Yaohua Weighing System XK3190-A12, XK3190-A12E User Manual

Digital Weighing Indicator
XK3190-A12(E)
User Manual
2008.10 OC Version
XK3190A12(E)
Content
CHAPTER 1 MAIN SPECIFICATIONS ………………….…….. …………..……….……....................-2-
1.1 TECHNICAL PARAMETERS OF INDICATOR
1.2 INTRODUCTION OF INDICATOR
CHAPTER 2 INSTALLATION ............... ............. ............. ............. ............................................. -5-
2.1 FRONT AND BACK VIEW OF INDICATOR
2.2 KEY FUNCTIONS
2.3 CONNECTION OF THE INDICATOR
CHAPTER 3 OPERATION ……………. ……………………..…………………….....................……-8-
3.1 POWER ON AND AUTO ZERO
3.2 SEMI-AUTOMATIC ZERO-SETTING
3.3 TARE
3.4 COUNTING
3.5 ACCUMULATING
3.6 USER FUNCTION SETTING
3.7 SERIAL COMMUNICATION
CHAPTER 4 CALIBRATION …….……… ……………………...…..…………….... ...............................-12-
CHAPTER 5 MAINTENANCE ….. ………………...…………...………………….............. ……............ .-14-
CHAPTER 6 ERROR INDICATION ……………………….……………………… ……………………-15-
XK3190A12(E)
Indicator
Model name
XK3190 - A12 XK3190 - A12E
Indicator type
Self-indicating instrument Non-automatic weighing indicator Software-embedded indicator
Application
Platform scales
●Commercial weighing and industrial weighing
●Instrument for not direct sales to the public
Metrological parameters
Accuracy Class
Class III
Value of the fractional error (pi)
0.5
Maximum number of verification scale intervals(n
ind
)
3000 Operating temperature range
0~40
Initial zero-setting range
±10%Max
Semi-automatic zero-setting range
±2% Max
Zero-tracking range
±2% Max
Subtractive tare T-
Max
Scale divisions
1/2/5/10/20/50 optional
Relative humidity
≤85%RH
Storage and transportation temperature
-2555
Power supply
AC 230V / 50Hz Battery DC 6V/2.8Ah
Load cell excitation voltage (Uexc)
DC 5V
Minimum signal voltage for dead load
0mV
DEAR CUSTOMERS, PLEASE READ THE USER GUIDE CAREFULLY BEFORE USING THE INDICATOR!
Chapter 1 Main Specifications
1.1 TECHNICAL PARAMETERS OF INDICATOR
XK3190A12(E)
Electrical parameters
Maximum signal voltage for dead load
14mV
Minimum input-voltage per verification scale interval (ΔUmin)
2µV
Measuring range minimum voltage (U
MRmin
)
6mV
Measuring range maximum voltage (U
MRmax
)
20mV Minimum load cell impedance (R
Lmin
)
87Ω
Maximum load cell impedance (R
Lmax
)
350Ω
Sense system
Six wires with compensation
A/D sample rate
10 times/s
Display
A12: 6 bits LCD7 state indicating signals A12E: 6 bits LED8 state indicating lightsLED lights
In the following chapter, indicating signal and indicating light will not be distinguished, both will be called indicating signal to indicate the position
Structure Housing
Anti flaming plastic housing
Specifications
248mm*140mm*156mm
Weight
2.2kg
Load cell interface
Six wires, connect 1-4 pieces 350 Ω load cells
Six-core shielded cable MaterialCopper
Max length100m/0.5mm
2
When the cable is longer than 3m, its strongly advised to add a magnetic loop (optional component) on the RS232 interface of the indicator to enhance the anti-interference ability
XK3190A12(E)
Interfaces
Serial communication interfaces RS232C:
Unidirectional communication. Only the indicator can transfer the data to computer, the
computer can not control the indicator.
Baud rate: 1200/2400/4800/9600 optional. Data for RS232 communication is in ASCII
code as below: The 1st bit is starting bit, the 10th bit is stopping bit, and the middle 8 bits are data bits. Even check.
Transmission distance<10m
Cable specifications: 2-core shielded cable
MaterialCopper Max length10m
When the cable is longer than 3m, its strongly advised to add a magnetic loop (optional component) on the RS232 interface of the indicator to enhance the anti-interference ability
Software: Third party softwares are optional, such as: COMPort, Debuger…etc
Software environment: Windows 2000Windows XP
1.2 INTRODUCTION OF INDICATOR
XK3190-A12(E) weighing indicator is non-automatic weighing indicator applied to platform scale. It can be connected to electric platform scale, electric floor scale, and can be connected 1-4 350 ohm load cells. It is mainly applied to enterprise with factory, wholesale market, bus station, port, mine, warehouse etc, all kinds of weighing occasions.
The indicator is composed of anti flaming ABS plastic housing, PCB, LED/LCD indicator, keypad, power transformer and AC power plug, fuse, load cell socket, RS232 interface, 6V/4AH rechargeable battery. A/D
conversion adopts MCU with excellent stability and anti interference ability.
A12(LCD) has three types of backlight user mode: still closed, still open and automatic. Under automatic mode, backlight will be off automatically after the indicating weight is stable for few seconds. When the weight changes, the backlight will be on automatically. It has automatic power off function.
XK3190A12(E)
Indicator will turn off automatically to save electricity consumption after no operations for a while. When the voltage of built in battery decreases to a certain value, indicator will alert and remind user to recharge in time.
A12(LED) has electricity saving function for the indication. After the weight is stable for a few seconds, it will indicate a record segment; When there is weight changes, indicator will turn on automatically and display the weight. When the voltage of built in battery decreases to a certain value, indicator will alert and remind user to recharge in time.
XK3190-A12(E) indicator has the calibration protection and software tamper-proof function. There is calibration switch on the hardware and is protected with lead seal. Only when you open the lead seal and softly touch the calibration switch, the data will be saved. Otherwise, after the restart-up, the calibration data will recovery to previously saved one. There is software version, software verifying code, calibration counter and parameter verifying code for legal metrology. This is for relevant metrology bureau to test and control.
CHAPTER 2 INSTALLATION
2.1 FRONT AND BACK VIEW OF THE INDICATOR
2.2 KEY FUNCTIONS
XK3190A12(E)
2-1 Housing
Key Function
#
Func
A12:【Accum A12E:【 *
Tare
Zero
ON/OFF
When turning on the indicator, keep pressing this key to enter the calibration mode When calibration, press this key to switch from parameters At weighing mode, press this key to start counting; and keep pressing this key for more than 5 seconds to enter the user parameters setting mode At weighing mode, press this key to accumulate the weight At counting mode, press this key for sample taking At user parameters setting mode, press this key to switch from parameters At weighing mode, press this key to tare At user parameters setting mode, press this key to change the parameters value At weighing mode, press this key to zero When calibration, press this key to change the parameters value Power on/off
Note: A12E has no such button, on and off controlled by switch at the back
XK3190A12(E)
2.3 CONNECTION OF THE INDICATOR
2.3.1 The definition of the connectors on the PCB
(2-2) A12 main board connecting figure
(2-3) A12E main board connecting figure
2.3.2 Load cell connection
indicator load cell
EX- S- SH EX+ Excitation
S+
EX+ S+ IN- IN+ IN-
S-
EX-IN+
SH shield
Excitation
Load cell output signal load cell output signal
XK3190A12(E)
▲. Indicator must be reliably connected to load cell, and the shielded-cable of load cell
must be reliably connected to underground . To protect the indicator and load cell, we cannot plug or withdraw the connector when the indicator is power on. . Both the load cell and indicator are static sensitive devices. You must adopt anti-static measures. The electric welding operation and other strong electric operation on the scale platform are strictly prohibited. In order to protect the operator, indicator, and relevant devices, you should install lightning rod in the thunderstorm frequently happen area
(2-4) Load cell connection
Please use 6-core shielded cable to connect the load cell. If uses 4-core cable, please short “EX+ and “S+“EX- and “S-. If the cable is longer
than 3m, its strongly advised to add a magnetic loop (optional component) on the RS232 interface of the indicator to enhance the anti-interference ability.
2.3.3 RS232 connection
Please use 2-core shielded cable. If the transmission distance is longer than 3 meter, its strongly advised to add a magnetic loop (optional component) on the RS232 interface of the indicator to ensure the reliability of the data transmission.
2-5 RS232 connection
XK3190A12(E)
Please warm the indicator up for 20 minutes before using to ensure the best performance
Chapter 3 Operation
3.1 POWER ON AND AUTO ZERO
3.1.1 After turning on the switch on the back housing, keep pressing ON/OFF】key to turn on the indicator.
3.1.2 The indicator displays all sections, and all the indication signs first, then the display is as follow (1)Model name: 【-A12-】
(2) Version of the software: such as【V 1.01】 (3) Sections check: 111111-999999
If press [Func] before the indicator displaying the version of software, the indicator will display as below (1)Model name: 【-A12-】 (2) Version of the software: such asV 1.01 (3)Calibration times of the indicator: such asn 10 (4)Verification code of metrology parameters: such asC 1A5D (5)Verification code of the softwaresuch asF 2B6C (6) Sections check: 111111-999999
3.1.3 When power on, if the scale deviates from the zero point, but its still within auto zero range (±10%Max), the indicator will zero automatically; if it is out of auto zero range, the indicator will displayErr 3as warning, then show the weight.
3.2 SEMI-AUTOMATIC ZERO-SETTING
3.2.1 At weighing mode, if there is any tolerance when unloaded, and the tolerance is within the semi-automatic
zero-setting range, press [Zero] to zero the scale. If the tolerance is beyond the semi-automatic zero-setting range, the [Zero] key is invalid(Err 7】), and you need to recalibrate the scale.
3.2.2 Only when the stable indication sign is on, zero operation is valid.
3.3 TARE
At weighing mode, when displaying stable positive weight, and the weight is no more than F.S. , press
the[Tare] key to deduct the weight as tare. The indicator displays net weight as “0”, and the Net indication sign is on. When the displaying weight is negative or zero, pressing [Tare] key can clear the saved tare weight, and get back to gross weight displaying mode. The Net indication sign will be off.
10
3.4 COUNTING
At weighing mode, press [Func] to enter the counting mode. Indicator displays “count”, and we can load
XK3190A12(E)
the scale. When stable indication sign is on, press [ move the small indication triangle directedly, the number corresponded with the small triangle means the sample quantity, and this number will increase by one when press [Zero] key. After inputting the sample quantity, press [
At counting mode, press [Func] key to return to weighing mode.
] key to start the counting. The counting indication sign will be on.
*
] key, indicator displays “C00000”, then press [Tare] to
*
Note:
When “count” is displayed after we pressing [Func] key at the weighing mode, press [
counting mode directly. Indicator will count according to last time sampling result.
In this process, if theErr 4appears, it means sampling failed, the indicator will keep the result of last
sampling.
] twice to enter
*
3.5 ACCUMULATING
At weighing mode, when the displaying value is positive and stable, press [ present weight. Press [ be performed after the weight return to zero. When the scale displaying zero, press [ accumulated weight.
At accumulating mode, press [Func] key to clear the accumulated weight in Memory.
] key again, it gets back to weighing mode. The next accumulating operation must
*
] key to accumulate the
*
] to display the
*
3.6 USER FUNCTIONS SETTING
At weighing mode, keeping pressing [Func] for more than 5 seconds, it enters user functions setting mode (mode P). There are 9 parameters. Press [ the value. After setting the parameters, press [Func] to save.
] to change the parameter, and press [Tare] to change
*
Note: The parameters P6, P7, P8, and P9 are metrological related. We need to press the calibration switch
on the back housing to save the settings.
Parameters are as below: 1P1 x Automatically power off (Default setting:1)
X=1: Close this function X=2: Auto power off 10 minutes after last operation
X=3 Auto power off 20 minutes after last operation X=4 Auto power off 30 minutes after last operation
2P2 x Baud rate setting (Default setting:4) X=1: 9600
X=2: 4800 X=3: 2400 X=4: 1200
3P3 x RS232 output content (Default setting:1)
XK3190A12(E)
11
X=1: Net weight output X=2: Gross weight output
4P4 x RS232 output mode (Default setting:1) X=1: No transmission (RS232 stop)
X=2: Continuous transmission X=3: Continuous transmission only when stable
5P5 x Backlight mode (Default setting:2) X=1: No backlight
X=2: Automatic backlight X=3: Constant backlight
6P6 x Zero-tracking range (Default setting:1)
X=1: 0.5e
X=2: No zero tracking
7P7 x Digital filtering intensity (Default setting:2)
X=1: High
X=2: Medium X=3: Low
8P8 x Stabling speed (Default setting:2) X=1: Fast
X=2: Medium X=3: Slow
9P9 x Stable range (Default setting:2) X=1: Wide
X=2: Medium X=3: Narrow
3.7 SERIAL COMMUNICATION (RS232)
Make sure that communication interfaces are correctly connected. If there is anything wrong with the
connection, damage may happen to output port of the indicator, input port of the computer, and even other peripherals may get involved.
Necessary computer technology and programming skills are required for computer communication,
which should be participated and instructed by professionals. Non-professional staff is supposed not to be involved in this regard.
Detailed connection instruction kindly check the 2.3.3 on page 8
Data for RS232 communication is in ASCII code. The format is as below: The 1st bit is starting bit, the 10th bit is stopping bit, and the middle 8 bits are data bits. Even check.
XK3190A12(E)
12
Communication mode as follows:
(1)At normal weighing mode, every group of data consists of 15 bytes as below: 1st byte: W 2nd byte: G (when the weighing result is gross weight) „N‟ (when the weighing result is net weight) 3rd -9th byte: Weighing result including the decimal point When there is no decimal point, the 9th bit is blank (0x20)
When the weight is negative, the 3rd bit is the negative sign (0x2d)
10th and 11th byte: Unit k g (0x6B, 0x67) 12th byte: the 4 high bits of checksum 13th byte: the 4 low bits of checksum 14th byte: 0x0d 15th byte: 0x0a
Note:
The checksum is from 1st byte to 11th byte If 4 high bits or 4 low bits of checksum is9, add 30h and transmit in ASCII code. For example, if the 4 high bits checksum is 6, then we add 30h, and transmit 36h in ASCII code 6.
If the checksum>9, then add 37h, and transmit in ASCII code. For example, if the 4 high bits checksum is B, then we add 37h, and transmit 42h in ASCII code B.
For example, if the weight is 4.139 Kg, the transmission format is as follow: ASCII: W G 0 0 4 . 1 3 9 k g 3 D Hex: 57 47 30 30 34 2E 31 33 39 6B 67 33 44 0D 0A
(2)When the scale is overloaded (G.W>Max+9e), the indicator will send OL23 in 15 bytes as follow: 1st to 9th byte: Blank (0x20) 10th byte: O (0x4f) 11th byte: „L (0x4c) 12th byte: the 4 high bits of checksum (0x32) 13th byte: the 4 low bits of checksum (0x33) 14th byte: 0x0d 15th byte: 0x0a
(3)When the weight is less than -20d, the indictor will send L023 in 15 bytes as follow: 1st to 9th byte: Blank (0x20) 10th byte: L (0x4c) 11th byte: „O (0x4f) 12th byte: the 4 high bits of checksum (0x32)
XK3190A12(E)
13
STEP
OPERATION
DISPLAY
NOTES
1
Press [TARE] to change the division, and press [#] to confirm
[d X ]
Division setting. For example: Press [#] when displayed【d 5】,
then the division is set to be5, and the indicator enters decimal point setting. Note: The 10, 20, 50 divisions are only valid when there is no decimal point. When there is a decimal point, the 10, 20, 50 divisions will be turned to 1, 2, and 5 automatically.
2
Press [TARE] to change the decimal point, and press[#] to confirm
[P X ]
Decimal point setting For example: Press [#] when displayed【P 0.000】, then the decimal point is set to be 0.000, and the indicator enters full capacity setting Note: When there is a decimal point, division 10, 20, 50 are invalid, and will be turned to 1, 2, and 5 automatically.
3
Press [TARE] to select the digit bit; Press [ZERO] to change the value;
Press [#] for confirm the
input of full capacity
[FULL ]
Full capacity setting For example: Press [#] when displayed【025000】, then the full capacity is set to be 25000, and the indicator enters zero point calibration
4
Make sure there is no load on scale, and press [#] when the stable indication sign is on
[nOLOAD]
Zero point calibration
13th byte: the 4 low bits of checksum (0x33) 14th byte: 0x0d 15th byte: 0x0a
Chapter 4 Calibration
4.1 CALIBRATION
Connect load cell properly, then turn on the indicator, keep pressing [#] key while it‟s initializing, it will enter
the calibration mode, and display d X .
XK3190A12(E)
14
5
Add full capacity load, press [TARE] to continue
[AdLOAD]
Calibrate the full capacity For example: Load 25000 weight on scale(as we set in step 3)
Use [TARE] and [ZERO] to change the value to be 25000.
When stable indication sign is on, press [#] to confirm.
Press [TARE] to select the digit bit; Press [ZERO] to change the value accordingly with the full capacity; Press [#] to confirm when the stable indication sign is on
[025000]
6
Press the calibration switch at the back housing of the indicator.
[ End]
It saves the calibration parameter and back to the weighing mode. Note: if no pressing the calibration switch, all the parameters wont be saved.
4.2 FAST CALIBRATION FOR ZERO POINT AND FULL CAPACITY
Connect load cell properly, then turn on the indicator, keep pressing [ # ] key while its initialization, it will enter into the calibration mode, and display d X .
4.2.1 Fast calibration of zero point: At any time before it showing [nOLOAD], press [FUNC]. Indicator will keep the original division, decimal point, full capacity, and enter the zero point calibration directly. Making sure there is no load on the scale, and the stable indication sign is on, press [ZERO] to re-calibrate the zero point. The indicator will display [End]. Press the calibration switch at the back of the indicator to save the setting and get back to the weighing status.
4.2.2 Fast calibration of full capacity: At any time before it showing [AdLOAD], press [*]. It keeps the original division, decimal point, full capacity, zero point, and enters into the full capacity calibration directly. After the full capacity is reset, press the calibration switch at the back of the indicator to save the setting and get back to the weighing status.
5.1 To ensure the clarity and service life of the indicator, it must be kept away from direct sunlight during using, and the ground where the indicator stands must be smooth.
5.2 It is improper to use this indicator in a dustful or vibrant or damp environment.
Chapter 5 Maintenance
XK3190A12(E)
15
5.3 The load cell and indicator need good connection. System must have a good ground connection, and kept away from strong electric field, strong magnetic field. The load cell and indicator must be kept away from strong corrosive substances and inflammable& explosive materials. Do not use it where inflammable gases or steams exist. Don‟t use it for canning system of compressive container. ! In the area where lightning and thunder happen frequently, reliable lightning arrester should be installed to ensure the personal safety and to prevent any damage to the indicator and relative equipment caused by lightning stroke. !The load cell and indicator are both static sensitive equipments, so anti-static measures must be taken during the use. It is strictly invalid to carry out welding operation or other operations with high current on the weighing platform. In the stormy season, lightening prevention measures must be taken reliably to prevent any damage to load cell and indicator caused by lightening stroke, and to guarantee the personal security of operators and safe running of weighing devices and relative equipments.
5.4 Strong solvents such as benzyl and nitro oils are forbidden for cleaning the housing
5.5 Don‟t inject any liquid or other conductive particles so as to avoid any damage of indicator and electric shock
5.6 Before plugging in or out of the connecting line between indicator and external equipment, the power of both indicator and equipment should be cut off
5.7 Advice of the company: our company is responsible for the indicator quality, but not responsible for the problems of the system where the indicator locates. Your attention is required when making purchase.
5.8 Please use the indicator outward interfaces strictly as per the operating instruction manual. Do not change the connection at random. If failure occurs in the using process, draw the plug immediately, and send it for professional factory for reparation. Non professional balance manufacturers are not supposed to do the repairing to avoid any worse damage. It is not allowed to open the indicator at will, or else, repairing will be refused.
5.9 If non artificial defects and failures happen after normal use within one year after the sale date, the users can mail the product and guarantee repair card ( with correct code) to the appointed reparation station or supplier. The manufacturer guarantees the life-time maintenance for the indicator.
Chapter 6 Error Indication
XK3190A12(E)
16
ERR 1
The AD value is too small when calibrating the full capacity Please change proper capacity load cell
ERR 2
The zero point is out of range when calibrating the zero point Please make sure no load is on the scale
ERR 3
The zero point is out of range when turning on Please make sure no load is on the scale when turning on
ERR 4
The input sample quantity is zero when sampling in counting mode Please input the right sample quantity
ERR 5
When full capacity calibrating, the weight input is zero Please input the weight in accordance with the load on scale
ERR 6
The unit weight is less than 0.25e when sampling in counting mode Please re-input the sample quantity
ERR 7
The weight is beyond the semi-automatic zero-setting range, the [Zero] key is invalid.
ERR 8
Overflow of the accumulated weight Please clear the accumulated weight
bAt-lo
Low power
Lo
The G.W. is less than -20e
oL
Overload, or load cell malfunction Please recalibrate the F.S., and check the connection of load cell