Wuhan Guide EasIR-4, EasIR-2, EasIR-1 User Manual

EasIR
TM
-4 / EasIR
TM
-2/ EasIR
TM
-1
Thermal Camera
Wuhan Guide Infrared Co., Ltd. No. 26 Shucheng Rd, Hongshan District, Wuhan 430070 P. R. China Telephone: +86-27-8767 1991 Facsimile: +86-27-8767 1927 Email: overseas@guide-infrared.com Internet: www.guide-infared.com © Wuhan Guide Infrared Co., Ltd., 2009 Version No: 2010-02-22
The Quality Management System of Wuhan Guide Infrared Co., Ltd. is approved to
ISO9001:2000 for the design and manufacturing, stockholding, in-house repair and site
servicing of non-contact temperature measuring instrumentation.
Wuhan Guide Infrared Co., Ltd. reserves the right to make changes and improvements on
any of the products described in this manual without prior notice.
EASIR
TM
-4/ EASIRTM-2/ EASIRTM-1 Thermal Camera complies with
current European directives relating to electromagnetic compatibility and
safety. (EMC directive 89/336/EEC; Low voltage directive 73/23/EEC).
Copyright
© Wuhan Guide Infrared Co., Ltd, 2009. All rights reserved worldwide. No parts of the
products may be reproduced, transmitted, transcribed or translated into any language or
computer language in any form or by any means, electronic, magnetic, optical, manual or
otherwise, without the prior written permission of Wuhan Guide Infrared Co., Ltd.
This manual must not, in whole or part, be copied, photocopied, reproduced, translated or
transmitted to any electronic medium or machine readable form without prior consent, in
writing, from Wuhan Guide Infrared Co., Ltd.
Table of the Contents
Table of the Contents .................................................................................................. 1
Introduction ................................................................................................................. 1
Precautions ................................................................................................................. 1
Maintenance................................................................................................................ 2
Calibration and Repair Philosophy .............................................................................. 2
Technical Support ........................................................................................................ 3
Feedback to Us ........................................................................................................... 3
System Overview ........................................................................................................ 3
System Configuration .................................................................................................. 3
Technical Specification ................................................................................................ 4
System Features ......................................................................................................... 6
Parts Described ........................................................................................................... 8
Charging Instruction .................................................................................................... 9
Buttons Introduction .................................................................................................. 10
Function operation .................................................................................................... 14
User Manual
1
Introduction
This publication provides the necessary information required to safely operate the
EASIR
TM
-4/ EASIRTM-2/ EASIRTM-1 Thermal Camera.
It is important to fully check all equipment with which you have been supplied.
The equipment should be used, maintained and serviced by suitably trained personnel,
capable of carefully following the procedures and guidelines given in this User Manual.
All User Manuals and leaflets should be read thoroughly before proceeding with operation
of the equipment.
It is also advisable that all User Manuals and Instruction Leaflets supplied are kept readily
available, for reference when the equipment is in general use.
Precautions
The following precautions must be adhered to at all times and must be considered in
addition to any advised precautions issued at the relevant worksite or work area.
Keep the EASIR
TM
-4/ EASIRTM-2/ EASIRTM-1 Thermal Camera steady during
operation.
Do not use the EASIRTM-4/ EASIRTM-2/ EASIRTM-1 Thermal Camera in
temperature exceeding its working and storage temperature ranges.
Do not direct the EASIRTM-4/ EASIRTM-2/ EASIRTM-1 Thermal Camera at very high
intensity radiation sources such as the sun, carbon dioxide lasers or arc welders
etc.
Do not expose the EASIR
TM
-4/ EASIRTM-2/EASIRTM-1 Thermal Camera to dust
and moisture. When operating the unit near water, ensure that the unit is
adequately guarded against splashes.
When the EASIR
TM
-4/ EASIRTM-2/EASIRTM-1Thermal Camera is not in use or is to
be transported, ensure that the unit and its accessories are stored in the protective
carry case.
Do not jam the holes or loudspeaker on the camera body.
Do not re-switch on the camera until 15 seconds later after switching it off.
Do not throw, knock or vibrate intensely the camera and its components in order to
keep them from damage.
Do not attempt to open the camera body, as this action will void the warranty.
Keep the SD memory card for the exclusive use of the camera.
During using, if move the IR camera from hot/cold place to cold/hot place, e.g.
from inside/outside to outside/inside of a air-conditioned car in winter, should
power off the camera firstly and leave it in the operation site for 20 minutes, then
power on the camera for normal operation and accurate temperature
measurement. Sharp and instant ambient temperature change may cause fault
temperature measurement and even damage camera’s IR detector.
User Manual
2
FPA setting: to ensure accurate temperature measurement, the FPA detector was
calibrated in different temperature points--7degree, 17degree, 27degree,
37degree. When power on the camera in 0 degree for example, after period of
working the temperature of the camera (the detector) will increase gradually. And
when the internal temperature of the camera beyond 7 degree, FPA setting will
take place and will last for about 30sec, and during the FPA setting the camera will
not response for any operations. If power on the camera in a temperature about 8
degree, only when the temperature of the camera beyond 17degree, this
phenomenon will happen again. Camera will adjust FPA setting itself
automatically to ensure its stable performance.
Maintenance
To ensure that the EASIRTM-4/ EASIRTM-2// EASIRTM-1 Thermal Camera is kept in good
working condition and remains fully operational, the following guidelines should be
adhered to at all times.
Non-optical surfaces
The non-optical surfaces of the camera can be cleaned when required, with a soft cloth
dampened with water and a mild detergent.
Optical surfaces
The lens of an IR camera is very expensive. Replacement or repair may be thousands of
dollars. The anti-reflective coating on the surface of the lens is the most expensive part of
the lens assembly (and is also critical to the radiometric capabilities of the system).
The optical surface should only be cleaned when visibly dirty. Care should be taken to
avoid oil, chemical dirt and touching the exposed lens surface, as skin acid left behind from
fingerprints can be damaging to coatings and lens substrates. After using the imager,
please close the lens cover.
Do not use dilution to clean the imager and its accessories, especially the optics. Use clean
soft dry tissue to clean the imager body, and the supplied lens cleaning tissue for lens.
Calibration and Repair Philosophy
To ensure the accuracy and reliability of the EASIRTM-4/ EASIRTM-2/EASIRTM-1 Thermal
Camera, it is highly recommended that the instrument be calibrated at 12 monthly
intervals.
Calibration or repair for the instrument can be obtained by either contacting the address/
telephone number on the cover of this User Manual, or by email to the following
address: overseas@guide-infrared.com
Caution
User Manual
3
The EASIRTM-4/ EASIRTM-2/EASIRTM-1Thermal Camera does not incorporate any user
serviceable parts. Never attempt to disassemble or modify the camera. Opening the unit
invalidates the warranty.
Technical Support
Technical support for your Wuhan Guide Thermal Imaging System can be obtained by
either contacting the address / telephone number on the cover of this User Manual or by
email to the following address: overseas@guide-infrared.com
Feedback to Us
We have tested and verified the information in this manual to the best of our abilities. Yet
as we are committed to continuous development and progress, you might find features of
the product have been changed since the time of printing. You are appreciated to let us
know about any error you find, and your suggestions for further editions by either
contacting the address/telephone number on the cover of this User Manual or by email to
the following address: overseas@guide-infrared.com
System Overview
EASIRTM-4/ EASIRTM-2/ EASIRTM-1 is a new infrared camera of Guide infrared, breaks
the IR world with its lowest price and high performance. Designed for tough work
environments and entry-level users, EASIRTM-4/ EASIRTM-2/ EASIRTM-1 is far more
robust and shock-resistant for any tough working environment and it is easy to operate
and allows for the learners to operate without being trained and take the inspection work
easily with one hand .Featured with latest InfraFusion technology, it helps you pinpoint the
problem exactly with the most efficiency. Power on the EASIR
TM
-4/ EASIRTM-2/
EASIR
TM
-1, let the 3.6" LCD bring you into the fresh IR world
System Configuration
Please ensure that the following items have been correctly supplied:
IR Camera with visual camera, laser locator
11mm IR lens & protection cover
3.6" TFT LCD with high resolution
2GB SD card & card reader
12 AA rechargeable batteries
AC Adapter & cable
USB extension cable
USB driver
User Manual
4
Guide IrAnalyser® Software
User manual
Wrist strap
Safety case
Soft bag
Options
30mm Tele lens
7mm wide angle lens
Extended temperature range up to +350
1200 high temperature filter
Sun Shield
Tripod mount
Technical Specification
Imaging Performance
THERMAL
EASIRTM-4 EASIRTM-2 EASIRTM-1
Detector type: Uncooled FPA microbolometer (160× 120 pixels, 25μm)
Spectral Range: 8-14μm
Thermal Sensitivity:
100mk at 30
Field of View/ Focus: 20.6° X 15.5°/ 11mm
Palette 8 6 4
Focus: Automatic or motorized
thermal & visual
Motorized (thermal) Motorized (thermal)
VISUAL
Built-in Digital Video: CMOS Sensor, 1600x1200 pixels, 224 true colors N/A
Image Presentation
External Display: 3.6" TFT LCD with high resolution
Video Output : PAL/ NTSC N/A
Live thermal video Recording via USB to PC N/A
Built-in Flash Yes N/A
Infra Fusion: Visual and IR blending
Man-Machine Communication
Buttons: Respond as per operators’ operation
User Manual
5
Menu: Microsoft® Windows style
Measurement
Temperature Range:
-20 to 250 (350,optional) -20 to 250
Accuracy:
±2 or ±2% of reading
Emissivity Correction: Variable from 0.01 to 1.00 (in 0.01 increment)
Measurement Features: Automatic correction based on distance, relative humidity, atmospheric transmission
and external optics
Optics Transmission Correction:
Auto, based on signals from sensors
Image Storage
Type: Removable 2GB SD card & built-in memory
File Format: JPG with analysis records
Voice Annotation : Up to 60 seconds N/A
Laser Locator
Classification Class 2 semiconductor laser N/A
Power System
Battery Type: AA rechargeable battery, field-replaceable, AA Alkaline battery also usable
Charging System: In camera or in battery charger
Battery Operating Time: Over 3 hours continuous operation
External Power Operation:
AC adapter 110/ 220 VAC, 50/ 60Hz
Environmental Specification
Operating Temperature:
-10 to 50
Storage Temperature:
-20 to 60
Humidity: Operating and storing 10% to 95%, non- condensing
Encapsulation: IP54 IEC 529 housing
Shock: Operational: 25G, IEC 68-2-29
Vibration: Operational: 2G, IEC 68-2-6
Interfaces
USB 2.0: For video and
image transfer
N/A
Physical Characteristics
Weight: 0.73KG(Excluding battery)
Size: 111mmx124mmx240mm
Colors Yellow & Grey interlaced or Red & Black interlaced alternative
User Manual
6
System Features
Features Advantages Benefit
Speedy 50 Hz or
60 Hz. IR
camera
makes sharp IR images
especially in cold outdoor climates, hand shakes
do not appear on IR images
Auto focus for
both IR and
visual image
fast (ZipZip) and best focus by
pressing one button only – quick
buttons 2 places on camera for
ergonomical fitness.
efficiently save your inspection time,
manual electric focus also possible.
Innovative AGT
(auto gate
technology)
not only functions as shutter but also lens cap, free from all disturbing heat
caused by optics and electronics
inside
ensures the most stable and accurate
temperature reading; protects the camera's
sensitive optics and detectors.
3.6” LCD screen
delivers better view, and clear menu
with large letter
clear image presentation and menu reading -
espcially for InfraFusion !
Navigation
interface
clear instructions for operation step
by step
easy operation for any users, especially entry
level users
Powerful
analysis
software
support image information read-in,
powerful further analysis, live trend
analysis, multipage generation by
over 9 IR images, real-time video
recording, report generation
analyzes the image with the help of various
analysis method and generate report.
Analog video
output
realtime PAL/NTSC thermal video
output
convenient for demonstration and online
monitoring, PAL/NTSC switchable on one camera
satisfies customers from different countries and
areas
Digital
temperature
data output
realtime recording of IR data stream (IR active video). Recording of up to
25 fps. of thermal image data via USB
2,0 to PC hard drives.
Temperature data stream recordings - object
realtime analyze in realtime by "endless"
recording on PC hard disc' - disc is the time limit.
Rugged
designed for
hard work
Approved for 2 mtr. drops
The camera last long time - years of field service
in nearly any environment.
2.0 Megapixel CMOS
offers high resolution visual image
better assists finding the trouble spot & know
better about inspection environment and target
Universal AA
battery
available everywhere
more convenient to buy the alternative batteries
locally.
Loading...
+ 20 hidden pages