Appendix 1: EC Declaration of conformity model UPT-2x ..................63
EN
Appendix 2: Menu tree, basic setting...................................64
Appendix 3: Menu tree, display........................................66
Appendix 4: Menu tree, diagnositic ....................................68
Appendix 5: Menu tree, detail setup....................................69
Appendix 6: Menu tree, info...........................................70
Declarations of conformity can be found online at www.wika.com
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
5
EN
1. General information
1. General information
■
The process transmitter described in the operating instructions has been designed
and manufactured using state-of-the-art technology. All components are subject
to stringent quality and environmental criteria during production. Our management
systems are certied to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on handling the instrument. Working safely requires that all safety instructions and work instructions are
observed.
■
Observe the relevant local accident prevention regulations and general safety regulations for the instrument's range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
■
The manufacturer's liability is void in the case of any damage caused by using the
product contrary to its intended use, non-compliance with these operating instructions, assignment of insuciently qualied skilled personnel or unauthorised modications to the instrument.
■
The general terms and conditions contained in the sales documentation shall apply.
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
2. Design and function
2. Design and function
2.1 Design
EN
Process connection, thread
Process connection, spanner ats
Sensor housing
Ex-relevant data
Case head
2.2 Description
The process transmitter processes the prevailing pressure and converts it into a current
signal. This current signal can be used for the evaluation, control and monitoring of the
process.
Push-on cap
Product label
Ground screw, outside
Electrical connection, cable gland
Second bore for cable gland (delivered sealed
with blind plug)
®
HART
(option)
The instrument version with HART
Measuring range scaling (turndown)
The start and end of the measuring range can be set within dened ranges.
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
®
can communicate with a controller (master).
7
2. Design and function
Display and operating unit (accessory)
The display and operating unit model DI-PT-U has a main and an additional display.
The main display and the additional display are able to be set in almost any way. In the
factory setting, the main display shows the pressure value of the output signal.
EN
The process transmitter can be congured via the display and operating unit.
Adaptable to mounting position
The process transmitter is tted with a case head which can be turned through 330°.
The display and operating unit can be attached in 90° steps. Thus the measured value
can be read irrespective of the mounting position.
Rotatable case head
330°
Displaceable display and operating unit
2.3 Scope of delivery
■
Pre-assembled process transmitter
■
Ordered accessories
■
Operating instructions
■
Measured value protocol
Cross-check scope of delivery with delivery note.
8
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
3. Safety
3. Safety
3.1 Explanation of symbols
DANGER!
... indicates a potentially dangerous situation in the hazardous area that can
result in serious injury or death, if not avoided.
WARNING!
... indicates a potentially dangerous situation that can result in serious injury
or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries
or damage to equipment or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for ecient and
trouble-free operation.
3.2 Intended use
The intrinsically safe process transmitter measures gauge pressure, absolute pressure
and vacuum. The physical quantity pressure is converted into an electrical signal.
EN
Only use the intrinsically safe process transmitter in applications that lie within its technical performance limits (e.g. max. ambient temperature, material compatibility, ignition
protection type, ...). Instruments with a ush process connection must not be used with
media which might damage the diaphragm of the process connection.
→ Performance limits see chapter 15 “Specications”.
ATEX and IECEx approval
Process transmitter approved for use in hazardous areas, type tested in accordance
with the following standards:
IEC 60079-0:2011 + Corr.1:2012 + Corr.2:2013 / EN 60079-0:2012 + A11:2013
IEC 60079-11:2011 + Corr.1:2012 / EN 60079-11:2012
IEC 60079-26:2006 + Corr.1:2009 / EN 60079-26:2007 + Corr.1:2009
Approval ratings ATEX and IECEx
■
Gases and mistMounting to zone 0 (EPL Ga/Gb)
Installation in zone 1 (EPL Gb) and zone 2 (EPL Gc)
■
Dusts
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
Mounting to zone 20 (EPL Da/Db); installation in zone 21 (EPL Db)
9
3. Safety
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
The manufacturer shall not be liable for claims of any type based on operation contrary
to the intended use.
EN
3.2.1 Special conditions for safe use
For the case “mounting to zone 0 (EPL Ga/Gb)”
The instrument's internal separation (diaphragm) from the wetted area has a wall thickness of < 0.2 mm in order to ensure proper functioning. In operation, it must be ensured
that any impairment of the separation, e.g. by aggressive media or through mechanical
damage, is eliminated.
3.3 Personnel qualication
WARNING!
Risk of injury should qualication be insucient!
Improper handling can result in considerable injury and damage to equipment.
▶
The activities described in these operating instructions may only be
carried out by skilled personnel who have the qualifications described
below.
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training,
knowledge of measurement and control technology and on their experience and knowl-
edge of country-specic regulations, current standards and directives, are capable of
carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive
media, compatibility of materials.
Special knowledge for working with instruments for hazardous areas:
The skilled personnel must have knowledge of ignition protection types, regulations and
provisions for equipment in hazardous areas.
10
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
3. Safety
3.4 Risks in hazardous areas
Observe the information given in the type examination certicate and the country-specic
regulations for installation and use in hazardous areas (e.g. IEC 60079-14, NEC, CEC). If
this is not observed, serious injuries and damage could occur.
Do not commission any damaged instruments within hazardous areas. Damaged instruments cannot reliably guarantee any explosion protection. Obvious damage must be
reported immediately.
3.5 Handling of hazardous media
For hazardous media such as oxygen, acetylene, ammable or toxic gases and liquids
and refrigeration plants, compressors, etc., in addition to all standard regulations, the
appropriate existing codes or regulations must also be followed.
Residual media in the dismounted process transmitter can result in a risk to personnel,
the environment and equipment. Take sucient precautionary measures.
3.6 Labelling, safety marks
Product label
EN
Pin assignment
Power supply
Output signal
Measuring range
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
Model
Model code
Date of manufacture YYYY-MM
S# Serial no.
P# Product no.
Permissible ambient temperature
Approval ratings
11
3. Safety
Symbols
Before mounting and commissioning the instrument, ensure you read the
operating instructions!
EN
Instruments bearing this mark comply with the relevant European directives.
ATEX European Explosion Protection Directive
(Atmosphère = AT, explosible = EX)
Instruments bearing this mark comply with the requirements of the
European directive 94/9/EC (ATEX) on explosion protection.
Output signal
Power supply
12
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
4. Transport, packaging and storage
4. Transport, packaging and storage
4.1 Transport
Check the process transmitter for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
4.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, return for calibration).
4.3 Storage
Permissible conditions at the place of storage:
■
Storage temperature: -40 ... +80 °C
■
Humidity: 35 ... 93 % relative humidity (no condensation)
Avoid exposure to the following factors:
■
Proximity to hot objects, when permissible storage temperature is exceeded by
radiation.
■
Mechanical vibration, mechanical shock (putting it down hard), when the permissible
values are exceeded, see chapter 15 “Specications”.
■
Soot, vapour, dust and corrosive gases
■
Hazardous areas and ammable atmospheres where the instruments are not suitable
for installation in or mounting to equipment in explosive atmospheres.
EN
Store the process transmitter in its original packaging in a location that fulls the conditions listed above. If the original packaging is not available, then store the instrument
in a container that is similar to the original packaging, so that the instrument cannot be
scratched and is protected against damage if dropped.
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
13
5. Commissioning, operation
5. Commissioning, operation
The process transmitter should only be commissioned and operated by skilled personnel.
The skilled personnel must have knowledge of ignition protection types, regulations and
provisions for equipment in hazardous areas.
EN
5.1 Mechanical mounting
DANGER!
Danger to life from explosion
Incorrect mounting and non-compliance with the contents of this chapter
can lead to a risk of explosion.
▶
Carefully read and comply with the following sub-chapters.
5.1.1 Requirements for mounting point
The process transmitter can be adjusted to the installation site.
→ see chapter 2.2 “Description”
■
Sucient space for a safe electrical installation.
■
Operating elements can be accessed following the mounting.
■
Permissible ambient and medium temperatures remain within the temperature
classes.
■
Consider possible restrictions on the ambient temperature range caused by mating
connector used.
■
Protect the process transmitter from heat sources (e.g. pipes or tanks).
Additional for instruments with cooling element:
■
Mount the process transmitter as horizontally as possible and ensure an unhindered
air circulation at the cooling element.
■
The cooling element should have as little soiling as possible, otherwise the cooling
action cannot be guaranteed. Ensure as much space as possible so that the cooling
element can be cleaned.
14
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
5. Commissioning, operation
5.1.2 Mounting the process transmitter
DANGER!
Damaged diaphragm of the ush process connection
If the diaphragm is damaged, the explosion protection is no longer guaranteed. Through any explosion resulting from this, there will be a high danger
to life.
▶
Before commissioning, the diaphragm should be checked for visible
damage. Leaking fluid is indicative of damage.
▶
Protect the diaphragm from contact with abrasive media and against any
impacts.
Sealing
Parallel threads
Seal the sealing face with at gaskets,
sealing rings or WIKA prole sealings.
Tapered threads
Wrap threads with sealing material,
e.g. PTFE tape.
EN
per EN 837per DIN 3852-E
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
NPT, R and PT
15
5. Commissioning, operation
Screwing in
CAUTION!
Improper mounting
Process transmitter is damaged.
▶
EN
Screw the process transmitter into the mounting location with a spanner using the
spanner ats.
The tightening torque depends on the dimensions of the process connection and the
gasket used (form/material).
Tighten the process transmitter using the spanner flats.
▶
Do not tighten using the sensor housing or the case head.
▶
Use a suitable open-ended spanner.
▶
Do not cross the threads.
16
Spanner ats
Sensor housing
Case head
For information on tapped holes and welding sockets, see Technical information IN 00.14 at www.wika.com.
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
5. Commissioning, operation
5.2 Electrical installation
DANGER!
Danger to life from explosion
Incorrect mounting and non-compliance with the contents of this chapter
can lead to a risk of explosion.
▶
Carefully read and comply with the following sub-chapters.
Observe the information given in the type examination certicate and the country-specic
regulations for installation and use in hazardous areas (e.g. IEC 60079-14, NEC, CEC)
this is not observed, serious injuries and damage could occur.
→ Type examination certicate is available for download at www.wika.com.
5.2.1 Safety instructions
■
Only connect with the voltage disconnected.
■
If any overvoltage is expected, install surge protection devices.
■
Exposed cables must not run close to bare metal components. Maintain a minimum
clearance of 5 mm.
■
Ensure that the cables are properly installed and the cable gland or plug connections
are securely closed and sealed.
. If
EN
5.2.2 Requirements for connection cable
■
Use and assemble connection cable that is suitable for the application. For cables
with exible wires, always use ferrules appropriate for the wire cross-section.
■
Where there is electromagnetic radiation above the test values per EN 61326, a
shielded connection cable must be used.
■
When using an M12 x 1 (4-pin) circular connector, the mating connector is supplied
by the customer. Ensure the matching design from the connector manufacturer.
Electrical connections
Cable gland M20 x 1.5 and spring-loaded
terminals
Angular connector DIN 175301-803A with
mating connector
Circular connector M12 x 1 (4-pin) without
mating connector
WIKA operating instructions process transmitter, model UPT-2x
Ingress protection: IP 66/67
Cable diameter: 7 ... 12 mm
Wire cross-section max. 2.5 mm
Single cable: 0.13 ... 2.5 mm²
End splices: 0.13 ... 1.5 mm²
For cable diameters outside of 7 ... 12 mm,
change the seal and cable gland
Ingress protection: IP 65
Cable diameter: 6 ... 8 mm
Wire cross-section: max. 1.5 mm²
Ingress protection: IP 65
Observe manufacturer's specications
2
(AWG 14)
17
5. Commissioning, operation
5.2.3 Requirements regarding intrinsic voltage supply
■
Power the process transmitter via an intrinsically safe circuit (Ex ia). Both the internal
capacitance and inductance must be considered (→ see chapter 15 “Specications”).
■
Provide the required voltage supply separation between Ex and non-Ex areas
with a certied isolated barrier or Zener barrier (suitable isolated barrier, model
EN
KFD2-STC4-Ex1).
■
For applications that require EPL Gb or Db, the power supply and signal circuit should
have a protection level of “ib”. Then the interconnections and the transmitter will have
a protection level of II 2G Ex ib IIC T4/T5/T6 Gb or II 2D Ex ib IIIC T4/T5/T6 Db, even if
the process transmitter is marked otherwise (→ see EN 60079-14 section 5.4).
5.2.4 Opening the case
CAUTION!
Ingress of moisture
Moisture can destroy the process transmitter.
▶
Protect the opened process transmitter against moisture.
▶
Screw off the case head cover by hand and pull out the display and operating unit or
push-on cap.
5.2.5 Shielding and grounding
The process transmitter must be shielded and grounded in accordance with the grounding concept of the plant.
▶
Connect the cable shield with the equipotential bonding.
▶
Connect the process connection or the external grounding screw with the equipotential bonding
18
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
5. Commissioning, operation
5.2.6 Connection
1. Pass the connection cable through the cable gland and connect it.
Ensure that no moisture can enter at the cable end.
→ For pin assignment see chapter 5.2.7 “pin assignments”.
2. Tighten the cable gland.
■
Recommended tightening torque 1.5 Nm
■
Check that the seals are correctly seated in order to guarantee the ingress
protection.
3. Perform a mounting correction.
→ Without display, see chapter 7.1 “Performing a mounting correction (oset)”
→ Via HART
®
, see chapter 7.2 “Conguring the HART® interface”
→ With display, see chapter 8.5 “Mounting correction (oset)”
4. Attach the push-on cap or display and operating unit and screw the case head cover
tight down to the stop.
EN
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
19
5. Commissioning, operation
1
2
3
4
3
1
2
5.2.7 Pin assignments
Spring-loaded terminal
Pin assignmentOutlet for connection cable
EN
Cable gland
Process connection
Positive power supply terminal +
Angular connector DIN 175301-803 A
+1
Negative power supply terminal –
Ground screw, inside (GND)
Circular connector M12 x 1 (4-pin)
+1
–2
ShieldGND
The shield connection is located on the inside of the instrument.
20
WIKA operating instructions process transmitter, model UPT-2x
–3
Shield4
14111166.02 04/2015 EN/DE
6. Display and operating unit, model DI-PT-U
6. Display and operating unit, model DI-PT-U
6.1 Design and description
The display and operating unit model DI-PT-U is available as an accessory.
It can be plugged into the instrument electronics at 90° increments. Thus the display can
be read, whether the process transmitter is mounted laterally or upside down.
Description
Direction key [▲]
Escape key [ESC]
Additional display
Unit
Main display
Trend indication
Bar graph with over/under limit
arrows and drag pointer function
Enter key [↵]
Direction key [▼]
EN
6.2 Accessing/exiting the operating menu
Accessing: Press [↵].
Exiting: Press [ESC] repeatedly until the menu has been exited.
If after 3 min. no entry is made, the menu will automatically be exited and the
last set display mode will be activated.
If there is an invalid entry, the message “Input error” will show in the display
for 2 seconds, and the previous menu will be accessed.
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
21
6. Display and operating unit, model DI-PT-U
6.3 Installation/Removal
CAUTION!
Ingress of moisture.
Moisture can destroy the process transmitter.
▶
EN
1. Screw o the case head cover by hand.
2. Installation
Pull out the push-on cap and attach the display and
operating unit into any of the locking positions (0°, 90°,
180°, 270°).
Removal
Pull out the display and operating unit and attach the
push-on cap.
Protect the opened process transmitter against moisture.
▶
Close the case head tightly.
3. Screw on the case head cover.
Ensure that the case head is tightly closed.
22
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
6. Display and operating unit, model DI-PT-U
6.4 Setting the main display
The main display can indicate the following values:
■
PressureApplied pressure is displayed.
■
LevelLevel is displayed.
■
VolumeVolume is displayed.
■
CurrentOutput signal is displayed.
■
PV percentOutput signal is displayed as a percentage.
■
Sensor temperature Temperature at the sensor is displayed.
■
PV (primary value) The value corresponding to the mode will be displayed.
If the mode is changed, then the main display will change.
1. Open the operating menu with [↵].
Select “Display” and conrm with [↵].
2. Select “Main display” and conrm with [↵].
EN
3. Select value and conrm with [↵].
» Main display indicates the selected value.
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
23
6. Display and operating unit, model DI-PT-U
6.5 Setting the additional display
The additional display can indicate the following values:
■
PressureApplied pressure is displayed.
■
EN
LevelLevel is displayed.
■
VolumeVolume is displayed.
■
CurrentOutput signal is displayed.
■
PV percentOutput signal is displayed as a percentage.
■
Sensor temperature Temperature at the sensor is displayed.
■
PV (primary value) The value corresponding to the mode will be displayed.
If the mode is changed, then the main display will change.
Drag pointer values
■
P
/ P
min
PV
T
min
min
/ T
max
/ PV
max
max
■
■
Further data
■
TAG short (max. 8 capital letters and gures)
■
TAG long (max. 32 alphanumeric characters)
■
Empty (additional display switched o)
1. Open the operating menu with [↵].
Select “Display” and conrm with [↵].
2. Select “Add. display” and conrm with [↵].
3. Select value and conrm with [↵].
» Additional display indicates the selected value.
24
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
7. Conguration without display and operating unit
7. Conguration without display and operating unit
7.1 Performing a mounting correction (oset)
The mounting correction corrects a zero oset in the output signal by dening a new
zero point. The zero oset is caused by the mounting position.
Correction range: ±20 % of maximum measuring range
Required tool: Multimeter (ammeter)
1. Screw o the case head cover and pull out the push-on cap.
2. Press [Zero] for approx. 2 s (e.g. with measuring tip of a multimeter).
» Mounting correction successful: Control diode lights up for 2 s.
» Mounting correction unsuccessful: Control diode blinks 5 times.
EN
3. Check the output signal as follows.
CAUTION!
Incorrect connection
A short circuit will destroy the process transmitter.
▶
Ensure that the multimeter does not come into contact with the
positive power supply terminal.
▶
Set the multimeter to current measurement.
▶
Connect the positive measuring line of the multimeter to the test contact.
▶
Connect the negative measuring line of the multimeter to the negative power
supply terminal.
» The result of the current measurement should give a value between 4 ... 20 mA
in the pressure-free condition. If the ambient atmospheric pressure is outside
the measuring range, the measured current can be < 4 mA or > 20mA.
Control diode
Mounting correction, [Zero] key
Test contact
Positive power supply terminal +
Negative power supply terminal –
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
25
7. Conguration without display and operating unit
7.2 Conguring over HART® interface
®
HART
-compatible process transmitters can be operated and congured with operat-
ing software (e.g. PACTware
®
), process control software (e.g. AMS or Simatic PDM) or
a hand-held device (e.g. FC475 from Emerson).
The operation of the respective menus is described in the associated online help.
EN
The device drivers are available for download from www.wika.com.
®
Connecting process transmitter to PC (HART
)
Any work should only be carried out in a safe area.
1. Connect HART
2. Connect HART
Ex-BereichSicherer Bereich
Hazardous areaNon-hazardous area
®
modem to process transmitter.
®
modem to PC or notebook..
HART-Widerstand 250 ohm
HART resistor 250 ohm
HART-Modem
HART modem
RS-232
=
~
Spannungsquelle
Voltage source
26
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
8. Conguration via display and operating unit
8. Conguration via display and operating unit
8.1 Conguring the measuring task
8.1.1 Conguring pressure measurement
1. Open the operating menu with [↵].
Select “Basic setting” and conrm with [↵].
2. Select “Application” and conrm with [↵].
3. Select “Pressure” and conrm with [↵].
4. Select “Unit” and conrm with [↵].
5. Select pressure unit and conrm with [↵].
» Pressure unit is set.
6. Go back one menu level using [ESC].
Select “Mode” and conrm with [↵].
EN
7. Select “Pressure” and conrm with [↵].
» Mode is set.
8. Scale the measuring range.
→ See chapter 8.3 “Scaling the measuring range”.
9. Perform a mounting correction.
→ See chapter 8.5 “Mounting correction (oset)”.
» Pressure measurement is congured.
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
27
8. Conguration via display and operating unit
8.1.2 Conguring level measurement
Requirement
■
Length unit for the lling height is known
■
The density of the medium is known
1. Open the operating menu with [↵].
EN
Select “Basic setting” and conrm with [↵].
2. Select “Application” and conrm with [↵].
3. Select “Level” and conrm with [↵].
4. Select “Unit” and conrm with [↵].
5. Select length unit and conrm with [↵].
» Length unit is set.
6. Select “Density” and conrm with [↵].
7. Select “Density unit” and conrm with [↵].
8. Select density unit and conrm with [↵].
» Density unit is set.
9. Select “Density value” and conrm with [↵].
10. Set digit using [▲] [▼] and conrm with [↵].
» The cursor moves to the next digit.
» Repeat for each digit.
» Density value is set.
11. Go back two menu levels using [ESC].
Select “Mode” and conrm with [↵].
12. Select “Level” and conrm with [↵].
» Mode is set.
13. Perform a mounting correction.
→ See chapter 8.5 “Mounting correction (oset)”.
» Level measurement is congured.
28
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
8. Conguration via display and operating unit
8.1.3 Conguring the volume measurement
Requirement
■
Length unit for the lling height is known
■
Medium density is known
■
Characteristic curve of the tank is known (→ see chapter 8.1.4
“Characteristic curves”)
1. Open the operating menu with [↵].
Select “Basic setting” and conrm with [↵].
2. Select “Application” and conrm with [↵].
3. Select “Level” and conrm with [↵].
4. Select “Unit” and conrm with [↵].
5. Select length unit and conrm with [↵].
» Length unit is set.
6. Select “Density” and conrm with [↵].
7. Select “Density unit” and conrm with [↵].
EN
8. Select density unit and conrm with [↵].
» Density unit is set.
9. Select “Density value” and conrm with [↵].
10. Setting the density of the medium.
Set digits using [▲] [▼] and conrm with [↵].
» The cursor moves to the next digit.
» Repeat for each digit.
» Density value is set.
11. Go back two menu levels using [ESC].
Select “Volume” and conrm with [↵].
12. Select “Scale in” and conrm with [↵].
14111166.02 04/2015 EN/DE
WIKA operating instructions process transmitter, model UPT-2x
29
8. Conguration via display and operating unit
13. Select “Low” and conrm with [↵].
14. Select “change” and conrm with [↵].
EN
15. Setting the start of the measuring range referred to the lling
height of the tank.
Set digits using [▲] [▼] and conrm with [↵].
» The cursor moves to the next digit.
» Repeat for each digit.
» Start of the measuring range is set.
16. Go back one menu level using [ESC].
Select “High” and conrm with [↵].
17. Select “change” and conrm with [↵].
18. Setting the end of the measuring range referred to the lling
height of the tank.
Set digits using [▲] [▼] and conrm with [↵].
» The cursor moves to the next digit.
» Repeat for each digit.
» End of the measuring range is set.
19. Go back two menu levels using [ESC].
Select “Characteristic” and conrm with [↵].
20. Select characteristic curve and conrm with [↵].
→ Explanation of characteristic curves, see chapter 8.1.4
“Characteristic curves”
21. Select “Scale out” and conrm with [↵].
22. Select “Unit” and conrm with [↵].
23. Select volume unit and conrm with [↵].
■
Volume unit: Standard units (e.g. litres, m³, ...)
■
Free input: Freely denable unit (selectable under
“Volume unit“)
» Volume unit is set.
30
WIKA operating instructions process transmitter, model UPT-2x
14111166.02 04/2015 EN/DE
Loading...
+ 110 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.