
Illustrations
Front of the device
Treatment tube connection

Illustrations
Screens and display
Arrow for navigation / changing parameters
Arrow for navigation / changing parameters
Arrow for navigation / changing parameters
Arrow for navigation / changing parameters
Airflow level display (graphical)
Airflow level display (numerical)
Treatment time display (graphical)
Treatment time display (numerical)

In the instructions for use this symbol indicates “Danger”.
Caution
In the instructions for use this symbol indicates “Caution” with regard to possible
damage of the device.
Applied part type B
do not push sideways
max. permitted load on glass plate 35 kg
Follow instructions for use.
Instructions for use
Serial number
Item number
Manufacturer
Date of manufacture

Front of the device
Rear of the device
1.
Indications / Contraindications
1.1 Indications / Contraindications for Cryo 6 Physio
1.2 Indications / Contraindications for Cryo 6 Derma
1
2
2.
Side effects 3
3.
Application information
3.1 General
3.2 Cryotherapy
4
5
4.
Warnings 6
5.
Cryo 6 – in brief 7
6.
Intended use 8
7.
Device set-up
7.1 Fitting
7.2 Fitting the supporting arm
9
11
8.
Default settings
8.1 Configuration menu
8.2 Device type and device configuration
8.3 Default settings and service
8.4 Defrosting
8.5 Own programs and favourites
8.6 Information and maintenance programs
9.
Operating instructions
9.1 Device description
9.2 Performing cryotherapy
9.3 Start-up screens
9.4 Pre-set parameters / change
21
22
23
24

10.
Technical information 25
11.
Cleaning / Disinfection 26
12.
CE mark / manufacturer 27
13.
Scope of delivery and accessories 28
14.
Device combinations 30
15.
Safety and maintenance
15.1 Safety
15.2 Maintenance
31
32
16.
Functional test 33
17.
Safety check / metrological check 34
18.
Error messages / troubleshooting / disposal 35
19.
EMC declaration 37
These instructions for use are an integral part of the device.
They must be stored with the device and kept accessible at all times for anyone authorised to operate this device.
The instructions for use are valid as of November 2016.

Indications / Contraindications
1.1 Indications / Contraindications for Cryo 6 Physio
Cryo 6 Physio is designed for the following uses:
x Reduction of pain and stiffness and to support treatment of acute and chronic
painful conditions of the musculoskeletal system
- Arthritis
- Bursitis
- Tendinitis
- Tenosynovitis
- Myositis
- Fibrositis
- Muscle tenseness
- Cervical syndrome
- Post-whiplash disorders
- Lumbar syndrome
- Muscle or joint injuries
x Pain reduction, improved mobility and reduction of joint stiffness in rheumatic
conditions
- Rheumatoid arthritis (progressive chronic arthritis)
- Post-synovectomy condition
x Reduction of spasticity in neurological disorders
- Multiple sclerosis
- Post-apoplectic hemiplegia
other possible uses:
x to support movement therapy by prior cooling
x in combination with compresses in acute injuries
x to prevent oedema and haematoma
x in sports, as preventive early-stage treatment before the development of pain
immediately after major exertion, such as after a competitive event
x to treat muscular trigger points in combination with stretching
Other indications must be discussed with the patient’s doctor or derived from
relevant specialist literature.
x Cold agglutinin disease and cold haemolysis
x Cold urticaria
x Parts of the body with impaired circulation
x Raynaud’s disease
x Parts of the body with impaired sensitivity
x Trophic disorders
x Hypersensitivity to cold
x When treating children, parts of the body that are not being treated must be
covered and kept warm.
x When treating the face, the eyes must be covered.
x Do not treat the face or trunk if the patient has severe arterial hypertension and
severe cardiac insufficiency.
x The patient should not become excessively cold during treatment.

Indications / Contraindications
1.2 Indications / Contraindications for Cryo 6 Derma
Cryo 6 Derma is designed for the following uses:
Reduction of pain and inflammation by cooling with cold air during and after
dermatological and cosmetic treatments
x Laser therapy
x Injections
The analgesia associated with cold air makes the treatment much more pleasant for
the patient.
Reduction in the side effects that are often related to laser treatment:
x less erythema
x fewer swellings and encrustations
x reduction of pain and thermal skin damage
x Cold agglutinin disease and cold haemolysis
x Cold urticaria
x Parts of the body with impaired circulation
x Raynaud’s disease
x Parts of the body with impaired sensitivity
x Trophic disorders
x Hypersensitivity to cold
x When treating children, parts of the body that are not being treated must be
covered and kept warm.
x When treating the face, the eyes must be covered.
x Do not treat the face or trunk if the patient has severe arterial hypertension and
severe cardiac insufficiency.
x The patient should not become excessively cold during treatment.
When using Cryo 6 with ablative lasers, make sure that the wound produced is given
appropriate antiseptic care after the treatment.

Side effects Cold-related skin damage such as skin reddening and even mild frostbite and
chilblains can occur, especially in sensitive patients.

Prior to using the device on a patient, the user should become familiar with the
instructions for use and individual treatment methods to be used as well as the
indications / contraindications, warnings and application information. Additional
sources of information about the treatment should be followed.
These instructions for use must always be stored with the device and kept
accessible at all times for anyone authorised to operate this device.
After the device has been in transit on its side, place it upright and keep it in an
upright position for at least 30 minutes before switching it on. Otherwise the
compressor will be damaged.
Before use, ensure that the device is powered via a properly grounded mains socket
(electrical installation according to DIN VDE 0100 Part 710). The device must only
be operated with the supplied power cable. The power cable must be protected
against mechanical stress.
Magnetic and electrical fields can affect the function of the device. For this reason,
do not operate Cryo 6 in the vicinity of devices which generate strong
electromagnetic fields (X-ray or diathermy equipment, MRI machines). Please keep
a safe distance of several meters.
Cryo 6 is not suitable for use in areas with an explosive, flammable or combustive
Do not place the device adjacent to heat sources (heating, hot mud products, sauna
etc.), and leave a gap of at least 50 cm between the device and a wall (to allow for
the supply of treatment products and cold air).
When used in combination with a laser device, the laser ventilation system must not
During use, the device is to be located in a position allowing direct access to the
device´s central mains supply so that it can be disconnected from the mains at any
To avoid the risk of electric shock, the plug must be disconnected from the power
supply before performing any cleaning or maintenance activities.
Caution
Inspect the device before use. If there is any damage, it must not be used.
Only accessories provided by Zimmer MedizinSysteme GmbH must be used.

Patients should be given an explanation of the aims and effects of cold air therapy
with Cryo 6 prior to treatment.
Tell patients that if they experience any discomfort, such as a sensation of extreme
cold during treatment, they must tell the therapist immediately.
During treatment, the therapist should ask patients if they are feeling all right.
The treatment parameters (airflow level and distance from area being treated)
should be adjusted if necessary.
Use the most appropriate nozzle for the various applications, from treatments
administered to large areas of the body through to trigger point treatment. Simply
unscrew a nozzle and screw in a different one when you want to switch.
Short distances between the skin and the nozzle are suitable for brief cooling:
x 1 cm distance for treatment times of up to 10 seconds,
x 5 cm distance for treatment times of up to 30 seconds.
Medium distances between the skin and the nozzle are suitable for dynamic cooling
of larger areas of skin and static cooling of smaller areas:
x 10-15 cm distance for treatment times of approximately 15-30 minutes.
Long distances between the skin and the nozzle are suitable for dynamic cooling of
large areas of skin:
x 15-20 cm distance for treatment times of over 30 minutes.
Longer treatment times are needed for joints and muscles, as otherwise only the
surface and upper layers of the skin are cooled. In patients with joint inflammation,
brief cooling causes reactive hyperaemia.
In patients with diseases of the musculoskeletal and support systems, best results
are obtained with a distance of 5-20 cm between the outlet nozzle and the surface of
the skin.
When used in conjunction with laser devices, a distance of 5 cm for a treatment area
of 10 cm² is recommended. The distance should be increased when treating larger
areas. A longer treatment period is needed to obtain adequate cooling.
Please note that the temperature of the air stream can increase during prolonged
treatment.
Cold air output can be reduced if the environmental conditions are unfavourable
(room temperature over 30°C and high air humidity).

The patient must not be left unattended during therapy.
Any treatment instructions regarding treatment location, duration and intensity
require medical knowledge and should be given by authorised physicians, therapists
and health paraprofessionals. These instructions must be followed.
Use in wet areas is not permitted and may in case of non-compliance lead to
considerable damage and endanger both the patient and the user.
Dispose of the packaging material properly. Make sure that it is not accessible to
children.
The cold air stream must not be applied to open wounds.
When using Cryo 6 to cool skin, do not increase the output of the laser device
beyond the level recommended by the manufacturer.
When applying cold to the face it is essential to protect the eyes.
The air stream should be directed evenly over the area to be treated. Avoid static or
excessively intensive cooling as this can lead to cold burns and hypothermia.
The use of the device out of the settings or applications specified in the instructions
for use may lead to hazard by the uncontrolled effects of cold.
Children can become too cold. Parts of the body that are not being treated should be
covered and kept warm. This is also recommended for adults undergoing prolonged
Frostbite can occur if the skin temperature falls to 0°C or below This can occur if the
nozzle is less than 10 cm away from the skin. If it is not possible to respect this
distance on therapeutic grounds, it is recommended that the nozzle be directed
dynamically over the area being treated.
Do not push the sides of the device that are marked with the warning symbol.
Do not lean on the device.

A compact cold-air device used to treat diseases of the support and
musculoskeletal systems.
Cryo 6 can also be used to cool the skin of patients undergoing dermatological laser
treatment to relieve pain and thermal skin damage.
It blows very cold air onto the parts of the body to be treated at various speed
settings.
It is very easy to use thanks to the pre-set program parameters and also very
powerful in prolonged use, matching the performance of much larger devices.
Its clear LCD display and ergonomic keyboard reflect the state of the art.
Therapy can be customised through six user-focused, pre-set programs as well as
three programs with free choice of settings and an individual start-up program.
concentration of cold
achieved?
Air volume can be set to one of nine levels depending on the size of the area to be
treated and how accessible it is.
Three nozzle sizes are supplied as standard for individual therapy (diameters of 5,
10 and 15 mm).
A special nozzle is also available for cooling the skin of patients undergoing
dermatological laser treatment.
Use of the device is reserved for medical professionals
(such as physicians, therapists, medical paraprofessionals).

The Cryo 6 cold air device is used to cool skin
- when treating injuries or diseases affecting the musculoskeletal system
- in order to reduce local pain
- in patients undergoing dermatological laser treatment to prevent thermal skin
1. Cryo 6 Physio to treat diseases of the support and musculoskeletal systems.
2. Cryo 6 Derma to cool the skin of patients undergoing dermatological laser
treatment to relieve pain and thermal skin damage.
The device types are preconfigured in the factory and differ in respect of the pre-set
fan settings and treatment periods as well as the start-up program.

Cryo 6 must be placed upright and kept in an upright position for at least 30 minutes
before being switched on if the device has been transported on its side or had fitting
work done on it.
Otherwise the compressor could be damaged.
Connect the power cable to the provided port (11) on the device and connect the
cable to the mains.
The device may only be connected to power outlets with a protective contact
The device is switched on using the toggle switch (10).
The device is switched off using the toggle switch (10).
To fully (all poles) disconnect the device from the mains, the power cable must be
All cables must be protected from pinching or other mechanical damage.
Insert the treatment tube into the connector (5) on the front of the device and lock it
into place.
The treatment tube can be stored in the curved groove above the control panel (3)
Fitting the castor
guards
Attach the castor guards (2) to the wheel holders.
Four spacing bolts are already fitted to the top of the device to allow the glass plate
to be attached.
Place a silicon intermediate disc onto each of the four spacing bolts.
Then put the glass plate on the spacing bolts.
Now secure the glass plate by screwing one of the clamp bearings provided into
each of the four spacing bolts.
You do not need to use a tool for this, in order to prevent damage to the glass plate.
Use your fingers to tighten the clamp bearings.

Carry out the steps described above in reverse order to dismantle the glass plate.
Never lift the device by the glass plate.
The treatment nozzle and the nozzle attachments can be kept in the storage box.

7.2 Fitting the supporting arm
The Cryo 6 device can be fitted with a supporting arm as an extra option.
This arm allows a particular part of the body to be cooled without
the therapist having to hold the tube.
The supporting arm accessory kit consists of:
1 supporting arm mounting bracket
2 M6x16 screws and allen key for mounting
1 anti-rotation element
1 supporting arm pole
1 rotating supporting arm
1 tube clip
1 tube holder
Also required:
Glass plate
supporting arm
Guide through
glass plate and
anti-rotation
element
Supporting arm
pole
Supporting arm securing element

7.2 Fitting the supporting arm
The glass plate must be attached to the device so that a supporting arm can be fitted
(see 7.1 Fitting) The glass plate contains an opening through which the supporting
The glass plate can be fitted so that it is symmetrical with the device. If a supporting
arm is also fitted, the glass plate is laid on the spacing bolts in such a way that the
hole for positioning the supporting arm is on the correct side of the device.
Step 1 - fitting the guide sleeve
Assemble the guide sleeve (supplied with the glass plate) in the hole made for this
purpose.
Step 2 - fitting the supporting arm securing element
Use the Allen key and the M6x16 screws to screw the mounting bracket to the base
of the device.
Threaded holes are already in place on the base of the device for this purpose.
Caution
The view from below is only presented for clarification, please DO NOT put the
device down for fitting!
Threaded
holes
Guide sleeve
Part 1
Guide sleeve
Part 2

7.2 Fitting the supporting arm
Step 3 - attaching the anti-rotation element and pole
Place the anti-rotation element on the glass plate spacer as shown in the picture.
Then pass the supporting arm pole through the guide sleeve and the anti-rotation
element from above, and lock it into place in the supporting arm securing element.
Anti-rotation
element

7.2 Fitting the supporting arm
Step 4 - attaching the supporting arm
Screw the supporting arm (a) onto the fitted pole. The recesses must be aligned in
order to ensure that the supporting arm can move freely up and down. They must lie
along an axis and allow the pole to bend.
Step 5 - attaching the tube clip / tube holder
Screw the tube clip (b) onto the rotating supporting arm, and hang the tube holder (c)
onto the fixing screw of the supporting arm.
The treatment tube can now be passed through the tube securing element and the
handpiece can be engaged in the tube clip.
- Recesses below
- Same bend orientation

The Cryo 6 performs a self-test once it has been switched on.
You can switch to the configuration menu during the self-test and subsequent pre-
Press the “Menu” key to select the menu.
You can use the menu to call up technical information, change factory settings,
select a favourite program or define your own programs.
You can choose from eleven menu items:
- Service
- Back to treatment
- Defrost
- Own programs
- Favourite
- Info
- Automatic maintenance program S01
- Automatic maintenance program S02
- Cryo device type
- Device configuration
- Basic settings
Use the arrows (14 and 15) to navigate among the individual menu items.
Press the “Select” key (17) to select a menu item.
Select the “Back to treatment” menu item to end the menu and return to the
treatment screen.

8.2 Device type and device configuration
Cryo 6 can be used as “Cryo 6 Physio” or “Cryo 6 Derma”.
Use the arrows (14 and 15) to navigate to the pre-set configuration that you want to
use. Then press the “Save” key to select it.
See chapter 9.3 for more information about device types and specific parameters.
This menu item offers various settings, such as languages or aspects of the start-up
process.
Use the arrows (14 and 15) to navigate through the sub-menu.
Press the “Select” key (17) to select an item.
The following languages are available:
German, English, French, Italian and Polish.
The “External control input”, “External control output” and “External start input” submenus are inactive.
Cryo 6 offers two different options here:
1. Direct start
2. Programs
Press the “Select” key to select the menu.
Press the “Change” key to switch between direct start and programs.
Press the “Save” key to save the start-up process you want to run and the “Yes” key
to confirm and apply your choice.

8.3 Default settings and service
You can restore factory settings under this menu item.
Press the “YES” key to reset all the parameters that have been changed back to the
factory settings.
Own programs and the favourite program that you have saved will not be lost.
Press the “NO” key to retain changed parameters.
This area is password-protected and is only accessible to people who have been
trained by Zimmer MedizinSysteme GmbH.
Press the “OK” key to quit the menu.

The “Defrost” menu item triggers defrosting of the heat exchanger.
Perform defrosting when the outgoing air stream is less strong than usual. This is
often caused by the Cryo 6 being placed in a humid environment such as a bathing
area or sauna. We also recommend defrosting the Cryo 6 and emptying the defrost
water container if it has not been used for a prolonged period, for instance during
Press the “Select” key to immediately begin the defrosting process. The defrosting
process is carried out in standby mode.
The treatment fan starts up and runs until the cooling unit’s defrosting temperature
has been reached. The blower then switches off.
The message “Defrost complete” appears on the display once the defrosting process
has been completed.
Press the “Back” key if you want to interrupt the defrosting process before it has

8.5 Own programs and favourites
The “Internal programs”, “Programming of 1 and 2-phase programs” and “Favourite
program” can only be activated if “Programs” is activated as the start-up process
Cryo 6 offers storage options for three own programs in addition to the six factory
pre-set programs which cannot be permanently changed.
Fan level and time can be allocated individually to a storage location.
The programs can be programmed either for one phase or for two consecutive
Use the arrows (14 and 15) to navigate among the three storage locations.
Press the “Select” key (17) to select the stored program that you want to use.
You can then use the arrows (15 and 16) to select the desired fan level. The fan
symbol (20) flashes.
Press the “Save” key to save the fan level.
The fan symbol now stays lit and the time symbol (22) flashes.
Use the arrows (14 and 15) to select the desired time.
Press the “Save” key to save the time.
Activate the “Back” key to stop programming the 1-phase program. Press the “Back”
Here you can arrange for two programs with different fan levels and times to run
consecutively in a single treatment session.
Start in the same way as when programming a 1-phase program. Now, instead of
pressing the “Back” key to quit the program, press the “Save” key again. The
parameters for phase 1 now appear in small font at the top of the display.
Here you can select a personal favourite program that is automatically loaded
whenever the device is started up. You can choose one of the six factory pre-set
programs P1-P6 or one of the own special programs S1-S3. Use the arrows (14 and
15) to navigate among the program locations. Choose the program you want to
define as the favourite and press the “Select” key to confirm.

8.6 Information and maintenance programs
Technical information about various device components is displayed under this
menu item.
No settings can be applied.
The S01 and S02 maintenance programs simulate long-term operation and logs the
correct condition of the device.
Any faults are diagnosed and recorded.
This program simulates two 15-minute treatment sessions and one defrosting
process. The program records key parameters such as temperature and duration.
The current temperatures of the compressor and evaporator can be checked during
The device cools down to - 43°C / - 38°C and starts a 15-minute treatment session
This program is repeated until the start/stop key is pressed.

As soon as Cryo 6 is switched on, the device starts to pre-cool to the minimum
temperature to be reached. The compressor and condenser fan work together during
this phase. At the same time the device performs a self-test. No data can be entered
via the display during the self-test or pre-cooling. The only active key is the “Menu” key
(1), which users can press to reach the configuration menu. When sufficient precooling has been carried out, the display automatically switches to the start screen.
Cryo 6 is now ready for use.
During treatment, the treatment fan blows cold air through the treatment tube. At the
same time, the compressor / condenser starts up automatically during treatment in
order to ensure constant cooling.
Cryo 6 is in standby mode when no treatment is being carried out. The compressor
automatically starts up as soon as a certain temperature is exceeded, in order to
ensure that sufficient cold air is available at all times. Treatment can be started
immediately from standby mode.
We recommend that about 10 minutes should be left between the end of pre-cooling
and the start of treatment. This ensures that the device has reached peak cold air
output.
We also recommend that Cryo 6 is only switched off when there are prolonged breaks
in treatment or at the end of the day.

9.2 Performing cryotherapy
Switch on the device at the mains switch (10), display lights up.
The current status (self-test, pre-cooling) can be read off at the display during
initialisation.
The factory settings mean that the start-up screen and start-up process are different
depending on whether you are using the device with the Cryo 6 Physio or Cryo 6
Derma pre-set parameters.
See the next page for illustrations of both start-up screens.
Program and treatment
time
The device is ready to use as soon as the treatment screen (in this case “Physio”)
10:00
P2
7
P
P
Use the program selection keys (14 and 15) to set the desired treatment program
and the timer keys (16 and 17) to set the desired treatment time.
The fan level cannot be altered before the start of treatment.
Press the start/stop (18) key to start the program running.
The fan level (14 and 15) and treatment time (16 and 17) can be altered by pressing
the corresponding keys during treatment.
7
05:30
If parameters are altered during treatment, the factory settings are restored after the
end of treatment (or the favourite program settings if a favourite program has been
An acoustic signal indicates the end of treatment, and the fan is automatically
switched off. This also applies to premature interruption of treatment via the

The factory settings mean that the start-up screens and start-up process of the
Physio and Derma versions are different.
Derma start-up screen / direct start
Direct start: Fan level and time can be selected directly.
Physio start-up screen / programs
Programs: direct selection option among various
pre-set programs and the use of
The device configuration menu item offers a free choice of the desired start-up
process. See chapter 8 for instructions.

9.4 Pre-set parameters / change
Cryo 6 Derma is factory-set for direct start operation. Use the “Device configuration”
menu to switch to program operation.
Cryo 6 offers options to adjust the fan level and treatment time to individual
requirements and to save these changes (see chapter 8.5).
The treatment fan can be set to nine levels using the keys (14 and 15).
Treatment time can be set to any duration between 00:00 and 99:59 minutes.
If an excessively high treatment time has been selected by mistake, press the
start/stop key (18) twice to return to the factory pre-set time.

100-120 V / 50 Hz / 60 Hz (9-12 A)
220-240 V / 50 Hz (7 A)
240 V / 60 Hz (7 A)
16 A circuit breaker in mains switch
H 645 mm x W 390 mm x D 680 mm
+10°C to +35°C, 20% to 80% relative humidity, without condensation
-10°C to +50°C, 10% to 90% relative humidity, without condensation
0°C to +40°C, 10% to 90% relative humidity, without condensation
Storage and transport only in original packaging.
(room temperature up to
25°C)
at the start of treatment: - 31°C
maximum: - 18°C (after 15 minutes of treatment)
Figures accurate to +/- 10%
The maximum weight and size of devices (such as laser devices) that can be
placed on the glass plate is 35 kg and 50 x 50 x 35cm (W x D x H).
Subject to technical changes.

- Before starting any maintenance and cleaning measures the device must always
be switched off at the main switch and the mains cable must be disconnected.
- Make sure that when cleaning and disinfecting the labelling of the device (such as
warnings, labels of control devices, identification plate) is not damaged.
- Make sure that during cleaning and disinfection no liquids penetrate the device. Do
not use sprays.
- If during cleaning or disinfecting liquid penetrates the device, please put the device
out of service, protect it from being used again and contact your service
representative.
- The device and its applied part are considered as uncritical in relation to hygiene
due to the use on non-injured and healthy skin.
Cleaning: In the event of visible contamination, the housing, cables and accessories
can be cleaned using commercially available alcohol-free plastic cleaners. Wipe the
surface until the dirt is removed, using a soft cloth soaked according to the
specifications of the manufacturer of the cleaning agent but not dripping wet.
Disinfection: We recommend that disinfection is to be carried out at least once a
week, as well as if there is any indication of contamination. Consult with your
hygiene specialist when doing so. Always perform cleaning prior to disinfection.
Housing, cables and accessories can be disinfected using disinfectant wipes. Use a
commercially available alcohol-free disinfectant for metal and plastic, with
bactericidal, virucidal and fungicidal properties. Observe the application instructions
of the manufacturer. Wipe all surfaces using a soft cloth soaked according to the
specifications of the manufacturer of the disinfectant, but not dripping, or with cloths
pre-impregnated with disinfectant (wipes).
Also observe requirements for drying or post-cleaning, where applicable.
Cleaning: The defrost water container (7) should be cleaned whenever it has been
emptied. Follow the procedure described under "Housing / Accessories".
Disinfection: Follow the procedure described under "Housing / Accessories".
The device may only be operated with the defrost water container in place.
The air filter (9) should be cleaned regularly, and in any event after no more than
200 operating hours (maintenance notice appears on the display).
This is done by vacuuming the air filter from the outside using a commercially
available domestic vacuum cleaner.
Use the device only in a hygienic environment.

in accordance with the EC directive on medical devices 93/42/EEC.
Zimmer MedizinSysteme GmbH
Junkersstraße 9
D-89231 Neu-Ulm
Tel.
+49 731. 9761-291
Fax +49 731. 9761-299

Scope of delivery / accessories
1 glass plate with handles
1 screw Set for glass shelf
4 plastic castors 75 mm
1 defrosted water container
1 treatment tube Light Neo
1 glass plate with handles
1 screw Set for glass shelf
4 plastic castors 75 mm
1 defrosted water container
Subject to technical changes.
* Individual mains cable available. Please contact your distributor.

Scope of delivery / accessories
1 defrosted water container
Blank plug for defrosted water container
Supporting arm for Physio tube
Glass plate without handles
Screw Set for glass shelf
Supporting arm for Derma tube
For safety reasons only use original accessories, as proper functioning cannot
otherwise be guaranteed.
Subject to technical changes.

Cryo 6 can be used in combination with various laser devices. Please comply with
the laser manufacturer’s instructions for use when doing so.
The party combining the devices and thus operating a medical system is
independently responsible for combining the devices correctly.
When using devices in combination, please comply with the safety regulations of
DIN EN 60601-1.

Cryo 6 is manufactured according to the DIN EN 60601-1 safety regulations.
As the manufacturer, Zimmer MedizinSysteme GmbH can only consider itself to be
responsible for the safety and reliability of the device if
x the device is operated using a proper power outlet with earth contact and the
electrical installation complies with DIN VDE 0100 part 710,
x the equipment is operated in accordance with the instructions for use,
x extensions, readjustments or modifications are carried out only by persons
authorised by Zimmer MedizinSysteme,
x the user has confirmed the functional safety and the proper operating condition
for him-/herself before use,
x the device is operated only by properly trained personnel,
x the device is not operated in areas at risk of explosion and / or a combustive
atmosphere, and
x the device is immediately disconnected from the mains when penetrated by
liquid.
The device does not contain any parts that can be repaired by the operator.

There is a filter element on the rear of the device (9) to ensure that the treatment air
and the cold air needed for the cooling machine are free from coarse dust particles.
It is normally sufficient to clean this regularly by vacuuming the whole of the outside
with a commercially available domestic vacuum cleaner, no later than when the
software issues a reminder after 200 operating hours. We recommend cleaning the
filter element more frequently in carpeted rooms or in areas where dust formation is
likely to be higher for other reasons.
The software message must be confirmed by pressing the “OK” key after the air filter
has been cleaned.
When Cryo 6 is switched off after use, or when the defrosting program is started, the
cooling system heats up and so produces defrost water.
The tank, which has a capacity of approximately 1 litre, can be removed by pulling it
out of the device and can then be emptied.
It should then be replaced in the device.
Please also follow the instructions for cleaning and disinfection (chapter 11).

After being switched on, Cryo 6 automatically performs a self-test and checks the
function of the technical components.
If necessary, the user can check the function of the cooling technology as described
below:
1. Switch on the device.
2. Wait until the device is ready for use.
This is the case when the device displays the program selection menu /
direct start menu.
3. Press the start/stop key (18) to start up Cryo 6.
4. Select the various airflow levels one after the other and check the strength
of the air stream and the cold air output.

In Germany, no safety checks according to section 6 of the MPBetreibV (Medical Device
Operator Ordinance) are required for the device Cryo 6. The device is not listed in
attachment 1 of the ordinance.
A metrological control (MTK) according to section 11 of the MPBetreibV (Medical device
operator ordinance) is also not required for the device Cryo 6. The device is not listed in
attachment 2 of the ordinance.
These requirements apply to the operation of the device in Germany. Please
consider divergent national regulations in your country.

Cryo 6 is fitted with a bipolar overload protection element integrated into the main
switch (10) to protect the device in the event of supply problems. If the fuse trips, the
device automatically switches off via the toggle switch (10). The device can only be
made ready for use again by switching it on via the toggle switch (10).
Please notify the customer service department if this fault occurs frequently.
output / reduction in
cold air output
Dirt in and around the defrost opening (8) may be the cause of a significant reduction
in cold air output and a weak air stream. Dust particles are deposited in the heat
exchanger by the treatment air. This can eventually lead to blockage of the defrost
opening and a build-up of defrost water.
The defrost opening is located above the defrost water container. The device must
be defrosted before cleaning the defrost opening (see page 18, “Performing
defrosting”). After defrosting, switch off the device and remove the plug from the
socket.
Remove the defrost water container and place a flat container underneath the device
to capture the defrost water. Slowly pull out the defrost opening, using a twisting
motion.
Clean the defrost opening under running water and screw it back into place.
Device components that are important to the running of programs are tested when
the device is switched on and sometimes also when it is in use. If an error is
detected during these tests, treatment is terminated, and an error message is
displayed in the top line of the display and an acoustic signal is produced.
Error messages are shown in plain text in the top line of the display. Treatment
cannot be continued, the device can only be switched off.
Errors that do not affect continued work with the device can be resolved by switching
the device off, waiting for five seconds and then switching it back on.
If the error message relates to excessively high temperature or pressure, you should
wait for 30 minutes before switching the device back on as the device needs time to
cool down. These error messages may be caused by high external temperatures
and room temperatures which impair cold air output.
OK
Start/ Stop
Excess pressure cold circuit

Please notify the customer service department if this fault occurs frequently. You
may get in touch with them via your sales representative or via the main office in
Neu-Ulm.
For other functional problems, contact your service representative.
Zimmer MedizinSysteme GmbH
Junkersstraße 9
D-89231 Neu-Ulm
Tel. +49 731. 9761- 291
Fax +49 731. 9761- 299
The device may only be returned to the factory in its original packaging. It must be
disposed of by the factory in Neu-Ulm.
In foreign (European) countries please refer to national regulations for disposal.
Contact your distributor if necessary.

Manufacturer´s EMC declaration
Medical electrical devices, such as Cryo 6, are subject to special precautionary measures with regard to EMC
(electromagnetic compatibility) and must be installed and put into operation according to the EMC information in
the instructions for use or accompanying documentation.
Portable and mobile HF communication devices (such as mobile phones, cell phones) can affect medical
electrical devices.
Cryo 6 may only be operated with the original power cable indicated in the list of the scope of delivery and
accessories.
Operation of the device with a different power cable can lead to increased emissions or reduced interference
immunity of the device!
Guidelines and manufacturer's declaration – Electromagnetic emissions
The Cryo 6 device is intended to be operated in an electromagnetic environment as indicated below. The customer
or user of the Cryo 6 should ensure that it is operated in such an environment.
Interference emission measurements
Electromagnetic environment - Guideline
HF emissions according to CISPR 11
The Cryo 6 device uses HF energy only for its
internal functioning. This means that its HF
emissions are very low, and it is very unlikely
that adjacent electronic devices would suffer
HF emissions according to CISPR 11
The Cryo 6 device is suitable for use in all
facilities, including those in a residential area,
and in those which are connected directly to
the public grid which also supplies buildings
used for residential purposes.
Harmonic emissions according to IEC
61000-3-2
Voltage fluctuations/flickers according to
IEC 61000-3-3
The device should not be used in the direct vicinity of or stacked with other devices. If operation near or stacked
with other devices is necessary, the device should be observed to check its proper operation in the arrangement
used.

Manufacturer´s EMC declaration
Guidelines and manufacturer's declaration – Electromagnetic immunity
The Cryo 6 device is intended to be operated in the electromagnetic environment as indicated below. The customer or user
of the Cryo 6 device should ensure that it is used in such an environment.
Electromagnetic environment - Guidelines
discharge (ESD)
according to IEC
61000-4-2
Floors should be made of wood or concrete or should be covered
with ceramic tiles. If the floor is covered with synthetic material,
the relative air humidity must be at least
30%.
electrical
disturbances/bursts
according to IEC
61000-4-4
power cables
± 1 kV for input
and output
cables
± 1 kV for input
and output
The quality of the supply voltage should correspond to that of a
typical business or hospital environment.
according to IEC
6100-4-5
differential mode
± 2 kV common
mode
The quality of the supply voltage should correspond to that of a
typical business or hospital environment.
interruptions and
fluctuations in the
supply voltage
according to IEC
61000-4-11.
T
(>95% dip in
the U
T
for ½
cycle)
40% U
T
(60% dip in the
U
T
for 5 cycles)
70% U
T
(30% dip in the
U
T
for 25
cycles)
<5% U
T
(>95% dip in
the U
T
for 5
T
(>95% dip in the
U
T
for ½ cycle)
40% U
T
(60% dip in the
U
T
for 5 cycles)
70% U
T
(30% dip in the
U
T
for 25 cycles)
<5% U
T
(>95% dip in the
U
T
for 5
seconds)
The quality of the supply should correspond to that of a typical
business or hospital environment. If the user of the Cryo 6
requires continued function even if interruptions in the power
supply occur, it is recommended to power Cryo 6 from an
uninterruptible power supply or a battery.
power supply
frequency (50/60
Hz) according to
Magnetic fields at the mains frequency should correspond to the
typical values found in a business or hospital environment.
Note: UTis the a.c. supply voltage prior to application of the test level.

Manufacturer´s EMC declaration
Key performance features of the Cryo 6 are: smooth output of cold air, smooth operation of all functions.
Guidelines and manufacturer's declaration – Electromagnetic immunity
The Cryo 6 device is intended to be operated in the electromagnetic environment as indicated below. The customer or user
of the Cryo 6 device should ensure that it is used in such an environment.
Electromagnetic environment - Guidelines
HF
disturbances
according to
IEC 610004-6
Radiated HF
disturbances
according to
IEC 610004-3
Effective value
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Effective value
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Portable and mobile radio equipment should not be used
at a distance away from the Cryo 6, including the cables,
that is less than the recommended safety distance which is
calculated according to the equation applicable to the
transmission frequency.
Recommended safety distance:
G ¥3
G ¥3IRU0+]WR0+]
G ¥3IRU0+]WR GHz
with P as the rated output of the transmitter in watts (w)
according to the information of the manufacturer of the
transmitter and d as the recommended safety distance in
meters (m).
At all frequencies, the field strength of stationary radio
transmitter should be less than the compliance level
b
according to on-site testing
a
In the environment of devices which bear the following
symbols, interferences are possible:
NOTE 1 At 80 Hz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not be applicable in all cases. The propagation of electromagnetic variables is influenced by
absorption and reflection from buildings, objects and people.

Manufacturer´s EMC declaration
a
The field strength of stationary transmitters, such as base stations for radio telephones and land mobile radios,
amateur radio stations, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment with regard to the stationary transmitters, an electromagnetic site survey is to be
recommended. If the measured field strength in the location in which the Cryo 6 device is used exceeds the above
compliance level, the Cryo 6 device must be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the Cryo 6.
b
Above a frequency range of 150 kHz to 80 MHz, the field strength should be less than 3 V/m.
Recommended safety distances between portable and mobile HF telecommunication devices and the Cryo 6 device
The Cryo 6 device is intended to be operated in an electromagnetic environment in which the HF disturbances are controlled.
The customer or user of the Cryo 6 device can help avoid electromagnetic interference by maintaining a minimum distance
between portable and mobile HF telecommunication devices (transmitters) and the Cryo 6 device – depending on the output
power of the communication device as indicated below.
Rated output of the transmitter
Safety distance depending on the transmission frequency
m
150 kHz to 80 MHz
G ¥3
80 MHz to 800 MHz
G ¥3
800 MHz to 2.5 GHz
G ¥3
For transmitters whose maximum rated output is not listed in the table above, the recommended safety distance d in metres
(m) can be determined using the equation applicable to the respective column, where P is the maximum rated output of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not be applicable in all cases. The propagation of electromagnetic variables is influenced by
absorption and reflection from buildings, objects and people.

Cryo 6
Zimmer MedizinSysteme GmbH
Junkersstrasse 9
D-89231 Neu-Ulm
Tel. +49 731. 97 61-291
Fax +49 731. 97 61-299
export@zimmer.de
www.zimmer.de
Instructions for Use
EN 10 101 665 UR 1216 I Version 4 I Right of modication reserved