X-Rite 968, 948 User Manual

X-Rite® 948/968
Color Measurement Instruments
Operation Manual
CAUTION: Operational hazard exists if AC adaptor other than X-Rite SE30-61 (115V) or SE30-62 (230V) is used.
VORSICHT: Es besteht Betriebsgefahr bei der Verwendung von einem Adapter außer X-Rite SE30-61 (115 U) oder SE30-62 (230 U).
AVISO: No use otro adaptador C.A. que no sea la pieza X-Rite SE30-61 (115V) o SE30-62 (230V), por el riesgo de mal funcionamiento del equipo.
ATTENTION: Ne pas utiliser un autre adaptateur que la piéce X-Rite SE30-61 (115V) ou SE30-62 (230V).
AVVISO: Non usare un altro adattatore C.A. che non è del pezzo X-Rite SE30-61 (115V) o SE30-62 (230V), per il rischio di malfunzionamento dell’apparecchio.
NOTE: Shielded interface cables must be used in order to maintain compliance with the desired FCC and European emission requirements.
USE ONLY: AA NICad batteries that are 600/700mAhr rated, six required. Other types may burst causing personal injury.
AUFGEPASST: Verwenden Sie nur AA Nicad Akkus von 600/700mAhr (Milliampere/Stunde) Nennstrom (6 Stück erforderlich). Mit anderen Akkus läuft die Gefahr von Explosion und Verletzung.
ATENCION: Use solamente las pilas de AA NiCad (se requiere seis) con condiciones normales de funcionamiento 600/700mAhr (horas miliamperios). Es posible que los otros tipos puedan estallar y causar daños corporales.
ATTENTION: Utiliser seulement les batteries AA NICad de 600/700mAhr (milliampère/heure) nominale (6 pièces nécessaire). Il y a danger d’explosion et des blessures avec des autres types.
ATTENZIONE: Usare solamente gli accumulatori al AA NiCad (si richiede sei) con le condizioni normali di funzionamento 600/700mAhr (ore milliamperi). E possibile che altri tipi possano scoppiare e causare danno personale.
WARNING: This instrument is not for use in explosive environment. WARNUNG: Das Gerät soll in einer explosiven Umgebung NICHT verwendet
werden.
ADVERTENCIA - NO use este aparato en los ambientes explosivos. ATTENTION: Cet instrument NE DOIT PAS être utilisé dans un environnent
explosif.
AVVERTIMENTO - NON usare questo apparecchio in ambienti esplosivi.
FCC
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Canada
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
The Manufacturer: X-Rite, Incorporated Der Hersteller: 3100 44th Street, S.W. El fabricante: Grandville, Michigan 49418 Le fabricant: Il fabbricante:
Declares that: Spectrophotometer gibt bekannt: 948, 968 advierte que: avertit que: avverte che:
is not intended to be connected to a public telecommunications network. an ein öffentliches Telekommunikations-Netzwerk nicht angeschlossen werden soll. no debe ser conectado a redes de telecomunicaciones públicas. ne doit pas être relié à un réseau de télécommunications publique. non deve essere connettuto a reti di telecomunicazioni pubblici.
DECLARATION OF CONFORMITY
EN50081-1: 1992 EN55022 Class B Radiated, Class B Conducted,
complies to
EN45014:1989
_____________________________________________________________________________________
Manufacturer’s Name: Manufacturer’s Address:
X-Rite, Incorporated. 3100 44th Street, S.W. Grandville, Michigan 49418 U.S.A.
European Contact Name: European Contact Address:
X-Rite GmbH Stollwerckstraße 32 51149 Köln, Deutschland (49) 2203-91450 FAX (49) 2203-914519
Model Name
Model No.
Year of Manufacture
Directive(s) Conformance
Test Standards Declared
Spectrophotometer
948 968
EMC
89/336/EEC
EN50082-1: 1992 IEC801-2,-3,-4 and EN60950
1995
LVD
73/23/EEC, 93/68/EEC
X-Rite, Incorporated and the undersigned, hereby declare under our sole responsibility that the product to which this declaration relates, conform to the above listed Directive(s) and Standard(s).
Grandville, Michigan
4-11-96
Location Date Mark P. Spliedt
Product Support Engineering Manager
NOTE: The device complies to the product specifications for the Low Voltage Directive when furnished with the 230VAC AC Adapter (X-Rite P/N SE30-62), and to UL Standards when furnished with the 115VAC AC Adapter ( X-Rite P/N SE30-61).
Dear Customer:
Congratulations! We at X-Rite, Incorporated are proud to present you with an X-Rite Color Measurement Instrument. This instrument represents the very latest in microcontrollers, integrated circuits, fiber optics, and display technology. As a result, your X-Rite 948/968 is a rugged and reliable instrument whose performance and design exhibit the qualities of a finely engineered instrument, which is not surpassed.
To fully appreciate and protect your investment, we suggest that you take the necessary time to read and fully understand this manual. As always, X-Rite stands behind your unit with a one year limited warranty, and a dedicated service organization. If the need arises, please don’t hesitate to call us.
Thank you for your trust and confidence.
Ted Thompson
Chairman and CEO
i
Table of Contents
Opening Letter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
Proprietary Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
General Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
What To Do First!. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
1. Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Packaging Check list. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Shoe Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Battery Charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. Positioning Techniques. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1 Keyswitch Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Function Selection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Illuminant/Observer Selection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.4 Absolute Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5 Difference Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.5.1 Entry of Reference Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.5.2 Taking A Difference Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.6 CMC Difference Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.7 Measurement Averaging Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.8 Store Data Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.9 Display Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4. Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.1 Positioning the Instrument on the X-Rite Calibration Standard . . . . . . . . . 29
4.2 Calibrating to a White Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5. Setting System Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.1 Averaging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2 Operation Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.3 RS232 I/O Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.4 Format Output Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6. Printing Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7. Changing Apertures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
8.1 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
8.2 Optics Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.3 Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8.4 Target Window Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
8.5 Lamp Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Appendix
A1 - Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
A2 - Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
A3 - Factory Presets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
A4 - Spectrophotometer Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
A5 - Color Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
ii
PROPRIETARY NOTICE: The information contained in this manual is derived from patent
and proprietary data of X-Rite, Incorporated. This manual has been prepared expressly for the purpose of assisting in the use and general maintenance of this instrument.
Publication of this information does not imply any rights to reproduce or use this manual for purposes other than installing, operating, or maintaining this instrument. No part of this manual may be reproduced, transcribed, transmitted, stored in a retrieval system, or translated into any language or computer language, in any form or by any means, electronic, magnetic, mechanical, optical, manual, or otherwise, without the prior written permission of an officer of X-Rite, Incorporated.
These provisions are intended to state all of the rights and responsibilities between X-Rite, Incorporated and the customer. They supersede all warranties, expressed or implied, and whether of merchantability, fitness or otherwise. The remedies contained in this manual are exclusive. Customer and X-Rite, Incorporated waive all other remedies, including but not limited to consequential damages.
This instrument is covered by one or more of the following U.S. and foreign patents:
U.S. patent #4,080,075, #4,591,978, and other patents pending.
Copyright 1990 by X-Rite, Incorporated
"ALL RIGHTS RESERVED"
Operating Keys
Target Window
Serial Number Label
LIMITED WARRANTY: X-Rite, Incorporated warrants each unit manufactured to be free
of defects in material and workmanship (excluding Ni-Cad batteries) for a period of twelve months. THERE ARE NO WARRANTIES OF MERCHANTABILITY OR FITNESS. THIS WARRANTY OBLIGATION IS LIMITED TO SERVICING THE UNIT RETURNED TO X-RITE, INCORPORATED or AN AUTHORIZED SERVICE DEALER FOR THAT PURPOSE. The unit shall be returned with transportation charges prepaid. If the fault has been caused by misuse or abnormal conditions of operations, repairs will be billed at a nominal cost. In this case, an estimate will be submitted before work is started, if requested. Always include serial number in any correspondence concerning the unit. The serial number is located on the bottom housing.
X-Rite, Incorporated offers a repair program for instruments out of warranty. For more information, contact X-Rite Instrument Services Department.
This agreement shall be interpreted in accordance with the laws of the State of Michigan and jurisdiction and venue shall lie with the courts of Michigan as selected by X-Rite, Incorporated.
2 x 16 Character Display
RS232 I/O Port
AC Adaptor Charger Jack
iii
General Description
The X-Rite 948 is a reflection spectrocolorimeter. The 948 measures spectral
reflectance from 400nm to 700nm in 20nm intervals. It has a 0° illumination angle, a 45° viewing angle, and features a dual-beam, single light pulse compensation method to insure accuracy.
The 948 calculates and displays colorimetric data. In addition, the 948 can output
(to a printer or computer) colorimetric data. An optional software package,
ColorStart®, can collect, sort, view, and analayze L*a*b* data using an IBM (or
100% compatible) computer.
The X-Rite 968 is a reflection spectrophotometer. The 968 measures spectral
reflectance from 400nm to 700 nm in 20nm intervals. It has a 0° illumination angle, a 45° viewing angle, and features a dual-beam, single light pulse compensation method to insure accuracy.
The 968 calculates colorimetric and spectrophotometric data. The 968 displays
colorimetric data, and can output (to a printer or computer) colorimetric and spectral data. An optional software package, SpectroStart®, can collect, sort, view, and analayze L*a*b* and spectral data using an IBM (or 100% compatible) computer.
The 948 and 968 measure spectral reflectance, absolute color or color difference,
and absolute index or index difference for Whiteness and Yellowness.
XYZ or RGB & XYZ or RGB (CIE 1931 or RGB = X%Y%Z%) > Z% is one method used to measure paper brightness. Refer to Section 5.2 to activate
RGB (X%Y%Z%).
Yxy & Yxy (CIE 1931)
L*a*b* or Lab & L*a*b* or Lab (CIE 1976 or HUNTER)
L*u*v* & L*u*v* (CIE 1976)
L*C*h° & L*C*H* (CIE 1976, ab, CMC, or uv)
Whiteness & Tint & Whiteness & Tint (CIE & ASTM E313)
Yellowness & Yellowness (ASTM E313 & ASTM D1925)
The 948 and 968 have eight different CIE Illuminants for calculations under varying lighting conditions. In addition, you can select the CIE 1931 2° Observer or the CIE 1964 10° Observer.
Illuminant C
Illuminant D65
Illuminant D
Illuminant A
Illuminant F2 (Cool White Fluorescent)
Illuminant F7 (Narrow-band White Fluorescent)
Illuminant F11 (TL84 / Wide-band Fluorescent)
Illuminant F12 (Ultralume 3000)
50
iv
General Description . . . continued
Twenty four different reference values (Ref 01 - Ref 24) can be entered into the 948 and 968 manually (numerically) or by measuring a reference sample. The
exceptions are that the reference values for Whiteness and Yellowness can only be entered by measurement. During reference entry all color spaces are automatically updated when a change has been made in one of them.
The 948 and 968 can function as color difference meters. By entering in your reference (by measurement or numerically), the units will indicate the amount of difference between the reference and the measured sample.
The store data feature allows the units to store up to five hundred measurements for transferring to a printer or computer (via RS232) at a later time.
The CMC difference feature provides a single numeric value (E
CMC
) which
describes the color difference between a sample and a standard.
The averaging feature allows the units to make several measurements from the same and/or different locations on a sample or reference. This will allow for a better overall average of a color.
The units are powered by six AA rechargeable batteries, or by the AC adaptor/charger provided. Additionally, the units retain calibration and reference values when turned off, or if the Ni-Cad batteries are discharged.
X-Rite®, ColorStart®, SpectroStart®, and QA-Master® are registered trademarks of X-Rite, Incorporated.
IBM is a registered trademark of International Business Machines Corp.
Macintosh is a registered trademark of Apple Computer, Incorporated.
All other logos, product names, registered trademarks, and trademarks mentioned are the property of their respective holders.
v
User Interface
This section will familiarize you with the typographical conventions, display functions, and general terms used in this manual.
In the text portion of this manual the 968 keys are shown with brackets on
both sides and in boldface. Ex.,
[FUNCTION], [ILLUM]
When a key is to be momentarily pressed, the statement "press" will be used.
Ex., Press
[FUNCTION].
When a key is to be held depressed until another action occurs, the statement
"hold down" will be used. Ex., Hold down
ILLUMINANT"
is displayed.
[ILLUM]
Information that will appear in the display
window will be shown with quotation marks on each side and in boldface. Ex.,
"SELECT ILLUMINANT"
FUNCTION ILLUM
, and
until
[DIF]
"SELECT
SELECT
ILLUMINANT
CAL
.
DIF
The symbols t and s represent the
blue and red arrows on the 968. In calibration and reference value entry the arrows indicate which direction a value can be changed. Holding
[DIF]
depressed
Y 2 . 8 7 Z 2 . 4 1
C
FUNCTION ILLUM
CAL
X 4 . 1 2
2X Y Z
DIF
and pressing [t ] decreases the value and
[s ] increases the value.
The term "cursor" represents a black
rectangle that will blink next to or over a
WHT
x . 3 0 9 0
C
2
CURSOR
Y 8 5 . 5 5 y . 3 1 8 1
value or function in the display. In most cases it means that value is active and can be altered via the [t] and [s] keys.
A "hand" indicates important notes and
possible operations that need to be performed before the normal operation.
When a procedure is continued on the next
page an arrow will appear in the bottom right hand corner of the page.
Illuminant/observer combinations are expressed by special notations (Ex.,
D
indicates illuminant D, 6500°k, and 2° observer).
652
vi
What To Do First!
See how to unlock the shoe and charge the batteries...read Section 1 ­Getting Started.
See how the positioning of the instrument during measurement affects the reading...read Section 2 - Positioning Techniques .
Calibrate your instrument...read Section 4 - Calibration.
Setup your instrument. You can set the operating parameters, RS232 parameters, and format output parameters of your unit. Remember, you can lock out of the display any function you do not want to use...read Section 5
- Setting System Parameters.
Learn the basic functions...read Section 3 - Operation.
vii
1. Getting Started
1.1 Packaging Check List
After removing the instrument from the shipping carton, inspect for possible damage. If any damage is noted, contact the transportation company immediately. Do nothing more until the carrier’s agent has inspected the damage.
If damage is not evident, check and make sure that all items are included (Refer to the parts list below, and following page for the packaging illustration).
Your unit was packaged in a specially designed carton to assure against damage. If reshipment is necessary, the instrument should be packaged in the original carton. If the original carton is not available, a new one can be obtained from X-Rite Inc. Refer to the packaging drawing on the following page (items 1, and 2).
PACKAGING PARTS LIST
1
PACKAGING
2
1.2 Shoe Lock
The shoe can be locked next to the housing for ease of storage.
Unlocking Shoe
1. Hold shoe [1] against unit.
2. Slide black lock button [2] on bottom of unit towards the back until it stops, then slowly release shoe [1].
Locking Shoe
1. Hold shoe [1] against unit.
2. Slide black lock button [2] towards the front until it stops, then release shoe [1] .
[1]
[2]
3
1.3 Battery Charging
> THE UNIT SHOULD BE CHARGED BEFORE USE. The unit can be operated while the
batteries are being charged. Before using make sure the voltage indicated on the AC adaptor complies with the AC line voltage in your area. If not contact your X-Rite dealer.
> The unit should be fully charged in 14 hours. Note: If your unit has not been used for
several weeks recharge for approximately 24 hours. > Charging the batteries for less than 14 hours will reduce the operating time of the unit. > The best method to obtain the maximum battery life is to:
- Always run the unit down to the point where the "BATTERIES LOW" message is displayed. Then charge the unit before the next message "CHARGE BATTERIES" is displayed.
- Leaving the unit plugged into the AC wall outlet for extended periods of time (over 48 hours) may shorten the battery life.
- If you are going to store the unit for an extended period of time (over 6 months) you should remove the Ni-Cad batteries from the unit.
1. Plug the small connector end of the adaptor into back of unit.
2. Plug the adaptor into AC wall outlet.
ADAPTOR
WALL OUTLET
SMALL CONNECTOR
> This unit retains calibration, reference values, and all other data when the unit turns off
after 45 seconds of non-use (in battery operation), or if the Ni-Cad batteries are discharged.
1.4 Applying Power
During battery operation power is applied automatically when a measurement is taken or if a key is pressed. It automatically shuts off (within 45 seconds) if no keys are pressed or no further measurements are taken. The unit will not automatically shut off if the AC adaptor is connected.
4
2. Positioning Techniques
The variety of items that the 948 and 968 can measure is almost endless. However, in order to obtain accurate and repeatable measurements, the bottom of the shoe must be:
Parallel with the surface to be measured if the surface is flat.
Tangent to the surface to be measured if the surface is curved.
The reason for this is that any movement during measurement can cause the reading to vary. To obtain the most accurate and repeatable measurements, there are a few guide rules you can follow.
If the item to be measured is smaller than the shoe, you may want to make a platform (the same height of the item) for the instrument to sit on. If the item to measure is curved, you may want to make a jig for the item to rest in.
Shown below and on the next page are some examples of methods used to accomplish this. Example 1 shows a platform for measuring a paint chip. Example 2 shows a jig being used to measure a cup. Example 3 shows a jig being used for measuring a small knob.
Note: A mounting fixture (jig) is available from X-Rite (P/N 968-80). The jig can measure objects that are a maximum of 4" inches in width, or 2" to the center of the object. Refer to Appendix A-4 for further information about this mounting fixture.
Example 1- Measuring a Paint Chip
Paint Chip
CORRECT
Platform
INCORRECT
5
Coffee Cup
Example 2 - Measuring a Coffee Cup
Jig
CORRECT
Example 3 - Measuring a Small
Knob
Jig
CORRECT
6
3. Operation
3.1 Key Descriptions
FUNCTION ILLUM
CAL
DIF
[FUNCTION]/[t]
Selects the function that will be used for measurement. Each
momentary depression will cause the display to page through XYZ, Yxy, L*a*b*, L*u*v*, L*C*h°, Whiteness, & Yellowness.
Decreases numeric values when used with the
[DIF]
key during
entry of reference or calibration values.
Selects Calibration when pressed with
[ILLUM]
key.
Selects various steps when in system setup procedures.
[ILLUM]/[s]
When held depressed activates illuminant/observer selection,
thereafter, momentary depressions select the illuminant to be used. The available illuminants are:
C A
D65D F7F2 F12F11
50
Note: the [DIF] key selects the Observer (CIE 1931 or CIE 1964 10°) to be used.
Increases numeric values when used with [DIF] key during entry
of reference or calibration values.
Displays the reference number when pressed momentarily.
Selects Calibrate when pressed with the
[FUNCTION]
key.
See Section
3.2
4.0
4.0
5.0
3.3
4.0
3.5
4.0
[DIF]/[ts]
When at function level, toggles Delta () On and Off with each
depression. Delta being the difference mode.
Decreases reference or calibration values when used with the
[FUNCTION]
Increases numeric values when used with the
key.
[ILLUM]
7
key.
3.5
4.0
4.0
3.2 Function Selection
Each momentary depression of [FUNCTION] will sequentially page through the available functions: CIE XYZ (or RGB), Yxy, L*a*b* (or Lab Hunter), L*u*v*,
a
L*C*h° (or
b
u
v
L*C*h°), Whiteness, & Yellowness.
If is active the following is displayed
2
YEL
W H T
C
2
C
W 9 7 . 8 7
. .
u
v
L C h
OR
. .
a
L C h
b
.
2
C
C 2 0 . 7 8
.
. .
L u v
.
u + 2 2 . 4 6
(Hunter)
2
C
Lab
OR
.
..
L a b
.
a + 1 8 . 5 9
Y x y
2
C
2
C
x .2 6 7 8
RGB
OR
2
X Y Z
C
Y 2 . 8 7
FUNCTION ILLUM
CAL
Y + 3 3 . 5 6
e
+ 8 7 . 9 7
Y d
W - 1 . 3 7
e
T - 1 1 5 . 8
.
L 2 3 . 4 5
h 2 5 . 7 9
.
L 1 2 . 3 3
.
v + 6 . 8 9
.
L 2 3 . 4 8
.
b + 8 . 7 5
Y 3 . 2 4
y .3 5 9 8
X 4 . 1 2
Z 2 . 4 1
DIF
YEL
C
W H T
+
W 9 7 . 8 7
L C H
E
OR
.
C 2 0 . 7 8
.
u + 2 2 . 4 6
(Hunter)
. . .
L C H
.
.
L u v
CCC
.
Lab
OR
a + 1 8 . 5 9
L a b
.
.
..
Y x y
x .2 6 7 8
+
RGB
OR
X Y Z
+
Y 2 . 8 7
2
Y + 3 3 . 5 6
e d
+ 8 7 . 9 7
Y
2
W - 1 . 3 7
C
T - 1 1 5 . 8
.
2
L 2 3 . 4 5
C
C
+
H 2 5 . 7 9
+
.
2
+
L 1 2 . 3 3
v + 6 . 8 9
.
2
L 2 3 . 4 8
C
b + 8 . 7 5
2
C
Y 3 . 2 4
y .3 5 9 8
2
C
X 4 . 1 2
e
+
+
+
+
+
Z 2 . 4 1
PRESS TO SELECT
THE NEXT FUNCTION
8
> The illuminant/observer will not be displayed in the difference mode during operation of
that function. It is only displayed during function selection.
> E will be displayed in place of L*a*b*, L*u*v*, and L*C*H* during difference
operation.
ILLUMINANT NOT DISPLAYED
DELTA E
Y E L
W H T
+
W 9 7 . 8 7
E
C
OR
E 41.89
.
C 2 0 . 7 8
E 23.67
.
u + 2 2 . 4 6
E 15.98
.
a + 1 8 . 5 9
Y + 3 3 .5 6e
Yd + 8 7 .9 7
W - 1 . 3 7
e
T - 1 1 5 . 8
.
L 2 3 . 4 5
+
+
H 2 5 . 7 9
.
+
L 1 2 . 3 3
v + 6 . 8 9
.
L 2 3 . 4 8
+
b + 8 . 7 5
Y 3 .2 4
Y x y
+
x . 2 8 9 0
X Y Z Y 2 . 8 7 Z 2 .4 1
9
+
y .3 6 4 5
+
X 4 .1 2
+ +
3.3 Illuminant/Observer Selection
There are eight different illuminants and two observers to choose from.
1) To enable illuminant
selection: Hold down
[ILLUM] until ILLUMINANT"
"SELECT
is displayed.
2) The active illuminant &
observer are displayed.
Press [ILLUM] to select a different illuminant.
Note: Illuminant F2 represents a cool white fluorescent lamp.
Illuminant F7 represents a broad-band daylight fluorescent lamp.
Illuminant F11 represents a narrow-band white fluorescent lamp (or illuminant TL84).
Illuminant F12 represents Ultralume 3000.
SELECT
ILLUMINANT
X Y Z
C
X 4 . 1 2
2
Y 2 . 8 7 Z 2 . 4 1
FUNCTION ILLUM
CAL
2
F12
Ultralume
2
F11
TL84
2
broad bandF7
2
F2
cool white
2
A
2
D
50
2
D
65
2
C
exit
FUNCTION ILLUM
CAL
HOLD
illm
PRESS
DIF
obs
DIF
Press [DIF] to select a different observer.
10
C
2
C
exit
FUNCTION ILLUM
CAL
10
illm
obs
DIF
PRESS
Illuminant/Observer Selection...continued
3) To exit press [FUNCTION].
When "Function Key" is pressed, the instrument will automatically convert the stored spectral references (if any) to tristimulus for the different illum/obs selected. The unit will also ask if you would like the references to be cleared. If you do not want the references cleared (or you do not have any entered), disregard the procedure below.
Clearing References
The "Clear Reference" function is useful if the Illum/Obs is being changed often, and many references exist in the instrument that are not being used. This will enable the operator to speed up the time it takes the instrument to recalculate to a different illum/obs.
X Y Z
C
X 4 . 0 8
10
Y 2 . 8 1 Z 2 . 3 7
UPDATING REFS # 1-24
CLEAR REFS
10
C
obsillmexit
FUNCTION
CAL
PRESS
ILLUM
DIF
Press [DIF] to enter clear reference function. After the "Dif Key" is pressed "CLEAR REFERENCES
NO YES" is displayed. Press
[DIF] to clear or [FUNCTION] not
to clear.
> You must press the "Dif" key
while the unit is recalculating the references to activate this function.
> If the clearing procedure is
carried out, all references (measured & numerically entered) will be removed.
> Refer to Section 3.5 for a more
detailed explanation on measured and numerically entered references.
11
3.4 Absolute Measurements
The 948 and 968 can perform absolute measurement in XYZ (RGB), Yxy, L*a*b*, L*u*v*, L*C*h°, Whiteness, & Yellowness.
> Helpful Hint: If you are going to measure something like a textile, you will receive
better results if you use the averaging function (see Section 3.6). Also, while measuring a textile, take several measurements in a circular format.
> If the is displayed, press [DIF] to activate absolute mode. > If the desired illuminant/observer is not displayed, refer to Section 3.3.
1) Press
[FUNCTION]
to select
desired measurement space.
2) Center the target window
over the area to be measured.
3) Lower unit to target window
and hold compressed.
"READING XYZ C2" (i.e., the
selected function & the illum/obs) will momentarily be displayed, and then the data. Release unit after data is displayed.
PRESS
TARGET WINDOW
X Y Z
Y 3 . 3 6
READING XYZ
2
C
2
X Y Z
Y 2 . 8 7 Z 2 . 4 1
FUNCTION ILLUM
X 1 . 2 4
Z 4 . 1 8
C
CAL
2
C
X 4 . 1 2
DIF
4) The display will show the
absolute values for XYZ.
FUNCTION
X Y Z X 1 . 2 4
Y 3 . 3 6
ILLUMINANT X VALUE
2
C
Z 4 . 1 8
Y VALUE Z VALUE
12
3.5 Difference Measurement
The 948 and 968 can measure the difference between a sample color and a reference. In order to measure these differences the reference must first be entered into memory. There are 24 different locations to store references.
The reference can be entered numerically using the key pad, or by measuring the reference color.
> The Whiteness and Yellowness reference color can only be entered by measurement.
Reference (Measurement)
Measured references are stored spectrally and the tristimulus values are recalculated each time a different illum/obs is selected. As long as references are measured, the instrument can automatically convert the spectral data to allow for viewing of difference measurements under various illum/obs sets.
Reference (Numeric)
Numerically entered references are saved with the selected illum/obs. A total of 7 sets of numerically entered tristimulus values (with different illum/obs) can be stored in each reference location.
If a different illum/obs is selected and no numerically entered reference supports that illum/obs, the instrument will turn "OFF" that reference location.
The instrument can not recalculate to a different illum/obs if it was not previously entered during numeric reference entry.
Operation Mode
The instrument can be set to operate in an automatic mode, where it will automatically select the closest reference; or in manual mode, where you have to manually select the reference locations you want to use. Refer to Section 5.2, Setting Operation Parameters for information on selecting manual or automatic reference operation.
In automatic mode, the instrument will select the reference that is closest to the measured color. If by some chance it is not the correct reference, simply press the
[ILLUM] key twice and the unit will select the next closest reference.
In manual mode, you must manually select the reference. Once you have selected a reference, that reference will be used until you select a different reference.
To activate the Difference mode, press
[DIF]
and the sign
will appear in the display. The Functions will have a "+" sign
in front of them for positive values instead of being implied as in the absolute function.
> L*a*b*, L*u*v*, & L*C*H* showE in function location.
DELTA SIGN PLUS SIGN
X Y Z
+
Y 2 . 8 7
X Y Z
Y 2 . 8 7 Z 2 . 4 1
FUNCTION ILLUM DIF
CAL
PRESS TO TOGGLE DIFFERENCE (DELTA) ON & OFF
2
C
X 4 . 1 2
+
Z + 2 . 4 1
X 4 . 1 2
DELTA ON
DELTA OFF
13
Difference Measurement...continued
> When entering references with any illuminant other than C2, Yellowness and Whiteness
(per ASTM E313) are not calculable.
> Once a reference has been set, changing between CIE Lab or Hunter Lab; Lch (ab) or
Lch (uv); and XYZ or X%Y%Z%, the references will not be automatically adjusted.
Note: You can set all reference locations to OFF by:
Activating difference mode.
Hold down [DIF] until reference entry is activated, then simultaneously press
[ILLUM] and [FUNCTION].
Ref 1 defaults to ON if all references are turned OFF.
Refer to Steps 2 and 3 below.
3.5.1 Entry of Reference Values
1) Select desired Illum/Obs (see
Sec. 3.3).
2) Press
[FUNCTION]
to select
desired measurement space.
> If is not displayed, press
[DIF] to activate Difference mode.
3) Hold down
REF VALUES - ENTER or MEASURE"
[DIF]
is displayed.
until
"SET
First displayed are the previous entered reference values for the last selected reference location.
The reference location is momentarily displayed.
Then "EXIT" is displayed.
> You can set all reference
locations to OFF by: holding down [DIF] then simultaneously pressing [ILLUM] and [FUNCTION]. Note, Ref 1 defaults to ON.
By Measurement
EXIT
Y 4 . 4 5
REF 03
Y 4 . 4 5
XYZ
Y 4 . 4 5
SET REF VALUES
ENTER or MEASURE
FUNCTION ILLUM
PRESS
C
CAL
2
2
X Y Z
Y 2 . 8 7
FUNCTION ILLUM
C
+
CAL
X 6 . 1 6
Z 2 . 6 9
X 6 . 1 6 Z 2 . 6 9
X 6 . 1 6 Z 2 . 6 9
DIF
HOLD DOWN
X + 4 . 1 2
Z + 2 . 4 1
DIF
14
Entry of Reference values - By measurement . . . continued
4) Select the reference location.
If the cursor is not blinking next to the reference location, press [DIF] until it does.
Hold down [DIF] and press [t] to decrement thru the reference locations or press [s] to increment.
5) Measure the reference color.
CURSOR+REFERENCE SPACE
REF 03
Y 4 . 4 5
FUNCTION ILLUM
DECREMENT
REF 03
Y 4 . 4 5
C
READING XYZ
CAL
2
X 6 . 1 6
Z 2 . 6 9
INCREMENT
X 6 . 1 6 Z 2 . 6 9
C
Reference Color
DIF
+
2
6) Press
[FUNCTION]
and
[ILLUM]
at the same time to exit to the main menu.
"REFERENCES UPDATED" is
momentarily displayed and the procedure is finalized.
> If no measurement or entry is
made the unit will display
"REFERENCES - NOT CHANGED."
EXIT
Y 2 . 8 7
FUNCTION ILLUM
CAL
PRESS AT SAME TIME
15
X 4 . 1 2
Z 2 . 4 1
DIF
Entry of Reference Values... By Numeric Entry
> Whiteness & Yellowness reference values can not be entered via the keyboard.
1) Select desired Illum/Obs (see
Sec. 3.3).
2) Press
[FUNCTION]
to select
desired measurement space.
> If is not displayed,
momentarily press activate Difference mode.
3) Hold down
REF VALUES - ENTER or MEASURE" is displayed.
[DIF] until
[DIF]
"SET
to
The unit will first display the previous entered reference values for the last selected reference location.
The reference location is momentarily displayed.
Then "EXIT" is displayed.
4) Select the reference location.
If the cursor is not blinking next to the references location, press [DIF] until it does.
X Y Z
Y 2 . 8 7
FUNCTION ILLUM
PRESS
EXIT
Y 4 . 4 5
REF 03
Y 4 . 4 5
XYZ
Y 4 . 4 5
SET REF VALUES
ENTER or MEASURE
FUNCTION ILLUM
CURSOR+REFERENCE SPACE
Y 4 . 4 5
FUNCTION ILLUM
CAL
REF 03
CAL
C
2
X 6 . 1 6
Z 2 . 6 9
X 6 . 1 6
Z 2 . 6 9
X 6 . 1 6 Z 2 . 6 9
X 6 . 1 6
Z 2 . 6 9
2
C
+
CAL
DIF
DIF
X + 4 . 1 2
Z + 2 . 4 1
DIF
HOLD DOWN
Hold down [DIF] and press [t] to decrement thru the reference locations or press [s] to increment.
DECREMENT
16
INCREMENT
+
Entry of Reference values - By Numeric Entry . . . continued
5) The cursor will blink over the
active value that can be edited. Enter the numbers for each value.
[DIF]
The
key advances the cursor to the next value to be edited.
[DIF]
The
conjunction with decrease and the value. Note: Hold
key is used in
[t]
to
[s]
to increase
[DIF]
depressed then press either
[t]
or [s] to change value.
6) Press
[FUNCTION]
and
[ILLUM]
at the same time to exit to the main menu.
"REFERENCES UPDATED" is
momentarily displayed and the procedure is finalized.
> If no measurement or entry
is made the unit will display "REFERENCES - NOT
CHANGED."
REF 03
Y 2 . 8 7
FUNCTION ILLUM
CAL
INCREASEDECREASE +
EXIT
Y 2 . 8 7
FUNCTION ILLUM
CAL
PRESS AT SAME TIME
CURSOR
X 4 . 1 2 Z 2 . 4 1
DIF
ADVANCE
+
X 4 . 1 2 Z 2 . 4 1
DIF
17
3.5.2 Taking A Difference Measurement
> If the is not displayed, press [DIF] to activate difference mode. > The reference color must first be entered into memory, refer to Section 3.5.1. > If reference selection is not set to auto, select the correct reference location before
following the procedure below (refer to Section 3.5.1).
> If the desired illuminant/observer is not displayed, refer to Sec. 3.3.
1) Press
[FUNCTION]
to select desired
measurement space.
2) Center target window over area to
be compared.
3) Lower unit to target window and
hold compressed. "READING XYZ
C2" (i.e., the selected function and
illum/obs) will momentarily be displayed, and then the reference location and data.
Release unit after data is displayed.
X Y Z
+
Y 2 . 8 7
FUNCTION ILLUM
PRESS
SAMPLE AREA TO BE COMPARED
REF 03
+
Y . 8 8 Z + . 4 8
READING XYZ
X + 2 . 8 2
2
C
X + 4 . 1 2
Z + 2 . 4 1
CAL
2
C
DIF
4) The display will show the
difference between the reference color and the area measured.
> If L*a*b*, L*u*v*, or L*C*H* are
selected, E value will be displayed in place of the Function & Illuminant.
> If you want to display the
selected reference location, press [ILLUM].
FUNCTION ILLUMINANT X VALUE
X Y Z
2
C
Y + . 8 8
Y VALUE Z VALUE
OR
DELTA E VALUE
E 2 . 9 9
a + . 8 8
18
+
X 2 . 8 2
+
Z . 4 8
+
L 2 . 8 2
+
a . 4 8
3.6 CMC Difference Operation
COMMERCIAL
CMC is an ellipsoidal tolerancing method which attempts to correlate small
measured color differences with visual assessment.
CMC provides a single numeric value (∆E
difference between a sample and a standard. This allows the use of a single tolerance value for comparing the acceptability of a color match.
E
is derived from the following formula:
CMC
2
C
CC =
LC =
Ec = [L
L
S
L
ι
+ C
C
cS
The CMC reference menu consists of the three factors that can be set: Lightness
factor (ι), Chromaticity factor (c), and Commercial factor.
The lightness factor is normally set to 2.00 (default) but other values may be
required when surface characteristics differ dramatically.
The chromaticity factor normally does not require any adjustment from the
default setting of 1.00.
The commercial factor is the tolerance limit that each sample is not to exceed. (e.g., if cf= 1.00 then any sample which has a ∆E
would be commercially unacceptable.)
) which describes the color
CMC
+ H
1/2
2
]
C
H
HC =
S
H
value greater than 1.00
CMC
2
C
L
To Setup CMC Difference:
> CMC Option is only available on L*C*H* difference function. > CMC option must first be selected in Operation Parameters (Sec. 5.2).
1) Follow Step 1 through 5 in Section 3.5.1 for entering a reference manually or by
measurement.
2) Press [DIF] until the CMC menu is
displayed. The cursor will blink over the active value that can be edited. Enter the number for each
CHROMATICITY FACTOR
CURSOR+REFERENCE SPACE
LIGHTNESS FACTOR
FACTOR
value.
The [DIF] key advances the
cursor to the next value to be edited.
REF 01
c 1 . 0 0
FUNCTION ILLUM
CAL
l 2 . 0 0
cf = 1 . 0 0
DIF
The [DIF] key is used in
conjunction with the [t] to decrease and [s] to increase the value. NOTE: Hold [DIF] depressed then press either [t] or [s] to change values.
DECREMENT
INCREMENT
+
19
CMC Difference Operation...continued
3) Press [FUNCTION] and [ILLUM]
at the same time to exit to the main menu. "REFERENCES UPDATED" is momentarily displayed and the procedure is finalized.
> If no measurement or entry is
made the unit will display
"REFERENCES NOT CHANGED."
EXIT
c 1 . 0 0
FUNCTION ILLUM
CAL
PRESS AT SAME TIME
cf = 1 . 0 0
Taking a CMC Difference Measurement
Before a CMC difference measurement can be made:
LCh CMC must be set to "ON" in modes (Sec. 5.2)
Reference must be entered (Sec. 3.5.1)
CMC difference operation must be setup (Sec. 3.6).
1) Select desired illuminant/observer (see Sec. 3.3).
l 2 . 0 0
DIF
2) Select L*C*h° difference mode function.
3) Center target window over area to be compared.
4) Lower unit to target window and hold compressed. "Reading LcCcHc C2" (i.e.,
the selected illum/obs) will momentarily be displayed, and then the reference location and data. Release the unit.
5) The display will show the difference
CMC Difference Values
between the reference color and sample, and the ∆E
CMC
value.
Pressing the [DIF] key will toggle
the display between the absolute
Absolute
measurement values and the CMC difference values.
> "CMC cf TOLERANCE
EXCEEDED" will display during a L*C*h° difference measurement when DECMC value
exceeds the cf (commercial factor) value.
> LcCcHc are values used for calculating ∆E
and are not L*, C*, and ∆H*.
cmc
20
3.7 Measurement Averaging Procedure
When averaging is activated in setup, averaging operation will occur on absolute
measurements, reference entry measurements, and difference measurements.
There are two averaging methods that are used: (refer Sec. 5.1 to activate
averaging):
SAMPLES AVERAGED - allows you to enter the number of measurements you will make (1-99) at various locations on a sample, to obtain an average value.
SUB AVERAGE - allows you to enter the number of measurements the unit will take (1-6) on a sample at a single location, to obtain an average value.
> When average and sub average are set to one, no averaging is performed and no
averaging messages will be displayed.
The following example was taken in the absolute measurement mode using: XYZ
function, two sub averages, and two measurement averages.
1) Press [FUNCTION] to select XYZ
measurement space.
2) Center the target window over first
area on the sample to be measured.
TARGET WINDOW
3) Lower unit to target window and
hold compressed. "READING XYZ
C2 1 OF 2", "2 OF 2", and the data
is display.
After releasing the read head,
"MEASURE SAMPLE 2 OF 2" is
displayed.
X Y Z
Y 2 . 8 7
FUNCTION ILLUM
PRESS
MEASURE SAMPLE
XYZ C
Y 65.03 Z 23.02
READING XYZ
READING XYZ
C
CAL
2
2
2 OF 2
2 OF 2
1 OF 2
X 4 . 1 2 Z 2 . 4 1
DIF
X 63.90
C
C
2
2
21
Measurement Averaging Procedure . . . continued
C
2
4) Center the target window over
second area on the sample to measure.
TARGET WINDOW
5) Lower unit to target window and
hold compressed. "READING XYZ
C2 1 OF 2" and "2 OF 2" will
momentarily be displayed, and then the averaged data.
6) The display will show the average
absolute values for XYZ.
X Y Z
X 63.91
Y 65.04 Z 23.02
AVERAGE
CALCULATED
READING XYZ
2 OF 2
READING XYZ
1 OF 2
FUNCTION
X Y Z X 63.91 Y 65.04
ILLUMINANT X VALUE
2
C
Z 23.02
2
C
2
C
> Do not jar or lift reading head
during sub averaging measurements.
> Refer to Section 5.1 averaging
setup procedure.
> When averaging is activated, this
basic procedure must also be followed during reference measurements and difference measurements.
Y VALUE Z VALUE
22
3.8 Store Data Operation
The data storage feature allows the unit to store measurement data for up to 500
readings. (Refer to Sec. 5.4 to activate.)
There are four basic operations of the store data function that can be utilized in
the store data function.
Group Selection - group tagging allows you to assign group numbers to the stored values. Group tagging is only accessable through RCI with a software package such as SpectroStart.
Send data - allows all of the stored data to be transferred to a computer or serial printer. Data is sent as determined by the function displayed and modes enabled.
Delete last - allows only the last measurement data to be deleted from stored memory.
Clear all - causes all of the measured data in stored memory to be deleted.
The following sample (made in XYZ absolute) illustrates the different functions
of the data storage feature.
MEASUREMENT
1) Press [FUNCTION] to select XYZ
measurement space.
2) Measure first sample.
"READING XYZ C2" is momentarily
displayed, then the measurement. Upon release of the read head,
"DATA STORED 1 OF 500" and then
the measurement is displayed.
2
C
X Y Z
Y 2 . 8 7 Z 2 . 4 1
FUNCTION ILLUM
CAL
PRESS
2
XYZ C
Y 3.47 Z 4.25
DATA STORED
1 OF 500
2
XYZ C
Y 3.47 Z 4.25
READING XYZ
X 3.41
X 3.41
X 4 . 1 2
DIF
2
C
23
Store Data Operation . . . continued
3) Measure second sample.
"READING XYZ C2" is momentarily
displayed, then the measurement. Upon release of the read head,
"DATA STORED 2 OF 500" and then
the measurement is displayed.
SEND DATA
4) Press both [ILLUM] and [DIF] at the
same time.
XYZ
X 3.45
2
C
Y 3.52 Z 4.29
DATA STORED
2 OF 500
XYZ C
2
X 3.45
Y 3.52 Z 4.29
READING XYZ
XYZ
X 3.45
2
C
Y 3.52 Z 4.29
FUNCTION ILLUM
CAL
2
C
DIF
5) Press [DIF] to transfer data (go to
Step 7). If "NO" is selected, press
[FUNCTION] and advance to Step 6.
Sample Printout
24
NEXT STEP
PRESS AT SAME TIME
DATA TRANSFER
COMPLETE
TRANSMITTING
2 OF 2
TRANSMITTING
1 OF 2
SEND DATA ?
NO
2 YES
PRESS
Store Data Operation . . . continued
DELETE LAST MEASUREMENT
6) Press [DIF] to delete the last
measurement made (this step can be repeated to delete several of the last measurements). If "NO" is selected, press
[FUNCTION] and advance to Step 7.
CLEAR ALL DATA
7) Press [DIF] to clear all stored
measurement data taken. If "NO" is selected, press
[FUNCTION] to return to normal
operation.
> The Send/Clear data function may
be exited at anytime by pressing [FUNCTION] and [ILLUM] at the same time.
> Refer to Section 5.4 to activate
the store data function.
> "STORAGE FULL" will display on
all measurements after 500 stored measurements has been reached
NEXT STEP
NORMAL OPERATION
DELETE LAST ?
NO
DELETE LAST ?
NO
Number of Measurements Stored
XYZ C
Y 3.52
NO
1 YES
LAST DELETED
2 YES
2
DATA CLEARED
CLEAR ALL ?
X 3.45
Z 4.29
2 YES
PRESS
PRESS
25
3.9 Display Messages
The most likely reasons for error messages to display are described below. If an error message is consistently displayed, contact X-Rite or an Authorized Service Center.
Power-Up Messages
"MEMORY LOST" is displayed when the instrument determines that the data in the
(battery backed up) RAM has been corrupted, if the internal lithium battery on the P.C.B. is bad, or if a new EPROM has been installed.
Operational Messages
"CMC cf TOLERANCE EXCEEDED" is displayed during a L*C*h° difference
measurement when ∆E
"C2 ONLY FOR - YELLOWNESS" is displayed when trying to select an
illuminant other than C2 for Yellowness.
"INVALID READING" is displayed when the unit is not held down long enough
during a measurement.
"NEED CALIBRATION" is displayed if the calibration procedure is not performed
for 24 hours, or if there is a 10°C change in temperature since the last calibration.
"READING COMPLETE" measurement has been taken and the instrument can be released. "READING XYZ C2" is displayed during a measurement. XYZ can be any of the
functions and C2 can be any of the illum/obs combinations.
"REFERENCES - NOT CHANGED" is displayed after leaving the Reference Entry
function, and no change was made.
"REFERENCES UPDATED" is displayed when
simultaneously pressed, when exiting reference mode.
"SELECT ILLUMINANT" is displayed when the
seconds, activating illuminant selection.
"SET REF VALUES - ENTER or MEASURE" is displayed if the difference mode
is on, and the
[DIF]
key has been held down for three seconds, activating
reference entry mode.
"SET REF VALUES - MEASURE ONLY" is displayed in WHT and YEL when
reference entry mode is enabled. You can only enter these reference values by measurement.
value exceeds the cf (commercial factor) value.
CMC
[FUNCTION]
[ILLUM]
key is held down for three
and
[ILLUM]
are
"UPDATING REFS ## - CLEAR REFS" when the illuminant/observer is
changed the stored references are automatically updated. This message also allows you to clear out all stored references (see Section 3.3 for more details).
"USE ANOTHER COLOR SPACE" is displayed when you try to enter the
reference value numerically in WHT and YEL. You can only enter the reference values by measurement.
Calibration Messages
"CALIBRATION - FAILED" is displayed when making a calibration measurement
and something is wrong (invalid reading or data is out of range).
26
Display Messages . . . continued
"CALIBRATION - NOT CHANGED" is displayed when the user decides to terminate
the calibration procedure by simultaneously pressing
[FUNCTION]
and
[ILLUM]
.
"CALIBRATION - UPDATED" is displayed after the calibration procedure has been
successfully completed.
"CAL ERROR ### - CELL CHANGE" make sure that calibration plaque is clean and
instrument is positioned exactly on the white spot. If the error persists, the number that is displayed should be reported to X-Rite or an authorized service center.
"CAL ERROR ### - MOTION DETECTED" calibration requires that the
instrument remain motionless during the five measurements of the white spot. If the error persists and is not due to movement, the number that is displayed should be reported to X-Rite or an authorized service center.
"INVALID ENTRY" is displayed when the numbers entered for the white values are
inappropriate.
"INSUFFICIENT READINGS" indicates that all sub-average readings were not made.
Miscellaneous Messages
"BATTERIES LOW" indicates that the batteries are getting low and will soon need
to be charged. It will only be displayed while the measurement is in progress. Once displayed you will have approximately 100-200 measurements remaining before charging is mandatory.
"BATTERIES MUST - BE CHARGED" indicates that the batteries are too low to
operate the unit. It will be displayed until you begin the recharge cycle, thereafter, the unit will be functional and all previous data will be accessible.
"I/O NOT READY - check I/O PIN 5" is displayed when I/O pin 5 is set to "CTS"
in modes and the unit is not connected to a printer or computer. Set handshake to "OFF" if unit is not connected to a printer or computer.
"LAMP FAILURE" measurement lamp is bad. The lamp should be replaced by
X-Rite or an authorized X-Rite service center. When this message occurs, you can get out of this condition by pressing
[FUNCTION]
then
; or waiting until the unit powers down.
"READING ERROR ##" reading error due to hardware problem. If this message
persists the number that is displayed should be reported to X-Rite or an authorized service center.
"REFL. EXCEEDED" calculation of reflectance for at least one of the 20nm
segments was more than 200%. This is usually caused by a bad calibration procedure. Recalibrate unit.
"THANKS! - I NEEDED THAT!" indicates that the charger has been plugged in and
the batteries are being charged. This is in response to the message "BATTERIES MUST BE CHARGED".
"WEAK LAMP - REPLACE SOON" indicates that the lamp is getting weak and
should be replaced in the near future. When this message occurs, you can get out of this condition by pressing
[FUNCTION]
waiting until unit powers down.
"X-Rite VER#### - COPYRIGHT 1990" is displayed when first activating
calibration. #### represents the datecode of the software.
[FUNCTION]
[ILLUM]
then
then
[ILLUM]
then
[FUNCTION]
; or
27
4. Calibration
The 948 & 968 should be calibrated to the X-Rite standard the first thing each day and every four hours of operation thereafter. However, a "NEED CALIBRATION" message will appear in the display:
If the calibration procedure has not been performed for 24 hours.
If there is a 10°C change in temperature since the last calibration.
If zero reflectance is measured improperly.
If the lamp output changes.
Whenever this message appears the calibration procedure should be performed before another measurement is taken to ensure accuracy.
Important Calibration Notes
If you change apertures you must change the aperture setting, read zero
reflectance, & calibrate the unit (see Section 4.2, steps 1, 2, & 5a thru 5d).
Dirt or dust in the optics area will cause an inaccurate calibration reading.
Refer to Section 8.2 for the optics cleaning procedure.
The ceramic Reflection Standard is dramatically affected by smudge
marks, dust, and finger prints. The standard should be cleaned using a mild
soap and warm water solution, thoroughly rinsed with warm water, and wiped dry with a lint free cloth. You must let the standard dry completely before taking a calibration reading.
If you are having linearity problems it’s possible that there is dust in the
optics or Zero Reflectance has drifted. If you improperly measure Zero Reflectance the unit can not automatically detect the drift. If you suspect this is the case, you should manually activate Read Zero Reflectance by performing Steps 1, 2, & 5 of Section 4.2.
Do not move the 948/968 while taking a calibration measurement. If motion
is detected an error message will be displayed and calibration aborted.
28
4.1 Positioning the 968 Onto the X-Rite Standard
You must set the 948/968 on the white standard so that the maximum amount of the bottom rubber pad of the shoe resides on the standard, and the target is centered on the circle. If you do not, the unit may rock slightly and cause an erroneous reading of the standard.
1) Center the target window on the White circle, making sure that the rubber pad is
completely on the standard and is down flat.
2) Take the measurement.
BOTTOM VIEW
29
Shown below is an IMPROPER METHOD of measuring the standard.
BOTTOM VIEW
INCORRECT!
30
X-Rite Reflection Standard
X-Rite
Reflection Standard P/N 968-62 Serial No. CXXXXXXXP-XXXXX Date XX/XX/XX
WAVE
[nm]
390 400 410 420 430 440 450 460 480 500 520 540 560 580 600 620 640 660 680 700 710
Refer to section in operation manual appendix for procedure to obtain reference values.
®
REFL
[%]
62.10
76.70
81.88
85.02
85.99
86.12
86.48
86.74
87.05
87.42
87.36
87.08
86.72
86.25
85.71
85.18
84.52
84.02
83.35
82.69
81.99
CALIBRATION
CHECK
C/2 (WHITE)
L* a*
b*
Instrument Model No.
Instrument Serial No.
COLOR CHECK
DATE: TEMP: ILLUM/OBS:
REFERENCES VALUES
STEP 1 (White) 2 (Blue) 3 (Brown)
APERT:
L* a* b*
Notes:
IMPORTANT!
The standard is dramatically affected by smudge marks and dust; and must be kept clean.
The Calibration Values for the white spot are affected by the environment
and cleaning method of the standard.
If the white spot does not measure correctly, it could be that the unit needs to be calibrated; there is dust in the optics; or the standard has smudge marks or is dirty.
The ceramic standard should be cleaned using a mild soap and warm water solution, thoroughly rinsed with warm water, and wiped dry with a clean, lint free cloth.
If you lose your envelope, you can obtain the calibration values from the back page of this manual, or, by looking at them in the calibration procedure. Otherwise, you can contact X-Rite with the serial number of your standard and get the values.
31
4.2 Calibrating to the White Standard
> If you are changing aperture size: Do steps 1, 2, & 5a thru 5d. Skip steps 3 & 4.
1) Press both
[ILLUM]
2) Select
[FUNCTION]
at the same time.
"YES"
, press
[DIF].
and
The copyright and software version are momentarily displayed.
3) "READ CAL WHITE" is displayed.
At this point you have two options:
You can measure the white
spot on the standard (go to step 4). This is standard operating procedure.
2
X Y Z
C
Y 2 . 8 7
FUNCTION ILLUM
CAL
PRESS AT SAME TIME
X-Rite
X 4 . 1 2
Z 2 . 4 1
Ver XXXX
Copyright
CALIBRATE ?
NO
FUNCTION ILLUM
CAL
READ CAL WHITE
5
SEC
READ CAL WHITE
EXIT CHANGE
FUNCTION ILLUM
CAL
DIF
1990
YES
DIF
PRESS
DIF
Or, you can press [CHANGE]
to change the calibration values for the 21 different spectral values or to change the aperture setting. This
procedure is only used when you want to calibrate to a different standard or you change apertures. Go to step 6.
32
Calibrating to a White Standard . . . continued
4) Measure the WHITE patch on the
standard. You must hold the unit depressed until all five readings have been completed.
> You must set the unit on the
standard so that the maximum amount of the bottom rubber pad of the shoe resides on the standard (refer to Section 4.1).
"CAL READING 1 OF 5" thru "CAL
READING 5 OF 5" is displayed; then "CALIBRATION UPDATED." The
procedure is finalized and the unit returns back to main menu.
> During the Cal Reading, the read
head must remain down and stable or an error message may occur.
> If "PLEASE WAIT XX (1-30)
SECONDS" is displayed during calibration, continue to hold read head down until calibration readings are over. This will only occur if the calibration procedure is performed within 30 seconds of a previous measurement.
> If "READ ZERO
REFLECTANCE" is displayed after reading the white spot, go
to step 5. This will only occur if the calibration values for zero reflectance have drifted.
CALIBRATION
UPDATED
CAL READING
5 OF 5
CAL READING
4 OF 5
CAL READING
3 OF 5
CAL READING
2 OF 5
CAL READING
1 OF 5
WHITE
THE FOLLOWING STEPS (5 & 6) ARE ONLY NECESSARY IF YOU WANT TO MEASURE ZERO REFLECTANCE, CHANGE SPECTRAL VALUES, or CHANGE APERTURE SIZE.
5) Steps 5a - 5d are only necessary if "READ ZERO REFLECTANCE" is displayed,
you want to manually activate Zero Reflectance Measurement, or YOU WANT TO CHANGE APERTURE SIZE.
Zero Reflectance is defined as, "measuring air with no ambient light". This can be accomplished by holding the unit depressed (take a reading) in a
dark room. In some cases, it is possible to just take a reading under a desk with no direct light.
To manually activate zero reflectance measurement or you want to change aperture size, see next page.
33
Zero Reflectance/Aperture size Setting . . . continued
5a) Press [CHANGE].
READ CAL WHITE
EXIT CHANGE
FUNCTION ILLUM
CAL
DIF
PRESS
5b) Press [NO].
5c) Press [NEXT] if the aperture size is
correct (go to step 5d).
If the aperture size is wrong press
[DIF] to change size.
> If you change aperture size you will be
sent back to step 5a after pressing
[NEXT]. At that point you must read Cal White, then Zero Reflectance.
5d) Measure Zero Reflectance.
> You must measure Cal White after
Zero Reflectance, if you manually selected Zero reflectance
6a) Steps 6a - 6c are only necessary if
you want to change the 21 spectral calibration values.
Press [YES]
NEW CAL VALUES?
NO YES
FUNCTION ILLUM
CAL
PRESS
APERTURE = 8.0mm
NEXT
FUNCTION ILLUM
CAL
READ ZERO REFL.
EXIT
NEW CAL VALUES?
NO YES
FUNCTION ILLUM
CAL
TOGGLE
DIF
DIF
4.0mm
8.0mm 20 mm
DIF
6b) Enter the value for 390nm.
[DIF]
> The
key advances the cursor to
the next value to be edited.
[DIF]
> Hold
to decrease value or
depressed then press
[
]
s
to increase
value.
6c) After entering value press
[FUNCTION] to advance.
Enter the value for 400nm. Repeat procedure for all values
After entering all 21 spectral values go back to step 4.
[t]
SET 390nm VALUE
FUNCTION ILLUM
DECREASE +
SET 400nm VALUE
NEXT
FUNCTION ILLUM
DECREASE
34
CAL
INCREASE
CAL
INCREASE
53.22NEXT
DIF
ADVANCE
+
53.22
DIF
ADVANCE
+ +
5. Setting System Parameters
5.1 Averaging
The averaging feature allows the unit to make several measurements from the same
and/or different locations on a sample. This will allow you to have a better overall average of a sample color. The averaging feature has two functions:
SAMPLES AVERAGED - allows you to enter the number of measurements
you will make (1-99) at various locations on the sample, to obtain an average value.
SUB AVERAGE - allows you to enter the number of measurements the unit
will take (1-6) on a sample at a single location, to obtain an average value.
> When average and sub average are set to one, no averaging is performed and no
averaging messages will be displayed.
1) Press both [FUNCTION] and [ILLUM]
at the same time.
> To exit Average Setup: Press
[FUNCTION] and [ILLUM] at the same time and the unit will return back to normal operation.
X Y Z
Y 2 . 8 7 Z 2 . 4 1
FUNCTION ILLUM
CAL
2
C
X 4 . 1 2
DIF
2) Select "NO", press [FUNCTION].
PRESS
35
PRESS AT SAME TIME
CALIBRATE ?
NO
FUNCTION ILLUM
CAL
YES
DIF
Averaging . . . continued
3) Select "YES", press [DIF].
4) Select the amount (1-99) of sample
measurements required.
Hold down [DIF] and press [t]
to decrease the sample measurement number or press [s] to increase the sample measure number.
Press [FUNCTION] to advance to
the Sub Average number entry.
5) Select the amount (1-6) of sub
average measurement required.
ADVANCE
SET AVERAGING ?
NO
FUNCTION ILLUM
CAL
SAMPLES AVERAGED
NEXT
FUNCTION ILLUM
CAL
INCREASEDECREASE +
SUB AVERAGE
NEXT TOGGLE
YES
DIF
PRESS
01
DIF
+
1
Press [DIF] to select sub
average number.
6) Press [NEXT] to return to "SET
AVERAGING?" Menu. Or, press
[FUNCTION] and [ILLUM] at the same time to exit to normal operation.
36
FUNCTION ILLUM
CAL
DIF
PRESS
1 - 6
5.2 Operation Parameters
Operation setup allows you to select various options, and individually turn on/off
certain functions. When a function is turned off it will not be displayed during operation.
CAL
2
C
X 4 . 1 2
Z 2 . 4 1
DIF
1) Press both
[FUNCTION]
at the same time.
> To exit Operation Setup:
Press [FUNCTION] and [ILLUM] at the same time and the unit will return back to normal operation.
and
[ILLUM]
X Y Z
Y 2 . 8 7
FUNCTION ILLUM
PRESS AT SAME TIME
2) Select "NO", press
[FUNCTION]
twice.
3) Select "YES", press
The procedure for the following steps is to press particular function on or off. Press
[DIF]
.
[NEXT]
to advance to the next step.
4) Select tone SOFT TONE, LOUD
TONE, or TONE OFF.
5) Select XYZ ON, X%Y%Z% (RBG) ON,
or OFF.
PRESS TWICE
[TOGGLE]
SOFT TONE
NEXT
XYZ ON
NEXT
CALIBRATE ?
NO
FUNCTION ILLUM
CAL
YES
DIF
OPERATION ?
NO
FUNCTION ILLUM
CAL
YES
DIF
PRESS
to change an option, or to turn a
TOGGLE
PRESS
TOGGLE
PRESS
(RGB) ON
XYZ OFF
SOFT LOUD
OFF
XYZ ON X%Y%Z%
37
Operation Parameters setup . . . continued
6) Select Yxy ON or OFF.
7) Select Lab (CIE), Lab (Hunter), or
OFF.
8) Select L*u*v* ON or OFF.
9) Select L*C*h° ab, CMC, uv, or OFF.
10) Select Whiteness ON or OFF.
11) Select Yellowness ON or OFF.
Yxy ON
NEXT
Lab (CIE) ON
NEXT
Luv ON
NEXT
LCh (ab) ON
NEXT
WHT ON
NEXT
YEL ON
NEXT
TOGGLE
TOGGLE
TOGGLE
TOGGLE
TOGGLE
TOGGLE
PRESS
PRESS
PRESS
PRESS
PRESS
PRESS
ON OFF
(CIE) ON (HUNTER) ON
OFF
ON OFF
(ab) ON (CMC) ON (uv) ON
OFF
ON OFF
ON OFF
12) Select Auto Reference ON or
OFF. When set to On, a
reference is automatically selected during a measurement. When set to Off, a reference must manually be selected before a measurement.
> Exit via [NEXT] OR press
[FUNCTION] & [ILLUM] at the
same time to return back to normal operation.
AUTO REF ON
NEXT
BACK TO STEP 3
38
TOGGLE
PRESS
ON OFF
5.3 RS232 I/O Parameters
Your instrument comes equipped with a serial port that allows data to be
transmitted/received to/from an external device. Listed below are the available I/O parameters.
"RCI ON/OFF" (Remote Control Interface) - enables or disables the ability
to externally control the 948/968 via the I/O port.
"BAUD" - determines the input/output rate (characters per second) of the
I/O port. Available outputs are: 300, 600, 1200, 2400, 4800, and 9600.
"CR WITH LF/CR WITH NO LF" (Carriage Return with Line
Feed/Carriage Return with no Line Feed) - varies the delimiter at the end of each line of data. When set to CR WITH NO LINE FEED, just a Carriage Return is sent at the end of a line of data. When set to CR WITH LF, a Carriage Return and then a Line Feed are sent at the end of a line of data.
"HANDSHAKE" is used for data transmission. Handshake may be set to
HANDSHAKE OFF (ASCII = 13H), PIN 5 CTS (Clear to Send), PIN 5 BUSY, or XON/XOFF ENABLED (ASCII = 11H). Handshake should normally be set to OFF when it is not being used.
1) Press both
[FUNCTION]
the same time.
> To exit I/O Setup: Press
[FUNCTION] and [ILLUM] at the same time and the unit will return back to normal operation.
2) Select "NO", press
times.
[ILLUM]
and
[FUNCTION]
three
at
39
PRESS THREE
TIMES
2
X Y Z
Y 2 . 8 7
FUNCTION ILLUM
PRESS AT SAME TIME
FUNCTION ILLUM
C
CAL
CALIBRATE ?
NO
CAL
X 4 . 1 2
Z 2 . 4 1
DIF
YES
DIF
RS232 I/O setup...continued
3) Select "YES", press
[DIF]
.
SETUP I/O ?
NO
FUNCTION ILLUM
CAL
> The basic procedure for the following steps is to press [TOGGLE] to set the I/O
parameter. Press [NEXT] to advance to the next step.
4) Select RCI ON or OFF.
5) Select the BAUD Rate.
6) Select Carriage Return with
or without Line Feed.
RCI ON
NEXT
BAUD 1200
NEXT
CR WITH LF
NEXT
TOGGLE
PRESS
TOGGLE
PRESS
TOGGLE
PRESS
YES
DIF
PRESS
ON OFF
1200 2400 4800 9600
300 600
CR WITH LF CR WITH NO LF
7) Select HANDSHAKE OFF, PIN
5 CTS, PIN 5 BUSY, or XON/OFF ENABLED.
Exit via [NEXT] OR press [FUNCTION] & [ILLUM] at
the same time to return back to normal operation.
HANDSHAKE OFF
NEXT
BACK TO STEP 3
40
TOGGLE
PRESS
HANDSHAKE OFF PIN 5 CTS PIN 5 BUSY XON/XOFF ENABLED
5.4 Format Output Parameters
This procedure allows you set the output format of the RS232 port. Refer to
Section 6 for sample print outs. Listed below are the available options.
AUTO PRINTOUT/STORE DATA/PRINTOUT OFF - determines if data
will be transmitted after each measurement, stored until transmitted, or not allow data transfer.
AUTO PRINTOUT - enables the unit to transmit data after each measurement. STORE DATA - allows the unit to store up to 500 measurements. PRINTOUT OFF - disables the ability to output data via the RS232 port from a
reading.
COMPUTER/PRINTER/SPECTRAL - selects the output format of the data
being transmitted.
COMPUTER FORMAT (used for wide printers) - transmits a group of data values per line of print (see Section 6, Print Example 2).
PRINTER FORMAT (used for narrow printers) - transmits a group of data values in a single column format (see Section 6, Print Example 1).
SPECTRAL FORMAT (968 ONLY) - transmits the measurement values for each wavelength (400nm - 700nm) in 10 nanometer increments (see Section 6, Print Examples 4 & 5).
ALL PRINT OFF/ON - determines which functions (XYZ, Yxy, etc.) are
transmitted. When set to OFF, only the active function will be transmitted. When set to ON, all the functions that are turned on in the "Operation Parameters" are transmitted.
PRINT HEADER OFF/ON - enables or disables the header (XYZ, Yxy,
etc.) during transmit.
PRINT REF OFF/ON (Reference) - disables or enables the Reference
values during transmit.
DEC. POINT ON/OFF (Decimal Point) - enables or disables the
decimal point during transmit.
CAL
2
C
X 4 . 1 2
Z 2 . 4 1
1) Press both
[ILLUM]
[FUNCTION]
at the same time.
> To exit Format Setup:
Press [FUNCTION] and [ILLUM] at the same time and the unit will return back to normal operation
and
X Y Z
Y 2 . 8 7
FUNCTION ILLUM
PRESS AT SAME TIME
41
DIF
Format Output setup . . . continued
SPECTRAL OUTPUT (968 only)
2) Select "NO", press
times.
3) Select "YES", press
[FUNCTION]
[DIF]
.
four
PRESS
FOUR
TIMES
CALIBRATE ?
NO
FUNCTION ILLUM
CAL
FORMAT OUTPUT ?
NO
FUNCTION ILLUM
CAL
> The basic procedure for the following steps is to press [TOGGLE] to set the format.
Press [NEXT] to advance to the next step.
4) Select Auto Printout, Store
Data or Printout Off.
5) Select Computer, Printer, or
Spectral.
AUTO PRINTOUT
NEXT TOGGLE
COMPUTER OUTPUT
NEXT
TOGGLE
PRESS
PRESS
AUTO PRINTOUT STORE DATA PRINTOUT OFF
COMPUTER OUTPUT PRINTER OUTPUT
YES
DIF
YES
DIF
PRESS
6) Select All Print OFF or ON.
7) Select Print Header OFF or
ON.
8) Select Print Reference OFF or
ON.
9) Select Decimal Point ON or
OFF.
> Exit via [NEXT] OR press
[FUNCTION] & [ILLUM] at
same time.
ALL PRINT OFF
NEXT
PRINT HEADER OFF
NEXT
PRINT REF OFF
NEXT
DEC. POINT ON
NEXT
BACK TO STEP 3
TOGGLE
TOGGLE
TOGGLE
TOGGLE
42
PRESS
PRESS
PRESS
PRESS
OFF ON
OFF ON
OFF ON
ON OFF
6. Printing Data
The RS232 parameters for the 948/968 serial interface are configured in Section
5.3. The variables are:
Remote Control Interface (On/Off)
Baud Rate (300, 600, 1200, 2400, 4800, or 9600)
Carriage Return (With or Without Line Feed)
Handshake (Off, Pin 5 CTS, Pin 5 Busy, or XON/XOFF Enabled)
The manner in which output data is arranged is determined by the Format Output
Parameter settings in Section 5.4. The variables are:
Printout (Auto, Store Data, Off)
Output Type (Computer, Printer, Spectral)
All Print (On/Off)
Print Header (On/Off)
Print References (On/Off)
Decimal Point (On/Off)
Data is transmitted from the serial port after each measurement if "Printout" is set
to Auto. The display will show "TRANSMITTING DATA" while transmission is taking place, and then return back to normal operation.
The following three pages contain some examples of the different types of outputs
that can be obtained according to the Format Output Parameter settings.
43
Print Example 1:
Printer ON
All Print ON
Header ON
References ON
Decimal Point ON
> If Header, References, or
Decimal Point are turned
OFF, they will not print.
> If All Print is OFF, only the
data for the active color space will print.
44
Print Example 2:
Computer ON
All Print ON
Header ON
References ON
Decimal Point ON
> If Header,
References, or Decimal Point are
turned OFF, they will not print.
> If All Print is OFF, only
the data for the active color space will print ­See Print Example 3, below.
Print Example 3:
Computer ON
All Print OFF
Header OFF
References OFF
Decimal Point ON
45
Spectral ON
All Print ON
Header ON
References ON
Decimal Point ON
> If Header,
References, or Decimal Point are
turned OFF, they will not print- See Print Example 5, below.
> All Print does not
affect the print out when Spectral in on.
Print Example 5:
Spectral ON
All Print ON
Header OFF
References OFF
Decimal Point ON
46
7. Changing Apertures
X-Rite has three different aperture kits available for the 948 & 968, and they are
listed below.
8mm/4mm Aperture Kit P/N 968-100-08
8mm UV Excluded Aperture Attachment P/N 968-61-08E
20mm Aperture Kit P/N 968-100-20
Each aperture kit includes the aperture, sensor nose, extra target windows, and a
specially designed aperture wrench.
APERTURE INSTALLATION
1) Unscrew the two thumb screws [1], then remove the nose piece [2].
2) To install the 4mm aperture attachment; screw the 4mm aperture attachment [3] onto the 8mm aperture [4]. The 4mm aperture attachment must be screwed on finger tight; then advance to Step 6 on following page.
To install the 8mm or 20mm aperture; advance to Step 3.
[4]
[3]
47
[2]
[1]
Changing Apertures . . . continued
3) Unscrew the existing aperture [5] with the aperture wrench [6].
Note: The aperture wrench tool has two sides, one side fits the 8mm aperture and
the other fits the 20mm aperture. The aperture wrench has two pins that fit into the holes in the aperture.
INSERT PINS INTO THE HOLES IN THE APERTURE, THEN TURN COUNTERCLOCKWISE TO REMOVE.
[6]
[5]
[7]
4) Screw in the new aperture using the aperture wrench.
IMPORTANT! The new aperture must be clean.
5) Attach the new nose piece to the housing with two thumb screws.
Note: When you tighten the thumb screws, be sure that the nose piece is flush
against the housing. Gently tighten one screw, then the other. Then finish tightening both screws.
IMPORTANT! The new nose piece must be clean.
6) Replace the existing target window [7] with the new target window.
Refer to Section 8.4, Target Window Replacement for this procedure.
Note: If you switch back and forth between the 4mm and 8mm aperture, it is possible to use the 8mm target window with the 4mm aperture.
If you’re very careful you can reuse the target window several times.
7) Change the aperture setting and recalibrate the unit.
Refer to Section 4.2, steps 1, 2, & 5a thru 5d for this procedure.
48
8. Maintenance
8.1. Troubleshooting
The X-Rite 948/968 is covered by a one year limited warranty (excluding ni-cad batteries) and should be referred to the factory or authorized service center for repair within the warranty period. Attempts to make repairs within this time frame may void the warranty.
X-Rite provides a factory repair service to their customers. Because of the complexity of the circuitry all circuit repairs should be referred to the factory or an authorized service center.
X-Rite will repair any 948/968 past warranty. Shipping costs to the factory or to an authorized service center shall be paid by the customer and the instrument shall be submitted in its original carton, as a complete unaltered unit.
CAUTION: DO NOT use any ketone solvents to remove ink from the unit. This will cause damage to the cover.
IMPORTANT! ALWAYS CHECK TO SEE IF READ LAMP IS WORKING BEFORE TROUBLESHOOTING.
-----TROUBLESHOOTING CHART------
A. If a wrong reading exists:
1. Recalibrate unit.
2. Clean Optics.
3 Contact authorized service center.
B. Unit does not turn on.
1. Check for low batteries.
2. Contact authorized service center.
C. Display not working.
1. Check for low batteries.
2. Contact authorized service center.
D. Reading Drifts.
1. Clean Optics.
2. Recalibrate unit.
3. Contact authorized service center.
E. Unit will not calibrate properly.
1. Dirty reference.
2. Optics dirty.
3. Contact authorized service center.
49
8.2. Optics Cleaning
The target window and optics should be cleaned once a week in normal environments; and more often in dirty or dusty environments.
Target Window
1. Remove dust and lint from target window by wiping it with a clean, lint free cloth,
slightly moistened with water.
Optics
1. Unscrew the two thumb screws [1] and remove the nose piece [2].
Remove dust from aperture with camel hair brush or a camera lens cleaner [3].
[2]
[1]
[3]
2. Blow short bursts of air into the optics opening using a camera lens cleaner (with
the brush removed) until all dust is removed. This can be done with the nose piece removed or attached.
WARNING! Do not use an air can that uses freon as a propellant.
Doing so could cause damage to the optics assembly.
50
8.3. Battery Replacement
1. Set the 948/968 on it’s side and lock shoe in place. The shoe must be locked.
2. Slide battery access door [1] toward bottom of unit and remove.
3. Disconnect plug [4] and pull battery pack [2] out of unit.
4. Remove old AA NI-CAD batteries [3], and install six fresh AA NI-CAD
(recognizing proper polarity).
5. Slide battery pack [2] into unit, and reconnect battery plug [4]. Reinstall battery access door [1].
6. Unlock shoe.
7. Unit should be charged for 24 hours after new battery installation is performed.
Batteries P/N SE15-19 (6 Required)
[3]
[1]
[4]
[2]
[3]
51
8.4. Target Window Replacement
1. Remove old target window [2] by pushing downward from top of shoe [1].
2. Place the instrument on end and align the target window so that the word "front"
runs parallel with the top edge of the shoe [1].
3. Insert one edge (top or bottom) of the new target window [2] in the opening of the shoe [1].
4. Place the other side of the target window [2] in the shoe [1] by snapping into position.
[1]
[2]
Target Window Part Numbers 968-121-20 (20mm) 968-121-08 (8mm) 968-121-04 (4mm)
Front Edge
52
8.5. Lamp Replacement
Due to the circuit complexity, alignment procedures, and test equipment required
- The read lamp should only be replaced by X-Rite or an authorized X-Rite
Service Center.
The lamp is monitored for intensity and failure warnings will be displayed if a
problem occurs.
The lamp should last approximately one million measurements and is covered by
a one year limited warranty. Refer to the Limited Warranty statement on page iii.
53
A1 Specifications
Measuring Functions: Absolute Difference Indices:
CIE XYZ ∆(XYZ) X%Y%Z% (X%Y%Z%) CIE Yxy ∆(Yxy) CIE LAB (L*a*b*) ∆E* Hunter LAB (Lab) ∆E CIE LUV (L*u*v*) ∆E* CIE LCH (L*C*H*) ∆E*(ab, CMC, or uv space)
Display: 2 row by 16 character Supertwist dot matrix LCD Measuring Geometry: 0°/45°, fiber optic pickup, multi-sensor array
Measuring Area: 8.0mm & 4.0mm (20mm optional) Light Source: Gas filled tungsten lamp, approx. 2856°Κ
(corrected for D65 illuminant)
Illuminant Types: C, D65, D
A, F2 (cool white fluorescent),
50,
F7 (broad-band white fluorescent), F11 (TL84), & F12 (Ultralume 3000)
Standard Observers: 2° & 10° Measurement Range: 0 to 200% reflectance Spectral Range: 400nm - 700nm Spectral Interval: 20nm (15nm bandwidth) Resolution: .01%
Inter-Instrument Agreement: 948: 0.40E*ab average (Based on average of 12 BCRA tiles)
968: 0.20E*ab average (Based on average of 12 BCRA tiles)
Short Term Repeatability: 948: 0.10 max E*ab on a white ceramic (20 measurements)
968: 0.05 max E*ab on a white ceramic (20 measurements)
Warm Up Time: None Measurements Per Charge: Approx. 1000 typical Measuring Time: Approx. 2 seconds Data Interface: Patented Bidirectional RS-232, 300 to 9600 baud
(user selectable), bipolar output
Power Supply: Six rechargeable AA NiCad batteries
7.2v total rated @ 600mAh (included)
Charge Time: Approx. 14 hours AC Adaptor Requirements: 948/968 90 - 130VAC, 50 - 60Hz, 18W Max.
948X/968X 180 - 260VAC, 50 - 60Hz, 20W Max. 12VDC @ 700ma: Positive Tip
Operating Temp. Range: 50° - 104°F (10° - 40°C)
ab
uv
Whiteness per ASTM E313 Whiteness & Tint per CIE Yellowness per ASTM E313 Yellowness per ASTM D1925
54
Specifications . . . continued
Storage Temp. Range: -4° - 122°F (-20° - 50°C) Weight: 2.3 lbs. (1050 grams) Dimensions: 3 3/16" H x 3" W x 7 3/4" L
(81mm H x 76mm W x 197mm L)
Accessories Provided: Calibration Standard
Operation Manual Reference Guide AC Adaptor Carrying Case
X-Rite standards are traceable to the National Institute of Standards and Technology.
This product covered by U.S. Patent 4,591,978 and other patents pending.
Specifications and design subject to change without notice.
A2 Optional Accessories
Part Number
QA-Master® (Quality Assurance software program) . . . . . 1255-00
ColorStart® (948 Quality Control software program) . . . . 1220-00
SpectroStart® (968 Quality Control software program) . . 1250-00
4/8mm Aperture Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 968-100-08
20mm Aperture Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 968-100-20
8mm Aperture Attachment (UV excluding). . . . . . . . . . . . 968-61-08E
948/968 Mounting Fixture . . . . . . . . . . . . . . . . . . . . . . . . . 968-80
Security Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 418-75
Portable Thermal Printer (115VAC). . . . . . . . . . . . . . . . . . 418-113
Portable Thermal Printer (230VAC). . . . . . . . . . . . . . . . . . 418X-113
Interconnect cable for Macintosh computers
with 8 pin mini-DIN connector . . . . . . . . . . . . . . . . . . . . . 418-79
Modular Interconnect Cable (requires adaptor below) . . . . . SE108-69
DB25P DCE (Null Modem) Interface Adaptor . . . . . . . . . 418-70
DB25S DCE (Null Modem) Interface Adaptor . . . . . . . . 418-71
DB25P DTE (Normal) Interface Adaptor . . . . . . . . . . . . . 418-80
DB25S DTE (Normal) Interface Adaptor . . . . . . . . . . . . . 418-81
DB9P Interface Adaptor . . . . . . . . . . . . . . . . . . . . . . . . . . 418-90
DB9S Interface Adaptor . . . . . . . . . . . . . . . . . . . . . . . . . . 418-91
4mm Target Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 968-121-04
8mm Target Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 968-121-08
20mm Target Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . 968-121-20
55
A3 Factory Presets
Shown below are the factory presets for the Averaging, Operation, I/O, and Format parameters.
AVERAGING PARAMETERS
Average - 01 Sub Average - 1
OPERATION PARAMETERS
Tone - SOFT XYZ - ON Yxy - ON L*a*b* - ON (CIE) L*u*v* - ON
L*C*h° - ON (ab space) Whiteness - ON Yellowness - ON Auto Reference - ON
RS232 I/O PARAMETERS
RCI - ON BAUD RATE - 1200 CR/LF - Carriage Return With Line Feed HANDSHAKE - OFF
FORMAT PARAMETERS
Printout - AUTO Comp/Print/Spectral - COMPUTER All Print - OFF Print Header - OFF Print Reference - OFF Decimal Point - ON
The unit is shipped from the factory with function "L*a*b*" and illuminant/observer
°
2
"C
." If the memory is reset, the function, illum/obs, and all parameters will return to
the settings described above.
56
A4 Spectrophotometer Stand
X-Rite has an optional mounting fixture available (P/N 968-80). The fixture can hold items that are a maximum of four inches wide, or two inches to the center of the object.
ITEM TO BE MEASURED
THE UNIT QUICKLY ATTACHES TO THE FIXTURE WITH A THUMB SCREW ADJUSTMENT
HOLDS UNIT CLOSED
SPRING LOADED HOLDING CLAMP
57
A5 Color Check
The color check procedure will help you track instrument color performance to assure measurement accuracy. Initially, the three patches on the standard should be measured once a day for a period of 10 days to determine the reference values. Thereafter, a periodic check (once a week depending on usage) is all that is required.
Located on the back of the reference standard envelope is a color check label. The reference data should be recorded there for quick reference.
Reference Data Procedure
To obtain reference values:
1) Clean instrument optics and reference standard if necessary.
2) Calibrate instrument according to procedure in operation manual. Note: The reference standard is dramatically affected by smudge marks and
dust; and must be kept clean.
3) Select L*a*b* absolute function and desired illuminant/observer. Note: The selected illum/obs must be the same whenever the color check is
performed.
4) Measure the White (1), Blue (2), and Brown (3) patches on the reference standard.
White Brown
5) Record the L*a*b* values for each color on a piece of paper.
6) Repeat steps 1-5 for 10 days.
Blue
58
Color Check . . . continued
7) After 10 days average the white, blue, and brown values and enter data in the designated place on the color check label.
8) Enter date, illuminant, observer, aperture size, temperature, instrument model, and serial number on color check label. Note: Temperature of the reference standard should be maintained (±2°C) throughout color check sequence.
Periodic Color Check Procedure
1) Copy reference data located on the color check label to the shaded areas on
the color check record located on the next page.
2) Perform steps 1-4 in reference data procedure.
3) Record the measured white, blue, and brown values and date on the color
check record on the next page. Note: You may want to make several copies of the color check record before entering any data. This will allow you to have additional pages available when they are needed.
4) Compare measured values to the reference values.
If values are off, clean optics and reference standard and remeasure.
Consistent L*, a*, or b* differences exceeding .30 (or obvious trends away
from the reference data) may require the unit to be serviced. Contact X-Rite or authorized service center if this is the case.
Always verify color performance with the same aperture size, illum/obs,
and temperature.
59
COLOR CHECK RECORD
Date: Illuminant: Observer: Temperature: Aperture Size: Model-S/N:
Reference White Blue Brown
Values L* a* b* L* a* b* L* a* b*
Date
60
X-Rite, Incorporated----World Headquarters
3100 44th Street, S.W. Grandville, Michigan 49418 U.S.A www.x-rite.com (616) 534-7663 Toll-Free U.S. Numbers Tel: 1-888-826-3042 Fax: 1-888-826-3043 Toll-Free International Numbers Tel: 1-888-826-3039 Fax: 1-888-826-3041
X-Rite GmbH
Stollwerckstraße 32 51149 Köln Germany Tel: (49) 2203-91450 Fax: (49) 2203-914519
X-Rite GmbH
Sochorova 705 CZ-682 11 Vyskov Czech Republic Tel: (42) 0507-328197 Fax: (42) 0507-328138
X-Rite Asia Pacific Ltd.
Room 808-10 Kornhill Metro Tower 1 Kornhill Road Hong Kong Tel: (852) 2-568-6283 Fax: (852) 2-885-8610
X-Rite Ltd.
The Acumen Centre First Avenue Poynton, Cheshire England Tel: 9-011-44-0-1625-871100 Fax: 9-011-44-0-1625-871444
X-Rite Méditerranée
Parc du moulin de Massy 35, rue du Saule Trapu 91300 Massy France Tel: 33-1-69.53.66.20 FAX 33-1-69.53.00.52
X-Rite Asia Pacific Limited-----Japan Office
1-7-11-506 Higashi-Gotanda, Shinagawa-ku, Tokyo 141 Japan Tel: 03-5447-1607 Fax: 03-5447-1608
P/N 968-500 Rev. V-5/11/99