Young Lin aquaMAX-Basic 360, aquaMAX-Basic 362, aquaMAX-Basic 363, aquaMAX-Basic 361 Series Manual

Contents
Chapter 1. Introduction
1-1. Introduction 1-2. Operation Principle 1-3. Flow diagram 1-4. Configuration 1-5. System Configuration 1-6. Specifications
Chapter 2. Installation
2-1. Packing 2-2. CE Specification 2-3. Installing Filters 2-4. The Connection of Feed Water 2-5. The Connection of the Produced Water 2-6. The Connection of the Rejection Water 2-7. How to Connect Feed Water Kit
1
1 2 3 4 5 7
9
9 12 12 13 14 16 17
2-8. The Connection of Reservoir Tank Sensor (Optional)
Chapter 3. Operation
3-1. Operation Mode 3-2. Initial Operation 3-3. Initial Flushing 3-4. Operation of the Keys 3-5. Normal Operation and Standby Mode 3-6. Measuring Point Display 3-7. Service Operation 3-8. Errors and its Display 3-9. System Shut Down
Chapter 4. Maintenance
4-1. Maintenance and Solutions 4-2. Filter Replacement 4-3. Trouble Shooting
17
18
18 19 19 20 22 23 24 25 26
27
27 29 30
4-4. Fuse Replacement
31
Figure Lists
Figure 1-1 ] Standard System of the Basic 360 Series 1Figure 1-2 ] Flow Diagram 3Figure 2-1Components of aquaMAX Basic. 9
Figure 2-2How to connect filters 12Figure 2-3The Connection of Feed Water 13
Figure 2-4The Connection of the Produced Water 1(aquaMAX 360, 362) 14Figure 2-5The Connection of the Produced Water 2(aquaMAX 361, 363) 14Figure 2-6The Connection of the Rejection Water 15Figure 4-1Filters in the aquaMAX Basic 27Figure 4-2Filter Replacement 29Figure 4-3The Fuse Replacement 31
Table Lists
Table 1-1System Configuration 5Table 1-2Water Contaminants and its Effective Removal 5Table 1-3Comparison of Water Quality according to Water Purification Method 6Table 1-4Water Classification According to CAP, ASTM Standards 6Table 1-5The Specification of aquaMAX 7Table 1-6Feed Water Requirements 7Table 1-7Dimension, Power and Weight 8Table 3-1Display and Causes of System Troubles 25
aquaMAXTM-Basic 360 Series
Chapter 1. Introduction
1-1. Introduction
The Young Lin Basic 360 Series is the water purification system that produces pure water from tap or
underground water. The purified water by the Basic 360 Series is adequate to use as a laboratory grade water or
supply for ultra-pure water production system. The components of the Basic 360 Series are as follows.
[ Figure 1-1 ] Standard System of the Basic 360 Series
1
aquaMAXTM-Basic 360 Series
1. Control panel
2. Door
3. RO Membrane Pack
4. RO Pre Pack
5. RO Post Pack
6. Switch of Tank Level
7. Power On/Off Switch
8. Fuse Holder
9. Power Socket
10. Name Tag
11. Connection part of feed water
12. Display Light
13. Keypad Name tag
14. LCD Display
1-2. Operation Principle
The Basic 360 Series Water Purification System produces decontaminated pure water by using reverse
osmosis to filter particles, inorganics, organics and bacteria from water supply. The water supply is fed through a
prefilter to get rid of the large particles and the pre-filtered water flows to a carbon filter to be removed chlorine.
Then the produced water is pressurized by a booster pump and delivered to the RO Membrane Pack to be
purified. The purified water is used as laboratory grade water or feed water for ultra-high pure water.
In use of the aquaMAX
TM
-Basic 361 or 362 System, it produces 2
anion through an ion exchange pack.
nd
grade distilled water removed cation and
2
1-3. Flow diagram
aquaMAXTM-Basic 360 Series
[ Figure 1-2 ] Flow Diagram
3
aquaMAXTM-Basic 360 Series
1-4. Configuration
1-4-1. R.O. Membrane Pack and Prefilter(PRE PACK)
The Water supply(tap water or treated municipal water) is fed through a pre-filter to be separated large and small
particles by filtration. The permeate water then enters the RO Pack housing and permeates through the
spiral-wound RO membrane. For the Basic 360 Series, when RO membrane is pressurized at 5.5 Kg/cm
approximately 70% of the input water is continuously withdrawn from the system through a reject valve and the
remaining 30% of the water permeates through the membrane to be filtered. The productivity of RO membrane
is varied depending on pressure and used time.
1-4-2. Booster Pump
The booster pump of the Basic 360 Series is a silent diaphragm pump. It can transfer water flow rate of
maximum 60 liter/hour, however the purified water can be produced at maximum 10 liter/hour when supplied
with RO pressure of 5.5 Kg/cm
2
. The amount of the purified water can be varied by temperature, used time and
pressure of RO membrane.
2
,
1-4-3. Pressure Switch
If the inlet pressure of a water supply is less than 0.8 Kg/m
booster pump to prevent the pump from damage and if it is more than 7.0 Kg/cm
prevent the filter from damage. The inlet pressure of water supply should be more than 1.0 Kg/cm
2
, the pressure switch automatically stops the
2
, it stops the operation to
2
, because the
pressure drops down after passing through the prefilter. In case the switch stops the booster pump at low
pressure (less than 0.8 Kg/m
2)
, you should raise the inlet pressure first to drive the booster pump again.
It is required to install an extra external pump if the pressure of water supply is less than average 0.8 Kg/m2..
1-4-4. Drain Valve and Automatic Flushing
The drain valve consisted of a needle valve and a solenoid valve is designed to control the backpressure of RO
membrane. The reverse osmosis is displayed on the pressure gauge mounted inside of the system. The Basic
360 Series automatically turns on the drain valve every 30 minutes for 1 minute so that the contaminants
collected on the RO membrane surface can be flushed away by a high velocity flush of water.
1-4-5. Water Quality Monitoring
The Basic 360 Series monitors water quality by measuring conductivity at 3 different points. The measured
value indicates ion drain ratio and time of filter replacement.
4
1-5. System Configuration
aquaMAX 360 : Standard Type (10L/h)
aquaMAX 361 : Ion Exchange Addition Type (10L/h)
aquaMAX 362 : Extended Capacity Type (20L/h)
aquaMAX 363 : Extended Capacity and Ion Exchange Addition Type(20L/h)
Table 1-1System Configuration
aquaMAXTM-Basic 360 Series
R.O Membrane 1 Pack (Standard Type for RO) R.O Membrane 2 Pack (Extended Capacity Type for RO) R.O Pre Pack (Prefilter, Carbon filter) R.O Post Pack (Ion Exchange Fileter)
Table 1-2Water Contaminants and its Effective Removal
Contaminant Excellent Good Bad
Inorganic ion Ion Exchange Distillation R.O
Organics R.O. Distillation Carbon
Particles Membrane R.O. Distillation
Bacteria Membrane R.O. Distillation
aquaMAX 360 aquaMAX 361 aquaMAX 362 aquaMAX 363
1 ea 1 ea * *
* * 1 ea 1 ea
1 ea 1 ea 1 ea 1 ea
* 1 ea * 1 ea
5
aquaMAXTM-Basic 360 Series
Table 1-3 Comparison of Water Quality according to Water Purification Method
Ion
Exchange
Water
Ultra Pure
Water
aquaMAX CAP/ASTM
Tap Water
Distilled
Water
Dissolved Inorganics (CaCO
, ppm)
3
Resistivity
(M·cm, 25 ℃ )
170 0.3 0.1 0.05 0.025 20 -
0.003 2 0.1 10 18.2 0.1 0.5 0.1 1.0
Silica(ppm) 1 0.05 <0.01 <0.01 0.1 <0.01
Heavy Metal (ppm)
Dissolved Organics(ppm)
Bacteria(cfu/)
Small Particles(>5 ㎛/㎖)
1 0.05 <0.01 <0.01 0.1 <0.01
- <1 <1 <1 <1 -
>100 <10 >100 * <10 -
>10,000 <100 >10,000 * <10 -
Table 1-4 Water Classification According to CAP, ASTM Standards
Water Quality
Resistivity
(M·cm)
Conductivity(μS)
aquaMAX R.O Type 0.1 0.5 2.0∼5
aquaMAX Ion Exchange addition Type 2.0 1∼ 0 0.1 0.5
Laboratory-Grade 0.1 1.0 1.0 10
Analytical-Grade 1.0 2.0 0.5 1.0
Reagent-Grade 10 0.05
CAP : The College of American Pathologists
ASTM : American Society for Testing and Materials
6
1-6. Specifications
Table 1-5The Specification of aquaMAX
aquaMAXTM-Basic 360 Series
Ion Removal 95-99%
Particle Rejection
Bacteria Removal
Pyrogen Reduction
The Young Lin aquaMAX can reject more than 99% of inorganic and organic contaminants in the feed water.
Thus, after permeation of the reverse osmosis membrane, the purified water contains less than 1% of the
contaminants present in the feed water. The aquaMAX uses 5.0 prefilter to filter large and small particles in
order to prevent clogging or fouling of the RO membrane.
Table 1-6Feed Water Requirements
pH Range 3.0 10.0 Pressure(Kg/cm2) 1 - 5
TDS
Conductivity
Residual Chlorine < 0.1 ppm
99%
99.99%
99% Material Polyamide Complex
500 ppm
800 μS
(After Carbon filter)
Organics Removal
Pure Water Production Rate (25℃,8Kg/cm2)
Temperature 4-35
Turbidity 1.0 NTU
The others If the feed water contains a
99%
Standard : 10L/hour
Extension : 20L/hour
lot of bacteria, it adversely affects the purified water quality and damages the RO membrane.
7
Table 1-7Dimension, Power and Weight
Width 435 mm
Dimension
Depth 370 mm
Height 580 mm
aquaMAX 360 24.0 Kg
aquaMAXTM-Basic 360 Series
Weight
Power
aquaMAX 361 24.5 Kg
aquaMAX 362 24.5 Kg
aquaMAX 363 25.0 Kg
Consumption 100 W
Power AC 110V/230V, 50/60Hz
Fuse
Littlefuse 250V 1A Part NO. 218001
8
aquaMAXTM-Basic 360 Series
Chapter 2. Installation
Handle with care not to change up side down at delivery.
2-1. Packing
The standard type of aquaMAX-Basic is consisted of following items. The filters are mounted inside of the
system.
Figure 2-1Components of aquaMAX Basic.
9
aquaMAXTM-Basic 360 Series
1. Water Purification System (Fuse 250V, 1A : mounted on the fuse holder) 1 system
: 1 ea of PRE PACK, RO PACK, POST PACK(Included aquaMAX
2. Seal Hose Kit(Ø6 tubing :3m, Kit for connection of tab water) 1 ea
3. Tubings for the produced water and the drain water Ø6 : 3m each 2 ea
4. Tool kit for filter replacement 1 kit
5. Filter Cartridge (PRE PACK, POST PACK, RO PACK) 1 ea
: Filters are separately packed to avoid damage on delivery. Mount the filters
after installation of system
6. Operating manual 1 ea
7. Power cord 1 ea
TM
-Basic 361/363 only)
2-2. CE Specification
2-2-1. Installation Requirements
Environmental conditions
- Indoor use only
- Altitude up to 2000m
- Temperature 5 to 40℃℃
- Maximum relative humidity 80% for temperature up to 30 decreasing linearly to 50% relative
humidity at 40
- Pollution degree 2
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
10
aquaMAXTM-Basic 360 Series
2-2-2. Safety Indication
An explanation of symbols related to safety, which are used on the equipment.
a)
-> Protective Earth
b) and -> Power On and Power Off
c)
-> Caution, risk of electric shock
2-2-3. WEED GUIDE
This marking shown on the product or its literature, indicates that it should not be disposed
with other household wastes at the end of its working life. To prevent possible harm to the
environment or human health from uncontrolled waste disposal, please separate this from
other types of wastes and recycle it responsibly to promote the sustainable reuse of
material resources.
Household users should contact either the retailer where they purchased this product, or their local government
office, for details of where and how they can take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of the purchase contract. This
product should not be mixed with other commercial wastes for disposal.
11
aquaMAXTM-Basic 360 Series
2-3. Installing Filters
Figure 2-2How to connect filters
1. Check there are 6 fittings for one-touch clip on upper part of where filters have been installed.
>Press down the clip connected to the fitting making a sound like "click".
: The fitting has been already pressed down at the first installation.
: If it's not pressed down, filter can't be fit in.
2. Insert the filter to fix into the marked fitting by making "click" sound.
>After inserted, check the installation by pressing down the up-side head of filter.
>There is a indicating mark at the lower part of filter not to install the filter upside down.
CAUTION : Filter installation upside down causes the system malfunction and water quality problem.
3. For the R.O membrane pack, the one-touch fitting is on the lower part.
Press down the clip by making a sound like "click" and connect the fitting to the external exposed tubing by
making "click" sound.
4. Follow same direction as above 1 and 2 to connect R.O. Pre Pack and R.O. Post Pack except R.O
membrane pack.
12
2-4. The Connection of Feed Water
aquaMAXTM-Basic 360 Series
Figure 2-3 The Connection of Feed Water
Cut the Ø6 seal hose tubing supplied with the system properly and connect to the one-touch type of FEED
fitting on the back-side. Then please connect the other side to the tap water.
2-5. The Connection of the Produced Water
For the aquaMAX 360, 362 type, close the one touch fitting on ION (PRODUCT 2) on the back side with a
supplied plug and insert Ø6 tube into the one touch fitting at R.O.(PRODUCT 1) as following Figure 2-3.
Connect the other side of tubing to the reservoir tank
For the aquaMAX 361, 363 type, connect tubings to both R.O.(PRODUCT 1) and ION (PRODUCT2) as following
Figure 2-3-1. It can produce either R.O water or De-ionized water by selection from a service menu. It’s
available to connect a sensor to check the level of water in a tank when you connect a reservoir tank.
13
aquaMAXTM-Basic 360 Series
Figure 2-4 The Connection of the Produced Water 1(aquaMAX 360, 362)
Figure 2-5 The Connection of the Produced Water 2(aquaMAX 361, 363)
14
aquaMAXTM-Basic 360 Series
2-6. The Connection of the Rejection Water
Connect by pushing the supplied Ø6 tubing to the REJECT on the back side and place the other side of the
tubing on the nearest waste hole.
Figure 2-6 The Connection of the Rejection Water
15
aquaMAXTM-Basic 360 Series
2-7. How to Connect Feed Water Kit
1. Purpose of Installation
On installation of water purification system, there are some difference at water pressure depending of location,
time and area. So, a decompression valve is built in the unit to prevent housing damage, pump error and fitting
separation. When you install water purification system, you must connect a tap water connector.
2. Details
1) Connection of tap water connector
a. Insert a bushing (1-1) of tap water connector(1) into a tap, and
then tighten a screw(+).
b. Turn a nipple (1-2) counterclockwise up to bushing completely.
-There is a rubber-ring to protect from leakage and helps installation.
3. Caution
The use with a booster pump is required when the water pressure of installation area is lower than
0.5kg/ ~ 1kg/ to meat the product flow rate on the specification.
16
aquaMAXTM-Basic 360 Series
2-8. The Connection of Reservoir Tank Sensor (Optional)
1. Purpose
With use of a reservoir tank, it protects from overflowing of the produced water.
2. Details
1) Install a tank sensor on the cap of 20L or 50L PE tank.
2) Connect one side of connector of Level Switch Outer Cable to the other connector of which connected to a
tank sensor. The direction of Level Switch Outer Cable doesn’t matter.
3) Connect the other side of connector of installed reservoir tank to Level S/W at the rear side of the system.
3. Caution
The system keeps on 「Standby」 mode when the level of water is up at installation of Level Switch. You’re
only able to re-operate the system after reduce the produced water in the tank.
17
aquaMAXTM-Basic 360 Series
Chapter 3. Operation
3-1. Operation Mode
To operate the aquaMAX, use a keyboard on upper of the front side. There are following modes available on
LCD.
Basic Mode
Operate/Standby Select the Operateor Standbymode.
Measure Use to measure following values during operation
- Temperature
- Conductivity of both the feed water and the produced water from RO
membrane
- Resistivity of Ion Exchange Filter / Ion Rejection Ratio of R.O membrane and
Ion Exchange Filter
Service Select the following function on「Standbymode.
- For the aquaMAX 361/363, select the rejection water(RO/ION) with Measure
button.
- Drain the low-grade water during initial flushing or operation.
- Use for decreasing inner pressure of filter.
- Notice for filter replacement.
Select Use to perform Measure or Service
LED Display
Power Confirm if power is supplied normally.
Operate Lights on 「Operate」 or 「Service」. Standby Lights on 「Standby」 or 「Service」.
Fault Lights when normal operation is not possible.
18
aquaMAXTM-Basic 360 Series
3-2. Initial Operation
After connection of all tubes described above, please follow instructions as below in order to normally operate
the system.
Check if the filters are installed properly. Check if there is any leakage.
(For the ion exchange type, Check the connection of ion exchange filter)
Check th e tube connection of the feed water and power supply
Open the valve of water supply and wait for the prefilter to be filled water.
Turn on, then the system is on the starts flushing for 5 min. for R.O filter. Pressing Operate/Standby key once more makes the system stop and
get back to
⑤ On
Standbymode, press Operate/Standby key to produce water. Press Operate/Standby key again to
stop and get back to
After theOperatemode, the RO pressure is gradually increased. Optimal ratio between the purified water
and the rejection water is around 30 : 70 and optimal RO pressure is around 5~6 Kg/cm
Standbymode. (For the filter, completing the flushing is required.)
Standbymode
Standbymode. After pressing Operate/Standby key, the instrument
2
.
3-3. Initial Flushing
When use a new RO membrane cartridge, perform the initial flushing for 4 hours after turn on the system in
order to flush the cartridge inside. During initial flushing, if the purified water quality does not reach to the
required purity, waste the produced water. After flushed the RO cartridge, push the Operate/Standby key to
stop producing the purified water and connect the purified water tubing to a reservoir tank. The performance of
the aquaMAX is varied depending on water quality of the feed water. Calculate the Ion Rejection Ratio(water
purity) from following formula.
aquaMAX Ion Rejection Ratio(%)
Example : 95.23% =
Feed Water Conductivity – Purified Water Conductivity
=
Feed Water Conductivity
210uS – 10uS
210uS
19
X 100(%)
X 100(%)
3-4. Operation of the Keys
On the front panel, there are 4 keys, LED for modes display and back light LCD.
aquaMAXTM-Basic 360 Series
Basic Mode
Operate/Standby
Select the 「Operate」or 「Standby」mode.
Measure Use to measure following values during operation
- Temperature
- Conductivity of both the feed water and the produced water from RO
membrane
- Resistivity of Ion Exchange Filter / Ion Rejection Ratio of R.O membrane and
Ion Exchange Filter
Service Select the following function on「Standbymode.
- For the aquaMAX 361/363, select the rejection water(RO/ION) with Measure
button.
- Drain the low-grade water during initial flushing or operation.
- Use for decreasing inner pressure of filter.
- Notice for filter replacement.
Select Use to perform Measure or Service
20
LED Display
Power Confirm if power is supplied normally.
aquaMAXTM-Basic 360 Series
Standby Lights on 「Standby」 or 「Service」.
Fault Lights when normal operation is not possible.
After Power On.
Pressing Operate/Standby key to flush for 5 minutes.
Operate Lights on 「Operate」 or 「Service」.
1. Self Test
BASIC VER 3.00
SELF TEST........OK
2. Display the model
* aquaMAXB360 *
YoungLin Ins.
Flush : Clean the R.O membrane while the pressure valve of R.O opened.
OPERATE  
FLUSH.....05min
Once you press Operate/Standby key or 5 min passes, the mode turns to be「Standby」 mode.
Standby : To protect the RO membrane automatically, it performs flushing every 30 minutes for 1
minute while it does not produce the purified water.
STANDBY
READY………………..
21
aquaMAXTM-Basic 360 Series
3-5. Normal Operation and Standby Mode
The system repeats Operate and Standby mode pressing Operate/Standby key. On normal operation mode,
the system checks water quality, pressure and temperature If any of them is not normal, it displays trouble
accordingly.
1. On Normal Operation : The message below is repeating.
2. On flushing the RO cartridge
AUTO FLUSH : Flushing every 30 minutes for 1 minute with no production of produce the purified
WATER TANK FULL
When the reservoir is fully filled with water, it turns to be「Standby」 mode and following is displayed.
At this moment, any key input will not work until the water level is lowered (however, auto flushing does
work).
OPERATE
Product : ION
water.
OPERATE
AUTO FLUSH .....
OPERATE
RUNNING .......
On Standby Mode Flushing R.O. Cartridge on Standby Mode
STANDBY   
WATER TANK FULL
  STANDBY  
FLUSH .........
AUTO FLUSH : Flushing every 30 minutes for 1 min.
22
aquaMAXTM-Basic 360 Series
3-6. Measuring Point Display
During normal operation, if you push Measure key continuously, the menus for water quality, temperature and ion rejection ratio are appeared. If want to see the values, push Select key to display. If 10 seconds pass or you push the Operate/Standby key, it displays the Normal Operation mode.
<Pressing of Measure Key> < Pressing of Select key on the left side of screen >
¢MEASURE¢
WATER TEM
¢MEASURE¢
R.O. WATER
¢MEASURE¢
ION EX.
¢MEASURE¢
24 -
¢MEASURE¢
2.4uS
¢MEASURE¢
10.3MΩ 99.1%
23
aquaMAXTM-Basic 360 Series
3-7. Service Operation
OnStandby」mode, pressing Service key, the menu for cleaning filters and tubing is displayed. On this, push Select key, then it performs「Service Operationand Operate LED is lightened on. After Service Operation is finished or you press the Operate/Standby key, it turns to the 「Standby」mode.
<Pressing SERVICE key > <Select by pressing Measure key> < Pressing Select key>
¢SERVICE¢
Select RO/ION
¢SERVICE¢
Product : RO
¢SERVICE¢
RO Selected
¢SERVICE¢
FLUSH
¢SERVICE¢
PRESSURE DOWN
¢SERVICE¢
CHANGE FILTER
¢SERVICE¢
WASH END
¢SERVICE¢
Wait ........OK!
¢SERVICE¢
CHANGE START
- Select RO/ION
Selection for the type of product water is available. This menu is shown in aquaMAX-Basic 361/363 type only.
The type of product water can be changed by Measure button and selected by Select button.
- Flush
On the first operation or flushing filters, this function is required and it drains off wastewater for 5 min.
Pressing Operate/Standby key once more makes the system stop and get back to
Standbymode.
- Pressure Down
It’s used to reduce inlet pressure when any parts of filters need to be replaced.
- Change Filter
The system showsChange Filtermessage if the filter life is finalized. And it reduces inlet pressure to
replace filter. At this point, filter life is reset.
24
aquaMAXTM-Basic 360 Series
3-8. Errors and its Display
If any of troubles is occurred and normal operation is not possible, the error message is displayed. On error
occurrence, it’s required to press the Operate/Standby key, to release the error message. During error display,
the flushing of the RO cartridge is performed as well except feedwater malfunction.
Table 3-1Display and Causes of System Troubles
Error Message Causes and Symptoms
¢FAULT¢
LOW FEED PRESS
¢FAULT¢
HIGH PRESS ERROR
¢OPERATE¢
R.O. WATER BAD
¢OPERATE¢
ION WATER BAD
¢OPERATE¢
CHANGE FILTER
¢STANDBY¢
WATER TANK FULL
- In case of overuse of filter life, the system stops the operation with FILTER time CHECK message when the
The feedwater pressure is low. The switch for the water supply is turned off and the pump is stopped.
If there is high pressure occurred in the system, decompressing gets started to protect filters and pump from damage. If the high pressure still remains even though the decompressing, the system is stopped to protect.
The water quality of the R.O. permeate is not good. Even after the Fault LED on, the operation keeps working.
The water quality of Ion Exchange filtrate is not good. Even after the Fault LED on, the operation keeps working.
An alarm message pops up when all filters and pump need to be replaced after lifetime.
The reservoir tank is fully filled with water. At this moment, the system is on lowered.
Standbymode until the water level is
power on. To get back to 「Standby」mode , press Operate/Standby button.
- To replace filters, run CHANGE FILTER function on Service menu.
- Normal operation is possible without filter replacement, but thefilter(to be replaced) change message
appears every 30 min.
Standby
FILTERtime CHECK RO Filter Change
25
Operate
aquaMAXTM-Basic 360 Series
3-9. System Shut Down
Emergency stop
To stop the system in case of trouble occurrence during operation, push the Operate/Standby key.
Shutdown within 48 hours
To stop the system within 48 hours, push the Operate/Standby Key. On the Standby mode, the system can restart to operate at any time with Operate/Standby Key. Pay attention not to stop the power or water supply
because RO membrane works as well.
26
aquaMAXTM-Basic 360 Series
Chapter 4. Maintenance
4-1. Maintenance and Solutions
The Young Lin aquaMAX system requires no special maintenance procedures. However, in order to ensure the
expected system performance and long-term trouble-free operation, it is essential to observe the following
points.
Filters in the aquaMAX Basic
Figure 4-1 Filters in the aquaMAX Basic
: There are total 3 types of filters used in aquaMAX Basic. (2 types when no ion exchange filter used)
The filter used at first is PRE PACK(The middle one out of three shown. The right one when there are only 2
filters). It’s consisted of a prefilter and carbon filter. Even though lifetime depends on
water
and amount of produced water, it usually requires replacement every 3 month ( 2~3 weeks with bad
water quality of feed
quality of feed water).
27
aquaMAXTM-Basic 360 Series
The second filter that water passes through is RO PACK (The very left one out of 3 filters or 2 filters) and it’s
consisted of a R.O.mebrane and carbon filter. It only needs the replacement every year and the right time to replace will be shown on the LCD. On operate mode, water quality can checked (Measure key Æ Select “ RO
water”). The right value for water quality is within 5uS-10uS, and if it’s higher than 10uS, filter has to be replaced.
The third one is POST PACK (The very right one.) and it’s consisted of an ion filter and carbon filter. Its life is
around 3 months and the right time to replace will be shown on the LCD.
On operate mode, water quality can checked (Measure key Æ Select “ ION water”).Water quality within 5MΩ ~10MΩ is normal, but if it’s lower than 5MΩ , filter has to be replaced.
NOTE : It is required to perform extra pretreatment to get expected filter life if feedwater is ground water or water from somewhere with a lot of turbidity or iron.
28
aquaMAXTM-Basic 360 Series
4-2. Filter Replacement
Figure 4-2 Filter Replacement
Close the feed water.
Push the Service key and Select key at Standby mode to perform “pressure down”. Perform it twice.( It takes
about 8 sec at one time) . Run “Change Filter” to replace filter when the filter life is done.
③ Open the door to replace filters.
* Take out the filter using a tool for use of filter replacement.
(Once hearing the "click" sound from the clip, grab the filter and slowly pull it out.)
* After separating the filters, press the housing by making sounds like “ click”.
* Connect the filters by each name. (The right connection makes sounds like “ click”.)
* For the R.O membrane pack, press down the fitting clip on the lower part of filter making a sound like "click"
to remove the connected fitting.
CAUTION : Put a beaker at the external end of drain tube in case small amount of waste water can be
drained.
Open the feedwater. Run “ Flush” on Standbymode.(Service key Æ Select “Flush”.)
29
aquaMAXTM-Basic 360 Series
4-3. Trouble Shooting
When the system is not normally operated, follow the procedures as described below.
In case the system doesn’t operate normally:
Check the system is operated at appropriate power and changes in pressure.
We don’t guarantee any malfunctions from operations by not following user manual.
In case the system doesn’t work:
Open the rear door and check if the fuse is short. There is a 250V and 1A fuse.
Low Permeate(Purified Water) Flow Rate
If the RO pressure is so low, it effects to the amount of purified water.
If the water temperature is too low, the flow rate can b e lowered.
If the RO membrane is blocked or contaminated, the flow rate drops and the pressure is also rapidly
increased.
30
aquaMAXTM-Basic 360 Series
4-4. Fuse Replacement
When it comes to replacing the fuses, replace to the new fuse supplied at the rear of power switch.
Figure 4-3The Fuse Replacement
31
System Check List
TYPE : SERIAL NUMBER :
aquaMAXTM-Basic 360 Series
1. PRE PACK
2. Booster Pump
3. Conductivity Cell for R.O. water supply
4. R.O. PACK
5. R.O. valve for drain
6. R.O. Conductivity cell for produced water from R.O.
7. POST PACK
8 Resistivity cell for produced water from ion exchange
Comments:
Measured by
Name Signature Date
Note) Please place its copy on the right side of system.
32
aquaMAXTM-Basic 360 Series
Part List
R.O. Water Purification System
aquaMAX 360 : Standard Type 3601011000
aquaMAX 361 : Ion Exchange Filter Addition Type 3601011001
aquaMAX 362 : Extended Capacity Type 3601011002
aquaMAX 363 : Ion Exchange Filter Addition and Extended Capacity Type 3601011003
Consumables
R.O PRE PACK 3101333512
R.O POST PACK 3101333522
R.O Membrane PACK 1 3101333502
R.O Membrane PACK 2 3101333503
Optional
PE Tank, 20 L 1002712300
PE Tank, 50 L 1002712310
Tank Sensor Assembly 3101012080
Young Lin Bldg., 899-6, Hogye, Anyang, Korea 431-836
Tel : +82-31-428-8700
Fax : +82-31-428-8779
Email :
Homepage : http://www.younglin.com
younglin@younglin.com
33
Loading...