Yamato WG270 Instruction Manual

Page 1
Water Purifier
SUPER AUTO STILL
Model
WG270
Instruction Manual
- First Edition -
Yamato Scientific America Inc
Santa ClaraCA
Printed on recycled paper
Thank you for purchasing "Auto Still, WG 270
of Yamato Scientific Co., Ltd.
To use this unit properly, read this "Instruction
Manual" thoroughly before using this unit. Keep this instruction manual around this unit for referring at anytime.
WARNING!: Carefully read and thoroughly understand the important warning items described in this manual before using this unit.
Page 2
Page 3
Contentes
1. Cautions in Using with Safety............................................................................................................
Explanation ........................................................................................................................................
Table of Illustrated Symbols ..............................................................................................................
Fundamental Matters of "WARNING!" and "CAUTION!" ...................................................................
2. Before Using this unit .........................................................................................................................
Requirements for Installation .............................................................................................................
3. Description and Function of Each Part......................................................................................... 15
Main Unit ........................................................................................................................................ 15
Water Sampling Table ................................................................................................................... 16
Piping System View ....................................................................................................................... 17
Principle of Operation .................................................................................................................... 18
Control Panel ................................................................................................................................. 22
4. Operation Method ............................................................................................................................ 23
Setup and Check before Use ......................................................................................................... 23
Operation Procedure ...................................................................................................................... 24
Sampling of Pure Water ................................................................................................................. 25
Collection of Measured Amount of Pure Water ............................................................................. 26
Collection of Measured Amount of Distilled Water ........................................................................ 27
Display of Water Quality ................................................................................................................ 29
Setting/Display of Submenu .......................................................................................................... 32
5. Handling Precautions ..................................................................................................................... 35
6. Maintenance Method ....................................................................................................................... 36
Inspection and Maintenance .......................................................................................................... 36
Replacement of pre-treatment cartridge ........................................................................................ 36
Replacement of ion exchange resin cartridge ............................................................................... 37
Washing of Distiller ........................................................................................................................ 38
Replacement of Heater .................................................................................................................. 42
Washing of Water Supply Hose Filter ............................................................................................ 43
Replacement of Hose .................................................................................................................... 43
Distilled water tank drain method ................................................................................................... 43
Replacing the TOC reduction UV lamp ......................................................................................... 44
Checking if the TOC reduction UV lamp goes ON ........................................................................ 45
Replacing the deaeration membrane module................................................................................ 46
Vacuum pump ................................................................................................................................ 47
7. Long storage and disposal ............................................................................................................ 49
Considerations on disposing .......................................................................................................... 50
8. In the Event of Failure… ................................................................................................................. 52
Failure indication and Its Contents ................................................................................................ 52
Remedy for Trouble ....................................................................................................................... 56
Trouble Shooting ............................................................................................................................ 57
Measures in emergency................................................................................................................. 57
9. After-sales Service .......................................................................................................................... 58
Page 4
10. Specifications ................................................................................................................................ 59
11. Wiring Diagram .............................................................................................................................. 60
WG270 ........................................................................................................................................... 60
12. Replacement Parts List ................................................................................................................ 62
13. Consumable parts ......................................................................................................................... 64
14. List of Dangerous Substances .................................................................................................... 65
Page 5
1. Cautions in Using with Safety
Explanation
MEANING OF ILLUSTRATED SYMBOLS
Various symbols are used in this safety manual in order to use the unit without danger of injury and damage of the unit. A list of problems caused by ignoring the warnings and improper handling is divided as shown below.Be sure that you understand the warnings and cautions in this manual before operating the unit.
WARNING!
If the warning is ignored, there is the danger of a problem that may cause a serious accident or even f atality.
CAUTION!
If the caution is ignored, there is the danger of a problem that may cause injury/damage to property or the unit itself.
Meaning of Symbols
This symbol indicates items that urge the warning (including the caution). A detailed warning message is shown adjacent to the sy mbol.
This symbol indicates items that are strictly prohibited. A detailed message is shown adjacent to the sy mbol with specific actions not to perform.
This symbol indicates items that should be always performed. A detailed message with instructions is shown adjacent to the symbol.
Illustrated Symbols
Page 6
1. Cautions in Using with Safety
Table of Illustrated Symbols
Warning
Warning, generally
Warning,
high voltage
Warning,
high
temperature
Warning,
drive train
Warning,
explosive
Caution
Caution,
generally
Caution,
electrical shock
Caution,
scald
Caution,
no road heating
Caution,
not to drench
Caution,
water only
Caution,
deadly poison
Prohibit
Prohibit,
generally
Prohibit,
inflammable
Prohibit,
to disassemble
Prohibit, to touch
Compulsion
Compulsion,
generally
Compulsion,
connect to the
grounding
terminal
Compulsion,
install on a flat
surface
Compulsion,
disconnect the
power plug
Compulsion,
periodical
inspection
Page 7
1. Cautions in Using with Safety
Fundamental Matters of "WARNING!" and "CAUTION!"
WARNING!
Do not use this unit in an area where there is flammable or explosive gas
Never use this unit in an area where there is flammable or explosive gas. This unit is not explosion-proof. An arc may be generated when the power switch is turned on or off, and fire/explosion may result. (Refer to Page 67"List of Dengerous Substance)
Be sure to connect grounding wire.
Connect to grounded plug socket. If no grounded plug socket is available, be sure to connect grounding lead by use of ground adapter attached in nonstandard. Failure to do so could cause electric shock or fire.
If a problem occurs
If smoke or strange odor should come out of this unit for some reason, turn off the power key right away, and then turn off the circuit breaker and the main power. Immediately contact a service technician for inspection. If this procedure is not followed, fire or electrical shock may result. Never perform repair work yourself, since it is dangerous and not recommended.
Do not use the power cord if it is bundled or tangled
Do not use the power cord if it is bundled or tangled. If it is used in this manner, it can overheat and fire may be caused.
Do not process, bend, wring, or stretch the power cord forcibly
Do not process, bend, wring, or stretch the power cord forcibly. Fire or electrical shock may result.
Do not disassemble or modify this unit
Do not disassemble or modify this unit. Fire or electrical shock or failure may be caused.
Do not touch hot portion
Boiler may be hot in some portion in operation or immediately after operation. Be aware of burns. When performing maintenance of heater etc., ensure that the boiler is cooled down beforehand.
Close the tap when unit is out of service
When unit is out of service (at night or on holiday), be sure to close the tap so as to avoid water leakage accident.
CAUTION!
During a thunder storm
During a thunderstorm, turn off the power key immediately, then turn off the circuit breaker and the main power. If this procedure is not followed, fire or electrical shock may be caused.
Exercise care in handling washing liquid (Orgazor)
Principal component of washing liquid (Orgazor) is sulfamic acid, which is acidic almost equal to water solution PH:1. Use protective tool (gloves, mask, and glasses) in handling. When it is touched by human body, immediately wash it away with clean water.
Page 8
2. Before Using this unit
Requirements for Installation
WARNING!
1. Be sure to connect grounding wire.
Always connect the equipment to the ground. If the equipment is not grounded, ground
leakage, if any, may not activate the ground leakage breaker, resulting in extremely-hazardous electric shock.
Connect the grounding wire either to the grounding wire or grounding terminal block of the
electric power equipment. If the grounding equipment is not available, consult the nerest electric constructor and carry out grounding according to Article 18 (Type III Grounding Work of 100Ω or less) of the Technical Standards for Electric Equipment.
Never connect the grounding wire to gas pipe or water supply pipe.
2. Choose a proper place for installation
Do not install this unit in a place where:
Rough or dirty surface. Flammable gas or corrosive gas is generated. Ambient temperature 35C and above or 5C and below. Ambient temperature fluctuates violently. There is direct sunlight. There is excessive humidity and dust. There is a constant vibration. Not horizontal surface.
The power source is instable.
Keep space around each product above the range shown below. Install units within sink equipment if possible.
10 ㎝ or more
10 ㎝ or more
10 ㎝ or more
Page 9
2. Before Using this unit
Requirements for Installation
WARNING!
3. Do not use this unit in an area where there is flammable or explosive gas
Never use this unit in an area where there is flammable or explosive gas. This unit is not explosion-proof. An arc may be generated when the power switch is turned ON or OFF, and fire/explosion may result.
To know about flammable or explosive gas, refer to Page 67 "List of Dangerous Substances".
Explosive gas
Flammable gas
4. Do not modify
Never disassemble this unit. This unit has high voltage inside in some portion, which may cause electric shock. Contact
a local dealer or Yamato sales office for adjusting or repairing inside.
In routine maintenance and inspection, follow the procedure described in the instruction
manual. Do avoid modification by customer because it may lead to trouble.
Modification
Page 10
2. Before Using this unit
Requirements for Installation
WARNING!
5. Installation on horizontal surface
Set this unit to the flattest place. Setting this unit on rough or slope place could cause the
unexpectible trouble or malfunction. The unit WG270 weighs 55kg gross. Two or more persons are required for carrying or setting these units.
6. Choose a correct power distribution board or receptacle
Use a plug socket conforming to electric capacity (capacity 15A or greater). When power capacity is insufficient, sampling of distilled water goes short, and normal
control is disabled by fall of power voltage. Connect to power equipment having sufficient power capacity.
Electric capacity WG270 AC220V Single phase 7A
7. Connection of power cord
Always ensure that breaker on power unit side is "Off" before connecting power cord.
Power plug of WG270 uses 3-core cord including grounding wire, and the plug is grounded type. If your plug socket is not compatible (2P), use a ground adapter attached in nonstandard. In using ground adapter, be sure to ground a grounding lead.
8. Handling of power code
Do not entangle the power cord. This will cause overheating and possibly a fire. Do not bend or twist the power cord, or apply excessive tension to it. This may cause a fire
and electrical shock.
Do not lay the power cord under a desk or chair, and do not allow it to be pinched in order to
prevent it from being damaged and to avoid a fire or electrical shock.
Keep the power cord away from any heating equipment such as a room heater. The cord's
insulation may melt and cause a fire or electrical shock.
If the power cord becomes damaged (wiring exposed, breakage, etc.), immediately turn off
the power at the rear of this unit and shut off the main supply power. Then contact your nearest dealer for replacement of the power cord. Leaving it may cause a fire or electrical shock.
Connect the power plug to the receptacle which is supplied appropriate power and voltage.
Page 11
2. Before Using this unit
Requirements for Installation
WARNING!
9. Observe the specified pressure range of raw water from waterworks
Apply the range of city water pressure between 0.5 - 5 X 100kPa (0.5 - 5kgf/cm2) including
nighttime.
Range of raw water pressure is the same when "Water Supply Port Unit" (optional
accessory) is used.
10. Connect the water supply hose securely
Take the supply hose (with connection port) out of attachments Install the unit on a
horizontal and stable place nearby tap and sink.
If not connecting the water supply hose securely, the water supply hose or connection port
may be disconnected, resulting in water leakage such as water bursting forth.
11. Connect the tap side
1. Slide the sleeve of socket on connection port side in the arrow direction, then
connection port body and water supply hose can be separated. There separate the two parts.
2. Once loosen the plug from the ring .
3. Tighten the 4 mounting screws uniformly while pressing the ring ① slightly and
uniformly to make the packing in flat contact with water tap. If the tap is a chemical tap, adjust the position so that the mounting screw is located at the bottom valley of tap nipple as shown.
4. Turn the plug clockwise to tighten securely. This will allow the tap and connection port
to be sealed by packing .
5. Insert the socket securely to the plug with the sleeve slid in the arrow direction.
The sleeve returns to the original position when released, and then connection is completed.
* In case that the socket is removed, the valve attached on the connection side of the main
unit stops the flow of water.
Tap
Connection port body
Socket on connection port side
Water supply hose
Water supply port
Sleeve on water
supply side
Tap
Sleeve
Water supply
hose
Connection Port Body
Ring Plug Socket Spacer Mounting Screw Packing Filter
One-touch Joint
Page 12
2. Before Using this unit
Requirements for Installation
WARNING!
12. Connection on body side
1. Remove the rubber cap from the
water supply port plug .
2. Insert the socket securely to
the plug on body side with the sleeve slid in the arrow direction. The sleeve returns to the original position when released, and then connection is completed. The socket contains a valve inside, which opens only when the socket is connected by plug; otherwise, water is not fed because this valve does not open.
Body side
13. Connect the water supply hose to the tap provided with sink equipment
If the water supply hose is
connected to a tap without sink equipment, flood damage may be caused when water supply hose is disconnected or damaged; therefore be sure to connect to a tap having sink equipment.
14. When the sink equipment is remote from water tap, use "Water Supply Port Unit" (optional accessory)
"Water Supply Port Unit" is designed to loosen the connection to the tap harder than the set
of standard water supply hoses when water pressure fluctuates.
15. In case that there is no tap, use appropriate joint shown in the optional "Coupler joints for feed water".
Tap
Water supply hose
Socket on connection port side
Connection port Body
Sink
Sleeve
Page 13
2. Before Using this unit
Requirements for Installation
WARNING!
16. Connect the drain hose securely
If the drain hose is not connected securely, it may be disconnected, leaking water in the unit,
or leading to trouble of system.
1. Pick up the drain hose (with elbow) and the hose clamp from the main unit accessories.
2. Always make sure that the earth leakage breaker of the unit is "OFF".
3. Remove the rubber stopper from the outlet of the drain port hose on the main unit.
4. Push the hose clamp into the elbow of the drain hose, and next, put the elbow into the drain outlet hose of the main unit, then tighten the hose clamp securely.
5. Pull the hose out from the drain hose-end outlet at right/left or backside of the main unit.
Hose clamp
Drain port hose on the main unit
Elbow
Drain hose (accessory)
17. Use care in routing of drain hose
Do avoid making bend or projection of drain hose. In case that the drain hose is bent and the drain cannot be performed, back-flow or breakage
of the cooling hose might occur.
Place the drain hose lower than the drain port of this unit. Further, avoid piping which
allows paddle in the hose or at the hose outlet, because it is a resistance against drain.
Place the end of drain hose where drain is allowed. When distilled water is being prepared,
cooling water is drained approx. 2 liters/min. Also drain further increases when boiler water is drained, and sufficient drain equipment is required.
Page 14
10
2. Before Using this unit
Requirements for Installation
WARNING!
18. Check the drain temperature of cooling water
Drain temperature may exceed 60 in drainage from boiler. Drain to a place remote from
working environment not to be touched easily because there is a danger of burns.
High-temperature cooling water could flow out. If vinyl chloride tube is used for the water
drain unit of the sink, such a tube could deteriorate. Water should be drained to a place away from the drain tube of the sink. Even if VP tube (JIS K6741) is used for the vinyl chloride tube, DV-RR joint is used for the joint, and even if insert socket (JIS K6739) is used, the water drain trap in the nonstandard options should be used when the control temp. cannot be lowered (60 or lower). Even if the drain temperature is 60 or lower, and if the above-mentioned tubes and joint are not used, the drain trap in the nonstandard options should be used.
19. When drain temperature of sink equipment does not fall under 60
Use a drain trap (optional accessory). Drain trap makes temperature fall by accumulating cooled drain water temporarily. Further,
it mixes city water and cooled-down drain water, makes mixed water temperature fall, then lets drain to sink equipment.
Contact your dealer or Yamato Scientific sales office for detail of drain trap.
Page 15
11
2. Before Using this unit
Requirements for Installation
WARNING!
20. Install firmly the ion exchange resin cartridge (CPC-N) and high-purity cartridge
(CPC-H).
Install the ion exchange resin cartridge (CPC-N) and high-purity cartridge (CPC-H) following
the procedure shown below.
Connect securely because insecure connection may cause water leakage.
1. Make sure that the earth leakage breaker of this unit is "Off" and that the tap is tightened.
2. Take the ion exchange resin cartridge and high-purity cartridge out of attachments to the unit.
3. Place the ion exchange resin cartridge and high-purity cartridge taken out on the receiver within the unit. (See Fig.1.)
4. Fix the ion exchange resin cartridge and high-purity cartridge with the band of receiver. (See Fig.2.)
Fig.1 Fig.2
5. Remove the rubber cap attached to the inlets and outlets of ion exchange resin cartridge and high-purity cartridge.
6. Fit in the coupler marked with (IN) to the inlet of ion exchange resin cartridge (left) until click is heard and repeat the procedure in the same way for high-purity cartridge. (See Fig.3.)
7. Fit in the coupler marked with (OUT) to the outlet of ion exchange resin cartridge (right) until click is heard and repeat the procedure in the same way for high-purity cartridge. (See Fig.3.)
IN
リセットスイッチ
I
N
O
U
T
OUT
Fig.3
Coupler may be hard at first. When applying force in inserting, do not make it curved
because insertion port may be broken.
8. Coupler can be removed easily from the ion exchange resin cartridge by pulling it toward yourself while pushing the black part of the coupler to the depth.
Page 16
12
2. Before Using this unit
Requirements for Installation
WARNING!
21. Secure the pre-treatment cartridge securely
Connect the hose in the body securely following the procedure shown below. Insecure connection may cause disconnection of connection hose, resulting in accident by
water leakage.
1. Make sure that the earth leakage breaker of this unit is "OFF" and that the tap is tightened.
2. Take the pre-treatment cartridge out of attachments to the body.
3. Inlet and outlet of the pre-treatment cartridge are provided with a cap, so remove it.
4. When the front door of this unit is opened, connection hose marked IN and OUT is found in the coupler there make connection matching them with IN and OUT on pre-treatment cartridge .
5. In connecting, mate the coupler and port of cartridge while sliding the blue portion of coupler toward the hose, push in, then release the blue portion .
6. When connection is finished, place the pre-treatment cartridge at the position shown on the right (near side on the left of distilled water tank) as paying attention to the bend of hose.
Fig.4
Fig.5
Page 17
13
2. Before Using this unit
Requirements for Installation
WARNING!
22. Install the membrane filter firmly.
Install the membrane filter as follows: Unless firm connection is made, water may leak from the threaded portion and may be mixed
into and contaminate the sampled pure water. Always ensure firm connection.
clockwise as viewed from the tape windind side two to three turns while pulling the tape slightly. Remaining tape should be cut away.
care not to crush threads. Check for water leak during sampling of pure water. If any, screw the membrane filter further.
Page 18
14
2. Before Using this unit
Requirements for Installation
WARNING!
23. Essential Points for Initial Use
In the case of WG270, the quality of distilled water stored in the distilled water tank may be
worse than the A4 level quality temporarily because of contact with the atmospheric air. WG270 is equipped with the function to recover the water quality of such pure water to the A4 level by means of circulation under ultraviolet ray irradiation and removal of CO2 gas.
It may take hours for the distilled water stored in the tank after initial start of WG270 to reach
the A4 level water quality. Carry out cleaning of the tank inside as follows: (1) Install and start operation of WG270 system. Keep operation till the tank is filled with
the distilled water. (2) Discharge the distilled water totally from the outlet of distilled water tank. (3) Start operation again and keep operation till the tank is filled with the distilled water. (4) When the tank is full of the distilled water, continue operation without turning OFF power
supply (keep the distilled water circulating). In four or more hours, use the distilled
water. (This step is necessary for initial operation only).
A guideline is as follows. If the specific resistance of distilled water is maintained at 1.2
Ω・cm or more for one hour or more continuously, the distilled water with the A4 level or higher water quality can always be sampled. If the specific resistance of distilled water is less than 1.2MΩ・cm, continue operation (keep the distilled water circulating) for approximately four hours or more before sampling the distilled water.
When the distilled water has been stored for two weeks or more in the tank, drain the water
totally and start operation. Take steps (3) and (4) above before using the distilled water.
24. After installation
This unit may be turned over under unexpected earthquake or impact. Provide adequate
measures against such turn-over to ensure the safety.
Page 19
15
3. Description and Function of Each Part
Main Unit
WG270
Boiler
Heater
Heater
terminal block
Ion exchange resin cartridge (CPC-N)
Drain port (2pcs each at right/left)
Earth leakage breaker
Ion exchange water
sampling port
Pre-treatment cartridge
Distilled water tank
Reset switch
Water sampling table
Drain switch for
water sampling table
Distilled water
sampling port
Leakage detection electrode
Drain cock
Control panel
Product number sticker
High-purity cartridge (CPC-H)
Water feeding port
Page 20
16
3. Description and Function of Each Part
Water Sampling Table
Water sampling table
Use the water sampling table by pulling it out frontward. Because coolant for condenser is used, so for draining the water in the following cases (and coolant is not flowed), press the drain switch specifically used for the water sampling table at right side of the unit.
Coolant flows for one minute, and then, the water sampling table is drained. After that, returns to the condition before pressing the switch.
When the tank is filled to capacity (Distillation is not operated.) When drawing pure water While standby time (the breaker is turned on, and POWER key is turned off.)
Drain switch for water sampling table
Water sampling table
Page 21
17
3. Description and Function of Each Part
Piping System View
1. Pressure reduction valve
2. Pressure switch
3. Cooling water solenoid valve
4. Raw water supply solenoid valve
5. Pre-treatment cartridge
6. Electromagnetic pump
7. Ion exchange resin cartridge (CPC-N)
8. Ion exchange water quality gauge electrode (for CPC-N)
9. High-purity cartridge (CPC-H)
10. Float cylinder
11. Float switch (1)
12. Float switch (2)
13. Boiler drain solenoid valve
14. Boiler drain cock
15. Boiler
16. Heater
17. Condenser
18. Ion exchange water quality gauge electrode (for CPC-H)
19. Check valve
20. Boiler water supply solenoid valve
21. Initial accumulated water drain solenoid valve
22. Distilled water quality gauge electrode
23. Deaeration membrane module
24. Solenoid valve for vacuum pump
25. Vacuum pump
26. TOC reducing UV lamp
27. Distilled water sampling pump
28. Distilled water tank drain port
29. Float switch (3)
30. Distilled water tank
31. Air filter
32. Check valve for circulation return
33. Ion exchange water sampling solenoid valve
34. Distilled water sampling solenoid valve
35. Ion exchange water flow sensor
36. Distilled water flow sensor
37. Ion exchange water sampling port
38. Distilled water sampling port
39. Membrane filter
40. Aspirator
Page 22
18
3. Description and Function of Each Part
Principle of Operation
Description of the operation mechanism of WG270.
(1) Drain of the boiler
Set the earth-leakage breaker to “ON”, and the standby screen appears in about four seconds. In
this state, press the POWER key, the boiler drain solenoid valve and cooling water solenoid valve open simultaneously for about 40 seconds. Boiler drain is made every five hours of distillation operation.
(2) Boiler Water Supply and Distilling Operation
When drain of boiler is finished, both the raw water supply solenoid valve and boiler water supply
solenoid valve open simultaneously, supplying water to the boiler . When the float switch (1) in the float cylinder detects the water level, the heater is energized to start distillation. Water supply to the boiler is controlled by opening/closing the raw water supply solenoid valve and boiler water supply solenoid valve with the fload switch(2)⑫. The supply-water circulation pump keeps circulation of the ion-exchanged water normally during distillation, ensuring stable water supply to the boiler .
Page 23
19
3. Description and Function of Each Part
Principle of Operation
(3) Flow of Cooling Water
During distillation, water is supplied and discharged in the order: pressure-reducing valve, cooling water solenoid valve and condenser . When the distilled water tank is full, or when ion exchanged water is sampled, distillation is stopped, and the cooling water is also stopped automatically.
(4) Sampling of Distilled Water
Distilled water condensed in the condenser is stored in the distilled water tank○30 for about 10
minutes from the start of distillation. In this case, the initial accumulated water drain solenoid valve○21 opens to discharge this water beforehand. When the float switch○29 at the top portion inside the tank is activated, the tank is judged to be full and distillation stops. When the distilled water is consumed to a certain level through its sampling, the unit generates the distilled water automatically. Stored distilled water is sampled, via TOC reduction UV○26, deaeration membrane module○23 , distilled water quality electrode○22 , distilled water sampling solenoid valve○34 , distilled water flow sensor○36 , distilled water sanpling port○38 , and membrane filter○39 , by the distilled water sampling pump○27 .
Page 24
20
3. Description and Function of Each Part
Principle of Operation
(5) Sampling Ion Exchanged Water
Ion exchange water is sampled by way of the pressure-reducing valve , raw water supply solenoid
valve , pre-treatment cartridge , supply-water circulation pump , ion exchange resin cartridge
(CPC-N), ion exchange water quality electrode CPC-N (front), high-purity cartridge (CPC-H) , ion exchange water quality electrode CPC-H (rear)18 , ion exchange water sampling solenoid
valve○33 , ion exchange water flow sensor ○35 , ion exchange water sampling port○37 , and membrane filter○39 . During ion exchange water sampling, the supply-water circulation pump runs to circulate the ion exchange water.
Page 25
21
3. Description and Function of Each Part
Principle of Operation
(6) TOC Reduction system
The TOC reduction system consists of returning the distilled water stored in the distilled water tank,
via TOC reduction UV, deaeration membrane module, distilled water quality electrode, and circulation return check valve, to the distilled water tank by means of the distilled water sampling pump.
With LED (the second LED from the bottom, which corresponds to the water amount of about 5.5
liters) indicating the distilled water level in the tank going ON in green, the TOC reduction system is activated, starting circulation. Almost simultaneously with passage of the distilled water through the deaeration membrane module, the deaeration vacuum pump is activated. The electric conductivity of deaerated distilled water decreases. When the electric conductivity of distilled water in the tank exceeds 0.83μScm, the TOC reduction UV lamp goes ON. The TOC value decreases along with increase in this electric conductivity. When the electric conductivity increases above 0.95μS/cm, the TOC reduction UV lamp goes OFF. Then, the distilled water is sampled. Circulation and vacuum pumps stop when the level LED (the LED in the bottom, which corresponds to thwater amount of about 2.5 liters). Under other conditions, both circulation and vacuum pumps are in operation normally while the TOC reduction UV lamp goes ON/OFF depending on the conditions. In this manner, the distilled water quality in the tank can be maintained.
To protect the vacuum pump (VP), the pump is open to the atmosphere for five minutes every two
hours by means of the solenoid valve. In this case, VP exhaust sound may become louder.
Page 26
22
3. Description and Function of Each Part
Control Panel
No.
Name
Operation/Action
POWER key
Turns on/off the power of the controller.
Message indication area
Indicates the measured value and setting value.
WATER LEVEL lamp
Indicates water level in the distilled water tank in five levels.
DITILLED WATER key
Starts/stops drawing distilled water.
▲▼key
Selects the value setting item.
PURE WATER key
Starts/stops drawing ion exchange water.
FILL key
Switches water collecting method. (Measured filling/Continuous filling)
FILL lamp
Lights up when either of MEASURED or CONTINUOUS is selected.
CONDUCTIVITY key
Switches conductivity indication unit. (S/m⇔Ω・m
M key
Used when entering submenu or maintenance mode. Also (confirms the setting) then shifts to the next setting item.
F key
(Cancels the setting) then returns to the previous setting item.
Pure water lamp
Lighting during pura water sampling
Distilled water lamp
Lighting duringdistilled water sampling
Water level lamp
This lamp indicates the storage amount of distilled water in the tank in five levels. When this red lamp lights, distilled water cannot be collected for empty drive prevention of the pump. When this red lamp lights, distilled water cannot be collected for empty drive prevention of the pump. When water keeps being stored, and a red lamp in the the lower is blinked, distilled water can be gathered. In addition, a red lamp in the the lower is turned off when the amount of storing water increases, a green lamp since the second step lights, and the amount of storing water can be confirmed.
FILL
CONDUCTIVITY
M
F
WATER LEVEL
MEASURED
CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
FULL
EMPTY
Lighting
FULL
EMPTY
FULL
EMPTY
FULL
EMPTY
FULL
EMPTY
FULL
EMPTY
Sampling pump operating range
No-water
Filled
Blinking
Lighitng
Lignting
Lighting
Lighting
⑫ ⑬ ⑤
Page 27
23
4. Operation Method
Setup and Check before Use
WARNING!
1. Check of water supply
Check that the water supply hose is securely connected. Open the tap. Check that water does not leak from connection of water supply hose.
2. Check of drain
Check that the drain hose is securely connected. Check that the drain hose is free from bend or projection. When the drain hose is bent or the like, system does not operate normally, and in addition, it
may lead to water leakage accident. Inspect from time to time, and ensure that water is drained properly.
3. Check of power supply
Check that the power cord is connected to appropriate plug socket.
4. Before operation
Turn on the earth leakage breaker, then, perform calibration before pressing POWER key.
Perform calibration operation at first-time using this unit (refer to page 24) and when changing the heater of the boiler (refer to page 42). Press POWER key while holding down PURE WATER key and DISTILLED WATER key. Perform calibration operation (at the measured values of heater temperature and power-supply voltage) for about five minutes, after then, distillation starts automatically. The calibrator is used to save the reference temperature (under normal operation state) of temperature sensor incorporated into the heater into the internal controller. If the temperature exceeds the reference temperature by +20 or more due to certain reasons, it is detected as abnormal. Key operation becomes disable while calibration operation.In case that power failure occurs while calibration operation, please perform calibration again.
5. Caution at initial operation
In sampling distilled water in initial energization and drain from distilled water storage tank, air
is contained in the pump and piping, and it takes time until sampling is started.
In sampling ion exchanged water immediately after changing pre-treatment cartridge or ion
exchange resin cartridge, it also takes time until sampling is started. Further, when each cartridge is changed, drain about 5 liters in order to remove initial impurities.
Page 28
24
4. Operation Method
Operation Procedure
When operation is set up, follow the procedure below for operation:
1. Turning on power
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
Yamato
Auto Still
WG250
2005/11/1 12:00
When first-time using the unit, perform calibration operation.
1. Turn on the earth leakage breaker.
2. Press POWER key while holding PURE WATER key and DISTILLED WATER key.
3. After five minutes passed, distillation starts automatically. (Clause 3)
From the second time using:
Turn on the earth leakage breaker. Press POWER key.
2. Drain of the boiler
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
Boiler Drainage
DW/DI Water
Not Available
Drain of the boiler starts. Drain of the boiler starts when turning on/off the earth
leakage breaker, and when five hours passed after distillation starts. (Drain of the boiler does not start if POWER key is not turned on.)
3. Distillation operation
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
In Distillation
Tank Empty
Distill-Water NA
Continuous
Distillation operation starts. When there is no water in the tank, the left screen is
displayed and sampling continues to be impossible after removal of the empty state Red LED flashing). When the level LED changes from flashing in red to continuous lighting in green, circulation of distilled water begins. The screen shown below appears.
In Distillation
DW-Conductivity
0.854×10-4S/m
Continuous
It takes approx. 4 hours till the tank is filled at the water collection available level. The above screen appears. When the tank is full, the below screen appears.
DW Tank Full
DW-Conductivity
0.854×10-4S/m Continuous
WG27 0
Page 29
25
4. Operation Method
Sampling of Pure Water
Continuous collection of pure water
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
DI sampling
DIS-Conductivity
0.06×10-4S/m
Continuous
DI sampling
DIH-Conductivity
0.06×10-4S/m Continuous
During distillation or when the distilled water tank is full, pressing the PURE WATER key while CONTINUOUS of the sampling selector lamp is ON causes the pure water lamp goes ON, enabling sampling of the pure water. If distillation is being made, the distillation
operation is stopped temporarily. Collecting water stops by pressing
PURE WATER key again. Then, PURE WATER lamp is turned on. After water collection is finished, it returns to distillation operation.
Continuous collection of distilled water
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
DI sampling
DIS-Conductivity
0.06×10-4S/m
Continuous
DW sampling
DW-Conductivity
0.854×10-4S/m Continuous
While the distilled water tank level LED is ON in green and when CONTINUOS of the sampling selector lamp is ON, pressing the DISTILLED WATER key causes the distilled water lamp to go ON, enabling sampling of the distilled water.
To stop sampling, press the DISTILLED WATER key again, and the distilled water lamp goes OFF. During sampling of distilled water, distillation is continued. Sampling is possible till the distilled water tank level LED goes ON in red.
Page 30
26
. Operation Method
Collection of Measured Amount of Pure Water
1. Switching water collection method
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
In Distillation
DW-Conductivity
0.85×10-4S/m
Measured: 0.1L
Switch to “MEASURED” amount water collection
method.
Switch from “CONTINUOUS” to “MEASURED” by
pressing FILL key.
2. Amount of collecting water setting
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
In Distillation
DW-Conductivity
0.85×10-4S/m
Measured: 20.0L
Set the amount of collecting pure water. The same setting screen is used for setting of pure water and distilled water.
1. Set the amount of pure water by pressing ▲▼ key. WG270: Setting can be made up to 20 liters
2. After setting, pressing PURE WATER key starts water collection.
3. Collecting
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
DI sampling
DIS-Conductivity
0.06×10-4S/m
Continuous
DI sampling
DIH-Conductivity
0.85×10-4S/m
Measured: 18.2L
Water collection of measured amount starts.
1. The counter starts decrementing from the setting value.
2. When collecting water is finished, message "Measured: 0.0L" appears on the window.
3. Go back to clause 2.
If pressing PURE WATER key while collecting water,
the operation stops and the setting value is reset.
Page 31
27
4. Operation Method
Collection of Measured Amount of Distilled Water
1. Switching water collection method
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
In Distillation
DW-Conductivity
0.85×10-4S/m
Measured: 0.1L
Switch to “MEASURED” amount water collection
method.
Switch from “CONTINUOUS” to “MEASURED” by
pressing FILL key.
2. Amount of collecting water setting
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
In Distillation
DW-Conductivity
0.85×10-4S/m
Measured: 20.0L
Set the amount of collecting distilled water. The same setting screen is used for setting of pure water and distilled water.
1. Set the amount of distilled water by pressing ▲▼ key. WG270: Setting can be made up to 20 liters
2. After setting, pressing DISTILLED WATER key starts water collection.
3. Collecting
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
DW sampling
DW-Conductivity
0.85×10-4S/m
Measured: 18.2L
Water collection of measured amount starts.
1. The counter starts decrementing from the setting value.
2. When collecting water is finished, message "Measured: 0.0L" appears on the window.
3. Go back to clause 2.
If pressing DISTILLED WATER key while collecting
water, the operation stops and the setting value is reset.
Page 32
28
4. Operation Method
Collection of Measured Amount of Distilled Water
In case that the tank becomes empty while collecting water
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
In Distillation
Tank Empty Distill-Water NA Measured: 15.0L
1. The left screen is displayed when the tank becomes empty during collecting water, and DISTILLED WATER lamp blinks and operation of water collection is suspended.
2. Distillation starts.
3. In case that "EMPTY" lamp of WATER LEVEL display does not light up, press
DISTILLED WATER key again to start collecting water. If PURE WATER key is pressed under this condition, the holding condition is released.
4. After collecting water is finished, "Measured: 0.0L" appears on the window, and return to the screen of Clause 2.
Cautions
1. During distillation and ion exchange, the supply water circulation solenoid pump is always running. When the distilled water tank is full, this pump stops. To sample pure water while this pump is stopped, set this pump normally ON in the sub-menu and sample pure water after stabilization of the water quality. This pump can be set to normal operation from the sub-menu while the distillation water tank is full.
2. If the supply water circulation pump has been set to normally ON, its setting is cancelled under following conditions:
Power key turned OFF Distilled water sampled, the storage tank water level lowering, and distillation resumed Abnormality
Page 33
29
4. Operation Method
Display of Water Quality
Switching conductivity unit
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
DW sampling
DW-Conductivity
0.85×10-4S/m
Continuous
Switch the conductivity unit. Pressing CONDUCTIVITY key for 2 seconds switches
the conductivity unit. Ex)
In case of switching the conductivity unit during collecting distilled water
DW sampling
DW-Conductivity
0.85×10-4S/m
Continuous
DW sampling
DW-Resistivity
1.17×104Ω・m
Continuous
Display is made in the following order: Conductivity Specific resistance → Conductivity.
Switching conductivity display
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
DW sampling
DW-Conductivity
0.85×10-4S/m
Continuous
Switch the conductivity display. The ion water conductivity is displayed durig sampling of pure water. In other cases, the conductivity of
distilled water is displayed.(Excluding the period up to start of circulation of distilled water inside the tank.
Pressing the Conductivity key enables switching of water quality display in the order of pure water CPC-H→ pure water CPC-N→ distilled water→ pure water CPC-H.
Example To display the pure water conductivity
during sampling of the distilled water:
DW sampling
DW-Conductivity
0.85×10-4S/m
Continuous
DW Sampling
DIH-Conductivity
0.06×10-4S/m Continuous
With out key operation for about 10 seconds, the conductivity returns to the original one.
Page 34
30
4. Operation Method
Display of Water Quality
Measurement of Electric Conductivity
The water conductivity meter on the control panel displays the conductivity at the outlets of the ion-exchange resin cartridge and the condenser for distilled water. The displayed value can be used as an index for replacing the ion-exchange resin cartridge. The value of conductivity is valid only when the electrode is fully damped with water, or when pure water is dripped. Correct value is not displayed in the following cases because the electrode is not damped or air bubbles are produced.
1. At beginning of operation and during halts
2. Just after changing pretreatment cartridge or ion-exchange resin cartridge
3. Just after distillation starts
Electric conductivity
Electric conductivity is a value indicating easiness of flowing of electricity. In the case of water,
electricity flows the more easily when the more electrolyte i.e. impurity is solved, so the value of conductivity is the greater; when the less electrolyte is solved, the smaller is the value.
When the value of electric conductivity is the smaller, the better is purity of pure water. Here,
electric conductivity indicates only electrolyte, and does not indicate content of non-electrolyte (such as organic substance, colloid substance, dissolved gas, and microorganism), and it is just an index indicating purity of pure water, and it does not represent all of purity.
Specific resistivity indicates the same contents as electric conductivity. Specific resistivity is
inversely related to electric conductivity, and when the value is the greater, the better is purity.
When obtaining specific resistivity from electric conductivity, where specific resistivity is R and
electric conductivityρ,
R[Ω・m]
1
  S m]
[ /
or
R[×10 Ω・m]
1
  ×10 S m]
-4
[ /
So the theoretical value of pure water is as follows: R=18.3×104Ω・m (18.3MΩ・cm) 25℃ (Take notice that the resistivity is displayed in integer form (not in decimal) in the range 18 to 1 x 104Ω・m.) ρ=0.055×10-4S/m (0.055μS/cm) 25
Quality of ion exchange water and distilled water
Ion exchange water and distilled water have the following features respectively. Distinguish them
as necessary in use. It is ideal to use pure water immediately after sampling; therefore be sure to drain water in distilled
water tank if it is out of use for a long time. If water has been stored in distilled water tank for a long time, drain once, then store in distilled water tank newly before use.
1. Ion exchange water Most of electrolyte in water is removed, and water with the lowest electric conductivity is obtained. However, non-electrolyte cannot be removed. In addition, slight fall of purity is found while resin is new and when water is fed again after halt of system.
2. Distilled water Electrolyte and non-electrolyte can be removed in average except for low boiling point substance such as ammonia. However, carbon dioxide gas in the atmosphere is absorbed and carbon oxide is generated in the process of manufacturing (condensing/storing), and so the electric conductivity is worse than ion exchange water, that is 1 to 2.5 X 10-4 S/m (1 to 2.5 μ
S/cm) at 25, and represents weak acid (pH5 to 6). See "2 - Common item (11) Water" of JIS K0102 (Plant drain test procedure) for removal of dissolved gas (oxygen and carbon dioxide) in pure water.
Page 35
31
4. Operation Method
Display of Water Quality
About TOC
TOC stands for “Total Organic Carbon.”
TOC represents contamination of the water by means of the amount of carbon contained in organics in wate. Of carbon compounds dissolved in water, the TOC value may be used to represent the total amount of carbon in organic compounds.
For the A4 water specified in JIS, this is 50ppb or less.JIS K 0557 For measurement, the UV oxidation + direct conductivity measurement methods are used. The UV
oxidation conductivity measurement method consists of measurement of the conductivity before and after organic decomposition and of calculation of TOC as the decomposition product from the difference.
TOC units: [mgC/L][μgC/L]、 [ppm] [ppb]
1[mgC/L]1000[μgC/L]=1[ppm]1000[ppb]
To obtain the repeatability of analytical result in the experiment, it is essential to use the super pure water of certain water quality. To check the water quality, it is necessary to control the water quality by means of “electric conductivity” and “TOC value.”
Item
Water quality standardJIS K 0557
A1
A2
A3
A4
Total organic carbon(TOC mgC/l
1 or less
0.5 or less
0.2 or less
0.05 or less Conductivity mS/m (25)
0.5 or less
0.1 or less
0.1 or less
0.1 or less
Zinc μgZn/l
0.5 or less
0.5 or less
0.1 or less
0.1 or less
Silica μgSiO2/l
-
50 or less
5.0 or less
2.5 or less
Chloride ion μgCl-/l
10 or less
2 or less
1 or less
1 or less
Sulfate ion μgSO
4
2-
/l
10 or less
2 or less
1 or less
1 or less
In the case of WG270 the quality of distilled water stored in the distilled water tank may be
worse than the A4 level quality temporarily because of contact with the atmospheric air. WG270 is equipped with the function to recover the water quality of such pure water to the A4 level by means of circulation under ultraviolet ray irradiation and removal of CO2 gas.
It may take hours for the distilled water stored in the tank after initial start of WG270 to
reach the A4 level water quality. Carry out cleaning of the tank inside as follows: (1) Install and start operation of WG270 system. Keep operation till the tank is filled with
the distilled water. (2) Discharge the distilled water totally from the outlet of distilled water tank. (3) Start operation again and keep operation till the tank is filled with the distilled water. (4) When the tank is full of the distilled water, continue operation without turning OFF
power supply (keep the distilled water circulating). In four or more hours, use the
distilled water. (This step is necessary for initial operation only).
A guideline is as follows. If the specific resistance of distilled water is maintained at 1.2
Ωcm or more for one hour or more continuously, the distilled water with the A4 level or higher water quality can always be sampled. If the specific resistance of distilled water is less than 1.2MΩcm, continue operation (keep the distilled water circulating) for approximately four hours or more before sampling the distilled water.
When the distilled water has been stored for two weeks or more in the tank, drain the
water totally and start operation. Take steps (3) and (4) above before using the distilled water.
Page 36
32
4. Operation Method
Setting/Display of Submenu
1. Setting/display of submenu
FILL
CONDUCTIVITY
M F
WATER
LEVEL
MEASURED CONTINUOUS
FULL
EMPTY
POWER
PURE WATER
DISTILLED WATER
Sub-Menu
1.Buzzer
2.Calendar
3.EXCH History
Perform setting/display of submenu.
1. Pressing M key for 2 seconds displays the submenu.
2. Select the item to be set or displayed by pressing ▲▼ key.
3. Pressing M key confirms the setting and shifts to the selection screen.
Press F key to (cancel the setting and) return the
previous setting screen.
If not pressing any keys for 2 minutes, the screen
returns to the previous display before setting/display of submenu.
Items of submenu are:
1) Buzzer
2) Calendar
3) EXCH History
4) M History
5) Error History
6) Outage History
7) Water Alarm
8) Pump Manual
9) Language
10) Resume FUNC
2. Setting/display of each item
2.1. Setting of ON/OFF of buzzer sound
Buzzer
1. Key Click
2. Error Buzz
To set a key clicking sound/alarm buzzer sound:
1. Select an item of submenu by pressing ▲▼ key.
2. Confirm the item by pressing M key. The selection screen appears.
3. Specify "ON" or "OFF" of each item by pressing ▲▼ key, and confirm the setting by pressing M key.
4. Pressing F key returns to the submenu screen.
"ON" is set to buzzer sound at default setting.
2.2. Calendar setting
Calendar 2006 year 9/27 month/day 12:00 h:min
To set the calendar:
1. Set the items (year/month/day/hour/minute)by pressing ▲▼ key.
2. Confirm the setting by pressing M key.
3. The set items are displayed. Check the set items and confirm your entry by pressing M key to go to the selection screen.
4. Pressing F key returns to the previous submenu screen.
①,③
Page 37
33
4. Operation Method
Setting/Display of Submenu
2. Setting/display of each item
2.3. Consumables replacement history
EXCH History
1. Hollow Filter
2. Ion-Exchange N
3. Ion-Exchange H
4. DW M-Filter
5. DI M-Filter
6. MHF
7.TOC-d UV Lamp
For display of the history of the pre-treatment filterion exchange resinN)・ Ion exchange resin (H)・ distilled water M filter・
ion change water M filterdeaeration membrane module
Cautions
1. During distillation and pure water sampling, the supply water circulation solenoid pump is always running. When the distilled water tank is full, this pump stops. To sample pure water while this pump is stopped, set this pump normally ON in the sub-menu and sample pure water after stabilization of the water quality. This pump can be set to normal operation from the sub-menu while the distillation water tank is full.
2. If the supply water circulation pump has been set to normally ON, its setting is cancelled under following conditions:
Power key turned OFF Distilled water sampled, the storage tank water level lowering, and distillation resumed Abnormality
When the distilled water tank is full, set the supply water circulation pump to “Normally ON” according to Setting and Displaying Submenu, “2.8 PUMP manual operation” in Page 34.
Page 38
34
4. Operation Method
Setting/Display of Submenu
2.4. Maintenance history Maintenance
History No.1
To display the maintenance history:
1. Confirm the maintenance history by pressing ▲▼ key.
2. Pressing F key returns to the previous screen.
Up to 20 records are displayed for each change history.
If it exceeds 20 records, the oldest record is deleted.
2.5. Error occurrence history
Error History History No.1 Leak error 2005/09/29 12:25
To display the error occurrence history:
1. Confirm the maintenance history by pressing ▲▼ key.
2. Pressing F key returns to the previous screen.
Up to 20 records are displayed for each change history.
If it exceeds 20 records, the oldest record is deleted.
For details of error alarms, refer to page 47.
2.6. Power failure occurrence history Outage History
History No.1
05/09/29 12:25 05/09/29 12:30
To display the power failure occurrence history:
1. Confirm the maintenance history by pressing ▲▼ key.
2. Pressing F key returns to the previous screen.
Up to 20 records are displayed for each change history.
If it exceeds 20 records, the oldest record is deleted.
2.7. Water cutoff occurrence history Water Alarm
History No.1
05/09/29 12:25 05/09/29 12:30
To display the water cutoff occurrence history:
1. Confirm the maintenance history by pressing ▲▼ key.
2. Pressing F key returns to the previous screen.
Up to 20 records are displayed for each change history.
If it exceeds 20 records, the oldest record is deleted.
2.8. Pump manual operation Pump manual
1.RUN
2.Cancel
To set the pump manual operation,
1. Select the submenu item to be set with ▲and ▼ keys.
2. Acknowledge with the M key and return to the submenu
screen.
Effective only when the distilled water tank is full
2.9. Language selection Language
1. English
2. Japanese
To select display language:
1. Select an item of submenu by pressing ▲▼ key.
2. Confirm the item by pressing M key. The submenu
screen appears.
2.10. Power failure recovery function Resume Function
1. No
2. Yes
To set the power failure recovery:
1. Select an item of submenu by pressing ▲▼ key.
2. Confirm the item by pressing M key. The submenu
screen appears.
This function is set to "invalid" at default setting.
3. Finishing of setting/display
Press M key for 2 seconds. The screen returns to the previous display before setting/display submenu.
Page 39
35
5. Handling Precautions
WARNING!
1. If a problem occurs
If smoke or strange odor should come out of this unit for some reason, turn off the power
key right away, and then turn off the circuit breaker and the main power. Immediately contact dealers or Yamato Scientific Co., Ltd. sales office for inspection. If this procedure is not followed, fire or electrical shock may result. Never perform repair work yourself, since it is dangerous and not recommended.
2. During a thunder storm
During a thunderstorm, turn off the power key immediately, then turn off the circuit breaker
and the main power.
If this procedure is not followed, fire or electrical shock may be caused.
3. Take enough care in handling detergent (liquid)
In storing detergent (liquid), store in enclosable container avoiding high temperature and
humidity.
Principal component of detergent (liquid) Orgazor 10 is sulfamic acid (acidic with pH of
water solution approx. 1).
In handling this detergent (liquid), use protective tools (gloves, mask, and glasses). When it is in contact with human body, wash it away with clean water. Neutralize the liquid with neutralizer (such as sodium hydroxide) after washing. Ensure neutralization with pH test paper, etc. Do not use empty container for beverage. Do not allow detergent to directly flow into agricultural irrigation canal or fields because it
causes withering of rice crop.
4. Do not step on this unit
Do not step on this unit. It will cause injury if this unit fall down or break.
5. Do not put anything on this unit
Do not put anything on this unit. It will cause injury if fall.
6. In power failure (Factory setting :no power failure recovery function)
When system has halted during operation due to power failure etc. and is provided with power again, system is brought to standby status. When restarting operation, start from (Page 24 "1. Turning on power ").
7.Do not look directly into the TOC reduction UV lamp while this lamp is ON.
While the TOC reduction system is operating, the TOC reduction UV lamp may go ON if the specific resistance of the distilled water in the tank increases to 1.05MΩ・cm or more. Never look directly into the UV lamp when opening the front door of this instrument. When replacing consumables, turn OFF the circuit breaker beforehand.
Page 40
36
6. Maintenance Method
Inspection and Maintenance
Timing of maintenance and inspection (Perform daily inspection for stable use of product.)
Maintenance/check
items
Reference for timing
Remarks
Replacement of pre-treatment cartridge
Approx. 6 months
Exchange message appears
on the indication area.
Throughput: About 5000 liters of city water in Tokyo. If the quality of raw water is not good, perform this maintenance as soon as possible.
Replacement of membrane filter
Approx. 3 months
Exchange message appears
on the indication area.
Throughput: approx. 1500 liters for pure water. Need to exchange if the amount of collection is less than the above value.
Replacement of ion exchange resin cartridge CPH-N
Exchange message appears
on the indication area.
(Displayed when the
conductivity of pure water is
more than 1 x 10-4S/m.)
Throughput: About 600 liters with the raw water of 200 x 10-4S/m.
Replacement of ion exchange resin cartridge CPC-H
Exchange message appears
on the indication area.
(Displayed when the
conductivity of pure water is
more than 0.2 x 10-4S/m.)
Throughput about 600 liters with the raw water of 200 x 10-4S/m By replacing CPC-N, the CPC-H service life can be extended substantially.
Replacement of deaeration membrane module
About one year
Exchange message appears in
the indication area.
Throughput:Maintains the specific resistance of distilled water to1.0 MΩ・cm or more
Replacement of TOC reduction UV lamp
About 4000 hours
Exchange message appears in
the indication area.
ThroughputTOC of the distilled water stored in the tank may be maintained at 50 ppb or less (equivalent to A4 water of JIS) for about week. If the tank has been left unused for a long period of time, drain first, and store the distilled water again before use.
Cleaning of distiller
3 months
If the quality of raw water is not good, perform cleaning early.
Cleaning of water supply hose filer
6 months
Replacement of hose
2 years
Connections shall be checked monthly.
Drainage of distilled water tank
3 months
If not used for a long time, drain water from the tank.
Inspection of the vacuum pump
One year
Change the consumable parts.
Replacement of pre-treatment cartridge
See Page 12 “Secure the pre-treatment cartridge securely " for replacement procedure. When the cartridge is put to use without replacement, life span of ion exchange resin cartridge
becomes short. Please dispose used cartridge as noncombustible. When it is sent back to our company, fill in the
designated invoice attached to a new cartridge, and send it with the used cartridge to us.
Page 41
37
6. Maintenance Method
Replacement of ion exchange resin cartridge
When a cartridge is stored for a long time, deterioration of water quality and fall of processing capacity are found; therefore prepare a spare cartridge in a planned manner for replacement timing. Standard for storage is about 4 months.
In replacement, see Page 11 "Install firmly the ion exchange resin cartridge (CPC-N) and high-purity cartridge (CPC-H)."
When a cartridge is used without replacement, much scale is deposited on boiler and heater, which causes decrease of distilled water sampling and damage to heater.
Please dispose used cartridge as nonconbustibles. When it is sent back to our company, fill in the designated invoice attached to a new cartridge, and send it with the used cartridge to us.
We promote the reasonable disposal, collection, and recycling of cartridge for environmental protection.
After replacement of pre-treatment cartridge・membrane filter・ion exchange resin cartridge, and high
purity cartridge, press the reset switch on the backside of front door, with power of the unit ON, for two seconds. The acceptance sound is issued. Alarm display is reset and registered in the consumables replacement history. In case of replacement of multiple components, register one component at a time in the order of replacement. Note that the history of ion exchange resin cartridge is registered last of all.
(Alarm for ion exchange resin and high-purity cartridges is automatically reset. Pressing the reset
switch causes recording of the replacement history.)
リセットスイッチ
Reset switch
Page 42
38
6. Maintenance Method
Washing of Distiller
Dismounting of Distiller
1. Turn "OFF" the earth leakage breaker of the unit.
2. Close the tap.
3. Check that the boiler is not hot (longer than 30 minutes after the breaker is turned "OFF"), then open the front door of the unit, and open the boiler water drain cock.
4. Disconnect the hose connected to the boiler and condenser . In disconnecting from the distilled water outlet and boiler water supply and drain port, turn the hose band by use of tool and displace the engaged portion (serrated portion). Take care in disconnecting because excessive force applied to glass may cause damage.
5. Disconnect the hole plug at left plate, remove four screws with a screw driver, then remove the left plate.
6. Loosen 4 screws on the right of terminal block located at the right top of the body frame with left side plate dismounted by use of Phillips screwdriver, and disconnect the heater lead terminal.
Cooling water outlet
Cooling water inlet
Distilled water outlet
Boiler water supply and drain port
Page 43
39
6. Maintenance Method
Washing of Distiller
7.
1) Disconnect the heater lead wire from grommet. Note that, do not bend or pull the heater lead wire more than necessary.
2) Remove the two screws of boiler securing band with a Phillips screwdriver, and take the boiler and condenser out of the body.
Grommet
3) Loosen the knurled screws (three) and remove the boiler and condenser.
Grommet
Connecting Hardware
Packing
Connection Port with Condenser
Boiler
Condenser
Washing of boiler
1. Adjust detergent liquid.
1) Prepare approx. 2 liters of hot water at 50 to 60C.
2) Add attached scale detergent (Orgazor) approx. 200g to hot water prepared in 1) and agitate well.
2. Seal the hose connection port at the bottom of boiler (boiler supply and drain port) by use of rubber stopper, etc.
3. Secure the boiler at a stable position to prevent washing liquid from spilling.
4. Pour in washing liquid through connection port with condenser with heater turned on. Most scale is removed in 4 to 5 hours approximately. Drain washing liquid in the boiler. If much scale is distiller deposited, pour in washing liquid newly, and repeat washing
1) When scale-removing work is finished, take the heater out of boiler and wash each of them
enough with city water. Here, in washing the heater with water, be sure to fill a larger beaker with water and wash the heater inside so that lead wire and its routing port are not wet by water. Avoid washing the heater directly with water from tap.
2) If solid scale distiller remains after washing by washing liquid, follow the remedy below:
Boiler: Scrub with brush etc. for removing.
Heater: Scrub with something soft such as wood piece or plastic. In this connection, remove scale on the heater uniformly in general, never leaving solid scale in part. In an extreme case, only such part has a great heat resistance, causing damage to the heater.
Knurled Screw
Page 44
40
6. Maintenance Method
Washing of Distiller
Washing of Condenser
1. Pour detergent liquid into the cooling pipe of condenser.
(See Page 39 "Washing of boiler" for formulating detergent liquid.)
Detergent Liquid
Hose Connection Port
2. If detergent liquid should flow out of hose connection port, seal with rubber stopper. Most fur can
be removed in 4 - 5 hours approximately.
3. Drain detergent liquid, and then wash enough with city water.
Handling of Detergent Liquid (also refer to Page 35 "Handling Precautions")
1. Wash the boiler and heater sooner. If the more scale is deposited, the more difficult is its
removal, which may cause decrease of distilled water sampling and damage to heater.
2. When washing is finished, drain detergent liquid out of the unit, and apply neutralization by
neutralizer (such as sodium hydroxide). In neutralization, check that it is neutral by use of pH test paper, etc. (Principal component of scale detergent: Sulfamic acid and pH of water solution: Acidic approximately 1)
3. In storing this detergent, seal the agent and store in cold and dark place avoiding high temperature
and humidity.
4. In handling this detergent, be sure to use protective tools (gloves, mask, and glasses).
5. When it is in contact with human body, wash it away with clean water.
6. Do not use empty container for beverage.
7. Do not allow detergent to directly flow into agricultural irrigation canal or fields because it causes
withering of rice crop.
Installation of boiler
1. Secure the boiler with the boiler securing band so that connection port of condenser is horizontal.
Check that the packing is contained in the cap nut, and then install the heater into the boiler with letters "YK-W-3" faced up.
Packing
Cap Nut
Heater
Boiler Securing Band
Connection Port with Condenser
Page 45
41
6. Maintenance Method
Washing of Distiller
2. Attach 4 heater lead terminals to the terminal block.
Heater Lead (wire diameter, thick)
Heater Lead (wire diameter, thin)
3. Install the left side plate on the body.
4. Insert the hose to the boiler water supply and drain port, and secure with the hose band.
Installation of condenser
1. Place packing in the connection port of boiler with condenser, and secure with connecting hardware
so that the boiler and condenser are placed in the same direction.
Inlet
Outlet
Boiler
Condenser
Connecting Hardware
2. Connect the hose respectively to the cooling water inlet, outlet, and distilled water outlet of
condenser.
3. Close the boiler water drain cock.
View from above
Boiler Water
Drain Cock
Open
Close
Page 46
42
6. Maintenance Method
Replacement of Heater
If the heater should be disconnected or damaged due to deposit of scale, replace it by the
procedure below. (Also refer to Pages 38 to 41"6. Maintenance Method " in working.)
1. Turn "OFF" the earth leakage breaker of this unit.
2. Close the tap.
3. Turn "OFF" earth leakage breaker, and when more than 30 minutes has passed, open the front
door of this unit, and open the boiler water drain cock.
4. Open the left side plate of the body, loosen the four screws on the right of the terminal block, and
disconnect the heater lead terminal.
5. Pull the heater lead out of the grommet.
6. Remove the cap nut of heater, and pull out the heater.
7. Remove the packing and cap nut from the damaged heater.
8. Install the packing and cap nut on the new heater. At that time, do not touch with bare hand in
order to prevent soiling by hand.
Packing
9. Install on the boiler so that "YK-W-3" mark of the heater is faced up.
10. Feed the heater lead wire through the grommet, check the heater lead wire attaching position, and
secure to the terminal block.
11. Mount the left side plate.
12. Close the boiler water drain cock.
13. Close the front door, and then open the tap.
14. Turn on the earth leakage breaker.
15. The standby screen appears at the display window of the control panel.
16. Perform calibrations by pressing POWER key while holding down PURE WATER key and
DISTILLED WATER key. (Refer to page 24.) (Calibration is performed for storing the standard temperature of the sensors in each heater under normal operation in the inner controller. Performing calibration detects errors when the temperature rises higher than the standard temperature + 20.)
17. During calibration, a message is displayed at the message window of the control panel.
18. After approx. 5 minutes passed after starting calibration, normal operation starts automatically.
Page 47
43
6. Maintenance Method
Washing of Water Supply Hose Filter
1. After turning off the earth leakage breaker
on this unit, turn the tap off, slide the sleeve to the direction of the arrow, then, remove the water supply hose from the connection port.
2. Remove the plug from ring as
turning.
3. Cleanse filter attached to the packing
with water.
4. Wash the filter with spray, etc.
5. Assemble by reversing the procedure.
Replacement of Hose
Be sure to use a hose specified by Yamato Scientific for replacement.
Distilled water tank drain method
1. Turn OFF the earth leakage breaker of the unit to “OFF.”
2. Take out thr hose in the lower portion on the backside of system.
3. There is a cap at the end of hose. Remove the cap and drain the distilled water tank.
4. When draining is over, be sure to reset the cap firmly. If not firmly provided, the can may
cause water leakage.
Connection Port Body
Ring Plug Socket Spacer Mounting Screw Packing Filter
One-touch Joint
④ ⑤ ⑥ ⑦ ①
Tap
Sleeve
Tap side
Water
Supply
Hose
Page 48
44
6. Maintenance Method
Replacing the TOC reduction UV lamp
1. Turn OFF the earth leakage breaker of the unit.
2. Close the tap.
3. Open the door of the main body. (See Photo 1)
Photo Photo 2
4. Remove the pre-treatment cartridge and ion exchange resin (CPC-N).
5. Remove a plate fixing the ion exchange resin (CPC-N) and remove a cover plate of the TOC reduction UV lamp.See Photos 2 and 3
Photo 3
6. Remove screws at four points on top of the TOC reduction UV lamp.
7. Raise the socket and remove the lamp.
8. Remove the lamp from the socket.
9. For installation, reverse the above procedure.
Never touch the quartz glass directly during replacement of the UV lamp.
TOC reduction UV lamp unit
Deaeration menbrance module
Vaccum pump for deaeration menbrance module
Page 49
45
6. Maintenance Method
Checking if the TOC reduction UV lamp goes ON
Checking if the TOC reduction UV lamp goes ON.
Open the front door of the unit. There are the switch to turn ON the TOC reduction UV lamp of Photo 1 and the TOC reduction UV lamp ON glow on top of the tank.
Normally, the TOC reduction UV lamp does not go ON unless the TOC reduction system is activated. Checking of the TOC reduction UV lamp and TOC reduction UV lamp ON glow can be made with ease by turning ON the TOC reduction UV lamp ON switch.
:The TOC reduction UV lamp ON switch is automatically reset to OFF when
released.
TOC reduction UV lamp switch
TOC reduction UV lamp ON glow
Page 50
46
6. Maintenance Method
Replacing the deaeration membrane module
1. Turn OFF the earth leakage breaker of the unit.
2. Close the tap.
3. Open the door of the main body. (See Photo 1)
Photo Photo 2
4. Remove the pre-treatment cartridge and ion exchange resin (CPC-N).
5. Remove a plate fixing the ion exchange resin (CPC-N).See Photos 2 and 3
Photo 3
6. Remove the piping connected to the deaeration membrane module as well as fixing screws.
7. For installation, reverse the above procedure.
TOC reduction UV lamp unit
Deaeration menbrance module
Vaccum pump for deaeration menbrance module
Page 51
47
6. Maintenance Method
Vacuum pump
1. Maintenance Check as follows at least once every three days during operation:
(1) Check for any abnormal sound. (2) Check if the pump is abnormally hot. (3) Check if exhaust is made correctly.
2. Regular inspection
(1) List of consumables
Parts
Quantity
Material
Reference life
Diaphragm
2
Synthetic rubber (EPDM)
10,000 H
Suction exhaust
valves
4
Synthetic rubber (FPM)
10,000 H
Head gasket
2
Synthetic rubber (EPDM)
10,000 H
Bearing
1 set
10,000 H
(2) Guidelines for replacement
Diaphragm Replace when the surface shows fine cracks, wear, hardening, deformation. Suctionexhaust valves Replace if they show any deformation, chipping, bend, etc. Head gasket Replace if it shows any hardening, crack, elongation. Bearing Request repair if grease shortage, abnormal sound, or abnormal motor vibration (chattering)
is detected.
3. Replacing the consumables CautionBe sure to turn OFF power supply to the unit before proceeding to replacement of
consumables. When replacing, refer to the exploded diagram. Depending on the diassembling portion, special tool may be necessary to handle a part of components.
(a) Replacing the diagram
Remove four cross-recessed pan-head machine screwsM4×12(with SW) of the pump head . Remove the pump head and remove one cross-recessed pan-head machine screw
M5×12 of the diaphragm holding plate .
Replace a total of two diaphragms , one for each pump head, with the new ones.
During reassembly, apply small amount (enough to cover three to four threads at the end) of an agent to prevent loosening (Locktight 242, etc.) to cross-recessed pan-head machine screws(M5×12)④. Tighten screws with a torque of 4.0 4.5
Nm.
Page 52
48
6. Maintenance Method
Vacuum pump
(b) Replacing the head gasket
Remove one cross-recessed pan-head machine screw M4×6)⑦ of head cover . There is a black head gasket under the head cover. Remove this gasket and replace the new one. (※) There are total of two hea gasket , one for each pump head.
Reassemble the head gasket and head cover in this order to the pump head.また、 During reassembly, apply small amount (enough to cover three to four threads at the end) of an agent to prevent loosening (Locktight 242, etc.) to cross-recessed pan-head machine screws(M4x6). Tighten screws with a torque of 1.2 ~ 1.3 N・m.
(c) Replacing suctionexhaust valvesspecial tool necessary
Remove one cross-recessed pan-head machine screw M4×6)⑦ of head cover , and remove the head gasket . Remove four cross-recessed pan-head machine screws M4×12(with SW))② of the pump head ①. Remove the screwM3×4)⑨ and valve holder on both sides of pump head and replace suctionexhaust valves . (※) There are a total of four suctionexhaust valves , two for each pump head.
During reassembly, apply small amount of an agent to prevent loosening (Locktight 242, etc.) to cross-recessed screwsM3×4)⑨. Tighten screws with a torque of 0.2 0.25 Nm.
Exploded diagram
Page 53
49
7. Long storage and disposal
When not using this unit for long term
If this unit is to be put out of service for a long time, be sure to turn of the earth leakage
breaker of this unit for safety, and close the tap. Water in the boiler and distilled water tank, if stored as it is, will deteriorate in quality due to generated bacteria or algae. Drain water by the procedure below:
Boiler Water Drain
1. In draining boiler water, turn off the earth
leakage breaker, ensure that the tap is closed, then wait for 30 minutes or more, and open the front door.
2. Open the boiler water drain cock.
3. Make sure that all water in the boiler and
float cylinder is drained.
4. Be sure to close the boiler water drain cock.
If boiler water drain cock is opened in next use, water is not fed into the boiler, and distillation is not started.
When the distilled water has been left stored in the tank for two weeks or more, drain the tank content completely and start operation. Start using the distilled water when the water is ready for sampling and when the specific resistance becomes 1.2MΩ・cm or more and after operation for one hour or more.
Close
Boiler Water Drain Cock
Open
Page 54
50
7. Long storage and disposal
WARNING!
1. When disposing…
Dispose of the unit as bulky waste.
When disposing, avoid a place visited by playing children frequently.
2. When the unit is out of service at night and on holidays
Turn off the earth leakage breaker. Be sure to close the tap. Fluctuation of city water pressure may cause unexpected accident such as water leakage. In use in winter at a severely cold place, beware of freezing in the tank, boiler, condenser,
etc. while system is stopped.
Considerations on disposing
Environmental protection should be considered
We request you to disassemble this unit as possible and recycle the reusable parts considering to the environmental protection. The feature components of this unit and materials used are listed below.
Component Name
Material
Main components of exterior
Exterior
Made of iron, bonded steel plate, melamine resin baking finish
Exterior rear plate
Made of iron, bonded steel plate, melamine resin baking finish
Door
Made of iron, bonded steel plate, melamine resin baking finish
Door rear plate
Stainless steel plate SUS 304
Mounting plate (painted)
Made of iron, bonded steel plate, melamine resin baking finish
Mounting plate (unpainted)
Stainless steel plate SUS 304
Electric parts mounting plate
Aluminum
Hinge
Stainless steel plate SUS
Rubber foot
Synthetic rubber
Mounting tabs
Stainless steel plate SUS 304
Production plates
Polyester
Main components of water circuit system
Boiler
Hard glass
Condenser
Hard glass
Float cylinder
Polypropylene
Pure water tank
Polyethylene
Drain port
Polypropylene
Float cylinder branch pipe
Polypropylene
Electrode holder
Polypropylene
Water sampling port
Polypropylene
Control panel
ABS resin
Water sampling table
ABS resin
Page 55
51
7. Long storage and disposal
Considerations on disposing
Component Name
Material
Main components of water circuit system
Resin cylinder case
Polypropylene
Ion exchange resin
Polystyrene Resin
Water quality gauge electrode
Titanium
Heater
Ceramic
Heater mounting nut
Teflon
TOC holder
Hard glass
UV lamp
Hard glass
Hard tube
Hard glass
TOC holder cover
Polyethylene
Deaeration membrane module
Polycarbonate
Main components of piping system
Water supply hose
Vinyl chloride
Drain hose
Ethylene propyne
Hose (transparent)
Vinyl
Hose (milky transparent)
Silicon
Hose clamp
Polyacetal
Hose nipple (resin black)
Polyamide
Hose nipple (resin white)
Polypropylene
Hose nipple (metal)
Brass
Main components of electric system
Pump
Casing: Polypropylene Impeller: Polypropylene Magnet: Ferrite magnet Motor case: Iron Rotor: Iron
Solenoid Valve
Made of metal: Body, brass Made of resin: Body, polyacetal
Float Switch
Polypropylene
Power Cord, Wiring Material, etc.
Wiring material and board coated by synthetic rubber and resin
Page 56
52
8. In the Event of Failure…
Failure indication and Its Contents
Alarm
Indication
Condition
Corrective measures
Water cutoff alarm
Feed Water Alarm
Check
Flow & Pressure
When raw water pressure decreases (0.5kg f/cm2), or when water cutoff occurs
Check raw water.
Pretreatment filter exchange notification
It's time to exchange Hollow Filter
When 6 months passed under continuous energizing condition
Change the pretreatment filter. Press and hold the reset switch on backside of the door till a beep is heard. Then, the alarm indication is reset.
Ion-exchange resin CPC-N exchange notification
Water Quality NG
It's time to exchange
Ion-EXCH Resin-N
When the conductivity of pure water exceeds 1×10-4 S/m
Change the ion-exchange resin. Press and hold the reset switch on backside of the door till a beep is heard. Then, the alarm indication is reset.
Ion-exchange resin CPC-H exchange notification
Water Quality NG
It's time to exchange
Ion-EXCH Resin-H
When the conductivity of pure water exceeds
0.2×10-4 S/m
Change the ion-exchange resin. Press and hold the reset switch on backside of the door till a beep is heard. Then, the alarm indication is reset.
Membrane filter for distilled water exchange notification
It's time to exchange
Distilled Water
Membrane Filter
When 3 months passed under continuous energizing condition
Change the membrane filter for distilled water. Press and hold the reset switch on backside of the door till a beep is heard. Then, the alarm indication is reset.
Membrane filter for pure water exchange notification
It's time to exchange
Ion-EXCH Water
Membrane Filter
When 3 months passed under continuous energizing condition
Change the membrane filter for pure water. Press and hold the reset switch on backside of the door till a beep is heard. Then, the alarm indication is reset.
TOC reduction UV lamp exchange message
It’s time
to exchange
TOC-d UV Lamp.
When concecutive power application hours of 4000 hours have passed
Replace the TOC reduction UV lamp. Keep pressing the reset switch on the door backside for an extended period of time. The confirmation sound is issued and the alarm display is reset.
Page 57
53
8. In the Event of Failure…
Failure indication and Its Contents
Deaeration membrane module exchange message
It’s time
to exchange
MHF
When concecutive water supply of one year has passed
Replace the deaeration membrane module. Keep pressing the reset switch on the door backside for an extended period of time. The confirmation sound is issued and the alarm display is reset.
Maintenance time notification
It's time
to maintenance
Service Call
When 3 years passed under continuous energizing condition
Perform maintenance of the unit. Press and hold the reset switch on backside of the door till a beep is heard. Then, the alarm indication is reset.
Page 58
54
8. In the Event of Failure…
Failure indication and Its Contents
Error
Indication
Cause
Symptom
Countermeasur
e
Controller
error
Controller Error
Service Call
When the setting value which is memorized in the memory chip cannot be read properly, or when an abnormal value is displayed
All controls of the heater and solenoid valve are turned OFF.
Turn the breaker on again. If the trouble persists, please call our customer service center.
Water
leakage
error
Leak Error
Service Call
When the resistance value of the water leakage sensor input becomes less than the water leakage error judgment value
Turn the breaker off and check the piping parts. For details, refer to page 56.
Overheat of
heater
Over Heat Error
Service Call
When temperature at the heater excesses the error judgment value, or when breakage or shortage occurs on the temperature sensor
Change the heater.
Burnout of
heater
Heater Error
Service Call
When temperature of the heater did not rise after certain time passed during distillation
Change the heater.
Tank water level meter
error
Tank Water
Sensor Error
Service Call
When the condition of the float contacting points in the tank water level meter becomes abnormal
Change the float switch.
Pure water
conductivity
meter (N)
error
Ion-Exchange-N
Sensor Error
Service Call
When the state of breakage or shortage of the thermistor sensor for pure water conductivity gauge continues longer than error judgment time
Change the pure water conductivity sensor.
Pure water
conductivity
meter (H)
error
Ion-Exchange-H
Sensor Error
Service Call
When the state of breakage or shortage of the thermistor sensor for pure water conductivity gauge continues longer than error judgment time
Change the pure water conductivity sensor.
Distilled
water
conductivity
meter error
Distilled Water
Sensor Error
Service Call
When the state of breakage or shortage of the thermistor sensor for distilled water conductivity gauge continues longer than error judgment time
Change the distilled water conductivity sensor.
Page 59
55
8. In the Event of Failure…
Failure indication and Its Contents
Others
Indication
Cause
Symptom
Countermeasure
Boiler drain error
Boiler Water
Waste Error Service Call
When the heater operation water level input kept OFF even if the time for required to evaluate the boiler water level error passed after starting water supply to the boiler
All controls of the
heater and
solenoid valve
are turned OFF.
Check the solenoid valve for drain and drain path.
Coolant error
A Coolant Error
Service Call
When the state that became boiler water overflow input ON in the float pipe continued longer than coolant error judgment time
Check the coolant solenoid valve and the coolant path.
Water level error of boiler
Boiler Water
Level Error
Service Call
When the heater operation water level input kept OFF even if the time for required to evaluate the boiler water level error passed after starting water supply to the boiler
Check the feedwater solenoid valve and feedwater path whether manual drain cock is opened or not.
Water level meter of boiler error
Boiler Water
Sensor Error
Service Call
When the condition of the float contacting points in the float pipe becomes abnormal
Change the float switch.
Circulation pump error
PUMP Erroe
Service Call
When the time has elapsed for the period necessary to determine error during ion water circulation
Turn ON the breaker again and sample the ion water for about one minute. If error cannot be
canceled, request repair.
Distilled water sampling pump error
DW PUMP Error
Service Call
When the time has elapsed for the period necessary to determine error during sampling of distilled water
Check the distilled water pump and the sampling route. Check the membrane filter for clogging. If error cannot be canceled, request repair.
When an error occurs, confirm the error message shown at the message window, and call the shop from
which you made a purchase or our customer support center.
Page 60
56
8. In the Event of Failure…
Remedy for Trouble
Countermeasure against water leakage error
1. Turn off the earth leakage breaker at left surface of the main unit.
2. After repairing the defective area, wipe water off at the bottom in the unit, and dry thoroughly for re-starting the operation.
3. If the electrode is removed, make sure to reinstall it.
4. Close the door.
5. Turn on the earth breaker, and press POWER key. Now the defect is repaired, and the unit starts normal operation.
<How to remove the lower front panel of WG270>
1. Remove the screws at left/right positions of the lower front panel.
2. Pull the panel frontward. Be careful of the cable of the drain switch specially used for water sampling table.
3. Disconnect the cable of the drain switch for water sampling table. Then, remove the lower front panel. For installation, perform the above steps in reverse order (from 3 to 1).
Water leakage detection electrode
Page 61
57
8. In the Event of Failure…
Trouble Shooting
When a trouble occurs
Condition
Check the following
Water is not supplied.
Defect of raw water supply solenoid valve. Insufficient city water pressure or water failure. Defect of pressure switch. Clogging of pre-treatment cartridge.
Water supply does not stop.
Defect of float switch. Defect of raw water supply solenoid valve.
No water is supplied to boiler.
Defect of boiler water supply solenoid valve. Defect of float switch. Clogging of pre-treatment cartridge. Opened boiler water drain cock.
Heater does not turn on.
Defect of float switch. Break in heater.
Cooling water does not flow.
Defect of cooling water solenoid valve.
The initial accumulated water is not drained.
Initial boiled water is not drained.
Distilled water is not accumulated.
Defect of initial boiled water drain solenoid valve. Defect of piping.
Distilled water is not stored.
Defect of float switch.
Water is not sampled.
Defect of Ion exchanged water/Distilled water sampling solenoid valve. Defect of distilled water sampling pipe. Defect of piping.
Measures in emergency
Error indications/causes
This unit has a self-diagnosis function. When a trouble occurs during operation, or when a failure occurs on the unit, an error message is displayed on the control panel. Turn off the earth leakage breaker when an alarm occurs, close the tap.
For details of error indications, refer to "Failure indication and Its Contents (pages 52 and 55)".
Measures
When the following error signs appear, memorize the sign and turn the tap off immediately. If an error occurs, part change or unit check becomes required. Please call the shop from which you made a purchase or our customer support center. In that case, please notify them of the error sign.
Page 62
58
9. After-sales Service
When requesting the repair
Guarantee card (attached separately
If any error should occur, record the error message and stop operation immediately. Turn OFF power supply and unplug the power cord from the receptacle. Then contact the shop from which you have purchased the product or our office.
(Data necessary for information)
Product type name
Serial No.
Year/month/day of purchase Check to find the date in the guarantee
card or the nameplate attached to the
unit. (See “Description and Function of
Each Part” in pages 13 and 14.)
Contents of trouble (as detailed as possible
Be sure to present the guarantee card when our service personnel visits your site.
The guarantee card will be delivered from the shop from which you have purchased our product or our office. Confirm that the shop name and the date of purchase are entered and read the card carefully. Keep the card with care.
The guarantee period is one year from the “date
of purchase.” Free repair may be available
according to the content of the card.
For repair after expiration of the guarantee period, consult the shop from which you have purchased our product or our office. If the repair can help maintaining the function, we will provide the repair service with charge if so requested by the customer.
Minimum hold period of performance parts
for repair
The minimum holding period of repair parts for
this unit is seven years after termination of the production. Repair parts mean those necessary to maintain the performance of the product.
Page 63
59
10. Specifications
Model
WG270
Feedwater method
One-touch coupler connecting resin hose/free hose connecting
Drain method
Left/right selection connecting method/hose connecting
Performance
Purified water
Distilled water/Ion exchange water
Production of distilled water
Approx. 1.8 L/h
Collection of distilled water
0.5 to 1.0 L/min
Collection of ion exchange water
0.5 to 1.0 L/min
Setting range of capacity
0.1 to 20 L
(Measured amount/continuous water collection)
Configuration
Condenser
Super hard glass
Heater
Ceramic heater
1.4kW
Raw water side filter
Pre-treatment cartridge (PWF-1),
Activated carbon + Hollow yarn film 0.1μm
Ion exchange resin cartridge
Ion exchange resin cartridge (CPC-N)
High-purity catrige (CPC-H)
Membrane filter
0.1μm×2
Water leakage detection
Water leakage detector forcefully shut off the feedwater solenoid valve
when water leakage is detected.
Distilled water storage tank
20 PE tank
Water sampling table
Slide-out type
Load-bearing capacity:10kg
For 5 beaker with handle
Water level detection
Lead switch, five level detection (also used for distillation control
function)
Standard
Raw water pressure range
0.5 to 5×100 kPa (0.5 to 5 kgf/cm2)
Ambient temp.
5℃~35
Power supply (50/60 Hz)
220V AC 7A
External dimension
(*1)
(Width X Depth X Height)
600×660×780 mm
Weight
Approx.55 kg
Accessories
WG270 : Water supply hose (with connection port), drain hose
assembly, and connection hose 1 respectively Operation manual: This manual Guarantee card: 1 Hose clamp: 1 Scale washing agent (1kg): 1 Pre-treatment cartridge: 1 Ion exchange resin cartridgeCPC-N: 1 High-purity catrige CPC-H: 1 Sealing tape: 1 Membrane filter: 2
The protrusion area is not included in the outside dimension.
Page 64
60
11. Wiring Diagram
WG270
Page 65
61
11. Wiring Diagram
WG270
Symbol
Part name
Symbol
Part name
ELB
Earth leakage breaker
X
Main relay
T1,T2,T3
Terminal block
MV1
Raw water solenoid valve
H
Heater
MV2
Boiler water supply solenoid valve
OH
Temperature sensor
MV3
Cooling water solenoid valve
E1
Distilled water quality gauge
MV4
Initial accumulated water drain solenoid valve
E2
Ion exchange water quality gauge CPC-N
MV5
Boiler drain solenoid valve
E3
Ion exchange water quality gauge CPC-H
MV6
Distilled water sampling solenoid valve
WL
Water leakage detector
MV7
Ion exchange water sampling solenoid valve
FL1
Distilled water flow gauge
MV9
Three-way solenoid valve for vaccum
FL2
Ion exchange water flow gauge
P1
Distilled water circulation/sampling pump
Fs-1
Control float switch
P2
Deaeration membrane vacuum pump
Fs-2
Control float switch
P3
Ion water circulation pump
Fs-3
Tank water level float switch
UV
TOC reduction UV lamp
SW1
Reset switch
UV-B
Ballast
SW2
Drain switch
UV-G
Glow
SW3
UVlamp turn-on switch
PIO
Display board
Ps
Pressure switch
CONT
PLANAR board
SSR
Solid state relay
TF
Transformer
Page 66
62
12. Replacement Parts List
WG270
Symbol
Parts
Code No.
Specifications
Manufacturer
ELB
Earth leakage breaker
2060050002
BJS2032N 20A
Panasonic
T1
Terminal board
LT00035672
MKH-250ABC-4P
Terminal
T2
Terminal board
A0050115
TB-20C 4P
Sakazume
Electric
T3
Terminal board
LT00035673
MKH-250ABC-5P
Terminal
HOH
Heater (temperature sensor)
2420016003
YK-W-3
Kowa Dennetsu
E1
Distilled water quality meter
1011890001
For WG252/220
Yamato Scientific
E2,E3
Pure water quality meter
5130000001
For WG240/260
Yamato Scientific
Fs-1
Float switch for control
LT00014440
WA050514-1
Yashima Sokki
Fs-2
Float switch for control
LT00014441
WA050514-2
Yashima Sokki
Fs-3
Tank water level float switch
LT00019425
YF4-2040
Yashima Sokki
SW1
Reset switch
2010010014
A2A-4W
Omron
SW2
Drain switch
LT00014520
AH164-LS1 1H1
Fuji Electric
SW3
UV lamp ON switch
LT00023393
M-2025
Nihon Kaiheiki
X
Main relay
2050000056
G7L-1A-TUB 100V
Omron
MV1
Raw water solenoid valve
LT00014451
AG3X-A300-100V
CKD
MV2
Boiler water supply solenoid valve
LT00014450
J241-811
CKD
MV3
Cooling water solenoid valve
LT00014453
AB2X-1242
CKD
MV4
Initial distilled water drain solenoid valve
A0060104
J241-120
CKD
MV5
Boiler drain solenoid valve
A0060109
J244-023
CKD
MV6
Distilled water sampling solenoid valve
A0060104
J241-120
CKD
MV7
Pure water sampling solenoid valve
A0060104
J241-120
CKD
MV9
Three-way solenoid valve for opening to the atmosphere
LT00014451
AG3X-A300-100V
CKD
P1
Distilled water sampling pump
LT00005462
MD-20RZ-N
Iwaki
P2
Deaeration membrane vacuum pump
LT00019426
DAP-6D
Ulvac
P3
Ion water circulation pump
LT00035062
WP-LM9 YR-2 (100V)
Iwaki
WL
Water leakage sensor
WG55005328
Lead wire 2,7m
Yamato Scientific
SSR
Solid state relay
2160000035
TRS5225
Toho
Ps
Pressure switch
2040040001
ST-B-BR1-N2
Sanyo Keiki
PIO
Display board
LT00019738
WG270
Yamato Scientific
CONT
Planer board
LT00013591
WG250
Yamato Scientific
FL1
Distilled water flow meter
1260010005
FHK-33
Yamato Scientific
Page 67
63
12. Replacement Parts List
FL2
Pure water flow meter
1260010005
FHK-33
Yamato Scientific
UV
TOC reduction UV lamp
LT00019420
SGL-180T4U Z
Nippo Electric
UV-B
Stabilizer
LT00019424
FBAB-11L-US9
NEC
UV-G
Glow
LT00024700
FE-7E/X
Matsushita
Electric Works
TF
Transformer
2180020010
AD21-02KB2
Toyozumi
Page 68
64
13. Consumable parts
WG270
Parts
Replacement timing
Quantity
Pre-treatment cartridge
About 6 months
1
Ion exchange cartridge(CPC-N)
About 600 liters
1
Ion exchange cartridge(CPC-H)
About1200 liters
1
Distilled water membrane filter
About3 months
1
Pure water membrane filter
About3 months
1
Deaeration membrane module
About1 year
1
Consumables of vacuum pump
10000 hours
1.Diaphragm
2
2.Suctionexhaust valve
4 3.Head basket
2
4.Bearing
1
TOC reduction UV lamp
About4000 hours
1
Glow
About1 year
1
Page 69
65
14. List of Dangerous Substances
Never process any explosive, flammable samples and also samples contained with those substances.
Explosive
Substance
Nitroglycol, Glycerine trinitrate, Cellulose Nitrate and other explosive nitrate
esters
Trinitrobenzen, Trinitrotoluene, Picric Acid and other explosive nitro compounds
Acetyl Hydroperoxide, Methyl Ethyl Ketone Peroxide, Benzoyl Peroxide and other
organic peroxides
Metallic Azide, including Sodium Azide, etc.
ExplosiveSsubstanc
es
Metal “Lithium” Metal “Potassium” Metal “Natrium” Yellow Phosphorus Phosphorus Sulfide Red PhosphorusPhosphorus Sulfide Celluloids, Calcium Carbide (a.k.a, Carbide)Lime PhosphideMagnesium
Powder Aluminum Powder Metal Powder other than Magnesium and Aluminum
Powder Sodium Dithionous Acid (a.k.a., Hydrosulphite)
Oxidizing Substances
Potassium Chlorate, Sodium Chlorate, Ammonium Chlorate, and other chlorates
Potassium Perchlorate, Sodium Perchlorate, Ammonium Perchlorate, and other
perchlorates
Potassium Peroxide, Sodium Peroxide, Barium Peroxide, and other inorganic
peroxides
Potassium Nitrate, Sodium Nitrate, Ammonium Nitrate, and other nitrates
Sodium Chlorite and other chlorites
Calcium Hypochlorite and other hypochlorites
Flammable Substances
Ethyl Ether, Gasoline, Acetaldehyde, Propylene Chloride, Carbon Disulfide, and
other substances with ignition point at a degree 30 or more degrees below zero.
n-hexane, Ethylene Oxide, Acetone, Benzene, Methyl Ethyl Ketone and other
substances with ignition point between 30 degrees below zero and less than zero.
Methanol, Ethanol, Xylene, Pentyl n-acetate, (a.k.a.amyl n-acetate) and other
substances with ignition point between zero and less than 30 degrees.
Kerosene, Light Oil, Terebinth Oil, Isopenthyl Alcohol(a.k.a. Isoamyl Alcohol),
Acetic Acid and other substances with ignition point between 30 degrees and less than 65 degrees.
Combustibl
e Gas
Hydrogen, Acetylene, Ethylene, Methane, Ethane, Propane, Butane and other gases combustible at 15 at one air pressure.
(Quoted from the separate table 1 in Article 6, the enforcement order of the Industrial Safety and
Health Law)
Page 70
Page 71
Responsibility
Please follow the instructions in this document when using this unit. Yamato Scientific has no responsibility for the accidents or breakdown of device if it is used with a failure to comply. Never conduct what this document forbids. Unexpected accidents or breakdown may result in.
Note
The contents of this document may be changed in future without notice.
Any books with missing pages or disorderly binding may be replaced.
Instruction Manual for Water Purifier, Super Auto Still
Model WG270 First Edition June 1, 2017
Yamato Scientific America Inc.
925 Walsh Avenue, Santa Clara, CA 95050 Phone: 800.292.6286 / 408.235.7725
http://www.yamato-usa.com