ABB SM1000 Operating Manual

User Guide
IM/SM1000 Issue 12
Videographic Recorder
SM1000
ABB
The Company
EN ISO 9001:2000
We are an established world force in the design and manufacture of instrumentation for industrial process control, flow measurement, gas and liquid analysis and environmental applications.
Cert. No. Q 05907
As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide.
EN 29001 (ISO 9001)
We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and support.
The quality, accuracy and performance of the Company's products result from over 100 years experience, combined with a continuous program of innovative design and development to
incorporate the latest technology.
The UKAS Calibration Laboratory No. 0255 is just one of the ten flow calibration plants operated
Stonehouse, U.K.
by the Company and is indicative of our dedication to quality and accuracy.
0255
Electrical Safety
This equipment complies with the requirements of CEI/IEC 61010-1:2001-2 'Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use'. If the equipment is used in a manner NOT specified by the Company, the protection provided by the equipment may be impaired.
Symbols
One or more of the following symbols may appear on the equipment labelling:
Warning – Refer to the manual for instructions Direct current supply only
Caution – Risk of electric shock Alternating current supply only
Protective earth (ground) terminal Both direct and alternating current supply
Earth (ground) terminal
Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual for any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of the Technical Publications Department.
Health and Safety
To ensure that our products are safe and without risk to health, the following points must be noted:
1. The relevant sections of these instructions must be read carefully before proceeding.
2. Warning labels on containers and packages must be observed.
3. Installation, operation, maintenance and servicing must only be carried out by suitably trained personnel and in accordance with the information given.
4. Normal safety precautions must be taken to avoid the possibility of an accident occurring when operating in conditions of high pressure and/or temperature.
5. Chemicals must be stored away from heat, protected from temperature extremes and powders kept dry. Normal safe handling procedures must be used.
6. When disposing of chemicals ensure that no two chemicals are mixed.
The equipment is protected through double insulation
Safety advice concerning the use of the equipment described in this manual or any relevant hazard data sheets (where applicable) may be obtained from the Company address on the back cover, together with servicing and spares information.
Videographic Recorder
SM1000 Contents
Contents
1 Introduction .....................................................................3
2 Operation .........................................................................5
2.1 Powering up the Instrument .....................................5
2.2 Displays and Controls ..............................................5
2.3 Chart Views .............................................................7
2.3.1 Electronic Signatures ..................................14
2.4 Bargraph Views ......................................................15
2.5 Digital Indicator View ..............................................17
2.6 Process View .........................................................20
2.7 Alarm Event Log .....................................................23
2.8 Totalizer Log ..........................................................25
2.9 Audit Log ...............................................................27
2.10 Status View ............................................................28
3 Setup .............................................................................29
3.1 Introduction ............................................................29
3.2 Accessing the Setup Level .....................................29
3.3 Password Entry ......................................................31
3.4 Setup Menu ...........................................................32
3.5 Archiving ................................................................34
3.5.1 Card Compatibility ......................................34
3.5.2 Media Status ..............................................34
3.5.3 Inserting and Removing Media ....................35
3.5.4 Archive File Types .......................................36
3.6 Text Format Archive Files .......................................37
3.6.1 Text Format Channel Data Files .................. 37
3.6.2 Text Format Filename Examples ................. 38
3.6.3 Text Format Log files ..................................38
3.6.4 Text Format Data File Examples ..................39
3.6.5 Text Format Data File Digital Signatures ......39
3.6.6 Text Format Data Verification and Integrity ..39
3.7 Binary Format Archive Files ....................................41
3.7.1 Binary Format Archive Filenames ................41
3.7.2 Binary Format Channel Data Files ...............41
3.7.3 Binary Format Log files ...............................41
3.7.4 Binary Format Data File Examples ...............41
3.7.5 Binary Format Data Verification
and Integrity ................................................41
3.8 Archiving Online/Offline ..........................................44
3.9 Backing-up Archived Data .....................................44
3.10 Archive Wrap .........................................................44
4 Configuration ................................................................ 45
4.1 Introduction ........................................................... 45
4.1.1 Configuration Level Security ....................... 45
4.1.2 Configuration Level Access ........................ 45
4.2 Overview of Configuration ...................................... 48
4.2.1 Adjusting Screen Contrast .......................... 48
4.3 Making Changes to Parameters ............................. 49
4.4 Common Configuration ......................................... 52
4.4.1 Setup ......................................................... 52
4.4.2 Screen ....................................................... 53
4.4.3 Time ........................................................... 54
4.4.4 Security ...................................................... 56
4.4.5 Users ......................................................... 58
4.4.6 Logs ........................................................... 60
4.4.7 Operator Messages .................................... 60
4.5 Process Group Configuration ................................. 61
4.5.1 Setting the Recording Parameters .............. 61
4.5.2 Configuring the Chart View ......................... 63
4.5.3 Configuring the Bargraph View ................... 66
4.5.4 Configuring the Process View ..................... 67
4.5.5 Configuring the Digital Indicator View .......... 68
4.5.6 Archiving .................................................... 70
4.6 Channel Configuration ........................................... 72
4.6.1 Recording Channel Setup .......................... 72
4.6.2 Analog Input Configuration ......................... 74
4.6.3 Digital Input Configuration .......................... 77
4.6.4 Alarm Configuration .................................... 78
4.6.5 Totalizer Configuration ................................ 84
4.7 Functions .............................................................. 87
4.7.1 Custom Linearizers .................................... 87
4.7.2 Real-time Alarms ........................................ 88
4.8 I/O Module Configuration ....................................... 89
4.8.1 Analog Inputs ............................................. 89
4.8.2 Relay Modules ........................................... 90
4.8.3 Ethernet Modules ....................................... 90
4.8.4 RS485 (Modbus™) Communications ......... 91
4.8.5 Comms Analog Input ................................. 92
4.8.6 Comms Digital Input ................................... 93
4.8.7 Hybrid Modules .......................................... 94
4.9 Modbus TCP ......................................................... 95
IM/SM1000 Issue 12 1
Videographic Recorder
SM1000 Contents
5 Installation .....................................................................96
5.1 Siting ......................................................................96
5.2 Mounting ................................................................97
5.3 Electrical Connections ............................................98
5.4 Analog Inputs .......................................................100
5.4.1 Current and Voltage ..................................100
5.4.2 Thermocouple ...........................................100
5.4.3 Resistance Thermometer (RTD) .................100
5.4.4 Transmitter Power Supply .........................100
5.5 RS422/485 Serial Communications ......................101
5.6 Mains Power Connections ....................................102
5.7 Relay Output Board Connections .........................102
5.8 Hybrid I/O Module Connections ............................103
5.8.1 Digital Output Connections .......................103
5.8.2 Digital Input Connections ..........................103
5.8.3 Analog Output Connections ......................103
Appendix A – Signal Sources ..........................................104
Appendix B – Modbus Guide ..........................................106
B.1 Introduction ..........................................................106
B.2 Setting Up ............................................................106
B.3 Modbus Commands Supported ...........................106
B.4 Modbus Exception Responses .............................106
B.5 Operating Mode Modbus Coils .............................106
B.6 Operating Mode Modbus Registers ......................112
B.7 Comms. Analog and Digital Inputs ........................114
Appendix C – Storage Capacity ..................................... 116
C.1 Internal Storage Capacity .....................................116
C.2 Archive Storage Capacity ..................................... 117
Appendix D – Default Settings .......................................119
D.1 Company Standard .............................................. 119
D.1.1 Common Configuration ............................ 119
D.1.2 Process Groups 1 to 6 .............................119
D.1.3 Recording Channels .................................120
D.1.4 I/O Modules ..............................................120
D.1.5 Functions .................................................121
D.2 QuickStart Templates ..........................................121
D.2.1 QSMilliAmp ............................................... 121
D.2.2 QSFlow .................................................... 121
D.2.3 QSTHC_C ................................................ 121
D.2.4 QSTHC_F .................................................121
D.2.5 QSRTD_C ................................................ 121
D.2.6 QSRTD_F .................................................121
D.2.7 QSDEMO ................................................. 122
D.2.8 QSDAIRY .................................................123
Appendix E – Spare Parts and Accessories .................. 124
Appendix F – Error & Diagnostics Information .............. 126
Appendix G – Symbols and Icons ..................................127
Index .................................................................................128
2 IM/SM1000 Issue 12
Videographic Recorder
SM1000 1Introduction

1Introduction

Functional Overview – Fig. 1.1
12 Recording Channels as standard, divided into 2 Process Groups, each with 6 Recording Channels.
Two Alarms and one Totalizer (if Totalizer option is enabled) are assigned to each Recording Channel.
Signal sources derived from universal analog inputs, the Modbus serial link, optional digital inputs or internal analog and digital
signals.
Any source can be assigned to any recording channel.
Data from assigned sources can be displayed in:
– Vertical or Horizontal Chart view format
– Vertical or Horizontal Bargraph view format
– Digital Indicator view format
– Process view format
Three instrument logs record alarm events, totalizer values (if totalizer option is enabled) and system/configuration changes.
Screen Capture facility – saves an image of any of the operator views to external archive media provided external archive media
with sufficient free space is inserted in the instrument. It is not necessary for archiving to be 'online'.
Shunt Resistors
(1 per analog input)
Mounting Clamps
Keys
Standard Accessories
Note. For optional accessories, refer to Appendix E on
page 124.
Compact Flash Card
(If fitted with a compact flash drive)
IM/SM1000 Issue 12 3
Videographic Recorder
SM1000 1Introduction
Signal Sources
Note. Signal sources
can be assigned to any recording channel in either process group.
Analog Inputs
A1
A2
A3
A4
A5
A6
B1
B2
B3
B4
B5
B6
Internal sources  see Appendix A
Comms. Analog Inputs
1
2
3
24
Comms. Digital Inputs
1
2
3
24
Digital Inputs
C1
C2
C3
C4
C5
C6
D1
D2
D3
D4
D5
D6
Process Group 1
1.1
1.2
1.3
1.4
1.5
 Two process groups as
standard provide 12 recording channels, irrespective of the number of external inputs.
Process Group 2
2.1
2.2
2.3
2.4
2.5
2.6
Recording Channels
2.1 to 2.6
Instrument Logs
Alarm/Event Log
Totalizer Log
Audit Log
Configuration Data
Configuration Data
Stores all instrument configuration, calibration and user preferences
Records all alarm transitions and all operator messages
Records all totalizer activity*
Records all system acfivity
*If totalizer option fitted
Recording
Channels
1.1 to 1.6
1.6
Digital Indicator View (Section 2.5) Process View (Section 2.6)
Chart View Bargraph View Digital Indicator View Process View
Alarm Event Log (Section 2.7)
Totalizer Log (Section 2.8)
Bargraph View (Section 2.4)Chart View (Section 2.3)
Audit Log (Section 2.9)
 Views provide a window on the stored data.
 Status View
Recorded Data
Recording Channels 1.1 to 1.6 (Process Group 1)
Recording Channels 2.1 to 2.6 (Process Group 2)
Instrument Logs
Configuration Data
Files stored in instrument's on-board flash memory. Newest data overwrites oldest.
Archive Data
Data saved to archive storage media
Archive Media
Fig. 1.1 Functional Overview
4 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation

2Operation

2.1 Powering up the Instrument
When power is first applied to the instrument, its processor carries out a number of self-tests and displays the start-up screen.
At the end of the start-up sequence the instrument displays the Operator View that was being displayed when the instrument was powered down.
2.2 Displays and Controls – Figs. 2.1 and 2.2
In normal day-to-day use, the instrument is operated via the Operator Keys located along the bottom of the screen.
Process Group Name
Short Channel Tag
Current Value
Alarm Status
Engineering Units
Channel not enabled
Operator Keys
Programming Socket
(Inside Door)
Door Lock
Group Key
Selects a different process group.
Process Group 1
View Key
Selects a different process view or log.
Chart View  see Section 2.3
Status Icons – see Appendix G, page 127
Current Date and Time
Digital ON or OFF tag, according to input status
Infrared Link
Media Door
Jacking Screw and Fitting for Tamper-evident Seal
Menu Key
Displays or hides the context-sensitive operator menu associated with each view.
Bargraph Views  see Section 2.4
Process Group 2
Instrument Status & Logs
Also cancels the menu without making a change or returns to the previous menu level.
Digital Indicator View  see Section 2.5
Process View  see Section 2.6
Up/Down Keys
Highlights menu items and scrolls through previously recorded data.
Enter Key
Selects the highlighted menu item. If 'Screen capture' is set to 'Enabled' (see Section 4.4.1, page 52) and external archive media is inserted in the instrument, the operator can save a snapshot of any operator view to the external media, if pressed when the menu is not open.
Fig. 2.1 Displays and Controls
IM/SM1000 Issue 12 5
Videographic Recorder
SM1000 2Operation
Process Group 1
(Recording Channels 1.1 to 1.6)
Chart View
Vertical Bargraph
Horizontal Bargraph
Process Group 2
(Recording Channels 2.1 to 2.6)
Vertical Bargraph
Horizontal Bargraph
Digital Indicator
Process View
Fig. 2.2 Overview of Operator Displays
Digital Indicator
Process View
Note. Only process groups and views that have been enabled are displayed – see Section 4.5, page 61.
6 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
2.3 Chart Views – Fig. 2.3
Units
Short Channel Tag
Current Value
(see Note 1 on Page 8)
Alarm Event Icon
(see Note 2 on Page 8)
Scale Bar
Chart Trace
Time Stamp
Operator Message Annotation
(see Note 3 on Page 8)
Alarm Event Annotation (see Note 3 on Page 8)
Selects next Process Group
(if enabled) or Status View
Group Tag
Selects the next enabled
view in the group
(Bar, Digital or Process)
Status Bar
Status Icons
Opens the Operator Menu for the current view
Global Alarm Status Icon (see Note 1 on Page 8)
Current Date and Time
Digital On/Off Tag
Channel Not Used
Newest Data
Screen Interval
Operator Message (appears for 1 second)
Oldest Data
Newest Data
For a description of the Status Icons (displayed in the Status Bar) and Alarm Event Icons
(displayed in the Bargraph, Digital Indicators, Process and Chart Views), see Appendix G, page 127.
Screen Interval
Oldest Data
Fig. 2.3 Horizontal and Vertical Chart Views
Note. Do not remove media while either of the media update in progress status icons ( or ) are displayed.
IM/SM1000 Issue 12 7
Videographic Recorder
SM1000 2Operation
Note.
1. Current Values
The Current Value, shown on the digital indicators at either the top (vertical chart view) or right hand side (horizontal chart view) of the screen, is the latest instantaneous value and its update rate is not affected by the recording sample rate.
If the current value in the digital indicator is displayed in red, recording has been stopped for that channel – see Section 3.4, page 32 and see Section 4.6.1, page 72.
Traces are shown only when that particular channel is being recorded. When a channel is set to Stop, its trace continues to be shown for up to one sample period.
2. Alarm Status
– Flashing red alarm event icon – alarm active and unacknowledged
– Continuous red alarm event icon – alarm active and acknowledged
If
any alarm in any process group is active, the Global Alarm status icon ( ) is displayed in the status bar – see Fig. 2.3. If
any active alarm in any process group is unacknowledged, the icon is surrounded by a red flashing border ( ).
3. Alarm Event and Operator Message Annotations
Alarm Event and Operator Message annotations are not shown on the chart unless enabled – see 'Chart Annotation' on page 11 and see Section 4.5.2, page 63.
If Alarm event annotation is enabled and an alarm becomes active, a red alarm event icon surrounded by a channel colored box is displayed at the point at which the alarm occurred, together with the alarm time and tag, e.g.
11:58:00 1.1A High Level
If more than one alarm occurs in the same sample period:
– and the second alarm on a channel becomes active, its icon is added behind the first.
– and more than one operator message is active (max. six), a second icon is added behind the first.
– the new alarm event icons appear to the left of earlier icons.
– the time and tag of the oldest alarm (right-most icon) only is displayed.
4. Screen Capture
If 'Screen capture' is set to 'Enabled' in Common Configuration (see Section 4.4.1, page 52) and an external archive media card is inserted in the instrument, an image of any Chart, Bargraph, Digital Indicator, Process, Instrument Status, Audit Log, Alarm Log or Totalizer Log view can be saved to the external media by pressing the key whenever the Operator Menu is not open.
8 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
Select the Configuration Level – see Section 4, page 45.
Select the Setup Level – see Section 3, page 29.
Note.
Available only if 'Security system' is set to 'Advanced' – see Section 4.4.4, page 56.
Sign the chart electronically – see Section 2.3.1, page 14.
Select Historical Review to view previously recorded data stored in the instrument's onboard memory.
Note. Use the and keys to move backwards and forwards through the recorded data.
Historical Review Active Invalid data  e.g. recording disabled
Digital
Indicators
Values displayed indicate the process status at the cursor position
Cursor
IM/SM1000 Issue 12 9
Videographic Recorder
SM1000 2Operation
Note.
While in Historical Review mode:
Recording of new data continues unless stopped from the Setup Menu – see
Section 3.4, page 32.
Invalid historical data (e.g. when recording has stopped) is denoted by '– – –' in the
digital indicator.
Where the trace at the cursor position represents more than one sample, the
indicators flash between the maximum and minimum values of those samples.
Menu options remain active – allowing the screen interval to be changed, different
scales and channels to be selected, etc.
Operator messages generated are added to the alarm event log at the present
time, not the time indicated by the cursor.
All data stored in the instrument's internal memory can be viewed.
The display can be scrolled back to the start of the oldest data.
Archiving to removable media does not occur but all data recorded in the internal
memory buffer during this time is archived on exiting Historical Review mode.
Select 'Goto' to move to data stored in the instrument's onboard memory that was recorded at a specific date and time.
Oldest data in
internal memory
'Goto' target date
'Goto' target time
Selected 'goto' target
date/time is within
daylight saving period
Moves cursor between
'goto' target date and
time boxes
Returns to Menu
Newest data in internal memory
Goes to the chart view at the selected 'goto' target date and time
Note.
If daylight saving is enabled (see Section 4.4.3, page 54) and the selected 'Goto'
target date/time is within the daylight saving period, 'Daylight Saving' is displayed on the dialog box.
Once internal memory becomes full, oldest data is overwritten by newest data. If
historical review has been selected for some time, the oldest data present may no longer be available.
The instrument exits Historical Review mode automatically after 15 minutes if no
key is pressed.
Select Exit to return to the real-time recording display.
Note. Pressing the key also exits Historical Review mode and displays the next enabled view.
10 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
Add one of 24 predefined Operator Messages (see Section 4.4.7, page 60) or one User­Defined Message to the alarm event log.
If '< user defined >' is selected, a data entry keyboard appears to enable the message to be entered (see Fig. 4.6, page 50).
The selected or user-defined message is displayed briefly on screen. If Operator Message annotation is selected (see 'Chart Annotation' below) the message is also added to the chart.
Note. When the instrument is in Historical Review mode, Operator Messages generated are added at the current time, not the time indicated by the cursor.
If an alarm or operator message is obscuring part of a chart trace, use the Chart Annotation option to hide or display alarms and messages on the screen. Select the annotation required. indicates the annotations selected.
The possible combinations are:
No annotation (Alarms and Operator Messages both disabled)
Alarm annotation only
Alarms and Operator Message annotation
Operator Message annotation cannot be enabled unless Alarm annotation is also enabled.
Note.
If more than 15 icons are present on the screen, chart annotation is disabled
automatically.
When chart annotation is disabled, new operator messages and alarms are still
added to the Alarm Event log – see Section 2.7, page 23.
IM/SM1000 Issue 12 11
Videographic Recorder
SM1000 2Operation
Use the 'Screen Interval' to change the amount of data displayed on the screen. A longer screen interval displays more data, a shorter screen interval displays data over a shorter time period, but in more detail. In both cases, the full trace is preserved by plotting the maximum and minimum samples for each display point.
Longer interval =
better overview
Shorter interval
= more detail
The screen intervals available in the menu are determined by the faster of the primary and secondary sample rates set for the process group during configuration – see Section 4.5.1, page 61. Unavailable screen intervals are greyed-out.
Table 2.1 shows the relationship between sample rate and screen interval for the Vertical and Horizontal Chart views.
Note.
A 'Please Wait' message appears in the status bar while the instrument retrieves
data from storage.
Selecting a different screen interval does not affect the rate at which data is
sampled.
When in historical review mode, changing the screen interval causes the time at
the cursor position to change.
Maximum Screen Interval
Sample Rate Setting Vertical Chart View horizontal Chart View
<1 second Up to 9 minutes/screen Up to 13 minutes/screen
Between 1 and 10 seconds Up to 1 hour/screen Up to 1.5 hours/screen
More than 10 seconds, less than 20 seconds Up to 12 hours/screen Up to 18 hours/screen
Between 20 and 40 seconds Up to 1 day/screen Up to 1.5 days/screen
Between 40 and 60 seconds Up to 2 days/screen Up to 3 days/screen
More than 60 seconds, less than 140 seconds Up to 3 days/screen Up to 4.5 days/screen
More than 140 seconds Up to 7 days/screen Up to 10 days/screen
Table 2.1 Sample Rates and Screen Intervals
12 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
Select a channel scale to be displayed in the scale bar at the top of the chart window. For digital channels, the On and Off tags are displayed at the corresponding position on the scale bar.
Select 'Auto Scroll' to display the scale for each enabled channel in turn for 36 seconds.
Hide individual channel traces to improve chart clarity.
Traces are identified by the Channel Number (e.g. Ch1.1) and its short tag.
Note. The recording of a channel's data is not affected by this operation and the instantaneous channel values are still shown on the indicators at the top of the screen.
To acknowledge a particular alarm, use the and keys to highlight it in the menu and press the key.
Note. Active unacknowledged alarms in the current process group are identified by a flashing red Alarm Event icon in the bottom left hand corner of the associated channel indicator. Active acknowledged alarms are identified by a continuous red Alarm Event icon in the bottom left hand corner of the indicator – see Fig. 2.3, page 7.
To acknowledge all active alarms in the current process group simultaneously, select 'All' and press the key.
Note. If an alarm in continues to be displayed in the status bar. If an active alarm in
the other process group is active, the Global Alarm status icon ( )
the other process group
is unacknowledged, the icon is surrounded by a red flashing border ( ).
If neither alarm on a particular channel is configured, the short channel tag is shown greyed-out in the menu.
Select 'AutoView Scroll' to display the chart for each configured process group in turn for 40 seconds. The AutoView Scroll status icon ( ) is displayed in the status bar when AutoView Scroll is active. Press any key to cancel AutoView Scroll.
IM/SM1000 Issue 12 13
Videographic Recorder
SM1000 2Operation
2.3.1 Electronic Signatures – Fig. 2.4
Entering an electronic signature is the equivalent to signing the chart of a conventional paper recorder. Local procedures may require the approval of a record by an authorized signatory; for this reason, an electronic signature is password protected.
Invalid Password
(see Note in Fig. 3.2, page 30)
Entered
Create a new signature
or
Select one of 7
previously
defined signatures
Note below)
(see
If more than one process group is
enabled (see Section 4.4.1, page 52),
assign the signature to one or more
groups
Enter Password
(see Section 3.3, page 31)
Create new signature or
edit an existing one.
Electronic signature is entered into alarm log.
If Operator Message annotation is enabled (see Page 11), the electronic signature is also displayed on the chart.
Valid Password
Entered
Fig. 2.4 Entering an Electronic Signature
Note. Up to 7 electronic signatures can be stored in the instrument’s memory. If 7 signatures exist and a new one is created, the oldest is overwritten.
14 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
2.4 Bargraph Views – Fig. 2.5
Short Channel Tag
Current Value
Alarm Event Icon 
see Rear fold-out
Engineering Units
Alarm Trip Level
(not shown for slow and
fast rate alarms)
Maximum Value
(solid color)
Instantaneous Value
Minimum Value
(white fill)
Selects the next Process Group
(if enabled) or Status View
Selects the next enabled view in the group
(Digital Indicator, Process or Chart)
Status Icons
Digital On or Off tag, according to input status
Engineering Range High Value
Digital Channel
Engineering Range Middle Value
Engineering Range Low Value
Minimum below engineering limit, maximum above engineering limit
Identical Maximum and Minimum Values
Opens the Operator Menu for
the current view
A  Vertical Bargraph View
Alarm Trip Level
(not shown for slow and
fast rate alarms)
Maximum Value
(solid color)
Minimum Value
(white fill)
Identical Maximum
and Minimum Values
Minimum below engineering limit, Maximum above
engineering limit
Engineering Range
(displayed in the Bargraph, Digital Indicators, Process and Chart Views), see Appendix G, page 127.
Global Alarm Status Icon
Low Value
Engineering Range
Middle Value
Instantaneous Value
Engineering Range High Value
B  Horizontal Bargraph View
For a description of the Status Icons (displayed in the Status Bar) and Alarm Event Icons
Short Channel Tag
Current Value
Alarm Event Icon
Fig. 2.5 Bargraph Views
IM/SM1000 Issue 12 15
Videographic Recorder
SM1000 2Operation
Select the Configuration Level – see Section 4, page 45.
Select the Setup Level – see Section 3, page 29.
Add one of 24 predefined Operator Messages (see Section 4.4.7, page 60) or one User­Defined Message to the alarm event log.
If '< user defined >' is selected, a data entry keyboard appears to enable the message to be entered (see Fig. 4.6, page 50). The selected or user-defined message is displayed briefly on-screen.
Reset the Maximum and Minimum value markers on one or all channels to the current value.
Note.
These Maximum and Minimum values are for display purposes only. They are not
saved or archived and are not connected to the Totalizer Maximum and Minimum Values displayed in the Process View.
These Maximum and Minimum values are reset whenever the current configuration
has been changed or is re-saved.
To acknowledge a particular alarm, use the and keys to highlight it in the menu and press the key.
Note. Active unacknowledged alarms in the current process group are identified by a flashing red Alarm Event icon in the bottom left hand corner of the associated indicator. Active acknowledged alarms are identified by a continuous red Alarm Event icon in the bottom left hand corner of the indicator – see Fig. 2.5, page 15.
To acknowledge all active alarms in the current process group simultaneously, select 'All' and press the key.
Note. If an alarm in continues to be displayed in the status bar. If an active alarm in
the other process group is active, the Global Alarm status icon ( )
the other process group
is unacknowledged, the icon is surrounded by a red flashing border ( ).
If neither alarm on a particular channel is configured, the short channel tag is shown greyed-out in the menu.
Select 'AutoView Scroll' to display the bargraph for each configured process group in turn for 40 seconds. The AutoView Scroll status icon ( ) is displayed in the status bar when AutoView Scroll is active. Press any key to cancel AutoView Scroll.
16 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
2.5 Digital Indicator View – Fig. 2.6
Short Channel Tag
Current Value
Alarm Event Icon
Engineering Units
Selects the next Process Group
(if enabled) or the Status View
Global Alarm Status Icon
Opens the Operator Menu for the current view
Selects the next enabled
view in the group
(Process, Chart or Bargraph)
Status Icons
Channel Units
Totalizer Units
Totalizer Value
For a description of the Status Icons (displayed in the Status Bar) and Alarm Event Icons
(displayed in the Bargraph, Digital Indicators, Process and Chart Views), see Appendix G, page 127.
Fig. 2.6 Digital Indicator View
Note.
Digital indicators resize automatically according to the number of channels displayed.
The totalizer option must be enabled and totalizers configured and enabled in the Configuration level before they are
displayed – see Section 4.6.5, page 84.
When a totalizer is not running (i.e. 'Stop' is selected), the corresponding totalizer value is shown in red.
IM/SM1000 Issue 12 17
Videographic Recorder
SM1000 2Operation
Select the Configuration Level – see Section 4, page 45.
Select the Setup Level – see Section 3, page 29.
Add one of 24 predefined Operator Messages (see Section 4.4.7, page 60) or one User­Defined Message to the alarm event log.
If '< user defined >' is selected, a data entry keyboard appears to enable the message to be entered (see Fig. 4.6, page 50). The selected or user-defined message is displayed briefly on-screen.
Hide or display individual channel indicators for the group currently displayed.
Note.
The digital indicators displayed are sized equally depending on the number of
channels selected. This has no effect on the recording of the channel.
Channels that have not been configured are greyed-out in the menu.
Three or Four
Channels Selected
Two Channels Selected One Channel Selected
18 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
Stop and start individual totalizers.
Channel totalizers that have not been enabled in the Configuration level are greyed-out.
Note.
Displayed only if the Totalizer option is enabled.
When a totalizer is not running (i.e. 'Stop' is selected), the corresponding totalizer
value is shown in red.
Reset the totalizer value to the totalizer preset value.
Channel totalizers that have not been enabled in the Configuration level are greyed-out.
Note. Displayed only if the Totalizer option is enabled.
To acknowledge a particular alarm, use the and keys to highlight it in the menu and press the key.
Note. Active unacknowledged alarms in the current process group are identified by a flashing red Alarm Event icon in the bottom left hand corner of the associated indicator. Active acknowledged alarms are identified by a continuous red Alarm Event icon in the bottom left hand corner of the indicator – see Fig. 2.6, page 17.
To acknowledge all active alarms in the current process group simultaneously, select 'All' and press the key.
Note. If an alarm in continues to be displayed in the status bar. If an active alarm in
the other process group is active, the Global Alarm status icon ( )
the other process group
is unacknowledged, the icon is surrounded by a red flashing border ( ).
If neither alarm on a particular channel is configured, the short channel tag is shown greyed-out in the menu.
Select 'AutoView Scroll' to display the digital indicator for each configured process group in turn for 40 seconds. The AutoView Scroll status icon ( ) is displayed in the status bar when AutoView Scroll is active. Press any key to cancel AutoView Scroll.
IM/SM1000 Issue 12 19
Videographic Recorder
SM1000 2Operation
2.6 Process View – Fig. 2.7
Note. The Process view is available only if the Totalizer option is enabled.
Totalizer not enabled
in configuration level
Totalizer
Tag Name
Totalizer
Value
Selects the next Process Group
(if enabled) or the Status View
(Bargraph, Digital Indicator or Chart)
Global Alarm Status Icon
Alarm Tag
Opens the Operator Menu for the current view
Selects the next enabled view
A  Process View (Alarms)
Alarm Trip Level
Alarm Acknowledged
Short Channel Tag
Instantaneous Value
Alarm Status
Channel Source Long Tag
Max
Values since last totalizer reset or totalizer wrap. Calculated only while
the totalizer is enabled. Updated only while the totalizer is running.
Min
Average
B  Process View (Statistics)
For a description of the Status Icons (displayed in the Status Bar) and Alarm Event Icons
(displayed in the Bargraph, Digital Indicators, Process and Chart Views), see Appendix G, page 127.
Fig. 2.7 Process View
Note.
Only totalizers that have been configured and enabled in the Configuration level are displayed.
When a totalizer is not running (i.e. 'Stop' is selected), the corresponding totalizer value is shown in red.
20 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
Select the Configuration Level – see Section 4, page 45.
Select the Setup Level – see Section 3, page 29.
Switch between the Alarm View and the Statistics View.
Note. If a channel's totalizer is not enabled in the Configuration level, Alarm trip levels are shown in place of the channel statistics. If no alarms are enabled, the long tag for that channel is displayed.
Add one of 24 predefined Operator Messages (see Section 4.4.7, page 60) or one User­Defined Message to the alarm event log.
If '< user defined >' is selected, a data entry keyboard appears to enable the message to be entered (see Fig. 4.6, page 50). The selected or user-defined message is displayed briefly on-screen.
Stop and start individual totalizers.
Channel totalizers that have not been enabled in the Configuration level are greyed-out.
Note. When a totalizer is not running (i.e. 'Stop' is selected), the corresponding totalizer value is shown in red.
IM/SM1000 Issue 12 21
Videographic Recorder
SM1000 2Operation
Reset the totalizer value to the totalizer preset value.
Channel totalizers that have not been enabled in the Configuration level are greyed-out.
To acknowledge a particular alarm, use the and keys to highlight it in the menu and press the key.
Note. Active unacknowledged alarms in the current process group are identified by a flashing red Alarm Event icon in the bottom left hand corner of the associated indicator. Active acknowledged alarms are identified by a continuous red Alarm Event icon in the bottom left hand corner of the indicator – see Fig. 2.6, page 17.
To acknowledge all active alarms in the current process group simultaneously, select 'All' and press the key.
Note. If an alarm in continues to be displayed in the status bar. If an active alarm in
the other process group is active, the Global Alarm status icon ( )
the other process group
is unacknowledged, the icon is surrounded by a red flashing border ( ).
If neither alarm on a particular channel is configured, the short channel tag is shown greyed-out in the menu.
Select 'AutoView Scroll' to display the process view for each configured process group in turn for 40 seconds. The AutoView Scroll status icon ( ) is displayed in the status bar when AutoView Scroll is active. Press any key to cancel AutoView Scroll.
22 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
2.7 Alarm Event Log – Fig. 2.8
Note. The Alarm Event log view provides an historical log of all alarm events in the sequence in which they occurred. To view the
current status of all alarms, choose the Process View – see Section 2.6, page 20.
Alarm becomes active
(Active transition)
Alarm becomes inactive
(Inactive transition)
Operator Message
and the Bargraph, Digital Indicators, Process and Chart Views), see Appendix G, page 127.
Short tag of the alarm's source
Selects Process Group 1
Selects the
Totalizer Log
For a description of the Alarm Event Icons (displayed in the Alarm Event Log
Opens the Operator Menu for the current view
Global Alarm Status Icon
View previous page of data
Oldest Data
Alarm Acknowledged
Newest Data
View next page of data
Fig. 2.8 Alarm Event Log
Note. When the number of entries in the Alarm Event log has reached that defined in 'Alarm log size' (see Section 4.4.6, page 60), the oldest data is overwritten by the newest. Entries are renumbered so that the number of the oldest entry is always 00.
IM/SM1000 Issue 12 23
Videographic Recorder
SM1000 2Operation
Select the Configuration Level – see Section 4, page 45.
Select the Setup Level – see Section 3, page 29.
Select the entries to be displayed in the log. Indicates entries currently displayed.
Note.
Hiding and displaying log entries does not affect the recording of events in the log.
All selected alarm event transitions (from inactive to active, from active to
acknowledged, from acknowledged to inactive, from active to inactive) appear in the sequence in which they occurred.
Selecting 'Active Transitions Only' displays entries for alarms when made active
and hides all acknowledged & inactive transitions.
The Alarm Event log is not group specific. To acknowledge a particular alarm, use the
, and keys to first select the relevant process group, followed by the alarm
to be acknowledged.
To acknowledge all active alarms in the selected process group simultaneously, select 'All' and press the key.
Note. If an alarm in continues to be displayed in the status bar. If an active alarm in
the other process group is active, the Global Alarm status icon ( )
the other process group
is unacknowledged, the icon is surrounded by a red flashing border ( ).
If neither alarm on a particular channel is configured, the short channel tag is shown greyed-out in the menu.
24 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
2.8 Totalizer Log – Fig. 2.9
Note.
The Totalizer log view is displayed only if the Totalizer option is enabled.
The Totalizer log view provides an historical log of totalizer activity. To view the current totalizer status, choose the Process
or Digital View.
When the number of entries in the Totalizer log has reached that defined in 'Totalizer log size' (see Section 4.4.6, page 60),
the oldest data is overwritten by the newest. Entries are renumbered so that the number of the oldest entry is always 00.
The logging of totalizer values can be triggered at pre-determined intervals and/or by digital signal – see 'Log update' on
page 85.
View previous
page of data
Totalizer Icon
Oldest
Log Entry Number
Data
Batch total at the
time of the event
Date/time at which
Max., min. and
average of the value
being totalized at the
time of the event
Selects Process Group 1
Selects the
Audit Log
For a description of the Totalizer Icons (displayed in the Totalizer Log
and the Process View), see Appendix G, page 127.
Opens the Operator Menu for the current view
View next page
max. and min. flowrates occured
Newest Data
of data
Fig. 2.9 Totalizer Log
Note. Maximum, Minimum and Average statistics are not shown unless enabled in the 'Filter' menu – see overleaf.
IM/SM1000 Issue 12 25
Videographic Recorder
SM1000 2Operation
Select the Configuration Level – see Section 4, page 45.
Select the Setup Level – see Section 3, page 29.
Select the entries to be displayed in the log. Indicates entries currently displayed. This does not affect which events are recorded in the log.
'Statistics' displays the maximum, minimum and average values of the analog value being totalized, together with the dates and times at which all maximum and minimum values occurred.
These values are reset when the totalizer is reset and are updated only when the totalizer is running.
26 IM/SM1000 Issue 12
Videographic Recorder
SM1000 2Operation
2.9 Audit Log – Fig. 2.10
Note.
The Audit log view provides an historical log of system activity.
When the number of entries in the Audit log has reached that defined in 'Audit log size' (see Section 4.4.6, page 60), the
oldest data is overwritten by the newest. Entries are renumbered so that the number of the oldest entry is always 00.
View previous
page of data
Oldest Data
Audit Log Icon
Fig. 2.10 Audit Log
Selects Process Group 1
Selects the Status View
Opens the Operator Menu for
the current log
For a description of the Audit Log Icons (displayed in the Audit Log) see Appendix G, page 127.
page of data
Select the Configuration Level – see Section 4, page 45.
Select the Setup Level – see Section 3, page 29.
Newest Data
View next
IM/SM1000 Issue 12 27
Videographic Recorder
SM1000 2Operation
2.10 Status View – Fig. 2.11
Note. The Status view provides an overview of the instrument's status.
Instrument Software Version
Operating System Version
Amount of memory used
on the archive media
currently installed
Approximate time left
before the current archive
media is full, assuming the
amount of data recorded
remains the same
Selects Process Group 1
Fig. 2.11 Status View
Selects the Alarm Event Log
Opens the Operator Menu for
the current view
Cold junction temperatures for Modules A & B; measured by the built-in cold junction circuitry.
'Not Used' is displayed if none of the inputs on the module are thermocouples.
Time Left Display Format
>=1 Day: Days, e.g. '5 Days' >=1 Hour, <1 Day: Hours, e.g. '10 hours' <1 Hour: Minutes, e.g. '25 minutes'
Select the Configuration Level – see Section 4, page 45.
Select the Setup Level – see Section 3, page 29.
28 IM/SM1000 Issue 12
Loading...
+ 104 hidden pages