WIKA DA10 User Manual

NI-209E
INSTRUCTION MANUAL
Rev. 4 12/02
DIFFERENTIAL PRESSURE SWITCHES
WEATHERPROOF: SERIES DW EXPLOSIONPROOF: SERIES DA
Models DW40, DW100, DW160
Models DA40, DA100, DA160
C
C
AB
AB
A Low pressure attachment B High pressure attachment
Weight 6,2 kg dimensions in mm Weight 7,2 kg dimensions in mm Models DW10
C
C Cable entry
A Low pressure attachment B High pressure attachment
Models DA10
C
C Cable entry
AB
A Low pressure attachment B High pressure attachment Weight 8,2 kg dimensions in mm Weight 10 kg dimensions in mm
NOTE: dimensions and weights are not binding unless released on certified drawings. CAUTION
Before installing, using or carrying out maintenance on the instrument it is necessary to read and understand the indications given in the attached Instruction Manual.
The instrument must only be installed and maintained by qualified personnel.
INSTALLATION IS TO BE CARRIED OUT ONLY AFTER CHECKING THAT INSTRUMENT CHARACTERISTICS
ARE CONSISTENT WITH PROCESS AND PLANT REQUIREMENTS
The functional features of the instrument and its degree of protection are shown on the identification plate fixed to the case.
SAFETY INSTRUCTIONS FOR USE IN HAZARDOUS ATMOSPHERES.
RECOMMENDATIONS FOR PRESSURE SWITCH SAFE USE.
All data, statements and recommendations supplied with this manual are based on information believed by us to be reliable. As
the conditions of effective use are beyond our control, our products are sold under the condition that the user himself evaluates such conditions before following our recommendations for the purpose or use foreseen by him.
C Cable entry
A Low pressure attachment B High pressure attachment
AB
C Cable entry
RELATED DOCUMENT
To authentified document with certificate
N° CESI 03 ATEX 004
INSTRUCTION MANUAL
1. GENERAL NOTES
1.1 FOREWORD
The wrong choice of a series or a model, as well as the incorrect installation, lead to malfunction and reduce instrument life. Failure to follow the indications given in this manual can cause damage to the instrument, the environment and persons.
1.2 ALLOWED OVERRANGE
Pressures exceeding the working range can be occasionally tolerated provided they remain within the limits stated in the instrument features (vacuum or proof pressure). Continuous pressures exceeding the working range can be applied to the instrument, provided they are clearly stated in the instrument features. The current and voltage values stated in the technical specifications and ratings must not be exceeded. Transitory overranges can have a destructive effect on the switch.
1.3 MECHANICAL VIBRATIONS
Can generally lead to the wearing of some parts of the instrument or cause false actuation. It is therefore recommended that the instrument be installed in a place where there are no vibrations. In cases where this is impossible it is advisable to take measures to lessen the effects (elastic supports, installation with the pin of the microswitch positioned at right angles to the vibration plane, etc.).
1.4 TEMPERATURE
Due to the temperature of both the environment and the process fluid, the temperature of the instrument could exceed the allowed limits (normally from -20° to +70°C). Therefore, in case it does, suitable measures (protection against heat radiation, fluid separators, cooling coils, heated lockers) must be taken. The process fluid or its impurities must not in any case solidify inside the instrument chambers.
2. OPERATING PRINCIPLE
The differential pressure, acting on the sensitive diaphragm element, determines its elastic deformation which is used to actuate one or two electric microswitches regulated at set point values. The microswitches are of the snap acting type with automatic reset. When the pressure moves away from the set values, returning towards the normal values, the switch is reset. The dead band (difference between the set point value and the reset value) can be set or adjustable (letter R in the contact codes).
NI-209E
Rev. 4 12/02
3. IDENTIFICATION PLATE AND MARKINGS
The instrument is fitted with a metal plate bearing all its functional characteristics and – in case of explosionproof execution (Series DA) – also the markings prescribed by standard CEI EN 50014. Fig. 1 shows the plate mounted on explosionproof instruments.
Fig. 1 - Explosionproof instruments plate
1 CE marking and identification number of the notified body
responsible for production surveillance.
2 Apparatus classification according to ATEX 94/9 CE
directive.
3 Type of protection and ambient temperature limits of
operation.
4 Notified body that issued the type certificate and number of
said certificate.
4. SET POINT REGULATION
Each microswitch is independent and can be regulated by means of a screw (for adjustment) in such a way that it is released when the pressure reaches (increasing or decreasing) the desired value (set point). The instrument is usually supplied with the switches set at the setting range value nearest to zero (factory calibration). The instrument is supplied with an adhesive rating plate showing the set point calibration value. With factory calibration the values are not indicated on the ratings as these are temporary and will be modified with the definitive values. Prior to installation the instrument must be calibrated and the definitive calibration values written on the adhesive rating plate using a suitable indelible ink pen.
If the instrument has been ordered with a specific calibration, it is a good rule to check the calibration values marked on the relevant adhesive label, prior to installation.
The position of the adjustment screws is given in fig. 2. The effect that the direction of rotation of the adjustment screws has is described on the adhesive plate.
INSTRUCTION MANUAL
Fig. 2 - Electrical connections and adjustment screws
One contact instruments Two contact instruments
NI-209E
Rev. 4 12/02
Microswitch electrical circuit: State of the contacts at atmospheric pressure
1 - Microswitch set point calibration screw 3 - Terminal block 4 - Electrical connection identification plate
5. SET POINT CALIBRATION
In order to proceed with the calibration and the periodical functional verification of the instrument a suitable calibration circuit (fig. 5) and an adequate pressure source is required.
5.1 PRELIMINARY OPERATIONS
Weatherproof pressure switches (Series DW) (Fig.
3)
Remove the blocking device fixed to the side of the instrument case and the adjustment screw access plate. Remove the cover be rotating it anticlockwise.
Fig. 3 -
and plumbing
Weatherproof pressure switch cover blocking device
1 – Microswitch 2 set point calibration screw 2 – Microswitch 1 set point calibration screw 3 – Terminal block 4 – Electrical connection identification plate
Designation of the contacts: C - common NA - Normally open NC - Normally closed
Loosen the locking headless screw situated on the cover using a 1,5 hexagonal key then unscrew the cover. Remove the internal blocking device inserted on the closure plugs and slide out the plugs.
Fig. 4 -
Explosionproof pressure switch cover blocking device
a) Plumbing wire b) Plumbing c) Blocking nut
Explosionproof pressure switches (Series DA) (Fig.
4)
CAUTION: do not open the cover of explosionproof
pressure switches (Series DA) when energized, in explosive atmospheres.
c) Blocking bracket e) Adjustment screw access plate
5.2 CALIBRATION CIRCUIT AND OPERATIONS
Prepare the control circuit as indicated in Fig.5. The warning lamps should be connected to contact 1 or 2 in the NO or NC position according to the required contact action.
Connection of C and NO terminals
•••• If the circuit is open at the working pressure, the switch
closes the circuit as the pressure increases when the desired value is reached.
• If the circuit is closed at the working pressure, the switch opens the circuit as the pressure decreases when the desired value is reached.
Connection of C and NC terminals
If the circuit is closed at the working pressure, the
switch opens the circuit as the pressure increases when the desired value is reached.
• If the circuit is open at the working pressure, the switch closes the circuit as the pressure decreases when the desired value is reached.
Loading...
+ 4 hidden pages