The compact precision pH meter TruLab pH/ISE 1320 enables you to carry out
pH measurements, ORP measurements and ion selective measurements
quickly and reliably.
The TruLab pH/ISE 1320 provides the maximum degree of operating comfort,
reliability and measuring certainty for all applications.
The USB interface can be used for data transmission to a PC and for software
updates of the meter.
1 Keypad
2 Display
3 Socket field
ba76157e0103/20147
Page 8
OverviewTruLab pH/ISE 1320
Print
er
modul
1
2
3
4
5
pH 1310P
1.2TruLab pH/ISE 1320P meter
The integrated printer of the TruLab pH/ISE 1320 P enables to document measurements according to GLP requirements.
The information concerning the printer of the TruLab pH/ISE 1320
P is given in a separate chapter (see
TRULABPH/ISE 1320 P), page 14).
section 4 PRINTER (ONLY
1 Keypad
2 Display
3 Socket field
4Printer
5 Control panel of the printer
1.3Sensors
A measuring system ready to measure consists of the TruLab pH/ISE 1320
meter and a suitable sensor.
Suitable sensors are pH electrodes, ORP electrodes and ion selective electrodes (ISE).
8ba76157e0103/2014
Page 9
TruLab pH/ISE 1320Safety
2Safety
2.1Safety information
2.1.1Safety information in the operating manual
This operating manual provides important information on the safe operation of
the meter. Read this operating manual thoroughly and make yourself familiar
with the meter before putting it into operation or working with it. The operating
manual must be kept in the vicinity of the meter so you can always find the infor
mation you need.
Important safety instructions are highlighted in this operating manual. They are
indicated by the warning symbol (triangle) in the left column. The signal word
(e.g. "Caution") indicates the level of danger:
WARNING
indicates a possibly dangerous situation that can lead to serious (irreversible) injury or death if the safety instruction is not
followed.
-
CAUTION
indicates a possibly dangerous situation that can lead to
slight (reversible) injury if the safety instruction is not fol
lowed.
NOTE
indicates a possibly dangerous situation where goods might be damaged
if the actions mentioned are not taken.
2.1.2Safety signs on the meter
Note all labels, information signs and safety symbols on the meter and in the
battery compartment. A warning symbol (triangle) without text refers to safety
information in this operating manual.
2.1.3Further documents providing safety information
The following documents provide additional information, which you should
observe for your safety when working with the measuring system:
•Operating manuals of sensors and other accessories
•Safety datasheets of calibration or maintenance accessories (such as buffer
solutions, electrolyte solutions, etc.)
-
ba76157e0103/20149
Page 10
SafetyTruLab pH/ISE 1320
2.2Safe operation
2.2.1Authorized use
This meter is authorized exclusively for pH, ORP and ISE concentrations
measurements in the laboratory.
Only the operation and running of the meter according to the instructions and
technical specifications given in this operating manual is authorized (see
section 14 TECHNICALDATA, page 94).
Any other use is considered unauthorized.
2.2.2Requirements for safe operation
Note the following points for safe operation:
•The meter may only be operated according to the authorized use specified
above.
•The meter may only be supplied with power by the energy sources
mentioned in this operating manual.
•The meter may only be operated under the environmental conditions
mentioned in this operating manual.
•The meter may only be opened if this is explicitly described in this operating
manual (example: Inserting the batteries).
2.2.3Unauthorized use
The meter must not be put into operation if:
•it is visibly damaged (e.g. after being transported)
•it was stored under adverse conditions for a lengthy period of time (storing
conditions, see
section 14 TECHNICALDATA, page 94).
10ba76157e0103/2014
Page 11
TruLab pH/ISE 1320Commissioning
3Commissioning
3.1Scope of delivery
 pH meter TruLab pH/ISE 1320
 4 batteries 1.5 V Mignon type AA
 Power pack
 USB cable (A plug on mini B plug)
 Stand
 Stand holder
 Comprehensive operating manual
 Short instructions
 CD-ROM with
– USB drivers
– detailed operating manual
– Software MultiLab Importer
3.2Power supply
The TruLab pH/ISE 1320 is supplied with power in the following ways:
 Mains operation with the supplied power pack
 Battery operation (4 x alkaline manganese batteries, type AA)
 USB operation via a connected USB-B cable
3.3Initial commissioning
Perform the following activities:
 Insert the supplied batteries
 For mains operation: Connect the power pack
 If necessary, mount a stand
 Switch on the meter
(see section 5.2 SWITCHINGONTHEMETER, page 19)
 Set the date and time
(see section 5.4.6 EXAMPLE 2 ONNAVIGATION: SETTINGTHEDATEANDTIME,
page 25)
ba76157e0103/201411
Page 12
CommissioningTruLab pH/ISE 1320
1
3.3.1Inserting the batteries
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
1 Battery compartment
1.Open the battery compartment (1) on the underside of the meter.
CAUTION
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the ± signs in
the battery compartment.
2.Place four batteries (type Mignon AA) in the battery compartment.
3.Close the battery compartment (1).
4.Set the date and time
(see section 5.4.6 EXAMPLE 2 ONNAVIGATION: SETTINGTHEDATEAND
TIME, page 25).
12ba76157e0103/2014
Page 13
TruLab pH/ISE 1320Commissioning
1
2
3.3.2Connecting the power pack
CAUTION
The line voltage at the operating site must lie within the input
voltage range of the original power pack (see
NICALDATA, page 94).
section 14 TECH-
CAUTION
Use original power packs only (see section 14 TECHNICALDATA,
page 94).
1.Connect the plug of the power pack to the socket for the power pack on
the TruLab
2.Connect the original power pack to an easily accessible power outlet.
pH/ISE 1320.
3.3.3Mounting the stand
The stand base can be mounted at the right side of the meter.
4.1Commissioning / switching the printer on or off
1.Connect the power pack to the TruLab pH/ISE 1320 P.
The LED lights up green. The printer is ready for operation.
or
When the printer had been switched off (LED flashing):
Switch on the printer with <ON/OFF>.
The LED lights up green. The printer is ready for operation.
If there is a USB-B connection at the same time (e.g. to a PC), the
data are output to the PC only.
Data are only output to the printer if the following requirements are met
 Data are transmitted manually or automatically (see section 11.1 OPTIONS
FORDATATRANSMISSION, page 84)
 The printer is switched on (LED illuminated)
 There is no USB connection.
4.3Printer settings
To open the Storage & config menu, press the <ENTER_> key in the measured
value display. After completing the settings, switch to the measured value dis
play with <M>.
Default settings are printed in bold.
-
Menu itemPossible set-
Description
ting
System / Printer /
Font size
12x20
8x16
7x16
Select the font size for the printer
A print sample (font of the printer) to
view the available font sizes can be
output with <ON/OFF_>.
System /
Printer /
Print test page
-The printer prints the meter informa-
tion from the menu, System / Ser-vice information.
The current printer settings are
used for the printout.
4.4Maintenance
4.4.1Changing the roll of paper (thermal paper)
1.Pull the lever (3) until the lid (4) of the paper compartment opens.
2.If necessary, remove any remains of the old roll of paper.
3.Insert the new roll of paper. The start of the paper should poke out of the
paper compartment.
4.Close the lid (4) of the paper compartment so that it clicks into place.
5.If necessary, move the printer paper forward with <FEED>.
Opens the menu for measurement settings / confirms
entries
Opens the menu for system settings
<AR>Freezes the measured value (HOLD function)
Switches the AutoRead measurement on or off
ba76157e0103/201417
Page 18
OperationTruLab pH/ISE 1320
HOLD AR
25.0
6.093
pH
°C
4
3
2
6
7
8
9
1
AutoCal
5
12.12.2012
12:12
USB output
YSI
5.1.2Display
1 Status information (sensor)
2 Measured value (with unit)
3 Measured parameter
4 Continuous measurement control (CMC function)
5 Channel display: Plug position of the sensor
6 Sensor symbol (calibration evaluation, calibration interval)
7 Measured temperature (with unit)
8 Status information (meter)
9 Softkeys and date + time
5.1.3Status information (meter)
ARStability control (AutoRead) is active
HOLDMeasured value is frozen (<AR> key)
Batteries are almost empty
18ba76157e0103/2014
Page 19
TruLab pH/ISE 1320Operation
1
2
3
7
8
9
4
5
6
5.1.4Socket field
1 pH electrode or ion selective electrode 1
2 Reference electrode 1
3 Temperature sensor 1
4 pH electrode or ion selective electrode 2
5 Reference electrode 2
6 Temperature sensor 2
7 Mini USB-B interface
8 Power pack
9 Service interface
CAUTION
Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current
circuit with current limiting).
Almost all customary sensors fulfill these conditions.
5.2Switching on the meter
1.Switch the meter on with <On/Off>.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test is being
performed.
The measured value display appears.
ba76157e0103/201419
Page 20
OperationTruLab pH/ISE 1320
pH
6.949
25.0
°C
12.12.2012
12:12
USB output
5.3Switching off
1.Switch the meter off with <On/Off>.
Automatic switch-offThe instrument has an automatic switch-off function in order to save the
batteries (see
switches off the measuring instrument if no key is pressed for an adjustable
period.
The automatic switchoff is not active
 if the communication cable is connected
 if the Automatic data storage function is active, or with automatic data trans-
mission
 if the power pack is connected
Display illuminationThe meter automatically switches off the display illumination if no key is
pressed for 30
again.
You can also generally switch the display illumination on or off (see section
9.3.1 SYSTEM, page 75).
section 9.3.1 SYSTEM, page 75). The automatic switchoff
seconds. The illumination is switched on with the next keystroke
5.4Navigation
The principles of navigation in menus and dialogs are explained in the following
sections.
5.4.1Operating modes
The meter has the following operating modes:
20ba76157e0103/2014
Page 21
TruLab pH/ISE 1320Operation
Operating
Description
mode
MeasuringThe measurement data of the connected sensor are shown
in the measured value display
CalibrationThe course of a calibration with calibration information, func-
tions and settings is displayed
Storage in
The meter stores measuring data automatically or manually
memory
Transmitting data
The meter transmits measuring data and calibration records
to a USB-B interface automatically or manually
SettingThe system menu or a sensor menu with submenus, set-
tings and functions is displayed
Only those displays and functions are available in the active operating mode
that are currently being required.
5.4.2Measured value display
In the measured value display, open the setting menus with <MENU/ENTER>.
The current functions of the softkeys are shown on the display.
 Use <MENU/ENTER> (short pressure) to open the menu for calibration and
measurement settings for the displayed measured parameter.
 Use <MENU/ENTER__> (long keystroke (approx. 2 s) to open the Storage
& config menu with the sensor-independent settings.
Use the keys of the keypad to carry out further functions such as storage or calibration (see section 5.1.1 KEYPAD, page 17). These functions are not available
in other operating situations.
5.4.3Menus and dialogs
The menus for settings and dialogs in procedures contain further submenus.
 To select a subelement, use the <><> keys. The current selection is dis-
played with a frame.
 To make further settings, switch to the next higher menu level with
<F1>[Back].
 Use <M> to return to the measured value display.
5.4.4Elements in menus and dialogs
 Submenus
The name of the submenu is displayed at the upper edge of the frame. Submenus are opened by confirming with <MENU/ENTER>. Example:
Calibration record
Calibration data storage
Serial number (sensor)
Buffer
:TEC
One point calibration:Yes
Calibration interval
:7 d
Unit for slope
:mV/pH
i
2.00 4.01 7.00 10.01
pH
12.12.2012
12:12
Back
 Settings
Settings are indicated by a colon. The current setting is displayed on the
right-hand side. The setting mode is opened with <MENU/ENTER>. Subse
quently, the setting can be changed with <><> and <MENU/ENTER>.
Example:
-
 Functions
 Messages
22ba76157e0103/2014
Functions are designated by the name of the function. They are immediately
carried out by confirming with <MENU/ENTER>.
Example: Display the Calibration record function.
Information is marked by the i symbol. They cannot be selected. Example:
Page 23
TruLab pH/ISE 1320Operation
Calibration record
Calibration data storage
Serial number (sensor)
Buffer
:TEC
One point calibration
:Yes
Calibration interval:7 d
Unit for slope
:mV/pH
i
2.00 4.01 7.00 10.01
pH
12.12.2012
12:12
Back
pH
6.949
25.0
°C
12.12.2012
12:12
USB output
System
Data storage
Storage & config
12.12.2012
12:12
Back
5.4.5Example 1 on navigation: Setting the language
1.Press the <On/Off> key.
The measured value display appears.
The instrument is in the measuring mode.
ba76157e0103/201423
2.Using <MENU/ENTER__>, open the Storage & config menu.
The instrument is in the setting mode.
3.Select the System submenu with <><>.
The current selection is displayed with a frame.
4.Open the System submenu with <MENU/ENTER>.
Page 24
OperationTruLab pH/ISE 1320
General
Interface
Clock
Service information
Reset
System
12.12.2012
12:12
Back
Language:Deutsch
Beep:Off
Illumination
:On
Contrast
:50 %
Switchoff time:1 h
Temperature unit
:°C
Stability control
:On
General
12.12.2012
12:12
Back
Language:Deutsch
Beep
:Off
Illumination
:On
Contrast
:50 %
Switchoff time
:1 h
Temperature unit
:°C
Stability control:On
General
12.12.2012
12:12
Back
5.Select the General submenu with <><>.
The current selection is displayed with a frame.
6.Open the Generalsubmenu with <MENU/ENTER>.
24ba76157e0103/2014
7.Open the setting mode for the Language with <MENU/
ENTER>.
8.Select the required language with <><>.
9.Confirm the setting with <MENU/ENTER>.
The meter switches to the measuring mode.
The selected language is active.
Page 25
TruLab pH/ISE 1320Operation
Date format:dd.mm.yyyy
Date
:12.12.2012
Time
:12:12:25
Clock
12.12.2012
12:12
Back
5.4.6Example 2 on navigation: Setting the date and time
The measuring instrument has a clock with a date function. The date and time
are shown in the measured value display.
When storing measured values and calibrating, the current date and time are
automatically stored as well.
The correct setting of the date and time and date format is important for the following functions and displays:
 Current date and time
 Calibration date
 Identification of stored measured values.
Therefore, check the time at regular intervals.
After a fall of the supply voltage (empty batteries), the date and time
are reset.
The date format can be switched from the display of day, month, year
(dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or mm.dd.yyyy).
1.In the measured value display:
Using <MENU/ENTER__>, open the Storage & config menu.
The instrument is in the setting mode.
2.Select and confirm the System / Clock menu with <><> and
<MENU/ENTER>.
The setting menu for the date and time opens up.
3.Select and confirm the Time menu with <><> and <MENU/ENTER>.
The hours are highlighted.
ba76157e0103/201425
4.Change and confirm the setting with <><> and <MENU/ENTER>.
The minutes are highlighted.
5.Change and confirm the setting with <><> and <MENU/ENTER>.
The seconds are highlighted.
6.Change and confirm the setting with <><> and <MENU/ENTER>.
The time is set.
Page 26
OperationTruLab pH/ISE 1320
7.If necessary, set the Date and Date format. The setting is made similarly
to that of the time.
8.To make further settings, switch to the next higher menu level with
[Back]<F1>.
or
Switch to the measured value display with <M>.
The instrument is in the measuring mode.
26ba76157e0103/2014
Page 27
TruLab pH/ISE 1320Operation
Channel display
Display of the plug position
for the respective parameter.
The filled bar indicates for
each connected sensor to
which plug position of the
meter it is connected.
pH
6.949
24.8 °C
12.12.2012
12:12
USB output
pH
6.966
25.0
°C
12.12.2012
12:12
USB output
5.5Channel display
The TruLab pH/ISE 1320 manages the connected sensors and displays which
sensor is plugged to which connection.
5.5.1Display of several sensors in the measuring mode
The measured values of the connected sensors can be displayed in the following ways:
 Clear display of all connected sensors
 Detailed display of one sensor
(e.g. incl. CMC feature with pH sensors)
With the softkey you can very easily switch between the two display types. The
suitable softkey is displayed depending on the operating situation.
ba76157e0103/201427
Page 28
pH valueTruLab pH/ISE 1320
pH
6.949
24.8 °C
12.12.2012
12:12
USB output
6pH value
6.1Measuring
6.1.1Measuring the pH value
NOTE
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as the values would be incorrect.
The USB interface is not galvanically isolated.
To ensure the high measurement accuracy of the measuring system, always measure with a calibrated electrode (see section 6.2
CALIBRATION, page 31).
1.Connect the pH electrode to the meter.
2.If necessary, select the measured parameter with <M>.
3.When measuring without temperature sensor:
Temper the test sample or measure the current temperature and enter it
with <>
4.Immerse the pH electrode in the test sample.
The measured value is checked for stability (automatic stability control).
The display of the measured parameter flashes.
5.Wait for a stable measured value.
The display of the measured parameter no longer flashes.
<>.
Stability control
(AutoRead)
& HOLD function
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc
ibility of measured values.
The measured parameter flashes on the display
 as soon as the measured value is outside the stability range
 when the automatic Stability control is switched off.
You can activate or switch off the automatic Stability control function (see
section 9.3.1 SYSTEM, page 75).
28ba76157e0103/2014
-
Page 29
TruLab pH/ISE 1320pH value
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the Stability control function and the HOLD func-
tion with <AR> or <M> at any time.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured value is recognized. The progress bar disappears and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
Criteria for a stable
measured value
You can prematurely terminate the Stability control function manually with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
Time intervalStability in the time interval
parameter
pH value15 seconds∆ : better than 0.01 pH
Temperature15 seconds∆ : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
ba76157e0103/201429
Page 30
pH valueTruLab pH/ISE 1320
6.1.2Measuring the temperature
For reproducible pH measurements, it is essential to measure the temperature
of the test sample.
You have the following options to measure the temperature:
 Automatic measurement of the temperature with the temperature sensor
30 or Pt 1000) integrated in the sensor.
(NTC
 Measurement by an external temperature sensor.
 Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
Temperatur
e sensor
Resolution of the
temp. display
Temp. measurement
yes0.1 °CAutomatic with temperature sensor
-1 °CManual
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
1.Measure the current temperature of the test sample.
2.Set the temperature value with <><>.
or
In the <MENU/ENTER>/pH/Man. temperature menu, set the tempera-
ture value with <><>.
30ba76157e0103/2014
Page 31
TruLab pH/ISE 1320pH value
6.2Calibration
6.2.1Why calibrate?
pH electrodes age. This changes the zero point (asymmetry) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope of the
electrode.
Thus, you should calibrate at regular intervals.
6.2.2When do you have to calibrate?
 After connecting a sensor
 Routinely within the framework of the company quality assurance
 When the calibration interval has expired
6.2.3Automatic calibration (AutoCal)
Make sure that in the sensor menu, Buffer menu, the buffer set is correctly
selected (see
9.1.1 SETTINGSFORPH MEASUREMENTS, PAGE 66).
Use any one to five buffer solutions of the selected buffer set in ascending or
descending order.
Below, calibration with Technical buffers (TEC) is described. When other buffer
sets are used, other nominal buffer values are displayed. Apart from that, the
procedure is identical.
If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed.
1.Connect the pH electrode to the meter.
2.Keep the buffer solutions ready.
When measuring without temperature sensor:
Temper the buffer solutions or measure the current temperature.
3.In the measured value display, select the measured parameter pH or mV
with <M>.
4.Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).
ba76157e0103/201431
Page 32
pH valueTruLab pH/ISE 1320
pH1
-180.0
24.8 °C
AutoCal TEC
12.12.2012
12:12
Buffer
mV
pH
-180.0
24.8 °C
AutoCal TEC
10.011
12.12.2012
12:12
AR
Buffer
mV
5.Thoroughly rinse the electrode with deionized water.
6.Immerse the electrode in the first buffer solution.
7.When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
8.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
9.Wait for the end of the measurement with stability control or accept the
10.If necessary, finish the calibration procedure as a single-point calibration
Continuing with two-
point calibration
32ba76157e0103/2014
11.Thoroughly rinse the electrode with deionized water.
12.Immerse the electrode in the second buffer solution.
13.When measuring without temperature sensor:
calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2
mV/pH at 25 °C) and determines the zero point of the elec-
trode.
Enter the temperature of the buffer with <><>.
Page 33
TruLab pH/ISE 1320pH value
pH
0.0
24.8 °C
AutoCal TEC
7.000
12.12.2012
12:12
AR
Buffer
mV
pH
180.0
24.8 °C
AutoCal TEC
4.010
12.12.2012
12:12
AR
Buffer
mV
14.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
15.Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
Continuing with
three- to five-point
calibration
16.If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.
17.Thoroughly rinse the electrode with deionized water.
18.Immerse the electrode in the next buffer solution.
19.When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
20.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
ba76157e0103/201433
Page 34
pH valueTruLab pH/ISE 1320
pH
0.0
24.8 °C
ConCal
1
12.12.2012
12:12
Buffer
mV
21.Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
22.If necessary, use <M> to finish calibration or
Continue calibrating using the next buffer with <ENTER>.
Calibration is automatically completed after the last buffer of a buffer set has been measured. Then the calibration record is displayed.
The calibration line is determined by linear regression.
6.2.4Manual calibration (ConCal)
Make sure that in the sensor menu, Buffer menu, the ConCal buffer set is
selected (see
Use any one to five buffer solutions in ascending or descending order.
If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed.
1.Connect the pH electrode to the meter.
The pH measuring window is displayed.
2.Keep the buffer solutions ready.
When measuring without temperature sensor:
Temper the buffer solutions or measure the current temperature.
3.In the measured value display, select the measured parameter pH or mV
with <M>.
4.Start the calibration with <CAL>.
The calibration display appears.
34ba76157e0103/2014
5.Thoroughly rinse the electrode with deionized water.
6.Immerse the electrode in the first buffer solution.
Page 35
TruLab pH/ISE 1320pH value
pH
0.0
24.8 °C
ConCal
7.000
12.12.2012
12:12
Buffer
mV
AR
pH
7.000
24.8 °C
ConCal
1
12.12.2012
12:12
Buffer
7.When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
8.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
9.Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the setting of the nominal buffer value
appears.
10.Set the nominal buffer value for the measured temperature with
<><>.
11.Accept the set calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
12.If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2
mV/pH at 25 °C) and determines the zero point of the elec-
Continuing with two-
point calibration
ba76157e0103/201435
13.Thoroughly rinse the electrode with deionized water.
trode.
Page 36
pH valueTruLab pH/ISE 1320
pH
4.035
24.8 °C
ConCal
2
12.12.2012
12:12
Buffer
14.Immerse the electrode in the second buffer solution.
15.When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
16.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
17.Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the setting of the nominal buffer value
appears.
Continuing with
three- to five-point
calibration
18.Set the nominal buffer value for the measured temperature with
<><>.
19.Accept the set calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
20.Finish the calibration procedure as a two-point calibration with <M>.
The calibration record is displayed.
21.Thoroughly rinse the electrode with deionized water.
22.Immerse the electrode in the next buffer solution.
23.When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
24.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
25.Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the setting of the nominal buffer value
appears.
36ba76157e0103/2014
Page 37
TruLab pH/ISE 1320pH value
pH
9.958
24.8 °C
ConCal
3
12.12.2012
12:12
Buffer
26.Set the nominal buffer value for the measured temperature with
<><>.
27.Accept the set calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
28.Use <M> to finish calibration or
Continue calibrating using the next buffer with <ENTER>.
After the fifth buffer has been measured the calibration is automatically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
6.2.5Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
Values to be
determined
1-pointAsy Zero point = Asy
2-pointAsy
Slp.
3-point to
5-point
Asy
Slp.
Displayed calibration data
 Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
 Zero point = Asy
 Slope = Slp.
 Zero point = Asy
 Slope = Slp.
The calibration line is calculated by linear regression.
You can display the slope in the units, mV/pH or % (see section
ba76157e0103/201437
9.1.1 SETTINGSFORPH MEASUREMENTS, page 66).
Page 38
pH valueTruLab pH/ISE 1320
6.2.6Calibration data
The calibration record is automatically transmitted to the interface
after calibrating.
Displaying the
calibration data
Displaying the
calibration data
storage
The calibration record of the last calibration is to be found under the menu item,
<MENU/ENTER> / Calibration / / Calibration record. To open it, press the
<CAL__> key in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with <F2>[USB output].
The calibration records of the last calibrations (up to 10) are given in the menu,
<MENU/ENTER>/Calibration / Calibration data storage.
Menu itemSetting/func-
Description
tion
Calibration /
-Displays the calibration record.
Calibration data storage /Display
Further options:
 Scroll through the calibration
records with <><>.
 Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].
 Quit the display with <F1>/
[Back] or <MENU/ENTER>.
Calibration
evaluation
 Switch directly to the mea-
sured value display with <M>.
Calibration /
Calibration data stor-
-Outputs the calibration records to
the interface.
age /
Output to USB
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
38ba76157e0103/2014
Page 39
TruLab pH/ISE 1320pH value
TruLab pH/ISE 1320
Ser. no. 11292113
CALIBRATIONpH
12.12.2012 15:55
Ser. no. 10501234
AutoCal TEC
Buffer 14.01
Buffer 27.00
Buffer 310.01
Voltage 1184.0 mV
Voltage 23.0 mV
Voltage 3-177.0 mV
Temperature 124.0 °C
Temperature 224.0 °C
Temperature 324.0 °C
Slope-60.2 mV/pH
Asymmetry4.0 mV
Sensor +++
DisplayCalibration
record
+++
++
+
-
Clean the electrode according to the electrode operating manual
ErrorError
Error elimination (see section 13 WHATTO
DOIF..., page 89)
Zero point
[mV]
Slope
[mV/pH]
- 15 ... + 15-60.5 ... -58.0
- 20 ... <-15
>-58.0 ... -57.0
or
>+15 ... + 20
- 25 ... <- 20
or
>+ 20 ... + 25
- 30 ... <- 25
or
->+25 ... + 30
-61.0 ... <-60.5
or
>-57.0 ... -56.0
-62.0 ... <-61.0
or
>-56.0 ... -50.0
- 30 ... + 30-62.0 ... -50.0
Calibration record
(example)
ba76157e0103/201439
Page 40
pH valueTruLab pH/ISE 1320
25.0
6.093
1
2
3
4
pH
°C
0
14
12.12.2012
12:12
USB output
6.2.7Continuous measurement control (CMC function)
The Continuous Measurement Control (CMC function) facilitates to evaluate
the current measured value instantly and definitely.
After each successful calibration the scale of the pH measuring range is
displayed in the measured value display. Here you can very clearly see
whether or not the current measured value is in the calibrated part of the
measuring range.
The following information is displayed:
1 Measuring range for which a valid calibration is available (white).
Measured values in this range are suitable for documentation.
2 Measuring range for which no valid calibration is available (shaded).
Measured values in this range are not suitable for documentation. If
necessary, calibrate the meter with buffers covering this measuring
range.
If the current measured value is outside the calibrated range, this
area is shaded stronger.
If a measured value is outside the measuring range pH 0 - 14, overflow arrows are displayed at the left or right edge of the measuring
range.
3 Currently measured pH value (needle)
4 Marking lines for all nominal buffer values used with the last valid cali-
bration
The limits of the calibrated range are determined by the buffers used for calibration:
Lower limit: Buffer with lowest pH value - 2 pH units
Upper limit: Buffer with highest pH value + 2 pH units
40ba76157e0103/2014
Page 41
TruLab pH/ISE 1320ORP voltage
U
157.0 mV
24.8 °C
12.12.2012
12:12
USB output
7ORP voltage
7.1Measuring
7.1.1Measuring the ORP
NOTE
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as the values would be incorrect.
The USB interface is not galvanically isolated.
1.Connect the ORP electrode to the meter.
2.If necessary, select the U (mV) display with <M>.
3.When measuring without temperature sensor:
Measure the temperature of the test sample and enter it with <> <>.
4.Rinse the ORP electrode and immerse it in the test sample.
The measured value is checked for stability (automatic stability control).
The display of the measured parameter flashes.
Stability control
(AutoRead )
5.Wait for a stable measured value.
The display of the measured parameter no longer flashes.
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc
ibility of measured values.
You can activate or switch off the automatic Stability control function (see
section 9.3.1 SYSTEM, page 75).
The measured parameter flashes on the display
-
 as soon as the measured value is outside the stability range
 when you switch over between the measured parameters with <M>.
 when the automatic Stability control is switched off.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
ba76157e0103/201441
Page 42
ORP voltageTruLab pH/ISE 1320
Manually starting
the stability control
Measured parame-
Time intervalStability in the time interval
ter
ORP voltage15 seconds∆ : better than 0.3 mV
Temperature15 seconds∆ : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
You can start the Stability control manually at any time, irrespective of the setting for automatic Stability control (see section 9.3.1 SYSTEM, page 75) in the
System menu.
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the mea
sured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears and the display of the
measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
Freezing the
measured value
(HOLD function)
You can prematurely terminate the Stability control function manually with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.Using <ENTER>, start a further measurement with Stability control.
or
Release the frozen measured value again with <AR>.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
With the HOLD function, you can freeze the current measured value. The displayed measured value stops changing until you switch the HOLD function off.
If the HOLD function is active, you can, e.g. start a manual measurement with stability control.
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
42ba76157e0103/2014
Page 43
TruLab pH/ISE 1320ORP voltage
ΔU
13.5 mV
24.8 °C
12.12.2012
12:12
USB output
2.Release the frozen measured value again with <AR>.
The HOLD function is switched off.
The [HOLD] status display disappears.
7.1.2Measuring the relative ORP
To measure the difference of the ORPs of two solutions, you have to define the
ORP of one solution as the zero point first.
pH and ORP electrodes can be used to determine the relative
ORP.
1.Connect the ORP electrode to the meter.
2.Prepare the reference solution for the determination of the reference
point.
3.Select the ΔU (mV display with <M>.
4.Immerse the ORP electrode in the reference solution.
5.Display the potential of the current zero point with <CAL>.
6.Press <ENTER> to measure the reference solution.
The measured value is checked for stability (automatic stability control). The display of the measured parameter flashes.
The measured potential is defined as the zero point.
or
Press <M> to terminate the display of the zero point.
7.Rinse the ORP electrode and immerse it in the test sample.
The measured value is checked for stability (automatic stability control). The display of the measured parameter flashes.
8.Wait for a stable measured value.
The display of the measured parameter no longer flashes.
ba76157e0103/201443
AutoRead The Stability control function checks whether the measured values are stable
within the monitored time interval.
Page 44
ORP voltageTruLab pH/ISE 1320
Measured parame-
Time intervalStability in the time interval
ter
ORP15 seconds∆ : better than 0.3 mV
Temperature15 seconds∆ : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
7.1.3Measuring the temperature
For reproducible ORP measurements, it is essential to measure the temperature of the test sample.
You have the following options to measure the temperature:
 Automatic measurement of the temperature with the temperature sensor
(NTC
30 or Pt 1000) integrated in electrode.
 Measurement by an external temperature sensor.
 Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
Temperature
sensor
Resolution of the
temp. display
Temp. measurement
yes0.1 °CAutomatic with temperature sensor
-1 °CManual
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
1.Measure the current temperature of the test sample.
2.Set the temperature value with <><>.
or
In the <MENU/ENTER>/U/Man. temperature menu, set the tempera-
ture value with <><>.
7.2ORP calibration
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.
44ba76157e0103/2014
Page 45
TruLab pH/ISE 1320Ion concentration
ISE
0.157 mg/l
25.0
°C
Ag
12.12.2012
12:12
USB output
8Ion concentration
8.1Measuring
8.1.1Measuring the ion concentration
NOTE
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as the values would be incorrect.
The USB interface is not galvanically isolated.
Incorrect calibration of ion selective electrodes will result in incorrect measured values. Calibrate regularly before measuring.
For precise ISE measurements the temperature difference
between measurement and calibration should not be greater that
2
K. Therefore, adjust the temperature of the standard and
measuring solutions accordingly. If the temperature difference is
greater the [TpErr] warning appears in the measured value display.
1.Connect the ISE combination electrode to the meter.
The pH/ISE measuring window is displayed.
2.If necessary, select the ISE display (unit, mg/l) with <M>.
3.If necessary, measure the temperature of the test sample with a thermometer.
4.Calibrate or check the meter with the electrode.
While no valid calibration is available, e.g. in the delivery condition,
"Error" appears in the measured value display.
5.Immerse the electrode in the test sample.
Stability control
ba76157e0103/201445
(AutoRead)
& HOLD function
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc
ibility of measured values.
The measured parameter flashes on the display
-
Page 46
Ion concentrationTruLab pH/ISE 1320
 as soon as the measured value is outside the stability range
 when the automatic Stability control is switched off.
You can activate or switch off the automatic Stability control function (see
section 9.1.1 SETTINGSFORPH MEASUREMENTS, page 66).
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the Stability control function and the HOLD func-
tion with <AR> or <M> at any time.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured value is recognized. The progress bar disappears and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manually with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
CriteriaThe AutoRead criteria affect the reproducibility of the measured values. The
following criteria can be adjusted:
 high: highest reproducibility
 medium: medium reproducibility
 low: lowest reproducibility
Increasing reproducibility also causes the response time to
increase until a measured value is evaluated as stable.
46ba76157e0103/2014
Page 47
TruLab pH/ISE 1320Ion concentration
8.1.2Measuring the temperature
For reproducible ion-selective measurements, it is essential to measure the
temperature of the test sample.
You have the following options to measure the temperature:
 Measurement by an external temperature sensor.
 Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
Temperatur
e sensor
Resolution of the
temp. display
Temp. measurement
yes0.1 °CAutomatic with temperature sensor
-1 °CManual
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
1.Measure the current temperature of the test sample.
2.Set the temperature value with <><>.
or
In the <MENU/ENTER>/ISE/Man. temperature menu, set the tempera-
ture value with <><>.
ba76157e0103/201447
Page 48
Ion concentrationTruLab pH/ISE 1320
ISE
0.0
24.8 °C
Ag 0.010mg/l
1
12.12.2012
12:12
Standard
mV
8.2Calibration
8.2.1Why calibrate?
Ion-selective electrodes age and are temperature-dependent. This changes
the slope. As a result, an inexact measured value is displayed. Calibration
determines the calibration line of the electrode and stores this value in the
meter.
Thus, you should calibrate before each measurement and at regular intervals.
8.2.2When to calibrate?
 Before any measurement if possible
 After connecting another ISE electrode
 When the sensor symbol flashes, e.g. after a voltage interruption (empty
batteries)
8.2.3Calibration (ISE Cal)
ISE Cal is the conventional two-point to seven-point calibration procedure
that uses 2 to 7 freely selectable standard solutions. The concentration
expected for the measurement determines the concentration of the calibration
standards.
1.Connect the ISE combination electrode to the meter.
The pH/mV/ISE measuring window is displayed.
2.Keep the standard solutions ready.
3.If necessary, measure the temperature of the standard solutions with a
thermometer.
4.If necessary, change the unit of the measurement result and calibration
standards in the ISE setup/Unit menu.
5.In the measured value display, select the ISE measuring window with
<> and <M>.
<>
6.Start the calibration with <CAL>.
The calibration display appears.
48ba76157e0103/2014
Page 49
TruLab pH/ISE 1320Ion concentration
ISE
0.0
24.8 °C
12.12.2012
12:12
0.01 mg/
mV
AR
ISE
0.0
24.8 °C
Ag 0.020mg/l
2
12.12.2012
12:12
Standard
mV
7.Thoroughly rinse the electrode with distilled water.
8.Immerse the electrode in standard solution 1.
9.When calibrating without temperature sensor:
 Measure the temperature of the standard solution using a thermom-
eter.
 Use <F2>/[ T ⇅ C ] to select the setting of the temperature.
 Use <> <> to set the temperature.
 Use <F2>/[ T ⇅ C ] to select the setting of the concentration.
10.Set the concentration of the standard solution with <> <> and
press <MENU/ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
Continuing with two-
point calibration
11.Wait for the end of the AutoRead measurement or accept the calibration value with <MENU/ENTER>.
The calibration display for the next standard solution appears.
12.Thoroughly rinse the electrode with distilled water.
13.Immerse the electrode in standard solution 2.
ba76157e0103/201449
Page 50
Ion concentrationTruLab pH/ISE 1320
ISE
8.9
24.8 °C
12.12.2012
12:12
0.01 m
mV
AR
ISE
20.5
24.8 °C
Ag 0.050mg/l
3
12.12.2012
12:12
Standard
mV
14.When calibrating without temperature sensor:
 Measure the temperature of the standard solution using a thermom-
eter.
 Use <F2>/[ T ⇅ C ] to select the setting of the temperature.
 Use <> <> to set the temperature.
 Use <F2>/[ T ⇅ C ] to select the setting of the concentration.
15.Set the concentration of the standard solution with <> <> and
press <MENU/ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
Continuing with
three- to seven-point
calibration
16.Wait for the end of the AutoRead measurement or accept the calibration value with <MENU/ENTER>.
The calibration display for the next standard solution appears.
17.Press <MENU/ENTER> to continue with three-point calibration.
or
Finish the calibration procedure as a two-point calibration with <M>.
The new calibration values are displayed.
Repeat the steps 8 to 13 in the same way with the third and further standard
solutions as necessary. The new calibration values are displayed after the last
calibration step was completed.
50ba76157e0103/2014
Based on the calibration data, the calibration curve is determined
in sections, according to the Nernst equation modified by Nikolski.
Page 51
TruLab pH/ISE 1320Ion concentration
8.2.4Calibration standards
Use two to seven different standard solutions. The standard solutions have to
be selected in either increasing or decreasing order.
Select the unit of the standard solution and measurement result in
the ISE setup/Unit menu.
Standard solution (Std 1 - 7)Values
Unit [mg/l]0.010 ... 500,000
Unit [mol/l]0.100 ... 5,000 µmol/l
10.00 ... 5,000 mmol/l
Unit [mg/kg]0.010 ... 500,000
Unit [ppm]0.010 ... 500,000
Unit [%]0.001 ... 50,000
Displaying the
calibration data
Displaying the
calibration data
storage
The measurement precision is also dependent on the selected
standard solutions. Therefore, the selected standard solutions
should cover the value range expected of the subsequent concen
-
tration measurement.
If the measured electrode potential is outside the calibrated range,
the [ISEErr] warning is displayed.
8.2.5Calibration data
The calibration record of the last calibration is to be found under the menu item,
<MENU/ENTER> / Calibration / / Calibration record. To open it, press the
<CAL__> key in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with <F2>[USB output].
The calibration records of the last calibrations (up to 10) are given in the menu,
<MENU/ENTER>/ ISE / Calibration / Calibration data storage.
ba76157e0103/201451
Page 52
Ion concentrationTruLab pH/ISE 1320
TruLab pH/ISE 1320
Ser. no. 12345678
CALIBRATION ISE
18.01.2013 08:09:10
Sensor
Ser. no. 12345678
Standard 10.010 mg/l
Standard 20.020 mg/l
Voltage 138.5 mV
Voltage 258.0 mV
Temperature 124.0 øC
Temperature 224.0 øC
Ion typeAg
Slope64.7 mV
Sensor+++
_____________________________________
Menu itemSetting/func-
tion
Calibration /
-Displays the calibration record.
Calibration data storage /Display
Calibration /
-Outputs the calibration records to
Calibration data storage /
Output to USB
Description
Further options:
 Scroll through the calibration
records with <><>.
 Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].
 Quit the display with <F1>/
[Back] or <MENU/ENTER>.
 Switch directly to the mea-
sured value display with <M>.
the interface.
Calibration record
Calibration
evaluation
(example)
After calibrating, the meter automatically evaluates the calibration.
DisplayCalibration
Magnitude of the slope [mV]
record
+++50.0 ... 70.0 or 25.0 ... 35.0
ErrorError< 50 or > 70
Error elimination (see section 13 WHATTO
DOIF..., page 89)
or
< 25 or > 35
52ba76157e0103/2014
Page 53
TruLab pH/ISE 1320Ion concentration
Calibration
Man. temperature
:25 °C
ISE setup
Method
:Standard addition
Start method
ISE
12.12.2012
12:12
Back
8.3Selecting the measuring method
The following methods are supported:
 Standard addition
 Standard subtraction
 Sample addition
 Sample subtraction
 Blank value addition
1.Connect the ISE combination electrode to the meter.
The pH/ISE measuring window is displayed.
2.If necessary, select the ISE display (unit, mg/l) with <M>.
3.If necessary, measure the temperature of the test sample with a thermometer.
4.Open the ISE menu with <MENU/ENTER>.
5.Thoroughly rinse the electrode with distilled water.
6.Adjust the temperature of the standard solutions.
7.Select Method with <> <> and confirm with <MENU/ENTER>.
8.Select a method with <> <> and confirm with <MENU/ENTER>.
9.Select Start method with <> <> and confirm with <MENU/ENTER>.
Measurement with the selected method begins (see section 8.3.1
STANDARDADDITION, page 54 ... section 8.3.5 STANDARDADDITIONWITH
Sample volume100.0 ml
ISA/TISAB vol.1.0 ml
Continue
Standard addition
12.12.2012
12:12
Back
8.3.1Standard addition
In the "Standard addition" procedure, a known amount of standard solution is
added to the sample.
The ion concentration in the sample is calculated from the change in potential.
1.Select the measuring method (see section 8.3 SELECTINGTHEMEASUR-
INGMETHOD, page 53).
2.Prepare two standard calibration solutions.
3.Perform a two-point calibration according to user guidance.
4.The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.
Measuring
5.Start the measurement with <F1>/[Continue].
An entry window appears.
6.Prepare the sample.
7.If necessary, add the ISA/TISAB solution to the sample.
8.Thoroughly rinse the electrode with deionized water.
9.Immerse the electrode in the sample.
54ba76157e0103/2014
10.Using <> <> and <MENU/ENTER>, select the values for the volume of the sample (Sample volume) and the volume of the ISA/TISAB
solution (ISA/TISAB vol.).
Page 55
TruLab pH/ISE 1320Ion concentration
i
Add standard!
Std. volume1.0 ml
Std. conc.1.0 mg/l
Continue
Standard addition
12.12.2012
12:12
Back
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Standard addition
mg/l
USB output
11.Select Continue with <> <> and start the measurement with
<MENU/ENTER>.
When the measurement is finished an entry window appears.
12.Add the standard solution to the sample.
13.Using <> <> and <MENU/ENTER>, enter the values for the vol-
ume of the standard solution (Std. volume) and concentration of the
standard solution (Std. conc.) .
14.Select Continue with <> <> and start the measurement with
<MENU/ENTER>.
The measurement result is displayed when the measurement is completed.
15.If necessary, start measuring further samples with <MENU/ENTER>.
Repeat steps 6 - 14 for all samples.
16.Terminate the measuring method with <M>.
A safety query appears.
17.Select Yes with <> <>.
18.Confirm Yes with <MENU/ENTER>.
Measurement with the selected method is completed.
Sample volume100.0 ml
ISA/TISAB vol.1.0 ml
Continue
Standard subtraction
12.12.2012
12:12
Back
8.3.2Standard subtraction
In the "Standard Subtraction" procedure, a known amount of standard solution
is added to the sample (as complexing agent or precipitating agent) and, thus,
the ion concentration lowered.
The ion concentration in the sample is calculated from the change in potential.
1.Select the measuring method (see section 8.3 SELECTINGTHEMEASUR-
INGMETHOD, page 53).
Calibration
Measuring
2.Prepare two standard calibration solutions.
3.Perform a two-point calibration according to user guidance.
4.The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.
5.Start the measurement with <F1>/[Continue].
An entry window appears.
6.Prepare the sample.
56ba76157e0103/2014
7.If necessary, add the ISA/TISAB solution to the sample.
8.Thoroughly rinse the electrode with deionized water.
9.Immerse the electrode in the sample.
Page 57
TruLab pH/ISE 1320Ion concentration
i
Add standard!
Ion typeS
Std. volume1.0 ml
Std. conc.1.0 mg/l
Continue
Standard subtraction
12.12.2012
12:12
Back
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Standard subtraction
mg/l
USB output
10.Using <> <> and <MENU/ENTER>, select the values for the vol-
ume of the sample (Sample volume) and the volume of the ISA/TISAB
solution (ISA/TISAB vol.).
11.Select Continue with <> <> and start the measurement with
<MENU/ENTER>.
When the measurement is finished an entry window appears.
12.Add the standard (complexing agent or precipitating agent) to the sample.
13.Using <> <> and <MENU/ENTER>, enter the values for the volume of the standard solution (Std. volume) and concentration of the
standard solution (Std. conc.) .
14.If necessary, set the ion type for the ion in the standard (Ion type) with
<> and <MENU/ENTER>.
<>
On selection of the ion type that can be defined (ION):
Using <> <> and <MENU/ENTER>, set the valence (Valency) and
molar weight (Molar mass) for the ion in the standard solution.
15.Select Continue with <> <> and start the measurement with
<MENU/ENTER>.
The measurement result is displayed when the measurement is completed.
ba76157e0103/201457
16.If necessary, start measuring further samples with <MENU/ENTER>.
Repeat steps 6 - 13 for all samples.
17.Terminate the measuring method with <M>.
A safety query appears.
18.Select Yes with <> <>.
Page 58
Ion concentrationTruLab pH/ISE 1320
19.Confirm Yes with <MENU/ENTER>.
Measurement with the selected method is completed.
Std. volume100.0 ml
Std. conc.1.0 mg/l
ISA/TISAB vol.1.0 ml
Continue
Sample addition
12.12.2012
12:12
Back
8.3.3Sample addition
In the "Sample addition" procedure, a known amount of sample is added to the
standard solution.
The ion concentration in the sample is calculated from the change in potential.
1.Select the measuring method (see section 8.3 SELECTINGTHEMEASUR-
INGMETHOD, page 53).
Calibration
Measuring
2.Prepare two standard calibration solutions.
3.Perform a two-point calibration according to user guidance.
4.The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.
5.Start the measurement with <F1>/[Continue].
An entry window appears.
6.Prepare the standard solution.
7.If necessary, add the ISA/TISAB solution to the standard solution.
8.Thoroughly rinse the electrode with deionized water.
ba76157e0103/201459
9.Immerse the electrode in the standard.
10.Using <> <> and <MENU/ENTER>, enter the values for the volume of the standard solution (Std. volume), the concentration of the
standard solution (Std. conc.) and the volume of the ISA/TISAB solu
tion (ISA/TISAB vol.).
-
Page 60
Ion concentrationTruLab pH/ISE 1320
i
Add sample!
Sample volume1.0 ml
Continue
Sample addition
12.12.2012
12:12
Back
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Sample addition
mg/l
USB output
11.Select Continue with <> <> and start the measurement with
<MENU/ENTER>.
When the measurement is finished an entry window appears.
12.Add the sample to the standard solution.
13.Using <> <> and <MENU/ENTER>, enter the value for the volume
of the sample (Sample volume).
14.Select Continue with <> <> and start the measurement with
<MENU/ENTER>.
The measurement result is displayed when the measurement is completed.
15.If necessary, start measuring further samples with <MENU/ENTER>.
Repeat steps 6 - 12 for all samples.
16.Terminate the measuring method with <M>.
A safety query appears.
17.Select Yes with <> <>.
18.Confirm Yes with <MENU/ENTER>.
Measurement with the selected method is completed.
With the "Sample subtraction" procedure, a known amount of sample is added
to the standard solution (complexing agent or precipitating agent).
The ion concentration in the sample is calculated from the change in potential.
The sample subtraction is one method for the indirect determination of ions. It
is used for the determination of ions for which no direct determination is possi
ble.
1.Select the measuring method (see section 8.3 SELECTINGTHEMEASUR-
INGMETHOD, page 53).
-
Calibration
2.If necessary, set the ion type for the ion in the standard (Ion type (sub))
with <>
<> and <MENU/ENTER>.
On selection of the ion type that can be defined (ION):
Using <> <> and <MENU/ENTER>, set the valence (Valency) and
molar weight (Molar mass) for the ion in the standard solution.
3.Select Continue with <> <> and confirm with <MENU/ENTER>.
4.Prepare two standard calibration solutions.
5.Perform a two-point calibration according to user guidance.
6.The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.
ba76157e0103/201461
Page 62
Ion concentrationTruLab pH/ISE 1320
i
Immerse sensor in standard
Std. volume100.0 ml
Std. conc.1.0 mg/l
ISA/TISAB vol.1.0 ml
Continue
Sample subtraction
12.12.2012
12:12
Back
i
Add sample!
Sample volume1.0 ml
Continue
Sample subtraction
12.12.2012
12:12
Back
Measuring
7.Start the measurement with <F1>Continue.
An entry window appears.
8.Prepare the standard solution.
9.If necessary, add the ISA/TISAB solution to the standard solution.
10.Thoroughly rinse the electrode with deionized water.
11.Immerse the electrode in the standard solution (complexing agent or
precipitating agent).
12.Using <> <> and <MENU/ENTER>, enter the values for the volume of the standard solution (Std. volume), the concentration of the
standard solution (Std. conc.) and the volume of the ISA/TISAB solu
tion (ISA/TISAB vol.).
-
13.Select Continue with <> <> and start the measurement with
<MENU/ENTER>.
When the measurement is finished an entry window appears.
14.Add the sample to the standard (complexing agent or precipitating
agent).
15.Using <> <> and <MENU/ENTER>, enter the value for the volume
of the sample (Sample volume).
16.Select Continue with <> <> and start the measurement with
<MENU/ENTER>.
The measurement result is displayed when the measurement is completed.
62ba76157e0103/2014
Page 63
TruLab pH/ISE 1320Ion concentration
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Sample subtraction
mg/l
USB output
17.If necessary, start measuring further samples with <MENU/ENTER>.
Repeat steps 8 - 16 for all samples.
18.Terminate the measuring method with <M>.
A safety query appears.
19.Select Yes with <> <>.
20.Confirm Yes with <MENU/ENTER>.
Measurement with the selected method is completed.
Calibration
8.3.5Standard addition with blank value correction (Blank value
addition)
In the "Standard addition with blank value correction" procedure, a known
amount of standard solution is added to the sample in two steps.
With the first addition, the ion concentration is raised to the linear range of the
electrode characteristic curve.
The second addition is equivalent to the standard addition. The ion concentration in the sample is calculated from the change in potential.
1.Select the measuring method (see section 8.3 SELECTINGTHEMEASUR-
INGMETHOD, page 53).
2.Prepare two standard calibration solutions.
3.Perform a two-point calibration according to user guidance.
4.The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.
Sample volume100.0 ml
ISA/TISAB vol.1.0 ml
BV volume100.0 ml
BV conc.1.000 mg/l
Continue
Blank value addition
12.12.2012
12:12
Back
i
Add standard!
Std. volume100.0 ml
Std. conc.1.0 mg/l
Continue
Blank value addition
12.12.2012
12:12
Back
Measuring
5.Start the measurement with <F1>Continue
An entry window appears.
6.Prepare the sample.
7.If necessary, add the ISA/TISAB solution to the sample.
8.Thoroughly rinse the electrode with deionized water.
9.Immerse the electrode in the sample that was supplemented with blank
value solution.
10.Using <> <> and <MENU/ENTER>, enter the values for the volume of the sample (Sample volume), the volume of the ISA/TISAB
solution (ISA/TISAB vol.), the volume of the blank value solution (BV volume) and the concentration of the blank value solution (BV conc.).
11.Select Continue with <> <> and start the measurement with
<MENU/ENTER>.
When the measurement is finished an entry window appears.
64ba76157e0103/2014
Page 65
TruLab pH/ISE 1320Ion concentration
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Blank value addition
mg/l
USB output
12.Add the standard solution to the sample.
13.Using <> <> and <MENU/ENTER>, enter the values for the vol-
ume of the standard solution (Std. volume) and concentration of the
standard solution (Std. conc.).
14.Select Continue with <> <> and start the measurement with
<MENU/ENTER>. The measurement result is displayed when the
measurement is completed.
15.If necessary, start measuring further samples with <MENU/ENTER>.
Repeat steps 6 - 14 for all samples.
16.Terminate the measuring method with <M>.
A safety query appears.
17.Select Yes with <> <>.
18.Confirm Yes with <MENU/ENTER>.
Measurement with the selected method is completed.
ba76157e0103/201465
Page 66
SettingsTruLab pH/ISE 1320
9Settings
9.1Measurement settings
9.1.1Settings for pH measurements
The settings for pH measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis
play the required measured parameter in the measured value display and
press the <MENU/ENTER> key. After completing the settings, switch to the
measured value display with <M>.
In the following table, only those settings are listed that concern the pH measurement.
Default settings are printed in bold.
-
Menu itemPossible set-
ting
Calibration /Calibration record
Calibration /
Calibration data
storage /Display
-Displays the calibration record of
-Displays the calibration record.
Description
the last calibration.
Further options:
 Scroll through the calibration
records with <><>.
 Output the displayed calibration
record to the interface with
<F2>/[USB output].
 Output all calibration records to
the interface with <F2__>[USB output].
 Quit the display with <F1>/
[Back] or <MENU/ENTER>.
 Switch directly to the measured
value display with <M>.
Calibration /
Calibration data
storage /
Output to USB
66ba76157e0103/2014
-Outputs the calibration records to
the interface.
Page 67
TruLab pH/ISE 1320Settings
Menu itemPossible set-
ting
Calibration/
-Entry of the series number of the
Serial number (sensor)
Calibration /BufferYSI
ConCal
NIST/DIN
...
Calibration /One
point calibration
Yes
No
Description
connected sensor.
The series number is output in the
calibration record.
 Change the contents of the high-
lighted position with <><>.
 Go to the next digit with <F2>/
[].
 After theseries number was
completely entered, confirm with
<ENTER>.
Buffer sets to be used for pH calibration (see section 6.2 CALIBRA-
TION, page 31).
Quick calibration with 1 buffer
Calibration /Cali-
1 ... 7 ... 999 dCalibration interval for the pH elec-
bration interval
Calibration /Unit for
slope
mV/pH
%
Man. temperature-25 ... +25 ...
+130
°C
Temperature unit°C
°F
Resolution pH0.001
0.01
0.1
trode (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring screen.
Unit of the slope.
The % display refers to the Nernst
slope of -59.2
mV/pH (100 x deter-
mined slope/Nernst slope).
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperature values are displayed with the selected unit.
Resolution of the pH display
ba76157e0103/201467
Page 68
SettingsTruLab pH/ISE 1320
Menu itemPossible set-
Description
ting
Limit monitoringWith the Limit monitoring function
you define the measured values
whose exceeding is signaled. An
acoustic signal sounds, and at the
same time a message is output to
the USB interface.
You can switch on or off the acoustic signal in the System menu (see
section 9.3.1 SYSTEM, page 75).
Switch on or off the limit signaling
device for the pH value.
Switch on or off the limit signaling
device for the temperature value.
-2 ... 20Upper limit. A message is output to
the USB interface if it is exceeded.
This menu item is only visible when
the pH monitoring setting is active.
-2 ... 20Lower limit. A message is output to
the USB interface if it is exceeded.
This menu item is only visible when
the pH monitoring setting is active.
-5 ... +105 °CUpper limit. A message is output to
the USB interface if it is exceeded.
This menu item is only visible when
the TP monitoring setting is active.
-5 ... 105 °CLower limit. A message is output to
the USB interface if it is exceeded.
This menu item is only visible when
the TP monitoring setting is active.
Reset-Resets all sensor settings to the
delivery condition (see
section
9.4.1 RESETTINGTHEMEASUREMENT
SETTINGS, page 76).
9.1.2Buffer sets for calibration
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.
68ba76157e0103/2014
Page 69
TruLab pH/ISE 1320Settings
No.Buffer set *pH valuesat
1YSI *4.000
25 °C
7.000
10.000
2ConCalAnyAny
3NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers
1.679
4.006
6.865
25 °C
9.180
12.454
4TEC
WTW Technical buffers
2.000
4.010
25 °C
7.000
10.011
5Merck 1*4.000
20°C
7.000
9.000
6Merck 2 *1.000
20°C
6.000
8.000
13.000
7Merck 3 *4.660
6.880
9.220
8Merck 4 *2.000
4.000
7.000
10.000
9Merck 5 *4.010
7.000
10.000
10DIN 192671.090
4.650
6.790
9.230
11Mettler Toledo USA *1.679
4.003
7.002
10.013
12Mettler Toledo EU *1.995
4.005
7.002
9.208
20°C
20°C
25 °C
25 °C
25 °C
25 °C
ba76157e0103/201469
Page 70
SettingsTruLab pH/ISE 1320
No.Buffer set *pH valuesat
13Fisher *2.007
4.002
7.004
10.002
14Fluka BS *4.006
6.984
8.957
15Radiometer *1.678
4.005
7.000
9.180
16Baker *4.006
6.991
10.008
17Metrohm *3.996
7.003
8.999
18Beckman *4.005
7.005
10.013
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
19Hamilton Duracal *4.005
7.002
10.013
20Precisa *3.996
7.003
8.999
21Reagecon TEC *2.000
4.010
7.000
10.000
22Reagecon 20 *2.000
4.000
7.000
10.000
13.000
25 °C
25 °C
25 °C
20°C
70ba76157e0103/2014
Page 71
TruLab pH/ISE 1320Settings
No.Buffer set *pH valuesat
23Reagecon 25 *2.000
25 °C
4.000
7.000
10.000
13.000
24Chemsolute *2.000
20°C
4.000
7.000
10.000
25USABlueBook *4.000
25 °C
7.000
10.000
* Brand names or trade names are trademarks of their respective owners protected
by law.
The buffers are selected in the menu, pH / <MENU/ENTER> / Calibration / Buffer (see 9.1.1 SETTINGSFORPH MEASUREMENTS, PAGE
66).
Setting the
calibration interval
9.1.3Calibration interval
The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
1.Open the menu for measurement settings with <MENU/ENTER>.
2.In the Calibration / Calibration interval menu, set the calibration interval
<><>.
with
3.Confirm the setting with <ENTER>.
4.Quit the menu with <M>.
ba76157e0103/201471
Page 72
SettingsTruLab pH/ISE 1320
9.1.4Settings for ORP measurements
The settings for ORP measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis
play the required measured parameter in the measured value display and
press the <MENU/ENTER> key. After completing the settings, switch to the
measured value display with <M>.
In the following table, only those settings are listed that influence the ORP measurement.
Default settings are printed in bold.
-
Menu itemPossible
Description
setting
Man. temperature-25 ... +25 ...
+130
°C
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Resolution mV0.1
Resolution of the mV display
1
Reset-Resets all sensor settings to the
delivery condition (see
section 9.4.1
RESETTINGTHEMEASUREMENTSET-
TINGS, page 76).
72ba76157e0103/2014
Page 73
TruLab pH/ISE 1320Settings
9.2Settings for ISE measurements
The settings are made in the measuring menu of the ISE measurement. To
open the settings, activate the relevant measuring window in the measured
value display and press the <MENU/ENTER> key
settings, switch to the measured value display with <M>.
The following settings are possible for ISE measurements:
Menu itemPossible settingDescription
shortly. After completing the
Calibration /
Calibration record
Calibration /
Calibration data storage /
Display
-Displays the calibration
record of the last calibration.
-Displays the calibration
record.
Further options:
 Scroll through the cali-
bration records with
<><>.
 Output the displayed cal-
ibration record to the
interface with <F2>/[USB output].
 Output all calibration
records to the interface
with <F2__>[USB output].
 Quit the display with
<F1>/[Back] or <MENU/
ENTER>.
 Switch directly to the
measured value display
with <M>.
-
Calibration /
Calibration data stor-
-Outputs the calibration
records to the interface.
age /
Output to USB
Calibration/
Serial number (sensor)
-Entry of the series number
of the connected sensor.
The series number is output
in the calibration record.
 Change the contents of
the highlighted position
with <><>.
 Go to the next digit with
<F2>/[].
 After theseries number
was completely entered,
confirm with <ENTER>.
ba76157e0103/201473
Page 74
SettingsTruLab pH/ISE 1320
Menu itemPossible settingDescription
Man. temperature-25 ... +25 ... +130 °CEntry of the manually deter-
mined temperature. For
measurements without tem
-
perature sensor only.
ISE setup /
AutoRead criterion
ISE setup /
Ion type
ISE setup /
Unit
low
medium
high
Ag, Br, Ca, Cd, Cl, CN,
Cu, F, I, K, Na, NO3,
Selection of the AutoRead
criteria (see
section 8.1.1
MEASURINGTHEIONCON-
CENTRATION, page 45).
Selection of the ion type to
be measured.
Pb, S, NH3, NH4*,
CO2, ION
An ion that is not included in
the list can be measured
with the setting, ION.
* Measuring with the NH 500 electrode:
The NH4 setting is not suitable for the gas-sensitive
electrode NH 500.
Select the following settings:
Ion type "ION", Valency "-1".
mg/l
µmol/l
mg/kg
ppm
Selection, with which unit
the measurement result and
calibration standards should
be displayed.
%
ISE setup /
Ion type / ION
Valency
Molar mass
Set the valence (Valency)
and molar weight (Molar
mass) for the ion.
ISE setup /
Density
0.001 ... 9.999 g/ml or
kg/l
Adjustable density of the
test sample (only with Unit:
mg/kg, ppm, %)
MethodStandard addition
Standard subtraction
Selection of the available
measuring methods.
Sample addition
Sample subtraction
Blank value addition
Start methodStart measurement with the
selected method.
74ba76157e0103/2014
Page 75
TruLab pH/ISE 1320Settings
9.3Sensor-independent settings
9.3.1System
To open the Storage & config menu, press the <MENU/ENTER__> key in the
measured value display. After completing the settings, switch to the measured
value display with <M>.
Default settings are printed in bold.
Menu itemPossible set-
ting
System / General /
Language
Deutsch
English
(further)
System / General /
Beep
System / General / Illumination
On
Off
Auto
On
Off
System / General /
0 ... 50 ... 100 %Changes the display contrast
Contrast
System /
General /
10 min ... 1h
... 24 h
Switchoff time
System /
General /
°C
°F
Temperature unit
Description
Selects the menu language
Switches on/off the beep on
keystroke
Switches the display illumination on/off
Adjusts the switch-off time
Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperature values are displayed with the selected unit.
System /
General /
Stability control
System /
Interface /
Baud rate
System /
Interface /
Output format
On
Off
1200, 2400,
4800, 9600,
19200
ASCII
CSV
Switches on or off the automatic stability control during
measurement
Baud rate of the data interface
Output format for data transmission (see section 11 TRANS-
MITTINGDATA (USB INTERFACE),
page 84)
System / Interface /
Decimal separator
System / Interface /
Output header
ba76157e0103/201475
Dot (xx.x)
Comma (xx,x)
Decimal separator
Output of a header for Output format: CSV
Page 76
SettingsTruLab pH/ISE 1320
Menu itemPossible set-
Description
ting
System / PrinterSettings for the integrated
printer of the TruLab
1320P.
ISE
pH/
For details, see section 4
PRINTER (ONLY TRULABPH/
ISE 1320 P), page 14
System /ClockDate format
Datum
Time
System /Service information
Time and date settings (see
section 5.4.6 EXAMPLE 2 ON
NAVIGATION: SETTINGTHEDATE
ANDTIME, page 25)
Hardware version and software
version of the meter are dis
-
played.
System /Reset-Resets the system settings to
the delivery condition (see
sec-
tion 9.4.2 RESETTINGTHESYS-
TEMSETTINGS, page 77).
pH
9.3.2Data storage
This menu contains all functions to display, edit and erase stored measured
values and calibration records (see
section 10 DATASTORAGE, page 78).
9.4Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
9.4.1Resetting the measurement settings
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
The following settings for pH measurements are reset to the default settings
with the Reset function:
SettingDefault settings
Buffer YSI
Calibration interval7 d
Unit for slopemV/pH
Measured parameterpH
76ba76157e0103/2014
Page 77
TruLab pH/ISE 1320Settings
SettingDefault settings
Resolution pH0.001
Resolution mV0.1
Asymmetry0 mV
Slope-59.2 mV
Man. temperature25 °C
One point calibrationNo
Stability controlOn
Temperature unit°C
The sensor settings are reset under the Reset menu item in the menu for calibration and measurement settings. To open it in the measured value display,
press the <MENU/ENTER> key.
9.4.2Resetting the system settings
The following system settings can be reset to the delivery status:
SettingDefault settings
LanguageEnglish
BeepOn
Baud rate4800 Baud
Output formatASCII
Decimal separatorDot (xx.x)
Contrast50 %
IlluminationAuto
Switchoff time1 h
The system settings are reset in the menu, Storage & config / System / Reset.
To open the Storage & config menu, press the <MENU/ENTER__> key in the
measured value display.
ba76157e0103/201477
Page 78
Data storageTruLab pH/ISE 1320
12.12.2012 11:24:16
pH 7.000 24.8 °C AR +++
ID number:1
Continue
Manual data storage4 von 200
12.12.2012
12:12
Back
10Data storage
You can transmit measured values (datasets) to the data storage:
 Manual data storage (see section 10.1 MANUALDATASTORAGE, page 78)
 Automatic storage at intervals (see section 10.2 AUTOMATICDATASTORAGE
ATINTERVALS, page 79)
With each data storage process, the current datasets of the sensors indicated
on the display are transmitted to the interface at the same time.
10.1Manual data storage
You can transmit a measurement dataset to the data storage as follows. With
each data storage process, the current datasets of the sensors indicated on the
display are transmitted to the interface at the same time.
1.Press the <STO> key shortly.
The menu for manual data storage appears.
If the storage is fullThe following window appears if all 200 storage locations are occupied:
78ba76157e0103/2014
2.If necessary, change and confirm the ID number (1 ... 10000) with
<><> and <MENU/ENTER>.
The dataset is stored. The meter switches to the measured value display.
Page 79
TruLab pH/ISE 1320Data storage
Data storage full. Erase?
Yes
No
Warning
12.12.2012
12:12
Back
1
ID number
1
Interval
30 s
Duration
180 min
Continue
0d03h00min
0 1d17h33min
Automatic data storage
12.12.2012
12:12
2
3
Back
You have the following options:
 To erase the entire storage, confirm Yes.
 To cancel the storage process and switch to the measured value display,
confirm No. Then you can e.g. transmit the stored data to a PC (see
section
10.3.1 EDITINGTHEMEASUREMENTDATASTORAGE, page 81) and subsequently erase the data storage (see section 10.3.2 ERASINGTHEMEASURE-
MENTDATASTORAGE, page 83).
Configuring the
automatic storage
function
10.2Automatic data storage at intervals
The storage interval (Interval) determines the time interval between automatic
data storage processes. With each data storage process, the current datasets
of the sensors indicated on the display are transmitted to the interface at the
same time.
1.Press the <STO__> key.
The menu for automatic data storage appears.
ba76157e0103/201479
1 Adjusted entire storage duration
2 Max. available storage duration
3 Graphical display of the storage usage
Page 80
Data storageTruLab pH/ISE 1320
1
2
pH
6.949
25.0
°C
0d03h00min
12.12.2012
12:12
0
14
Eco mode
SettingsYou can configure the automatic data storage function with the following
settings:
Menu itemPossible set-
ting
ID number1 ... 10000ID number for the dataset series.
Interval1 s, 5 s, 10 s,
s, 1 min,
30
5
min,
min,
10
min,
15
min,
30
60
min
Duration1 min ... x minStorage duration.
Description
Storage interval.
The lower limit of the storage
interval can be restricted by the
number of free storage locations.
The upper limit is restricted by the
storage duration.
Specifies after which time the
automatic data storage should be
terminated.
The lower limit of the storage
duration is restricted by the stor
age interval.
The upper limit is restricted by the
number of free storage locations.
-
Starting the
automatic storage
function
To start the automatic storage function, select Continue with <><> and
confirm with <MENU/ENTER>. The meter switches to the measured value dis
play.
1 Remaining storage duration
2 Graphical display of the storage duration
The active automatic storage function can be recognized by the progress bar
in the function display. The progress bar indicates the remaining storage dura
tion.
-
-
If the automatic storage function is activated, only the following
keys are active: Softkeys, <M>, <STO__> and <On/Off>. The
80ba76157e0103/2014
other keys and the automatic switch-off function are deactivated.
Page 81
TruLab pH/ISE 1320Data storage
Stop automatic storage?
Yes
No
Warning
12.12.2012
12:12
Back
Energy saving mode
([Eco mode])
Terminating the
automatic storage
function
prematurely
If the automatic storing function is active, the meter provides an energy saving
mode ([Eco mode]) to avoid unnecessary energy consumption. The energy
saving mode switches off functions of the meter that are not required for the
automatic storage of measurement data (such as the display). By pressing any
key the energy saving mode is switched off again.
Proceed as follows to switch off the automatic data storage function before the
adjusted storage duration has expired:
1.Press the <STO__> key.
The following window appears.
Editing the data
storage
2.Using <><>, select Yes and confirm with <MENU/ENTER>.
The meter switches to the measured value display.
The automatic data storage function is terminated.
10.3Measurement data storage
10.3.1 Editing the measurement data storage
The contents of the manual or automatic measurement data storage can be
shown on the display and output to the interface.
Each of the measurement data storages has a function to erase the entire
contents.
The storage is edited in the menu, Storage & config / Data storage. To open
the Storage & config menu, press the <MENU/ENTER__> key in the measured
value display.
Open the manual or automatic storage directly with the <RCL> or <RCL__>
key.
The settings are explained here using the manual data storage as
an example. The same settings and functions are available for the
automatic data storage.
pH1 6.012 24.8 °C, AR, S: +++
pH2 6.056 24.8 °C, AR, S: +++
________________________________________
Settings
Menu itemSetting/
function
Data storage /
-Displays all measurement datasets
Manual data storage
/ Display
Data storage /
-Erases the entire manual meaManual data storage
/ Erase
Data storage /
-Outputs all stored measurement
Manual data storage
/ Output to USB
Description
page by page.
Further options:
 Scroll through the datasets with
<><>.
 Output the displayed dataset to
the interface with <F2>/[USB output].
 Quit the display with <F1>/
[Back].
surement data storage.
All calibration data remain stored
when this action is performed.
data to the interface.
Display presentation
of a dataset
Representation of a
(USB output)
dataset
82ba76157e0103/2014
Page 83
TruLab pH/ISE 1320Data storage
Quitting the displayTo quit the display of stored measurement datasets, you have the following
options:
 Switch directly to the measured value display with <M>.
 Quit the display and move to the next higher menu level with <F1>/[Back].
10.3.2 Erasing the measurement data storage
Erasing the measurement data storage (see section 10.3.1 EDITINGTHE
MEASUREMENTDATASTORAGE, page 81).
10.3.3 Measurement dataset
A complete dataset includes:
 ID number
 Date/time
 Measured values of the connected sensors
 Measured temperature value of the connected sensors or manually set tem-
perature
 AutoRead info: The AR indicator appears with the measured value if the
AutoRead criterion was met while storing (stable measured value). Other
-
wise, there is no AR indicator.
 Calibration evaluation: +++, ++, +, -, or no evaluation
10.3.4 Storage locations
The TruLab pH/ISE 1320 meter has two measurement data storages. The
measured values recorded either manually or automatic are stored separately
in individual measurement data storages.
StorageMaximum number of datasets
Manual data storage200
Automatic data storage5000
ba76157e0103/201483
Page 84
TruLab pH/ISE 1320Transmitting data (USB interface)
11Transmitting data (USB interface)
11.1Options for data transmission
Via the USB interface you can transmit data to a PC. The following table shows
which data are transmitted to the interface in which way:
DataControlOperation / description
Current
datasets of the
sensors
indicated on
the display
Stored measured values
Manual With <F2>/[USB output].
 Simultaneously with every manual
data storage process (see
section
10.1 MANUALDATASTORAGE, page
78).
Automatic, at
intervals
 With <F2__>/[USB output].
Then you can set the transmission
interval.
 Simultaneously with every auto-
matic data storage process (see
section 10.2 AUTOMATICDATA
STORAGEATINTERVALS, page 79).
Manual Displayed dataset with <F2>[USB
output] after calling up from the
storage.
 All datasets with the Output to USB
function.
(see section 10.3.1 EDITINGTHE
MEASUREMENTDATASTORAGE,
page 81).
Calibration
records
The following rule applies: With the exception of the menus, shortly
pressing the <F2>[USB output] key generally outputs the display
contents to the interface (displayed measured values, measure
ment datasets, calibration records).
Manual Calibration record with <F2>/[USB
output]
(see section 6.2.6 CALIBRATION
DATA, page 38).
Automatic At the end of a calibration proce-
dure.
-
ba76157e0103/201484
Page 85
TruLab pH/ISE 1320Transmitting data (USB interface)
11.2Connecting a PC
Connect the TruLab pH/ISE 1320 to the PC via the USB interface.
NOTE
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as the values would be incorrect.
The USB interface is not galvanically isolated.
Installation of the
USB driver on the
PC
System requirements of the PC for installation of the USB driver:
 PC with Pentium processor or higher with at least one free USB connection
and CD-ROM drive
 Windows XP, Windows 7.
1.Insert the supplied installation CD in the CD drive of your PC.
2.Install the driver from the CD.
Follow the Windows installation instructions as necessary.
3.Connect the TruLab pH/ISE 1320 to the PC via the USB interface.
The meter is listed as a virtual COM interface among the connections in
the Windows instrument manager.
11.3MultiLab Importer
With the aid of the MultiLab Importer software, you can record and evaluate
measurement data with a PC.
More detailed information can be found in the MultiLab Importer
operating manual.
ba76157e0103/201485
Page 86
Transmitting data (USB interface)TruLab pH/ISE 1320
86ba76157e0103/2014
Page 87
TruLab pH/ISE 1320Maintenance, cleaning, disposal
1
12Maintenance, cleaning, disposal
12.1Maintenance
12.1.1 General maintenance activities
The only maintenance activity required is replacing the batteries.
See the relevant operating manuals of the electrodes for instructions on maintenance.
12.1.2 Replacing the batteries
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
1 Battery compartment
1.Open the battery compartment (1) on the underside of the
meter.
2.Remove the four batteries from the battery compartment.
3.Place four new batteries (type Mignon AA) in the battery compartment.
CAUTION
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the ± signs in
the battery compartment.
ba76157e0103/201487
Page 88
Maintenance, cleaning, disposalTruLab pH/ISE 1320
4.Close the battery compartment (1).
5.Set the date and time (see section 5.4.6 EXAMPLE 2 ONNAVIGATION:
SETTINGTHEDATEANDTIME, page 25)
When the batteries are nearly discharged, the status indicator is displayed.
Dispose of used batteries according to the local regulations of your
country.
End users within the European Union are obligated to return used
batteries (even ecologically compatible ones) to a collection point
set up for recycling purposes.
Batteries are marked with the crossed-out waste container symbol.
Therefore, they may not be disposed with the domestic waste.
12.2Cleaning
Occasionally wipe the outside of the measuring instrument with a damp, lintfree cloth. Disinfect the housing with isopropanol as required.
CAUTION
The housing is made of synthetic material (ABS). Thus, avoid
contact with acetone or similar detergents that contain sol
vents. Remove any splashes immediately.
-
12.3Packing
This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the
meter against damage during transport.
12.4Disposal
At the end of its operational lifetime, the meter must be returned to the disposal
or return system statutory in your country. If you have any questions, please
contact your supplier.
88ba76157e0103/2014
Page 89
TruLab pH/ISE 1320What to do if...
13What to do if...
13.1pH/ORP
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Error message
OFL, UFL
Error message,
Error
The measured value is outside the measuring range.
CauseRemedy
pH electrode:
– Air bubble in front of the junction– Remove air bubble
– Air in the junction– Extract air or moisten junction
– Cable broken– Replace electrode
– Gel electrolyte dried out– Replace electrode
– Measured value outside the mea-
suring range
CauseRemedy
pH electrode:
– The values determined for zero
point and slope of the electrode
are outside the allowed limits.
– pH value not stable– Measure with air excluded if
necessary
– Temperature not stable– Temper if necessary
Electrode + test sample:
– Conductivity too low– Use a suitable electrode
Obviously incorrect
measured values
– Temperature too high– Use a suitable electrode
– Organic liquids– Use a suitable electrode
CauseRemedy
pH electrode:
– pH electrode unsuitable– Use a suitable electrode
– Temperature difference between
buffer and test sample too great
– Measurement procedure not suit-
– Adjust the temperature of buf-
fer or sample solutions
– Follow a special procedure
able
90ba76157e0103/2014
Page 91
TruLab pH/ISE 1320What to do if...
13.2ISE
Error message OFL
Obviously incorrect
measured values
Error message,
Error (invalid
calibration)
CauseRemedy
– Measuring range exceeded– Dilute the test sample
CauseRemedy
– Electrode not connected– Connect the electrode
– Cable broken– Replace the electrode
CauseRemedy
ISE electrode:
– Moisture in the plug– Dry the plug
– Electrode obsolete– Replace the electrode
– Electrode not suitable for the range
– Use a suitable electrode
to be measured
– Electrode not suitable for the
selected ion
– Use a suitable electrode or
select a suitable ion
Warning [TpErr]
Warning [ISEErr]
– The gas-sensitive electrode NH 500
was calibrated with the NH4 setting
– Select the following settings:
Ion type "ION", Valency "-1"
– Socket damp– Dry the socket
Calibration procedure:
– Wrong sequence of standards for
– Select the correct sequence
three point calibration
– Calibration standards do not have
the correct temperature (max.
± 2 °C
– Adjust the temperature of the
calibration standards
temperature difference)
CauseRemedy
– Temperature difference between
measurement and calibration
greater than 2
K.
– Adjust the temperature of the
test sample
CauseRemedy
– Electrode voltage outside calibrated
– Recalibrate
range
ba76157e0103/201491
Page 92
What to do if...TruLab pH/ISE 1320
92ba76157e0103/2014
Page 93
TruLab pH/ISE 1320What to do if...
13.3General information
Sensor symbol
flashes
display
Meter does not react
to keystroke
You want to know
which software
version is in the
meter
CauseRemedy
– Calibration interval expired– Recalibrate the measuring
CauseRemedy
– Batteries almost empty– Replace the batteries (see
CauseRemedy
– Operating condition undefined or
EMC load unallowed
CauseRemedy
– E. g., a question by the service
department
system
section 12.1 MAINTENANCE,
page 87)
– Processor reset:
Press the <ENTER> and <On/Off> key simultaneously
– Switch on the meter.
Open the menu, <ENTER_> /
Storage & config / System /
Service information. The
instrument data are displayed.
ba76157e0103/201493
Page 94
Technical dataTruLab pH/ISE 1320
14Technical data
14.1Measuring ranges, resolution, accuracy
14.1.1 pH/ORP
Measuring ranges,
resolution
ParameterMeasuring rangeResolution
pH-2.0 ... +20.00.1
-2.00 ... +20.000.01
- 2.000 ... + 19.9990.001
U [mV]-1200.0 ... +1200.00.1
-2500 ... +25001
T [°C]-5.0 ... +105.00.1
T [°F]23.0 ... +221.00.1
Manual
temperature input
ParameterRangeIncrement
T
T
[°C]-25 ... +1301
manual
[°F]-13 ... +2661
manual
Accuracy (± 1 digit)ParameterAccuracyTemperature of the test
sample
pH / range *
-2.0 ... +20.0± 0.1+15 °C ... +35 °C
-2.00 ... +20.00± 0.01+15 °C ... +35 °C
- 2.000 ... + 19.999± 0.005+15 °C ... +35 °C
U [mV] / range
-2500 ... +2500± 1+15 °C ... +35 °C
-1200.0 ... +1200.0± 0.3+15 °C ... +35 °C
T [°C] / temperature sensor
NTC 30± 0.1
PT 1000± 0.1
* when measuring in a range of ± 2 pH around a calibration point
The accuracy values specified here apply exclusively to the meter.
The accuracy of the electrodes and buffer solutions has to be taken
into account additionally.
94ba76157e0103/2014
Page 95
TruLab pH/ISE 1320Technical data
14.1.2 ISE
Measuring ranges,
resolution
ParameterMeasuring rangeResolution
ISE [mg/l]0.000 ... 9.999
10.00 ... 99.99
100.0 ... 999.9
1000 ... 999999
ISE [µmol/l]0.000 ... 9.999
10.00 ... 99.99
100.0 ... 999.9
1000 ... 9999
0.001
0.01
0.1
1
0.001
0.01
0.1
1
[mmol/l]1000 ... 9999991
ISE [mg/kg]0.000 ... 9.999
10.00 ... 99.99
100.0 ... 999.9
1000 ... 999999
ISE [ppm]0.000 ... 9.999
10.00 ... 99.99
100.0 ... 999.9
1000 ... 999999
ISE [%]0.000 ... 9.999
10.00 ... 99.99
100.0 ... 999.9
1000 ... 999999
0.001
0.01
0.1
1
0.001
0.01
0.1
1
0.001
0.01
0.1
1
Manual
temperature input
ParameterRangeIncrement
T
[°C]- 20 ... + 1301
manual
ba76157e0103/201495
Page 96
Technical dataTruLab pH/ISE 1320
14.2General data
DimensionsTruLab pH/ISE 1320:Approx. 230 x 190 x 80 mm
Batteries4 x 1.5 V alkali-manganese batteries, type
AA
Rechargeable batteries
4 x 1.2 V NiMH rechargeable batteries,
AA
type
(no charging function)
Operational lifeUp to 1000 h without / 150 h with illumination
Power pack Kuantech Co. Ltd. KSAC 0900110W1UV-1
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 270 mA
Output: 9 V = / 1.1 A
Connection max. overvoltage category II
Primary plugs contained in the scope of
delivery: Euro, US, UK and Australian.
Sensor inputInput resistance> 5 * 1012 ohm
Input current< 1 * 10
-12
A
USB interfaceTypeUSB 1.1
USB B (device), data output
Baud rateAdjustable:
1200, 2400, 4800, 9600, 19200 Baud
Data bits8
Stop bits2
ParityNone
HandshakeRTS/CTS
Cable length Max. 3 m
96ba76157e0103/2014
Page 97
TruLab pH/ISE 1320Technical data
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with the limits
for a Class A digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely
to cause harmful interference in which case the user will be required to
correct the interference at his own expense.
Changes or modifications not expressly approved by the manufacturer
could void the user‘s authority to operate the equipment.
Applicable directives
and standards
EMCEC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Meter safetyEC directive 2006/95/EC
EN 61010-1
UL 61010-1
CAN/CSA-C22.2#61010-1
IP protection classEN 60529
ba76157e0103/201497
Page 98
Firmware updateTruLab pH/ISE 1320
15Firmware update
Available firmware updates are provided on the Internet.
With the firmware update program and a PC you can update the firmware of the
TruLab
For the update you have to connect the meter to a PC.
For the update via the USB interface, the following is required:
 a free USB interface (virtual COM port) on the PC
 the driver for the USB interface (on the enclosed CD-ROM)
 the USB cable (included in the scope of delivery of the TruLab pH/
1.Install the downloaded firmware update on a PC.
pH/ISE 1320 to the newest version.
1320).
ISE
An update folder is created in the Windows start menu.
If an update folder already exists for the meter (or meter type), the new
data is displayed there.
2.In the windows start menu, open the update folder and start the firmware
update program.
3.Using the USB interface cable, connect the TruLab pH/ISE 1320 to a
USB interface (virtual COM port) of the PC.
4.Switch on the TruLab pH/ISE 1320.
5.In the firmware update program, start the update process with OK.
6.Follow the instructions of the firmware update program.
During the programming process, a corresponding message and a
progress bar (in %) are displayed.
The programming process takes approx. three minutes. A terminatory
message is displayed after a successful programming process. The
firmware update is completed.
7.Disconnect the TruLab pH/ISE 1320 from the PC.
The TruLab pH/ISE 1320 is ready for operation again.
After switching the meter off and on you can check whether the meter has taken
over the new software version (see
VERSIONISINTHEMETER, PAGE 93).
YOUWANTTOKNOWWHICHSOFTWARE
98ba76157e0103/2014
Page 99
TruLab pH/ISE 1320Glossary
16Glossary
pH/ORP/ISE
Asymmetrysee zero point
Electromotive force of
a electrode
JunctionThe junction is a porous body in the housing wall of reference elec-
ORP voltageThe ORP is caused by oxidizing or reducing substances dissolved in
pH valueThe pH value is a measure of the acidic or basic effect of an aqueous
PotentiometryName of a measuring technique. The signal (depending on the mea-
SlopeThe slope of a linear calibration function.
The electromotive force U of the electrode is the measurable electromotive force of an electrode in a solution. It equals the sum of all the galvanic voltages of the electrode. Its dependency on the pH results in the
electrode function, which is characterized by the parameters, slope and
zero point.
trodes or electrolyte bridges. It arranges the electrical contact between
two solutions and makes the electrolyte exchange more difficult. The
expression, junction, is also used for ground or junction-less transitions.
water if these substances become effective on an electrode surface (e.
g. a gold or platinum surface).
solution. It corresponds to the negative decadic logarithm of the molal
hydrogen ions activity divided by the unit of the molality. The practical
pH value is the value of a pH measurement.
sured parameter) of the electrode is the electrical potential. The electrical current remains constant.
Zero pointThe zero point of a pH electrode is the pH value at which the electromo-
tive force of the pH electrode at a specified temperature is zero. Normally, this is at 25 °C.
General information
AdjustingTo manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or a
value that is regarded as correct, or that the difference remains within
the tolerance.
AutoRangeName of the automatic selection of the measuring range.
CalibrationComparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct.
this expression is also used when the measuring system is adjusted at
the same time (see adjusting).
ChannelA channel is a display indication that corresponds to a physical connec-
tion on the meter.
Measured parameterThe measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D.O. concentration.
Often,
ba76157e0103/201499
Page 100
GlossaryTruLab pH/ISE 1320
Measured valueThe measured value is the special value of a measured parameter to be
determined. It is given as a combination of the numerical value and unit
(e. g.
3 m; 0.5 s; 5.2 A; 373.15 K).
MolalityMolality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
vent.
Reset Restoring the original condition of all settings of a measuring system.
ResolutionSmallest difference between two measured values that can be dis-
played by a meter.
Stability control (Au-
Function to control the measured value stability.
toRead )
Standard solutionThe standard solution is a solution where the measured value is known
by definition. It is used to calibrate a measuring system.
Temperature functionName of a mathematical function expressing the temperature behavior
of a test sample, a sensor or part of a sensor.
Test sampleDesignation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample and
original sample are identical if the test sample was not processed.
100ba76157e0103/2014
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.