X-Rite Ci7500, Ci7800, Ci7860 User Manual

Ci7XX0
Benchtop Spectrophotometer
User Guide
Ci7 XX0 BENCHTOP S P ECTROPHOTOMETER
Consult this documentation in all cases where the Attention symbol appears. This symbol is used to inform you of any potential HAZARD or actions that may require your attention.
CE Declaration
Hereby, X-Rite, Incorporated, declares that this Ci7XX0 series is in compliance with the essential requirements and other relevant provisions of Directive(s) 2014/35/EU (LVD), 2014/30/EU (EMC), and RoHS 2011/65/EU.
US & Canadian Approvals
This product conforms to ANSI/UL 61010-1:2012 and is listed by Intertek; Control No. 3050828 This product is certified to CAN/CSA C22.2 No. 1010.1
Federal Communications Commission Notice
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Industry Canada Compliance Statement
CAN ICES-3 (A) / NMB-3 (A)
Equipment Information
Use of this equipment in a manner other than that specified by X-Rite, Incorporated may compromise design integrity and become unsafe.
WARNING: This instrument is not for use in explosive environments.
Do not look directly into the measurement optics when the instrument is on.
For indoor use only.
CLASS 1 LASER PRODUCT
IEC 60825-1:2007
Notice: This device emits class 1 laser radiation. Class 1 laser products are classified to be safe during normal operation. Laser emission specifications: continuous wave 532 nm, ≤ 0.39 mw
No user serviceable parts in this product. All warranty and non warranty repairs should be referred to an authorized X-Rite service center. CAUTION– Class 3B Laser Radiation from internal laser module is present when instrument cover is removed. Avoid exposure to the beam.
Instructions for disposal: Please dispose of Waste Electrical and Electronic Equipment (WEEE) at designated collection points for the recycling of such equipment.
1
Ci7XX0 BENCHTOP SPECTROPHOTOMETER
Proprietary Notice
The information contained in this manual is copyrighted information proprietary to X-Rite, Incorporated.
Publication of this information does not imply any rights to reproduce or use it for purposes other than installing, operating, or maintaining this instrument described herein. No part of this manual may be reproduced, transcribed or translated into any language or computer language in any form or by any means: electronic, magnetic, mechanical, optical, manual, or otherwise; without the prior written permission of an authorized officer of X-Rite, Incorporated.
Patents: www.xrite.com/ip “© 2018, X-Rite, Incorporated. All rights reserved”
X-Rite® is a registered trademark of X-Rite, Incorporated. All other logos, brand names, and product names mentioned are the properties of their respective holders.
Warranty Information
X-Rite warrants this Product against defects in material and workmanship for a period of twelve (12) months from the date of shipment from X-Rite’s facility, unless mandatory law provides for longer periods. During such time, X-Rite will either replace or repair at its discretion defective parts free of charge.
X-Rite’s warranties herein do not cover failure of warranted goods resulting from: (i) damage after shipment, accident, abuse, misuse, neglect, alteration or any other use not in accordance with X-Rite’s recommendations, accompanying documentation, published specifications, and standard industry practice; (ii) using the device in an operating environment outside the recommended specifications or failure to follow the maintenance procedures in X-Rite’s accompanying documentation or published specifications; (iii) repair or service by anyone other than X-Rite or its authorized representatives; (iv) the failure of the warranted goods caused by use of any parts or consumables not manufactured, distributed, or approved by X-Rite; (v) any attachments or modifications to the warranted goods that are not manufactured, distributed or approved by X-Rite. Consumable parts and Product cleaning are also not covered by the warranty.
X-Rite’s sole and exclusive obligation for breach of the above warranties shall be the repair or replacement of any part, without charge, which within the warranty period is proven to X-Rite’s reasonable satisfaction to have been defective. Repairs or replacement by X-Rite shall not revive an otherwise expired warranty, nor shall the same extend the duration of a warranty.
Customer shall be responsible for packaging and shipping the defective product to the service center designated by X-Rite. X-Rite shall pay for the return of the product to Customer if the shipment is to a location within the region in which the X-Rite service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations. Proof of purchase in the form of a bill of sale or receipted invoice which is evidence that the unit is within the Warranty period must be presented to obtain warranty service. Do not try to dismantle the Product. Unauthorized dismantling of the equipment will void all warranty claims. Contact the X-Rite Support or the nearest X-Rite Service Center, if you believe that the unit does not work anymore or does not work correctly.
THESE WARRANTIES ARE GIVEN SOLELY TO BUYER AND ARE IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED
2
Ci7 XX0 BENCHTOP S P ECTROPHOTOMETER
WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION, AND NON-INFRINGEMENT. NO EMPLOYEE OR AGENT OF X-RITE, OTHER THAN AN OFFICER OF X­RITE, IS AUTHORIZED TO MAKE ANY WARRANTY IN ADDITION TO THE FOREGOING.
IN NO EVENT WILL X-RITE BE LIABLE FOR ANY OF BUYER’S MANUFACTURING COSTS, OVERHEAD, LOST PROFITS, GOODWILL, OTHER EXPENSES OR ANY INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES BASED UPON BREACH OF ANY WARRANTY, BREACH OF CONTRACT, NEGLIGENCE, STRICT TORT, OR ANY OTHER LEGAL THEORY. IN ANY EVENT OF LIABILITY, X-RITE’S MAXIMUM LIABILITY HEREUNDER WILL NOT EXCEED THE PRICE OF THE GOODS OR SERVICES FURNISHED BY X-RITE GIVING RISE TO THE CLAIM.
3
Ci7XX0 BENCHTOP SPECTROPHOTOMETER
Package
libxml2
2.7.7
Copyright (C) 1998-2003 Daniel Veillard
ncurses
fontconfig
2.8.0
Copyright © 2001,2003 Keith Packard
giflib
pixman
0.20.2
Copyright 1987, 1988, 1989, 1998 The Open Group
Software Code
This X-Rite product includes software code developed by third parties. As applicable, the terms of the code and information on obtaining access to the code used in this product, are available to you at: www.xrite.com/opensourcelicensing
Where such specific license terms entitle you to the source code of such software, X-Rite will provide upon written request via email and/or paper mail the applicable source code files for a nominal cost to cover shipping and media charges (order part number CIA-700-DVD). The source code is available from X-Rite for a period of three years after the last shipment of this product.
Please direct all inquiries for source code to the following:
Email:
opensourcelicensing@xrite.com
Mail:
Attn: Order Entry X-Rite, Incorporated 4300 44
th
Street SE
Grand Rapids, Michigan 49512
Phone:
888-826-3059, option 1
GPL and LGPL
The GPL code and LGPL code used in this product are distributed WITHOUT ANY WARRANTY and is subject to the copyrights of one or more authors. For details, see the GPL code and the LGPL code for this product and the terms of the GPL and LGPL.
http://www.gnu.org/copyleft/gpl.html http://www.gnu.org/licenses/lgpl.html
MIT
This product uses software under the MIT license:
Version
5.9 Copyright (c) 1998-2010,2011 Free Software Foundation, Inc.
4.1.6 Copyright (c) 1997 Eric S. Raymond
Owner
Copyright 1987, 1988, 1989 Digital Equipment Corporation Copyright 1999, 2004, 2008 Keith Packard
Copyright 2000 SuSE, Inc. Copyright 2000 Keith Packard, member of The XFree86 Project,
Inc. Copyright 2004, 2005, 2007, 2008 Red Hat, Inc. Copyright 2004 Nicholas Miell Copyright 2005 Lars Knoll & Zack Rusin, Trolltech
4
Ci7 XX0 BENCHTOP S P ECTROPHOTOMETER
Copyright 2005 Trolltech AS
Mesa
7.10.3
Copyright (C) 1999-2005 Brian Paul
libICE
1.0.7
Copyright 1993, 1998 The Open Group
libpciaccess
0.12.1
Copyright IBM Corporation 2006, 2007
libSM
1.2.0
Copyright 2002 Sun Microsystems, Inc.
libX11
1.4.4
Copyright (C) 2003-2006,2008 Jamey Sharp, Josh Triplett
libXau
1.0.6
Copyright 1988, 1993, 1994, 1998 The Open Group
libxcb
1.7
Copyright (C) 2001-2006 Bart Massey, Jamey Sharp, and Josh
libXcomposite
0.4.3
Copyright © 2001,2003 Keith Packard libXdamage
1.1.3
Copyright © 2001,2003 Keith Packard libXdmcp
libXext
1.2.0
Copyright 1986, 1987, 1988, 1989, 1994, 1998 The Open Group
Copyright 1991,1993 by Digital Equipment Corporation, Maynard,
Copyright 2007 Luca Barbato Copyright 2008 Aaron Plattner, NVIDIA Corporation Copyright 2008 Rodrigo Kumpera Copyright 2008 André Tupinambá Copyright 2008 Mozilla Corporation Copyright 2008 Frederic Plourde Copyright 2009 Sun Microsystems, Inc.
Copyright Eric Anholt 2006 Copyright (c) 2007, 2008, 2009, Oracle and/or its affiliates. Copyright 2009 Red Hat, Inc.
Copyright 1993, 1998 The Open Group
Copyright © 2009 Red Hat, Inc. Copyright 1990-1992,1999,2000,2004,2009,2010 Oracle and/or
its affiliates
Triplett.
Copyright (c) 2006, 2007, Oracle and/or its affiliates. All rights reserved.
Copyright © 2007 Eric Anholt
1.1.0 Copyright 1989, 1998 The Open Group
Copyright (c) 1996 Digital Equipment Corporation, Maynard,
Massachusetts.
Copyright (c) 1997 by Silicon Graphics Computer Systems, Inc. Copyright 1992 Network Computing Devices
Massachusetts, and Olivetti Research Limited, Cambridge, England Copyright 1986, 1987, 1988 by Hewlett-Packard Corporation
Copyright (c) 1994, 1995 Hewlett-Packard Company Copyright Digital Equipment Corporation, 1996
Copyright 1999, 2005, 2006 Sun Microsystems, Inc. All rights
reserved. Copyright (c) 1989 X Consortium, Inc. and Digital Equipment
Corporation.
Copyright (c) 1992 X Consortium, Inc. and Intergraph
Corporation. Copyright (c) 1993 X Consortium, Inc. and Silicon Graphics, Inc.
5
Ci7XX0 BENCHTOP SPECTROPHOTOMETER
Copyright (c) 1994, 1995 X Consortium, Inc. and Hewlett­libXft
2.2.0
Copyright © 2001,2003 Keith Packard
libXpm
3.5.9
Copyright (C) 1989-95 GROUPE BULL
libXrandr
1.3.2
Copyright © 2000, Compaq Computer Corporation,
libXrender
0.9.6
Copyright © 2001,2003 Keith Packard
libXt
1.1.1
Copyright © 2001,2003 Keith Packard
libXxf86vm
1.1.1
Copyright (c) 1995 Kaleb S. KEITHLEY
libdrm
2.4.26
Copyright 1999 Precision Insight, Inc., Cedar Park, Texas.
xtrans
1.2.6
Copyright 1993, 1994, 1998, 2002 The Open Group
libpthread-stubs
0.3
Copyright (C) 2006 Diego Pettenò
qextserialport
1.2
Copyright (c) 2000-2003 Wayne Roth
Packard Company.
libXfixes
4.0.5 Copyright © 2001,2003 Keith Packard Copyright (c) 2006, Oracle and/or its affiliates. All rights
reserved.
Copyright (C) 1998 Arnaud LE HORS Copyright (C) 19896 Lorens Younes
Copyright © 2002, Hewlett Packard, Inc. Copyright © 2000 Compaq Computer Corporation, Inc. Copyright © 2002 Hewlett-Packard Company, Inc. Copyright © 2006 Intel Corporation Copyright © 2008 Red Hat, Inc. Copyright © 2000 Compaq Computer Corporation, Inc. Copyright © 2002 Hewlett Packard Company, Inc. Copyright © 2006 Keith Packard
http://opensource.org/licenses/MIT
Copyright © 2000 SuSE, Inc.
Copyright (c) 1993, 2011, Oracle and/or its affiliates. All rights
reserved. Copyright 1987, 1988 by Digital Equipment Corporation Copyright 1987, 1988, 1998 The Open Group
Copyright 2000 VA Linux Systems, Inc., Sunnyvale, California
Copyright 1993, 1994 NCR Corporation - Dayton, Ohio, USA Copyright (c) 2002, 2005, Oracle and/or its affiliates. All rights
reserved. Copyright 1996 by Sebastien Marineau and Holger Veit Copyright © 2003 Keith Packard, Noah Levitt
Copyright (c) 2004-2007 Stefan Sander Copyright (c) 2007 Michal Policht Copyright (c) 2008 Brandon Fosdick Copyright (c) 2009-2010 Liam Staskawicz Copyright (c) 2011 Debao Zhang
6
BSD
Package
Owner
zlib
1.2.5
Copyright (C) 1995-2010 Jean-loup Gailly and Mark Adler
expat
2.0.1
Copyright (c) 1998, 1999, 2000 Thai Open Source Software Center
2005, 2006 Expat maintainers.
pcre
8.2
Copyright (c) 1997-2011 University of Cambridge
Copyright(c) 2010-2011 Zoltan Herczeg
orc
0.4.14
Copyright 2002 - 2009 David A. Schleef
freetype
2.4.7
Copyright 2001, 2002 by David Turner, Robert Wilhelm, and Werner Lemberg
libjpeg
8c
Copyright (C) 1994-2010, Thomas G. Lane, Guido Vollbeding
libtiff
Silicon Graphics, Inc.
libogg
1.2.2
Copyright (c) 2002, Xiph.org Foundation
libvorbis
libtheora
1.1.1
Copyright (C) 2002-2009 Xiph.org Foundation
OpenCV
2.3.1
Copyright (C) 2000-2008, Intel Corporation, all rights reserved. Copyright (C) 2009, Willow Garage Inc., all rights reserved.
Package
Owner
Copyright 2007 - 2009 Martin Heinrich
Copyright: 2012 Nikolaos Ftylitakis
This product uses software under the BSD license:
Version
Ltd and Clark Cooper Copyright (c) 2001, 2002, 2003, 2004,
Copyright (c) 2007-2011, Google Inc.
3.9.5 Copyright (c) 1988-1997 Sam Leffler Copyright (c) 1991-1997
1.3.2 Copyright (c) 2002-2008 Xiph.org Foundation
Ci7 XX0 BENCHTOP S P ECTROPHOTOMETER
Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:
1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.
2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.
THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS "AS IS" AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE COPYRIGHT HOLDER OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
Apache
This product contains Apache licensed software.
Version
Log4qt 0.3
QZXing 1.2
http://www.apache.org/licenses/LICENSE-2.0.txt
7
Ci7XX0 BENCHTOP SPECTROPHOTOMETER
Table of Contents
Overview 9
Packaging Information 10 Instrument Setup 10 Powering the Instrument 10 Sample Preview Method 11 Front Panel 11 Sample Shelf 14 Sample Holder 14 Sample Door 15 Specular Control 15 UV Control 16 Aperture Control 16
Calibrating 18
Calibration Notes 18 Reflectance Calibration Procedure 18 Green Tile Color Check (optional) 19 Transmission Calibration Procedure (excludes Ci7500 series) 20
Transmission Calibration Setup 20 Total Calibration 21 Direct Calibration 22 R/T Calibration 23
UV Calibration Procedure 24
Measuring 25
Reflectance Measurements 25 Transmission Measurements (excludes Ci7500 series) 26
Total Transmission Measurement 26 Direct Transmission Measurement 28 R/T (Reflection/Total Transmission) Measurement 30 Liquid Measurements 32 Haze Measurements 33
Appendices 34
Service Information 34 Cleaning the Instrument 35
Cleaning Quick Reference Table 35 General Exterior Cleaning 36 Cleaning the Transmission Compartment 36 Cleaning the Calibration Tiles 36 Cleaning the Black Plastic Blocking Panel 36 Cleaning the Sphere 36 Cleaning the Aperture Plates 37 Cleaning the Transmission White Plaque 37 Cleaning the Black Trap 37
Cleaning the UV Calibration Plaque 38 Replacing the Fuse 39 Adjusting the Sample Holder Dampening 39 Troubleshooting 40 Specifications 41 Spare Parts and Accessories 43 System Repacking Instructions 44
8
Ci7 XX0 BENCHTOP S P ECTROPHOTOMETER
Sample holder
Button and indicator
membrane label (refer to
for a
Transmission
Accessory drawer (white tile, green tile, black trap, aperture plates, UV cal plaque)
Ci7XX0 instrument can
Communication,
Front lifting grip
Rear lifting grip
Locking pin
Sample door
OVERVIEW
The Ci7XX0 is a reflectance / transmittance sphere benchtop spectrophotometer offering agreement with these X-Rite/GretagMacbeth instruments: 7000A, Color i7, and Color i5. The Ci7XX0 Spectrophotometer includes these features:
Multiple areas of sample view in both reflectance and transmittance.
Self-adjusting, dual zoom lens to manage the measurement area and eliminate
configuration errors between aperture plates and lens position.
External video monitor viewing, preview video in the software application, and hinged sample door preview for measurement accuracy and targeting purposes.
Haze measurement mode (excludes Ci7500 series).
Automated ultra violet (UV) adjustment for measuring fluorescent or optically brightened
samples.
USB interface to PC and software applications.
Indicator panel display with dual buttons for standard and sample measurements, and
warning indicators for calibration interval.
Dampened sample arm holder that prevents sample damage.
Embedded white sample holder ceramic. Ideal for reflectance measurements that require a
white backing, such as textiles, plastics or other less than opaque samples (Standard on Ci7800/Ci7860 instrument and optional on the other models).
Supports embedded NetProfiler.
access cover
Front Panel
description)
line cord and power switch
also operate in the upright position with optional kit.
9
Ci7XX0 BENCHTOP SPECTROPHOTOMETER
Packaging Information
If you are reading this documentation, you have already followed the instructions detailed in the Ci7XX0 installation sheet that was included in the shipping carton. Please keep the installation sheet for reference. Keep your shipping carton in case you need to return your instrument to the factory for service.
Your instrument packaging should contain all the items listed below. If any of these items are missing or damaged, contact X-Rite or your Authorized Representative.
Software box containing the Ci7XX0 Driver and Documentation CD along with any other optional software you may have ordered such as Color iControl and NetProfiler.
Ci7500 Series, Ci7600 Series, or Ci7800 Series Spectrophotometer
Certificate of performance
Installation instructions sheet
USB cable
AC power cord
Sample shelf with thumb screws
Transmission kit (optional on Ci7800 Series & Ci7600 Series)
The following accessories are stored in the instrument accessory drawer.
Round white calibration tile
Round green performance tile
Black trap for calibration
Ci7800 Series aperture plates: 25 mm, 17 mm, 10 mm, 6 mm, (3.5 mm optional)
Ci7600 Series aperture plates: 25 mm, 10 mm, 6 mm, (3.5 mm optional), (17 mm optional) Ci7500 Series aperture plates: 25 mm, 10 mm
White UV plaque
Instrument Setup
Please follow the instructions detailed in the Ci7XX0 Installation Instructions sheet found as the first item in your shipping box when you opened it. A PDF file of the Installation Instructions also can be found on the Ci7XX0 Driver and Documentation CD.
Powering the Instrument
The power on-off AC switch is located on the back of the instrument. Press the top of the rocker switch to turn on the instrument and wait 50 seconds for the instrument to boot up. If the instrument does not power up after approximately 50 seconds, check the power connections to the unit and main power availability (breakers, fuses). If these connections are fine, check the instrument’s fuse. Refer to “Fuse Replacement” in the Appendix.
A power “standby” button is located on the front panel of the instrument. This button is used to power down the instrument into a low power state, or wake the instrument after it goes into a power down state. A power on condition is designated by a solid green light indicator. Simply press the power “standby” button to wake or power down the instrument into “standby” mode.
Discontinue use if AC cord is damaged. Ensure AC cord ratings meet or exceed the instrument ratings (see Specifications section in the Appendices).
10
Ci7 XX0 BENCHTOP S P ECTROPHOTOMETER
Specular Included
Specular Excluded
Lens
Standard button and indicator
Preview button and indicator
Trial button and indicator
Calibration
Transform
NetProfiler
Total Transmission
Direct Transmission
Haze
D65 Illuminate
UV Excluded
400, 420, 460 nm
cut filters
Custom UV (user defined)
Power “standby” button and indicator
Aperture plate
Sample Preview Method
The instrument has three sample preview methods:
Sample Door
This method enables you to open the sample door and look at the sample position in the viewport. You can make manual adjustments if necessary to obtain optimum sample alignment. Refer to the instructions on using the Sample Door Preview Method.
External Video Monitor Feature
This method requires an external preview video monitor. The monitor is plugged into the video connector located on the back of the instrument. After the monitor is connected, the Preview button on the spectrophotometer can be pushed to turn live preview video on and off. Video on/off is the default functionality of the Preview button when using a reflectance measurement mode. However, your software application may assign a different behavior to the button for various reasons.
Live Preview Video in PC Software Application
This method requires a software application, such as Color iQC to provide a means to view video within the application. Follow the instructions in your application to preview video for sample alignment.
Note: DO NOT power down the instrument into “Standby” mode when preview video is actively streaming in the application. Doing so may result in an unstable operation condition.
Front Panel
The front panel is used to initiate measurements and also indicate a variety of instrument conditions, such as calibration status, measurement status, etc. Below is description for each indicator and button.
11
Ci7XX0 BENCHTOP SPECTROPHOTOMETER
Lens and Aperture Plate Indicators (*, 3.5, 6, 10, 17, 25)
The top row of LEDs illuminates to indicate the current lens position. The “*” illuminates when the lens is in a custom position. The LED will light red if the unit is configured with a lens position larger than the installed aperture plate size.
The bottom row of LEDs illuminates to indicate the currently installed aperture plate. The “*” illuminates for custom plates and transmission aperture plates (with white reflective ring), and blinks if the instrument could not detect the installed plate. If auto is enabled for the lens, then the installation of an aperture plate automatically sets the lens to match the aperture plate.
Standard Button and Indicator
The standard button is used to initiate a standard measurement. Note: The button must be supported by your application software and function as programmed.
The indicator illuminates green when programmed to perform a measurement. It may also be programmed to flash green to support features in your application.
Preview Button and Indicator
The preview button is used in conjunction with the external video monitor preview feature.
The indicator illuminates green when it is activated.
The video preview is toggled when a reflectance measurement mode is selected. When a
transmission measurement mode is selected, the button will toggle the green transmission targeting laser on and off.
The behavior of the button may also be programmed in your software application.
Trial Button and Indicator
The trial button is used to initiate a trial measurement. Note: The button must be supported by your software application and function as programmed.
The indicator illuminates green when programmed to perform a measurement. It may also be programmed to flash green to support features in your application.
NetProfiler
Indicator Off: NetProfiler feature is not enabled.
Solid Green: NetProfiler subscription is currently activated.
Solid Amber: The profile has expired and updating is recommended.
Calibration
Solid Red: Calibration is required
Solid Green: White, black, and UV calibration are not required at this time
Solid Amber: White and black calibration are not needed, but UV calibration has expired.
Measurements can still be taken in this condition. It is up to the user to decide if the UV calibration should be updated at this time.
Transform
Indicator Off: Transform feature in not enabled
Solid Green: Transform feature is activated
Specular Included (SCI)
Illuminates when a specular included measurement is selected.
12
Ci7 XX0 BENCHTOP S P ECTROPHOTOMETER
Specular Excluded (SCE)
Illuminates when a specular excluded measurement is selected.
Total (T) – excludes Ci7500 series
Illuminates when a total transmission measurement is selected.
Direct (D) - excludes Ci7500 series
Illuminates when a direct transmission measurement is selected.
Haze (H) - excludes Ci7500 series
Illuminates when a haze measurement is selected.
D65 Illuminate
Illuminates when it is UV calibrated to a D65 UV illumination condition.
UV Excluded (
Illuminated when one of the UV filters fully blocks the UV component in the lamp to fully exclude UV illumination.
Custom (
Illuminated when a user-defined D65 or other UV condition is selected.
400
The 400 nm UV cutoff filter is in use. This filter is adjustable and can be used to calibrate
420 - excludes Ci7500 series
The 420 nm UV cutoff filter is in use. This optional filter is adjustable and can be used to
460 - excludes Ci7500 series
The 460 nm UV cutoff filter is in use. This optional filter is not adjustable and can only be
Power “standby” button and indicator
Refer to Powering the Instrument earlier in this section.
*
UV conditions.
calibrate UV conditions.
set to fully exclude UV illumination.
)
)
13
Ci7XX0 BENCHTOP SPECTROPHOTOMETER
Screw holes
Thumbscrew
Sample shelf
Thumbscrew
Sample Shelf
A sample shelf is provided to help align a sample with the measurement port. It is also useful for providing consistent sample placement when multiple samples require a measurement at the same location.
To install the sample shelf, position the shelf’s elongated slots over the two holes in the sample door and attach it with the thumb screws provided.
Adjust the shelf by loosening the thumbscrews and sliding the shelf up or down.
Sample Holder
The sample holder can be locked in a down position. This is convenient when large samples are measured, or when the aperture plate is changed or the sample holder is removed.
Simply open the sample holder to its maximum position to lock into position.
14
Ci7 XX0 BENCHTOP S P ECTROPHOTOMETER
Pull from top Sample door
Observe sample here through viewport opening
Sample holder
Sample Door
The sample door can be opened for a completely unobstructed view of the sample presented at the view port. When completely open, the sample door allows you to adjust the sample at the view port for optimal placement.
To use the sample door:
1. Open the sample holder by pulling it towards you. Place your sample at the view port and close the sample holder.
2. Open the door from the top by using your fingers, and gently lower the door to the complete open position. See the figure below.
3. With the sample door completely open, view the sample at the view port and make sure the sample is in an optimal position for measuring.
4. Close the sample door and prepare for the measurement.
Specular Control
You can change the specular component setting on the spectrophotometer for reflectance measurements using your software application. Select the desired specular component measurement mode; Included (SCI), Excluded (SCE), or Dual Mode (SCE/SCI). The indicator for
15