2WIKA operating instructions digital gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
Contents
Contents
1.General information4
2.Safety6
3.Specications9
4.Design and function13
5.Transport, packaging and storage13
6.Commissioning, operation14
7.Maintenance and cleaning25
8.Faults26
9.Dismounting, return and disposal27
10. Accessories28
Appendix 1: EC Declaration of conformity model DG-1055
GB
Declarations of conformity can be found online at www.wika.com.
11250313.03 09/2012 GB/D/F/E
3WIKA operating instructions digital pressure gauge, model DG-10
1. General information
1. General information
■
The digital pressure gauge described in the operating instructions has been designed and
GB
manufactured using state-of-the-art technology. All components are subject to stringent quality and
environmental criteria during production. Our management systems are certied to ISO 9001 and
ISO 14001.
■
These operating instructions contain important information on handling the instrument. Working
safely requires that all safety instructions and work instructions are observed.
■
Observe the relevant local accident prevention regulations and general safety regulations for the
instrument's range of use.
■
The operating instructions are part of the product and must be kept in the immediate vicinity of the
instrument and readily accessible to skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the operating instructions prior to
beginning any work.
■
The manufacturer's liability is void in the case of any damage caused by using the product contrary
to its intended use, non-compliance with these operating instructions, assignment of insuciently
qualied skilled personnel or unauthorised modications to the instrument.
■
The general terms and conditions contained in the sales documentation shall apply.
4WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
1. General information
Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury or death, if not
avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or damage to
the equipment or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for ecient and trouble-free
operation.
GB
11250313.03 09/2012 GB/D/F/E
5WIKA operating instructions digital pressure gauge, model DG-10
2. Safety
2. Safety
WARNING!
GB
2.1 Intended use
The digital pressure gauge acts as the display for the applied pressure. The pressure value is displayed
via a digital indicator.
The instrument has been designed and built solely for the intended use described here, and may only
be used accordingly.
The technical specications contained in these operating instructions must be observed. Improper
handling or operation of the digital pressure gauge outside of its technical specications requires the
digital pressure gauge to be taken out of service immediately and inspected by an authorised WIKA
service engineer.
Before installation, commissioning and operation, ensure that the appropriate digital
pressure gauge has been selected in terms of measuring range, design and specic
measuring conditions.
Non-observance can result in serious injury and/or damage to the equipment.
WARNING!
■
Open the connections only after the system has been depressurised.
■
Observe the working conditions in accordance with chapter 3 "Specications".
■
Always operate the digital pressure gauge within the overpressure limit.
Further important safety instructions can be found in the individual chapters of these
operating instructions.
The manufacturer shall not be liable for claims of any type based on operation contrary to the intended
use.
6WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
2. Safety
2.2 Personnel qualication
WARNING!
Risk of injury should qualication be insucient!
Improper handling can result in considerable injury and damage to equipment.
The activities described in these operating instructions may only be carried out by skilled
personnel who have the qualications described below.
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training, knowledge
of measurement and control technology and on their experience and knowledge of country-specic
regulations, current standards and directives, are capable of carrying out the work described and
independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive media.
2.3 Special hazards
WARNING!
For hazardous media such as oxygen, acetylene, ammable or toxic gases or liquids,
and refrigeration plants, compressors, etc., in addition to all standard regulations, the
appropriate existing codes or regulations must also be followed.
WARNING!
Residual media in dismounted digital pressure gauges can result in a risk to persons, the
environment and equipment.
Take sucient precautionary measures.
GB
11250313.03 09/2012 GB/D/F/E
7WIKA operating instructions digital pressure gauge, model DG-10
2. Safety
2.4 Labelling / safety marks
Product label
GB
Measuring range
Power supplyP# Product no.
If the serial number becomes illegible due to mechanical damage or overpainting, traceability will no
longer be possible.
Explanation of symbols
Before mounting and commissioning the instrument, ensure you read the operating
instructions!
CE, Communauté Européenne
Instruments bearing this mark comply with the relevant European directives.
S# Serial no.
8WIKA operating instructions digital pressure gauge, model DG-10
EMC directive 2004/108/EC, EN 61326 emission (group 1, class B) and immunity (industrial application)
3.5 Voltage supply
see chapter 6.2 "Voltage supply"
3.6 Reference conditions
Temperature
0 ... 60 °C
Atmospheric pressure
860 ... 1,060 mbar
Humidity
< 90 % rel., non-condensing
Nominal position
Calibrated in vertical mounting position with pressure connection facing downwards.
Power supply
2 x 1.5 V AA cell
10WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
3. Specications
3.6 Accuracy
Accuracy at reference conditions
≤ ±0.5 % of span ±1 digit
Including non-linearity, hysteresis, zero oset and end value deviation (corresponds to measured error
per IEC 61298-2).
Zero error for the zero signal
≤ 0.1 % of span (power-up reset)
Adjustability of zero point
≤ 20 % of span (via tare function with model DG-10-E)
Temperature error at 0 ... 60 °C
■
Mean temperature coecient of zero point
≤ 0.15 % of span/10 K
■
Mean temperature coecient of span
≤ 0.15 % of span/10 K
Long-term stability
≤ 0.2 % of span/year
GB
11250313.03 09/2012 GB/D/F/E
11WIKA operating instructions digital pressure gauge, model DG-10
3. Specications
3.7 Operating conditions
Ingress protection (per IEC 60529)
IP 65
GB
Temperatures
■
Medium:
- Measuring range < 100 bar:-30 ... +85 °C
- Measuring range ≥ 100 bar:-30 ... +100 °C
■
Ambient:-10 ... +60 °C
■
Storage:-20 ... +70 °C
3.8 Materials
Wetted parts
■
Process connection:Stainless steel 1.4571
■
Sensor element ≤ 50 bar:Ceramic Al2O3 96 %
NBR (option: EPDM)
■
Sensor element > 100 bar: XM-13 (stainless steel 1.4534)
Non-wetted parts
■
Case:Stainless steel 1.4301
■
Front foil and membrane: Polyester
For special model numbers, e.g. DG-10000, please note the specications stated on the delivery note.
For further specications see WIKA data sheet PE 81.66 and the order documentation.
12WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
4. Design and function / 5. Transport, packaging and storage
4. Design and function
4.1 Description
The prevailing pressure is measured at the sensor element through the deformation of a diaphragm. By
supplying power, this deformation of the diaphragm is converted into an electrical signal. This signal is
displayed via a digital indicator.
4.2 Scope of delivery
Cross-check scope of delivery with delivery note.
5. Transport, packaging and storage
5.1 Transport
Check the digital pressure gauge for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in installation
site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
■
Storage temperature: -20 ... +70 °C
■
Humidity: < 90 % rel., non-condensing
GB
WARNING!
Before storing the digital pressure gauge (following operation), remove any residual media.
This is of particular importance if the medium is hazardous to health, e.g. caustic, toxic,
carcinogenic, radioactive, etc..
11250313.03 09/2012 GB/D/F/E
13WIKA operating instructions digital pressure gauge, model DG-10
6. Commissioning, operation
6. Commissioning, operation
CAUTION!
GB
6.1 Mechanical mounting
■
The sealing faces at the instrument have to be
undamaged and clean.
■
When screwing the instrument in, the force
required to do this must not be applied through
the case, but only through the spanner ats
provided for this purpose and using a suitable
tool.
The correct torque depends on the dimensions
of the process connection and the gasket used
(form/material).
Do not overwind the process connection, do not exceed the permissible torques.
■
When screwing in, do not cross the threads.
■
For information on tapped holes and welding sockets, see Technical information IN 00.14
at www.wika.com.
Prior to commissioning, the digital pressure gauge must be subjected to a visual inspection.
Only use the digital pressure gauge if it is in perfect condition with respect to safety.
14WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
6. Commissioning, operation
Sealing of the process connection
Parallel threadsTapered threads
per EN 837per DIN 3852-ENPT, R and PT
GB
Correct sealing of the process connections with
parallel threads at the sealing face must be
made using suitable at gaskets, sealing rings or
WIKA prole sealings.
For further information on sealings see WIKA data sheet AC 09.08 or under www.wika.com.
Rotatability of the case
The case of the model DG-10-E can be rotated through 300°, thus the
display can be read from almost any position.
Do not turn the case forcibly past the stop point.
11250313.03 09/2012 GB/D/F/E
For sealing process connections with tapered
threads, the sealing must be made in the threads
using additional sealing material, e.g. PTFE tape
(EN 837-2).
300°
15WIKA operating instructions digital pressure gauge, model DG-10
6. Commissioning, operation
6.2 Voltage supply
The power supply for the digital pressure gauge is provided via two 1.5 V AA cells.
The operating time is approx. 4,000 hours, when using batteries with a capacity of 2,000 mAh.
GB
Battery status
The battery status is shown by a battery symbol in the display (see
chapter 6.3 "Description of the digital indicator").
Change the batteries as follows, as soon as the battery symbol
shows no bars and is blinking.
Change batteries
Take the cover o.
16WIKA operating instructions digital pressure gauge, model DG-10
Remove the battery compartment with the help of the tab and
change the batteries.
Put the instrument back together
in the reverse order.
11250313.03 09/2012 GB/D/F/E
6. Commissioning, operation
6.3 Description of the digital indicator
Model DG-10-SModel DG-10-E
Type of indication7-Segment LCD7-segment LCD
■
Digits4-digit4 ½-digit
■
Character size11 mm11 mm and 7 mm
■
Indication range-1999 ... 9999-1999 ... 19999
■
Update200 ms200 ms
■
Backlightingnoyes
■
Bargraph with drag pointer functionyesyes
■
Rotatabilityno300°
Model DG-10-SModel DG-10-E
14-segment LCD (2. Display)
GB
Max. value
Min. value
Main display
Battery status
11250313.03 09/2012 GB/D/F/E
Bargraph display
Full scale
Unit
Max. value
Min. value
Main display
Battery status
Bargraph display
Special functionsTare activated
2. Display
Full scale
Unit
17WIKA operating instructions digital pressure gauge, model DG-10
6. Commissioning, operation
Backlighting (model DG-10-E)
The model DG-10-E features backlighting so that the display can
be easily read, even under poor lighting conditions.
GB
After pressing any key, the display will light up for 7 seconds.
The backlighting can be switched on and o via the menu item
"LIGH", see chapter 6.4 "Menu navigation in the programming
mode, model DG-10-E".
The additional integrated bargraph with drag-pointer function within the display indicates the trend of
the current pressure directly, independently from the digital indicator. Using the drag-pointer function,
the max. memory value is also displayed in the bargraph display in the form of a bar segment, as well
as in the digital indicator. On resetting the MIN/MAX value, this bargraph is then updated to the current
measured value.
18WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
6. Commissioning, operation
MIN/MAX memory
The MIN/MAX memory is reset to the current measured value with each measuring cycle.
Tare function "TARE" (model DG-10-E)
In this mode, the positive deviation of the measured value from the tared value (max. 20 % of the
measuring range) is shown on the main display. The non-tared value is shown on the bargraph display,
so that you can always read the current utilisation of the digital pressure gauge.
The following possibilities can be activated for the tare function:
■
Select the tare function via the "TARE" menu item. Use the arrow keys to adjust the value to be tared.
The adjusted value to be tared will be stored in the memory as tare.
■
Load the digital pressure gauge to the pressure to be tared. The main display will show the current
measured value. Via the menu, select the tare function, "TARE", and press the two arrow keys
simultaneously. The current measured value will be stored as tare. The main and secondary display
will be set to "0000".
The tare function can be reset in both variants by setting the tare value to "0000".
GB
11250313.03 09/2012 GB/D/F/E
19WIKA operating instructions digital pressure gauge, model DG-10
6. Commissioning, operation
Key function in the operating mode
■
Switches the digital pressure gauge on or o.
■
GB
Switches the instrument to the programming mode (pressing for longer than 3 seconds).
On pressing, the display indicates the max. value.
On pressing, the display indicates the min. value.
+
+
20WIKA operating instructions digital pressure gauge, model DG-10
Resets the max. value to 0.
Resets the min. value to 0.
11250313.03 09/2012 GB/D/F/E
6. Commissioning, operation
Menu navigation in the programming mode, model DG-10-S
Operating mode
Software
version
Display of the
measuring
range
Selection of the
unit
"UNIT"
Saving the
settings
"STOR"
< 3 seconds
MPa
PSI
BAR
YES
NO
GB
Factory setting "BAR"
Factory setting "NO"
11250313.03 09/2012 GB/D/F/E
21WIKA operating instructions digital pressure gauge, model DG-10
6. Commissioning, operation
Menu navigation in the programming mode, model DG-10-E
GB
Software
version
Display of the
measuring
range
Password
status
"P.STA"
"P.CHA" (next page)
< 3 seconds
if P.STA "OFF"
Operating mode
Password
request
"PASS"
ON
OFF
Password
wrong
▼ key changes the digit
▲ key sets the numerical value
Factory setting "0000"
Password
correct
"ON" - password is requested.
"OFF" - password is not requested,
"PASS" menu item is not shown in menu
Factory setting "OFF"
22WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
6. Commissioning, operation
Changing the password
"P.CHA"
Selection of the unit
"UNIT"
Tare function
"TARA"
Backlighting
"LIGH"
"A.OFF" (next page)
Key changes the digit
Key sets the numerical
value
MPa
PSI
BAR
see point "Tare function" on
page 19
ON
OFF
Factory setting "0000"
Factory setting "BAR"
Factory setting "0000"
Factory setting "OFF"
GB
11250313.03 09/2012 GB/D/F/E
23WIKA operating instructions digital pressure gauge, model DG-10
6. Commissioning, operation
Automatic power-o
GB
"A.OFF"
Reset to factory settings
"FAC"
Saving the settings
"STOR"
"P.STA"
OFF
15, 30, 60, 120 minutes
YES
NO
YES
NO
Operating mode
Factory setting "OFF"
Factory setting "NO"
Factory setting "NO"
24WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
7. Maintenance and cleaning
7. Maintenance and cleaning
7.1 Maintenance
The digital pressure gauge is maintenance-free.
Repairs must only be carried out by the manufacturer.
7.2 Cleaning
CAUTION!
■
Before cleaning, correctly disconnect the digital pressure gauge from the pressure
supply.
■
Clean the instrument with a moist cloth.
■
Wash or clean the dismounted instrument before returning it, in order to protect persons
and the environment from exposure to residual media.
■
Residual media in dismounted instruments can result in a risk to persons, the
environment and equipment. Take sucient precautionary measures.
■
Do not use any pointed or hard objects for cleaning.
For information on returning the instrument see chapter 9.2 "Return".
GB
11250313.03 09/2012 GB/D/F/E
25WIKA operating instructions digital pressure gauge, model DG-10
8. Faults
8. Faults
In the event of any faults, rst check whether the digital pressure gauge is mounted correctly, mechanically
GB
and electrically.
FaultsCausesMeasures
No displayBatteries discharged
Batteries inserted the wrong way around
Display remains on 0, despite
there being pressure
Constant display signal upon
change in pressure
Signal span too small
Signal span variesEMC interference sources in the environment
Signal span inaccurateOperating temperature too high/lowObserve the permissible
Deviating zero point signalMedium or ambient temperature too high/lowObserve the permissible
Instrument turned too far
Sensor wire broken
Mechanical overload caused by overpressureReplace instrument
Mechanical overload caused by overpressure
Operating temperature too high/low
If complaint is unjustied, we will charge you the complaint processing fees.
CAUTION!
If faults cannot be eliminated by means of the measures listed above, shut down the digital
pressure gauge immediately, and ensure that pressure and/or signal are no longer present,
and secure the instrument from being put back into operation inadvertently. In this case,
contact the manufacturer. If a return is needed, please follow the instructions given in
chapter 9.2 "Return".
Change batteries
Insert the batteries with the
polarity correct
Replace instrument
Replace instrument
Remove sources of interference
Observe the permissible
temperatures
temperatures
temperatures
26WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
9. Dismounting, return and disposal
9. Dismounting, return and disposal
WARNING!
Residual media in dismounted pressure transmitters can result in a risk to persons, the
environment and equipment.
Take sucient precautionary measures.
9.1 Dismounting
Only disconnect the digital pressure gauge once the system has been depressurised!
9.2 Return
WARNING!
Strictly observe the following when shipping the digital pressure gauge:
All digital pressure gauges delivered to WIKA must be free from any kind of hazardous
substances (acids, bases, solutions, etc.).
When returning the instrument, use the original packaging or a suitable transport package.
Information on returns can be found under the heading "Service" on our local website.
9.3 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally compatible way and
in accordance with the country-specic waste disposal regulations.
GB
11250313.03 09/2012 GB/D/F/E
27WIKA operating instructions digital pressure gauge, model DG-10
10. Accessories
10. Accessories
GB
DescriptionOrder no.
Rubber protection cap (black)11262010
Minimess connector
Maximum pressure strength 630 bar
Sealing for process connection per DIN 3852-E
Material NBR
Thread size
G ¼ A1537857
9072314
28WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
Inhalt
Inhalt
1.Allgemeines30
2.Sicherheit32
3.Technische Daten35
4.Aufbau und Funktion39
5.Transport, Verpackung und Lagerung39
6.Inbetriebnahme, Betrieb40
7.Wartung und Reinigung51
8.Störungen52
9.Demontage, Rücksendung und Entsorgung53
10. Zubehör54
Anlage 1: EG-Konformitätserklärung Typ DG-1055
Konformitätserklärungen nden Sie online unter www.wika.de.
11250313.03 09/2012 GB/D/F/E
D
29WIKA Betriebsanleitung Digitalmanometer, Typ DG-10
1. Allgemeines
1. Allgemeines
■
Das in der Betriebsanleitung beschriebene Digitalmanometer wird nach dem aktuellen Stand der
Technik konstruiert und gefertigt.
Alle Komponenten unterliegen während der Fertigung strengen Qualitäts- und Umweltkriterien.
Unsere Managementsysteme sind nach ISO 9001 und ISO 14001 zertiziert.
D
■
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit dem Gerät. Voraussetzung
für sicheres Arbeiten ist die Einhaltung aller angegebenen Sicherheitshinweise und
Handlungsanweisungen.
■
Die für den Einsatzbereich des Gerätes geltenden örtlichen Unfallverhütungsvorschriften und
allgemeinen Sicherheitsbestimmungen einhalten.
■
Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer Nähe des Gerätes für das
Fachpersonal jederzeit zugänglich aufbewahrt werden.
■
Das Fachpersonal muss die Betriebsanleitung vor Beginn aller Arbeiten sorgfältig durchgelesen und
verstanden haben.
■
Die Haftung des Herstellers erlischt bei Schäden durch bestimmungswidrige Verwendung,
Nichtbeachten dieser Betriebsanleitung, Einsatz ungenügend qualizierten Fachpersonals sowie
eigenmächtiger Veränderung am Gerät.
■
Es gelten die allgemeinen Geschäftsbedingungen in den Verkaufsunterlagen.