Abbott Afinion 2 User Manual

Page 1
US User Manual
1116 554 , 1116 66 3
1116752 Rev. A 2018/09
Page 2
Dear Customer,
Congratulations on the purchase of your Anion™ 2 Analyzer.
Upon arrival of your Anion 2 Analyzer we recommend that the serial number along with the software version be recorded in the table provided below. The additional rows in the table are to be utilized if a software upgrade is performed on your Anion 2 Analyzer. The recorded information will be of great value if and when a question is reported, or the desire to add a new Anion Test to your analyzer arises.
Serial number
(for serial number (SN), see label on the rear side of the analyzer or on the transport container)
Software records
Date
Upon receipt
1. SW upgrade
2. SW upgrade
3. SW upgrade
4. SW upgrade
5. SW upgrade
* See start-up menu when you power on the analyzer (see “How to power on the analyzer”, page 11).
Software version* Anion™ Tests available
Notes
_________________________________________________________________________________________________________________________________________________
_________________________________________________________________________________________________________________________________________________
Technical Support
Call 1.866.216.9505
2 | US
AFINION™ 2 User Manual
Page 3
Intended use of the AFINION™ 2 System
Anion 2 System, consisting of the Anion 2 Analyzer and the Anion Test Cartridges, is for in vitro diagnostic use only. Anion 2 Analyzer is a compact multi-assay analyzer for point-of-care testing and is designed to analyze the Anion Test Cartridges.
CLIA Statements - Waived Anion™ Tests
Anion HbA1c is waived under the Clinical Laboratory Improvement Amendment of 1988 (CLIA`88). A CLIA Certicate of Waiver is needed to perform testing in a waived setting.
If the laboratory does not have a Certicate of Waiver, the Application for Certication (Form CMS-116), can be obtained at https://www.cms.gov/cmsforms/downloads/cms116.pdf.
The form should be mailed to the address of the local State Agency of the State in which the laboratory resides (https://www.cms.gov/CLIA/12_State_Agency_&_Regional_Ofce_CLIA_Contacts.asp).
If the laboratory modies the Anion Test or Anion 2 Analyzer system instructions, the test no longer meets the requirements for waived categorization. A modied test is considered to be highly complex and is subject to all applicable CLIA requirements.
Conformity to directives and standards
European IVD directive and RoHS 2 directive (CE marking)
The Anion 2 Analyzer meets all provisions in the Directive 98/79/EC on in vitro diagnostic (IVD) medical devices and in the Directive 2011/65/EU on the restriction of the use of certain hazardous substances in electrical and electronic equipment (RoHS 2).
North American product safety standards (CNUS mark)
The Anion 2 Analyzer has been tested and found to be in conformity with North American safety standards. See list of safety standards below.
Safety standards
The Anion 2 Analyzer has been tested and found to be in conformity with standards for Safety requirements for electrical equipment for measurement, control, and laboratory use (IEC 61010-1:2010 , UL 61010-1:2012, CAN/CSA-C22.2: 61010-1 -12) and standard for Particular requirements for in vitro diagnostic (IVD) medical equipment (IEC 61010-2-101:2015).
EMC standards
The Anion 2 Analyzer has been tested and found to be in conformity with standards for Electrical equipment for measurement, control, and laboratory use – EMC requirements (EN 61326-1:2013, EN 61326-2-6:2006, EN 61326-2-6:2013 and CFR 47 Telecommunications, Chapter I- FCC Part 15 – Radio Frequency Devices – Subpart B: unintentional radiators).
AFINION™ 2 User Manual
US | 3
Page 4
4 | US
AFINION™ 2 User Manual
Page 5
Table of Contents
Table of contents
Introduction About this user manual 7
Examining the package contents 7
Analyzer System Description Description of the AFINION™ 2 Analyzer 8
Description of the Anion™ Test Cartridge 8
How the AFINION™ 2 System works 9
Internal process control 9
The analyzer self-test 9
The fail-safe mechanisms 9
External process control 9
Patients ID 9
Operator ID 9
Quality Control lockout 9
Calibration 9
Getting Started Installing your analyzer 10
Connecting power supply 10
Connecting additional equipment 10
Connectivity 10
How to power ON the analyzer 11
How to power OFF the analyzer 11
How to operate the analyzer 11
Conguration The AFINION™ 2 menus 12
Setting the conguration 13
Patient ID conguration 13
Patient ID enable/disable 13
Operator conguration 14
Operator ID enable/disable 14
Operator login expiration 14
Operator list management 14
Choosing language 15
Adjusting screen/beeper settings 15
Setting date and time 15
QC lockout conguration 16
General settings 17
Erase all contents and conguration 17
Analyzer network settings 17
Connectivity settings 18
Quality Control Why quality control testing? 19
Choosing control material 19
Handling and testing controls 19
Frequency of control testing 19
Table of contents continues on next page
AFINION™ 2 User Manual
US | 5
Page 6
Table of Contents
Testing Procedures Operating precautions 20
When operating the analyzer 20
When handling the test cartridge 20
Preparing for an AFINION™ 2 Analysis 20
Collecting a sample 21
Analysing a patient/control sample 21
Using the operator ID function 22
Entering operator ID 22
Using the patient ID function 22
Entering patient ID 22
Using the control ID function 23
Entering control ID 23
Using the QC lockout function 23
QC lockout status 23
Running controls with enabled QC lockout function 24
Patient and control results records 25
View, print and export patient and control results 25
Information Codes and Troubleshooting
Maintenance Cleaning and maintenance 29
Warranty 30
Technical Specications AFINION™ 2 Analyzer 31
Gallery of Icons The touch buttons and their function 32
Symbols and Abbrevations 35
When an information code appears 26
Information codes caused by test-specic limitations 26
Information codes caused by sample or test cartridge 27
Information codes and messages caused by analyzer failure 27
Other information codes 28
Service information 28
Cleaning the exterior 29
Cleaning the cartridge chamber 29
Disposal of the analyzer 29
Software upgrade 29
Additional equipment 31
Other symbols and signs 34
6 | US
AFINION™ 2 User Manual
Page 7
Introduction
About this user manual
This user manual will guide you through installation, operation and maintenance of your Anion 2 Analyzer. The user manual also explains how the analyzer works, describes the quality assurance system and assists you in troubleshooting.
For analysing patient samples or controls, please also read the test specic information given in the package inserts following the Anion Test Kits. The quick reference guides, available from your local Anion supplier, highlight the main steps of the test procedures.
It is recommended that you become familiar with these user instructions before you start operating the Anion 2 Analyzer.
Some of the information in this user manual is accompanied with a symbol that points you to the following particulars:
Warnings and precautions
References to the package inserts for the specic Anion Tests and control kits
Examining the package contents
When unpacking, check the contents against the list below and examine the components for signs of shipping damage.
The Anion 2 package unit includes:
• Anion 2 Analyzer
• Power cable
• Power supply, 24 VDC
• User manual
• Quick guides for the available Anion Tests
If the package unit is found incomplete, please report missing items or shipping damage to your supplier. It is recommended to keep the shipping box in case of later transportation of the analyzer.
AFINION™ 2 User Manual
US | 7
Page 8
Analyzer System Description
Description of the AFINION™ 2 Analyzer
Figure 1 shows the main exterior parts of the Anion 2 Analyzer.
1
2
3
4
5
Figure 1
1 ON/OFF button: Turns the power to the analyzer on and off. 2 Red and green LEDs: Light emitting diodes (LEDs) that indicates whether the analyzer is busy or not. 3 Touch screen: Allows you to communicate with the analyzer through touch buttons and messages. 4 The lid: Covers and protects the cartridge chamber. 5 Connectors: For connecting to mains power supply. Options for printer, barcode reader and/or LIS/HIS/EMR.
Do not open the lid manually.
Description of the Anion™ Test Cartridge
The Anion Test Cartridge is unique for each analyte to be measured, as the reagent composition, reagent volumes and the integrated devices are test specic. The test cartridge label has a colour unique for the test. The test cartridges are separately packed in foil pouches to protect the reagents and plastic devices from light, dirt and humidity. A single test cartridge contains all necessary reagents for one test and is ready to use. An integrated sampling device is used for collection of the patient sample or control. The test cartridge cannot be reused. Figure 2 illustrates an Anion Test Cartridge with its functional parts:
Left side Right side
Figure 2
1 Sampling device: For collection of patient sample or control (1a - closed position, 1b - lifted position). 2 Capillary: Capillary to be lled with sample material. 3 Reaction wells: Contain all necessary reagents for one test. 4 Handle: For correct nger grip. 5 Barcode label: Contains assay and lot-specic information for the analyzer. 6 Optical reading area: Area for transmission measurement. 7 ID area: Space for written or labelled sample identication.
8 | US
AFINION™ 2 User Manual
Page 9
Analyzer System Description
How the AFINION™ 2 System works
The Anion 2 System uses different chemical and mechanical assay methods combined with advanced, computerized processing and measuring technology.
A test cartridge with patient sample or control is placed in the cartridge chamber of the analyzer. By manually closing the lid, the test cartridge is transported into the analysis compartment of the analyzer. Test and lot-specic information is obtained from the barcode label (Figure 2). When the test cartridge enters the analyzer, the integrated camera reads the barcode which then initiates the processing of the test cartridge. The sample and reagents are automatically transferred between the wells. An internal camera monitors the entire process. Light-emitting diodes (LEDs) illuminate the reaction area, which can be either a coloured membrane or a reaction well. The camera detects the reected or transmitted light, which is converted to a test result and displayed on the touch screen. When the user accepts the result, the lid covering the cartridge chamber opens automatically and the used test cartridge can be removed and discarded. The analyzer is then ready for the next run.
Internal process control
The analyzer self-test
A self-test is performed during start-up of the analyzer to ensure that the instrument is operating according to established specications. The self-test validates:
• Hardware and software integrity
• Test cartridge transport system
• Liquid transport system
• Camera vision system
If the self-test fails at any point, the red LED will start ashing and an information code will be displayed on the touch screen (see “Information codes and troubleshooting”, page 26-28).
When the analyzer is powered on for a longer period, it will automatically restart once a day to ensure that a self-test is done regularly. This procedure does not interrupt any analysis of the test cartridge.
The fail-safe mechanisms
Fail-safe mechanisms are included to secure safe processing. The integrated camera inspects the test cartridges initially before the process starts and during the assay. If defects are detected (e.g. broken capillary, the cartridge is used past its expiry date), the test cartridge is rejected and an information code is displayed. During processing vital functions and components (e.g. pumps, heater) are supervised. When problems are detected by the built-in safety mechanism, the process will be aborted and an information code will be displayed.
External process control
Patient ID
The Anion 2 patient ID functionality will, if congured, allow up to four patient ID elds to be entered. The patient ID will be stored with each patient test result in the result records.
Operator ID
The Anion 2 operator functionality will, if congured, require the operators to login before testing. The functionality may also prevent unauthorized operators to login, perform tests and conguration. The operator ID will be stored with each test result in the result records.
Quality Control lockout
The Anion 2 QC lockout function allows you to congure the instrument to automatically enforce your local required frequency of control testing. If the required control test has not been performed or the control result is outside the acceptable range, the instrument will disable patient testing for this assay. For manufacturer recommen­dations (see “Frequency of control testing” page 19).
For more information regarding these functionalities, see “Conguration” page 12–18.
Calibration
The Anion 2 Analyzer has been manufactured to deliver reliable and accurate results. During manufacturing, the analyzers are calibrated against a reference sys­tem. This procedure has been established to ensure that all analyzers operate within identical tolerance limits.
Test specic calibration data are established for each lot of test cartridges and then stored in the barcode label (Figure 2). When the test cartridge enters the analyzer, the integrated camera reads the barcode. The calibration data for the actual lot are transferred to the instrument and used for calculating the results. Calibration by the operator is thus not required.
AFINION™ 2 User Manual
US | 9
Page 10
Getting Started
Installing your analyzer
Place your Anion 2 Analyzer on a dry, clean, stable and horizontal surface. Make sure that the analyzer is located with sufcient surrounding airspace, at least 5 inches on each side. Placement of Anion 2 Analyzer should allow easy disconnection from the wall outlet at any time. Acclimate the analyzer to ambient operating temperature (15-32°C, 59-89°F) before use.
The analyzer might be impaired by:
• Condensing humidity and water
• Heat and large temperature variations
• Direct sunlight
Connecting power supply
- Connect the power cable to the power supply.
- Insert the plug from the power supply into the power socket (Figure 3) in the back of the analyzer.
- Plug in the power supply to a wall outlet.
Only use the power supply and cable supplied with Anion 2 Analyzer. Any other power supplies or cables can damage the analyzer and may cause possi­ble hazards.
Figure 3
1
1 Ethernet port for connection to LIS/HIS/EMR systems. Use shielded cable.
2
2 USB-A connectors for printer, USB ash and barcode reader.
3
3 Power input for power supply connection
• Vibrations (e.g. from centrifuges and dishwashers)
• Electromagnetic radiation
• Movement of the analyzer during processing of a test cartridge
Connecting additional equipment
Optional equipment, not provided with your Anion 2 Analyzer are:
• External barcode reader – for reading barcoded sample or operator identication.
• Printer – for optional print out of test results.
For additional information regarding barcode reader and printer specications, please contact your local Anion 2 supplier.
Connecting the equipment should be done while the analyzer is powered off.
All equipment connected to the USB and/or Ethernet ports must have double or reinforced insulation from mains to prevent the risk of electric shock.
Connectivity
Anion 2 Analyzer can reliably transfer test information to an information system. Use the Ethernet cable to interface the Anion 2 Analyzer to an information system. Anion 2 Analyzer automatically transfers patient and control results to a connected LIS/HIS/EMR system via TCP/IP networking using the protocols POCT1-A, HL7, ASTM 1381-85 (low level) or ASTM 1394-97 (high level), selectable by conguration. ASTM and HL7 protocols support the transfer of patient and QC results. POCT1-A protocol supports in addition functions such as device lockout and operator list management. Operator conguration allows for protection of connectivity settings. When operator conguration is set to operator ID with verication, the conguration of connectivity will only be available for operators at supervisor level. For relevant information, see chapter “Operator conguration”, page 14.
When you export data that contains patient information, it is your responsibility to comply with your local regulations on protection of personal health information.
Anion 2 Analyzer POCT1-A, ASTM and HL7 communication protocols are available at www.alere.com or by contacting your local Anion 2 supplier.
10 | US
AFINION™ 2 User Manual
Page 11
GETTING STARTED
How to power ON the analyzer
1
Power on the analyzer by pressing the ON/OFF button (Figure1). An automatic start-up procedure will be initiated. Please wait. Do not open the lid manually.
Getting Started
2
3
The automatic start-up procedure will be initiated shortly after the analyzer has been powered on. The red light on the top of the analyzer will turn on, indicating that the analyzer is busy. The analyzer is ready for use when the start-up menu is displayed and the green indicator light turns on.
Start-up menu
The analyzer’s software version (SW X.XX) will appear in the upper left corner of the Start-up menu screen. The temperature displayed in the Start-up menu is the operating analyzer temperature. Make sure that the operating temperature is within the recommended range for your Anion Test (see the package insert for the Anion Test). If the analyzer fails during the start-up procedure, an information code will appear referring to a message given in the section “Information codes and troubleshooting”, page 26-28.
How to power OFF the analyzer
Switch off the analyzer by pressing the ON/OFF button (Figure 1). The analyzer should be powered off after the end of a working day.
The analyzer can only be powered off when the cartridge chamber is empty and the lid is closed. If the ON/OFF button is pressed and the lid is open, the message ”Close lid” will appear on the screen.
How to operate the analyzer
The Anion 2 Analyzer has two main user interfaces, the touch screen and the cartridge chamber. The analyzer is easily operated using the touch buttons that appear on the screen. When a button is touched, its function will be activated. Text messages that appear on the screen will help guide you through the testing procedure. The functions of the touch buttons are explained in the section “Gallery of icons”, page 32–34.
The other main operative part of the Anion 2 Analyzer is the cartridge chamber. The cartridge chamber is designed to receive the test cartridge in one orientation only. The lid must be manually closed, but opens automatically. When a new test cartridge is placed in the chamber, manually closing the lid will initiate the analysis. When the analysis is complete the lid will open automatically. The lid protects the cartridge chamber from dust, dirt, light and humidity during processing and when the analyzer is not in use.
• The lid must be manually closed, but opens automatically. Do not open the lid manually.
• Use the ngertips only on the touch screen. Do not use pens or other sharp instruments.
Figure 4
1
2
Screen saver
The screen saver will turn on after 3 minutes, if the touch screen is not in use. To reactivate, touch the screen.
Light signals (the red and green LEDs)
The red diode is illuminated when the analyzer is busy. A ashing red light is seen when an information code is displayed. The green diode is illuminated when the analyzer is ready for use. A ashing green light indicates completion of an analysis.
Sound signals
A short beep indicates completion of an analysis. Two beeps mean that an information code or message is displayed.
1 Text message 2 Touch buttons
3
3 The cartridge chamber with a test cartridge
4
4 The lid in open position
AFINION™ 2 User Manual
US | 11
Page 12
Conguration
The AFINION™ 2 menus
Main menuStart-up menu
Patient ID conguration menu
Language setting
Date/Time menu
Conguration menu
Operator conguration menu
Screen/beeper menu
QC lockout conguration menu
12 | US
General settings menu
AFINION™ 2 User Manual
Page 13
Conguration
Setting the conguration
Before using your Anion 2 Analyzer you should set the conguration according to your needs. To enter the Conguration menu, do the following:
1
Start-up menu
Touch to enter Main menu.
2
3
Main menu
Touch to enter Conguration menu.
Conguration menu
Select an item for conguration (see following pages).
Patient ID conguration
Patient ID enable/disable
The patient identication (ID) function can be enabled or disabled. The patient ID function is enabled as a default setting by the manufacturer. When the patient ID function is enabled, the patient ID must be entered for each test cartridge to be analyzed. If the patient ID function is disabled, a run number will automatically replace the patient ID and be displayed in the upper left corner of the screen. This numbering is reset each day at midnight.
Touch
in the conguration menu to enter the patient ID on/off option.
Select to disable the patient ID function.
Select to enable the patient ID function.
Touch to accept and return to the Conguration menu.
AFINION™ 2 User Manual
US | 13
Page 14
Conguration
Operator conguration
The operator ID function is disabled as a default setting by the manufacturer. Touch
Operator ID enable/disable
Touch
Operator login expiration
Touch
in the Conguration menu to enter the Operator conguration menu.
in Operator conguration menu to enable/disable operator ID.
Select to disable the operator ID function.
Select to enable operator ID. Any operator ID is accepted.
Select to enable operator ID with verication.
· To enable this function at least one supervisor is required to be present in the operator list.
· When operator ID with verication is enabled, analyzer conguration will only be available to the supervisors.
· To log in, the operator ID entered is required to be present in the operator list. See “Operator list management”, page 14.
Touch to accept and return to the Conguration menu.
in the Operator conguration menu to set automatic logout of the operator.
Enter the number of minutes before automatic logout of operator.
The operator will automatically be logged out after the congured number of minutes after ended test.
Touch to conrm and return to previous view.
Operator list management
Touch
1
2
3
in Operator conguration menu to enter operator list.
Touch to add new operator. Touch desired operator ID and touch to delete or to edit the highlighted operator.
Copy operator list
It is possible to copy an existing operator lists between analyzers using a USB ash drive. Insert USB ash in the analyzer USB port. Touch to export operator list from instrument to USB ash. Move USB to the new analyzer and touch existing operator list on the analyzer will be deleted.
Enter new/edit operator ID
Enter new/edit operator ID and touch to enter. Both letters and numbers can be entered (maximum 16 characters). If a barcode reader is connected to the analyzer, a barcoded operator ID can be entered.
Congure the operator level
Select USER to congure user access.
Select SUPERVISOR to congure supervisor access.
Congure tests accessible:
Select the test accessible for this operator.
Touch to return and edit the operator ID.
Touch to accept and store new operator in the operator list. The operator list can store 1000 operator IDs.
Supervisors will be marked with * in the operator list. When analyzer is congured to Operator ID with verication, conguration of the analyzer will only be available to the supervisors.
to import operator list. Any
14 | US
AFINION™ 2 User Manual
Page 15
Conguration
Choosing language
Touch in the Conguration menu to enter the language setting. The default setting by manufacturer is English. Other languages are available.
1
Touch the arrow in the window to view other options. Scroll down until you nd the preferred language.
Touch to accept and return to the Conguration menu.
Adjusting screen/beeper settings
Touch in the Conguration menu to enter the Screen/beeper menu.
Touch to enter the Screen alignment setting.
Touch to enter the Beeper volume setting.
A + sign is shown. Use a blunt pencil and tap the center of the +. Repeat tapping the center of the + each time it is shown. When the process is nished, the previous screen will return.
Adjust the beeper volume by touching or
Touch to conrm and return to the previous view.
Setting date and time
The correct date and time should always be set because the date and time for the analyses are stored and displayed in the patient and control records. The date format is YYYY:MM:DD, where YYYY is the year, MM is the month (01 to 12), and DD is the day (01 to 31). The time format is hh:mm, where hh is the hour from 00 to 23 and mm is minutes from 00 to 59.
Touch in the Conguration menu to enter Date/time menu.
1
2
Touch to enter Date setting.
Touch to enter Time setting.
Enter today’s date or time.
Touch to conrm and return to the previous view.
AFINION™ 2 User Manual
US | 15
Page 16
Conguration
QC lockout conguration
Touch in the Conguration menu to enter the QC lockout conguration menu.
Touch to congure QC lockout for the assay selected.
Touch to congure QC lockout interval. Touch to view/add/delete stored control lots in the control lot database.
1
HbA1c
ACR
2
3
4
Select assay for QC lockout conguration
Touch the arrow in the window to open the drop down menu.
Touch the assay to select.
QC lockout
Select to disable the QC lockout function. No QC runs will be required for this assay. Select to enable the QC lockout function. It is required to run ONE passed control,
control level C I OR C II, to reset the QC lockout interval. Select to enable the QC lockout function. It is required to run TWO passed controls,
both control level C I AND C II, to reset the QC lockout interval. Touch
QC lockout interval
Select Select to congure QC lockout interval by hours.
Touch
Touch
Control lot database
To add a control to the control lot database the Anion Control Data is required. The Anion Control Data is a numeric data string which contains all lot specic data:
• Anion Control lot number
• Control type (assay)
• Control level (C I or C II)
to conrm and return to the previous view.
to congure QC lockout interval by number of runs.
to enter/edit number of runs/hours to QC lockout.
displays the number of runs/hours congured in the QC lockout interval.
to conrm and return to the previous view.
• Control expiry date
• Acceptable control range
• CRC (check sum to validate the previous data)
16 | US
The Anion Control Data and its accompanying barcode is found in the Anion Control Package Insert. If the Anion Control Data is not available, contact your local supplier.
Touch and either manually enter the Control Data or if a barcode reader is connected to the analyzer (recommended), scan the barcode. The Anion Control Data may also be entered before, during or after a control run. The control lot will automatically be stored in the database. See page 25.
Select lot number and touch to delete a control from the list. When a control lot has reached its expiry date, the control will automatically be deleted from the instrument control database. The control lot database can store 100 control lots.
AFINION™ 2 User Manual
Page 17
General settings
Touch
in the Conguration menu to enter the General settings menu.
Conguration
Touch
Touch
Touch
to erase all content and congurations.
to enter Instrument network settings.
to enter Connectivity settings.
Erase all contents and congurations
Touch
in General settings menu to erase all contents and congurations.
Touch
Touch
to erase all content and congurations.
to cancel and return to General settings menu.
Analyzer network settings
See Table 1 for description of the available analyzer network settings.
Touch
to enter Instrument network settings view.
Touch
to congure the network.
Enter the IP Address. Touch
Enter the Gateway. Touch
Enter the Network mask. Touch
Enter the Hostname. Touch to conrm and return to Instrument network settings view.
Touch
Table 1 Description of the available analyzer network settings
Consult your network administrator and LIS/HIS/EMR administrator for required network settings.
DHCP DHCP is turned on/off by selecting “DHCP”.
IP address Insert xed IP address [0-255/0-255/0-255/1-254]
Gateway Insert Gateway [0-255/0-255/0-255/1-254]
Network Mask Insert Network mask [0-255/0-255/0-255/0-255]
Host name Insert Host name. Valid characters are [A-Z], [0-9], [-]. The length can be from 1-16 characters
to accept and return to the General settings menu.
When using DHCP the instrument’s IP address will be assigned by the DHCP Server. No other network settings are necessary. NB! If DHCP is activated, only the hostname setting can be edited.
to conrm and continue to Gateway.
to conrm and continue to Network mask.
to conrm and return to Hostname.
AFINION™ 2 User Manual
US | 17
Page 18
Conguration
Connectivity settings
See Table 2 for description of the available Connectivity settings.
Touch
Touch to enter page 2 of the conguration or to return to the General settings menu
Touch to enter Server IP and Port number, Receiving application (available for ASTM HL, ASTM LL and HL7 only) and Receiving facility (available for HL7 only).
Enter the server IP address: Press
Enter the server port number setting: Touch
Enter the Receiving application setting: Press
Use the button to select Patient ID as (available for HL7 only):
HIS Patient ID
Visit Number
in General settings to enter Connectivity settings
Select appropriate communication protocol
ASTM HL
ASTM LL
HL7
POCT1-A
Communication protocol is disabled as default. Select which results to be transferred to LIS/HIS/EMR by selecting the appropriate button
Patient results only
Patient and control results
Select “New results only” and previous obtained results will not be transferred to the LIS/HIS/EMR
to continue to the Port number setting.
to continue to the Receiving application setting or to close the text input.
to continue to the Receiving facility setting or to close the text input.
Touch
Table 2 Connectivity settings
Consult your network administrator and LIS/HIS/EMR administrator for required connectivity settings.
Protocol ASTM HL ASTM High Level: The communication protocol is based on ASTM E 1394 - 97
Results Patient results only Only patient results will be transferred to the LIS/HIS/EMR
Server IP Insert the IP address of the receiving system [0-255.0-255.0-255.1-254]
Port [0-65535] (0 = not set)
Receiving Application
Receiving Facility (HL7 only)
Patient ID As HIS Patient ID (HL7 only)
For further information about the connectivity settings, see the Anion 2 data sheets for POCT1-A, ASTM and HL7 which can be obtained at www.Alere.com or through your local Anion supplier.
to return to the General Settings menu
ASTM LL ASTM Low Level: The communication protocol is based on ASTM E 1381 - 95
HL7 HL7: The communication protocol is based on HL7 version 2.4
POCT1-A POCT1-A: The communication protocol is based on CLSI: POCT01-A2 Point-of-Care Connectivity; Approved
Disabled Data connectivity is disabled
Patient and quality control
Visit Number (HL7 only)
Standard – Second Edition
Both patient and QC results will be transferred to the LIS/HIS/EMR
(ASTM HL, ASTM LL and HL7 only) Receiving application name (0 – 30 characters)
Receiving facility name (0-30 characters)
18 | US
AFINION™ 2 User Manual
Page 19
Quality Control
Why quality control testing?
Quality control testing should be done to conrm that your Anion 2 System is working properly and provides reliable results. Accurate results for patient samples can only be assured when controls are used routinely and the values are within the acceptable ranges.
Choosing control material
Controls supplied by Alere Technologies AS are recommended for use with the Anion 2 System. These control kits contain control materials with established acceptable ranges for the Anion 2 System.
If you decide to use controls from another supplier, you will need to determine their precision and to establish acceptable control ranges for the Anion 2 System.
Handling and testing controls
Consult the package insert that comes with each control kit for detailed instructions on handling and storage of the control material.
To run a control, follow the procedure in the section “Testing procedures”, page 20-25.
The measured value should be within the acceptable range stated on the control vial label or in the control package insert. If the control results are within the acceptable ranges, patient samples may be tested and results reported.
If the result obtained for a control is out of range, make sure that:
- The control vial has not passed its expiration date.
- The control vial has not passed the declared stability for opened vials.
- The control vial and Anion Test Cartridges have been stored according to recommendations.
- There is no evidence of bacterial or fungal contamination of the control vial.
Correct any procedural error and retest the control material. If no procedural errors are detected, it is recommended to examine the laboratory’s quality control record to investigate the frequency of control failures. Ensure that there is no trend in out-of-range quality control results. Retest the control material using a new control vial.
Patient results must be declared invalid when controls do not perform as expected. Contact your Technical service representative (1.866.216.9505) for advice before analyzing patient samples.
Frequency of control testing
It is recommended that controls are analyzed:
• When starting up an Anion 2 Analyzer for the rst time.
• With each new shipment of Anion Test Kits.
• With each new lot of Anion Test Kits.
• Anytime an unexpected patient test result is obtained.
• When training new personnel in the correct use of the Anion 2 System.
• If national or local regulations require more frequent testing of control materials, perform quality control in compliance with the regulations for your facility.
• Users with a low frequency of testing should analyze controls at least every 30 days.
The controls should always be analyzed if an unexpected test result is obtained (see the Anion Test Package Insert, section Test result reporting). If local, state and/or federal regulations require more frequent testing of control materials, then quality control should be performed in compliance with these regulations. Each laboratory site can benet from establishing a quality control plan. The laboratory director should determine whether additional testing is appropriate for their laboratory.
AFINION™ 2 User Manual
US | 19
Page 20
Testing Procedures
Operating precautions
When operating the analyzer:
• Use your ngertip to operate the touch screen. Do not use pens or other objects that may scratch or damage the screen. Exception: If the screen alignment function is required, you will need to use a blunt pencil.
• The lid opens automatically, but must be closed manually. Do not try to open the lid manually.
• The lid protects the cartridge chamber from dust, dirt, light and humidity. Empty the cartridge chamber and keep the lid closed when the analyzer is not in use.
• If an information code appears on the screen during the analysis, please consult the “Information codes and troubleshooting” section, page 26-28.
• Do not move the analyzer when a test cartridge is being processed.
When handling the test cartridge:
• Do not use test cartridges after the expiration date, or if the test cartridges are not stored in accordance with the recommendations.
• Do not touch the test cartridge optical reading area. Hold the test cartridge by the handle. (Figure 2).
• Do not use the test cartridge if the foil pouch, the desiccant bag or the test cartridge itself is damaged.
• The test cartridges must reach recommended operating temperature before use.
• Do not open the foil pouch until just before use. Once opened, the test cartridge has limited stability.
• Handle and dispose the test cartridges and sample collection equipment as potential biohazardous materials. Use gloves.
• Do not reuse any part of the test cartridge.
Consult the package insert that comes with each Anion Test Kit for assay specic information.
Preparing for an AFINION™ 2 analysis
- Allow the Anion Test Cartridges to reach the recommended operating temperature before use.
- Power on your Anion 2 Analyzer so it is ready for the day’s rst analysis.
- Enter the operator ID (optional). See procedure on page 22.
- The patient ID, control ID or Anion Control Data can be entered before or during processing of the test cartridge in the analyzer. See procedures on page 22-25.
Consult the package insert that comes with each Anion Test Kit for assay specic information.
Tear strip
1
Open the foil pouch. Grip the handle and remove the test cartridge from the pouch. Discard the desiccant bag and foil pouch in suitable waste containers.
When rst opened, the test cartridge has limited stability.
If a barcode reader is connected to the analyzer, a barcoded patient ID, control ID or Anion Control Data can be entered.
2 3
Inspect the cartridge. Do not use the test cartridgeif it is damaged or if loose desiccant particles are found on the test cartridge.
Use the handle to avoid touching the optical reading area.
Mark the test cartridge with the patient or control ID. Use the ID area on the test cartridge. An ID label can also be used.
Do not write on the barcode label or allow it to become wet, dirty or scratched. If an ID label is used, this must t into the ID area.
20 | US
AFINION™ 2 User Manual
Page 21
Testing Procedures
Collecting a sample
• The patient sample material and control material to be used is specic for each Anion Test.
• The length of the capillary in the sampling device, and thereby the sample volume, might also vary for the different Anion Tests.
• The time from lling the capillary until analysing the test cartridge must be as short as possible.
• Do not use the test cartridge if dropped on the bench or oor after the sample has been collected.
Consult the package insert that comes with each Anion Test Kit for assay specic information.
Examples:
1 2 3
Remove the sampling device from the test cartridge.
Use the handle to keep the test cartridge steady against the table and pull the sampling device straight up.
Fill the capillary; hold the sampling device almost horizontally and bring the tip of the capillary in surface contact with the sample. Make sure that the capillary lls completely. It is not possible to overll. Do not wipe off the capillary.
Avoid air bubbles and excess sample on the outside of the capillary.
Immediately and carefully replace the sampling device into the test cartridge.
The time from lling the capillary until analysing the test cartridge must be as short as possible.
Analysing a patient/control sample
1 2 3
Touch to enter the patient sample mode.
Touch to enter the control mode.
The lid opens automatically.
Insert the test cartridge with the barcode label facing left.
Be sure that the test cartridge is correctly placed in the cartridge chamber.
A “C” in the upper left corner indicates that the analyzer is in the control mode.
4 5 6
Touch and enter the patient ID.
Touch to conrm.
Touch and enter the control ID or Anion Control Data.
Touch to conrm.
Entering the patient ID, control ID or Anion Control Data will not interrupt the processing.
Record the result, then touch to accept.
If a printer is connected, touch to print the result.
The lid opens automatically.
The result will be saved in the result records.
Close the lid manually.The analyzer will start processing the test cartridge.
The processing time depends on the test in use.
Remove the used test cartridge from the cartridge chamber and discard it in a suitable waste container.
Insert a new test cartridge or close the lid manually.
Keep the lid closed to protect the cartridge chamber when the analyzer is not in use.
Consult the package insert that comes with each Anion Test Kit for assay specic information.
AFINION™ 2 User Manual
US | 21
Page 22
Testing Procedures
Using the operator ID function
Entering operator ID
If enabled, the operator’s identication (ID) is required before processing an Anion Test Cartridge (see “Operator conguration”, page 14). Both letters and numbers can be entered (maximum 16 characters). The operator ID will be displayed with the result and stored along with the other specic data for this run (see “Patient and control results records”, page 25). If congured to “enabled with verication” the operator ID entered is required to be present in the operator ID list (see “Operator conguration”, page 14).
Enter the operator ID by numbers and/or touch to enter letters. If a barcode reader is connected to the analyzer, a bar coded operator ID can be entered.
Touch to conrm and return to previous view.
The operator will be automatically logged out according to the conguration (see “Operator conguration” page 14).
The operator may also manually logout by using the operator logout button displayed in the Start-up menu.
Using the patient ID function
The patient ID function is enabled as a default setting. As long as this function is enabled, the patient ID must be entered for each patient sample to be analyzed. The patient ID function can be disabled (see “Patient ID conguration”, page 13).
Entering patient ID
It is recommended to enter the patient ID during processing of the test cartridge in the analyzer. Entering the patient ID will not interrupt the processing. It is also possible to enter the patient ID before the processing.
1
2
3
The P-ID 1 will be stored in the memory and displayed along with the other specic data for this run (see “Patient ID conguration” page 13). Patient ID 2-4 will not be displayed in the result records but will be stored in the memory and appear on printouts and data transferred to data management systems.
Touch to enter the patient ID option.
It is possible to enter up to four patient ID entries for each patient, P-ID 1 to 4. When enabled, P-ID 1 is required to be entered. Scrolling between the patient IDs is done with the and .
Enter patient ID by numbers and/or touch to enter letters (maximum 16 characters). If a barcode reader is connected to the analyzer, a barcoded patient ID can be entered.
Touch to conrm and return to previous view.
The entered P-ID 1 will appear on the screen.
The patient ID touch button will remain in the view and it is possible to make corrections.
22 | US
AFINION™ 2 User Manual
Page 23
Testing Procedures
Using the control ID function
In quality control testing, a suitable control ID must always be entered. The lot number of the control material is recommended as a suitable control ID. The control ID function cannot be disabled.
Entering Control ID
It is recommended to enter the control ID during processing of the test cartridge in the analyzer. Entering the control ID will not interrupt the processing. It is also possible to enter the control ID before processing. Both letters and numbers can be entered (maximum 16 characters). The control ID will be stored in the memory and displayed along with the other specic data for this run.
To enter the control ID during processing, do the following:
1
2
3
Touch to enter the control ID option.
Enter control ID by numbers and/or touch to enter letters. Touch to conrm and return to the previous view.
The entered control ID will appear on the screen. The control ID touch button will remain in the view and make corrections possible.
Using the QC lockout function
When the QC lockout function is enabled for one or more assays, approved control testing is required within the congured interval. If the interval expires, patient testing for the assay will be locked. A passed control run must be performed according to conguration to reset the interval or to unlock the assay for patient testing. A failed control run will disable patient testing (see “QC lockout conguration”, page 16).
QC lockout status
The status of the active QC lockouts is presented with a QC lockout status button (padlock symbol) visible in the Start-up menu. This gives the operator the status of QC lockout before he attempts to run any tests. The padlock symbol will only be visible if QC lockout function is enabled for one or more assay types.
The padlock symbols used are:
Enabled-unlocked
All controls are within the congured interval. It is possible to run patient tests for all assays.
Warning-unlocked
All controls are within the congured interval. When one or more of the assays has 10 % or less of the congured interval remaining the
warning icon will be displayed. It is possible to run patient tests for all assays.
Expired-locked
One or more controls have expired according to the congured interval. Patient testing on the expired assay has been locked.
AFINION™ 2 User Manual
US | 23
Page 24
Testing Procedures
Touch the QC lockout status button (padlock symbol) in the Start-up menu to enter the QC lockout status view.
Status
The information is displayed as a list.
Only the assays with QC lockout activated are displayed in this window.
Red text indicates expired assays and orange text indicates assays within warning period.
Control level
How to reset QC lockout interval and/or unlock expired assays.
If no control level is specied, it is required to run ONE passed control, control level C I OR C II, to reset the QC lockout interval and unlock the assay for patient testing.
E.g.
HbA1c: #0
If the control level is specied it is required to run TWO passed controls, both control level C I AND C II, to reset the QC lockout interval and unlock the assay for patient testing.
E.g.
ACR C I: 00.00.00
ACR C II: 00.00.00
Remaining time/runs
Remaining time (dd:hh:mm) or number of runs for each assay with active QC lockout is shown. dd is the number of days, hh is the number of hours, and mm is the number of minutes until the assay will be locked. # is number of patient tests.
Running controls with enabled QC lockout function
When running controls with the QC lockout function enabled, the Anion Control Data is required to be entered or previously stored in the instrument control lot database. See “QC lockout conguration”, page 16.
1) The Anion Control Data is entered before, during or after the control run. If a barcode scanner is connected (recommended) the control data barcode may be scanned. The control lot will automatically be stored in the instrument control database.
2) If the Anion Control Data is previously stored in the instrument control database, the operator will simply need to enter the 8 digit control lot number before, during or after the control run.
If the instrument is congured to QC lockout and the control lot number is not found in the Anion Control database or the Anion Control Data entered is not valid, the instrument will present an option to retry the input or discard the control run result. If discarded, the result will not be stored in the instrument result records.
Passed (result within the acceptable control range)
The result of the control is checked against the acceptable ranges for the corresponding lot number.
If the result is within the limits, a pass symbol is displayed on the screen and the QC lockout interval is reset according to the QC lock­out conguration.
If QC lockout is congured to require two control levels (C I and C II), both levels must pass to reset the lockout interval. Only the interval for the control level used in the test is reset.
Failed (result above or below the acceptable control range)
When a control result is not within the acceptable ranges specied for the control lot, a failed symbol is shown on the screen. The result is stored in the instrument and is sent to the data management system if connected. The QC lockout interval will not be reset.
The arrow symbol will specify whether the result is above or below the acceptable ranges. See “Handling and testing controls”, page 19.
24 | US
AFINION™ 2 User Manual
Page 25
Testing Procedures
Patient and control results records
The patient and control results are stored in the memory of the Anion 2 Analyzer. The last 500 patient results and the last 500 control results are saved in separate records. When exceeding the capacity of 500 results, the oldest result will be deleted. The following parameters are listed for each run: Date and time, run number, patient ID/control ID, operator ID, lot number of test cartridge and the test result.
View, print and export patient and control results
1
Main menu
Touch to enter patient results.
Touch to enter control results.
2
Result records may be exported if an USB ash (FAT 32 formatted) is inserted to the Anion 2 USB port.
Touch to export the results. The results will be stored on the USB in a .txt le for each assay tested on the Anion 2 Analyzer. These les may be opened in e.g. Microsoft Excel for further processing.
When you export data that contains patient information, it is your responsibility to comply with your local regulations on protection of personal health information.
The last patient result or control is displayed
To view more results touch
If a printer is connected, touch
or
to print the result.
AFINION™ 2 User Manual
US | 25
Page 26
Information Codes and Troubleshooting
When an information code appears
Information codes that might appear during use of the Anion 2 Analyzer refer to specic information or error messages. The code numbers, the possible causes and actions to take are listed below.
If the analyzer detects a problem during processing of a test cartridge, the test will automatically be aborted and the test cartridge will be safely moved to the cartridge chamber. Proceed as follows:
1
2
3
Do not reuse a test cartridge that has been rejected by the analyzer. Collect a new sample and repeat the test with a new test cartridge.
Record the code number (#) and touch to accept.
The lid opens automatically.
Remove the test cartridge.
If the test cartridge is not ejected, restart the analyzer.
Do not reuse the test cartridge.
Look up the possible cause from the table below, and take actions to solve the problem.
If the problem persists, contact your local Anion supplier (see “Service information” page 28).
Information codes caused by test-specic limitations
[#] Cause Action to take
103 Hemoglobin too low Consult the Anion HbA1c Package Insert.
104 Hemoglobin too high Consult the Anion HbA1c Package Insert.
105 HbA1c too low Consult the Anion HbA1c Package Insert.
106 HbA1c too high Consult the Anion HbA1c Package Insert.
107 Creatine too high Consult the Anion ACR Package Insert.
108 Blood in urine Consult the Anion ACR Package Insert.
26 | US
AFINION™ 2 User Manual
Page 27
Information Codes and Troubleshooting
Information codes caused by sample or test cartridge
[#] Cause Action to take
201 Insufcient sample volume:
- Empty capillary
- Air bubble in capillary
- Capillary incompletely lled
202 Excess sample on the sampling
device exterior
203 Wrong sample material Repeat the test with a new sample and test cartridge. Ensure that proper sample material is used (see package insert
204 Coagulated sample Repeat the test with a new sample and test cartridge.
Hemolysed blood sample or poor sample quality
Test cartridge or analyzer failure Repeat the test with a new sample and test cartridge.
205 Capillary cracked or damaged Repeat the test with a new sample and test cartridge.
206 Barcode label not readable
(dirty or damaged)
207 - No sampling device inserted Repeat the test with a new sample and test cartridge. Ensure that the correct sampling device is in place and that the
- Sampling device belongs to another Anion Test
- Label on sampling device not readable (dirty or damaged)
208 Test cartridge previously used Repeat the test with a new sample and test cartridge.
209 Test cartridge has passed
expiration date
The date in the analyzer is incorrectly set
210 Test cartridge temperature
too low
211 Test cartridge temperature
too high
212 Software upgrade is required
to run this test.
213
Test cartridge or analyzer failure Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.
214
215 Test cartridge or analyzer failure Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.
Hemolysed blood sample or poor sample quality (Anion HbA1c)
217 Hemolysed blood sample or
poor sample quality (Anion HbA1c)
218 Condensation detected on
cartridge
Repeat the test with a new sample and test cartridge. Ensure that the capillary is completely lled with no air bubbles (see package insert for the Anion Test in use).
Repeat the test with a new sample and test cartridge. Ensure that only the tip of the capillary is in contact with the sample (see package insert for the Anion Test in use).
for the Anion Test in use, section “Specimen collection and storage”).
The time from lling the capillary until analyzing the test cartridge should be as short as possible.
Consult the Anion Package Insert. Repeat the test with a new sample and test cartridge.
If the problem persists, restart the analyzer and run controls.
Inspect the sampling device before use and handle with care.
Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.
sampling device label is clean.
Check expiry date on the foil pouch or kit container. Repeat the test using a new sample and a new test cartridge fromanother lot.
Check the date in the analyzer to make sure it is set correctly. Repeat the test with a new sample and test cartridge.
Repeat the test with a new sample and a new test cartridge. Ensure that the operating temperature is within acceptable range (see package insert for the Anion Test in use).
Repeat the test with a new sample and a new test cartridge. Ensure that the operating temperature is within acceptable range (see package insert for the Anion Test in use).
Contact your local supplier for assistance.
Consult the Anion HbA1c Package Insert. Repeat the test with a new sample and test cartridge.
Consult the Anion HbA1c Package Insert. Repeat the test with a new sample and test cartridge.
Run a new test cartridge. Ensure that the cartridge is equilibrated to room temperature before the foil pouch is opened.
Information codes and messages caused by analyzer failure
[#] Cause Action to take
27 2829Start-up procedure failed Contact your local supplier for assistance.
Self-test
error.
Analyzer
in non-
operative
mode
Analyzer failure Restart analyzer. If the problem persists, contact your local Anion 2 supplier.
301 Self-test failed Restart the analyzer.
302 Analyzer failure Restart the analyzer and run controls. Repeat the test with a new sample and test cartridge.
303 Analyzer temperature is too
high
304 Analyzer temperature is
too low
Ensure that the operating temperature is within recommended range (15-32ºC, 59-89 °F). Wait until the analyzer has cooled down. Repeat the test with a new sample and test cartridge.
Ensure that the operating temperature is within recommended range for the Anion Test in use (see package insert). The analyzer temperature is displayed in the Start-Up menu (see page 11). Repeat the test with a new sample and test cartridge.
AFINION™ 2 User Manual
US | 27
Page 28
Information Codes and Troubleshooting
[#] Cause Action to take
305 Printer improperly connected
Malfunction of the printer
Touch
screen
error
Touch screen failure/ Touch screen buttons do not respond accurately
Other information codes
[#] Cause Action to take
401 No registered supervisors in
operator list
402 Cannot delete last
supervisor
403 This assay type is not
accessible to the operator
404 Operator ID is not found in
operator list
[#] Cause Action to take
501 The control lot has passed
expiration date
502 Anion Control Data is not
recognised and is not stored in control lot database
503 Control verication aborted. The Anion Control Data entered was not recognised. The control test was aborted by the operator. Test result was
504 Required control test interval
has expired. Patient testing is disabled for this assay.
Switch off the analyzer, reconnect the printer and restart the analyzer. If the message persists, see the printer user manual.
Restart analyzer and realign screen.
At least one supervisor is required in the operator list when the analyzer is congured to operator ID veried, (see page 14).
At least one supervisor is required in the operator list when the analyzer is congured to operator ID veried, (see page 14).
The operator logged in does not have access to run this assay type. Please contact your supervisor.
When Operator ID with verication is enabled, operator ID entered is required to be present in the operator list, (see page 14). Please contact your supervisor.
Check the expiration date on the control lot package insert or kit box. Repeat the test using a sample from a new control lot.
Reenter the Anion Control Data, (see page 16).
not stored. Run new control test to reset QC lockout interval.
A passed control run must be performed according to conguration to unlock this assay for patient testing.
[#] Cause Action to take
601 Operator list or control lot
database is full
The operator list can store 1000 operators and the control lot database can store 100 control lots. Delete an operator or control lot to enter a new item
Service information
The laboratory must notify the manufacturer of this test system of any performance, perceived or validated, that does not meet the performance specications as outlined in the instructions.
The manufacturer provides a toll free line for technical support: 1.866.216.9505
The toll free number is available for use only in the United States of America.
Before asking for assistance, please record the following information:
• Anion 2 serial number (SN) – see label on the backside of the analyzer
• Software version number – see Start-up menu.
• Anion Test type
• Test cartridge or kit lot number – see foil pouch or kit container
• Control identication and lot number – see vial label
• Control results obtained
• Description of the problem with reference to information codes or messages
28 | US
AFINION™ 2 User Manual
Page 29
Maintenance
Cleaning and maintenance
No maintenance of the Anion 2 Analyzer is required other than cleaning the exterior and cartridge chamber.
Cleaning the exterior
Cleaning the exterior of the Anion 2 Analyzer should be performed whenever necessary. Most spills and stains can be removed with water or a mild detergent.
- Power off the analyzer. Unplug the power supply when the shut down procedure is completed.
- Clean the outside of the analyzer and the touch display with a clean, lint-free and non-abrasive cloth dampened in water or a mild detergent.
- To disinfect the exterior of the analyzer use a 1:10 solution of household bleach (i.e., 0.5% sodium hypochlorite), 2% glutaraldehyde solution or 70% alcohol solution. The surface of the analyzer should be exposed to the disinfectant for at least 10 minutes.
- Allow the analyzer to air dry.
- Plug in the power supply and power on the analyzer.
• The analyzer must be powered off and unplugged before cleaning.
• Do not use any cleaning liquid or equipment other than those recommended above.
• Do not immerse the analyzer in water or other liquids.
Cleaning the cartridge chamber
The Cleaning Kit (
The cartridge chamber should be cleaned immediately if materials or liquids are spilled in the cartridge chamber. For regular maintenance (removal of dust particles etc.), the cartridge chamber should be cleaned every 30 days.
- Touch to open the lid.
- Unplug the power supply.
- Wet a Cleaning Swab with 3 drops of water and gently rinse the cartridge chamber. To disinfect the surface, use a 1:10 solution of household bleach (i.e., 0.5% sodium hypochlorite), 2% glutaraldehyde solution or 70% alcohol solution). Do not soak.
- Carefully remove spills and particles from the cartridge chamber using the moistened swab.
- To disinfect the cartridge chamber, the surface of the chamber should be exposed to the disinfectant for at least 10 minutes1.
- Wipe off any residual liquid from the cartridge chamber using a new, dry Cleaning Swab.
- Plug in the power supply, and power on the analyzer by pressing the on/off button.
- The lid will close automatically during the self-test. If it doesn’t, then close it manually and restart the analyzer.
1116048) should always be used for cleaning the cartridge chamber.
• The analyzer must be unplugged before cleaning.
• Do not use any cleaning liquid or equipment other than those recommended above.
• Do not allow liquid to drip off the Cleaning Swab into the analyzer. If liquid drips into the analyzer, optics can be destroyed.
• Do not immerse the analyzer in water or other liquids.
• Do not move or tilt the analyzer when cleaning the cartridge chamber.
Disposal of the analyzer
For correct disposal according to the Directive 2012/19/EU (WEEE), contact your local Anion 2 supplier.
Software upgrade
Consult the Anion USB Flash Drive Package Insert for software upgrade.
1
Clinical and Laboratory Standards Institute (CLSI) Guideline M29-A3: ”Protection of Laboratory Workers From Occupationally
Acquired Infections; Approved Guideline - Third Edition”. ISBN 1-56-238-567-4
AFINION™ 2 User Manual
US | 29
Page 30
Warranty
Alere Technologies AS warrants solely to the Buyer that the Anion 2 Analyzer will be free from defects in materials and workmanship, when given normal, proper and intended usage, and will perform in accordance with Alere Technologies AS’s specications for a period of twelve months from the date of delivery.
At its expense, Alere Technologies AS agrees to repair, or at Alere Technologies AS’s option, replace with a new or reconditioned unit, any Anion 2 Analyzer which is under warranty and not performing substantially in accordance with applicable product specications, provided that the Buyer has given Alere Technologies AS notication of such warranty claim within the warranty period. If Alere Technologies AS is unable after reasonable efforts to repair or replace the Anion 2 Analyzer not performing substantially in accordance with applicable product specications, the Buyer’s sole remedy shall be the refund of an amount not to exceed the actual purchase price paid by the Buyer for the Anion 2 Analyzer. All repairs will be done during normal working hours. All replaced parts shall become Alere Technologies AS’s property. Alere Technologies AS may require the Buyer to ship the Anion 2 Analyzer to Alere Technologies AS or elsewhere at Alere Technologies AS’s expense, for warranty service to be performed.
Notwithstanding the foregoing, Alere Technologies AS shall have no obligation to make repairs, replacements or corrections which result, in whole or in part, from (i) an act of God or other unforeseen catastrophe, (ii) any error, omission or negligence of the Buyer, (iii) improper or unauthorized use of the Anion 2 Analyzer, (iv)operating errors or the disregard of warnings and pre-cautions described in this Anion 2 Analyzer User Manual; (v) repairs performed to the Anion 2 Analyzer by any person other than an authorized Alere Technologies AS service representative; (vi) use of the Anion 2 Analyzer in a manner for which it was not designed, (vii) causes external to the Anion 2 Analyzer such as, but not limited to, power failure or electric power surges, or (viii) use of the Anion 2 Analyzer in combination with equipment, components or software not supplied by Alere Technologies AS.
EXCEPT AS STATED IN THIS SECTION OF THE USER MANUAL, ALERE TECHNOLOGIES AS DISCLAIMS ALL WARRANTIES, WHETHER EXPRESS OR IMPLIED, WRITTEN OR ORAL, WITH RESPECT TO THE AFINION 2 ANALYZER, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. ALERE TECHNOLOGIES AS’S MAXIMUM LIABILITY ARISING OUT OF THE SALE OF THE AFINION 2 ANALYZER OR ITS USE, WHETHER BASED UPON WARRANTY, CONTRACT, TORT OR OTHERWISE, SHALL NOT EXCEED THE ACTUAL PURCHASE PRICE PAID BY THE BUYER FOR THE AFINION 2 ANALYZER. IN NO EVENT SHALL ALERE TECHNOLOGIES AS BE LIABLE FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOT LIMITED TO, LOSS OF PROFITS, LOSS OF DATA OR LOSS OF USE DAMAGES, ARISING HEREUNDER OR FROM THE SALE OF THE AFINION 2 ANALYZER. THIS WARRANTY MAY NOT BE TRANSFERRED BY THE BUYER.
The acknowledgement of claims shall be reported to your Technical Care Specialist at 1.866.216.9505
30 | US
AFINION™ 2 User Manual
Page 31
AFINIONTM 2 Analyzer
Analyzer
Size 200 mm W x 186 mm H x 328 mm D
Weight 3.4 kg
Display Standard LCD colour display with back light and integrated touch panel.
Resolution: 240 x 320 pixels. Visible area: 58 x 77 mm.
Camera 640 x 480 pixels
Capacity of result records 500 patient results and 500 control results
Capacity of operator list 1000 operators
Capacity of control lot database 100 control lots
SW update Via USB ash drive
Communication interface USB 2.0 High Speed, Ethernet 10/100 Mbps
Power supply
Power supply Separate AC to DC power supply. Double insulated.
Input 100-240 VAC, 50-60 Hz
Output 24 VDC ± 5% , 1.75 A, 42 W
Output connector 0.2 x 0.1 in. / 5.5 x 2.5 mm plug. Positive (+) on inner pin.
Operating conditions
Temperature 15-32°C (59-89°F)
Relative humidity 10-80%, non-condensing
Altitude Max 4000 MASL
Location Dry, clean, horizontal surface. Avoid direct sunlight.
Test cartridge temperature According to specications for the Anion Assay in use.
Storage and transport (in the original container)
Temperature -40 to 70ºC (-40 to 158°F)
Relative humidity 10-93 % at 40ºC
Technical Specications
Additional equipment
For information regarding recommended barcode reader, printer, the Anion Analyzer Cleaning Kit or USB ash drive, please call 1.866.216.9505.
AFINION™ 2 User Manual
US | 31
Page 32
Gallery of Icons
The touch buttons and their function
Touching a button on the screen will activate the function of this button. All the touch buttons that may appear during operation of the Anion 2 Analyzer are explained below by their function.
Start-up menu
Main menu
Conguration menu
Patient sample mode Select patient sample mode.
Control mode Select control mode.
Main menu Enter Main menu (operator ID, patient records, control records and conguration menu).
QC lockout status
QC lockout status
QC lockout status
Operator logout button
Patient records View patient result records. View, print or export patient results.
Control records View control result records. View, print or export control results.
Conguration menu Enter conguration menu (language, patient ID on/off, date/time and screen/volume).
Patient ID conguration menu
Operator conguration menu
Language setting Enter language conguration.
Enabled-unlocked All controls are within the congured interval. It is possible to run patient tests for all assays.
Warning-unlocked All controls are within the congured interval. When one or more of the assays has 10 % or less of the congured interval remaining the warning icon will be displayed. It is possible to run patient tests for all assays.
Expired-locked One or more controls have expired according to the congured interval. Patient testing on the expired assay has been locked.
Manual operator logout button.
Congure patient ID function.
Congure operator function.
Patient ID conguration menu
Operator conguration menu
Patient and control records
Universial buttons
Screen/Volume menu Congure screen and volume settings (screen contrast, screen adjustment and beeper volume).
Date/Time menu Enter date/time settings (date and time).
QC lockout conguration menu
General settings menu Enter the general settings menu.
Patient ID disabled Patient ID disabled.
Patient ID enabled Patient ID enabled and required.
Operator ID conguration
Automatic operator logout
Operator list Manage operator list. View, add, edit and delete operators.
Print Print result on connected printer.
Result records export Export result records to connected USB ash.
Patient ID Enter patient ID.
Control ID Enter control ID.
Enter Enter and return to previous view.
Backspace Delete previous character.
Congure QC lockout function.
Congure operator ID function.
Congure number of minutes before automatic logout of operator.
32 | US
Increase Increase contrast/volume.
AFINION™ 2 User Manual
Page 33
Menu Touch button Name Function
Decrease Decrease contrast/volume.
Scroll up View previous
Scroll down View next
Exit Exit current menu and return to previous screen view.
Accept Accept (a setting or a test result).
Abort Abort the test result or cancel operation.
Add button Add new operator or control lot.
Delete button Delete operator or control lot.
Edit button Edit QC lockout interval or operator ID.
Operator ID conguration
Language settings
Screen/Beeper menu
Operator ID disabled Operator ID function is disabled.
Operator ID enabled Operator ID is required to be entered to run an Anion Test Cartridge
Operator ID enabled with verication
Language Enter language conguration.
Screen alignment Enter screen alignment function.
Beeper volume Enter beeper volume setting.
Operator ID is required to be entered to run an Anion Test Cartridge. The operator ID is veri­ed against the analyzer operator list.
Gallery of Icons
Date/Time menu
General settings menu
QC lockout conguration menu
Operator list Operator list export Export operator list from analyzer to USB ash.
QC lockout QC lockout disabled QC lockout is disabled for this test.
QC lockout interval
Date Enter date setting.
Time Enter time setting.
Erase Erase all content and congurations.
Instrument network settings
Connectivity settings Enter connectivity settings
QC lockout Enable/disable QC lockout function.
QC lockout interval Congure QC warning and lockout interval.
Control lot information View, add or delete control lots stored on analyzer.
Operator list import Import operator list from USB ash to analyzer.
QC lockout enabled One passed control run of either C I or C II is required to reset QC lockout interval.
QC lockout enabled Two passed control runs, C I and C II are required to reset QC lockout interval.
Interval by number of patient tests
Interval by number of hours
Enter analyzer network settings
QC reminder and lockout active after a congured set of patient tests.
QC reminder and lockout active after a congured set of hours.
AFINION™ 2 User Manual
US | 33
Page 34
Gallery of Icons
Other symbols and signs
Other symbols, signs and abbreviations that may appear during operation of the Anion 2 Analyzer are explained below. These symbols or signs are only inform­ative and can not be activated like the buttons.
Symbol Meaning Appears when?
Wait! Hour-glass icon that appears in the start-up procedure.
Information code
Operator ID Icon illustrates the operator ID.
Patient ID Icon illustrates the patient ID.
Control ID Icon illustrates the control ID.
Connected
Quality control pass Control result is within acceptable range.
Quality control failed Control result is outside acceptable range.
Result is above acceptable range
Result is below acceptable range
C Control The letter C will appear on the screen when the control mode is selected.
O-ID Operator ID Abbreviation used in the patient and control records.
P-ID Patient ID Abbreviation used in the patient records.
C-ID Control ID Abbreviation used in the control records.
RUN# Run number
LOT# Lot number Abbreviation used in the patient and control records for the lot number of the test cartridge.
Icon used along with a code number [#] that corresponds to code specic information messages [#] (see “Information codes and troubleshooting”).
The instrument is connected to the LIS/HIS/EMR server. When no symbol, the instrument is not connected to the LIS/HIS/EMR server.
The displayed control result is above acceptable range.
The displayed control result is below acceptable range.
Abbreviation used in the patient and control records for the run number of the analysis. This numbering is reset each day at midnight.
USER User Operator with user privileges.
SUPERVISOR Supervisor Operator with supervisor privileges.
34 | US
AFINION™ 2 User Manual
Page 35
Symbols and Abbreviations
LOT
TEST CARTRIDGE
CONTROL C I
CONTROL C II
CLEANING KIT
N
CAN/CSA 22.2 No. 61010-1
The following symbols and abbreviations are used in the product labeling of the Anion 2 System.
Symbol/Abbreviation Explanation
The product conforms to all applicable EC Directives and Regulations
IVD
In vitro diagnostic medical device
Catalogue number
Lot number
Serial number
Test cartridge
Symbol/Abbreviation Explanation
UL 61010-1
Conformity to the RoHS 2 directive
Conformity to the technical regulations for EurAsian Conformity Mark
Conformity to the North American product safety standards
Direct current
USB port
Control C I
Control C II
Cleaning kit
Waste electrical and electronic equipment (WEEE)
Biological risks
Contents sufcient for ”n” number of tests
Expiration date
Storage temperature limitations
Manufacturer
Date of manufacture
Fragile, handle with care
Keep away from sunlight
Keep dry
Warnings and precautions
Consult the Anion User Instructions
Ethernet port
Double insulation
LED Light emitting diode
PC Personal computer
ID Identication
HIS Hospital information system
LIS Laboratory information system
LCD Liquid crystal display
AC Alternating current
DC Direct current
ASTM American Society for Testing and Materials
HL7 Health Level Seven
POCT1-A
EMR Electronic medical record
DHCP Dynamic host conguration protocol
IP Internet protocol
Point-of-Care Connectivity; approved standard
AFINION™ 2 User Manual
US | 35
Page 36
Alere Technologies AS
Kjelsåsveien 161 P.O. Box 6863 Rodeløkka NO-0504 Oslo, Norway
www.alere.com
ISO 13485 certied company
© 2018 Abbott.All rights reserved.All trademarks referenced are trademarks of either the Abbott group of companies or their respective owners.
1116752 Rev. A 2018/09
Loading...