X-Rite 890 Operation Manual

Page 1
890
AUTOMATIC STRIP READING DENSITOMETER
Operation Manual
Page 2
Page 3
Congratulations! We at X-Rite, Incorporated are proud to present you with X-Rite 890 Auto Strip Reading, Color Photographic Densitometer. This instrument represents the very latest in microcontrollers, integrated circuits, optics, and display technology. As a result, your X-Rite instrument is a rugged and reliable instrument whose performance and design exhibit the qualities of a finely engineered instrument, which is not surpassed.
To fully appreciate and protect your investment, we suggest that you take the necessary time to read and fully understand this manual. As always, X-Rite stands behind your instrument with a one year limited warranty, and a dedicated service organization. If the need arises, please don’t hesitate to call us.
Thank you for your trust and confidence.
X-Rite, Incorporated
i
Page 4
I N T R O D U C T I O N
FCC
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Canada
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations. Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
NOTE: Shielded interface cables must be used in order to maintain compliance with the desired FCC and European emission requirements.
CAUTION: Operational hazard exists if AC adapter other than X-Rite SE30-77 (100-240V) is used.
VORSICHT: Betriebsgefahr besteht bei Gebrauch von anderen Adaptern als X-Rite SE30-77 (100-240 V).
AVISO: No use otro adaptador C.A. que no sea la pieza X-Rite SE30-77 (100­240V), por el riesgo de mal funcionamiento del equipo.
ATTENTION: Ne pas utiliser d’adaptateur autre que SE30-77 (100-240V) de X­Rite au risque de mauvais fonctionnement de l’appareil.
AVVISO: Non usare un altro adattatore C.A. che non è del pezzo X-Rite SE30­77 (100-240V), per il rischio di malfunzionamento dell'apparecchio.
ii
The Manufacturer: X-Rite, Incorporated Der Hersteller: 4300 44th Street, S.E. El fabricante: Grand Rapids, Michigan 49512 Le fabricant: Il fabbricante:
Declares that: Densitometer gibt bekannt: 890 advierte que: avertit que: avverte che:
is not intended to be connected to a public telecommunications network. an ein öffentliches Telekommunikations-Netzwerk nicht angeschlossen werden soll. no debe ser conectado a redes de telecomunicaciones públicas. ne doit pas être relié à un réseau de télécommunication publique. non deve essere connettuto a reti di telecomunicazioni pubblici.
Page 5
U S E R I N F O R M A T I O N
CE DECLARATION
Manufacturer's Name: X-Rite, Incorporated Manufacturer's Address: 4300 44 Grand Rapids, Michigan 49512 U.S.A.
Model Name: Densitometer Model No.: 890
Directive(s) Conformance: EMC 89/336/EEC LVD 73/23/EEC
th
Street, S.E.
WEEE
As of August 13, 2005, X-Rite products meet the European Union – Waste Electrical and Electronic Equipment (WEEE) directive. Please refer to www.xrite.com X-Rite’s compliance with the WEEE directive.
Warning: This is a class A product. In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures.
for more information on
PROPRIETARY NOTICE
The information contained in this manual is derived from patent and proprietary data of X-Rite, Incorporated. This manual has been prepared solely for the purpose of assisting in the use and general maintenance of this instrument.
The contents of this manual are the property of X-Rite, Incorporated and are copyrighted. Any reproduction in whole or part is strictly prohibited. Publication of this information does not imply any rights to reproduce or use this manual for any purpose other than installing and operating the equipment described herein.
No part of this manual may be reproduced, transcribed, transmitted, stored in a retrieval system, or translated into any language or computer language, in any form or by any means, electronic, magnetic, mechanical, optical, manual, or otherwise, without the prior written permission of an officer of X-Rite, Incorporated.
This instrument may be covered by one or more patents. Refer to the instrument for actual patent numbers.
Copyright © 2007 by X-Rite Incorporated “ALL RIGHTS RESERVED”
X-Rite® is a registered trademark of X-Rite, Incorporated All other logos, product names, and trademarks mentioned are the property of their respective holders.
iii
Page 6
I N T R O D U C T I O N
LIMITED WARRANTY
X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to be free of defects in material and workmanship (excluding battery pack) for a period of 12 months. This warranty shall be fulfilled by the repair or replacement, at the option of X-Rite, of any part or parts, free of charge including labor, F.O.B. its factory or authorized service center.
This warranty shall be voided by any repair, alteration, or modification, by persons other than employees of X-Rite, or those expressly authorized by X­Rite to perform repairs, and by any abuse, misuse, or neglect of the product, or by use not in accordance with X-Rite’s published instructions.
X-Rite reserves the right to make changes in design and /or improvements to its products without any obligation to include these changes in any products previously manufactured. Correction of defects by repair or replacement shall constitute fulfillment of all warranty obligations on the part of X-Rite.
THIS WARRANTY IS EXPLICITLY IN LIEU OF ANY OTHER EXPRESSED OR IMPLIED WARRANTIES, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. THIS WARRANTY OBLIGATION IS LIMITED TO REPAIR OR REPLACEMENT OF THE UNIT RETURNED TO X-RITE OR AN AUTHORIZED SERVICE CENTER FOR THAT PURPOSE.
This agreement shall be interpreted in accordance with the laws of the State of Michigan and jurisdiction and venue shall lie with the courts of Michigan as selected by X-Rite, Incorporated..
iv
Page 7
TABLE OF CONTENTS
This manual is organized into eight sections and four appendices. In order to make the best use of your instrument, it is recommended that you read all sections and appendices.
SECTION ONE - Getting Started Unpacking and Inspection__________________________________1-1 Packing Drawing and Parts List_______________________1-1 Product Description_______________________________________1-2 Applying Power _________________________________________1-3 User Interface ___________________________________________1-4 Channel and Function Menus_________________________1-4 Display and Keyswitch Definition_____________________ 1-5 Menu Keys_______________________________________1-5 Help Messages ____________________________________1-5 Adjusting the Paper Guides_________________________________1-6 Strip Measurement Techniques______________________________1-7 Strip Information __________________________________1-7 Paper Strip _______________________________________1-7 Film Strip ________________________________________ 1-10 Printer Balance Strip _______________________________ 1-10
SECTION TWO - Instrument Calibration
Calibration Information____________________________________2-1 Frequency of Calibration ____________________________ 2-1 Auto Calibration Procedure_________________________________ 2-2 Manual Calibration Procedure_______________________________2-3
SECTION THREE - Setting System Configuration
Instrument Configuration __________________________________3-1 Page 1 Configuration Options ________________________3-1 Page 2 Configuration Options ________________________3-2 Page 2a Configuration Options _______________________3-2 Page 2b Configuration Options _______________________3-3 Page 2c Configuration Options _______________________3-3 Page 2d Configuration Options _______________________3-3 Configuration Setup Procedure _______________________ 3-3
v
Page 8
I N T R O D U C T I O N
SECTION FOUR - Strip Operation
Measuring Strips _________________________________________ 4-1 View Strip Data ___________________________________4-3 Manually Transmitting Strip Data____________________________4-5
SECTION FIVE - Service and General Maintenance
Repair Information _______________________________________5-1 Cleaning the Instrument ___________________________________5-2 General Cleaning __________________________________ 5-2 Cleaning the Optics ________________________________ 5-2 Cleaning the Calibration Strip ________________________ 5-3 Replacing the Read Lamp __________________________________5-4 Troubleshooting Tips _____________________________________ 5-5
APPENDIX A - Technical Specification APPENDIX B - Display Messages APPENDIX C - Term Abbreviations APPENDIX D - Parts List and Packaging Drawing APPENDIX E - 890 Instrument Firmware Update
vi
Page 9
S E C T I O N O N E
Getting Started
This section covers unpacking, inspection, and general set up of your instrument. Product description, paper guide adjustments, and measurement techniques are also include.
Section One Contents
Unpacking and Inspection
Product Description
Applying Power
User Interface
Adjusting the Paper Guides
Strip Measurement Techniques
UNPACKING AND INSPECTION
After removing the instrument from the shipping carton, inspect for possible damage. If any damage occurred during shipping, immediately contact the transportation company. Do not proceed with installation until the carrier’s agent has inspected the damage.
Your instrument was packaged in a specially designed carton to assure against damage. If reshipment is necessary, the instrument should be packaged in the original carton. If the original carton is not available, contact X-Rite to have a replacement shipped to you.
Packaging Drawing and Parts List
Check your packaging contents against your packing list and your original order. Detailed packaging drawing and parts list is included in this manual as Appendix D.
1-1
Page 10
SECTION ONE
PRODUCT DESCRIPTION
The X-Rite 890 automated instruments measures paper, film, and printer balance control strips. Simply insert the strip into the instrument for motorized, automatic measurements. Red, green, and blue density data is sorted for fields such as HD, LD, and Stain, then simultaneously displayed and transmitted to a minilab printer for analysis.
To accommodate the different size control strips, the instrument has adjustable paper guides on each side of the control strip entrance. Adjustment is made by simply sliding the paper guides to the settings displayed for the selected strip.
The instrument communicates through a standard RS-232 interface. If you wish to remotely control your instrument, you must use the remote control interface protocol discussed in the RS-232 Interface Manual—available from X-Rite, Incorporated.
Operating Key llll
Operating Key lll
Operating Key ll
2 x 16 Character Display
Operating Key l
Paper Guide
35mm Strip Entrance
Paper Guide
RS-232 Serial Interface
Power Input
1-2
Page 11
GETTING STARTED
APPLYING POWER
To apply power to the instrument:
1. Verify that the voltage indicated on the adapter complies with the AC
line voltage in your area. If not, contact X-Rite or your authorized representative.
2. Insert the small plug from the adapter into the power-input connector on
the instrument.
3. Plug the detachable line cord into the adapter.
4. Plug the line cord into an AC wall receptacle. IMPORTANT: To extend the life of the internal memory battery, it is
recommended that the instrument remain plugged into AC power. If you are using a modem, AC power must be applied in order to transmit or receive data.
Detachable Line Cord
At initial power-up, the instrument performs a “boot memory test.” The testing sequence is displayed on the instrument screen. If all internal tests are OK, Page 1 Main Menu displays—see next page.
Power Input
Adapter
Testing Program boot memory...3
2 1
XRite 890 vXXXX Copyrt 1989—1997
1-3
Page 12
SECTION ONE
USER INTERFACE
The 890 instrument incorporates a three page menu system. Menu pages are displayed by continually pressing keyswitch l (p#) on the instrument. The first page contains all the predefined control strips. The last two pages provide access to the instrument setup options and strip measurement data.
Main Menu
pap (paper) - accesses the paper measurement channel. The format
film - access the film measurement channel. The format selected remains
bal (printer balance) - accesses the printer balance measurement channel.
Function Menus
view - used to view data of the last strip measured.
xmit (transmit)- used to manually transmit data of the last strip
cal (calibration) - used for calibration.
cnfg (configuration) - used to access system configuration options.
load - for updating the instrument’s firmware (see Appendix E).
selected remains selected until changed.
selected until changed.
The format selected remains selected until changed.
measured—via RS-232 port.
MAIN MENU
p1
pap
film bal
FUNCTION MENU
p2
view
xmit
FUNCTION MENU
cnfg load
p3
cal
1-4
Page 13
GETTING STARTED
Display and Keyswitch Definition
The characters in the bottom row of the display dictate which function is selected or which action takes place when a corresponding keyswitch is pressed. Normally, upper-case lettering in the top row of the display is used for messages, and lower-case lettering in the bottom row is used for menu options that are selectable.
Each keyswitch is embossed with identification marks signifying its number. The left-most key is labeled “I”, center-left key is labeled “ll”, and so on. Keyswitches that are used to select a function or perform an action are tinted (grayed) throughout this manual.
MAIN MENU
p1
pap
film
bal
Messages (top row) Functions (bottom row)
Keyswitch identification marks
Menu Keys
Keyswitch lll and llll have the word “MENU” printed above them. Pressing these two keys simultaneously at any location in the menu structure causes the display to return to page 1 of the main menu.
MAIN MENU
p1
pap
film
bal
MENU
Press keys simultaneously
Help Messages
Built-in help messages are available for most functions. Help messages are activated by pressing and holding down a key until the message appears. To temporarily pause a message, press the key again.
MAIN MENU
pap
filmp1
MENU
Press and hold keyswitch to see help message
bal
PAPer
"pap" key select
1-5
Page 14
SECTION ONE
ADJUSTING THE PAPER GUIDES
To accommodate various size paper strips, the instrument has an adjustable paper guide on each side of the strip entrance. Each paper guide has a pointer (triangle) that is aligned with a number on the guide rail. The guide setting numbers are shown on the instrument display when a paper strip is selected. The paper guide rail has setting marks that range from “9-30” to the left and right of the center diamond. Each mark (stop) is 1/10 increment—10 stops equal one inch. The instrument automatically displays the proper guides settings for the selected strip (see below).
30
Paper Guide
The paper guides are moved by sliding them to the left or right.
Center Alignment Line
To accommodate 35mm wide strips, the instrument has a special stationary guide located above the diamond
Instrument Display
20
READ k:RA–4 ALL
AT<18
10
Center Diamond
18>other
10
Paper Strip
20 30
Pointer
Stationary Guide
35mm Slot
1-6
Film Strip
35mm Film
Page 15
GETTING STARTED
STRIP MEASUREMENT TECHNIQUES
All control strip types should have at least a 30.5mm (1.25) leader before the first measurable patch. However, strips with less than a 30.5mm leader can also be measured reliably if required. Refer to your Control Strip and Balance Print Format Guide for leaderless strip measurement procedure.
Supported strip definitions and strip insertion direction information is available in your Control Strip and Balance Print Format Guide shipped with your instrument.
Strip Information
If a strip becomes “jammed” in your instrument during a measurement,
simultaneously press key lll and key llll labeled “MENU.” This should cause the instrument to feed-out the strip. If this method does not work, slowly pull strip out from the front of the instrument.
A strips is inserted into the instrument until it rests against the drive
rollers.
After a strip is initially inserted into the instrument, a one second delay
occurs before the drive mechanism is activated to allow time for proper alignment.
Due to the amount of variation in printer balance strips, no guide setting
Due to the configuration of the instrument drive mechanism, 16mm wide
The instrument does measure strips that are creased.
Paper Strip
When a particular paper strip is selected, one or both of the displayed paper guide setting numbers flash. If only one guide setting number is flashing, this indicates the guide that the strip rests on first when measuring. In most cases this only occurs on a strip that required multiple passes such as, Fuji CP-21. Both guide setting numbers flash on single pass strips.
Indicates ALL patches are
Both guide setting number
flash, indicating a single
read on one pass
Indicates strip type
Single
Pass
READ k:RA–4 ALL
AT<18
pass strip
numbers appear in the display in the printer balance categories.
film control strips cannot be measured.
Indicates CYN patches are
read one the first pass
Indicates strip type
READ f:CP21 CYN
18>other
One guide setting number
flashes, indicating strip
placement on this guide first
AT<17
26>other
Multi Pass
1-7
Page 16
SECTION ONE
Single Pass
Single pass strips require the strip to be inserted into the instrument once. In this example, the strip—Kodak RA4—can be inserted in either direction. Refer to your Control Strip and Balance Print Format Guide for information on all strips supported.
Indicates ALL patches are measured
18>other
READ k:RA–4 ALL
AT<18
20 30
10
10
20
30
Multi Pass
Typically, a multi pass strip requires three pass through the instrument to obtain measurement data. The instrument displays what color patches are measured and what guide is used to insert the strip from. In this example, the cyan—Fuji CP21—row of patches is measured first, followed by the magenta and yellow. Refer to your Control Strip and Balance Print Format Guide for information on strip insertion direction.
Indicates cyan patches are measured
20 30
10
26>other
10
20
AT<17
READ f:CP21 CYN
30
Guide number flashes, indicting strip positioning
PASS ONE
1-8
Page 17
GETTING STARTED
Indicates magenta patches are
20 30
10
26>other
10
20
AT<17
READ f:CP21 MAG
Indicates yellow patches are measured
30
Guide number flashes, indicting strip positioning
20 30
10
26>other
PASS TWO
PASS THREE
AT<17
READ f:CP21 YEL
10
20
30
Guide number flashes, indicting strip positioning
The third pass requires the strip to be rotated to take the measurement.
1-9
Page 18
SECTION ONE
Film Strip
The Film category requires that the film strip be inserted into the 35mm slot located above the center diamond. In this example—Kodak C41—the strip can be inserted in either direction. Refer to your Control Strip and Balance Print Format Guide for information on all strips supported.
Indicates ALL patches are measured
20 30
10
10
20
AT 35mmSlot other
READ k:C41 ALL
30
Indicates the strip is inserted into the 35mm slot
Printer Balance Strip
The Printer Balance category in the instrument has five basic strip formats: Wht-Eye, Blk-Eye, No-Ring, K:3510, and Small-BE. Each category has six print options to choose from. Refer to your Printer Balance and Control Guide for detailed information on all aspects of the printer balance strips. Due to the amount of variation between printer balance strips, no guide setting numbers appear in the display. In this example—Blk-Eye, UNO 1-Pass—the Under patch is inserted first.
Indicates the under patch is inserted first
20 30
10
10
20
U
UNDER
NORMAL
N
O
OVER
UNO 1–Pass other
READ Blk–Eye UND
30
Indicates the strip type to measure
1-10
Page 19
S E C T I O N T W O
Instrument Calibration
This section covers the calibration procedures for your instrument.
Section Two Contents
Calibration Information
Auto Calibration Procedure
Manual Calibration Procedure
CALIBRATION INFORMATION
The instrument has two types of calibration procedures available, automatic and manual. AUTO is the standard method of calibration. If you want your instrument to measure the same as another densitometer that has the same response, you would use the MANUAL procedure. When the instrument automatically calibrates, the density values are set precisely to that of the paper measured for reflection, and to no film (air) for transmission.
The instrument has many self checking algorithms built in to verify the accuracy of calibration. If you wish to maintain accuracy verification, we suggest that you obtain a C41 reference strip and a paper reference strip. Measure each strip and record the density values. Place the strips in an envelope and store them in a dry, cool environment. Periodically, measure them to verify accuracy. Note, reference strips tend to drift in density with time, consult the manufacturer's specifications for expected density changes.
If your Auto-Cal strip gets ruined or worn, you can order a replacement from X-Rite or your local dealer. Order Part Number 880-100.
Frequency of Calibration
Under normal operating conditions, the instrument should be calibrated once a week or when instrument displays a message regarding calibration.
2-1
Page 20
SECTION TWO
AUTO CALIBRATION PROCEDURE
NOTE: Make sure the Auto-Cal strip is free of dust, dirt, and smudgemarks. Refer to Section Eight for cleaning procedure. Handle cal strip by the edges.
1. Press key lll (cal) located on Function Menu page (p3) to initiate calibration.
The instrument screen momentarily displays CALIBRATING
FUNCTION MENU
p3 cnfg cal
TRANSMISSION
(reading air), and
during automatic transmission calibration
CALIBRATING MOTOR SPEED.
load
Auto must be
displayed in upper-
case. If not, press key l to toggle on.
35mm slot
CALIBRATING
TRANSMISSION
CALIBRATING
MOTOR SPEED...
After this is complete, the instrument screen asks you to INSERT CAL
. Make sure the key labeled AUTO is upper-case lettering. If not,
STRIP
press key l (AUTO) to toggle to upper-case. Note, lower-case lettering denotes manual calibration (MANUAL) is activated.
INSERT CAL STRIP
AUTO
2. Insert the designated end of the Auto-Cal strip into the 35mm slot until it comes to rest against the drive rollers.
X-RITE Auto-Cal Strip P/N 880-100 Patent Pending
INSERT THIS EN D
10 10
manual
SERIAL #: XXX-XXX
CAL R= 0.04 G= 0.07 B= 0.08
Note insertion direction
READING is momentarily displayed on the screen followed by calibration density values. Auto calibration is now complete and the instrument screen returns to the Main menu.
NOTE: If when a cal strip is measured, the strip may be dirty—refer to Section Eight for procedure. However, if cleaning the strip does not resolve the problem, enter calibration manually then re-measure the strip.
2-2
UNRECOGNIZEABLE STRIP momentarily appears on the screen
Page 21
INSTRUMENT CALIBRATION
MANUAL CALIBRATION PROCEDURE
The manual calibration procedure is used to correlate the low densities of the instrument to another densitometer. Doing this allows the instrument to measure approximately the same as another densitometer that has Status A reflection response, and has been calibrated to ANSI Standards.
1. Using the instrument's own reference, calibrate the densitometer that you want this instrument to correlate with.
2. Using the same densitometer, measure the white area below the first black bar on the X-Rite Auto-Cal strip and record the Red, Green, and Blue values.
X-RITE Auto-Cal Strip P/N 880-100 Patent Pending
SERIAL #: XXX-XXX
INSERT THIS END
Measure here using other instrument.
3. Press key lll (cal) located on Function Menu page (p3) to initiate calibration.
CAL R= 0.04 G= 0.07 B= 0.08
FUNCTION MENU
p3 cnfg cal
The instrument screen momentarily displays CALIBRATING
TRANSMISSION
(reading air), and
during automatic transmission calibration
CALIBRATING
TRANSMISSION
4. After this is complete, the instrument screen asks you to INSERT CAL
STRIP
. DO NOT INSERT STRIP AT THIS TIME.
Press key llll (manual) to advance to calibration reference values screen.
load
CALIBRATING MOTOR SPEED.
CALIBRATING
MOTOR SPEED...
INSERT CAL STRIP
AUTO
manual
2-3
Page 22
SECTION TWO
5. Enter the Red, Green, and Blue values you previously recorded into the designated fields. Press key llll (
) to move the cursor to the desired
location, and to advance to the next step after editing the values. Press key ll (
) to increase value and key lll () to decrease value.
RED GREEN BLUE
Cursor
35mm slot
0.04
_
CAL R
0.07
0.08
6. INSERT CAL STRIP re-appears on the screen. Insert the designated end of the Auto-Cal strip into the 35mm slot until it comes to rest against the drive rollers.
INSERT CAL STRIP
auto
X-RITE Auto-Cal Strip P/N 880-100 Patent Pending
INSERT THIS END
10 10
MANUAL
Note insertion direction
SERIAL #: XXX-XXX
CAL R= 0.04 G= 0.07 B= 0.08
READING is momentarily displayed on the screen followed by calibration density values. Manual calibration is now complete and the instrument screen returns to the Main menu.
2-4
Page 23
S E C T I O N T H R E E
Setting System Configuration
The system configuration allows you to customize your instrument to meet your requirements. The configuration should be viewed and edited as needed before any measurements are taken.
Section Three Contents
Instrument Configuration
INSTRUMENT CONFIGURATION
The configuration functions are contained in two main screen pages. Page one contains setup functions for instrument beeper tone, and network. Page 2 is broken down into five menu pages, four of which are submenus—p2a, p2b, and so on. Page two menus contain serial interface control options.
Descriptions for all configuration options are listed below, followed by set up procedures.
Page 1 Configuration Options
TONE
Use to adjust the instruments beeper volume. Available settings are: OFF, SOFT, and LOUD.
NET (Networking) Enables the instrument to automatically send and receive data via the modem at a predetermined time that is selected by the QC computer. Available settings are: OFF, OUT, IN, and I&O.
OFF - disables the instrument from automatically sending or receiving data.
OUT - enables unit to call out via modem from QC host.
IN - enables unit to answer incoming call on modem from QC host.
I&O - enables both automatic transmission and reception of data.
3-1
Page 24
SECTION THREE
Page 2 Configuration Options
IO preset This options provides several QC system selections. All the necessary RS-232 interface parameters are automatically set when a predefined I/O preset is selected. A “CUSTOM” preset is available which allows you manually set all available I/O options. Below is a chart that contains all available I/O presets and their parameter settings.
NOTE: Certain parameters are mandatory for the I/ O preset and cannot be changed. They are labeled with an “L” for locked. Other parameters are not considered mandatory and are labeled with a “U” for unlocked. Finally, parameters that remain unchanged from previous settings—and can be changed—are shown as an asterisk (*).
I/O Preset RCI PIN5 XON DPT COMP ALF DEL AXMT BAUD
REPORT Off * * On (L) On (L) On (U) Off (L) Off (U) *
SPRD-SHT * * * * Off (L) * * On (L) *
k:TNetA Off (L) Off (L) Off (L) Off (L) On (L) On (L) On (L) On (L) 300(L)
k:TNetXT Off (L) CTS(L) Off (L) Off (L) On (L) On (L) Off (L) On (L) *
K:C.A.P. Off (L) CTS(U) Off (L) Off (U) On (U) On (U) Off (L) On (L) 9600(L)
k:SYS25-75 Off (L) CTS(L) Off (L) Off (L) On (L) On (L) Off (L) On (L) *
f:TECOM Off (L) Off (L) Off (U) Off (U) Off (U) Off (U) Off (U) On (U) 1200(U)
c:I/O#1 On (L) Off (L) Off (L) Off (U) On (U) On (U) On (U) On (L) 9600(L)
c:I/O#2 On (U) Off (U) Off (U) Off (U) On (U) On (U) On (U) On (U) 9600(U) MITSY#1 Off (U) Off (U) Off (U) Off (L) Off(L) Off(L) Off (U) On (U) 300(L) MITSY#2 Off (L) Off (L ) On (U) Off (L) Off(L) Off(L) Off (L) On (L) 300(U)
NORITSU Off (L) CTS(U) Off (L) Off (L) On (L) On (L) Off (L) On (L) 2400(U)
n:QSSnet Off (U) Off (U) Off (U) Off (U) Off (U) Off (U) Off (U) On (U) 2400(U)
n:Micro * * * * * * * * *
kn:Net TTL (U) Off (U) Off (U) Off (U) On (U) Off (U) Off (U) Off (U) 2400(U) a:LC2000 Off (U) Off (U) Off (U) Off (U) On (U) On (U) On (U) On (U) 1200(U) CUSTOM # # # # # # # # #
Windense On (U) Off (U) Off (U) Off (U) On (U) Off (U) Off (U) On (U) 9600 (U)
Kodatel On (U) Off (U) Off (U) Off (U) On (U) Off (U) Off (U) On (U) 9600 (U) Sienna On (U) Off (U) Off (U) On (U) On (U) On (U) Off (U) On (U) 4800 (U)
k: (Kodak), f: (Fuji), c: (Copal), kn: (Konica), n: (Noritsu), a: (Agfa)
3-2
Page 2a Configuration Options
RCI (Remote Control Interface) Controls the ability of the instrument to be operated remotely. Available settings are: OFF, ON, TTL.
OFF - sets the instrument into a receive mode at which time the instrument
receives data.
• ON - enables data to be received via RCI command.
TTL (Time Tagged Log) - causes all data received into the serial port to be placed into the log buffer with a time stamp.
PIN5
Determines the status of the handshaking input on Pin 5 of the RS-232 I/O port. Pin 5 may be interpreted as BUSY, CTS (clear-to-send), or OFF (ignored).
Page 25
SETTING SYSTEM CONFIGURATION
XON
When set to On (XON), this enables bi-directional transmit On/Off protocol. When set to Off (xon), the instrument ignores XON/OFF codes. Note, usage is limited to output at this point.
Page 2b Configuration Options
DPT (Decimal Point)
Controls the decimal point availability during data output. When off (dtp), no decimal is output with data. When on (DTP), decimal points are included with the output data.
COMP (Compact)
Varies the output format of the data. When off (comp), data for each color is delimited with a carriage return—or CR LF if ALF is enabled. When on (COMP), a space and delimiter is transmitted after each set of RGB data values.
ALF (Automatic Line Feed)
Controls the line feed availability during data output. When off (alf), no line feed is output with data. When on (ALF), a line is included with the output data.
Page 2c Configuration Options
DEL (Delay)
Controls a transmitted delay between each set if RGB data output. When off (del), no delay is sent between each set of RGB data. When on (DEL), a one second delay is sent between each set of RGB data.
AXMT (Automatic Transmit) Controls automatic transmission of data after a measurement. When off (axmt), no data is transmitted out the I/O port after a measurement. When on (AXMT), data is automatically sent out the I/O after a measurement.
Page 2d Configuration Options
BAUD (Rate) Determines the data output rate (characters per second) of the I/O port. Available settings are: 110, 300, 600, 1200, 2400, 4800, and 9600.
LOCK
Controls the ability to change the configuration options. When off (lock), configuration options are available for change. When on (LOCK), configuration options cannot be changed. Note, an additional step is required to change this option, refer to the following procedure for details.
Configuration Setup Procedure
Options that toggle on and off display in upper-case (on) or lower-case
(off). For example, COMP = on and comp = off.
3-3
Page 26
SECTION THREE
You can save and exit the configuration at anytime without editing all
Configuration can be exited at anytime without saving changes by
1. Press key ll (cnfg) located on the Function Menu screen (p3) to enter
2. Page 1 Configuration Screen
Press key lll (TONE) to page through and select a beeper tone option
Press key llll (NET) to page through and select a networking option
After editing Page 1 Configuration options, press key l (P1) to advance
options. This is accomplished by simultaneously pressing key l and key ll when SAVE is displayed in the screen.
simultaneously pressing key lll and key llll (MENU keys).
Page1 Configuration screen.
FUNCTION MENU
p3 cnfg cal
(OFF, SOFT, and LOUD).
(OFF, OUT, IN, and I&O. Note, only set if network modem is used.
to Page 2 Configuration screen.
load
SAVE
P1
3. Page 2 Configuration Screen
Press key ll or key lll (IOpreset) once to enter I/O preset menu.
CNFG
TONE NET
CNFG
P2 IOpreset
Press key ll or key lll to page through and select a preset options
(CUSTOM, REPORT, SPRD-SHT, k:TNetA, etc.).
After selecting I/O preset option, press key llll (P2) to load preset and
advance to Page 2a Configuration screen.
If you do not want to load a preset, press key l (P2) to advance to
Page 2a Configuration screen.
CNFG
P2 CUSTOM
load
3-4
Page 27
SETTING SYSTEM CONFIGURATION
4. Page 2a Configuration Screen
Press key ll (RCI) to page through and select a remote control interface
option (OFF, ON, TTL).
Press key lll (PIN5) to page through and select a handshake option (OFF,
BUSY, and CTS).
Press key llll (XON) to select bi-directional transmit On/Off protocol
option. Each key depression alternates between xon (off) and XON (on).
After editing Page 2a Configuration options, press key l (P2a) to advance
to Page 2b Configuration screen.
SAVE
P2a
5. Page 2b Configuration Screen
Press key ll (DPT) to select decimal point output option. Each key
depression alternates between dtp (off) and DPT (on).
Press key lll (COMP) to select data output format option. Each key
depression alternates between comp (off) and COMP (on).
Press key llll (ALF) to select automatic line feed option. Each key
depression alternates between alf (off) and ALF (on).
After editing Page 2b Configuration options, press key l (P2b) to
advance to Page 2c Configuration screen.
RCI
SAVE
P2b
6. Page 2c Configuration Screen
Press key ll (DEL) to select delay option. Each key depression alternates
between del (off) and DEL (on).
Press key lll (AXMT) to select automatic transmission option. Each key
depression alternates between axmt (off) and AXMT (on).
After editing Page 2b Configuration options, press key l (P2b) to
advance to Page 2c Configuration screen.
DPT
CNFG
PIN5 XON
CNFG
COMP ALF
P2c
SAVE
DEL
CNFG
AXMT
3-5
Page 28
SECTION THREE
7. Page 2d Configuration Screen
Press key ll (baud) to page through and select a baud rate option (300,
Insert a piece of paper into the paper slot at least 1-¼ to activate the read
After editing Page 2d Configuration options, simultaneously press
600, 1200, 2400, 4800, 9600, 19200, 38400, 57600).
switch—heavy paper or film works best. Press key llll (lock) to select lock option. Each key depression alternates between lock (off) and LOCK (on). A “LOCK” selection (padlock icon closed) prevents any option changes from occurring .
key l (P2d) and key ll (baud) to save configuration settings. Pressing key l (P2d) alone causes Page 1 Configuration screen to reappear.
SAVE
P2d
baud
CNFG
lock
Padlock icon displays “closed” when options are locked.
3-6
Page 29
S E C T I O N F O U R
Strip Operation
It is important to make sure control strips and reference strips used are from the same batch number.
Always inspect control strips for damage before measuring. Care must also be taken when handing strips. Fingerprints on measurable patches can affect density values.
IMPORTANT! When inserting strips into the instrument, there should be at least a 30.5mm (1.25) leader before the outside edge of the first measurable target or the first target may not be detected. For information on measuring leaderless strips (30.5mm), refer to your Control Strip and Balance Print Format Guide, Section Five.
Section Four Contents
Measuring Strips
Manually Transmitting Strip Data
MEASURING STRIPS
Refer to your Control Strip and Balance Print Format Guide to located your control strip for proper insertion direction.
1. Select strip channel from Main Menu page 1.
MAIN MENU
papp1 film
2. If format displayed is not correct, press key llll (other) to select correct
strip format.
READ k:RA—4 ALL AT
<18 18>
bal
other
4-1
Page 30
SECTION FOUR
3. Measure strip
For paper strips, adjust the paper guides according to the number
For film strips, note insertion direction and insert strip into the 35mm
For printer balance strips, note insert direction and center bull’s-eye
displayed on the screen. If multiple pass, note what color is to be measured first. Insert strip until it comes to rest against the drive rollers.
slot until it comes to rest against the drive rollers.
over middle diamond. Slide paper guides next to strip and insert strip until it comes to rest against the drive rollers.
Paper Guide
30
35mm Slot
20
10
Paper Guide
Center Diamond
10
20 30
The densitometer indicates that processing is taking place and the
measurement status. If a paper strip is measured that has multiple passes, the display indicates the pass to insert.
PROCESSING...
PASS #1 of 1 OK!
NOTE: If INVALID READING, UNRECOGNIZABLE STRIP, or BUFFER
OVERFLOW messages display after a measurement, re-read strip. If the same
message appears after re-reading strip, refer to Appendix B of this manual.
Transmitting Data momentarily appears in the display screen (if auto
transmit is enabled in Configuration Options), indicating measurement data is being transmitted out the I/O port.
The instrument screen displays the RGB measurement results. Refer to
next page for procedure to view field results.
r.rr g.gg b.bb Stain exit
4-2
TRANSMITTING
DATA
Page 31
STRIP OPERATION
View Strip Data
Control strip data consists of the “actual” values measured by the instrument. Control strip data is viewed after a strip measurement or by selection from
the Main Menu (p2). Viewing Data after Control Strip Measurement
1. Press key l (Stain) to page through RGB data fields of the strip.
2. After viewing data, press key llll (exit) to exit.
Red Green Blue
0.29 0.34 0.54
Stain
Viewing Control Strip Data from View Menu
1. Press key ll (view) located on Function Menu page (p2) to enter View
Menu screen.
exit
FUNCTION MENU
p2
view xmit
2. Press key ll (pap), key lll (film), or key llll (bal) to select strip category.
VIEW MENU
balpap film
3. Press key ll or key lll (Stain) to page through RGB data fields of the strip. After viewing data, press key llll (exit) to exit to the View Menu, or simultaneously press key lll and key llll (MENU) to exit to the Main Menu.
Red Green Blue
VIEW
NOTE: CHANNEL IS EMPTY appears in the display if no control strip data is available to view.
0.29 0.34 0.54
>
Stain
exit
4-3
Page 32
SECTION FOUR
MANUALLY TRANSMITTING STRIP DATA
The manual transmit function allows you to transmit data for the last measurement taken in the paper, film, and printer balance categories.
1. Press key lll (xmit) located on Main Menu page (p2) to enter Xmit Menu
2. Press key ll (pap), key lll (film), or key llll (bal) to select strip category.
screen.
p2
MAIN MENU
view xmit
XMIT MENU
balpap film
NOTE: CHANNEL IS EMPTY appears in the display if no control strip data is available in the selected channel.
3. Transmit strip data.
Press key l (xmit) to transmit all strip data. (“xmit” changes to upper-case during transmission.)
After transmitting data, press key llll (exit) to exit to the Xmit Menu, or simultaneously press key lll and key llll (MENU) to exit to the Main Menu.
film
XMIT
( k:C41 )
All)
(
exit
4-4
Page 33
S E C T I O N F I V E
Service and General
Maintenance
This section covers repair information, cleaning, general maintenance, and troubleshooting tips for your instrument.
Section Five Contents
Repair Information
Cleaning the Instrument
Replacing the Read Lamp
Troubleshooting Tips
REPAIR INFORMATION
The X-Rite 890 is covered by a one-year limited warranty and should be referred to the factory or an authorized service center for repairs within the warranty period. Attempts to make repairs within this time frame may void the warranty.
X-Rite provides a factory repair service to their customers. Because of the complexity of the circuitry, all repairs should be referred to the factory or an authorized service center (call: 1-888-826-3044).
X-Rite will repair any 890 instrument past warranty. Shipping cost to the factory or authorized service center shall be paid by the customer, and the instrument shall be submitted in the original carton, as a complete unaltered unit.
5-1
Page 34
SECTION FIVE
CLEANING THE INSTRUMENT
Your instrument requires very little maintenance to achieve years of reliable operation. However, to protect your investment and maintain reading accuracy, a few simple cleaning procedures should be performed from time to time.
General Cleaning
Whenever required, the exterior of the instrument may be wiped clean with a cloth dampened in water or a mild cleaner.
NOTE: DO NOT use any ketone solvents to clean the unit, this will cause damage to the cover.
Cleaning the Optics
The optics and drive wheel assembly should be cleaned once a week in normal environments, and more often in dirty or dusty environments.
1. Carefully lift instrument and insert tube from canned air into “front” strip insertion slot.
2. With back and forth motion, spray clean, dry air from one side to the other—do this several times. This should remove any accumulated dust and lint from the optics and wheel assembly.
WARNING: DO NOT invert cans that use freon as a propellant, doing so could cause damage to the optics assembly.
5-2
Page 35
SERVICE AND GENERAL MAINTENANCE
Cleaning the Calibration Strip (Part Number 880-100)
The Auto-Cal strip can be cleaned with a mild soap detergent, and wiped dry with a clean, lint-free cloth. You must let the calibration strip dry completely before taking a calibration measurement.
DO NOT clean below the bottom triangle. The attached label is not coated and smears when moisture is applied. DO NOT get fingerprints on any portion of the strip, handle by the edge.
X-RITE Auto-Cal Strip P/N 880-100 Patent Pending
INSERT THIS END
DO NOT CLEAN BELOW THIS LINE!
CAL R= 0.04 G= 0.07 B= 0.08
SERIAL #: XXX-XXX
5-3
Page 36
SECTION FIVE
REPLACING THE READ LAMP (PART NUMBER 880-07)
NOTICE: New lamp may appear bent, DO NOT attempt to straighten.
1. Remove four screws (2) securing the bottom cover (3) with a phillips-
2. Holding top (1) and bottom (3) covers in place, turn unit over so it rests
3. Locate optics assembly (4) and remove screw and washer (5) in the
4. Lift out old lamp assembly (6) and discard.
5. Install new lamp assembly (6) by carefully inserting lamp (6) into
6. Secure lamp screw and washer (5) in place.
7. Carefully clean any dust or plastic chips off circuit board and top cover (1)
8. Holding top and bottom covers in place, turn unit over so that it rests on
9. Remove bottom cover (3). Clean circuit board and bottom cover (3) with
10. Secure bottom cover (3) to instrument with four screws (2) using a
head screwdriver. Leave bottom cover (3) on unit.
on the bottom cover (3). Remove top cover (1).
middle of lamp assembly P.C.B. (6).
housing (7) and lamp pins (8) into lamp connector. Press down gently to make sure connector pins (8) are properly seated.
using moisture free compressed air. Place top cover back on instrument.
top cover (1).
compressed air, then place bottom cover (3) back on instrument.
phillips-head screwdriver.
1
5
6
8
4
7
3
2
5-4
Page 37
SERVICE AND GENERAL MAINTENANCE
TROUBLESHOOTING TIPS
Reflection measurement incorrect:
Recalibrate instrument.
Clean reflection cal strip or replace if bad.
Replace read lamp.*
Contact X-Rite or Authorized Service Center.
Transmission measurement incorrect:
Recalibrate instrument.
Replace read lamp.*
Contact X-Rite or Authorized Service Center.
Transmission and Reflection measurement incorrect:
Recalibrate instrument.
Replace read lamp.*
Contact X-Rite or Authorized Service Center.
Measurements drift:
Recalibrate instrument.
Replace read lamp.*
Contact X-Rite or Authorized Service Center.
Unit will not calibrate:
Clean or replace cal strip.
Read lamp not working.
Replace read lamp.*
Contact X-Rite or Authorized Service Center.
Measurements unrepeatable/incorrect:
Re-insert strip.
Use different strip.
Contact X-Rite or Authorized Service Center.
* The instrument has a failure monitor that in most cases automatically indicates when the read lamp requires replacement.
5-5
Page 38
SECTION FIVE
5-6
Page 39
A P P E N D I X A
Technical Specifications
Transmission Process Control
Film Width 35mm fixed slot or 1.4–6.0in. adjustable Measurement Speed 1.1 to 1.4 in./sec. (1.2 in./sec. typical) Spectral Response Status M Density Range 0–4.0D Density Accuracy ±.02D (0–3.0D), ±1% (3.1–3.4D)
±3% (3.5–4.0D)
Density Repeatability ±.01D (0–3.0D) Control Strip Measurement Area 0.375” (length) x 0.5” (wide) minimum
Reflection (paper) Process Control and Printer Balance
Paper Width 1.4–6.0in. adjustable slot Measurement Speed 1.3in./sec. Control Strip Measurement Area 0.375” L x 0.5” W minimum Printer Balance Measuring Area 0.75” diameter minimum Spectral Response Status A Density Range 0–2.5D Density Accuracy ±.02D Density Repeatability ±.01D
General Specifications
AC Adapter 12VDC @ 0.7amp 115VAC (P/N SE30-61) 230VAC (P/N SE30-62)
Instrument Dimensions 7.2” W x 6.0” D x 2.75” H (182.8mm W x 152.4mm D x 69.mm H)
Specifications and design subject to change without notice.
A-1
Page 40
APPENDIX A
A-2
Page 41
A P P E N D I X B
Error Messages
Below is a list of typical error messages that can appear on your instrument’s display screen. If any error messages listed, or not listed should appear, make a note of it and take the appropriate steps to try to correct it. If an error message is consistently displayed, contact X-Rite or an authorized service center.
Message Reason Possible Cause Solution
INVALID READING ­PLEASE RE-READ!
or
UNRECOGNIZABLE STRIP
or
BUFFER OVERFLOW
(during measurement)
Unit needs calibration Calibrate instrument, Sec. 2. Measurement region
One or more
Motor drive roller
Lamp failure (weak or
Unit did not recognize strip. Strip did not have a
Strip not inserted in the
Wrong strip selection. Select correct format.
30.5mm (1.2”) leader before first target.
correct direction.
not aligned with center diamond, or strip not tracking properly.
measurement patches are cloudy, have excessive gradients, or have flecks.
NOTE: If first read causes invalid error message, try re-reading strip again. Measurement tolerances across the patch are opened slightly on second read.
slippage due to restraint or obstruction, or contamination of rollers from reading wet strips.
bad).
Use strip with leader or refer to the Control Strip & Balance Print Format Guide for leaderless strip insertion. Refer to your Control Strip and Balance Print Format Guide for insertion direction.
Make sure measurement patches are centered with diamond and center line. Usually setting the paper guides to the numbers indicated on the display takes care of the problem. Make sure strip feed straight through unit and does not curve toward one side. Process and then measure a new strip.
Remove restraint/obstruction or dry drive rollers with air.
Perform transmission calibration to test lamp, Sec. 2.
B-1
Page 42
APPENDIX B
Message Reason Possible Cause Solution
UNRECOGNIZABLE AUTO-CAL STRIP!
(during reflection cal) Cal strip is dirty. Clean cal strip, Sec. 8.
STRIP RESTRAINED RE-INSERT STRIP!
(during reflection cal) Motor drive roller
WARNING MOTOR ERROR!
(during reflection cal) Motor brush wear. Return unit for service.
PRESET MEMORY PLEASE CALIBRATE
(during power-up)
WARNING LAMP MARGINAL!
(during trans cal)
WARNING REPLACE LAMP
(during trans cal)
Unit did not feed consistently.
Cal strip did not feed consistently.
Unit senses motor abnormality.
Memory data detected in unit is not valid. Lamp output is less than 50% of its peak intensity (but still able to read). Lamp output is less than required intensity. Measurement accuracy of unit is questionable at this point.
Strip inserted in
Insert strip correctly, Sec. 1. backwards or upside­down.
Strip path is blocked by
Clean strip path, Sec. 8. debris keeping cal strip from feeding properly.
Remove restraint/obstruction slippage due to restraint or obstruction, or contamination of rollers
or dry drive rollers with air. IF
problem persists, return unit
for service. from reading wet strips. Strip was pulled out from back during
DO NOT pull on strip during
measurement. calibration.
NOTE: If motor error message constantly displays, unit should be returned for proper service. Recalibrate unit, Sec. 2.
Lamp has aged close to end of its useful life.
Useful lamp life has expired.
Order new lamp. Replace at
convenient time.
Replace lamp immediately,
Sec. 8.
B-2
Page 43
A P P E N D I X C
Term Abbreviations
ALF Automatic Line Feed ALL Strip is one pass and may be read in either direction (strip measurement mode) AXMT Automatic Transmit bal Printer Balance baud Varies unit of data transmission speed. BLK Black patch must be read first (strip measurement mode) cal Calibration c:I/O#1 Copal, I/O #1 c:I/O#2 Copal, I/O #2 cnfg Configuration COMP Compact CTS Clear to Send CYN Cyan patch must be read first (strip measurement mode) def Default del Delete DEL Delay (in configuration) DPT Decimal Point f:TECOM Fuji, Tecom System HD High Density patch must be read first (strip measurement mode) k:C.A.P. Kodak, Create a Print
2
k:SYS
7
Kodak, Systems 25 through 75
5
5
k:TNetA Kodak, Technet A k:TNetXT Kodak, Technet XT LANG Language MIN Minimum Density patch must be read first (strip measurement mode) NRM Normal patch must be read first (strip measurement mode) NUO 1-Pass Normal/Under/Over in 1 pass pap Paper PIN5 Pin5 of RS232 port can be set to Off, Busy, or CTS p1 through p3 Page #1 through Page #3 p2a through p2d Page #2a through Page #2d
C-1
Page 44
APPENDIX C
Q.C. Quality Control RCI Remote Control Interface SPRD-SHT Spread Sheet UND Under patch must be read first (strip measurement mode) UNO 1-Pass Under/Normal/Over in 1 pass UNO 3-Pass Under/Normal/Over in 3 passes xmit Transmit YEL Yellow patch must be read first (strip measurement mode)
Increase amount Decrease amount Advances cursor
C-2
Page 45
A P P E N D I X D
Parts List and
Packaging Drawings
Parts List
D-1
Page 46
APPENDIX D
Packaging Drawing
D-2
Page 47
A P P E N D I X E
890 Instrument
Firmware Update
The instrument’s firmware is upgraded using Interface Cable P/N SE108-70 and DB Adapter P/N 881-71, both available from X-Rite.
The following items are required to perform a 890 instrument firmware update:
Computer with an available serial port
890 instrument with AC adapter
Update firmware disk
Interface cable (P/N SE108-70) and DB25 adapter (P/N 881-71)
NOTE: Computer screen savers must be disabled during the update procedure. Once the update has been started, the old firmware version is permanently erased.
Step 1 Cabling Connection Procedure
Insert one end of the interface cable P/N SE108-70 into the I/O port on the
instrument. Modular connector only inserts in one direction.
Insert the other end of the interface cable into the DB25 Adapter P/N 881-71.
Attach the DB25 adapter to an available “RS-232” I/O port on the computer.
890 Instrument
I/O Serial Port
Interface Cable P/N SE108-70
E-1
DB25 Adapter P/N 881-71
Page 48
A P P E N D I X E
Step 2 Setting Instrument to Receive Update
Apply AC power to the instrument.
Enter the Load Menu by pressing key ll (netwk) on page 3 (p3) Function
Press key lll (load), Waiting for Host is displayed.
Step 3 Firmware Update Procedure
If you are running DOS under Windows 3.x, you may need to exit Windows to run the Loader program reliably. You can open an MS DOS window using Win95 or NT.
Insert 3-1/2" Update disk into your computer 3-1/2" drive.
Change DOS prompt to appropriate drive (A or B).
Type: 89X and press [ENTER] key. The program now establishes
After the firmware is updated the unit goes through a normal startup routine.
Firmware update is now complete, remove disk and cabling.
NOTE: Make sure modem baud is set to "57.6K." To check baud, from P3 Function menu, press "Load" key, and then press "cfg" key. Next press "mod" key, and then press "baud" key. Adjust baud rate if required.
Menu.
communication with the instrument and begins update.
Main Menu Fail Message
A Main Menu Fail message will appear on the instrument display if an interruption in communication occurred during the update. Possible cause for this message is a power failure, cabling disconnect, etc. When this occurs, the instrument can only be set to the Waiting for Host mode from the Loader menu. Refer to procedure that follows.
From the Main Menu Fail display, press key llll (ok). The instrument displays
Press key III (load) and then press key IIII (recv) on Select Function menu.
Continue with Step 3.
the Loader menu.
Waiting for Host is re-displayed on the instrument.
E-2
Page 49
Page 50
Page 51
Page 52
Corporate Headquarters - USA 4300 44th Street SE Grand Rapids, Michigan 49512 Phone 1 800 248 9748 or 1 616 803 2100 Fax 1 800 292 4437 or 1 616 803 2705
Corporate Headquarters - Europe Althardstrasse 70 8105 Regensdorf Switzerland Phone (+41) 44 842 24 00 Fax (+41) 44 842 22 22
Corporate Headquarters - Asia Room 808-810 Kornhill Metro Tower, 1 Kornhill Road Quarry Bay, Hong Kong Phone (+852) 2 568 6283 Fax (+852) 2 885 8610
Please visit www.xrite.com
for a local office near you.
P/N 890-500 Rev. N
Loading...