ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Lire le mode d'emploi avant de commencer toute opération !
A conserver pour une utilisation ultérieure !
¡Leer el manual de instrucciones antes de comenzar cualquier trabajo!
¡Guardar el manual para una eventual consulta!
WIKA operating instructions model DIH102
Page
Página
47 - 68
69 - 90
14089810.01 03/2014 GB/D/F/E
Contents
Contents
1. General information4
2. Safety6
3. Specications10
4. Design and function12
5. Transport, packaging and storage15
6. Commissioning, operation16
7. Maintenance and cleaning20
8. Faults21
9. Dismounting, return and disposal22
Appendix: EC declaration of conformity24
Declarations of conformity can be found online at www.wika.com.
GB
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH103
1. General information
1. General information
■
The instrument described in these operating instructions has been
GB
designed and manufactured using state-of-the-art technology. All
components are subject to stringent quality and environmental
criteria during production. Our management systems are certied to
ISO 9001 and ISO 14001.
■
These operating instructions contain important information on
handling the instrument. Working safely requires that all safety
instructions and work instructions are observed.
■
Observe the relevant local accident prevention regulations and
general safety regulations for the instrument's range of use.
■
The operating instructions are part of the product and must be kept
in the immediate vicinity of the instrument and readily accessible to
skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the
operating instructions prior to beginning any work.
■
The manufacturer's liability is void in the case of any damage
caused by using the product contrary to its intended use,
non-compliance with these operating instructions, assignment
of insuciently qualied skilled personnel or unauthorised
modications to the instrument.
■
The general terms and conditions contained in the sales
documentation shall apply.
... indicates a potentially dangerous situation that can result
in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can
result in light injuries or damage to equipment or the
environment, if not avoided.
Information
... points out useful tips, recommendations and information
for ecient and trouble-free operation.
DANGER!
... identies hazards caused by electrical power. Should
the safety instructions not be observed, there is a risk of
serious or fatal injury.
WARNING!
... indicates a potentially dangerous situation that can result
in burns, caused by hot surfaces or liquids, if not avoided.
GB
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH105
2. Safety
2. Safety
GB
2.1 Intended use
The model DIH10 digital indicator is an all-purpose, configurable
loop-powered indicator for single resistance thermometers (RTD),
thermocouples (TC) and resistance and sensor transmitters.
It is especially applicable where a standard signal of 4 ... 20 mA is
already available. The current-loop display is built into an appropriate
connection head or wall-mounted housing, and can thus be used as a
local display in all sectors of industry.
The instrument has been designed and built solely for the intended
use described here, and may only be used accordingly.
The technical specications contained in these operating instructions
must be observed. Improper handling or operation of the instrument
outside of its technical specications requires the instrument to be
taken out of service immediately and inspected by an authorised
WIKA service engineer.
If the instrument is transported from a cold into a warm environment,
the formation of condensation may result in instrument malfunction.
Before putting it back into operation, wait for the instrument
temperature and the room temperature to equalise.
WARNING!
Before mounting, commissioning and operation, make sure
that the display is suitable for the application.
Non-observance can result in serious injury and/or damage
to the equipment.
Further important safety instructions can be found in the
individual chapters of these operating instructions.
6WIKA operating instructions model DIH10
14089810.01 03/2014 GB/D/F/E
2. Safety
The manufacturer shall not be liable for claims of any type based on
operation contrary to the intended use.
2.2 Personnel qualication
WARNING!
Risk of injury should qualication be insucient!
Improper handling can result in considerable injury and
damage to equipment.
■
The activities described in these operating instructions
may only be carried out by skilled personnel who have
the qualications described below.
■
Keep unqualied personnel away from hazardous areas.
Skilled personnel
Skilled personnel are understood to be personnel who, based on
their technical training, knowledge of measurement and control
technology and on their experience and knowledge of country-specic
regulations, current standards and directives, are capable of carrying
out the work described and independently recognising potential
hazards.
Special operating conditions require further appropriate knowledge,
e.g. of aggressive media.
2.3 Special hazards
WARNING!
For hazardous media such as oxygen, acetylene,
ammable or toxic gases or liquids, and refrigeration
plants, compressors, etc., in addition to all standard
regulations, the appropriate existing codes or regulations
must also be followed.
GB
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH107
GB
2. Safety
WARNING!
Protection from electrostatic discharge (ESD) required!
The proper use of grounded work surfaces and personal
wrist straps is required when working with exposed circuitry
(printed circuit boards), in order to prevent static discharge
from damaging sensitive electronic components.
To ensure safe working on the instrument, the operating
company must ensure
DANGER!
Danger of death caused by electric current
Upon contact with live parts, there is a direct danger of
death.
WARNING!
Residual media in dismounted instruments can result in
a risk to persons, the environment and equipment. Take
sucient precautionary measures.
■
that suitable rst-aid equipment is available and aid is
provided whenever required.
■
that the operating personnel are regularly instructed
in all topics regarding work safety, rst aid and
environmental protection and know the operating
instructions and in particular, the safety instructions
contained therein.
■
Electrical instruments may only be installed and
connected by skilled electrical personnel.
■
Operation using a defective power supply unit (e.g. short
circuit from the mains voltage to the output voltage) can
result in life-threatening voltages at the instrument!
Do not use this instrument in safety or emergency stop
devices. Incorrect use of the instrument can result in injury.
8WIKA operating instructions model DIH10
14089810.01 03/2014 GB/D/F/E
2. Safety
Should a failure occur, aggressive media with extremely
high temperature and under high pressure or vacuum may
be present at the instrument.
2.4 Labelling, safety marking
■
Product labels when tted into a connection head or housing
- Head/housing labelling
Year of manufacture
- Head/housing product label
■
Display module product label
Model
1
GB
Model
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH109
2
3
4
Year of manufacture
2. Safety / 3. Specications
Explanation of symbols
GB
Before mounting and commissioning the instrument,
ensure you read the operating instructions!
CE, Communauté Européenne
Instruments bearing this mark comply with the relevant
European directives.
3. Specications
SpecicationsModel DIH10 (display module)
Input
Current loop4 ... 20 mA
Input impedance Ri at 20 mA < 160 Ω (U ≤ 3.2 V)
Accuracy
Resolution-999 ... 9999 digit
Measuring error±0.2 % of measuring range, ±1 digit
Temperature drift100 ppm/K
Display
PrincipleLED, rotatable in 90° steps
Measured value of display7 segment, 8 mm high, red, 4-digit =
Overrange/underrangeto HI / to LO
Display rate0.1 s - 1 s - 10 s (adjustable)
Parameters
Maximum voltage U
Ambient conditions
Operating temperature0 ... 80 °C
Storage temperature-20 ... +80 °C
i
Display 9999
DC 30 V
10WIKA operating instructions model DIH10
14089810.01 03/2014 GB/D/F/E
Ø 62,5
3. Specications
Module housing
DimensionsØ 43.0 / 48.5 / 62.5 mm x 37 mm
Housing materialPolycarbonate
Ingress protectionFront: IP 67
Weightapprox. 55 g
Electrical connectionPlug-in terminal strip, up to 1.5 mm²
Programmable featuresScale range, display time, decimal
Housing designs
(see chapter 4.2)
CE conformity
EMC directive
For further specications see WIKA data sheet AC 80.11 and order
documentation.
Dimensions in mm for the model DIH10 display module
Connection: IP 20 or IP 00
open cable ends
point, unit (°C/°F), zero point
stabilisation, programming lock,
reference points, TAG number
■
Connection head (standard)
■
Wall-mount housing
■
Housing with pipe support
2004/108/EC, EN 61326 emission
(group 1, class A) and interference
immunity (industrial application)
Ø 43
Ø 48,5
GB
43 mm
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH1011
40 mm
4. Design and function
4. Design and function
4.1 Description
GB
The model DIH10 is a universal, microprocessor-controlled display,
monitoring and control instrument. It has an input, to which 4 … 20 mA
standard signals can be connected.
With its simple mounting and open programmability via the easily
accessible push-buttons, it can also be retrotted quickly to a pressure
transmitter which is already operating.
The display is loop-powered, and so no additional voltage supply is
needed. The 8 mm high LED display ensures good readability, even in
poor lighting conditions.
Model DIH10 display module, operating side
Connector pins for factory programming
▲
▼
P
12WIKA operating instructions model DIH10
Buttons for programming
Coupler connector
14089810.01 03/2014 GB/D/F/E
4. Design and function
4.2 Housing designs
4.2.1 BSZ-H connection head (standard design), DIH10-Z-*A
Weight approx. 350 g
SpecicationsStandard version
Connection to thermometer dM24 x 1.5 / ½ NPT
MaterialAluminium black, lacquered
Cable gland1 x M20 x 1.5 / 1 x ½ NPT
Ingress protectionIP 65 (option: IP 67)
4.2.2 Wall-mount housing, DIH10-Z-*C
BSZ-H upper body special version
1)
GB
14053621.0114053621.01
Weight approx. 500 g
SpecicationsStandard version
MaterialAluminium black, lacquered
Lower bodyAluminium black, lacquered
Cable gland2 x M16 x 1.5
Ingress protectionIP 65 (option: IP 67)
1) RAL 5022
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH1013
1)
1)
95
20
25
136
91
20
2x M16x1,5
95
20
25
136
91
4. Design and function
4.2.3 Housing with pipe support, DIH10-Z-*D
BSZ-H upper body special version
GB
95
SpecicationsStandard version
MaterialAluminium black, lacquered
Lower bodyAluminium black, lacquered
Pipe supportStainless steel
Pipe diameter R1" ... 2"
Cable gland2 x M16 x 1.5
Ingress protectionIP 65 (option: IP 67)
1) RAL 5022
4.3 Scope of delivery
Cross-check scope of delivery with delivery note.
25
2x M16x1,5
Weight approx. 950 g
136
91
1)
1)
14078928.01
14WIKA operating instructions model DIH10
14089810.01 03/2014 GB/D/F/E
5. Transport, packaging and storage
5. Transport, packaging and storage
5.1 Transport
Check the instrument for any damage that may have been caused by
transport. Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during
transport (e.g. change in installation site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
■
Storage temperature: -20 ... +80 °C
■
Humidity: 35 ... 85 % relative humidity (no condensation)
Avoid exposure to the following factors:
■
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
Store the instrument in its original packaging in a location that fulls
the conditions listed above. If the original packaging is not available,
pack and store the instrument as described below:
1. Wrap the instrument in an antistatic plastic lm.
2. Place the instrument along with shock-absorbent material in the
packaging.
3. If stored for a prolonged period of time (more than 30 days), place a
bag containing a desiccant inside the packaging.
WARNING!
Before storing the instrument (following operation), remove
any residual media. This is of particular importance if
the medium is hazardous to health, e.g. caustic, toxic,
carcinogenic, radioactive, etc.
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH1015
GB
6. Commissioning, operation
6. Commissioning, operation
GB
WARNING!
■
Never connect directly to a voltage supply (e.g. 24 V) as
this will damage the display.
■
Only use a transmitter that will not exceed the
permissible electrical limit values.
6.1 Operating a current loop
Power
supply
unit
DC 24 V
WARNING!
The display is only ever operated in a 4 ... 20 mA current
loop. A direct connection to a voltage supply (e.g. DC 24 V)
will lead to the display being damaged and will invalidate
the warranty.
For a functional check, the display must be powered from a
mA source (e.g. mA transmitter or calibrator).
In normal operation, the display is connected in series with
a 4 ... 20 mA transmitter or connected to the 4 ... 20 mA
analogue output of an instrument.
internal
bridge
DIH10
display
Housing
16WIKA operating instructions model DIH10
14089810.01 03/2014 GB/D/F/E
6. Commissioning, operation
6.2 Model DIH10 connection
Coupler connector
Internal bridge
2341
+
-
+-
-
Transmitter
+
Power
supply
unit
The 4-pin plug-in terminal strip for the connection of the current-loop display is
located on the rear of the display and is accessible once the housing is open.
Terminal 1: Positive terminal of the current loop
Terminal 2: Negative terminal of the current loop
Terminals 3 and 4 are connected together on the PCB and enable the conven-
ient connection of a transmitter built into the connection head.
GB
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH1017
6. Commissioning, operation
6.3 Programming the display
6.3.1 Programming table for programming the display
GB
PNDescriptionRangeFactory
0Calibration mode
0 = Sensor calibration (signal must be
applied, only for factory setting)
1 = Rescaling (display value at 4 ... 20 mA)
1Full scale value (programming the display
value at 20 mA, e.g. 600)
2Initial value (programming the display value
at 4 mA, e.g. 100)
3Decimal point or unit selection
(for programming a unit, the value shifts to
the left)
4
Display and averaging time in 1/10 of a second
5 **Zero point stabilisation (the ± range in which
0000 will be displayed)
50 *** Denition of PIN code > 0000 (for
programming lock)
51 *** Authentication by PIN code (removal of
programming lock)
100Number of reference points (reference points
for sensor calibration, reference points
reduce the measuring rate)
101 ...
Reference points (as programmed under
130
PN100)
200
TAG number
* For factory setting
** For programming a value >1 a hysteresis of 0.1 % will be activated. This prevents
the display jumping.
*** optional
(If no PIN code has been dened, PN50/PN51 are not available. A PIN code can
only be initially programmed via the serial interface for the factory programming. If a
PIN code has been dened (display after segment test: Pin), the programming lock
must be released, using the PIN code dened via PN50, through PN51 under input,
as otherwise, any changes made under the other PN's will not be able to be saved.
In order to reactivate the PIN code via PN50 on completion of any programming,
the programming mode must be left (by the input of 0000 under PN50 or by briey
interrupting the current loop).
18WIKA operating instructions model DIH10
0/11
-999...9999 150
-999...9999 0
0 / 0.0 / 0.00
/ 0.000 /
°F / °C
5.0 ... 100.0 10
0 ... 1002
0000...9999 0000
0000...9999 0000
0 ... 300
-999 ... 9999 0
0000 ... 9999 0
default *
°C
14089810.01 03/2014 GB/D/F/E
6. Commissioning, operation
6.3.2 Programming the display
1. Connect instrument in accordance with connection diagram
2. Switch on the current loop supply (current between 4 ... 20 mA)
An initialisation and segment test is carried out. Then DIH10 and
subsequently the rmware version number (e.g. F1.24) is displayed.
Following this, the display switches into operating mode.
3. Press the program button P, the program number P 0 will be
displayed
4. By pressing P and the button or P and the button
simultaneously, change the program number
5. Under the selected program number, switch to the stored value with
the P button
6. By pressing P briey, the required digit can be selected;
the selected digit is changed using or
7. The entry is saved by pressing the P button for approx. 2 seconds;
the value has been saved once the display shows horizontal bars
8. If no button is pressed, after approx. 7 seconds the instrument
returns to operating mode
Additional functions in normal operation to retrieve the recorded
Min/Max values
■
Using the button, the recorded value of the Max memory is
shown in the display for some seconds
■
Using the button, the recorded value of the Min memory is
shown in the display for some seconds
■
By simultaneously pressing the and the buttons, the Min and
Max values are deleted
GB
CAUTION!
If the measuring range setting is changed, to maintain the
measuring accuracy, the display must be recalibrated.
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH1019
6. Commissioning ... / 7. Maintenance, cleaning
6.3.3 Display information overrange/underrange
Range:4 ... 20 mA
Displayed range:3.9 ... 20.1 mA
GB
Usable range:3.6 ... 21.5 mA
Advance warning of underrange: 3.6 ... < 3.9 mA
Advance warning of overrange:> 20.1 ... 21.5 mA
Underrange displayed:< 3.6 mA
Overrange displayed:> 21.5 mA
With the advance warning, the display blinks (normal display
alternates with horizontal bars).
For values under 3.6 mA, a horizontal bar alternates with the display
“undr”. For values over 21.5 mA, a horizontal bar alternates with the
display “over”.
7. Maintenance and cleaning
7.1 Maintenance
The instrument is maintenance-free.
Repairs must only be carried out by the manufacturer.
7.2 Cleaning
CAUTION!
■
Before cleaning, correctly disconnect the instrument
from the pressure supply, switch it o and disconnect it
from the mains.
■
Clean the instrument with a moist cloth.
■
Electrical connections must not come into contact with
moisture.
■
Wash or clean the dismounted instrument before
returning it, in order to protect persons and the
environment from exposure to residual media.
■
Residual media in dismounted instruments can result in
a risk to persons, the environment and equipment. Take
sucient precautionary measures.
20WIKA operating instructions model DIH10
14089810.01 03/2014 GB/D/F/E
7. Maintenance and cleaning / 8. Faults
For information on returning the instrument see chapter 9.2
“Return”.
8. Faults
Fault-tree
DisplayMeasured value
GB
No display
Horizontal bar with
display “over.”
A horizontal bar
alternates with the
display “undr.”
I = 0 mA
I = > 20 mA
I = 4 mA
4 mA < I < 20 mA
wrong value
Current loop disconnected
Transmitter not connected
Power supply incorrectly
connected (reverse polarity)
Process temperature out of
range
Sensor break or short circuit
Sensor connected incorrectly
Wrong transmitter conguration
CAUTION!
If faults cannot be eliminated by means of the measures
listed above, shut down the instrument immediately, and
ensure that pressure and/or signal are no longer present,
and secure the instrument from being put back into
operation inadvertently.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in
chapter 9.2 “Return”.
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH1021
9. Dismounting, return and disposal
9. Dismounting, return and disposal
GB
9.1 Dismounting
Only disconnect the digital indicator once the system has been
depressurised!
9.2 Return
When returning the instrument, use the original packaging or a
suitable transport package.
To avoid damage:
1. Wrap the instrument in an antistatic plastic lm.
2. Place the instrument along with shock-absorbent material in the
packaging.
Place shock-absorbent material evenly on all sides of the transport
packaging.
3. If possible, place a bag containing a desiccant inside the packaging.
4. Label the shipment as carriage of a highly sensitive measuring
instrument.
WARNING!
Residual media in dismounted instruments can result in
a risk to persons, the environment and equipment. Take
sucient precautionary measures.
WARNING!
Strictly observe the following when shipping the
instrument:
All instruments delivered to WIKA must be free from any
kind of hazardous substances (acids, bases, solutions,
etc.).
Information on returns can be found under the heading
“Service” on our local website.
22WIKA operating instructions model DIH10
14089810.01 03/2014 GB/D/F/E
9. Dismounting, return and disposal
9.3 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an
environmentally compatible way and in accordance with the country-
specic waste disposal regulations.
GB
14089810.01 03/2014 GB/D/F/E
WIKA operating instructions model DIH1023
Appendix: EC declaration of conformity
GB
24WIKA operating instructions model DIH10
14089810.01 03/2014 GB/D/F/E
Inhalt
Inhalt
1. Allgemeines26
2. Sicherheit28
3. Technische Daten32
4. Aufbau und Funktion34
5. Transport, Verpackung und Lagerung37
6. Inbetriebnahme, Betrieb38
7. Wartung und Reinigung42
8. Störungen43
9. Demontage, Rücksendung und Entsorgung 44
Anlage: EG-Konformitätserklärung46
Konformitätserklärungen nden Sie online unter www.wika.de.
D
14089810.01 03/2014 GB/D/F/E
WIKA Betriebsanleitung Typ DIH10
25
1. Allgemeines
1. Allgemeines
■
Das in der Betriebsanleitung beschriebene Gerät wird nach dem
aktuellen Stand der Technik konstruiert und gefertigt.
Alle Komponenten unterliegen während der Fertigung strengen
Qualitäts- und Umweltkriterien. Unsere Managementsysteme sind
D
nach ISO 9001 und ISO 14001 zertiziert.
■
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit
dem Gerät. Voraussetzung für sicheres Arbeiten ist die Einhaltung
aller angegebenen Sicherheitshinweise und Handlungsanweisungen.
■
Die für den Einsatzbereich des Gerätes geltenden örtlichen Unfallverhütungsvorschriften und allgemeinen Sicherheitsbestimmungen
einhalten.
■
Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer Nähe des Gerätes für das Fachpersonal jederzeit zugänglich
aufbewahrt werden.
■
Das Fachpersonal muss die Betriebsanleitung vor Beginn aller
Arbeiten sorgfältig durchgelesen und verstanden haben.
■
Die Haftung des Herstellers erlischt bei Schäden durch bestimmungswidrige Verwendung, Nichtbeachten dieser Betriebsan-
leitung, Einsatz ungenügend qualizierten Fachpersonals sowie
eigenmächtiger Veränderung am Gerät.
■
Es gelten die allgemeinen Geschäftsbedingungen in den Verkaufsunterlagen.
… weist auf eine möglicherweise gefährliche Situation hin,
die zum Tod oder zu schweren Verletzungen führen kann,
wenn sie nicht gemieden wird.
VORSICHT!
… weist auf eine möglicherweise gefährliche Situation hin,
die zu geringfügigen oder leichten Verletzungen bzw. Sachund Umweltschäden führen kann, wenn sie nicht gemieden
wird.
Information
… hebt nützliche Tipps und Empfehlungen sowie Infor-
mationen für einen ezienten und störungsfreien Betrieb
hervor.
GEFAHR!
…kennzeichnet Gefährdungen durch elektrischen Strom.
Bei Nichtbeachtung der Sicherheitshinweise besteht die
Gefahr schwerer oder tödlicher Verletzungen.
WARNUNG!
… weist auf eine möglicherweise gefährliche Situation
hin, die durch heiße Oberächen oder Flüssigkeiten zu
Verbrennungen führen kann, wenn sie nicht gemieden wird.
D
14089810.01 03/2014 GB/D/F/E
WIKA Betriebsanleitung Typ DIH10
27
2. Sicherheit
2. Sicherheit
WARNUNG!
Vor Montage, Inbetriebnahme und Betrieb sicherstellen,
dass die Anzeige für den Anwendungsfall geeignet ist.
D
2.1 Bestimmungsgemäße Verwendung
Die Digitalanzeige Typ DIH10 ist eine universelle, konfigurierbare
Schleifenanzeige für Einzel-Widerstandsthermometer (RTD), Thermoelemente (TC) sowie Widerstands- und Sensorgeber.
Sie ist besonders dort einsetzbar, wo schon ein Standardsignal von
4 ... 20 mA vorhanden ist. Die Stromschleifen-Anzeige ist eingebaut
in geeignete Anschlussköpfe oder Wandgehäuse, und kann so als
Vor-Ort-Anzeige im gesamten industriellen Bereich eingesetzt werden.
Das Gerät ist ausschließlich für den hier beschriebenen bestimmungsgemäßen Verwendungszweck konzipiert und konstruiert und darf nur
dementsprechend verwendet werden.
Die technischen Spezikationen in dieser Betriebsanleitung sind
einzuhalten. Eine unsachgemäße Handhabung oder ein Betreiben
des Gerätes außerhalb der technischen Spezikationen macht die
sofortige Stilllegung und Überprüfung durch einen autorisierten WIKAServicemitarbeiter erforderlich.
Bei Nichtbeachten können schwere Körperverletzungen
und/oder Sachschäden auftreten.
Weitere wichtige Sicherheitshinweise benden sich in den
einzelnen Kapiteln dieser Betriebsanleitung.
Wird das Gerät von einer kalten in eine warme Umgebung transportiert, so kann durch Kondensatbildung eine Störung der Gerätefunktion eintreten. Vor einer erneuten Inbetriebnahme die Angleichung der
Gerätetemperatur an die Raumtemperatur abwarten.
28WIKA Betriebsanleitung Typ DIH10
14089810.01 03/2014 GB/D/F/E
Loading...
+ 64 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.