3M VH2O2 BI Reference Guide

3M™ Attest™ Mini Auto-reader 490M System* Quick Reference Guide
VH2O2 BI Preparation and Use
2
Inspect
1
Remove BI from pouch and reseal. Verify media ampoule is intact and indicator stripes on cap are blue.
Pouch
Place BI and chemical indicator (CI) in peel pouch indicated for use in vaporized hydrogen peroxide (VH2O2) sterilization processes.
STEAM and VH2O2 BI Activation and Incubation
Crush
1
Squeeze BI in activator to close cap and crush media ampoule.
Flick
2
Remove BI and flick wrist to ensure media flows to growth chamber.
3M™ Attest™ Mini Auto-reader 490M
Positive Control BI
For both Steam and VH2O2, each day a processed BI is incubated, activate and incubate an unprocessed BI as a positive control. The control BI should be from the same lot code as the processed BI.
Results After Incubation
Control BI: Positive (+) Acceptable Processed BI: Negative (-) Acceptable Sterilization Processed BI: Positive (+) Failed Sterilization
3
Process
Seal pouch and place in sterilizer with white side facing up. Process load as recommended by sterilizer manufacturer.
Inspect
3
Verify media has transferred to BI growth chamber at the vial bottom.
4
Identify
After completion of cycle, inspect and verify media ampoule is intact and indicator stripes on cap have turned towards pink. Identify the BI by writing on the label.
Incubate
4
Place activated BI in any well. Display shows remaining minutes of incubation or test results.
DO NOT REMOVE BI until the (+) or (-) symbol indicates the test is complete.
NOTES:
• Activate and incubate 1295 BI within 1 hour of completion of the VH2O2 sterilization cycle.
• Gloves are recommended when activating 1295 BI.
• Allow 1491 and 1492V BIs to cool 10 minutes prior to activation.
Remaining Incubation Minutes
Touch Button (alarm off)
Interpretation of 3M™ Sterilization Assurance Core Products
STEAM Indicator Unexposed Acceptable Exposed
3M
Attest™ BI
1491 an d 1492V
1491 1492V
3M™ Comply™ Indicator Tap e 1 322
3M™ Comply™ Indicator Tap e 1 355
STEAM Indicator Unprocessed Acceptable Processed PASS Unacceptable Processed FAIL
3M™ Attest™ Steam Chemical Integrator 1243
3M™ Comply™ Bowie-Dick Plus Test Pac k 00135LF
3M™ Comply™ Early Warning Test Sheet (above)
Comply™ Bowie-Dick
3M Test Sheet (below)
After processing, indicator on cap turns light brown or darker.
After processing, indicator stripes turn dark-brown/black.
After processing, indicator stripes turn dark-brown/black.
1491
After processing, indicator bar crosses into the ACCEPT region. If the color bar is in the REJECT region or on the line, it is considered a fail.
1492V
PASS
When indicator ink circle on Early Warning Test sheet is lighter than the color standard, this may indicate that sterilizer maintenance should be scheduled.
VH2O2 Indicator Unexposed Acceptable Exposed Color Range
3M
Attest™ BI 129 5
After processing, indicator stripes turn toward pink.
3M™ Comply™ Indicator Tap e 1 228
After processing, indicator stripes turn toward pink.
3M™ Attest™ Tri-Metric Chemical Indicator 1348/1348E
After processing, a color change towards pink should have entered the ACCEPT window.
FAIL
FAIL
FAIL
PASS
FAIL
with a lighter colored area, usually near the center.
FAIL
3M Company
3M Center, Building 275-4E-01 St. Paul, MN 55144 USA
© 2020 3M. All rights reserved. 3M, Attest and Comply are marks and/or registered marks of 3M. Unauthorized use prohibited. 70-2011-7924-2 Rev 9/20
PASS
After processing, a uniform color change to dark brown/black is a pass. Fail result is indicated by a non-uniform color development,
* Refer to 3M™ Attest™ Mini Auto-reader 490M Operator Manual and 3M™ Attest™ Biological Indicator 1295, 1491 and 1492V Instructions for Use for complete
instructions and other important information. Refer to the indications for use provided in the biological indicator Instructions for Use. The indications for use should align with the sterilization cycle to be monitored. Refer to the Instructions for Use for complete instructions and other important information related to the use of all other 3M products included in this Guide.
Health Care Help Line 1-800-228-3957 Learn more about 3M™ Sterilization Assurance products at go.3M.com/sterilization
Loading...