Abbott Flexiflo Quantum User manual

Title: Ross “Flexiflo” Quantum By Ross/Abbott Labs, Columbus, OH. File = Flexiflo-Quantum.doc
DATE: 7-29-2010
DISCLAIMER: THIS PROCEDURE PROVIDED "AS IS" AND WITH POSSIBLE FAULTS. USER MUST VERIFY BEFORE USE. NEITHER PROVIDER NOR WEBSITE ASSUMES ANY RESPONSIBILITY FOR ITS USE.
2. Reference Documents
FLEXIFLO® QUANTUM™ Enteral Pump Manual
3. Tools / Fixtures / Labels
Safety Testor Cassette: See picture. Scale with 0.1 gm or better resolution to support a collection cup (preferred) or -­Burette Note: Don’t trust the first 10ml interval at bottom.
Fill this with 10mL and start from there.
“PM next due” label
4. Basic PM Procedure
4.1. Physical: Check line cord, cassette door, and case for cracks.
4.2. Electrical Safety: There is a small metallic stud on lower back for gnd reference. The
maximum leakage is 300 uA. Line cord ground resistance <= 0.1 ohms.
4.3. Turn unit on via the knob. Listen for a normally two segment beep.
4.4. Tubing Setup: At this point have tubing set positioned
into the scale or burette. Note: If a flush line is part of the cassette assembly, pinch it “off”.
4.5. Flow/volume Setup: Set Rate for 120mL/hr and Set Dose
= 10mL. If a flush bag is part of setup, clamp it off.
4.6. Turn knob to Run. After 5.0 minutes the unit should stop,
leaving a 10mL increase. Acceptable delivery is 9 to 11 mL. Leave unit beeping and …
4.7. Audio Volume Button: While the unit beeps indicating
end of dose, press the audio button and note it changes audio volume from normal to somewhat muted.
4.8. Occlusion: Unplug unit and run on battery. Start unit and pinch output to verify an
occlusion is detected.
4.9. Check remaining functions of selector switch. They should be self-explanatory.
4.10. Re-label: Replace old PM label with the new dated (per preference of the institution).
Optional: Protect by applying clear cover label.
5. Extended PM Procedure
One known service issue: the sense pin/switch (shiniest) can electrically get stuck in “cassette is in” position. Spray contact cleaner into the opening around this pin, and gently pull it in and out. Repeat.