While this information is presented in good faith and believed to be accurate, Yamatake
disclaims the implied warranties of merchantability and fitness for a particular purpose
and makes no express warranties except as may be stated in its written agreement with
and for its customer.
In no event shall Yamatake Corporation be liable to anyone for any indirect, special or
consequential damages. The information and specifications in this document are subject
to change without notice.
®
Microsoft
ration.
Windows™ and Excel™ are trademarks of Microsoft Corporation.
, Windows®, and MS-DOS® are registered trademarks of Microsoft Corpo-
NI-FBUS™ is a trademark of NATIONAL INSTRUMENTS CORPORATION
®
.
Safety
Safety symbols
Be sure to correctly operate the model HGC303 while strictly observing the safety
precautions provided in this manual-especially the Warnings and Cautions indicated
by the symbols as shown below.
The descriptions of the Warning and Caution signs used in this manual are as follows.
The Warning sign means that serious personal injury, even death, could result if the
instructions given are not strictly observed.
The Caution sign means that light personal injury and/or equipment damage could
result if the instructions given are not correctly observed.
WARNING
CAUTION
Model HGC303 - Heat Value Gas Chromatographi
SafetyYamatake Corporation
Hazardous Areas Certifications
The model HGC303 complies with the type of protection, which based on the
following standards.
(1) ISSeP/ATEX Flameproof Certification
0344
II 2 GD
EEx d II C T6 -10°C<Tamb<+50°C IP65
ISSeP01ATEX003X
Special conditions for safe use (symbol X)
The fastening screws of this apparatus are made of stainless steel and have a
yield stress of 500 N/mm².
(2) FM Explosionproof / Flameproof Approval
Explosionproof for Class I, Division 1, Groups C and D, T4
Flameproof for Class I, Zone 1, AEx d IIB T4
Dust-ignitionproof for Class II and III, Division 1, Groups E, F and G, T4
(3) TestSafe Flameproof Certification
Ex d IIC T6
DIP A21 TA T6
The vent plug must be removed and 1/4 NPT fitting must be connected for piping.
Therefore, the model HGC303 can be installed in various hazardous locations.
However, an explosion-protected electrical apparatus requires special care. Please read
all instruction and safety notes before installation.
WARNING
NEVER open the terminal box cover while the model HGC303 is energized in a
hazardous location.
CAUTION
Use the model HGC303 only in an ambient temperature of -10 to 50°C (14 to 122°F)
CAUTION
Take precautions to prevent corrosion, deformation or damage to the housing or
terminal box cover.
CAUTION
See that all conduits are properly sealed. Otherwise, the model HGC303 cannot
withstand the pressure that can result from explosion of an explosive gas inside the
housing. Also, the model HGC303 cannot prevent the explosion of any external
explosive gas.
iiModel HGC303 - Heat Value Gas Chromatograph
Yamatake CorporationSafety
(1) Installation for ISSeP/ATEX Flameproof Apparatus
1. General
1.1 The apparatus protected by the flameproof enclosure in accordance with EN
50018 can be installed in such hazardous areas, for which the apparatus has been
certified, as an explosive atmosphere containing flammable substances in the form
of gas, vapour, mist or dust may be present.
~NoteThe apparatus has been certified to comply with EN 50281-1-1 (dust
ignition protection).
1.2 The apparatus enclosure must be kept closed in the hazardous areas when the
apparatus is energized because the internal circuit of the apparatus is capable of
igniting the explosive atmosphere. (Never connect any hand-held communicator to
the apparatus terminals by opening the cover, except while no explosive
atmosphere is present.)
1.3 It is required to connect the external earthing terminal of the apparatus to the equipotential bonding system which includes protective conductors, metal
conduits, metal cable sheaths, steel wire armouring and metallic parts of structures,
but does not include the neutral conductors of the power systems.
~NoteThe protective conductor to which exposed conductive parts of equipment
(machines, apparatus, devices, components and instrumentation thereof)
are connected, must be separated in the hazardous area from the neutral
conductor, and must be connected to the power systems earth point in the
non-hazardous area, if the power system is directly earthed.
For external earthing and bonding of the apparatus it is recommended to use a
cable lug so that the conductor is secured against loosening and twisting and that
the contact pressure is permanently secured.
1.4 Either cable systems (cable entry systems) or conduit systems can be employed
for wiring of the apparatus in the hazardous areas (see 2 or 3).
1.5 Non-sheathed single core cables are not permitted for live conductors unless
they are installed inside enclosures or conduit systems.
1.6 Conduits and, in special cases, cables (for example, where there is a pressure
difference) must be sealed so as to prevent the passage of the explosive
atmosphere.
1.7 Further information concerning installation and maintenance of apparatus is
given by relevant clauses of the following documents.
EN 60079-14 Electrical apparatus for explosive gas atmospheres
Part 14: Electrical installations in hazardous areas other than mines
EN 60079-17 Part 17: Inspection and maintenance of electrical installations in
hazardous areas.
EN 50281-1-2 Electrical apparatus for use in the presence of combustible dust
Part 1-2: Electrical apparatus protected by enclosures
-- Selection, installation and maintenance
Model HGC303 - Heat Value Gas Chromatographiii
SafetyYamatake Corporation
2. Cable systems
2.1 Thermoplastic sheathed cables, thermosetting sheathed cables, or elastomeric
sheathed cables can be selected for fixed wiring in the hazardous areas.
2.2 Flameproof cable entry devices (cable glands) certified to comply with EN 50018
and appropriate to the type of cable employed, must be used for the connection of
cables to the apparatus.
3. Conduit systems
For conduit systems, relevant national standards or codes of practice are followed
prior to the following recommendations.
3.1 Screwed heavy gauge steel, solid drawn or seam welded conduit, or flexible conduit for protection of cables in explosive atmospheres (see ISO 10807) can
be selected for fixed wiring in the hazardous areas.
3.2 Conduit must be threaded for connection to permit the full engagement of five
threads.
3.3 Either conduit entry devices or sealing devices such as stopping boxes are
provided at the wall of the apparatus enclosure to limit the pressure piling effect
and to prevent hot gases from entering the conduit system from the enclosure
containing a source of ignition. Each type of both the devices must be certified
to comply with EN 50018.
3.4 The stopping boxes, if used, are filled with a compound which does not shrink
or setting and is impervious to, and unaffected by, chemicals found in the
hazardous area. The depth of the compound in the stopping box is at least equal
to the internal diameter of the conduit, but in no case less than 10 mm.
3.5 When the conduit contains three or more non-seathed single or multi-core cables, the total cross-sectional areas of cables, including insulation, are not more
than 40% of the cross-sectional area of the conduit.
4. Installation in explosive atmospheres caused by air / dust mixtures
4.1 Conduit or cable glands, if employed to connect cables to the apparatus, must be
selected and used in such a way that an IP6X protection (dust-tight) is guaranteed.
4.2 It is recommended to maintain the apparatus so that the dust layer will not exceed a thickness of 5 mm.
~NoteWhere the ignition temperature of a dust layer up to 5 mm thickness is
equal to, or higher than, the value that is obtained by adding 75K to the
maximum surface temperature of the enclosure “T...°C” as marked on the
apparatus, the apparatus is incapable of causing ignition of the dust
layer. (T...°C is based on the maximum ambient temperature)
ivModel HGC303 - Heat Value Gas Chromatograph
Yamatake CorporationSafety
(2) Installation for FM Explosionproof / Flameproof Apparatus (in
accordance with NEC)
CAUTION
•Install the apparatus only in hazardous (classified) locations for which the
apparatus has been approved.
•Seal each conduit entering the apparatus enclosure within 18 in.(457 mm)
from the enclosure.
•Do not open the apparatus enclosure when an explosive atmosphere is
present.
Figure S-1 An example of conduit seal (with stopping plug)
Figure S-2 An example of conduit seals (without stopping plug)
Model HGC303 - Heat Value Gas Chromatographv
SafetyYamatake Corporation
1. Class I, Division 1 locations
1.1 Wiring methods
• Threaded rigid metal conduit, threaded steel intermediate metal conduit, or
Type MI cable with termination fittings approved for the location, can be
employed
• Threaded joints must be made up with at least five threads fully engaged.
• Boxes, fittings, and joints must be approved for Class I, Division 1.
1.2 Sealing
• Each conduit entering the apparatus enclosure is required to be sealed
within 18 in. (457 mm) from the enclosure.
• The sealing of each conduit can be provided with a sealing fitting approved for
class I locations.
• Sealing compound must be approved and must not have a melting point of less
than 93° (200°F).
• The minimum thickness of the sealing compound should not be less than the
trade size of the conduit and, in no case, less than 5/8 in.(16 mm).
• Splices and taps cannot be made in the fittings.
2. Class I, Division 2 locations
2.1 Wiring methods
• Threaded rigid metal conduit, threaded steel intermediate metal conduit,
enclosed gasketed busways, or Type PLTC cable in accordance with the
provisions of remote-control, signaling, and power-limited circuits (see NEC,
Article 725), or Type ITC cable in cable trays, in raceways, supported by
messenger wire, or directly buried where the cable is listed for this use; Type MI, MC, MV, or TC cable with approved termination fittings can be
employed.
• Boxes, fittings, and joints are not required to be explosionproof.
2.2 Sealing
• Each conduit entering the apparatus enclosure is required to be sealed as shown in 1.2.
3. Class II, Division 1 locations
3.1 Wiring methods
• Threaded rigid metal conduit, threaded steel intermediate metal conduit, or
Type MI cable with termination fittings approved for the location, can be
employed.
• Boxes and fittings must be dusttight.
3.2 Sealing
• Where a raceway provides communication between the apparatus enclosure and
an enclosure that is not required to be dust-ignitionproof, suitable means must be
provided to prevent the entrance of dust into the dust-ignitionproof enclosure
through the raceway. One of the following means can be used: (1) a permanent
and effective seal; (2) a horizontal raceway not less than 10 ft (3.05 m) long; or
(3) a vertical raceway not less than 5 ft (1.52 m) long and extending downward
from the dust-ignitionproof enclosure.
• Seals are not required to be explosionproof.
viModel HGC303 - Heat Value Gas Chromatograph
Yamatake CorporationSafety
4. Class II, Division 2 locations
4.1 Wiring methods
• Rigid metal conduit, intermediate metal conduit, electrical metallic tubing,
dust-tight wireways, or Type MC or MI cable with approved termination
fittings, or Type PLTC in cable trays, or Type ITC in cable trays, or Type
MC or TC cable installed in ladder, ventilated trough, or ventilated channel
cable trays in a single layer, with a space not less than the larger cable
diameter between the two adjacent cables, can be employed.
• All boxes and fittings must be dusttight.
4.2 Sealing
•Sealing means must be provided as shown in 3.2.
5. Class III, Division 1 locations
5.1 Wiring methods
• Rigid metal conduit, rigid non-metallic conduit, intermediate metal conduit,
electrical metallic tubing, dust-tight wireways, or Type MC or MI cable with
approved termination fittings, can be employed.
• All boxes and fittings must be dusttight.
5.2 Sealing
• Sealing means are not required.
6. Class III, Division 2 locations
6.1 Wiring methods
• Wiring methods must comply with 5.1.
6.2 Sealing
• Sealing means are not required.
Model HGC303 - Heat Value Gas Chromatographvii
SafetyYamatake Corporation
viiiModel HGC303 - Heat Value Gas Chromatograph
Table of Contents
Chapter 1: Introduction
1-1: Definition of terms ......................................................................................... 1-1
1-2: General ......................................................................................................... 1-3
1-3: Model HGC303 measuring system ............................................................... 1-4
1-4: Model HGC303 Structure.............................................................................. 1-5
1-5: Fieldbus communication system................................................................... 1-6
Chapter 2: Installation
2-1: Unpacking and storing .................................................................................. 2-1
2-2: Selecting an interface and installation .......................................................... 2-3
2-2-1: Installing the model HFA100 ........................................................... 2-3
2-2-2: Installing the National Instruments Fieldbus-PC Interface............... 2-7
Table A-1An example of the calculated normalized concentration
and heating value ................................................................................. A-2
Table A-2Physical constants of selected hydrocarbons
and non-hydrocarbons ......................................................................... A-7
Chapter 1 : Introduction
0
1-1 : Definition of terms
Heat Value Gas Chromatograph (Model HGC303)
The Heat Value Gas Chromatograph measures process gases (N2,
CO2, C1~C6+) that are mainly contained in natural gas, calculates
heat value, density, Wobbe index and compressibility factor, and
converts them into a Fieldbus signal in the field and transmits the
signal to a receiver.
Parameters can all be remotely set, adjusted, and self-diagnosed by
using the HGM.
Measuring and calculating methods comply with ISO 6974 Part 4 and
ISO 6976.
HGC Modbus Interface Unit (Model HMU303)
The model HMU303 Modbus Interface can convert a
Fieldbus signal into a Modbus signal.
HGC000
Monitoring of such data as heat value can be done using a
Modbus signal as well as a Fieldbus signal.
See the model HMU303 User's manual for more details.
HGC Data Manager (Model HDM303)
Model HDM303 is Modbus interface unit for model HGC303.
Model HDM303 covers all the function of model HMU303.
Model HDM303 also has a powerful functions.
The functions are local display, data storage function, multi
Modbus serial port, multi stream switching, and analog output.
HMU can not be connected together with HDM in the same FB
loop. Only one HMU can be connected in one FB loop with the
HGC. Two or more HDM can be connected in the same FB
loop. For this application, the HDM must be configured first.
Please refer to the model HDM303 User's Manual for more details.
Heat Value Gas Chromatograph Fieldbus Adaptor (Model HFA100)
HFA is an interface used to connect the HGM (HGC
monitor), Windows-based PC application, to
Yamatake’s state of the art analyzer, HGC (Heat value
Gas Chromatograph) that operates on FOUNDATION™
fieldbus H1 network. Users are able to configure,
monitor and maintain the HGC all from the PC by
simply connecting the HFA to the Fieldbus network.
Model HGC303 - Heat Value Gas Chromatograph1-1
IntroductionYamatake Corporation
HGC Monitor (HGM)
HGM software is provided as a standard accessory
with the model HGC303.
The model HGC303 Monitor allows the user to configure and calibrate the model HGC303 as well as
allowing one to monitor a heat value-trend graph.
Moreover, HGM also has a report function for concise management.
The model HGC303 is a gas chromatograph designed to analyze natural gas and is
able to transmit a process variable via a Fieldbus signal.
One can easily adjust configuration data and monitor values such as the heat value by
using the HGM.
The heat value monitoring system, which can be controlled from both the model
HGC303 and HGM, will substantially minimize time, cost and maintenance.
This chapter first describes the measuring system and structure of the model HGC303.
After that, the characteristics and the specifications of Fieldbus are described in detail.
First time users of the model HGC303 should read this chapter carefully and thoroughly.
Components of the model HGC303 system
Before installing the model HGC303, the following components must be prepared:
Hardware
Model HGC303
Model HMU303
Model HFA100
Power supply (24 V DC, 4A min.), Power supply cable
Fieldbus cable (See “2-4-1 : Fieldbus requirements” on page 2-11)
Flow meter for process gas
(A flow meter for methane should be used scale: 0 - 200 ml/min.)
Laptop or desktop PC
(See “2-3-1 : Computer system requirements” on page 2-8 for detail)
The National Instruments Fieldbus-PC interface*
(See “2-2-2 : Installing the National Instruments Fieldbus-PC Interface” on page 2-7 for
detail)
Helium gas for carrier gas and valve operating gas
Calibration gas
1/8 inch stainless steel (SS) tubing
Fitting for piping (1/4 NPT male, 5 pieces)
Software
Microsoft Windows 98 / Me / NT / 2000 / XP
HGM
~Note *: This is not necessary if using the model HFA100. However, the model
HFA100 can only run on Windows 2000 / XP. If you are running an
operating system other than Windows 2000 / XP, purchase the National
Instruments Fieldbus-PC Interface.
Model HGC303 - Heat Value Gas Chromatograph1-3
IntroductionYamatake Corporation
1-3 : Model HGC303 measuring system
Modbus
RS2320
HMU-
Modbus
Converter
PS
POWER
DC
24V
Manual
*
Switch
(for manual
calibration)
: Terminator
T
Required for manual calibration.
*
(at least one of them)
or
RS422/RS485
T
ALARM50LRELAY
FB
Safe AreaField
AIR/N2/He He Standard
350 - 450 kPa
350 - 450 kPa
HGM
Gas
Fieldbus
Solenoid
valve
(24V DC)
Sampling
flow meter
3-way
*
valve
(for manual
calibration)
Inline
Filter
AIR
CARR
INLET
50ml/mh
Needle
valve
Block
valve
Block
valve
POWER
T
HGC303
Bypass
flow meter
Needle
valve
Sample Gas
OUTPUT
AIR VENT
VENT (Open to Air)
TCD-VENT
OUTLET
RETURN to
low pressure
line
NG
PIPE LINE
Figure 1-1 Model HGC303 measuring system diagram
CAUTION
A block valve is a kind of air actuator valve. It is used mainly for the protection of
the TCD and columns.
It works as sample shut-off valve when the pressure of the carrier gas or air supply
is lower than approximately 294 kPa.
Yamatake recommends that it should be installed.
The Heat Value Gas Chromatograph measures process gases (N2, CO2, C1~C6+) that
are mainly contained in natural gas, calculates heat value, density, Wobbe index and
compressibility factor, and converts them into a Fieldbus signal in the field and transmits the signal to receivers.
Parameters can be remotely set, adjusted, and self-diagnosed with the HGM.
1-4Model HGC303 - Heat Value Gas Chromatograph
Yamatake CorporationIntroduction
1-4 : Model HGC303 Structure
Terminal housing
Terminal cover
Analyzer unit housing
Manifold
Oven cover
HGC000004000D
Figure 1-2 Main parts of Model HGC303
1 Terminal housing ..........Terminal box for wiring.
2 Analyzer unit housing...Micro valve, solenoid valve, TCD sensor are located here.
3 Manifold........................Connection parts for gas inlet and outlet line
4 Oven cover....................Analyzer valve and column system are found inside the
cover.
Model HGC303 - Heat Value Gas Chromatograph1-5
IntroductionYamatake Corporation
1-5 : Fieldbus communication system
The model HGC303 uses FOUNDATION™ fieldbus technology to transfer information
between other devices.
The FOUNDATION™ fieldbus is an open, 2-wire, multi-drop, two-way digital commu-
nication system which interconnects field equipment such as sensors, actuators and
controllers.
The FOUNDATION™ fieldbus is supported by a worldwide network of customers and
manufacturers in Europe, North America and Asia Pacific.
This chapter guides you through the procedures for installing of your hardware and
software.
2-1 : Unpacking and storing
Unpacking the model HGC303
Your model HGC303 is a precision instrument and should be handled with care to prevent any damage to it or breaking it.
After unpacking the model HGC303, verify that the following items are included:
Package items
(1)Model HGC303
(2)Mounting bracket set
(3)Wrench for seal plug
(4)CD-ROM including HGM software and user's manual
(5)EC declaration of conformity and safety instructions
CAUTION
Exposing the model HGC303 to the atmosphere might cause deterioration of the
column. Therefore, the model HGC303 has been packed and shipped in a protective
bag with a desiccant. Install and operate the model HGC303 immediately after
breaking the seal.
Model HGC303 - Heat Value Gas Chromatograph2-1
InstallationYamatake Corporation
Inquires
If you have any questions regard-
(1)
ing the specifications of your
model HGC303, contact one of the
Yamatake products service offices
listed at the back of this user's
manual or contact your nearest
(4)
(5)
Yamatake representative.
When making an inquiry, make
sure to provide the model number
and product number of your model
HGC303.
(3)
(2)
HGC000002000P
2-2Model HGC303 - Heat Value Gas Chromatograph
Yamatake CorporationInstallation
Storing the model HGC303
The model HGC303 should be stored:
- indoor at storage temperature (-40 to 70°C); humidity (up to 95%RH)
- in a place safe from vibration or shock.
- in the same packing as it was shipped in.
Model HGC303 that has been used should be stored by following procedures below.
StepAction
1Makes sure no process gas remains in the model HGC303.
2Purge the model HGC303 with helium gas.
Insert metal plugs into all the inlets and outlets for carrier
3
gas, valve operating gas and process gas except VENT
(valve operating gas outlet) in order to keep moisture out.
4Pack it as it was when it was originally received.
5
Store the model HGC303 indoors at normal temperature
and humidity in a place safe from vibration or shock.
Model HGC303 - Heat Value Gas Chromatograph2-3
InstallationYamatake Corporation
2-2 : Selecting an interface and installation
The HGM requires a data converter to convert the data that is being received from the
HGC. You may use either 1) the Yamatake Heat Value Gas Chromatograph Fieldbus
Adaptor (model HFA100) or 2) the National Instruments Fieldbus-PC Interface.
CAUTION
Model HFA100 operates only on Windows2000 / XP. Purchase the National Instruments Fieldbus-PC Interface if you are running an operating system other than Windows 2000/XP.
~Note1. A PC with a USB1.0 port is required to run the model HFA100.
2. For the National Instruments Fieldbus-PC Interface, you additionally
have to prepare a floppy disk drive and a PC having either a type-II
PCMCIA socket or a 16-bit ISA slot.
2-2-1 : Installing the model HFA100
Installing the hardware:
Refer to the model HFA100 instruction manual.
Installing the software:
In order to use the model HFA100, following two softwares must be installed.
1. HFA Config.exe (Installed at the same time the HGM is installed)
2. HFBADRV (An USB driver for model HFA100)
Procedure for software installation
(1) First install the HGM by following instructions found in section 2-3 “HGM Instal-
lation” below.
An installation CD-ROM comes supplied with your model HGC303. Take note that
there are two types of “Setup HGM” folders on the CD-ROM: (1) Setup HGM for
HFA and (2) Setup HGM for NI.
To use the model HFA100, you must select the {setup} icon that is in the “Setup
HGM for HFA” folder. If you by chance install from the “Setup HGM for NI”
folder, uninstall it, and then install from the “Setup HGM for HFA” folder.
(2) Once HGM installation has completed, install 'HFBADRV'.
CAUTION
When connecting the model HFA100 to a USB port, an installation wizard will
appear on your screen. Follow the instructions that the installation wizards give
you. You will be asked for your CD-ROM that came with the model HGC303
where you should select the folder “HFBADRV”. (Once the file ‘HFBADRV’ is
installed, cannot be uninstalled.)
2-4Model HGC303 - Heat Value Gas Chromatograph
Yamatake CorporationInstallation
(3) Checking after installation
Connect the model HFA100 to the PC, and then start HFA Config.exe (Default
directory: C:\Program file\HGM), which is in the HGM folder. A dialog box like
that shown in Figure 2-1 will appear.
Figure 2-1 HFA configuration initial display
Model HGC303 - Heat Value Gas Chromatograph2-5
InstallationYamatake Corporation
Click on the Read Parameter button. If PD_TAG, Node Address and Device Class
appear on the display as they do below, the driver has been installed correctly. (See
Figure 2-2).
Figure 2-2 Model HFA100 operation check display
Default Setting
PD_TAG:HFA100
Node Address:0×FC
Device Class:Basic Device
This completes your model HFA100 installation. Refer to “3-3 : HGM operation”
when using the model HFA100 connected to a HGM.
2-6Model HGC303 - Heat Value Gas Chromatograph
Yamatake CorporationInstallation
~Note
*Installing and Uninstalling the model HFA100
Installing:Connect the model HFA100 to the PC’s USB port. An new icon will
appear on the taskbar.
Uninstalling:Select the “Stop HFBADRV (HFA100 FB Adapter)” from the task-
bar.
* Default setting of HGC system device (normally no change required)
Model No.PD_TAGNode address
HGC303HGC0×F7
HMU303HMU0×10
HDM303HDM3030×10
HFA100HFA1000×FC
Usable setting range
PD_TAGFirst three characters must be model No.
0×10 - 0×13
HMU system
Node address
HDM system
FB initialization should be done when either PD_TAG or Node Address is
changed.
0×E0 - 0×F7
0×FC - 0×FF
0×10 - 0×24
0×E0 - 0×F7
0×FC - 0×FF
CAUTION
* The device has been factory set for the user's application therefore the default
value should suit your process. However, when changing the settings, have a thorough understanding of the parameters beforehand.
* “Fieldbus device configuration” is only for authorized service personnel. If the
“Fieldbus device configuration” button is clicked, shutdown at once and then restart
“HFA Config.exe”.
Model HGC303 - Heat Value Gas Chromatograph2-7
InstallationYamatake Corporation
2-2-2 : Installing the National Instruments Fieldbus-PC Interface
For instructions on installing the National Instruments Fieldbus-PC Interface, refer to
the instruction manual packed with the product. Once installation is completed, proceed to “2-3 : HGM installation” in this manual for HGM installation procedures. To
install the HGM, select the {setup} icon from the folder “Setup HGM for NI” found
on the installation CD-ROM.
Reference:
The National Instruments Fieldbus-PC interface
Manufacturer:National Instruments
http://www.natinst.com/
Requirements
(1) PCMCIA Type
Product name:PCMCIA-FBUS (1 port)
Model number:777282-01
Includes:1) Fieldbus Interface for PCMCIA
2) NI-FBUS Communication Manager Software
3) PCMCIA-FBUS Cable (1 port)
4) Documentation Kit
(2) ISA board Type
Product name:AT-FBUS (1 port)
Model number:777281-01
Includes:1) Fieldbus Interface ISA board
2) NI-FBUS Communication Manager Software
3) Documentation Kit
2-8Model HGC303 - Heat Value Gas Chromatograph
Yamatake CorporationInstallation
2-3 : HGM installation
2-3-1 : Computer system requirements
(1) Computer: DOS/V compatible,
Minimum 200 MHz Pentium processor.
(2) System memory: 64 MB minimum
(3) Disk storage: 100 MB free space
(4) CD-ROM drive
(5) Operating system: Windows 98 / Me / NT / 2000 / XP
(6) Video Monitor: Microsoft Windows 98 / Me / NT / 2000 / XP, 100% compatible,
including a display resolution of 1024×768 or higher and high color of 16 bit or
higher.
(7) Printer: Must be 100% compatible with Microsoft Windows 98 / Me / NT / 2000 /
XP.
2-3-2 : Windows installation
Before installing the HGM, Windows 98 / Me / NT / 2000 / XP must be installed and
running as an operating system on your computer.
~NoteInstall the latest service pack of Windows.
Model HGC303 - Heat Value Gas Chromatograph2-9
InstallationYamatake Corporation
2-3-3 : HGM software installation
Installing the HGM
(1) Make sure Windows 98 / Me / NT / 2000 / XP has been installed.
(2) Insert the CD-ROM into the CD-ROM drive.
(3) Double-click on the [Setup HGM] folder.
(4) Double-click on the [setup] icon. The following screen will appear:
HGC Monitor Version 2.00 Setup
Welcome to the HGC Monitor Version 2.00 installation
program.
Setup cannot install system files or update shared files if they are in
use. Before proceeding, we recommend that you close any applications
you may be running.
OK
Exit Setup
Figure 2-3 Setup message
Click on [OK] and the following display will appear:
HGC Monitor Version 2.00 Setup
Begin the installation by clicking the button below.
Click this button to install HGC Monitor Version 2.00 software to the specified
destination directory.
Directory:
C:\Program Files\hgm\
Exit Setup
Change Directory
Figure 2-4 HGM installation location
Default directory: C:\Program Files\hgm
If you want to use another drive or directory, select an optional drive or directory by
pressing the [Change Directory] button.
2-10Model HGC303 - Heat Value Gas Chromatograph
Loading...
+ 100 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.