INSTALLATION
AND OPERATING
INSTRUCTIONS
SECTION 2.0
A.T.S. 750 TOURNIQUET SYSTEM
2.1 INITIAL INSPECTION
Unpack the A.T.S. 750 Tourniquet upon receipt and
inspect the unit for any obvious damage that may have
occurred during shipment. We recommend that this
inspection be performed by a qualified biomedical
engineer or other person thoroughly familiar with
electronic medical devices. If the unit is damaged, notify
the carrier and your Zimmer representative immediately.
If the initial inspection results are satisfactory, a
functional and calibration check should be performed.
2.2 CONTROLS, INDICATORS, AND
CONNECTORS
Refer to Figure 2.1 for the locations of the unit’s controls,
indicators, and connectors.
1. ON/STANDBY Touch-Switch
Turns the unit on or sets the unit to STANDBY.
This switch will not set the unit to STANDBY
when the cuff pressure is at a non-zero value.
Ensure the cuff is fully defl ated and has been
removed from the patient prior to setting the
unit to STANDBY.
Note: During STANDBY, power to the A.T.S.
750 instrument and all instrument functions
(i.e. infl ation, defl ation, etc) are off but power
continues to supply the battery charging
circuitry anytime ~ (AC) power is present.
2. ROTARY knob
Changes the value of set or default time and
pressure. Turn knob clockwise to increase a value;
turn knob counterclockwise to decrease value.
3. PRESSURE switch
Press to verify or modify set pressure.
4. TIME switch
Press to verify or modify set time.
5. INFLATE switch
Infl ation of the cuff is initiated by depressing the
INFLATE switch.
6. DEFLATE switch
Defl ation of the cuff is initiated by depressing
the DEFLATE switch. For greater safety, the
DEFLATE switch has a delay and, therefore, must
be held for approximately 2 seconds before the unit
will allow a cuff to defl ate.
7. ALARM SILENCE switch
The ALARM SILENCE switch will silence most
audible alarms for 30 seconds after the switch is
depressed. When an alarm sounds because of an
internal circuit malfunction, the alarm cannot be
silenced.
NOTE: The alarm messages will continue to fl ash
on the displays until the alarm condition is
corrected.
8. AC INDICATOR light
The AC INDICATOR light indicates that the unit
is plugged in and is being powered by alternating
current.
9. PRESSURE display
During normal operation with no switches being
depressed, the PRESSURE display will show the
monitored cuff pressure. At other times, depending
on alarm conditions and switches, this display may
communicate other information such as alarm
messages, set pressure, or default set pressure.
10. TIME display
During normal operation with no switches being
depressed, the TIME display will show elapsed
infl ation time of the cuff in 1 minute increments.
At other times, depending on alarm conditions and
switches, this display may communicate other
information such as alarm messages, set time limit,
or default time limit.
11. CUFF connector
The CUFF connector is the port used to connect the
unit to cuff hose.
12. POLE clamp
The POLE clamp is used to mount the unit on an
I.V. pole.
NOTE: Do not hang articles on the tourniquet
pole which are not related to tourniquet use.
5