The IDS WA System is an accessory for IDS measuring systems enabling a wireless connection between any IDS sensor with plug head connector (variant W) and
your IDS meter.
Two adapters, one at the IDS meter (IDS WA-M) and one at the sensor (IDS WAS), replace the sensor cable with an energy-saving Bluetooth LE radio connection.
The sensor is supplied with power by a rechargeable battery in the IDS WA-S
adapter. The battery is charged with the charging device included in the IDS WA
System (IDS WA-S Charger).
To upgrade the IDS WA System, equip further IDS sensors with the IDS WA-S
adapter, which is available as an accessory.
1IDS WA-M adapter (on the meter)
With an IDS WA-M adapter connected, the IDS meter registers up to 6
wireless sensors within a radius of up to 10 meters. The administering of
the sensors is done in the Sensors menu.
2Wireless sensor
Unit of IDS WA-S adapter (3) and IDS sensor (4)
3IDS WA-S adapter (connected to the sensor)
4IDS sensor with plug head connector (variant W)
5Sensors registered and selected for the measured value display (max. 3,
corresponding to the number of sensor connections / channels on the
meter).
6Sensors registered but not selected for the measured value display.
ba76199e0308/20185
SafetyIDS WA System
2Safety
2.1Safety information
2.1.1 Safety information in the operating manual
This operating manual provides important information on the safe operation of the
system. Read this operating manual thoroughly and make yourself familiar with the
system before putting it into operation or working with it. The operating manual
must be kept in the vicinity of the system so you can always find the information
you need.
2.1.2 Further documents providing safety information
The following documents provide additional information, which you should observe
for your safety when working with the measuring system:
• Operating manuals of meters, sensors and other accessories
• Safety datasheets of calibration or maintenance accessories (such as buffer
solutions, electrolyte solutions, etc.)
2.2Safe operation
2.2.1 Authorized use
The authorized use of the IDS WA System is exclusively the running of a wireless
connection beween IDS sensors with IDS meters.
Only the operation and running of the analyzer according to the instructions and
technical specifications given in this operating manual is authorized (see section 9
ECHNICALDATA).
T
Any other use is considered unauthorized.
2.2.2 Requirements for safe operation
Note the following points for safe operation:
• The IDS WA System may only be operated according to the authorized use
specified above.
• The IDS WA System may only be supplied with power by the energy sources
mentioned in this operating manual.
• The IDS WA System may only be operated under the environmental conditions
6ba76199e0308/2018
IDS WA SystemCommissioning
mentioned in this operating manual.
• The individual components of the IDS WA System must not be opened.
3Commissioning
3.1Scope of delivery
Depending on the order number, the scope of delivery consists of the whole IDS
WA Kit, or one individual item out of the IDS WA Kit.
•IDS WA Kit
– IDS WA-M adapter for the meter
– IDS WA-S adapter for the sensor
– IDS WA-S Charger to charge the battery in the IDS WA-S adapter
– USB cable to connect the IDS WA-S Charger to a USB socket
– Power pack with USB connection and country-specific line adapters
• Operating manual
3.2System requirements
• IDS sensor with plug head connector (variant W)
Available sensors, see Internet
• WA capable meter
– MultiLab 4010(P)-1W IDS, MultiLab 4010(P)-1 IDS (from software version
V2.00)
– MultiLab 4010-2W IDS, MultiLab 4010-3W IDS
The meters with older software listed here can be retrofitted with the
function for the administration of wireless sensors via a software update
(see operating manual of your meter).
ba76199e0308/20187
CommissioningIDS WA System
1
2
4
5
3
3.3Charging the battery with the IDS WA-S Charger
The wireless transmission of sensor data from sensor to meter is only possible if
the battery in the IDS WA-S adapter is charged.
Charge the battery with the IDS WA-S Charger.
1Charging slot for the IDS WA-S adapter
2LED indicating the operating condition of the IDS WA-S Charger
3USB-B connection for the power supply of the IDS WA-S Charger
4Charging contacts for the IDS WA-S adapter
5Retaining clip for the IDS WA-S adapter
The IDS WA-S Charger is used both to charge and store the IDS WA-S adapters
(e.g. overnight). The sensors may remain plugged to the IDS WA-S adapters while
being charged.
3.3.1 Power supply of the IDS WA-S Charger
1.Plug the USB plug of the USB cable supplied into the USB connector (3) of the
IDS WA-S Charger.
2.Plug the other USB plug of the cable to the USB connector of the power pack
or to the USB connector (USB 3.0 or higher) of a PC.
The status LED of the IDS WA-S Charger lights up green. The IDS WA-S
Charger is ready for operation.
8ba76199e0308/2018
IDS WA SystemCommissioning
A USB 2.x connector on the PC supplies up to 200 mA.
A USB 3.x connector on the PC supplies up to 900 mA.
If several USB devices are connected to a PC or a USB hub without extra
power supply, the available current is distributed to several connections.
One device consuming a lot of power may cause other devices to receive
not enough power so that they cannot be operated.
• Operate the IDS WA-S Charger on the USB 3.0 or higher interfaces of
your PC.
• Operate only one IDS WA-S Charger per PC to avoid possible conflicts
with other USB devices about power resources.
3.3.2 Charging the IDS WA-S battery
1.Open the retaining clip (5) on the IDS WA-S Charger.
2.Insert the IDS WA-S adapter in the IDS WA-S Charger. The charging contacts
of the adapter have to be positioned on the charging contacts (4) of the IDS
WA-S Charger. The function keys of the adapter point upwards.
3.Close the retaining clip (5) of the IDS WA-S Charger until it clicks into place.
The status LED (2) of the IDS WA-S Charger lights up red. The battery is being
charged.
3.3.3 Status LED of the IDS WA-S Charger
Color StatusMeaning
GREENIllumi-
nated
with IDS WA-S adapter:
The battery is fully charged. The charging current is
switched off.
without IDS WA-S adapter:
The IDS WA-S Charger is ready for operation.
REDIllumi-
The battery is being charged.
nated
GREEN/
FlashesCharging error (see section 7 W
HATTODOIF ...)
RED
Off-The IDS WA-S Charger is not supplied with enough power