
GBF
Operating instructions
Betriebsanleitung
Mode d’emploi
WIKA Alexander Wiegand GmbH & Co. KG
Alexander-Wiegand-Straße 30
63911 Klingenberg/ Germany
Tel. (+49) 93 72/132-295
Fax (+49) 93 72/132-706
E-Mail support-tronic@wika.de
www.wika.de
UT-10 / UT-11
UT-10
UT-11
GB
GB
Universal Transmitter for various applications
DUniversaltransmitter für vielfältige Einsatzgebiete
Transmetteur unviersel pour utilisations multiples

S-Nr. 218 4125.04 Ausgabedat um 16.02.04
Universal-Drucktransmitter
Universal Pressure Transmitter
Transmetteur de Pression Universel
UniTrans
®
Betriebsanleitung / Manual / Manuel

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Contents
Subject t o change due to te chnical modific ations. © Copyright WIKA A lexander W iegand GmbH & Co . KG / BRD
and GmbH & Co. KG · Al exander-W iegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
42
Contents
1 General Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
2 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
2.1 Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
2.1.1 Pressure Transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
2.1.2 Processin
Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
2.1.3 Displa
Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
2.2 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2.2.1 Functions of Transmitters without Displa
s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2.2.2 Functions of Transmitters with Displa
s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2.3 Installation Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
3 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
3.1 Input-values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
3.2 Output-values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
3.3 Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
3.4 Auxilliar
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.5 Ambient Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.6 Process Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.7 Identification Plates
example) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.1 Pressure Transmitter Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.2 Displa
Unit Upgrades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Reconfiguration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.4 Electrical Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
4.5 Pressure Compensation when usin
a Relative Pressure Sensor . . . . . . . . . . . . 57
5 Operation of Transmitters without Display. . . . . . . . . . . . . . . . . . . . 58
5.1 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.2 Ke
Functions (only available for transmitters without display) . . . . . . . . . . . . . . 58
5.3 Calibration with Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.1 Zero Point Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.2 Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.4 Calibration without Pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.4.1 Zero Point Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.4.2 Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.4.3 Mountin
correction of the sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.5 Inte
ration Time (Damping) Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.6 Reset to Default . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6 Operation of Transmitters with Display. . . . . . . . . . . . . . . . . . . . . . . 63
6.1 The Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.3 The Pro
ramming Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

C
Universal Pressure Transmitter UniTrans
Contents
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 6391 1 Klingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
43
6.4 Default Data (factory settings). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
6.5 Main Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
6.5.1 Main Menu: Displa
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
6.5.2 Main Menu: Calibration of zero and span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
6.5.3 Main Menu: Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
6.5.4 Main Menu: Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
6.5.5 Main Menu: Lan
e. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
6.5.6 Main Menu: Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
7 Diagnostics and Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
8 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9 Appendix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.1 Dimension Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
9.3 Warrant
Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
9.4 Glossar
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
9.5 Units of Pressure Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Contents
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. K G / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
44
1 General Safety Instructions
All pressure connections may only be opened after the system is
without pressure!
Observe the national regulations about safety and accident prevention , as
well as the safety instructions in this operating manual when operating the
pressure transmitter.
Any operation not described in the following instructions must not be carried out.
If a failure cannot be repaired, the transmitter must be switched off. The
operator must then make sure that it is only switched on again after the
failure has been repaired.
Prior to installing, starting and operating a pressure measuring instrument,
the user must ensure that the appropriate instrument has been selected
with regard to scale range and performance and that the wetted parts
material is suitable for the specific measuring conditions of the respective
application.
Serious injuries and/or damage can occur should the relevant regulations
not be observed.
Dangerous pressure media such as oxygen, acetylene, flammable gases
or liquids and toxic gases or liquids as well as instruments for refrigeration
plants or compressors etc. require attention above the standard regulations. Here the specific safety codes or regulations must be considered.
Ramaining pressure medium contained in the pressure element may be
hazardous or toxic. This should be considered when handling and storing
the removed pressure measuring instrument.
Repairs should only be carried out by the manufacturer. All other repairs or
modifications of the transmitter are unauthorized.
Other important safety guidelines can be found in the different sections of
this instruction manual.
Warnin

C
Universal Pressure Transmitter UniTrans
Product Description
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
45
2 Product Description
The UniTrans pressure transmitter can be used in level control applications as well as
for pressure measurement applications in process industr
. A variety of process con-
nections, measurement ran
es, main boards and display options result in a product
e of applications.
2.1 Construction
The UniTrans consists of a pressure sensor, a control interface unit and a housin
cover with optional displa
. Due to this modular design, different transmitter versions
can be mounted (see “Model Ke
2.1.1 Pressure Transducer
The pressure transducer has a piezo-resistive or thinfilm measurement cell dependin
on the pressure range. The sensors are temperature compensated, and have a
welded membrane which is "helium" leak-tested. The pressure trans-
ducers do not have internal sealin
elements.
Pressure transducers further distin
uish themselves from one another based on their
pressure ran
es and the different materials of wetted parts. Different process
connections can be selected to serve a wide ran
e of applications.
Never exceed the overpressure limit of the respective pressure transducer.
Direct Mounted
Processing Unit Housing

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Product Description
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
46
2.1.2 Processing Unit
The processin
unit, which is integrated in the housing contains the terminal compart-
pad used for programming the transmitter. The four keys must be
activated (unlocked) before use. Durin
normal operation the keypad is locked to pro-
tect data and functions previousl
entered. The keypad is automatically locked when
is hit for 10 minutes. The processing unit converts the digitalized signal from
unit into a standard 4...20 mA current signal.
2.1.3 Display Unit
The indicator has four di
its (7-segment display) + symbols. Below it, line 1 (16-seg-
) is used to display error codes and the signal’s unit of measure. The unit
of measure can be selected b
the operator. Measurements over 9999 can not be cor-
rectl
displayed. Please note this when choosing the unit (e.g. 9999 Pascal = 0,09999
bar). Additional information is displa
ed in lines 2 and 3 (16-segment display). The
operator can enter commands in the pro
ramming mode on the display unit by means
uided, clear-text prompts.
Processing Unit
CALIBRATION
RANGE

C
Universal Pressure Transmitter UniTrans
Product Description
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
47
Transmitters with displays offer a larger number of programming and processing options. These options include alarm status, dampin
, signal inversion, tank lineariza-
2.2 Function
The mode of operation for si
nal conversion works in the same way for all versions.
The pressure transducer converts the existin
pressure into an electrical signal. Mi-
croelectronics further process the input si
nal and produce a proportional 4-20 mA
standard si
-version allows programming (parameterization) and the display of ex-
tended functions such as inversion, dampin
, alarm status and linearization.
2.2.1 Functions of Transmitters without Displays
• Calibration of zero and span under pressure (see 5.3)
• Calibration of zero and span without pressure (dr
adjustment) (see 5.4)
• Settin
the dampening / integrating the output signal 0-40 s (see 5.5)
• Reset to manufacturer’s default values (see 5.6)
• Mountin
correction of the sensor (beginning with software version 1.05)
(see section 5.4.3)
2.2.2 Functions of Transmitters with Displays
• Settable units of measurement (mbar, bar, psi, mA, %, m, mm WS) (see 6.5.1)
• Temperature and Min/Max values shown in displa
(see 6.5.1)
• Nominal pressure ran
e of the sensor shown in display (see 6.5.1)
• Zero and span calibration (with/without pressure) (see 6.5.2)
• Settin
of damping / integration of output signal 0-40 s (see 6.5.3)
• Inversion of the output current si
the output current value in case of alarm (3.6 mA or 21 mA) (see 6.5.3)
• Settin
the limits of the output signal (see 6.5.3)
• Offset of the output si
nal (see 6.5.3)
• Mountin
correction of the sensor
• Measurin
circuit test function (see 6.5.4)
• Reset functions (see 6.5.4)
• Password activation (see 6.5.4)
• Selectin
the language of the display (see 6.5.5)
of a table function for the linearization of the output signal (see 6.5.6)
the medium density (see 6.5.6)
Display units can be easily upgraded (see chapter 4.2).

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Product Description
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
48
2.3 Installation Examples
The UniTrans is primaril
used to monitor the pressure in pipes, technical equipment
and tanks. Dependin
on the pressure range pressures between 20 mbar up to 1000
bar can be measured. The pressure is measured usin
absolute (against a vacuum)
ainst external or air pressure) measurement depending on the type of
sensor selected.
The UniTrans is also used for h
drostatic pressure measurement within liquid filled
pipes and containers.
Process Pressure Measurement: Process Pressure Measurement:
Used to measure pressure of liquids Used to measure container pressure.
or gases in pipelines.
Process Pressure Measurement: Process Pressure Measurement:
Installed behind feed pumps for Installed in front of and behind the filter.
process control or monitoring of pump Uses the pressure differential for moni-
functions. toring the function or accumulation of dirt
in the filter. Both output signals are pro-
cessed by a PLC or signal converter.

C
Universal Pressure Transmitter UniTrans
Product Description
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
49
Level Control: Level Control:
Externally mounted Combined pressure and head pressure
(with front flat diaphragm) are measured by two externally mounted
pressure transducers. The two signals
are analyzed and the differential is cal-
culated by a PLC or suitable signal con-
verter.

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Technical Data
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
50
3 Technical Data
3.1 Input-values
3.2 Output-values
Pressure Ranges
(Absolute pressure upon request)
/ overpressure limit / burst pressure
0 ... 0.4 bar 2 2.4
0 ... 1,6 bar 10 12
0 ... 6 bar 35 42
0 ... 16 bar 80 96
0 ... 40 bar 80 400
0 ... 100 bar 200 800
0 ... 250 bar 500 1200
0 ... 600 bar 1200 2400
1)
0 ... 1.000 bar 1500 3000
0 ... 1,600 bar 2000 4000
0 ... 2,500 bar 3000 5000
0 ... 4,000 bar 4400 7000
-1 ... +0 bar* 2 2.4
-1 ... +0,6 bar* 10 12
-1 ... +3 bar* 35 42
-1 ... +5 bar* 35 42
-1 ... +15 bar* 80 96
*onl
relative pressure
Do not exceed the nominal pressure!
1)For flush diaphragm version: The value specified in the table
applies only when sealing is realised with the sealing ring underneath the hex. Otherwise max. 1500 bar applies.
Output signal 4 ... 20 mA
, hysteresis, repeatability)
<
0.10 at ranges < 1000 bar
<
0.30 at ranges > 1000 bar
Turn down behavior: (1/k)
up to 1 : 5
1 : 5 to 1 : 20
no changes of deviation
the accurac
must be multiplied by the
factor (turn down / 5)
example for TD = 1:15, (k = 15)
accurac
= 0.10 * (15/5) = 0.3
Overall deviation
(at +10 °C ... +40 °C)
<
0.15 % (limit point calibration)
< 0.6 % for pressure ran
es of > 1000
bar
Load RA < (UB–12 V)/0.023 A
with R
A
in Ohm and UB in Volt
Fault si
nal 3.6 mA or 21 mA, programmable

C
Universal Pressure Transmitter UniTrans
Technical Data
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
51
3.3 Construction
Integration time 0s, 1s, 5s, 20s, 40s, programmable
Adjustment of the span Up to Turn down 1 : 20
Integrated lightning protection optional
Zero point adjustment 2.5 ... 99 %
Process connections
Model UT-10
Model UT-11
Model UT-11 EHEDG version
G 1/2 B per DIN 16288 (1/2 NPT)
M 16 x 1,5 with sealin
cone
>
1600 bar
3/8-24 UNF LH male
>
1600 bar
1/4"-28 UNF LH M 250-C
>
1600 bar
G 1B flush diaphra
es:
0 ... 0.4 up to 0 ... 1.6 bar)
G 1/2 B flush diaphra
es:
0 ... 6 bis 0 ... 600 bar)
G 1 1/2 flush diaphra
es:
0 ... 0.4 bis 0 ... 16 bar)
G 1 flush diaphra
es:
0...0.4 bis 0...16 bar)
G 1 flush diaphra
es:0...0.4 bis 0...16 bar)
Materials
Housin
Wetted parts (UT-10)
(UT-11)
Internal transmission fluid
hi
resistive, fiberglass-enforced
aluminium
CrNi-steel 1.4571 and 2.4711
CrNi-steel 1.4571, o-rin
: NBR {Viton or
EPDM}; {Hastello
C4}
Standard {Halocarbon oil for ox
en-
applications}; {FDA-approved}

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Technical Data
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
52
3.4 Auxilliary Power
3.5 Ambient Conditions
3.6 Process Conditions
Electrical connection
per EN 60 529/ IEC529
M 20 x 1.5 cable gland with internal terminal block (see 4.4)
M12x1 plu
, 4-pin (pin allocation: 1+ 3-)
[3/4" NPT female conduit (onl
with alu-
minium case)]
Electric protection Reverse polarit
, overload and short cir-
cuit protection
Power suppl
12 ... 36 V DC
Ambient temperature – 40 °C ... + 85 °C
(– 20 °C ... 70 °C with displa
e temperature – 40 °C ... + 85 °C
(– 35 °C ... 80 °C with displa
)
Climate class D per DIN IEC 654-1
In
ress protection per EN 60 529 IP 65 with plastic case
IP 67 with aluminum case
a
-conformit
Interference emission and immunity
see EN 61326
(also fulfills NAMUR NE 21)
97/23 EG Pressure Equipment Directive (Module H)
Medium temperature
G 1 1/2
G 1 EHEDG with coolin
element
– 30 °C ... + 105 °C
(up to 30 min. 140 °C at an ambient
temperature of < 50 °C
-30 °C ... + 150 °C
°F = (°C * 1.8) + 32
°F = (°C * 1.8) + 32

C
Universal Pressure Transmitter UniTrans
Technical Data
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
53
3.7 Identification Plates (example)

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Installation
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
54
4 Installation
The device should be installed/operated in accordance with the regulations of ElexV,
the Device Safet
Regulation, this operating manual and generally recognized indus-
standards.
4.1 Pressure Transmitter Installation
Installation Using a Weld-on Adapter:
• Insert a filler piece (a pressure transmitter dumm
) into the weld-on adapter.
• Weld the adapter into the container/pipe wall (section-weld process).
• Remove the filler piece.
• Install the pressure transmitter in the weld-on adapter.
4.2 Display Unit Upgrades
The displa
unit can be easily upgraded at any time.
cover and the supporting string.
unit’s supporting string to the same place.
the display unit’s connector into the appropriate jack.
unit can be mounted at 90° angles.
unit with screws.
The pressure transmitter’s diaphram should not come into contact with
hard or sharp objects.
When installing the display unit, make sure that the connection cable
and the supporting string are not kinked or pinched.
Attention
Attention

C
Universal Pressure Transmitter UniTrans
Installation
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
55
•
All functions are pro
rammable once the pressure transmitter has been upgraded
unit. The adjusted parameters are stored after the display unit is re-
unit can be rotated in 300°, so that it can be read under various installation
conditions. The housin
cover with built-in display can be fastened to the housing at
all four side positions.
4.3 Housing Reconfiguration
Rotate the housin
of the display unit in order to be able to read the display from
above when the pressure transmitter is installed in an upri
ht position.
• Loosen the 4 internal hexa
htly lift off the housing with the display unit.
turn the housing by 180°.
hten the screws.
When tightening the 4
hollow screws,
make sure that they are adequately and securely seated in order to ensure that the transmitter is
properly sealed.
Connection Cable
Supporting String

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Installation
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
56
4.4 Electrical Connection
The suppl
voltage is between 12 and 36 V DC. The power supply and the output signal are transmitted via a two-wire cable (max. 12 mm outer diameter, max. 14 AWG)
and connected in accordance with the pin confi
voltage can be supplied by a power unit, a transmitter power supply or by
means of a PLC connection.
It is su
ested to use a model with integrated lightning protection for preventing dam-
Please observe local installation regulations (Germany: VDE-Standard).
The terminal voltage should not exceed 36 V.
Attention

C
Universal Pressure Transmitter UniTrans
Installation
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
57
Terminal Configuration
Ground
L- suppl
minus (to connect the negative signal line)
plus (to connect the positive signal line)
I Test circuit; connect the ampere meter between terminals L+ and I
The unit must be properl
rounded in order to guarantee EMC resistance.
4.5 Pressure Compensation when using a Relative Pressure Sensor
A Goretex diaphram is used to compensate for the atmospheric pressure under the
IP 65 Protection Method.
A special cable with capillaries for relative pressurization is used for In

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters without Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
58
5 Operation of Transmitters without Display
5.1 Preparation
This unit can be pro
rammed before or after installation.
• Connect an ampere meter to the device’s output (between terminals I and L+).
• Note that after each action, a brief oscillation/deflection of 20 mA occurs (verifica-
tion of a successful action).
The followin
functions can be programmed without a display unit:
• Zero point adjustment with a full or empt
container (with/without pressure)
• Span adjustment with a full or empt
container (with/without pressure)
• Inte
ration time
• Reset to manufacturer’s defaults
5.2 Key Functions (only available for transmitters without display)
An error signal is caused by a current surge (21 mA or 3.6 mA; 5 sec) when
the zero point or span setting fall outside of the sensor’s nominal pressure
range during adjustments with existing pressure. No values are stored.
The keypad becomes inactive after 10 min. of disuse. All settings will default to previously stored values. Only settings that have been confirmed
with the "OK" function are stored.
Function 1 Function 2
Basic settin
,
store span (2 s)
Action: upward,
increase value
Basic settin
,
store zero point
(2 s)
Action: downward,
decrease value
Exit ke
for 2 sec.)
Verification (store)
(2 s)
Mountin
correction of the sensor
(push simultanousl
for
2 sec.)
Reset to default
(push simultaneousl

C
Universal Pressure Transmitter UniTrans
Operation of Transmitters without Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
59
meas
Zero point calibration
Calibration Value
Confirm
Function
(2 sec.)
(2 sec.)
(2 sec.)
024-DS-GB
meas
Calibration of span
Calibration Value
Confirm
Function
(2 sec.)
(2 sec.)(2 sec.)
025-DS-GB
5.3 Calibration with Pressure
5.3.1 Zero Point Calibration
5.3.2 Span Calibration
Calibration of the measurement ran
e (span).
Make sure that the pressure to be used as the zero point (P 0 %), is
present at the transmitterer diaphragm before calibration.
Make sure that the pressure to be used as the span end-point (P 100 %) is
present at the transmitter diaphragm. The measuring range between zero
and end value of span is saved as span.
A change in the zero point has no effect on the calibrated span.
However, if the span end-point is higher than the peak value of the sensor’s nominal pressure range, then span end-point is fixed at this peak value and the span is reduced accordingly.
A change in the span setting has no effect on the zero point. The zero point
and span end-point must fall within the sensor’s nominal pressure range.
A mounting correction must not be carried out when making an adjustment
with pressure (wet adjustment). Otherwise, the mounting correction must
be carried out before saving the zero point and the end value of span.

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters without Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
60
5.4 Calibration without Pressure
Determine the current reference values for the zero point and the span to be entered
in the transmitter before calibration. This is done as follows:
5.4.1 Zero Point Calibration
• Determine the h
drostatic pressure of the liquid’s surface that meets the zero point.
• Adjust this pressure in proportion to the sensor’s nominal pressure ran
this proportion by 16 mA and add 4 mA to the result.
This produces the calculated current (value I
calc
), which is entered in the transmitter
and used to pro
ramm the zero point (0%).
Example:
A pressure transducer with 0 ... 400 mbar (nominal pressure) needs to be pro-
rammed.The liquid’s surface (with a density of 1) is 1 m above the diaphragm at the
a pressure of 100 mbar.
This means that the device’s current value must be set to 8 mA when performin
) calibration.
5.4.2 Span Calibration
• Determine the h
drostatic pressure of the liquid’s surface, which corresponds to the
span end-point.
• Calculate the difference of the pressure value between span end-point and zero
point and divide this difference b
the nominal pressure range of the sensor.
this proportion by 16 mA and add 4 mA to the result.
This produces the calculated current (value I
calc
), which is entered in the transmitter
and used to pro
ram the span end-point (100%).
The measurement ran
e between zero point and span end-point will be stored as
span.
I
calc
Zero point pressure (0%) 100 mbar
Sensors nominal pressure 400 mbar
----------------------------------------------- -------------------------------------------------
16 mA 4 mA
+
⋅ 8 mA
==
Calibration Value
calc
Calibration of zero point, without display, without pressure
Calibration Value
... to I
meas=Icalc
... to I
meas=Icalc
(2 sec.)
(2 sec.)
(2 sec.)
026-DS-GB

C
Universal Pressure Transmitter UniTrans
Operation of Transmitters without Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
61
Example:
A pressure transmitter with 0 ... 400 mbar (nominal pressure) is to be programmed.
The liquid’s surface (with a densit
of 1) is 1 m above the diaphragm at the zero point.
The maximum (span end-point) should be 3 m. The measurin
mbar.
This means that the output must be set to 12 mA durin
5.4.3 Mounting correction of the sensor
The position of the measurin
cell ist adjusted by simultaneously pressing (2 sec.) the
"zero" and "esc" buttons.
A change in the zero point has no effect on the adjusted span.
However, if the span end-point is higher than the peak value of the
transmitter’s nominal pressure range, then the span end-point is fixed
at this peak value and the span is reduced accordingly.
A change in the span setting has no effect on the zero point. The zero
point and span end-point must fall within the transmitter’s nominal
pressure range.
A test / correction of the zero point is suggested after adjusting the
span in order to maintain optimum accuracy.
A mounting correction should be carried out before or after making an
adjustment without pressure (dry adjustment) (see 5.4.3). The transmitter must therefore be placed in the reference position for the measurement (installation site) without pressure on the diaphragm.
I
pressure difference (span) (300 mbar -100 mbar)
Sensors nominal pressure 400 mbar
--------------------------------------------------------------------------------------------------------------------------------
Enter
Value
Access
Function
... to I
meas=Icalc
Activate
Keypad
(2 sec.)
Calculate
Calibration Value
Calibration of span, without display, without pressure
027-DS-GB
current value decreases
Confirm
Calibration Value
... to I
meas=Icalc
current value increases
I
calc
=
(2 sec.)
(2 sec.)
Mounting correction of the sensor
049-DS-GB
Calibration Value
Function
(2 sec.)
(2 sec.)
(2 sec.)

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters without Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
62
5.5 Integration Time (Damping) Adjustment
The followin
integration time settings can be used: 0, 1, 5, 20 and 40 s.
The sensor’s measured values can then be avera
ed using the adjusted integration
time.
5.6 Reset to Default
All default data settin
s are restored by simultaneously pressing the "zero", "esc" and
the "ok" buttons for 2 seconds (see chapter 6.4).
Calibrated special measurement ranges i. e. 4 bar on a 6 bar transmitter can be adjusted by factory pre-setting. A reset to default will reset
the sensor back to its nominal range (i. e. 6 bar). The factory pre-setting
gets lost.
Important
Reset to default
050-DS-GB
Calibration Value
Function
(2 sec.)
(2 sec.)
(2 sec.)
Calibration Value
Calibration Value
Calibration of the integration time
(2 sec.)
(2 sec.)(2 sec.)
4 times
028-DS-GB

C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
63
6 Operation of Transmitters with Display
6.1 The Display
In order to pro
ram the device, remove the display with a screwdriver and re-attach it
as shown in the diagram below.

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
64
6.2 Key Functions
6.3 The Programming Mode
The transmitter can be pro
rammed before or after installation.
The ke
pad is activated and the device can be programmed by simultaneously press-
the "esc" and "ok" keys (for 2 sec.). This method is used to access the main
menus. Each main menu has one or more sub-menus and each sub-menu, ma
have
its own sub-menus.
Button Functions
Main Menu Sub-menu Edit Functions
back to the previous
menu option
back to the previous
menu option
increase value
forward to next menu
option
forward to next menu
option
decrease value
back to value displa
back to main menu
without savin
to the sub-menu to the edit functions save value
activate ke
pad (push simultaneously; 2 s)
The keypad becomes inactive after 10 min. of disuse. All settings will default to previously stored values. Only settings that have been confirmed
with the "OK" function are stored.
A change in the starting measurement (zero point) has no effect on the
measurement span. Likewise, a change in the span has no effect on the
starting measurement.
An error signal occurs when the zero point or span settings fall outside of
the sensor’s nominal pressure range during calibration with pressure.
Nothing is saved.

C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
65
6.4 Default Data (factory settings)
Function Defaults
Displa
Unit of measurement
(Line 1)
Line 2
Line 3
Pressure display (in bar)
(in °C)
Sensor’s nominal pressure ran
e (in bar)
Calibration zero 4 mA
span 20 mA
nom. pressure ran
e start
nom. pressure ran
Inversion
Fault
Limits
I-offset
0 s
no
21 mA (upscale)
3.8 ... 20.5 mA
0 mA
Service password no active password
Service mountin
correction not activated
Lan
lish
Evaluation linear
densit
/cm
3
Calibrated special measurement ranges i. e. 4 bar on a 6 bar transmitter can be adjusted by factory pre-setting. A reset to default will reset
the sensor back to its nominal range (i. e. 6 bar). The factory pre-setting
gets lost.
Important

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
66
6.5 Main Menu
DISPLAY
OPTIONS
CALIBRATION
RANGE
OUTPUT
DEFINITION
EVALUATION
FUNCTION
LANGUAGE
OPTIONS
SERVICE
FUNCTIONS

C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
67
6.5.1 Main Menu: Display
DISPLAY
OPTIONS
UNIT
DISPLAYED
DISPLAY
ROW 2
ROW 2
MEASURE
ROW 2
BLANK
ROW 2
MIN VALUE
ROW 2
MEASUREMENT
ROW 2
TEMPERATUR
TEMPERATURE
TEMPERATU
ROW 2
MAX VALUE
ROW 2
MEASUREMENT
ROW 2
TEMPERATUR
TEMPERATU
TEMPERATURE
ROW 2
TEMPERATURE
ROW 2
CURRENT in mA
TEMPERATURE
in C
o
TEMPERATURE
in F
o
ROW 2
P-RANGE
Second line
shows measured
value in %
Second line
remains empty
Second line
shows minimum
values
Second line
shows minimum
Temp. in C
o
Second line
shows minimum
Temp. in F
o
Second line
shows maximum
values
Second line
shows maximum
Temp. in C
o
Second line
shows maximum
Temp. in F
o
Second line
shows current
Temp. in C
o
Second line
shows current
Temp. in F
o
Second line
shows present
current values
Merger of the
sensor's nominal
pressure range
Units are set for the
measured value.
UNIT
UNIT
mmHG
see *)
Units are set for the
volume-related value.
UNIT
UNIT
lb
see **)
l
VOL_REF
100% = 0.0
OK
VOL_REF
100% = 0.0
OK
Units are set for the
measured value.
Units are set for the
volume-related value.
031a1-DS-GB
see chapter 2.1.3
display unit
*) The density of a medium must
be entered to calculate the correct
fill-level when displaying or adjusting the level in height units (e.g.
mm, m, feet, inch) (see 6.5.6). For
ranges which require over 4 digits
see 2.1.3
**) For volume-based units it is
necessary to enter the reference
value (100% = 0.0,
value range 0 ... 3000).
*)
mbar
bar
PSI
at
kg/cm²
mA
%
mm
m starting with softinch ware version 1.05
feet
Pa
hPa
kPa
MPa
mmWS
mWS
mmHG
**)
l
kg
t starting with softm³ ware version 1.05
gal
lb

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
68
DISPLAY
ROW 3
ROW 3
MEASURE %
ROW 3
BLANK
ROW 3
MIN VALUE
ROW 3
MEASUREMENT
ROW 3
TEMPERATURE
TEMPERATURE
in C
o
TEMPERATURE
in F
o
ROW 3
MAX VALUE
ROW 3
MEASUREMENT
ROW 3
TEMPERATURE
TEMPERATURE
in C
o
TEMPERATURE
in F
o
TEMPERATURE
in C
o
TEMPERATURE
in F
o
ROW 3
TEMPERATURE
ROW 3
CURRENT in mA
ROW 3
P-RANGE
Third line
shows measured
value in %
Third line
remains empty
Third line shows
minimum values
Third line shows
minimum Temp.
in C
o
Third line shows
minimum Temp.
in F
o
Third line shows
maximum values
Third line shows
maximum Temp.
in C
o
Third line shows
maximum Temp.
in F
o
Third line
shows current
Temp. in C
o
Third line
shows current
Temp. in F
o
Third line
shows present
current value
Merger of the
sensors's nominal
pressure range

C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
69
6.5.2 Main Menu: Calibration of zero and span
A single pressure value is set for the zero point or the span end-point
within the sensor’s nominal pressure range, and assigned to the associated output current signal when making adjustments with existing pressure. An error signal occurs when the existing pressure lies outside of
the sensor’s nominal pressure range. The value is not saved in this
case.
A mounting correction should be performed before or after making an
adjustment without pressure (dry adjustment) (see 6.5.6). The sensor
must therefore be placed in the reference position for the measurement
(installation site) without pressure on the diaphragm.
A mounting correction is unnecessary when making an adjustment with
pressure (wet adjustment). Otherwise, the mounting correction must be
performed before saving the zero point and span end- point.
A test / correction of the zero point is suggested after adjusting the
span in order to maintain optimum accuracy.
MAIN MENU
CALIBRATION
CALIBRATE
WITH PRESSURE
CALIBRATE
RANGE
CALIBRATE
SET ZERO
CALIBRATE
SET SPAN
CALIBRATE
WITHOUT
PRESSURE
CALIBRATE
DEFINE ZERO
CALIBRATE
DEFINE SPAN
DEFINE ZERO
DEFINE SPAN
Adjustment
Zero Point
Adjustment
Span
Adjustment zero point via entry
of a pressure value within the
nominal pressure range
Adjustment span via entry
of a pressure value within the
nominal pressure range

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
70
6.5.3 Main Menu: Output
DAMPING
Damping
0 s active
DAMPING
DAMPING
DAMPING
DAMPING
MAIN MENU
OUTPUT
OUTPUT
DEFINITION
OUTPUT
DAMPING
OUTPUT
INVERT: OFF
OUTPUT
INVERT: ON
OUTPUT
ALARM/ERROR
OUTPUT
LIMITS
OUTPUT
CURRENT OFFSET
LIMITS
LIMITS
The current runs between 4 and 20 mA
Alarm condition when limits are reached;
It is necessary to restart device by
using reset or by shutting down power
(see 8.5.6).
Damping
40 s active
Damping
20 s active
Damping
5 s active
Damping
1 s active
Output is inverted (20 mA ... 4 mA)
Output is not inverted (4 mA ... 20 mA)
Alarm at 3.6 mA
(downscale)
Alarm at 21 mA
(upscale)
Measured value runs between 3.8 mA
and 20.5 mA. The current holds at the
limit when the sensing range is exceeded.
Query routine for
additional safety.
The output current is combined with an
assigned offset, which can have a maximum
of 0.4 mA.
033-DS-GB
OUTPUT
INVERSION

C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
71
6.5.4 Main Menu: Evaluation
Enter hei
ht values, which are each assigned a volumetric value of measure for tank
linearization. The linearization and the assi
nment of the 4 ... 20 mA output signal
areconverted into tank volumes usin
this value pair.
When the evaluations function is activated the Turn down becomes
inactive
MAIN MENU
EVALUATION
EVALUATION
FUCTIONS
EVALUATION
REF. TABLE
REF. POINT
INSERT
REF. POINT
DELETE
INSERT
INSERT
DELETE
DELETE
DELETE
The level value and volume value
always be adjusted or verified in
order to store a value pair.
TABLE
EDIT
TABLE
ACTIVATE ?
EVALUATION
LINEAR
LINEAR
ACTIVATE ?
EVALUATION
DENSITY
Entry of a new value pair
for the table used in
tank linearization
Entry of the level as a %
Entry of the associated
volume as a %
It is possible to enter up to
30 value pairs
The value pair for 0% (P0)
and 100% (P31) are fixed
The selected
tag (value pair)
is deleted
The upper limit
can only be verified, not altered
The lower limit
can only be verified, not altered
Selected tag
(value pair:
level/volume)
is altered
The preset
linearization table
is activated
The linear correlation
between input and
output is set
The density of
the medium is set
l
d=

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
72
Example:
Level 100 %: 4000 mm
Densit
/cm
3
Density correction: 0.9 g/cm
3
Span end point:
The span (end-point) must be re-calibrated (with or without pressure) to 4000 mm in
order to prevent a 4000 mm level tank from bein
overfilled.
6.5.5 Main Menu: Language
Please check the following if "Wrong Entry" appears in the Evaluation
menu:
• whether or not more than 32 value pairs are entered in the table for tank
linearization (please note: P 0 and P 31 are fixed at 0% and 100% respectively)
• whether or not an existing height value was tried to be stored again
Please enter correct values.
A change or correction in the density causes a change in the span endpoint’s unit of measure (mm, m, inch, feet). The span end-point must be
re-calibrated when changing the medium to be measured (due to a
change in density).
4000 mm 1 g/cm
/cm
3
-------------------------------------------- ------
4444 mm
=
Important
MAIN MENU
LANGUAGE
LANGUAGE
OPTIONS
LANGUAGE
GERMAN
LANGUE
FRANCAIS
LANGUAGE
ENGLISH
All displays will
be in German
All displays will
be in English
All displays will
be in French
IDIOMA
ESPANOL
All displays will
be in Spanish
LINGUA
ITALIANO
All displays will
be in Italian

C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
73
6.5.6 Main Menu: Service
MAIN MENU
SERVICE
MOUNTING
CORRECTION
Mounting correction is performed;
sensor must be correctly positioned /
mounted and without pressure.
SERVICE
FUNCTIONS
SERVICE
LOOPTEST
SERVICE
DEVICE DATA
SERVICE
TIMER
RESET
TIMER
RESET
MIN/MAX VALUES
All values are reset
to the factory setting
(see 6.4)
SERVICE
PASSWORD
PASSWORD
(DE)ACTIVE
ENTRY
CIRCUIT TEST
Hrs.-TOTAL
Hrs.-CALIBRATE
Hrs.-RESET
Hrs.-SENSOR
RESET
ALL
RESET
ALARM
The set current value is
used as the test signal until the
"esc" button is pressed
Total number of
operating hours.
Number of hours
since the last
calibration.
Number of hours
since the last
system Reset.
Number of hours
since the last
Reset to zero.
Number of hours
that sensor was in
operation.
Min/Max values
are reset.
Reset the alarm after exceeding
the 4/20 mA limit, necessary when
setting fixed limits 4/20 mA FIX (see 6.5.3)
A digital value between
0000 and 9999 is set as
the password.
SERVICE
RESET
RESET ALL
???

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Diagnostics and Service
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
74
7 Diagnostics and Service
The following error messages can appear on devices with displays
(see chapter 2.1.3):
8 Disposal
If a failure cannot be repaired, the transmitter must be switched off.The operator then must make sure, that it is only switched on again after the
failure has been repaired.
Repairs should only be carried out by the manufacturer. All other repairs or
modifications are unauthorized.
Error Code Error Error Correction Measures
E00 ROM-error Return device to manufacturer
E01 Power suppl
E03
E
2
PROM communications error
Disconnect and reconnect power
suppl
E04 Sensor’s temperature range was
exceeded
Return sensor’s temperature to
specified limits
E06 Sensor reco
nition Disconnect and reconnect power
suppl
E07 General communications error
between the sensor and the control interface unit
Check the connection between
the sensor and the control interface unit
E08
Error E
2
PROM
send in transmitter for service
Please observe local guidlines and regulations when disposing of transmitters that are no longer serviceable.
Please turn any recycleable components in to the appropriate local organizations.
Attention
Important

C
Universal Pressure Transmitter UniTrans
Appendix
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
75
9Appendix
9.1 Dimension Diagrams

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Appendix
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
76

C
Universal Pressure Transmitter UniTrans
Appendix
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
77

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Appendix
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
78

C
Universal Pressure Transmitter UniTrans
Appendix
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
79
9.2 Model Key
1) with accuracy 0.5 % only; max. Turn down 2:1
2) please make sure to consider the max. pressure admissible for your respective high pressure tube
(see manufacturers specification for the high pressure tube)
Unit
B
bar
1S
bar absolut
bis 16 bar abs
Pressure range
CA
-1 bar ... 0 bar
BM
0 bar ... 40 bar
CD
-1 bar ... 0,6 bar
BO
0 bar ... 100 bar
CH
-1 bar ... 3 bar
BQ
0 bar ... 250 bar
CK
-1 bar ... 5 bar
BT
0 bar ... 600 bar
CP
-1 bar ... 15 bar
BU
0 bar ... 1000 bar
BB
0 bar ... 0,4 bar / bar absolut
BV
0 bar ... 1600 bar
1)
only with aluminum case
BE
0 bar ... 1,6 bar / bar absolut
BX
0 bar ... 2500 bar
1)
only with aluminum case
BH
0 bar ... 6 bar / bar absolut
BZ
0 bar ... 4000 bar
1)
only with aluminum case
2BK
0 bar ... 16 bar / bar absolut
Process connection
GD
G ½ B
ND
½ NPT
ML
M16 x 1,5 female, with sealing cone
2)
> 1600 bar
VS
3/8-24 UNF LH male
> 1600 bar
3CS
chemical seal
prices and designs according to chemical seals product range
Special design features
Z
without
E
oil and grease free
A
oxygen, oil and grease free
up to 1600 bar abs, max. medium temperature 60°C
G
suitable for food
4O
overvoltage protection according to IEC 801-5
Case material
M
highly resistive, fiberglass-enforced plastic (PBT)
Ingress protection IP 65
5A
Aluminium
Ingress protection IP 67
Electrical connection
A
cable gland M20x1.5 with internal terminal block
standard
6M
4 pin locking plug M12x1
Digital display
Z
without
7A
with integrated 4-digit LCD-display
Approvals
Z
without
8?
others
Additional order info
YES NO
91Z
quality certificates
10 T Z
additional text
Order code:
12 3 45 678 910
UT-10 - A - - - S -

Publishe d 16.2.04
C
Universal Pressure Transmitter UniTrans
Appendix
Subject t o change du e to technical mo dificatio ns. © Copyright WIKA Alex ander Wie gand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiega nd-Str. · 63911 Kling enberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
80
Unit
B
bar
1S
bar absolut
bis 16 bar abs
Pressure range
CA
-1 bar ... 0 bar
BH
0 bar ... 6 bar / bar absolut
CD
-1 bar ... 0,6 bar
BK
0 bar ... 16 bar / bar absolut
CH
-1 bar ... 3 bar
BM
0 bar ... 40 bar
CK
-1 bar ... 5 bar
BO
0 bar ... 100 bar
CP
-1 bar ... 15 bar
BQ
0 bar ... 250 bar
BB
0 bar ... 0,4 bar / bar absolute
BT
0 bar ... 600 bar
2BE
0 bar ... 1,6 bar / bar absolute
Process connection
85
G 1 B, flush diaphragm with O-ring
up to 1.6 bar
86
G ½ B flush diaphragm with O-Ring
> 1.6 bar
G6
G 1 ½ B flush diaphragm
up to 16 bar
83
G 1 flush diaphragm according to EHEDG
1)
up to 16 bar
384
G 1 flush diaphragm up to +150 °C according to EHEDG
1)
up to 16 bar
Material of wetted parts
1
stainless steel and O-ring from NBR
G 1 1/2 without O-ring
L
stainless steel and O-ring from Viton
B
stainless steel and O-ring from EPDM
4S
Hastelloy C4
Special design features
Z
without
E
oil and grease free
A
oxygen, oil and grease free
up to 100 bar, max. medium temperature 60°C
G
suitable for food
5O
overvoltage protection according to IEC 801-5
Case material
M
highly resistive, fiberglass-enforced plastic (PBT)
Ingress protection IP 65
6A
Aluminium
Ingress protection IP 67
Electrical connection
A
cable gland M20x1.5 with internal terminal block
standard
7M
4 pin locking plug M12x1
Digital display
Z
without
8A
with integrated 4-digit LCD-display
Approvals
Z
without
9?
other
Additional order info
YES NO
10 1 Z
quality certificates
11 T Z
additional text
1) not with ’’Special design features’’ Code A
Order code:
1 2 3 456 789 1011
UT-11 - A - - - S -

C
Universal Pressure Transmitter UniTrans
Appendix
Publishe d 16.2.04
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-W iegand-Str. · 63911 K lingenber g · (09372) 1 32 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
81
9.3 Warranty Conditions
The pressure transmitter has a 24 month warrant
in accordance with the WIKA
General Terms of Deliver
.
9.4 Glossary
Adjustment Allocation of the si
nal output range (4 ... 20 mA) to the desired
e or level measurement range.
ration Also damping: timely communication of the measurement sig-
nal; rise time of the current output si
Inversion Conversion of the output si
nal from 4 ... 20 mA to 20 ... 4 mA
Nom. pressure ran
e The operating pressure range for which the sensor was de-
ned
Zero point Start of the pressure measurement ran
e
Parameterization Also confi
uration: programming of the relevant parameters
and the pressure measurement ran
e specific to the application
and measurement location.
Span The pro
rammed pressure measurement range
hest pressure value of the programmed measurement
span (end-point of the span)
Tank linearization Determination of approximate volume/pressure ratio values
with non-linear correlations based on var
For example, a non-linear correlation exists between the fill level and the volume in spherical containers. Durin
linearization,
the non-linear volume is assi
ned the 4 ... 20 mA output signal
from a table of values (proximit
process by means of up to 32
support points).
Defaults The sensor parameters are pre-pro
turer
9.5 Units of Pressure Measurement
1 atm (atmospheres) = 760 mm H
= 760 Torr
= 1.033 kp/cm
2
= 0.1013 MPa
1 Torr = 133.3 Pa
1 kp/mm
2
= 9.81 N/mm2 = 9.81 MPa
1 bar = 0.1 MPa
1 mbar = 1 hPa (Hektopascal)
1 psi (pound per square inch) = 6.895 · 10
3
Pa
1 bar = 33.5 feet of water
1 PA = 1.0 x 10
-5
bar
1 mmHG = 1.333 mbar
Repairs may only be carried out by the manufacturer. All other repairs and
device modifications are unauthorized and will void the warranty.
Attention