WPI LWCC-M-10, LWCC-M-100, LWCC-M-50 Instruction Manual

Page 1
Instrumenting scientific ideas
WORLD PRECISION INSTRUMENTS
LWCC-M-xxx
Measure low volume samples with UV/VIS absorbance spectroscopy
Serial No._____________________
www.wpiinc.com
INSTRUCTION MANUAL
111717
Page 2
Page 3
LWCC-M
World Precision Instruments i
CONTENTS
ABOUT THIS MANUAL ................................................................................................................... 1
INTRODUCTION .............................................................................................................................. 1
General Warnings and Cautions ........................................................................................... 2
INSTRUMENT DESCRIPTION ........................................................................................................ 2
Parts List ...................................................................................................................................... 2
Unpacking ................................................................................................................................... 2
Hardware Description .............................................................................................................. 3
Required But Not Provided (see Accessories) .............................................................. 3
Assembly ............................................................................................................................... 3
OPERATING INSTRUCTIONS ......................................................................................................... 4
Transmission Verication ........................................................................................................ 4
Noise Determination ................................................................................................................ 4
Refractive Index ......................................................................................................................... 4
Measuring in a Continuous Flow ........................................................................................... 5
Measuring Discrete Samples with WPI’s Injection System .............................................. 5
INSTRUMENT MAINTENANCE ...................................................................................................... 6
Standard Cleaning .................................................................................................................... 6
Advanced Cleaning .................................................................................................................. 6
Storage ........................................................................................................................................ 7
ACCESSORIES................................................................................................................................... 7
TROUBLESHOOTING ..................................................................................................................... 8
Typical Contamination Eects ................................................................................................ 8
Additional Information on Contamination .......................................................................... 9
SPECIFICATIONS ............................................................................................................................ 10
Related Patents .......................................................................................................................10
INDEX ............................................................................................................................................... 11
WARRANTY .....................................................................................................................................12
Claims and Returns ................................................................................................................12
Copyright © 2017 by World Precision Instruments, LLC. All rights reserved. No part of this publication may be reproduced or translated into any language, in any form, without prior written permission of World Precision Instruments, Inc.
Page 4
ii World Precision Instruments
Page 5
LWCC-M
World Precision Instruments 1
ABOUT THIS MANUAL
The following symbols are used in this guide:
This symbol indicates a CAUTION. Cautions warn against actions that can cause damage to equipment. Please read these carefully.
This symbol indicates a WARNING. Warnings alert you to actions that can cause personal injury or pose a physical threat. Please read these carefully.
NOTES and TIPS contain helpful information.
Fig. 1—The LWCC-M is available with 10, 50 and 100mm pathlengths.
INTRODUCTION
The Micro Liquid Waveguide Capillary Cell (WPI #LWCC-M-xxx) is a ber optic, low volume ow cell for UV/VIS/NIR absorbance analysis; xxx represents the pathlength in mm. The LWCC-M is available with 10mm, 50mm and 100mm pathlengths. Based on WPI’s established liquid core waveguide technology, the analyte solution functions as the core of a uid-lled light waveguide. Wetted parts in the sample cell light path are PEEK, fused silica and Teon. Optical bers are used to transport light to and from the sample cell. The cell can be used in biochemistry for DNA, RNA and protein quantication, colorimetric nutrient and trace metal analysis, drug discovery and dissolution testing,
process control and HPLC analysis.
Similar to optical bers, light is conned within the (liquid) core of an LWCC-M by total internal reection at the core/wall interface. Ultra-sensitive absorbance measurements can be performed in the ultraviolet (UV), visible (VIS) and near-infrared (NIR) to detect
low sample concentrations in a laboratory or process control environment.
Page 6
2 World Precision Instruments
General Warnings and Cautions
NOTE: Fluid pressure uctuation, cross-contamination and the introduction of small air
bubbles can cause baseline variations and measurement inaccuracies. Use the optional Sample Injector Kit (WPI #58006) or the LWCC Injection System (WPI #89372) to minimize the development of small air bubbles.
NOTE: WPI’s Waveguide Cleaning Kit (WPI #501609) is recommended for cleaning the LWCC-M between uses and sample runs.
INSTRUMENT DESCRIPTION
Parts List
After unpacking, verify that there is no visible damage to the instrument. Verify that all
items are included: (1) LWCC-M (1) Instruction Manual (1) Quality Control documentation
Unpacking
Upon receipt of this instrument, make a thorough inspection of the contents and check
for possible damage. Missing cartons or obvious damage to cartons should be noted on the delivery receipt before signing. Concealed damage should be reported at once to the carrier and an inspection requested. Please read the section entitled “Claims and Returns” on page 13 of this manual. Please contact WPI Customer Service if any parts are missing at 941.371.1003 or customerservice@wpiinc.com.
Returns: Do not return any goods to WPI without obtaining prior approval (RMA # required) and instructions from WPI’s Returns Department. Goods returned
(unauthorized) by collect freight may be refused. If a return shipment is necessary, use the original container, if possible. If the original container is not available, use a suitable
substitute that is rigid and of adequate size. Wrap the instrument in paper or plastic surrounded with at least 100mm (four inches) of shock absorbing material. For further
details, please read the section entitled “Claims and Returns” on page 13 of this manual.
Page 7
LWCC-M
World Precision Instruments 3
Hardware Description
Required But Not Provided (see Accessories)
Fiber Optic Cables (2)
Detection System including either a spectrophotometer or a spectrometer and light source
Assembly
CAUTION: Unlike electrical cables, ber optic cables are fragile as they contain
glass and are subject to breakage. Avoid sharp bends in the cables and protect them from impact or permanent damage may result.
External Fiber Optic Cable Connections
Light Input
500µm SMA
Light Output
500µm SMA
Fluid Input
800827 Connector
Fluid Output
800827 Connector
1.38”
1.00"
Fig. 2—LWCC-M-10 is shown. The LWCC-M-50 and LWCC-M-100 are longer.
The LWCC-M (10, 50, 100mm) has SMA 905 terminated (WPI #FO-500-SMA1M) ber optic connections. The light source and detector can be connected to the LWCC-M via two SMA- terminated ber optic cables with a core diameter of 500µm. The ber optic connections are interchangeable in that either connector can be used to connect to the light source or to the spectrometer. Use these ber optic connections to connect the LWCC-M to a light source and spectrometer (detector) module of your spectrophotometer system.
Liquid Ports
As with the ber optic connectors, the liquid ports are also interchangeable. It makes no dierence which port is the inow and which port serves as the outow. However, if
Page 8
4 World Precision Instruments
an in-line lter is to be installed, connect it on the input side of the liquid ow path. Use standard HPLC type 10-32 coned PEEK 1/16th tubing ttings (WPI #800827) to make
connections to the pump and waste receptacle.
OPERATING INSTRUCTIONS
CAUTION: Materials exposed to uid in the LWCC-M are PEEK, FEP and fused
silica. Any chemical that could attack these substances should not be used in the LWCC-M. For example, hydrouoride (HF) will dissolve silica and PEEK will be
damaged by concentrated sulfuric and nitric acids (40% w/w or greater).
CAUTION: Keeping the LWCC-M clean is essential for a stable result. See “Instrument Maintenance” on page 6.
Transmission Verication
1. Connect one 500µm diameter ber optic assembly to a light source and one 500µm diameter ber optic assembly to the detector.
2. Use an SMA bulkhead adapter (WPI #13395) to butt-couple the bers. This is the reference for the measurement of transmission.
3. Remove the SMA bulkhead adapter and insert the LWCC-M.
4. Record the percent transmission. Values for the 254nm and 540nm should exceed 40% and 50% respectively.
Noise Determination
Noise is determined using the ASTM E685-93 standard as reapproved in 2000. Use methanol as the solvent at a ow rate of 1.0mL/min. Using the WPI test setup, noise values are below 35µAU. See www.astm.org/Standards/E685.htm to obtain a copy of
the standards document.
Refractive Index
The refractive index between methanol and cyclohexane is measured using a ow rate of 3.0mL/min. Because of the immiscibility of the solvents, ethanol is used between the
solvents to prevent unwanted interaction between methanol and clyclohexane. At the
280nm wavelength, RI values shown below are acceptable:
LWCC-M-10 0–7
LWCC-M-50 0–14
LWCC-M-100 0–25
Page 9
LWCC-M
World Precision Instruments 5
Measuring in a Continuous Flow
This is the recommended method.
1. To help prevent clogging the ports, lter all samples using a 0.2µm vacuum ltration
disc before injection into the LWCC-M.
2. Using ber optic cables, connect the LWCC-M to a light source and a detector.
3. Clean the LWCC-M using the standard cleaning procedure described in “Instrument Maintenance” on page 6.
4. Connect the liquid source to the LWCC-M system. A pressure of approximately
1.5–3.0 PSI is necessary to run liquid through the LWCC-M.
5. At a rate of 1mL/min., ush the LWCC-M with de-ionized water or experimental buer
solution using a pump or a syringe. Observe the light intensity or absorbance baseline
on the detector. Continue ushing until the signal is stable. See “Troubleshooting”
on page 8 if the signal does not stabilize.
Measuring Discrete Samples with WPI’s Injection System
The sample measurement procedure can be improved using the optional LWCC Injection system (WPI # 89372)
as discrete, reproducible
sample volumes can be introduced to the LWCC-M with a seamless transition from the baseline solution to sample at a continuous flow rate.
The LWCC Injection System,
together with a pump and liquid waveguide capillary cell (LWCC-M), provides an ecient, continual flow for injecting a sample through the LWCC-M.
Fig. 3—The LWCC Injection System is shown with MINISTAR peristaltic pump and LWCC-M. Pump and LWCC-M are sold separately.
Page 10
6 World Precision Instruments
INSTRUMENT MAINTENANCE
Thorough and consistent cleaning routines are essential for maintaining the instrument and ensuring optimal operation.
Standard Cleaning
With normal use, the LWCC-M-10 will rarely need cleaning. If measurement results diminish, clean your LWCC-M-10. Use the cleaning kit for liquid waveguides (WPI
#501609).
1. Connect the exit tubing (for example, silicon or equivalent) from a uid port (output)
of the LWCC-M to a waste container.
2. Rinse the cell thoroughly using Ultra Pure water. Obtain a new reference intensity and take a baseline absorbance reading.
3. Inject 1cc “Cleaning Solution 1” into the opposite LWCC-M uid port (input).
4. Inject 1cc “Methanol Solution 2” into the uid port (input) of the LWCC-M.
5. Inject 1cc “HCl Solution 3” into the uid port (input) of the LWCC-M.
6. Flush out the cleaning solutions with distilled, Ultra Pure, reverse osmosis or
equivalent quality water and take an absorbance reading.
7. Repeat these cleaning cycles until a stable absorbance signal can be obtained.
8. If the UV readings are still low after cleaning, swab the ber optic surfaces of the sample cell with a ber optic Q-tip dipped in isopropyl alcohol.
NOTE: The sample injector kit (WPI #58006) can be used to conveniently ll all LWCC-M models. Using the sample injector, the injection volume of the cleaning solutions per
cycle can be reduced to the internal volume of the LWCC-M (for example, 0.26mL for a 100cm LWCC-M).
Advanced Cleaning
1. Connect the exit tubing (for example, silicon or equivalent) from a uid port (output)
of the LWCC-M to a waste container.
2. Rinse the cell thoroughly using Ultra Pure water. Obtain a new reference intensity and take a baseline absorbance reading.
3. Fill the LWCC-M with a 10% solution of Contrad-NF (available from Decon Labs, Inc.–www.deconlabs.com part #6002). Allow the solution to sit inside the cell for 20 minutes.
4. Flush out the Contrad-NF solution with distilled, Ultra Pure, reverse osmosis or equivalent quality water and take an absorbance reading.
Page 11
LWCC-M
World Precision Instruments 7
5. Repeat steps 3–4 until a stable absorbance signal can be obtained.
NOTE: For removal of persistent contamination, the 10% Contrad-NF solution can be heated to 60ºC prior to injection into the LWCC-M.
Storage
CAUTION: Do not leave the LWCC-M “half dried” and open to the air. Oxygen
in the air may facilitate the growth of microorganisms inside the device.
To store the instrument, clean the LWCC-M and then remove all liquid using a peristaltic pump or syringe. Seal the uid ports using either the caps provided or an alternative.
ACCESSORIES
Table 1: Accessories Part Number Description
FO-500AS-SMA 1M, SMA/500µm core, anti-solarization FO-500-SMA1M 1M, SMA/500µm core, UV-enhanced 15807 Cleaning solution concentrate–100g (Solution 1) 58450 External syringe adapters 89372 LWCC Injection System 501609 Waveguide cleaning kit 500320 Silicone tubing for Peri-Star pump. 1.6mm ID x 1.6W 800490 Fiber optic foam swabs (5) MINISTAR Miniature DC Peristaltic Pump PERIPRO-8LS Peri-Star™ 8-channel, low rate peristaltic pump PERIPRO-4LS Peri-Star™ 4-channel, low rate peristaltic pump TIDAS-1 Tidas I high performance photodiode array
spectrometer module
TIDAS-BASIC Tidas E high performance ber optic based photodiode
array Spectrophotometer
D4H D4H™ deuterium/halogen light source FO-6000 Tungsten light source
Page 12
8 World Precision Instruments
TROUBLESHOOTING
The LWCC-M is a highly sensitive device, and it is extremely important to keep it clean. This is especially important when working in the ultraviolet range, where unexpected
results may often be produced by contamination of the experimental solution. The high sensitivity of the LWCC-M may create some problems that can be easily overcome with care and forethought — the user may need to develop new skills in handling both the equipment and the samples being examined.
Typical Contamination Eects
Issue Possible Cause Solution
Transmission in both UV and VIS ranges
becomes low or very unstable
A contamination layer (such as biofilm) sticking to the LWCC-M wall
Flush cell for 30 seconds each of each of the three cleaning solvents in the wave­guide cleaning kit (WPI Part #501609).
Prepare a 10% Surfactant using Contrad- NF concentrate followed by HPLC grade Methanol and HPLC grade 2N HCl solution.
A particle trapped in the
LWCC-M
Attach 1–5cc syringes with luer ttings to the
SAMPLE IN and SAMPLE OUT ports on the LWCC-M. Fill one syringe with distilled water and inject it into the LWCC-M. Use the other syringe to push the water back through the LWCC-M in the opposite direction. Continue using both syringes to move the liquid rapidly through the LWCC-M in alternating directions.
UV transmission is
low. VIS is ne and
stable.
Optic ber and silica tubing
are coated by a layer of metal corrosion.
Flush with 1N HCl.
Optic ber and silica tub-
ing is coated by a layer of organics.
Flush with an organic solvent, such as
acetonitrile.
Transmission
below 250nm is
low. VIS is ne
and stable
Contamination of ber optic cable end-faces with a metal lm generated during
repeated connection attempts
Wipe all ber optic end-faces, ber
connections on the LWCC-M using a ber optic foam swab (WPI #800490) dipped in methyl alcohol. NOTE: Do NOT use cotton swabs (Q-Tips).
Page 13
LWCC-M
World Precision Instruments 9
Additional Information on Contamination
The following notes regarding contamination have been collated at WPI during the development and testing of the LWCC-M:
Most syringe lters contain some contaminants that absorb UV. More than likely, this is caused by the (plastic) mold release agent used in manufacturing. The rst few milliliters of solution coming from a new lter will have some absorption in the UV range.
The rst two loads of solution from most new plastic syringes often have some
contamination that absorbs UV. In addition, when plastic syringes are used to transfer organic solvents, the rubbery gasket material in the plunger absorbs some of the chemical. If the syringe is later used to transfer aqueous solution, the chemical will slowly leach out. Since most organic solvents have an absorbance in the UV range, the liquid initially released from the syringe might be found to have a dierent spectrum than last of the liquid in the syringe, the latter having been contaminated
by the chemical in the plunger. Some commonly used organic solvents which have “relatively low” UV absorption and are suitable for UV detection in conventional spectrometers might not be problem-free when used in LWCC-M.
A beaker of freshly ltered water sitting overnight in open air will probably have an increased absorbance in the UV range because of dust from the air or growth of microorganisms.
Some plastic tubing will release a substance that absorbs UV. In WPI’s lab, silicone tubing used in a peristaltic pump constantly released a contaminant even after a week of washing.
A bubble in the LWCC-M will result in unstable readings. Additional liquid circulating through the device will usually push the bubble out. If the bubble doesn’t clear easily, try introducing a larger bubble followed by liquid. This will usually pick up a small bubble that may cling and cause problems.
Avoid introducing particulate into the LWCC-M. If trapped in the LWCC-M, particles can scatter light and may cause unstable spectrometer readings. The LWCC-M
contains two potential “bottlenecks” at the ber-capillary interface. Due to the diverse applications of LWCC-M, no in-line lter can be installed which will t all users’ needs. It is imperative, therefore, that a proper in-line lter be added to the LWCC-M if
the solution contains large particles. When the LWCC-M is directly connected a
chromatography column, a lter might not be necessary.
NOTE: If you have a problem/issue with that falls outside the definitions of this troubleshooting section, contact the WPI Technical Support team at 941.371.1003 or
technicalsupport@wpiinc.com.
Page 14
10 World Precision Instruments
SPECIFICATIONS
This instrument conforms to the following specications:
M-10 M-50 M-100
Optical Pathlength 10 mm 50 mm 100 mm Internal Volume 2.4 µL 12 µL 24 µL Wavelength Range 200 – 1000 nm Fiber Connection (SMA) 500 µm Transmission @ 254 nm *
> 25% > 20% >15% Transmission @500 nm * > 40% > 35% >30% Maximum Pressure 1000 PSI
Refractive Index @ 280 nm** < 15 mAU <20 mAU <30 mAU Wetted Material PEEK, Fused Silica, PTFE
* Reference: Two 600 µm bers, butt-coupled ** Measured using ASTM E 685 - 93, includes intrinsic solvent absorbance
WPI U.S. Patents: 5,444,807; 5,570,447; 5,604,587; 6,603,556; 6,385,380.
Fig. 4—This graph show the percentage of light transmission at various wavelengths. The higher the light transmission percentage, the greater the resolution.
Related Patents
Micro Chemical Analysis Employing Flow Through Detectors, 1995, U.S. Patent No. 5,444,807.
Aqueous Fluid Core Waveguide, 1996, U.S. Patent No. 5,507,447.
Long Capillary Waveguide Raman Cell, 1997, U.S. Patent No. 5,604,587.
Chemical Sensing Techniques Employing Liquid-Core Optical Fibers, U.S. Patent No. 6,016,372
Page 15
LWCC-M
World Precision Instruments 11
WARRANTY
WPI (World Precision Instruments, Inc.) warrants to the original purchaser that this equipment, including its components and parts, shall be free from defects in material and workmanship for a period of one year* from the date of receipt. WPI’s obligation under this warranty shall be limited to repair or replacement, at WPI’s option, of the equipment or defective components or parts upon receipt thereof f.o.b. WPI, Sarasota,
Florida U.S.A. Return of a repaired instrument shall be f.o.b. Sarasota.
The above warranty is contingent upon normal usage and does not cover products which have been
modied without WPI’s approval or which have been subjected to unusual physical or electrical stress or on which the original identication marks have been removed or altered. The above warranty will not apply if adjustment, repair or parts replacement is required because of accident, neglect, misuse, failure of electric power, air conditioning, humidity control, or causes other than normal and ordinary usage.
To the extent that any of its equipment is furnished by a manufacturer other than WPI, the foregoing
warranty shall be applicable only to the extent of the warranty furnished by such other manufacturer. This
warranty will not apply to appearance terms, such as knobs, handles, dials or the like.
WPI makes no warranty of any kind, express or implied or statutory, including without limitation any warranties of merchantability and/or tness for a particular purpose. WPI shall not be liable for any damages, whether direct, indirect, special or consequential arising from a failure of this product to operate
in the manner desired by the user. WPI shall not be liable for any damage to data or property that may be caused directly or indirectly by use of this product.
Claims and Returns
Inspect all shipments upon receipt. Missing cartons or obvious damage to cartons should be noted on the delivery receipt before signing. Concealed loss or damage should be reported at once to the carrier and an inspection requested. All claims for shortage or damage must be made within ten (10) days after receipt of shipment. Claims for lost shipments must be made within thirty (30) days of receipt of invoice
or other notication of shipment. Please save damaged or pilfered cartons until claim is settled. In some instances, photographic documentation may be required. Some items are time-sensitive; WPI assumes no extended warranty or any liability for use beyond the date specied on the container
Do not return any goods to us without obtaining prior approval and instructions from our Returns Department. Goods returned (unauthorized) by collect freight may be refused. Goods accepted for restocking will be exchanged or credited to your WPI account. Goods returned which were ordered by customers in error are subject to a 25% restocking charge. Equipment which was built as a special order cannot be returned.
Repairs
Contact our Customer Service Department for assistance in the repair of apparatus. Do not return goods until instructions have been received. Returned items must be securely packed to prevent further damage
in transit. The Customer is responsible for paying shipping expenses, including adequate insurance on all items returned for repairs. Identication of the item(s) by model number, name, as well as complete description of the diculties experienced should be written on the repair purchase order and on a tag
attached to the item.
* Electrodes, batteries and other consumable parts are warranted for 30 days only from the date on which the customer receives these items.
Page 16
12 World Precision Instruments
Page 17
LWCC-M
World Precision Instruments 13
INDEX
15807 .............................7
58006 ..........................2, 6
58450 .............................7
89372 ....................... 2, 5, 7
500320 .............................7
501609 ....................... 2, 7, 8
800490 ..........................7, 8
A
accessories .........................7
B
bubble .............................9
bubbles ............................2
C
cables .............................3
chromatography ..................2, 9
contaminants .......................9
contamination ......................8
continuous ow .....................5
Contrad-NF ......................6, 8
cotton swabs ........................8
D
D4H .............................7
dust .............................9
F
FEP .............................4
ber optic cables ....................3
lter ..........................4, 9
FO-500AS-SMA ......................7
FO-500-SMA1M .....................7
FO-6000 ............................7
H
hydrouoride .......................4
I
injection system .....................5
L
LEDspec ............................2
light source ................... 3, 5, 7
liquid ports .........................3
M
microorganisms .....................9
MINISTAR ...........................5
mold release agent ..................9
N
nitric acid ...........................4
O
organic solvents .....................9
P
particle .............................8
particulate ..........................9
parts list ............................2
PEEK .............................4
Q
Q-Tips .............................8
R
returns .............................2
S
specications ..................... 10
spectrophotometer ..............1, 3
sulfuric acid .........................4
T
TIDAS-BASIC ........................7
Tidas E .............................7
TIDAS E .............................2
troubleshoot ........................7
U
unpacking ..........................2
Page 18
14 World Precision Instruments
Page 19
LWCC-M
World Precision Instruments 15
Page 20
USA
International Trade Center, 175 Sarasota Center Blvd., Sarasota FL 34240-9258
Tel: 941-371-1003 • Fax: 941-377-5428 • E-mail: sales@wpiinc.com
UK
1 Hunting Gate, Hitchin, Hertfordshire SG4 0TJ
Tel: 44 (0)1462 424700 • Fax: 44 (0)1462 424701 • E-mail: wpiuk@wpi-europe.com
Germany
Zossener Str. 55, 10961 Berlin
Tel: 030-6188845 • Fax: 030-6188670 • E-mail: wpide@wpi-europe.com
China & Hong Kong
WPI Shanghai Trading Co., Ltd.
Rm 25e, No8 Dongfang Rd., Pudong District, Shanghai, 200120 PR China
Tel: +86 21 6888 5517 • E-mail:chinasales@china.wpiinc.com
Brazil
Av. Conselheiro Nébias, 756 sala 2611, Santos-CEP: 11045-002, São Paulo Brazil •
Tel: (013) 406-29703 • E-mail: info@brazil.wpiinc.com
Internet
www.wpiinc.com • www.wpi-europe.com • www.wpiinc.cn
Loading...