3M 290 User Manual

3M™ Attest
290 Auto-reader
Operator’s Manual
Read and understand all safety information before using this product.
INTENDED USE
3
2
1
9
4
5
6
1. Cover
2. Incubation/Reader Wells
3. Crusher Well
4. Red Light Positive (+) Indicator
5. Yellow Light (•) Test In Progress Indicator
Intended Use
The 3M incubate and automatically read the 3M 1291 Rapid Readout Biological Indicators (RRBI) and/or 3M Biological Indicators at 60 fluorescence reading at 1 hour for 1291 or 3 hours for 1292.
The 3M designed to allow for further incubation of the 3M 1292 RRBI for a final negative, visual pH color change of the growth media at 24 hours for 1291 and 48 hours for 1292.
TM
AttestTM 290 Auto-reader is designed to
TM
AttestTM 1292 Rapid Readout
TM
AttestTM 290 Auto-reader is also
TM
AttestTM 1291 RRBI and/or 3MTM AttestTM
o
C for a final negative
TM
AttestTM
8
7
6. Green Light Negative (-) Indicator
7. Buzzer Alarm OFF Button
8. Digital Display Panel
9. Time Remaining Button
A positive fluorescence reading or a visual color change indicates a steam sterilization process failure.
Other biological indicators (BIs), are not compatible with this device and cannot be used.
The unit is designed to be used only with the Power Supply module supplied by 3M.
3M © 2003 78-8078-8977-5 Page 1 of 9 Attest 290 Auto-reader
p
SAFETY INFORMATION
WARNING
To avoid the risk of electric shock, which if not avoided could result in serious injury or death:
- Do not immerse the Attest 290 Auto-
reader in any liquid.
- Use indoors only.
- Disconnect power cord before
cleaning.
- To clean, use only a mild detergent
and wipe with a damp cloth.
- There are no user serviceable parts.
- The Attest 290 Auto-reader must be
returned to the manufacturer for repair.
- Make sure a properly rated power
cord is installed on the Attest 290 Auto-reader power supply.
To reduce the risk of using incompletely sterilized loads:
- Match the color on the exterior label
of the incubator with the same color cap on the RRBI.
- Do not remove RRBI until incubation
cycle is complete.
- Do not use on a vibrating surface.
There is a glass ampule inside the plastic vial of the RRBI. To avoid the risk of injury from flying debris due to a ruptured RRBI, which if not avoided could result in serious injury or death:
- Allow the RRBI to cool for the
recommended time period before crushing.
- Crushing or excessive handling of
the biological indicator before cooling may cause the glass ampule to burst.
- Wear safety glasses and gloves
when removing the RRBI from the sterilizer.
- Wear safety glasses when crushing
the RRBI.
- Handle the RRBI
by the cap when crushing and tapping.
- Do not use your fingers to crush the
glass ampule.
- Do not roll the RRBI between your
WARNING
WARNINGWARNING
fingers to wet the spore strip.
CAUTION
To avoid a potentially hazardous situation, which if not avoided may result in minor or moderate injury:
- Avoid contact with the hot surface of the metal incubator block.
To avoid possible environmental contamination, dispose of the Attest 290 Auto-reader according to applicable governmental regu l ation s.
To avoid RRBI vial from absorbing fluorescent residue from a chemical indicator or tape, place the RRBI vial so it does not come in direct contact with chemical indicators or ta
CAUTION
CAUTIONCAUTION
e.
Explanation of symbols:
Caution – Risk of electric shock
Attention – Refer to accompanying documents
3M © 2003 78-8078-8977-5 Page 2 of 9 Attest 290 Auto-reader