The compact precision pH meter TruLab pH/ISE 1320 enables you to carry out
pH measurements, ORP measurements and ion selective measurements
quickly and reliably.
The TruLab pH/ISE 1320 provides the maximum degree of operating comfort,
reliability and measuring certainty for all applications.
The USB interface can be used for data transmission to a PC and for software
updates of the meter.
1 Keypad
2 Display
3 Socket field
ba76157e0103/20147
OverviewTruLab pH/ISE 1320
Print
er
modul
1
2
3
4
5
pH 1310P
1.2TruLab pH/ISE 1320P meter
The integrated printer of the TruLab pH/ISE 1320 P enables to document measurements according to GLP requirements.
The information concerning the printer of the TruLab pH/ISE 1320
P is given in a separate chapter (see
TRULABPH/ISE 1320 P), page 14).
section 4 PRINTER (ONLY
1 Keypad
2 Display
3 Socket field
4Printer
5 Control panel of the printer
1.3Sensors
A measuring system ready to measure consists of the TruLab pH/ISE 1320
meter and a suitable sensor.
Suitable sensors are pH electrodes, ORP electrodes and ion selective electrodes (ISE).
8ba76157e0103/2014
TruLab pH/ISE 1320Safety
2Safety
2.1Safety information
2.1.1Safety information in the operating manual
This operating manual provides important information on the safe operation of
the meter. Read this operating manual thoroughly and make yourself familiar
with the meter before putting it into operation or working with it. The operating
manual must be kept in the vicinity of the meter so you can always find the infor
mation you need.
Important safety instructions are highlighted in this operating manual. They are
indicated by the warning symbol (triangle) in the left column. The signal word
(e.g. "Caution") indicates the level of danger:
WARNING
indicates a possibly dangerous situation that can lead to serious (irreversible) injury or death if the safety instruction is not
followed.
-
CAUTION
indicates a possibly dangerous situation that can lead to
slight (reversible) injury if the safety instruction is not fol
lowed.
NOTE
indicates a possibly dangerous situation where goods might be damaged
if the actions mentioned are not taken.
2.1.2Safety signs on the meter
Note all labels, information signs and safety symbols on the meter and in the
battery compartment. A warning symbol (triangle) without text refers to safety
information in this operating manual.
2.1.3Further documents providing safety information
The following documents provide additional information, which you should
observe for your safety when working with the measuring system:
•Operating manuals of sensors and other accessories
•Safety datasheets of calibration or maintenance accessories (such as buffer
solutions, electrolyte solutions, etc.)
-
ba76157e0103/20149
SafetyTruLab pH/ISE 1320
2.2Safe operation
2.2.1Authorized use
This meter is authorized exclusively for pH, ORP and ISE concentrations
measurements in the laboratory.
Only the operation and running of the meter according to the instructions and
technical specifications given in this operating manual is authorized (see
section 14 TECHNICALDATA, page 94).
Any other use is considered unauthorized.
2.2.2Requirements for safe operation
Note the following points for safe operation:
•The meter may only be operated according to the authorized use specified
above.
•The meter may only be supplied with power by the energy sources
mentioned in this operating manual.
•The meter may only be operated under the environmental conditions
mentioned in this operating manual.
•The meter may only be opened if this is explicitly described in this operating
manual (example: Inserting the batteries).
2.2.3Unauthorized use
The meter must not be put into operation if:
•it is visibly damaged (e.g. after being transported)
•it was stored under adverse conditions for a lengthy period of time (storing
conditions, see
section 14 TECHNICALDATA, page 94).
10ba76157e0103/2014
TruLab pH/ISE 1320Commissioning
3Commissioning
3.1Scope of delivery
pH meter TruLab pH/ISE 1320
4 batteries 1.5 V Mignon type AA
Power pack
USB cable (A plug on mini B plug)
Stand
Stand holder
Comprehensive operating manual
Short instructions
CD-ROM with
– USB drivers
– detailed operating manual
– Software MultiLab Importer
3.2Power supply
The TruLab pH/ISE 1320 is supplied with power in the following ways:
Mains operation with the supplied power pack
Battery operation (4 x alkaline manganese batteries, type AA)
USB operation via a connected USB-B cable
3.3Initial commissioning
Perform the following activities:
Insert the supplied batteries
For mains operation: Connect the power pack
If necessary, mount a stand
Switch on the meter
(see section 5.2 SWITCHINGONTHEMETER, page 19)
Set the date and time
(see section 5.4.6 EXAMPLE 2 ONNAVIGATION: SETTINGTHEDATEANDTIME,
page 25)
ba76157e0103/201411
CommissioningTruLab pH/ISE 1320
1
3.3.1Inserting the batteries
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
1 Battery compartment
1.Open the battery compartment (1) on the underside of the meter.
CAUTION
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the ± signs in
the battery compartment.
2.Place four batteries (type Mignon AA) in the battery compartment.
3.Close the battery compartment (1).
4.Set the date and time
(see section 5.4.6 EXAMPLE 2 ONNAVIGATION: SETTINGTHEDATEAND
TIME, page 25).
12ba76157e0103/2014
TruLab pH/ISE 1320Commissioning
1
2
3.3.2Connecting the power pack
CAUTION
The line voltage at the operating site must lie within the input
voltage range of the original power pack (see
NICALDATA, page 94).
section 14 TECH-
CAUTION
Use original power packs only (see section 14 TECHNICALDATA,
page 94).
1.Connect the plug of the power pack to the socket for the power pack on
the TruLab
2.Connect the original power pack to an easily accessible power outlet.
pH/ISE 1320.
3.3.3Mounting the stand
The stand base can be mounted at the right side of the meter.
4.1Commissioning / switching the printer on or off
1.Connect the power pack to the TruLab pH/ISE 1320 P.
The LED lights up green. The printer is ready for operation.
or
When the printer had been switched off (LED flashing):
Switch on the printer with <ON/OFF>.
The LED lights up green. The printer is ready for operation.
If there is a USB-B connection at the same time (e.g. to a PC), the
data are output to the PC only.
Data are only output to the printer if the following requirements are met
Data are transmitted manually or automatically (see section 11.1 OPTIONS
FORDATATRANSMISSION, page 84)
The printer is switched on (LED illuminated)
There is no USB connection.
4.3Printer settings
To open the Storage & config menu, press the <ENTER_> key in the measured
value display. After completing the settings, switch to the measured value dis
play with <M>.
Default settings are printed in bold.
-
Menu itemPossible set-
Description
ting
System / Printer /
Font size
12x20
8x16
7x16
Select the font size for the printer
A print sample (font of the printer) to
view the available font sizes can be
output with <ON/OFF_>.
System /
Printer /
Print test page
-The printer prints the meter informa-
tion from the menu, System / Ser-vice information.
The current printer settings are
used for the printout.
4.4Maintenance
4.4.1Changing the roll of paper (thermal paper)
1.Pull the lever (3) until the lid (4) of the paper compartment opens.
2.If necessary, remove any remains of the old roll of paper.
3.Insert the new roll of paper. The start of the paper should poke out of the
paper compartment.
4.Close the lid (4) of the paper compartment so that it clicks into place.
5.If necessary, move the printer paper forward with <FEED>.
Opens the menu for measurement settings / confirms
entries
Opens the menu for system settings
<AR>Freezes the measured value (HOLD function)
Switches the AutoRead measurement on or off
ba76157e0103/201417
OperationTruLab pH/ISE 1320
HOLD AR
25.0
6.093
pH
°C
4
3
2
6
7
8
9
1
AutoCal
5
12.12.2012
12:12
USB output
YSI
5.1.2Display
1 Status information (sensor)
2 Measured value (with unit)
3 Measured parameter
4 Continuous measurement control (CMC function)
5 Channel display: Plug position of the sensor
6 Sensor symbol (calibration evaluation, calibration interval)
7 Measured temperature (with unit)
8 Status information (meter)
9 Softkeys and date + time
5.1.3Status information (meter)
ARStability control (AutoRead) is active
HOLDMeasured value is frozen (<AR> key)
Batteries are almost empty
18ba76157e0103/2014
TruLab pH/ISE 1320Operation
1
2
3
7
8
9
4
5
6
5.1.4Socket field
1 pH electrode or ion selective electrode 1
2 Reference electrode 1
3 Temperature sensor 1
4 pH electrode or ion selective electrode 2
5 Reference electrode 2
6 Temperature sensor 2
7 Mini USB-B interface
8 Power pack
9 Service interface
CAUTION
Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current
circuit with current limiting).
Almost all customary sensors fulfill these conditions.
5.2Switching on the meter
1.Switch the meter on with <On/Off>.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test is being
performed.
The measured value display appears.
ba76157e0103/201419
OperationTruLab pH/ISE 1320
pH
6.949
25.0
°C
12.12.2012
12:12
USB output
5.3Switching off
1.Switch the meter off with <On/Off>.
Automatic switch-offThe instrument has an automatic switch-off function in order to save the
batteries (see
switches off the measuring instrument if no key is pressed for an adjustable
period.
The automatic switchoff is not active
if the communication cable is connected
if the Automatic data storage function is active, or with automatic data trans-
mission
if the power pack is connected
Display illuminationThe meter automatically switches off the display illumination if no key is
pressed for 30
again.
You can also generally switch the display illumination on or off (see section
9.3.1 SYSTEM, page 75).
section 9.3.1 SYSTEM, page 75). The automatic switchoff
seconds. The illumination is switched on with the next keystroke
5.4Navigation
The principles of navigation in menus and dialogs are explained in the following
sections.
5.4.1Operating modes
The meter has the following operating modes:
20ba76157e0103/2014
TruLab pH/ISE 1320Operation
Operating
Description
mode
MeasuringThe measurement data of the connected sensor are shown
in the measured value display
CalibrationThe course of a calibration with calibration information, func-
tions and settings is displayed
Storage in
The meter stores measuring data automatically or manually
memory
Transmitting data
The meter transmits measuring data and calibration records
to a USB-B interface automatically or manually
SettingThe system menu or a sensor menu with submenus, set-
tings and functions is displayed
Only those displays and functions are available in the active operating mode
that are currently being required.
5.4.2Measured value display
In the measured value display, open the setting menus with <MENU/ENTER>.
The current functions of the softkeys are shown on the display.
Use <MENU/ENTER> (short pressure) to open the menu for calibration and
measurement settings for the displayed measured parameter.
Use <MENU/ENTER__> (long keystroke (approx. 2 s) to open the Storage
& config menu with the sensor-independent settings.
Use the keys of the keypad to carry out further functions such as storage or calibration (see section 5.1.1 KEYPAD, page 17). These functions are not available
in other operating situations.
5.4.3Menus and dialogs
The menus for settings and dialogs in procedures contain further submenus.
To select a subelement, use the <><> keys. The current selection is dis-
played with a frame.
To make further settings, switch to the next higher menu level with
<F1>[Back].
Use <M> to return to the measured value display.
5.4.4Elements in menus and dialogs
Submenus
The name of the submenu is displayed at the upper edge of the frame. Submenus are opened by confirming with <MENU/ENTER>. Example:
Calibration record
Calibration data storage
Serial number (sensor)
Buffer
:TEC
One point calibration:Yes
Calibration interval
:7 d
Unit for slope
:mV/pH
i
2.00 4.01 7.00 10.01
pH
12.12.2012
12:12
Back
Settings
Settings are indicated by a colon. The current setting is displayed on the
right-hand side. The setting mode is opened with <MENU/ENTER>. Subse
quently, the setting can be changed with <><> and <MENU/ENTER>.
Example:
-
Functions
Messages
22ba76157e0103/2014
Functions are designated by the name of the function. They are immediately
carried out by confirming with <MENU/ENTER>.
Example: Display the Calibration record function.
Information is marked by the i symbol. They cannot be selected. Example:
TruLab pH/ISE 1320Operation
Calibration record
Calibration data storage
Serial number (sensor)
Buffer
:TEC
One point calibration
:Yes
Calibration interval:7 d
Unit for slope
:mV/pH
i
2.00 4.01 7.00 10.01
pH
12.12.2012
12:12
Back
pH
6.949
25.0
°C
12.12.2012
12:12
USB output
System
Data storage
Storage & config
12.12.2012
12:12
Back
5.4.5Example 1 on navigation: Setting the language
1.Press the <On/Off> key.
The measured value display appears.
The instrument is in the measuring mode.
ba76157e0103/201423
2.Using <MENU/ENTER__>, open the Storage & config menu.
The instrument is in the setting mode.
3.Select the System submenu with <><>.
The current selection is displayed with a frame.
4.Open the System submenu with <MENU/ENTER>.
OperationTruLab pH/ISE 1320
General
Interface
Clock
Service information
Reset
System
12.12.2012
12:12
Back
Language:Deutsch
Beep:Off
Illumination
:On
Contrast
:50 %
Switchoff time:1 h
Temperature unit
:°C
Stability control
:On
General
12.12.2012
12:12
Back
Language:Deutsch
Beep
:Off
Illumination
:On
Contrast
:50 %
Switchoff time
:1 h
Temperature unit
:°C
Stability control:On
General
12.12.2012
12:12
Back
5.Select the General submenu with <><>.
The current selection is displayed with a frame.
6.Open the Generalsubmenu with <MENU/ENTER>.
24ba76157e0103/2014
7.Open the setting mode for the Language with <MENU/
ENTER>.
8.Select the required language with <><>.
9.Confirm the setting with <MENU/ENTER>.
The meter switches to the measuring mode.
The selected language is active.
TruLab pH/ISE 1320Operation
Date format:dd.mm.yyyy
Date
:12.12.2012
Time
:12:12:25
Clock
12.12.2012
12:12
Back
5.4.6Example 2 on navigation: Setting the date and time
The measuring instrument has a clock with a date function. The date and time
are shown in the measured value display.
When storing measured values and calibrating, the current date and time are
automatically stored as well.
The correct setting of the date and time and date format is important for the following functions and displays:
Current date and time
Calibration date
Identification of stored measured values.
Therefore, check the time at regular intervals.
After a fall of the supply voltage (empty batteries), the date and time
are reset.
The date format can be switched from the display of day, month, year
(dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or mm.dd.yyyy).
1.In the measured value display:
Using <MENU/ENTER__>, open the Storage & config menu.
The instrument is in the setting mode.
2.Select and confirm the System / Clock menu with <><> and
<MENU/ENTER>.
The setting menu for the date and time opens up.
3.Select and confirm the Time menu with <><> and <MENU/ENTER>.
The hours are highlighted.
ba76157e0103/201425
4.Change and confirm the setting with <><> and <MENU/ENTER>.
The minutes are highlighted.
5.Change and confirm the setting with <><> and <MENU/ENTER>.
The seconds are highlighted.
6.Change and confirm the setting with <><> and <MENU/ENTER>.
The time is set.
OperationTruLab pH/ISE 1320
7.If necessary, set the Date and Date format. The setting is made similarly
to that of the time.
8.To make further settings, switch to the next higher menu level with
[Back]<F1>.
or
Switch to the measured value display with <M>.
The instrument is in the measuring mode.
26ba76157e0103/2014
TruLab pH/ISE 1320Operation
Channel display
Display of the plug position
for the respective parameter.
The filled bar indicates for
each connected sensor to
which plug position of the
meter it is connected.
pH
6.949
24.8 °C
12.12.2012
12:12
USB output
pH
6.966
25.0
°C
12.12.2012
12:12
USB output
5.5Channel display
The TruLab pH/ISE 1320 manages the connected sensors and displays which
sensor is plugged to which connection.
5.5.1Display of several sensors in the measuring mode
The measured values of the connected sensors can be displayed in the following ways:
Clear display of all connected sensors
Detailed display of one sensor
(e.g. incl. CMC feature with pH sensors)
With the softkey you can very easily switch between the two display types. The
suitable softkey is displayed depending on the operating situation.
ba76157e0103/201427
pH valueTruLab pH/ISE 1320
pH
6.949
24.8 °C
12.12.2012
12:12
USB output
6pH value
6.1Measuring
6.1.1Measuring the pH value
NOTE
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as the values would be incorrect.
The USB interface is not galvanically isolated.
To ensure the high measurement accuracy of the measuring system, always measure with a calibrated electrode (see section 6.2
CALIBRATION, page 31).
1.Connect the pH electrode to the meter.
2.If necessary, select the measured parameter with <M>.
3.When measuring without temperature sensor:
Temper the test sample or measure the current temperature and enter it
with <>
4.Immerse the pH electrode in the test sample.
The measured value is checked for stability (automatic stability control).
The display of the measured parameter flashes.
5.Wait for a stable measured value.
The display of the measured parameter no longer flashes.
<>.
Stability control
(AutoRead)
& HOLD function
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc
ibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can activate or switch off the automatic Stability control function (see
section 9.3.1 SYSTEM, page 75).
28ba76157e0103/2014
-
TruLab pH/ISE 1320pH value
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the Stability control function and the HOLD func-
tion with <AR> or <M> at any time.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured value is recognized. The progress bar disappears and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
Criteria for a stable
measured value
You can prematurely terminate the Stability control function manually with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
Time intervalStability in the time interval
parameter
pH value15 seconds∆ : better than 0.01 pH
Temperature15 seconds∆ : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
ba76157e0103/201429
pH valueTruLab pH/ISE 1320
6.1.2Measuring the temperature
For reproducible pH measurements, it is essential to measure the temperature
of the test sample.
You have the following options to measure the temperature:
Automatic measurement of the temperature with the temperature sensor
30 or Pt 1000) integrated in the sensor.
(NTC
Measurement by an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
Temperatur
e sensor
Resolution of the
temp. display
Temp. measurement
yes0.1 °CAutomatic with temperature sensor
-1 °CManual
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
1.Measure the current temperature of the test sample.
2.Set the temperature value with <><>.
or
In the <MENU/ENTER>/pH/Man. temperature menu, set the tempera-
ture value with <><>.
30ba76157e0103/2014
Loading...
+ 76 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.