ASSEMBLING
Unpack the package, put the platter on the top of the loader.
Put the balance on a steady flat surface away from vibration, direct sunshine, air
blow or strong magnetic disturbance.
Battery locker is at the bottom of the balance, switch the locker to to turn on the
balance, please switch the locker to during moving or transportation
TURN ON THE BALANCE
Connect the balance to power supply.
Press function key ON/OFF to turn on the balance, the balance will in turns display
following figures,
8.8.8.8.8.8.
Battery Voltage U=
Maximum capacity
-----------Finally, the balance stays in a standby state, showing 0.00g on the display
Notice: the lasting time of --------- will be decided by the stability of the loadcell, thus, the
balance must not be located on an unstable surface or near windows or air flow. If symbol
flashes in the display, it means the balance is not in a stable condition, calibration and
counting operations is not allowed.
CALIBRATION
The purpose for this operation is to calibrate the balance, to achieve the best
performance in case there is obvious tolerance error in weighing or the balance is
located in different gravity due to different latitude.
it is suggested to warm up the balance over half an hour before calibration.
Remove all loads from the platter, press function key TARE to clear the readings
to 0.00g.
Press and hold function key CAL for 3 seconds, release function key CAL when “---
CAL---“ is shown in the display, a figure of standard weight will flash in the display,
put a standard weight on the platter accordingly, standby state “------“ will be shown in
the display for a few seconds before the standard weight figure is shown in the
display, remove the weight, “------“ standby state will remain for a few seconds before
the balance enter stable weighing mode, showing 0.00g.
It is recommended to make the operations twice to achieve the best calibration
result.