11.1 Spare parts column stand .................................................................................. 46
11.2 Spare parts adjustable end piece ...................................................................... 46
11.3 Spare parts column body 70 – 100 mm ID ........................................................ 47
11.4 Spare parts column body 140 – 200 mm ID ...................................................... 48
12 Index ....................................................................................................................... 49
Page 5
YMC Europe GmbH
Information on manual
YMC Pilot
5 / 49
PLUS
low pressure chromatography column
DANGER
WARNING
CAREFUL
ATTENTION
1 Information on manual
1.1 General
This manual includes important massages and information for the intended use of this
pilot column. The manual must be kept accessible for the user at all times.
In order to avoid injuries and damages, read this manual carefully prior to use and act
on the recommendations given.
1.2 Explanation of warnings
Warnings are highlighted by coloured signal word fields. Always read the full text of
the warning in order to effectively protect yourself from damage!
The following signal word fields marked by different colours and signal words distinct
security levels:
Disregarding this warning will result in death or serious injuries.
Disregarding this warning may result in death or serious injuries.
Disregarding this warning may result in minor or moderate injuries.
Disregarding this warning can result in damage to surrounding property.
Warnings are always in the same structure. They include signal word, type and source
of danger, consequences of neglect and security measures / avoidance.
Page 6
YMC Europe GmbH
Information on manual
YMC Pilot
6 / 49
PLUS
low pressure chromatography column
Example for additional information
1.3 Structure of instructions
The instructions relate directly to an activity. They are structured in a task oriented
manner. Always perform the individual steps in the order in which they are given.
Instructions are structured as follows and are marked by the corresponding levels:
Aim of the instruction
1. Action step
Impact of the action step to check whether the step has been performed
correctly.
2. Another action step
Figures within instructions:
A figure is only related to the action step which is listed directly below the figure.
1.4 Guidelines
The following guidelines are available in this manual:
• Contents at the beginning
• Index in the end
• Name of the main Chapter in the header
• Links to other Chapters, for example “Information in this manual“ ( Chapter 1, S.
5)
• Additional information and hints indicated with an i-symbol
1.5 Figures
Figures can differ in the graphical presentation from the product.
Page 7
YMC Europe GmbH
Safety
YMC Pilot
7 / 49
PLUS
low pressure chromatography column
2 Safety
This section gives an overview of all the important safety aspects for the protection of
personnel and for the safe and trouble-free handling of the YMC Pilot
Other task-specific safety requirements are contained in the sections relating to the
respective operation. In addition, the national and local legal and other applicable
safety regulations must be observed.
2.1 Intended use
The pilot columns are only to be used for the separation of substances via
chromatographic techniques.
The intended use of this product must be performed with regards to the specifications
Chapter 3.2, S. 9. The product is only for use with liquids that have a pressure <0.5
bar above the atmospheric pressure at the intended maximum temperature (40°C).
Every use which is outside these specifications is defined as not intended use. The
intended use includes the compliance with the advice and terms in the manual.
PLUS
column.
2.2 General safety instructions
In order to avoid accidents and injuries,
• operate the pilot column only within its intended use.
• follow all general and special safety instructions as well as all warnings in this
manual.
• follow danger signs and sign giving instructions.
• follow the rules and regulations for prevention of accidents at the site of operation.
• comply with all inspection and maintenance intervals.
2.3 Personnel requirements
The various tasks described in this manual make different demands on the
qualifications and training of the persons carrying them out.
The user must:
• have competent and proficient training.
• be trained in dealing with pressure devices.
• read and fully understand this manual before using this YMC Pilot
PLUS
column.
Page 8
YMC Europe GmbH
Safety
YMC Pilot
8 / 49
PLUS
low pressure chromatography column
Symbol
PPE
Phase of life
Action
Protective gloves
Operation
• Column packing
• Column unpacking
Initiation
• Cleaning
Safety goggles
Operation
• All actions
Initiation
• Cleaning
• Pressure testing
Dust mask
Operation
• Slurry preparation
Protective clothing
Operation
• All actions
Initiation
• Cleaning
• Pressure testing
2.4 Personnel protective equipment
In order to reduce the risks during working on the column, always use the
recommended personnel protective equipment.
The YMC Pilot
is compatible with common BioLC separation modes (IEX, SEC, Affinity, IMAC, HIC,
MMC) and aqueous conditions. Due to the metal-free and the minimal dead volumes,
the column is perfectly suitable for the separation of biomolecules (e.g. proteins,
oligonucleotides or nucleic acids).
3.2 Specifications
PLUS
pilot column is especially designed for pilot scale use. The column
*Pressure Equipment Directive
1
Depending on packing material and packing method
Page 10
YMC Europe GmbH
Specifications and overview
YMC Pilot
10 / 49
PLUS
low pressure chromatography column
If the desired solvent is not listed, YMC will be pleased to advise you.
Material
Borosilicate
glass 3.3
FEP
75
Fluoro-
pren
XP41
PE
PTFE
PEEK
Lösungsmittel
Acetone
Good
Good
Unstable
Good
Good
Good
Acetonitrile
Good
Good
Limited
Good
Good
Good
Dihydrogen phosphate
Good
Good
Good
Good
Good
Good
Cyclohexane
Good
Good
Good
Limited
Good
Good
Dichloromethan
Good
Good
Limited
Unstable
Good
Unstable
0.1M EDTA (3%)
Good
Good
Good
Good
Good
Good
1M Acetic acid(6%)
Good
Good
Good
Good
Good
Good
Ethanol
Good
Good
Good
Good
Good
Good
Ethyl acetate
Good
Good
Unstable
Good
Good
Good
n-hexane
Good
Good
Good
Limited
Good
Good
Isopropanol
Good
Good
Good
Good
Good
Good
Methanol
Good
Good
Good
Good
Good
Good
2M Sodium hydroxide
(8%)
Good
Good
Good
Good
Good
Good
1M Hydrochloric acid
(4%)
Good
Good
Good
Good
Good
Good
8M Urea (36%)
Good
Good
Good
Good
Good
Good
1M NaCl
Good
Good
Good
Good
Good
Good
0,5M Na2SO4
Good
Good
Good
Good
Good
Good
3.3 Chemical Resistance
The chemical resistance depends on the wetted materials used Chapter 3.2, S. 9.
YMC recommends avoiding the following chemicals:
• Strong oxidizing agents
• Halogenated compounds
• Chlorinated solvents
• Aromatic compounds (e.g. toluene)
• High concentrations of strong acids
• THF
Page 11
YMC Europe GmbH
Specifications and overview
YMC Pilot
11 / 49
PLUS
low pressure chromatography column
A parts list with all wetted parts and non-wetted parts is in the appendix.
Adjustable end piece
Column body
Column stand
3.4 Overview
The YMC Pilot
PLUS
column consists of the adjustable end piece, the column body and
the column stand. These parts can be assembled whilst operating the column without
any tools.
Page 12
YMC Europe GmbH
Specifications and overview
YMC Pilot
12 / 49
PLUS
low pressure chromatography column
1
Tubing
4
Star screw
2
Spindle
5
Closing disc (AEP)
3
Rotary Handle
6
Connecting flange
1
Spindle
4
Frit
2
Connecting flange
5
Profile gasket
3
Plunger (AEP)
3.4.1 Adjustable end piece
The adjustable end piece consists of a closing disc and the piston. The piston consists
of the spindle, the connecting flange, the plunger, the frit and the profile gasket. A
tubing goes through the spindle to the connecting flange. With the rotary handles in
the closing disc, the piston can be adjusted up and down. The closing disc is attached
with star screws to the column body.
Piston front
Page 13
YMC Europe GmbH
Specifications and overview
YMC Pilot
13 / 49
PLUS
low pressure chromatography column
1
Cylinder screw
6
Connector
2
Column disc
7
Glass body
3
Connector drainage tubing
8
Rod 4 Drainage tubing
9
Holding bar
5
Closing disc (CB)
3.4.2 Column body
The glass body is fixed between the column disc and the closing disc by the rods. The
glass body has a visible scale. Excessive solvent is removed during piston insertion
via the connector drainage tubing. The fixed end piece is assembled with the closing
disc (CB). The fixed end piece consists of the connector, the sealing washer, the
plunger (CB), the ring gasket B, the frit (CB) and the ring gasket A. The column body
can be assembled in the stand in two different positions.
Page 14
YMC Europe GmbH
Specifications and overview
YMC Pilot
14 / 49
PLUS
low pressure chromatography column
1
Connector without cap nut
4
Frit (CB)
2
Plunger (CB)
5
Ring gasket A
3
Ring gasket B
1
Levelling feet
3
Frame disc
2
Main pillar
4
Locking bolts
Frit column body
3.4.3 Column stand
With the locking bolts in the frame disc the column body is attached to the column
stand. With the adjustable levelling feet, the pilot column can be aligned vertically.
Alternatively, the column stand is available with casters.
Page 15
YMC Europe GmbH
Delivery
YMC Pilot
15 / 49
PLUS
low pressure chromatography column
YMC recommends retaining and storing the original packaging and the
foam cushion so that the column can always be shipped safely.
4 Delivery
4.1 Delivery contents
• Pilot column
• Documentation folder:
– Brochure YMC Glass Columns
– Manual
– Overview drawings
– YMC Quality Control / Leakage Tests
– Certificate of conformity FDA/BSE/TSE/USP Class VI
4.2 Delivery inspection
Delivery inspection
Delivery inspection is performed as follows:
1. Check the delivery for visible transport damage.
2. Check the delivery for the completeness of parts.
3. Compare the column description on the glass body with the data of the delivery
note.
Transport damages
Procedure, if you observe any visible transport damage:
1. Refuse or reserve acceptance of the delivery.
2. Record the transport damage on the delivery note of the transporter or in the
transport documents.
3. Inform YMC Europe GmbH and/or the responsible supplying agent.
Page 16
YMC Europe GmbH
Requirements for operation
YMC Pilot
16 / 49
PLUS
low pressure chromatography column
WARNING
WARNING
5 Requirements for Operation
5.1 Pilot column set up
Required devices:
• Drainage tubing
• Collecting vessel
The pilot column is set up as follows:
1. Make sure that the pilot column is standing on fixed ground.
2. Make sure that the pilot column is standing vertical in the final operating place.
3. Connect the drainage tubing to the drainage tubing connector.
4. Put the drainage tubing in the collecting vessel.
5.2 Pressure test
Operation of pressure tests
Eye or skin injuries possible due to solvent leakage or incorrectly sealed connectors.
► Perform the pressure test behind a safety screen.
► During work on the pilot column, wear the prescribed personnel protective
equipment.
Pilot column under high pressure
Eye or skin injuries possible due to remaining pressure in the tubing or in the pilot
column.
► Stop the pump during work on the pilot column.
► During work on the pilot column, wear the prescribed personnel protective
equipment.
A pressure test is used to check the tightness of both the seals in the pilot column and
also the tubing connectors. Before using the column for chromatographic applications,
it is recommended that a pressure test is performed to verify the correct assembly.
Prior to delivery of the column, the column has been tested according to the standard
procedure listed below. The relevant certificate is located in the documentation folder.
Page 17
YMC Europe GmbH
Requirements for operation
YMC Pilot
17 / 49
PLUS
low pressure chromatography column
1
Pilot column
4
Pressure gauge
2
Back pressure regulator
5
Gate valve
3
Gate valve
6
Pump
If a back pressure regulator is used, the maximum pressure is reached
easier.
Leakages at the piston of the adjustable end piece are often related to
impurities trapped in the seal.
Standard procedure pressure test
The pressure test should be performed as follows:
1. Assemble the system according to the schematic overview of a pressure test.
2. Turn the piston of the adjustable end piece in the DOWN direction as far as it
will go.
3. Fill the pilot column with isopropanol.
4. Increase the testing pressure in the pilot column by slow and stepwise
adjustment of the flow rate at the pump.
The testing pressure is the pressure limit of the column plus 10% (70
mm ID = 17.6 bar | 100 mm ID = 11 bar | 140 mm ID = 7.7 bar | 200
mm ID = 5.5 bar)
5. Keep the testing pressure constant for 10 minutes.
6. If leakages occur, check the connections. If required, replace tubing and/or
connectors.
If the testing pressure was constant for 10 minutes, the column is ready for use.
Page 18
YMC Europe GmbH
Operation
YMC Pilot
18 / 49
PLUS
low pressure chromatography column
WARNING
6 Operation
Risk of explosion
Severe injuries possible due to glass splinters.
► Never put the glass column under gas pressure!
In order to guarantee safe operation, the pilot column must be used for the intended
use only. ( Chapter 2.1, S. 7).
6.1 Removal of adjustable end piece
Required devices:
• Clean and soft surface
The adjustable end piece is removed as follows:
1. Turn the piston of the adjustable end piece in the UP direction as far as it will
go.
2. Remove the star screws from the closing disc (AEP).
3. Remove the adjustable end piece from the glass body and put it on a clean
and soft surface.
Page 19
YMC Europe GmbH
Operation
YMC Pilot
19 / 49
PLUS
low pressure chromatography column
CAREFUL
CAREFUL
The use of a more dilute liquid slurry simplifies the column packing. The
total volume of the slurry should not exceed the maximum volume of the
column.
The removal of fines decreases the column back pressure and so
improves the performance of the packed column.
6.2 Pilot column packing
Packing material
Eye and skin injuries possible through contact with packing material.
► Wear the prescribed personnel protective equipment.
Two different procedures are common for slurry packing the column:
• under flow,
• with sedimentation.
6.2.1 Slurry preparation
Formation of dust with dry packing material
Lung irritation possible if packing material is inhaled.
► Read the material safety data sheet from the packing material manufacturer.
► Wear a dusk mask.
The slurry is prepared as follows:
1. Mix the solvent or buffer with the appropriate amount of packing material in a
suitable container (see manufacturer’s recommendations).
The total volume of the slurry must be less than the maximum column volume.
The amount of packing material determines the final bed length.
2. Mix the slurry with a stirring rod until the slurry is homogenous and avoid
introduction of air bubbles.
3. Check if the material is suitable for ultrasonication (see manufacturer’s
recommendations).
If the material is suitable for ultrasonication, degas the slurry in an ultrasonic
bath.
If the material is not suitable for ultrasonication, let the material settle.
4. Let the degassed slurry settle.
5. If fines are visible on the surface of the slurry, remove them.
Page 20
YMC Europe GmbH
Operation
YMC Pilot
20 / 49
PLUS
low pressure chromatography column
6.2.2 Slurry column packing under flow
The column is packed with the slurry method under flow as follows:
1. Remove the adjustable end piece Chapter 6.1, S. 18.
2. Check the drainage tubing for correct connection.
3. Check the drainage tubing outlet for correct position in the collecting vessel.
4. Connect the pump with the tubing at the column body.
5. Start the pump.
6. Fill the glass body with solvent or buffer until 2-3 cm above the frit in the
column body (bottom-up).
This ensures that no air is trapped under or is inside the frit.
7. Close the valve in the tubing on the column body.
8. Connect the pump with the adjustable end piece.
9. Flush the piston of the adjustable end piece with slurry solvent until the frit of
the adjustable end piece is wetted completely.
10. Disconnect the pump from the adjustable end piece.
11. Pour the homogenous and/or degassed slurry into the glass body.
12. Rinse the inner wall of the column body with slurry solvent for removal of
particles.
By removing particles from the inner wall of the column body, a correct seal
with the profile gasket is guaranteed.
13. Fill the pilot column up to the edge of the drain function with slurry solvent.
14. Turn the closing disc of the adjustable end piece in the UP direction as far as it
will go.
Page 21
YMC Europe GmbH
Operation
YMC Pilot
21 / 49
PLUS
low pressure chromatography column
15. Turn the closing disc of the adjustable end piece half turn in the DOWN
direction.
The adjustable end piece is ready for insertion into the column body.
16. Put the piston of the adjustable end piece in a smooth angle opposite the drain
function on the column body.
17. Bring the piston of the adjustable end piece slowly to the vertical position.
Excessive slurry solvent will flow into a reservoir through the drainage tubing.
Page 22
YMC Europe GmbH
Operation
YMC Pilot
22 / 49
PLUS
low pressure chromatography column
18. Align the marks on the closing disc (AEP) and column disc.
19. Screw the 4 star screws continuously in a diagonal pattern to each other.
20. Open the valve in the tubing on the adjustable end piece.
21. Turn the piston of the adjustable end piece in the DOWN direction until the
tubing is completely filled with liquid.
The tubing is now free of air.
22. Connect the pump with the tubing of the adjustable end piece.
23. Open the valve in the tubing on the column body.
Page 23
YMC Europe GmbH
Operation
YMC Pilot
23 / 49
PLUS
low pressure chromatography column
24. Start the flow (top down).
25. Increase the flow rate stepwise until the desired pressure / flow rate is
reached. Wait after every step until the pressure is constant. Afterwards, the
flow rate can be further increased.
Consider the manufacturer’s recommendations for the packing material.
The flow rate should be considerably (> 20%) higher than the operating flow
rate.
The column bed is visually compressed.
26. Check and mark the bed length frequently.
27. If the final bed length is reached, stop the pump.
28. Turn the piston of the adjustable end piece in the DOWN direction until the
column bed is reached.
29. Start the pump.
30. Check, if the column bed is further compressed. A dead volume would occur.
31. If a dead volume occurs, stop the pump and repeat step 28 to 30.
32. If no dead volume occurs, the pilot column is released Chapter 6.3, S. 27.
33. Remove liquid above the connecting flange through the two holes in the
adjustable end piece.
YMC recommends flushing this area if high salt concentrations are used.
34. Equilibrate the pilot column with the desired eluent and test the column
performance.
Page 24
YMC Europe GmbH
Operation
YMC Pilot
24 / 49
PLUS
low pressure chromatography column
6.2.3 Slurry column packing with sedimentation
The column is packed via the sedimentation slurry method as follows:
1. Remove the adjustable end piece Chapter 6.1, S. 18.
2. Check the drainage tubing for correct connection.
3. Check the drainage tubing outlet for correct position in the collecting vessel.
4. Connect the pump to the tubing of the column body.
5. Start the pump.
6. Fill the glass body with solvent or buffer until 2-3 cm above the frit in the
column body (bottom-up).
This ensures that no air is trapped under or is inside the frit.
7. Close the valve in the tubing of the column body.
8. Disconnect the pump from the column body.
9. Connect the pump to the adjustable end piece.
10. Flush the piston of the adjustable end piece with slurry solvent until the frit in
the adjustable end piece is wetted completely.
11. Disconnect the pump from the adjustable end piece.
12. Pour the homogenous and/or degassed slurry into the glass body.
13. Rinse the inner wall of the column body with slurry solvent in order to remove
particles.
By removing particles from the inner wall of the glass body, a correct seal with
the profile gasket is guaranteed.
14. Do not move the pilot column anymore.
YMC recommends covering the open column body during sedimentation with
some plastic film.
15. Wait until the packing material has settled (sedimentation).
Depending on the packing material, this can take some hours.
16. Fill the pilot column up to the edge of the drain function with slurry solvent.
17. Turn the closing disc of the adjustable end piece in the UP direction as far as it
will go.
Page 25
YMC Europe GmbH
Operation
YMC Pilot
25 / 49
PLUS
low pressure chromatography column
18. Turn the closing disc of the adjustable end piece half turn in the DOWN
direction.
The adjustable end piece is ready for insertion into the column body.
19. Put the piston of the adjustable end piece in a smooth angle opposite the drain
function on the column body.
20. Bring the piston of the adjustable end piece slowly to the vertical position.
Excessive slurry solvent will flow into a reservoir through the drainage tubing.
Page 26
YMC Europe GmbH
Operation
YMC Pilot
26 / 49
PLUS
low pressure chromatography column
21. Align the marks on the closing disc (AEP) and column disc.
22. Screw the 4 star screws continuously in a diagonal pattern to each other.
23. Open the valve in the tubing on the adjustable end piece.
24. Turn the piston of the adjustable end piece in the DOWN direction until the
column bed is reached.
25. Connect the pump to the tubing of the adjustable end piece.
26. Open the valve in the tubing at the column body.
27. Start the flow (top-down).
28. Increase the flow rate stepwise until the desired pressure / flow rate is
reached. Wait after every step until the pressure is constant. Afterwards, the
flow rate can be further increased.
Consider the manufacturer’s recommendations for the packing material.
The flow rate should be considerably (> 20%) higher than the operating flow
rate.
The column bed is visually compressed.
Page 27
YMC Europe GmbH
Operation
YMC Pilot
27 / 49
PLUS
low pressure chromatography column
29. Check and mark the column bed length frequently.
30. If the final bed length is reached, stop the pump.
31. Turn the piston of the adjustable end piece in the DOWN direction until the
column bed is reached.
32. Start the pump.
33. Check if the column bed is further compressed. A dead volume would occur.
34. If a dead volume occurs, stop the pump and repeat step 31 to 33.
35. If no dead volume occurs, the pilot column is released. Chapter 6.3, S. 27.
36. Remove liquid above the connecting flange through the two holes in the
adjustable end piece.
YMC recommends flushing this area if high salt concentrations are used.
37. Equilibrate the pilot column with the desired eluent and test the column
performance.
6.3 Column test and release criteria
The column test and the release criteria are selected with regards to the later
application field. The preparation of the required specifications for the final usage of a
method or for the purification of a target molecule is determined individually. YMC will
gladly support you.
Page 28
YMC Europe GmbH
Operation
YMC Pilot
28 / 49
PLUS
low pressure chromatography column
WARNING
CAREFUL
6.4 Pilot column unpacking
Pilot column under high pressure
Eye or skin injuries possible due to remaining pressure in the tubing or in the pilot
column.
► Stop the pump during working on the pilot column.
► While working on the pilot column, wear the prescribed personnel protective
equipment.
Packing material
Eye and skin injuries possible through contact with packing material.
► Wear the prescribed personnel protective equipment.
The column is unpacked under flow as follows:
1. Close the valve in the tubing of the column body.
2. Disconnect the pump from the column body.
3. Ensure that the valve at the adjustable end piece is open.
4. Disconnect the pump from the adjustable end piece.
5. Turn the piston of the adjustable end piece in the UP direction until the profile
gasket does not touch the glass body anymore.
The loosening of the profile gasket from the glass body is noticeable.
6. Remove the adjustable end piece Chapter 6.1, S. 18.
7. Put a collecting vessel below the column stand.
Page 29
YMC Europe GmbH
Operation
YMC Pilot
29 / 49
PLUS
low pressure chromatography column
8. Loosen the locking bolts and lock them by turning slightly.
9. Lift the column body vertical from the column stand using two holding bars
which are vertically above each other.
10. Pour any unpacked material and/or solvent into a collecting vessel.
11. If no material now remains in the column body, clean the column body
Chapter 7.3, S. 33.
12. If material remains in the column body, put the column body into the stand with
the open end facing down.
13. Turn the column body slowly in the frame disc until the locking bolts lock.
You will hear a noticeable “click” when the locking bolts are locked.
14. Connect the pump with the tubing at the column body.
15. Open the valve in the tubing of the column body.
16. Start the pump.
Because of the flow, the packing material is pressed out of the column body
into the collecting vessel.
17. Clean the pilot column Chapter 7.3, S. 33.
Page 30
YMC Europe GmbH
Cleaning and maintenance
YMC Pilot
30 / 49
PLUS
low pressure chromatography column
WARNING
WARNING
WARNING
Consumable parts lists are included in the appendix.
7 Cleaning and maintenance
Incorrect maintenance
Severe injuries when used after incorrect maintenance is possible, if the column or
column parts do not withstand the operating pressure.
► Only authorised and qualified personnel should carry out the maintenance.
Incorrect replacement of spare parts
Severe injuries due to breakage possible if unauthorised spare parts are used.
► Use only the manufacturer authorised spare parts.
Faulty maintenance of the pilot column
Severe injuries due to deterioration possible.
► Meet maintenance intervals.
► Perform the maintenance as described.
Page 31
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
31 / 49
PLUS
low pressure chromatography column
CAREFUL
ATTENTION
7.1 Notes for Cleaning-In-Place (CIP)
Packing material
Eye and skin injuries possible through contact with packing material.
► Wear the prescribed personnel protective equipment.
Cleaning with inappropriate solvents
Material damage possible due to corrosion, leaching or deformation.
► Check the chemical resistance of the wetted parts for the cleaning method before
the CIP procedure.
With the cleaning-in-place procedure, a packed pilot glass column can be cleaned
without unpacking or dismantling. Depending on the contamination, three different
procedures can be used:
1. Regeneration,
2. Sterilisation,
3. Depyrogenation.
Regeneration
Regeneration removes inorganic and organic contamination, which binds to the
chromatography matrix and substantially affects the capacity and resolution of the
column. Commonly these are: lipids, pyrogens, protein aggregates, pigments,
polyphenols or metal complexes.
Sterilisation
Sterilisation is the removal and/or destruction of microorganisms and spores that will
contaminate the purified product. The most common method is the treatment with
sodium hydroxide, acetic acid or ethanol solution.
Depyrogenation
Depyrogenation includes the removal of endotoxins, which can be bound on the
chromatographic material or the column hardware (frits, tubing, etc.) and which can
contaminate the purified product.
Page 32
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
32 / 49
PLUS
low pressure chromatography column
Treatment
Type of
contamination
Sterilisation
Depyrogenation
Pepsin, pH 1.5 - 2
Proteolysis of adsorbed
proteins
No
No
Non-ionic detergents (Triton X100; Tween 80)
Removal of hydrophobic
proteins and lipids
No
No
Cationic detergents, pH 9 - 11
Removal of hydrophobic
proteins and lipids
No
Partly
Urea, 6 - 8 M
Removal of protein
aggregates
No
Unknown
1 - 100 mM EDTA, pH
depending on the stationary
phase
Removal of metal
complexes
No
No
0.1 - 1 M NaOH
Removal of bounded
hydrophobic proteins,
lipopolysaccharides
Good
Good
0.5 - 1 M acetic acid in 60%
ethanol
Removal of lipids,
pigments,
lipopolysaccharides, and
other lipophilic
substances
Very good
Good
Interval
Maintenance
After every operation
• Clean the pilot column.
Daily
• Check the column bed, adjust the piston if necessary.
• Check all tubing and connections regarding leakage Chapter 5.2,
S. 16.
• Fix loosened fixing screws again or replace tubing and connectors.
Every 3 months
• Check the profile gasket in the adjustable end piece.
If damaged, replace it.
• Check the ring gasket of the column body.
If damaged, replace it.
Every 6 months
• Check the frit in the adjustable end piece.
If damaged, replace it Chapter 7.4.1, S. 36.
• Check the frit in the column body.
If damaged, replace it Chapter 7.4.3, S. 40.
Cleaning methods for CIP
7.2 Cleaning and maintenance plan
The specified maintenance plan here is to be regarded as a general recommendation.
Depending on the frequency of use and specific application, the appropriate
maintenance plan and the intervals must be defined locally.
Page 33
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
33 / 49
PLUS
low pressure chromatography column
WARNING
CAREFUL
7.3 Cleaning
Pilot column under high pressure
Eye or skin injuries possible due to remaining pressure in the tubing or in the pilot
column.
► Stop the pump while working on the pilot column.
► When working on the pilot column, wear the prescribed personnel protective
equipment.
Packing material
Eye and skin injuries possible through contact with packing material.
► Wear the prescribed personnel protective equipment.
Required devices:
• Wetted and lint-free rags
• Collecting vessel
• Solvent for column body and adjustable end piece cleaning Chapter 3.3, S. 9
• Cleaning agent for cleaning the outer wall of the column body
Authorised cleaning agents for cleaning the outer wall of the
column body:
• Commercially available dishwashing cleaners
• Aliphatic hydrocarbons
• Aliphatic alcohols
• Aromatic hydrocarbons
• Aldehydes
• Diluted and weak acids
• Bases
Page 34
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
34 / 49
PLUS
low pressure chromatography column
The pilot column is cleaned as follows:
1. Remove the adjustable end piece Chapter 6.1, S. 18.
2. Clean the outer wall of the column body.
3. Clean the column body with an authorised cleaning agent and a wetted, lintfree rag.
4. Put a collecting vessel below the column stand.
5. Loosen the locking bolts and lock them by turning slightly.
6. Lift the column body vertical from the column stand using two holding bars
which are vertically above each other.
7. Put the column body with the open end facing downwards in the column stand.
8. Turn the column body slowly in the frame disc until the locking bolts lock.
You will hear a noticeable “click” when the locking bolts are locked.
Page 35
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
35 / 49
PLUS
low pressure chromatography column
For a rigorous cleaning, dismantle the pilot column.
9. Connect the pump to the tubing in the column body.
10. Start the pump.
11. Flush the column body with an appropriate solvent and/or 1 M sodium
hydroxide.
12. Stop the pump.
13. Disconnect the pump from the tubing in the column body.
14. Loosen the locking bolts and lock them by turning slightly.
15. Lift the column body vertical from the column stand using two holding bars
which are vertically above each other.
16. Insert the column body in the starting position again.
17. Turn the column body slowly in the frame disc until the locking bolts lock.
You will hear a noticeable “click” when the locking bolts are locked.
18. Clean the adjustable end piece with an appropriate solvent and/or 1 M sodium
hydroxide.
Page 36
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
36 / 49
PLUS
low pressure chromatography column
ATTENTION
7.4 Replacement of spare parts
Faulty assembly
Damage to the glass body possible due to faulty assembly.
► Take care during the assembly of the glass body that the maximum torque of 5
Nm for columns with 70 and 100 mm ID respectively 10 Nm for column with 140
mm and 200 mm ID are not exceeded.
► Assemble the cylinder screws in a diagonal pattern and always tighten on the
opposite sites.
► Take care during the assembly of the glass body that all parts are correctly
aligned.
7.4.1 Frit replacement in the adjustable end piece
The frit in the adjustable end piece is replaced as follows.
1. Remove the adjustable end piece Chapter 6.1, S. 18.
2. Remove the profile gasket.
3. Remove the frit.
4. Insert the new frit. The distribution profile faces the plunger.
5. Reattach the profile gasket and press firmly in position.
Page 37
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
37 / 49
PLUS
low pressure chromatography column
ATTENTION
7.4.2 Replacing the glass body
Faulty assembly
Damage to the glass body possible due to faulty assembly.
► Take care during assembly of the glass body that the maximum torque of 5 Nm
for columns with 70 and 100 mm ID respectively 10 Nm for column with 140 mm
and 200 mm ID are not exceeded.
► Assemble the cylinder screws in a diagonal pattern and always tighten on the
opposite sites.
► Take care during assembly of the glass body that affected parts are correctly
aligned
Required devices:
• Cylinder screw key / torque wrench
The glass body is dismantled as follows:
1. Remove the adjustable end piece Chapter 6.1, S. 18.
2. Loosen the cylinder screws stepwise half turn. Do it diagonally: Loosen
cylinder screws always on opposite sites.
3. Remove cylinder screws.
4. Remove column disc from the column body.
Page 38
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
38 / 49
PLUS
low pressure chromatography column
5. Remove flat gasket.
6. Remove glass body carefully.
7. Position the new glass body on the ring gasket A.
8. Check that the glass body is concentric with the gaskets.
9. Put the flat gasket corresponding to the structure on the top edge of the glass
body.
Page 39
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
39 / 49
PLUS
low pressure chromatography column
10. Adjust the column disc over the rods and lower it onto the glass body.
11. Check that the flat gasket is in the groove in the column disc.
12. Check that the glass body is concentric with the column disc.
13. ATTENTIONDamage to the glass body is possible due to faulty
assembly. Note the maximum torque during the assembly of the glass
body.
Insert the cylinder screws in a diagonal pattern and tighten them by gradually
increasing the torque to screws on the opposite sides. Adjust to a maximum
torque of 5 Nm for columns with 70 and 100 mm ID respectively 10 Nm for
columns with 140 and 200 mm ID.
Page 40
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
40 / 49
PLUS
low pressure chromatography column
ATTENTION
7.4.3 Frit replacement column body
Faulty assembly
Damage to the glass body is possible due to faulty assembly.
► Take care during assembly of the glass body that the maximum torque of 3 Nm is
not exceeded.
► Assemble the cylinder screws in a diagonal pattern and always tighten on the
opposite sites.
► Take care during assembly of the glass body that affected parts are correctly
aligned.
Required devices:
• Cylinder key/torque wrench
The frit in the column body is replaced as follows:
1. Remove the adjustable end piece Chapter 6.1, S. 18.
2. Dismantle the glass body Chapter 7.4.2, S. 37.
3. Remove the union nut of the connector.
4. Press the fixed end piece with the connector out of the closing disc (CB).
Page 41
YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
41 / 49
PLUS
low pressure chromatography column
5. Remove the ring gasket A.
6. Remove the frit.
7. Insert the frit. The distribution profile faces the plunger.
8. Reattach the ring gasket A and press firmly in position.
9. Replace the fixed end piece into the closing disc (CB).
10. Fix the union nut at the connector.
11. Assemble the glass body Chapter 7.4.2, S. 37.
Page 42
YMC Europe GmbH
Troubleshooting
YMC Pilot
42 / 49
PLUS
low pressure chromatography column
WARNING
WARNING
Fault description
Possible cause
Solution
Dead volume occurred
Exceeding the maximum
allowable pressure of the
column packing
► If dead volume occurs, proceed as
follows:
1. Stop the pump.
3. Readjust the piston of the
adjustable end piece.
Stressed column packing
Peak shape of the eluted
substances deteriorates
Dead volume at the column
inlet
See fault description “Dead volume
occurred”
Partially clogged inlet frit
Frit replacement Chapter 7.4.1, S. 36
Partially clogged outlet frit
Frit replacement Chapter 7.4.3, S. 40
Separation performance of
the stationary phase altered
by impurities
Wash the stationary phase. Contact
also the manufacturer of the packing
material Chapter 7.1, S. 31
Stationary phase
mechanically destroyed
Replace packing material
Air in the column
Degassing or evaporation of
solvent during storage
New column conditioning. Possibly
new packing material is necessary.
Exceptional pressure rise
during operation
Valve circuit incorrect
Check valve positions
Clogged frit
Frit replacement Chapter 7.4, S. 36
Over-tightened fixing
screws
Replace fixing screws and clamping
tubing, cut the end of the tubing
Pressure drop during
operation
Leakage of tubing or fitting
between pump and column
Check tubing and connections
Depleted solvent reservoir
Refill solvents
8 Troubleshooting
Incorrect maintenance
Severe injuries during operation after incorrect maintenance possible if the column or
column parts do not withstand the operating pressure.
► Only authorised and qualified personnel should carry out the maintenance.
Incorrect replacement of spare parts
Severe injuries due to breakage possible, if not authorised spare parts are used.
► Use only the manufacturer authorised spare parts.
Page 43
YMC Europe GmbH
Troubleshooting
YMC Pilot
43 / 49
PLUS
low pressure chromatography column
Fault description
Possible cause
Solution
Solvent drips from the
column
Column body leaks
► If the column body leaks, proceed
as follows:
1. Visual check, if solvent leaks
between glass body and flat
gasket.
2. Check the contact surface of the
glass body and the sealing
surface for impurities.
3. Check the torque and the
concentric arrangement of the
parts.
Tubing connection leaks
► If the tubing connection leaks,
proceed as follows:
1. Tighten the fixing screw.
2. Check tightness of tubing
connection.
3. New column conditioning.
If faults occur which are not listed here, please contact YMC for support.
Page 44
YMC Europe GmbH
Storage, packaging and transport
YMC Pilot
44 / 49
PLUS
low pressure chromatography column
WARNING
For storage of packed columns, the chemical resistance of the wetted
parts and of the packed stationary phase (please contact the
manufacturer) should be considered.
9 Storage, packaging and transport
Solvents in the pilot column
Dangers possible due to liquids leaking from the column during transportation.
► Unpack and clean the column before transport.
9.1 Storage
The pilot column should be stored as follows:
• unpacked
• cleaned
• dry
• empty
9.2 Transport
We recommend the use the original wooden box for transportation. If you are no
longer in possession of the box, please pack the column in a similar container.
The column is packaged as follows:
1. Unpack the pilot column Chapter 6.4, S. 28.
2. Clean the pilot column Chapter 7.3, S. 33.
3. Insert the adjustable end piece.
4. Wrap the glass body in a bubble wrap.
5. If existing, fix the wheels to the base of the wooden box.
6. Make sure that there is a minimum distance of 100 mm from the outer wall.
7. Fix the pilot column in the wooden box with bubble wrap.
8. Attach appropriate warning signs to the box (e.g. “Caution breakable glass”
and “Top”).
Page 45
YMC Europe GmbH
Disposal
YMC Pilot
45 / 49
PLUS
low pressure chromatography column
10 Disposal
For the disposal of the pilot column it has to be disassembled and disposed according
to the national and other applicable environmental regulations.