day
Enter value using keypad
2003
10
31
Figure 1-5. Setting the Time and Date
21
Enter
Cancel/
Back
00K13004
Technical Service Manual1 - 13430-95424-001 (Rev. 09/03)
Page 25
SECTION 1 INTRODUCTION
BIOMED SETTINGS
Common IV Parameters
Continue Rate
Deliver Together
Enable Delay/Standby
KVO
Concurrent
Yes
Callback Default
Select using Change Value
Change
Value
Figure 1-6. Common IV Parameters
Enter
Cancel/
BIOMED SETTINGS
Macro IV Parameters
No
Back
02K03004
Default Distal Press
Max Rate
Enter Value using keypad
Enter
Figure 1-7. Macro IV Parameters
6.0 psi
999 mL/hr
Cancel/
Back
00K13005
430-95424-001 (Rev. 09/03)1 - 14Plum A+3 Infusion System
Page 26
Section 2
WARRANTY
Subject to the terms and conditions herein, Abbott Laboratories, herein referred to as
Abbott, warrants that (a) the product shall conform to Abbott's standard specifications
and be free from defects in material and workmanship under normal use and service for
a period of one year after purchase, and (b) the replaceable battery shall be free from
defects in material and workmanship under normal use and service for a period of 90 days
after purchase. Abbott makes no other warranties, express or implied, as to
merchantability, fitness for a particular purpose, or any other matter.
Purchaser's exclusive remedy shall be, at Abbott's option, the repair or replacement of the
product. In no event shall Abbott's liability arising out of any cause whatsoever (whether
such cause be based in contract, negligence, strict liability, other tort, or otherwise) exceed
the price of such product, and in no event shall Abbott be liable for incidental,
consequential, or special damages or losses or for lost business, revenues, or profits.
Warranty product returned to Abbott must be properly packaged and sent freight prepaid.
The foregoing warranty shall be void in the event the product has been misused, damaged,
altered, or used other than in accordance with product manuals so as, in Abbott's
judgment, to affect its stability or reliability, or in the event the serial or lot number has
been altered, effaced, or removed.
The foregoing warranty shall also be void in the event any person, including the Purchaser,
performs or attempts to perform any major repair or other service on the product without
having been trained by an authorized representative of Abbott and using Abbott
documentation and approved spare parts. For purposes of the preceding sentence, "major
repair or other service" means any repair or service other than the replacement of accessory
items such as batteries, flow detectors, detachable AC power cords, and patient pendants.
In providing any parts for repair or service of the product, Abbott shall have no
responsibility or liability for the actions or inactions of the person performing such repair
or service, regardless of whether such person has been trained to perform such repair or
service. It is understood and acknowledged that any person other than an Abbott
representative performing repair or service is not an authorized agent of Abbott.
Technical Service Manual2 - 1430-95424-001 (Rev. 09/03)
Page 27
SECTION 2 WARRANTY
This page intentionally left blank.
430-95424-001 (Rev. 09/03)2 - 2Plum A+3 Infusion System
Page 28
Section 3
SYSTEM OPERATING MANUAL
A copy of the system operating manual is included with every Plum A+3 infusion system.
Insert a copy here for convenient reference. If a copy of the system operating manual is
not available, contact Abbott LaboratoriesTechnical Support Operations
(see Section 6.1, Technical Assistance).
Technical Service Manual3 - 1430-95424-001 (Rev. 09/03)
Page 29
SECTION 3 SYSTEM OPERATING MANUAL
This page intentionally left blank.
430-95424-001 (Rev. 09/03)3 - 2Plum A+3 Infusion System
Page 30
Section 4
THEORY OF OPERATION
This section describes the Plum A+3 infusion system theory of operation. Related drawings
are provided in Section 9, Drawings. The theory of operation details the general
description, electronic subsystem overview, printed wiring assemblies (PWA), remote
mounted peripherals, and mechanical overview of the infusion pump.
4.1
GENERAL DESCRIPTION
The infusion system consists of three component pumps which are designated line 1,
line 2, and line 3. Each line includes the following features:
-Dose calculation
-Loading dose
-Multi-step programming
-Therapy selection
-Nurse call
-Delayed start setting
-Standby mode
-Drug label library
-Piggyback and concurrent
delivery modes
-Titration
-0.1-99.9 mL/hr flow rate range
for both lines
(in 0.1 mL/hr increments)
-100-999 mL/hr flow rate range
for both lines
(in 1.0 mL/hr increments)
-Anti free-flow protection
-Air removal/backpriming
-Battery gauge
-Long battery life (6 hours) for
emergency backup and temporary
portable operation
-Air detection (proximal and distal)
-Serial communication
-Alarm history
-Plug-in barcode reader
for drug identification (optional)
-Volumes infused
(A, B, total volumes)
-KVO at dose end (1.0 mL/hr
or less depending on delivery rate)
or continue rate to continue
-Variable distal pressure setting
-Nonpulsatile volumetric accuracy
-Microprocessor control
-Large LCD
-Panel back illumination
on mains power
-Lockout switch
-Standard fullfill, partfill, syringe,
and vial use
-Enteral and parenteral fluid delivery
-Blood and blood product delivery
-Wide range of standard and
specialty administration sets
Technical Service Manual4 - 1430-95424-001 (Rev. 09/03)
Page 31
SECTION 4 THEORY OF OPERATION
Alarms include the following:
-Distal occlusion
-Proximal occlusion
-Proximal air-in-line
-Distal air-in-line
-Low battery
-Door open while pumping
Battery
Pressure Sensors
Plunger, LS, & IO Motors
MECHANISM
SYSTEM 1
Power Supply PWA
APP PWA
Switches PWA
Flex FSR
Driver
PWA
-Lockout violation
-VTBI complete
-Valve/cassette test failure
-Nurse call
-No action alarm
-Infuser idle for two minutes
Peripheral PWA
LCD Display
CPU
PWA
Keypad
LEDs
On/Off Switch
FRONT PANEL
Main Piezo Buzzer
Battery
Pressure Sensors
SYSTEM 2
Battery
Pressure Sensors
SYSTEM 3
PLUM A+3
Power Supply PWA
APP PWA
Switches PWA
Flex FSR
Plunger, LS, & IO Motors
MECHANISM
Power Supply PWA
APP PWA
Switches PWA
Flex FSR
Plunger, LS, & IO Motors
MECHANISM
Driver
PWA
Driver
PWA
CPU
PWA
CPU
PWA
Main Piezo Buzzer
Interface/Peripheral PWA
Figure 4-1. Electronic Functional Diagram
Peripheral PWA
LCD Display
Keypad
LEDs
On/Off Switch
FRONT PANEL
LCD Display
Keypad
LEDs
On/Off Switch
FRONT PANEL
Main Piezo Buzzer
Lockout Switch
Volume Control
Nurse Call Jack
Barcode Reader
DataPort
Driver
PWA
01K07015
430-95424-001 (Rev. 09/03)4 - 2Plum A+3 Infusion System
Page 32
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2
ELECTRONIC SUBSYSTEM OVERVIEW
This section describes the function and electronic circuitry (see Figure 4-1, Electronic
Functional Diagram) of three main subsystems in the infusion pump: CPU subsystem,
power supply subsystem, and mechanism subsystem. Schematic diagrams of subsystem
PWAs are in Section 9, Drawings.
Note: An asterisk (*) denotes an active low or negative true logic signal.
4.2.1
CPU SUBSYSTEM
The CPU subsystem contains the main microcontroller, which is responsible for controlling
the display/keyboard interface, external communications interfaces, barcode reader
interface, and system management (see Figure 9-12, Peripheral PWA Schematic,
and Figure 9-14, CPU PWA Schematic).
The CPU subsystem provides the following functions:
-External memory devices access
-LCD interfaces
-Real-time clock generator interface
-System watchdog
-Analog-to-digital and digital-to-analog converter interface
-Keypad interfaces
-Control and monitor status signals, such as LEDs, audible alarms,
volume control, nurse call switch, and lockout switch
-Serial communication with host computer (DataPort) and barcode reader
-Power supply subsystem interface
-Mechanism subsystem interface
4.2.1.1
CPU
The central processing unit is a Motorola MC68302 CPU. The CPU has a closely coupled
16 bit data bus and 24 bit address bus, MC68000 microprocessor core, a system
integration block for peripherals, and an RISC communications processor. The MC68302
is packaged in a 144 pin thin quad flat pack (TQFP) package and operates from a 3.3 V
power supply.
DC
The on-chip peripheral devices are isolated from the system through the dual port RAM.
The 1152 byte dual port RAM has 576 bytes of system RAM and 576 bytes of parameter
RAM, which contains various peripheral registers, parameters, and the buffer descriptors
for each of the three serial communication controller (SCC) channels and the serial
communication port (SCP) channels. The 24 bit address bus is capable of accessing up to
16 MB of data.
Technical Service Manual4 - 3430-95424-001 (Rev. 09/03)
Page 33
SECTION 4 THEORY OF OPERATION
4.2.1.2
SYSTEM MEMORY ADDRESS MAP
The CPU has a 24 bit address bus when combined with UDS*/A0. The address bus is a
bi-directional, three state bus capable of addressing 16 MB of data that is configured as
16 bits per word (including the IMP internal address space).
Each of the four programmable chip-select lines has two registers that define the starting
address of a particular address space and the block size.
4.2.1.3
PROGRAMMABLE READ-ONLY MEMORY
The CPU subsystem has two 512 K x 8 bit programmable read-only memory (PROM)
memory devices, which provide a total of 1024 KB. The PROM space is expandable up to
2 MB. The PROM memory devices operate off the 3.3 V
The CPU chip-select 0 pin (CS0*), is connected to the PROM chip-enable (CE*) pin (signal
CSROM*). This special chip-select signal can support bootstrap operation after reset.
The interface to the CPU is the 16 bit data bus, and a 19 bit address bus. The address
bus is connected to the ADDR<19:1> lines, and the data bus is connected to the
DATA<15:0> lines.
supply.
DC
4.2.1.4
STATIC RANDOM ACCESS MEMORY
There are two 512 K x 8 bit CMOS static random access memory (SRAM) devices, which
provide a total of 1024 KB of data memory. During an SRAM read or write cycle, the
chip-enable (CE*) is controlled by the CPU chip-select pin 1 (CS1*, signal name (CSRAM*)).
The SRAM space is expandable up to 2 MB. The SRAM operates off the 3.3 V
The CPU subsystem includes the additional SRAM for video buffer and real-time clock.
See Section 4.2.1.6, LCD Controller, and Section 4.2.1.9, Real-Time Clock.
4.2.1.5
supply.
DC
CONTROL LOGIC
The CPU PWA uses field programmable gate arrays (FPGA), which are high density, high
speed, I/O intensive general purpose devices. They are used to implement all the digital
control functions, including: memory-map address decoding; memory read-write enable;
direct memory access (DMA) request; I/O status signals; chip-select control; motor control;
sensor select; and power up/system reset control.
4.2.1.6
LCD CONTROLLER
The liquid crystal display (LCD) controller is used to interface the LCD to the CPU.
The device displays layered text and graphics, scrolls the display in any direction,
and partitions the display into multiple screens. It stores bit-mapped graphic data in
external frame buffer memory. The display controller functions include: transferring data
from the controlling microprocessor to the buffer memory; reading memory data;
converting data to display pixels; and generating timing signals for the buffer memory and
LCD panel.
430-95424-001 (Rev. 09/03)4 - 4Plum A+3 Infusion System
Page 34
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
The LCD controller accesses 32 KB of frame buffer SRAM (video) via the controller’s video
address and data busses (VA<14:0> and VD<7:0>). The LCD controller external clock
frequency is 8 MHz.
The interface to the CPU is through the lower 8 bits of the data bus, which is connected
to DATA<7:0> lines, address line A1, and LCD chip-select signal CSLCD* (CS2*).
This controller is also configured as 8080 family compatible interface device with all the
control signals, such as WRLCD* (WR*) and RDLCD* (RD*), generated by the FPGA logic.
The LCD controller and the display memory are operated off the 3.3 V
signal levels are shifted up to 5 V
by buffers for interface with the 5 VDC LCD panel.
DC
supply. The output
DC
4.2.1.7
LCD BACKLIGHT CONTROL
The LCD panel is backlit by a cold cathode fluorescent tube (CCFT) lamp. The CCFT lamp
requires 300 V
to deliver a current regulated sine wave to the lamp. A switching regulator regulates the
CCFT current by monitoring feedback pin 3, and varies its output duty cycle to drive a
DC/AC inverter. Intensity control is achieved by superimposing a DC control signal with
the feedback signal. The DC control signal is sourced by a voltage divider consisting of
a digitally controlled non-volatile potentiometer and three series diodes.
The CPU can adjust LCD backlight intensity by selecting the digitally controlled
non-volatile potentiometer and controlling TUBU/D and TUBINC* signals.
The potentiometer has a five bit up/down counter with non-volatile memory. It is used to
store one of 31 settings of the potentiometer. Each count represents 323 Ω with a range
of 323 to 10 KΩ. The current counter value is stored in non-volatile memory after CSTUB*
is returned high while the TUBINC* input is also high. The current counter value is not
stored if CSTUB* is returned high and TUBINC* is low. The CCFT intensity is directly
proportional to the CCFT current, where 0 mA
is maximum intensity. The CCFT current is inversely proportional to the counter value.
to operate; a current controlled DC to AC voltage inverter circuit is used
rms
is minimum intensity and 5 mA
rms
rms
4.2.1.8
LCD CONTRAST CONTROL
A digitally adjustable LCD bias supply is used to control the LCD contrast over a range
of -24
converter (DAC). The CPU provides two signals, LCDADJ (ADJ) and LCDCTL (CTL),
to interface with this device. On power up or after a reset, the counter sets the DAC output
to the mid-range value.
Each rising edge of LCDADJ increments the DAC output. When incremented beyond full
scale, the counter rolls over and sets the DAC to the minimum value. Therefore, a single
pulse applied to LCDADJ increases the DAC set point by one step, and 63 pulses decrease
the set point by one step.
Technical Service Manual4 - 5430-95424-001 (Rev. 09/03)
to -8 VDC. It is digitally adjustable in 64 equal steps by an internal digital-to-analog
Page 35
SECTION 4 THEORY OF OPERATION
4.2.1.9
REAL-TIME CLOCK
The watchdog timekeeper chip includes a complete real-time clock/calendar (RTC),
watchdog timer, alarm, and interval timer. The time/date information includes
hundredths of seconds; seconds; minutes; hours; day; date; month; and year. The date
at the end of the month is automatically adjusted for months with less than 31 days,
including correction for leap year. The watchdog timekeeper operates in either 24-hour or
12-hour format with an AM/PM indicator. The device can be programmed to set up an
interval timer, and it can generate an alarm every day, hour, or minute. These alarm
functions may be used to schedule real-time related activities. A parallel resonant 32,768
Hz crystal oscillator drives the internal time base.
The external interface is a separate (non-multiplexed) 8 bit data bus and 6 bit address
bus, with a contiguous address space of 64 bytes. When system power is turned off,
a battery voltage input is available, which makes the RTC data non-volatile. The address
bus is connected to the ADDR<6:1> lines, and the data bus is connected to DATA<7:0>
lines. Since the CPU accesses are 16 bits wide, the RTC data is on the lower byte of the word.
The RTC chip-enable pin (CE*) is active low enabled for read and write operations.
It is driven by the FPGA control logic, chip-select RTC signal (CSRTC*), which involves
address decoding circuitry (see Section 4.2.1.2).
4.2.1.10
VOLTAGE MONITOR WATCHDOG TIMER
It is important to protect the system during power transitions, and the CPU is reset after
the V
automatic reset output during power up, power down, or brownout conditions. When the
V
CC
and holds the microprocessor in reset for approximately 200 ms after V
threshold. The supervisory circuit includes a chip-select inhibit circuit, which is used to
disable access to the real-time clock’s non-volatile SRAM during power transitions and
power down mode.
This device also provides a watchdog timer function to monitor the activity of the
microprocessor. To service the watchdog timer immediately after reset, the device has a
longer time-out period (1.6 second minimum) right after a reset. The normal time-out
period (70 ms minimum) is effective after the first transition of watchdog input (WDI) after
RESET* is inactive. If the microprocessor does not toggle WDI within the time-out period,
both RESET* and watchdog out (WDO*) outputs are asserted low. The RESET* remains
active low for a minimum of 140 ms and it resets the CPU. The WDO* remains low as long
as the WDI remains either high or low for longer than the watchdog time-out period.
After a reset, the software reads this memory-mapped bit to determine if the latest reset
was a watchdog time-out.
power supply is applied. The microprocessor supervisory circuit generates an
CC
falls below the reset threshold voltage of 2.90 VDC, the reset signal (RESET*) goes low
rises above the
CC
430-95424-001 (Rev. 09/03)4 - 6Plum A+3 Infusion System
Page 36
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.1.11
ANALOG-TO-DIGITAL CONVERTER
The analog-to-digital converter (ADC) monitors the proximal pressure sensor, distal
pressure sensor, proximal air sensor, distal air sensor, battery charge/discharge current,
battery voltage, buzzer test signal, LCD contrast voltage, CCFT test signal, and two chopper
motor drive reference voltages. The ADC is an advanced 10 bit accurate, 11 channel,
switched-capacitor, successive-approximation device. It has three inputs and a three-state
output (chip-select, I/O clock, address input, and data out) that provide a direct four-wire
interface to the serial communication port of the CPU. The ADC is designed to be used in
conjunction with multiple serial devices on a common bus; consequently, the data-out
pin is driven only when the chip-select (CS*) pin is asserted. Figure 4-2, Serial Interface
to ADC, illustrates the serial interface between the ADC and the CPU.
In addition to a high-speed ADC and versatile control capability, this device has an on-chip
14 channel multiplexer that can select any one of 11 analog inputs or any one of three
internal self test voltages. The sample-and-hold function is automatic.
The end-of-conversion (EOC) output goes high to indicate that conversion is complete.
The CPU polls the EOC signal.
Channel selection and conversion results are transferred through the SCP pins. A serial
transfer synchronizing clock (SPCLK) must be fed into the I/O clock input pin when the
CS* pin is driven low. The address to be converted is serially transmitted into the address
pin, and the conversion results are serially shifted out the data-out pin. Typical access
time is 21 microseconds. The air, pressure, and pin (APP) PWA is the source of the 2.5 V
reference voltage.
DC
The analog inputs are selected by the channel multiplexer according to the input address
(see Table 4-1, Analog Inputs). The input multiplexer is a break-before-make type to reduce
input-to-input noise injection resulting from channel switching.
Technical Service Manual4 - 7430-95424-001 (Rev. 09/03)
Page 37
SECTION 4 THEORY OF OPERATION
ANALOG
INPUTS
U4
A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
GND
CS*
I/O CLOCK
ADDRESS
DATA OUT
EOC
REF+
REF-
GDIG
Figure 4-2. Serial Interface to ADC
CPU
I/O
SPCLK
SPTXDSCP
SPRXD
I/O
(OR INTERRUPT)
2.5V
GANA
98K01022
Table 4-1. Analog Inputs
Signal NameAnalog InputAddress (hex)Description
PRPRSA0$00Proximal pressure sensor
DIPRSA1$01Distal pressure sensor
PXAIRA2$02Proximal air sensor
DIAIRA3$03Distal air sensor
IBATTA4$04Battery current
VBATTA5$05Battery voltage
BUZTSTA6$06Buzzer test voltage
LCDTSTA7$07LCD contrast test voltage
TUBTSTA8$08CCFT intensity test voltage
MI_STAA9$09Motor current A control
MI_STBA10$0AMotor current B control
$0B(V
$0CV
ref(+)
ref(-)
- V
ref(-)
) / 2
$0DV
430-95424-001 (Rev. 09/03)4 - 8Plum A+3 Infusion System
ref(+)
Page 38
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.1.12
DIGITAL-TO-ANALOG CONVERTER
The dual 8 bit digital-to-analog converter (DAC) generates two analog signals to control
the phase A and phase B motor coil currents. The interface between the DAC device and
the CPU is the 8 bit data bus, which is connected to DATA15:8. All the control signals for
this DAC are generated by FPGA logic devices. Buffer amplifier/ground compensation
circuits (U6 and U7) condition the DAC outputs.
4.2.1.13
FRONT PANEL KEYPAD MATRIX
A 5 x 5 membrane switch keypad matrix is located on the front panel. The keypad column
lines (COL4:0) are driven by open collector type memory mapped input ports, while the
keypad row lines (ROW4:0), are read by memory mapped input ports. The keypad strobing,
scanning, and switch de-bouncing is accomplished by software. The keypad interface is
designed with ESD protection. Refer to Table 4-2, Keypad Map.
Table 4-2. Keypad Map
COL 0COL 1COL 2COL 3COL 4
Row 4Softkey 1Softkey 2Softkey 3Softkey 4
Row 3Start123[
Row 2Stop456
Row 1789[
Row 0On/OffClear0
4.2.1.14
.
]
]
Silence
FRONT PANEL [ON/OFF] KEY
The [ON/OFF] key on the front panel provides a start up (STRTUP) signal to wake up the
power supply when the system is shutdown. When activated during normal operation,
the [ON/OFF] key interrupts (STRUPD*) the CPU, signaling a request for shutdown.
4.2.1.15
FRONT PANEL LED INDICATORS
The CPU drives the three light emitting diode (LED) indicators embedded in the front panel.
Two memory mapped I/O signals activate the two LED lights used to indicate which
channel is in delivery mode (LEDAE*, LEDBE*). The AC power on LED indicates the status
of AC power (LEDAC) and that the system is in the battery charge mode. A buffered AC
on signal (BACON) drives the LED and is active only when AC power is present.
Technical Service Manual4 - 9430-95424-001 (Rev. 09/03)
Page 39
SECTION 4 THEORY OF OPERATION
4.2.1.16
KEYPAD LOCKOUT INTERFACE
A lockout switch (SW1) on the peripheral/interface PWA locks the front panel keypad for
all three infusers. A memory mapped input port (LOTSW*) reads the switch. The switch
serves as a lockout request and software performs the lockout.
4.2.1.17
NURSE CALL INTERFACE
A nurse call relay switch on the peripheral/interface PWA indicates alarm conditions to
a remote operator. A memory-mapped output signal (NURSE) activates the relay during
alarm conditions. The relay has both normally open and normally closed contacts.
A jumper on the peripheral/interface board selects the contact type. The factory setting
is normally open.
4.2.1.18
AUDIBLE INDICATORS
There are two audible indicators on the CPU subsystem. Three loud, main audible
indicators are mounted on the main chassis, one per infuser. This main alarm is used for
alerting the operator to alarm conditions. A keypad beeper (LS1), with lower power and a
distinctly different tone, is used to provide audible feedback to the operator. The keypad
beeper is driven by a memory-mapped output (KEYALM). It is used to indicate keypad
activation, and confirmation to the operator.
The main alarm has an adjustable volume control on the peripheral/interface PWA (R2),
mounted on the rear of the instrument. The main alarm can be activated by either a
memory-mapped control (MAINALM), the reset pulse(s), or by a power failure alarm latch.
The main alarm will sound a chirp for every reset pulse sent by the watchdog timer IC.
Continuous chirping indicates a stuck processor.
The alarm is activated continuously during power failure. If the control software does not
shut down power in a proper sequence, a latch on the CPU PWA (U2), powered by a backup
supply (0.1 F supercap), will activate a continuous alarm. This continuous alarm sounds
until either the backup supply is discharged or the user resets the latch by pressing the
front panel [ON/OFF] key. Reliable operation of the main alarm is assured by software
monitoring of a buzzer test signal (FBUZTST) via the ADC.
4.2.1.19
BARCODE READER INTERFACE
The CPU communicates with a barcode wand that is connected to the peripheral/interface
PWA from the rear of the infusion device. The barcode wand reads and decodes a Code 128
barcode symbology and outputs the barcode data via an RS-232 port using an
asynchronous, serial ASCII format.The software controls power to the barcode reader and
to the interface circuits via memory-mapped outputs BARPWR and COMPWR*.
The barcode reader is isolated from the main system by an optical data path on the
peripheral PWA (U10, U11, and U13) and an isolated power supply (U3 and T1).
430-95424-001 (Rev. 09/03)4 - 10Plum A+3 Infusion System
Page 40
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.1.20
DATAPORT INTERFACE
The CPU communicates with an external computer by way of a DataPort interface.
The DataPort interface provides for remote monitoring of up to four pumps using a host
computer with a modified RS-232-D serial interface. Pumps are either connected directly
to the host or in a daisy chain configuration using junction boxes that provide a 5 bit hard
ID via DIP switches on the junction box. The DIP switches are buffered (peripheral PWA
U8) and read by the CPU via the memory-mapped input/output (MMIO) port.
The DataPort system conforms to the EIA-232-D standard, with the following exceptions:
-DataPort uses non-standard DB-15 and 6 pin modular connectors in addition
to the standard DB-25 and DB-9 connectors
-With DataPort, more than one pump is allowed on the line
-The minimum line impedance is 2 KΩ (EIA-232-D standard: 3 KΩ min.)
-The maximum line impedance is 30 KΩ (EIA-232-D standard: 7 KΩ max.)
-The maximum line capacitance is 13 nF (EIA-232-D standard: 2,500 pF)
The communications default is 1200 BAUD, no parity, 8 data bits and 1 stop bit.
The Plum A+3 BAUD rate is selectable (1200, 2400, 4800, and 9600). The data format on
the serial port is a 10 bit frame with asynchronous start and stop. The CTS line is held
high and the RTS line is disconnected.
The DataPort is isolated from the main system by an optical data path on the peripheral
PWA (U10, U11, and U13) and an isolated power supply (U3 and T1).
4.2.1.21
POWER SUPPLY INTERFACE
The CPU subsystem interfaces the power supply subsystem by providing the MMIO signals
needed for power control and battery management. Additionally, the CPU subsystem
measures the battery terminal voltage and charge/discharge current via the ADC
(see Table 4-3, CPU-Power Supply Interface).
Table 4-3. CPU-Power Supply Interface
Signal NameTyp eDescription
PWRHLDD, OHolds system power on
STRTUPA, IStartup pulse from the [ON/OFF] key
STRUPD*D, IDigital startup pulse, used as interrupt to the CPU
V3_3P3.3 volt system power
V5_0/VANAP5.0 volt analog and interface power
VMOTPRaw, unregulated charger voltage or battery voltage
V2_7P2.7 volt backup power for RTC and non-volatile SRAM
VSCPFull time 5 volt supply, backed up by supercap
V12_0P12 volt, low current supply for audio alarm
Technical Service Manual4 - 11430-95424-001 (Rev. 09/03)
Page 41
SECTION 4 THEORY OF OPERATION
Table 4-3. CPU-Power Supply Interface
Signal NameTyp eDescription
OVRVLT*D, ISignal that indicates overvoltage, regulation problem
on the power supply main regulator
BACOND, IBuffered AC on signal
IBATTA, IVoltage proportional to integration of battery
charge/discharge current
VBATTA, IDivided battery terminal voltage
CHG*D, OBattery charger enable
VFLOAT*D, OSet the main regulator voltage to battery float charge level
ITGRSTD, OReset the charge current integrator
Legend: P = Power; A = Analog; D = Digital; I = Input; O = Output
4.2.1.22
MECHANISM INTERFACE
The CPU subsystem provides the MMIO ports for interface to the mechanism subsystem,
in addition to the analog interface mentioned in the ADC and DAC sections
(see Section 4.2.1.11 and Section 4.2.1.12). Refer to Table 4-4 forCPU-mechanism interface
signals.
Table 4-4. CPU-Mechanism Interface Signals
Signal NameTypeDescription
MI_STAA, OMotor current set for phase A
MI_STBA, OMotor current set for phase B
GDACA, OGround signal from chopper (for compensation)
M_PHAD, OMotor phase A
M_PHBD, OMotor phase B
M_SEL1, M_SEL0D, OMotor select bits
FLCAMED, OI/O and L/S cam flag sensors enable
FLPINED, OL/S pin motion detectors enable
FLPLED, OPlunger motor sensor pair enable
FLLS_CD, IFlag, L/S valve cam sensor
FLIO_CD, IFlag, I/O valve cam sensor
FLLS_AD, IFlag, L/S valve A pin detector
FLLS_BD, IFlag, L/S valve B pin detector
FLPLROD, IFlag, plunger rotation sensor
430-95424-001 (Rev. 09/03)4 - 12Plum A+3 Infusion System
Page 42
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
Table 4-4. CPU-Mechanism Interface Signals
Signal NameTypeDescription
FLPLTRD, IFlag, plunger translation sensor
PXPRED,OProximal pressure sensor enable
PXPRSA, IProximal pressure sensor
DIPRED, ODistal pressure sensor enable
DIPRSD, ODistal pressure sensor
PXARED, OProximal air sensor enable
PXAIRA, IProximal air sensor
DIARED, ODistal air sensor enable
DIAIRA, IDistal air sensor
CASPR*D, ICassette present
CASS2*, CASS1*, CASSO*D, ICassette type coding: Macro (111), Micro (010),
all others are invalid
SPCLKD, OSCP clock output
SPRXDD, ISCP receive data
SPTXDD, OSCP transmit data
CSSEP*D, OChip select, EEPROM
V5_0P5.0 volt supply for interface power
V3_3P3.3 volt supply for logic power
GDIGPDigital ground
VANAP5.0 volt supply for analog power
GANAPAnalog ground
VMOT, GMOTPMotor power is directly from power supply PWA
V2_5A, IReference voltage for ADC and DAC
Legend: P = Power; A = Analog; D = Digital; I = Input; O = Output
Technical Service Manual4 - 13430-95424-001 (Rev. 09/03)
Page 43
SECTION 4 THEORY OF OPERATION
4.2.2
POWER SUPPLY SUBSYSTEM
The power supply subsystem provides DC power to system circuits and interface software
controlled power and battery management (see Figure 9-11, Power Supply PWA Schematic).
The power supply subsystem provides for the following functions:
-Main switching regulator
-AC power detection
-Main regulator fault detection
-System power (secondary regulators)
-Auxiliary supplies
-Power control
-Battery charging circuitry
-Battery terminal voltage measurement
-Battery charge/discharge current measurement
4.2.2.1
MAIN SWITCHING REGULATOR
The main source of power for the Plum A+3 is the AC line. The main switching regulator
is a pulse width modulated, AC-to-DC converter which provides the system an isolated
DC voltage of 6.9 V
preceded by: line fuses F1 and F2, surge suppressor VR1, and a line filter (T3, T4, C54-56).
The bridge rectifier U14 and capacitors C52 and C53 provide the DC voltage required for
the switching circuit. Voltage regulator U13 provides the pulse width modulator (PWM)
device U12 startup supply voltage. After startup, supply voltage for U12 is supplied by
half wave rectifier circuitry CR14, R76, and C51.
The PWM oscillation frequency is approximately 40 kHz, determined by external resistor
R72 and capacitor C45. U12 controls the power delivered by varying the duty cycle of the
power metal-oxide-semiconductor field-effect transistor (MOSFET) Q9, which drives T2.
A half-wave rectifier (CR9 and C37-C41) rectifies the transformer’s secondary voltage,
which provides the raw DC voltage for the battery charger and system power. There are
three feedback mechanisms that maintain control: a main loop for normal control,
a secondary loop for overvoltage protection, and a current limit loop.
4.2.2.1.1
(or 7.5 VDC in battery charger boost mode). The main regulator is
DC
Main Loop
The main loop uses an optical feedback path to regulate the charger voltage (BATPOS)
at 6.9 V
control of the VFLOAT* line). A shunt regulator and opto-isolator provide feedback to the
PWM error amplifier.
430-95424-001 (Rev. 09/03)4 - 14Plum A+3 Infusion System
(except during boost charge, when the limit is raised to 7.5 VDC by software
DC
Page 44
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.2.1.2
Secondary Loop
Diode CR10 and opto-isolator U10 provide overvoltage protection. CR10 conducts and
activates U10 when secondary voltage exceeds approximately 10 V
U12 is reduced until the excessive voltage is removed.
4.2.2.1.3
. The duty cycle of
DC
Current Limit Loop
The current limit loop is activated when the primary current, sensed by R71, exceeds
3.0 A. Resistor R70 and capacitor C46 filter the voltage across R71 and feed it back to the
current sense input (1.5 V
excessive load is removed.
4.2.2.2
threshold) of U12. The duty cycle of U12 is reduced until the
DC
MAIN REGULATOR FAULT DETECTION
If the switching regulator’s main loop fails, the secondary voltage limit loop takes over.
However, the battery charger and motors must be disabled, and an alarm must be
generated. A comparator is used to monitor the raw DC (+BUSS) for overvoltage. A 3.3 V
logic signal (OVRVLT*) is provided to the CPU subsystem.
DC
4.2.2.3
SYSTEM POWER
Along with the unregulated VMOT supply, a secondary switching regulator provides
system power. The secondary switching regulator includes IC U4, transformer T1, and
transistors Q4 and Q5. The regulator is a triple output, wide supply range, fly-back
converter that provides regulated 3.3 V
winding transformer T1. The regulator operates over an input range of 4 to 10 V
provides output current limit as well as voltage overshoot limit. Primary feedback is
metered through a bias arrangement on transistor Q3. A Schottky rectifier diode CR4
provides feedback in the event of V3_3 or V12_0 failure, and transistor Q10 provides
feedback in the event of V5_0 failure. The positive terminal of the battery provides the raw
DC voltage, VMOT, for the motors and backlight of the display.
4.2.2.4
, 5.0 VDC, and 12.0 VDC outputs from the five
DC
DC
and
AUXILIARY SUPPLIES
The power supply subsystem provides full time 5.0 V
active when battery or AC voltage is present. The full time 5.0 V
linear low dropout voltage regulator U6, whose power source is directly from the battery
and is backed up by a 0.1 F capacitor. VSC is used for the ON/OFF switch and a power
failure alarm latch. The full time 2.7 V
to supply the ultra-low current needed to power the real-time clock and non-volatile SRAM
during shutdown.
supply (V2_7) is derived from VSC and is used
DC
and 2.7 VDC supplies, which are
DC
supply (VSC) uses a
DC
Technical Service Manual4 - 15430-95424-001 (Rev. 09/03)
Page 45
SECTION 4 THEORY OF OPERATION
4.2.2.5
POWER CONTROL
The infusion pump will operate in one of three modes: normal, standby, or shutdown.
During normal operation, the user interface is active and either on battery or AC line
power. During standby mode the user interface is inactive while the CPU is still operating,
servicing the battery management and waiting for a startup interrupt. Shutdown mode is
when system power is off. Shutdown mode only occurs during battery operation; otherwise,
+BUSS holds the system power on.
The infusion pump is activated when the [ON/OFF] key is pressed or the AC line is
plugged in. The [ON/OFF] key activates the STRTUP signal, triggering a three second
one-shot circuit (C3, R10, CR1, and Q1) that will temporarily turn the system power on.
This three second one-shot period allows the CPU enough time to power up, initialize, and
turn on the PWRHLD signal. The CPU monitors the STRTUP signal, via interrupt, to signal
a user request for turning off the infuser.
Figure 4-3, System Startup and Shutdown Timing, Battery Powered illustrates the system
startup/shutdown sequence while battery powered. System power is always on while
AC powered.
STRTUP
3 SEC
ONE-SHOT
V3_3, V5_0, V12V
PWRHLD
98K01021
Figure 4-3. System Startup and Shutdown Timing, Battery Powered
4.2.2.6
BATTERY VOLTAGE MEASUREMENT
The battery terminal voltage (BATPOS - BATNEG) is measured with a differential amplifier
consisting of U1, R1, R2, R4, R7, and R8. It has a gain of 0.317 to generate a single ended
VBATT signal. The VBATT signal is then provided to the CPU A/D converter as input for
the battery management algorithms.
430-95424-001 (Rev. 09/03)4 - 16Plum A+3 Infusion System
Page 46
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.2.7
BATTERY CHARGE/DISCHARGE CURRENT MEASUREMENT
The battery management algorithms measure battery charge/discharge current for battery
capacity estimation and charger control. The charge/discharge current is measured by
integrating the voltage across current sense resistor R57. An operational amplifier
(op-amp) integrator circuit, consisting of U2, C5, R12, R13, R19, and R20, provides a
voltage proportional to the integration of battery current (IBATT) over a CPU controlled
measurement period. The IBATT signal is fed to the CPU A/D converter, where it is sampled
at the end of the measurement period. The battery management algorithm further
accumulates the charge/discharge current for battery capacity estimation. The op-amp
integrator is reset by the CPU system at the beginning of each measurement period by
parallel analog switches U3, controlled by the CPU’s ITGRST signal. The battery
management algorithm periodically calibrates the op-amp integrator.
4.2.2.8
BATTERY CHARGER
The software battery management algorithm controls the battery charger. The charging
scheme is a current limit/two stage voltage limit charger. The charge current is limited to
1.3 A and the voltage is limited to either 6.9 V
or 7.5 VDC.
DC
The source of the charge current is power MOSFET transistor Q7 operating in the linear
mode. Charge current passes through a current sense resistor R57, where it develops a
feedback signal for the charger control amplifier consisting of U7, Q6, and associated
parts. The feedback signal is compared against a 2.5 V
A .5 A fuse (F4) protects against damage due to a short circuit. The battery management
algorithm maintains on/off control of the charger by the charger enable signal CHG*.
When set high, CHG* activates a comparator U7, which overrides the feedback signal and
disables the charger. Excessive voltage on the BATNEG terminal indicates that there is
a shorted battery cell, and will disable the charger through the same comparator.
voltage reference U8.
DC
4.2.3
MECHANISM SUBSYSTEM
The mechanism subsystem includes the electronics and electromechanical components
that interface the Plum A+3 pumping mechanism (see Figure 9-15, Driver PWA Schematic;
Figure 9-16, Switch PWA Schematic; and Figure 9-17, APP PWA Schematic). Refer to
Table 4-4 for mechanism interface signals.
The mechanism subsystem provides the following functions:
-Chopper motor drive for three stepper motors (plunger, L/S valve, I/O valve)
-Four motor position sensors (flag detectors)
-Precision voltage reference
-Two air sensors (distal, proximal)
-Two pressure sensors (distal, proximal)
-Cassette presence and type detection
-Serial electrically erasable PROM (EEPROM)
Technical Service Manual4 - 17430-95424-001 (Rev. 09/03)
Page 47
SECTION 4 THEORY OF OPERATION
4.2.3.1
MOTORS/MOTOR DRIVE
The Plum A+3 infusion system uses three stepper motors for pumping: one for fluid
displacement and two for cassette valve actuation. The stepper motors are driven, under
step-by-step control from software, by a unipolar chopper drive.
4.2.3.1.1
Stepper Motors
Each motor is named by its function:
-Plunger motor for driving the plunger screw
-I/O valve motor for moving the input-output valve pins
-L/S valve motor for moving the line select valve pins A and B
All three motors are four phase stepper types. One electrical revolution is accomplished
after four motor steps (phases) are completed. The step-angle (the number of steps per
shaft revolution) resolutions are 3.6 degrees/step (100 steps/rev) for the plunger motor,
and 7.5 degrees/step (48 steps/rev) for the I/O and L/S valve motors.
The unipolar motor windings have a center tap connected on each of the two coils as shown
in Figure 4-4, Stepper Motor Coils. Unidirectional current enters the center tap and is
steered to one end of the coil or the other end by the driver electronics, creating positive
or negative flux lines in the motor coil. With two coils each with a choice of flux polarity,
four electrical combinations or phases are possible.
A
ACOM
A
BCOM
B B
98K01020
Figure 4-4. Stepper Motor Coils
430-95424-001 (Rev. 09/03)4 - 18Plum A+3 Infusion System
Page 48
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.3.1.2
Chopper Motor Drive
The Plum A+3 stepper motor drive is a chopper drive, which is a pulse width modulation
of the coil current in each motor winding. Current is switched on and off to maintain a
predetermined coil current independent of supply voltage and motor speed. The motor
winding inductance acts as a filter to smooth out the switching currents, slowing the
current rise when turned on and storing a decaying current when turned off.
Each motor coil is modulated independently, allowing different coil currents in the two
motor windings. The coil current is sensed and compared to a reference input for each
winding. Modulation circuits correct for any error between the sensed current and the
reference. This reference input can be changed to set a different coil current.
4.2.3.2
MOTOR POSITION SENSORS
Motor position is estimated by counting the motor steps, relative to a position reference.
Optical switches and flags serve as position references, which are used to find the motor
home positions and to verify proper motion. Flag positions are anticipated by software.
Optical switch flag sensors are used for tracking the following:
-Plunger motor rotational position (coupler flag)
-Plunger translational (linear) position
-I/O valve motor rotational position (cam flag)
-L/S valve motor rotational position (cam flag)
Each optical switch consists of an infrared LED, which shines through a rectangular
aperture, across a slot, to illuminate a photo-transistor. The photo-transistor is activated
as long as the beam is on and not blocked (by a flag in the slot). The optical switches are
distributed throughout the mechanism, near their associated flags. The motor rotational
optical switches (U5, U9, and U10) are mounted on the driver PWA along with the control
circuitry. The plunger translational optical switch is mounted remotely on the switch PWA.
The switches are used intermittently to save power. There are two control signals that
enable associated switch pairs:
-FLCAME flag, valve motor cam sensor enable
-FLPLE flag, plunger motor rotation and translation sensors enable
Each of these control signals enables a constant current source which turns on the
associated switch’s infrared LEDs. The photo transistor states are sensed by Schmidt
trigger inverters (U11 on driver PWA) which provide a 3.3 volt logic high when the optical
path is blocked or a logic low when the optical path is clear. The Schmidt trigger output
is high when the sensor is disabled. The following output signals are provided to the CPU
subsystem:
-FLIO_C flag I/O valve motor cam sensor
-FLLS_C flag L/S valve motor cam sensor
-FLPLRO flag plunger motor rotation sensor
-FLPLTR flag plunger motor transition sensor
Technical Service Manual4 - 19430-95424-001 (Rev. 09/03)
Page 49
SECTION 4 THEORY OF OPERATION
4.2.3.3
V2_5 REFERENCE VOLTAGE
A precision 2.50 VDC reference voltage is generated on the APP PWA for use by the pressure
sensor excitation circuits, the air sensor amplifier circuits, and the ADC and DAC reference
voltage. The precision 2.50 V
reference (U1) is buffered by a voltage follower (U3).
DC
The signal name is V2_5.
4.2.3.4
AIR SENSORS
The mechanism subsystem includes two air sensors, used to detect air passage into
(proximal) or out of (distal) the cassette. Both sensors are piezoelectric crystal transmitter
receiver pairs. Liquid between the transmitter and receiver will conduct the ultrasonic
signal, while air will not (see Figure 4-5, Air Sensor Block Diagram).
VSO
CPU
XTL
TX
VCO/PLL
A/D
G_TX
OUTPUT
AMP
Figure 4-5. Air Sensor Block Diagram
XTL
RX
G_RX
PEAK
DETECTOR
98K01019
430-95424-001 (Rev. 09/03)4 - 20Plum A+3 Infusion System
Page 50
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.3.4.1
Transmitter Circuitry
The transmitter circuitry consists of a voltage sweep oscillator, a voltage-controlled
oscillator (VCO), and a transmitter amplifier, and are located on the APP PWA.
The voltage sweep oscillator circuit (U10B, R24, C12, and part of U9) oscillates at
approximately 12 kHz at 50 percent duty cycle. The output of the sweep oscillator is
between +2 V
sensor’s peak coupling frequency, which is between 3.0 and 6.0 MHz. A resistor and
capacitor (R28 and C13) are used to configure the VCO center frequency. The VCO is
enabled when the CPU asserts either DIARE or PXARE control signals.
The transmitter amplifier consists of a push-pull, emitter-follower, complementary pair
of transistors (Q15 and Q16). The transmitter amplifier drives both proximal and distal
sensors simultaneously.
4.2.3.4.2
and +3 VDC, and is used to sweep the VCO. The VCO sweeps through the
DC
Receiver Circuitry
When the cassette’s test port is filled with fluid, the transmitted signal will be coupled to
an identical piezoelectric crystal, where it is amplified and detected by the receiver
circuitry. The receiver circuitry consists of an amplifier, a peak detector, and an adjustable
gain buffer stage. There is a separate, symmetrical receiver circuit for each channel
(proximal and distal). Component references (called out in this design description) will be
made to the distal channel only.
The first amplifier includes two, directly coupled common emitter stages (Q5 and Q7),
biased from the V2_5 supply. DIARE and PXARE are used to enable the distal and proximal
sensors, respectively.
The detector stage consists of an emitter follower (Q3), charging a 400 microsecond time
constant, refreshed every 40 microseconds (twice per VCO sweep).
The peak detector output is buffered by an op-amp (U7) configured as a basic non-inverting
amplifier with a trimming potentiometer (R31) for gain adjustment. Each sensor has an
independent gain adjustment. The two air sensor, gain-trimming potentiometers are
accessible for calibration in an assembled mechanism.
These final signals are read by the CPU subsystem via the ADC.
-PXAIRProximal air sensor output
-DIAIR Distal air sensor output
4.2.3.5
PRESSURE SENSORS
The mechanism subsection contains two strain gauge-type pressure sensors, one at the
proximal and the other at the distal cassette ports. Electrically, the strain gauge is a
Wheatstone bridge made of four strain gauge resistors. When the bridge is electrically
excited, the bridge will output a millivolt level signal proportional to the applied pressure.
The output signal is amplified and offset adjusted before being read by the ADC.
Each pressure sensor circuit includes an excitation voltage supply, sensor amplifiers,
and a low pass filter.
Technical Service Manual4 - 21430-95424-001 (Rev. 09/03)
Page 51
SECTION 4 THEORY OF OPERATION
The pressure sensor circuitry is on the APP PWA. Each of the two channels has an identical
topology, but different gain and filter response. A block diagram of this circuit is shown
in Figure 4-6, Pressure Sensor Excitation and Amplifier Block Diagram.
Note: Component references are made to the distal channel only.
PRESSURE
SENSOR
ENABLE
(FROM CPU)
BRIDGE
EXCITATION
3.75 V
REFERENCE
2.5 V
WHEATSTONE
DIFFERENTIAL
AMPLIFIER
AND OFFSET
ADJUST
OUTPUT
AMPLIFIER
AND FILTER
BRIDGE
PRESSURE
SIGNAL OUTPUT
(TO CPU)
98G01018
Figure 4-6. Pressure Sensor Excitation and Amplifier Block Diagram
4.2.3.5.1
Bridge Excitation Supply
The bridge excitation voltage is 3.75 VDC, and is derived from the 2.5 VDC reference signal
(V2_5), gained 1.5 times by amplifier (U8A, Q13). The CPU subsystem may independently
enable power to each pressure sensor bridge.
These enable signals are active high 3.3 volt logic level inputs:
-PXPREProximal pressure sensor enable
-DIPREDistal pressure sensor enable
430-95424-001 (Rev. 09/03)4 - 22Plum A+3 Infusion System
Page 52
4.2 ELECTRONIC SUBSYSTEM OVERVIEW
4.2.3.5.2
Amplifier and Low Pass Filter
The pressure sensor amplifiers include a high gain differential pre-amplifier (U4), followed
by a second stage non-inverting amplifier (U6B) with low gain. A trimming potentiometer
(R48) is adjusted to minimize any offset in the impedance of the bridge.
A two-pole filter is used to filter the pressure signals. The first pole is formed by a capacitor
(C39, multiplied by 230 due to Miller effect) and a Thevenin resistance (seen at U4-2).
The second pole is the RC filter at the ADC input, which is located on the CPU PWA.
These output signals to the A/D converter in the CPU PWA are:
-PXPRSProximal pressure signal
-DIPRSDistal pressure signal
4.2.3.6
PRESSURE SENSOR CALIBRATION
Pressure sensors are calibrated for offset and gain during mechanism calibration.
A trimming potentiometer is used to adjust the initial, zero pressure offset. The proximal
and distal pressure sensors have independent offset adjustments.
The final system gain (cassette pressure to corrected amplifier output) is adjusted in
software. During mechanism calibration, each channel’s gain (amplifier output/cassette
pressure) will be measured, and stored in the serial EEPROM on the driver PWA.
4.2.3.7
CASSETTE TYPE/PRESENCE SELECTION
The mechanism subsystem includes four force sensing resistor (FSR) switches, which are
coupled to the cassette. Three FSRs are used for cassette type decoding and one is used
for cassette present detection.
The FSR is a polymer thick film device, which exhibits a decrease in resistance with any
increase in force applied to the active surface. The FSRs have a resistance that is either
very large (> 1 MΩ) or relatively small (< 100 KΩ). The large resistance is defined as a logical
‘0’, and the small resistance is defined as logical ‘1’. Each FSR is arranged in a voltage
divider configuration with a fixed resistor, followed by a comparator with hysteresis.
The comparator circuits are located on the CPU PWA. The comparators (CPU PWA: U8 and
associated passives) are designed to trip as the FSR’s resistance falls below 120 KΩ.
4.2.3.8
SERIAL EEPROM
The driver PWA holds the 8 K x 8 bit, serial EEPROM, which is used to store event, alarm,
malfunction, and calibration data specific to the pumping mechanism. It is accessed
through a serial peripheral interface (SPI) compatible interface, which is a high-speed serial
interface to the CPU. The CPU PWA accesses this device through its SCP serial interface.
This interface is a subset of the SPI, and consists of clock (SPCLK), data in (SPRXD),
and data out (SPTXD) pins. This device is in the driver PWA to allow the calibration data
to stay with the mechanism.
Technical Service Manual4 - 23430-95424-001 (Rev. 09/03)
Page 53
SECTION 4 THEORY OF OPERATION
4.3
PRINTED WIRING ASSEMBLIES
The Plum A+3 electronics are packaged into six printed wiring assemblies (PWA) and
several remote mounted peripherals (see Section 4.4, Remote Mounted Peripherals).
The following sections provide a brief description of the functional interfaces of each PWA.
4.3.1
POWER SUPPLY PWA
The power supply PWA (see Figure 9-11) contains the following functions of the power
supply subsystem:
-Main switching regulator
-AC power detection
-Main regulator fault detection
-System power
-Auxiliary supplies
-Power control
-Battery management
The power supply PWA is a four-layer printed wiring board (PWB), with primarily surface
mount technology (SMT) components. The board is fully testable from the bottom side.
An insulating tape covers the back of the power supply PWA. Open system troubleshooting
should be done under battery power. If connection to the AC line is required, an isolation
transformer should be used since AC line potentials are present on the power supply PWA.
See Section 4.2.2 for a functional description, and refer to Table 4-5 for power supply PWA
interface connections.
Table 4-5. Power Supply PWA Interface Connections
ConnectorTypeInterface
P230 pin receptacleBoard-to-board connection to CPU PWA
J164 pin headerMotor power connection to driver PWA
J213 pin receptacleAC power cord connection
J222 pin headerBattery cable connection
430-95424-001 (Rev. 09/03)4 - 24Plum A+3 Infusion System
Page 54
4.3 PRINTED WIRING ASSEMBLIES
4.3.2
PERIPHERAL PWA
The peripheral PWA (see Figure 9-13) contains part of the CPU subsystem circuitry,
including system program and data memories (PROM and SRAM), and external
communication interface circuits. The peripheral PWA is designed to be field replaceable,
to facilitate software upgrades or additional external interfaces. The peripheral PWA is
a four-layer PWB, including one ground plane, one power plane, and two signal layers.
In its initial configuration, all of the components are mounted on the top side.
See Section 4.2.1 for a functional description, and refer to Table 4-6 for peripheral PWA
interface connections.
Table 4-6. Peripheral PWA Interface Connections
ConnectorTypeInterface
P196 pin receptacleBoard-to-board connection to CPU PWA
J2615 pin D-subDataPort
J279 pin D-subBarcode reader connection
J283 pin phone jackNurse call jack
4.3.3
PERIPHERAL/INTERFACE PWA
The peripheral/interface PWA (see Figure 9-12) contains part of the CPU subsystem
circuitry, including system program and data memories (PROM and SRAM), external
communication interface circuits, and rear instrument user controls. The peripheral/
interface PWA has internal logic to control selection of the barcode reader and lock out
the selection of the barcode reader by another infuser until the operator completes the
selection in process.
Refer to Table 4-7 for peripheral/interface PWA interface connections.
P196 pin receptacleBoard-to-board connection to CPU PWA
J29 and J3050 pin plugBoard-to-board connection to peripheral PWA
J2615 pin D-subDataPort
J279 pin D-subBarcode reader connection
J283 pin phone jackNurse call jack
Technical Service Manual4 - 25430-95424-001 (Rev. 09/03)
Page 55
SECTION 4 THEORY OF OPERATION
4.3.4
CPU PWA
The CPU PWA (see Figure 9-14) contains most of the CPU subsystem functions, with the
exception of main memory and communications ports, which are located on the peripheral
PWA. The CPU PWA also accommodates system interconnect. The CPU PWA is an
eight-layer PWB, with one ground plane, one power plane, and six signal layers. The CPU
PWA primarily contains SMT components. Most of the components are on the top side,
while the bottom side holds wave-solder compatible SMT resistors and capacitors.
See Section 4.2.1 for a functional description, and refer to Table 4-8 for CPU PWA interface
connections.
Table 4-8. CPU PWA Interface Connections
ConnectorTypeInterface
J796 pin headerConnection to peripheral PWA (CPU bus, rear
The driver PWA (see Figure 9-15) contains the mechanism subsystem’s motor drive
circuitry, motor position sensors, and serial EEPROM. The driver PWA is mounted in the
mechanism sub-chassis. The driver PWA is a four-layer PWB, with one ground plane, one
power plane and two signal layers. The driver PWA primarily uses SMT components. Most
of the components are located on the top side of the board, while the bottom side holds
wave-solder compatible resistors and capacitors.
See Section 4.2.3 for a functional description, and refer to Table 4-9 for driver PWA interface
connections.
430-95424-001 (Rev. 09/03)4 - 26Plum A+3 Infusion System
Page 56
4.3 PRINTED WIRING ASSEMBLIES
Table 4-9. Driver PWA Interface Connections
ConnectorTypeInterface
J76 pin headerPlunger motor
J86 pin headerInput/output motor
J96 pin headerLine select motor
J1020 pin SMTFlat flex cable to APP PWA
J1150 pin headerRibbon cable to CPU PWA
J126 pin SMTFSR flex circuit
J134 pin headerMotor power, from power supply PWA
J148 pin SMTFlat flex cable to switch PWA
4.3.6
SWITCH PWA
The switch PWA (see Figure 9-16) contains the plunger translation position sensor, which
is one of four position sensors in the system. The switch PWA is located at the side of the
mechanism sub-chassis, and connects to the driver PWA (see Section 4.2.3.2).
4.3.7
APP PWA
The APP (air, pressure, and pin) PWA (see Figure 9-17) is mounted in the mechanism
sub-chassis. The APP PWA contains the following mechanism subsystem circuitry:
-Proximal and distal air sensors and circuitry
-Proximal and distal pressure sensor amplifiers and excitation
-V2_5 precision voltage reference
-Pin detector optical switch module
The APP PWA is a four-layer PWB, with one ground plane, one power plane and two signal
layers. The APP PWA uses SMT components, mounted on both sides of the board. The air
sensors and the pin detector module are board mounted.
See Section 4.2.3 for a functional description, and refer to Table 4-10 for APP PWA interface
connections.
Table 4-10. APP PWA Interface Connections
ConnectorTypeInterface
J1520 pin SMTFlat flex cable to driver PWA
J1110 pin SMTPressure sensor connector
Technical Service Manual4 - 27430-95424-001 (Rev. 09/03)
Page 57
SECTION 4 THEORY OF OPERATION
4.4
REMOTE MOUNTED PERIPHERALS
The following sections describe the major remote mounted peripherals.
4.4.1
LCD
The infusion pump uses a graphic LCD module with a CCFT. The CCFT provides a
backlight source for the LCD. The LCD requires a nominal -16 V
control, which is controlled by the CPU. The pump’s graphic display data is shifted out to
the LCD by the CPU LCD controller, which interfaces directly with the CPU
(see Section 4.2.1.6). The display is configured as a 240 x 240 dot matrix with a viewing
angle of approximately 60 degrees.
4.4.2
SEALED LEAD ACID BATTERY
supply for contrast
DC
The infusion pump uses a nominal 6.0 VDC rechargeable sealed lead acid battery with 4.0
amp-hour capacity.
4.4.3
BARCODE READER WAND
The barcode reader (BCR) wand connects to the BCR port J27 on the peripheral/interface
PWA (see Figure 9-12). A PLD serving as a serial multiplexer installed in the peripheral/
interface PWA determines which one of the three systems can use the BCR. The BCR wand
interfaces through the infusion pump’s optically isolated, TTL logic level, asynchronous
interface. The BCR wand is also capable of interfacing at RS-232 levels. The infusion pump
provides an isolated +5.0 V
of the BCR wand is swiped across a barcode label, the reflected light is scanned and
processed. After a successful scan, the data is sent over the communication interface to
the CPU.
regulator to power the BCR wand. When the LED at the tip
DC
4.5
MECHANICAL OVERVIEW
The principal mechanical elements of the infusion pump include the cassette and
the mechanism assembly. When a cassette is locked into the operating position and the
[ON/OFF] switch is pressed, the infusion pump performs a self test to verify the integrity
of the internal systems. The operation of the mechanism assembly moves a plunger,
causing a pumping action. A valve motor selects the A or B valve, depending on the
command. An additional valve motor alternately opens and closes an inlet valve and outlet
valve to control fluid flow through the cassette pumping chamber.
The following sections detail the cassette and the mechanism assembly.
430-95424-001 (Rev. 09/03)4 - 28Plum A+3 Infusion System
Page 58
4.5 MECHANICAL OVERVIEW
4.5.1
CASSETTE
The cassette operates on a fluid displacement principle to volumetrically deliver fluid
(see Figure 4-7, Major Elements of the Dual-Channel Cassette and Figure 4-8, Fluid Path
in the Cassette). Refer to the system operating manual for a description of the major
cassette functions.
The pumping cycle begins when the outlet valve is opened and the inlet valve is closed.
The plunger extends to deflect the cassette diaphragm and expel fluid. At the end of the
pumping stroke, the outlet valve is closed, the inlet opens, the appropriate A or B valve
opens, and the plunger retracts to allow fluid to refill the pumping chamber. After the
pumping chamber is filled, the inlet and outlet valves are reversed, the A and B valves are
closed, and the cycle repeats.
The cassette contains two chambers: an upper air trap chamber and a pumping chamber.
The two chambers are separated by an inlet valve (see Figure 4-7 and Figure 4-8)
and operate together to detect air. The air trap chamber receives fluid from the intravenous
(IV) container through either the A or B valve. The air trap chamber collects air bubbles
from the IV line and container to prevent them from entering the pumping chamber and
can collect a substantial amount of air.
A proximal air-in-line sensor (bubble detector) is located between the A/B valves and the
upper air-trap chamber. The proximal air-in-line sensor detects air entering the upper
air-trap chamber and initiates an audible alarm if the predetermined air collection
threshold is exceeded. Similarly, a second air-in-line sensor located distal to the pumping
chamber initiates an audible alarm if a predetermined amount of air is detected.
The infusion pump expels air from the cassette.
The pumping chamber receives fluid from the upper air-trap chamber through an inlet
valve. A pressure sensor located in the upper air-trap chamber monitors pressure on the
proximal side of the cassette. When the diaphragm covering the pumping chamber is
deflected by the plunger, the pumping chamber expels fluid through an outlet valve.
A pressure sensor located distal to the pumping chamber monitors pressure on the distal
side of the cassette.
A flow regulator is incorporated into the cassette distal end. This flow regulator is used to
manually control flow when the cassette is not inserted in the infusion pump. When the
cassette is properly inserted into the pump and the door is closed, a mechanism opens
the flow regulator to allow the pump to control fluid flow. When the door is opened,
the same mechanism closes the flow regulator to disable fluid flow.
Technical Service Manual4 - 29430-95424-001 (Rev. 09/03)
Page 59
SECTION 4 THEORY OF OPERATION
SECONDARY PORT
PRESSURE SENSOR
RIGHT VIEW
(Y-RESEAL OR
LOCKING CAP)
B VALVE
AIR-IN-LINE
SENSOR
(PROXIMAL)
(PROXIMAL)
AIR TRAP
CHAMBER
INLET VALVE
FROM PRIMARY
CONTAINER
REAR VIEW
PRECISION GRAVITY
(CONTROL NOT SHOWN)
A VALV E
AIR-IN-LINE
SENSOR
(DISTAL)
PRESSURE SENSOR
(DISTAL)
OUTLET VALVE
PUMPING CHAMBER
FLOW REGULATOR
FINGER
GRIP
OUTLET
TO
PATIENT
LEFT VIEW
98G01017
Figure 4-7. Major Elements of the Dual-Channel Cassette
430-95424-001 (Rev. 09/03)4 - 30Plum A+3 Infusion System
Page 60
LINE A
4.5 MECHANICAL OVERVIEW
A VALV E
AIR TRAP CHAMBER
INLET VALVE
OUTLET VALVE
LINE B
B VALVE
AIR-IN-LINE SENSOR (PROXIMAL)
PRESSURE SENSOR (PROXIMAL)
PUMPING CHAMBER
PRESSURE SENSOR (DISTAL)
AIR-IN-LINE SENSOR (DISTAL)
PRECISION GRAVITY
FLOW REGULATOR (AND SHUT OFF)
98G01001
Figure 4-8. Fluid Path in the Cassette
4.5.2
MECHANISM ASSEMBLY
The mechanism assembly is a fully self-contained unit consisting of the motor and valve
assemblies, A/B valve subsystem, inlet/outlet valve subsystem, plunger drive subsystem,
air bubble (ultrasonic) sensor assemblies, cassette door, and pressure sensor assemblies.
The motor and valve assemblies, A/B valve subsystem, inlet/outlet valve subsystem,
and plunger drive subsystem are detailed in the following sections.
During pump operation, the mechanism assembly plunger motor drives a lead screw that
is coupled to the plunger. The motor action and lead screw move the plunger forward to
cause the delivery of approximately 0.33 mL of fluid per cycle. The plunger motion is
synchronized to the valve motors to provide controlled fluid delivery.
4.5.2.1
MOTOR AND VALVE ASSEMBLIES
The mechanism assembly pumping action is controlled by three stepper motors. The first
stepper motor, in conjunction with an associated valve assembly, activates the A or the B
valve of the cassette, depending on the command. The second stepper motor alternately
opens and closes the inlet and outlet valve to control fluid delivery through the cassette
pumping chamber. A third stepper motor controls plunger movement.
Technical Service Manual4 - 31430-95424-001 (Rev. 09/03)
Page 61
SECTION 4 THEORY OF OPERATION
4.5.2.2
A/B VALVE SUBSYSTEM
The A/B valve subsystem includes a motor designed to rotate a cam (see Figure 4-9,
Mechanism Valve Pins and Sensor Locations). When the cam is positioned at
top-dead-center (home position), both valves are closed. Clockwise rotation (when viewed
from the motor side) from the home position opens the A valve, while the B valve remains
closed. Counterclockwise rotation opens the B valve, while the A valve remains closed.
The A/B valve subsystem consists of a stepper motor with attached cam and integral cam
flag, A and B rockers and valve pins, and a pin detector assembly. The cam flag passes
through an interrupter module as it rotates with the cam. Valve home position is
determined by this cam flag/interrupter module combination through predetermined
factory calibration data. During operation, if the cam flag passes through the interrupter
module at the incorrect time sequence, a motor phase loss is detected. The rocker is the
connecting link between the cam and the valve pin.
A VALV E
REGULATOR ACTUATOR
AIR-IN-LINE SENSOR
(DISTAL)
PRESSURE SENSOR
(DISTAL)
CASSETTE LOCATOR
OUTLET VALVE
Figure 4-9. Mechanism Valve Pins and Sensor Locations
PLUNGER
INLET VALVE
B VALVE
AIR-IN-LINE SENSOR
(PROXIMAL)
PRESSURE SENSOR
(PROXIMAL)
FORCE SENSING
RESISTOR
02K01023
430-95424-001 (Rev. 09/03)4 - 32Plum A+3 Infusion System
Page 62
4.5 MECHANICAL OVERVIEW
4.5.2.3
INLET/OUTLET VALVE SUBSYSTEM
The inlet/outlet valve subsystem is similar in function and build to the A/B valve
subsystem (see Section 4.5.2.2).
4.5.2.4
PLUNGER DRIVE SUBSYSTEM
The main components of the plunger drive subsystem are: plunger, lead screw and coupler,
and stepper motor. When the pump is turned on, the plunger moves from the retracted,
PARK position to the HOME position. The cassette diaphragm is engaged. The stepper
motor rotates approximately 1 2/3 revolutions per pump cycle to permit a 0.33 mL fluid
displacement every pump cycle. The stepper motor then reverses and the plunger returns
to HOME position. This cycle repeats for the duration of fluid administration.
The screw/coupler assembly links the motor and the plunger. This assembly includes
a flag that passes through an interrupter module. This screw/coupler, flag/interrupter
module combination is used in conjunction with predetermined factory calibration data
to determine the plunger position. During operation, if the screw/coupler flag passes
through the interrupter module at the incorrect time sequence, a motor phase loss
is detected.
Technical Service Manual4 - 33430-95424-001 (Rev. 09/03)
Page 63
SECTION 4 THEORY OF OPERATION
This page intentionally left blank.
430-95424-001 (Rev. 09/03)4 - 34Plum A+3 Infusion System
Page 64
Section 5
MAINTENANCE AND SERVICE
TESTS
A complete maintenance program promotes infusion pump longevity and trouble-free
instrument operation. Such a program should include routine maintenance, periodic
maintenance inspection, and following any repair procedure, performance verification
testing.
5.1
ROUTINE MAINTENANCE
Routine maintenance consists of basic inspection and cleaning procedures. As a minimum
requirement, inspect and clean the infusion pump after each use. In addition, establish
a regular cleaning schedule for the infusion pump.
5.1.1
INSPECTING THE INFUSION PUMP
Inspect the infusion pump periodically for signs of defects such as worn accessories,
broken instrument connections, or damaged cables. In addition, inspect the infusion
pump after repair or during cleaning. Replace any damaged or defective external parts.
See Section 5.2.2, Inspection, for a detailed list of areas to be inspected.
5.1.2
CLEANING THE INFUSION PUMP
The following procedures are designed to maintain the infusion pump, sustain system
longevity, and promote trouble-free instrument operation.
Follow hospital protocol for establishing the infusion pump cleaning schedule.
WARNING:
CAUTION: Do not immerse the infusion pump in liquids. Immersion could damage
the instrument. Do not allow liquids to enter the infusion pump electronics
compartment.
DISCONNECT THE INFUSION PUMP FROM AC POWER PRIOR TO
CLEANING THE INSTRUMENT. FAILURE TO COMPLY WITH THIS
WARNING COULD RESULT IN ELECTRICAL SHOCK.
CAUTION: Do not spray cleaning solutions toward any openings in the infusion
pump.
Technical Service Manual5 - 1430-95424-001 (Rev. 09/03)
Page 65
SECTION 5 MAINTENANCE AND SERVICE TESTS
CAUTION: Certain cleaning and sanitizing compounds may slowly degrade
components made from some plastic materials. Using abrasive cleaners or cleaning
solutions not recommended by Abbott Laboratories may result in product damage
and, potentially, void the product warranty. Do not use compounds containing
combinations of isopropyl alcohol and dimethyl benzyl ammonium chloride.
CAUTION: Clean the exposed surfaces of the infusion pump with a soft, lint-free
cloth dampened with one of the cleaning solutions listed in Table 5-1, CleaningSolutions, or a mild solution of soapy water. Remove soap residue with clear water.
Do not use solvents that are harmful to plastic, such as isopropyl alcohol or acetone.
Do not use abrasive cleaners.
CAUTION: To avoid infusion pump damage, cleaning solutions should be used only
as directed in Table 5-1. The disinfecting properties of cleaning solutions vary;
consult the manufacturer for specific information.
Table 5-1. Cleaning Solutions
Cleaning SolutionManufacturerPreparation
Coverage
TM
HBSteris CorporationPer manufacturer's
recommendation
Dispatch
TM
Caltech IndustriesPer manufacturer's
recommendation
Formula C
TM
JohnsonDiverseyPer manufacturer's
recommendation
Manu-Klenz
®
Steris CorporationPer manufacturer's
recommendation
Precise
TM
Caltech IndustriesPer manufacturer's
recommendation
Sporicidin
®
Sporicidin InternationalPer manufacturer’s
recommendation
Household bleachVariousPer hospital procedures;
do not exceed one part
bleach in ten parts water
5.1.3
SANITIZING THE INFUSION PUMP
Sanitize the external surfaces of the infusion pump using a cleaning solution listed
in Table 5-1.
Note: Not all cleaning solutions are sanitizers. Check product labeling.
CAUTION: Do not sterilize the infusion pump using heat, steam, ethylene oxide
(ETO), or radiation. These methods may cause the instrument to malfunction.
430-95424-001 (Rev. 09/03)5 - 2Plum A+3 Infusion System
Page 66
5.2 PERFORMANCE VERIFICATION TEST
5.2
PERFORMANCE VERIFICATION TEST
The Performance Verification Test (PVT) consists of the tests described in the following
sections. The PVT can be used for diagnostic purposes during the troubleshooting of
a malfunctioning infusion pump. The PVT should be used for performance verification
before an infusion pump is placed back in service after repair. If any malfunction
is detected as a result of the PVT, refer to Table 6-3, Troubleshooting with the PVT.
Note: The PVT must be performed exactly as described in this manual to assure
effective and reliable product evaluation information.
Note: When performing the PVT, all lines must be tested. However, if appropriate,
the test may be performed on all lines concurrently.
5.2.1
EQUIPMENT REQUIRED
The PVT requires the following equipment (or equivalents):
❏ Graduated cylinder, 25 mL, with 0.2 mL graduations (Type A)
❏ Sterile water or tap water in an IV bag/container
❏ Digital pressure meter (DPM), 0 to 50 psi (Fluke
❏ Three-way stopcock, latex-free (List No. 3233-01, or equivalent)
❏ IV Set (List No. 11419, or equivalent)
❏ 21-gauge butterfly needle, latex-free (List No. 4492-01, or equivalent),
or 18-gauge blunt cannula
❏ Safety analyzer (Fluke Biomedical 232D)
❏ Stopwatch
❏ Digital multimeter (DMM), (Fluke Biomedical 8012A) (optional)
❏ Barcode directory (optional)
❏ Nurse call test cable (P/N 561-88416-001, or equivalent) (optional)
®
Biomedical DPM3)
Technical Service Manual5 - 3430-95424-001 (Rev. 09/03)
Page 67
SECTION 5 MAINTENANCE AND SERVICE TESTS
5.2.2
INSPECTION
Inspect the infusion pump periodically for signs of defects such as worn accessories
or damaged cables. Also, inspect the infusion pump after repair or during cleaning.
Replace any damaged or defective external parts.
Inspect the following areas for missing or damaged parts:
-Labels
-AC power cord
-Velcro
-Rubber foot pads
-Door assembly and handle
-Keypad
®
retainer straps
-External screws
-Pole clamp assembly
-Front and rear enclosures
-Battery access cover
-LCD
-LEDs
5.2.3
TEST SETUP
WARNING:
To set up the infusion pump for the PVT, proceed as follows:
1. Confirm the infusion pump and appropriate accessories are assembled.
2. Hang two sterile water containers at a height of 18 ± 6 inches (46 ± 15.3 cm) above
the pumping chamber of the device.
3. Connect the infusion pump to AC power. Conduct all tests with the pump connected
to AC power unless otherwise specified.
4. Press [ON/OFF] to turn on the pump.
5. Verify the infusion pump is in the unlocked mode. Toggling the [LOCKOUT] switch
alternates between unlocked [DOWN] and locked [UP] modes.
6. Press [ON/OFF] to turn off the pump.
A PATIENT SHOULD NEVER BE CONNECTED TO THE INFUSION
PUMP DURING DEVICE TESTING.
5.2.4
SELF TEST
CAUTION: Do not place the infusion pump in service if the self test fails.
Note: If an alarm condition occurs during the self test, cycle the power and repeat
the self test. If the alarm condition recurs, note the message and take corrective action
(see Section 6, Troubleshooting). Repeat the self test. If the alarm condition recurs,
remove the Plum A+3 infusion system from service and contact Abbott Laboratories.
Note: Screen representations are examples only, and do not necessarily reflect the
most current software version.
430-95424-001 (Rev. 09/03)5 - 4Plum A+3 Infusion System
Page 68
5.2 PERFORMANCE VERIFICATION TEST
To perform the self test, refer to Figure 5-1, LCD and Keypad, and proceed as follows:
1. Connect the AC power cord to a grounded AC outlet. Verify the charging/line
indicator CHARGE illuminates and an alarm beep sounds.
2. Without a cassette installed, press [ON/OFF] to turn on the pump.
3. The LCD screen briefly displays the SELF TEST screen. Verify the screen display
matches Figure 5-1.
Note: If the SELF TEST screen does not appear, contact Abbott Laboratories.
4. After the self test is complete, the message “INSERT PLUM SET CLOSE LEVER”
appears. Press the decimal [.] key, then [START].
5. Using the [SELECT] key, select Set Time and Date, and press the [Choose] softkey.
6. Verify the time and date. To set the time and date, refer to Section 1.8.3, Setting the
Time and Date.
7. Press [ON/OFF] to exit the SET TIME AND DATE screen.
8. Press [ON/OFF] to turn the pump back on.
9. Open the cassette door and insert a primed cassette. Close the cassette door.
The cassette test is complete when the “CASSETTE TEST IN PROGRESS” message
disappears.
Note: The message “MECHANISM INITIALIZATION IN PROGRESS” may briefly
appear prior to the “CASSETTE TEST IN PROGRESS” message.
10. If previously entered programming exists, the “CLEAR SETTINGS?” message
appears. Press the [Yes] softkey to clear the settings.
Technical Service Manual5 - 5430-95424-001 (Rev. 09/03)
Page 69
SECTION 5 MAINTENANCE AND SERVICE TESTS
ABBOTT Plum A+
A
LINE FLOW
INDICATORS
B
STATUS
REGION
MESSAGE
REGION
SOFTKEY
LABEL REGION
LINE
INDICATOR
START
STOP
CHARGE
ON OFF
Release XX.X - MM/DD/YY
Copyright Abbott Laboratories
2002
System Self Test
ogress
In Pr
1
4
7
CLEAR
2
5
8
0
3
6
9
.
WORKING
REGION
SOFTKEYS
SELECT
KEYPAD
SILENCE
98G01002
Figure 5-1. LCD and Keypad
5.2.5
CASSETTE ALARM TEST
To perform the cassette alarm test, proceed as follows:
1. Verify the infusion pump is on. Insert an empty cassette and close the door.
2. Verify the “CASSETTE TEST FAIL” message is flashing on the display and the alarm
sounds after the cassette test is complete.
3. Open the door and remove the cassette.
4. Press [ON/OFF] to turn off the pump.
430-95424-001 (Rev. 09/03)5 - 6Plum A+3 Infusion System
Page 70
5.2 PERFORMANCE VERIFICATION TEST
5.2.6
FREE FLOW TEST
To perform the free flow test, proceed as follows:
1. With a primed cassette installed, press [ON/OFF] to turn on the pump.
2. After the self test, press [Yes] to clear settings.
3. Place the distal end of tubing into a collection container a minimum of 36 inches
below the cassette.
4. With the cassette door closed, check the distal end of the tubing for fluid flow.
Verify a minimal flow of fluid occurs (a few drops maximum).
5. Open the cassette door and check the distal end of the tubing for fluid flow.
Verify a minimal flow of fluid occurs (a few drops maximum).
Note: A small amount of fluid may be expelled from the cassette when opening
or closing the door.
6. Close the cassette door.
5.2.7
DISPLAY TEST
To perform the display test, refer to Figure 5-1, then proceed as follows:
1. Verify the LCD backlight is illuminated and the display is clearly legible at eye level
from approximately 18 inches.
2. With the pump in the DELIVERY screen, press the [Options/Vol Inf] softkey
to select the OPTIONS screen.
3. Using the [SELECT] key, select Lighting/Contrast, and press [Choose].
4. Press the [Decrease Setting] and [Increase Setting] softkeys to change backlight
intensity. Verify intensity decreases and increases.
5. Using the [SELECT] key, select Display Contrast.
6. Press [Decrease Setting] and [Increase Setting] to change display contrast. Verify the
display contrast decreases and increases.
7. Press the [Cancel] softkey to return to the OPTIONS screen.
8. Press the [Back] softkey to return to the DELIVERY screen.
5.2.8
KEYPAD VERIFICATION/FUNCTIONAL TEST
To perform the keypad verification/functional test, refer to Figure 5-1, then proceed
as follows:
1. With the pump in the DELIVERY screen, press the [A] softkey to select Line A.
2. Verify the PROGRAM screen is displayed.
3. Using the numeric keypad, enter a rate of 123 mL/hr.
4. Using the [SELECT] key, select VTBI.
5. Using the numeric keypad, enter a VTBI of 4567 mL.
6. Press [START]. Verify fluid is pumping, the message “PUMPING” is displayed in the
line A status bar, and the line A LED flashes.
Technical Service Manual5 - 7430-95424-001 (Rev. 09/03)
Page 71
SECTION 5 MAINTENANCE AND SERVICE TESTS
7. Press [STOP], then press and hold the [Backprime] softkey.
8. Verify the “BACKPRIMING” and “RELEASE BACKPRIME TO STOP” messages
are displayed, and verify the pump is actually backpriming.
9. Release the [Backprime] softkey, press [START], and verify normal pumping
operation.
10. Press the [B] softkey.
11. Verify PIGGYBACK is the displayed delivery mode. If necessary, change the delivery
mode by pressing the [Change Mode] softkey.
12. Using the [SELECT] key, select Rate.
13. Using the numeric keypad, enter a rate of 890 mL/hr.
14. Using the [SELECT] key, select VTBI.
15. Using the numeric keypad, enter a VTBI of 2.0 mL.
16. Press [START] and verify fluid is pumping. Verify the message “PUMPING”
is displayed in the line B status bar, the line B LED flashes, and line A goes into
Delayed mode.
17. After 20 seconds, verify pumping has switched to line A.
18. Press [STOP].
19. Press [Options/Vol Inf]. Select Volume Infused, and press [Choose].
20. Using the [SELECT] key, select Line A.
21. Press the [CLEAR] key. Verify the line A volume is 0.0 mL and press [Enter].
5.2.9
ALARM LOUDNESS TEST
To perform the alarm loudness test, refer to Figure 5-2, Rear Enclosure and Peripheral/
Interface Assembly, then proceed as follows:
1. Press the [A] softkey to select Line A.
2. Enter a rate of 400 mL/hr and a VTBI of 1 mL, then press [START].
3. Verify the alarm sounds when the dose has been delivered.
4. Turn the volume control knob between HIGH and LOW and verify the alarm
loudness changes. The volume control knob is located on the peripheral/interface
assembly.
5. Press the [SILENCE] key, and verify the alarm is silenced.
6. Press [STOP].
430-95424-001 (Rev. 09/03)5 - 8Plum A+3 Infusion System
Page 72
VOLUME
CONTROL
KNOB
LOCKOUT
SWITCH
NURSE
CALLBACK
BARCODE
DATAPORT
PERIPHERAL/
INTERFACE
ASSEMBLY
5.2 PERFORMANCE VERIFICATION TEST
REAR
ENCLOSURE
01K07016
Figure 5-2. Rear Enclosure and Peripheral/Interface Assembly
5.2.10
LOCKOUT SWITCH TEST
To perform the lockout switch test, refer to Figure 5-2, then proceed as follows:
1. Press the [A] softkey to select Line A.
2. Enter a rate of 400 mL/hr and a VTBI of 50 mL.
3. Press [START], and verify the pump is operating.
4. Toggle the lockout alarm switch up (ON) to engage the alarm. The lockout switch
is located on the peripheral/interface assembly.
5. Press any key except [STOP], and verify an alarm sounds and the “HARD LOCKOUT
ENABLED” message is displayed. Verify the pump continues to operate until [STOP]
is pressed.
6. Press [STOP] and verify the “HARD LOCKOUT VIOLATION” message appears.
7. Toggle the lockout alarm switch down (OFF). Verify the “HARD LOCKOUT
VIOLATION” message disappears and the alarm stops.
8. Press [START].
9. Open the door and verify the “DOOR OPEN WHILE PUMPING” message is displayed
and the audio alarm activates.
10. Close the cassette door.
11. Press [No] at the “CLEAR SETTINGS?” prompt.
Technical Service Manual5 - 9430-95424-001 (Rev. 09/03)
Page 73
SECTION 5 MAINTENANCE AND SERVICE TESTS
5.2.11
PROXIMAL OCCLUSION TEST
To perform the proximal occlusion test, proceed as follows:
1. Press the [A] softkey to select Line A.
2. Enter a rate of 400 mL/hr and a VTBI of 50 mL.
3. Press [START].
4. After several pumping cycles, clamp line A tubing proximal to the cassette.
Verify the “PROX OCCL A/AIR” message flashes and the alarm sounds before three
pumping cycles are completed.
5. Press the [SILENCE] key and verify the alarm stops while the message on the display
continues to flash.
6. Unclamp the proximal line and press [START]. Verify pumping resumes.
7. Press [STOP].
5.2.12
PROXIMAL AIR-IN-LINE TEST
To perform the proximal air-in-line alarm test, refer to Figure 5-3, Special Cassettes
with Bubble Sensor Tips Removed, then proceed as follows:
1. Install the special cassette marked proximal, and close the cassette door.
Note: Confirm the special cassette proximal bubble sensor tips are removed
(see Figure 5-3).
2. Press [Yes] to clear settings.
3. Press the [A] softkey to select Line A.
4. Enter a rate of 400 mL/hr and a VTBI of 50 mL.
5. Press [START].
6. Before 1 mL of fluid is delivered, verify the alarm sounds and the “PROX AIR
A. BACKPRIME” message is flashing on the display.
7. Press [STOP], open the door, and remove the special cassette.
430-95424-001 (Rev. 09/03)5 - 10Plum A+3 Infusion System
Page 74
PROXIMAL BUBBLE SENSOR BULB TIPS
(REMOVED FOR PROXIMAL-AIR-IN-LINE ALARM TEST)
5.2 PERFORMANCE VERIFICATION TEST
(REMOVED FOR DISTAL-AIR-IN-LINE ALARM TEST)
CASSETTE CENTERING DEVICE
DISTAL BUBBLE SENSOR BULB TIPS
98G01024
Figure 5-3. Special Cassettes with Bubble Sensor Tips Removed
5.2.13
DISTAL AIR-IN-LINE TEST
To perform the distal air-in-line alarm test, refer to Figure 5-3, then proceed as follows:
1. Install the special cassette marked distal, and close the cassette door.
Note: Confirm the special cassette distal bubble sensor tips are removed
(see Figure 5-3).
2. Press [Yes] to clear settings.
3. Press the [A] softkey to select Line A.
4. Enter a rate of 400 mL/hr and a VTBI of 50 mL.
5. Press [START].
6. Before 1 mL of fluid is delivered, verify the alarm sounds and the “DISTAL AIR”
message is flashing on the display.
7. Press [STOP], open the door, and remove the special cassette.
Technical Service Manual5 - 11430-95424-001 (Rev. 09/03)
Page 75
SECTION 5 MAINTENANCE AND SERVICE TESTS
5.2.14
DISTAL OCCLUSION TEST
To perform the distal occlusion test, refer to Figure 5-4, Distal Occlusion Test Setup,
then proceed as follows:
1. Install the cassette and connect the distal tubing to the DPM through a three-way
stopcock as illustrated in Figure 5-4. Close the cassette door.
Note: A reflux valve may be attached between the stopcock and the DPM
to keep moisture out of the DPM.
Note: The height of the DPM must be 0 ± 12 inches from the midline of the
pumping chamber.
2. Press [ON/OFF] to turn on the pump.
3. Press [Yes] to clear settings.
4. Press [Options/Vol Inf] to select the OPTIONS screen.
5. Using the [SELECT] key, select Pressure/Post Infusion Rate, and press [Choose].
6. Verify the distal pressure Limit is set at 6.0 psi. If the pressure limit is not 6.0 psi,
use the numeric keypad to enter 6.0 and press [Enter].
7. Press the [A] softkey to select Line A.
8. Enter a rate of 40 mL/hr and a VTBI of 50.0 mL.
9. Open the three-way stopcock to air.
10. Press [START] and allow the infusion pump to stabilize for one minute. Verify all
air is cleared from the tubing.
11. Set the three-way stopcock to measure pressure.
12. Verify the distal occlusion audible alarm occurs at 6.0 ± 2.0 psi. Verify the “DISTAL
OCCLUSION” message is flashing on the screen.
13. Open the three-way stopcock to air.
14. Open and close the door. Press [No] at the “CLEAR SETTINGS?” prompt.
15. Press [Options/Vol Inf] to select the OPTIONS screen.
16. Using the [SELECT] key, select Pressure/Post Infusion Rate, and press [Choose].
17. Using the [SELECT] key, select Distal Pressure Limit.
18. Using the numeric keypad, enter 10.0 psi, and press [Enter].
19. Set the three-way stopcock to measure pressure, then press [START].
20. Verify the distal occlusion audible alarm occurs at 10.0 ± 2.0 psi. Verify the “DISTAL
OCCLUSION” message is flashing on the screen.
21. Open the door and remove the cassette.
430-95424-001 (Rev. 09/03)5 - 12Plum A+3 Infusion System
Page 76
5.2 PERFORMANCE VERIFICATION TEST
OFF
mmHg
cmOFHO
-13.5TO15
-13.5TO75
PSI
INCHESOFHO
UNIVERSAL
PRESSUREMETER
PRESSUREINPUT
OFF
mmHg
cmOFHO
-13.5TO15
-13.5TO75
PSI
INCHES OFH O
UNIVERSAL
PRESSUREMETER
PRESSUREINPUT
FROM FLUID
CONTAINERS
DPM
THREE-WAY
STOPCOCK
UNIVERSAL
UNIVERSAL
PRESSUREMETER
PRESSUREMETER
0.00
0.00
INCHESOFHO
INCHES OFH O
cmOFHO
2
cmOFHO
2
2
2
mmHg
-13.5TO15
mmHg
-13.5TO15
PSI
PSI
OFF
OFF
-13.5TO75
1
23
PRESSUREINPUT
PRESSUREINPUT
-13.5TO75
01K07018
Figure 5-4. Distal Occlusion Test Setup
5.2.15
DELIVERY ACCURACY TEST
Note: Accuracy testing is for informational purposes only, and is not to be used
as a re-release test. If there is any concern as to infusion pump accuracy,
contact Abbott Laboratories.
CAUTION: Do not remove the protective cover from the 21-gauge needle.
To perform the delivery accuracy test, proceed as follows:
1. Install an 18-gauge blunt cannula or a 21-gauge needle to the distal end of the
tubing. Verify the fluid container is 18 to 24 inches above the pumping chamber.
Verify all lines are unclamped.
2. Place the distal output end of tubing into the graduated cylinder.
3. Press the [A] softkey to select Line A.
4. Enter a rate of 200 mL/hr and a VTBI of 10 mL. Start the stopwatch and press
[START] simultaneously.
5. Press the [B] softkey to select Line B.
6. Verify the pump is in the PIGGYBACK delivery mode. Press [Change Mode] to
change the delivery mode, if necessary.
7. Enter a rate of 200 mL/hr and a VTBI of 10 mL.
Technical Service Manual5 - 13430-95424-001 (Rev. 09/03)
Page 77
SECTION 5 MAINTENANCE AND SERVICE TESTS
8. Press [START], and verify the pump switches to line B.
9. Verify the “KVO” message flashes on the display and an audible alarm sounds when
total delivery is complete on line A.
Note: When the “KVO” message appears, stop the stopwatch and press
[STOP].
10. Verify the volume delivered is 20 mL ± 0.8 mL.
11. Use the following formula to calculate the delivery accuracy.
[(Total volume delivered (mL) x (18 sec/mL)]
Total delivery time (sec)
An example:
[(20.4 mL) x (18 sec/mL)]
360 sec
12. Verify the accuracy is 1.0 ± 0.04.
=
1.02
5.2.16
ELECTRICAL SAFETY TEST
To perform the electrical safety test, proceed as follows:
1. Connect the infusion pump AC power cord to a safety analyzer.
2. Connect the safety analyzer ground lead to the infusion pump ground test-point
located on the rear of the pump.
3. Check the leakage current with the safety analyzer. Leakage current (both open
and closed ground) must not exceed 100 microamperes AC
4. Measure the resistance of the AC connector ground lug with the safety analyzer.
Resistance should not exceed 0.1 ohm.
rms
.
5.2.17
END OF THE PVT
If all performance verification tests have been successful, proceed as follows:
1. Press [Options/Vol Inf]. Select Volume Infused, and press [Choose].
2. Press [CLEAR] to clear the volume infused.
3. Press [Enter].
4. Press the [A] softkey.
5. Press the [Clear Program] softkey.
6. Press [Yes] at the “CLEAR LINE A SETTINGS?” prompt.
7. Press [Cancel/Back] to return to the delivery screen.
8. Press the [B] softkey.
9. Press the [Clear Program] softkey.
10. Press [Yes] at the “CLEAR LINE B SETTINGS?” prompt
430-95424-001 (Rev. 09/03)5 - 14Plum A+3 Infusion System
Page 78
5.3 BARCODE READER WAND TEST (OPTIONAL)
11. Reset the infusion pump to the original configuration.
12. Turn off the pump and return it to service.
Note: If any tests fail, refer to Section 6, or contact Abbott Laboratories.
5.3
BARCODE READER WAND TEST
(OPTIONAL)
Note: The barcode reader wand test may be bypassed if the barcode reader function
is not used.
To perform the barcode reader wand test, proceed as follows:
1. Remove the plastic cover from the 9-pin connector on the peripheral/interface
assembly, and connect the barcode reader wand to the 9-pin connector.
2. Insert a primed cassette and close the door.
3. Press [ON/OFF] to turn on the infusion pump. Press [Yes] to clear settings.
4. Set the rate to 100 mL and the VTBI to 50 mL.
5. Verify the “WAND ACTIVE” message appears on the display.
6. Scan a barcode label from the barcode directory, and verify the corresponding drug
name is displayed.
5.4
NURSE CALL TEST (OPTIONAL)
Note: The nurse call test may be bypassed if the nurse call function is not used.
To perform the nurse call test, attach the nurse call test cable and proceed as follows:
1. Set the primary delivery rate to 400 mL/hr, and the primary dose limit to 1 mL.
2. Connect the DMM to the nurse call test cable.
3. Press [START] and verify pumping action.
4. After “DOSE END” and “KVO” appear on the display, observe a short circuit on the
DMM (approximately 1 ohm on a scale of 0 to 100 ohms).
5.5
PERIODIC MAINTENANCE INSPECTION
Periodic maintenance inspections should be performed per hospital procedures for
compliance to accreditation requirements. It is recommended that JCAHO and/or hospital
protocol be followed for establishing an infusion pump periodic maintenance inspection
schedule. Product specifications for this inspection are listed in Section 8, Specifications.
To perform the periodic maintenance inspection, complete the PVT in Section 5.2.
Technical Service Manual5 - 15430-95424-001 (Rev. 09/03)
Page 79
SECTION 5 MAINTENANCE AND SERVICE TESTS
5.6
BATTERY OPERATION OVERVIEW
The Plum A+3 infusion pump is intended to operate on battery power on an exception
basis only, such as emergency backup or temporary portable operation. Examples
of emergency backup include AC power failure or inadvertent disconnection of the AC
power cord. An instance of temporary portable operation includes patient transfer from
one location to another.
The infusion pump should be connected to AC power whenever possible to allow the battery
to remain fully charged. The line power indicator turns off and the BATTERY legend
illuminates when the infusion pump is operating on battery power. The backlight
extinguishes after one minute of pump operation on battery power.
Factors that most commonly affect battery life are the depth and frequency of discharge
and the length of the recharge period. As a general rule, the more often the battery is
discharged and recharged, the sooner it will need replacement. The primary cause of
damage is leaving the battery in a less than fully charged state for any period of time.
Battery damage can occur in a matter of hours and cause a permanent loss of battery
capacity. The amount of lost capacity depends on the degree of discharge, the storage
temperature, and the length of time the battery was stored in a discharged state.
Note: A permanently damaged battery cannot be recharged to full capacity.
When the battery discharges below the acceptable level while the infusion pump is
operating, the alarm sounds and the “LOW BATTERY” message displays. Although it is
not recommended to continue operating the device on battery power at this point, the
battery continues providing power until discharged. At this point, the infusion pump
enters the battery discharged mode, a continuous audible alarm sounds and, after three
minutes, operation ceases.
CAUTION: As soon as the LOW BATTERY alarm occurs, connect the infusion pump
to AC power.
Recharging occurs any time the infusion pump is connected to AC power.
It is recommended that the pump be connected to AC power whenever practical to
maximize available battery charge during transport or ambulation. The device does not
have to be on for the battery to recharge. Recharging while the infusion pump is operating
is rate dependent.
Note: The infusion pump should be operated on battery power for six continuous
hours at least once every six months for optimum battery performance and life.
430-95424-001 (Rev. 09/03)5 - 16Plum A+3 Infusion System
Page 80
Section 6
TROUBLESHOOTING
This section contains information on technical assistance, alarm messages and error
codes, and troubleshooting procedures for the Plum A+3 infusion system.
6.1
TECHNICAL ASSISTANCE
For technical assistance, product return authorization, and to order parts, accessories,
or manuals within the United States, contact Abbott LaboratoriesTechnical Support
Operations.
1-800-241-4002
For additional technical assistance, including Technical Service Bulletins, technical
training, and product information, visit the website at:
www.abbotthpd.com/service
Send all authorized, prepaid returns within the United States to the following address:
Abbott Laboratories
Technical Support Operations
755 Jarvis Drive
Morgan Hill, California 95037
For technical assistance, product return authorization, and to order parts, accessories,
or manuals from outside the United States, contact the nearest Abbott Laboratories
sales office.
6.2
ALARM MESSAGES AND ERROR CODES
Under most alarm conditions the infusion pump ceases normal operation, generates
an audible alarm, and displays an alarm message or error code on the LCD screen.
There are two types of alarm conditions:
-alarm codes that can be cleared by the operator
-error codes that require qualified service personnel
See Table 6-1, Operational Alarm Messages and Corrective Actions, and Table 6-2,
Error Codes Requiring Technical Service.
Technical Service Manual6 - 1430-95424-001 (Rev. 09/03)
Page 81
SECTION 6 TROUBLESHOOTING
6.2.1
OPERATIONAL ALARM MESSAGES
Table 6-1 lists infusion pump alarm codes that can be cleared by the operator. Also listed
in Table 6-1 are the alarm messages, descriptions, possible causes, and corrective actions.
Note: Operational alarm messages are displayed on the LCD screen. Associated error
codes are displayed in the alarms log (see Section 1.8.2).
Note: If an alarm message persists, contact Abbott Laboratories.
Note: Air-in-line on either line applies to both lines.
Table 6-1. Operational Alarm Messages and Corrective Actions
Alarm
Code
N100
or
E100
(URC)
N101
or
E101
(NAA)
N102
or
E102
(RL)
N103
or
E103
(SEEP
CRC)
N104
or
E104
(NC2)
AlarmDescriptionPossible Cause
Unrecognizable
cassette
No actionNo operator action
Infuser idle
2 minutes
NV RAM lost
thrpy data
Nurse Callback BDelivery line B
Incorrect cassette
type
and no delivery for
two minutes during
delivery
parameters entry
Infuser in reset
or idle for over
two minutes
Therapy data is lostInfuser did not
has changed
(if alarm is enabled)
An incorrect
cassette is
inserted
Interruption or
a partial change
to a program
Programming
set without start
for two minutes
complete the
previous
non-volatile
memory write
successfully
End of delivery
step on line B
other than VTBI
complete while
callback is
enabled
Corrective
Action
Insert proper
cassette
Complete
programming
of the pump
Press [START]
Re-enter all
programmed
data
Press [SILENCE]
N105
or
E105
(NC1)
N160
or
E160
(VTB2)
430-95424-001 (Rev. 09/03)6 - 2Plum A+3 Infusion System
Nurse Callback ADelivery line A
has changed
(if alarm is enabled)
Line B VTBI
complete
Programmed
volume to be
infused completed
on line B
End of delivery
step on line A
other than VTBI
complete while
callback is
enabled
VTBI is complete
on line B
Press [SILENCE]
Press [SILENCE]
and replace
IV bag, and
restart line B
Page 82
6.2 ALARM MESSAGES AND ERROR CODES
Table 6-1. Operational Alarm Messages and Corrective Actions
Alarm
Code
N161
or
E161
(VTB1)
N180
or
E180
(OD1)
N181
or
E181
(OD1)
N182
or
E182
(OP2)
AlarmDescriptionPossible Cause
Line A VTBI
complete
Programmed
volume to be
VTBI is complete
on line A
infused completed
on line A
Distal OcclPeak distal
occlusion,
non-delivery
Distal OcclNegative distal
occlusion,
non-delivery
Prox. Occl BNegative proximal
occlusion B,
non-delivery
Distal occlusion
detected during
non-delivery
Distal occlusion
detected during
non-delivery
Proximal
occlusion
detected on
line B during
non-delivery
Corrective
Action
Press [SILENCE]
and replace
IV bag, and
restart line A
Backprime the
cassette and
restart pump
Backprime the
cassette and
restart pump
Backprime the
cassette and
restart line B
or
Stop all lines,
backprime the
cassette, and
restart all lines
N183
or
E183
(OP2)
N184
or
E184
(OP1)
N185
or
E185
(OP1)
Prox. Occl BPeak proximal
occlusion B,
non-delivery
Prox. Occl ANegative proximal
occlusion A,
non-delivery
Prox. Occl APeak proximal
occlusion A,
non-delivery
Proximal
occlusion
detected on
line B during
non-delivery
Proximal
occlusion
detected on
line A during
non-delivery
Proximal
occlusion
detected on
line A during
non-delivery
Backprime the
cassette and
restart line B
or
Stop all lines,
backprime the
cassette, and
restart all lines
Backprime the
cassette and
restart line A
or
Stop all lines,
backprime the
cassette, and
restart all lines
Backprime the
cassette and
restart line A
or
Stop all lines,
backprime the
cassette, and
restart all lines
N186
or
E186
(OD1)
Technical Service Manual6 - 3430-95424-001 (Rev. 09/03)
Distal OcclPeak distal
occlusion,
delivery
Distal occlusion
detected during
delivery
Fix occlusion
(closed clamps;
kinked tubing),
and restart pump
Page 83
SECTION 6 TROUBLESHOOTING
Table 6-1. Operational Alarm Messages and Corrective Actions
Alarm
Code
N187
or
E187
(OD1)
N188
or
E188
(OP2)
N189
or
E189
(OP2)
AlarmDescriptionPossible Cause
Distal OcclNegative distal
occlusion,
delivery
Prox. Occl BNegative proximal
occlusion B,
delivery
Distal occlusion
detected during
delivery
Proximal
occlusion
detected during
delivery on line B
Prox. Occl BPeak proximal
occlusion B,
delivery
Proximal
occlusion
detected during
delivery on line B
Corrective
Action
Fix occlusion
(closed clamps;
kinked tubing),
and restart pump
Fix occlusion
(closed clamps;
kinked tubing),
and restart line B
or
Stop all lines,
fix occlusion
(closed clamps;
kinked tubing),
and restart pump
Fix occlusion
(closed clamps;
kinked tubing),
and restart line B
or
Stop all lines,
fix occlusion
(closed clamps;
kinked tubing),
and restart pump
N190
or
E190
(OP1)
N191
or
E191
(OP1)
Prox. Occl ANegative proximal
occlusion A,
delivery
Prox. Occl APeak proximal
occlusion A,
delivery
Proximal
occlusion
detected during
delivery on line A
Proximal
occlusion
detected during
delivery on line A
Fix occlusion
(closed clamps;
kinked tubing),
and restart line A
or
Stop all lines,
fix occlusion
(closed clamps;
kinked tubing),
and restart pump
Fix occlusion
(closed clamps;
kinked tubing),
and restart line A
or
Stop all lines,
fix occlusion
(closed clamps;
kinked tubing),
and restart pump
430-95424-001 (Rev. 09/03)6 - 4Plum A+3 Infusion System
Page 84
6.2 ALARM MESSAGES AND ERROR CODES
Table 6-1. Operational Alarm Messages and Corrective Actions
Alarm
Code
N230
or
E230
(APT)
N231
or
E231
(APB)
N232
or
E232
(APA)
AlarmDescriptionPossible Cause
Prox. Air TotalProximal
air-in-line
total
Prox. Air on BProximal
air-in-line
on line B
500 microliters
of air entered
the cassette
500 microliters
of air entered
the cassette
on line B
Prox. Air on AProximal
air-in-line
on line A
500 microliters
of air entered
the cassette
on line A
Corrective
Action
Backprime
the cassette
and restart pump
or
Remove and
manually reprime
the cassette,
and restart pump
Backprime
the cassette
and restart line B
or
Remove and
manually reprime
the cassette
and restart pump
Backprime
the cassette
and restart line A
or
Remove and
manually reprime
the cassette
and restart pump
N233
or
E233
(ADC)
N234
or
E234
(ADB)
N250
or
E250
(DCO1)
N251
or
E251
(CS1)
N252
or
E252
(BDP)
Distal Air
Cumulative
Distal air
cumulative
500 microliters
of air detected
in the last 5.3 mL
of fluid delivered
Distal Air BolusDistal air bolus100 microliters
bolus of air
detected at
distal sensor
Door opened
while pumping
Valve/Cass
Test Fail
Door opened
while pumping
Valve/cassette
test failure
Door opened
while pumping
Valve/cassette
fails the leak test
Depleted BatteryDischarged batteryThe battery is
discharged to the
recommended
maximum
discharge
condition
Remove and
manually reprime
the cassette
and restart pump
Remove and
manually reprime
the cassette
and restart pump
Turn off the pump
or
Insert the
cassette and
close the door
Replace cassette
and retest or
backprime
and retest
Connect the
pump to
AC power
or
Recharge
or replace
the battery
Technical Service Manual6 - 5430-95424-001 (Rev. 09/03)
Page 85
SECTION 6 TROUBLESHOOTING
Table 6-1. Operational Alarm Messages and Corrective Actions
Alarm
Code
N253
or
E253
(LOV)
N254
or
E254
(FPL)
N255
or
E255
(SLV)
N256
or
E256
(SLE)
AlarmDescriptionPossible Cause
Lockout ViolationLockout violationThe use of the
[STOP] key or an
attempt to open
the door while
lockout switch
is locked
Lockout EnabledKeypad lockedAny action not
resulting in
stopping of the
delivery while
the lockout
switch is locked
Soft lockout
violation
Soft lockout
enabled
Soft lockout
violation
Soft lockout
enabled
The use of the
[STOP] key or an
attempt to open
the door while
soft lockout
is locked
Any action not
resulting in
stopping of the
delivery while
the soft lockout
locked
Corrective
Action
Unlock the
lockout switch
Unlock the
lockout switch
Enter .963
key code
Enter .963
key code
6.2.2
ERROR CODES REQUIRING TECHNICAL SERVICE
Table 6-2 lists infusion pump error codes that require technical service. Also listed
in Table 6-2 are malfunction descriptions, possible causes, and corrective actions.
Table 6-2. Error Codes Requiring Technical Service
Error
Code
E300ADC failureAnalog to digital
E301Audio alarm
MalfunctionPossible CauseCorrective Action
Replace CPU PWA
failure
converter failure
Piezo is off but sensed on
or
Piezo is on but sensed off
(see Section 7.2.14.6)
Reset time and date,
if required
(see Section 1.8.3)
Turn power off, then on,
to reset the pump
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date,
if required
(see Section 1.8.3)
430-95424-001 (Rev. 09/03)6 - 6Plum A+3 Infusion System
Page 86
Error
Code
6.2 ALARM MESSAGES AND ERROR CODES
Table 6-2. Error Codes Requiring Technical Service
MalfunctionPossible CauseCorrective Action
E302Backlight failureBacklight (CCFT tube)
is not at the expected range
E320Battery charge
current out
Battery charge current is
out of range after six hours
of range
E321Battery not
charging
Battery charging timed out
Complete battery discharge
has occurred
E322Battery current
calibration value
Battery integrator calibration
value is out of range
out of range
E323Battery trickle
charge current
Battery trickle charge
current is out of range
out of range
Turn power off, then on,
to reset the pump
Replace LCD
(see Section 7.2.14.3)
and/or CPU PWA
(Section 7.2.14.6)
Reset time and date,
if required
(see Section 1.8.3)
Replace battery
(see Section 7.2.4)
Replace power supply PWA
(see Section 7.2.14.1)
Reset time and date,
if required
(see Section 1.8.3)
Charge battery for
additional six hours
Replace power supply PWA
(see Section 7.2.14.1)
Reset time and date,
if required
(see Section 1.8.3)
E324Supply
overvoltage
E325Battery
overvoltage
E326Battery
disconnected
E340Critical
instruction
failure
An overvoltage condition
is detected in the charging
circuit
An overvoltage condition
is detected in the battery
Battery disconnected while
pump is powered on
Power-up CPU
register test failed
(no malfunction
message displayed)
Check for loose battery
connections
Replace battery
(see Section 7.2.4)
Reset time and date,
if required
(see Section 1.8.3)
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date,
if required
(see Section 1.8.3)
Technical Service Manual6 - 7430-95424-001 (Rev. 09/03)
Page 87
SECTION 6 TROUBLESHOOTING
Table 6-2. Error Codes Requiring Technical Service
Error
Code
MalfunctionPossible CauseCorrective Action
E341Critical data
memory failure
Critical data memory failureReplace mechanism
assembly
(see Section 7.2.14.8)
Reset time and date,
if required
(see Section 1.8.3)
E342Display failureDefective displayReplace LCD
(see Section 7.2.14.3)
and/or CPU PWA
(Section 7.2.14.6)
Reset time and date,
if required
(see Section 1.8.3)
E343Distal air sensor
failure 1
E344Distal air sensor
failure 2
E345Distal pressure
sensor failure 1
With the cassette removed,
the distal air sensor self test
detects liquid
With the cassette inserted,
the distal air sensor self test
detects sensor out of range
Distal pressure sensor failed
while pump is OFF
Replace mechanism
assembly
(see Section 7.2.14.8)
Reset time and date,
if required
(see Section 1.8.3)
E346Distal pressure
sensor failure 2
E347Hardware
watchdog
failure
E371I/O valve motor
failure 1
E372I/O valve motor
failure 2
E373L/S valve motor
failure 1
E374L/S valve motor
failure 2
E375Motor position
sensor failure
E376Plunger synch
failure 1
Distal pressure sensor failed
while pump is ON
Hardware watchdog failure
I/O valve motor malfunction
when a total of four
resynchronizations failed
I/O valve motor malfunction
when three consecutive
resynchronizations failed
L/S valve motor malfunction
when a total of four
resynchronizations failed
L/S valve motor malfunction
when three consecutive
resynchronizations failed
Motor position sensor failure
Plunger malfunction when
resynchronization has failed
a total of four times
430-95424-001 (Rev. 09/03)6 - 8Plum A+3 Infusion System
Page 88
Error
Code
6.2 ALARM MESSAGES AND ERROR CODES
Table 6-2. Error Codes Requiring Technical Service
MalfunctionPossible CauseCorrective Action
E377Plunger synch
failure 2
E378I/O valve
phase loss
E379L/S valve
phase loss
E380Plunger motor
phase loss
E430Proximal
air sensor
failure 1
Plunger malfunction when
three consecutive
resynchronizations failed
Replace mechanism
assembly
(see Section 7.2.14.8)
Reset time and date,
if required
(see Section 1.8.3)
Generic I/O valve failureTurn power off, then on,
to reset infusion pump
Replace mechanism
assembly
(see Section 7.2.14.8)
Reset time and date,
if required
(see Section 1.8.3)
Generic L/S valve failure
Generic plunger motor failure
Proximal air sensor ongoing
test detects liquid with
cassette removed
Replace mechanism
assembly
(see Section 7.2.14.8)
Reset time and date,
if required
(see Section 1.8.3)
E431Proximal
air sensor
failure 2
E432Proximal
pressure
Proximal air sensor self test
detects liquid with cassette
removed
Proximal pressure sensor
failed while pump is OFF
sensor 1
E433Proximal
pressure
Proximal pressure sensor
failed while pump is ON
sensor 2
E434RAM failureRAM failureTurn power off, then on,
to reset infusion pump
E435RTC failReal-time clock failure
E436ROM failureROM checksum failure
E437Software failureGeneric software failure
E438Stack
Stack out-of-range failure
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date,
if required
(see Section 1.8.3)
out-of-range
failure
Technical Service Manual6 - 9430-95424-001 (Rev. 09/03)
Page 89
SECTION 6 TROUBLESHOOTING
Table 6-2. Error Codes Requiring Technical Service
Error
Code
MalfunctionPossible CauseCorrective Action
E439Stuck keyA key is sensed as pressed
for over two minutes
Replace LCD
(see Section 7.2.14.3)
and/or CPU PWA
E440Power hold stuckPower hold signal stuck
Power cannot be turned off
(Section 7.2.14.6)
Reset time and date,
if required
(see Section 1.8.3)
E441Valve self test
failure
I/O or L/S valve self test failedReplace mechanism
assembly
(see Section 7.2.14.8)
Reset time and date,
if required
(see Section 1.8.3)
E443LCD failureLCD bias is out of rangeReplace LCD
(see Section 7.2.14.3)
and/or CPU PWA
(Section 7.2.14.6)
Reset time and date,
if required
(see Section 1.8.3)
E444CPU timebase
inaccurate
E445RTC memory
failure
CPU timer 2 and RTC
measured times disagree
Real-time clock memory
is corrupt
Turn power off, then on,
to reset infusion pump
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date,
E446CPU timer failureCPU timer 1 and timer 2
measured times disagree
if required
(see Section 1.8.3)
E447Battery ADC
reading failure
16 consecutive readings
have been either all zero
or the max value
E448SEEP write
SEEP data write failed
failure
E449SEEP calibration
data corrupted
E450MMIO port
Calibration data block
corrupted
I/O port read/write failure
read/write
failure
E451Inaccurate
Over/underdelivery detected
delivery
E452Software failureMiscellaneous software
failure
430-95424-001 (Rev. 09/03)6 - 10Plum A+3 Infusion System
Page 90
Error
Code
6.3 TROUBLESHOOTING PROCEDURES
Table 6-2. Error Codes Requiring Technical Service
MalfunctionPossible CauseCorrective Action
E453Two SEEP CRC
errors
E454NVRAM over
capacity
NVRAM data block corruptedTurn power off, then on,
to reset infusion pump
Software trying to write into
non-existent NVRAM space
Replace CPU PWA
(see Section 7.2.14.6)
Reset time and date,
if required
(see Section 1.8.3)
Note: Some error codes include sub-ID codes. These sub-ID codes are intended
for Abbott Laboratories internal use only, and should be included when contacting
Abbott Laboratories Technical Support Operations(see Section 6.1).
6.3
TROUBLESHOOTING PROCEDURES
This section details recommended procedures for problems not associated with
malfunction alarms. Before performing any troubleshooting procedure, turn the infusion
pump off, then on. Allow the self test to complete and proceed as follows:
1. If a malfunction exists, carefully inspect the infusion pump for damage as described
in Section 5.2.2.
2. If an infusion pump inspection has not disclosed a malfunction, perform the PVT
in Section 5.2. Refer to Table 6-3, Troubleshooting with the PVT, for section reference,
probable cause, and corrective actions.
3. If, after completing Steps 1 and 2, a malfunction has not been located, or if the
infusion pump persistently fails, contact Abbott LaboratoriesTechnical Support
Operations.
Table 6-3. Troubleshooting with the PVT
Test FailureProbable CauseCorrective Action
Self test
Section 5.2.4
Cassette alarm test
Section 5.2.5
Technical Service Manual6 - 11430-95424-001 (Rev. 09/03)
Cassette not properly
installed
Defective CPU PWAReplace CPU PWA
Cassette not properly
seated
Defective cassetteReplace cassette
Reseat cassette
(see Section 7.2.14.6)
Reseat cassette
Page 91
SECTION 6 TROUBLESHOOTING
Table 6-3. Troubleshooting with the PVT
Test FailureProbable CauseCorrective Action
Free flow test
Section 5.2.6
Display test
Section 5.2.7
Keypad test
Section 5.2.8
Alarm loudness test
Section 5.2.9
Cassette not properly
Reseat cassette
seated
Defective cassetteReplace cassette
Defective or dirty valve pinsClean valve pins
Replace mechanism assembly
(see Section 7.2.14.8)
Defective display/CPU
assembly
Replace LCD
(see Section 7.2.14.3)
and/or CPU PWA
(Section 7.2.14.6)
Defective display/CPU
assembly
Replace LCD
(see Section 7.2.14.3)
and/or CPU PWA
(Section 7.2.14.6)
Defective CPUReplace CPU PWA
(see Section 7.2.14.6)
Defective peripheral PWAReplace peripheral PWA
(see Section 7.2.8)
Defective piezo alarm
assembly
Replace piezo alarm assembly
(see Section 7.2.14.7)
Lockout switch test
Section 5.2.10
Proximal occlusion test
Section 5.2.11
Proximal air-in-line test
Section 5.2.12
Distal air-in-line test
Section 5.2.13
Defective peripheral PWAReplace peripheral PWA
(see Section 7.2.8)
Closed proximal clampOpen clamp
Cassette not properly
Re-prime cassette
primed
Defective cassetteReplace cassette
Dirty sensor pinClean sensor pin
Defective APP PWAReplace mechanism assembly
(see Section 7.2.14.8)
Defective special cassetteReplace special cassette
Dirty sensorsClean sensors
Defective APP PWAReplace mechanism assembly
(see Section 7.2.14.8)
Defective special cassetteReplace special cassette
Dirty sensorsClean sensors
Defective APP PWAReplace mechanism assembly
(see Section 7.2.14.8)
430-95424-001 (Rev. 09/03)6 - 12Plum A+3 Infusion System
Page 92
6.3 TROUBLESHOOTING PROCEDURES
Table 6-3. Troubleshooting with the PVT
Test FailureProbable CauseCorrective Action
Distal occlusion test
Section 5.2.14
Delivery accuracy test
Section 5.2.15
Electrical safety test
Section 5.2.16
Cassette not properly
Re-prime cassette
primed
Defective cassetteReplace cassette
Dirty sensor pinClean sensor pin
Defective APP PWAReplace mechanism assembly
(see Section 7.2.14.8)
Set not properly primedRe-prime cassette
Damaged or faulty cassetteReplace cassette
Defective mechanism
assembly
Replace mechanism assembly
(see Section 7.2.14.8)
Defective AC power cordReplace AC power cord
(see Section 7.2.5)
Technical Service Manual6 - 13430-95424-001 (Rev. 09/03)
Page 93
SECTION 6 TROUBLESHOOTING
This page intentionally left blank.
430-95424-001 (Rev. 09/03)6 - 14Plum A+3 Infusion System
Page 94
Section 7
REPLACEABLE PARTS AND
REPAIRS
This section itemizes all parts and subassemblies of the Plum A+3 infusion system that
are repairable within the scope of this manual. In addition, this section details replacement
procedures for all listed parts.
7.1
REPLACEABLE PARTS
Replaceable parts for the Plum A+3 are itemized in the spare parts price list and are
identified in Figure 9-1, Illustrated Parts Breakdown. Table 9-2, IPB for the Infusion Pump
identifies each part by an index number that correlates to Figure 9-1.
To request a copy of the current spare parts price list, contact Abbott Laboratories
(see Section 6.1, Technical Assistance), or to view the catalog online, visit the website at:
www.abbotthpd.com/parts
For convenient reference, insert a copy of the spare parts price list here.
Technical Service Manual7 - 1430-95424-001 (Rev. 09/03)
Page 95
SECTION 7 REPLACEABLE PARTS AND REPAIRS
This page intentionally left blank.
430-95424-001 (Rev. 09/03)7 - 2Plum A+3 Infusion System
Page 96
7.2 REPLACEMENT PROCEDURES
7.2
REPLACEMENT PROCEDURES
This section contains safety and equipment precautions, required tools and materials,
and step-by-step procedures for replacing parts in the infusion pump. Unless otherwise
stated, always perform the PVT after a replacement procedure.
7.2.1
SAFETY AND EQUIPMENT PRECAUTIONS
Before opening the front enclosure of the infusion pump, take all necessary precautions
for working on high-voltage equipment.
WARNING:POSSIBLE EXPLOSION HAZARD EXISTS IF THE INFUSION PUMP
WARNING:UNLESS OTHERWISE INDICATED, DISCONNECT THE INFUSION
CAUTION: Use proper ESD grounding techniques when handling components.
Wear an antistatic wrist strap and use an ESD-protected workstation. Store PWAs
in antistatic bags before placing them on any surface.
IS SERVICED OR REPAIRED IN THE PRESENCE OF FLAMMABLE
ANESTHETICS.
PUMP FROM AC POWER BEFORE PERFORMING ANY
REPLACEMENT PROCEDURE.
7.2.2
REQUIRED TOOLS AND MATERIALS
The following tools and materials, or equivalents, are required for the replacement
procedures in this section. In addition, the beginning of each procedure lists tools and
materials required for that specific procedure.
❏ Set of standard and metric nutdrivers
❏ Set of flat blade screwdrivers
knife
®
screwdrivers
❏ Set of Phillips
❏ Custom nutdriver (P/N 519-95056-001)
❏ Long needle nose pliers
❏ X-acto
❏ Mild solvent
®
Technical Service Manual7 - 3430-95424-001 (Rev. 09/03)
Page 97
SECTION 7 REPLACEABLE PARTS AND REPAIRS
7.2.3
RUBBER FOOT PAD REPLACEMENT
The recommended tool for this procedure is a No. 2 Phillips screwdriver.
To replace the rubber foot pads, refer to Figure 7-1, Bottom View, then proceed as follows:
1. Press [ON/OFF] to turn off the infusion pump, and disconnect the pump from
AC power.
2. Place the pump face down on a soft flat surface.
3. Using the Phillips screwdriver, remove the 6-32 x 1/2 screw securing the rubber
foot pad.
4. Install the replacement rubber foot pad in the exact reverse order of removal.
5. Connect the infusion pump to AC power.
Replacement of the rubber foot pads is routine maintenance and no verification procedure
is normally required. However, if the infusion pump may have been damaged during a
rubber foot pad replacement, perform the PVT in Section 5.2.
Figure 7-1. Bottom View
RUBBER FOOT (4)
BATTERY DOOR (3)
6-32 x 1/2 SCREW (4)
01K06006
430-95424-001 (Rev. 09/03)7 - 4Plum A+3 Infusion System
Page 98
7.2 REPLACEMENT PROCEDURES
7.2.4
BATTERY ASSEMBLY, BATTERY DOOR, AND BATTERY
DOOR PAD REPLACEMENT
Recommended tools for this procedure are a medium size flat blade screwdriver, an X-acto
knife, and mild solvent.
To replace the battery, battery door, and battery door pad, refer to Figure 7-1 and
Figure 7-2, AC (Mains) Power Cord, Power Cord Retainer, Velcro Retainer Strap, and Battery
Assembly Replacement, then proceed as follows:
1. Press [ON/OFF] to turn off the infusion pump, and disconnect the pump from
AC power.
2. Place the pump face down on a soft flat surface.
3. Using the flat blade screwdriver, remove the 6-32 x 1/2 screw that attaches the
battery door to the pump, and remove the door.
4. Inspect the battery door and door pad for damage. Replace the door, if necessary.
5. If the battery door pad is defective, remove it and clean the door with mild solvent.
Dry the battery door thoroughly, and install the replacement pad on the door.
6. Disconnect the battery harness from the charger circuit cable. Carefully pull the
battery harness wires and connector outside the enclosure, and remove the battery.
7. Connect the replacement battery harness to the charger circuit cable, and insert
the replacement battery into the enclosure.
Note: The cable connectors are keyed so that cables cannot be connected
incorrectly.
Note: Confirm the battery harness is not pinched between the battery and
the enclosure.
8. Replace the battery door using the screw that was removed in Step 3.
To verify successful replacement of the battery, press [ON/OFF] with the infusion pump
disconnected from AC power, and verify the front panel battery symbol illuminates.
Replacement of the battery door and battery door pad is routine maintenance and no
verification procedure is normally required. However, if the infusion pump may have been
damaged during these procedures, perform the PVT in Section 5.2.
Technical Service Manual7 - 5430-95424-001 (Rev. 09/03)
Page 99
SECTION 7 REPLACEABLE PARTS AND REPAIRS
COTTER RING
BATTERY
ASSEMBLY (3)
DOOR PAD (3)
RETAINING RING (3)
BATTERY DOOR (3)
6-32 x 1/2
SCREW (3)
HANGER ASSEMBLY
VELCRO STRAP
POWER CORD
4-40 X 3/8
SCREW
POWER CORD
RETAINER
01H06001
Figure 7-2. AC (Mains) Power Cord, Power Cord Retainer, Velcro Retainer Strap,
and Battery Assembly Replacement
7.2.5
AC (MAINS) POWER CORD, POWER CORD RETAINER,
AND VELCRO RETAINER STRAP REPLACEMENT
The recommended tool for this procedure is a No. 2 Phillips screwdriver.
Note: For Velcro retainer strap replacement only, proceed to Step 5.
To replace the AC power cord, power cord retainer, and Velcro retainer strap, refer to
Figure 7-2, then proceed as follows:
1. Press [ON/OFF] to turn off the infusion pump, and disconnect the pump from
AC power.
2. Using the Phillips screwdriver, remove the 4-40 x 3/8 screw from the power cord
retainer. Turn the power cord retainer approximately 1/8 turn counterclockwise.
3. Unplug the power cord, and slide the plug through the retainer.
Note: Remove the AC power cord from its receptacle by grasping the plug.
Do not pull the cord.
4. Remove the Velcro strap from the power cord. Inspect the Velcro strap for wear and
replace the strap, if necessary.
5. Attach the Velcro strap to the replacement power cord.
430-95424-001 (Rev. 09/03)7 - 6Plum A+3 Infusion System
Page 100
7.2 REPLACEMENT PROCEDURES
6. Install the replacement AC power cord in the exact reverse order of removal.
7. Connect the infusion pump to AC power. Press [ON/OFF] and verify the pump
powers on.
To verify successful AC power cord, power cord retainer, and Velcro retainer strap
replacement, perform the PVT in Section 5.2.
7.2.6
SEPARATING THE FRONT ENCLOSURE, REAR
ENCLOSURE, AND MAIN CHASSIS ASSEMBLY
The recommended tool for this procedure is a No. 2 Phillips screwdriver.
CAUTION: Use proper ESD grounding techniques when handling components.
Wear an antistatic wrist strap and use an ESD-protected workstation. Store PWAs
in antistatic bags before placing them on any surface.
Note: The front enclosure consists of an upper assembly and a lower assembly.
The main chassis consists of an upper assembly and a lower assembly.
To separate the front enclosure, rear enclosure, and main chassis assembly, refer to
Figure 7-3, Separating the Front Enclosure, Rear Enclosure, and Main Chassis Assembly
then proceed as follows:
1. Remove the battery doors and batteries as described in Section 7.2.4.
2. Using the Phillips screwdriver, remove the five 6-32 x 1 1/4 inch screws; three
8-32 x 3 1/2 inch screws; and one 6-32 x 1/2 inch screw from the rear enclosure
(see Figure 7-4, Screw Placement Sequence).
3. Remove the rear enclosure by lifting it up and to the side.
4. Disconnect the three internal power connectors.
5. Using the Phillips screwdriver, remove the two 6-32 x 2 3/4 inch screws from the
bottom corners of the center mechanism.
6. Set the pump upright and remove the front upper enclosure by pulling it away from
the upper chassis assembly.
7. Remove the lower front enclosure by tilting the pump back approximately
10 degrees, being careful not to damage the peripheral/interface PWA. Pull the
lower front enclosure away from the lower chassis assembly.
8. Reassemble the front enclosure, rear enclosure, and main chassis assembly in the
exact reverse order of disassembly. Follow the screw placement sequence in
Figure 7-4.
Note: When reassembling the upper front enclosure, lift all three door
handles first.
To verify successful assembly, perform the PVT in Section 5.2.
Technical Service Manual7 - 7430-95424-001 (Rev. 09/03)
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.