X-Rite DensiEye 100, DensiEye 700 Reference Manual

Page 1
DensiEye
Reference Manual
E N G L I S H
Page 2
X-Rite DensiEye 100 and 700
Contents
1. General ................................................................................................................................... 3
1.1. Introduction ........................................................................................................................... 3
1.2. Safety instructions ................................................................................................................. 4
1.3. Package contents .................................................................................................................. 5
1.4. Packaging and transport ....................................................................................................... 5
2. Operation ................................................................................................................................ 6
2.1. Functional elements .............................................................................................................. 6
2.2. Release and lock measuring head ........................................................................................ 6
2.2.1. Release measuring head ............................................................................................................... 6
2.2.2. Lock measuring head .................................................................................................................... 6
2.3. Factory settings ..................................................................................................................... 7
2.4. Standard buttons ................................................................................................................... 8
2.5. Special buttons and button combinations ............................................................................. 8
2.6. Symbols and their functionality ............................................................................................. 9
2.7. Display and Pass/Fail indicator ........................................................................................... 10
2.7.1. Display layout .............................................................................................................................. 10
2.7.2. Message box ............................................................................................................................... 10
2.7.3. Pass/Fail indicator ....................................................................................................................... 10
2.7.4. Pass/Fail information ................................................................................................................... 11
2.8. Function selection ............................................................................................................... 11
2.9. Symbol selection ................................................................................................................. 11
2.10. Positioning and measurement ............................................................................................. 11
3. Measurement Functions ..................................................................................................... 12
3.1. Auto Mode ........................................................................................................................... 12
3.2. Density ................................................................................................................................ 14
3.2.1. Density (Paper) ............................................................................................................................ 14
3.2.2. Density (Absolute) ....................................................................................................................... 15
3.3. Dot Gain .............................................................................................................................. 16
3.4. Trapping .............................................................................................................................. 18
3.5. Gray Balance ...................................................................................................................... 19
3.5.1. Gray Balance (Paper) .................................................................................................................. 19
3.5.2. Gray Balance (Absolute) .............................................................................................................. 20
3.6. Dot Area .............................................................................................................................. 21
3.7. Print Characteristic .............................................................................................................. 22
4. References ........................................................................................................................... 24
4.1. Select a reference set ......................................................................................................... 24
4.2. Solid reference .................................................................................................................... 25
4.2.1. Measure Solid reference values (Paper) ...................................................................................... 25
4.2.2. Edit Solid reference and tolerance values (Paper) ...................................... ................................. 25
4.2.3. Measure Solid reference values (Absolute) ................................................................................. 26
4.2.4. Edit Solid reference and tolerance values (Absolute) ................................................................... 26
4.3. Halftone and Dot Gain ......................................................................................................... 27
4.4. Gray Balance ...................................................................................................................... 28
4.4.1. Measure Gray Balance reference values (Paper) ........................................................................ 28
4.4.2. Edit Gray Balance reference and tolerance values (Paper) ......................................................... 29
4.4.3. Measure Gray Balance reference values (Absolute) .................................................................... 29
4.4.4. Edit Gray Balance reference and tolerance values (Absolute) ..................................................... 30
5. Calibration ............................................................................................................................ 31
5.1. Calibration (paper white base) ............................................................................................ 31
5.1.1. Select paper white base ............................................................................................................... 31
5.1.2. Calibration (Paper) ....................................................................................................................... 32
5.1.3. Check Calibration (Paper) ............................................................................................................ 32
5.2. Calibration (absolute white base) ........................................................................................ 33
5.2.1. Select absolute white base .......................................................................................................... 33
5.2.2. Calibration (Absolute) .................................................................................................................. 34
5.2.3. Check Calibration (Absolute) ....................................................................................................... 34
5.3. Density calibration reference ............................................................................................... 35
DensiEye Ref erence Manual 4 1
Page 3
X-Rite DensiEye 100 and 700
6. Settings ................................................................................................................................. 36
6.1. Function settings ................................................................................................................. 36
6.1.1. Pass / Fail indicator ..................................................................................................................... 36
6.1.2. White base ................................................................................................................................... 37
6.1.3. Scale graduation .......................................................................................................................... 38
6.1.4. Decimal places ............................................................................................................................ 38
6.2. Device settings .................................................................................................................... 39
6.2.1. Turn display ................................................................................................................................. 39
6.2.2. Instrument type ............................................................................................................................ 39
6.2.3. Language ..................................................................................................................................... 40
7. USB Interface ....................................................................................................................... 41
7.1. USB Driver Installation ........................................................................................................ 41
7.2. USB Interface Protocol ........................................................................................................ 41
8. DensiEye 100 to 700 Upgrade ............................................................................................. 41
9. Maintenance and Care ......................................................................................................... 43
9.1. Reset ................................................................................................................................... 43
9.2. Total Reset .......................................................................................................................... 43
9.3. Instrument type information ................................................................................................. 44
9.4. Replacing the batteries ....................................................................................................... 45
9.5. Firmware upgrade ............................................................................................................... 46
9.6. Warranty registration ........................................................................................................... 47
9.7. Recertification ..................................................................................................................... 47
9.8. Service ................................................................................................................................ 47
10. Specifications ...................................................................................................................... 48
10.1. Functions ............................................................................................................................. 48
10.2. Technical Specification ....................................................................................................... 48
2 DensiEye Reference Manual 4
Page 4
X-Rite DensiEye 100 and 700
1. General
1.1. Introduction
Dear Customer
Congratulations! You have just acquired the reflection densitometer DensiEye.
The DensiEye features an easy to use graphical user interface and a high level of automation. Its high measurement accuracy makes this a truly unique instrument.
In this reference manual, all functionality of the DensiEye 100 and DensiEye 700 is explained in step by step detail. For a quick first start, please refer to the quick start guide, which enables you to use your new instrument immediately.
Your X-RiteTeam
DensiEye Ref erence Manual 4 3
Page 5
X-Rite DensiEye 100 and 700
1.2. Safety instructions
The X-Rite DensiEye is not for use in an explosive environment
Do not expose the DensiEye to strong electromagnetic fields.
For safety reasons, the DensiEye should be operated at ambient temperatures between 10°C / 50°F and 40°C / 104°F with a relative air humidity of 20% to 80% (non-condensing).
For safety reasons, the DensiEye should be stored at ambient temperatures between ­20°C /- 4°F and 70°C / 158°F with a relative air humidity of 5% to 90% (non-condensing).
Do not expose the DensiEye to direct sunlight.
Protect the DensiEye against chemical substances, corrosive vapors, strong vibrations, and mechanical influences.
Always use the original packaging to transport the DensiEye.
Do not attempt to repair the DensiEye. Opening the instrument will void your warranty.
Only use genuine X-Rite accessories and spare parts
DensiEye should only be used by trained personnel
Always place samples on a stable measuring surface. Make sure the bottom of the shoe is flat against the measurement surface.
The plastic device casing can be cleaned using a damp cloth and a small amount of soap
Never short-circuit the batteries!
Always replace both batteries at the same time
Dispose of batteries in accordance with local regulations
4 DensiEye Reference Manual 4
Page 6
X-Rite DensiEye 100 and 700
1.3. Package contents
Item Name
1 DensiEye 2 Calibration card 3 USB cable 4 Quick start guide 5 Device certificate 6 Registration card 7 CD with operating instructions and software 8 Device case
1.4. Packaging and transport
Always use the original packaging to transport the DensiEye.
DensiEye Ref erence Manual 4 5
Page 7
r
X-Rite DensiEye 100 and 700
2. Operation
2.1. Functional elements
PASS/FAIL Indicator
Display
Reset button
Measuring diaphragm
Batteries
Up button
Locking button
Ente
button
Down button
USB Interface
2.2. Release and lock measuring head
2.2.1. Release measuring head
Slide the locking button on the left side to the back to release the measuring head.
After approximately 30 seconds of inactivity, the DensiEye goes into power saving
mode.
Press any button to wake up the instrument.
2.2.2. Lock measuring head
Lower the measuring head and slide the locking button forward to lock the measuring head
After approximately 30 seconds of inactivity, the DensiEye goes into power saving mode
6 DensiEye Reference Manual 4
Page 8
X-Rite DensiEye 100 and 700
2.3. Factory defaults
Absolute/Difference values: Absolute
White base: Paper
Reference set: 1 reference set
Filter/color selection: Auto
Pass/Fail indicator: Off
Scale graduation for print characteristic: 10%
Number of decimal places for densities: 2
Language: English
Left/right-handed operation: Right-handed operation
DensiEye Ref erence Manual 4 7
Page 9
X-Rite DensiEye 100 and 700
2.4. Standard buttons
Button Description Functionality
Up button Moves cursor up Increase value
Open / close submenu
Enter button
Toggle options
Down button Moves cursor down Decrease value
Confirm value
2.5. Special buttons and button combinations
Button(s) Description
Enter + Measure
Reset Auto mode (DensiEye 700) or Density (DensiEye 100) is selected in
Reset + Enter 2 Auto Mode function is pre-selected in the main menu
1
White measurement
the main menu
All settings are retained
All settings are reset to factory settings
The reference- and tolerance values are retained
The calibration is reset. The device must be re-calibrated
Enter (3 sec) Return to Function Selection on the main menu
Enter (10 sec) Displays the language selection menu
Note:
1
Sequence:
Press the Enter button
Execute a white measurement
Release the Enter button
2
Sequence:
Press the Reset button
Press the Enter button
Release the Reset button
Release the Enter button
8 DensiEye Reference Manual 4
Page 10
X-Rite DensiEye 100 and 700
2.6. Symbols and their functions
Symbol Name Description
Return Return to next menu level up, accepts data
Paper
Filter selection Select density filter (automatic or manual)
Reference selection Select one of four reference sets
Solid Measure solid patch / Edit or measure solid reference
Halftone 1 Measure halftone 1 / Edit halftone 1 reference
Halftone 2 Measure halftone 2 / Edit halftone 2 reference
Halftone 3 Measure halftone 3 / Edit halftone 3 reference
Gray balance patch
Measure paper white / Measure absolute white reference (calibration)
Measure gray balance patch / Edit or measure gray balance reference
Any patch Measure any patch using Auto Mode function
View / Edit mode View and edit halftone measurements
Absolute / Difference mode
First solid patch Measure first solid patch (trapping)
Second solid patch Measure second solid patch (trapping)
Trapping Measure trapping patch
Black Edit black value (calibration)
Cyan Edit cyan value (calibration)
Magenta Edit magenta value (calibration)
Yellow Edit yellow value (calibration)
Pass/Fail Toggle Pass/Fail indicator on and off
White base Set white base to paper white or absolute white
Toggle between absolute and difference values
Scale graduation Set scale graduation for print characteristics
Decimal places Set accuracy for density measurements
Screen orientation Change screen orientation for right or left hand use
Instrument type Display device type information
Language Select language
DensiEye Ref erence Manual 4 9
Page 11
X-Rite DensiEye 100 and 700
2.7. Display and Pass/Fail indicator
2.7.1. Display layout
The display is divided into 5 areas:
Header -> Display the name of the basic function
Help area -> Display help on the specific function
Symbol area -> Display symbols for navigation via options and menus
Output area -> Display the measurement results
Status area -> Display status messages (variable length):
o Selected reference set for difference display o Selected density filter setting: Auto, Cyan, Magenta,
Yellow or Black
Symbol area
Header
Output area
Help area
Status area
2.7.2. Message box
A message box displays information and error messages for example ”White measurement done”.
2.7.3. Pass/Fail indicator
The DensiEye has an integrated Pass/Fail indicator. If you take a density, dot gain, or gray balance measurement, the measurement values are compared to references and tolerances. The red LED indicates that the measurement is out of tolerance while a green light indicates the measurement is OK. Settings for the Pass/Fail indicator are in the function setup.
10 DensiEye Reference Manual 4
Page 12
X-Rite DensiEye 100 and 700
2.7.4. Pass/Fail information display
After setting up reference and tolerance values for density, dot gain (DensiEye 700 only) and gray balance, the Pass/Fail function can indicate corrective measures to the right of the measurement values.
If the measured value is below the specified tolerance range, the instrument displays an arrow pointing up.
If the measured value is within tolerance, the instrument displays a check mark behind the value.
If the measured value is above the specified tolerance range, the instrument displays an arrow pointing down.
2.8. Function selection
Use the Up/Down buttons to select a function from the main menu.
Press the Enter button in the main menu to open the function screen
Press the Enter button in a popup window to activate the selected option
2.9. Symbol selection
Use the Up/Down buttons to select a symbol from the main menu.
The symbols have different functions (see section Symbols and their functions). Depending
on the function, you need to press the Enter button or initiate a measurement.
2.10. Positioning and measurement
Always place the printed sheet on a stable measuring surface. Make sure the bottom of the shoe is flat against the measurement surface.
Use measuring diaphragm to position the DensiEye properly on the measuring patch
Lower the measuring head to initiate a measurement
After the measurement, the instrument displays the measurement values and you can
release the measuring head
DensiEye Ref erence Manual 4 11
Page 13
X-Rite DensiEye 100 and 700
3. Measurement Functions
3.1. Auto Mode (DensiEye 700 only)
The instrument identifies Solid, Halftone, Trapping or Gray Balance patches automatically and the display shows the corresponding measurement values
The Paper White measurement is performed automatically if the White Base is set to Absolute Auto or Paper Auto in the Function Settings.
The default setting for filter/color detection is the Automatic color/filter detection
Select reference values in the References menu
Select the Auto Mode in the main menu
o The status displays the current reference set o The status displays “A” for automatic filter/color detection
Select Absolute/Difference
Select Absolute for absolute measurement values
o The output area displays D
Select Difference for difference measurement values
o The output area displays D
Select Reference for difference readings
The pop-up window shows a list with reference sets
Select a reference set
o The status area displays the selected reference set
Select Paper and measure paper white
After the measurement, the message box briefly displays “White Measurement
done”
Measure solids in correct printing sequence
The press icons in the output area show values for the solid densities and the
corresponding colors starting from right to left
The output area shows absolute or difference values for solid density and the corresponding color of the last patch:
12 DensiEye Reference Manual 4
Page 14
X-Rite DensiEye 100 and 700
Measure any patch (Solid, Halftone, Trapping or Gray Balance)
o The output area shows the measurement values of the last patch:
Absolute or difference values for solid density and the corresponding
color
Halftone reference and dot gain value and the corresponding color
Trapping value and the corresponding printing sequence
Absolute or difference values for density for Cyan, Magenta and Yellow
Select Return to return to the main menu
DensiEye Ref erence Manual 4 13
Page 15
X-Rite DensiEye 100 and 700
3.2. Density
3.2.1. Density (Paper)
Set White Base to Paper or Paper Auto in the Function Settings
Make sure the instrument is calibrated for paper white base
Set reference values in the References menu
Select Density in the main menu
o The status displays the current reference set if the Pass/Fail detection is activated
or Absolute/Difference is set to Difference
o The status displays the current filter/color
Select Filter
The pop-up window shows a list with automatic or manual filter/color selections
Set the automatic or manual filter/color selection
o The status displays the current filter/color
Select Absolute/Difference
Select Absolute for absolute density values
o The output area displays D
Select Difference for difference density values
o The output area displays D
Select Reference for difference readings
The pop-up window shows a list with reference sets
Select a reference set
o The status area displays the selected reference set
Select Paper and measure paper white
After the measurement, the message box briefly displays “White Measurement
done”
Measure Solid
The output area shows absolute or difference density values and the corresponding
color of the last patch
Select Return to return to the main menu
14 DensiEye Reference Manual 4
Page 16
X-Rite DensiEye 100 and 700
3.2.2. Density (Absolute)
Set White Base to Absolute or Absolute Auto in the Function Settings
Make sure the instrument is calibrated for absolute white
Set reference values in the References menu
Select Density in the main menu
o The status displays the current reference set if the Pass/Fail detection is activated
or Absolute/Difference is set to Difference
o The status displays the current filter/color
Select Filter
The pop-up window shows a list with automatic or manual filter/color selections
Set the automatic or manual filter/color selection
o The status displays the current filter/color
base
Select Absolute/Difference
Select Absolute for absolute density values
o The output area displays D
Select Difference for difference density values
o The output area displays D
Select Reference for difference readings
The pop-up window shows a list with reference sets
Select a reference set
o The status area displays the selected reference set
Measure Solid
The output area shows absolute or difference density values and the corresponding
color of the last patch
Select Return to return to the main menu
DensiEye Ref erence Manual 4 15
Page 17
X-Rite DensiEye 100 and 700
3.3. Dot Gain (DensiEye 700 onl y)
Dot gain is a densitometric function based on paper white
Make sure the instrument is calibrated for paper white base
Set reference values in the References menu
Deactivate halftone reference values in the References
Select Dot Gain in the main menu
o The status displays the current reference set o The status displays the current filter/color
Select Filter selection
The pop-up window shows a list with automatic or manual filter/color selections
Set the automatic or manual filter/color selection
o The status displays the current filter/color
Select Reference for difference readings
The pop-up window shows a list with reference sets
Select a reference set
Select Paper and measure paper white
After the measurement, the message box briefly displays “White Measurement
o The status area displays the selected reference set
done”
Measure Solid
o The output area displays solid density values
Measure 1. Halftone (if activated)
o The output field displays solid density, the 1. halftone reference and dot gain
values
16 DensiEye Reference Manual 4
Page 18
X-Rite DensiEye 100 and 700
Measure 2. Halftone (if activated)
o The output field displays solid density, 1. and 2. halftone reference and dot gain
values
Measure 3. Halftone (if activated)
o The output field displays 1., 2. and 3. halftone reference and dot gain values
Select Return to return to the main menu
DensiEye Ref erence Manual 4 17
Page 19
X-Rite DensiEye 100 and 700
3.4. Trapping (DensiEye 700 only)
Trapping is a densitometric function based on paper white
Make sure the instrument is calibrated for paper white base
Select Trapping in the main menu
o The status displays the current filter/color
Select Filter selection
The pop-up window shows a list with automatic or manual filter/color selections
Set the automatic or manual filter/color selection
o The status displays the current filter/color
Select Paper and measure paper white
After the measurement, the message box briefly displays “White Measurement
done”
Measure 1st Solid Patch
o The output field displays the density values of the 1. solid patch
Measure 2nd Solid Patch
o The output field displays the density values of the 1. and 2. solid patch
Measure Trapping Patch
o The output field displays the density values of the 1. and 2. solid patch and %
Trapping
Select Return to return to the main menu
18 DensiEye Reference Manual 4
Page 20
X-Rite DensiEye 100 and 700
3.5. Gray Balance
3.5.1. Gray Balance (Paper)
Set White Base to Paper or Paper Auto in the Function Settings
Make sure the instrument is calibrated for paper white base
Set reference values in the References menu
Select Gray Balance in the main menu
o The status displays the current reference set if the Pass/Fail detection is activated
or Absolute/Difference is set to Difference
Select Absolute/Difference
o Select Absolute for gray balance absolute values
The output area displays DC, DM and DY
o Select Difference for gray balance difference readings
The output area displays ∆DC, DM and DY
Select Reference for difference readings
The pop-up window shows a list with reference sets
Select a reference set
o The status area displays the selected reference set
Select Paper and measure paper white
After the measurement, the message box briefly displays “White Measurement
done”
Measure Gray Balance Patch
o The output area displays absolute or difference values for density for Cyan,
Magenta and Yellow
Select Return to return to the main menu
DensiEye Ref erence Manual 4 19
Page 21
X-Rite DensiEye 100 and 700
3.5.2. Gray Balance (Absolute)
Set White Base to Absolute or Absolute Auto in the Function Settings
Make sure the instrument is calibrated for absolute white
Set reference values in the References menu
Select Gray Balance in the main menu
o The status displays the current reference set if the Pass/Fail detection is activated
or Absolute/Difference is set to Difference
Select Absolute/Difference
o Select Absolute for gray balance readings
The output area displays DC, DM and DY
o Select Difference for gray balance difference readings
The output area displays ∆DC, DM and DY
base
Select Reference for difference readings
The pop-up window shows a list with reference sets
Select a reference set
o The status area displays the selected reference set
Measure Gray Balance Patch
o The output area displays absolute or difference values for density for Cyan,
Magenta and Yellow
Select Return to return to the main menu
20 DensiEye Reference Manual 4
Page 22
X-Rite DensiEye 100 and 700
3.6. Dot Area (DensiEye 700 only)
Dot Area is a densitometric function based on paper white
Make sure the instrument is calibrated for paper white base
Select Dot Area in the main menu
o The status displays the current filter/color
Select Filter selection
The pop-up window shows a list with automatic or manual filter/color selections
Set the automatic or manual filter/color selection
o The status displays the current filter/color
Select Paper and measure paper white
After the measurement, the message box briefly displays “White Measurement
done”
Measure Solid
Measure Halftone
Select Return to return to the main menu
o The output area displays solid density values
o The output field displays solid density and dot area of halftone
DensiEye Ref erence Manual 4 21
Page 23
X-Rite DensiEye 100 and 700
3.7. Print Characteristic (DensiEye 700 only)
Print Characteristic is a densitometric function based on paper white
Make sure if instrument is calibrated for paper white base
Set Scale Graduation (e.g. 10%) in the Function Settings menu
Select Print Characteristic in the main menu
o The output area displays an empty print characteristic diagram o The status displays the current filter/color
Select Filter selection
The pop-up window shows a list with automatic or manual filter/color selections
Set the automatic or manual filter/color selection
o The status displays the current filter/color
Select Paper and measure paper white
After the measurement, the message box briefly displays “White Measurement
Measure Solid
The output area displays solid density values
Measure halftone patches e.g.10% patch up to 90% patch
done”
The output field displays solid density, the last dot area reference, the corresponding dot area sample value and the print characteristic curve
After the last measurement, the help area displays “Measurement done”
22 DensiEye Reference Manual 4
Page 24
X-Rite DensiEye 100 and 700
Select View / Edit
o A vertical bar points to the last measured halftone o Use the Up/Down button to select and view the desired halftone o Re-measure the desired halftone o Press the Enter button to exit the View / Edit mode
Select Return to return to the main menu
DensiEye Ref erence Manual 4 23
Page 25
X-Rite DensiEye 100 and 700
4. References
Set White Base to Paper or Paper Auto and Absolute or Absolute Auto in the Function Settings
Make sure that the instrument is calibrated for paper or absolute white
Select References in the main menu
o The header displays the current reference set o The symbol area shows the symbols for Reference selection, Solid, Halftone 1 …3
(DensiEye 700 only) and Gray Balance
o If paper white base is set, the Paper symbol appears
4.1. Select a reference set
The DensiEye includes 6 reference sets
A reference set contains:
o Solid density references and tolerance values o Halftone and dot gain reference values, dot gain tolerance values
(DensiEye 700 only)
o Gray balance density references and tolerance values
base
Select Reference
Select a reference set
o The header area displays the selected reference set o If paper white base is set, the help area displays the Paper symbol and
“Measure Paper White”
Measure Paper White
After the measurement, the message box briefly displays “White Measurement
done”
24 DensiEye Reference Manual 4
Page 26
X-Rite DensiEye 100 and 700
4.2. Solid reference
Reference values are referenced to paper white or absolute white base
Measured reference values appear in a table with solid density reference and
tolerance values for cyan, magenta, yellow and black
You can edit or measure these reference values and edit the tolerance values for each color in this table
4.2.1. Measure Solid reference values (Paper)
Set White Base to Paper or Paper Auto in the Function Settings
Make sure that the instrument is calibrated for paper white base
Select Paper and measure paper white
After the measurement, the message box briefly displays “White Measurement
done”
Select Solid
o The output area displays a table with solid density reference and tolerance
values for cyan, magenta, yellow and black
o Measure solid reference values
The output area displays measured solid density reference values
Select Return to return to the main menu
4.2.2. Edit Solid reference and tolerance values (Paper)
Set White Base to Paper or Paper Auto in the Function Settings
Make sure that the instrument is calibrated for paper white base
Select Solid
o Highlight solid density reference and tolerance values for cyan o Press the Enter button to open the edit or measure screen
Edit or measure solid density reference values
Edit solid density tolerance value
The output area displays the measured or edited solid density
reference and tolerance values
Continue editing the other colors as described
Select Return to return to the main menu
DensiEye Ref erence Manual 4 25
Page 27
X-Rite DensiEye 100 and 700
4.2.3. Measure Solid reference values (Absolute)
Set White Base to Absolute or Absolute Auto in the Function Settings
Make sure that the instrument is calibrated for absolute white base
Select Solid
o The output area displays a table with solid density reference and tolerance
values for cyan, magenta, yellow and black
o Measure solid reference values
The output area displays measured solid density reference values
Select Return to return to the main menu
4.2.4. Edit Solid reference and tolerance values (Absolute)
Set White Base to Absolute or Absolute Auto in the Function Settings
Make sure that the instrument is calibrated for absolute white base
Select Solid
o Highlight solid density reference and tolerance values for cyan o Press the Enter button to open the edit or measure screen
Edit or measure solid density reference values
Edit solid density tolerance value
The output area displays the measured or edited solid density
reference and tolerance values
Continue editing the other colors as described
Select Return to return to the main menu
26 DensiEye Reference Manual 4
Page 28
X-Rite DensiEye 100 and 700
4.3. Halftone and Dot Gain (DensiEye 700 only)
Halftone reference values, dot gain reference and tolerance values can be edited
Select Halftone 1, Halftone 2, or Halftone 3
o The output area displays a table with halftone and dot gain reference values and
dot gain tolerance values for cyan, magenta, yellow and black
Press the Enter button
o Highlight dot area and dot gain reference and dot gain tolerance values for cyan o Press the Enter button to open the edit screen
Edit dot area reference value
Edit dot gain reference value
Edit dot gain tolerance value
Continue editing the other colors as described
Select Return to return to the main menu
DensiEye Ref erence Manual 4 27
Page 29
X-Rite DensiEye 100 and 700
4.4. Gray Balance
Reference values are referenced to paper white or absolute white base
Measured reference values appear in a table with solid density reference and
tolerance values for cyan, magenta, yellow and black
You can edit or measure these reference values and edit the tolerance values for each color in this table
4.4.1. Measure Gray Balance reference values (Paper)
Set White Base to Paper or Paper Auto in the Function Settings
Make sure that the instrument is calibrated for paper white base
Select Paper and measure paper white
After the measurement, the message box briefly displays “White Measurement
done”
Select Gray Balance
o The output area displays a table with gray balance density reference and
tolerance values for cyan, magenta, yellow and black
o Measure gray balance reference values
The output area displays measured gray balance density reference
values
Select Return to return to the main menu
28 DensiEye Reference Manual 4
Page 30
X-Rite DensiEye 100 and 700
4.4.2. Edit Gray Balance reference and tolerance values (Paper)
Set White Base to Paper or Paper Auto in the menu Function Settings
Make sure that the instrument is calibrated for paper white base
Select Gray Balance
o Highlight gray balance reference and tolerance values for cyan o Press the Enter button to open the edit screen
Edit or measure gray balance density reference value
Edit gray balance density tolerance value
The output area displays the measured or edited gray balance
Continue editing the other colors as described
density reference values and the edited gray balance density tolerance values
Select Return to return to the main menu
4.4.3. Measure Gray Balance reference values (Absolute)
Set White Base to Absolute or Absolute Auto in the menu Function Settings
Make sure that the instrument is calibrated for absolute white base
Select Paper and measure paper white
After the measurement, the message box briefly displays “White Measurement
done”
Select Gray Balance
o The output area displays a table with gray balance density reference and
tolerance values for cyan, magenta, yellow and black
o Measure gray balance reference values
The output area displays measured gray balance density reference
values
Select Return to return to the main menu
DensiEye Ref erence Manual 4 29
Page 31
X-Rite DensiEye 100 and 700
4.4.4. Edit Gray Balance reference and tolerance values (Absolute)
Set White Base to Absolute or Absolute Auto in the Function Settings
Make sure that the instrument is calibrated for absolute white base
Select Gray Balance
o Highlight gray balance reference and tolerance values for cyan o Press the Enter button to open the edit or measure screen
Edit or measure gray balance density reference value
Press the Enter button to open the edit window for gray balance density
tolerance value
Edit gray balance density tolerance value
The output area displays the measured or edited gray balance
density reference values and the edited gray balance density tolerance values
Continue editing the other colors as described
Select Return to return to the main menu
30 DensiEye Reference Manual 4
Page 32
X-Rite DensiEye 100 and 700
5. Calibration
5.1. Calibration (paper white base)
In the graphic arts industry, measurements with paper white base are the
preferred operating mode. All measurements are relative to paper white
Check the calibration once a month and recalibrate the instrument if
necessary
For precise measurements, make sure to only use the calibration reference
you received with your DensiEye
The density standard and the filter type on the calibration card must
correspond with the device information
5.1.1. Select paper white base
Select Function Settings in the main menu
Select White base
o The output area displays the current settings
Press the Enter button to open the pop-up screen
Set the white base to paper or paper auto
o The output area displays the current settings
Select Return to return to the main menu
DensiEye Ref erence Manual 4 31
Page 33
X-Rite DensiEye 100 and 700
5.1.2. Calibration (Paper)
Select Calibration in the main menu
Measure Paper White on the calibration card
After the measurement, the message box briefly displays “White Measurement
done”
Measure the solid patches Black, Cyan, Magenta and Yellow on the calibration card
Press the Enter button to open the edit or measure screen for Black
o Edit density value
Continue editing the other colors as described
Select Return to return to the main menu
5.1.3. Check Calibration (Paper)
Select Density in the main menu
Select Paper
Measure Paper White on the calibration card
o After the measurement, the message box briefly displays “White Measurement
done”
Measure Solid Patch B, C, M, Y on the calibration card
o The difference between reference value and the measured value must be within
± 0.01
Select Return to return to the main menu
32 DensiEye Reference Manual 4
Page 34
X-Rite DensiEye 100 and 700
5.2. Calibration (absolute white base)
Measurements with absolute white base are required for applications which specifically call for absolute white as white base
Measurement values are referenced to absolute white for density and gray balance
In this mode, the calibration has to be checked once a week, or whenever the ambient temperature has changed by more than 10 °C / 18 °F. Recalibrate the instrument if necessary
The density standard and the filter type on the calibration card must correspond with the device information
5.2.1. Select absolute white base
Select Function Settings in the main menu
Select White Base
o The output area displays the current setting
Press the Enter button to open the pop-up screen
Set White Base to Absolute or Absolute Auto
o The output area displays the current settings
Select Return to return to the main menu
DensiEye Ref erence Manual 4 33
Page 35
X-Rite DensiEye 100 and 700
5.2.2. Calibration (Absolute)
Select Calibration in the main menu
Measure Paper White on the calibration card
Measure the solid patches Black, Cyan, Magenta and Yellow on the calibration card
o Instrument displays measurement values
Press the Enter button to open the edit screen for Black
o Edit the white value for black according to the value on the calibration card o Edit black density value according to the value on the calibration card
Continue editing the other colors as described
Select Return to return to the main menu
5.2.3. Check Calibration (Absolute)
Select Gray Balance in the main menu
Measure White Patch on the calibration card
o The difference between the white reference values and the measured values for
Cyan, Magenta and Yellow must be within ± 0.01
Select Return to return to the main menu
Select Density in the main menu
Measure White Patch on the calibration card
o The difference between the white reference value and the measured value for
black must be within ± 0.01
Measure Solid Patch B, C, M, Y on the calibration card
o The difference between the color reference values and the measured values
must be within ± 0.01
Select Return to return to the main menu
34 DensiEye Reference Manual 4
Page 36
X-Rite DensiEye 100 and 700
5.3. Density calibration reference
Do not use the X-Rite calibration cards after the expiration date
Always keep the X-Rite calibration card in its protective cover in a dry place
Treat the calibration card with care and make sure not to scratched or soiled the
measurement patches
Clean the calibration fields with a soft cloth
DensiEye Ref erence Manual 4 35
Page 37
X-Rite DensiEye 100 and 700
6. Settings
6.1. Function settings
Select Function Settings in the main menu
o The symbol area displays symbols for Pass/Fail, white base, scale gradation,
and decimal places
6.1.1. Pass / Fail indicator
The Pass/Fail indicator can be switched On and Off
A green light indicates that the measured value is within tolerance
A red light indicates that the measured value is out of tolerance
Select Pass/Fail
o The output field displays the current settings
Press the Enter button to open the pop-up screen
Toggle Pass/Fail Indicator On or Off
o The output field displays the current settings
Select Return to return to the main menu
36 DensiEye Reference Manual 4
Page 38
X-Rite DensiEye 100 and 700
6.1.2. White base
Depending on the application, you can select Paper or Paper Auto and Absolute or Absolute Auto as white base
In Auto Mode, Paper Auto causes an automatic paper white measurement
In Auto Mode, Absolute Auto causes an automatic paper white measurement
only for Dot Gain and Trapping
White Base Paper Auto corresponds to White Base Paper
White Base Absolute Auto corresponds to White Base Auto
Select White Base
o The output field displays the current setting
Press the Enter button to open the pop-up screen
Set White Base to Absolute, Paper, Absolute Auto or Paper Auto
o The output field displays the current setting
Select Return to return to the main menu
DensiEye Ref erence Manual 4 37
Page 39
X-Rite DensiEye 100 and 700
6.1.3. Scale graduation (DensiEye 700 only)
Set an application related scale graduation before measuring a Print Characteristic
Select Scale Graduation
o The output field displays the current setting
Press the Enter button to open the pop-up screen
Set Scale Graduation (5% … 50%) for Print Characteristic
o The output field displays the current setting
Select Return to return to the main menu
6.1.4. Decimal places
The number of decimal places for density values can be set to 2 or 3 digits
Select Decimal places
o The output field displays the current setting
Press the Enter button to open the pop-up screen
Set Decimal Places to 2 or 3 digits
o The output field displays the current setting
Select Return to return to the main menu
38 DensiEye Reference Manual 4
Page 40
X-Rite DensiEye 100 and 700
6.2. Device settings
Select Device Settings in the main menu
o The symbol area displays the symbols for Display Orientation, Instrument Type
and Language
6.2.1. Display orientation
The display can be rotated for right or left handed operation
Select Display orientation
o The output field displays the current setting
Confirm to rotate display for left or right hand operation
o The output field displays the current setting
Select Return to return to the main menu
6.2.2. Instrument type
This screen displays important information about your instrument like instrument type, firmware version, serial number, densitometric standard and filter
Select Instrument type
o The output field displays instrument type Information
Type: Instrument type
Firmware: Firmware version
S/N: Serial number
Standard: Densitometric standard (ISO E, ISO I or ISO T)
Filter: Pol or No (polarized or unpolarized)
Select Return to return to the main menu
DensiEye Ref erence Manual 4 39
Page 41
X-Rite DensiEye 100 and 700
6.2.3. Language
The texts can be displayed in different languages
Select Language
o The output area displays the current language
Press the Enter button to open the pop-up screen
Select Language
o The output area displays the current language
Select Return to return to the main menu
40 DensiEye Reference Manual 4
Page 42
X-Rite DensiEye 100 and 700
7. USB Interface
7.1. USB Driver installation
Procedure:
Insert DensiEye Application CD into the CD-Drive
Connect the USB cable to the DensiEye and the computer
The Plug&Play window appears. Select the driver on the CD
Confirm all messages to install the driver
The USB-Interface is available as a COM port and can be used like a serial
interface.
The assigned COM port can be changed in the property sheet of the Windows Device Manager as follows:
Open the Control Panel and select System. The System Properties opens.
Select the Hardware tab and click on Device Manager.
From the View menu, select Devices by type.
Select Ports (COM & LPT)
o Select USB serial port o Click Properties. o Select the Port Settings tab o Click Advanced o Select the desired COM port number from the list o Click OK
7.2. USB interface protocol
The device automatically reports all measurement to an external host. The report strings are in ASCII text and terminated by a carriage return <CR> and a line feed <LF>, (ASCII 13, ASCII 10).
Please refer to the document “DensiEye Data Interface Description” for detailed information.
DensiEye Ref erence Manual 4 41
Page 43
X-Rite DensiEye 100 and 700
8. Upgrade DensiEye 100 to 700
You can upgrade the DensiEye 100 to the full functionality of the DensiEye 700 by entering a special 4 digit code (for example: 1679). Please contact your X-Rite dealer with the serial number of your instrument to purchase the upgrade code.
Upgrade Procedure:
Press and hold the Reset button.
Press and hold the Down button.
Release the Reset button.
Release the Down button.
o The DensiEye is now in programming mode. The display shows 0000.
Use the Up/Down button to set the code for the first digit.
Press the Enter button to move the cursor to the next digit.
Repeat steps 5 and 6 until all digits are entered.
After the last digit is entered, the additional functions are available and can be selected
from the main menu.
42 DensiEye Reference Manual 4
Page 44
X-Rite DensiEye 100 and 700
9. Maintenance and Care
9.1 Reset
If the instrument is unresponsive, the microprocessor may be locked up. This can happen after changing batteries or another disruption. Press the Reset button (red button on the bottom side of the measuring head). The instrument restarts and displays the the status of the DensiEye as follows:
In the main menu, the settings Auto Mode (DensiEye 700) or Density (DensiEye
100) are selected
All settings remain the same
Procedure:
Press the Reset button for approximately 3 seconds
9.2 Total Reset
After a Total Reset the status of the DensiEye is as follows:
In the main menu, the settings Auto Mode (DensiEye 700) or Density (DensiEye
100) are selected
All settings are reset to factory defaults
The reference and tolerance values remain the same
The calibration is reset
The device must be recalibrated
Procedure:
Press the Reset button
Press the Enter button
Release the Reset button
Release the Enter button
DensiEye Ref erence Manual 4 43
Page 45
X-Rite DensiEye 100 and 700
9.3 Instrument type information
If you have technical questions always mention device type, serial number and firmware version
Procedure:
Select Device Settings in the main menu
Select Instrument Type
o The output field displays instrument information
Type: Instrument type
Firmware: Firmware version
S/N: Serial number
Standard: Densitometric standard (ISO E, ISO I or ISO T)
Filter: Pol or No (polarized or unpolarized)
Select Return to return to the main menu
44 DensiEye Reference Manual 4
Page 46
X-Rite DensiEye 100 and 700
9.4 Replacing the batteries
The batteries must be replaced after approximately 500,000 measurements
The DensiEye monitors the battery voltage. If the batteries are low, the
instrument displays an empty batteries symbol
Procedure:
Remove the cover from the battery compartment
Remove the two old batteries
Insert the new batteries while taking the polarity into account. The polarity and the
mounting position are illustrated in the battery compartment
Place the battery cover back on
Dispose of the old batteries in accordance with local regulations
Press the Reset button
Replace the batteries as soon as possible
Push to release
Always replace both batteries at the same time
If you are not using the instrument for an extended period of time, please
remove the batteries from the battery compartment.
DensiEye Ref erence Manual 4 45
+
+
Page 47
X-Rite DensiEye 100 and 700
9.5. Firmware upgrade
Required:
Download Software ”X-Rite Firmware Downloader”
DensiEye Firmware
USB cable
Procedure:
Connect the DensiEye with the data cable
Press the Reset button and keep it pressed
Press and release the Up button
Release the Reset button. The instrument is now in programming mode (the screen is
empty)
Start the ”X-Rite Firmware Downloader” application
Click the COM port in the Settings menu
Click Connect
Click Prog and select the firmware file. The firmware downloads within a few seconds
Close the Firmware Downloader Software window
Press and release the Reset button
Disconnect the cable
46 DensiEye Reference Manual 4
Page 48
X-Rite DensiEye 100 and 700
9.6. Warranty registration
You need to register your DensiEye to receive technical support
You can register the DensiEye online at http://www.xrite.com
9.7. Recertification
X-Rite recommends annual recertification of the DensiEye reflectance densitometer
Contact your X-Rite dealer or your nearest X-Rite Service Center for more information on the recertification process
9.8. Service
Do not attempt to repair the X-Rite DensiEye yourself
Opening the instrument without authorization will void your warranty
Contact X-Rite Support or your nearest X-Rite Service Center if you experience
problems with your instrument
DensiEye Ref erence Manual 4 47
Page 49
X-Rite DensiEye 100 and 700
10. Specifications
10.1. Functions
Automatic color recognition (CMYK)
Automatic patch type recognition (Solid, Halftone, Trapping, Gray Balance) (DensiEye 700
only)
Automatic paper white measurement
Density, Density difference
Dot Gain (shadow, half tone, highlight) (DensiEye 700 only)
Trapping (DensiEye 700 only)
Gray Balance, Gray Balance difference
Dot Area (Murray Davis) (DensiEye 700 only)
Print Characteristic (DensiEye 700 only)
Pass/Fail Indicator with correction advice
Reference Sets (predefined but customizable)
Density Calibration
Absolute White and Paper White base switchable
10.2. Technical Specification
Specification
Illumination 3 LED (red, green, blue)
Measuring geometry 45°/0° (according to DIN 5033)
Measurement aperture 3 mm or 1.6 mm (ex factory)
Density Standards ISO Status I / E / T (according to ISO 5-3) (ex factory)
Polarization filter With / without (ex factory)
Measurement range 0.00 – 3.00 D
Repeatability ± 0.01 D or ± 1%
Inter instrument agreement ± 0.02 D or ± 2%
Linearity ± 0.01 D or ± 1%
Measuring time 0.5 seconds
Graphical display 160 x 80 Pixels, 4 gray scales
Interface USB
Measurements per battery set Over 500.000
Power source 2 Alkaline 1.5 V size AA
Dimension (H x W x L) 50 x 75 x 175 mm
Weight [gr.] 400
48 DensiEye Reference Manual 4
Page 50
Corporate Headquarters - USA 4300 44th Street SE Grand Rapids, Michigan 49512 Phone 1 800 248 9748 or 1 616 803 2100 Fax 1 800 292 4437 or 1 616 803 2705
Corporate Headquarters - Europe Althardstrasse 70 8105 Regensdorf Switzerland Phone (+41) 44 842 24 00 Fax (+41) 44 842 22 22
Corporate Headquarters - Asia Room 808-810 Kornhill Metro Tower, 1 Kornhill Road Quarry Bay, Hong Kong Phone (+852) 2 568 6283 Fax (+852) 2 885 8610
Please visit www.xrite.com
for a local office near you.
2/2009
Loading...