The X-Rite 882 was designed to meet the quality control needs of the professional
photofinisher processing E-6, as well as the other common processes.
The 882 is very much like the X-Rite 881, except that the 881 has only Status M for
transmission operation, and the 882 has Status A and M for transmission operation.
Transmission Status A is used for checking Positive transparencies (i.e., E-6 films) and
Status M for checking Negative transparencies (i.e., C41 negatives). The 881 Operation
Manual explains all of the functions and operation of the 882. However, there are a few
differences in operation and they are explained in this 882 addendum.
The 882 has the additional capability of:
•Measuring E-6 (Kodak & Fuji), Q-Lab, AP44, and P-5 film strips
•Measuring 2 different types of Internegatives
The additional film strips are selected and measured in the same manner as the other
film strips explained in the 881 Operation Manual. The procedure for selecting and
measuring internegative strips is explained in Section 1 of this addendum. Refer to
Section 2 for definition on the additional film strips and internegatives.
The following are registered trademarks: Kodak; Fuji; Agfa; Ilford
All products mentioned are trademarks of their respective companies.
P/N 882-500 Rev. F-07/16/93
1. Measuring an Internegative
The 882 allows measurements of internegatives. It will optional compute the changes in exposure
and filter pack necessary to balance the internegative using the Density Difference method.
Step 1 - Press [film].
Step 2 - Press [other] until "Ineg Step 1" is
displayed.
Step 3 - Select the interneg type (see Section 2
for type definition).
Each depression of the [type] key will
alternate between "type 1" and "type 2"
(for this example "type 2" is selected).
Step 4 - Select interneg film type.
CHANNEL MENU
p1papfilmpbM
Process Type
READ
Typ 1other
READ
Typ 2
Interneg
Selection
READ
Typ 2
Ineg Step 1
6011
Ineg Step 1
6011
Ineg Step 1
4114
other
Other
Film Strips
other
Press either center key to select the film
type. Select "6011" if you are using film
similar to Kodak 6011 or 4112 Interneg
film. Select "4114" if you are using film
similar to Kodak 4114.
Step 5 - Insert strip into the 35mm slot (step 1
inserted first) until it rests against the
drive rollers.
Film Type
Selection
35mm SLOT
30
203010
10
20
1
Measuring an Internegative - continued
Step 6 - When the reading is completed, the
display will indicate which step pair will be
used on subsequent calculations.
•Pressing the [view] key allows viewing
of density difference numbers (refer to
Section 1.1).
•Pressing the [+Step-] key allows
viewing of the individual steps on the
strip (refer to Section 1.2)
•Pressing the [auto] key performs filter
pack and exposure calculations
automatically (refer to Section 1.3).
› If the strip is severely over exposed, the warning message
"STRIP TOO FAR OVEREXPOSED" will appear briefly. You will
then be allowed to view the individual steps using [+] or [-]
step keys, but will not be able to use the view and auto
functions. If this happens, reduce your exposure by three or
four stops and try another strip.
1.1 View (continued from Step 5 in Section 1)
STEPPAIR
view+ step -auto
See Section 1.1See Section 1.2See Section 1.3
= 3, 13
The View function allows you to view the density difference values and recommended
filter pack changes. Density difference values are calculated by simply subtracting the
density values of the higher numbered step from the densities of the lower numbered
step. These are used to look up your filter pack changes from the table.
Step 6 - Press [∆fp] key to get the filter pack
changes from the internal table and
D = 0.62
0.750.80
fpexit
display them.
› These numbers are changes (indicated by the Greek letter
∆) and should be added to the filter pack which was used
for exposing this internegative.
Step 7 - Press [∆den] key to display the density
FP -03C
+05M+02Y
denexit
difference values again.
Step 8 - Press [exit] to return to the Step Pair
Menu so that further viewing or
calculations can be performed.
To Step Pair Menu
To Step Pair Menu
2
1.2 +Step- (continued from Step 5 in Section 1)
The +Step- function allows viewing of the individual steps on the strip.
Step 6 - Press either the [+] or [-] key to view
data on each step of the internegative.
•[+] increments to the next step (step 21 will
advance to step 1).
•[-] decrements to the next step (step 1 will
advance to step 21).
Step 7 - Press [exit] to return to the Step Pair
Menu so that further viewing or
calculations can be performed.
› For "Type 1" internegatives (4 x 5 format), all of the even
numbered steps are computed internally using a curve
fitting algorithm because the actual internegative contains
only the odd numbered steps.
1.3 Auto (continued from Step 5 Section 1)
0.20
0.470.83
+ step -exit#21
Next
Step
Previous
Step
To Step
Pair Menu
The Auto function initiates the automatic
exposure and filter pack compensation.
Step 6 - "Enter Starting Filter Pack..." is
momentarily displayed. Enter the values
of the filter pack used to make this
internegative.
•[↑] increases the value marked by cursor.
•[↓] decreases the value marked by cursor.
•[→] moves the cursor to the next position.
Press [cont] when finished.
Step 7 - "Enter Exposure" is momentarily
displayed. Enter your exposure time and
aperture.
› The "f" stop values increase or decrease in 1/6 stop steps
and values can range from f/4.0 to f/32.0.
•[↑] increases the value marked by cursor.
Moves
Cursor
Enter Starting
Filter Pack . . .
FP
EXP
00C
Enter Exposure
10 sec
30M30Y
DecreasesIncreases
f/8.0
cont
To Next
Step
cont
•[↓] decreases the value marked by cursor.
•[→] moves the cursor to the next position.
Press [cont] when finished.
Moves
Cursor
DecreasesIncreases
To Next
Step
3
Loading...
+ 9 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.