WIKA CTR3000 Specifications

Multi-functional precision thermometer CTR3000
Applications
Pharmaceutical industry
Industry (laboratory, workshop and production)
Temperature sensor and transmitter manufacturers
Calibration service companies and service industry
Special features
High accuracy
Innovative and intuitive user interface
Versatile applications by measuring thermocouples and resistance thermometers
Logger and scan functions
Up to 44 channels possible
Calibration technology
WIKA data sheet CT 60.15
Description
Application
The model CTR3000 precision thermometer provides a complete measurement and control interface for users wishing to make high-accuracy temperature measurements or calibrate thermometers. It supports a wide range of
thermometer types including 25 Ω SPRTs, 100 Ω PRTs,
thermistors and thermocouples. The CTR3000 is a high-accuracy instrument designed for laboratory and industrial temperature measurement and calibration applications.
Functionality
The instrument will operate with all 3- and 4-wire (S)PRTs (25 Ω, 100 Ω) platinum resistance thermometers as well as
most standard international thermocouple types and NTC thermistors. The following temperature measurement units
are selectable: °C, °F, K. Base measuring units mV and Ω
are also displayed. The temperature values will be calculated through common conversion of the base measurement.
01/2017
Due to the wide range of this instrument it makes individual instruments needless and makes the calibration
cost-eective.
Features included:
Dual capability for both thermocouple and resistance thermometer measurements
Input channels can be expanded up to 44
Large graphic touchscreen for temperature measurement
values as well as conguration settings and statistical
results
Logger and log-data transfer to USB stick or
communication interface
Scan function with a live screen and graph
Communication interfaces available for automated monitoring and calibration applications
Page 1 of 9WIKA data sheet CT 60.15
Data sheets showing similar products: Precision thermometer; model CTR5000; see data sheet CT 60.20
Specications
Model CTR3000
Multi-functional precision thermometer
Input
Input channels 4
Channels 1 + 2 Resistance thermometers with 5-pin DIN connector
Channels 3 + 4 Thermocouple with standard miniature 2-pin thermocouple plug
Scanner box
Input connections 5-pin DIN plug or bare wires (resistance thermometer or thermistor)
Data entry format
Measuring ranges
1)
PRT/SPRT Measuring range 0 ... 500 Ω
Thermocouple Measuring range ±100 mV
Thermistor 0 ... 500 kΩ
1) Depending on sensor type
up to 4 modules
max. 44 channels (in total)
each module has 10 channels
Standard miniature 2-pin thermocouple plug or bare wires (thermocouple)
ITS-90 and CvD for calibrated resistance thermometers; or EN 60751 standard conversion for
uncalibrated resistance thermometers
TC polynomial for calibrated thermocouples; or EN 60584 standard conversion for uncalibrated
thermocouple
Steinhart and Hart for NTC thermistors
-200 … +962 °C (-328 ... +1,764 °F)
3- and 4-wire measurement
-210 … +1,820 °C (-346 ... +3,308 °F) Type B, E, J, K, N, R, S, T in accordance with EN 60584
Accuracies
2)
Resistance thermometer
Base measurement 0.002 Ω
Temperature accuracy 4-wire
3-wire
±0.005 K ±0.03 K
Temperature conversions Standard EN 60751, CvD, ITS-90
Sensor currents 1 mA, 2 mA and √2
Keep warm currents R
< 50 Ω
0
R0 ≥ 50 Ω
0 ... 125 Ω 0 ... 500 Ω
2 mA 1 mA
Measurement time 3 seconds update rate
Thermocouple
Base measurement
3)
±% of reading + µV ±0.004 % + 2 µV
Temperature accuracy Type B
Type E Type J
Type K Type N Type R
Type S
Type T
±0.09 °C + ±0.025 % of reading ±0.05 °C + ±0.031 % of reading ±0.07 °C + ±0.030 % of reading ±0.09 °C + ±0.035 % of reading
±0.08 °C + ±0.035 % of reading
±0.27 °C + ±0.020 % of reading ±0.27 °C + ±0.020 % of reading ±0.09 °C + ±0.025 % of reading
Temperature conversions Standard EN 60584, polynomial
Measurement time 3 seconds update rate
Cold junction compensation Internal, external or channel
Accuracy internal cold junction compensation ±0.15 K
2) The accuracy in K denes the deviation between the measured value and the reference value. (Only valid for indicating instruments.)
3) In a range of -20 mV ... +100 mV
Page 2 of 9 WIKA data sheet CT 60.15 ∙ 01/2017
Accuracies
Thermistor
Accuracy 0 ... 400 Ω
Temperature conversions Steinhart-Hart, Polynomial
Sensor currents 0 ... 450 Ω
Measurement time 3 seconds update rate
2) The accuracy in K denes the deviation between the measured value and the reference value. (Only valid for indicating instruments.)
3) In a range of -20 mV ... +100 mV
2)
±0.006 Ω 400 Ω ... 50 kΩ 50 ... 500 kΩ
400 Ω ... 45 kΩ 40 ... 500 kΩ
±0.01 % of reading
±0.02 % of reading
1 mA
10 μA
3 μA
Digital indicator
Display
Screen Colour TFT display including projective capacitive touchscreen with a resolution of 800 x 480 pixels
Resolution 0.0001 K / 0.00001 Ω / 0.00001 mV
Display units °C, °F, K, mV and Ω
Functions
Real-time clock Integrated clock with date
Voltage supply
Power supply AC 100 ... 240 V, 47 ... 63 Hz; universal input on rear panel
Permissible ambient conditions
Operating temperature 0 ... 50 °C (32 ... 122 °F)
Maximum achievable accuracy within 17 ... 23 °C (63 ... 73 °F)
Relative humidity 0 … 70 % r. h. (non-condensing)
Storage temperature -20 ... +50 °C (-4 ... +122 °F)
Communication
Interfaces Standard: USB host, USB device and Ethernet
Optional: RS-232
Case
Dimensions (W x H x D) 314 x 176 x 322 mm (12.4 x 6.9 x 12.7 in)
Weight 6 kg (13.2 lbs)
Approvals
Logo Description Country
EU declaration of conformity
EMC directive
Low voltage directive
European Union
Certicates
Certicate
Calibration
Recommended recalibration interval 1 year (dependent on conditions of use)
4) System calibration with a temperature probe and/or electrical calibration of the CTR3000 itself
Approvals and certicates, see website
WIKA data sheet CT 60.15 ∙ 01/2017 Page 3 of 9
4)
Standard: 3.1 calibration certicate per DIN EN 10204 Option: DKD/DAkkS calibration certicate