Yamatake HGC303 User Manual

Heat Value Gas Chromatograph
Model: HGC303
User’s Manual
HGC000001000P
CM2-HGC100-2001
First issue: Apr. 2001, Rev.12: Oct. 2004
Copyright, Notices and Trademarks
Printed in Japan - © 2001-2004 Yamatake Corporation ALL RIGHTS RESERVED
In no event shall Yamatake Corporation be liable to anyone for any indirect, special or consequential damages. The information and specifications in this document are subject to change without notice.
®
Microsoft ration.
Windows™ and Excel™ are trademarks of Microsoft Corporation.
, Windows®, and MS-DOS® are registered trademarks of Microsoft Corpo-
NI-FBUS™ is a trademark of NATIONAL INSTRUMENTS CORPORATION
®
.

Safety

Safety symbols

Be sure to correctly operate the model HGC303 while strictly observing the safety precautions provided in this manual-especially the Warnings and Cautions indicated by the symbols as shown below.
The descriptions of the Warning and Caution signs used in this manual are as follows.
The Warning sign means that serious personal injury, even death, could result if the instructions given are not strictly observed.
The Caution sign means that light personal injury and/or equipment damage could result if the instructions given are not correctly observed.
WARNING
CAUTION
Model HGC303 - Heat Value Gas Chromatograph i
Safety Yamatake Corporation

Hazardous Areas Certifications

The model HGC303 complies with the type of protection, which based on the following standards.
(1) ISSeP/ATEX Flameproof Certification
0344
II 2 GD
EEx d II C T6 -10°C<Tamb<+50°C IP65
ISSeP01ATEX003X
Special conditions for safe use (symbol X)
The fastening screws of this apparatus are made of stainless steel and have a yield stress of 500 N/mm².
(2) FM Explosionproof / Flameproof Approval
Explosionproof for Class I, Division 1, Groups C and D, T4 Flameproof for Class I, Zone 1, AEx d IIB T4 Dust-ignitionproof for Class II and III, Division 1, Groups E, F and G, T4
(3) TestSafe Flameproof Certification
Ex d IIC T6 DIP A21 TA T6 The vent plug must be removed and 1/4 NPT fitting must be connected for piping.
Therefore, the model HGC303 can be installed in various hazardous locations. However, an explosion-protected electrical apparatus requires special care. Please read all instruction and safety notes before installation.
WARNING
NEVER open the terminal box cover while the model HGC303 is energized in a hazardous location.
CAUTION
Use the model HGC303 only in an ambient temperature of -10 to 50°C (14 to 122°F)
CAUTION
Take precautions to prevent corrosion, deformation or damage to the housing or terminal box cover.
CAUTION
See that all conduits are properly sealed. Otherwise, the model HGC303 cannot withstand the pressure that can result from explosion of an explosive gas inside the housing. Also, the model HGC303 cannot prevent the explosion of any external explosive gas.
ii Model HGC303 - Heat Value Gas Chromatograph
Yamatake Corporation Safety

(1) Installation for ISSeP/ATEX Flameproof Apparatus

1. General
1.1 The apparatus protected by the flameproof enclosure in accordance with EN 50018 can be installed in such hazardous areas, for which the apparatus has been certified, as an explosive atmosphere containing flammable substances in the form of gas, vapour, mist or dust may be present.
~Note The apparatus has been certified to comply with EN 50281-1-1 (dust
ignition protection).
1.2 The apparatus enclosure must be kept closed in the hazardous areas when the apparatus is energized because the internal circuit of the apparatus is capable of
igniting the explosive atmosphere. (Never connect any hand-held communicator to the apparatus terminals by opening the cover, except while no explosive atmosphere is present.)
1.3 It is required to connect the external earthing terminal of the apparatus to the equipotential bonding system which includes protective conductors, metal conduits, metal cable sheaths, steel wire armouring and metallic parts of structures, but does not include the neutral conductors of the power systems.
~Note The protective conductor to which exposed conductive parts of equipment
(machines, apparatus, devices, components and instrumentation thereof) are connected, must be separated in the hazardous area from the neutral conductor, and must be connected to the power systems earth point in the non-hazardous area, if the power system is directly earthed.
For external earthing and bonding of the apparatus it is recommended to use a cable lug so that the conductor is secured against loosening and twisting and that the contact pressure is permanently secured.
1.4 Either cable systems (cable entry systems) or conduit systems can be employed for wiring of the apparatus in the hazardous areas (see 2 or 3).
1.5 Non-sheathed single core cables are not permitted for live conductors unless they are installed inside enclosures or conduit systems.
1.6 Conduits and, in special cases, cables (for example, where there is a pressure difference) must be sealed so as to prevent the passage of the explosive atmosphere.
1.7 Further information concerning installation and maintenance of apparatus is given by relevant clauses of the following documents.
EN 60079-14 Electrical apparatus for explosive gas atmospheres
Part 14: Electrical installations in hazardous areas other than mines
EN 60079-17 Part 17: Inspection and maintenance of electrical installations in
hazardous areas.
EN 50281-1-2 Electrical apparatus for use in the presence of combustible dust
Part 1-2: Electrical apparatus protected by enclosures
-- Selection, installation and maintenance
Model HGC303 - Heat Value Gas Chromatograph iii
Safety Yamatake Corporation
2. Cable systems
2.1 Thermoplastic sheathed cables, thermosetting sheathed cables, or elastomeric sheathed cables can be selected for fixed wiring in the hazardous areas.
2.2 Flameproof cable entry devices (cable glands) certified to comply with EN 50018 and appropriate to the type of cable employed, must be used for the connection of cables to the apparatus.
3. Conduit systems
For conduit systems, relevant national standards or codes of practice are followed prior to the following recommendations.
3.1 Screwed heavy gauge steel, solid drawn or seam welded conduit, or flexible conduit for protection of cables in explosive atmospheres (see ISO 10807) can be selected for fixed wiring in the hazardous areas.
3.2 Conduit must be threaded for connection to permit the full engagement of five threads.
3.3 Either conduit entry devices or sealing devices such as stopping boxes are provided at the wall of the apparatus enclosure to limit the pressure piling effect and to prevent hot gases from entering the conduit system from the enclosure containing a source of ignition. Each type of both the devices must be certified to comply with EN 50018.
3.4 The stopping boxes, if used, are filled with a compound which does not shrink or setting and is impervious to, and unaffected by, chemicals found in the hazardous area. The depth of the compound in the stopping box is at least equal to the internal diameter of the conduit, but in no case less than 10 mm.
3.5 When the conduit contains three or more non-seathed single or multi-core cables, the total cross-sectional areas of cables, including insulation, are not more than 40% of the cross-sectional area of the conduit.
4. Installation in explosive atmospheres caused by air / dust mixtures
4.1 Conduit or cable glands, if employed to connect cables to the apparatus, must be selected and used in such a way that an IP6X protection (dust-tight) is guaranteed.
4.2 It is recommended to maintain the apparatus so that the dust layer will not exceed a thickness of 5 mm.
~Note Where the ignition temperature of a dust layer up to 5 mm thickness is
equal to, or higher than, the value that is obtained by adding 75K to the maximum surface temperature of the enclosure “T...°C” as marked on the apparatus, the apparatus is incapable of causing ignition of the dust layer. (T...°C is based on the maximum ambient temperature)
iv Model HGC303 - Heat Value Gas Chromatograph
Yamatake Corporation Safety
(2) Installation for FM Explosionproof / Flameproof Apparatus (in
accordance with NEC)
CAUTION
Install the apparatus only in hazardous (classified) locations for which the apparatus has been approved.
Seal each conduit entering the apparatus enclosure within 18 in.(457 mm) from the enclosure.
Do not open the apparatus enclosure when an explosive atmosphere is present.
Figure S-1 An example of conduit seal (with stopping plug)
Figure S-2 An example of conduit seals (without stopping plug)
Model HGC303 - Heat Value Gas Chromatograph v
Safety Yamatake Corporation
1. Class I, Division 1 locations
1.1 Wiring methods
Threaded rigid metal conduit, threaded steel intermediate metal conduit, or Type MI cable with termination fittings approved for the location, can be
employed
Threaded joints must be made up with at least five threads fully engaged.
• Boxes, fittings, and joints must be approved for Class I, Division 1.
1.2 Sealing
• Each conduit entering the apparatus enclosure is required to be sealed
within 18 in. (457 mm) from the enclosure.
• The sealing of each conduit can be provided with a sealing fitting approved for
class I locations.
Sealing compound must be approved and must not have a melting point of less
than 93° (200°F).
• The minimum thickness of the sealing compound should not be less than the trade size of the conduit and, in no case, less than 5/8 in.(16 mm).
• Splices and taps cannot be made in the fittings.
2. Class I, Division 2 locations
2.1 Wiring methods
Threaded rigid metal conduit, threaded steel intermediate metal conduit, enclosed gasketed busways, or Type PLTC cable in accordance with the
provisions of remote-control, signaling, and power-limited circuits (see NEC, Article 725), or Type ITC cable in cable trays, in raceways, supported by messenger wire, or directly buried where the cable is listed for this use; Type MI, MC, MV, or TC cable with approved termination fittings can be employed.
• Boxes, fittings, and joints are not required to be explosionproof.
2.2 Sealing
• Each conduit entering the apparatus enclosure is required to be sealed as shown in 1.2.
3. Class II, Division 1 locations
3.1 Wiring methods
Threaded rigid metal conduit, threaded steel intermediate metal conduit, or Type MI cable with termination fittings approved for the location, can be
employed.
• Boxes and fittings must be dusttight.
3.2 Sealing
• Where a raceway provides communication between the apparatus enclosure and an enclosure that is not required to be dust-ignitionproof, suitable means must be provided to prevent the entrance of dust into the dust-ignitionproof enclosure through the raceway. One of the following means can be used: (1) a permanent and effective seal; (2) a horizontal raceway not less than 10 ft (3.05 m) long; or (3) a vertical raceway not less than 5 ft (1.52 m) long and extending downward from the dust-ignitionproof enclosure.
• Seals are not required to be explosionproof.
vi Model HGC303 - Heat Value Gas Chromatograph
Yamatake Corporation Safety
4. Class II, Division 2 locations
4.1 Wiring methods
Rigid metal conduit, intermediate metal conduit, electrical metallic tubing,
dust-tight wireways, or Type MC or MI cable with approved termination fittings, or Type PLTC in cable trays, or Type ITC in cable trays, or Type MC or TC cable installed in ladder, ventilated trough, or ventilated channel cable trays in a single layer, with a space not less than the larger cable diameter between the two adjacent cables, can be employed.
• All boxes and fittings must be dusttight.
4.2 Sealing
Sealing means must be provided as shown in 3.2.
5. Class III, Division 1 locations
5.1 Wiring methods
Rigid metal conduit, rigid non-metallic conduit, intermediate metal conduit, electrical metallic tubing, dust-tight wireways, or Type MC or MI cable with approved termination fittings, can be employed.
• All boxes and fittings must be dusttight.
5.2 Sealing
• Sealing means are not required.
6. Class III, Division 2 locations
6.1 Wiring methods
• Wiring methods must comply with 5.1.
6.2 Sealing
• Sealing means are not required.
Model HGC303 - Heat Value Gas Chromatograph vii
Safety Yamatake Corporation
viii Model HGC303 - Heat Value Gas Chromatograph
Table of Contents
Chapter 1: Introduction
1-1: Definition of terms ......................................................................................... 1-1
1-2: General ......................................................................................................... 1-3
1-3: Model HGC303 measuring system ............................................................... 1-4
1-4: Model HGC303 Structure.............................................................................. 1-5
1-5: Fieldbus communication system................................................................... 1-6
Chapter 2: Installation
2-1: Unpacking and storing .................................................................................. 2-1
2-2: Selecting an interface and installation .......................................................... 2-3
2-2-1: Installing the model HFA100 ........................................................... 2-3
2-2-2: Installing the National Instruments Fieldbus-PC Interface............... 2-7
2-3: HGM installation............................................................................................ 2-8
2-3-1: Computer system requirements ...................................................... 2-8
2-3-2: Windows installation ........................................................................ 2-8
2-3-3: HGM software installation................................................................ 2-9
2-4: Fieldbus installation ...................................................................................... 2-11
2-4-1: Fieldbus requirements ..................................................................... 2-11
2-4-2: Fieldbus wiring................................................................................. 2-13
2-5: Model HGC303 installation ........................................................................... 2-14
2-5-1: Installation site................................................................................. 2-14
2-5-2: Model HGC303 dimensions............................................................. 2-15
2-5-3: Model HGC303 installation example ............................................... 2-16
2-5-4: Model HGC303 piping ..................................................................... 2-17
2-5-5: Model HGC303 wiring ..................................................................... 2-19
Chapter 3: Operation
3-1: Starting up the model HGC303..................................................................... 3-1
3-1-1: Secondary pressure and flow set .................................................... 3-1
3-1-2: Piping leak check............................................................................. 3-1
3-1-3: Power on ......................................................................................... 3-2
3-1-4: Model HGC303 leak check.............................................................. 3-3
3-2: Stopping the model HGC303 ........................................................................ 3-4
3-3: HGM operation.............................................................................................. 3-4
3-3: Introduction ......................................................................................... 3-4
Functions .................................................................................................. 3-4
3-3-1: Getting started with model HFA100................................................. 3-5
3-3-2: Getting started with PCMCIA-FBUS................................................ 3-8
3-3-3: HGM Main menu ............................................................................. 3-14
Table of Contents
3-3-4: Set up HGM..................................................................................... 3-15
3-3-5: User's mode menu and commands ................................................. 3-22
3-3-6: Main displays of HGM ..................................................................... 3-23
3-3-7: Report.............................................................................................. 3-28
3-3-8: Configuration mode ......................................................................... 3-31
3-3-9: HGM shut down............................................................................... 3-38
3-4: Calibration..................................................................................................... 3-39
3-4-1: Calibration gas requirement ............................................................ 3-39
3-4-2: Calibration procedure ...................................................................... 3-40
3-4-3: Calibration function.......................................................................... 3-44
3-4-4: Description of component data table ............................................... 3-46
3-4-5: Report.............................................................................................. 3-47
3-4-6: Calibration methods......................................................................... 3-48
3-5: GPA mode .................................................................................................... 3-50
3-5-1: Setting the HGM to GPA ................................................................. 3-50
3-5-2: Data save ........................................................................................ 3-51
3-5-3: Data edit .......................................................................................... 3-51
3-5-4: File auto saving ............................................................................... 3-52
3-5-5: Configuration mode ......................................................................... 3-53
3-5-6: User's mode (GPA).......................................................................... 3-58
3-5-7: Main display panels of HGM (GPA)................................................. 3-60
3-5-8: Report (GPA)................................................................................... 3-65
Chapter 4: Maintenance
4-1: Checking and changing the carrier gas ........................................................ 4-1
4-2: Checking and changing the filters in model HGC303 ................................... 4-1
4-3: Overhaul ....................................................................................................... 4-1
Chapter 5: Troubleshooting
Appendix
GPA calculation.................................................................................................... A-1
Description of normalization method.......................................................... A-1
Formulas .................................................................................................... A-3
ISO calculation ..................................................................................................... A-8
Description of normalization method.......................................................... A-8
Formulas .................................................................................................... A-10
List of replacement parts...................................................................................... A-12
Drawings .............................................................................................................. A-13
List of Figure
Figure S-1 An example of conduit seal (with stopping plug) ............................... v
Figure S-2 An example of conduit seals (without stopping plug) ........................ v
Figure 1-1 Model HGC303 measuring system diagram...................................... 1-4
Figure 1-2 Main parts of Model HGC303 ............................................................ 1-5
Figure 2-1 HFA configuration initial display ........................................................ 2-4
Figure 2-2 Model HFA100 operation check display ............................................ 2-5
Figure 2-3 Setup message.................................................................................. 2-9
Figure 2-4 HGM installation location................................................................... 2-9
Figure 2-5 Group name....................................................................................... 2-10
Figure 2-6 Complete installation ......................................................................... 2-10
Figure 2-7 Example of Type A fieldbus cable structure ...................................... 2-11
Figure 2-8 Example of cable finish...................................................................... 2-13
Figure 2-9 Model HGC303 dimension ................................................................ 2-15
Figure 2-10 Example of model HGC303 installation with mounting bracket......... 2-16
Figure 2-11 Piping location ................................................................................... 2-17
Figure 2-12 Wiring location ................................................................................... 2-19
Figure 3-1 Leak check ........................................................................................ 3-3
Figure 3-2 Model HGC303-HGM connection example ....................................... 3-5
Figure 3-3 Model HGC303-HGM connection example ....................................... 3-8
Figure 3-4 Main menu......................................................................................... 3-14
Figure 3-5 Set up HGM display........................................................................... 3-15
Figure 3-6 Example of saved data files (.hv1) .................................................... 3-19
Figure 3-7 User's mode display .......................................................................... 3-22
Figure 3-8 Trend graph of SCV and the total concentration (Raw)..................... 3-24
Figure 3-9 Trend chromatogram (online) ............................................................ 3-25
Figure 3-10 Zoom box........................................................................................... 3-26
Figure 3-11 Trend graph of carrier pressure and oven temp. control ................... 3-27
Figure 3-12 User report entry form ....................................................................... 3-28
Figure 3-13 User report......................................................................................... 3-29
Figure 3-14 Configuration mode display ............................................................... 3-31
Figure 3-15 Calibration setting panel .................................................................... 3-44
Figure 3-16 Component data table ....................................................................... 3-46
Figure 3-17 Preview screen of report.................................................................... 3-47
Figure 3-18 Setup HGM display (GPA)................................................................. 3-50
Figure 3-19 Normalization method setting (GPA) ................................................. 3-50
Figure 3-20 An Example of saved data files (.hv2) ............................................... 3-51
Figure 3-21 Configuration mode display (GPA) .................................................... 3-53
Figure 3-22 Output configuration panel (GPA) ..................................................... 3-54
Figure 3-23 User’s mode display (GPA) ............................................................... 3-58
Figure 3-24 BTU trend graph and the total of raw concentration.......................... 3-62
Figure 3-25 Trend Chromatogram (online) ........................................................... 3-63
Figure 3-26 Trend graph of carrier pressure and oven temp. control ................... 3-64
Figure 3-27 Report entry form (GPA).................................................................... 3-65
Figure 3-28 User report (GPA mode).................................................................... 3-66
List of Table
Table 2-1: Fieldbus cable description.................................................................... 2-11
Table 2-2: Conduit type ......................................................................................... 2-15
Table 2-3: Piping description................................................................................. 2-17
Table 2-4: Gas specifications ................................................................................ 2-18
Table 2-5: Wiring description................................................................................. 2-19
Table 3-1 Gas specifications ................................................................................ 3-1
Table 3-2 The procedure to start up the model HGC303 system......................... 3-2
Table 3-3 Model HGC303 leak test procedure ..................................................... 3-3
Table 3-4 Stopping model HGC303 operation ..................................................... 3-4
Table 3-5 Main menu description ......................................................................... 3-14
Table 3-6 Set up HGM description ....................................................................... 3-16
Table 3-7 Set up online mode .............................................................................. 3-17
Table 3-8 Analyzer status and available functions ............................................... 3-18
Table 3-9 Save data description........................................................................... 3-18
Table 3-10 Description of user’s mode display....................................................... 3-22
Table 3-11 Description of the indication panel ....................................................... 3-23
Table 3-12 Trend graph of SCV and total raw concentration description............... 3-24
Table 3-13 Chromatogram description................................................................... 3-25
Table 3-14 Description of trend graph of carrier gas pressure
and oven temperature control .............................................................. 3-27
Table 3-15 Description of user report..................................................................... 3-30
Table 3-16 Description of configuration mode display............................................ 3-32
Table 3-17 Possible configurations of PV12-20 ..................................................... 3-34
Table 3-18 Stopping the HGM................................................................................ 3-38
Table 3-19 Example of calibration action ............................................................... 3-43
Table 3-20 Calibration factor setting....................................................................... 3-45
Table 3-21 Description of component data table.................................................... 3-46
Table 3-22 Manual calibration procedure............................................................... 3-48
Table 3-23 Operating auto calibration procedure................................................... 3-49
Table 3-24 The contents of each line of .hv2 ......................................................... 3-52
Table 3-25 Description of configuration mode display (GPA)................................. 3-53
Table 3-26 Possible configurations of PV12-20 (GPA) .......................................... 3-54
Table 3-27 Description of user’s mode display (GPA)............................................ 3-58
Table 3-28 The difference between ISO and GPA mode in user’s mode............... 3-59
Table 3-29 Indication panel description (GPA)....................................................... 3-60
Table 3-30 Trend graph of BTU and Total raw concentration description.............. 3-62
Table 3-31 Chromatogram description................................................................... 3-63
Table 3-32 Trend graph of Carrier gas pressure and Oven temperature control ... 3-64
Table 5-1 Model HGC303 self-diagnostics........................................................... 5-1
Table A-1 An example of the calculated normalized concentration
and heating value ................................................................................. A-2
Table A-2 Physical constants of selected hydrocarbons
and non-hydrocarbons ......................................................................... A-7

Chapter 1 : Introduction

0

1-1 : Definition of terms

Heat Value Gas Chromatograph (Model HGC303)
The Heat Value Gas Chromatograph measures process gases (N2, CO2, C1~C6+) that are mainly contained in natural gas, calculates heat value, density, Wobbe index and compressibility factor, and converts them into a Fieldbus signal in the field and transmits the signal to a receiver.
Parameters can all be remotely set, adjusted, and self-diagnosed by using the HGM.
Measuring and calculating methods comply with ISO 6974 Part 4 and ISO 6976.
HGC Modbus Interface Unit (Model HMU303)
The model HMU303 Modbus Interface can convert a Fieldbus signal into a Modbus signal.
HGC000
Monitoring of such data as heat value can be done using a Modbus signal as well as a Fieldbus signal.
See the model HMU303 User's manual for more details.
HGC Data Manager (Model HDM303)
Model HDM303 is Modbus interface unit for model HGC303. Model HDM303 covers all the function of model HMU303. Model HDM303 also has a powerful functions.
The functions are local display, data storage function, multi Modbus serial port, multi stream switching, and analog output.
HMU can not be connected together with HDM in the same FB loop. Only one HMU can be connected in one FB loop with the HGC. Two or more HDM can be connected in the same FB loop. For this application, the HDM must be configured first.
Please refer to the model HDM303 User's Manual for more details.
Heat Value Gas Chromatograph Fieldbus Adaptor (Model HFA100)
HFA is an interface used to connect the HGM (HGC monitor), Windows-based PC application, to Yamatake’s state of the art analyzer, HGC (Heat value Gas Chromatograph) that operates on FOUNDATION™ fieldbus H1 network. Users are able to configure, monitor and maintain the HGC all from the PC by simply connecting the HFA to the Fieldbus network.
Model HGC303 - Heat Value Gas Chromatograph 1-1
Introduction Yamatake Corporation
HGC Monitor (HGM)
HGM software is provided as a standard accessory with the model HGC303.
The model HGC303 Monitor allows the user to con­figure and calibrate the model HGC303 as well as allowing one to monitor a heat value-trend graph.
Moreover, HGM also has a report function for con­cise management.
HGM000001000P
SP (Set Point)
The set value of each variable.
PV (Process variable)
The present value of each variable.
SCV
ICV
TCD
URV
LRV
Total (Raw)
Component name
Superior Calorific Value
Inferior Calorific Value
Thermal Conductivity Detector
Upper Range Value
Lower Range Value
Total of raw concentration
C6+: Hexane and heavier gas C3H8: Propane i-C4H10: i-Butane n-C4H10: n-Butane neo-C5H12: neo-Pentane i-C5H12: i-Pentane n-C5H12: n-Pentane N2: Nitrogen CH4: Methane CO2: Carbon dioxide C2H6: Ethane
1-2 Model HGC303 - Heat Value Gas Chromatograph
Yamatake Corporation Introduction

1-2 : General

The model HGC303 is a gas chromatograph designed to analyze natural gas and is able to transmit a process variable via a Fieldbus signal.
One can easily adjust configuration data and monitor values such as the heat value by using the HGM.
The heat value monitoring system, which can be controlled from both the model HGC303 and HGM, will substantially minimize time, cost and maintenance.
This chapter first describes the measuring system and structure of the model HGC303.
After that, the characteristics and the specifications of Fieldbus are described in detail.
First time users of the model HGC303 should read this chapter carefully and thor­oughly.
Components of the model HGC303 system
Before installing the model HGC303, the following components must be prepared:
Hardware
Model HGC303
Model HMU303
Model HFA100
Power supply (24 V DC, 4A min.), Power supply cable
Fieldbus cable (See “2-4-1 : Fieldbus requirements” on page 2-11)
Flow meter for process gas
(A flow meter for methane should be used scale: 0 - 200 ml/min.)
Laptop or desktop PC
(See “2-3-1 : Computer system requirements” on page 2-8 for detail)
The National Instruments Fieldbus-PC interface*
(See “2-2-2 : Installing the National Instruments Fieldbus-PC Interface” on page 2-7 for detail)
Helium gas for carrier gas and valve operating gas
Calibration gas
1/8 inch stainless steel (SS) tubing
Fitting for piping (1/4 NPT male, 5 pieces)
Software
Microsoft Windows 98 / Me / NT / 2000 / XP
HGM
~Note *: This is not necessary if using the model HFA100. However, the model
HFA100 can only run on Windows 2000 / XP. If you are running an operating system other than Windows 2000 / XP, purchase the National Instruments Fieldbus-PC Interface.
Model HGC303 - Heat Value Gas Chromatograph 1-3
Introduction Yamatake Corporation

1-3 : Model HGC303 measuring system

Modbus
RS2320
HMU-
Modbus Converter
PS
POWER
DC 24V
Manual
*
Switch (for manual calibration)
: Terminator
T
Required for manual calibration.
*
(at least one of them)
or RS422/RS485
T
ALARM 50L RELAY
FB
Safe Area Field
AIR/N2/He He Standard
350 - 450 kPa
350 - 450 kPa
HGM
Gas
Fieldbus
Solenoid valve (24V DC)
Sampling flow meter
3-way
*
valve (for manual calibration)
Inline Filter
AIR
CARR
INLET
50ml/mh
Needle valve
Block valve
Block valve
POWER
T
HGC303
Bypass flow meter
Needle valve
Sample Gas
OUTPUT
AIR VENT
VENT (Open to Air)
TCD-VENT
OUTLET
RETURN to low pressure line
NG
PIPE LINE
Figure 1-1 Model HGC303 measuring system diagram
CAUTION
A block valve is a kind of air actuator valve. It is used mainly for the protection of the TCD and columns.
It works as sample shut-off valve when the pressure of the carrier gas or air supply is lower than approximately 294 kPa.
Yamatake recommends that it should be installed.
The Heat Value Gas Chromatograph measures process gases (N2, CO2, C1~C6+) that are mainly contained in natural gas, calculates heat value, density, Wobbe index and compressibility factor, and converts them into a Fieldbus signal in the field and trans­mits the signal to receivers.
Parameters can be remotely set, adjusted, and self-diagnosed with the HGM.
1-4 Model HGC303 - Heat Value Gas Chromatograph
Yamatake Corporation Introduction

1-4 : Model HGC303 Structure

Terminal housing
Terminal cover
Analyzer unit housing
Manifold
Oven cover
HGC000004000D
Figure 1-2 Main parts of Model HGC303
1 Terminal housing ..........Terminal box for wiring.
2 Analyzer unit housing...Micro valve, solenoid valve, TCD sensor are located here.
3 Manifold........................Connection parts for gas inlet and outlet line
4 Oven cover....................Analyzer valve and column system are found inside the
cover.
Model HGC303 - Heat Value Gas Chromatograph 1-5
Introduction Yamatake Corporation

1-5 : Fieldbus communication system

The model HGC303 uses FOUNDATION™ fieldbus technology to transfer information between other devices.
The FOUNDATION™ fieldbus is an open, 2-wire, multi-drop, two-way digital commu- nication system which interconnects field equipment such as sensors, actuators and controllers.
The FOUNDATION™ fieldbus is supported by a worldwide network of customers and manufacturers in Europe, North America and Asia Pacific.
FOUNDATION™ fieldbus http://www.fieldbus.org/
FOUNDATION™ fieldbus literature
(1) FOUNDATION™ fieldbus Technical overview (FD-043)
(2) Fieldbus Installation & Planning Guide (AG-165)
(3) FOUNDATION™ fieldbus Application Guide
31.25kbit/s Wiring and Installation (AG-140)
(4) FOUNDATION™ fieldbus Application Guide
31.25kbit/s Intrinsically Safe Systems (AG-163)
1-6 Model HGC303 - Heat Value Gas Chromatograph

Chapter 2 : Installation

This chapter guides you through the procedures for installing of your hardware and software.

2-1 : Unpacking and storing

Unpacking the model HGC303

Your model HGC303 is a precision instrument and should be handled with care to pre­vent any damage to it or breaking it.
After unpacking the model HGC303, verify that the following items are included:
Package items
(1) Model HGC303
(2) Mounting bracket set
(3) Wrench for seal plug
(4) CD-ROM including HGM software and user's manual
(5) EC declaration of conformity and safety instructions
CAUTION
Exposing the model HGC303 to the atmosphere might cause deterioration of the column. Therefore, the model HGC303 has been packed and shipped in a protective bag with a desiccant. Install and operate the model HGC303 immediately after breaking the seal.
Model HGC303 - Heat Value Gas Chromatograph 2-1
Installation Yamatake Corporation
Inquires
If you have any questions regard-
(1)
ing the specifications of your model HGC303, contact one of the Yamatake products service offices listed at the back of this user's manual or contact your nearest
(4)
(5)
Yamatake representative.
When making an inquiry, make sure to provide the model number and product number of your model HGC303.
(3)
(2)
HGC000002000P
2-2 Model HGC303 - Heat Value Gas Chromatograph
Yamatake Corporation Installation

Storing the model HGC303

The model HGC303 should be stored:
- indoor at storage temperature (-40 to 70°C); humidity (up to 95%RH)
- in a place safe from vibration or shock.
- in the same packing as it was shipped in.
Model HGC303 that has been used should be stored by following procedures below.
Step Action
1 Makes sure no process gas remains in the model HGC303.
2 Purge the model HGC303 with helium gas.
Insert metal plugs into all the inlets and outlets for carrier
3
gas, valve operating gas and process gas except VENT (valve operating gas outlet) in order to keep moisture out.
4 Pack it as it was when it was originally received.
5
Store the model HGC303 indoors at normal temperature and humidity in a place safe from vibration or shock.
Model HGC303 - Heat Value Gas Chromatograph 2-3
Installation Yamatake Corporation

2-2 : Selecting an interface and installation

The HGM requires a data converter to convert the data that is being received from the HGC. You may use either 1) the Yamatake Heat Value Gas Chromatograph Fieldbus Adaptor (model HFA100) or 2) the National Instruments Fieldbus-PC Interface.
CAUTION
Model HFA100 operates only on Windows2000 / XP. Purchase the National Instru­ments Fieldbus-PC Interface if you are running an operating system other than Win­dows 2000/XP.
~Note 1. A PC with a USB1.0 port is required to run the model HFA100.
2. For the National Instruments Fieldbus-PC Interface, you additionally have to prepare a floppy disk drive and a PC having either a type-II PCMCIA socket or a 16-bit ISA slot.

2-2-1 : Installing the model HFA100

Installing the hardware:
Refer to the model HFA100 instruction manual.
Installing the software:
In order to use the model HFA100, following two softwares must be installed.
1. HFA Config.exe (Installed at the same time the HGM is installed)
2. HFBADRV (An USB driver for model HFA100)
Procedure for software installation
(1) First install the HGM by following instructions found in section 2-3 “HGM Instal-
lation” below.
An installation CD-ROM comes supplied with your model HGC303. Take note that there are two types of “Setup HGM” folders on the CD-ROM: (1) Setup HGM for HFA and (2) Setup HGM for NI. To use the model HFA100, you must select the {setup} icon that is in the “Setup HGM for HFA” folder. If you by chance install from the “Setup HGM for NI” folder, uninstall it, and then install from the “Setup HGM for HFA” folder.
(2) Once HGM installation has completed, install 'HFBADRV'.
CAUTION
When connecting the model HFA100 to a USB port, an installation wizard will appear on your screen. Follow the instructions that the installation wizards give you. You will be asked for your CD-ROM that came with the model HGC303 where you should select the folder “HFBADRV”. (Once the file ‘HFBADRV’ is installed, cannot be uninstalled.)
2-4 Model HGC303 - Heat Value Gas Chromatograph
Yamatake Corporation Installation
(3) Checking after installation
Connect the model HFA100 to the PC, and then start HFA Config.exe (Default directory: C:\Program file\HGM), which is in the HGM folder. A dialog box like that shown in Figure 2-1 will appear.
Figure 2-1 HFA configuration initial display
Model HGC303 - Heat Value Gas Chromatograph 2-5
Installation Yamatake Corporation
Click on the Read Parameter button. If PD_TAG, Node Address and Device Class appear on the display as they do below, the driver has been installed correctly. (See Figure 2-2).
Figure 2-2 Model HFA100 operation check display
Default Setting
PD_TAG: HFA100
Node Address: 0×FC
Device Class: Basic Device
This completes your model HFA100 installation. Refer to “3-3 : HGM operation” when using the model HFA100 connected to a HGM.
2-6 Model HGC303 - Heat Value Gas Chromatograph
Yamatake Corporation Installation
~Note
*Installing and Uninstalling the model HFA100
Installing: Connect the model HFA100 to the PC’s USB port. An new icon will
appear on the taskbar.
Uninstalling:Select the “Stop HFBADRV (HFA100 FB Adapter)” from the task-
bar.
* Default setting of HGC system device (normally no change required)
Model No. PD_TAG Node address
HGC303 HGC 0×F7
HMU303 HMU 0×10
HDM303 HDM303 0×10
HFA100 HFA100 0×FC
Usable setting range
PD_TAG First three characters must be model No.
0×10 - 0×13
HMU system
Node address
HDM system
FB initialization should be done when either PD_TAG or Node Address is changed.
0×E0 - 0×F7 0×FC - 0×FF
0×10 - 0×24 0×E0 - 0×F7 0×FC - 0×FF
CAUTION
* The device has been factory set for the user's application therefore the default value should suit your process. However, when changing the settings, have a thor­ough understanding of the parameters beforehand. * “Fieldbus device configuration” is only for authorized service personnel. If the “Fieldbus device configuration” button is clicked, shutdown at once and then restart “HFA Config.exe”.
Model HGC303 - Heat Value Gas Chromatograph 2-7
Installation Yamatake Corporation

2-2-2 : Installing the National Instruments Fieldbus-PC Interface

For instructions on installing the National Instruments Fieldbus-PC Interface, refer to the instruction manual packed with the product. Once installation is completed, pro­ceed to “2-3 : HGM installation” in this manual for HGM installation procedures. To install the HGM, select the {setup} icon from the folder “Setup HGM for NI” found on the installation CD-ROM.
Reference:
The National Instruments Fieldbus-PC interface
Manufacturer: National Instruments
http://www.natinst.com/
Requirements
(1) PCMCIA Type
Product name: PCMCIA-FBUS (1 port)
Model number: 777282-01
Includes: 1) Fieldbus Interface for PCMCIA
2) NI-FBUS Communication Manager Software
3) PCMCIA-FBUS Cable (1 port)
4) Documentation Kit
(2) ISA board Type
Product name: AT-FBUS (1 port)
Model number: 777281-01
Includes: 1) Fieldbus Interface ISA board
2) NI-FBUS Communication Manager Software
3) Documentation Kit
2-8 Model HGC303 - Heat Value Gas Chromatograph
Yamatake Corporation Installation

2-3 : HGM installation

2-3-1 : Computer system requirements

(1) Computer: DOS/V compatible,
Minimum 200 MHz Pentium processor.
(2) System memory: 64 MB minimum
(3) Disk storage: 100 MB free space
(4) CD-ROM drive
(5) Operating system: Windows 98 / Me / NT / 2000 / XP
(6) Video Monitor: Microsoft Windows 98 / Me / NT / 2000 / XP, 100% compatible,
including a display resolution of 1024×768 or higher and high color of 16 bit or higher.
(7) Printer: Must be 100% compatible with Microsoft Windows 98 / Me / NT / 2000 /
XP.

2-3-2 : Windows installation

Before installing the HGM, Windows 98 / Me / NT / 2000 / XP must be installed and running as an operating system on your computer.
~Note Install the latest service pack of Windows.
Model HGC303 - Heat Value Gas Chromatograph 2-9
Installation Yamatake Corporation

2-3-3 : HGM software installation

Installing the HGM
(1) Make sure Windows 98 / Me / NT / 2000 / XP has been installed.
(2) Insert the CD-ROM into the CD-ROM drive.
(3) Double-click on the [Setup HGM] folder.
(4) Double-click on the [setup] icon. The following screen will appear:
HGC Monitor Version 2.00 Setup
Welcome to the HGC Monitor Version 2.00 installation program.
Setup cannot install system files or update shared files if they are in use. Before proceeding, we recommend that you close any applications you may be running.
OK
Exit Setup
Figure 2-3 Setup message
Click on [OK] and the following display will appear:
HGC Monitor Version 2.00 Setup
Begin the installation by clicking the button below.
Click this button to install HGC Monitor Version 2.00 software to the specified destination directory.
Directory:
C:\Program Files\hgm\
Exit Setup
Change Directory
Figure 2-4 HGM installation location
Default directory: C:\Program Files\hgm
If you want to use another drive or directory, select an optional drive or directory by pressing the [Change Directory] button.
2-10 Model HGC303 - Heat Value Gas Chromatograph
Loading...
+ 100 hidden pages