The compact TruLab pH 1310 precision pH meter enables you to perform pH
measurements rapidly and reliably.
The TruLab pH 1310 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The USB interface can be used for data transmission to a PC and for software
updates of the meter.
1 Keypad
2 Display
3 Socket field
6ba76156e0103/2014
Page 7
TruLab pH 1310Overview
1
2
3
4
5
pH 1310P
1.2TruLab pH 1310P meter with integrated printer
The integrated printer of the TruLab pH 1310P enables to document measurements according to GLP requirements.
The information concerning the printer of the TruLab pH 1310P is
given in a separate chapter (see
TRULABPH 1310P), page 55).
section 10 PRINTER (ONLY
1 Keypad
2 Display
3 Socket field
4Printer
5 Control panel of the printer
1.3Sensors
A measuring system ready to measure consists of the TruLab pH 1310 meter
and a suitable sensor.
Suitable sensors are pH electrodes and ORP electrodes.
ba76156e0103/20147
Page 8
SafetyTruLab pH 1310
2Safety
2.1Safety information
2.1.1Safety information in the operating manual
This operating manual provides important information on the safe operation of
the meter. Read this operating manual thoroughly and make yourself familiar
with the meter before putting it into operation or working with it. The operating
manual must be kept in the vicinity of the meter so you can always find the infor
mation you need.
Important safety instructions are highlighted in this operating manual. They are
indicated by the warning symbol (triangle) in the left column. The signal word
(e.g. "CAUTION") indicates the level of danger:
WARNING
indicates a possibly dangerous situation that can lead to serious (irreversible) injury or death if the safety instruction is not
followed.
-
CAUTION
indicates a possibly dangerous situation that can lead to
slight (reversible) injury if the safety instruction is not fol
lowed.
NOTE
indicates a possibly dangerous situation where goods might be damaged
if the actions mentioned are not taken.
2.1.2Safety signs on the meter
Note all labels, information signs and safety symbols on the meter and in the
battery compartment. A warning symbol (triangle) without text refers to safety
information in this operating manual.
2.1.3Further documents providing safety information
The following documents provide additional information, which you should
observe for your safety when working with the measuring system:
•Operating manuals of sensors and other accessories
•Safety datasheets of calibration or maintenance accessories (such as buffer
solutions, electrolyte solutions, etc.)
-
8ba76156e0103/2014
Page 9
TruLab pH 1310Safety
2.2Safe operation
2.2.1Authorized use
This meter is authorized exclusively for pH and ORP measurements in the
laboratory.
Only the operation and running of the meter according to the instructions and
technical specifications given
section 13 TECHNICALDATA, page 63).
Any other use is considered unauthorized.
2.2.2Requirements for safe operation
Note the following points for safe operation:
•The meter may only be operated according to the authorized use specified
above.
•The meter may only be supplied with power by the energy sources
mentioned in this operating manual.
•The meter may only be operated under the environmental conditions
mentioned in this operating manual.
•The meter may only be opened if this is explicitly described in this operating
manual (example: Inserting the batteries).
in this operating manual is authorized (see
2.2.3Unauthorized use
The meter must not be put into operation if:
•it is visibly damaged (e.g. after being transported)
•it was stored under adverse conditions for a lengthy period of time (storing
conditions, see
section 13 TECHNICALDATA, page 63).
ba76156e0103/20149
Page 10
CommissioningTruLab pH 1310
3Commissioning
3.1Scope of delivery
pH meter TruLab pH 1310
4 batteries 1.5 V Mignon type AA
Power pack
USB cable
Stand
Stand holder
Comprehensive operating manual
Short instructions
CD-ROM with
– USB drivers
– comprehensive operating manual
– software MultiLab Importer
3.2Power supply
The TruLab pH 1310 is supplied with power in the following ways:
Mains operation with the supplied power pack
Battery operation (4 x alkaline manganese batteries, type AA)
USB operation via a connected USB-B cable
3.3Initial commissioning
Perform the following activities:
Insert the supplied batteries
For mains operation: Connect the power pack
If necessary, mount a stand
Switch on the meter (see section 4.2 SWITCHINGONTHEMETER, page 15)
Set the date and time (see section 4.4.6 EXAMPLE 2 ONNAVIGATION: SETTING
THEDATEANDTIME, page 20)
10ba76156e0103/2014
Page 11
TruLab pH 1310Commissioning
1
3.3.1Inserting the batteries
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
1 Battery compartment
1.Open the battery compartment (1) on the underside of the meter.
CAUTION
Make sure that the poles of the batteries are positioned correctly.
The ±-signs on the batteries must correspond to the ± signs in
the battery compartment.
2.Place four batteries (type Mignon AA) in the battery compartment.
3.Close the battery compartment (1).
4.Set the date and time
(see section 4.4.6 EXAMPLE 2 ONNAVIGATION: SETTINGTHEDATEAND
TIME, page 20).
ba76156e0103/201411
Page 12
CommissioningTruLab pH 1310
1
2
3.3.2Connecting the power pack
CAUTION
The line voltage at the operating site must lie within the input
voltage range of the original power pack (see
NICALDATA, page 63).
section 13 TECH-
CAUTION
Use original power packs only (see section 13 TECHNICALDATA,
page 63).
1.Connect the plug of the power pack to the socket for the power pack on
the TruLab
2.Connect the original power pack to an easily accessible power outlet.
pH 1310.
3.3.3Mounting the stand
The stand base can be mounted at the right side of the meter.
12ba76156e0103/2014
Page 13
TruLab pH 1310Operation
F1
F2
STOSTO
RCL
ARAR
4Operation
4.1General operating principles
4.1.1Keypad
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <ENTER>) generally indicates a short keystroke (under
sec) in this operating manual. A long keystroke (approx. 2 sec) is indicated
2
by the underscore behind the key symbol (e.g. <ENTER__>).
Opens the menu for measurement settings / confirms
entries
Opens the menu for system settings
<AR>Freezes the measured value (HOLD function)
Switches the AutoRead measurement on or off
ba76156e0103/201413
Page 14
OperationTruLab pH 1310
HOLD AR
25.0
6.093
pH
°C
4
3
2
5
6
7
8
1
AutoCal
01.02.2014
08:00
MenuUSB output
YSI
4.1.2Display
1 Status information (sensor)
2 Measured value (with unit)
3 Measured parameter
4 Continuous measurement control (CMC function)
5 Sensor symbol (calibration evaluation, calibration interval)
6 Measured temperature (with unit)
7 Status information (meter)
8 Softkeys and date + time
4.1.3Status information (meter)
ARStability control (AutoRead) is active
HOLDMeasured value is frozen (<AR> key)
Batteries are almost empty
14ba76156e0103/2014
Page 15
TruLab pH 1310Operation
1
2
3
4
5
6
pH
6.949
25.0
°C
01.02.2014
08:00
Menu
USB output
4.1.4Socket field
1 pH electrode
2 Reference electrode
3 Temperature sensor
4 Mini USB-B interface
5 Power pack
6 Service interface
CAUTION
Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current
circuit with current limiting).
Almost all customary sensors meet these requirements.
4.2Switching on the meter
1.Press the <On/Off> key.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test is being
performed.
The measured value display appears.
ba76156e0103/201415
Page 16
OperationTruLab pH 1310
4.3Switching off
1.Press the <On/Off> key.
Automatic switch-off
function
Display illuminationThe meter automatically switches off the display illumination if no key is
The instrument has an automatic switch-off function in order to save the
batteries (see
switches off the measuring instrument if no key is pressed for an adjustable
period.
The automatic switchoff is not active
if the communication cable is connected
if the Automatic data storage function is active, or with automatic data trans-
mission
if the power pack is connected
pressed for 30
again.
You can also generally switch the display illumination on or off (see section
7.3.1 SYSTEM, page 43).
section 7.3.1 SYSTEM, page 43). The automatic switchoff
seconds. The illumination is switched on with the next keystroke
4.4Navigation
The principles of navigation in menus and dialogs are explained in the following
sections.
4.4.1Measured value display
In the measured value display, open the setting menus with the <F1> softkey.
The current functions of the softkeys are shown on the display.
Use <F1>/[Menu] (short pressure) to open the menu for calibration and
measurement settings for the displayed measured parameter.
Use <F1__>/[Menu] (long
fig menu with the sensor-independent settings.
Use the keys of the keypad to carry out further functions such as storage or calibration (see section 4.1.1 KEYPAD, page 13). These functions are not available
in other operating situations.
4.4.2Operating modes
The meter has the following operating modes:
pressure (approx. 2 s) to open the Storage & con-
16ba76156e0103/2014
Page 17
TruLab pH 1310Operation
General
Interface
Clock
Service information
Reset
System
01.02.2014
08:00
Back
Operating
Description
mode
Measuring
The measurement data of the connected sensor are shown
in the measured value display
CalibrationThe course of a calibration with calibration information, func-
tions and settings is displayed
Storage in
The meter stores measuring data automatically or manually
memory
Transmitting data
The meter transmits measurement data and calibration
records to the USB interface automatically or manually.
SettingThe system menu or a sensor menu with submenus, set-
tings and functions is displayed
4.4.3Menus and dialogs
The menus for settings and dialogs in procedures contain further subelements.
To select a subelement, use the <><> keys. The current selection is dis-
played with a frame.
To make further settings, switch to the next higher menu level with
<F1>[Back].
Use <M> to return to the measured value display.
4.4.4Elements in menus and dialogs
Submenus
The name of the submenu is displayed at the upper edge of the frame. Submenus are opened by confirming with <MENU/ENTER>. Example:
ba76156e0103/201417
Settings
Settings are indicated by a colon. The current setting is displayed on the
right-hand side. The setting mode is opened with <MENU/ENTER>. Subse-
quently, the setting can be changed with <><> and <MENU/ENTER>.
Example:
Page 18
OperationTruLab pH 1310
Language:English
Beep
:Off
Illumination:On
Contrast
:50 %
Switchoff time
:1 h
General
01.02.2014
08:00
Back
Calibration record
Calibration data storage
Buffer
:TEC
One point calibration
:Yes
Calibration interval
:7 d
Unit for slope:mV/pH
pH
01.02.2014
08:00
Back
Calibration record
Calibration storage
Buffer
:TEC
One point calibration
:Yes
Calibration interval
:7 d
Unit for slope
:mV/pH
i
2.00 4.01 7.00 10.01
pH
01.02.2014
08:00
Back
Functions
Functions are designated by the name of the function. They are immediately
carried out by confirming with <MENU/ENTER>.
Example: Display the Calibration record function.
Messages
Information is marked by the i symbol. It cannot be selected. Example:
4.4.5Example 1 on navigation: Setting the language
1.Press the <On/Off> key.
The measured value display appears.
The instrument is in the measuring mode.
18ba76156e0103/2014
Page 19
TruLab pH 1310Operation
pH
6.949
25.0
°C
01.02.2014
08:00
MenuUSB output
System
Data storage
Storage & config
01.02.2014
08:00
Back
General
Interface
Clock
Service information
Reset
System
01.02.2014
08:00
Back
2.Using <F1__>/[Menu], open the Storage & config menu.
The instrument is in the setting mode.
3.Select the System submenu with <><>.
The current selection is displayed with a frame.
4.Open the System submenu with <MENU/ENTER>.
5.Select the General submenu with <><>.
The current selection is displayed with a frame.
6.Open the Generalsubmenu with <MENU/ENTER>.
ba76156e0103/201419
Page 20
OperationTruLab pH 1310
Language:English
Beep
:Off
Illumination:On
Contrast
:50 %
Switchoff time
:1 h
General
01.02.2014
08:00
Back
Language:English
Beep:Off
Illumination
:On
Contrast
:50 %
Switchoff time:1 h
General
01.02.2014
08:00
Back
7.Open the setting mode for the Language with <MENU/ENTER>.
8.Select the required language with <><>.
9.Confirm the setting with <MENU/ENTER>.
The meter switches to the measuring mode.
The selected language is active.
4.4.6Example 2 on navigation: Setting the date and time
The measuring instrument has a clock with a date function. The date and time
are shown in the measured value display.
When storing measured values and calibrating, the current date and time are
automatically stored as well.
The correct setting of the date and time and date format is important for the following functions and displays:
Current date and time
Calibration date
Identification of stored measured values.
Therefore, check the time at regular intervals.
20ba76156e0103/2014
Page 21
TruLab pH 1310Operation
Date format:dd.mm.yyyy
Date:01.02.2014
Time
:08:00:25
Clock
01.02.2014
08:00
Back
After a fall of the supply voltage (empty batteries), the date and time
are reset.
The date format can be switched from the display of day, month, year
(dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or mm.dd.yyyy).
1.In the measured value display:
Using <F1__>/[Menu], open the Storage & config menu.
The instrument is in the setting mode.
2.Select and confirm the System / Clock menu with <><> and
<MENU/ENTER>.
The setting menu for the date and time opens up.
3.Select and confirm the Time menu with <><> and <MENU/ENTER>.
The hours are highlighted.
4.Change and confirm the setting with <><> and <MENU/ENTER>.
The minutes are highlighted.
5.Change and confirm the setting with <><> and <MENU/ENTER>.
The seconds are highlighted.
6.Change and confirm the setting with <><> and <MENU/ENTER>.
The time is set.
7.If necessary, set the Date and Date format. The setting is made similarly
to that of the time.
8.To make further settings, switch to the next higher menu level with
[Back]<F1>.
or
Switch to the measured value display with <M>.
The instrument is in the measuring mode.
ba76156e0103/201421
Page 22
pH valueTruLab pH 1310
pH
6.949
24.8 °C
01.02.2014
08:00
Menu
USB output
5pH value
5.1Measuring
5.1.1Measuring the pH value
When connecting an earthed PC, measurements cannot be performed in
earthed media as the values would be incorrect.
The USB interface is not galvanically isolated.
To ensure the high measurement accuracy of the measuring system, always measure with a calibrated electrode (see section 5.2
C
ALIBRATION, page 25).
1.Connect the pH electrode to the meter.
2.If necessary, select the measured parameter with <M>.
3.When measuring without temperature sensor:
Temper the test sample or measure the current temperature and enter it
with <>
<>.
Stability control
(AutoRead)
& HOLD function
4.Immerse the pH electrode in the test sample.
The measured value is checked for stability (automatic stability control).
The display of the measured parameter flashes.
5.Wait for a stable measured value.
The display of the measured parameter no longer flashes.
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can activate or switch off the automatic Stability control function (see
section 7.1.1 CHANGINGTHESETTINGSFORPH MEASUREMENTS, page 38).
22ba76156e0103/2014
Page 23
TruLab pH 1310pH value
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the Stability control function and the HOLD func-
tion with <AR> or <M> at any time.
2.
Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured value is recognized. The progress bar disappears and the display
of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
Criteria for a stable
measured value
You can prematurely terminate the Stability control function manually with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
Using <ENTER>, start a further measurement with stability control.
3.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured
Time intervalStability in the time interval
parameter
pH value15 seconds∆ : Better than 0.01 pH
Temperature15 seconds∆ : Better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
ba76156e0103/201423
Page 24
pH valueTruLab pH 1310
5.1.2Measuring the temperature
For reproducible pH and ORP measurements, it is essential to measure the
temperature of the test sample.
You have the following options to measure the temperature:
Automatic measurement of the temperature with the temperature sensor
(NTC 30 or Pt 1000) integrated in the sensor.
Measurement with an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
Temperature
sensor
Resolution of the
temp. display
Temp. measurement
yes0.1 °CAutomatic with tempera-
ture sensor
-1 °CManual
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
1.Measure the current temperature of the test sample.
2.Set the temperature value with <><>.
or
In the <F1>/[Menu]/Man. temperature menu, set the temperature value
with <><>.
24ba76156e0103/2014
Page 25
TruLab pH 1310pH value
5.2Calibration
5.2.1Why calibrate?
pH electrodes age. This changes the zero point (asymmetry) and slope of the
pH
electrode. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope of the
electrode.
Thus, you should calibrate at regular intervals.
5.2.2When do you have to calibrate?
After connecting a sensor
Routinely within the framework of the company quality assurance
When the calibration interval has expired
5.2.3Automatic calibration (AutoCal)
Make sure that in the sensor menu, Buffer menu, the buffer set is correctly
selected (see
Use any one to five buffer solutions of the selected buffer set in ascending or
descending order.
Below, calibration with YSI buffers (YSI) is described. When other buffer sets
are used, other nominal buffer values are displayed. Apart from that, the proce
dure is identical.
If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed.
1.Connect the pH electrode to the meter.
2.Keep the buffer solutions ready.
When measuring without temperature sensor:
Temper the buffer solutions or measure the current temperature.
3.In the measured value display, select the measured parameter pH or mV
with <M>.
-
4.Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).
ba76156e0103/201425
Page 26
pH valueTruLab pH 1310
pH1
-180.0
24.8 °C
AutoCal YSI
01.02.2014
08:00
Buffer
mV
pH
-180.0
24.8 °C
AutoCal YSI
10.00
01.02.2014
08:00
AR
Buffer
mV
5.Thoroughly rinse the electrode with deionized water.
6.Immerse the electrode in the first buffer solution.
7.When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
8.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
Continuing with two-
point calibration
26ba76156e0103/2014
9.Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
10.If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the electrode.
11.Thoroughly rinse the electrode with deionized water.
12.Immerse the electrode in the second buffer solution.
13.When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
Page 27
TruLab pH 1310pH value
pH
0.0
24.8 °C
AutoCal YSI
7.000
01.02.2014
08:00
AR
Buffer
mV
pH
180.0
24.8 °C
AutoCal YSI
4.00
01.02.2014
08:00
AR
Buffer
mV
14.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
15.Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
Continuing with
three- to five-point
calibration
16.If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.
17.Thoroughly rinse the electrode with deionized water.
18.Immerse the electrode in the next buffer solution.
19.When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
20.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
ba76156e0103/201427
Page 28
pH valueTruLab pH 1310
pH
0.0
24.8 °C
ConCal
1
01.02.2014
08:00
Buffer
mV
21.Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
22.If necessary, use <M> to finish calibration or
Continue calibrating using the next buffer with <ENTER>.
Calibration is automatically completed after the last buffer of a buffer set has been measured. Then the calibration record is displayed.
The calibration line is determined by linear regression.
5.2.4Manual calibration (ConCal)
Make sure that in the sensor menu, Buffer menu, the ConCal buffer set is
selected (see
page 38).
Use any one to five buffer solutions in ascending or descending order.
If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed.
1.Connect the pH electrode to the meter.
The pH measuring window is displayed.
2.Keep the buffer solutions ready.
When measuring without temperature sensor:
Temper the buffer solutions or measure the current temperature.
3.In the measured value display, select the measured parameter pH or
mV with <M>.
4.Start the calibration with <CAL>.
The calibration display appears.
5.Thoroughly rinse the electrode with deionized water.
28ba76156e0103/2014
Page 29
TruLab pH 1310pH value
pH
0.0
24.8 °C
ConCal
7.000
01.02.2014
08:00
Buffer
mV
AR
pH
7.000
24.8 °C
ConCal
1
01.02.2014
08:00
Buffer
6.Immerse the electrode in the first buffer solution.
7.When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
8.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
9.Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the setting of the nominal buffer value appears.
10.Set the nominal buffer value for the measured temperature with
<><>.
11.Accept the set calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
12.If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
Continuing with two-
point calibration
ba76156e0103/201429
13.Thoroughly rinse the electrode with deionized water.
(-59.2 mV/pH at 25 °C) and determines the zero point of the electrode.
Page 30
pH valueTruLab pH 1310
pH
4.035
24.8 °C
ConCal
2
01.02.2014
08:00
Buffer
14.Immerse the electrode in the second buffer solution.
15.When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
16.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
17.Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the setting of the nominal buffer value appears.
Continuing with
three- to five-point
calibration
18.Set the nominal buffer value for the measured temperature with
<><>.
19.Accept the set calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
20.Finish the calibration procedure as a two-point calibration with <M>.
The calibration record is displayed.
21.Thoroughly rinse the electrode with deionized water.
22.Immerse the electrode in the next buffer solution.
23.When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
24.Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
25.Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the setting of the nominal buffer value appears.
30ba76156e0103/2014
Page 31
TruLab pH 1310pH value
pH
9.958
24.8 °C
ConCal
3
01.02.2014
08:00
Buffer
26.Set the nominal buffer value for the measured temperature with
<><>.
27.Accept the set calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
28.Use <M> to finish calibration or
Continue calibrating using the next buffer with <ENTER>.
After the fifth buffer has been measured the calibration is automatically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
5.2.5Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
Determined
values
1-pointAsy Zero point = Asy
2-pointAsy
Slp.
3-point to
5-point
Asy
Slp.
Displayed calibration data
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
Zero point = Asy
Slope = Slp.
Zero point = Asy
Slope = Slp.
The calibration line is calculated by linear regression.
You can display the slope in the units, mV/pH or % (see section
ba76156e0103/201431
7.1.1 C
HANGINGTHESETTINGSFORPH MEASUREMENTS, page 38).
Page 32
pH valueTruLab pH 1310
5.2.6Calibration data
The calibration record is automatically transmitted to the interface
after calibrating.
Displaying the cali-
bration data
Displaying the cali-
bration data storage
The calibration record of the last calibration is to be found under the menu item,
<F1> / [Menu]Calibration
/ Calibration record. To open it, press the <CAL__>
key in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with <F2>[USB output].
The calibration records of the last calibrations (up to 10) are available in the
menu, <F1>/[Menu] / Calibration / Calibration data storage and in the menu,
<F1__>/[Menu] / Storage & config/Data
Menu itemSetting/func-
storage / Calibration data storage.
Description
tion
Calibration /
-Displays the calibration record.
Calibration data storage /Display
Further options:
Scroll through the calibration
or
Data storage /
Calibration data storage /Display
records with <><>.
Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].
Quit the display with <F1>/
[Back] or <MENU/ENTER>.
Calibration evalua-
tion
Switch directly to the mea-
sured value display with <M>.
Calibration /
Calibration data stor-
-Outputs the calibration records to
the interface.
age /
Output to USB
or
Data storage /
Calibration data storage /Output to USB
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
Voltage 23.0 mV 24.0 °C
Voltage 3-177.0 mV 24.0 °C
Slope-60.2 mV/pH
Asymmetry 4.0 mV
Sensor+++
etc...
DisplayCalibration
record
+++
++
+
-
Clean the electrode according to the
electrode operating manual
ErrorError
Error elimination (see section 12 WHAT
TODOIF..., page 60)
Zero point
[mV]
Slope
[mV/pH]
- 15 ... + 15-60,5 ... -58,0
- 20 ... <-15
>-58,0 ... -57,0
or
>+15 ... + 20
- 25 ... <- 20
or
>+ 20 ... + 25
- 30 ... <- 25
or
->+25 ... + 30
-61,0 ... <-60,5
or
>-57,0 ... -56,0
-62,0 ... <-61,0
or
>-56,0 ... -50,0
- 30 ... + 30-62,0 ... -50,0
Calibration record
(example)
ba76156e0103/201433
Page 34
pH valueTruLab pH 1310
25.0
6.093
1
2
3
4
pH
°C
0
14
01.02.2014
08:00
Menu
USB output
5.2.7Continuous measurement control (CMC function)
The Continuous Measurement Control (CMC function) facilitates to evaluate
the current measured value instantly and definitely.
After each successful calibration the scale of the pH measuring range is
displayed in the measured value display. Here you can very clearly see
whether or not the current measured value is in the calibrated part of the
measuring range.
The following information is displayed:
1 Measuring range for which a valid calibration is available (white).
Measured values in this range are suitable for documentation.
2 Measuring range for which no valid calibration is available (shaded).
Measured values in this range are not suitable for documentation. If
necessary, calibrate the meter with buffers covering this measuring
range.
If the current measured value is outside the calibrated range, this
area is shaded stronger.
If a measured value is outside the measuring range pH 0 - 14, overflow arrows are displayed at the left or right edge of the measuring
range.
3 Currently measured pH value (needle)
4 Marking lines for all nominal buffer values used with the last valid cali-
bration
The limits of the calibrated range are determined by the buffers used for calibration:
Lower limit: Buffer with lowest pH value - 2 pH units
Upper limit: Buffer with highest pH value + 2 pH units
34ba76156e0103/2014
Page 35
TruLab pH 1310ORP voltage
U
157.0 mV
24.8 °C
01.02.2014
08:00
Menu
USB output
6ORP voltage
6.1Measuring
6.1.1Measuring the ORP
When connecting an earthed PC, measurements cannot be performed in
earthed media as the values would be incorrect.
The USB interface is not galvanically isolated.
1.Connect the ORP electrode to the meter.
2.If necessary, select the mV display with <M>.
3.When measuring without temperature sensor:
Measure the temperature of the test sample and enter it with <> <>.
4.Immerse the ORP electrode in the test sample.
The measured value is checked for stability (automatic stability control).
The display of the measured parameter flashes.
Stability control
(AutoRead )
5.Wait for a stable measured value.
The display of the measured parameter no longer flashes.
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values.
You can activate or switch off the automatic Stability control function (see
section 7.1.1 CHANGINGTHESETTINGSFORPH MEASUREMENTS, page 38).
The measured parameter flashes on the display
Criteria for a stable
measured value
ba76156e0103/201435
as soon as the measured value is outside the stability range
when you switch over between the measured parameters with <M>.
when the automatic Stability control is switched off.
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Page 36
ORP voltageTruLab pH 1310
Manually starting
the stability control
Measured parame-
Time intervalStability in the time interval
ter
ORP voltage15 seconds∆ : Better than 0.3 mV
Temperature15 seconds∆ : Better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
You can start the Stability control manually at any time, irrespective of the setting for automatic Stability control (see section 7.1.1 C
FORPH MEASUREMENTS, page 38) in the Measurement menu.
HANGINGTHESETTINGS
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
2.Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the measured
parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears and the display of the
measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
Freezing the mea-
sured value (HOLD
function)
You can prematurely terminate the Stability control function manually with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
Using <ENTER>, start a further measurement with Stability control.
3.
or
Release the frozen measured value again with <AR>.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
With the HOLD function, you can freeze the current measured value. The displayed measured value stops changing until you switch the HOLD function off.
If the HOLD function is active, you can, e.g. start a manual measurement with stability control.
1.Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
36ba76156e0103/2014
Page 37
TruLab pH 1310ORP voltage
2.Release the frozen measured value again with <AR>.
The HOLD function is switched off.
The [HOLD] status display disappears.
6.1.2Measuring the temperature
For reproducible ORP measurements, it is essential to measure the temperature of the test sample.
You have the following options to measure the temperature:
Automatic measurement of the temperature by the temperature sensor
(NTC 30 or Pt 1000) integrated in electrode.
Measurement by an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
Temperature
sensor
Resolution of the
temp. display
Temp. measurement
yes0.1 °CAutomatic with temper-
ature sensor
-1 °CManual
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
1.Measure the current temperature of the test sample.
2.Set the temperature value with <><>.
or
In the <F1>/[Menu]/Man. temperature menu, set the temperature value
with <><>.
6.2ORP calibration
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.
ba76156e0103/201437
Page 38
SettingsTruLab pH 1310
7Settings
7.1Measurement settings
7.1.1Changing the settings for pH measurements
The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required parameter in the measured value display and press the <F1>/[menu] or <MENU/ENTER> key. After completing the settings, switch to the measured value display with <M>.
Default settings are printed in bold.
Menu itemPossible set-
ting
Calibration /Calibration record
Calibration /
Calibration data
storage /Display
-Displays the calibration record of
-Displays the calibration record.
Description
the last calibration.
Further options:
Scroll through the calibration
records with <><>.
Output the displayed calibration
record to the interface with
<F2>/[USB output].
Output all calibration records to
the interface with <F2__>[USB output].
Quit the display with <F1>/
[Back] or <MENU/ENTER>.
Switch directly to the measured
value display with <M>.
Calibration / Calibration data storage / Output to USB
Calibration / Serial
number (probe)
38ba76156e0103/2014
-Outputs the calibration records to
the interface.
Entry of the serial number of the
connected sensor.
The serial number is output in the
calibration record.
Change the content of the high-
lighted position with <><>.
Go to the next position with
<F2>/[].
When the serial number has
been completely entered, confirm with <ENTER>.
Page 39
TruLab pH 1310Settings
Menu itemPossible set-
ting
Calibration /BufferYSI
ConCal
NIST/DIN
...
Calibration /One
point calibration
Calibration /Cali-
Yes
No
1 ... 7 ... 999 dCalibration interval for the pH elec-
bration interval
Calibration /Unit for
slope
mV/pH
%
Man. temperature-25 ... +25 ...
+130 °C
Description
Buffer sets to be used for pH calibration (see section 5.2 CALIBRA-
TION, page 25).
Quick calibration with 1 buffer
trode (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring window.
Unit of the slope.
The % display refers to the Nernst
slope of -59.2
mV/pH (100 x deter-
mined slope/Nernst slope).
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Temperature unit°C
°F
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperatures are displayed
with the selected unit.
Resolution pH0.001
Resolution of the pH display
0.01
0.1
Stability controlOn
Off
Switches on or off the automatic
stability control during measure
ment (see section 5.1.1 MEASUR-
INGTHEPH VALUE, page 22)
Reset-Resets all sensor settings to the
delivery condition (see
section
7.4.1 RESETTINGTHEMEASURE-
MENTSETTINGS, page 45).
7.1.2Buffer sets for calibration
-
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.
ba76156e0103/201439
Page 40
SettingsTruLab pH 1310
No.Buffer set *pH valuesat
1YSI *4.000
25 °C
7.000
10.000
2ConCalAnyAny
3NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers
1.679
4.006
6.865
25 °C
9.180
12.454
4TEC
Technical buffers
2.000
4.010
25 °C
7.000
10.011
5Merck 1*4.000
20°C
7.000
9.000
6Merck 2 *1.000
20°C
6.000
8.000
13.000
7Merck 3 *4.660
6.880
9.220
8Merck 4 *2.000
4.000
7.000
10.000
9Merck 5 *4.010
7.000
10.000
10DIN 192671.090
4.650
6.790
9.230
11Mettler Toledo USA *1.679
4.003
7.002
10.013
12Mettler Toledo EU *1.995
4.005
7.002
9.208
20°C
20°C
25 °C
25 °C
25 °C
25 °C
40ba76156e0103/2014
Page 41
TruLab pH 1310Settings
No.Buffer set *pH valuesat
13Fisher *2.007
4.002
7.004
10.002
14Fluka BS *4.006
6.984
8.957
15Radiometer *1.678
4.005
7.000
9.180
16Baker *4.006
6.991
10.008
17Metrohm *3.996
7.003
8.999
18Beckman *4.005
7.005
10.013
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
19Hamilton Duracal *4.005
7.002
10.013
20Precisa *3.996
7.003
8.999
21Reagecon TEC *2.000
4.010
7.000
10.000
22Reagecon 20 *2.000
4.000
7.000
10.000
13.000
25 °C
25 °C
25 °C
20°C
ba76156e0103/201441
Page 42
SettingsTruLab pH 1310
No.Buffer set *pH valuesat
23Reagecon 25 *2.000
25 °C
4.000
7.000
10.000
13.000
24Chemsolute *2,000
20 °C
4,000
7,000
10,000
25USABlueBook *4,000
25 °C
7,000
10,000
*Brand names or trade names are trademarks of their respective owners pro-
tected by law.
The buffers are selected in the menu, pH / <F1>/[Menu] / Calibra-tion / Buffer (see 7.1.1 C
MENTS, PAGE 38).
HANGINGTHESETTINGSFORPH MEASURE-
Setting the calibra-
tion interval
7.1.3Calibration interval
The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
1.Open the menu for measurement settings with [Menu]<F1>.
2.In the Calibration / Calibration interval menu, set the calibration interval
<><>.
with
3.Confirm the setting with <ENTER>.
4.Quit the menu with <M>.
7.2Settings for ORP measurements
The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required parame-
42ba76156e0103/2014
Page 43
TruLab pH 1310Settings
ter in the measured value display and press the <F1>/[menu] or <MENU/
ENTER> key. After completing the settings, switch to the measured value dis-
play with <M>.
Default settings are printed in bold.
Menu itemPossible
Description
setting
Man. temperature-25 ... +25
... +130
°C
Entry of the manually determined temperature. For measurements without
temperature sensor only.
Temperature unit°C
°F
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperatures are displayed with the
selected unit.
Resolution mV0.1
Resolution of the mV display
1
Stability controlOn
Off
Switches on or off the automatic stability
control during measurement (see
section Stability control (AutoRead) &
HOLD function, page 22 )
Reset-Resets all sensor settings to the delivery
condition (see
THEMEASUREMENTSETTINGS, page 45).
section 7.4.1 RESETTING
7.3Sensor-independent settings
7.3.1System
To open the Storage & config menu, press the <F1__>[Menu] key in the measured value display. After completing the settings, switch to the measured
value display with <M>.
Default settings are printed in bold.
Menu itemPossible set-
ting
System / General /
Language
Deutsch
English
(further)
System / General /
Beep
System / General /
Illumination
On
Off
Auto
On
Off
System / General /
0 ... 50 ... 100 %Changing the display contrast
Contrast
Description
Selects the menu language
Switches on/off the beep on
keystroke
Switches the display illumination on/off
ba76156e0103/201443
Page 44
SettingsTruLab pH 1310
Menu itemPossible set-
Description
ting
System / General /
Switchoff time
System / Interface /
Baud rate
10 min ... 1h ...
24 h
1200, 2400,
4800, 9600,
Adjusts the switch-off time
Baud rate of the data interface
19200
System / Interface /
Output format
ASCII
CSV
Output format for data transmission (see section 9 TRANS-
MITTINGDATA (USB INTERFACE),
page 53)
System / Interface /
Decimal separator
System / Interface /
Output header
Dot (xx.x)
Comma (xx,x)
Decimal separator
Output of a header for Output
format: CSV
System / PrinterSettings for the integrated prin-
ter of the TruLab pH 1310 P.
For details, see section 10
PRINTER (ONLY
TRULABPH 1310P), page 55
System /ClockDate format
Date
Time
System /Service
information
Time and date settings (see
section 4.4.6 EXAMPLE 2 ON
NAVIGATION: SETTINGTHEDATE
ANDTIME, page 20)
Hardware version and software
version of the meter are dis
-
played.
System /Reset-Resets the system settings to
the delivery condition (see
sec-
tion 7.4.2 RESETTINGTHESYS-
TEMSETTINGS, page 45).
7.3.2Data storage
This menu contains all functions to display, edit and erase stored measured
values and calibration records (see
section 8 DATASTORAGE, page 47).
7.4Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
44ba76156e0103/2014
Page 45
TruLab pH 1310Settings
7.4.1 Resetting the measurement settings
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
The following settings for pH measurements are reset to the default settings
with the Reset function:
SettingDefault settings
Buffer YSI
Cal. interval7 d
Unit for slopemV/pH
Measured parameterpH
Resolution pH0.001
Resolution mV0.1
Asymmetry0 mV
Slope-59.2 mV
Man. temperature25 °C
One point calibrationNo
Stability controlOn
Temperature unit°C
The sensor settings are reset under the Reset menu item in the menu for calibration and measurement settings. To open it, press the <F1>/[Menu] or
<MENU/ENTER> key in the measured value display.
7.4.2Resetting the system settings
The following system settings can be reset to the default condition:
SettingDefault settings
LanguageEnglish
BeepOn
Baud rate4800 Baud
Output formatASCII
Contrast50 %
IlluminationAuto
Switchoff time1 h
ba76156e0103/201445
Page 46
SettingsTruLab pH 1310
The system settings are reset in the menu, Storage & config / System / Reset.
To open the Storage & config menu, press the <F1__>[Menu] key in the
measured value display.
46ba76156e0103/2014
Page 47
TruLab pH 1310Data storage
01.02.2014 11:24:16
pH 7.000 24.8 °C AR +++
ID number:1
Continue
Manual data storage4 From 500
01.02.2014
08:00
Back
Data storage full. Erase?
Yes
No
Warning
01.02.2014
08:00
Back
8Data storage
You can transmit measured values (datasets) to the data storage:
Manual storage (see section 8.1 M
Automatic storage at intervals (see section 8.2 A
INTERVALS, page 48)
ANUALSTORAGE, page 47)
UTOMATICSTORAGEAT
Each storage process transmits the current dataset to the interface at the same
time.
8.1Manual storage
You can transmit a measurement dataset to the data storage as follows. The
dataset is at the same time output to the interface:
1.Press the <STO> key shortly.
The menu for manual storage appears.
If the storage is fullThe following window appears if all 500 storage locations are occupied:
ba76156e0103/201447
2.If necessary, change and confirm the ID number (1 ... 10000) with
<><> and <MENU/ENTER>.
The dataset is stored. The meter switches to the measured value display.
You have the following options:
Page 48
Data storageTruLab pH 1310
1
ID number
1
Interval
30 s
Duration
180 min
Continue
0d03h00min
0 1d17h33min
Automatic data storage
01.02.2014
08:00
2
3
Back
To erase the entire storage, confirm Yes.
To cancel the storage process and switch to the measured value display,
confirm No. Then you can e.g. transmit the stored data to a PC (see section
8.3.1 E
erase the storage (see section 8.3.2 E
AGE, page 52).
DITINGTHEMEASUREDVALUESTORAGE, page 50) and subsequently
RASINGTHEMEASUREMENTDATASTOR-
8.2Automatic storage at intervals
The storage interval (Interval) determines the time interval between automatic
storage processes. Each storage process transmits the current dataset to the
interface at the same time.
Configuring the
automatic storage
function
1.Press the <STO__> key.
The menu for automatic storage appears.
1 Adjusted entire storage duration
2 Max. available storage duration
3 Graphical display of the storage usage
48ba76156e0103/2014
Page 49
TruLab pH 1310Data storage
1
2
pH
6.949
25.0
°C
0d03h00min
01.02.2014
08:00
0
14
USB outputEco mode
SettingsYou can configure the automatic storage function with the following settings:
Menu itemPossible set-
ting
ID number1 ... 10000ID number for the dataset series.
Interval1 s, 5 s, 10 s,
s, 1 min,
30
min,
5
10
min,
min,
15
min,
30
min
60
Duration1 min ... x minStorage duration.
Description
Storage interval.
The lower limit of the storage
interval can be restricted by the
number of free storage locations.
The upper limit is restricted by the
storage duration.
Specifies after which time the
automatic storage should be ter
minated.
The lower limit of the storage
duration is restricted by the stor
age interval.
The upper limit is restricted by the
number of free storage locations.
-
-
Starting the auto-
matic storage func-
tion
To start the automatic storage function, select Continue with <><> and
confirm with <MENU/ENTER>. The meter switches to the measured value display.
1 Remaining storage duration
2 Graphical display of the storage duration
The active automatic storage function can be recognized by the progress bar
in the function display. The progress bar indicates the remaining storage dura
tion.
-
Energy saving mode
ba76156e0103/201449
([Eco mode])
If the automatic storage function is activated, only the following
keys are active: Softkeys, <M>, <STO__> and <On/Off>. The
other keys and the automatic switch-off function are deactivated.
If the automatic storing function is active, the meter provides an energy saving
mode ([Eco mode]) to avoid unnecessary energy consumption. The energy
Page 50
Data storageTruLab pH 1310
Stop automatic storage?
Yes
No
Warning
01.02.2014
08:00
Back
saving mode switches off functions of the meter that are not required for the
automatic storage of measurement data (such as the display). By pressing any
key the energy saving mode is switched off again.
Terminating the
automatic storage
function prema
turely
Proceed as follows to switch off the automatic storage function before the
adjusted storage duration has expired:
-
1.Press the <STO__> key.
The following window appears.
2.Using <><>, select Yes and confirm with <MENU/ENTER>.
The meter switches to the measured value display.
The automatic storage function is terminated.
Editing the data stor-
age
8.3Measurement data storage
8.3.1Editing the measured value storage
The contents of the manual or automatic measurement data storage can be
shown on the display and output to the interface.
Each of the measurement data storages has a function to erase the entire
contents.
The storage is edited in the menu, Storage & config / Data storage. To open
the Storage & config menu, press the <F1__>[Menu] key in the measured
value display.
Open the manual or automatic storage directly with the <RCL> or <RCL__>
key.
The settings are explained here using the manual storage as an
example. The same settings and functions are available for the
automatic storage.
pH 6.012 24.8 °C, AR, S: +++
________________________________________
etc...
Settings
Menu itemSetting/
function
Data storage /
-Displays all measurement datasets
Manual data storage /
Display
Data storage /
-Erases the entire manual measure-
Manual data storage /
Erase
Data storage /
-Outputs all stored measurement
Manual data storage /
Output to USB
Description
page by page.
Further options:
Scroll through the datasets with
<><>.
Output the displayed dataset to
the interface with <F2>/[USB output].
Quit the display with <F1>/
[Back].
ment data storage.
All calibration data remains stored
when this action is performed.
data to the interface.
Display presentation
of a dataset
Representation of a
dataset
(USB output)
ba76156e0103/201451
Page 52
Data storageTruLab pH 1310
Quitting the displayTo quit the display of stored measurement datasets, you have the following
options:
Switch directly to the measured value display with <M>.
Quit the display and move to the next higher menu level with <F1>/[Back].
8.3.2Erasing the measurement data storage
Erasing the measurement data storage (see section 8.3.1 EDITINGTHE
MEASUREDVALUESTORAGE, page 50).
8.3.3Measurement dataset
A complete dataset consists of:
ID number
Date/time
Measured value of the connected sensor
Measured temperature value of the connected sensor or manually set tem-
perature
AutoRead info: The AR indicator appears with the measured value if the
AutoRead criterion was met while storing (stable measured value). Otherwise, there is no AR indicator.
Calibration evaluation: +++, ++, +, -, or no evaluation
8.3.4Storage locations
The TruLab pH 1310 meter has two measurement data storages. The
measured values recorded either manually or automatic are stored separately
in individual measurement data
storages.
StorageMaximum number of datasets
Manual data storage500
Automatic data storage5000
52ba76156e0103/2014
Page 53
TruLab pH 1310Transmitting data (USB interface)
9Transmitting data (USB interface)
9.1Options for data transmission
Via the USB interface you can transmit data to a PC. The following table shows
which data are transmitted to the interface in which way:
DataControlOperation / description
Current
measured values of all connected
sensors
Stored measured values
Calibration
records
Manual With <F2>/[USB output].
Simultaneously with every manual
storage process (see section 8.1
M
ANUALSTORAGE, page 47).
Automatic, at
intervals
With <F2__>/[USB output].
Then you can set the transmission
interval.
Simultaneously with every auto-
matic storage process (see section
8.2 A
UTOMATICSTORAGEATINTER-
VALS, page 48).
Manual Displayed dataset with <F2>[USB
output] after calling up from the
storage.
All datasets with the Output to USB
function.
(see section 8.3.1 E
MEASUREDVALUESTORAGE, page
DITINGTHE
50).
Manual Calibration record with <F2>/[USB
output]
(see section 5.2.6 C
DATA, page 32).
ALIBRATION
Automatic At the end of a calibration proce-
dure.
The following rule applies: With the exception of the menus, shortly
pressing the <F2>[USB output] key generally outputs the display
contents to the interface (displayed measured values, measurement datasets, calibration records).
ba76156e0103/201453
Page 54
Transmitting data (USB interface)TruLab pH 1310
9.2Connecting a PC
Connect the TruLab pH 1310 to the PC via the USB interface.
NOTE
When connecting an earthed PC, measurements cannot be performed in
earthed media as the values would be incorrect.
The USB interface is not galvanically isolated.
Installation of the
USB driver on the
PC
System requirements of the PC for installation of the USB driver:
PC with Pentium processor or higher with at least one free USB connection
and CD-ROM drive
Windows 2000, XP, Vista.
1.Insert the supplied installation CD in the CD drive of your PC.
2.Install the driver from the CD.
Follow the Windows installation instructions as necessary.
3.Connect the TruLab pH 1310 to the PC via the USB interface.
The meter is listed as a virtual COM interface among the connections in
the Windows instrument manager.
9.3MultiLab Importer
With the aid of the MultiLab Importer software, you can record and evaluate
measuring data with a PC.
More detailed information can be found in the MultiLab Importer
operating manual.
54ba76156e0103/2014
Page 55
TruLab pH 1310Printer (only TruLab pH 1310P)
2
1
3
4
10Printer (only TruLab pH 1310P)
1 Operating LED
– Illuminated: Printer is switched on
Switching on the
printer
– Flashing: Printer is switched off
2 Control panel with the keys <ON/FEED> and <OFF>
3 Lever to open the paper compartment
4 Lid of the paper compartment
10.1Commissioning / switching the printer on or off
1.Connect the power pack to the TruLab pH 1310 P.
The LED lights up green. The printer is ready for operation.
or
If the printer was switched off (LED flashing):
Switch on the printer with <OFF>.
The LED lights up green. The printer is ready for operation.
If there is a USB connection (such as to a PC) the data are output
only to the PC.
Switching off the
printer
ba76156e0103/201455
1.Switch off the printer with <OFF>.
The LED flashes. The printer is switched off.
Page 56
Printer (only TruLab pH 1310P)TruLab pH 1310
10.2Operation / printing
Data are only output to the printer if the following conditions are met
Data are transmitted manually or automatically (see section 9.1 O
FORDATATRANSMISSION, page 53)
PTIONS
the printer is switched on (LED illuminated)
there is no USB connection.
10.3Printer settings
To open the Storage & config menu, press the <F1_>[Menu] key in the measured value display. After completing the settings, switch to the measured
value display with <M>.
Default settings are printed in bold.
Menu itemPossible set-
ting
System / Printer /
Font size
12x20
8x16
7x16
Description
Select the font size for the printer
A print sample (character set of the
printer) to view the available font
sizes can be printed with <OFF_>.
System / Printer /
Print test page
-The printer prints the meter informa-
tion from the System / Service infor-mation menu.
The current printer settings are
used for the printout.
10.4Maintenance
10.4.1 Changing the roll of paper (thermal paper)
1.Pull the lever (3) until the lid (4) of the paper compartment opens.
2.If necessary, remove any remains of the old roll of paper.
3.Insert the new roll of paper. The start of the paper should poke out of the
paper compartment.
4.Close the lid (4) of the paper compartment so that it clicks into place.
5.If necessary, move the printer paper forward with <ON/FEED>.
56ba76156e0103/2014
Page 57
TruLab pH 1310Printer (only TruLab pH 1310P)
10.5What to do if... / printer
CauseRemedy
Integrated printer
does not print
– Printer switched off
(LED flashing)
– No power pack connected– Connect the power pack
– USB cable connected– Disconnect the USB cable from the
– Switch on the printer
(LED illuminated)
meter
Printer operating -
paper not being
printed
Integrated printer
prints automatically
– Function "automatic storing at
intervals" with long intervals is
switched on
– Switch off the function (see section
UTOMATICSTORAGEATINTER-
8.2 A
VALS, page 48)
– No paper available– Insert a roll of paper
CauseRemedy
– Paper inserted the wrong way up– Turn the roll of paper around and
insert it with the other side upwards
CauseRemedy
– The function "Automatic storage
at intervals" or "Transmit data
automatic at intervals" is
switched on
– Switch off the function (see section
UTOMATICSTORAGEATINTER-
8.2 A
VALS, page 48 or
section 9.1 O
TRANSMISSION, page 53)
PTIONSFORDATA
ba76156e0103/201457
Page 58
Maintenance, cleaning, disposalTruLab pH 1310
1
11Maintenance, cleaning, disposal
11.1Maintenance
11.1.1 General maintenance activities
The only maintenance activity required is replacing the batteries.
See the relevant operating manuals of the electrodes for instructions on maintenance.
11.1.2 Replacing the batteries
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
1 Battery compartment
1.Open the battery compartment (1) on the underside of the
meter.
2.Remove the four batteries from the battery compartment.
3.Place four new batteries (type Mignon AA) in the battery compartment.
CAUTION
Make sure that the poles of the batteries are positioned correctly.
The ±-signs on the batteries must correspond to the ± signs in
the battery compartment.
58ba76156e0103/2014
Page 59
TruLab pH 1310Maintenance, cleaning, disposal
4.Close the battery compartment (1).
The date (day) flashes on the display.
5.Set the date and time (see section 4.4.6 EXAMPLE 2 ONNAVIGATION:
SETTINGTHEDATEANDTIME, page 20)
Dispose of used batteries according to the local regulations of your
country.
End users within the European Union are obligated to return used
batteries
set up for recycling
Batteries are marked with the crossed-out waste container
symbol. Therefore, they may not be disposed with the domestic
waste.
(even ecologically compatible ones) to a collection point
purposes.
11.2Cleaning
Occasionally wipe the outside of the measuring instrument with a damp, lintfree cloth. Disinfect the housing with isopropanol as required.
CAUTION
The housing is made of synthetic material (ABS). Thus, avoid
contact with acetone or similar detergents that contain sol
vents. Remove any splashes immediately.
-
11.3Packing
This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the
meter against damage during transport.
11.4Disposal
At the end of its operational lifetime, the meter must be returned to the
disposal or return system statutory in your country. If you have any questions,
please contact your supplier.
ba76156e0103/201459
Page 60
What to do if...TruLab pH 1310
12What to do if...
12.1pH/ORP
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Error message
OFL, UFL
Error message,
Error
The measured value is outside the measuring range.
CauseRemedy
pH electrode:
– Air bubble in front of the junction– Remove air bubble
– Air in the junction– Extract air or moisten junction
– Cable broken– Replace electrode
– Gel electrolyte dried out– Replace electrode
– Measured value outside the mea-
suring range of the meter
CauseRemedy
pH electrode:
– The values determined for zero
point and slope of the electrode
are outside the allowed limits.
Batteries4 x 1.5 V alkali-manganese batteries, type
AA
Rechargeable batteries
4 x 1.2 V NiMH rechargeable batteries,
AA
type
(no charging function)
Operational lifeUp to 1000 h without / 150 h with illumination
Power pack Kuantech Co. Ltd. KSAC 0900110W1UV-1
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 270 mA
Output: 9 V = / 1.1 A
Connection max. overvoltage category II
Primary plugs contained in the scope of
delivery: Euro, US, UK and Australian.
Sensor inputInput resistance> 5 * 1012 ohm
Input current< 1 * 10
-12
A
USB interfaceTypeUSB 1.1
USB B (device), data output
Baud rateAdjustable:
1200, 2400, 4800, 9600, 19200 Baud
Data bits8
Stop bits2
ParityNone
HandshakeRTS/CTS
Cable length max. 3 m
64ba76156e0103/2014
Page 65
TruLab pH 1310Technical data
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with the limits
for a Class A digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely
to cause harmful interference in which case the user will be required to
correct the interference at his own expense.
Changes or modifications not expressly approved by the manufacturer
could void the user‘s authority to operate the equipment.
Guidelines
and norms used
EMCEC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Meter safetyEC directive 2006/95/EC
EN 61010-1
UL 61010-1
CAN/CSA-C22.2#61010-1
IP protection classEN 60529
ba76156e0103/201465
Page 66
Firmware updateTruLab pH 1310
14Firmware update
Available firmware updates are provided on the Internet.
With the firmware update program and a PC you can update the firmware of the
TruLab
For the update you have to connect the meter to a PC.
For the update via the USB interface, the following is required:
a free USB interface (virtual COM port) on the PC
the driver for the USB interface (on the enclosed CD-ROM)
the USB cable (included in the scope of delivery of the TruLab pH 1310).
1.Install the downloaded firmware update on a PC.
pH 1310 to the newest version.
An update folder is created in the Windows start menu.
If an update folder already exists for the meter (or meter type), the new
data is displayed there.
2.In the windows start menu, open the update folder and start the firmware
update program.
3.Using the USB interface cable, connect the TruLab pH 1310 to a USB
interface (virtual COM port) of the PC.
4.Switch on the TruLab pH 1310.
5.In the firmware update program, start the update process with OK.
6.Follow the instructions of the firmware update program.
During the programming process, a corresponding message and a
progress bar (in %) are displayed.
The programming process takes approx. three minutes. A terminatory
message is displayed after a successful programming process. The
firmware update is completed.
7.Disconnect the TruLab pH 1310 from the PC.
The TruLab pH 1310 is ready for operation again.
After switching the meter off and on you can check whether the meter has taken
over the new software version (see
VERSIONISINTHEMETER, PAGE 62).
YOUWANTTOKNOWWHICHSOFTWARE
66ba76156e0103/2014
Page 67
TruLab pH 1310Glossary
15Glossary
pH/ORP
Asymmetrysee zero point
Electromotive force of
an electrode
JunctionThe junction is a porous body in the housing wall of reference elec-
ORP voltageThe ORP is caused by oxidizing or reducing substances dissolved in
pH valueThe pH value is a measure of the acidic or basic effect of an aqueous
PotentiometryName of a measuring technique. The signal (depending on the mea-
The electromotive force U of the combination electrode is the measurable electromotive force of an electrode in a solution. It equals the
sum of all the galvanic voltages of the combination electrode. Its
dependency on the pH results in the electrode function, which is characterized by the parameters, slope and zero point.
trodes or electrolyte bridges. It arranges the electrical contact
between two solutions and makes the electrolyte exchange more difficult. The expression, junction, is also used for ground or junctionless transitions.
water if these substances become effective on an electrode surface
(e. g. a gold or platinum surface).
solution. It corresponds to the negative decadic logarithm of the molal
hydrogen ions activity divided by the unit of the molality. The practical
pH value is the value of a pH measurement.
sured parameter) of the electrode is the electrical potential. The electrical current remains constant.
SlopeThe slope of a linear calibration function.
Zero pointThe zero point of a pH combination electrode is the pH value at which
the electromotive force of the pH combination electrode at a specified
temperature is zero. Normally, this is at 25 °C.
General information
AdjustingTo manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains
within the tolerance.
AutoRangeName of the automatic selection of the measuring range.
CalibrationComparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct.
Often, this expression is also used when the measuring system is
adjusted at the same time (see adjusting).
Measured parameterThe measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D.O. concentration.
ba76156e0103/201467
Page 68
GlossaryTruLab pH 1310
Measured valueThe measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3
m; 0.5 s; 5.2 A; 373.15 K).
MolalityMolality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
vent.
Reset Restoring the original condition of all settings of a measuring system.
ResolutionSmallest difference between two measured values that can be dis-
played by a meter.
Stability control
Function to control the measured value stability.
(AutoRead )
Standard solutionThe standard solution is a solution where the measured value is
known.
It is used to calibrate a measuring system.
Temperature functionName of a mathematical function expressing the temperature behav-
ior of a test sample, a sensor or part of a sensor.
Test sampleDesignation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
(937) 767-7241
Monday through Friday, 8:00 AM to 5:00 PM ET
Fax:(937) 767-1058
Email:environmental@ysi.com
Mail:YSI Incorporated
1725 Brannum Lane
Yellow Springs, OH 45387
USA
Internet:www.ysi.com
When placing an order please have the following information available:
YSI account number (if available)Name and Phone Number
Model number or brief descriptionBilling and shipping address
QuantityPurchase Order or Credit Card
17.2Service Information
YSI has authorized service centers throughout the United States and
Internationally. For the nearest service center information, please visit
www.ysi.com and click ‘Support’ or contact YSI Technical Support
directly at 800-897-4151.
When returning a product for service, include the Product Return form
with cleaning certification. The form must be completely filled out for an
YSI Service Center to accept the instrument for service. The Product
Return form may be downloaded at
‘Support‘ tab.
www.ysi.com and clicking on the
ba76156e0103/2014
71
Page 72
Contact InformationTruLab pH 1310
72
ba76156e0103/2014
Page 73
Page 74
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We're 12,500 people unified in a common purpose: creating innovative solutions
to meet our world's water needs. Developing new technologies that will improve
the way water is used, conserved, and re-used in the future is central to our work.
We move, treat, analyze, and return water to the environment, and we help people
use water efficiently, in their homes, buildings, factories and farms. In more than
150 countries, we have strong, long-standing relationships with customers who
know us for our powerful combination of leading product brands and applications
expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com
Xylem |'m|zīlə
YSI
1725 Brannum Lane
Yellow Springs, OH 45387
Tel: +1 937-767-7241; 800-765-4974
Fax: +1 937-767-1058
Email: environmental@ysi.com
Web: www.ysi.com