Zimmer A.T.S. 3000 Operators & Service Manual

Page 1
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Operator
& Service Manual
Zimmer A.T.S.® 3000
AUTOMATIC TOURNIQUET SYSTEM
REF 60-3000-101-00
Page 2
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
SCOPE OF LIMITED WARRANTY
Zimmer, Inc. warrants the Product (A.T.S. 3000 Tourniquet System) for one year from date of purchase. During the warranty period, Zimmer will repair or replace, at its option, any product which is defective in materials or workmanship or which fails to meet the published specification for that model. This Limited Warranty is made only to the original purchaser of the product and is non-transferable. The remedies described in this Limited Warranty are the exclusive remedies for breach of warranty. THIS
WARRANTY SHALL NOT APPLY TO ANY PRODUCT WHICH HAS BEEN ALTERED OR MODIFIED IN ANY WAY, OR WHICH HAS BEEN SUBJECTED TO MISUSE OR ABUSE.
DISCLAIMER OF IMPLIED WARRANTIES
The forgoing of Express Limited Warranty is given in lieu of any and all other express or implied warranties. ZIMMER MAKES NO OTHER WARRANTIES INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
LIMITATION OF REMEDIES
In no case shall Zimmer, Inc. be liable for any special, incidental, or consequential damages whether based on breach of warranty or other legal theory whether or not such damages are foreseeable. Some states do not allow limitations on warranties or on remedies for breach in certain transactions. In such states, the limits in this paragraph and the preceding paragraph do not apply.
WARRANTY CLAIMS
In the event of a warranty claim within the warranty period please take the following steps:
1. Notify Customer Service Department, Zimmer Surgical, at 1-800-348-2759 or contact your local Zimmer representative. Please provide details about the nature of the problem and include the product serial number. Upon receipt of this information, Zimmer will provide a date for service or a return shipping authorization.
2. Upon receipt of the shipping authorization, forward the equipment, freight prepaid, to the location specified in the shipping authorization.
Your compliance with these steps will help assure that you receive prompt warranty service for your product.
WARRANTY (OUTSIDE U.S.A.)
Please contact your local Zimmer Representative for warranty information.
Unit Serial Number _______________________
2
Page 3
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
ENGLISH
TABLE
OF
CONTENTS
ZIMMER A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
SECTION TITLE PAGE
1.0 GENERAL INFORMATION
1.1 Specifications ....................................................................................................................................... 3
1.2 Intended Use ........................................................................................................................................ 6
1.3 Contraindications ................................................................................................................................. 6
1.4 Precautions in Use ............................................................................................................................... 6
1.5 Adverse Effects .................................................................................................................................... 7
2.0 INSTALLATION AND OPERATING INSTRUCTIONS
2.1 Initial Inspection .................................................................................................................................. 8
2.2 Controls, Indicators, and Connectors .................................................................................................. 8
2.3 Initial Setup .......................................................................................................................................... 9
2.4 Functional and Calibration Check ....................................................................................................... 9
2.5 Pressure and Time Defaults ................................................................................................................. 11
2.6 Limb Occlusion Pressure (LOP) Determination ................................................................................. 11
2.7 Single Cuff Operation .......................................................................................................................... 13
2.8 Dual Cuff Operation ............................................................................................................................ 14
2.9 Bier Block Cuff Operation (IVRA) ..................................................................................................... 15
2.10 Speaker Volume Setting ....................................................................................................................... 15
2.11 Contrast Display Setting ...................................................................................................................... 15
2.12 Alarm Conditions ................................................................................................................................. 15
3.0 MAINTENANCE
3.1 General Maintenance Information ....................................................................................................... 24
3.2 Access to Parts ..................................................................................................................................... 24
3.3 Limb Occlusion Pressure (LOP) Sensor Cleaning and Disinfecting .................................................. 24
3.4 Periodic Maintenance .......................................................................................................................... 24
3.5 Calibration ............................................................................................................................................ 25
3.6 Leak Testing ......................................................................................................................................... 26
3.7 Battery Voltage and Battery Service .................................................................................................... 27
3.8 Fuse and Fuse Drawer Replacement ................................................................................................... 27
3.9 Unscheduled Maintenance ................................................................................................................... 28
3.10 Troubleshooting Guide ........................................................................................................................ 28
3.11 Expected Test Point Readings ............................................................................................................. 28
3.12 Replacement Parts ................................................................................................................................ 32
3.13 Storage ................................................................................................................................................. 32
3.14 Warnings, Cautions, and Symbology ................................................................................................... 39
1
Page 4
TABLES
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
2.1 Alarm Conditions ................................................................................................................................. 16
2.2 LOP Alarm Conditions ........................................................................................................................ 18
2.3 Error Codes .......................................................................................................................................... 18
2.4 EMC Guidance and Declaration - EMC Emissions ............................................................................ 20
2.5 EMC Guidance and Declaration - EMC Immunity/Disturbances ....................................................... 20
2.6 EMC Guidance and Declaration - EMC Immunity/RF ....................................................................... 22
2.7 EMC Guidance and Declaration - EMC Immunity/Separation Distances .......................................... 23
3.1 Board Plug Designators ....................................................................................................................... 24
3.2 Troubleshooting ................................................................................................................................... 29
3.3 Expected Test Point Readings ............................................................................................................. 31
3.4 Parts List .............................................................................................................................................. 32
ILLUSTRATIONS
1 A.T.S. 3000 Overview – front ............................................................................................................. 33
2 A.T.S. 3000 Overview – rear ............................................................................................................... 34
3 Main and Second Calibration Kit Setup .............................................................................................. 35
4 Battery Removal Overview ................................................................................................................. 35
5 Rear Power Cord Cover Removal ....................................................................................................... 36
6 Rear Clamp Removal ........................................................................................................................... 36
7 Wiring View of Control Board ............................................................................................................ 36
8 Rear Case Separation ........................................................................................................................... 37
9 Control Board Layout .......................................................................................................................... 38
2
Page 5
GENERAL INFORMATION
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
SECTION 1.0
ZIMMER A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
1.1 SPECIFICATIONS
Mains Line Voltage Range:
100–240 V ~ (AC), 50/60 Hz. Auto switching
Line Current:
670 mA RMS @ 120 V ~ (AC)
Input Power:
53 Watts typical
Battery Type:
Rechargeable, 12 VDC sealed lead acid,
4.0 amp hours
Battery Discharge Time:
Unit will operate on battery power for 240 minutes minimum with a fully charged battery.
Battery Recharge Time:
24 hours Unit should be plugged in 24 hours before initial use. In the event of a deep battery discharge that cannot be recovered in the first 24 hours, a second 24 hour charging period may be required. In this event, the A.T.S. unit should be unplugged for 60 seconds and reconnected to AC power prior to starting the second charging session.
Power Cord:
Type SJT or international equivalent of HO5VV-F,
AWG 16 (1.31mm2), 14 ft. (4.27 m)
Power Plug:
Hospital grade, 3 prong straight blade, 15 amp or proper
equivalent power plug
Line Protection:
2 time delayed 1.0 amp 250 volt fuses
CONTROLS:
PRESSURE Button:
PRESSURE
Used in conjunction with the shuttle knob to adjust the pressure set point. Can also be pressed to verify the set point. The Main Cuff and Second Cuff have separate Pressure buttons.
TIME Button:
TIME
Used in conjunction with the Shuttle Knob to adjust the time alarm set point. Can also be pressed to verify the set point. The Main Cuff and Second Cuff have separate Time buttons.
Cuff INFLATE / DEFLATE Buttons:
INFLATE
DEFLATE
Controls inflation or deflation of the respective cuff. The Main Cuff and Second Cuff have separate DEFLATE
buttons.
INFLATE /
ON/STANDBY Button:
Turns the unit ON / sets unit to STANDBY.
ALARM SILENCE Button:
Allows operator to manually silence most alarms for 30 seconds.
3
Page 6
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Limb Occlusion Pressure (LOP) Button:
Cuff Pressure Range:
LOP
Controls the LOP feature. When the pulse sensor is in place and cuff applied, pressing this button will start the process to measure the patient’s LOP and give the user a recommended tourniquet pressure (RTP).
AC Indicator Light Icon (Green LED):
50–475 mmHg, 1 mmHg increments
Pressure Accuracy:
±3 mmHg (50–475 mmHg)
Pressure Regulation:
±4 mmHg of set point (10 second average under non-transient conditions without external leaks)
Maximum Pressure:
475 mmHg cuffs
Time Alarm Set Range:
5–240 minutes; 1 minute increments
Timer Accuracy:
0.25% of elapsed time
Indicates unit is operating on AC Mains. This is the normal means of operation (battery power is only intended for emergency power loss or patient transport).
Battery Indicator Light Icon (Orange LED):
Indicates unit is operating on backup battery. This indicator always flashes.
Alarm Indicator Light Icon (Red LED):
Visual indicator to show the unit is in an alarm condition.
LOP Heart Indicator Icon (Yellow LED):
Internal Diagnostics:
Program, memory, watchdog timer, transducer calibration, improper valve actuation.
SIZE: Height:
13.0 in. (33.02 cm)
Width:
9.5 in. (24.1 cm)
Depth:
10.375 in. (26.35 cm) (including clamp and ports)
Weight:
16.3 lbs. (7.4
DISPLAYS:
The A.T.S. 3000 uses a backlit 1/4 panel LCD.
Pressure Display: Displays pressure setting, sensed cuff pressure, and hardware failure conditions / other messages.
Time Display: Displays time alarm set point, elapsed time, and hardware failure conditions / other messages.
kg)
Visual indicator to show the LOP function has been invoked.
4
Page 7
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
UL60601-1 Classification:
Type of protection against electric shock: Class I or Internally Powered Equipment*
Degree of protection against electric shock: Type BF applied part
Classification according to the degree of IPX0 protection against ingress of water:
Mode of operation: Continuous operation
*When the unit is operating on backup battery, the type of protection against electric shock changes to internally powered equipment.
This device is not suitable for use in the presence of flammable anesthetic or gases.
Emissions/Immunity:
The A.T.S. 3000 Tourniquet System complies with EMC criteria set forth in EN 60601-1-2. The user of this device should be aware that precautions should be taken in regards to EMC. The device should be installed and used according to the EMC information provided in the instructions for use. See EMC Guidance Tables included in this manual.
Cables
Zimmer LOP sensor cable
Mains power cord
WARNING: use of an LOP sensor cable or mains power cord with a length other than those specified may result in increased emissions and decreased immunity.
Maximum length
98 inches (249 cm)
170 inches (432 cm)
5
Page 8
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
1.2 INTENDED USE
The A.T.S. 3000 Tourniquet System is intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient’s extremities during surgical procedures on those extremities. Tourniquets have been found useful in producing a bloodless operation field in surgical procedures involving the extremities including:
Reduction of certain fractures Kirschner wire removal Tumor and cyst excisions Subcutaneous fasciotomy Nerve injuries Tendon repair Bone grafts Total wrist joint replacement Replacement of joints in the fingers Knee joint replacements Amputations Replantations
WARNING: Do not use tourniquet cuffs to control the distal flow of CO2 or any other gases used as a distention media. Tourniquet cuffs have not been evaluated for safety or effectiveness in controlling gas flow beyond the surgical site during arthroscopic insufflation procedures. Possible effects of using a tourniquet cuff in this manner include serious subcutaneous emphysema proximal to the cuff.
1.3 CONTRAINDICATIONS
The medical literature lists the following as possible contraindications. However, in every case, the final decision whether to use a tourniquet rests with the attending physician.
Open fractures of the leg Post-traumatic lengthy hand reconstruction Severe crushing injuries Elbow surgery (where there is excess swelling) Severe hypertension Skin grafts in which all bleeding points must be
readily distinguished
Compromised vascular circulation, e.g., peripheral
artery disease Diabetes mellitus The presence of sickle cell disease is a relative
contraindication. (See PRECAUTIONS IN USE.)
A tourniquet should also be avoided in patients who are undergoing secondary or delayed procedures after immobilization.
1.4 PRECAUTIONS IN USE
The tourniquet system must be kept well calibrated and in operable condition. Accessories should be checked regularly for leaks and other defects.
The tourniquet cuff must never be punctured; therefore towel clips used near the system must be handled with special care. Cuffs with inner rubber bladders must be completely enclosed by the outer envelope to preclude ballooning and possible rupture of the bladder. Cleaning and assembly instructions of the cuff manufacturer should be followed carefully.
Do not use an elastic bandage for exsanguination in cases where this will cause bacteria, exotoxins, or malignant cells to spread to the general circulation, or where it could dislodge thromboemboli that may have formed in the vessels.
The tourniquet cuff must be applied in the proper location on the limb, for a “safe” period of time, and within an appropriate pressure range. Never apply a tourniquet over the area of the peroneal nerve or over the knee or ankle. Do not readjust an already inflated cuff by rotating it because this produces shearing forces which may damage the underlying tissue.
Prolonged ischemia may lead to temporary or permanent damage to tissues, blood vessels, and nerves. Tourniquet paralysis may result from excessive pressure. Insufficient pressure may result in passive congestion of the limb with possible irreversible functional loss. Prolonged tourniquet time can also produce changes in the coagulability of the blood with increased clotting time.
Inflation should be done rapidly to occlude arteries and veins as near simultaneously as possible.
Careful and complete exsanguination reportedly prolongs pain free tourniquet time and improves the quality of Intravenous Regional Anesthesia, (Bier Block anesthesia). In the presence of infection and painful fractures, after the patient has been in a cast, and in amputations because of malignant tumors, exsanguination before tourniquet application may be done without the use of an elastic bandage by elevating the limb for 3 to 5 minutes.
In case of failure, the tourniquet cuff must be fully deflated and the limb exsanguinated again before reinflation. Reinflation over blood-filled vasculature may lead to intravascular thrombosis.
Tourniquet users must be familiar with the inflation­deflation sequence when using a dual-cuff tourniquet or two tourniquet cuffs together for IVRA (Bier Block anesthesia), so that the wrong tourniquet will not be released accidentally.
Test for hemoglobin type and level before using a tourniquet on patients with sickle-cell anemia. When the tourniquet is used for these patients, the limb should be carefully exsanguinated and the PO closely monitored.
Select the proper cuff size to allow for an overlap of about 3 to 6 in. (7.6–15 cm). Too much overlap may cause cuff rolling and telescoping, and may lead to undesired
and pH should be
2
6
Page 9
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
pressure distribution on the limb. The skin under the tourniquet cuff must be protected from mechanical injury by
smooth, wrinkle-free application of the cuff. If the tourniquet cuff is applied over any material that may shed loose fibers (such as Webril) the fibers may become embedded in the contact closures and reduce their effectiveness. As an under padding, a section of stockinette may be used. The deflated cuff and any underlying
bandage or protective sleeve should be completely removed as soon as tourniquet pressure is released. After the cuff has been fully deflated and removed from the patient, the unit can be set to STANDBY. Even the
slightest impedance of venous return may lead to congestion and pooling of blood in the operative field.
If skin preparations are used preoperatively, they should not be allowed to flow and collect under the cuff where they may cause chemical burns.
Whenever the tourniquet cuff pressure is released, the wound should be protected from blood surging back by applying pressure dressings and, if necessary, elevating the limb. Transient pain upon tourniquet pressure release can be lessened by elevation of the limb. If full color does not return within 3 to 4 minutes after release, the limb should be placed in a position slightly below body level.
Whenever IVRA, Bier Block anesthesia, is used, it is recommended that the tourniquet remain inflated for at least 20 minutes from the time of injection.
WARNING: Cuffs will not deflate in STANDBY mode. Ensure cuffs are fully deflated before setting the unit to STANDBY.
1.5 ADVERSE EFFECTS
A dull aching pain (tourniquet pain) may develop throughout the limb following use. Pathophysiologic changes due to pressure, hypoxia, hypercarbia, and acidosis of the tissues occur and become significant after about 1 1/2 hours of tourniquet use. Symptoms of tourniquet paralysis are motor paralysis and loss of sense of touch, pressure, and proprioceptive responses. Intraoperative bleeding may be caused:
1. By the slight impeding effect exerted by an unpressurized cuff (and its padding, if used), which prevents venous return at the beginning of the operation.
2. By blood remaining in the limb because of insufficient exsanguination.
3. By inadequate tourniquet pressure (between systolic and diastolic blood pressure of the patient), or slow inflation and deflation, all of which allow arterial blood to enter while preventing venous return.
4. By blood entering through the nutrient vessels of the long bones, such as the humerous.
7
Page 10
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
INSTALLATION
AND OPERATING
INSTRUCTIONS
SECTION 2.0
ZIMMER A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
2.1 INITIAL INSPECTION
Unpack the A.T.S. 3000 Tourniquet upon receipt and inspect the unit for any obvious damage that may have occurred during shipment. We recommend that this inspection be performed by a qualified biomedical engineer or other person thoroughly familiar with electronic medical devices. If the unit is damaged, notify the carrier and your Zimmer representative immediately. If the initial inspection results are satisfactory, a functional and calibration check should be performed after a 24-hour charge. The attention label covering the ON/STANDBY button can be removed and discarded after the 24-hour charge.
2.2 CONTROLS, INDICATORS, AND CONNECTORS
Refer to Figure 1 in the back of the manual for the locations of the unit’s controls, indicators, and connectors.
1. ON/STANDBY button
Turns the unit ON or sets the unit to STANDBY. This
button will not set the unit to STANDBY when the cuff pressure is at a non-zero value. Ensure both
cuffs are fully deflated and have been removed from the patient as well as all underlying bandages or protective sleeve prior to setting the unit to STANDBY.
NOTE: During STANDBY, the power to the A.T.S.
3000 instrument and all instrument functions (i.e. inflation, deflation, etc.) are OFF but power continues to supply the battery charging circuitry anytime ~ (AC) power (Mains) is present.
2. Shuttle Knob
Changes the value of set time or default time and set
pressure or default pressure. Turn knob clockwise to increase the value; turn knob counterclockwise to decrease the value.
3. PRESSURE button
Press to verify or modify set pressure.
4. TIME button
Press to verify or modify set time.
5. INFLATE buttons
Inflation of the respective cuff is initiated by depressing
the INFLATE button.
6. DEFLATE buttons
Deflation of the respective cuff is initiated by depressing
the DEFLATE button. For greater safety, the DEFLATE button has a delay and, therefore must be held for approximately 2 seconds before the unit will allow a cuff to deflate.
7. ALARM SILENCE button
The ALARM SILENCE button will silence most audible
alarms for 30 seconds after the button is pressed. When an alarm sounds because of an internal hardware malfunction, the alarm cannot be silenced.
NOTE: The alarm messages will continue to flash on
the displays until the alarm condition is corrected.
8. AC INDICATOR light
The AC INDICATOR light indicates that the unit is
plugged in and is being powered by AC Mains. This is the normal means of operation (battery power is only intended for emergency power loss or patient transport).
9. BATTERY INDICATOR light
The BATTERY INDICATOR light indicates that the
unit is operating on backup battery. The light will flash continuously while the unit is running on battery backup power.
10. PRESSURE displays (independent)
During normal operation with no buttons being pressed,
the independent PRESSURE display areas will show the monitored cuff(s) pressure. At other times, depending on alarm conditions and buttons pressed, the display may communicate other information such as alarm messages, set pressure, or default set pressure.
11. TIME displays (independent)
During normal operation with no buttons being pressed,
the independent TIME display area will show elapsed inflation time of each cuff. At other times, depending on alarm conditions and buttons pressed, the display may communicate other information such as alarm messages, set time, or default set time.
8
Page 11
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
A.T.S. 3000 TOURNIQUET SYSTEM
NOTE: The elapsed inflation time can be “zeroed” at
any point in the procedure by pressing the TIME and PRESSURE buttons for the respective cuff simultaneously.
Cuff connector ports
12.
The
13. Pole Clamp The Pole Clamp is used to mount the unit on an
NOTE: Do not hang articles on the tourniquet pole that
14. Hose hangers
The A.T.S. 3000 is equipped with hose hangers that
NOTE: Do not hang articles on the tourniquet’s hose
2.3 INITIAL SETUP
Inspect to ensure the correct fuse drawer with the appropriately rated fuses is present. The 100–120 V unit uses the gray fuse drawer with 1.0 A time delay fuses. The 220–240 V unit uses the black fuse drawer with 1.0 A time delay fuses. See Section 3.8 Fuse and Fuse Drawer Replacement. The power cord should be plugged into the power entry module on the back of the unit. The unit should
be plugged into ~ (AC) power (Mains) for 24 hours before initial use. During shipping and storage, the unit’s battery could become weak. Always charge 24 hours before any initial use including any calibration checking procedures, initial checks, tests and any institutional performed biomedical evaluations.
2.4 FUNCTIONAL AND CALIBRATION CHECK
The unit shall produce the results explained in the following steps exactly as indicated. Failure to do so indicates that a problem may exist and the device is not to be used until necessary repair or calibration has been made.
Cuff connectors are the ports used to connect the unit to the cuff hoses. Please note that the Main Cuff is the RED ports and the Second Cuff is the BLUE ports. The A.T.S. 3000 is designed and tested for use with Zimmer dual port cuffs. Zimmer does not recommend the use of any cuff other than Zimmer dual port cuffs. Do not use single port cuffs with the A.T.S. 3000.
I.V. pole.
are not related to tourniquet use. For stability reasons, do not use an I.V. pole with a base less than 27.56 inches (70 cm) in diameter.
pull out of the unit’s handle. The cuff hoses can be temporarily hung on the hangers for transport or when disconnecting from the cuff.
hangers that may cause the tourniquet to become unstable. Cuff hoses or the LOP sensor should be the only item to utilize the hose hangers.
1. Connect the power plug of the unit to a properly polarized and grounded power source with voltage and frequency characteristics compatible with the specifications listed in Section 1.1.
Observe that the green AC Mains indicator light turns on.
2. Turn the unit ON by pressing the ON/STANDBY button and observe the following: a) “Zimmer” along with the circle “Z” appears on the
LCD display.
b) The unit displays “SELF TEST” below the
“Zimmer” name. The unit is self-testing specific
system hardware and software. c) Emits tones while “SELF TEST” is displayed. d) The front panel “Alarm”, “LOP”, and “Battery”
indicators flash on and off while “SELF TEST”
is displayed. e) “CAL” is displayed in the PRESSURE display areas
during the calibration check. f) “0” is displayed in the PRESSURE and TIME
display areas after the startup routing is complete.
If a number other than zero is displayed in the
PRESSURE display, the unit should be calibrated.
3. Test the PRESSURE set point system as follows: a) Press either PRESSURE button. b) The PRESSURE display should read “*250” (the
factory default set point) for approximately 2 seconds.
c) Within the 2 second time frame, rotate the
SHUTTLE KNOB to change the pressure set point (clockwise to increase, counter-clockwise to decrease). The set pressure can be maintained between 50 mmHg and 475 mmHg in increments of 1 mmHg.
d) Repeat step 3 for the other PRESSURE set point.
4. Test the TIME set point system as follows: a) Press either TIME button. b) The main TIME display should read “* 60” (the
factory default set point) for 2 seconds.
c) Within the 2 second time frame, rotate the
SHUTTLE KNOB to change the time set point (clockwise to increase, counter-clockwise to decrease). The set time can be maintained between 5 and 240 minutes in increments of 1 minute.
d) Repeat step 4 for the other TIME set point.
NOTE: Anytime an asterisk (*) appears in the left most display digit, the data being displayed is the set point. Set pressure and time will revert back to the default pressure and time settings when the unit is set to STANDBY.
9
Page 12
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
5. Calibration Check
NOTE: During the power-up diagnostic self-test described above, the unit will test the calibration. Should an out of calibration condition be detected, the unit will display either “CAL” “FAIL”, “CAL M” “FAIL” or “CAL 2” “FAIL” in the PRESSURE and TIME display areas. Even though the unit performs this check at every power-up, the following quantitative check is recommended at regular intervals.
a) Verify the unit is in the STANDBY mode. b) Enter the calibration mode by pressing then
releasing the ON/STANDBY button then immediately depress and hold in the Main Cuff INFLATE and Main Cuff DEFLATE buttons during power on self test. The unit will enter calibration mode after momentarily displaying “ZIMMER” “SELF TEST”. When “CALIBRATION” is momentarily displayed, release the Main Cuff INFLATE and Main Cuff DEFLATE buttons. The unit will also display the software revision level in the lower left corner. The software revision level can be recorded for future reference.
NOTE: The calibration is only being checked in this section. For complete calibration, see Maintenance Section 3.0.
c) Connect a calibrated 0 to 700 mmHg pressure meter
(minimum requirement) to the calibration hose assembly. The calibrated meter will be used as the pressure standard (See Figure 3 in the back of the manual).
d) Connect a pressure source capable of supplying 0 to
700 mmHg of pressure, minimum.
e) Insert one end of the calibration hose assembly
connector into the Main Cuff sense port on the unit (red port). The sense port is the second port over from the left side of the unit.
f) Insert the other end of the calibration hose assembly
connector into the Second Cuff sense port on the unit (blue port). The sense port is the fourth port over from the left side of the unit.
NOTE: The unit will be displaying “0” where the cuff pressures are normally displayed and alternating “CAL” and “0” where the cuff times are normally displayed.
g) Increase the pressure in the calibration hose
assembly to 50 mmHg. Both PRESSURE displays should read 50±5 mmHg when compared to the calibrated meter.
h) Increase the pressure to 250 mmHg. Both
PRESSURE displays should read 250±5 mmHg.
i) Increase the pressure to 475 mmHg. Both
PRESSURE displays should read 475±5 mmHg.
j) Decrease the pressure to 0 mmHg and remove
the calibration hose assembly from the unit. The PRESSURE displays should return to 0 mmHg.
k) Press and hold in the Main Cuff INFLATE and
Main Cuff DEFLATE buttons to advance to the reservoir calibration check.
NOTE: At this point the reservoir pressure, if pressurized, will be exhausted through the source ports and the Main Cuff PRESSURE display should go to “0” while the Main Cuff TIME display should continue to display “CAL” “0” as described earlier.
l) Insert one end of the calibration hose assembly
connector into the Main Cuff source port on the unit (red port). The source port is the first port from the left side of the unit.
m) Insert the other end of the calibration hose assembly
connector into the Second Cuff source port on the unit (blue port). The source port is the third port over from the left side of the unit.
n) Increase the pressure in the calibration hose assembly
to 250 mmHg. The display should read 250±5 mmHg.
o) Increase the pressure to 475 mmHg. The display
should read 475±5 mmHg.
p) Increase the pressure to 700 mmHg. The display
should read 700±5 mmHg.
NOTE: If any reading is off by more than ±5 mmHg, the entire unit must be recalibrated by following the calibration procedure as outlined in Maintenance Section 3.0.
q) To complete this procedure, turn the unit OFF by
holding in the ON\STANDBY button until the unit is set to STANDBY.
r) Calibration check is complete.
6. Low Pressure Alarm Check –
Turn the unit ON by pressing the ON/STANDBY button.
Connect a cuff and standard length hose set to the Main Cuff ports. Inflate the cuff to 250 mmHg. Create a leak in the cuff by partially detaching the hose (either port) from the unit while a cuff is inflated. Make the leak large enough that the pressure drops more than 15 mmHg below set point. Observe:
a) A 1.5 second delay is instituted to reduce nuisance alarms. b) The PRESSURE display flashes between “LO-P” and
the monitored pressure (if a substantial leak has been present for more than 9 seconds, the PRESSURE display will show “CUFF” “LEAK”).
c) An audible tone will sound and the red alarm indicator
will illuminate announcing the alarm condition.
10
Page 13
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
d) Stop the leak and observe the monitored pressure
returns to regulated state, the audible tone stops, the red alarm indicator turns off, and the alarm message is no longer displayed.
Repeat this procedure with the Second Cuff ports.
2.5 PRESSURE AND TIME DEFAULTS
To modify the default pressure or time limits for either cuff, follow the following steps.
1. Default Pressure a) The Default Pressure is selected by depressing and
holding either PRESSURE button for 2 seconds. When the default mode is entered, the audible alarm beeps once and a “D” is displayed in the first position on the selected cuff PRESSURE display.
b) The Default Pressure is modified via the
and can be set between 50 and 475 mmHg in
Knob increments of 1 mmHg.
c) After the correct value is selected, it is saved by
momentarily depressing the PRESSURE button or it will be saved automatically in 3 seconds.
d) The new default value will be displayed for
1.5 seconds and the audible alarm will beep once signifying a new default value has been stored.
e) The new default pressure will be stored and remains
the default every time the machine is turned on.
2. Default Time Limit a) The Default Time Limit is selected by pressing and
holding either TIME button for 2 seconds. When the default mode is entered the audible alarm beeps and a “D” is displayed in the first position on the selected cuff TIME display.
b) The Default Time Limit is modified via the Shuttle
Knob and can be set between 5 and 240 minutes in increments of 1 minute.
c) After the correct value is selected, it is saved by
momentarily depressing the TIME button or it will be saved automatically in 3 seconds.
d) The new default value will be displayed for
1.5 seconds and the audible alarm will beep once signifying a new default value has been stored.
e) The new time limit default will be stored and
remains the default every time the machine is turned ON.
NOTE: The elapsed inflation time can be “zeroed” at any point in the procedure by pressing the TIME and PRESSURE buttons simultaneously for each respective cuff time.
Shuttle
2.6 LIMB OCCLUSION PRESSURE (LOP) DETERMINATION
NOTE: Limb Occlusion Pressure (LOP) determination is not
intended for use in pediatric procedures.
The patient’s limb occlusion pressure is the lowest pressure required to stop the flow of blood in the extremity. The A.T.S. 3000 has the ability to estimate the patient’s limb occlusion pressure based on their physiological characteristics. The A.T.S. 3000 will also take into account anticipated changes in blood pressure during the procedure by adding an additional pressure margin to the LOP measurement at the end of the LOP determination. This additional pressure margin added to the LOP measurement is referred to as the Recommended Tourniquet Pressure or RTP. The RTP is calculated using the LOP with the following:
LOP 100–130 mmHg LOP + 50 mmHg = RTP
LOP 131–190 mmHg LOP + 75 mmHg = RTP
LOP 191–300 mmHg LOP + 100 mmHg = RTP
The RTP can be accepted or rejected based on the physician’s discretion. The RTP value is presented at the end of the LOP determination.
When deciding to accept the RTP value or not the physician may take into account other factors such as the patient’s blood pressure, anesthetic technique to be used, expected procedure duration, cuff location, cuff type, cuff width, snugness of cuff application and surgical procedure to be performed. The physician may also choose to use an alternative method such as the Doppler stethoscope to manually determine the patient’s LOP, or to confirm the LOP determined by the A.T.S. 3000. The accuracy of the automatic determination of LOP can be verified manually, by employing a Doppler stethoscope and carefully following the published technique for manual LOP determination.
The A.T.S. 3000 will suggest the RTP as the lowest pressure for the extremity to ensure the field will remain clear even during changes in blood pressure. However large changes in the patient’s blood pressure during surgery may result in reduced visibility in the field. The pressure may need to be adjusted slightly to improve visual quality. The RTP may be overridden at any time simply by changing the pressure set point.
11
Page 14
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
It should be noted that certain physiologic conditions in some patients may prevent the A.T.S. 3000 from making a determination of LOP, in which case the instrument will display an appropriate message and will terminate the attempt to measure LOP. In that event, the physician’s judgment should be used to set tourniquet pressure in the absence of LOP and RTP values. The determination of LOP is intended as additional, supplementary information for the physician responsible for selecting the tourniquet pressure to be employed for a specific patient and procedure. The physician’s best judgment should always be paramount in the selection of tourniquet pressure.
NOTE: The pulse sensor appears very similar to other sensors used for pulse measurements. It should be noted, the A.T.S. 3000’s pulse sensor does not measure oxygen saturation nor can it be modified to do so. The A.T.S. 3000 uses a custom sensor. Non Zimmer sensors will not work with the A.T.S. 3000. Use of a non Zimmer sensor could damage the A.T.S. 3000 or cause unpredictable operation. Never use any sensor other than approved Zimmer pulse sensors.
2.6.1 Single Bladder Cuff LOP Measuring
NOTE: If the LOP determination is performed on the Main
Cuff, all readings and recommendations are for the Main Cuff only. If the LOP determination is performed on the Second Cuff, all readings and recommendations are for the Second Cuff only.
c) Connect the LOP pulse sensor to the LOP socket in the
front of the A.T.S. 3000.
d) Prepare the patient in accordance with your established
procedures and cuff manufacturer’s instructions. The precautions of Section 1 and the following are offered as a guide to assist in this process.
In most cases a tourniquet cuff should be applied to
the widest part of the limb to allow as much tissue as possible to lie between the cuff and any nerves or vascular structures susceptible to damage. The optimum positions are the upper arm and the proximal third of the thigh. In certain cases of fore-foot surgery, the tourniquet cuff can be applied around the calf or to the area proximal to the malleoli. The valve port and hose connections should be placed so that the hose will not be kinked when the limb is positioned for surgery.
e) Attach the LOP pulse sensor to the patient’s index
finger or second toe on which the tourniquet cuff has been applied.
NOTE: LOP determination temporarily inflates then deflates the tourniquet cuff automatically to obtain the patient’s LOP.
a) Press the ON/STANDBY button to turn the unit ON.
The unit will execute a self-check diagnostic test as described in Section 2.4 of this manual. Successful completion of the self-check indicates the unit is ready for use.
CAUTION: If a connected cuff is pressurized to
50 mmHg or more during power-up, the A.T.S. 3000 Tourniquet will declare it an abnormal start-up sequence. It will assume that a surgical procedure is in process, and will adopt the pressure sensed in the cuff as the new set point. It will automatically go into the regulate mode on the cuff. To alert the operator of this condition, the unit will sound a tone and display a “CUFF” “INFL” alarm. The operator should immediately check the pressure set point and readjust to the proper set point if necessary. The alarm will be cleared as soon as the set point is examined, (press the correct pressure button).
b) Connect a dual port cuff to the unit at the Main Cuff or
Second Cuff port connectors. The Main Cuff and Second Cuff both have the ability to perform the LOP function.
NOTE: The LOP Pulse Sensor is applied to a finger or toe on the operative limb.
f) Ensure the sensor is fully engaged to achieve the best
possible and most accurate reading.
g) With the cuff and sensor applied to the patient, press the
corresponding LOP icon to start the LOP determination. The A.T.S. 3000 will begin to inflate the cuff incrementally while continuously analyzing the patient’s pulse. If the sensor or cuff were not properly installed, the unit will display alarm messages. The meanings to the alarm messages are found in Table 2.2.
h) The LOP determination will last approximately
30 seconds depending on the quality of pulse sensed.
i) At the end of the LOP determination, the A.T.S. 3000
will beep and display the LOP and RTP pressures in the lower display area for that cuff. The unit will automatically display the RTP in the cuff pressure display area preceded by a “*”.
NOTE: The RTP is the summation of the pressure margin and the LOP which ensures the field visibility remains clear during the procedure.
j) To accept the RTP and return to normal operation
press the corresponding PRESSURE button. To reject
12
Page 15
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
the RTP value and enter a user defined pressure value, turn the Section 2.4.3.c.
k) To retry the LOP procedure, repeat section 2.6 steps
f thru i.
NOTE: The A.T.S. 3000 will not automatically inflate the cuff at the end of the LOP determinations. It is the user’s responsibility to accept or reject the RTP.
l) If the RTP is acceptable and no other adjustments are
required, pressing the respective INFLATE button will inflate the cuff to the RTP.
m) In the event the quality of visibility is reduced by
an increase in patient blood pressure, the tourniquet pressure can be increased manually. To increase the tourniquet pressure manually, momentarily press the corresponding PRESSURE button and rotate the Knob clockwise to increase the pressure set point.
n) After the LOP has been measured and the unit has
offered an RTP, the unit is ready to be used. However, the user need not perform a LOP measurement to use the tourniquet system.
NOTE: The LOP is only used to obtain patient LOP prior to tourniquet use. Once the patient LOP measurement is complete, remove the LOP sensor from the patient and store.
2.6.2 Dual Bladder Cuff LOP Measuring
NOTE: LOP determination temporarily inflates then deflates
the tourniquet cuff automatically to obtain the patient’s LOP.
LOP measurement for a dual bladder cuff is identical to a single bladder cuff except for the following points:
1. A dual bladder dual port cuff is connected to the unit (Reminder: Main Cuff is the Red ports, Second Cuff is the Blue ports).
2. At the end of the LOP determination using the first bladder, press the corresponding PRESSURE button to accept the RTP for that cuff.
3. Initiate an LOP determination using the second bladder by pressing the corresponding LOP button.
4. At the end of the LOP determination using the second bladder, the RTP will appear in the corresponding pressure display area preceded by a “*”.
5. Compare the RTP currently being displayed with the RTP that was accepted in step 2 above.
NOTE: The RTP accepted in step 2 above is now the pressure set point for that cuff. To verify the RTP, momentarily press the corresponding PRESSURE button. During this activity the RTP in step 4 will temporarily
Shuttle Knob and follow the procedure in
Shuttle
be “cleared” from the display but will return following a 3-second delay.
6. If the RTP from the first LOP determination is higher than the second RTP, turn the Shuttle Knob and follow the procedure in Section 2.4.3.c to adjust the second RTP to equal the first RTP.
7. If the RTP from the first LOP determination is lower than the second RTP, press the corresponding PRESSURE button to accept the RTP from the second LOP determination. Next, follow the procedure in Section 2.4.3.c to adjust the first RTP to equal the second RTP.
2.7 SINGLE CUFF OPERATION
1. Press the ON/STANDBY button to turn the unit ON. The unit will execute a self-check diagnostic test as described in Section 2.4 of this manual. Successful completion of the self-check indicates the unit is ready for use.
CAUTION: If a connected cuff is pressurized to 50 mmHg or more during power-up, the A.T.S. 3000 Tourniquet will declare it an abnormal start-up sequence. It will assume that a surgical procedure is in process, and will adopt the pressure sensed in the cuff as the new set point. It will automatically go into the regulate mode on the cuff. To alert the operator of this condition, the unit will sound a tone and display a “CUFF” “INFL” alarm. The operator should immediately check the pressure set point and readjust to the proper set point if necessary. The alarm will be cleared as soon as the cuff set point is examined, (press the correct PRESSURE button).
2. Connect a dual port cuff to the unit at the Main Cuff connectors (red ports).
3. The default settings for cuff pressure and time limit are retrieved from the nonvolatile memory during power up.
For each patient, tourniquet pressure required to occlude
blood flow to the operative site should be set to the minimum effective pressure. The minimum effective pressure should be determined by factors such as: whether the cuff is to be applied to an upper or lower limb; whether the limb is normal, hypertrophied, or obese; the patient’s preoperative systolic pressure; and the maximum anticipated rise in systolic pressure during the procedure. The A.T.S. 3000 has the unique ability to estimate the minimum effective pressure. This pressure is referred to as the Limb Occlusion Pressure or LOP. Refer back to Section 2.6 for using the LOP measuring feature.
13
Page 16
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
4. Prepare the patient in accordance with your established procedures and cuff manufacturer’s instructions. The precautions of Section 1 and the following are offered as a guide to assist in this process.
In most cases a tourniquet cuff should be applied to
the widest part of the limb to allow as much tissue as possible to lie between the cuff and any nerves or vascular structures susceptible to damage. The optimum positions are the upper arm and the proximal third of the thigh. In certain cases of fore-foot surgery, the tourniquet cuff can be applied around the calf or to the area proximal to the malleoli. For emergency surgery of the hand, a sufficiently small tourniquet can be fitted around the wrist.
Apply a leak-free tourniquet cuff smoothly without
wrinkles. The valve port and hose connections should be placed so that the hose will not be kinked when the limb is positioned for surgery. The viability of the skin and deeper tissues should be established prior to exsanguination of the limb and tourniquet inflation. Exsanguinate the limb by elevating it for a minimum of 2 minutes and wrapping it, distal to proximal, using an Esmarch, Martin, or elastic bandage. The bandage should come up approximately to 1 in. (2.5 cm) from the edge of the tourniquet cuff. The elastic bandage is removed following inflation of the cuff. If regional anesthesia is being used, the anesthetic agent or nerve block is then administered. The tourniquet time depends greatly on the patient’s anatomy, age, and absence of vascular disease. The surgeon will determine:
1) When the tourniquet is to be inflated;
2) What pressure is applied;
3) How long the tourniquet is applied;
4) Whether to allow for intermittent aeration of tissue
by deflating the cuff for 10 to 15 minutes;
5) To what point in the operation the tourniquet should
be released.
In many operating rooms, it is customary to prominently
note the time of inflation, and to warn the surgeon after a certain time has elapsed. This will allow the surgeon to assess the need for further tourniquet time.
There is a general agreement that, for reasonably healthy
adults, 2 hours should not be exceeded without releasing the tourniquet to allow the underlying tissue to breathe. During this time, the limb should be elevated to about 60 degrees, and steady pressure should be applied to the incision with sterile dressings.
5. The cuff is inflated by pressing the red Main Cuff INFLATE button. The unit will pressurize the Main Cuff to the set pressure and start the elapsed inflation
time alarm clock. The Main Cuff inflation information will be displayed on the LCD screen. If the unit cannot pressurize the cuff to within 15 mmHg of the set point in less than 30 seconds, a leak alarm will be sounded. See Section 2.12 for information about possible alarm conditions. Once the cuff is inflated, the time display will track elapsed inflation time.
6. At the end of the procedure, deflate the cuff by pressing the Main Cuff DEFLATE button for minimum of 2 seconds. The Main Cuff PRESSURE display will show the deflation of the cuff and the elapsed inflation time alarm clock will stop.
NOTE: The elapsed inflation time can be “zeroed” at any point by pressing the TIME and PRESSURE buttons simultaneously.
7. Remove the tourniquet cuff and any underlying
bandages or protective sleeve immediately following final deflation. The time of tourniquet cuff removal
should be noted, and the circulation of the limb should be checked.
8. After the cuff has been removed, disconnect the cuff from the A.T.S. 3000.
9. During normal use, the A.T.S. 3000 should not be set to STANDBY if pressure is present in either cuff. Once the cuff has been properly deflated, removed from the patient, and disconnected from the A.T.S. 3000, the unit can be set to STANDBY.
2.8 DUAL CUFF OPERATION
Operation of the unit is identical to Single Cuff operation (see Section 2.7) except for the following points:
1. Two dual port cuffs are connected to the unit (Reminder: Main Cuff is the Red ports, Second Cuff is the Blue ports).
2. The Main and Second LCD screen section will display inflation information and begin timing the cuff inflation.
3. Deflation of one cuff will not be permitted while the other cuff is inflating.
4. When inflating a second cuff with the other cuff already inflated, the unit will continuously check the original cuff to ensure that the pressure is within allowable limits. The unit will stop its inflation and maintain the original cuff to within 10 mmHg of the set point before returning to the inflating cuff. This ensures that at least one cuff maintains occlusion at all times. If there is a significant leak in the original cuff, this feature could cause the inflation rate of the subsequent cuff to be longer and perhaps even cause the 30-second inflation alarm to sound.
14
Page 17
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
5. When both cuffs are inflated, the LCD screen displays independent information for each cuff. That is, the PRESSURE and TIME displays are independently operated and controlled for each cuff.
6. In order to deflate the final cuff, a sequence must be
followed to prevent accidental deflation: a) Press and hold the DEFLATE button on the cuff to
be deflated;
b) When the “CUFF” “DEFL” alarm is active,
release the DEFLATE button;
c) Within 5 seconds of the alarm discontinuing, press
the DEFLATE button once again;
d) The cuff will deflate. This safety feature is
particularly useful when using the unit for Bier Block Cuff Operation, (IVRA).
2.9 BIER BLOCK CUFF OPERATION (IVRA)
Review Sections 2.7 and 2.8, SINGLE CUFF OPERATION and DUAL CUFF OPERATION.
1. The following are suggested cuff connections: a. The proximal cuff connected to the Red Main Cuff
ports using the white/red cuff tubing;
b. The distal cuff connected to the Blue Second Cuff
ports using the white/blue cuff tubing.
2. Follow the cuff inflation sequence adopted by your institution or requested by the surgeon.
3. Deflation of a cuff is not possible while the other is inflating.
4. When requested, the first can be deflated simply by pressing and holding the DEFLATE button for a minimum of 2 seconds.
5. In order to deflate the final cuff, a sequence must be
followed to prevent accidental deflation: a) Press and hold the DEFLATE button on the cuff to
be deflated;
b) When the “CUFF” “DEFL” alarm is active,
release the DEFLATE button;
c) Within 5 seconds of the alarm discontinuing, press
the DEFLATE button once again;
d) The cuff will deflate.
6. For Bier Block procedures follow the cuff inflation/ deflation sequence adopted by your institution or requested by the surgeon.
2.10 SPEAKER VOLUME SETTING
An operator may want to change the speaker volume setting from the default. The following steps will allow a user to customize the speaker volume default setting.
1. Press the ON/STANDBY button to turn the unit ON. The unit will execute a self-check diagnostic test as described in Section 2.4.
2. Press and hold both the Main Cuff TIME and Main Cuff PRESSURE buttons for approximately 3 seconds.
3. When the volume selection screen appears, the user can adjust the volume up or down by turning the
. Clockwise rotation increases volume while a
Knob counterclockwise rotation decreases volume. With every increment or decrement the speaker will sound to give the user feedback of the volume level setting selected. After 5 seconds of inactivity the system will accept the volume setting as displayed and return the display back to its normal mode of operation.
4. The user can accept the volume setting by pressing the Main Cuff PRESSURE button which stores the setting and returns the display to its normal mode. Or, wait approximately 5 seconds as noted above to allow the system to accept the new volume setting.
2.11 CONTRAST DISPLAY SETTING
An operator may want to change the display contrast setting from the default. The following steps will allow a user to customize the display contrast default setting.
1. Press the ON/STANDBY button to turn the unit ON. The unit will execute a self-check diagnostic test as described in Section 2.4.
2. Press and hold both the Second Cuff TIME and Second Cuff PRESSURE buttons for approximately 3 seconds.
3. When the contrast control setting selection screen appears, the user can adjust the display contrast by turning the Shuttle Knob. Clockwise rotation increases display contrast while a counterclockwise rotation decreases display contrast. With every increment or decrement the display will show the new contrast setting selected. After approximately 5 seconds of inactivity the system will accept the contrast control setting as displayed and return the display back to its normal mode of operation.
4. The user can accept the new setting by pressing the Second Cuff PRESSURE button which stores the setting and returns the display to its normal mode. Or, wait approximately 5 seconds as noted above to allow the system to accept the setting.
2.12 ALARM CONDITIONS
There are a number of conditions for which the A.T.S. 3000 Tourniquet will produce a visual and/or audible alarm. Those conditions, indications, and appropriate actions are shown in Table 2.1. The appropriate actions indicated are based on the most probable causes and should only be used as a guide. Other causes of alarm conditions may indicate a need for other actions. In addition to the conditions shown in Table 2.1, it is conceivable that a malfunction could occur for which the indications are unintelligible and unpredictable. In this
Shuttle
15
Page 18
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
situation, it is likely that the valves will be disabled causing the system to hold cuff pressure. It is also likely that a tone will sound under these conditions. Most audible alarm tones may be silenced for 30 seconds by pressing the ALARM SILENCE button. The tone will be re-enabled at the end of the silenced period. Pressing the ALARM SILENCE button will cause the alarm tone to be silenced again. The A.T.S. 3000 Tourniquet will also provide Error Code information for critical alarms as shown in Table 2.3. To minimize nuisance alarms (i.e. “HI-P”, “LO-P”) that can be caused by vigorous movement of the patient’s limbs, a
1.5-second delay has been designed into the alarm actuation. Under certain conditions, such as when a FAIL indication appears in the TIME display or the information that appears in the TIME and PRESSURE displays is unintelligible, the operator should conclude that a hardware failure has occurred, rendering the unit unusable. The appropriate action is to set the unit to STANDBY by pressing the ON/ STANDBY button. Since this removes power from the
internal instrument circuitry, all instrument functions, commands to the valves and pump will cease. This
Table 2.1 Alarm Condition
CONDITION PRESSURE
DISPLAY
will cause the cuff to hold pressure (in the absence of leaks). Clamp the cuff lines with hemostats and replace the
tourniquet unit.
2.12.1 PRESSURE ALARMS
A pressure alarm will occur when the pressure in a cuff is more than 15 mmHg from the pressure set point. It is also possible for a cuff to have a leak that is substantial but which the unit can compensate for by continual pumping. This type of leak could be due to a pin hole in a cuff bladder, or a loose pneumatic fitting. This type of leak could progress into a total failure of a cuff to hold pressure. To alert the operator that a substantial leak is present, a pressure alarm is declared when this type of leak is continuously present for more than 9 seconds. If a pressure alarm occurs, and the displayed pressure is not more than 15 mmHg from the set point, then this type of substantial leak has been detected and all cuffs and pneumatic fittings should be checked for leaks.
TIME
DISPLAY
APPROPRIATE ACTION/REMARKS
CUFF PRESSURE LOW The pressure in the cuff is 15 mmHg below set point
CUFF PRESSURE HIGH The pressure in the cuff is 15 mmHg above set point
CUFF SIDE LEAK A leak has been present for at least 9 seconds.
RESERVOIR LEAK A leak is present between the pump and valves
INFLATION TIME IN EXCESS OF SETTING The cuff has been inflated beyond the set time limit
CUFF INFLATION ON POWER UP Cuff pressurized to 50 mmHg or greater at power up
CUFF NOT DEFLATED Pressure in deflated cuff is a non­zero value
LO-P normal This condition is generally caused by a leak in
the system, or a hose occlusion. All lines and connections should be checked.
HI-P normal Normally caused by transient conditions such as
patient movement, controller overshoot, or hose occlusion. This condition, for an extended period,
would indicate a hardware failure and the A.T.S. 3000 unit should be replaced.
CUFF LEAK normal A substantial leak has been present for more
the 9 seconds. All lines and connections should be checked.
RES LEAK normal Do not use the unit. Service the unit.
normal TIME UP Surgeon should be warned of time up condition.
Only on the direction of the surgeon, time should be set to new value.
CUFF INFL The system assumes that a procedure is in progress
and adopts the sensed pressure as the new set point. The operator should immediately check the set value to determine if it needs reset.
normal CUFF NOT
DEFL
Check for kinks in hose. If alarm persists, disconnect hose from cuff. If attempting to set the unit to STANDBY, ensure that cuff is fully deflated.
16
Page 19
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 2.1 Alarm Condition
CONDITION PRESSURE
LINE OCCLUSION An occlusion is present in the cuff tubing
LOW BATTERY VOLTAGE Low battery voltage
CHECK CUFF An increase in cuff pressure has not been sensed within 1 second after depressing the inflate key
BATTERY FAILURE Battery voltage is too low to ensure proper operation
CALIBRATION OUT OF SPEC The transducer calibration is out of specification
CALIBRATION OUT OF SPEC A transducer calibration is out of specification
AMPLIFIER FAILURE Amplifier is out of range
MATH FAILURE Result of math operation was out of range
ROM FAILURE Microprocessor failed a ROM memory check
RAM FAILURE Microprocessor failed a RAM memory check
VALVE FAILURE Improper valve combination occurred
WATCHDOG FAILURE Windowing watchdog system detected a malfunction
SYSTEM FAILURE SYSTEM FAIL Cycle the ON/STANDBY Button. If problem
Continued
TIME
DISPLAY
LINE OCCL normal Check for hose kinks or other defects.
normal BATT LOW
CHK CUFF Check to see that a cuff and hose assembly is
BATTERY FAIL Plug unit in and cycle the ON/STANDBY Button.
CAL M or CAL 2
CAL FAIL Indicates general calibration fail. Calibrate the unit.
AMP FAIL Cycle the ON/STANDBY Button. If problem persists,
MATH
VA LV E
DISPLAY
PLUG IN
FAIL
FAIL
ROM
FAIL
RAM
FAIL
FAIL
WDT
FAIL
APPROPRIATE ACTION/REMARKS
Plug unit in. If the unit is not plugged in, a battery failure condition will occur and the unit will shut down in a fail safe mode closing all valves. While running
with a Low Battery Voltage Alarm Condition other alarm conditions can not be guaranteed.
connected to the proper ports on the unit.
CAL M (CALIBRATION MAIN) or CAL 2 (CALIBRATION SECOND) indicates a cuff
transducer circuitry is out of calibration. Pressure in error by at least 6 mmHg will cause these failures. Calibrate the unit.
service the unit.
Cycle the ON/STANDBY Button. If problem persists, service the unit.
Cycle the ON/STANDBY Button. If problem persists, service the unit.
Cycle the ON/STANDBY Button. If problem persists, service the unit.
Cycle the ON/STANDBY Button. If problem persists, service the unit.
Cycle the ON/STANDBY Button. If problem persists, service the unit.
persists, service the unit.
17
Page 20
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 2.2 LOP Alarm Condition
CONDITION PRESSURE
DISPLAY
No LOP pulse sensor detected during LOP
Message displayed at bottom
of display area.
“CONNECT LOP SENSOR”
LOP pulse sensor not properly secured to patient normal
normal
normal TIME OUT
Cuff is leaking or not connected to unit or connected to incorrect
normal CUFF INFL
port during LOP function
Patient LOP is too high for LOP measurement normal HIGH LOP
LOP pulse sensor attached to incorrect limb normal
LOP procedure stopped
normal STOP
TIME
DISPLAY
CHECK
SENSOR
LOW
SIGNAL
CHECK
SENSOR
APPROPRIATE ACTION/REMARKS
Plug LOP pulse sensor in. Reattach LOP pulse sensor and retry. Replace LOP pulse sensor with known good working sensor and retry. If problem persists, service the unit.
Check LOP pulse sensor and retry. Replace LOP pulse sensor with known good working sensor and retry. If problem persists, service the unit.
Check LOP pulse sensor and retry. Replace LOP pulse sensor with known good working sensor and retry. If problem persists, service the unit.
Check LOP pulse sensor and retry. Replace LOP pulse sensor with known good working sensor and retry. If problem persists, service the unit.
Check cuff for leaks and retry. Connect cuff to correct port and retry. Connect cuff to unit and retry. If problem persists, service the unit.
Check LOP pulse sensor and retry. Replace LOP pulse sensor with known good working sensor and retry. Do not use the LOP function, follow normal tourniquet procedures.
Attach LOP pulse sensor to correct limb and retry. Replace LOP pulse sensor with known good working sensor and retry. If problem persists, service the unit.
A front panel button may have been pressed while in the LOP function. Wait for the message to clear and retry. If problem persists, service the unit.
NOTE: 1 Some Error Codes are followed by a four-digit numeric code. This code represents detailed
Table 2.3 Error Codes
information related to the failure. Contact the manufacturer for information concerning those specific codes.
ERROR CODE CRITICAL FAILURE EXPLANATION/APPROPRIATE ACTIONS
A timing error was detected during self-test. Cycle the ON/
E001 XXXX
1
SYSTEM FAIL
STANDBY button. If problem persists, note error code and contact manufacturer.
E002 XXXX
1
SYSTEM FAIL
A Safety Monitor signal error was detected. Cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
A pneumatic system error was detected during self-test. Cycle the
E003 SYSTEM FAIL
ON/STANDBY button. If problem persists, note error code and contact manufacturer.
A Safety Monitor timing error was detected. Cycle the ON/
E004 WDT FAIL
STANDBY button. If problem persists, note error code and contact manufacturer.
The Safety Monitor detected an illegal valve state. Cycle the ON/
E005 VALVE FAIL
STANDBY button. If problem persists, note error code and contact manufacturer.
18
Page 21
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 2.3 Error Codes
ERROR CODE CRITICAL FAILURE EXPLANATION/APPROPRIATE ACTIONS
E006 SYSTEM FAIL
E007
E008 SYSTEM FAIL
E010 ROM FAIL
E011 ROM FAIL
E012 RAM FAIL
E013 BATTERY FAIL
E014 BATTERY FAIL
E015 CAL FAIL
E016 CAL FAIL
E017 CAL FAIL
E018 CAL M FAIL
E019 CAL 2 FAIL
E020 XXXX
E021 XXXX
E022 XXXX
E023 XXXX
E024 XXXX
E025 XXXX
E026 XXXX
E027 XXXX
1
1
1
1
1
1
1
1
Continued
MATH FAIL
SYSTEM FAIL
AMP FAIL
AMP FAIL
AMP FAIL
AMP FAIL
AMP FAIL
AMP FAIL
SYSTEM FAIL
A Safety Monitor detection error was detected. Cycle the ON/ STANDBY button. If problem persists, note error code and contact manufacturer.
A processor error was detected. Cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
The processor received an unknown interrupt. Cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
A non-volatile memory value was detected during power-up. Cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
ROM check error detected a self-test. Cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
A microprocessor RAM error was detected during self-test. Cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
Battery voltage below minimum threshold. Plug unit in and cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
A battery charging circuit failure was detected at power-up. Verify a battery is connected and cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
Incorrect zero pressure calibration point was detected. Calibrate unit. If problem persists, note error code and contact manufacturer.
Incorrect calibration point was detected. Calibrate unit. If problem persists, note error code and contact manufacturer.
Calibration error was detected. Calibrate unit. If problem persists, note error code and contact manufacturer.
Main cuff transducer calibration error was detected. Calibrate unit. If problem persists, note error code and contact manufacturer.
Second cuff transducer calibration error was detected. Calibrate unit. If problem persists, note error code and contact manufacturer.
A software exception error was detected. Cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
A transducer reference voltage is out of range. Calibrate unit. If problem persists, note error code and contact manufacturer.
Reservoir transducer is out of range. Calibrate unit. If problem persists, note error code and contact manufacturer.
The Main cuff transducer is out of range. Calibrate unit. If problem persists, note error code and contact manufacturer.
The Second cuff transducer is out of range. Calibrate unit. If problem persists, note error code and contact manufacturer.
A voltage level is out of range. Cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
A voltage level is out of range. Cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
A speaker or alarm tone generation circuitry failure was detected. Verify that the speaker is connected and cycle the ON/STANDBY button. If problem persists, note error code and contact manufacturer.
19
Page 22
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 2.4 EMC Guidance and Declaration - EMC Emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The A.T.S. 3000 is intended for use in the electromagnetic environments specified below. The customer or the user of the A.T.S. 3000 should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
RF emissions
Group 1
The A.T.S. 3000 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ Flicker emissions
IEC 61000-3-3
Table 2.5 EMC Guidance and Declaration - EMC Immunity/Disturbances
Guidance and manufacturer’s declaration – electromagnetic immunity
The A.T.S. 3000 is intended for use in the electromagnetic environments specified below. The customer or the user of the A.T.S. 3000 should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Class A
Class A
Complies
Compliance level Electromagnetic environment – guidance
± 6 kV contact (230 Vac 50 Hz / battery)
± 8 kV air (230 Vac/ battery)
± 2 kV for power supply lines (120 Vac 60 Hz / 230 Vac 50 Hz)
± 1 kV for input/output lines
± 1 kV line(s) to line(s) (120 Vac 60 Hz / 230 Vac 50 Hz)
± 2 kV line(s) to earth (120 Vac 60 Hz / 230 Vac 50 Hz)
The A.T.S. 3000 is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic
purposes.
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic materials, the relative humidity should be at least 30%
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
In the event of a power surge, the A.T.S. 3000’s LCD screen may malfunction, but all other functions should continue to operate. Power down the device to reset. Consult the Operator and Service Manual for additional specifications.
20
Page 23
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 2.5 EMC Guidance and Declaration - EMC Immunity/Disturbances Continued
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz) magnetic field
<5% U
T
(>95% dip in UT) For 0.5 cycle
40% U
T
(60% dip in UT)
70% U
T
(30% dip in UT)
<5% U
T
(>95% dip in UT) for 5 s
3 A/m
<5% U
T
(>95% dip in UT) For 0.5 cycle (120 Vac 60 Hz / 230 Vac 50 Hz)
40% U
T
(60% dip in UT) (120 Vac 60 Hz / 230 Vac 50 Hz)
70% U
T
(30% dip in UT) (120 Vac 60 Hz / 230 Vac 50 Hz)
<5% U
T
(>95% dip in UT) for 5 s (120 Vac 60 Hz / 230 Vac 50 Hz)
3 A/m
Main power quality should be that of a typical commercial or hospital environment. If the user of the A.T.S. 3000 requires continued operation during power mains interruptions, it is recommended that the A.T.S. 3000 be powered from an uninterruptible power supply or a battery.
The A.T.S. 3000 is equipped with a backup battery in the event of power mains interruptions. Consult the Operator and Service Manual for additional specifications.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
IEC 61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
21
Page 24
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 2.6 EMC Guidance and Declaration - EMC Emissions/RF
Guidance and manufacturer’s declaration – electromagnetic immunity
The A.T.S. 3000 is intended for use in the electromagnetic environments specified below. The customer or the user of the A.T.S. 3000 should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance level Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used no closer to any part of the A.T.S. 3000, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to 2.5 GHz
3V
3 V/m
Recommended separation distance
d = 1.2√P
d = 1.2P 80 MHz to 800 MHz
d = 2.3P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance levels in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
b
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephone and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the A.T.S. 3000 is used exceeds the application RF compliance levels above, the A.T.S. 3000 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the A.T.S. 3000.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
22
Page 25
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 2.7 EMC Guidance and Declaration - EMC Immunity/Separation Distances
Recommended separation distances between portable and mobile RF communication equipment and the A.T.S. 3000
The A.T.S. 3000 is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the A.T.S. 3000 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the A.T.S. 3000 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
150 kHz to 80 MHz
Separation distance according to frequency transmitter
m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
d = 1.2√P
d = 2.3√P
23
Page 26
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
MAINTENANCE
SECTION 3.0
ZIMMER A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
3.1 GENERAL MAINTENANCE INFORMATION
While the A.T.S. 3000 Tourniquet has been designed and manufactured to high industry standards, it is recommended that regular inspection and calibration be performed to ensure continual safe and effective operation. This section contains information to assist in the effort as well as serve as a guide to expediting unscheduled maintenance.
3.2 ACCESS TO PARTS
CAUTION: Be sure that the unit is set to STANDBY and the power plug is unplugged before disassembly. To remove mains power from unit, disconnect power cord from wall or rear of unit. Many of the parts on the control board are
static sensitive. Take precaution when servicing the board.
To gain access to all internal parts, remove: a) Rear – 4 (#6) nuts b) Rear – 2 large pole clamp screws c) Rear – 2 carrying handle screws d) Bottom – 4 foot pad screws
See illustration 4 through 8 in the back of the manual.
When opening, take care not to damage any of the wire harnesses or pneumatic tubing. The control board is attached to the front housing therefore the harnesses and tubing will need to be disconnected for full disassembly. Follow the table below to reassemble.
Table 3.1 Board Plug Designators
Component Board Plug ID
ALARM SILENCE P1
AC Mains P2
LCD Panel Display P5
Valve Harness Second P6
Valve Harness Main P7
Membrane Panel P8
LOP Harness P9
Backup Battery P10
ON/STANDBY P11
Factory Test Port P12
Speaker P13
Pneumatic Pump P14
To reduce the risk of damage, tubing should not be disconnected at the transducer.
NOTE: Failure to plug the electrical or pneumatic components into the correct associated receptacle can result in damage to the control board.
When reassembling the unit, be extremely careful not to pinch any wiring or tubing.
3.3 LIMB OCCLUSION PRESSURE (LOP) SENSOR CLEANING AND DISINFECTING
Clean or disinfect the sensor before attaching to a new patient.
1. Cleaning
Unplug the LOP sensor from the A.T.S. 3000 Tourniquet
before cleaning. Clean the sensor and patient contact surfaces with a soft cloth moistened in water or a mild soap solution.
2. Disinfecting
Unplug the LOP sensor from the A.T.S. 3000 Tourniquet
before disinfecting. Disinfect the sensor by wiping the sensor and patient contact surfaces with disinfecting solution. Isopropyl alcohol is recommended as a disinfecting solution. If another commercially available disinfectant is used, follow the manufacturer’s recommendation for use.
3.4 PERIODIC MAINTENANCE
Test and inspect as per this section at minimum every six months.
1. Cleaning
The exterior of the unit may be cleaned with a cloth that
has been dampened (not dripping) with a mild detergent. The interior of the unit may be vacuumed or blown out as required. The exterior of the cuff hose may be cleaned using a mild detergent solution or alcohol. The interior of the cuff hoses should not be cleaned. Tourniquet cuffs should be cleaned in accordance with their cuff package insert instructions.
2. Inspection
The unit should be inspected at regular intervals. It
is recommended that a qualified technician perform a visual inspection at least once every six months. Inspection points are: a) Obvious internal or external damage. b) Condition of the power cord. c) Tightness of pneumatic fittings.
24
24
Page 27
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
d) Condition of internal tubing. e) Accumulation of dust or dirt within the unit. f) Mating integrity of internal connectors. g) Integrity of the pump. h) Security of circuit board. i) Security of the membrane panel.
3. Functional and Calibration Checks
It is recommended that the functional and calibration
checks described in Section 2.4 are performed at least once every three months.
3.5 CALIBRATION
Calibration should be performed every six months, or after any unscheduled maintenance.
Calibration of the A.T.S. 3000 Tourniquet allows the output signal from the pressure transducers to be compared against a calibrated pressure source. The difference between the known pressure and the pressure measured by the transducers is recorded at each of four set points for cuffs (0 mmHg, 50 mmHg, 250 mmHg, and 475 mmHg) and 4 set points for the reservoir (0 mmHg, 250 mmHg, 475 mmHg, and 700 mmHg). These calibration points are used to correct the signal from the pressure transducers during normal operation. The calibration points and a checksum are stored in non­volatile memory.
EQUIPMENT REQUIRED:
A.T.S. 3000 calibration hose (supplied). Calibrated 0 to 700 mmHg pressure meter (minimum requirement). Adjustable 0 to 700 mmHg pressure source (minimum required pressure source).
CAUTION: The following steps must be taken in the exact order to calibrate the unit. Failure to do so may result in incorrect pressure readings while the unit is in operation.
CUFF AND RESERVOIR CALIBRATION Below is a step-by-step procedure for calibrating both cuff transducers as well as the reservoir transducer. The calibration procedure will not be complete until both cuff transducers are calibrated as well as the reservoir transducer.
1. To enter the calibration mode, press and hold the Main Cuff INFLATE and Main Cuff DEFLATE buttons while powering the unit ON. When the unit displays “CALIBRATION” release the Main Cuff INFLATE and Main Cuff DEFLATE buttons. Calibration mode is indicated by alternating “CAL” and “0” in the TIME display windows and indicates the unit is now ready to be calibrated.
2. The unit will display “0” in the PRESSURE displays and alternating “CAL” and “0” in the TIME displays. Throughout this procedure, the TIME display(s) will indicate the pressure in which the user is calibrating.
3. For zero, allow the port to be open to atmospheric pressure so the unit can sense the zero point (i.e. when setting the zero point, nothing should be connected to the cuff ports). Press the Main PRESSURE button, the unit will calibrate the zero pressure. The unit will beep to let the user know the set point was taken.
4. Connect the calibration hose, calibrated pressure meter and adjustable pressure source to the Main and Second Cuff sense ports. See Figure 3 for more details. a. The Main Cuff sense port is the second port over
from the left side of the unit.
b. The Second Cuff sense port is the fourth port over
from the left side of the unit.
5. Once the zero point is calibrated, press the Main Cuff INFLATE button to advance the unit to the next pressure level. The unit’s TIME displays will now be alternating between “CAL” and “50”. Apply 50 mmHg of pressure. Once the pressure has stabilized, press the Main Cuff PRESSURE button so the unit can calibrate the 50 mmHg point. The unit will beep to let the user know the set point was taken.
6. Once the 50 mmHg point is calibrated, press the Main Cuff INFLATE button to advance the unit to the next pressure level. The unit’s TIME displays will now be alternating between “CAL” and “250”. Increase the pressure to 250 mmHg. Once the pressure has stabilized, press the Main Cuff PRESSURE button so the unit can calibrate the 250 mmHg point. The unit will beep to let the user know the set point was taken.
7. Once the 250 mmHg point is calibrated, press the Main Cuff INFLATE button to advance the unit to the next pressure level. The unit’s TIME displays will now be alternating between “CAL” and “475”. Increase the pressure to 475 mmHg. Once the pressure has stabilized, press the Main Cuff PRESSURE button so the unit can calibrate the 475 mmHg point. The unit will beep to let the user know the set point was taken.
8. After the 475 mmHg set point has been taken, press the Main Cuff INFLATE and Main Cuff DEFLATE buttons to advance to the reservoir transducer calibration step. When in the reservoir calibration process the Second Cuff display area goes blank and the Main Cuff display area switches to “RESERVIOR”. Calibration mode is indicated by alternating “CAL” and “0” in the lower display area and indicates the unit is now ready to calibrate the reservoir transducer.
25
Page 28
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
NOTE: At this point the reservoir pressure, if pressurized, will be exhausted through the source ports and the lower portion of the display should continue to display “CAL” “0” as described earlier.
9. For zero, allow the port to exhaust to atmospheric pressure so the unit can sense the zero point (i.e. when setting the zero point nothing should be connected to any cuff port). Press the Main Cuff PRESSURE button, the unit will calibrate the reservoir zero pressure, the unit will beep to let the user know the set point was taken.
10. Connect the calibration hose, calibrated pressure meter and adjustable pressure source to the Main and Second Cuff source ports. See Figure 3 for more details. a. The Main Cuff (red port) source port on the unit is the
first port from the left side of the unit.
b. The Second Cuff (blue port) source port on the unit
is the third port over from the left side of the unit.
11. Press the Main Cuff INFLATE button to advance the unit to the next reservoir pressure level. The display will now be alternating between “CAL” and “250”. Apply 250 mmHg of pressure. Once the pressure has stabilized, press the Main Cuff PRESSURE button so the unit can calibrate the 250 mmHg point. The unit will beep to let the user know the set point was taken.
12. Once the 250 mmHg point is calibrated, press the Main Cuff INFLATE button to advance the unit to the next reservoir pressure level. The display will now be alternating between “CAL” and “475”. Apply 475 mmHg of pressure. Once the pressure has stabilized, press the Main Cuff PRESSURE button so the unit can calibrate the 475 mmHg point. The unit will beep to let the user know the set point was taken.
13. Press the Main Cuff INFLATE button to advance the unit to the next reservoir pressure level. The display will now be alternating between “CAL” and “700”. Apply 700 mmHg of pressure. Once the pressure has stabilized, press the Main Cuff PRESSURE button so the unit can calibrate the 700 mmHg point. The unit will beep to let the user know the set point was taken.
14. At this point, press the Main Cuff INFLATE and Main Cuff DEFLATE buttons. The unit will display “CALIBRATION” “COMPLETE”. The message tells the user the calibration set points have been calibrated to the unit and have now been saved into the non-volatile memory.
NOTE: If the “CALIBRATION” “COMPLETE” message is not displayed, the calibration is incomplete and the adjustments will not be saved. Be certain to end the calibration session by simultaneously pressing the Main Cuff INLFATE and Main Cuff DEFLATE buttons and verify the “CALIBRATION” “COMPLETE” message is displayed.
If any pressure setting is off by more than 15 mmHg or adjusted incorrectly, a “CAL” “FAIL” alarm will be generated and service or calibration to the user’s pressure meter or pressure source is recommended.
If the pressure signal from the internal transducer requires more than a 15 mmHg correction to equal the applied pressure, a “CAL” “FAIL” alarm will also be generated. Service to the unit is recommended.
15. The unit remains in calibration mode until it is set to STANDBY.
16. The stored calibration factors are retrieved from the non­volatile memory during the power-up sequence. If the checksum is invalid, a “CAL” “FAIL” alarm is generated in the displays. The alarm will persist until the unit is set to STANDBY. Re-calibration is required if this occurs.
17. It is recommended to check the calibration by following the steps in Section 2.4 Step 5 “Calibration Check” before using this unit on a patient.
3.6 LEAK TESTING
The A.T.S. 3000 Tourniquet is capable of keeping a cuff with a substantial leak inflated. Naturally it is desirable to keep plumbing leaks to an absolute minimum. For this reason, a check for significant leakage is recommended at regular intervals as well as following any service procedure. After verifying the operation of the A.T.S. 3000 Tourniquet per Section 2.4, connect a 24 in. (61 cm) (or larger) cuff which is known to be leak free to the Main of A.T.S. 3000 Tourniquet System. Adjust the Main Cuff set point to 475 mmHg. Ensure that all external connections are tight. Inflate the Main Cuff and allow the pressure to stabilize. At this point, the unit must be set to STANDBY. Under normal use, the unit cannot be set to STANDBY with a non-zero pressure value displayed in either cuff. However for leak testing purposes, a bypass feature has been incorporated. Press the ON/STANDBY button until the alarm message “CUFF” “NOT” “DEFL” appears. Release the ON/STANDBY button and within 5 seconds of the alarm discontinuing, press and hold the ON/STANDBY button again. The button must be held in for an additional 10 seconds before the unit will be set to STANDBY.
Cuff ports (Red)
26
Page 29
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
NOTE: During the 10 seconds, the alarm message will be displayed, the alarm will continue to sound and the ALARM SILENCE button will not silence the alarm.
Once the unit is set to STANDBY, wait for approximately 10 minutes and turn the unit back on. Operation will resume under cuff inflated start-up conditions (See Section 2.7 Part 1 for explanation). Cancel the alarm using the ALARM SILENCE button. Display the set point by activating the PRESSURE button and view the current (new) pressure set point. The set point is always displayed with an asterisk in the far left position. The current set point for the cuff should be at least 400 mmHg or more. Values less than this indicate an unacceptable leak rate and the source of the leak should be traced and corrected. The first connection to check should be the connections of the cuff. Different cuffs and/or cuff hoses may be tried to determine if the leak is internal or external of the unit.
Repeat the test for the Second cuff ports.
3.7 BATTERY VOLTAGE AND BATTERY SERVICE
NOTE: This section assumes that the unit has been
charged for at least 24 hours. The unit’s bottom battery compartment must be removed to measure battery voltage. See Section 3.2 “Access to Parts” and be sure to follow cautionary statements.
1. Battery Voltage Check Be sure the unit is unplugged. Measure the battery
voltage. The battery voltage should not be lower than 12 volts while the unit is unplugged and set to STANDBY. If, after 1 minute, the voltage reads less than 12 volts, the integrity of the battery should be suspect and should be replaced.
2. Battery Service The 12-volt sealed lead acid battery is charged using
lead acid charging technology. The charging circuit is active anytime the unit is plugged into an acceptable AC Mains outlet. The charger automatically sequences through several charge states based on the battery voltage and charging current conditions. Based on a charger test, the best charge mode is selected. No maintenance is required of the battery charging circuit.
The life of the battery depends on the type of service
and the storage method. Battery replacement will need to be more frequent with continued cycles of deep discharge and/or storage in a high temperature environment. Infrequent short-term use of the battery and storage in a room temperature environment will result in maximum life. It is recommended that the
battery in the A.T.S. 3000 Tourniquet System be replaced
annually. As a reminder, the A.T.S. 3000 System should be plugged in 24 hours before initial use.
NOTICE – BATTERY DISPOSAL
THE BATTERY IS OF A LEAD ACID TYPE.
BATTERY MUST BE RECYCLED OR DISPOSED OF PROPERLY.
3.8 FUSE AND FUSE DRAWER REPLACEMENT Depending on region, A.T.S. systems are factory installed with two fuses in one of these types:
1A, 250V Time Delay Fuse ¼ inch x 1 ¼ inch
1A, 250V Time Delay Fuse 5 x 20 mm
NOTE: When using the ¼ inch x 1 ¼ inch fuses, the GRAY fuse drawer must be used and the BLACK fuse drawer will not be needed. When using the 5 x 20 mm fuses, the BLACK fuse drawer must be used and the GRAY fuse drawer will not be needed.
Fuse drawer holding clip
Fuse Drawer (removed from device)
To remove and install a fuse drawer:
1. Locate the power entry module at the rear of the device where the power cord attaches.
2. If still attached, remove the power cord to ensure no power is being supplied to the device.
3. Using a tool such as a slotted small screwdriver, gently pry the holding clip of the fuse drawer to release the latch. The fuse drawer and fuses will release and can be removed.
4. Install the proper fuses in the appropriate fuse drawer – the ¼ inch x 1 ¼ inch fuses fit the GRAY fuse drawer; the 5 x 20 mm fuses fit the BLACK fuse drawer.
27
Page 30
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
5. Push the fuse drawer into the power entry module until the fuse drawer locks into place (an audible click will be heard).
6. Reconnect power cord to power entry module, plug in to AC and verify device AC power is active.
3.9 UNSCHEDULED MAINTENANCE
The A.T.S. 3000 Tourniquet is designed with several specific self-test features to assist in fault isolation. These features are designed to show messages in the Pressure and Time displays. The meanings of these messages are delineated in Tables 2.1, 2.2 and 2.3. Another mode of failure that may occur is when an audible alarm occurs that cannot be silenced by the ALARM SILENCE button. The valves and pump will be disabled which seals off the cuff to prevent pressure loss. The displays may show random characters. Should this occur, the watchdog timer circuit of the safety processor has detected a problem. The microprocessor may not be executing reliable instructions and is not able to display the correct failure message. The unit should be serviced if this occurs. The calibration error message “CAL” “FAIL”, “CAL M” “FAIL” or “CAL 2” “FAIL” may be due to defective circuitry or may simply indicate the need for calibration.
3.10 TROUBLE SHOOTING GUIDE
To aid in unscheduled maintenance, Table 3.2 delineates a number of possible malfunctions that could occur with the unit. The most likely causes are shown for each symptom. While it is not practical to enumerate every conceivable malfunction and all possible causes, the table will assist in isolating the most common problems.
3.11 EXPECTED TEST POINT READINGS
To expedite unscheduled maintenance, Table 3.3, Expected Test Point Readings, has been incorporated into this manual. This table, as well as Table 3.2, Troubleshooting, should give a qualified technician a good starting point from which to locate and repair most problems that could occur during the life of the unit. Unless noted, all measurements are to be made at room temperature with the cuffs disconnected, and the unit plugged in. All voltage measurements are with respect to ground and are to be made with the unit ON.
28
Page 31
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 3.2 Troubleshooting
SYMPTOM POSSIBLE CAUSES
1. Main Cuff or Second Cuff will not inflate.
2. Main Cuff or Second Cuff will not deflate.
a) Membrane Panel not properly plugged into P8. b) Tubing inside unit may be pinched or
improperly connected. c) Deflate valve is stuck open. d) Pump not properly plugged into P14. e) Pump’s electrical harness damaged. f) INFLATE button not working. g) Valve’s electrical harness damaged. h) Defective valve driver circuitry.
a) Membrane Panel not properly plugged into P8. b) DEFLATE button not pressed long enough
(at least 2 seconds). c) Deflate valve is stuck shut. d) DEFLATE button not working. e) Valve’s electrical harness damaged. f) Defective valve driver circuitry.
3. No green AC Indicator light.
4. No flashing orange Battery Indicator light.
5. ALARM SILENCE button not working.
6. Alarm indicator light not working.
a) Unit not plugged into wall outlet. b) No c) Mains AC harness not properly plugged into P2. d) Blown fuse(s). e) Membrane Panel not properly plugged into P8. f) Defective AC indicator. g) Defective AC indicator circuitry.
a) Unit running on AC. b) Membrane Panel not properly plugged into P8. c) Defective battery indicator. d) Defective battery indicator circuitry.
a) ALARM SILENCE button not properly plugged
b) Non-silenceable alarm (System Failure). c) ALARM SILENCE button defective. d) Defective alarm silence circuitry.
a) Membrane b) Defective alarm indicator. c) Defective alarm indicator circuitry.
power at wall outlet.
into P1.
Panel not properly plugged into P8.
29
Page 32
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 3.2 Troubleshooting
SYMPTOM POSSIBLE CAUSES
a) Transducer amplifier not working.
7. No cuff pressure reading.
8. Pump will not stop running.
9. Battery Fail alarm/message.
10. Backup battery not charging.
b) Internal tubing kinked. c) Transducer tubing on incorrect transducer.
a) Leak in internal hose or connector. b) Internal tubing kinked. c) Transducer(s) not working. d) Transducer tubing on incorrect transducer.
a) Blown battery fuse (board mounted F1). b) Broken battery wire harness. c) Dead or depleted battery.
a) Blown battery fuse (board mounted F1). b) Battery not properly plugged into P10. c) Unit not plugged into wall outlet (verify the
green AC indicator is illuminated). d) Mains AC harness not properly plugged into P2
(verify the green AC indicator is illuminated). e) Unit was not permitted to charge for at least
24 hours. f) Defective battery. g) Defective battery charging circuitry.
11. AMP FAIL alarm.
12. Unit cannot be set to STANDBY.
13. Unit does not turn ON.
a) Transducer(s) amplifier out of range. b) Battery fully depleted or defective. c) Extremely high pressure exerted on transducers.
a) ON/STANDBY
into P11. b) Pressure sensed in the Main or Second Cuff
(unit will alarm “CUFF” “NOT” “DEFL”). c) ON/STANDBY not fully pressed. d) ON/STANDBY button defective.
a) Membrane Panel not properly plugged into P8. b) ON/STANDBY button defective. c) Blown fuse(s). d) Unit not plugged in and battery fully depleted.
30
membrane not properly plugged
Page 33
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 3.3 Expected Test point Readings
Board Location Nominal-Reading Tolerance Description / Comments
TP1 5.0 Vdc
0.0 Vdc
TP2 26.9 Vdc
TP3 N/A N/A Not used for any measurement
TP4 DC Common +50 mV DC common supply
TP5 5.0 Vdc
TP6 13.6 Vdc
TP7 14.0 Vdc +1.0 Bulk system voltage
TP8 N/A N/A Development programmable pins only
TP9 N/A N/A Development programmable pins only
TP10 5.0 Vdc +0.2 Nominal voltage under normal use
TP11 4.1 Vdc +0.2 ON / STANDBY switch logic
TP12 5.0 Vdc +0.3 5 Volt DC regulator supply
TP13 Removed Removed Not used
TP14 5 Vdc 5 Vdc Transducer reference voltage supply
TP15 Removed Removed Not used
TP16 13.6 Vdc Nominal system voltage
TP17 4.095 Vdc
+0.2 5 Volt DC supply for battery charge circuit
0 Vdc while on battery operation
+1.5 Battery DC bulk charging voltage
12 Vdc nominal while on battery operation
Voltage will vary while on battery operation and condition
of battery
(high voltage DC common only)
+0.2
+1.0 Nominal system voltage
Voltage will vary while on battery operation and condition
Voltage goes to 0 Vdc under system reset condition
Voltage will vary while on battery operation and condition
+50 mV Precision reference voltage supply
5 Volt DC switch mode supply
of battery
0 Vdc while on battery operation
of battery
NOTE: USE TP4 (DC COMMON) AS THE VOLTAGE REFERENCE FOR ALL MEASUREMENTS LISTED ABOVE.
CAUTION: HIGH VOLTAGE ELECTRICAL HAZARD. HIGH VOLTAGE WILL BE
PRESENT ON THE POWER INPUT MODULE AND CONTROL BOARD. ALL SERVICE WORK MUST BE COMPLETED BY QUALIFIED TECHNICIANS.
Note: If the unit is locked in a failure, not all test voltages will be valid. Above voltages are listed for normal readings because failure readings will likely be unpredictable.
31
31
Page 34
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
3.12 REPLACEMENT PARTS
The following is a list of field replacement parts that can be ordered from Zimmer. To obtain part or additional information regarding your unit, write or phone:
MAIL: Zimmer Surgical 200 West Ohio Avenue Dover, Ohio 44622 U.S.A.
PHONE: 1-330-343-8801 or 1-800-830-0970
You can also contact your Zimmer distributor. To ensure prompt service, please include the following information with your order:
Model Number
Serial Number
Description of Part
Part Number (if known)
Quantity Desired
Shipping Address
Shipping Means (if any)
We strongly recommend that all repairs be done by Zimmer staff.
Parts marked with an “*” are commonly used maintenance and/or preventative maintenance parts.
Table 3.4 Parts List
Replacement Part
Number 06001304883 1A Time Delay Fuse, 5 x 20 mm * 60200011000 Calibration Hose Assembly * 60236000100 Pole Clamp Knob 60300010500 LOP Sensor Kit * 60300070200 Battery Kit * 60300071100 Pole Clamp Kit 60460000100 Power Cord Kit – Central Europe / Korea 60460000400 Power Cord Kit – Denmark 60460000500 Power Cord Kit – Italy 61210000200 Cord/Pole Clamp Cover 62113700100 Fuse Drawer, 1/4 x 1-1/4 in. 62113800100 Fuse Drawer, 5 x 20 mm 62117900100 1A Time Delay Fuse, 1/4 x 1-1/4 in. * 62210001900 Battery Plate 62210002000 Battery Clamp 62210009600 Operator & Service Manual (English) 622100096DE Operator & Service Manual (German)
Description
Replacement Part
Number 622100096DK Operator & Service Manual (Danish) 622100096ES Operator & Service Manual (Spanish) 622100096FI Operator & Service Manual (Finnish) 622100096FR Operator & Service Manual (French) 622100096GR Operator & Service Manual (Greek) 622100096IT Operator & Service Manual (Italian) 622100096NL Operator & Service Manual (Dutch) 622100096PT Operator & Service Manual (Portuguese) 622100096SE Operator & Service Manual (Swedish) 622100096TR Operator & Service Manual (Turkish) 62460000200 Power Cord – UK / Hong Kong 62460000300 Power Cord – Switzerland 62460000600 Power Cord – Australia 62460000700 Power Cord – Japan / Taiwan 62460000800 Power Cord – India / South Africa 62800001000 Power Cord – North America *
Description
3.13 STORAGE
The A.T.S. 3000 Tourniquet System has an operating range of 50˚F to 100˚F (10˚C to 38˚C). The following are environmental conditions for transportation and storage:
A. Ambient temperature range ..............................................................1˚F to 149˚F (–17˚C to 65˚C).
B. Relative humidity range ...................................................................10% to 80%
C. Atmospheric pressure range .............................................................500 hPA to 1060 hPA.
32
32
Page 35
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
1
19
23
21
25
1
2
24
3
4
5
22
6
7
20
8
18
17
16
15
1. Carrying Handle
2. ALARM SILENCE Button
3. Display Window
4. Second Cuff TIME Button
5. Second Cuff PRESSURE Button
6. Second Cuff LOP Button
7. Green AC MAINS Indicator
8. ON/STANDBY Button
9. Second Cuff DEFLATE Button
10. Second Cuff INFLATE Button
11. LOP Sensor Holder
12. Orange Backup Battery Indicator
13. Second Cuff Ports
12
9
10
11
13
14
FRONT CASE VIEW
14. LOP Port Socket
15. Main Cuff Ports
16. Quick Reference Cards
17. Main Cuff DEFLATE Button
18. Main Cuff INFLATE Button
19. Red Alarm Indicator
20. Yellow LOP Indicator
21. Main Cuff LOP Button
22. Main Cuff TIME Button
23. Main Cuff PRESSURE Button
24. Shuttle Knob
25. Pull Out Hose Hanger – One on each side of unit
33
33
Page 36
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
2
7
6
3
1
8
1. Power Cord
2. Cord / Pole & Clamp Cover
3. Potential Equalization Conductor Stud
4. Factory Test Port
5
REAR CASE VIEW
5. Pole Clamp
6. Mains Fuse Block
7. Rear Case and Handle Screws
8. Bottom Feet Case Screws
34
34
4
2
Page 37
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
3
3
TO MAIN CUFF PORT
TO SECOND CUFF PORT
1
Main and Second Calibration Kit Setup
1. Calibrated Pressure Meter with a minimum range of 0 to 700 mmHg.
2. Pressure Regulator / Source adjustable from 0 to 700 mmHg minimum.
3. Calibration Hose Kit included with unit.
2
4a
Removing the Battery – Easiest if unit is mounted
to stand.
1) Remove two (2) screws from Battery Plate.
4b
Removing the Battery Continued
2) Carefully slide the bottom Battery Plate to the left and away from front of unit. Battery will drop out.
Caution: battery is heavy
35
35
Page 38
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
4c
Removing the Battery Continued
3) Disconnect battery at the cord plug connection.
4) Replace battery by removing Battery Clamp.
5) Make note of wire color to battery polarity.
6) Reverse the process to reassemble.
5
Removing the Rear Power Cord Clamp Cover
1) Remove the 4 #6 nuts holding the Cord / Pole Clamp Cover.
6
Removing the Pole Clamp
1) Remove the two (2) Pole Clamp screws from the rear.
7
Disconnecting wiring from control board.
1) Use caution when disconnecting wiring from printed circuit board headers.
2) Mark wires before disconnecting and see Section 3.2 Table 3.1 for proper re-connect of Board Plug Designators.
36
36
Page 39
8
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Rear case separation
1) All components are easily accessed when the rear case is removed.
2) Remove the Cord / Pole Clamp Cover and Pole Clamp before proceeding. See illustrations 5 and 6.
3) Remove the power cord from the back of the unit.
4) Remove the two screws from the rear bottom feet.
5) Remove the rear handle screws, one on each side of the handle.
6) Slide the rear cover away from the front cover.
7) Remove mufflers from inside of rear case.
8) Reverse the process to reassemble.
Be extremely careful not to pinch any wiring or tubing when reassembling!
37
37
Page 40
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
13
10
9
12
15
14
1
2
11
16
8
7 9
Control Board Layout
1. LCD Panel Header (P5)
2. Main Cuff Valve Header (P7)
3. Factory Test Port Header (P12)
4. Speaker Header (P13)
5. LOP Sensor Socket Header (P9)
6. Shuttle Knob Retaining Washer/Screw (not shown)
7. Front Membrane Header (P8)
8. Battery Header (P10)
3
4
5
6
9. Battery Fuse (F1)
10. Second Cuff Valve Header (P6)
11. ON / STANDBY Header (P11)
12. ALARM SILENCE Header (P1)
13. Not Used - Development Ports Only (P4)
14. AC Power Input Header (P2)
15. Power Supply Area
16. Pneumatic Pump (P14)
38
38
Page 41
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
3.14 Warnings, Cautions, and Symbology
YYYY-MM
M
h f
This product contains a Lead Acid Battery that must be recycled.
Do not dispose of battery in trash.
UL / C-UL Classification mark Medical Equipment with respect to electric shock, fire, and mechanical hazards only, in accordance with UL 60601-1, CAN/CSA C22.2 No. 601.1, and IEC 60601-1
Declaration that a product meets the EMC regulations per marking requirements.
39
39
P
Contains or presence of Phthalates.
Replace fuse as marked
CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician.
This product contains electrical or electronic materials. The presence of these materials may, if not disposed of properly, have potential adverse affects on the environment. Presence of this label on the product means it must not be disposed of in normal household waste and must be disposed of separately. To find out how to properly dispose of this product, please contact your local Zimmer Representative.
This product contains one or more toxic or hazardous substances or elements. The Environmental Protection Use Period on the logo refers to the period in years which toxic or hazardous substances or elements contained in the product will not leak or mutate under normal operating conditions.
d
Page 42
Page 43
Page 44
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
M
Revision: 12-10 © 2005, 2010 Zimmer Surgical, Inc. Printed in U.S.A. 62210009600 ENG
Zimmer Surgical, Inc. 200 West Ohio Avenue Dover, Ohio 44622 U.S.A.
P
0086
Zimmer U.K. LTD. 9 Lancaster Place South Marston Park Swindon, Wiltshire SN3 4FP United Kingdom
Loading...