Zimmer NexGen Surgical Techniques

Page 1
Zimmer®
Patient Specific
Instruments (CT)
Surgical Techniques
®
for NexGen
Knee Solution
Page 2
Page 3
Zimmer® Patient Specific Instruments (CT) – Surgical Techniques
Introduction
General Considerations
Zimmer® PSI is not designed for use with the Zimmer
MIS Quad-Sparing
Zimmer PSI should be used in conjunction with a femur first technique.
Verify stability and fixation of all pins. When securing the conventional tibia cutting guides, avoid the use of spring pins as these may cause stress / unwanted shift on the tibia cutting guide.
Total Knee Procedure.
for NexGen® Complete Knee Solution
INTRO
INTRO.1
Page 4
TOC
Zimmer
®
Patient Specific Instruments (CT) – Surgical Techniques
for NexGen® Complete Knee Solution
Table of Contents
SECTION PAGE
1A
NexGen® Conventional Instruments -
Femoral Technique
1
NexGen Conventional Instruments ­Femoral Technique
SECTION PAGE
1B
SECTION PAGE
2A
NexGen Conventional Instruments ­Tibial Technique
NexGen Conventional Instruments ­Tibial Technique
NexGen Posterior Referencing Instruments - Femoral Technique
NexGen Posterior Referencing Instruments - Femoral Technique
1
4
4
8
8
TOC.1
SECTION PAGE
2B
NexGen Posterior Referencing Instruments - Tibial Technique
NexGen Posterior Referencing Instruments - Tibial Technique
10
10
Page 5
NexGen® Conventional Instruments – Femoral Technique
NexGen Conventional Instruments – Femoral Technique
Position Femoral Pin Guide
Do not remove osteophytes from the femur.
Position the Pin Guide on the distal femur (Fig. 1).
Pin Guide will be captured by the anterior and proximal ridge of the femur. The guide fits over the edge on the proximal bone to increase locking.
The A/P axis reference line on the Pin Guide can be used to assess alignment. The most distal end of the guide is parallel to the epicondylar axis and can also be used as reference.
Apply anterior proximal pressure to the Pin Guide and verify a secure fit. Applying posterior pressure may destabilize the Pin Guide.
SECTION
1A
Fig. 1
Drill and Pin the Anterior Pin Holes
Drill and pin the medial and lateral anterior pin holes of the Pin Guide using the Universal Disposable Drill (2001-00-000) (Fig. 2).
Drill Distal Pin Holes
Drill the distal pin holes of the Pin Guide (do not place pins)
(Fig. 3).
Instruments
Fig. 2
Fig. 3
PSI Femoral
Pin Guide
00-5970-015/
018-00
Universal Pin
and Drill Kit
2001-00-000
1
Page 6
SECTION
NexGen® Conventional Instruments – Femoral Technique
1A
Place Distal Cut Guide and Cut Femur
Remove the Pin Guide by sliding it off the pins, leaving the anterior pins in the bone.
Secure the MIS Distal Cut Guide (00-5967-036-00).
Verify the distal resection on the medial and lateral side with a resection guide (00-5977-084-00). Check alignment if desired, and make the cut (Fig. 4).
TECHNIQUE TIP 1A.A
When performing the cuts, excessive force (considering the pressure being applied to the guide) on the medial or lateral side of the cut guide could influence the amount of varus/valgus in the cut.
Fig. 4
Remove Distal Cut Guide and Place Distal Pins
Locate the drilled pin holes.
Remove the anterior pins
Place the distal pins in drilled holes of the distal femur
(Fig. 5).
TECHNIQUE TIP 1A.B
In the event the drilled distal holes are covered up following the distal cut, clean the area. Alternatively, prior to removing the anterior pins, place the Pin Guide back over the anterior pins and locate the distal pins through the distal Pin Guide holes (Fig. 6).
Fig. 5
Fig. 6
Instruments
Instruments
MIS Distal Cut Guide
00-5967-036-00
Resection Guide
00-5977-084-00
PSI Femoral
Pin Guide
00-5970-015/
018-00
2
Page 7
NexGen® Conventional Instruments – Femoral Technique
Place Femoral Finishing Guide
Place the NexGen 4-in-1 Femoral Finishing Guide (silver) or the Flex Femoral Finishing Guide (gold) over the distal pins.
Adjust the M/L positioning for appropriate placement.
Secure the Finishing Guide and remove the pins. Verify resections using a resection guide and make the cuts (Fig. 7).
SECTION
1A
Fig. 7
Follow 4-in-1 Femoral Surgical Technique
Following the 4-in-1 cuts, proceed with the surgery as suggested in the Zimmer MIS Multi-Reference Femoral Instrumentation Surgical Technique.
®
4-in-1
Instruments
NexGen MIS
4-in-1 Femoral
Finishing Guide
00-5983-031/38-00
-OR-
NexGen MIS
4-in-1 Flex Femoral
Finishing Guide
00-5983-041/48-00
Resection Guide
00-5977-084-00
3
Page 8
SECTION
1B
NexGen® Conventional Instruments – Tibial Technique
NexGen Conventional Instruments – Tibial Technique
Position Tibial Pin Guide
Remove the meniscus.
Do not remove osteophytes from the tibia. Clean the bone around the ACL and anterior tibia where the guide fits. Remove any soft tissues that may prevent proper placement of the Pin Guide.
Position the Pin Guide on the tibia (Fig. 8).
The rotation cylinders do not make contact with bone surface Both cylinders will be above the tibia surface. All other contact with the Pin Guide should ensure a proper fit.
The mechanical axis and proximal resection lines may be used to asses the alignment. Additionally, the alignment arch (00-5977-024-00) and rod (00-5785-080-00) may be used to verify alignment. Additionally, the alignment adapter (00-5970-024-00) and rod (00-5785-080-00) may be used to verify alignment.
Fig. 8
.
Drill and Pin Anterior Pin Holes
Drill and pin the medial and lateral anterior holes of the Pin Guide (Fig. 9).
Instruments
Fig. 9
PSI Tibia
Pin Guide
00-5970-015/
018-00
Universal Pin
and Drill Kit
2001-00-000
Alignment Adapter
00-5970-024-00
Alignment Rod
00-5785-080-00
4
Page 9
NexGen® Conventional Instruments – Tibial Technique
Establish Tibia Rotation (Drill Proximal Holes)
Tibia rotation may be established through the Pin Guide per surgeon discretion.
Drill the proximal medial and lateral holes of the Pin Guide (do not place pins) (Fig. 11).
TECHNIQUE TIP 1B.A
When drilling the proximal holes, excessive side loads through the proximal guide cylinders could alter the planned hole position.
SECTION
1B
Fig. 11
Remove Tibial Pin Guide
Remove the Pin Guide by sliding anteriorly off the pins
(Fig. 12). Do not pry the Pin Guide from the tibia as this
could shift the pins.
Fig. 12
Instruments
PSI Tibia
Pin Guide
00-5970-015/
018-00
Universal Pin
and Drill Kit
2001-00-000
5
Page 10
SECTION
1B
NexGen® Conventional Instruments – Tibial Technique
Secure Cut Guide and Cut Tibia
Secure the NexGen Tibial Cut Guide. Verify the proximal resection on the medial and lateral side with a resection guide.
Make tibial cut (Fig. 14).
Use the appropriate right or left tibia cut guide.
0° Left Cut Guide 00-5997-075-00
0° Right Cut Guide 00-5997-076-00
Insert the Alignment Arch into the Tibial Cut Guide, insert the Alignment Rod and check alignment.
Note: Two options are available for visually verifying the tibia alignment.
The first option uses the standard alignment arch (00­5977-024-00). The pin placement of the tibial Pin Guide is offset 20 degrees medially to avoid soft tissues. Noting this offset placement, ensure the alignment rod is parallel with respect to the tibial shaft (Fig. 15).
The second option uses the alignment adapter which accounts for the medial offset and provides a more traditional observance (aligned to the tibia shaft) for the visual alignment check.
Fig.14
Alignment Adapter 00-5970-024-00
Alignment Arch 00-5977-024-00
Alignment Rod 00-5785-080-00
TECHNIQUE TIP 1B.B
When fixating the Cut Guide, excessive force from screws can alter the planned location and could influence the amount of varus/valgus in the cut.
TECHNIQUE TIP 1B.C
When performing the cuts, excessive force (considering the pressure from leaning on the guide) on the medial or lateral side of the cut guide could influence the amount of varus/valgus in the cut.
Instruments
Fig. 15
NexGen Tibial
Cut Guide
00-5997-075/
076-00
Alignment Adapter
00-5970-024-00
Alignment Rod
00-5785-080-00
6
Page 11
NexGen® Conventional Instruments – Tibial Technique
Position Tibial Sizing Plate
Position the tibial sizing plate to align with the drilled proximal holes (Fig. 16).
TECHNIQUE TIP 1B.D
In the event the drilled proximal holes are covered up following the proximal cut, clean the area. Alternatively, prior to removing the anterior pins, place the Pin Guide back over the anterior pins and locate the proximal holes through the Tibial Pin Guide.
SECTION
1B
Fig. 16
3o Fluted Tibial Plate (00-5999-060-01_02)
o
7
Fluted Tibial Plate (00-5997-060-01_02)
Trabecular Metal
(00-5953-012-02_08)
Tibial Plate
o
Fluted Tibila Plate (00-5999-060-03_08) Stemmed Tibial Plate (00-5951-019-01_10)
3
o
7
Fluted Tibial Plate (00-5997-060-03_10) A/P Wedged Tibial Plate (00-5987-043_048-00)
Trabecular Metal Monoblock Tibial Plate
(00-5887-043_048-00)
Follow 4-in-1 Femoral Surgical Technique
Following the positioning of the tibial plate, proceed with the surgery as suggested in the Zimmer MIS Multi-Reference 4-in-1 Femoral Instrumentation Surgical Technique.
Pegged Tibial Plate (00-5977-015-01_10)
Instruments
NexGen Tibial
Sizing Plate
(see above)
NexGen Locking
Tibial Tray
Provisional Handle
00-5977-096-00
-OR-
NexGen PRI
Offset Sizing Plate Handle
00-5953-096-00
Alignment Rod
00-5785-080-00
7
Page 12
SECTION
2A
NexGen® Posterior Referencing Instruments – Femoral Technique
NexGen Posterior Referencing Instruments – Femoral Technique
Position Femoral Pin Guide
Do not remove osteophytes from the femur.
Position the Pin Guide on the distal femur (Fig. 17).
Pin Guide will be captured by the anterior and proximal ridge of the femur. The guide fits over the edge on the proximal bone to increase locking.
The A/P axis reference line on the Pin Guide can be used to assess alignment. The most distal end of the guide is parallel to the epicondylar axis and can also be used as reference.
Apply anterior proximal pressure to the Pin Guide and verify a secure fit. Applying posterior pressure may destabilize the Pin Guide.
Fig. 17
Drill and Pin the Anterior Pin Holes
Drill and pin the medial and lateral pin holes of the Pin Guide using the Universal Disposable Drill (2001-00-00)
(Fig. 18).
Drill Distal Pin Holes
Drill the distal pin holes of the Pin Guide (do not place pins)
(Fig. 19).
Fig. 18
Fig. 19
Instruments
PSI Femoral
Pin Guide
00-5970-015/
018-00
Universal Pin
and Drill Kit
2001-00-000
8
Page 13
NexGen® Posterior Referencing Instruments – Femoral Technique
Place Distal Cut Guide and Cut Femur
Remove the Pin Guide by sliding it off the pins, leaving the anterior pins in the bone.
Secure the NexGen Posterior Referencing Distal Femoral Resection Guide (00-5901-064-00)
Verify the distal resection on the medial and lateral side with a resection guide (00-5977-084-00). Check alignment if desired, and make the cut (Fig. 20).
TECHNIQUE TIP 2A.A
When performing the cuts, excessive force (considering the pressure being applied to the guide) on the medial or lateral side of the cut guide could influence the amount of varus/valgus in the cut.
SECTION
2A
Fig. 20
Remove Distal Cut Guide
Locate the drilled pin holes.
Remove the anterior pins (Fig. 21).
TECHNIQUE TIP 2A.B
TECHNIQUE TIP
In the event the drilled distal holes are covered up following the distal cut, clean the area. Alternatively, prior to removing the anterior pins, place the Pin Guide back over the anterior pins and locate the distal pins through the distal Pin Guide holes.
Place Posterior Referencing Femoral Cut Guide
Secure the NexGen Posterior Referencing 4-in-1 Flex Femoral cut guide, verify resections with a resection guide and finish the femur following the instructions f rom the Zimmer NexGen CR-Flex and LPS-Flex Knees Surgical Technique with Posterior Referencing Instrumentation (Fig. 22).
Fig. 21
Instruments
NexGen Posterior
Referencing
Distal Femoral
Resection Guide
00-5901-064-00
Resection Guide
00-5977-084-00
NexGen Posterior
Referencing
4-in-1 Flex Femoral
Cut Guide
00-5901-041/048-00
Fig. 22
9
Page 14
SECTION
2B
NexGen® Posterior Referencing Instruments – Tibial Technique
NexGen Posterior Referencing Instruments – Tibial Technique
Position Tibial Pin Guide
Remove the meniscus.
Do not remove osteophytes from the tibia. Clean the bone around the ACL and anterior tibia where the guide fits.
Remove any soft tissues that may prevent proper placement of the Pin Guide.
Position the Pin Guide on the tibia (Fig. 23).
The rotation cylinders do not make contact with bone surface Both cylinders will be above the tibia surface. All other contact with the Pin Guide should ensure a proper fit.
Fig. 23
.
The mechanical axis and proximal resection reference lines on the Pin Guide can be used to assess alignment.
Drill and Pin Anterior Pin Holes
Drill and pin the medial and lateral anterior holes of the Pin Guide (Fig. 24).
Fig. 24
10
Instruments
Instruments
PSI Tibia
Pin Guide
00-5970-015/
018-00
Universal Pin
and Drill Kit
2001-00-000
Page 15
NexGen® Posterior Referencing Instruments – Tibial Technique
Establish Tibia Rotation (Drill Proximal Holes)
Tibia rotation may be established through the Pin Guide per surgeon discretion.
Drill the proximal medial and lateral holes of the Pin Guide (do not place pins) (Fig. 26).
SECTION
2B
TECHNIQUE TIP 2B.A
When drilling the proximal holes, excessive side loads through the proximal guide cylinders could alter the planned hole position.
Remove Tibial Pin Guide
Remove the Pin Guide by sliding anteriorly off the pins
(Fig. 27). Do not pry the Pin Guide from the tibia as this could
shift the pins.
Fig. 26
Fig. 27
Instruments
Instruments
PSI Tibia
Pin Guide
00-5970-015/
018-00
Universal Pin
and Drill Kit
2001-00-000
11
Page 16
SECTION
2B
NexGen® Posterior Referencing Instruments – Tibial Technique
Secure Cut Guide and Cut Tibia
Secure the NexGen Posterior Referencing Tibial Cut Guide. Verify the proximal resection on the medial and lateral side with a resection guide.
Make tibial cut (Fig. 29).
Use the appropriate right or left tibia cut guide.
0° PRI Left Cut Guide 00-5901-075-00
0° PRI Right Cut Guide 00-5901-076-00
Insert the Alignment Adapter into the Tibial Cut Guide, insert the Alignment Rod and check alignment.
Fig. 29
Note: If you wish to check alignment, use of the Tibia Alignment Rod with the NexGen Posterior Referencing Tibia Cut Guides allows for a standard observance of the tibia alignment rod (Fig. 30).
PRI Alignment Adapter 00-5901-086-00
Alignment Rod 00-5785-080-00
TECHNIQUE TIP 2B.B
When fixating the Cut Guide, excessive force from screws can alter the planned location and could influence the amount of varus/valgus in the cut.
TECHNIQUE TIP 2B.C
When performing the cuts, excessive force (considering the pressure from leaning on the guide) on the medial or lateral side of the cut guide could influence the amount of varus/valgus in the cut.
Fig. 30
12
Instruments
Instruments
NexGen PRI
Tibial Cut Guide
00-5901-075/
076-00
PRI Alignment
Adapter
00-5901-086-00
Alignment Rod
00-5785-080-00
Page 17
NexGen® Posterior Referencing Instruments – Tibial Technique
Position Tibial Sizing Plate
Position the tibial sizing plate to align with the drilled proximal holes (Fig. 31).
TECHNIQUE TIP 2B.D
In the event the drilled proximal holes are covered up following the proximal cut, clean the area. Alternatively, prior to removing the anterior pins, place the Pin Guide back over the anterior pins and locate the proximal holes through the Tibial Pin Guide.
SECTION
2B
Fig. 31
3o Fluted Tibial Plate (00-5999-060-01_02)
o
7
Fluted Tibial Plate (00-5997-060-01_02)
Trabecular Metal Tibial Plate
(00-5953-012-02_08)
o
Fluted Tibila Plate (00-5999-060-03_08) Stemmed Tibial Plate (00-5951-019-01_10)
3
7o Fluted Tibial Plate (00-5997-060-03_10) A/P Wedged Tibial Plate (00-5987-043_048-00)
Trabecular Metal Monoblock Tibial Plate
(00-5887-043_048-00)
Follow NexGen Posterior Referencing Instrumentation Surgical Technique
Following the positioning of the tibial component, proceed with the surgery as suggested in the Zimmer NexGen CR-Flex and LPS-Flex Knees Surgical Technique with Posterior Referencing Instrumentation.
Pegged Tibial Plate (00-5977-015-01_10)
Instruments
Instruments
NexGen Tibial
Sizing Plate
(see above)
NexGen Locking
Tibial Tray
Provisional Handle
00-5977-096-00
-OR-
NexGen PRI
Offset Sizing Plate Handle
00-5953-096-00
Alignment Rod
00-5785-080-00
13
Page 18
Page 19
Page 20
Note: Materialise and the Materialise logo are trademarks of Materialise NV
This documentation is intended exclusively for physicians and is not intended for laypersons. Information on the products and procedures contained in this document is of a general nature and does not represent and does not constitute medical advice or recommendations. Because this information does not purport to constitute any diagnostic or therapeutic statement with regard to any individual medical case, each patient must be examined and advised individually, and this document does not replace the need for such examination and/or advice in whole or in part. Please refer to the package inserts for important product information, including, but not limited to, contraindications, warnings, precautions, and adverse effects.
Contact your Zimmer representative or visit us at www.zimmer.com
The CE mark is valid only if it is also printed on the product label.
97-5970-075-00 1203-K02 ©2012 Zimmer, Inc.
Loading...