YSI 1310P User Manual

Page 1
OPERATIONS MANUAL
pH 1310P
pH 1310
pH 1310P
ba76156e01 03/2014
TruLab pH 1310(P)
pH METER
Page 2
TruLab pH 1310
For the most recent version of the manual, please visit www.ysi.com.
Contact YSI
1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241
800-765-4974 Email: environmental@ysi.com Internet: www.ysi.com
Copyright © 2014 Xylem Inc.
2 ba76156e01 03/2014
Page 3
TruLab pH 1310 Contents
TruLab pH 1310 - Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.1 TruLab pH 1310 meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2 TruLab pH 1310P meter with integrated printer . . . . . . . . . . . . . 7
1.3 Sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1.1 Safety information in the operating manual . . . . . . . . . . . . 8
2.1.2 Safety signs on the meter . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1.3 Further documents providing safety information . . . . . . . . 8
2.2 Safe operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.1 Authorized use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2 Requirements for safe operation . . . . . . . . . . . . . . . . . . . . 9
2.2.3 Unauthorized use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Initial commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3.1 Inserting the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3.2 Connecting the power pack . . . . . . . . . . . . . . . . . . . . . . . 12
3.3.3 Mounting the stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1 General operating principles . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1.1 Keypad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.1.3 Status information (meter) . . . . . . . . . . . . . . . . . . . . . . . . 14
4.1.4 Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Switching on the meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.3 Switching off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.4 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.4.1 Measured value display. . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.4.2 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.4.3 Menus and dialogs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.4.4 Elements in menus and dialogs. . . . . . . . . . . . . . . . . . . . 17
4.4.5 Example 1 on navigation: Setting the language . . . . . . . 18
4.4.6 Example 2 on navigation: Setting the date and time . . . . 20
ba76156e01 03/2014 3
Page 4
Contents TruLab pH 1310
5 pH value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.1.1 Measuring the pH value . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.1.2 Measuring the temperature . . . . . . . . . . . . . . . . . . . . . . 24
5.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.2.1 Why calibrate? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.2.2 When do you have to calibrate?. . . . . . . . . . . . . . . . . . . 25
5.2.3 Automatic calibration (AutoCal) . . . . . . . . . . . . . . . . . . . 25
5.2.4 Manual calibration (ConCal) . . . . . . . . . . . . . . . . . . . . . . 28
5.2.5 Calibration points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.2.6 Calibration data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.2.7 Continuous measurement control (CMC function) . . 34
6 ORP voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.1.1 Measuring the ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.1.2 Measuring the temperature . . . . . . . . . . . . . . . . . . . . . . 37
6.2 ORP calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
7.1 Measurement settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
7.1.1 Changing the settings for pH measurements . . . . . . . . . 38
7.1.2 Buffer sets for calibration . . . . . . . . . . . . . . . . . . . . . . . . 39
7.1.3 Calibration interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.2 Settings for ORP measurements . . . . . . . . . . . . . . . . . . . . . . . 42
7.3 Sensor-independent settings . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.3.1 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.3.2 Data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.4 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.4.1 Resetting the measurement settings . . . . . . . . . . . . . . . 45
7.4.2 Resetting the system settings. . . . . . . . . . . . . . . . . . . . . 45
8 Data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8.1 Manual storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8.2 Automatic storage at intervals . . . . . . . . . . . . . . . . . . . . . . 48
8.3 Measurement data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.3.1 Editing the measured value storage . . . . . . . . . . . . . . . . 50
8.3.2 Erasing the measurement data storage . . . . . . . . . . . . . 52
8.3.3 Measurement dataset. . . . . . . . . . . . . . . . . . . . . . . . . . . 52
8.3.4 Storage locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
9 Transmitting data (USB interface). . . . . . . . . . . . . . . . . . . . . 53
9.1 Options for data transmission . . . . . . . . . . . . . . . . . . . . . . . . . 53
9.2 Connecting a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
9.3 MultiLab Importer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
10 Printer (only TruLab pH 1310P) . . . . . . . . . . . . . . . . . . . . . . . 55
10.1 Commissioning / switching the printer on or off . . . . . . . . . . . . 55
4 ba76156e01 03/2014
Page 5
TruLab pH 1310 Contents
10.2 Operation / printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
10.3 Printer settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
10.4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
10.4.1 Changing the roll of paper (thermal paper). . . . . . . . . . . 56
10.5 What to do if... / printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
11 Maintenance, cleaning, disposal. . . . . . . . . . . . . . . . . . . . . . 58
11.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
11.1.1 General maintenance activities . . . . . . . . . . . . . . . . . . . 58
11.1.2 Replacing the batteries. . . . . . . . . . . . . . . . . . . . . . . . . . 58
11.2 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
11.3 Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
11.4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
12 What to do if.... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
12.1 pH/ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
12.2 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
13 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
13.1 Measuring ranges, resolution, accuracy . . . . . . . . . . . . . . . . . 63
13.2 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
14 Firmware update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
15 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
16 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
17 Contact Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
17.1 Ordering & Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . 71
17.2 Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
ba76156e01 03/2014 5
Page 6
Overview TruLab pH 1310
1
2
3
pH 1310

1 Overview

1.1 TruLab pH 1310 meter

The compact TruLab pH 1310 precision pH meter enables you to perform pH measurements rapidly and reliably.
The TruLab pH 1310 provides the maximum degree of operating comfort, reli­ability and measuring certainty for all applications.
The USB interface can be used for data transmission to a PC and for software updates of the meter.
1 Keypad
2 Display
3 Socket field
6 ba76156e01 03/2014
Page 7
TruLab pH 1310 Overview
1
2
3
4
5
pH 1310P

1.2 TruLab pH 1310P meter with integrated printer

The integrated printer of the TruLab pH 1310P enables to document measure­ments according to GLP requirements.
The information concerning the printer of the TruLab pH 1310P is given in a separate chapter (see TRULAB PH 1310P), page 55).
section 10 PRINTER (ONLY
1 Keypad
2 Display
3 Socket field
4Printer
5 Control panel of the printer

1.3 Sensors

A measuring system ready to measure consists of the TruLab pH 1310 meter and a suitable sensor.
Suitable sensors are pH electrodes and ORP electrodes.
ba76156e01 03/2014 7
Page 8
Safety TruLab pH 1310

2 Safety

2.1 Safety information

2.1.1 Safety information in the operating manual

This operating manual provides important information on the safe operation of the meter. Read this operating manual thoroughly and make yourself familiar with the meter before putting it into operation or working with it. The operating manual must be kept in the vicinity of the meter so you can always find the infor mation you need.
Important safety instructions are highlighted in this operating manual. They are indicated by the warning symbol (triangle) in the left column. The signal word (e.g. "CAUTION") indicates the level of danger:
WARNING
indicates a possibly dangerous situation that can lead to seri­ous (irreversible) injury or death if the safety instruction is not followed.
-
CAUTION
indicates a possibly dangerous situation that can lead to slight (reversible) injury if the safety instruction is not fol lowed.
NOTE
indicates a possibly dangerous situation where goods might be damaged if the actions mentioned are not taken.

2.1.2 Safety signs on the meter

Note all labels, information signs and safety symbols on the meter and in the battery compartment. A warning symbol (triangle) without text refers to safety information in this operating manual.

2.1.3 Further documents providing safety information

The following documents provide additional information, which you should observe for your safety when working with the measuring system:
Operating manuals of sensors and other accessories
Safety datasheets of calibration or maintenance accessories (such as buffer solutions, electrolyte solutions, etc.)
-
8 ba76156e01 03/2014
Page 9
TruLab pH 1310 Safety

2.2 Safe operation

2.2.1 Authorized use

This meter is authorized exclusively for pH and ORP measurements in the laboratory.
Only the operation and running of the meter according to the instructions and technical specifications given section 13 TECHNICAL DATA, page 63).
Any other use is considered unauthorized.

2.2.2 Requirements for safe operation

Note the following points for safe operation:
The meter may only be operated according to the authorized use specified above.
The meter may only be supplied with power by the energy sources mentioned in this operating manual.
The meter may only be operated under the environmental conditions mentioned in this operating manual.
The meter may only be opened if this is explicitly described in this operating manual (example: Inserting the batteries).
in this operating manual is authorized (see

2.2.3 Unauthorized use

The meter must not be put into operation if:
it is visibly damaged (e.g. after being transported)
it was stored under adverse conditions for a lengthy period of time (storing conditions, see
section 13 TECHNICAL DATA, page 63).
ba76156e01 03/2014 9
Page 10
Commissioning TruLab pH 1310

3 Commissioning

3.1 Scope of delivery

pH meter TruLab pH 1310
4 batteries 1.5 V Mignon type AA
Power pack
USB cable
Stand
Stand holder
Comprehensive operating manual
Short instructions
CD-ROM with
– USB drivers
– comprehensive operating manual
– software MultiLab Importer

3.2 Power supply

The TruLab pH 1310 is supplied with power in the following ways:
Mains operation with the supplied power pack
Battery operation (4 x alkaline manganese batteries, type AA)
USB operation via a connected USB-B cable

3.3 Initial commissioning

Perform the following activities:
Insert the supplied batteries
For mains operation: Connect the power pack
If necessary, mount a stand
Switch on the meter (see section 4.2 SWITCHING ON THE METER, page 15)
Set the date and time (see section 4.4.6 EXAMPLE 2 ON NAVIGATION: SETTING
THE DATE AND TIME, page 20)
10 ba76156e01 03/2014
Page 11
TruLab pH 1310 Commissioning
1

3.3.1 Inserting the batteries

You can operate the meter either with normal batteries or with rechargeable batteries (Ni-MH). In order to charge the batteries, an external charging device is required.
1 Battery compartment
1. Open the battery compartment (1) on the underside of the meter.
CAUTION
Make sure that the poles of the batteries are positioned cor­rectly. The ±-signs on the batteries must correspond to the ± signs in the battery compartment.
2. Place four batteries (type Mignon AA) in the battery compartment.
3. Close the battery compartment (1).
4. Set the date and time
(see section 4.4.6 EXAMPLE 2 ON NAVIGATION: SETTING THE DATE AND
TIME, page 20).
ba76156e01 03/2014 11
Page 12
Commissioning TruLab pH 1310
1
2

3.3.2 Connecting the power pack

CAUTION
The line voltage at the operating site must lie within the input voltage range of the original power pack (see
NICAL DATA, page 63).
section 13 TECH-
CAUTION
Use original power packs only (see section 13 TECHNICAL DATA, page 63).
1. Connect the plug of the power pack to the socket for the power pack on
the TruLab
2. Connect the original power pack to an easily accessible power outlet.
pH 1310.

3.3.3 Mounting the stand

The stand base can be mounted at the right side of the meter.
12 ba76156e01 03/2014
Page 13
TruLab pH 1310 Operation
F1
F2
STOSTO
RCL
ARAR

4Operation

4.1 General operating principles

4.1.1 Keypad

In this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. <ENTER>) generally indicates a short keystroke (under
sec) in this operating manual. A long keystroke (approx. 2 sec) is indicated
2 by the underscore behind the key symbol (e.g. <ENTER__>).
<F1>: <F1__>: <F2>:
Softkeys providing situation dependent functions, e.g.:
<F1>/[Menu]: Opens the menu for measurement settings <F1__>/[Menu]: Opens the menu for system settings
<F2__>:
<On/Off>: Switches the meter on or off
<M>: Selects the measured parameter / Quits the settings
<CAL>: <CAL__>:
<STO>: <STO__>:
<RCL>: <RCL__>:
Calls up the calibration procedure Displays the calibration data
Saves a measured value manually Opens the menu for the automatic save function
Displays the manually stored measured values Displays the automatically stored measured values
Menu control, navigation
<><>: <__><__>:
<ENTER>: <ENTER__>:
Increments, decrements values Increments, decrements values continuously
Opens the menu for measurement settings / confirms entries Opens the menu for system settings
<AR> Freezes the measured value (HOLD function)
Switches the AutoRead measurement on or off
ba76156e01 03/2014 13
Page 14
Operation TruLab pH 1310
HOLD AR
25.0
6.093
pH
°C
4
3
2
5 6 7 8
1
AutoCal
01.02.2014 08:00
Menu USB output
YSI

4.1.2 Display

1 Status information (sensor)
2 Measured value (with unit)
3 Measured parameter
4 Continuous measurement control (CMC function)
5 Sensor symbol (calibration evaluation, calibration interval)
6 Measured temperature (with unit)
7 Status information (meter)
8 Softkeys and date + time

4.1.3 Status information (meter)

AR Stability control (AutoRead) is active
HOLD Measured value is frozen (<AR> key)
Batteries are almost empty
14 ba76156e01 03/2014
Page 15
TruLab pH 1310 Operation
1
2
3
4
5
6
pH
6.949
25.0
°C
01.02.2014 08:00
Menu
USB output

4.1.4 Socket field

1 pH electrode
2 Reference electrode
3 Temperature sensor
4 Mini USB-B interface
5 Power pack
6 Service interface
CAUTION
Only connect sensors to the meter that cannot return any volt­ages or currents that are not allowed (> SELV and > current circuit with current limiting). Almost all customary sensors meet these requirements.

4.2 Switching on the meter

1. Press the <On/Off> key. The meter performs a self-test. The display shows the manufacturer's logo while the self-test is being performed. The measured value display appears.
ba76156e01 03/2014 15
Page 16
Operation TruLab pH 1310

4.3 Switching off

1. Press the <On/Off> key.
Automatic switch-off
function
Display illumination The meter automatically switches off the display illumination if no key is
The instrument has an automatic switch-off function in order to save the batteries (see switches off the measuring instrument if no key is pressed for an adjustable period.
The automatic switchoff is not active
if the communication cable is connected
if the Automatic data storage function is active, or with automatic data trans-
mission
if the power pack is connected
pressed for 30 again.
You can also generally switch the display illumination on or off (see section
7.3.1 SYSTEM, page 43).
section 7.3.1 SYSTEM, page 43). The automatic switchoff
seconds. The illumination is switched on with the next keystroke

4.4 Navigation

The principles of navigation in menus and dialogs are explained in the following sections.

4.4.1 Measured value display

In the measured value display, open the setting menus with the <F1> softkey. The current functions of the softkeys are shown on the display.
Use <F1>/[Menu] (short pressure) to open the menu for calibration and
measurement settings for the displayed measured parameter.
Use <F1__>/[Menu] (long
fig menu with the sensor-independent settings.
Use the keys of the keypad to carry out further functions such as storage or cali­bration (see section 4.1.1 KEYPAD, page 13). These functions are not available in other operating situations.

4.4.2 Operating modes

The meter has the following operating modes:
pressure (approx. 2 s) to open the Storage & con-
16 ba76156e01 03/2014
Page 17
TruLab pH 1310 Operation
General Interface Clock Service information Reset
System
01.02.2014 08:00
Back
Operating
Description
mode
Measuring
The measurement data of the connected sensor are shown in the measured value display
Calibration The course of a calibration with calibration information, func-
tions and settings is displayed
Storage in
The meter stores measuring data automatically or manually
memory
Transmit­ting data
The meter transmits measurement data and calibration records to the USB interface automatically or manually.
Setting The system menu or a sensor menu with submenus, set-
tings and functions is displayed

4.4.3 Menus and dialogs

The menus for settings and dialogs in procedures contain further subelements.
To select a subelement, use the <><> keys. The current selection is dis-
played with a frame.
To make further settings, switch to the next higher menu level with
<F1>[Back].
Use <M> to return to the measured value display.

4.4.4 Elements in menus and dialogs

Submenus
The name of the submenu is displayed at the upper edge of the frame. Sub­menus are opened by confirming with <MENU/ENTER>. Example:
ba76156e01 03/2014 17
Settings
Settings are indicated by a colon. The current setting is displayed on the right-hand side. The setting mode is opened with <MENU/ENTER>. Subse- quently, the setting can be changed with <><> and <MENU/ENTER>. Example:
Page 18
Operation TruLab pH 1310
Language: English Beep
: Off
Illumination: On Contrast
: 50 %
Switchoff time
: 1 h
General
01.02.2014 08:00
Back
Calibration record Calibration data storage Buffer
: TEC
One point calibration
: Yes
Calibration interval
: 7 d
Unit for slope: mV/pH
pH
01.02.2014 08:00
Back
Calibration record Calibration storage Buffer
: TEC
One point calibration
: Yes
Calibration interval
: 7 d
Unit for slope
: mV/pH
i
2.00 4.01 7.00 10.01
pH
01.02.2014 08:00
Back
Functions
Functions are designated by the name of the function. They are immediately carried out by confirming with <MENU/ENTER>. Example: Display the Calibration record function.
Messages
Information is marked by the i symbol. It cannot be selected. Example:

4.4.5 Example 1 on navigation: Setting the language

1. Press the <On/Off> key. The measured value display appears. The instrument is in the measuring mode.
18 ba76156e01 03/2014
Page 19
TruLab pH 1310 Operation
pH
6.949
25.0
°C
01.02.2014 08:00
Menu USB output
System Data storage
Storage & config
01.02.2014 08:00
Back
General Interface Clock Service information Reset
System
01.02.2014 08:00
Back
2. Using <F1__>/[Menu], open the Storage & config menu. The instrument is in the setting mode.
3. Select the System submenu with <><>. The current selection is displayed with a frame.
4. Open the System submenu with <MENU/ENTER>.
5. Select the General submenu with <><>. The current selection is displayed with a frame.
6. Open the General submenu with <MENU/ENTER>.
ba76156e01 03/2014 19
Page 20
Operation TruLab pH 1310
Language: English Beep
: Off
Illumination: On Contrast
: 50 %
Switchoff time
: 1 h
General
01.02.2014 08:00
Back
Language: English Beep: Off Illumination
: On
Contrast
: 50 %
Switchoff time: 1 h
General
01.02.2014 08:00
Back
7. Open the setting mode for the Language with <MENU/ ENTER>.
8. Select the required language with <><>.
9. Confirm the setting with <MENU/ENTER>. The meter switches to the measuring mode. The selected language is active.

4.4.6 Example 2 on navigation: Setting the date and time

The measuring instrument has a clock with a date function. The date and time are shown in the measured value display. When storing measured values and calibrating, the current date and time are automatically stored as well.
The correct setting of the date and time and date format is important for the fol­lowing functions and displays:
Current date and time
Calibration date
Identification of stored measured values.
Therefore, check the time at regular intervals.
20 ba76156e01 03/2014
Page 21
TruLab pH 1310 Operation
Date format: dd.mm.yyyy Date: 01.02.2014 Time
: 08:00:25
Clock
01.02.2014 08:00
Back
After a fall of the supply voltage (empty batteries), the date and time are reset.
The date format can be switched from the display of day, month, year (dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or mm.dd.yyyy).
1. In the measured value display: Using <F1__>/[Menu], open the Storage & config menu. The instrument is in the setting mode.
2. Select and confirm the System / Clock menu with <><> and <MENU/ENTER>. The setting menu for the date and time opens up.
3. Select and confirm the Time menu with <><> and <MENU/ ENTER>. The hours are highlighted.
4. Change and confirm the setting with <><> and <MENU/ENTER>. The minutes are highlighted.
5. Change and confirm the setting with <><> and <MENU/ENTER>. The seconds are highlighted.
6. Change and confirm the setting with <><> and <MENU/ENTER>. The time is set.
7. If necessary, set the Date and Date format. The setting is made similarly to that of the time.
8. To make further settings, switch to the next higher menu level with [Back]<F1>. or Switch to the measured value display with <M>. The instrument is in the measuring mode.
ba76156e01 03/2014 21
Page 22
pH value TruLab pH 1310
pH
6.949
24.8 °C
01.02.2014 08:00
Menu
USB output

5 pH value

5.1 Measuring

5.1.1 Measuring the pH value

When connecting an earthed PC, measurements cannot be performed in earthed media as the values would be incorrect. The USB interface is not galvanically isolated.
To ensure the high measurement accuracy of the measuring sys­tem, always measure with a calibrated electrode (see section 5.2 C
ALIBRATION, page 25).
1. Connect the pH electrode to the meter.
2. If necessary, select the measured parameter with <M>.
3. When measuring without temperature sensor: Temper the test sample or measure the current temperature and enter it with <>
<>.
Stability control
(AutoRead)
& HOLD function
4. Immerse the pH electrode in the test sample. The measured value is checked for stability (automatic stability control). The display of the measured parameter flashes.
5. Wait for a stable measured value. The display of the measured parameter no longer flashes.
The stability control function (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproduc­ibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can activate or switch off the automatic Stability control function (see section 7.1.1 CHANGING THE SETTINGS FOR PH MEASUREMENTS, page 38).
22 ba76156e01 03/2014
Page 23
TruLab pH 1310 pH value
1. Freeze the measured value with <AR>. The [HOLD] status indicator is displayed. The HOLD function is active.
You can terminate the Stability control function and the HOLD func- tion with <AR> or <M> at any time.
2.
Using <ENTER>, activate the Stability control function manually. The [AR] status indicator appears while the measured value is assessed as not stable. A progress bar is displayed and the display of the measured parameter flashes. The [HOLD][AR] status indicator appears as soon as a stable mea­sured value is recognized. The progress bar disappears and the display of the measured parameter stops flashing. The current measurement data is output to the interface. Measurement data meeting the stability control criterion is marked by AR.
Criteria for a stable
measured value
You can prematurely terminate the Stability control function manu­ally with <ENTER> at any time. If the Stability control function is prematurely terminated, the current measurement data are output to the interface without the AutoRead info.
Using <ENTER>, start a further measurement with stability control.
3. or Release the frozen measured value again with <AR> or <M>. The [AR] status display disappears. The display switches back to the previous indication.
The Stability control function checks whether the measured values are stable within the monitored time interval.
Measured
Time interval Stability in the time interval
parameter
pH value 15 seconds : Better than 0.01 pH
Temperature 15 seconds : Better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer.
ba76156e01 03/2014 23
Page 24
pH value TruLab pH 1310

5.1.2 Measuring the temperature

For reproducible pH and ORP measurements, it is essential to measure the temperature of the test sample. You have the following options to measure the temperature:
Automatic measurement of the temperature with the temperature sensor
(NTC 30 or Pt 1000) integrated in the sensor.
Measurement with an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring mode:
Temperature
sensor
Resolution of the
temp. display
Temp. measurement
yes 0.1 °C Automatic with tempera-
ture sensor
- 1 °C Manual
If you wish to measure (or calibrate) without temperature sensor, proceed as follows:
1. Measure the current temperature of the test sample.
2. Set the temperature value with <><>. or In the <F1>/[Menu]/Man. temperature menu, set the temperature value with <><>.
24 ba76156e01 03/2014
Page 25
TruLab pH 1310 pH value

5.2 Calibration

5.2.1 Why calibrate?

pH electrodes age. This changes the zero point (asymmetry) and slope of the pH
electrode. As a result, an inexact measured value is displayed. Calibration determines and stores the current values of the zero point and slope of the electrode. Thus, you should calibrate at regular intervals.

5.2.2 When do you have to calibrate?

After connecting a sensor
Routinely within the framework of the company quality assurance
When the calibration interval has expired

5.2.3 Automatic calibration (AutoCal)

Make sure that in the sensor menu, Buffer menu, the buffer set is correctly selected (see
7.1.1 CHANGING THE SETTINGS FOR PH MEASUREMENTS, PAGE 38).
Use any one to five buffer solutions of the selected buffer set in ascending or descending order.
Below, calibration with YSI buffers (YSI) is described. When other buffer sets are used, other nominal buffer values are displayed. Apart from that, the proce dure is identical.
If single-point calibration was set in the menu, the calibration proce­dure is automatically finished with the measurement of buffer solu­tion 1 and the calibration record is displayed.
1. Connect the pH electrode to the meter.
2. Keep the buffer solutions ready.
When measuring without temperature sensor: Temper the buffer solutions or measure the current temperature.
3. In the measured value display, select the measured parameter pH or mV
with <M>.
-
4. Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).
ba76156e01 03/2014 25
Page 26
pH value TruLab pH 1310
pH 1
-180.0
24.8 °C
AutoCal YSI
01.02.2014 08:00
Buffer
mV
pH
-180.0
24.8 °C
AutoCal YSI
10.00
01.02.2014 08:00
AR
Buffer
mV
5. Thoroughly rinse the electrode with deionized water.
6. Immerse the electrode in the first buffer solution.
7. When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
8. Start the measurement with <ENTER>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes.
Continuing with two-
point calibration
26 ba76156e01 03/2014
9. Wait for the end of the measurement with stability control or accept the calibration value with <ENTER>. The calibration display for the next buffer appears (voltage display).
10. If necessary, finish the calibration procedure as a single-point calibration with <M>. The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the zero point of the elec­trode.
11. Thoroughly rinse the electrode with deionized water.
12. Immerse the electrode in the second buffer solution.
13. When measuring without temperature sensor: Measure the temperature of the buffer and enter it with <><>.
Page 27
TruLab pH 1310 pH value
pH
0.0
24.8 °C
AutoCal YSI
7.000
01.02.2014 08:00
AR
Buffer
mV
pH
180.0
24.8 °C
AutoCal YSI
4.00
01.02.2014 08:00
AR
Buffer
mV
14. Start the measurement with <ENTER>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes.
15. Wait for the measurement with stability control to be completed or ter­minate the stability control and take over the calibration value with <ENTER>. The calibration display for the next buffer appears (voltage display).
Continuing with
three- to five-point
calibration
16. If necessary, finish the calibration procedure as a two-point calibration with <M>. The calibration record is displayed.
17. Thoroughly rinse the electrode with deionized water.
18. Immerse the electrode in the next buffer solution.
19. When measuring without temperature sensor: Measure the temperature of the buffer and enter it with <><>.
20. Start the measurement with <ENTER>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes.
ba76156e01 03/2014 27
Page 28
pH value TruLab pH 1310
pH
0.0
24.8 °C
ConCal
1
01.02.2014 08:00
Buffer
mV
21. Wait for the measurement with stability control to be completed or termi­nate the stability control and take over the calibration value with <ENTER>. The calibration display for the next buffer appears (voltage display).
22. If necessary, use <M> to finish calibration or Continue calibrating using the next buffer with <ENTER>.
Calibration is automatically completed after the last buffer of a buf­fer set has been measured. Then the calibration record is dis­played.
The calibration line is determined by linear regression.

5.2.4 Manual calibration (ConCal)

Make sure that in the sensor menu, Buffer menu, the ConCal buffer set is selected (see page 38).
section 7.1.1 CHANGING THE SETTINGS FOR PH MEASUREMENTS,
Use any one to five buffer solutions in ascending or descending order.
If single-point calibration was set in the menu, the calibration proce­dure is automatically finished with the measurement of buffer solu­tion 1 and the calibration record is displayed.
1. Connect the pH electrode to the meter. The pH measuring window is displayed.
2. Keep the buffer solutions ready. When measuring without temperature sensor: Temper the buffer solutions or measure the current temperature.
3. In the measured value display, select the measured parameter pH or mV with <M>.
4. Start the calibration with <CAL>. The calibration display appears.
5. Thoroughly rinse the electrode with deionized water.
28 ba76156e01 03/2014
Page 29
TruLab pH 1310 pH value
pH
0.0
24.8 °C
ConCal
7.000
01.02.2014 08:00
Buffer
mV
AR
pH
7.000
24.8 °C
ConCal
1
01.02.2014 08:00
Buffer
 
6. Immerse the electrode in the first buffer solution.
7. When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
8. Start the measurement with <ENTER>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes.
9. Wait for the end of the measurement with stability control or accept the calibration value with <ENTER>. The calibration display for the setting of the nominal buffer value appears.
10. Set the nominal buffer value for the measured temperature with <><>.
11. Accept the set calibration value with <ENTER>. The calibration display for the next buffer appears (voltage display).
12. If necessary, finish the calibration procedure as a single-point calibration with <M>. The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
Continuing with two-
point calibration
ba76156e01 03/2014 29
13. Thoroughly rinse the electrode with deionized water.
(-59.2 mV/pH at 25 °C) and determines the zero point of the elec­trode.
Page 30
pH value TruLab pH 1310
pH
4.035
24.8 °C
ConCal
2
01.02.2014 08:00
Buffer
 
14. Immerse the electrode in the second buffer solution.
15. When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
16. Start the measurement with <ENTER>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes.
17. Wait for the measurement with stability control to be completed or termi­nate the stability control and take over the calibration value with <ENTER>. The calibration display for the setting of the nominal buffer value appears.
Continuing with
three- to five-point
calibration
18. Set the nominal buffer value for the measured temperature with <><>.
19. Accept the set calibration value with <ENTER>. The calibration display for the next buffer appears (voltage display).
20. Finish the calibration procedure as a two-point calibration with <M>. The calibration record is displayed.
21. Thoroughly rinse the electrode with deionized water.
22. Immerse the electrode in the next buffer solution.
23. When measuring without temperature sensor: Measure the temperature of the buffer and enter it with <><>.
24. Start the measurement with <ENTER>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes.
25. Wait for the measurement with stability control to be completed or termi­nate the stability control and take over the calibration value with <ENTER>. The calibration display for the setting of the nominal buffer value appears.
30 ba76156e01 03/2014
Page 31
TruLab pH 1310 pH value
pH
9.958
24.8 °C
ConCal
3
01.02.2014 08:00
Buffer
 
26. Set the nominal buffer value for the measured temperature with
<><>.
27. Accept the set calibration value with <ENTER>. The calibration display for the next buffer appears (voltage display).
28. Use <M> to finish calibration or Continue calibrating using the next buffer with <ENTER>.
After the fifth buffer has been measured the calibration is automat­ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.

5.2.5 Calibration points

Calibration can be performed using one to five buffer solutions in any order (single-point to five-point calibration). The meter determines the following values and calculates the calibration line as follows:
Determined values
1-point Asy Zero point = Asy
2-point Asy
Slp.
3-point to 5-point
Asy Slp.
Displayed calibration data
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
Zero point = Asy
Slope = Slp.
Zero point = Asy
Slope = Slp.
The calibration line is calculated by lin­ear regression.
You can display the slope in the units, mV/pH or % (see section
ba76156e01 03/2014 31
7.1.1 C
HANGING THE SETTINGS FOR PH MEASUREMENTS, page 38).
Page 32
pH value TruLab pH 1310

5.2.6 Calibration data

The calibration record is automatically transmitted to the interface after calibrating.
Displaying the cali-
bration data
Displaying the cali-
bration data storage
The calibration record of the last calibration is to be found under the menu item, <F1> / [Menu]Calibration
/ Calibration record. To open it, press the <CAL__>
key in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface, e. g. to a PC, with <F2>[USB output].
The calibration records of the last calibrations (up to 10) are available in the menu, <F1>/[Menu] / Calibration / Calibration data storage and in the menu,
<F1__>/[Menu] / Storage & config/Data
Menu item Setting/func-
storage / Calibration data storage.
Description
tion
Calibration /
- Displays the calibration record.
Calibration data stor­age /Display
Further options: Scroll through the calibration
or
Data storage / Calibration data stor­age /Display
records with <><>.
Output the displayed calibra-
tion record to the interface with <F2>/[USB output].
Quit the display with <F1>/
[Back] or <MENU/ENTER>.
Calibration evalua-
tion
Switch directly to the mea-
sured value display with <M>.
Calibration / Calibration data stor-
- Outputs the calibration records to the interface.
age / Output to USB
or
Data storage / Calibration data stor­age /Output to USB
After calibrating, the meter automatically evaluates the calibration. The zero point and slope are evaluated separately. The worse evaluation of both is taken into account. The evaluation appears on the display and in the calibration record.
32 ba76156e01 03/2014
Page 33
TruLab pH 1310 pH value
inoLab TruLab pH 1310 Ser. no. 11292113
CALIBRATION pH
01.02.2014 15:55
Ser. no. 10501234 YSI Buffer 1 4.00 Buffer 2 7.00 Buffer 3 10.00 Voltage
1 184.0 mV 24.0 °C
Voltage 2 3.0 mV 24.0 °C Voltage 3 -177.0 mV 24.0 °C Slope -60.2 mV/pH Asymmetry 4.0 mV Sensor +++
etc...
Display Calibration
record
+++
++
+
-
Clean the electrode according to the electrode operating manual
Error Error
Error elimination (see section 12 WHAT
TO DO IF..., page 60)
Zero point [mV]
Slope [mV/pH]
- 15 ... + 15 -60,5 ... -58,0
- 20 ... <-15
>-58,0 ... -57,0 or >+15 ... + 20
- 25 ... <- 20 or >+ 20 ... + 25
- 30 ... <- 25 or
->+25 ... + 30
-61,0 ... <-60,5
or
>-57,0 ... -56,0
-62,0 ... <-61,0
or
>-56,0 ... -50,0
- 30 ... + 30 -62,0 ... -50,0
Calibration record
(example)
ba76156e01 03/2014 33
Page 34
pH value TruLab pH 1310
25.0
6.093
1
2
3
4
pH
°C
0
14
01.02.2014 08:00
Menu
USB output

5.2.7 Continuous measurement control (CMC function)

The Continuous Measurement Control (CMC function) facilitates to evaluate the current measured value instantly and definitely.
After each successful calibration the scale of the pH measuring range is displayed in the measured value display. Here you can very clearly see whether or not the current measured value is in the calibrated part of the measuring range.
The following information is displayed:
1 Measuring range for which a valid calibration is available (white).
Measured values in this range are suitable for documentation.
2 Measuring range for which no valid calibration is available (shaded).
Measured values in this range are not suitable for documentation. If necessary, calibrate the meter with buffers covering this measuring range. If the current measured value is outside the calibrated range, this area is shaded stronger. If a measured value is outside the measuring range pH 0 - 14, over­flow arrows are displayed at the left or right edge of the measuring range.
3 Currently measured pH value (needle)
4 Marking lines for all nominal buffer values used with the last valid cali-
bration
The limits of the calibrated range are determined by the buffers used for cali­bration:
Lower limit: Buffer with lowest pH value - 2 pH units Upper limit: Buffer with highest pH value + 2 pH units
34 ba76156e01 03/2014
Page 35
TruLab pH 1310 ORP voltage
U
157.0 mV
24.8 °C
01.02.2014 08:00
Menu
USB output

6 ORP voltage

6.1 Measuring

6.1.1 Measuring the ORP

When connecting an earthed PC, measurements cannot be performed in earthed media as the values would be incorrect. The USB interface is not galvanically isolated.
1. Connect the ORP electrode to the meter.
2. If necessary, select the mV display with <M>.
3. When measuring without temperature sensor:
Measure the temperature of the test sample and enter it with <> <>.
4. Immerse the ORP electrode in the test sample. The measured value is checked for stability (automatic stability control). The display of the measured parameter flashes.
Stability control
(AutoRead )
5. Wait for a stable measured value. The display of the measured parameter no longer flashes.
The stability control function (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproduc­ibility of measured values.
You can activate or switch off the automatic Stability control function (see section 7.1.1 CHANGING THE SETTINGS FOR PH MEASUREMENTS, page 38).
The measured parameter flashes on the display
Criteria for a stable
measured value
ba76156e01 03/2014 35
as soon as the measured value is outside the stability range
when you switch over between the measured parameters with <M>.
when the automatic Stability control is switched off.
The Stability control function checks whether the measured values are stable within the monitored time interval.
Page 36
ORP voltage TruLab pH 1310
Manually starting
the stability control
Measured parame-
Time interval Stability in the time interval
ter
ORP voltage 15 seconds : Better than 0.3 mV
Temperature 15 seconds : Better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer.
You can start the Stability control manually at any time, irrespective of the set­ting for automatic Stability control (see section 7.1.1 C
FOR PH MEASUREMENTS, page 38) in the Measurement menu.
HANGING THE SETTINGS
1. Freeze the measured value with <AR>. The [HOLD] status indicator is displayed.
2. Using <ENTER>, activate the Stability control function manually. The [AR] status indicator appears while the measured value is assessed as not stable. A progress bar is displayed and the display of the measured parameter flashes. The [HOLD][AR] status indicator appears as soon as a stable measured value is recognized. The progress bar disappears and the display of the measured parameter stops flashing. The current measurement data is output to the interface. Measurement data meeting the stability control criterion is marked by AR.
Freezing the mea-
sured value (HOLD
function)
You can prematurely terminate the Stability control function manu­ally with <ENTER> at any time. If the Stability control function is prematurely terminated, the current measurement data are output to the interface without the AutoRead info.
Using <ENTER>, start a further measurement with Stability control.
3. or Release the frozen measured value again with <AR>. The display switches to the measured value display. The [AR][HOLD] status display disappears.
With the HOLD function, you can freeze the current measured value. The dis­played measured value stops changing until you switch the HOLD function off.
If the HOLD function is active, you can, e.g. start a manual mea­surement with stability control.
1. Freeze the measured value with <AR>. The [HOLD] status indicator is displayed.
36 ba76156e01 03/2014
Page 37
TruLab pH 1310 ORP voltage
2. Release the frozen measured value again with <AR>. The HOLD function is switched off. The [HOLD] status display disappears.

6.1.2 Measuring the temperature

For reproducible ORP measurements, it is essential to measure the tempera­ture of the test sample. You have the following options to measure the temperature:
Automatic measurement of the temperature by the temperature sensor
(NTC 30 or Pt 1000) integrated in electrode.
Measurement by an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring mode:
Temperature
sensor
Resolution of the
temp. display
Temp. measurement
yes 0.1 °C Automatic with temper-
ature sensor
- 1 °C Manual
If you wish to measure (or calibrate) without temperature sensor, proceed as follows:
1. Measure the current temperature of the test sample.
2. Set the temperature value with <><>. or In the <F1>/[Menu]/Man. temperature menu, set the temperature value with <><>.

6.2 ORP calibration

ORP electrodes are not calibrated. You can, however, check ORP electrodes by measuring the ORP of a test solution and comparing the value with the nominal value.
ba76156e01 03/2014 37
Page 38
Settings TruLab pH 1310

7 Settings

7.1 Measurement settings

7.1.1 Changing the settings for pH measurements

The settings are made in the menu for calibration and measurement settings of the pH/ORP measurement. To open the settings, display the required parame­ter in the measured value display and press the <F1>/[menu] or <MENU/ ENTER> key. After completing the settings, switch to the measured value dis­play with <M>.
Default settings are printed in bold.
Menu item Possible set-
ting
Calibration /Cali­bration record
Calibration / Calibration data storage /Display
- Displays the calibration record of
- Displays the calibration record.
Description
the last calibration.
Further options: Scroll through the calibration
records with <><>.
Output the displayed calibration
record to the interface with <F2>/[USB output].
Output all calibration records to
the interface with <F2__>[USB output].
Quit the display with <F1>/
[Back] or <MENU/ENTER>.
Switch directly to the measured
value display with <M>.
Calibration / Cali­bration data stor­age / Output to USB
Calibration / Serial number (probe)
38 ba76156e01 03/2014
- Outputs the calibration records to the interface.
Entry of the serial number of the connected sensor. The serial number is output in the calibration record.
Change the content of the high-
lighted position with <><>.
Go to the next position with
<F2>/[].
When the serial number has
been completely entered, con­firm with <ENTER>.
Page 39
TruLab pH 1310 Settings
Menu item Possible set-
ting
Calibration /Buffer YSI
ConCal NIST/DIN ...
Calibration /One point calibration
Calibration /Cali-
Yes
No
1 ... 7 ... 999 d Calibration interval for the pH elec-
bration interval
Calibration /Unit for slope
mV/pH
%
Man. temperature -25 ... +25 ...
+130 °C
Description
Buffer sets to be used for pH cali­bration (see section 5.2 CALIBRA-
TION, page 25).
Quick calibration with 1 buffer
trode (in days). The meter reminds you to calibrate regularly by the flashing sensor symbol in the measuring window.
Unit of the slope. The % display refers to the Nernst slope of -59.2
mV/pH (100 x deter-
mined slope/Nernst slope).
Entry of the manually determined temperature. For measurements without temperature sensor only.
Temperature unit °C
°F
Temperature unit, degrees Celsius or degrees Fahr­enheit. All temperatures are displayed with the selected unit.
Resolution pH 0.001
Resolution of the pH display
0.01
0.1
Stability control On
Off
Switches on or off the automatic stability control during measure ment (see section 5.1.1 MEASUR-
ING THE PH VALUE, page 22)
Reset - Resets all sensor settings to the
delivery condition (see
section
7.4.1 RESETTING THE MEASURE-
MENT SETTINGS, page 45).

7.1.2 Buffer sets for calibration

-
You can use the buffer sets quoted in the table for an automatic calibration. The pH values are valid for the specified temperature values. The temperature dependence of the pH values is taken into consideration during the calibration.
ba76156e01 03/2014 39
Page 40
Settings TruLab pH 1310
No. Buffer set * pH values at
1 YSI * 4.000
25 °C
7.000
10.000
2 ConCal Any Any
3 NIST/DIN
DIN buffers according to DIN 19266 and NIST Traceable Buffers
1.679
4.006
6.865
25 °C
9.180
12.454
4 TEC
Technical buffers
2.000
4.010
25 °C
7.000
10.011
5 Merck 1* 4.000
20°C
7.000
9.000
6 Merck 2 * 1.000
20°C
6.000
8.000
13.000
7 Merck 3 * 4.660
6.880
9.220
8 Merck 4 * 2.000
4.000
7.000
10.000
9 Merck 5 * 4.010
7.000
10.000
10 DIN 19267 1.090
4.650
6.790
9.230
11 Mettler Toledo USA * 1.679
4.003
7.002
10.013
12 Mettler Toledo EU * 1.995
4.005
7.002
9.208
20°C
20°C
25 °C
25 °C
25 °C
25 °C
40 ba76156e01 03/2014
Page 41
TruLab pH 1310 Settings
No. Buffer set * pH values at
13 Fisher * 2.007
4.002
7.004
10.002
14 Fluka BS * 4.006
6.984
8.957
15 Radiometer * 1.678
4.005
7.000
9.180
16 Baker * 4.006
6.991
10.008
17 Metrohm * 3.996
7.003
8.999
18 Beckman * 4.005
7.005
10.013
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
19 Hamilton Duracal * 4.005
7.002
10.013
20 Precisa * 3.996
7.003
8.999
21 Reagecon TEC * 2.000
4.010
7.000
10.000
22 Reagecon 20 * 2.000
4.000
7.000
10.000
13.000
25 °C
25 °C
25 °C
20°C
ba76156e01 03/2014 41
Page 42
Settings TruLab pH 1310
No. Buffer set * pH values at
23 Reagecon 25 * 2.000
25 °C
4.000
7.000
10.000
13.000
24 Chemsolute * 2,000
20 °C 4,000 7,000 10,000
25 USABlueBook * 4,000
25 °C 7,000 10,000
* Brand names or trade names are trademarks of their respective owners pro-
tected by law.
The buffers are selected in the menu, pH / <F1>/[Menu] / Calibra- tion / Buffer (see 7.1.1 C
MENTS, PAGE 38).
HANGING THE SETTINGS FOR PH MEASURE-
Setting the calibra-
tion interval

7.1.3 Calibration interval

The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It is still possible to measure.
To ensure the high measuring accuracy of the measuring system, calibrate after the calibration interval has expired.
The calibration interval is set to 7 days in the factory. You can change the interval (1 ... 999 days):
1. Open the menu for measurement settings with [Menu]<F1>.
2. In the Calibration / Calibration interval menu, set the calibration interval
<><>.
with
3. Confirm the setting with <ENTER>.
4. Quit the menu with <M>.

7.2 Settings for ORP measurements

The settings are made in the menu for calibration and measurement settings of the pH/ORP measurement. To open the settings, display the required parame-
42 ba76156e01 03/2014
Page 43
TruLab pH 1310 Settings
ter in the measured value display and press the <F1>/[menu] or <MENU/ ENTER> key. After completing the settings, switch to the measured value dis-
play with <M>.
Default settings are printed in bold.
Menu item Possible
Description
setting
Man. temperature -25 ... +25
... +130
°C
Entry of the manually determined tem­perature. For measurements without temperature sensor only.
Temperature unit °C
°F
Temperature unit, degrees Celsius or degrees Fahrenheit. All temperatures are displayed with the selected unit.
Resolution mV 0.1
Resolution of the mV display
1
Stability control On
Off
Switches on or off the automatic stability control during measurement (see
sec­tion Stability control (AutoRead) & HOLD function, page 22 )
Reset - Resets all sensor settings to the delivery
condition (see
THE MEASUREMENT SETTINGS, page 45).
section 7.4.1 RESETTING

7.3 Sensor-independent settings

7.3.1 System

To open the Storage & config menu, press the <F1__>[Menu] key in the mea­sured value display. After completing the settings, switch to the measured value display with <M>.
Default settings are printed in bold.
Menu item Possible set-
ting
System / General / Language
Deutsch
English
(further)
System / General / Beep
System / General / Illumination
On
Off
Auto
On Off
System / General /
0 ... 50 ... 100 % Changing the display contrast
Contrast
Description
Selects the menu language
Switches on/off the beep on keystroke
Switches the display illumina­tion on/off
ba76156e01 03/2014 43
Page 44
Settings TruLab pH 1310
Menu item Possible set-
Description
ting
System / General / Switchoff time
System / Interface / Baud rate
10 min ... 1h ... 24 h
1200, 2400, 4800, 9600,
Adjusts the switch-off time
Baud rate of the data interface
19200
System / Interface / Output format
ASCII
CSV
Output format for data trans­mission (see section 9 TRANS-
MITTING DATA (USB INTERFACE),
page 53)
System / Interface / Decimal separator
System / Interface / Output header
Dot (xx.x)
Comma (xx,x)
Decimal separator
Output of a header for Output
format: CSV
System / Printer Settings for the integrated prin-
ter of the TruLab pH 1310 P. For details, see section 10 PRINTER (ONLY TRULAB PH 1310P), page 55
System /Clock Date format
Date Time
System /Service information
Time and date settings (see section 4.4.6 EXAMPLE 2 ON
NAVIGATION: SETTING THE DATE AND TIME, page 20)
Hardware version and software version of the meter are dis
-
played.
System /Reset - Resets the system settings to
the delivery condition (see
sec-
tion 7.4.2 RESETTING THE SYS-
TEM SETTINGS, page 45).

7.3.2 Data storage

This menu contains all functions to display, edit and erase stored measured values and calibration records (see
section 8 DATA STORAGE, page 47).

7.4 Reset

You can reset (initialize) all sensor settings and sensor-independent settings separately from each other.
44 ba76156e01 03/2014
Page 45
TruLab pH 1310 Settings

7.4.1 Resetting the measurement settings

The calibration data are reset to the default settings together with the measuring parameters. Recalibrate after performing a reset.
The following settings for pH measurements are reset to the default settings with the Reset function:
Setting Default settings
Buffer YSI
Cal. interval 7 d
Unit for slope mV/pH
Measured parameter pH
Resolution pH 0.001
Resolution mV 0.1
Asymmetry 0 mV
Slope -59.2 mV
Man. temperature 25 °C
One point calibration No
Stability control On
Temperature unit °C
The sensor settings are reset under the Reset menu item in the menu for cali­bration and measurement settings. To open it, press the <F1>/[Menu] or
<MENU/ENTER> key in the measured value display.

7.4.2 Resetting the system settings

The following system settings can be reset to the default condition:
Setting Default settings
Language English
Beep On
Baud rate 4800 Baud
Output format ASCII
Contrast 50 %
Illumination Auto
Switchoff time 1 h
ba76156e01 03/2014 45
Page 46
Settings TruLab pH 1310
The system settings are reset in the menu, Storage & config / System / Reset. To open the Storage & config menu, press the <F1__>[Menu] key in the measured value display.
46 ba76156e01 03/2014
Page 47
TruLab pH 1310 Data storage
01.02.2014 11:24:16 pH 7.000 24.8 °C AR +++
ID number: 1 Continue
Manual data storage 4 From 500
01.02.2014 08:00
Back
Data storage full. Erase?
Yes No
Warning
01.02.2014 08:00
Back

8 Data storage

You can transmit measured values (datasets) to the data storage:
Manual storage (see section 8.1 M
Automatic storage at intervals (see section 8.2 A
INTERVALS, page 48)
ANUAL STORAGE, page 47)
UTOMATIC STORAGE AT
Each storage process transmits the current dataset to the interface at the same time.

8.1 Manual storage

You can transmit a measurement dataset to the data storage as follows. The dataset is at the same time output to the interface:
1. Press the <STO> key shortly. The menu for manual storage appears.
If the storage is full The following window appears if all 500 storage locations are occupied:
ba76156e01 03/2014 47
2. If necessary, change and confirm the ID number (1 ... 10000) with <><> and <MENU/ENTER>. The dataset is stored. The meter switches to the measured value dis­play.
You have the following options:
Page 48
Data storage TruLab pH 1310
1
ID number
1
Interval
30 s
Duration
180 min
Continue
0d03h00min
0 1d17h33min
Automatic data storage
01.02.2014 08:00
2
3
Back
To erase the entire storage, confirm Yes.
To cancel the storage process and switch to the measured value display,
confirm No. Then you can e.g. transmit the stored data to a PC (see section
8.3.1 E erase the storage (see section 8.3.2 E
AGE, page 52).
DITING THE MEASURED VALUE STORAGE, page 50) and subsequently
RASING THE MEASUREMENT DATA STOR-

8.2 Automatic storage at intervals

The storage interval (Interval) determines the time interval between automatic storage processes. Each storage process transmits the current dataset to the interface at the same time.
Configuring the
automatic storage
function
1. Press the <STO__> key. The menu for automatic storage appears.
1 Adjusted entire storage duration
2 Max. available storage duration
3 Graphical display of the storage usage
48 ba76156e01 03/2014
Page 49
TruLab pH 1310 Data storage
1
2
pH
6.949
25.0
°C
0d03h00min
01.02.2014 08:00
0
14
USB outputEco mode
Settings You can configure the automatic storage function with the following settings:
Menu item Possible set-
ting
ID number 1 ... 10000 ID number for the dataset series.
Interval 1 s, 5 s, 10 s,
s, 1 min,
30
min,
5 10
min, min,
15
min,
30
min
60
Duration 1 min ... x min Storage duration.
Description
Storage interval.
The lower limit of the storage interval can be restricted by the number of free storage locations. The upper limit is restricted by the storage duration.
Specifies after which time the automatic storage should be ter minated.
The lower limit of the storage duration is restricted by the stor age interval. The upper limit is restricted by the number of free storage locations.
-
-
Starting the auto-
matic storage func-
tion
To start the automatic storage function, select Continue with <><> and confirm with <MENU/ENTER>. The meter switches to the measured value dis­play.
1 Remaining storage duration
2 Graphical display of the storage duration
The active automatic storage function can be recognized by the progress bar in the function display. The progress bar indicates the remaining storage dura tion.
-
Energy saving mode
ba76156e01 03/2014 49
([Eco mode])
If the automatic storage function is activated, only the following keys are active: Softkeys, <M>, <STO__> and <On/Off>. The other keys and the automatic switch-off function are deactivated.
If the automatic storing function is active, the meter provides an energy saving mode ([Eco mode]) to avoid unnecessary energy consumption. The energy
Page 50
Data storage TruLab pH 1310
Stop automatic storage?
Yes No
Warning
01.02.2014 08:00
Back
saving mode switches off functions of the meter that are not required for the automatic storage of measurement data (such as the display). By pressing any key the energy saving mode is switched off again.
Terminating the
automatic storage
function prema
turely
Proceed as follows to switch off the automatic storage function before the adjusted storage duration has expired:
-
1. Press the <STO__> key. The following window appears.
2. Using <><>, select Yes and confirm with <MENU/ENTER>. The meter switches to the measured value display. The automatic storage function is terminated.
Editing the data stor-
age

8.3 Measurement data storage

8.3.1 Editing the measured value storage

The contents of the manual or automatic measurement data storage can be shown on the display and output to the interface.
Each of the measurement data storages has a function to erase the entire contents.
The storage is edited in the menu, Storage & config / Data storage. To open the Storage & config menu, press the <F1__>[Menu] key in the measured value display.
Open the manual or automatic storage directly with the <RCL> or <RCL__> key.
The settings are explained here using the manual storage as an example. The same settings and functions are available for the automatic storage.
50 ba76156e01 03/2014
Page 51
TruLab pH 1310 Data storage
Manual data storage 3 of 64
01.02.2014 11:24:16 ID number: 2
pH 7.000 24.8 °C AR +++
01.02.2014 08:00
Back USB output
01.02.2014 09:56:20 inoLab TruLab pH 1310 Ser. no. 08502113
ID number 2
pH 6.012 24.8 °C, AR, S: +++ ________________________________________
01.02.2014 10:56:20 inoLab TruLab pH 1310 Ser. no. 08502113
ID number 2
pH 6.012 24.8 °C, AR, S: +++ ________________________________________
etc...
Settings
Menu item Setting/
function
Data storage /
- Displays all measurement datasets
Manual data storage / Display
Data storage /
- Erases the entire manual measure-
Manual data storage / Erase
Data storage /
- Outputs all stored measurement
Manual data storage / Output to USB
Description
page by page.
Further options: Scroll through the datasets with
<><>.
Output the displayed dataset to
the interface with <F2>/[USB output].
Quit the display with <F1>/
[Back].
ment data storage. All calibration data remains stored when this action is performed.
data to the interface.
Display presentation
of a dataset
Representation of a
dataset
(USB output)
ba76156e01 03/2014 51
Page 52
Data storage TruLab pH 1310
Quitting the display To quit the display of stored measurement datasets, you have the following
options:
Switch directly to the measured value display with <M>.
Quit the display and move to the next higher menu level with <F1>/[Back].

8.3.2 Erasing the measurement data storage

Erasing the measurement data storage (see section 8.3.1 EDITING THE
MEASURED VALUE STORAGE, page 50).

8.3.3 Measurement dataset

A complete dataset consists of:
ID number
Date/time
Measured value of the connected sensor
Measured temperature value of the connected sensor or manually set tem-
perature
AutoRead info: The AR indicator appears with the measured value if the
AutoRead criterion was met while storing (stable measured value). Other­wise, there is no AR indicator.
Calibration evaluation: +++, ++, +, -, or no evaluation

8.3.4 Storage locations

The TruLab pH 1310 meter has two measurement data storages. The measured values recorded either manually or automatic are stored separately in individual measurement data
storages.
Storage Maximum number of datasets
Manual data storage 500
Automatic data storage 5000
52 ba76156e01 03/2014
Page 53
TruLab pH 1310 Transmitting data (USB interface)

9 Transmitting data (USB interface)

9.1 Options for data transmission

Via the USB interface you can transmit data to a PC. The following table shows which data are transmitted to the interface in which way:
Data Control Operation / description
Current measured val­ues of all con­nected sensors
Stored mea­sured values
Calibration records
Manual With <F2>/[USB output].
Simultaneously with every manual
storage process (see section 8.1 M
ANUAL STORAGE, page 47).
Automatic, at intervals
 With <F2__>/[USB output].
Then you can set the transmission interval.
Simultaneously with every auto-
matic storage process (see section
8.2 A
UTOMATIC STORAGE AT INTER-
VALS, page 48).
Manual Displayed dataset with <F2>[USB
output] after calling up from the storage.
All datasets with the Output to USB
function. (see section 8.3.1 E
MEASURED VALUE STORAGE, page
DITING THE
50).
Manual Calibration record with <F2>/[USB
output] (see section 5.2.6 C
DATA, page 32).
ALIBRATION
Automatic At the end of a calibration proce-
dure.
The following rule applies: With the exception of the menus, shortly pressing the <F2>[USB output] key generally outputs the display contents to the interface (displayed measured values, measure­ment datasets, calibration records).
ba76156e01 03/2014 53
Page 54
Transmitting data (USB interface) TruLab pH 1310

9.2 Connecting a PC

Connect the TruLab pH 1310 to the PC via the USB interface.
NOTE When connecting an earthed PC, measurements cannot be performed in earthed media as the values would be incorrect. The USB interface is not galvanically isolated.
Installation of the
USB driver on the
PC
System requirements of the PC for installation of the USB driver:
PC with Pentium processor or higher with at least one free USB connection
and CD-ROM drive
Windows 2000, XP, Vista.
1. Insert the supplied installation CD in the CD drive of your PC.
2. Install the driver from the CD. Follow the Windows installation instructions as necessary.
3. Connect the TruLab pH 1310 to the PC via the USB interface. The meter is listed as a virtual COM interface among the connections in the Windows instrument manager.

9.3 MultiLab Importer

With the aid of the MultiLab Importer software, you can record and evaluate measuring data with a PC.
More detailed information can be found in the MultiLab Importer operating manual.
54 ba76156e01 03/2014
Page 55
TruLab pH 1310 Printer (only TruLab pH 1310P)
2
1
3
4

10 Printer (only TruLab pH 1310P)

1 Operating LED
– Illuminated: Printer is switched on
Switching on the
printer
– Flashing: Printer is switched off
2 Control panel with the keys <ON/FEED> and <OFF>
3 Lever to open the paper compartment
4 Lid of the paper compartment

10.1 Commissioning / switching the printer on or off

1. Connect the power pack to the TruLab pH 1310 P. The LED lights up green. The printer is ready for operation. or If the printer was switched off (LED flashing): Switch on the printer with <OFF>. The LED lights up green. The printer is ready for operation.
If there is a USB connection (such as to a PC) the data are output only to the PC.
Switching off the
printer
ba76156e01 03/2014 55
1. Switch off the printer with <OFF>. The LED flashes. The printer is switched off.
Page 56
Printer (only TruLab pH 1310P) TruLab pH 1310

10.2 Operation / printing

Data are only output to the printer if the following conditions are met
Data are transmitted manually or automatically (see section 9.1 O
FOR DATA TRANSMISSION, page 53)
PTIONS
the printer is switched on (LED illuminated)
there is no USB connection.

10.3 Printer settings

To open the Storage & config menu, press the <F1_>[Menu] key in the mea­sured value display. After completing the settings, switch to the measured value display with <M>.
Default settings are printed in bold.
Menu item Possible set-
ting
System / Printer / Font size
12x20
8x16 7x16
Description
Select the font size for the printer
A print sample (character set of the printer) to view the available font sizes can be printed with <OFF_>.
System / Printer / Print test page
- The printer prints the meter informa-
tion from the System / Service infor- mation menu. The current printer settings are used for the printout.

10.4 Maintenance

10.4.1 Changing the roll of paper (thermal paper)

1. Pull the lever (3) until the lid (4) of the paper compartment opens.
2. If necessary, remove any remains of the old roll of paper.
3. Insert the new roll of paper. The start of the paper should poke out of the paper compartment.
4. Close the lid (4) of the paper compartment so that it clicks into place.
5. If necessary, move the printer paper forward with <ON/FEED>.
56 ba76156e01 03/2014
Page 57
TruLab pH 1310 Printer (only TruLab pH 1310P)

10.5 What to do if... / printer

Cause Remedy
Integrated printer
does not print
– Printer switched off
(LED flashing)
– No power pack connected – Connect the power pack
– USB cable connected – Disconnect the USB cable from the
– Switch on the printer
(LED illuminated)
meter
Printer operating -
paper not being
printed
Integrated printer
prints automatically
– Function "automatic storing at
intervals" with long intervals is switched on
– Switch off the function (see section
UTOMATIC STORAGE AT INTER-
8.2 A
VALS, page 48)
– No paper available – Insert a roll of paper
Cause Remedy
– Paper inserted the wrong way up – Turn the roll of paper around and
insert it with the other side upwards
Cause Remedy
– The function "Automatic storage
at intervals" or "Transmit data automatic at intervals" is switched on
– Switch off the function (see section
UTOMATIC STORAGE AT INTER-
8.2 A
VALS, page 48 or
section 9.1 O
TRANSMISSION, page 53)
PTIONS FOR DATA
ba76156e01 03/2014 57
Page 58
Maintenance, cleaning, disposal TruLab pH 1310
1

11 Maintenance, cleaning, disposal

11.1 Maintenance

11.1.1 General maintenance activities

The only maintenance activity required is replacing the batteries.
See the relevant operating manuals of the electrodes for instruc­tions on maintenance.

11.1.2 Replacing the batteries

You can operate the meter either with normal batteries or with rechargeable batteries (Ni-MH). In order to charge the batteries, an external charging device is required.
1 Battery compartment
1. Open the battery compartment (1) on the underside of the meter.
2. Remove the four batteries from the battery compartment.
3. Place four new batteries (type Mignon AA) in the battery com­partment.
CAUTION
Make sure that the poles of the batteries are positioned cor­rectly. The ±-signs on the batteries must correspond to the ± signs in the battery compartment.
58 ba76156e01 03/2014
Page 59
TruLab pH 1310 Maintenance, cleaning, disposal
4. Close the battery compartment (1). The date (day) flashes on the display.
5. Set the date and time (see section 4.4.6 EXAMPLE 2 ON NAVIGATION: SETTING THE DATE AND TIME, page 20)
Dispose of used batteries according to the local regulations of your country.
End users within the European Union are obligated to return used batteries set up for recycling
Batteries are marked with the crossed-out waste container symbol. Therefore, they may not be disposed with the domestic waste.
(even ecologically compatible ones) to a collection point
purposes.

11.2 Cleaning

Occasionally wipe the outside of the measuring instrument with a damp, lint­free cloth. Disinfect the housing with isopropanol as required.
CAUTION
The housing is made of synthetic material (ABS). Thus, avoid contact with acetone or similar detergents that contain sol vents. Remove any splashes immediately.
-

11.3 Packing

This meter is sent out in a protective transport packing. We recommend: Keep the packing material. The original packing protects the meter against damage during transport.

11.4 Disposal

At the end of its operational lifetime, the meter must be returned to the disposal or return system statutory in your country. If you have any questions, please contact your supplier.
ba76156e01 03/2014 59
Page 60
What to do if... TruLab pH 1310

12 What to do if...

12.1 pH/ORP

More information and instructions on cleaning and exchange of sensors are given in the documentation of your sensor.
Error message
OFL, UFL
Error message,
Error
The measured value is outside the measuring range.
Cause Remedy
pH electrode:
– Air bubble in front of the junction – Remove air bubble
– Air in the junction – Extract air or moisten junction
– Cable broken – Replace electrode
– Gel electrolyte dried out – Replace electrode
– Measured value outside the mea-
suring range of the meter
Cause Remedy
pH electrode:
– The values determined for zero
point and slope of the electrode are outside the allowed limits.
– Use suitable electrode
– Recalibrate
– Junction contaminated – Clean junction
– Electrode broken – Replace electrode
Buffer solutions:
– Wrong buffer solutions – Change calibration procedure
– Buffer solutions too old – Use only once.
Note the shelf life
– Buffer solutions depleted – Change solutions
60 ba76156e01 03/2014
Page 61
TruLab pH 1310 What to do if...
No stable measured
Cause Remedy
value
pH electrode:
– Junction contaminated – Clean junction
– Membrane contaminated – Clean membrane
Test sample:
– pH value not stable – Measure with air excluded if
necessary
– Temperature not stable – Temper if necessary
Electrode + test sample:
– Conductivity too low – Use suitable electrode
Obviously incorrect
measured values
– Temperature too high – Use suitable electrode
– Organic liquids – Use suitable electrode
Cause Remedy
pH electrode:
– pH electrode unsuitable – Use suitable electrode
– Temperature difference between
buffer and test sample too great
– Measurement procedure not suit-
– Adjust temperature of buffer or
sample solutions
– Follow special procedure
able
ba76156e01 03/2014 61
Page 62
What to do if... TruLab pH 1310

12.2 General information

Sensor symbol
flashes
Display
Instrument does not
react to keystroke
You want to know
which software
version is in the
meter
Cause Remedy
– Calibration interval expired – Recalibrate the measuring
Cause Remedy
– Batteries almost empty – Replace the batteries (see
Cause Remedy
– Operating condition undefined or
EMC load unallowed
Cause Remedy
– E. g., a question by the service
department
system
section 11.1 MAINTENANCE, page 58)
– Processor reset:
Press the <ENTER> and <On/ Off> key simultaneously
– Switch on the meter.
Open the menu, <F1_>/ [Menu] / Storage & config / SystemService information. The instrument data are dis played.
-
62 ba76156e01 03/2014
Page 63
TruLab pH 1310 Technical data

13 Technical data

13.1 Measuring ranges, resolution, accuracy

Measuring ranges, resolution
Parameter Measuring range Resolution
pH -2.0 ... +20.0 0.1
-2.00 ... +20.00 0.01
-2.000 ... +19.999 0.001
U [mV] -1200.0 ... +1200.0 0.1
-2500 ... +2500 1
T [°C] -5.0 ... +105.0 0.1
T [°F] 23.0 ... +221.0 0.1
Manual temperature input
Parameter Range Increment
T
T
[°C] -25 ... +130 1
manual
[°F] -13 ... +266 1
manual
Accuracy (± 1 digit) Parameter Accuracy Temperature of the test
sample
pH / range *
-2.0 ... +20.0 ± 0.1 +15 °C ... +35 °C
-2.00 ... +20.00 ± 0.01 +15 °C ... +35 °C
-2.000 ... +19.999 ± 0.005 +15 °C ... +35 °C
U [mV] / range
-2500 ... +2500 ± 1 +15 °C ... +35 °C
-1200.0 ... +1200.0 ± 0.3 +15 °C ... +35 °C
T [°C] / temperature sensor
NTC 30 ± 0.1
PT 1000 ± 0.1
* when measuring in a range of ± 2 pH around a calibration point
The accuracy values specified here apply exclusively to the meter. The accuracy of the
electrodes and buffer solutions has to be taken
into account additionally.
ba76156e01 03/2014 63
Page 64
Technical data TruLab pH 1310

13.2 General data

Dimensions TruLab pH 1310: Approx. 230 x 190 x 80 mm
TruLab pH 1310 P: Approx. 280 x 230 x 80 mm
Weight TruLab pH 1310: Approx. 0.8 kg
TruLab pH 1310 P: Approx. 1.0 kg
Mechanical structure Type of protection IP 43
Electrical safety Protective class III
Test certificates CE, cETLus
Ambient conditions
Power supply
Storage -25 °C ... +65 °C
Operation +5 °C ... +55 °C
With the power pack connected: +5 °C ... +40 °C
Admissible relative humidity
Yearly mean: < 75 % 30 days/year: 95 % Other days: 85 %
Batteries 4 x 1.5 V alkali-manganese batteries, type
AA
Rechargeable batter­ies
4 x 1.2 V NiMH rechargeable batteries,
AA
type (no charging function)
Operational life Up to 1000 h without / 150 h with illumination
Power pack Kuantech Co. Ltd. KSAC 0900110W1UV-1
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 270 mA Output: 9 V = / 1.1 A Connection max. overvoltage category II Primary plugs contained in the scope of delivery: Euro, US, UK and Australian.
Sensor input Input resistance > 5 * 1012 ohm
Input current < 1 * 10
-12
A
USB interface Type USB 1.1
USB B (device), data output
Baud rate Adjustable:
1200, 2400, 4800, 9600, 19200 Baud
Data bits 8
Stop bits 2
Parity None
Handshake RTS/CTS
Cable length max. 3 m
64 ba76156e01 03/2014
Page 65
TruLab pH 1310 Technical data
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. Changes or modifications not expressly approved by the manufacturer could void the user‘s authority to operate the equipment.
Guidelines and norms used
EMC EC directive 2004/108/EC
EN 61326-1 EN 61000-3-2 EN 61000-3-3 FCC Class A
Meter safety EC directive 2006/95/EC
EN 61010-1 UL 61010-1 CAN/CSA-C22.2#61010-1
IP protection class EN 60529
ba76156e01 03/2014 65
Page 66
Firmware update TruLab pH 1310

14 Firmware update

Available firmware updates are provided on the Internet. With the firmware update program and a PC you can update the firmware of the TruLab
For the update you have to connect the meter to a PC.
For the update via the USB interface, the following is required:
a free USB interface (virtual COM port) on the PC
the driver for the USB interface (on the enclosed CD-ROM)
the USB cable (included in the scope of delivery of the TruLab pH 1310).
1. Install the downloaded firmware update on a PC.
pH 1310 to the newest version.
An update folder is created in the Windows start menu. If an update folder already exists for the meter (or meter type), the new data is displayed there.
2. In the windows start menu, open the update folder and start the firmware
update program.
3. Using the USB interface cable, connect the TruLab pH 1310 to a USB
interface (virtual COM port) of the PC.
4. Switch on the TruLab pH 1310.
5. In the firmware update program, start the update process with OK.
6. Follow the instructions of the firmware update program.
During the programming process, a corresponding message and a progress bar (in %) are displayed. The programming process takes approx. three minutes. A terminatory message is displayed after a successful programming process. The firmware update is completed.
7. Disconnect the TruLab pH 1310 from the PC.
The TruLab pH 1310 is ready for operation again.
After switching the meter off and on you can check whether the meter has taken over the new software version (see
VERSION IS IN THE METER, PAGE 62).
YOU WANT TO KNOW WHICH SOFTWARE
66 ba76156e01 03/2014
Page 67
TruLab pH 1310 Glossary

15 Glossary

pH/ORP
Asymmetry see zero point
Electromotive force of
an electrode
Junction The junction is a porous body in the housing wall of reference elec-
ORP voltage The ORP is caused by oxidizing or reducing substances dissolved in
pH value The pH value is a measure of the acidic or basic effect of an aqueous
Potentiometry Name of a measuring technique. The signal (depending on the mea-
The electromotive force U of the combination electrode is the measur­able electromotive force of an electrode in a solution. It equals the sum of all the galvanic voltages of the combination electrode. Its dependency on the pH results in the electrode function, which is char­acterized by the parameters, slope and zero point.
trodes or electrolyte bridges. It arranges the electrical contact between two solutions and makes the electrolyte exchange more dif­ficult. The expression, junction, is also used for ground or junction­less transitions.
water if these substances become effective on an electrode surface (e. g. a gold or platinum surface).
solution. It corresponds to the negative decadic logarithm of the molal hydrogen ions activity divided by the unit of the molality. The practical pH value is the value of a pH measurement.
sured parameter) of the electrode is the electrical potential. The elec­trical current remains constant.
Slope The slope of a linear calibration function.
Zero point The zero point of a pH combination electrode is the pH value at which
the electromotive force of the pH combination electrode at a specified temperature is zero. Normally, this is at 25 °C.
General information
Adjusting To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or a value that is regarded as correct, or that the difference remains within the tolerance.
AutoRange Name of the automatic selection of the measuring range.
Calibration Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Often, this expression is also used when the measuring system is adjusted at the same time (see adjusting).
Measured parameter The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D.O. concentration.
ba76156e01 03/2014 67
Page 68
Glossary TruLab pH 1310
Measured value The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and unit (e. g. 3
m; 0.5 s; 5.2 A; 373.15 K).
Molality Molality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
vent.
Reset Restoring the original condition of all settings of a measuring system.
Resolution Smallest difference between two measured values that can be dis-
played by a meter.
Stability control
Function to control the measured value stability.
(AutoRead )
Standard solution The standard solution is a solution where the measured value is
known. It is used to calibrate a measuring system.
Temperature function Name of a mathematical function expressing the temperature behav-
ior of a test sample, a sensor or part of a sensor.
Test sample Designation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample and original sample are identical if the test sample was not processed.
68 ba76156e01 03/2014
Page 69
TruLab pH 1310 Index

16 Index

A
Automatic switch-off . . . . . . . . . . . . . . . . . . 16
AutoRead
ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
B
Battery compartment . . . . . . . . . . . . . . 11, 58
C
Calibration
pH . . . . . . . . . . . . . . . . . . . . . . . . . 25, 37
Calibration evaluation
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Calibration interval . . . . . . . . . . . . . . . . . . . 42
Calibration points
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Connecting a PC . . . . . . . . . . . . . . . . . . . . 54
D
Dataset . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Date and time . . . . . . . . . . . . . . . . . . . . . . . 20
Default settings
Measured parameter . . . . . . . . . . . . . . 45
System settings . . . . . . . . . . . . . . . . . . 45
Measuring
ORP voltage . . . . . . . . . . . . . . . . . . . . 35
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Menus (navigation) . . . . . . . . . . . . . . . . . . 17
Messages . . . . . . . . . . . . . . . . . . . . . . . . . 18
P
pH buffer sets . . . . . . . . . . . . . . . . . . . . . . 39
Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Printer (only pH 7310P) . . . . . . . . . . . . . . . 55
R
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
S
Sample display . . . . . . . . . . . . . . . . . . . . . 14
Scope of delivery . . . . . . . . . . . . . . . . . . . . 10
Single-point calibration
pH . . . . . . . . . . . . . . . . . . . . . . . . . 26, 29
Slope
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Socket field . . . . . . . . . . . . . . . . . . . . . . . . 15
Storage interval . . . . . . . . . . . . . . . . . . . . . 48
Storing in memory . . . . . . . . . . . . . . . . . . . 47
Automatic . . . . . . . . . . . . . . . . . . . . . . 48
Manual . . . . . . . . . . . . . . . . . . . . . . . . 47
E
Energy saving mode . . . . . . . . . . . . . . . . . 49
F
Firmware update . . . . . . . . . . . . . . . . . . . . 66
I
Initial commissioning . . . . . . . . . . . . . . . . . 10
Initialize . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Interval for calibration . . . . . . . . . . . . . . . . . 42
K
Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
M
Measured value display . . . . . . . . . . . . . . . 16
Measurement accuracy . . . . . . . . . . . . . . . 42
Measurement data storage
Edit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Erase . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Measurement dataset . . . . . . . . . . . . . . . . 52
T
Temperature measurement
pH . . . . . . . . . . . . . . . . . . . . . . . . . 24, 37
Three-point calibration
pH . . . . . . . . . . . . . . . . . . . . . . . . . 27, 30
Transmitting data . . . . . . . . . . . . . . . . . . . 53
automatic . . . . . . . . . . . . . . . . . . . . . . . 53
manual . . . . . . . . . . . . . . . . . . . . . . . . 53
Transmitting measured values . . . . . . . . . 53
Two-point calibration
pH . . . . . . . . . . . . . . . . . . . . . . . . . 26, 29
Z
Zero point of pH combination electrode . . 25
ba76156e01 03/2014 69
Page 70
Index TruLab pH 1310
70 ba76156e01 03/2014
Page 71
TruLab pH 1310 Contact Information

17 Contact Information

17.1 Ordering & Technical Support

Telephone: (800) 897-4151
(937) 767-7241 Monday through Friday, 8:00 AM to 5:00 PM ET
Fax: (937) 767-1058
Email: environmental@ysi.com
Mail: YSI Incorporated
1725 Brannum Lane Yellow Springs, OH 45387 USA
Internet: www.ysi.com
When placing an order please have the following information available:
YSI account number (if available) Name and Phone Number Model number or brief description Billing and shipping address Quantity Purchase Order or Credit Card

17.2 Service Information

YSI has authorized service centers throughout the United States and Internationally. For the nearest service center information, please visit
www.ysi.com and click ‘Support’ or contact YSI Technical Support
directly at 800-897-4151.
When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at ‘Support‘ tab.
www.ysi.com and clicking on the
ba76156e01 03/2014
71
Page 72
Contact Information TruLab pH 1310
72
ba76156e01 03/2014
Page 73
Page 74
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We're 12,500 people unified in a common purpose: creating innovative solutions to meet our world's water needs. Developing new technologies that will improve the way water is used, conserved, and re-used in the future is central to our work. We move, treat, analyze, and return water to the environment, and we help people use water efficiently, in their homes, buildings, factories and farms. In more than 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com
Xylem |' m|zīlə
YSI 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241; 800-765-4974 Fax: +1 937-767-1058 Email: environmental@ysi.com Web: www.ysi.com
©Xylem Inc
Loading...