Zimmer A.T.S.1200 User manual

Page 1
Operator
& Service Manual
A.T.S.
1200
TOURNIQUET SYSTEM
REF 60-1200-101-00
Page 2
LIMITED ONE YEAR WARRANTY (U.S.A.)
Zimmer, Inc. warrants the Product (A.T.S. 1200 Tourniquet System) for one year from date of purchase. During the warranty period, Zimmer will repair or replace, at its option, any product which is defective in materials or workmanship or which fails to meet the published specification for that model. This Limited Warranty is made only to the original purchaser of the product and is non-transferable. The remedies described in this Limited Warranty are the exclusive remedies for breach of warranty. THIS
WARRANTY SHALL NOT APPLY TO ANY PRODUCT WHICH HAS BEEN ALTERED OR MODIFIED IN ANY WAY, OR WHICH HAS BEEN SUBJECTED TO MISUSE OR ABUSE.
DISCLAIMER OF IMPLIED WARRANTIES
The forgoing of Express Limited Warranty is given in lieu of any and all other express or implied warranties. ZIMMER MAKES NO OTHER WARRANTIES INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
LIMITATION OF REMEDIES
In no case shall Zimmer, Inc. be liable for any special, incidental, or consequential damages whether based on breach of warranty or other legal theory whether or not such damages are foreseeable. Some states do not allow limitations on warranties or on remedies for breach in certain transactions. In such states, the limits in this paragraph and the preceding paragraph do not apply.
WARRANTY CLAIMS
In the event of a warranty claim within the warranty period please take the following steps:
1. Notify Customer Service Department, Zimmer Orthopaedic Surgical Products, at 1-800-348-2759 or contact your local Zimmer representative. Please provide details about the nature of the problem and include the product serial number. Upon receipt of this information, Zimmer will provide a date for service or a return shipping authorization.
2. Upon receipt of the shipping authorization, forward the equipment, freight prepaid, to the location specified in the shipping authorization.
Your compliance with these steps will help assure that you receive prompt warranty service for your product.
WARRANTY (OUTSIDE U.S.A.)
Please contact your local Zimmer Representative for warranty information.
Unit Serial Number _______________________
2
Page 3
TABLE
OF
CONTENTS
A.T.S. 1200 TOURNIQUET SYSTEM
SECTION TITLE PAGE
1.0 GENERAL INFORMATION
1.1 Specifications............................................................................................................. 3
1.2 Intended Use .............................................................................................................. 5
1.3 Contraindications ....................................................................................................... 5
1.4 Precautions in Use ..................................................................................................... 5
1.5 Adverse Effects.......................................................................................................... 6
2.0 INSTALLATION AND OPERATING INSTRUCTIONS
2.1 Initial Inspection ........................................................................................................ 7
2.2 Controls, Indicators, and Connectors ........................................................................ 7
2.3 Initial Setup................................................................................................................8
2.4 Functional and Calibration Check ............................................................................. 8
2.5 Pressure and Time Defaults....................................................................................... 9
2.6 Single Cuff Operation................................................................................................ 10
2.7 Dual Cuff Operation .................................................................................................. 11
2.8 Bier Block Cuff Operation (IVRA)........................................................................... 11
2.9 Alarm Conditions....................................................................................................... 11
3.0 MAINTENANCE
3.1 General Maintenance Information............................................................................. 15
3.2 Access to Parts........................................................................................................... 15
3.3 Periodic Maintenance ................................................................................................ 15
3.4 Calibration..................................................................................................................15
3.5 Leak Testing............................................................................................................... 17
3.6 Battery Voltage and Battery Service.......................................................................... 17
3.7 Unscheduled Maintenance......................................................................................... 18
3.8 Troubleshooting Guide .............................................................................................. 18
3.9 Expected Test Point Readings ................................................................................... 18
3.10 Replacement Parts...................................................................................................... 22
3.11 Storage ....................................................................................................................... 22
3.12 Warnings, Cautions, and Symbol Definitions ........................................................... 29
ENGLISH
1
Page 4
TABLES
2.1 Alarm Conditions....................................................................................................... 13
3.1 Board Plug Designators ............................................................................................. 15
3.2 Troubleshooting ......................................................................................................... 19
3.3 Expected Test Point Readings ................................................................................... 21
3.4 Parts List .................................................................................................................... 22
ILLUSTRATIONS
1 A.T.S. 1200 Tourniquet Overview – front ................................................................ 23
2 A.T.S. 1200 Tourniquet Overview – rear ................................................................. 24
3 Calibration Setup Overview ...................................................................................... 25
4 Disassembly – front case opening ............................................................................. 26
5 Disassembly – front case removal............................................................................. 26
6 Disassembly – disconnecting wiring ......................................................................... 26
7 Disassembly – rear case removal .............................................................................. 26
8 Disassembly – overview ............................................................................................ 27
9 Electronic circuit layout............................................................................................. 28
10 Labels......................................................................................................................... 30
2
Page 5
3
GENERAL INFORMATION
SECTION 1.0
A.T.S. 1200 TOURNIQUET SYSTEM
1.1 SPECIFICATIONS
Mains Line Voltage Range:
100–240 V ~ (AC), 50/60 Hz. Auto switching
Line Current:
670 mA RMS @ 120 V ~ (AC)
Input Power:
53 W typical
Battery Type:
Rechargeable, 12 VDC sealed lead acid,
2.3 A hours
Battery Discharge Time:
Unit will operate on battery power for 45 minutes minimum with a fully charged battery.
Battery Recharge Time:
24 hours Unit should be plugged in 24 hours before initial use. In the event of a deep battery discharge that cannot be recovered in the first 24 hours, a second 24 hour charging period may be required. In this event, the A.T.S. unit should be unplugged for 60 seconds and reconnected to AC power prior to starting the second charging session.
Power Cord:
Type SJT, AWG 16, 14 ft. (4.27 m)
Power Plug:
Hospital grade, 3 prong straight blade, 15 A
Line Protection:
2 time delayed 1.0 A 250 V fuses
CONTROLS: ON/STANDBY Button:
Turns the unit on/sets unit to STANDBY.
PRESSURE Button:
Used in conjunction with the ROTARY knob to adjust the pressure set point. Can also be pressed to verify the set point.
TIME Button:
Used in conjunction with the ROTARY knob to adjust the time alarm set point. Can also be pressed to verify the set point.
MAIN CUFF INFLATE/DEFLATE Buttons:
Controls inflation or deflation of the Main cuff. Orange LED bar indicates inflation of the main cuff or pressure alarm in the main cuff when flashing in conjunction with the PRESSURE display.
SECOND CUFF INFLATE/DEFLATE Buttons:
Controls inflation or deflation of the Second cuff. Orange LED indicator bar indicates inflation of the second cuff or pressure alarm in the second cuff when flashing in conjunction with the PRESSURE display.
ALARM SILENCE Button:
Allows operator to manually silence most alarms for 30 seconds.
AC MAINS Indicator Light (Green LED):
Indicates unit is operating on AC Mains. This is the normal means of operation (battery power is only intended for emergency power loss or patient transport).
Page 6
4
BATTERY Indicator Light (Orange LED):
Indicates unit is operating on backup battery. This indicator always flashes.
Cuff Pressure Range:
50-475 mm Hg, 5 mm Hg increments
Pressure Accuracy:
±5 mm Hg (50–475 mm Hg)
Pressure Regulation:
±6 mm Hg of set point (10 second average under non-transient conditions without external leaks)
Maximum Pressure:
475 mm Hg (Normal Operation)
Time Alarm Set Range:
5–240 minutes; 5 minute increments
Timer Accuracy:
0.25 % of elapsed time
Internal Diagnostics:
Program, memory, watchdog timer, transducer calibration, improper valve actuation.
SIZE:
Height:
12.50 in. (31.75 cm)
Width:
10.25 in. (26 cm)
Depth:
8.1 in. (20.6 cm) (including clamp and ports)
Weight:
11.2 lbs. (5.1 Kg)
DISPLAYS:
PRESSURE Display:
Red 14-segment light emitting diodes (LED). Displays pressure setting, sensed cuff pressure, and hardware failure conditions/other messages.
TIME Display:
Red 14-segment light emitting diodes (LED) Displays time alarm set point, elapsed time, and hardware failure conditions/other messages.
UL 60601-1 Classification:
*When the unit is operating on backup battery, the type of protection against electric shock changes to internally powered equipment.
This device is not suitable for use in the presence of flammable anesthetic or gases.
Emissions/Immunity
:
The A.T.S. 1200 Tourniquet System complies with EMC criteria set forth in EN 60601-1-2.
Mode of operation: Continuous operation
Classification according to the degree of IPX0 protection against ingress of water:
Degree of protection against electric shock: Type BF applied part
Type of protection against electric shock: Class I or Internally Powered Equipment*
A.T.S. 1200 TOURNIQUET SYSTEM
Page 7
1.2 INTENDED USE
The A.T.S. 1200 Tourniquet System is intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient’s extremities during surgical procedures on those extremities. Tourniquets have been found useful in producing a bloodless operation field in surgical procedures involving the extremities including:
Reduction of certain fractures Kirschner wire removal Tumor and cyst excisions Subcutaneous fasciotomy Nerve injuries Tendon repair Bone grafts Total wrist joint replacement Replacement of joints in the fingers Knee joint replacements Amputations Replantations
WARNING: Do not use tourniquet cuffs to control the distal flow of CO2or any other gases used as a distention media. Tourniquet cuffs have not been evaluated for safety or effectiveness in controlling gas flow beyond the surgical site during arthroscopic insufflation procedures. Possible effects of using a tourniquet cuff in this manner include serious subcutaneous emphysema proximal to the cuff.
1.3 CONTRAINDICATIONS
The medical literature lists the following as possible contraindications. However, in every case, the final decision whether to use a tourniquet rests with the attending physician.
Open fractures of the leg Post-traumatic lengthy hand reconstruction Severe crushing injuries Elbow surgery (where there is excess swelling) Severe hypertension Skin grafts in which all bleeding points must be
readily distinguished
Compromised vascular circulation, e.g., peripheral
artery disease Diabetes mellitus The presence of sickle cell disease is a relative
contraindication. (See PRECAUTIONS IN USE.)
A tourniquet should also be avoided in patients who are undergoing secondary or delayed procedures after immobilization.
1.4 PRECAUTIONS IN USE
N The tourniquet system must be kept well calibrated and
in operable condition. Accessories should be checked regularly for leaks and other defects.
N The tourniquet cuff must never be punctured; therefore
towel clips used near the system must be handled with special care. Cuffs with inner rubber bladders must be completely enclosed by the outer envelope to preclude ballooning and possible rupture of the bladder. Cleaning and assembly instructions of the cuff manufacturer should be followed carefully.
N Do not use an elastic bandage for exsanguination in
cases where this will cause bacteria, exotoxins, or malignant cells to spread to the general circulation, or where it could dislodge thromboemboli that may have formed in the vessels.
N The tourniquet cuff must be applied in the proper
location on the limb, for a “safe” period of time, and within an appropriate pressure range. Never apply a tourniquet over the area of the peroneal nerve or over the knee or ankle. Do not readjust an already inflated cuff by rotating it because this produces shearing forces which may damage the underlying tissue.
N Prolonged ischemia may lead to temporary or permanent
damage to tissues, blood vessels, and nerves. Tourniquet paralysis may result from excessive pressure. Insufficient pressure may result in passive congestion of the limb with possible irreversible functional loss. Prolonged tourniquet time can also produce changes in the coagulability of the blood with increased clotting time.
N Inflation should be done rapidly to occlude arteries and
veins as near simultaneously as possible.
N Careful and complete exsanguination reportedly
prolongs pain free tourniquet time and improves the quality of Intravenous Regional Anesthesia (Bier Block anesthesia). In the presence of infection and painful fractures, after the patient has been in a cast, and in amputations because of malignant tumors, exsanguination before tourniquet application may be done without the use of an elastic bandage by elevating the limb for 3 to 5 minutes.
N In case of failure, the tourniquet cuff must be fully
deflated and the limb exsanguinated again before reinflation. Reinflation over blood-filled vasculature may lead to intravascular thrombosis.
N Tourniquet users must be familiar with the inflation-
deflation sequence when using a dual-cuff tourniquet or two tourniquet cuffs together for IVRA (Bier Block anesthesia), so that the wrong tourniquet will not be released accidentally.
N Test for hemoglobin type and level before using a
tourniquet on patients with sickle-cell anemia. When the tourniquet is used for these patients, the limb should be carefully exsanguinated and the PO2and pH should be closely monitored.
N Select the proper cuff size to allow for an overlap of
about 3 to 6 in. (7.6 cm – 15 cm). Too much overlap may cause cuff rolling and telescoping, and may lead to undesired pressure distribution on the limb. The skin under the tourniquet cuff must be protected from mechanical injury by smooth, wrinkle-free application of the cuff. If the tourniquet cuff is applied over any material that may shed loose fibers
5
A.T.S. 1200 TOURNIQUET SYSTEM
Page 8
(such as Webril) the fibers may become embedded in the contact closures and reduce their effectiveness. As an under padding, a section of stockinette may be used. The deflated
cuff and any underlying bandage or protective sleeve should be completely removed as soon as tourniquet pressure is released. After the cuff has been fully deflated and removed from the patient, the unit can be set to STANDBY. Even the slightest impedance of venous return
may lead to congestion and pooling of blood in the operative field.
N If skin preparations are used preoperatively, they should
not be allowed to flow and collect under the cuff where they may cause chemical burns.
N Whenever the tourniquet cuff pressure is released, the
wound should be protected from blood surging back by applying pressure dressings and, if necessary, elevating the limb. Transient pain upon tourniquet pressure release can be lessened by elevation of the limb. If full color does not return within 3 to 4 minutes after release, the limb should be placed in a position slightly below body level.
N Whenever IVRA Bier Block anesthesia is used, it is
recommended that the tourniquet remain inflated for at least 20 minutes from the time of injection.
N WARNING: Cuffs will not deflate in STANDBY
mode. Ensure cuffs are fully deflated before setting the unit to STANDBY.
1.5 ADVERSE EFFECTS
A dull aching pain (tourniquet pain) may develop throughout the limb following use.
Pathophysiologic changes due to pressure, hypoxia,
hypercarbia, and acidosis of the tissues occur and become significant after about 1.5 hours of tourniquet use. Symptoms of tourniquet paralysis are motor paralysis and loss of sense of touch, pressure, and proprioceptive responses.
Intraoperative bleeding may be caused:
1. By the slight impeding effect exerted by an unpressurized cuff (and its padding, if used), which prevents venous return at the beginning of the operation;
2. By blood remaining in the limb because of insufficient exsanguination;
3. By inadequate tourniquet pressure (between systolic and diastolic blood pressure of the patient), or slow inflation and deflation, all of which allow arterial blood to enter while preventing venous return;
4. By blood entering through the nutrient vessels of the long bones, such as the humerus.
6
A.T.S. 1200 TOURNIQUET SYSTEM
Page 9
2.1 INITIAL INSPECTION
Unpack the A.T.S. 1200 Tourniquet upon receipt and inspect the unit for any obvious damage that may have occurred during shipment. We recommend that this inspection be performed by a qualified biomedical engineer or other person thoroughly familiar with electronic medical devices. If the unit is damaged, notify the carrier and your Zimmer representative immediately. If the initial inspection results are satisfactory, a functional and calibration check should be performed after a 24
-hour char
ge. The attention label covering the ON/STANDBY button can be removed and discarded after the 24-hour charge.
2.2 CONTROLS, INDICATORS,
AND CONNECTORS
Refer to Figure 1 and 2 in the back of the manual for the locations of the unit’s controls, indicators, and connectors.
1. ON/STANDBY Button
Turns the unit ON or sets the unit to STANDBY. This button will not set the unit to STANDBY when the cuff pressure is at a non-zero value. Ensure both cuffs are
fully deflated and have been removed from the patient prior to setting the unit to STANDBY. NOTE: During STANDBY, the power to the A.T.S. 1200 instrument and all instrument functions (i.e. inflation, deflation, etc.) are OFF but power continues to supply the battery charging circuitry anytime ~ (AC) power (Mains) is present.
2. ROTARY knob
Changes the value of set time or default time and set pressure or default. Turn knob clockwise to increase the value; turn knob counterclockwise to decrease the value.
3. PRESSURE button
Press to verify or modify set pressure.
4. TIME button
Press to verify or modify set time.
5. MAIN CUFF INFLATE button
Inflation of the Main cuff is initiated by depressing the red INFLATE button.
6. MAIN CUFF DEFLATE button
Deflation of the Main cuff is initiated by depressing the red DEFLATE button. For greater safety, the DEFLATE button has a delay and, therefore, must be held for approximately 2 seconds before the unit will allow a cuff to deflate.
7. SECOND CUFF INFLATE button
Inflation of the Second cuff is initiated by depressing the blue INFLATE button.
8. SECOND CUFF DEFLATE button
Deflation of the Second cuff is initiated by depressing the blue DEFLATE button. For greater safety, the DEFLATE button has a delay and, therefore, must be held for approximately 2 seconds before the unit will allow a cuff to deflate.
9. ALARM SILENCE button
The ALARM SILENCE button will silence most
audible alarms for 30 seconds after the button is pressed. When an alarm sounds because of an internal hardware malfunction, the alarm cannot be silenced.
NOTE: The alarm messages will continue to flash on the displays until the alarm condition is corrected.
10. AC MAINS indicator light
The AC MAINS indicator light indicates that the unit is plugged in and is being powered by AC Mains. This is the normal means of operation (battery power is only intended for emergency power loss or patient transport).
11. BATTERY indicator light
The BATTERY indicator light indicates that the unit is operating on backup battery. The light will flash continuously while the unit is running on battery backup power.
12. PRESSURE display
During normal operation with no buttons being pressed, the PRESSURE display will show the monitored cuff(s) pressure. At other times, depending on alarm conditions and buttons pressed, this display may communicate other information such as alarm messages, set pressure, or default set pressure.
13. TIME display
During normal operation with no buttons being pressed, the TIME display will show elapsed inflation time of the first cuff that was inflated in 1-minute increments. At other times, depending on alarm conditions and buttons pressed, this display may communicate other information such as alarm messages, set time, or default set time.
NOTE: The elapsed inflation time can be “zeroed” at any point in the procedure by pressing the TIME and PRESSURE buttons simultaneously.
INSTALLATION
AND OPERATING
INSTRUCTIONS
SECTION 2.0
A.T.S. 1200 TOURNIQUET SYSTEM
7
Page 10
14. CUFF connector ports
The CUFF connectors are the ports used to connect the unit to the cuff hoses. Please note that the MAIN CUFF is the red port and the SECOND CUFF is the blue port. The A.T.S. 1200 Tourniquet is designed and tested for use with Zimmer single port cuffs. Zimmer does not recommend the use of any cuff other than Zimmer single port cuffs. Do not use dual port cuffs with the A.T.S. 1200 Tourniquet.
15. CUFF indicators
The CUFF indicators illuminate steady when stable cuff pressure is sensed in the cuff. The CUFF indicators will also flash during inflation or to indicate an alarm condition in the respective cuff.
16. Pole clamp
The pole clamp is used to mount the unit on an I.V. pole.
NOTE: Do not hang articles on the tourniquet pole that are not related to tourniquet use. For stability reasons, do not use an I.V. pole with a base less than 27.27 inches (70 cm) in diameter.
2.3 INITIAL SETUP
Inspect to ensure the correct fuse drawer with the appropriately rated fuses is present. The 100–120 V unit uses the gray fuse drawer with 1.0 A time delay fuses. The 220–240 V unit uses the black fuse drawer to 1.0 A time delay fuses. The power cord should be plugged into the power entry module on the back of the unit. The unit should
be plugged into ~ (AC) power (Mains) for 24 hours before initial use. During shipping and storage, the unit’s battery could become weak. Always charge 24 hours before any initial use including any calibration checking procedures, initial checks, tests and any institutional performed biomedical evaluations.
2.4 FUNCTIONAL AND CALIBRATION CHECK
The unit shall produce the results explained in the following steps exactly as indicated. Failure to do so indicates that a problem may exist and the device is not to be used until necessary repair or calibration has been made.
1. Connect the power plug of the unit to a properly polarized and grounded power source with voltage and frequency characteristics compatible with the specifications listed in Section 1.1. Observe that the green AC MAINS indicator light turns on.
2. Turn the unit ON by pressing the ON/STANDBY button and observe the following:
a) A “0*0*”/“*0*0” sequence appears on the
PRESSURE and TIME displays followed by the text “ATS” “1200” on the respective display.
b) The unit emits tones when “0*0*”/“*0*0” and text
is displayed.
c) The unit will display “SELF” “TEST” for an instant.
The unit is self-testing specific system hardware and software.
d) “CAL” is displayed in the PRESSURE displays
during the calibration check.
e) “0” is displayed in the PRESSURE and TIME
display after the startup routing is complete. If a number other than zero is displayed in the PRESSURE display, the unit should be calibrated.
3. Test the PRESSURE set point system as follows: a) Press the PRESSURE button. b) The PRESSURE display should read “*250” (the
default set point) for 2 seconds.
c) Within the 2-second time frame, rotate the
ROTARY knob to change the pressure set point (clockwise to increase, counter-clockwise to decrease). The set pressure can be maintained between 50 mm Hg and 475 mm Hg in increments of 5 mm Hg.
4. Test the TIME set point system as follows: a) Press the TIME button. b) The main TIME display should read “*60” (the
default set point) for 2 seconds.
c) Within the 2-second time frame, rotate the
ROTARY knob to change the time set point (clockwise to increase, counter-clockwise to decrease). The set time can be maintained between 5 and 240 minutes in increments of 5 minutes.
NOTE: Anytime an asterisk (*) appears in the left display digit, the data being displayed is the set point. Set pressure and time will revert to the default pressure and time when the unit is set to STANDBY.
5. Calibration Check
NOTE: During the power-up diagnostic self-test described above, the unit will test calibration. Should an out of calibration condition be detected, the unit will display either “CAL” “FAIL”, “CALM” “FAIL” or “CAL2” “FAIL” in the PRESSURE and TIME displays. Even though the unit performs this check at every power-up, the following quantitative check is recommended at regular intervals.
a) Verify the unit is in the STANDBY mode. b) Enter the calibration mode by pressing and holding
the Main cuff Inflate and Deflate buttons while powering the unit ON. The unit will enter the calibration mode when the Pressure and Time display momentarily displays “CAL” “MODE”. After the “CAL” “MODE” is displayed, the unit will display the software revision level. The software revision level can be recorded for future reference.
NOTE: The calibration is only being checked in this section.
For complete calibration, see Maintenance Section 3.0.
8
A.T.S. 1200 TOURNIQUET SYSTEM
Page 11
c) Connect a calibrated pressure meter, with a minimum
range of 0 to 500 mm Hg, to the calibration hose. The calibrated meter will be used as the pressure standard (see Figure 3 in the back of the manual).
d) Connect a pressure source capable of supplying
500 mm Hg of pressure.
e) Insert the calibration hose connector into the Main
Cuff port on the unit (red port).
NOTE: The unit will be displaying “0” in the Pressure display and alternating “CAL” and “0” in the Time display. During the calibration check, the unit will attempt to deflate any applied pressure if the adjustment point displayed is set to zero. This will result in an inaccurate calibration check. Press the MAIN CUFF INFLATE button once to change the adjustment point to 50 mm Hg in order to close the deflate valve.
f) Apply 50 mm Hg of pressure to the cuff port. The
PRESSURE display should read 50±5 mm Hg when compared to the calibrated meter.
g) Increase the pressure to 250 mm Hg. The
PRESSURE display should read 250±5 mm Hg.
h) Increase the pressure to 475 mm Hg. The
PRESSURE display should read 475±5 mm Hg.
i) Remove the calibration hose setup from the unit.
The PRESSURE display should now read 0 mm Hg.
j) At this point, use the MAIN CUFF DEFLATE
button to back the pressure level down to the point where the Time display is alternating between “CAL” and “0”. Then, press the MAIN CUFF INFLATE and the MAIN CUFF DEFLATE buttons simultaneously to exit the Main cuff calibration check and to enter the Second cuff calibration check.
k) Connect the previously used setup to the Second
Cuff port (blue port).
NOTE: The unit will be displaying “0” in the Pressure display and alternating “CAL” and “0” in the Time display. During the calibration check, the unit will attempt to deflate any applied pressure if the adjustment point displayed is set to zero. This will result in an inaccurate calibration check. Press the SECOND CUFF INFLATE button once to change the adjustment point to 50 mm Hg in order to close the deflate valve.
l) Apply 50 mmHg of pressure to the cuff port. The
PRESSURE display should read 50±5 mmHg when compared to the calibrated meter.
m) Increase the pressure to 250 mmHg. The
PRESSURE display should read 250±5 mmHg.
n) Increase the pressure to 475 mmHg. The
PRESSURE display should read 475±5 mmHg.
o) Remove the calibration hose from the unit. The
PRESSURE display should now read 0 mmHg.
NOTE: If any reading is off by more than 5 mm Hg during the calibration check, the unit must be calibrated. See CALIBRATION in MAINTENANCE, Section 3.
p) At this point, use the SECOND CUFF DEFLATE
button to back the pressure level down to the point where the Time display is alternating between “CAL” and “0”. Exit the calibration check by simply setting the unit to STANDBY.
6. Low Pressure Alarm Check
Connect a cuff and standard length hose to the Main cuff port. Inflate the cuff to 250 mm Hg. Create a leak in the cuff by partially detaching the hose from the unit while a cuff is inflated. Make the leak large enough that the pressure drops more than 15 mm Hg below set point. Observe:
a) A 1.5 second delay is instituted to reduce
nuisance alarms.
b) The PRESSURE display flashes between “LO-P”
and the monitored pressure (if the leak is large enough, the PRESSURE display may show “LEAK”).
c) An audible tone will sound and the red ALARM
indicator will illuminate announcing the alarm condition.
d) Stop the leak and observe the monitored pressure
returns to regulated state, the audible tone stops, the red ALARM indicator turns OFF, and the alarm message is no longer displayed.
Repeat this procedure with the Second cuff port.
2.5 PRESSURE AND TIME DEFAULTS
To modify the default pressure or time limits, follow the following steps.
1. Default Pressure a) The Default Pressure is selected by depressing and
holding the PRESSURE button for at least 2 seconds. When the default mode is entered, the audible alarm beeps once and a “D” is displayed in the first position on the PRESSURE display.
b) The Default Pressure is modified via the ROTARY
knob and can be set between 50 and 475 mm Hg in increments of 5 mm Hg.
c) After the correct value is selected, it is saved by
momentarily depressing the PRESSURE button or it will be saved automatically in 3 seconds.
d) The new default value will be displayed for
1.5 seconds and the audible alarm will beep once signifying a new default value has been stored.
e) The new default pressure will be stored and remains
9
A.T.S. 1200 TOURNIQUET SYSTEM
Page 12
the default every time the machine is turned on.
2. Default Time Limit a) The Default Time Limit is selected by pressing and
holding the TIME button for at least 2 seconds. When the default mode is entered the audible alarm beeps and a “D” is displayed in the first position on the TIME display.
b) The Default Time Limit is modified via the
ROTARY knob and can be set between 5 and 240 minutes in increments of 5 minutes.
c) After the correct value is selected, it is saved by
momentarily depressing the TIME button or it will be saved automatically in 3 seconds.
d) The new default value will be displayed for 1.5
seconds and the audible alarm will beep once
signifying a new default value has been stored. The new time limit default will be stored and remains the default every time the machine is turned on.
NOTE: The elapsed inflation time can be “zeroed” at any point in the procedure by pressing the TIME and PRESSURE buttons simultaneously.
2.6 SINGLE CUFF OPERATION
1. Press the ON/STANDBY button to turn the unit on. The unit will execute a self-check diagnostic test as described in Section 2.4 of this manual. Successful completion of the self-check indicates the unit is ready for use. CAUTION: If a connected cuff is pressurized to 50 mm Hg or more during power-up, the A.T.S. 1200 Tourniquet will declare it an abnormal start-up sequence. It will assume that a surgical procedure is in process, and will adopt the pressure sensed in the cuff as the new set point. It will automatically go into the regulate mode on the cuff. To alert the operator of this condition, the unit will sound a tone and display a “CUFF” “INFL” alarm. The operator should immediately check the pressure set point and readjust to the proper set point if necessary. The alarm will be cleared as soon as the set point is examined (PRESSURE button pressed).
2. Connect a single port cuff to the unit at the Main cuff connector (red port).
3. The default settings for cuff pressure and time limit are retrieved from the nonvolatile memory during power up.
For each patient, tourniquet pressure required to occlude blood flow to operative site should be set to the minimum effective pressure. The minimum effective pressure should be determined by factors such as: whether the cuff is to be applied to an upper or lower limb; whether the limb is normal, hypertrophied, or obese; the patient’s preoperative systolic pressure; and the maximum anticipated rise in systolic pressure during the procedure.
4. Prepare the patient in accordance with your established procedures and cuff manufacturer’s instructions. The precautions of Section 1 and the following are offered as a guide to assist in this process.
In most cases a tourniquet cuff should be applied to the widest part of the limb to allow as much tissue as possible to lie between the cuff and any nerves or vascular structures susceptible to damage. The optimum positions are the upper arm and the proximal third of the thigh. In certain cases of fore-foot surgery, the tourniquet cuff can be applied around the calf or to the area proximal to the malleoli. For emergency surgery of the hand, a sufficiently small tourniquet can be fitted around the wrist.
Apply a leak-free tourniquet cuff smoothly without wrinkles. The valve port and hose connections should be placed so that the hose will not be kinked when the limb is positioned for surgery. The viability of the skin and deeper tissues should be established prior to exsanguination of the limb and tourniquet inflation. Exsanguinate the limb by elevating it for a minimum of 2 minutes and wrapping it, distal to proximal, using an Esmarch, Martin, or elastic bandage. The bandage should come up approximately to 1 in. (2.5 cm) from the edge of the tourniquet cuff. The elastic bandage is removed following inflation of the cuff. If regional anesthesia is being used, the anesthetic agent or nerve block is then administered. The tourniquet time depends greatly on the patient’s anatomy, age, and absence of vascular disease. The surgeon will determine:
1) When the tourniquet is to be inflated;
2) What pressure is applied;
3) How long the tourniquet is applied;
4) Whether to allow for intermittent aeration of tissue by deflating the cuff for 10 to 15 minutes;
5) To what point in the operation the tourniquet should be released.
In many operating rooms, it is customary to prominently note the time of inflation, and to warn the surgeon after a certain time has elapsed. This will allow the surgeon to assess the need for further tourniquet time.
There is a general agreement that, for reasonably healthy adults, 2 hours should not be exceeded without releasing the tourniquet to allow the underlying tissue to breathe. During this time, the limb should be elevated to about 60 degrees, and steady pressure should be applied to the incision with sterile dressings.
5. The cuff is inflated by pressing the red MAIN CUFF INFLATE button. The unit will pressurize the Main cuff to the set pressure and start the time limit (inflation) clock. The MAIN CUFF INFLATION indicator will
10
A.T.S. 1200 TOURNIQUET SYSTEM
Page 13
illuminate to indicate cuff activation. If the unit cannot pressurize the cuff to within 15 mm Hg of the set point in less than 13 seconds, a leak alarm will be sounded. See Section 2.8 for information about possible alarm conditions. Once the cuff is inflated, the TIME display will track elapsed inflation time.
6. At the end of the procedure, deflate the cuff by pressing the MAIN CUFF DEFLATE button for minimum of 2 seconds. The MAIN CUFF bar indicator will go out, the PRESSURE display will show the deflation of the cuff, and the time limit (inflation) clock will stop.
NOTE: The elapsed inflation time can be “zeroed” at any point in the procedure by pressing the TIME and PRESSURE buttons simultaneously.
7. Remove the tourniquet cuff and any underlying
bandages or protective sleeve immediately following final deflation. The time of tourniquet cuff removal
should be noted, and the circulation of the limb should be checked.
8. After the cuff has been removed, disconnect the cuff from the A.T.S. 1200 Tourniquet.
9. During normal use, the A.T.S. 1200 Tourniquet should not be set to STANDBY if pressure is present in either cuff. Once the cuff has been properly deflated, removed from the patient and disconnected from the A.T.S. 1200 Tourniquet, the unit can be set to STANDBY.
2.7 DUAL CUFF OPERATION
Operation of the unit is identical to Single Cuff operation (see Section 2.6) except for the following points:
1. Both single port cuffs are connected at the bottom of the unit (Reminder: Main cuff is the red port, Second cuff is the blue port).
2. The MAIN CUFF and SECOND CUFF indicator lights will illuminate to indicate cuff activation (i.e. if both cuff indicator lights are illuminated then both cuffs are inflated).
3. During a pressure alarm a flashing CUFF indicator light warns which cuff(s) to check.
4. Deflation of one cuff will not be permitted while the other cuff is inflating.
5. When inflating a second cuff with the other cuff already inflated, the unit will continuously check the original cuff to ensure that the pressure is within allowable limits. The unit will stop its inflation and maintain the original cuff to within 10 mm Hg of the set point before returning to the inflating cuff. This ensures that at least one cuff maintains occlusion at all times. If there is a significant leak in the original cuff, this feature could cause the inflation rate of the subsequent cuff to be longer and perhaps even cause the 30-second inflation alarm to sound. The display shows the pressure in the
inflated cuff so as to allow the operator to view the progress of inflation.
6. When both cuffs are inflated, the display indicates an average of the pressures sensed in the two cuffs. If one cuff has a sustained leak or a significant leak (as described in Section 2.9.1) a pressure alarm may be sounded.
7. When deflating a cuff with the other cuff remaining inflated, the display shows the pressure in the cuff that remains inflated.
8. In order to deflate the final cuff, a sequence must be
followed to prevent accidental deflation:
a) Press and hold the DEFLATE button on the cuff
to be deflated.
b) When the “CUFF” “DEFL” alarm is active,
release the DEFLATE button.
c) Within 5 seconds of the alarm discontinuing,
press the DEFLATE button once again.
d) The cuff will deflate. This safety feature is
particularly useful when using the unit for Bier Block Cuff Operation (IVRA).
2.8 BIER BLOCK CUFF OPERATION (IVRA)
Review Sections 2.6 and 2.7, SINGLE CUFF OPERATION and DUAL CUFF OPERATION.
1. The following are suggested cuff connections:
a) The proximal cuff connected to the red Main cuff
port using the white/red cuff tubing;
b) The distal cuff connected to the blue Second cuff
port using the white/blue cuff tubing.
2. Follow the cuff inflation sequence adopted by your institution or requested by the surgeon.
3. Deflation of a cuff is not possible while the other is inflating.
4. When requested, the first can be deflated simply by pressing and holding the DEFLATE button for a minimum of 2 seconds.
5. In order to deflate the final cuff, a sequence must be
followed to prevent accidental deflation:
a) Press and hold the DEFLATE button on the cuff
to be deflated.
b) When the “CUFF” “DEFL” alarm is active,
release the DEFLATE button.
c) Within 5 seconds of the alarm discontinuing,
press the DEFLATE button once again.
d) The cuff will deflate.
6. For Bier Block procedures follow the cuff inflation/ deflation sequence adapted by your institution or requested by the surgeon.
2.9 ALARM CONDITIONS
There are a number of conditions for which the A.T.S. 1200 Tourniquet will produce a visual and/or audible alarm. Those conditions, indications and appropriate actions are shown in Table 2.1. The appropriate actions indicated are based on the
11
A.T.S. 1200 TOURNIQUET SYSTEM
Page 14
most probable causes and should only be used as a guide. Other causes of alarm conditions may indicate a need for other actions.
In addition to the conditions shown in Table 2.1, it is conceivable that a malfunction could occur for which the indications are unintelligible and unpredictable. In this situation, it is likely that the valves will be disabled causing the system to hold cuff pressure. It is also likely that a tone will sound under these conditions.
Most audible alarm tones may be silenced for 30 seconds by depressing the ALARM SILENCE button. At the end of the silenced period, tones will be reenabled. Depressing the ALARM SILENCE button will cause the alarm tone to be silenced again.
To minimize nuisance alarms (i.e. “HI-P”, “LO-P”) that can be caused by vigorous movement of the patient’s limbs, a 1.5-second delay has been designed into the alarm actuation.
Under certain conditions, such as when a FAIL indication appears in the TIME display or the information that appears in the TIME and PRESSURE displays is unintelligible, the operator should conclude that a hardware failure has occurred, rendering the unit unusable. The appropriate action is to set the unit to STANDBY by pressing the ON/STANDBY button.
Since this removes power from the internal instrument circuitry, all instrument functions, commands to the valves and pump will cease. This will cause the cuff to hold pressure (in the absence of leaks). Clamp the cuff line with
hemostats and replace the tourniquet unit.
2.9.1 PRESSURE ALARMS
A Pressure alarm will occur when the pressure in a cuff is more than 15 mm Hg from the pressure set point. It is also possible for a cuff to have a leak that is substantial but which the unit can compensate for by continual pumping. This type of leak could be due to a pin hole in a cuff bladder, or a loose pneumatic fitting. This type of leak could progress into a total failure of a cuff to hold pressure. To alert the operator that a substantial leak is present, a pressure alarm is declared when this type of leak is continuously present for more than 9 seconds. If a pressure alarm occurs, and the displayed pressure is not more than 15 mm Hg from the set point, then this type of substantial leak has been detected and all cuffs and pneumatic fittings should be checked for leaks.
12
A.T.S. 1200 TOURNIQUET SYSTEM
Page 15
Table 2.1 Alarm Conditions
CONDITION PRESSURE TIME APPROPRIATE ACTION/REMARKS
DISPLAY DISPLAY
CUFF PRESSURE LOW LO-P normal This condition is generally caused by a leak in the The pressure in the cuff system, or a hose occlusion. All lines and is 15 mm Hg below set point. connections should be checked.
CUFF PRESSURE HIGH HI-P normal Normally caused by transient conditions such as The pressure in the cuff is patient movement, controller overshoot, or hose 15 mm Hg above set point. occlusion. This condition, for an extended period,
would indicate a hardware failure and the A.T.S. 1200 unit should be replaced.
CUFF SIDE LEAK LEAK normal A substantial leak has been present for more than A leak has been present for 7 seconds. All lines and connections should be at least 7 seconds. checked.
INFLATION TIME IN normal TIME UP Surgeon should be warned of time up condition. EXCESS OF SETTING Only on the direction of the surgeon, time should The cuff has been inflated be set to new value. beyond the set time limit.
CUFF INFLATION ON CUFF INFL The system assumes that a procedure is in progress POWER UP and adopts the sensed pressure as the new set point. Cuff pressurized to 50 mm Hg The operator should immediately check the set value or greater at power up. to determine if it needs resetting.
CUFF NOT DEFLATED normal CUFF Check for kinks in hose. If alarm persists, Pressure in deflated cuff is NOT disconnect hose from cuff. If attempting to set the a non-zero value. DEFL unit to STANDBY, ensure that cuff is fully deflated.
LINE OCCLUSION LINE OCCL normal Check for hose kinks or other defects. An occlusion is present in the cuff tubing.
LOW BATTERY VOLTAGE normal BAT LOW Unit needs to be plugged in. Low battery voltage. PLUG IN
BATTERY FAILURE BATT FAIL Plug unit in and cycle the ON/STANDBY Button. Battery voltage is too low to ensure proper operation.
CALIBRATION OUT OF SPEC CALM FAIL CALM (CALIBRATION MAIN) indicates the The transducer calibration or CAL2 main cuff transducer circuitry is out of calibration. is out of specification. CAL2 (CALIBRATION SECOND) indicates the
second cuff transducer circuitry is out of calibration. Pressure in error by at least 6 mm Hg will cause these failures. Calibrate the unit.
CALIBRATION OUT OF SPEC CAL FAIL Indicates general calibration fail. Calibrate the unit. The transducer calibration is out of specification.
AMPLIFIER FAILURE AMP FAIL Cycle the ON/STANDBY Button. If problem Amplifier is out of range. persists, service the unit.
MATH FAILURE MATH FAIL Cycle the ON/STANDBY Button. If problem Result of math operation persists, service the unit. was out of range.
13
A.T.S. 1200 TOURNIQUET SYSTEM
Page 16
ROM FAILURE ROM FAIL Cycle the ON/STANDBY Button. If problem Microprocessor failed a ROM persists, service the unit. memory check.
RAM FAILURE RAM FAIL Cycle the ON/STANDBY Button. If problem Microprocessor failed a RAM persists, service the unit. memory check.
VALVE FAILURE VALV FAIL Cycle the ON/STANDBY Button. If problem Improper valve persists, service the unit. combination occurred.
WATCHDOG FAILURE WDT FAIL Cycle the ON/STANDBY Button. If problem Windowing watchdog system persists, service the unit. detected a malfunction.
SYSTEM FAILURE SYS FAIL Cycle the ON/STANDBY Button. If problem Internal diagnostic detected an error. persists, service the unit.
OVERPRESSURE OVER PRES Cycle the ON/STANDBY Button. If problem The PRESSURE button has detected persists, service the unit. This occurrence a pressure higher than the range could indicate a runaway pump. of the unit.
14
A.T.S. 1200 TOURNIQUET SYSTEM
Page 17
3.1 GENERAL MAINTENANCE INFORMATION
While the A.T.S. 1200 Tourniquet has been designed and manufactured to high industry standards, it is recommended that regular inspection and calibration be performed to ensure continual safe and effective operation. This section contains information to assist in the effort as well as serve as a guide to expediting unscheduled maintenance.
3.2 ACCESS TO PARTS
CAUTION: Be sure that the unit is set to STANDBY and
the power plug is unplugged before disassembly. Many of the parts on the control board are static sensitive. Take precaution when servicing the board.
To gain access to all internal parts, remove: a) Rear – 5 screws b) Rear – 2 large pole clamp screws c) Bottom – 4 foot pad screws
See Illustration 4 through 8 in the back of the manual.
When opening, take care not to damage any of the wire harnesses or pneumatic tubing. The control board is attached to the front housing therefore the harnesses and tubing will need to be disconnected for full disassembly. Follow the table below to reassemble.
Table 3.1 Board Plug Designators
Component Board Plug Location
AC Mains P1 Overpressure Switch P2 Speaker P3 Pneumatic Pump P5 Valve Harness P6 Backup Battery P7 Membrane Panel P8 Factory Test Port P9
Main Cuff Port Tube U17 Second Cuff Port Tube U21
To reduce the risk of damage, the tubing should be disconnected at the plastic tee fitting and not the transducer.
NOTE: Failure to plug the electrical or pneumatic components into the correct associated receptacle can result in damage to the control board.
When r
eassembling the unit, be extremely car
eful not to
pinch any wiring or tubing
.
3.3 PERIODIC MAINTENANCE
Test and inspect as per this section at minimum every six months.
1. Cleaning
The exterior of the unit may be cleaned with a cloth that has been dampened (not dripping) with a mild detergent. The interior of the unit may be vacuumed or blown out as required. The exterior of the cuff hose may be cleaned using a mild detergent solution or alcohol. The interior of the cuff hoses should not be cleaned. Tourniquet cuffs should be cleaned in accordance with their cuff package inset instructions.
2. Inspection
The unit should be inspected at regular intervals. It is recommended that a qualified technician perform a visual inspection at least once every six months.
Inspection points are: a) Obvious internal or external damage. b) Condition of the power cord. c) Condition of the power cord-retaining clip. d) Tightness of pneumatic fittings. e) Condition of internal tubing. f) Accumulation of dust or dirt within the unit. g) Mating integrity of internal connectors. h) Security of the EPROM and safety processor.
See Illustration 9 for location. i) Integrity of the pump. j) Security of circuit board. k) Security of the membrane panel.
3. Functional and Calibration Checks
It is recommended that the functional and calibration checks described in Section 2.4 are performed at least once every three months.
3.4 CALIBRATION
Calibration should be performed every six months, or after any unscheduled maintenance.
Calibration of the A.T.S. 1200 Tourniquet allows the output signal from the pressure transducer to be compared against a calibrated pressure source. The difference between the known pressure and the pressure measured by the transducers is recorded at each of four set points (0 mm Hg, 50 mm Hg, 250 mm Hg, and 475 mm Hg). These four calibration factors are used to correct the signal from the pressure transducers during normal operation. The calibration factors and a checksum are stored in non-volatile memory.
15
MAINTENANCE
SECTION 3.0
A.T.S. 1200 TOURNIQUET SYSTEM
Page 18
EQUIPMENT REQUIRED:
A.T.S. 1200 calibration hose (supplied). Calibrated 0 to 500 mm Hg pressure meter. Adjustable 0 to 500 mm Hg pressure source.
CAUTION: The following steps must be taken in the exact order to calibrate the unit. Failure to do so may result in incorrect pressure readings while the unit is in operation.
MAIN CUFF CALIBRATION The Main cuff transducer is always calibrated first. Below is a step-by-step procedure for calibrating the Main cuff transducer. The procedure continues with the Second cuff. Both cuff transducers should always be calibrated. The calibration will not be complete unless both cuff transducer calibrations are performed.
1. To enter the calibration mode, press and hold the MAIN CUFF INFLATE and DEFLATE buttons while powering the unit ON. The unit will enter the calibration mode when the PRESSURE and TIME display momentarily displays “CAL” “MODE”. After the “CAL” “MODE” is displayed, the unit will display the software revision level. The software revision level can be recorded for future reference.
2. After displaying the software revision level, the unit will display “MAIN” “CUFF” in the PRESSURE and TIME displays respectively. This is to indicate that the unit is now ready to calibrate the Main cuff transducer.
3. The unit will now display “0” in the PRESSURE display and alternating “CAL” and “0” in the TIME display. Throughout this procedure, the TIME display will indicate the pressure in which the user is calibrating.
4. For zero, allow the port to be open to atmospheric pressure so the unit can sense the zero point (i.e. when setting the zero point, nothing should be connected to the cuff port). By pressing the PRESSURE button, the unit will calibrate the pressure. The unit will beep to let the user know the set point was taken.
5. Connect the calibration hose, calibrated pressure meter and adjustable pressure source to the Main Cuff port. See Figure 3 for more details.
6. Once the zero point is calibrated, press the MAIN CUFF INFLATE button to advance the unit to the next pressure level. The unit’s TIME display will now be alternating between “CAL” and “50”. Apply 50 mm Hg to the unit’s Main cuff port. Once the pressure has stabilized, press the PRESSURE button so the unit can calibrate the 50 mm Hg point. The unit will beep to let the user know the set point was taken.
7. Once the 50 mm Hg point is calibrated, press the MAIN CUFF INFLATE button to advance the unit to the next pressure level. The unit’s TIME display will now be alternating between “CAL” and “250”. Apply 250 mm Hg to the unit’s Main cuff port. Once the pressure has
stabilized, press the PRESSURE button so the unit can calibrate the 250 mm Hg point. The unit will beep to let the user know the set point was taken.
8. Once the 250 mm Hg point is calibrated, press the MAIN CUFF INFLATE button to advance the unit to the next pressure level. The unit’s TIME display will now be alternating between “CAL” and “475”. Apply 475 mm Hg to the unit’s Main cuff port. Once the pressure has stabilized, press the PRESSURE button so the unit can calibrate the 475 mm Hg point. The unit will beep to let the user know the set point was taken.
9. At this point, use the MAIN CUFF DEFLATE button to back the pressure level down to the point where the TIME display is alternating between “CAL” and “0”. Then, press the MAIN CUFF INFLATE and the MAIN CUFF DEFLATE buttons simultaneously to exit the Main cuff calibration and to enter the Second cuff calibration.
SECOND CUFF CALIBRA
TION The Second cuff transducer is always calibrated last. Below is a step-by-step procedure for calibrating the Second cuff transducer. The Second cuff transducer calibration is performed the same way the Main cuff transducer calibration. Following the completion of the Second cuff transducer calibration, be certain to perform the techniques described to save the calibration in non­volatile memory.
10. Once the unit enters the Second cuff transducer calibration, the unit will beep and momentarily display “SCND” “CUFF” in the PRESSURE and TIME window respectively. This is to indicate that the unit is now ready to calibrate the Second cuff transducer.
11. The unit will now display “0” in the PRESSURE display and alternating “CAL” and “0” in the TIME display. Throughout this procedure, the TIME display will indicate the pressure in which the user is calibrating.
12. For zero, allow the port to be open to atmospheric pressure so the unit can sense the zero point (i.e. when setting the zero point, nothing should be connected to the cuff port). By pressing the PRESSURE button, the unit will calibrate the pressure. The unit will beep to let the user know the set point was taken.
13. Connect the calibration hose, calibrated pressure meter and adjustable pressure source to the Second Cuff port. See Figure 3 for more details.
14. Once the zero point is calibrated, press the SECOND CUFF INFLATE button to advance the unit to the next pressure level. The unit’s TIME display will now be alternating between “CAL” and “50”. Apply 50 mm Hg to the unit’s Second cuff port. Once the pressure has stabilized, press the PRESSURE button so the unit can calibrate the 50 mm Hg point. The unit will beep to let the user know the set point was taken.
16
A.T.S. 1200 TOURNIQUET SYSTEM
Page 19
15. Once the 50 mm Hg point is calibrated, press the SECOND CUFF INFLATE button to advance the unit to the next pressure level. The unit’s TIME display will now be alternating between “CAL” and “250”. Apply 250 mm Hg to the unit’s Second cuff port. Once the pressure has stabilized, press the PRESSURE button so the unit can calibrate the 250 mm Hg point. The unit will beep to let the user know the set point was taken.
16. Once the 250 mm Hg point is calibrated, press the SECOND CUFF INFLATE button to advance the unit to the next pressure level. The unit’s TIME display will now be alternating between “CAL” and “475”. Apply 475 mm Hg to the unit’s Second cuff port. Once the pressure has stabilized, press the PRESSURE button so the unit can calibrate the 475 mm Hg point. The unit will beep to let the user know the set point was taken.
17. At this point, use the SECOND CUFF DEFLATE button to back the pressure level down to the point where the TIME display is alternating between “CAL” and “0”. Then, press the SECOND CUFF INFLATE and the SECOND CUFF DEFLATE buttons simultaneously to exit the Second cuff calibration. The unit will display “CAL” “DONE” in the PRESSURE and TIME windows respectively. The “CAL” “DONE” message tells the user that the calibration set points that have been calibrated to the unit have now been saved into the non-volatile memory.
NOTE: If the “CAL” “DONE” message is not displayed, the calibration is incomplete and the adjustments will not be saved. Be certain to end the calibration session by simultaneously pressing the SECOND CUFF INFLATE and DEFLATE buttons and verifying the “CAL” “DONE” message.
If the pressure settings were adjusted incorrectly on the Main or Second cuff and an attempt is made to have the unit calibrate the incorrect value that is off by more than 15 mm Hg, a “CAL” “FAIL” alarm will be generated. Service or calibration to the user’s pressure meter or pressure source is recommended.
If the pressure signal from the internal transducer requires more then a 15 mm Hg correction to equal the applied pressure, a “CAL” “FAIL” alarm will also be generated. Service to the unit is recommended.
18. The unit remains in calibration mode until it is set to STANDBY.
19. The stored calibration factors are retrieved from the non­volatile memory during the power-up sequence. If the checksum is invalid, a “CAL” “FAIL” alarm is generated in the displays. The alarm will persist until the unit is set to STANDBY. Re-calibration is required if this occurs.
20. It is recommended to check the calibration by following the steps in Section 2.4 Step 5 “Calibration Check” before using this unit on a patient.
3.5 LEAK TESTING
The A.T.S. 1200 Tourniquet is capable of keeping a cuff with a substantial leak inflated. Naturally it is desirable to keep plumbing leaks to an absolute minimum. For this reason, a check for significant leakage is recommended at regular intervals as well as following any service procedure.
After verifying the operation of the A.T.S. 1200 Tourniquet per Section 2.4, connect a 24 in. (61 cm) (or larger) cuff which is known to be leak free to the Main cuff port (red) of A.T.S. 1200 Tourniquet System. Adjust the Main cuff set point to 475 mm Hg. Ensure that all external connections are tight. Inflate the Main cuff and allow the pressure to stabilize. At this point, the unit must be set to STANDBY. Under normal use, the unit cannot be set to STANDBY with a non-zero pressure value displayed in either cuff. However for leak testing purposes, a bypass feature has been incorporated. Press the ON/STANDBY button until the alarm message “CUFF” “NOT” “DEFL” appears. Release the ON/STANDBY button and within 5 seconds of the alarm discontinuing, press and hold the ON/STANDBY button again. The button must be held in for an additional 10 seconds before the unit will be set to STANDBY.
NOTE: During the 10 seconds, the alarm message will be displayed, the alarm will continue to sound and the ALARM SILENCE button will not silence the alarm.
Once the unit is set to STANDBY, wait for approximately 10 minutes and turn the unit back on. Operation will resume under cuff inflated start-up conditions (See Section 2.6 Part 1 for explanation). Cancel the alarm using the ALARM SILENCE button. Display the set point by activating the PRESSURE button and view the current (New) pressure set point. The set point is always displayed with an asterisk in the far left position. The current set point for the cuff should be at least 400 mm Hg or more. Values less than this indicate an unacceptable leak rate and the source of the leak should be traced and corrected. The first connection to check should be the connections of the cuff. Different cuffs and/or cuff hoses may be tried to determine if the leak is internal or external of the unit.
Repeat the test for the Second cuff port.
3.6 BATTERY VOLTAGE AND BATTERY
SERVICE
NOTE: This section assumes that the unit has been charged
for at least 24 hours. The unit’s enclosure must be removed to measure battery voltage. See Section 3.2 “Access to Parts” and be sure to follow cautionary statements.
1. Battery Voltage Check
Be sure the unit is unplugged. Measure the battery voltage. The battery voltage should not be lower than 12 V while the unit is unplugged and set to STANDBY.
17
A.T.S. 1200 TOURNIQUET SYSTEM
Page 20
If, after 1 minute, the voltage reads less than 12 V, the integrity of the battery should be suspect and should be replaced.
2. Battery Service The 12 V sealed lead acid battery is charged using lead acid charging technology. The charging circuit is active anytime the unit is plugged into an acceptable AC Mains outlet. The charger automatically sequences through several charge states based on the battery voltage and charging current conditions. Based on a charger test, the best charge mode is selected. No maintenance is required of the battery charging circuit.
The life of the battery depends on the type of service and the storage method. Battery replacement will need to be more frequent with continued cycles of deep discharge and/or storage in a high temperature environment.
Infrequent short-term use of the battery and storage in a room temperature environment will result in maximum life. It is recommended that the battery in the A.T.S. 1200 Tourniquet System be replaced annually. As a reminder, the A.T.S. 1200 System should be plugged in 24 hours before initial use.
3.7 UNSCHEDULED MAINTENANCE
The A.T.S. 1200 Tourniquet is designed with several specific self-test features to assist in fault isolation. These features are designed to show messages in the PRESSURE and TIME displays. The meanings of these messages are delineated in Table 2.1.
Another mode of failure that may occur is when an
audible alarm occurs that cannot be silenced by the ALARM SILENCE button. The valve and pump will be disabled which seals off the cuff to prevent pressure loss. The displays may show random characters. Should this occur, the watch dog timer circuit of the safety processor has detected a problem. The microprocessor may not be executing reliable instructions and is not able to display the correct failure message. This unit should be serviced if this occurs.
The calibration error message “CAL” “FAIL”, “CALM” “FAIL” or “CAL2” “FAIL” may be due to defective circuitry or may simply indicate the need for calibration.
3.8 TROUBLE SHOOTING GUIDE
To aid in unscheduled maintenance, Table 3.2 delineates a number of possible malfunctions that could occur with the unit. The most likely causes are shown for each symptom. While it is not practical to enumerate every conceivable malfunction and all possible causes, the table will assist in isolating the most common problems.
3.9 EXPECTED TEST POINT READINGS
To expedite unscheduled maintenance, Table 3.3, Expected Test Point Readings, has been incorporated into this manual. This table, as well as Table 3.2, Troubleshooting, should give a qualified technician a good starting point from which to locate and repair most problems that could occur during the life of the unit. Unless noted, all measurements are to be made at room temperature with the cuffs disconnected, and the unit plugged in. All voltage measurements are with respect to ground and are to be made with the unit on.
18
A.T.S. 1200 TOURNIQUET SYSTEM
Page 21
Table 3.2 Troubleshooting
SYMPTOM POSSIBLE CAUSES
1. Main cuff or Second cuff will not inflate. a) Membrane Panel not properly plugged into P8. b) Tubing inside unit may be pinched or
improperly connected. c) Deflate valve is stuck open. d) Pump not properly plugged into P5. e) Pump’s electrical harness damaged. f) INFLATE button not working. g) Valve’s electrical harness damaged (P6). h) Defective valve driver circuitry.
2. Main cuf
f or Second cuff will not deflate. a) Membrane Panel not properly plugged into P8.
b) DEFLATE button not pressed long enough
(at least 2 seconds). c) Deflate valve is stuck shut. d) DEFLATE button not working. e) Valve’s electrical harness damaged (P6). f) Defective valve driver circuitry.
3. No green AC MAINS indicator light. a) Unit not plugged into wall outlet. b) No Power at wall outlet. c) Mains AC harness not properly plugged into P1. d) Blown fuse(s). e) Membrane Panel not properly plugged into P8. f) Defective AC MAINS indicator. g) Defective AC MAINS indicator circuitry.
4. No flashing orange BATTERY indicator light. a) Unit running on AC. b) Membrane Panel not properly plugged into P8. c) Defective BATTERY indicator. d) Defective BATTERY indicator circuitry.
5. ALARM SILENCE button not working. a) Membrane panel not properly plugged into P8. b) Non-silenceable alarm (System Failure). c) ALARM SILENCE button defective. d) Defective alarm silence circuitry.
6. ALARM indicator light not working. a) Membrane panel not properly plugged into P8. b) Defective ALARM indicator. c) Defective ALARM indicator circuitry.
7. No cuff pressure reading. a) Transducer amplifier not working. b) Internal tubing kinked. c) Transducer tubing on incorrect transducer.
8. Pump will not stop running. a) Leak in internal hose or connector. b) Internal tubing kinked. c) Transducer(s) not working. d) Transducer tubing on incorrect transducer.
9. BATTERY FAIL alarm/message. a) Blown battery fuse (board mounted F1). b) Broken Battery wire harness. c) Dead or depleted battery.
19
A.T.S. 1200 TOURNIQUET SYSTEM
Page 22
10. Backup battery not charging. a) Blown battery fuse (board mounted F1). b) Battery not properly plugged into P7 c) Unit not plugged into wall outlet (verify that the green
AC MAINS indicator is illuminated).
d) Mains AC harness not properly plugged into P1
(verify that the green AC MAINS indicator is illuminated). e) Unit was not permitted to charge for at least 24 hours. f) Defective battery. g) Defective battery charging circuitry.
11. AMP FAIL alarm. a) Transducer(s) amplifier out of range. b) Battery fully depleted or defective. c) Extremely high pressure exerted on transducers.
12. Unit cannot be set to STANDBY. a) Membrane panel not properly plugged into P8. b) Pressure sensed in the Main or Second cuff
(unit will be alarming “CUFF” “ NOT” “DEFL”). c) ON/STANDBY not fully pressed. d) ON/STANDBY button defective.
13. Unit does not turn on. a) Membrane panel not properly plugged into P8. b) ON/STANDBY button defective. c) Blown Fuse(s). d) Unit not plugged in and battery fully depleted.
20
A.T.S. 1200 TOURNIQUET SYSTEM
Page 23
Table 3.3 Expected Test Point Readings
Board Location Nominal-Reading Tolerance Description/Comments
TP1 0 VDC N/A GND – Digital ground
TP2 14 VDC ±1 VDC Main DC supply voltage
TP3 AC Line Voltage N/A AC power supply common
TP7 4.4 VDC ±0.5 VDC Audio amp output resting voltage
TP8 11 to 14 VDC ±1.0 VDC Power ON/STANDBY – unit ON
0 VDC N/A Power ON/STANDBY – unit STANDBY
TP9 5 VDC ±0.5 VDC Digital 5 V power
TP11 5 VDC ±0.5 VDC Supply voltage for U12 and U8
TP12 4.096 VDC ±0.1 VDC Transducer(s) reference voltage
TP13 5 VDC ±0.5 VDC Transducer(s) voltage supply
TP14 12 to 14 VDC 1.0 VDC Battery/AC supply output
TP15 5 VDC ±0.5 VDC Reset – U14 or U7
TP16 11 to 14 VDC ±1.0 VDC Pneumatics control ON/OFF
Y1 PIN 3 32.768 kHz N/A Oscillator square wave
TP21 26.4 to 27.5 VDC ±1.0 VDC Boost supply for battery charger
TP22 5 VDC ±0.5 VDC Supply voltage for battery charger
TP23 5 VDC ±0.5 VDC ON/STANDBY switch = OPEN
0 VDC N/A ON/STANDBY switch = CLOSED
Q27 PIN 1 4 to 5 VDC ±0.5 VDC Power ON control signal to Q10 = ON
0 VDC Power ON control signal to Q10 = STANDBY
NOTE: TEST POINTS AND SCHEMATICS ARE SUBJECT TO CHANGE AND ARE DEPENDENT ON BOARD REVISION LEVEL.
CAUTION: HIGH VOLTAGE ELECTRICAL HAZARD. HIGH VOLTAGE WILL BE PRESENT ON THE POWER INPUT MODULE AND CONTROL BOARD. ALL SERVICE WORK MUST BE COMPLETED BY QUALIFIED TECHNICIANS.
NOTE: If the unit is locked in a failure, not all test voltages will be valid. For instance if the unit is experiencing valve trouble, the pneumatic voltage will not be the normal reading but rather a post failure reading which could be erratic. Above voltages are listed for normal readings because failure readings will likely be unpredictable.
21
A.T.S. 1200 TOURNIQUET SYSTEM
Page 24
3.10 REPLACEMENT PARTS
The following is a list of field replacement parts that can be ordered from Zimmer. To obtain parts or additional information regarding your unit, write or phone:
MAIL: Zimmer Orthopaedic Surgical Products
200 West Ohio Avenue Dover, Ohio 44622 U.S.A.
PHONE: 1-330-343-8801 or 1-800-830-0970
You can also contact your local Zimmer distributor. To ensure prompt service, please include the following information with your order:
Model Number
Serial Number
Description of Part
Part Number (if known)
Quantity Desired
Shipping Address
Shipping Means (if any)
Parts marked with an “*” are commonly used maintenance and/or preventive maintenance parts.
Table 3.4 Parts List
3.11 STORAGE
The A.T.S. 1200 Tourniquet System has an operating range of 50 °F to 100 °F (10 °C to 38 °C). The following are environmental conditions for transportation and storage:
A. Ambient temperature range…………………………………… 1 °F to 149 °F (–17 °C to 65 °C) B. Relative humidity range………………………………………. 10 % to 80 % C. Atmospheric pressure range………………………………….. 500 hPa to 1060 hPa
We strongly recommend that all repairs be done by Zimmer staff.
22
A.T.S. 1200 TOURNIQUET SYSTEM
Zimmer Description Replacement Part Number
0600-1304883 1 A Fuse, 5 x 20 mm* 0600-1501263 Calibration Hose* 0600-2002018 Plumbing Assembly 60-0938-003-00 A.T.S. Stand 5-Wheel Caster Kit 60-2360-001-00 A.T.S. Pole Clamp Knob w/Screw 60-3000-001-00 A.T.S. Cal & Regulator Kit 60-7000-027-00 Backup Battery 60-8000-013-00
A.T.S. 1200 Housing w/EMI Suppression Core 60-8000-014-00 A.T.S. 1200 Feet and Case Screws 60-8000-020-00 A.T.S. 1200 Digital Encoder 60-8000-026-00 A.T.S. 1200 Pole Clamp Assy 60-8000-102-00 A.T.S. 1200 P.C.B. Calibrated & Tested 61-7308-000-00 PLC O-Rings Kit, 25 each 62-1137-001-00 Fuse Drawer, 1/4 in. x 1-1/4 in. 62-1138-001-00 Fuse Drawer, 5 x 20 mm 62-1167-001-00 Power Entry Ground Wire 62-1179-001-00 1 A Time Delay 1-1/4 in. Glass Fuse* 62-1193-001-00 Power Entry Module
Zimmer Description Replacement Part Number
62-1340-001-00 Valve Muffler 62-1711-001-00 12 V Pump 62-1714-001-00 ROTARY Knob 62-1726-001-00 Overpressure Switch 62-2671-001-00 A.T.S. Female Pneumatic Coupling 62-2879-001-00 Male PLC 62-2879-002-00 Female PLC 62-8000-001-00 Pneumatic Manifold 62-8000-004-00 Battery Bracket 62-8000-005-00 Membrane Panel 62-8000-006-00 Clippard Valve (yellow) 62-8000-007-00 Burkert Valve (black) 62-8000-008-00 Overpressure Switch Harness 62-8000-009-00 Battery Harness 62-8000-010-00 Power Cord* 62-8000-015-00 Operator’s and Service Manual 62-8000-018-00 Mains Wire Harness 62-8000-019-00 Cord Retaining Clip*
Page 25
A) Carrying Handle B) PRESSURE ADJUST Activation Buttons C) ALARM SILENCE Button D) Red ALARM Indicator E) TIME/PRESSURE Adjustment Knob F) TIME ADJUST Activation Button G) ON/STANDBY Button H) Green AC MAINS Indicator I) Orange BACKUP BATTERY Indicator
J) SECOND CUFF INFLATE/DEFLATE
Buttons K) Second Cuff Port L) Main Cuff Port M) MAIN CUFF INFLATE/DEFLATE Buttons N) MAIN CUFF INFLATION Indicator O) SECOND CUFF INFLATION Indicator P) TIME Window (displayed in minutes) Q) PRESSURE Window (displayed in mm Hg)
23
1
Page 26
R) Power Cord S) Cord Retaining Clip T) Factory Test Port U) Pole Clamp V) Mains Fuse Block
24
2
Page 27
W) Calibrated Pressure Meter with a range of 0 to 500 mm Hg X) Pressure Regulator/Source adjustable from 0 to 500 mm Hg Y) Calibration Hose Included with Unit
25
3
Page 28
26
Opening the front case
1) Remove 3 screws on top, 2 at side and 2 feet on bottom.
Removing the front case
1) Carefully slide the front off the rear.
Disconnecting the wiring
1) Use caution when disconnecting the wiring. Mark the wires before disconnecting or see Section 3.2 for proper re-connect.
Removing the rear case
1) Remove 2 pole clamp screws on rear and 2 feet on bottom.
4
5
6
7
Page 29
Rear Case separation
1) All components are easily accessed when the rear case is removed. Reverse the process to reassemble.
Be extremely careful not to pinch any wiring or tubing when reassembling!
8
27
Page 30
28
9
Control Board Layout
Page 31
3.12 WARNINGS, CAUTIONS, AND SYMBOL DEFINITIONS (See Illustration 10)
Type BF equipment
Alternating Current (AC)
Protective earth ground
Direct Current (DC)
Refer to instruction manual
Electrical hazard dangerous voltage
Year of manufacture
Replace fuse as marked
Conformity Marking of the Council of the European Community (BSi Product Services, UK)
UL/C-UL Classification mark Medical Equipment with respect to electric shock, fire and mechanical hazards only, in accordance with UL 60601-1, CAN/CSA C22.2 No. 601.1, and IEC 60601-1
This product contains electrical or electronic materials. The presence of these materials may, if not disposed of properly, have potential adverse affects on the environment. Presence of this label on the product means it must not be disposed of in normal household waste and must be disposed of separately. To find out how to properly dispose of this product, please contact your local Zimmer Representative.
29
A.T.S. 1200 TOURNIQUET SYSTEM
This product contains one or more toxic or hazardous substances or elements. The Environmental Protection Use Period on the logo refers to the period in years which toxic or hazardous substances or elements contained in the product will not leak or mutate under normal operating conditions.
Page 32
30
10
Page 33
Revised: 10-07 Zimmer Orthopaedic Surgical Products ©2002, 2007 Zimmer Orthopaedic Surgical Products, Inc. 200 West Ohio Avenue Printed in U.S.A. P.O. Box 10 62-8000-015-00 ENG Dover, Ohio 44622 U.S.A.
Loading...