4. CREATE A NEW SENSOR MANUALLY
• Use the button to switch to the Main menu.
• Select “Sensors” and press the
button to edit the sensor
list. A window appears; select New Sensor and press
.
• Select Manual and press
.
• In the Sensor Settings you have to choose the Sensor type
of your new sensor (NomaSense O
2
P300: PSt3 ; NomaSense
O2 P6000: PSt3 or PSt6). Then move to the upper right of
the screen to the Next button and press
.
• Select Humid and press
. Then change the calibration
values to the values stated on the sensor´s Final Inspection
Protocol.
• When you have adjusted all the calibration values move to
the upper right of the screen to the Next button and press
.
• A keyboard screen opens. Use the navigation arrows to type
in the new Sensor Name.
• When you have finished typing go to the Done button and
press
.
• The new sensor shows in the sensor list. Press
once more
to select the new sensor and the display will switch to the
measurement screen automatically.
5. TUNE THE MEASUREMENT SETTINGS
• Select “Meas. Settings” in the Main menu and change
Measurement Settings:
• Change settings for Temperature and Pressure
compensation. Use the arrow buttons to navigate between
input fields. Press
to get into editing mode and change
the respective setting or value (one digit at a time) with the
and buttons. To cancel and leave the editing mode
press . When you have adjusted an input field press
the button again to save the changes.
• Select Single Scan for a single measurement or Interval
for continuous measurement at a certain time interval
Fig. 8 Main menu screen –Sensors selected
Fig. 9 Calibration Data screen
Fig. 10 Measurement Settings screen