WIKA UT-10 Operating Instructions Manual

Page 1
GBF
Operating instructions Betriebsanleitung Mode d’emploi
WIKA Alexander Wiegand GmbH & Co. KG
Alexander-Wiegand-Straße 30 63911 Klingenberg/ Germany Tel. (+49) 93 72/132-295 Fax (+49) 93 72/132-706 E-Mail support-tronic@wika.de www.wika.de
UT-10 / UT-11
UT-10
UT-11
GB
GB
Universal Transmitter for various applications
DUniversaltransmitter für vielfältige Einsatzgebiete
Transmetteur unviersel pour utilisations multiples
Page 2
S-Nr. 218 4125.04 Ausgabedat um 16.02.04
Universal-Drucktransmitter Universal Pressure Transmitter Transmetteur de Pression Universel
UniTrans
®
Betriebsanleitung / Manual / Manuel
Page 3
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Contents
Subject t o change due to te chnical modific ations. © Copyright WIKA A lexander W iegand GmbH & Co . KG / BRD
WIKA Alexander Wie
g
and GmbH & Co. KG · Al exander-W iegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
42
Contents
1 General Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
2 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
2.1 Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
2.1.1 Pressure Transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
2.1.2 Processin
g
Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
2.1.3 Displa
y
Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
2.2 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2.2.1 Functions of Transmitters without Displa
y
s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2.2.2 Functions of Transmitters with Displa
y
s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2.3 Installation Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
3 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
3.1 Input-values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
3.2 Output-values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
3.3 Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
3.4 Auxilliar
y
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.5 Ambient Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.6 Process Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.7 Identification Plates
(
example) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.1 Pressure Transmitter Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.2 Displa
y
Unit Upgrades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.3 Housin
g
Reconfiguration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.4 Electrical Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
4.5 Pressure Compensation when usin
g
a Relative Pressure Sensor . . . . . . . . . . . . 57
5 Operation of Transmitters without Display. . . . . . . . . . . . . . . . . . . . 58
5.1 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.2 Ke
y
Functions (only available for transmitters without display) . . . . . . . . . . . . . . 58
5.3 Calibration with Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.1 Zero Point Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.2 Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.4 Calibration without Pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.4.1 Zero Point Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.4.2 Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.4.3 Mountin
g
correction of the sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.5 Inte
g
ration Time (Damping) Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.6 Reset to Default . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6 Operation of Transmitters with Display. . . . . . . . . . . . . . . . . . . . . . . 63
6.1 The Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
6.2 Ke
y
Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.3 The Pro
g
ramming Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Page 4
C
Universal Pressure Transmitter UniTrans
Contents
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 6391 1 Klingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
43
6.4 Default Data (factory settings). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
6.5 Main Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
6.5.1 Main Menu: Displa
y
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
6.5.2 Main Menu: Calibration of zero and span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
6.5.3 Main Menu: Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
6.5.4 Main Menu: Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
6.5.5 Main Menu: Lan
guag
e. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
6.5.6 Main Menu: Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
7 Diagnostics and Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
8 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9 Appendix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.1 Dimension Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.2 Model Ke
y
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
9.3 Warrant
y
Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
9.4 Glossar
y
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
9.5 Units of Pressure Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Page 5
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Contents
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. K G / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
44
1 General Safety Instructions
All pressure connections may only be opened after the system is without pressure!
Observe the national regulations about safety and accident prevention , as well as the safety instructions in this operating manual when operating the pressure transmitter.
Any operation not described in the following instructions must not be car­ried out.
If a failure cannot be repaired, the transmitter must be switched off. The operator must then make sure that it is only switched on again after the failure has been repaired.
Prior to installing, starting and operating a pressure measuring instrument, the user must ensure that the appropriate instrument has been selected with regard to scale range and performance and that the wetted parts material is suitable for the specific measuring conditions of the respective application.
Serious injuries and/or damage can occur should the relevant regulations not be observed.
Dangerous pressure media such as oxygen, acetylene, flammable gases or liquids and toxic gases or liquids as well as instruments for refrigeration plants or compressors etc. require attention above the standard regula­tions. Here the specific safety codes or regulations must be considered.
Ramaining pressure medium contained in the pressure element may be hazardous or toxic. This should be considered when handling and storing the removed pressure measuring instrument.
Repairs should only be carried out by the manufacturer. All other repairs or modifications of the transmitter are unauthorized.
Other important safety guidelines can be found in the different sections of this instruction manual.
Warnin
g
Warnin
g
Warnin
g
Attention
Warnin
g
Warnin
g
Warnin
g
Warnin
g
Attention
Page 6
C
Universal Pressure Transmitter UniTrans
Product Description
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
45
2 Product Description
The UniTrans pressure transmitter can be used in level control applications as well as for pressure measurement applications in process industr
y
. A variety of process con-
nections, measurement ran
g
es, main boards and display options result in a product
for a wide ran
g
e of applications.
2.1 Construction
The UniTrans consists of a pressure sensor, a control interface unit and a housin
g
cover with optional displa
y
. Due to this modular design, different transmitter versions
can be mounted (see Model Ke
y
on page 79).
2.1.1 Pressure Transducer
The pressure transducer has a piezo-resistive or thinfilm measurement cell depend­in
g
on the pressure range. The sensors are temperature compensated, and have a
hermeticall
y
welded membrane which is "helium" leak-tested. The pressure trans-
ducers do not have internal sealin
g
elements.
Pressure transducers further distin
g
uish themselves from one another based on their
pressure ran
g
es and the different materials of wetted parts. Different process
connections can be selected to serve a wide ran
g
e of applications.
Never exceed the overpressure limit of the respective pressure transducer.
Direct Mounted
Processing Unit Housing
Attention
Page 7
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Product Description
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
46
2.1.2 Processing Unit
The processin
g
unit, which is integrated in the housing contains the terminal compart-
ment and the ke
y
pad used for programming the transmitter. The four keys must be
activated (unlocked) before use. Durin
g
normal operation the keypad is locked to pro-
tect data and functions previousl
y
entered. The keypad is automatically locked when
no ke
y
is hit for 10 minutes. The processing unit converts the digitalized signal from
the measurin
g
unit into a standard 4...20 mA current signal.
2.1.3 Display Unit
The indicator has four di
g
its (7-segment display) + symbols. Below it, line 1 (16-seg-
ment displa
y
) is used to display error codes and the signals unit of measure. The unit
of measure can be selected b
y
the operator. Measurements over 9999 can not be cor-
rectl
y
displayed. Please note this when choosing the unit (e.g. 9999 Pascal = 0,09999
bar). Additional information is displa
y
ed in lines 2 and 3 (16-segment display). The
operator can enter commands in the pro
g
ramming mode on the display unit by means
of menu
g
uided, clear-text prompts.
Processing Unit
CALIBRATION RANGE
Page 8
C
Universal Pressure Transmitter UniTrans
Product Description
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
47
Transmitters with displays offer a larger number of programming and processing op­tions. These options include alarm status, dampin
g
, signal inversion, tank lineariza-
tion and dia
g
nostic messages.
2.2 Function
The mode of operation for si
g
nal conversion works in the same way for all versions.
The pressure transducer converts the existin
g
pressure into an electrical signal. Mi-
croelectronics further process the input si
g
nal and produce a proportional 4-20 mA
standard si
g
nal.
The displa
y
-version allows programming (parameterization) and the display of ex-
tended functions such as inversion, dampin
g
, alarm status and linearization.
2.2.1 Functions of Transmitters without Displays
Calibration of zero and span under pressure (see 5.3)
Calibration of zero and span without pressure (dr
y
adjustment) (see 5.4)
Settin
g
the dampening / integrating the output signal 0-40 s (see 5.5)
Reset to manufacturers default values (see 5.6)
Mountin
g
correction of the sensor (beginning with software version 1.05)
(see section 5.4.3)
2.2.2 Functions of Transmitters with Displays
Settable units of measurement (mbar, bar, psi, mA, %, m, mm WS) (see 6.5.1)
Temperature and Min/Max values shown in displa
y
(see 6.5.1)
Nominal pressure ran
g
e of the sensor shown in display (see 6.5.1)
Zero and span calibration (with/without pressure) (see 6.5.2)
Settin
g
of damping / integration of output signal 0-40 s (see 6.5.3)
Inversion of the output current si
g
nal (see 6.5.3)
Settin
g
the output current value in case of alarm (3.6 mA or 21 mA) (see 6.5.3)
Settin
g
the limits of the output signal (see 6.5.3)
Offset of the output si
g
nal (see 6.5.3)
Mountin
g
correction of the sensor
Measurin
g
circuit test function (see 6.5.4)
Reset functions (see 6.5.4)
Password activation (see 6.5.4)
Selectin
g
the language of the display (see 6.5.5)
Enterin
g
of a table function for the linearization of the output signal (see 6.5.6)
Enterin
g
the medium density (see 6.5.6)
Display units can be easily upgraded (see chapter 4.2).
Page 9
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Product Description
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
48
2.3 Installation Examples
The UniTrans is primaril
y
used to monitor the pressure in pipes, technical equipment
and tanks. Dependin
g
on the pressure range pressures between 20 mbar up to 1000
bar can be measured. The pressure is measured usin
g
absolute (against a vacuum)
or relative (a
g
ainst external or air pressure) measurement depending on the type of
sensor selected.
The UniTrans is also used for h
y
drostatic pressure measurement within liquid filled
pipes and containers.
Process Pressure Measurement: Process Pressure Measurement:
Used to measure pressure of liquids Used to measure container pressure.
or gases in pipelines.
Process Pressure Measurement: Process Pressure Measurement:
Installed behind feed pumps for Installed in front of and behind the filter.
process control or monitoring of pump Uses the pressure differential for moni-
functions. toring the function or accumulation of dirt
in the filter. Both output signals are pro-
cessed by a PLC or signal converter.
Page 10
C
Universal Pressure Transmitter UniTrans
Product Description
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
49
Level Control: Level Control:
Externally mounted Combined pressure and head pressure
(with front flat diaphragm) are measured by two externally mounted
pressure transducers. The two signals
are analyzed and the differential is cal-
culated by a PLC or suitable signal con-
verter.
Page 11
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Technical Data
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
50
3 Technical Data
3.1 Input-values
3.2 Output-values
Pressure Ranges (Absolute pressure upon request)
/ overpressure limit / burst pressure
0 ... 0.4 bar 2 2.4 0 ... 1,6 bar 10 12 0 ... 6 bar 35 42 0 ... 16 bar 80 96 0 ... 40 bar 80 400 0 ... 100 bar 200 800 0 ... 250 bar 500 1200 0 ... 600 bar 1200 2400
1)
0 ... 1.000 bar 1500 3000 0 ... 1,600 bar 2000 4000 0 ... 2,500 bar 3000 5000 0 ... 4,000 bar 4400 7000
-1 ... +0 bar* 2 2.4
-1 ... +0,6 bar* 10 12
-1 ... +3 bar* 35 42
-1 ... +5 bar* 35 42
-1 ... +15 bar* 80 96 *onl
y
relative pressure
Do not exceed the nominal pressure!
1)For flush diaphragm version: The value specified in the table applies only when sealing is realised with the sealing ring under­neath the hex. Otherwise max. 1500 bar applies.
Output signal 4 ... 20 mA
Accurac
y
[% of span]
(linearit
y
, hysteresis, repeatability)
<
0.10 at ranges < 1000 bar
<
0.30 at ranges > 1000 bar
Turn down behavior: (1/k) up to 1 : 5 1 : 5 to 1 : 20
no changes of deviation the accurac
y
must be multiplied by the factor (turn down / 5) example for TD = 1:15, (k = 15) accurac
y
= 0.10 * (15/5) = 0.3
Overall deviation (at +10 °C ... +40 °C)
<
0.15 % (limit point calibration)
< 0.6 % for pressure ran
g
es of > 1000
bar
Load RA < (UB–12 V)/0.023 A
with R
A
in Ohm and UB in Volt
Fault si
g
nal 3.6 mA or 21 mA, programmable
Page 12
C
Universal Pressure Transmitter UniTrans
Technical Data
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
51
3.3 Construction
Integration time 0s, 1s, 5s, 20s, 40s, programmable
Adjustment of the span Up to Turn down 1 : 20
Integrated lightning protection optional
Zero point adjustment 2.5 ... 99 %
Process connections Model UT-10
Model UT-11
Model UT-11 EHEDG version
G 1/2 B per DIN 16288 (1/2 NPT) M 16 x 1,5 with sealin
g
cone
>
1600 bar
3/8-24 UNF LH male
>
1600 bar
1/4"-28 UNF LH M 250-C
>
1600 bar
G 1B flush diaphra
g
m with o-ring
(Ran
g
es:
0 ... 0.4 up to 0 ... 1.6 bar)
G 1/2 B flush diaphra
g
m with o-ring
(Ran
g
es:
0 ... 6 bis 0 ... 600 bar)
G 1 1/2 flush diaphra
g
m with o-ring
(Ran
g
es:
0 ... 0.4 bis 0 ... 16 bar)
G 1 flush diaphra
g
m with o-ring
(Ran
g
es:
0...0.4 bis 0...16 bar)
G 1 flush diaphra
g
m with o-ring
and coolin
g
element (Ran-
g
es:0...0.4 bis 0...16 bar)
Materials Housin
g
Wetted parts (UT-10) (UT-11)
Internal transmission fluid
hi
ghly
resistive, fiberglass-enforced
plastic (PBT); optionall
y
aluminium CrNi-steel 1.4571 and 2.4711 CrNi-steel 1.4571, o-rin
g
: NBR {Viton or
EPDM}; {Hastello
y
C4}
Standard {Halocarbon oil for ox
yg
en-
applications}; {FDA-approved}
Page 13
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Technical Data
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
52
3.4 Auxilliary Power
3.5 Ambient Conditions
3.6 Process Conditions
Electrical connection per EN 60 529/ IEC529
M 20 x 1.5 cable gland with internal ter­minal block (see 4.4) M12x1 plu
g
, 4-pin (pin allocation: 1+ 3-)
[3/4" NPT female conduit (onl
y
with alu-
minium case)]
Electric protection Reverse polarit
y
, overload and short cir-
cuit protection
Power suppl
y
12 ... 36 V DC
Ambient temperature – 40 °C ... + 85 °C
(– 20 °C ... 70 °C with displa
y
)
Stora
g
e temperature – 40 °C ... + 85 °C
(– 35 °C ... 80 °C with displa
y
)
Climate class D per DIN IEC 654-1
In
g
ress protection per EN 60 529 IP 65 with plastic case
IP 67 with aluminum case
a
-conformit
y
Interference emission and immunity see EN 61326 (also fulfills NAMUR NE 21) 97/23 EG Pressure Equipment Direc­tive (Module H)
Medium temperature G 1 1/2
G 1 EHEDG with coolin
g
element
– 30 °C ... + 105 °C (up to 30 min. 140 °C at an ambient temperature of < 50 °C
-30 °C ... + 150 °C
°F = (°C * 1.8) + 32
°F = (°C * 1.8) + 32
Page 14
C
Universal Pressure Transmitter UniTrans
Technical Data
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
53
3.7 Identification Plates (example)
Page 15
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Installation
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
54
4 Installation
The device should be installed/operated in accordance with the regulations of ElexV, the Device Safet
y
Regulation, this operating manual and generally recognized indus-
tr
y
standards.
4.1 Pressure Transmitter Installation
Installation Using a Weld-on Adapter:
Insert a filler piece (a pressure transmitter dumm
y
) into the weld-on adapter.
Weld the adapter into the container/pipe wall (section-weld process).
Remove the filler piece.
Install the pressure transmitter in the weld-on adapter.
4.2 Display Unit Upgrades
The displa
y
unit can be easily upgraded at any time.
Remove the housin
g
cover and the supporting string.
Attach the displa
y
units supporting string to the same place.
Plu
g
the display units connector into the appropriate jack.
The displa
y
unit can be mounted at 90° angles.
Fasten the displa
y
unit with screws.
The pressure transmitters diaphram should not come into contact with hard or sharp objects.
When installing the display unit, make sure that the connection cable and the supporting string are not kinked or pinched.
Attention
Attention
Page 16
C
Universal Pressure Transmitter UniTrans
Installation
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
55
All functions are pro
g
rammable once the pressure transmitter has been upgraded
with a displa
y
unit. The adjusted parameters are stored after the display unit is re-
moved.
The displa
y
unit can be rotated in 300°, so that it can be read under various installation
conditions. The housin
g
cover with built-in display can be fastened to the housing at
all four side positions.
4.3 Housing Reconfiguration
Rotate the housin
g
of the display unit in order to be able to read the display from
above when the pressure transmitter is installed in an upri
g
ht position.
Loosen the 4 internal hexa
g
onal screws.
Li
g
htly lift off the housing with the display unit.
Carefull
y
turn the housing by 180°.
Re-ti
g
hten the screws.
When tightening the 4
hollow screws,
make sure that they are ade­quately and securely seated in order to ensure that the transmitter is properly sealed.
Connection Cable
Supporting String
Attention
Page 17
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Installation
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
56
4.4 Electrical Connection
The suppl
y
voltage is between 12 and 36 V DC. The power supply and the output sig­nal are transmitted via a two-wire cable (max. 12 mm outer diameter, max. 14 AWG) and connected in accordance with the pin confi
g
uration.
Suppl
y
voltage can be supplied by a power unit, a transmitter power supply or by
means of a PLC connection.
It is su
gg
ested to use a model with integrated lightning protection for preventing dam-
a
g
e due to voltage peaks.
Please observe local installation regulations (Germany: VDE-Standard). The terminal voltage should not exceed 36 V.
Attention
Page 18
C
Universal Pressure Transmitter UniTrans
Installation
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
57
Terminal Configuration
Ground
L- suppl
y
minus (to connect the negative signal line)
L+ suppl
y
plus (to connect the positive signal line)
I Test circuit; connect the ampere meter between terminals L+ and I
The unit must be properl
y g
rounded in order to guarantee EMC resistance.
4.5 Pressure Compensation when using a Relative Pressure Sensor
A Goretex diaphram is used to compensate for the atmospheric pressure under the
IP 65 Protection Method.
A special cable with capillaries for relative pressurization is used for In
g
ress Protec-
tion IP 67.
Page 19
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters without Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
58
5 Operation of Transmitters without Display
5.1 Preparation
This unit can be pro
g
rammed before or after installation.
Connect an ampere meter to the devices output (between terminals I and L+).
Note that after each action, a brief oscillation/deflection of 20 mA occurs (verifica-
tion of a successful action).
The followin
g
functions can be programmed without a display unit:
Zero point adjustment with a full or empt
y
container (with/without pressure)
Span adjustment with a full or empt
y
container (with/without pressure)
Inte
g
ration time
Reset to manufacturers defaults
5.2 Key Functions (only available for transmitters without display)
An error signal is caused by a current surge (21 mA or 3.6 mA; 5 sec) when the zero point or span setting fall outside of the sensors nominal pressure range during adjustments with existing pressure. No values are stored. The keypad becomes inactive after 10 min. of disuse. All settings will de­fault to previously stored values. Only settings that have been confirmed with the "OK" function are stored.
Function 1 Function 2
Basic settin
g
,
store span (2 s)
Action: upward, increase value
Basic settin
g
, store zero point (2 s)
Action: downward, decrease value
Exit ke
y
or pro-
g
ramming mode
(2 s)
Activate ke
y
s (push simulta­neousl
y
for 2 sec.)
Verification (store) (2 s)
Mountin
g
correc­tion of the sensor (push simul­tanousl
y
for 2 sec.)
Basic settin
g
Inte
g
ration time/
dampin
g
(push
simultaneousl
y
for
2 sec.)
Reset to default (push simulta­neousl
y
for 2 sec.)
Page 20
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters without Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
59
meas
Zero point calibration
Calibration Value
Confirm
Function
(2 sec.)
(2 sec.)
(2 sec.)
024-DS-GB
meas
Calibration of span
Calibration Value
Confirm
Function
(2 sec.)
(2 sec.)(2 sec.)
025-DS-GB
5.3 Calibration with Pressure
5.3.1 Zero Point Calibration
5.3.2 Span Calibration
Calibration of the measurement ran
g
e (span).
Make sure that the pressure to be used as the zero point (P 0 %), is present at the transmitterer diaphragm before calibration.
Make sure that the pressure to be used as the span end-point (P 100 %) is present at the transmitter diaphragm. The measuring range between zero and end value of span is saved as span.
A change in the zero point has no effect on the calibrated span. However, if the span end-point is higher than the peak value of the sen­sors nominal pressure range, then span end-point is fixed at this peak val­ue and the span is reduced accordingly. A change in the span setting has no effect on the zero point. The zero point and span end-point must fall within the sensors nominal pressure range. A mounting correction must not be carried out when making an adjustment with pressure (wet adjustment). Otherwise, the mounting correction must be carried out before saving the zero point and the end value of span.
Page 21
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters without Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
60
5.4 Calibration without Pressure
Determine the current reference values for the zero point and the span to be entered in the transmitter before calibration. This is done as follows:
5.4.1 Zero Point Calibration
Determine the h
y
drostatic pressure of the liquids surface that meets the zero point.
Adjust this pressure in proportion to the sensors nominal pressure ran
g
e.
Multipl
y
this proportion by 16 mA and add 4 mA to the result.
This produces the calculated current (value I
calc
), which is entered in the transmitter
and used to pro
g
ramm the zero point (0%).
Example:
A pressure transducer with 0 ... 400 mbar (nominal pressure) needs to be pro-
g
rammed.The liquids surface (with a density of 1) is 1 m above the diaphragm at the
zero point producin
g
a pressure of 100 mbar.
This means that the devices current value must be set to 8 mA when performin
g
a dry
(empt
y
) calibration.
5.4.2 Span Calibration
Determine the h
y
drostatic pressure of the liquids surface, which corresponds to the
span end-point.
Calculate the difference of the pressure value between span end-point and zero point and divide this difference b
y
the nominal pressure range of the sensor.
Multipl
y
this proportion by 16 mA and add 4 mA to the result.
This produces the calculated current (value I
calc
), which is entered in the transmitter
and used to pro
g
ram the span end-point (100%).
The measurement ran
g
e between zero point and span end-point will be stored as
span.
I
calc
Zero point pressure (0%) 100 mbar
Sensors nominal pressure 400 mbar
----------------------------------------------- -------------------------------------------------
16 mA 4 mA
+
8 mA
==
Calibration Value
calc
Calibration of zero point, without display, without pressure
Calibration Value
... to I
meas=Icalc
... to I
meas=Icalc
(2 sec.)
(2 sec.)
(2 sec.)
026-DS-GB
Page 22
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters without Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
61
Example:
A pressure transmitter with 0 ... 400 mbar (nominal pressure) is to be programmed. The liquids surface (with a densit
y
of 1) is 1 m above the diaphragm at the zero point.
The maximum (span end-point) should be 3 m. The measurin
g
range (span) is 200
mbar.
This means that the output must be set to 12 mA durin
g
programming.
5.4.3 Mounting correction of the sensor
The position of the measurin
g
cell ist adjusted by simultaneously pressing (2 sec.) the
"zero" and "esc" buttons.
A change in the zero point has no effect on the adjusted span. However, if the span end-point is higher than the peak value of the transmitters nominal pressure range, then the span end-point is fixed at this peak value and the span is reduced accordingly. A change in the span setting has no effect on the zero point. The zero point and span end-point must fall within the transmitters nominal pressure range.
A test / correction of the zero point is suggested after adjusting the span in order to maintain optimum accuracy. A mounting correction should be carried out before or after making an adjustment without pressure (dry adjustment) (see 5.4.3). The trans­mitter must therefore be placed in the reference position for the mea­surement (installation site) without pressure on the diaphragm.
I
calc
pressure difference (span) (300 mbar -100 mbar)
Sensors nominal pressure 400 mbar
--------------------------------------------------------------------------------------------------------------------------------
--
16 mA 4 mA
+
12 mA
==
Enter Value
Access
Function
... to I
meas=Icalc
Activate
Keypad (2 sec.)
Calculate
Calibration Value
Calibration of span, without display, without pressure
027-DS-GB
current value decreases
Confirm
Calibration Value
... to I
meas=Icalc
current value increases
I
calc
=
(2 sec.)
(2 sec.)
Important
Mounting correction of the sensor
049-DS-GB
Calibration Value
Function
(2 sec.)
(2 sec.)
(2 sec.)
Page 23
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters without Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
62
5.5 Integration Time (Damping) Adjustment
The followin
g
integration time settings can be used: 0, 1, 5, 20 and 40 s.
The sensors measured values can then be avera
g
ed using the adjusted integration
time.
5.6 Reset to Default
All default data settin
g
s are restored by simultaneously pressing the "zero", "esc" and
the "ok" buttons for 2 seconds (see chapter 6.4).
Calibrated special measurement ranges i. e. 4 bar on a 6 bar transmit­ter can be adjusted by factory pre-setting. A reset to default will reset the sensor back to its nominal range (i. e. 6 bar). The factory pre-setting gets lost.
Important
Reset to default
050-DS-GB
Calibration Value
Function
(2 sec.)
(2 sec.)
(2 sec.)
Calibration Value
Calibration Value
Calibration of the integration time
(2 sec.)
(2 sec.)(2 sec.)
4 times
028-DS-GB
Page 24
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
63
6 Operation of Transmitters with Display
6.1 The Display
In order to pro
g
ram the device, remove the display with a screwdriver and re-attach it
to the housin
g
as shown in the diagram below.
Page 25
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
64
6.2 Key Functions
6.3 The Programming Mode
The transmitter can be pro
g
rammed before or after installation.
The ke
y
pad is activated and the device can be programmed by simultaneously press-
in
g
the "esc" and "ok" keys (for 2 sec.). This method is used to access the main
menus. Each main menu has one or more sub-menus and each sub-menu, ma
y
have
its own sub-menus.
Button Functions
Main Menu Sub-menu Edit Functions
back to the previous menu option
back to the previous menu option
increase value
forward to next menu option
forward to next menu option
decrease value
back to value displa
y
without savin
g
back to main menu without savin
g
back without savin
g
to the sub-menu to the edit functions save value
activate ke
y
pad (push simultaneously; 2 s)
The keypad becomes inactive after 10 min. of disuse. All settings will de­fault to previously stored values. Only settings that have been confirmed with the "OK" function are stored. A change in the starting measurement (zero point) has no effect on the measurement span. Likewise, a change in the span has no effect on the starting measurement. An error signal occurs when the zero point or span settings fall outside of the sensors nominal pressure range during calibration with pressure. Nothing is saved.
Page 26
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
65
6.4 Default Data (factory settings)
Function Defaults
Displa
y
Unit of measurement (Line 1) Line 2 Line 3
Pressure display (in bar)
Temperature displa
y
(in °C)
Sensors nominal pressure ran
g
e (in bar)
Calibration zero 4 mA
span 20 mA
nom. pressure ran
g
e start
nom. pressure ran
g
e end
Output Dampin
g
Inversion Fault Limits I-offset
0 s no 21 mA (upscale)
3.8 ... 20.5 mA 0 mA
Service password no active password
Service mountin
g
correction not activated
Lan
guag
eEn
g
lish
Evaluation linear
densit
y
y
es
1
g
/cm
3
Calibrated special measurement ranges i. e. 4 bar on a 6 bar transmit­ter can be adjusted by factory pre-setting. A reset to default will reset the sensor back to its nominal range (i. e. 6 bar). The factory pre-setting gets lost.
Important
Page 27
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
66
6.5 Main Menu
DISPLAY OPTIONS
CALIBRATION RANGE
OUTPUT DEFINITION
EVALUATION FUNCTION
LANGUAGE OPTIONS
SERVICE FUNCTIONS
Page 28
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
67
6.5.1 Main Menu: Display
DISPLAY OPTIONS
UNIT DISPLAYED
DISPLAY ROW 2
ROW 2 MEASURE
ROW 2 BLANK
ROW 2 MIN VALUE
ROW 2 MEASUREMENT
ROW 2 TEMPERATUR
TEMPERATURE
TEMPERATU
ROW 2 MAX VALUE
ROW 2 MEASUREMENT
ROW 2 TEMPERATUR
TEMPERATU
TEMPERATURE
ROW 2 TEMPERATURE
ROW 2 CURRENT in mA
TEMPERATURE in C
o
TEMPERATURE in F
o
ROW 2 P-RANGE
Second line shows measured value in %
Second line remains empty
Second line shows minimum values
Second line shows minimum Temp. in C
o
Second line shows minimum Temp. in F
o
Second line shows maximum values
Second line shows maximum Temp. in C
o
Second line shows maximum Temp. in F
o
Second line shows current Temp. in C
o
Second line shows current Temp. in F
o
Second line shows present current values
Merger of the sensor's nominal pressure range
Units are set for the measured value.
UNIT
UNIT mmHG
see *)
Units are set for the volume-related value.
UNIT
UNIT lb
see **)
l
VOL_REF 100% = 0.0
OK
VOL_REF 100% = 0.0
OK
Units are set for the measured value.
Units are set for the volume-related value.
031a1-DS-GB
see chapter 2.1.3 display unit
*) The density of a medium must be entered to calculate the correct fill-level when displaying or adju­sting the level in height units (e.g. mm, m, feet, inch) (see 6.5.6). For ranges which require over 4 digits see 2.1.3
**) For volume-based units it is necessary to enter the reference value (100% = 0.0, value range 0 ... 3000).
*) mbar bar PSI at kg/cm² mA % mm m starting with soft­inch ware version 1.05 feet Pa
hPa kPa MPa mmWS
mWS mmHG
**) l kg t starting with soft­m³ ware version 1.05 gal lb
Page 29
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
68
DISPLAY ROW 3
ROW 3 MEASURE %
ROW 3 BLANK
ROW 3 MIN VALUE
ROW 3 MEASUREMENT
ROW 3 TEMPERATURE
TEMPERATURE in C
o
TEMPERATURE in F
o
ROW 3 MAX VALUE
ROW 3 MEASUREMENT
ROW 3 TEMPERATURE
TEMPERATURE in C
o
TEMPERATURE in F
o
TEMPERATURE in C
o
TEMPERATURE in F
o
ROW 3 TEMPERATURE
ROW 3 CURRENT in mA
ROW 3 P-RANGE
Third line shows measured value in %
Third line remains empty
Third line shows minimum values
Third line shows minimum Temp. in C
o
Third line shows minimum Temp. in F
o
Third line shows maximum values
Third line shows maximum Temp. in C
o
Third line shows maximum Temp. in F
o
Third line shows current Temp. in C
o
Third line shows current Temp. in F
o
Third line shows present current value
Merger of the sensors's nominal pressure range
Page 30
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
69
6.5.2 Main Menu: Calibration of zero and span
A single pressure value is set for the zero point or the span end-point within the sensors nominal pressure range, and assigned to the associ­ated output current signal when making adjustments with existing pres­sure. An error signal occurs when the existing pressure lies outside of the sensors nominal pressure range. The value is not saved in this case.
A mounting correction should be performed before or after making an adjustment without pressure (dry adjustment) (see 6.5.6). The sensor must therefore be placed in the reference position for the measurement (installation site) without pressure on the diaphragm.
A mounting correction is unnecessary when making an adjustment with pressure (wet adjustment). Otherwise, the mounting correction must be performed before saving the zero point and span end- point.
A test / correction of the zero point is suggested after adjusting the span in order to maintain optimum accuracy.
MAIN MENU
CALIBRATION
CALIBRATE WITH PRESSURE
CALIBRATE RANGE
CALIBRATE SET ZERO
CALIBRATE SET SPAN
CALIBRATE WITHOUT PRESSURE
CALIBRATE DEFINE ZERO
CALIBRATE DEFINE SPAN
DEFINE ZERO
DEFINE SPAN
Adjustment Zero Point
Adjustment Span
Adjustment zero point via entry of a pressure value within the nominal pressure range
Adjustment span via entry of a pressure value within the nominal pressure range
Important
Important
Page 31
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
70
6.5.3 Main Menu: Output
DAMPING
Damping 0 s active
DAMPING
DAMPING
DAMPING
DAMPING
MAIN MENU
OUTPUT
OUTPUT DEFINITION
OUTPUT DAMPING
OUTPUT INVERT: OFF
OUTPUT INVERT: ON
OUTPUT ALARM/ERROR
OUTPUT LIMITS
OUTPUT CURRENT OFFSET
LIMITS
LIMITS
The current runs between 4 and 20 mA Alarm condition when limits are reached;
It is necessary to restart device by using reset or by shutting down power
(see 8.5.6).
Damping 40 s active
Damping 20 s active
Damping 5 s active
Damping 1 s active
Output is inverted (20 mA ... 4 mA)
Output is not inverted (4 mA ... 20 mA)
Alarm at 3.6 mA (downscale)
Alarm at 21 mA (upscale)
Measured value runs between 3.8 mA and 20.5 mA. The current holds at the limit when the sensing range is exceeded.
Query routine for additional safety.
The output current is combined with an assigned offset, which can have a maximum of 0.4 mA.
033-DS-GB
OUTPUT INVERSION
Page 32
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
71
6.5.4 Main Menu: Evaluation
Enter hei
g
ht values, which are each assigned a volumetric value of measure for tank
linearization. The linearization and the assi
g
nment of the 4 ... 20 mA output signal
areconverted into tank volumes usin
g
this value pair.
When the evaluations function is activated the Turn down becomes inactive
MAIN MENU
EVALUATION
EVALUATION FUCTIONS
EVALUATION REF. TABLE
REF. POINT INSERT
REF. POINT DELETE
INSERT
INSERT
DELETE
DELETE
DELETE
The level value and volume value always be adjusted or verified in order to store a value pair.
TABLE EDIT
TABLE ACTIVATE ?
EVALUATION LINEAR
LINEAR ACTIVATE ?
EVALUATION DENSITY
Entry of a new value pair for the table used in tank linearization
Entry of the level as a % Entry of the associated
volume as a % It is possible to enter up to
30 value pairs The value pair for 0% (P0)
and 100% (P31) are fixed
The selected tag (value pair) is deleted
The upper limit can only be veri­fied, not altered
The lower limit can only be veri­fied, not altered
Selected tag (value pair: level/volume) is altered
The preset linearization table is activated
The linear correlation between input and output is set
The density of the medium is set
l
d=
Page 33
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
72
Example:
Level 100 %: 4000 mm
Densit
y
:1
g
/cm
3
Density correction: 0.9 g/cm
3
Span end point:
The span (end-point) must be re-calibrated (with or without pressure) to 4000 mm in order to prevent a 4000 mm level tank from bein
g
overfilled.
6.5.5 Main Menu: Language
Please check the following if "Wrong Entry" appears in the Evaluation menu:
whether or not more than 32 value pairs are entered in the table for tank linearization (please note: P 0 and P 31 are fixed at 0% and 100% re­spectively)
whether or not an existing height value was tried to be stored again Please enter correct values.
A change or correction in the density causes a change in the span end­points unit of measure (mm, m, inch, feet). The span end-point must be re-calibrated when changing the medium to be measured (due to a change in density).
4000 mm 1 g/cm
3
0.9
g
/cm
3
-------------------------------------------- ------
4444 mm
=
Important
MAIN MENU LANGUAGE
LANGUAGE OPTIONS
LANGUAGE GERMAN
LANGUE FRANCAIS
LANGUAGE ENGLISH
All displays will be in German
All displays will be in English
All displays will be in French
IDIOMA ESPANOL
All displays will be in Spanish
LINGUA ITALIANO
All displays will be in Italian
Page 34
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
73
6.5.6 Main Menu: Service
MAIN MENU
SERVICE
MOUNTING CORRECTION
Mounting correction is performed; sensor must be correctly positioned / mounted and without pressure.
SERVICE FUNCTIONS
SERVICE LOOPTEST
SERVICE DEVICE DATA
SERVICE TIMER
RESET TIMER
RESET MIN/MAX VALUES
All values are reset to the factory setting (see 6.4)
SERVICE PASSWORD
PASSWORD (DE)ACTIVE
ENTRY
CIRCUIT TEST
Hrs.-TOTAL
Hrs.-CALIBRATE
Hrs.-RESET
Hrs.-SENSOR
RESET ALL
RESET ALARM
The set current value is used as the test signal until the "esc" button is pressed
Total number of operating hours.
Number of hours since the last calibration.
Number of hours since the last system Reset.
Number of hours since the last Reset to zero.
Number of hours that sensor was in operation.
Min/Max values are reset.
Reset the alarm after exceeding the 4/20 mA limit, necessary when setting fixed limits 4/20 mA FIX (see 6.5.3)
A digital value between 0000 and 9999 is set as the password.
SERVICE RESET
RESET ALL ???
Page 35
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Diagnostics and Service
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
74
7 Diagnostics and Service
The following error messages can appear on devices with displays (see chapter 2.1.3):
8 Disposal
If a failure cannot be repaired, the transmitter must be switched off.The op­erator then must make sure, that it is only switched on again after the failure has been repaired. Repairs should only be carried out by the manufacturer. All other repairs or modifications are unauthorized.
Error Code Error Error Correction Measures
E00 ROM-error Return device to manufacturer
E01 Power suppl
y
error Check power suppl
y
E03
E
2
PROM communications error
Disconnect and reconnect power suppl
y
E04 Sensors temperature range was
exceeded
Return sensors temperature to specified limits
E06 Sensor reco
g
nition Disconnect and reconnect power
suppl
y
E07 General communications error
between the sensor and the con­trol interface unit
Check the connection between the sensor and the control inter­face unit
E08
Error E
2
PROM
send in transmitter for service
Please observe local guidlines and regulations when disposing of trans­mitters that are no longer serviceable. Please turn any recycleable components in to the appropriate local orga­nizations.
Attention
Important
Page 36
C
Universal Pressure Transmitter UniTrans
Appendix
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
75
9Appendix
9.1 Dimension Diagrams
Page 37
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Appendix
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
76
Page 38
C
Universal Pressure Transmitter UniTrans
Appendix
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
77
Page 39
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Appendix
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
78
Page 40
C
Universal Pressure Transmitter UniTrans
Appendix
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
79
9.2 Model Key
1) with accuracy 0.5 % only; max. Turn down 2:1
2) please make sure to consider the max. pressure admissible for your respective high pressure tube (see manufacturers specification for the high pressure tube)
Unit
B
bar
1S
bar absolut
bis 16 bar abs
Pressure range
CA
-1 bar ... 0 bar
BM
0 bar ... 40 bar
CD
-1 bar ... 0,6 bar
BO
0 bar ... 100 bar
CH
-1 bar ... 3 bar
BQ
0 bar ... 250 bar
CK
-1 bar ... 5 bar
BT
0 bar ... 600 bar
CP
-1 bar ... 15 bar
BU
0 bar ... 1000 bar
BB
0 bar ... 0,4 bar / bar absolut
BV
0 bar ... 1600 bar
1)
only with aluminum case
BE
0 bar ... 1,6 bar / bar absolut
BX
0 bar ... 2500 bar
1)
only with aluminum case
BH
0 bar ... 6 bar / bar absolut
BZ
0 bar ... 4000 bar
1)
only with aluminum case
2BK
0 bar ... 16 bar / bar absolut
Process connection
GD
G ½ B
ND
½ NPT
ML
M16 x 1,5 female, with sealing cone
2)
> 1600 bar
VS
3/8-24 UNF LH male
> 1600 bar
3CS
chemical seal
prices and designs according to chemical seals product range
Special design features
Z
without
E
oil and grease free
A
oxygen, oil and grease free
up to 1600 bar abs, max. medium temperature 60°C
G
suitable for food
4O
overvoltage protection according to IEC 801-5
Case material
M
highly resistive, fiberglass-enforced plastic (PBT)
Ingress protection IP 65
5A
Aluminium
Ingress protection IP 67
Electrical connection
A
cable gland M20x1.5 with internal terminal block
standard
6M
4 pin locking plug M12x1
Digital display
Z
without
7A
with integrated 4-digit LCD-display
Approvals
Z
without
8?
others
Additional order info
YES NO
91Z
quality certificates
10 T Z
additional text
Order code:
12 3 45 678 910
UT-10 - A - - - S -
Page 41
Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Appendix
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
80
Unit
B
bar
1S
bar absolut
bis 16 bar abs
Pressure range
CA
-1 bar ... 0 bar
BH
0 bar ... 6 bar / bar absolut
CD
-1 bar ... 0,6 bar
BK
0 bar ... 16 bar / bar absolut
CH
-1 bar ... 3 bar
BM
0 bar ... 40 bar
CK
-1 bar ... 5 bar
BO
0 bar ... 100 bar
CP
-1 bar ... 15 bar
BQ
0 bar ... 250 bar
BB
0 bar ... 0,4 bar / bar absolute
BT
0 bar ... 600 bar
2BE
0 bar ... 1,6 bar / bar absolute
Process connection
85
G 1 B, flush diaphragm with O-ring
up to 1.6 bar
86
G ½ B flush diaphragm with O-Ring
> 1.6 bar
G6
G 1 ½ B flush diaphragm
up to 16 bar
83
G 1 flush diaphragm according to EHEDG
1)
up to 16 bar
384
G 1 flush diaphragm up to +150 °C according to EHEDG
1)
up to 16 bar
Material of wetted parts
1
stainless steel and O-ring from NBR
G 1 1/2 without O-ring
L
stainless steel and O-ring from Viton
B
stainless steel and O-ring from EPDM
4S
Hastelloy C4
Special design features
Z
without
E
oil and grease free
A
oxygen, oil and grease free
up to 100 bar, max. medium temperature 60°C
G
suitable for food
5O
overvoltage protection according to IEC 801-5
Case material
M
highly resistive, fiberglass-enforced plastic (PBT)
Ingress protection IP 65
6A
Aluminium
Ingress protection IP 67
Electrical connection
A
cable gland M20x1.5 with internal terminal block
standard
7M
4 pin locking plug M12x1
Digital display
Z
without
8A
with integrated 4-digit LCD-display
Approvals
Z
without
9?
other
Additional order info
YES NO
10 1 Z
quality certificates
11 T Z
additional text
1) not with ’’Special design features’’ Code A
Order code:
1 2 3 456 789 1011
UT-11 - A - - - S -
Page 42
C
Universal Pressure Transmitter UniTrans
Appendix
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
81
9.3 Warranty Conditions
The pressure transmitter has a 24 month warrant
y
in accordance with the WIKA
General Terms of Deliver
y
.
9.4 Glossary
Adjustment Allocation of the si
g
nal output range (4 ... 20 mA) to the desired
pressure measurement ran
g
e or level measurement range.
Inte
g
ration Also damping: timely communication of the measurement sig-
nal; rise time of the current output si
g
nal after a signal surge
Inversion Conversion of the output si
g
nal from 4 ... 20 mA to 20 ... 4 mA
Nom. pressure ran
g
e The operating pressure range for which the sensor was de-
si
g
ned
Zero point Start of the pressure measurement ran
g
e
Parameterization Also confi
g
uration: programming of the relevant parameters
and the pressure measurement ran
g
e specific to the application
and measurement location.
Span The pro
g
rammed pressure measurement range
Span end point The hi
g
hest pressure value of the programmed measurement
span (end-point of the span)
Tank linearization Determination of approximate volume/pressure ratio values
with non-linear correlations based on var
ying
container designs
For example, a non-linear correlation exists between the fill lev­el and the volume in spherical containers. Durin
g
linearization,
the non-linear volume is assi
g
ned the 4 ... 20 mA output signal
from a table of values (proximit
y
process by means of up to 32
support points).
Defaults The sensor parameters are pre-pro
g
rammed by the manufac-
turer
9.5 Units of Pressure Measurement
1 atm (atmospheres) = 760 mm H
g
= 760 Torr
= 1.033 kp/cm
2
= 0.1013 MPa
1 Torr = 133.3 Pa
1 kp/mm
2
= 9.81 N/mm2 = 9.81 MPa
1 bar = 0.1 MPa
1 mbar = 1 hPa (Hektopascal)
1 psi (pound per square inch) = 6.895 · 10
3
Pa
1 bar = 33.5 feet of water
1 PA = 1.0 x 10
-5
bar
1 mmHG = 1.333 mbar
Repairs may only be carried out by the manufacturer. All other repairs and device modifications are unauthorized and will void the warranty.
Attention
Loading...