Page 16 of 28 802006UK BCS Residential User Guide - Aug 2015- V2.01
After replacing the filter, the counters should be reset.
The Zip HydroTap
®
G4 is equipped with a self-calibrating program. On start
up, the controls take the system through a calibration process. Once this
mode is completed the system reverts back to normal operation. If you would
like to calibrate the system anytime after installed, please follow the steps
below.
• Press the [MENU] button for main
menu.
• Press the [Install] button.
• Press the [Boiling Calibration]
button.
• In the next screen, press [Calibrate]
if you want to start calibration.
Boiling calibration screen
3. Boiling calibration
Caution
In general, boiling water units may vent steam from the tap spout.
Take care to avoid personal injury whenever this occurs.
• Press the [MENU] button for main
menu.
• Press the [Install] button.
• Press the [Booster] button.
• In the next screen, press button
A
to enable the booster.
• Dispense boiling water for 30 secs
and check the outlet hose from the
booster is warm.
This function should only be used when a booster unit is installed.
Booster screen
4. Booster
Note
Failing to make the correct selection for [Booster] will affect product
performance.
2. Filter reset
LCD screen operation - section C
• Press the [MENU] button for main
menu.
• Press the [Install] button.
• Press the [Filter Reset] button.
• In the next screen, press [YES]
button if you want to flush the filter.
Follow the filter flush steps to start
filter flush or press [NO] to cancel the
filter reset program. To set the filter
life or edit filter usage see page 21.