WOSON TANCO Operation Manual

WOSON Medical System
Technical Publications
ASS0003 REV- B
TANCO Steam Sterilizer Operation Manual
Copyright © 2011 By Woson Medical Instrument Co., Ltd.
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 1 of 44
Regulatory Requirement
This product complies with regulatory requirements of the following European Directive 93/42/EEC concerning medical devices.
Ningbo Jiangbei Woson Medical Instrument Co., Ltd
No.25, Lane 300, Jinshan Road, Jiangbei District, Ningbo 315032, China
Tel: 86-574-83022668 Fax:86-574-87639376 Website: www.woson.com.cn E-mail: overseas@woson.com.cn
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 2 of 44
Revision History
REV
ISSUE DATE
REASON FOR CHANGE
Rev-A
2010.01.25
First Issue
Rev-B
2011.07.31
Version Change
Please verify that you are using the latest revision of this document. Information pertaining to this document is maintained on manufactory. If you need to know the latest revision, please contact your distributor, sales representative, or our service dept.
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 3 of 44
Regulatory Requirements
Conformance Standards
The content of this instruction is fit for sterilizers. Above sterilizer accord with the requirements of European Class B: 93/42/EEC 97/23/EC EN 61010-1 EN 61010-2-040 EN 13060 EN 61326-1
European Authorized Representative
DTF Technology s.r.l.
Address: via De Sanctis 32, 20141 milano Tel: 39 02 84893641 Fax: 39 02 84718594
This product complies with the regulatory requirement of the following:
• Council Directive 93/42/EEC concerning medical devices: The CE label affixed to the product testifies compliance to the Directive. The location of the CE marking is shown in this manual.
Certifications
Manufactory is EN ISO 9001 and EN ISO 13485 certified.
Original Documentation
The original document was written in English.
Declaration of Conformity
Council Directive 93/42/EEC concerning medical devices:
The CE label affixed to the product testifies compliance to the Directive.
The location of the CE marking is shown in this manual. In this manual there are present the CE
certification and the Conformity. Check appendixes.
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 4 of 44
Table of Content
Regulatory Requirement .................................................................................................................... 1
Revision History ............................................................................................................................... 2
Regulatory Requirements .................................................................................................................. 3
Table of Content ................................................................................................................................ 4
Chapter 1 Introduction ...................................................................................................................... 6
1.1 Attention .............................................................................................................................. 6
1.2 Usage Indications ................................................................................................................ 6
1.3 Contraindication .................................................................................................................. 6
Chapter 2 Safety ................................................................................................................................ 7
2.1 Explanation Symbol ............................................................................................................ 7
2.2 General Safety Recommendations ...................................................................................... 8
2.3 Safty Parts ........................................................................................................................... 9
2.3 Operation Risk .................................................................................................................. 10
2.4 Protection Device .............................................................................................................. 10
Chapter 3 Receiving And Installation ............................................................................................. 11
3.1 Check the Package ............................................................................................................ 11
3.2 Unpacking the Accessories ............................................................................................... 11
3.3 Optional Accessories ......................................................................................................... 12
3.4 Installation Enviroment ..................................................................................................... 12
3.5 Set ..................................................................................................................................... 12
3.6 Power Connection ............................................................................................................. 13
Chapter 4 Description and Specification ........................................................................................ 14
4.1 Front View of the Sterilizer ............................................................................................... 14
4.2 Rear View of Sterilizer ...................................................................................................... 14
4.3 Opened Vew of Sterilizer .................................................................................................. 15
4.4 Outside Size ...................................................................................................................... 16
4.5 Loading Size ..................................................................................................................... 16
4.6 Specification ...................................................................................................................... 16
4.7 Sterilization Cycle ............................................................................................................. 18
Chapter 5 Panel and Functions ...................................................................................................... 19
5.1 Controlling Panel .............................................................................................................. 19
5.2 Menu ................................................................................................................................. 20
5.3 Window of Sterilization Process ....................................................................................... 22
Chapter 6 Operation Process .......................................................................................................... 24
6.1 Power On........................................................................................................................... 24
6.2 Distilled Water Adding ...................................................................................................... 24
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 5 of 44
6.3 Alarming If Used Water Reservoir Is Full ......................................................................... 25
6.4 Selecting Sterilizing Program ........................................................................................... 25
6.5 Loading Articles ................................................................................................................ 25
6.6 Closing the Door ............................................................................................................... 26
6.7 Starting A Sterilization Program ....................................................................................... 28
6.8 End Of The Sterilization Working Cycle. ......................................................................... 29
6.9 Power Off .......................................................................................................................... 29
6.10 Abnormal Exiting ............................................................................................................ 30
6.11 Sudden Power Off ........................................................................................................... 30
Chapter 7 Essential Information ..................................................................................................... 31
7.1 Please Ensure The Following.... ........................................................................................ 31
7.2 And Please Do Not.... ........................................................................................................ 31
Chapter 8 Maintenance ................................................................................................................... 32
8.1 Maintenance Schedule Chart ............................................................................................. 32
8.2 Daily Maintenance ............................................................................................................ 32
8.3 Weekly Maintenance ......................................................................................................... 33
8.4 Monthly Maintenance ....................................................................................................... 34
8.5 Other Maintenance ............................................................................................................ 35
Chapter 9 Servicing By the Approved Technician ........................................................................... 36
Chapter 10 Transportation and Storage ......................................................................................... 37
10.1 Preparation before Transportation and Storage ............................................................... 37
10.2 Draining .......................................................................................................................... 37
10.3 Conditions For Transportation And Storage .................................................................... 37
10.4 Package ........................................................................................................................... 37
Appendix 1 Instrument Which Need To Sterilize ............................................................................. 38
Appendix 2 Error Code List ............................................................................................................ 39
Appendix 3 Electric And Hydraulic Drawings ................................................................................ 40
Hydraulic Drawing .................................................................................................................. 40
Electric Drawing ..................................................................................................................... 41
Appendix 4 The Standards Of Testing ............................................................................................. 42
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 6 of 44
Chapter 1 Introduction
1.1 Attention
This manual contains necessary and sufficient information to operate the system safely,
like optimal use, a safe and reliable operation, regular and correct servicing
requirements.
Read and understand all instructions in this manual before attempting to use the product. Keep this manual with the equipment at all times. Periodically review the procedures for
operation and safety precautions.
1.2 Usage Indications
Application to all wrapped or non-wrapped, solid, hollow load products type A and porous products or related articles.
This sterilizer can be used for dental clinic, laboratory, surgical room, emergency room, ophthalmology, gynecology and steam, cosmetic hospital and so on, by doctors and professionals.
1.3 Contraindication
There is no any contraindication of this equipment.
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 7 of 44
Chapter 2 Safety
2.1 Explanation Symbol
Device Icon Description
“ATTENTION” – Consult accompanying documents” is intended to alert the user to refer to the operator manual or other instructions when complete information cannot be provided on the label.
Pay attention to the high temperatures in the chamber, and to the sterilizer exterior when exhausting system is running.
“Protective Earth” indicates the protective earth (grounding) terminal.
“CAUTION” - Dangerous voltage”
Be aware of electric shock hazards.
Label Description
Symbol for "SERIAL NUMBER"
Symbol for "MANUFACTURER"
Symbol for "CATALOGUE NUMBER"
Symbol for "AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY
Symbol for "DATE OF MANUFACTURE"
Symbol for "CAUTION"
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 8 of 44
Manual Prompt
Note
Indicates that concerning information which easier or helpful in operation
Caution
Indicates that a potential hazard may exist which through inappropriate conditions or actions will or can cause:
• Minor injury
• Property damage.
• Damage to machine
Warning
Indicates that a specific hazard is known to exist which through inappropriate conditions or actions may cause:
• Severe personal injury
• Substantial property damage
Substantial damage to machine
NOTE: Please apply the precautions or recommendations indicated in the Operation Manual.
2.2 General Safety Recommendations
The user is responsible for proper operation and maintenance of the sterilizer in
accordance with the instructions listed in this manual.
The sterilizer could not be used for liquid. The sterilizer has not been designed to operate in the presence of gas or explosive
vapors.
The trays and the load will still be hot at the end of each cycle. Use the tray holder to
remove each tray from the chamber.
Do not open the door of the chamber during the sterilization programs. Do not put you hands or face on the cover of the water tank when the sterilizer is running. Do not remove the instruction plate or any label from the sterilizer. Do not pour water or any other liquid over the sterilizer. Do not fill the caustic liquid into the water tank. Do not fill the caustic matter in the chamber. Use only high quality distilled water. Unplug the mains lead before inspecting or servicing the machine. Only an approved technician using original spare parts can carry out repairs and
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 9 of 44
maintenance.
In case of transport, drain both water tanks completely, allow the sterilization chamber to
cool down and preferably use the original packaging.
The articles under sterilization should be picked-up by special handle of getting plate
when the temperature shows over 40℃.
Picking-up the sterilizing plate should use the special tools provided. During the transport, the 4 bottoms of sterilizer should be carried by two people in case of
turning over.
Notice! This product can’t be put on the place that is not easy to cut off power supply. ❖ Prohibit covering the cover of water tank during usage.
2.3 Safety Parts
Temperature Protection
Part Name
Function
Temperature Protector (Steam Generator)
Cut off current when the steam generator temperature is too high.
Temperature Protector (Heating Ring)
Cut off current when the heating ring temperature is too high.
Electronic Protection
Part Name
Function
Double Fuse
It will be broken circuit if the connected power is too high or unstable.
Electronic Filter
It will filter the electromagnetic interference during working
Mechanical Protection
Part Name
Function
Jiggle Switch
To ensure the door is closed completely for avoiding the risk of safety
Tray tong
when removing articles from the chamber
Control Part
Part Name
Function
Temperature Sensor (Inside Chamber)
To measure temperature inside the chamber Temperature Sensor
To measure temperature of the heating ring
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 10 of 44
(Heating Ring)
Temperature Sensor (Steam Generator)
To measure temperature of the steam generator Pressure Sensor
To measure pressure of the chamber
PCB Control
Control system for controlling all the process of sterilization
CAUTION
MANUFACTURER IS NOT HELD RESPONSIBLE FOR ANY ARBITRARY DISASSEMBLY, AMENDMENT WITH THE UNIT, BY UNAUTHORIZED PERSON OR UNPROFESSIONAL TECHNICIAN
2.4 Operation Risk
Please take attention on avoid operation risk during operation.
Scald risk
Every time open the door after finish the cycle, please keep an appropriate distance,
because the chamber still have rudimental steam with high temperature, avoid to be scalded.
Every time open the door after finish cycle, please not touch the main door and chamber,
because of high temperature, and avoid to be scalded.
Pollution risk
Please clean the chamber after every time use to avoid rudimental contamination left inside chamber.
2.5 Protection Device
Device Name
Function
Plastic or fabric glove
Useful during load and remove articles, avoid scald.
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 11 of 44
Chapter 3 Receiving and Installation
3.1 Check the Package
Please check package carefully when you receive the product, any damage of package maybe damage the product inside.
Front
Flank
Item
8L Model
12L Model
Packing Size
652×549×486
652×549×486
Packing Weight
42kg
48kg
3.2 Unpacking the Accessories
Open the package and take the product out, remove the plastic bag then open the door to take all the accessories and check as followed:
No.
Part Name
Qty
1 2 6 4 3 5
1
Tray frame
1 piece
2
Trays
3 sets
3
Tray removal tong
1 piece
4
Draining tube
1 piece
5
Power supply cable
1 piece
6
Door gasket
1 piece
7
Operation manual
1 piece
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 12 of 44
3.3 Optional Accessories
Name
Model
QTY
Picture
Mini Printer
PINTER10
1
3.4 Installation Environment
The sterilizer should be set in a place which at least has 10cm distance with each side (20cm of the top) as followed:
The sterilizer should be set in a place with good ventilation. The temperature of enviroment:5-40℃. The humidity of environment: ≤85% Atmosphere pressure:860Hpa1060Hpa An earth connection is essential
CAUTION
DO NOT PUT ANY STUFF WHICH EASILY MELTS NEAR THE STERILIZER.
3.5 Set
The sterilizer should be set on level table or place; the front-end should be a little bit
higher then back-end.
The sterilizer cooling and vent area should not be jammed or blocked. Do not put any stuff on the top of the sterilizer. Do not put any stuff in front of the door, to avoid accident when open the door. Do not put any corrosive stuff near the sterilizer to avoid accident or risk.
Steam Sterilizer Operation Manual
TANCO Model REV-B Page 13 of 44
3.6 Power Connection
The sterilizer should be connected with a stable and separated power source Power socket is at the back of the sterilizer Please confirm the connection power is complied with specification of nameplate at the
back of sterilizer
CAUTION
DO NOT BEND THE POWER WIRE TO AVOID DAMAGE OF POWER WIRE DO NOT PUT ANY HEAVY STUFF ON THE POWER WIRE TO AVOID DAMGE OF POWER WIRE DO NOT USE OTHER POWER WIRE TO AVOID DAMAGE OF STERILIZER DO NOT TRY TO ADD THE POWER WIRE TO AVOID ACCIDENT AND RISK
Loading...