WIKA CTP9000 Specifications

Standard thermometer Models CTP2000 and CTP9000
Applications
Comparative calibration in dry well calibrators, tube furnaces and liquid baths
Special features
Low drift, long service life
Wide temperature range
Calibration technology
WIKA data sheet CT 61.10
Description
Calibration by means of an external standard thermometer
The WIKA standard thermometers are particularly suited for applications in industrial laboratories. They enable easy comparative calibration in our baths, in tube furnace and in dry well calibrators.
The use of an external reference thermometer is recommended, particularly for the calibration of short temperature sensors. Thus the errors due to the radial and axial temperature distribution in the temperature conditioning unit are considerably reduced.
Platinum resistance thermometer model CTP2000
For calibration, the test items and the standard thermometer are brought to the same temperature in a temperature conditioning unit.
As soon as a stable temperature is reached, the test items are read or their output signals are measured (resistance, thermoelectric voltage, standard signal) and compared with the standard thermometer. Using this comparison method, the measuring uncertainty can be considerably reduced because not only the display of the temperature conditioning unit is taken into consideration.
Data sheets showing similar products:
Reference thermometer; model CTP5000; see data sheet CT 61.20
Page 1 of 4WIKA data sheet CT 61.10 ∙ 01/2015
Models
Platinum resistance thermometers model CTP2000
Features
The measuring resistor consists of a platinum winding of highest purity. All parts are pre-aged in order to remove contamination and distortions.
Measuring technology
The 4-wire design oers an optimum connection solution for resistance thermometers. The measuring result is aected
neither by the lead resistances nor the temperature-
dependant uctuations.
The electrical connections are welded in order to minimise the transition resistance. The connecting wires are bound in
a 2 m long, shielded connection cable.
Principle of a platinum winding
Thermocouple model CTP9000
Features
The standard thermometer is a type S element whose nominal composition consists of Platinum-10 % Rhodium (positive leg) against Platinum (negative leg) and belongs to the group of noble thermocouples. It is characterised by its high stability. The quality of the thermowell used is essential for stabil ity at high temperatures. For this reason, the high-purity aluminium oxide ceramic C 799 is used. The type S thermocouple, besides the low ageing drift, also
oers the advantage of a low basic tolerance.
Measuring technology
During measurement it must be ensured that the compensating leads from the measuring point to the cold junction consist of substitute materials which have, in a limited temperature range, the same thermoelectric properties as the materials of the thermocouple. Therefore, at this transition, there is no thermoelectric voltage. This voltage is only generated at the point where the compensating leads are connected to normal copper leads.
-
Thermocouple model CTP9000
Calibration
Your standard thermometer should be calibrated once a year. If it is subject to high mechanical stresses, it should be calibrated immediately to guarantee the measuring uncertainty.
Page 2 of 4 WIKA data sheet CT 61.10 ∙ 01/2015