WIKA CTR2000 Specifications

Precision thermometer Model CTR2000
Applications
Precision thermometer for very accurate temperature measurements in a range of -200 ... +962 °C
Self-contained, complete system, also suitable for on-site measurements/calibrations
Calibration technology
WIKA data sheet CT 60.10
Special features
Temperature range -200 ... +962 °C
Accuracy: 0.01 °C
2 channels as standard with optional 8 channels
4-wire measurement, optional 3-wire
SMART probe review and editor
Description
The model CTR2000 precision thermometer comprises two measurement versions: One version accepts only 4-wire PRTs, the other version accepts both 3- and 4-wire PRTs. The high-performance 2- or 8-channel, 3- and 4-wire precision digital thermometers is suitable for use with calibrated and uncalibrated Pt100 probes. It is characterised by its extended features for high usability and low life costs.
Using calibrated probes with the CTR2000 thermometer series choose between storing calibration data into memory in the instrument or into the probe’s proprietary SMART connector. Calibration data stays permanently with the probe in this connector, which the CTR2000 series instantly recognises, ensuring best measurement accuracy.
Precision thermometer, model CTR2000
Using a built-in calibration routine, the CTR2000 instrument range can be recalibrated using a calibrated, traceable reference resistor. This feature is password protected.
Other features have been included for CTR2000 range when using SMART probes, like the temperature watchdog and the calibration status monitor. The temperature watchdog continuously monitors the temperature of SMART probes and, if any are used outside
their specied working range, the probe history record is
updated.
With this way, it can be seen at a glance if the probes have ever been used outside their calibrated range, as this may
aect their calibration validity. The calibration status monitor
alerts with an on-screen message when a SMART probe is due for recalibration.
Data sheets showing similar products: Precision thermometer; model CTR5000; see data sheet CT 60.20 Reference thermometer; model CTP5000; see data sheet CT 61.20
Page 1 of 6WIKA data sheet CT 60.10 ∙ 07/2014
Specications
Model CTR2000
Precision thermometer
Probe types Industrial platinum resistance thermometers (PRTs) and standard platinum resistance ther-
mometers (SPRTs) with Ro = 25 Ω and 100 Ω up to an alpha of 0.00392
Measuring inputs 2 or 8
Data entry format ITS 90 and CVD for calibrated probes; or EN 60751 for uncalibrated probes
Measuring ranges
Probe current DC 1 mA switched polarity
Temperature range -200 ... +962 °C, depending on thermometer probe
Accuracy
1) The accuracy in K denes the deviation between the measured value and the reference value. (Only valid for indicating instruments.)
Digital indicator
Display
Screen large graphic VFD display screen
Resolution 0.001 °C
1)
0.01 K (4-wire)
0.03 K (3-wire, do not include uncertainties due to variations in sensor cables)
Functions
Real-time clock integrated clock with date
Voltage supply
Power supply AC 90 ... 264 V, 47 ... 63 Hz; universal rear input on rear panel
Power consumption max. 30 VA
Permissible ambient conditions
Operating temperature 15 ... 25 °C
Storage temperature -20 ... +50 °C
Communication
Interface RS-232
Case
Dimensions 168 x 108 x 215 mm (W x H x D)
Weight 1 kg
CE conformity, certicates
CE conformity
EMC directive 2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity (portable test
and measuring equipment)
Certicates
Calibration
Recommended calibration interval 1 year (depending on the conditions of usage)
2) Only system calibration
2)
Standard: 3.1 calibration certicate per DIN EN 10204 Option: DKD/DAkkS calibration certicate
Approvals and certicates, see website
Page 2 of 6 WIKA data sheet CT 60.10 ∙ 07/2014
Recommended temperature probes
Temperature probe
Resistance thermometer
Spring strain relief
2 metres
Engraved line
300 mm maximum insertion depth
350 or 450
Resistance thermometer (immersion probe) Temperature range
Pt100, d = 6 mm, l = 350 mm -50 ... +250 °C
Pt100, d = 6 mm, l = 350 mm -70 ... +450 °C
Pt100, d = 6 mm, l = 450 mm -70 ... +650 °C
ø 6 or ø 7
nominal
Centreline of sensor
Resistance thermometer
Brazed weld
3 metres
Resistance thermometer (immersion probe) Temperature range
Pt100, d = 3 mm, l = 30 mm -50 ... +200 °C
Page 3 of 6WIKA data sheet CT 60.10 ∙ 07/2014
Resistance thermometer connection, 4-wire (5-pin DIN connector)
View towards front panel connector
Shield
Options
With DIN plug or SMART plug
With ASL's SMART connector on the probes, storing the data is needed only once - in the connector! The calibration data stays with the probe - permanently. It can even be used on another instrument without any further action.
Viewed from top panel
Shield
The SMART connector saves time and reduces error. If there are existing calibrated or uncalibrated probes, no problem, CTR2000 automatically registers if a probe is SMART or normal.
Features of the precision thermometer
Easy to use
Large graphic VFD display screen
2 channels as standard with optional 8 channels
4-wire measurement, optional 3-wire
SMART probe review and editor
1
2
3
8-Channel precision thermometer, model CTR2000
1 Inputs 2 VFD display with status channel 3 Soft keys
Page 4 of 6 WIKA data sheet CT 60.10 ∙ 07/2014
Operation
The instrument keypad consists of ve function keys and
three menu keys.
In temperature measurement mode the function keys are used to directly control the measurement operation of the instrument.
In instrument conguration mode the function keys operate
as soft keys along with the menu keys to enable the operator
to congure the instrument.
Instrument function keys
Key symbol Description Function
Selecting input channels
Ch Select input channel 1 to 8 Selects and displays measurement channel 1 to 8
A-B
Setting up measurement options
Select dierential measurement
ChA - ChB
Relative measurement function displays the dierence between selected ChA and
ChB inputs
Zero Zero display measurement Toggles the display zero (null) function
Hold Hold display measurement Toggles the display hold (measurement continues)
Units Select display units Rotates through the display units (Ω, °C, °F or K)
Menu functions
Menu Select menu options Cycles through the main menu functions of the instrument
Clear Clear data entry Clears any data entry errors or min/max statistics
Enter Save entry Saves data entry and returns to previous menu
Page 5 of 6WIKA data sheet CT 60.10 ∙ 07/2014
Scope of delivery
Model CTR2000 precision thermometer incl. power cord
Choice of model CTP5000 temperature probes
Option
DKD/DAkkS calibration certicate
UKAS calibration certicate
1) Only system calibration
1)
1)
Accessories
Expansion
Upgrade from 2 to 8 channels
Temperature probe model CTP5000
Immersion probe
Customer-specic probes are available on request
2-Channel precision thermometer, model CTR2000
Ordering information
CTR2000 / Number of channels / Resistance measurement / Additional order information
© 2013 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modications to the specications and materials.
Page 6 of 6 WIKA data sheet CT 60.10 ∙ 07/2014
WIKA Alexander Wiegand SE & Co. KG
Alexander-Wiegand-Straße 30 63911 Klingenberg/Germany Tel. +49 9372 132-0 Fax +49 9372 132-406 info@wika.de www.wika.de
07/2014 GB
Loading...