Zimmer® PSI is not designed for use with the Zimmer
MIS Quad-Sparing
•
Zimmer PSI should be used in conjunction with a femur
first technique.
•
Verify stability and fixation of all pins. When securing the
conventional tibia cutting guides, avoid the use of spring
pins as these may cause stress / unwanted shift on the
tibia cutting guide.
Position the Pin Guide on the distal femur (Fig. 1).
•
Pin Guide will be captured by the anterior and proximal
ridge of the femur. The guide fits over the edge on the
proximal bone to increase locking.
•
The A/P axis reference line on the Pin Guide can be used
to assess alignment. The most distal end of the guide
is parallel to the epicondylar axis and can also be used
as reference.
•
Apply anterior proximal pressure to the Pin Guide
and verify a secure fit. Applying posterior pressure
may destabilize the Pin Guide.
SECTION
1A
Fig. 1
Drill and Pin the Anterior Pin Holes
•
Drill and pin the medial and lateral anterior pin holes of
the Pin Guide using the Universal Disposable Drill
(2001-00-000) (Fig. 2).
Drill Distal Pin Holes
•
Drill the distal pin holes of the Pin Guide (do not place pins)
Remove the Pin Guide by sliding it off the pins, leaving the
anterior pins in the bone.
•
Secure the MIS Distal Cut Guide (00-5967-036-00).
•
Verify the distal resection on the medial and lateral side
with a resection guide (00-5977-084-00). Check alignment
if desired, and make the cut (Fig. 4).
TECHNIQUE TIP 1A.A
When performing the cuts, excessive force (considering the pressure
being applied to the guide) on the medial or lateral side of the cut
guide could influence the amount of varus/valgus in the cut.
Fig. 4
Remove Distal Cut Guide and Place Distal Pins
•
Locate the drilled pin holes.
•
Remove the anterior pins
•
Place the distal pins in drilled holes of the distal femur
(Fig. 5).
TECHNIQUE TIP 1A.B
In the event the drilled distal holes are covered up following the distal
cut, clean the area. Alternatively, prior to removing the anterior pins,
place the Pin Guide back over the anterior pins and locate the distal
pins through the distal Pin Guide holes (Fig. 6).
Do not remove osteophytes from the tibia. Clean the bone
around the ACL and anterior tibia where the guide fits.
Remove any soft tissues that may prevent proper placement
of the Pin Guide.
•
Position the Pin Guide on the tibia (Fig. 8).
•
The rotation cylinders do not make contact with bone surface
Both cylinders will be above the tibia surface. All other
contact with the Pin Guide should ensure a proper fit.
•
The mechanical axis and proximal resection lines may be
used to asses the alignment. Additionally, the alignment
arch (00-5977-024-00) and rod (00-5785-080-00) may
be used to verify alignment. Additionally, the alignment
adapter (00-5970-024-00) and rod (00-5785-080-00)
may be used to verify alignment.
Fig. 8
.
Drill and Pin Anterior Pin Holes
•
Drill and pin the medial and lateral anterior holes of the
Pin Guide (Fig. 9).
Secure the NexGen Tibial Cut Guide. Verify the
proximal resection on the medial and lateral side
with a resection guide.
•
Make tibial cut (Fig. 14).
•
Use the appropriate right or left tibia cut guide.
0° Left Cut Guide00-5997-075-00
0° Right Cut Guide00-5997-076-00
•
Insert the Alignment Arch into the Tibial Cut Guide, insert
the Alignment Rod and check alignment.
Note: Two options are available for visually verifying the tibia
alignment.
•
The first option uses the standard alignment arch (005977-024-00). The pin placement of the tibial Pin Guide
is offset 20 degrees medially to avoid soft tissues. Noting
this offset placement, ensure the alignment rod is parallel
with respect to the tibial shaft (Fig. 15).
•
The second option uses the alignment adapter which
accounts for the medial offset and provides a more
traditional observance (aligned to the tibia shaft) for the
visual alignment check.
Fig.14
Alignment Adapter00-5970-024-00
Alignment Arch00-5977-024-00
Alignment Rod00-5785-080-00
TECHNIQUE TIP 1B.B
When fixating the Cut Guide, excessive force from screws can alter the
planned location and could influence the amount of varus/valgus in
the cut.
TECHNIQUE TIP 1B.C
When performing the cuts, excessive force (considering the pressure
from leaning on the guide) on the medial or lateral side of the cut
guide could influence the amount of varus/valgus in the cut.
Position the tibial sizing plate to align with the drilled
proximal holes (Fig. 16).
TECHNIQUE TIP 1B.D
In the event the drilled proximal holes are covered up following the
proximal cut, clean the area. Alternatively, prior to removing the
anterior pins, place the Pin Guide back over the anterior pins and
locate the proximal holes through the Tibial Pin Guide.
Following the positioning of the tibial plate, proceed with the
surgery as suggested in the Zimmer MIS Multi-Reference 4-in-1
Femoral Instrumentation Surgical Technique.
Position the Pin Guide on the distal femur (Fig. 17).
•
Pin Guide will be captured by the anterior and proximal
ridge of the femur. The guide fits over the edge on the
proximal bone to increase locking.
•
The A/P axis reference line on the Pin Guide can be used
to assess alignment. The most distal end of the guide is
parallel to the epicondylar axis and can also be used as
reference.
•
Apply anterior proximal pressure to the Pin Guide and verify
a secure fit. Applying posterior pressure may destabilize the
Pin Guide.
Fig. 17
Drill and Pin the Anterior Pin Holes
•
Drill and pin the medial and lateral pin holes of the Pin
Guide using the Universal Disposable Drill (2001-00-00)
(Fig. 18).
Drill Distal Pin Holes
•
Drill the distal pin holes of the Pin Guide (do not place pins)
Remove the Pin Guide by sliding it off the pins, leaving the
anterior pins in the bone.
•
Secure the NexGen Posterior Referencing Distal Femoral
Resection Guide (00-5901-064-00)
•
Verify the distal resection on the medial and lateral side
with a resection guide (00-5977-084-00). Check alignment
if desired, and make the cut (Fig. 20).
TECHNIQUE TIP 2A.A
When performing the cuts, excessive force (considering the pressure
being applied to the guide) on the medial or lateral side of the cut
guide could influence the amount of varus/valgus in the cut.
SECTION
2A
Fig. 20
Remove Distal Cut Guide
•
Locate the drilled pin holes.
•
Remove the anterior pins (Fig. 21).
TECHNIQUETIP2A.B
TECHNIQUE TIP
In the event the drilled distal holes are covered up following the distal
cut, clean the area. Alternatively, prior to removing the anterior pins,
place the Pin Guide back over the anterior pins and locate the distal
pins through the distal Pin Guide holes.
Place Posterior Referencing Femoral Cut Guide
•
Secure the NexGen Posterior Referencing 4-in-1 Flex
Femoral cut guide, verify resections with a resection
guide and finish the femur following the instructions f
rom the Zimmer NexGen CR-Flex and LPS-Flex Knees
Surgical Technique with Posterior Referencing
Instrumentation (Fig. 22).
Do not remove osteophytes from the tibia. Clean the bone
around the ACL and anterior tibia where the guide fits.
•
Remove any soft tissues that may prevent proper placement
of the Pin Guide.
•
Position the Pin Guide on the tibia (Fig. 23).
•
The rotation cylinders do not make contact with bone surface
Both cylinders will be above the tibia surface. All other
contact with the Pin Guide should ensure a proper fit.
Fig. 23
.
•
The mechanical axis and proximal resection reference lines
on the Pin Guide can be used to assess alignment.
Drill and Pin Anterior Pin Holes
•
Drill and pin the medial and lateral anterior holes of the
Pin Guide (Fig. 24).
Secure the NexGen Posterior Referencing Tibial Cut Guide.
Verify the proximal resection on the medial and lateral side
with a resection guide.
•
Make tibial cut (Fig. 29).
•
Use the appropriate right or left tibia cut guide.
0° PRI Left Cut Guide00-5901-075-00
0° PRI Right Cut Guide00-5901-076-00
•
Insert the Alignment Adapter into the Tibial Cut Guide,
insert the Alignment Rod and check alignment.
Fig. 29
Note: If you wish to check alignment, use of the Tibia
Alignment Rod with the NexGen Posterior Referencing Tibia
Cut Guides allows for a standard observance of the tibia
alignment rod (Fig. 30).
PRI Alignment Adapter00-5901-086-00
Alignment Rod00-5785-080-00
TECHNIQUE TIP 2B.B
When fixating the Cut Guide, excessive force from screws can alter the
planned location and could influence the amount of varus/valgus in
the cut.
TECHNIQUE TIP 2B.C
When performing the cuts, excessive force (considering the pressure
from leaning on the guide) on the medial or lateral side of the cut
guide could influence the amount of varus/valgus in the cut.
Position the tibial sizing plate to align with the
drilled proximal holes (Fig. 31).
TECHNIQUE TIP 2B.D
In the event the drilled proximal holes are covered up following the
proximal cut, clean the area. Alternatively, prior to removing the
anterior pins, place the Pin Guide back over the anterior pins and
locate the proximal holes through the Tibial Pin Guide.
Following the positioning of the tibial component,
proceed with the surgery as suggested in the ZimmerNexGen CR-Flex and LPS-Flex Knees Surgical Technique
with Posterior Referencing Instrumentation.
Pegged Tibial Plate (00-5977-015-01_10)
Instruments
Instruments
NexGen Tibial
Sizing Plate
(see above)
NexGen Locking
Tibial Tray
Provisional Handle
00-5977-096-00
-OR-
NexGen PRI
Offset Sizing
Plate Handle
00-5953-096-00
Alignment Rod
00-5785-080-00
13
Page 18
Page 19
Page 20
Note: Materialise and the Materialise logo are trademarks of Materialise NV
This documentation is intended exclusively for physicians and is not intended for laypersons.
Information on the products and procedures contained in this document is of a general nature
and does not represent and does not constitute medical advice or recommendations. Because
this information does not purport to constitute any diagnostic or therapeutic statement with
regard to any individual medical case, each patient must be examined and advised individually,
and this document does not replace the need for such examination and/or advice in whole or
in part. Please refer to the package inserts for important product information, including, but not
limited to, contraindications, warnings, precautions, and adverse effects.
Contact your Zimmer representative or visit us at www.zimmer.com
The CE mark is valid only if it is also printed on the product label.