YSI 1310P User Manual [en, fr, es]

OPERATING
MANUAL
TruLine ORP 50
MODE
INSTRUCCIONES DE
T
D’EMPLOI
OPERACIÓN
814241 04/2014
TruLine
TruLine
Operating
Technical data ................................................................................................. 5
Commissioning, measuring, testing .................................................................................. 6
Storage ............................................................................................................................. 8
Aging ................................................................................................................................. 8
Maintenance and cleaning .............................................................................. 9
Activating platinum electrodes ........................................................................................ 10
Wear parts and accessories .................................................................................. 11
Mode d'emploi ...................................................................................................................... 13
Caractéristiques techniques ............................................................................................ 15
Mise en service, mesure, verification .............................................................................. 16
Stockage ......................................................................................................................... 18
Vieillissement .................................................................................................................. 18
Maintenance et nettoyage ............................................................................................... 19
Activation d'électrodes de platine .................................................................................... 20
Pièces d'usure et accessoires ......................................................................................... 21
Instrucciones de
Especificaciones técnicas ............................................................................................... 25
Puesta en funcionamiento, medición, verificación .......................................................... 26
Almacenamiento ............................................................................................................. 28
Envejecimiento .............................................................................................. 28
Mantenimiento y limpieza ...................................................................................... 29
Activar electrodos de platino ........................................................................................... 30
Piezas de desgaste y accesorios .................................................................................... 31
Contact Information......................................................................................... 35
Manual .......................................................................................................................
operación ....................................................................................................
3
23
TruLine ORP 50
OPERATING
T
814241 04/2014
MANUAL
TruLine
ORP ELECTRODE
TruLine
For the most recent version of the manual, please visit www.ysi.com.
Contact YSI
1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241
800-765-4974 Email: environmental@ysi.com Internet: www.ysi.com
Copyright © 2014 Xylem Inc.
TruLine
Technical data
T 814241
04/2014
5
Model
Reference electrolyte
Electrode material and
Junction
TruLine ORP 50
3 mol/l KCl, Ag+-free
Platinum / Round piece
Ceramic
Model
Allowed
Typical application
TruLine ORP 50
0 ... 100 °C
Laboratory
Model
Shaft Electrical connection
Length
Ø [mm]
Material
Combination
Meter con-
Cable
TruLine ORP 50
120
12 Glass
Fixed cable
BNC
1 m
Technical data
General data
Measurement and application characteristics
Shaft dimensions, shaft material, electrical connection
shape
temperature range
[mm]
electrode connection
nection
length
Commissioning,
measuring, testing
TruLine
6
T 814241
04/2014
Commissioning,
measuring, testing
Note
For ORP electrodes with platinum electrode, please observe the instructions in section A
Commissioning Prepare the combination electrode for measuring as follows:
CTIVATING PLATINUM ELECTRODES.
TruLine ORP 50: Open the refilling opening for the reference electrolyte
solution.
The refilling opening must always be open during measurement!
Refilling opening
Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics and can easily be removed with deionized water.
General rules for
measuring
Note
Please keep the watering cap. It is required for storing the combination electrode. Always keep the watering cap clean.
Connect the combination electrode to the meter. Measure with the combination electrode according to the operating man-
ual of the meter and observe the following rules while doing so:
TruLine ORP 50: Make sure the refilling opening for the reference electro-
lyte solution is open.
Avoid displacement of the sample solution from one measurement to the
next by taking the following measures: – Briefly rinse the calibration beakers with the solution the beakers are to
be filled with next.
– Between measurements, rinse the combination electrode with the so-
lution specified below. Alternatively, you can also rinse the combination electrode with deionized water and then carefully dab it dry.
Immerse the combination electrode in the solution in a vertical or slightly
tilted position.
T 814241
04/2014
7
TruLine
Commissioning,
measuring, testing
T (°C)
U
[mV]
T (°C)
U
[mV]
0 +224
35
+200
Make sure the immersion depth is correct. The junction must be complete-
ly submersed in the solution. The junction is in the area of the bottom end of the shaft (see arrow).
Conversion to the
standard hydrogen electrode potential
The level of the reference
electrolyte
must be at least 2 cm above the level
of the solution.
UH = U
Meas
+ U
Ref
where: U
U U
= ORP voltage, relative to the
H
standard hydrogen electrode
= measured ORP voltage
Meas
= voltage of the reference system relative to the
Ref
standard hydrogen electrode
U
is temperature dependent and can be taken from the following table
Ref
(also refer to DIN 38404-6):
Ref
TruLine ORP 50
Ref
TruLine ORP 50
5 10 15 20 25 30
+221
40
+217
45
+214
50
+211
55
+207
60
+203
+196 +192 +188 +184 +180
8
T 814241
04/2014
Storage, Aging
TruLine
TruLine ORP 50
KCl-250 (250 ml)
During short
measuring breaks
Overnight
longer
or
Storage
Immerse the electrode in the reference electrolyte with the refilling opening open.
Combination electrode
Reference electrolyte YSI Model
(see page 11)
3 mol/l KCl, Ag+-free
Prior to the next measurement, briefly rinse the combination electrode with the test sample or deionized water.
Insert the clean electrode into the watering cap filled with reference electrolyte and shut the refilling opening.
Note During longer storage periods, salt deposits may develop on the watering cap. They do not affect the measuring characteristics and can easily be re­moved with deionized water when the combination electrode is put into oper­ation again.
Aging
Every ORP combination electrode undergoes a natural aging process. Ex­treme operating conditions can considerably shorten the lifetime of the com­bination electrode. These are:
Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro-
mides, sulfides, iodides, proteins
High temperatures Great changes in pH and temperature.
The warranty does not cover failure caused by measuring conditions and mechanical damage.
T 814241
04/2014
9
TruLine
Maintenance and cleaning
Refilling reference
electrolyte
Maintenance
and cleaning
During operation, a small amount of reference electrolyte leaks through the junction from the combination electrode into the test sample. If the level of reference electrolyte becomes too low with time, refill it through the refilling opening.
Refilling is very easy using a dropper bottle. To do so, proceed as follows: Cut off the tip of the dropping bottle at a right angle until the opening in the
tip can be seen
Open the refilling opening of the combination electrode Press the tip of the dropper bottle into the refilling opening while turning it
slightly
Pump several small quantities of the reference electrolyte into the stem
using the dropper bottle
Pull the dropper bottle out of the refilling opening while turning it slightly if
necessary.
KCl-250
10
T 814241
04/2014
Activating platinum electrodes
TruLine
Lime and hydroxide deposits
Rinse with citric acid (10 % by weight)
CleaningIL Remove water-soluble contamination by rinsing with deionized water. Re-
After cleaning Rinse the combination electrode with deionized water.
First-time
activation
installation and
required
Activation during
very long
set-up times
during
as
move other contamination as follows:
Contamination Cleaning procedure Fat and oil Rinse with water containing household wash-
ing-up liquid
Activating platinum electrodes
For the first-time activation use the activation powder from the SORT/RH reagent set (component of the SORT/RH reagent set). Immerse the moist (but not dripping) platinum electrode into the activation powder and rotate the electrode several times in the powder. Then remove the activation powder under flowing water with a soft brush (e.g. toothbrush).
When changing from oxidizing to reducing test solutions and vice versa this can result in set-up times that can take significantly more than an hour. In this case pretreatment (activation) of the platinum surface can shorten the set-up time. The type of pretreatment (reducing or oxidizing) is based on the pH val­ue and the ORP voltage (U timated for the first measurement.
The type of pretreatment can then be determined using the following diagram where U
U
in mV
H
is based on the normal hydrogen electrode:
H
1000
800
600
400
200
reducing pretreatment
0
0 2 4 6 8 10 12 14
) of the test solution where the latter must be es-
H
oxidizing
pretreatment
pH value
T 814241
04/2014
11
TruLine
Wear parts and accessories
Description
Model
Order no.
Reference electrolyte solution 3 mol/l KCl, A
KCl-250
109 705Y
Reagent set for regenerating ORP platinum e po
SORT/RH
109 730Y
Zobell ORP buffer solution (125 ml)
3682
061320
Oxidizing pretreatment
Reducing pretreatment
Immerse the platinum electrode for two to three days in a sulfuric acid chlorine solution. Chlorine powder for producing the solution is included in the SORT/RH reagent set.
Note: The diaphragm must not be immersed in the chlorine solution!
When the electrode is ready for the test immerse it in the RH 28 ORP buffer solution and wait for a sta­ble measured value.
Wear parts and accessories
g+-free (250 ml)
lectrodes, consisting of 10 g activation
wder and 30 g chlorine powder
Loading...
+ 25 hidden pages